Home

User guide for phase rotation meter TM603

image

Contents

1. 400 Hz Test Currents In per phase less than 3 5 mA 3 5mA Accessories included E User manual a 9V battery NEDA 1604 IEC 6F 22 JIS OO6P 1 carrying case a Test leads should be 4 feet long 3 UL 120 4 feet model FCTL108 a Test leads should be removable and have alligator clips 3 model E230751 TENTAMEN 8 The following symbols are used C AUTION refer to the instruction manual an improper use may damage the instrument or its components Danger high voltage risk of electric shock Double insulated meter AC voltage or current es OC voltage or current Earth ll TENMARS TM 603 9 Safety and maintenance standards The instrument complies with UL61010 1 and CSA C22 2 No 61010 1 between ground and input terminal and EN 61010 1 2001 and EN61557 E Operating altitude below 2 000m E Operating environment for indoor use expose to pollution level II a This is a precision device During use or storage do not go beyond its spec to prevent any possible damage or danger a Do not put this device in direct sunlight or where it is hot and or damp a Remember to turn OFF the power after use For long storage remove the battery to prevent the battery from leaking to cause damage to the parts inside Clean the device with a dry soft cloth Wet cloths liquid and water are prohibited 12 TENMARS TM 603 warranty E Repairs made necessary by impr
2. If the motor turns clockwise and counterclockwise and the red lamp remains ON the red and green lamps do not go ON alternately this suggests one wire is disconnected or not connected well Please measure the motor s 3 wires with a multimeter and check if there is a disconnection between any 2 wires If the motor is disconnected change it 4 The green lamp goes ON when the motor tums clockwise and the red lamp tums ON when it turns counterclockwise This suggests the wiring of L1 L2 and L3 is nomal In this case skip to step 8 If the red lamp goes ON when the motor turns clockwise and the green lamp goes ON when it turns counterclockwise this indicates that the wiring sequence must be corrected Go to step 5 5 Exchange L2 and L3 The green lamp goes ON when the motor tums clockwise and the red lamp goes ON when it tums counterclockwise this suggests the wiring of L1 L2 and L3 is normal In this case skip to step 8 If the red lamp goes ON when the motor tums clockwise and the green lamp goes ON when it turns 5 TENMARS TM 603 5 3 Easy test 3 1 Motor turning test TENMARS TM 603 Corr Son Onot defined hail eee Rotate right 0 amp Q a E Power on Sea OTO O Rotate left al IEJ Q L1 missing oO E f L2 missing 0 IOI Q L3 missing 6 oO 8 TENMARS TM 603 7 Safety precautions and procedu
3. TENMARS 3 Phase Source Motor Tester TM 603 User s Manual CE HB2T M6030000 TENMARS TM 603 1 Preface Thank you for buying our excellent product Products available on the market are not very user friendly their test AC phase and motor phase are separated from each other have two sockets to be plugged out before use etc When you use the TM 603 you will not have these troubles anymore and time is saved So enjoy it 2 Application Ja Test for poles of the 3 phase power source L1 A L2 B and L3 C and identify which phase for which wire 2 Determine the 3 phase motor s direction of rotation clockwise or counterclockwise and the circuit status of the motor wires TENMARS TM 603 CONTENTS E I EE E a I E E EAN E ers Application Features Panel display How to operate Electrical specifications Oe SOG 1 NO 10 Battery Replacement 11 SOWICE cc6ccccaess ARRE TENGON IO a a E E 1 TENMARS TM 603 3 Features Power button Waterproof resist IP67 Check the 3 wires of the power cord L1 L2 and L3 through the 3 red lamps which indicate if each wire is connected well a Check if L1 L2 and L3 are in place the 3 motor wires the motor s direction of rotation and the ER red green OK indicators In each test hold the power button and check the panel indicator a When the battery is low an indicator will turn ON to re
4. ctrical shocks Tum OFF the meter and disconnect the test leads from the input tenninals a Unscrew the battery cover and remove the battery Insert a new battery of the same type 9V NEDA1604 JISOO6P IEC6F22 observing the proper polarity re screw the battery cover and reposition the protective holster Conditions This instrument is guaranteed for one year against material or production defects in accordance with our general sales conditions During the warranty period the manufacturer reserves the right to decide either to repair or replace the product Should you need for any reason to return back the instrument for repair or replacement take prior agreements with the local distributor from whom you bought it Do not forget to enclose a report describing the reasons for returning detected fault Use only original packaging Any damage occurred in transit due to non original packaging will be charged anyhow to the customer The warranty doesn t apply to E Accessories and batteries not covered by 13 TENMARS TM 603 the reasons for returning detected fault Use only original packaging Any damage occurred in transit due to non original packaging will be charged anyhow to the customer The manufacturer will not be responsible for any damage to persons or things 12 End of life Caution this symbol indicates that equipment and its accessories shall be subject to a separate collection and correct di
5. mind you to replace it 4 Panel display 3 red lamps L1 L2 and L3 ER red green OK indicators Low battery indicator see step 10 TENMARS TM 603 5 How to operate 5 1 Howto test the 3 phases of AC Let your crocodile clip clamp the 3 phase power terminal In each test hold the power button and check the indicators on the panel T If L1 L2 and L3 are lit it means normal In this case skip to step 2 If one of them remains unlit L1 for example it means the wire for L1 is disconnected or in power failure mode You may check visually or with a multimeter if the wire is connected well 2 Next check if the red or green lamp is lit If it is the green lamp it suggests L1 L2 and L3 are correct for their phases In this case skip to step 6 If it is the ER red lamp it means L1 L2 and L3 are wired incorrectly You must correct them before proceeding to step 3 3 Exchange L2 and L3 If the green lamp goes ON it means the 3 phases are correct In this case skip to step 6 If the red lamp ER goes ON it means the 3 phases are wired incorrectly You must correct them before going to the next step 4 Exchange L1 and L3 If the green lamp goes ON it means the 3 phases are correct In this case skip to step 6 If the red lamp ER goes ON it means the 3 phases are wired incorrectly You must correct them before going to the next step 5 Exchange L1 and L2 If the green lamp goes ON it means the 3 phases are cor
6. oper use including adaptation to particular applications not foreseen in the instructions manual or improper combination with incompatible accessories or equipment E Repairs made necessary by improper shipping material causing damages in transit E Repairs made necessary by previous attempts for repair carried out by non skilled or unauthorized personnel E Instruments for whatever reason modified by the customer himself without explicit authorization of our Technical Dept The contents of this manual may not be reproduced in any form whatsoever without the manufacturer s authorization Our products are patented and our logotypes registered We reserve the right to modify specifications and prices in view of technological improvements or developments which might be necessary 11 Service Shouldn t the instrument work properly before contacting your distributor make sure that batteries are correctly installed and working check the test leads and replace them if necessary Should you need for any reason to return back the instrument for repair or replacement take prior agreements with the local distributor from whom you bought it Do not forget to enclose a report describing 14 TENMARS TM 603 10 Battery Replacement Battery Replacement When the symbol Low battery LED is displayed batteries need replacement Before replacing batteries disconnect the test leads from any energized circuits to avoid ele
7. rect In this case skip to step 6 6 Note the definition for L1 L2 and L3 each TENMARS TM 603 counterclockwise this indicates the wiring sequence must be corrected Go to siep 6 6 Exchange L1 and L3 If the green lamp goes ON when the motor tums clockwise and the red lamp goes ON when it tums counterclockwise skip to step 8 If the red lamp goes ON when the motor turns clockwise and the green lamp goes ON when it turns counterclockwise this indicates the wiring sequence must be corrected Go to step 7 T Exchange L1 and L2 If the green lamp goes ON when the motor turns clockwise and the red lamp goes ON when it turns counterclockwise this suggests the wiring of L1 L2 and L3 is normal In this case skip to step 8 8 Once the green lamp goes ON when the motor turns clockwise and the red lamp goes ON when it turns counterclockwise the 3 phases are correct Then the motor s R S T can ve connected to the power system s L1 L2 and L3 Now your motor can run correctly TENMARS TM 603 5 2 Howto test the motor s direction of rotation ls Make sure the motor s power source has been removed 2 Connect the test wire to the motor terminal In each test press the power button then check the indicators on the panel when the motor is not tuming the red lamp is ON 3 The green lamp goes ON when the motor tums clockwise and the red lamp goes ON when it tums counterclockwise Nomal skip to step 8
8. res This instrument conforms with safety Standard EN 61010 1 related to electronic measuring instruments For your own safety and to avoid damaging the instrument follow the procedures described in this instruction manual and read carefully all notes preceded by this symbol A For your own safety and to avoid damaging the instrument follow the procedures described in this instruction manual and read carefully all notes preceded by this symbol When taking measurements E Avoid doing that in humid or wet places Avoid doing that in rooms where explosive gas combustible gas steam or excessive dust is present a Keep you insulated from the object under test Do not touch exposed metal parts such as test lead ends sockets fixing objects circuits etc a Avoid doing that if you notice anomalous conditions such as breakages deformations fractures leakages of battery liquid blind display etc E Be particularly careful when measuring voltages exceeding 20V to avoid risks of electrical shocks 10 TENMARS TM 603 6 Electrical specifications Battery life about 50 hours Operating Temperature 5 C 40 C 80 RH Storage temp amp RH 10 C 60 C gt 70 RH Overvoltage category CAT IV 600V CAT III 1000V Dimensions with holster 143 x 68 x35 mm 5 6 x 2 9 x 1 4 inch Weight including battery about 200g Motor POWER 6W 160W Inter phase voltage 45V AC up to 600 AC Signal Frequency Range 45 to
9. sposal TENMARS ELECTRONICS CO LTD 6F 586 RUI GUANG ROAD NEIHU TAIPEI 114 TAIWAN E mail service tenmars com http www tenmars com E mail service tenmars com 15