1. 10 DRILL Micromotor On Off 11 PUMP to set the peristaltic pump flow 3 values available min mid Max Off with Off No flow 12 CALIBRATION to calibrate the contrangle N B To reset the instrument software Switch Off and than switch on the unit while pressing the Program button on the keyboard After resetting the software all the programs will show the manufacturer s default settings TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 6 3 REAR PANEL and PERISTALTIC PUMP Rear panel is shown in the following figures with the spring closed Fig 2 and with the spring open Fig 3 hd epi Ieri f EE TECHNOIMPLANT brushless Serial n 054613 230VAC 115VAC 50 60Hz 80 VA FUSE 230V 2x800mAT 115V 1 6AT Ton Toff 30sec 10sec CE 0476 A 2013 Manufactured by RTC Sistemi s n c Via A Fauli 60 PRATO ITALY camai TECHNOIMPLANT brushless Serial n 054613 230VAC 115VAC 50 60 Hz 80 VA FUSE 230V 2x800mAT 115V 1 6AT Ton Toff 30sec 10sec CE 0476 A x 4 1 2013 Manufactured by RTC Sistemi s n c Via A Fauli 60 PRATO ITALY eins TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 To insert the irrigation set tube disengage the spring from the tubing bed and position it as shown in the pictures Fig 4 Fig 5 Fig 6 Fig 10 TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January
2. TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 TECHNOIMPLANT USER MANUAL TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 2 P 10 11 12 13 14 15 16 INDEX HTroOdUCHON sesseereranni an A EERE 3 Sale ty PrECAULIONS eeror aE 4 NOTICG ON WANS icre Re EO E 5 TCAE aNd a AE E 6 Package COMEN assesi a Aoa 6 PSS ClO UO Verono iana EO E A EN T oL PRONE Dalle enea E a T 6 2 Control panel fUNCTIONS 0 ccc cece eee e eee e eee eeeeeeeeeeeeeeeees 8 6 3 Rear panel and peristaltic DUMP cc cece ee eeee eee eeeeeeeeeees 9 CA PCG ali oaen N E EE 11 WAYS CAMO AE A E E T E A A E ners cote 12 COMMS CHONG eerren E E 12 Use with button GESCIIPTIONS cc cece cece eee e eee e ee ee eee eeeeeeeeteeeeeas 12 EE e E E E E T nese oem tee oaaeee decease taae 15 SS MINIZ AIO ee ere e EEEE 15 MAMEN ANC Oreretara e Aa tien 15 Note ON precaUloNSseronpperin EnaA ANSETE 15 Waran yeeie ee 16 CE Declaration Of CONTORMINY so cececocanarumenearrenan EE 17 ELE FIRA TeSt REDOM orcii T A aces 18 TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 TECHNOIMPLANT Brushless SURGICAL IMPLANTOLOGY MICROMOTOR WITH PHYSIODISPENSER SPEED AND TORQUE MICROPROCESSOR CONTROLLED CE 0476 ECC 93 42 CLASS lla 1 INTRODUCTION Your new Technoimplant 3za24 ese is a surgical micromotor with physiodispenser that combines simplicity to high performance It is entirely des
3. W S0MHz 800MHz d 23 P d 1 2 JP d 1 2 VP For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be determined using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer Notes 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 20
4. CaICWIated Om NE SOU aon applicare EN 61000 4 3 80MHz to frequency of the transmitter 2 5GHz 2 5 Hz peter ined separation distance d 1 2 VP from 150kKHz to 80MHz d 1 2 xF from 80 MHz to 800 MHz d 2 3 P from 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed AF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interlerence may occur in the vicinity of equipment marked with the following symbol Ka 19 TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 Test Report R EM 244 0214 014 Recommended separation distances between portable and mobile RF communications equipment and the Device The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the device as recommended below according to the maximum output power of the communications equipment Rated maximum Separation distance accordino to frequency of the transmitter m output power of transmitter From 150KHz to From 80MHz to From 800MHz to 2GHz
5. for any unapproved purposes Place the unit away from any direct or indirect heat source Avoid humid environments or contact with any liquids Allow air to circulate around the unit to cool it oh ges oe To It is forbidden and dangerous to modify in any way this equipment in this case RTC Sistemi will not be responsible for any eventual damages Environmental conditions Altitude below 2000 m 6500 ft Temperature 18 40 C 64 104 F Relative Humidity lt 80 Mains variations should not exceed 10 of nominal voltage The Micromotor should not be working continuously for more than 3 minutes 8 CONNECTIONS Set the supply voltage in accordance with the local supply Fig 2 Voltage selector Connect the power supply cord Connect the motor plug Connect the pedal plug Insert the physiological solution support rod Insert the physiological solution NY OD UW BR U N e Insert the peristaltic pump irrigation set as described on page 10 8 Switch on the unit O Off I On Fig 2 9 USE When switching on the instrument the parameters are the same for every program default parameters Reduction Rate 20 1 RPM 2000 Forward Torque Max Peristaltic pump flow MAX The unit in its default state is ready to set 5 programs each program includes the following functions which will be automatically performed when the appropriate program number is selected 12 TE
6. 2014 6 4 PEDAL Fig 8 1 To sets the peristaltic pump flow 4 values are available 20 30 45 and 60 ml min Off No flow 2 In the five programs mode scroll the 5 available programs 1 2 3 4 and 5 In the PHASE mode scroll PHASE 1 2 3 and 4 inside the selected program 3 To set the micromotor clockwise or anticlockwise rotation At first push the micromotor is ready to run anticlockwise at the preset speed A beep warns that the micromotor is set in REVERSE and REW appears on the LCD display 4 To Start the peristaltic pump and the motor with rheostatic effect 11 TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 7 INSTALLATION The unopened package can be stored at a temperature from 5 C 41 F to 65 C 149 F with relative humidity below 90 Open the package by cutting the adhesive tape remove first any packaging materials then the instrument and its accessories and place the unit on a flat surface Allow at least 150 mm free space around the unit for ventilation Before installing make sure no damage has occurred in transit Any such damage should be reported to the dealer within 24 hours from receipt Verify the contents according to paragraph 5 PACKAGE CONTENTS When installing Technoimplant please observe the following 1 This equipment must be used only by specialized personnel Do not allow unauthorized persons to touch the unit or play with it nor to use it
7. CHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 Reduction Rate RPM Forward Reverse Torque Peristaltic pump flow rate Clockwise and anti clockwise rotation 9 1 To select the preferred configuration As we have said in the Control Panel Functions TECHNOIMPLANT Brushless gives two modes in the program management and the possibility to use or not the Rheastat effect on starting the motor from the pedal Pressing the Config button the configuration menu appears on the display At this point pressing the Program button it is possible to choose one of the four combinations between Rheostat and Phases mode Rheostat ON and Phases OFF Rheostat OFF and Phases OFF Rheostat ON and Phases ON Rheostat OFF and Phases ON As soon as the desired combination is obtained press the Config button again to memorize the selected choice and exit the configuration menu 9 2 To select the program Press Program on the keyboard or the specific button on the pedal 2 on Fig 3 5 programs are scrolling up 9 3 To select the reduction rate Press REDUCT on the keyboard Available reduction rates are 1 1 16 1 20 1 20 1 Wh 32 1 64 1 80 1 Reduction rate 20 1Wh is for the W amp H models e g WI 75 E KM model or equival ent 9 4 To select the rpm Press RPM on the keyboard to increase or decrease the rpm The speeds will ap pear on the display by keeping it pres
8. CHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 2 SAFETY PRECAUTIONS The device must be used only by competent dentists qualified in implantology The device meets the requirements of the applicable electromagnetic compatibility standards fixed by the norms in force However the user should ensure that a possible interference doesn t provoke a risk in certain equipments in the proximities of the tool It could for instance occur interferences on cardiac pacemaker or other similar devices 1 RTC Sistemi cannot accept responsibility direct or indirect for lesions or resultant damages from improper uses consequential particularly from the not observance of the operational instructions or maintenance of the product 2 Before using the equipment read this manual carefully to become completely attentive to all operating instructions 3 he component motor cable connector and autoclavable micromotor support is furnished not sterile Therefore before the use sterilization is recommended 4 The use of accessories of other builders is exclusively left to the responsibility of the operator 5 Servicing must be performed by RTC Sistemi or by authorized personnel 6 Operating with this unity follow these guidelines never touch drills burs or other handpiece tips when they are still rotating insert contrangles when the motor has stopped running 7 Do not install the instrument where there is a risk of an explosion Thi
9. OR AUTOCLAVE STERILIZATION OF MICROMOTOR 8 POWER SUPPLY CORD TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 6 DESCRIPTION 6 1 FRONT PANEL 7 2 TECHNOIMPLANT 2 2244 PROGRAM 1 RPM 258 TREE MAA ae 5e j R R 2031 FWD Ae PUMP OFF MOT OFF Calibration Pump Drill Reverse Reduct HEES Shot sien oe he Y a ee Config 12 11 10 8 6 Fig 1 TECHNOIMPLANT Brushless has two program management modes Program Mode and Phase Mode selectable from Config button The five programs mode this is the default mode In PROGRAM Mode there is the choice of 5 programs that can be customized with different parameters Torque Speed Contrangle reduction ratio Irrigation flow and Rotation direction The PHASE mode In the PHASE mode within each of the 5 programs selected by the front panel PROGRAM button there are 4 phases which can be selected by the upper central button 2 of the pedal obtaining in this way 4 working phases for each of the 5 programs To each of the 5 programs may be matched a particular implant and the 4 phases may be programmed for instance as Pre drilling Drilling Tapping or Screwing 20 programs are thus available for the user TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 6 2 CONTROL PANEL FUNCTIONS 1 DISPLAY easily read alphanumeric multifunction screen to visualize all functions on 4 lines 2 RPM UP to set the micromotor speed with scrol
10. d they are CE marked according to UE directive 93 42 Medical Device Its usage is allowed only in a hospital environment in clinics and dental surgeries It should be used only by expert qualified dentistry physicians The manufacturer accepts no responsibility in the event of a actions or repairs carried out by persons not authorized by the manufacturer or importer b use on an electrical system not incompliance with the terms of regulation CEI 64 4 c any use different from the one specified in this manual TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 4 TECHNICAL DATA RTC Sistemi s n c PRATO ITALY TECHNOIMPLANT Brushless 255 x 255 x 110 mm oe Central unit ABS and aluminium Material Motor case aluminium Torque Max of the motor Fuse 230 V size 5x20mm 2x0 8AT Slo Blo Fuse 115 V size 5x20mm 2x1 6AT SIlo Blo Type aa Classification Medical device lla Rule IX 93 42 CEE Intermittent 30s ON and 10s OFF Temperature of use From 15 C to 40 C Not suitable for use in the presence of inflammable anaesthetics or oxygen Storage temperature From 5 C to 70 C Temperature of use From 15 C to 40 C 5 Package contents 1 CENTRAL UNIT 2 MULTIFUNCTION PEDAL 3 AUTOCLAVABLE MICROMOTOR WITH CABLE AND CONNECTOR 4 PHYSIOLOGICAL SOLUTION SUPPORT ROD stainless steel 5 AUTOCLAVABLE MICROMOTOR SUPPORT 6 SINGLE USE STERILE IRRIGATION KIT one way 3 pieces 7 PROTECTION CAP F
11. equipment It may be necessary to take mitigation measures such as re orienting or relocating the equipment or its cables or shielding the location Harmonic emissions Class A It is possible to use the device in all IEG 61000 3 2 Conforms establishments including domestic establishments and those directly connected to Voltage fluctuations Conforms ie ete eee peer A ee ee oe supplies buildings used for flicker emissions domestic purposes IEC 61000 3 3 The device is intended for use in the electromagnetic environment specified below The customer or the user of the device should assure that it is used in such an environment Immunity test Test level Compliance Level Electromagnetic environement EN 60601 1 2 guide Electrostatic discharge 6kV contact 6kKV contact Floors should be wood concrete or ESD 8kV air 8kV air ceramic tile If floors are covered with EN 1000 4 2 synthetic material the relative humidity should be at least 30 o Burst Fast Transient 2kV power supply 2kV power supply Mains power quality should be that of EN 61000 4 4 a typical commercial or hospital environment Surge differential 1kV differential Mains power quality should be that of EN 61000 4 5 a typical commercial or hospital common tekV common environment 18 TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 Test Report R EM 244 0214 01A The device is intended for use in the electromagnetic environmen
12. gt The duration of the warranty is one year from the date of the invoice gt Any carrier s expenses are excluded from the warranty gt The warranty shall be invalidated in case of repairs carried out by anybody not authorized by RTC SISTEMI gt From the guarantee any defects shall be excluded which are gt caused during transport gt due to normal wear and tear gt caused by fires or atmospheric events such as lightening discharge In case of breakdown of the equipment please return it only to RTC SISTEMI Snc Via A FAULI 60 59100 PRATO ITALY 16 TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 15 CE DECLARATION OF CONFORMITY Manufacturer RIC Sistemi snc Via Anselmo Fauli 60 I 59100 PRATO Italy Product Micromotor for implantology identified with the name TECHNOIMPLANT Brushless Serial NUMDEr 0ccccceceeceeeeees The above mentioned product conforms to the requirements of Directive CEE 93 42 and successive integrations relative to Medical Devices acknowledged in Italy with the Decree Law N 46 97 and successive integrations and to Directive 2007 47 CE acknowledged in Italy with Decree Law N 37 of 25 01 2010 This product is classified as Class Ila relative to the Directive CEE 93 42 and the Directive CEE 2007 47 CE and has been certified by the CERMET notified body with identification number 0476 In addition the device conforms to the following harmonized re
13. gulations CEI EN 60601 1 Electromedical Devices Part 1 General safety rules EN 60601 1 2 Electromedical Devices Part 1 General safety rules 2 Electromagnetic compatibility EMC Prato ALY iscacccuttientcomeeeesdeies RTC Sistemi snc for the Executive Management Riccardo Rosati TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 Test Report R EM 244 0214 01A ANNEX 1 EMC TABLES TO BE INSERTED IN THE INSTRUCTION MANUAL Following as a guide for the Manufacturer are brought the EMC immunity and emission tables These were filled according to clause 6 of EN 60601 1 2 based on test results This information must be written in the documents accompanying the product Emissiontest Conformity Electromagnetic Environment guidance a Group 1 The appliance use RF energy only for its internal RF Emissions j een Therefore its RF emlsslans are very Cispr 11 low and are not likely to cause any interference in nearby electronic equipment us Class A The appliance is suitable for use in all RF Emissions establishments other than domestic and may be Cispr 11 used in domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes provided the following warning is heeded Warming This equipment is intended for use by healthcare professionals only This equipment may cause radio interference or may disrupt the operation of nearby
14. igned and built by RTC Sistemi both for hardware and software Equipped with a unique design pedal of one centimetre thickness that represents an absolute novelty lt uses a brushless induction autoclavable micromotor that reaches the speed of 40 000 RPM Revolution Per Minute to which can be coupled every type E contrangle that follows the standard ISO 3964 The peristaltic pump is mounted externally from the central unit with the purpose to prevent any possible flooding of the equipmentt in case of break of the irrigation pipe The pump entirely built by RTC Sistemi in a simple approach entirely of metal and this assures a great sturdiness It is of the type provided with a spring loaded tubing bed that ensures perfect tightness of the silicone tube with the pump off The pressure operated by the spring can be regulated simply and easily by hand in such a way that the perfect holding of the liquid can be assured The pedal is extremely innovative denominated solid state because it doesn t have any mech anical moving element while allowing all the functions of the other pedals Built on a strong stainless steel plate and oval shaped for elegance of design covered entirely with a resilient synthetic rubber which makes it resistant to any leaking fluids it rises only10 mm from the floor equipped with a force sensor that determines the speed of the motor in progressive action ac cording to the pressure practiced with the foot TE
15. ling up of 21 values from 400 to 40000 rpm with reduction rate 1 1 from 25 to 2500 rpm with reduction rate 16 1 from 20 to 2000 rpm with reduction rate 20 1 from 12 to 1250 rpm with reduction rate 32 1 from 6 to 625 rpm with reduction rate 64 1 and from 5 to 500 with reduction rate 80 1 3 PROGRAM to set 5 programs with scrolling up values 1 2 3 4 5 For each program the following can be set reduction rate rom torque peristaltic pump speed clockwise and anti clockwise direction 4 5 TORQUE to set the micromotor torque value by scrolling it up and down With the reduction rate 20 1 24 values can be selected from 3 to 69 Ncm in 3 Ncm steps In the MAX position there is no torque limitation 6 CONFIG It shows the configuration menu It is possible to choose 4 combinations between Phase and Rheostat pressing the Program button As soon as the wanted combination is set press again the Config button and data will be memorized and the configuration menu will be leaved 7 RPM DOWN to set the micromotor speed by scrolling down the values 8 REDUCT to set the contrangle reduction or multiplication rate Available reduction rates are 1 1 16 1 20 1 20 1WH 32 1 64 1 and 80 1 9 REVERSE to set the micromotor clockwise or anticlockwise rotation At first pushing the micromotor is ready to run anticlockwise at the preset speed A beep warns that the micromotor is set in REVERSE and REW appears on the LCD display
16. nt Technoimplant is protected by two fuses located in the power entry module To replace the fuses proceed as follows Unplug the unit from the mains Extract the fuse drawer fuse holder with a screwdriver Replace the fuse with one of the same type and value The fuse ratings are 0 8 Ampere for 230 V and 1 6 Ampere for 115 V in each case of the T type Time delay or Time lag The letter F or T represents the time current characteristic of a Fast acting or Time delay fuse type On Technoimplant a type T Time lag fuse called also Slo Blo should be used 12 2 Contrangle Regarding the contrangle please follow the manufacturer s instructions 13 NOTE Keep the instrument out of the reach of children It is to be used exclusively for oral cavities as described in this text The equipment may be used only in accordance with the specific instructions for its use RTC Sistemi does not accept any responsibility for damages resulting from failure to comply with such instructions or from uses other than those foreseen by the field of application for the product The user is responsible for experimenting with materials for uses not explicitly described in these instructions The descriptions and data reported herein are approximate and not binding The contents of this user manual are subject to modifications without notice 15 TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 14 Warranty
17. s device is not for operating in presence of inflammable anaesthetics or gases 8 With the purpose to assure the best accuracy of torque and speed of the motor it is recommended that calibration is performed every change of contrangle TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 9 Contrangles have intrinsic inefficiencies that can lead to torque variations If it is desired a further and more accurate verification of the torque it is suggested to use a torque wrench 10 Always follow the instructions of the producers that accompany the contrangle with respect to the maximum allowed speed couple direction reverse forward and other instructions of USE 11 The system of irrigation is conceived for use with a saline solution or sterile water Always follow the instructions recommended by the implants manufacturers 12 Connect line power cable to a plug provided with earth protection 13 The motor is sensitive to shock and may be damaged if allowed to fall on a hard surface 14 Do not lubricate the motor Lubricate only the contrangle and make sure that the lubricant doesn t penetrate into the motor 15 Do not abruptly fold up the motor cable for placing it in autoclave 3 NOTICE AN This symbol warns the user to refer to the manual Technoimplant is a surgical implantology micromotor with physiodispenser suitable for every type of dental implant burs and handpieces with INTRA plug provide
18. sed they scroll quickly otherwise step by step It is possible to scroll 21 values from 400 to 40000 rpm with reduction rate 1 1 from 25 to 2500 rpm with reduction rate 16 1 from 20 to 2000 rpm with re duction rate 20 1 from 12 to 1250 rpm with reduction rate 32 1 from 6 to 625 rpm with reduction rate 64 1 and from 5 to 500 with reduction rate 80 1 13 TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 9 5 To select the torque Press Torque on the keyboard to increase or decrease the value The torque values will appear on the display by keeping it pressed they scroll quickly otherwise step by step It is possible to scroll 24 values from 3 to 69 Ncm in steps of 3 Ncm with the reduction rate 20 1 In the MAX position there is no limiting torque Torque vs Motor speed Torque Max Motor speed From 3 to 12 Ncm 1250 RPM From 15 to 36 Ncm 1750 RPM From 39 to 69 Ncm 2000 RPM 9 6 To select the peristaltic pump flow Press Pump on the keyboard or the specific button on the pedal 1 on Fig 3 3 dif ferent peristaltic pump flows are available min mid Max Off Off no flow 9 7 To select Forward Reverse Press Reverse on the keyboard or the specific button on the pedal 3 on Fig 3 A beep warns that the micromotor is set in REVERSE and REV appears on the LCD display Reverse 9 8 Contrangle calibration The resistance of a contrangle
19. t specified below The customer or the user of the device should assure that it is used in such an environment Immunity test Test level Compliance Level Electromagnetic environnement EN 60601 1 2 guide Voltage dips short 5 U lt 5 Uh Mains power quality should be that of interruptions and gt 95 dip in U gt 95 dip in U a typical commercial or hospital voltage variations on fo 0 5 cycles fo 0 5 cycles environment If the user of the device power supply input lines requires continued operation during ENSIDAN 40 Ur Lore u oe a gg 60 dip in Uy 60 dip in Ur powered from an uninterruptible power for 5 cycles for 5 cycles supply or a battery 30 dip in Uy 30 dip in Uy for 25 cycles for 25 cycles z lt 5 Uy lt 5 LL gt 95 dip in U gt 95 dip in U for 5 seconds for 5 seconds Power frequency 3 A m Magnetic power frequency fields magnetic field should be that of a typical commercial EN 61000 4 8 or hospital environment Immunity aspects at r f The devices is intended for use in the electromagnetic environment specified below The customer or the user of the device should assure that it is used in such an electromagnetic environment EN 60601 1 2 Level a Vett fram Portable and mobile RF communications EN 61000 4 6 150kHz to 150kHz to equipment should be used no closer to any part 80MIHz s0MHz of the device including cables than the RF radiated 3 Ver from 3 Vek WOM separation distance y
20. varies according to the contrangle model its age and condition the degree of wear on the contrangle gear and so on This function recog nizes the level of resistance and calibrates the micromotor so that it will rotate to the specific speed and torque required gt Insert the contrangle on the micromotor gt Press Calibration on the keyboard The motor will automatically start to run When the resistance diagnosis is com pleted the micromotor will stop The contrangle calibration is now completed N B Any drill must be removed from the contrangle before starting the calibration process Calibration 9 9 To start the motor Press Drill on the keyboard or the specific button on the pedal 4 on Fig 8 14 TECHNOIMPLANT brushless USER MANUAL Rev 10Wh January 2014 10 CLEANING Clean the external surfaces of the case using a clean soft cloth dampened with water and or alcohol or another disinfectant Avoid any splashes the instrument case is not watertight 11 STERILIZATION STERILISABLE PARTS Micromotor and cable Physiological solution rod Micromotor stand STERILISING Apparatus water or chemical vapour autoclave Time 40 min at 134 C 273 F 2 bar Where possible we suggest to sterilize with BIDISTILLED WATER at 121 C 250 F It is essential to apply the protection cap to the motor before it goes into the autoclave 12 MAINTENANCE 12 1 Fuse replaceme