User`s Manual CFM/CMM Thermo Anemometer Model AN100
Contents
1. A single beep will sound and the HOLD icon will appear in the display The average reading will be displayed and number of readings measured will appear in the upper right hand corner of the display After 5 seconds the display will return to the current reading Repeat steps 2 3 until the desired number of points have been measured To return to standard airflow measuring mode press and hold AVG button until meter beeps twice Note In the standard velocity measuring mode press the AVG button once to recall the previous average The average will be cleared when you enter the Averaging Mode again 5 Model AN100 Version 2 0 02 06 Data Hold Air Velocity Air Flow 1 While taking measurements press the HOLD button to freeze the air velocity air flow reading The HOLD indicator will appear in the bottom of the LCD display Press HOLD again to return to normal operation MAX MIN AVG Record Air Velocity Air Flow This allows the user to record and view the highest MAX lowest MIN and average AVG readings Press the button MAX MIN button The MAX indicator and RECORD indicator along with the Max reading will appear on the LCD display and the meter will begin keeping track of the MAX MIN and Average values Press the MAX MIN button again to view the minimum reading The MIN indicator along with the minimum reading will appear on the LCD display Press the MAX MIN button again to view the average reading The AVG indic
2. User s Manual EXTECH INSTRUMENTS CFM CMM Thermo Anemometer Model AN100 Introduction Congratulations on your purchase of the Extech AN100 CFM CMM Thermo Anem ometer This instrument measures Air Velocity Air Flow volume and Temperature The large easy to read backlit LCD includes primary and secondary displays plus numerous status indicators The meter is shipped fully tested and calibrated and with proper use will provide years of reliable service Meter Description 1 LCD Display 2 Vane 3 Rubber Boot 4 Meter 5 Power On Off 6 Area 7 AVG 8 Backlight 9 Temperature Buttons 10 Airflow Velocity Buttons Keypad AUTO POWER orc Press to turn the meter ON or OFF MAX MIN Used to record and store the highest lowest and average airflow or velocity readings LEFT also serves as change decimal point button in AREA mode UNITS Press to select the mode of operation In FLOW mode the meter displays air volume In VELOCITY mode the meter displays air speed A UP also serves as increase number button in AREA mode AVG Used to Average multiple readings in FLOW or VELOCITY mode Up to 20 readings can be averaged HOLD Press to freeze the displayed reading Press again to unlock display This button also functions as the RIGHT scroll button in AREA mode and RECALL mode AREA Press and hold to manually enter the area of a duct in CFM or CMM mode NEXT In AREA mode used to sele
3. ator along with the average reading will appear on the LCD display Press the MAX MIN button again to display current readings NOTE the meter will keep recording MAX MIN AVG readings To clear and stop MAX MIN AVG recording and return to normal operation press and hold the MAX MIN button until the meter beeps twice Automatic Power OFF To conserve battery life the meter automatically turns off after 20 minutes To defeat this feature ths Turn the meter OFF 2 Press and hold thet Backlight key while turning the meter ON 3 dis APO will appear in the display The AUTO POWER OFF feature will now be disabled 4 Note that AUTO POWER OFF is re enabled each time the meter is turned on 5 Also note that AUTO POWER OFF is disabled in CFM CMM or Average mode Battery Replacement When appears on the LCD the 9V battery must be replaced gy a a Disconnect the sensor Remove the meter s rubber protective jacket Use a Phillips screwdriver to open the rear battery compartment Replace the 9V battery Close the battery compartment and replace the meter s protective jacket 6 Model AN100 Version 2 0 02 06 Specifications Air Velocity m s meters per sec km h kilometers hour ft min feet per minute mph miles per hour knots nautical MPH CMM cubic meters min mem medube mra 0 9999 m min 0 to 9 999m CFM cubic ft min 0 9999 ft min 0 to 9 9998 Air Temperature Circuit Displa
4. ce the sensor in the air stream Ensure that the air enters the vane as indicated by the arrow sticker placed inside the vane Refer to the diagram The large main LCD display shows the Air Velocity reading The upper right LCD sub display shows the temperature reading The meter has 16 memory locations 8 for CFM and 8 for CMM that can be used to store commonly used area sizes that you can recall at anytime 1 Press the AREA button until meter beeps twice A memory location number will appear in the top right of the display indicating the memory location Push the NEXT button to scroll thru and select the desired location Once you have selected the desired memory location enter your dimension Use the A UP button to change the flashing digit Use the lt LEFT button to move the decimal Use gt RIGHT button to select the other digits After all of the digits are entered press and hold the AREA button until it beeps twice to save the area into memory and return to CFM or CMM measuring mode To select and use a previously stored dimension press and hold the AREA button until it beeps twice Press NEXT to scroll thru the 8 memory locations Press and hold the AREA button until it beeps twice to return to CFM or CMM measuring mode Air Flow Averaging Mode 1 To enter 20 Point Averaging Mode press and hold the AVG button until it beeps twice The AVG icon will be displayed Take a measurement and press the AVG button
5. ct memory locations 1 8 ro Press to turn the backlight on off MAX MIN Temperature Used to record and store the highest lowest readings for air temperature C F HOLD Temperature Press to freeze the displayed temperature reading Press again to unlock the display Press and hold for 3 seconds to switch between C and F Meter will beep twice to indicate change Battery compartment located on the back of the instrument The rubber protective jacket must be removed from the meter to access the compartment 2 Model AN100 Version 2 0 02 06 Display Layout 4880 e e a LILI ooo MAX MIN AVG HOLD MAX top of LCD Max Hold function engaged for the Air Temperature function HOLD top of LCD Data Hold function engaged for the Air Temperature function VEL indicates that meter is in air velocity mode FLOW indicates that meter is in air flow mode MAX bottom of LCD Max Hold for the IR Temperature and RH function HOLD bottom of LCD Data Hold for the IR Temperature function and RH function C F Temperature units of measure CFM CMM airflow units of measure Ft m units for area dimensions m s ft min km h MPH knots air velocity units of measure X10 X100 multipliers for air flow readings AVG air averaging mode RECORD indicates that min max function is running top for temp bottom for air Large LCD digits at center of display for Relative Hu
6. e minute product information visit our website www extech com Extech Instruments Corporation 285 Bear Hill Rd Waltham MA 02451 Copyright 2006 Extech Instruments Corporation All rights reserved including the right of reproduction in whole or in part in any form 8 Model AN100 Version 2 0 02 06 Useful Equations and Conversions Area equation for rectangular or square ducts Height H Width W Area A Width W x Height H Area equation for circular ducts Area A zx r Where T 3 14 and r radius x radius Cubic equations CFM ft3 min Air Velocity ft min x Area ft CMM m3 min Air Velocity m sec x Area m x 60 NOTE Measurements made in inches must be converted to feet or meters before using the above formulae Unit of Measure Conversion Table m s ft min knots km h MPH 1km h 0 54 0 6222 1MPH 0 4464 87 89 0 8679 1 6071 1 9 Model AN100 Version 2 0 02 06
7. midity and IR Temperature Smaller LCD digits at top right of display for Probe Temperature Low battery indicator 3 Model AN100 Version 2 0 02 06 Operation Connecting the Vane 1 The vane plug is inserted in the meter s sensor jack at the top of the meter The plug and jack are keyed so that the plug can only fit in the jack one way 2 Turn the plug carefully until it lines up with the jack and then firmly push the plug in place Do not apply undue force or try to twist the plug side to side 3 Ifthe vane is not connected to the meter or if the sensor is defective the LCD display will indicate OL in place of a Temperature reading Air Velocity Measurements Single Point Side view of Vane 1 Turn on the meter using the oN orF button 2 Press UNITS button to select the desired unit of Arrow measure NOTE At power up the meter will display the last unit of measure previously Vd entered 3 Place the sensor in the air stream Ensure that lt airflow the air enters the vane as indicated by the arrow Sticker placed inside the vane 4 View the readings on the LCD Display The large main LCD display shows the Air Velocity reading The upper right LCD sub display shows the temperature reading Air Velocity Averaging Mode 1 To enter 20 Point Averaging Mode press and hold the AVG button until it beeps twice The AVG icon will be displayed Take a measurement and press the AVG button A single beep
8. tion A Return Authorization RA number must be issued before any product is returned to Extech The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct indirect incidental or consequential damages Extech s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Calibration and Repair Services Extech offers repair and calibration services for the products we sell Extech also provides NIST certification for most products Call the Customer Service Department for information on calibration services available for this product Extech recommends that annual calibrations be performed to verify meter performance and accuracy ry Support line 781 890 7440 Technical support Extension 200 E mail support extech com Repair amp Returns Extension 210 E mail repair extech com Product specifications subject to change without notice For the latest version of this User s Guide Software updates and other up to th
9. will sound and the HOLD icon will appear in the display The average reading will be displayed and number of readings measured will appear in the upper right hand corner of the display After 5 seconds the display will return to the current reading Repeat steps 2 3 until the desired number of points have been measured To return to standard velocity measuring mode press and hold AVG button until meter beeps twice Note In the standard velocity measuring mode press the AVG button once to recall the previous average The average will be cleared when you enter the Averaging Mode again 4 Model AN100 Version 2 0 02 06 Air Flow Measurements CMM CFM 1 2 Turn on the meter using the ON OFF button Press the UNITS button to select the desired air flow units CMM cubic meters per minute or CFM cubic Side view of Vane feet per minute NOTE At power up the meter will Arrow display the last unit of measure previously entered Pd To begin entering the area in m or f press and lt airflo hold the AREA button until it beeps twice The leftmost digit of the bottom display will begin to flash Use the A UP button to change the flashing digit Use the lt LEFT button to move the decimal Use gt RIGHT button to select the other digits After all of the digits are entered press and hold the AREA button until meter beeps twice to save the area into memory and return to CFM or CMM measuring mode Pla
10. y Sampling rate Sensors Automatic Power off Operating Temperature Storage Temperature Operating Humidity Storage Humidity Operating Altitude Battery Battery life Battery Current Weight Dimensions 14 140 F 10 60 C 0 1 F C 4 0 F 2 0 C Custom LSI microprocessor circuit Dual function 0 5 13 mm 4 digit LCD 1 reading per second approx Air velocity flow sensor Conventional angled vane arms with low friction ball bearing Temperature sensor NTC type precision thermistor Auto shut off after 20 minutes to preserve battery life 32 F to 122 F 0 C to 50 C 14 to 140 F 10 to 60 C lt 80 RH lt 80 RH 2000 meters 7000ft maximum One 9 volt NEDA 1604 battery 80 hours approx if the Backlight is used continuously battery life will be reduced significantly 8 3 mA DC approx 1 6 Ibs 725g including battery amp probe Main instrument 7 0 x 2 9 x1 2 178 x 74 x 33mm Sensor Head 2 75 70mm Diameter 7 Model AN100 Version 2 0 02 06 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment a six month limited warranty applies to sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authorization or visit our website www extech com for contact informa