HYDRIM® L110w - Sterilizer Autoclave
Contents
1. 3 4 14 5 145 psi Cold Inlet 19m 6ft 1 5m 5 ft 1 10 bar 3 4 14 5 145 psi Drain 1 5m 5 ft 3 4 Electrical 1 8m AWG 14 3 Installation If the Hydrim is installed in a Sterilization Center the manufacturer of the Sterilization Center should allow a 10mm 1 2 space at the top back and both sides of the Hydrim This will facilitate installa tion and service access to the Hydrim Do not move the Hydrim into place by maneuvering the open wash chamber door This may cause to door to become misaligned and leak Leveling the Hydrim The unit is standing on three supports rollers wheels at the back and two legs at the front 1 Remove the kickplate Push the Hydrim into place while lifting the strap at the front to allow the unit to roll on the rollers 2 Adjust the front legs as required until the Hydrim is level Access the legs from inside the unit 3 The rear two legs are used only if the floor is uneven or cannot provide support to the rollers power cord cold water inlet hose hot water inlet hose drain hose rollers 4 Installation Instructions Connecting the water inlet hoses Connect the hot and cold water inlet hoses to the hot and cold water supply Drain requirements Connect the drain hose to the drain outlet The drain hose can be attached to existing drain lines using a 3 5 cm 1 5 or larger standpipe P trap combination If the hose is connected dire
2. The door cannot be e The cycle is not complete opened e The cycle was aborted and the unit has not completed draining yet Operating cycles appear e The water is not hot enough Check that the incoming to be taking too long hot water temperature is in the recommended range to complete e Check that the hot and cold hoses are not reversed No detergent message e Check for a kink in chemical tubing comes on when there is still chemical left in the container Drying Performance is e Increase the drying time user menu inadequate NOTE Fuses should NOT be replaced by the operator Call an authorized service technician for assistance 21 7 Troubleshooting Error Messages CF1 Water Heating The water does not reach the Call for service failure required temperature in the specified time CF2 Chamber The water does not fill the Check that the water hoses Filling Failure chamber within the specified time are not kinked Open the water shut off valves CF3 Chamber Faulty sensor Call for service temperature reading failure CF4 Draining Water cannot drain Check that the drain hose is failure from the unit not kinked Check that the filter in the bottom of the chamber is not obstructed CF9 Software or Cycle length is more than Call for service PCB Failure 2hr 30 min Replace detergent box if empty If the problem persists call for service CF14 Fault Flow Flow switch is stuck closed Call for serv
3. Turn the power switch OFF 2 Open the cleaning solution door 3 Disconnect the cleaning solution connector Remove any dried chemical by immersing the connector in a cup of water 4 Remove the empty cleaning solution container and discard or recycle 5 Connect new container and insert into the cleaning solution compartment Follow the instructions printed on each cleaning solution container 6 Close the door 7 Turn the power switch ON 6 Care and Maintenance 6 2 Refilling the Water Softening Salt Reservoir When the error message salt level low appears the salt reservoir is in need of refilling Ensure that your authorized technician has set the water softener according to the hardness level in your area To fill the salt reservoir follow these steps 1 Unscrew the salt container lid 2 Fill the salt container to the top maximum 1 kg 2 2 lbs 3 Screw the salt container lid on tightly 4 Run a complete cycle with no instruments 6 3 Using Rinse Aid Rinse aid is not required when using the Hydrim L110w with HIP cleaning solution There is a rinse aid container in the Hydrim L110w but it is not active and should not be filled 6 Care and Maintenance 6 4 Cleaning the Hydrim L110w To clean outer surfaces and the decal covering the touchscreen use a soft cloth A moistened with soap and water or a surface disinfectant Do not use harsh cleaning chemicals If preparing the
4. Use the up and down arrows to select A oO the desired language The default language is English Touch the back arrow to return to the previous menu Default English Setup Menu Set Regeneration Set Regeneration Set the salt regeneration according to the local water hardness Use the up and down arrows to change the value The default setting is 0 See Installation Instructions for further information OT a 5 Instructions for Use Setup Menu Set Button Beep Set Button Beep The user can select between having the unit beep or not beep whenever a screen button is touched Turn the beep on or off by using the up and down arrows Note that the default setting is ON Touch the back arrow to return to the previous menu Default On Screen Saver Setup Menu Screen Saver The user can adjust the time that elapses before the screen 4min goes into screen saver mode Touch Screen Saver Touch the up and down arrows to select the desired screen saver setting The default is 4 minutes Touch the back arrow to return to Default 4min the previous menu Temperature C F Setup Menu Temperature C F The user can choose to have information displayed in degrees Celsius or degrees Fahrenheit Touch the up and down arrows to select the desired choice The default is degrees Celsius Touch the back arrow to return to the previous menu Default LCD Contrast Setup Menu LCD Contrast The user can adjust the LCD contrast To
5. and Wash Arm Maintenance 20 7 Troubleshooting 21 8 Material Safety Data Sheet 23 9 Specifications 26 10 Serial Number Plate 26 For all service and repair inquiries Canada 1 800 870 7777 United States 1 800 572 1211 EU 49 0 821 278 93400 International 1 416 445 1600 Email HYDRIM and STATIM are registered trademarks of SciCan Ltd SysTM is a trademark of SciCan Ltd All other trademarks referred to in this manual are the property of their respective owners techservice ca scican com 1 Introduction Thank you for selecting the Hydrim L110w instrument washer The Hydrim complements the Statim autoclave by quickly and hygienically preparing soiled instruments for sterilization To ensure years of safe trouble free service carefully review this Operators Manual before operating the unit The Hydrim L110w is suitable only for the applications listed in this manual Using the Hydrim for other purposes may be dangerous The manufacturer cannot be held responsible for damage caused by improper use Operational maintenance and replacement instructions must be followed for the product to perform as designed All trademarks referred to in this manual are the property of their respective owners Contents of this manual are subject to change without notice to reflect changes and improvements to the Hydrim product The Hydrim L110w should only be installed and serviced by a q
6. 110w Menus SciCan HYDRIM SH11MR301 ji Technician Set Drying Time Cycle Count Language Selection Set Regeneration Set Button Beep Screen Saver Temperature C F LCD Contrast 5 Instructions for Use 5 7 User Menu To access the user menu follow these steps Turn the Hydrim OFF and then ON An i information will appear briefly at the bottom right hand of screen SciCan HYDRIM SH11MR301 i nl Touch the i to move to the Menu screen Menu User Menu Select User The user menu consists of three choices Set Drying Time P2 Wash User Menu Set Drying Time The drying time for P2 and P3 can be set independently P1 does not have a drying cycle Default 10min Use the up and down arrows to select the desired drying time Touch default if you wish to return to the factory default Use the back arrow to return to the previous screen Touch the cycle for which you want to set the drying time 5 Instructions for Use User Menu Error History User Menu Error History Printout 1 Page 1 2 This allows the user to see the last PMDE COS three errors Press the down arrow to Unit No 00 move to the next page error Press the Error History on Cycle Number 000032 E X in the lower left hand corner to clear p the error history Press the back arrow to move to the previous menu Set Drying Time Cycle Count x OA User Menu Cycle Count User Menu Cycl
7. 72 No reportable ingredients Toxic Substances All ingredients are listed on the Chemical Substances Inventory Control Act TSCA or exempt Canadian Domestic Substance List DSL All ingredients are listed on the DSL or exempt 12 0 PREPARATION INFORMATION Date November 9 2005 MSDS Prepared by Virox Technologies Inc Telephone 905 813 0110 Disclaimer Information for this material safety data sheet was obtained from sources cosidered technically accurate and reliable While every effort has been made to ensure full disclosure of product hazards in same cases data is not available and is so stated Since conditions of actual product use are beyond control of the supplier it is assumed that users of this material have been fully trained No warranty expressed or implied is made and supplier will not be liable for any losses injuries or consequential damages which may result from the use of or reliance on any information contained in this form If user requires independent information on ingredients in this or any other material we recommend contact with the Canadian Center for Occupational Health and Safety CCOHS in Hamilton Ontario 1 905 877 4400 or the local office of US EPA 25 9 Specifications Height freestanding 850 mm 33 5 Height built in 810 mm 32 Width 600 mm 23 75 Depth 600 mm 23 75 Depth with door open 1200 mm 47 Weight 80 kg 176 lbs Running Noise 60 dB A Hot and cold wate
8. 966S 5 Instructions for Use 5 2 Cycle Description Chart P1 Rinse and P2 Regular P3 Heavy Duty Hold Cycle Cycle Cycle Description Use to prevent Use for moderately Use for heavily soil from drying soiled loose soiled instruments on instruments instruments and cassettes when they will not be washed within one hour Cold Prewash lt 45 C 113 F lt 45 C 113 F Wash N A 50 C 122 F 50 C 122 F 5 minutes 9 minutes Rinse 60 C 140 F 60 C 140 F 60 C 140 F Dry N A 0 20 min 0 20 min 10 min default 10 min default Total Time 9 minutes 16 minutes 25 minutes without drying Water 11 5L 3 gallons 19L 4 75 gallons 25L 6 25 gallons Consumption This is not a wash cycle Always run a wash cycle following the rinse and hold cycle Cycle times depend on the temperature and pressure of incoming water Website www yourCEBA com Customer Service Phone 1 916 335 6113 Email Support yourCEBA com Product Information and Technical Support Phone 509 747 5027 email Info yourCEBA com 5 Instructions for Use 5 3 Baskets and Accessories Observe the instrument manufacturer s recommendations for cleaning and care When loading the unit arrange loose instruments so that they do not touch one another Instruments with concave surfaces or cavities should be placed so that water can drain freely The following are included with the Hydrim Rack for five 8 x11 cassette
9. Review federal provincial and local government requirements prior to disposal 7 0 EXPOSURE CONTROLS PERSONAL PROTECTION Gloves Natural or butyl rubber nitrile or neoprene golves Eye Protection Safety Glasses Respiratory Protection Not normally required if good ventilation is maintained and exposure guidelines are not exceeded Other protective equipment As required by employer code 24 8 Material Safety Data Sheet 8 0 PHYSICAL CHEMICAL CHARACTERISTICS Physical State Liquid Boiling Point No Data pH 9 5 10 5 Specific Gravity 1 14 1 15 25 C Solubility in Water 100 soluble in water Appearance and Color Clear colorless to slight yellow Melting Point No Data 9 0 PHYSICAL HAZARDS REACTIVITY DATA Stability Stable under normal conditions of handling use and transportation Incompatibility Materials to Avoid Acids Hazardous Decomposition Products Combustion produces carbon dioxide and carbon monoxide Hazardous polymerization Will Not Occur 10 0 TRANSPORTATION T D G D O T Classifications Not regulated 11 0 REGULATORY INFORMATION Occupational health and safety Regulations WHMIS Class Class D Division 2B OSHA amp WHMIS MSDS prepared pursuant to the Hazard Communication Standard CFR29 1910 1200 and Canadian WHMIS regulations Controlled Products Regulations under the Hazardous Products Act Environmental SARA Section 313 Toxic Chemical Release Reporting Regulatory Lists 40 CFR 3
10. T within the cycle is shown beneath the bar eg Filling Rinsing FILLING The estimated time remaining in minutes is always displayed Please note that the time will increase or decrease depending on the projected cycle completion 36 min Gr When the cycle is complete the program name and Cycle P2 Wash Complete is displayed and the unit beeps three times Press gt to open the door Use gloves to unload the Cycle Complete instruments After all cycles exercise caution as the instruments trolley baskets and cassettes may be hot Inspect the instruments for debris prior to packaging sterilizing om 5 5 Interrupting Aborting a Cycle P2 Wash The door of the Hydrim is locked during the cycle If the STOP Instruments button is pressed the cycle is aborted and cannot be re started not processed The message Draining Please wait will appear The unit will drain Then the message Instruments not processed will appear Press OK Open the door by pressing Use gloves if handling the instruments Exercise caution as instruments trolley baskets and cassettes may be hot The cycle must be repeated to ensure that instruments are properly processed Note that if the drying phase of the cycle is interrupted the critical parts of the cycle are complete and the message displaying the program name and Cycle Complete Drying interrupted will appear 5 Instructions for Use 5 6 Overview of Hydrim L
11. Tract Sensitization Non hazardous by WHMIS criteria Carcinogenicity Non hazardous by WHMIS criteria Teratogenicity Non hazardous by WHMIS criteria Mutagenicity Non hazardous by WHMIS criteria Reproductive Effects Non hazardous by WHMIS criteria 23 8 Material Safety Data Sheet 4 0 FIRST AID MEASURES Eye Immediately flush with water for 15 minutes while lifting eyelids and rolling eyes Get medical attention if irritation develops or persists Skin Wash promptly with soap and water Ingestion Get medical help Do NOT induce vomiting If conscious and medical help not readily available give water Inhalation Not a normal route of exposure If symptoms develop move to fresh air If breathing is difficult get medical attention 5 0 FIRE FIGHTING MEASURES Flash point Not Available Auto lgnition temperature Not Available Extinguisher Media Water spray dry chemical foam carbon dioxide Special Fire Fighting Firefighters should have eye protection and wear Procedures self contained breathing apparatus 6 0 SPECIAL PRECAUTIONS AND SPILL LEAK PROCEDURES Precautions to be taken in handling and storage Store in a cool dry well ventilated location Other Precautions Avoid skin contact Keep containers tightly closed No smoking or eating in handling area Steps to be taken in case Material is Released or Spilled Flush into sewage system in accordance with local regulations Waste Disposable Methods
12. Website www yourCEBA com Customer Service Phone 1 916 335 6113 Email Support yourCEBA com Product Information and Technical Support Phone 509 747 5027 email Info yourCEBA com jenueyy sioye1odO 4 HYDRIM L110w e Operator s Manual page 1 26 e Manuel de l utilisateur page 26 54 SciCan A HIGHER STANDARD Table of Contents 1 Introduction ceee Cee eee wd 1 2 Important Information 2 3 Disclaimers cscs cc cccccne ewe we wews 3 4 Installation Instructions 4 5 Instructions for Use 6 5 1 Instrument Reprocessing Recommendations 6 5 2 Cycle Description Chart 7 5 3 Baskets and Accessories 8 5 4 Activating the Hydrim L110w 9 5 5 Interrupting Aborting a Cycle 10 5 6 Overview of the Hydrim L110w Menus 11 5 7 User Menu 12 5 8 Technician Menu 13 5 9 Setup Menu 14 5 10 Opening the Door in Case of Power Failure 17 Manufactured by SciCan Ltd 1440 Don Mills Road Toronto ON M3B 3P9 CANADA Phone 416 445 1600 Fax 416 445 2727 Toll free 1 800 667 7733 CE 6 Care and Maintenance 18 6 1 Replacing the HIP Cleaning Solution 18 6 2 Refilling the Water Softener Salt Reservoir Sas su s desde 19 6 3 Using Rinse Aid 19 6 4 Cleaning the Hydrim L110w 20 6 5 Filter
13. ctly to the drain line fittings and adapters should not reduce water flow The drain hose should be attached to the main drain at a point no more than 1 metre 3 ft above the base of the Hydrim A floor drain is acceptable check local codes Electrical requirements Connect the electrical cord to the power supply The Hydrim must have a dedicated circuit equipped with a 15A fuse The outlet must be grounded Additional Information The Hydrim unit is heavy Use a forklift or dolly and exercise caution when moving it The Hydrim should be serviced on site The Hydrim is equipped with an air gap anti suction device to prevent backflow of dirty water into the water supply No other air gap device is necessary If you need to extend the water inlet and drain hoses ensure that you use commercial grade plumbing hose The maximum length of the drain hose is 3 3 m 13 ft Installing Cleaning Solution Install HIP Cleaning Solution as follows Open the cleaning solution drawer Put a new cleaning solution box into the drawer upside down Tear open the tab Testing water hardness and setting the water softener salt Hydrim is equipped with a built in water softening system which needs to be adjusted according to the local water hardness The Hydrim water test kit includes three water hardness test strips in bags Take a water sample from the location where the machine will be installed Open one of the bags and r
14. e Count Set Drying Time This allows the user to see how many cycles have been run on the Hydrim Error History 2 Touch cycle count and the number of cycles completed and aborted will Cycle Count be displayed This cannot be reset Touch the back arrow to return to the previous menu 5 8 Technician Menu Technician This menu is only available to authorized technicians 5 Instructions for Use 5 9 Setup Menu To access the Setup Menu proceed to the menu screen by turning the Hydrim OFF and then ON An i information will appear briefly at the bottom right hand corner Menu User SciCan Technician HYDRIM Touch the i to move to the Menu Setup screen and select Setup SH11MR301 i Setup Setup Unit No Set Button Beep The Setup Menu consists of seven choices as shown Using the up and down arrows to move Set Regeneration Temperature C F through the choices touch the desired option to select it For more A G CZ detail on each setup menu option refer to the following sections Language Selection Screen Saver LCD Contrast Ov 5 Instructions for Use Setup Menu Unit Number If you have more than one Hydrim you can designate aa them Unit 1 Unit 2 etc Use the up and down arrow OO a keys to select the unit number The default is O Language Selection Setup Menu Language Selection This allows the user to select among available languages English for the Hydrim
15. e to the previous menu Up Arrow used to adjust a value up or to move between screens in a menu Down Arrow used to adjust a value down or to move between screens in a menu Unlock used to unlock the door Used to clear the error history Info menu allows access to User Setup and Technician menus 3 Disclaimers Do not permit any person other than authorized personnel to service maintain or supply parts for your Hydrim L110w SciCan shall not be liable for incidental special or consequential damages caused by any maintenance or services performed on the Hydrim L110w by unauthorized personnel or for the use of equipment or parts manufactured by a third party including lost profits any commercial loss economic loss or loss arising from personal injury Never remove the covers of the unit and never insert objects through holes or openings in the cabinetry Doing so may damage the unit and or pose a hazard to the operator If the unit is used in a manner other than that specified the protection provided by the equipment may be impaired Avoid contact with eyes and mouth Never lean on the open door The unit may tip forward causing injury Always turn the unit OFF before adding softener salt adding solutions or performing routine maintenance to the unit The owner shall not allow any personnel other than the trained and authorized personnel to operate the unit For further information or questions about the Hydrim L110
16. emove the test strip Dip the strip in the water Compare the color of the strip with the chart on the back of the bag Determine the water hardness according to the chart on the water test kit envelope Power the unit on Touch the i in the lower right hand corner of the sceen Select Setup and then Set Regeneration 0 180 0 Using the upand 190 210 1 down arrows set 220 250 2 260 300 3 the water softener regeneration level according to this table BREL 4 Ci 5040 5 BE LE VE consider using an additional gt 890 Additional water water treatment treatment required Unscrew the salt container lid and pour 1 litre 1 quart of water into the water softener Add 1 kg 2 2 Ibs of water softening salt in the same manner Screw the salt container cap on tightly salt container Place the box in the cleaning solution drawer and then close Pull out the spout and remove the protective cap Connect the hose Installation Test Turn on the shut off valves Run a cycle checking for leaks in the plumbing connections 5 5 Instructions for Use 5 1 Instrument Reprocessing Recommendations The Hydrim L110w is designed to clean general dental and medical instruments SciCan recommends terminal sterilization after processing in the Hydrim L110w When used correctly the Hydrim L110w is effective for the removal of debris on instruments that may
17. ice Switch CF15 Chamber Overflow switch did not turn off after Check if drain is blocked overflow 30 sec of running the drain pump Call for service CF11 No detergent Flow switch did not detect detergent during wash phase 22 8 Material Safety Data Sheet 1 0 PRODUCT AND COMPANY IDENTIFICATION Product Name HIP Hydrim Cleaning Solution With Instrument Protection Product Use Cleaner Manufacturer Supplier Virox Technolgies Inc Address 2815 Bristol Circle Unit 4 Oakville Ontario L6H 6X5 Phone 905 813 0110 W HMIS Classification Health 2 Flammability 0 Reactivity 0 Personal Protection B Gloves Goggles 2 0 INFORMATION ON ACTIVE INGREDIENTS Ingredient CAS Wt OSHA PEL ACGIH TLV 1 Hydroxyethane 2809 21 4 5 10 N E N E 1 1 Diphosphonic acid K salt 1 2 propylene glycol 57 55 6 15 20 N E N E Xylenesulfonic acid K salt 1300 72 7 5 10 N E N E N Octyl 2 Pyrrolidone 2687 94 7 3 5 N E N E 3 0 HEALTH HAZARD IDENTIFICATION Rout of Entry Skin contact Eye Ingestion Signs and Symptoms Eye contact Causes irritation Skin contact Causes irritation Ingestion Like any product not designed to be ingested this product may cause stomach distress if ingested in large quantities Inhalation Not a normal route of exposure Acute overexposure Corrosive May cause burn or irritation Chronic Overexposure Prolonged or repeated exposure can cause drying defatting and dermatitis Respiratory
18. r connections 3 4 Inlet water pressure 1 10 bar Incoming hot water temperature 50 70 C 122 158 F Drain 3 4 Water softener 1 kg 2 2 lbs salt capacity Filling System 5 5L 1 5 gallon safety maximum Drying System Heater 1 kW Electrical Connection 208 240 V 60 Hz 15A Equipment pollution degree Pollution Degree 2 Equipment installation category Installation category II Maximum relative humidity 80 for temp up to 31 C 88 F 50 for temp up to 40 C 104 F Operating temperature range 5 C to 40 C 23 104 F Max altitude 2000m 6561 7 ft Mains supply 10 of nominal 10 Serial Number Plate When ordering supplies spare parts or requesting service please ensure that the information contained on the serial number plate is available Model number serial number etc The serial number plate is located at the bottom left on the rear panel of the Hydrim L110w unit A small label is located on the left hand side of the chemical door 26
19. result in the insulation of microorganisms during terminal sterilization Prior to placing instruments in the Hydrim consult the instrument manufacturer s reprocessing instructions All critical and semi critical instruments can be considered clean after processing in the Hydrim L110w Non critical instruments can be terminally processed in the Hydrim L110w Examples of medical and dental instruments suitable for cleaning and disinfecting in the Hydrim L110w include e Stainless steel scalers spatulas and hemostats e Stainless steel retractors needle holders skin hooks tissue forceps scissors curettes and other stainless steel instruments free of deep cavities For best results instruments processed in the Hydrim L110w should have the following properties e Heat resistance to a temperature of up to 70 C 158 F e Corrosion resistance in the presence of heat and alkalinity Burrs dental handpieces microkeratomes phaco tubing fiberoptics and electrical equipment should not be processed in the Hydrim HIP Hydrim Cleaning Solution with Instrument Protection has been carefully formulated to provide outstanding material compatibility with a wide range of metals and coatings However some manufacturers instruments may not be compatible All hinged instruments should be placed in open baskets in P3 Heavy Duty cycle When processing small or light objects in the Hydrim use the basket with lid SciCan part number 01 109
20. s for upper or lower trolley Full size 1 1 Part number 01 109963S Rack for five medium size cassettes for upper or lower trolley Half size 1 2 Part number 01 109964S Basket with lid Quarter size 1 4 Part number 01 109966S Hygiene basket for loose instruments Quarter size 1 4 Part number 01 109967S The following are optional Rack for four 11 x 14 cassettes for upper or lower trolley Full size 1 1 Part number 01 109965S Hinged instrument rack Part number 15793 00 Statim 2000 basket Part number 01 107240 Statim 5000 basket Part number 01 107241 Long Basket Part number 01 108232 5 Instructions for Use 5 4 Activating the Hydrim L110w Prior to activating the washer check that the water supply is on and follow these steps Open the chemical door and ensure that a box of HIP Cleaning Solution has been installed Touch the 2 HYDRIM to begin Turn the SH11MR301 unit on l Press to unlock door P2 Wash P3 HD Wash Select the desired cycle The cycle will start automatically Use a heavy duty cycle for Load instruments into the cassettes hinged instruments baskets or cassettes place i or heavily soiled instruments them in the machine and close the door 5 Instructions for Use The cycle will start by draining the unit for 20 seconds The bar P2 Wash on the screen indicates the progression of the cycle The stage L
21. ualified contractor as it is an Installation Category 2 device Before connecting the unit the installer should check that the voltage and frequency of the electrical supply correspond with the details in the installation instructions The unit must only be plugged into a grounded outlet Please note that this unit is completely isolated from the electrical supply only when it is unplugged or the circuit breaker is turned OFF This must be done before any repair work is carried out The Hydrim L110w uses HIP Cleaning Solution exclusively The unit will not perform aD as described if other detergents cleaning products are used Other detergents cleaning products may damage the unit the instruments cause excessive foaming and void the warranty Do not use solvents in this unit Do not allow small children or unauthorized personnel access to this unit or its controls Website www yourCEBA com Customer Service Phone 1 916 335 6113 Email Support yourCEBA com Product Information and Technical Support Phone 509 747 5027 email Info yourCEBA com 2 Important Information e gt x DIDCUEG The following symbols appear in this book Caution there is a potential hazard to the operator This situation may lead to a mechanical failure This is important information The following symbols appear on the touchscreen Used to stop the cycle Forward Arrow used to move to the next menu Back Arrow used to mov
22. uch the up and down arrows to select the desired contrast The default is 50 Touch the back arrow to return to the previous menu Default 50 5 Instructions for Use 5 10 Opening the Door in Case of Power Failure with a screwdriver and lift the kickplate up and toward you Locate the ring and pull it to open the door Exercise caution There may be fluid remaining in the unit and the instruments may be hot Instruments that have not completed the cycle should not be used and should be reprocessed When replacing the kickplate ensure that the grounding wire is attached If there is a power failure remove the kickplate of the Hydrim Remove the screw Grounding wire Kickplate Screw 6 Care and Maintenance 6 1 Replacing the HIP Cleaning Solution The Hydrim L110w uses a 3 8L 1 gallon container of HIP Cleaning Solution Part number CS HIPL Use only the recommended cleaning solution and read the MSDS before inserting into the unit To order additional HIP Cleaning Solution contact your local dealer When the message no detergent appears on the screen the cleaning solution container is empty and is in need of changing To avoid spillage replace the solution container only when the message appears and be sure to wear gloves NOTE Each cleaning solution container will last approximately 30 cycles depending on which cycles are used To replace the cleaning solution container follow these steps 1
23. unit for transport or storage waste water should be removed from the sump Waste water in the unit may contain biological contaminants Use a mechanical means to siphon the contents Wear disposable rubber gloves Dispose of absorbent material according to biological waste disposal regulations 6 5 Filter and Wash Arm Maintenance Filter Maintenance Inspect the filter in the bottom of the chamber daily for debris and clean if necessary To clean remove the filter rinse under a tap and replace Ensure that the filter is firmly locked into position when replaced Wash arms lt S 4 Wash Arm Maintenance If the wash arms are not turning easily remove them rinse under a tap and reassemble To remove the upper or middle arm unscrew the collar counterclockwise To remove the lower arm unscrew the collar clockwise Filter 20 7 Troubleshooting Problem Possible Causes and Solutions Instruments are e Do not overload cassettes or instrument baskets not clean Do not load too many cassettes into the Hydrim e Use the Heavy Duty cycle for cassettes and hinged instruments e Check filters and wash arms for debris Touchscreen e Ensure that the power button is turned ON Unit does not come on e Ensure that the plug is properly inserted into the wall socket e Check if a fuse or circuit is blown call service for assistance The wash arms e Remove the wash arms clean and reinstall them do not rotate
24. w contact your authorized dealer or SciCan Ltd 1440 Don Mills Road Toronto ON M3B 3P9 CANADA Phone 416 445 1600 Fax 416 445 2727 Toll free 1 800 667 7733 SciCan Inc 701 Technology Drive Canonsburg PA 15317 USA Phone 412 494 0181 Fax 412 494 4794 Toll free 1 800 572 1211 SciCan Medtech Alpenstrasse 16 6300 Zug Switzerland Tel 41 0 41 727 7027 Fax 41 0 41 727 7029 EC Representative BHT Hygienetechnik GmbH MesserschmittstraBe 11 86368 Gersthofen Germany Tel 49 0 821 2 78 93 400 Fax 49 0 821 78 93 401 For all service and repair inquiries Canada 1 800 870 7777 United States 1 800 572 1211 EU 49 0 821 278 93400 International 1 416 445 1600 Email techservice ca scican com 4 Installation Instructions To open the wash chamber door plug in the unit turn it on and press the unlock symbol Alternatively remove the screw in the kickplate and pull the ring Tools and supplies required for installation e Slot screwdriver Channel locks Specifications Voltage 208 240 V Height 850 mm 33 5 Frequency 60 Hz Width 600 mm 23 75 Depth 600 mm 23 75 Rated load 2 KW Depth with Circuit door open 1200 mm 47 breaker 15A per phase Weight 80 kg 176 Ibs Running Noise 60 dBA Utility Hook ups Hose Length Max Distance Water Cord Diameter Distance from Pressure inlet drain optimal Hot inlet 19m 6ft 1 5m 5ft 1 10 bar