1. DASCOR Th al f NEWSLETTER October 2007 TOPICS l Newsletter Mail List 2 SENSOR BREAKAGE PLEASE READ 3 SOLID STATE pH Probes A CLEANING pH Probes for extended life 5 The Latest on Ammonia Methane Carbon Dioxide and Pressure 6 Hardware Upgrades in the Current Design 7 NEW SOFTWARE RELEASE SOON V6 10 8 DASCOR Contact Information NEWSLETTER MAIL LIST This newsletter is being sent to everyone who has contacted us regarding DASCOR s LRCpH Ruminal Data Loggers and provided an email address The newsletter will be used primarily to bring you up to date on our current technology and marketing efforts and announce changes in performance new software releases and technical notes on relevant topics It will be sent on an irregular as needed basis Contributions and requests for information are more than welcome However if you are not interested in receiving it please let us know SENSOR BREAKAGE PLEASE READ According to the pH probe manufacturer Sensorex more than 98 of pH electrodes will reach you in excellent working condition Probes that fail during normal handling or during first use may be returned for possible warranty replacement Sensorex will examine the glass fracture surfaces and can easily identify and differentiate failures caused by manufacturing stress crack defects or by mechanical causes such as impacts So if you feel a probe failure was premature and was not ca
2. S655 s have a flat glass electrode while the 971761 s have a hemispherical glass electrode Over the past few years approximately 15 sensors have been returned to DASCOR of which 1 or 2 were covered by warranty replacements This is in keeping with a lt 2 failure rate The remainder were clearly damaged in use and not covered by warranty The good news is that we have identified a possible cause and cure for electrode glass breakage Please examine the photographs below Note that if the wire ties connecting the weights to the shield are around the bottom ring there is a strong probability that the weights can swing into the sensor area and impact the glass We believe this has been the major cause of previous glass breakage The cure is simply to move the attachment to the vertical struts keeping the connecting loop fairly short The weights will have adequate freedom of motion and will continue to hold the logger in the proper position but will not be able to impact the glass aE a _ E _ ma jr s D lt lt a DO ATTACH THE WEIGHTS THIS WAY DASCOR Newsletter 2007 10x doc www dascor com Page 2 11 28 2007 kborsum dascor com of 6 DASCOR Another possibility for impacting the glass is dropping the logger assembly onto the bench or hitting the bottom of the beaker holding the calibration solution The logger is hea
3. and better resolution for pH and ORP REDOX NEW SOFTWARE RELEASE V6 10 After a long wait we are almost ready to issue the final release of the applications package for the LRCpH Loggers All of the problems that have been brought to our attention over the last two years have been corrected and support for new features has been added From the User s standpoint the individual changes are small but DASCOR Newsletter 2007 10x doc www dascor com Page 5 11 28 2007 kborsum dascor com of 6 weeds SCOR j cumulative combine to make the package smoother to use and less prone to locking up and exhibiting other inexplicable behavior Until the new release is available please continue to use version 5 x software In the past one of the major problem areas has been incompatibility between versions of the firmware programs inside the logger Data Formats referred to as header data and the software that runs on the PC These have been resolved in the new release Appropriate warnings will be displayed upon first use of the new version with each logger After all loggers have been automatically updated old versions of the software should be uninstalled from all computers used with the loggers Once a logger has been run with the new software the header will be updated and the warning notices should stop Combined with improvements to the electronics and sensors overall drift and accuracy has been sig
4. ater is not necessary Follow with a good rinse in clean water then storage solution and finally place the cap partially filled with storage solution over the end of the sensor A toothpick or syringe needle can be used to burp air out of the cap The sensor may be stored in any position http www phmeters com faqs html 09 Here s the link directly to the Alconox website it includes links to world wide distributors http www alconox com static techbull techbull_ tergazyme asp Phe storage solution liquid provided in the cap by Sensorex is standard pH 4 buffer solution but has some additional KCI added for storage DASCOR can provide bottles of storage solution Please let us know if you are interested DASCOR Newsletter 2007 10x doc www dascor com Page 4 11 28 2007 kborsum dascor com of 6 DASCOR P pee el fo if The Latest on Ammonia Methane Carbon Dioxide and Pressure It appears that small infrared reflective or absorption measurement systems can be designed for measuring these three compounds in both liquid and gas phases but not at the same time with the same sensor and they should be able to survive in the Rumen for extended periods The initial cost to design and test such a sensor is between 25K and 40K Once proven the sensors are expected to be about 1 5K each and about 3K for all three in one housing They are expected to be small enough to add to the existing LRCpH logger
5. housings The major hurdle will be preventing particulate matter from getting into the measurement cell and partially blocking the light path Off the shelf more or less solutions are also being investigated These will most likely include a miniaturized near Infrared spectrometer covering 300 1000 nM We are working with two vendors to get evaluation prototypes and should have some more news in a couple of months We have received five custom pressure sensors and will be mounting them in evaluation loggers After field testing these units will be available for purchase If you are interested in supporting this project with funding field testing or other technical support please contact kborsum dascor com HARDWARE UPGRADES To date 30 of the new design Type 4 have been shipped and we have commitments for another 20 After the final scheduled deliveries we expect to have 5 or so extra loggers available on the shelf for rapid delivery After they are gone deliveries will be 8 weeks or so after receipt of your order From a hardware standpoint the new loggers Mod 4 shipped after the 2007 San Antonio show will include the following features e Increased memory Improved pH Sensor connector system and seals Easier access to the battery and PC connector Removal of real time reading errors when connected to the PC Improved drift and accuracy of measurements New sensor design to improve resistance to breakage Tighter range
6. nificantly improved Although calibration procedures for the pH sensor are still critical the Real Time display function now operates without excessive noise or causing a shift in the pH readings when the loggers are connected to the PC As always the logger should NOT be held by hand when calibrating or at least consistent contact should be maintained Changes in the proximity of the electric field of the human body will affect the readings momentarily With the help of Karen Beauchemin and Greg Penner Bev Farr has been writing a new User s Manual for the LRCpH Data Loggers written from the researcher s perspective We are looking forward to having this available for download in the near future DASCOR CONTACT INFORMATION Mail P O Box 462885 Escondido CA 92046 2885 Shipping 862 Ball Ave Escondido CA 92026 Voice 1 760 796 7788 Fax 1 760 796 7785 Panic 1 619 794 7788 Kelly s cell Website www DASCOR com Email websales DASCOR com or kborsum DASCOR com We have been getting an incredible amount of spam in recent months and have had to tighten up our email filters at the risk of losing your emails Please call if you don t get an expected response in a reasonable amount of time DASCOR Newsletter 2007 10x doc www dascor com Page 6 11 28 2007 kborsum dascor com of 6
7. used by dropping or impacting the weights rocks or tramp metal in the rumen then return the cleaned sensor to DASCOR freight prepaid and we will have it examined A replacement will be provided by Sensorex if the failure is determined to be manufacturing related such as stress fracture electrical short or open or other workmanship or material issues The most fragile shape for a pH sensor is the spherical lab style Second in line is the flat face The most robust shape is the hemisphere There are minor trade offs between glass thickness sensitivity and surface area that give the hemispherical shape roughly the same performance as the flat glass electrodes but with improved strength Sensorex tells me that the increase in strength is roughly 5 1 HOWEVER and this is important any glass shape can be broken if it is struck by a hard object or has too much pressure applied to it Also with the hemispherical probes there is more exposure of the glass and the potential for impact is increased DASCOR Newsletter 2007 10x doc www dascor com Page 1 11 28 2007 kborsum dascor com of 6 In an ongoing effort to increase the life expectancy of the pH sensors and on the recommendation of the manufacturer all of the sensors that we have shipped recently have been changed to a new part number 971761 The new sensors are identical in function to the S655CD HT s we have been delivering and include the gel reference electrolyte However the
8. ver a few days and high cost 300 Abrasive particles in the rumen appear to be responsible in part for mechanically modifying the sensor coatings and causing the drift According to Malcom Mitchel at IQ Sensors Monsanto promoted the original work but most of the players dropped out several years ago Also the sensor still requires the wet chemistry reference electrode which has been a major cause of premature failure in the older sensors that used the liquid electrolyte CLEANING pH Probes Why clean The rumen tends to leave a protein film over the glass surface of the sensor and on the white frit forming the contact with the reference electrode chemistry In time this will build up in the pores in the glass and frit cause shifts in readings and eventually the failure of the probe After each use and definitely prior to storage pH sensors should be cleaned There are several ways to clean them and several ways NOT to clean them First the NOT Do not use chemical solvents used for cleaning paintbrushes such as acetone paint thinner etc However dilute isopropyl alcohol is OK as is soap and water or dilute acids Many labs have an enzyme based cleaner for lab ware that works very well and is recommended by the sensor manufacturer Tergazyme or Terg Y Zyme by Alconox Using the Alconox product clean the sensor as you would any lab glassware being very careful not to get the connector wet Cool or warm water is OK Hot w
9. vy and enough weight on the sensor will cause breakage whether it has the flat or hemispherical glass The shield ring when properly installed will prevent contact with the glass under normal handling so it should always be in place except when changing or cleaning the electrodes If you encounter problems and feel that the shield needs to extend further out please let us know and we can provide spacer rings Here is a suggested method of handling the calibration buffers please note that the hemispherical glass does not touch the bottom of the buffer container If you follow these procedures and continue to have glass breakage please contact me immediately kborsum dascor com or on my cell phone 1 619 794 7788 DASCOR Newsletter 2007 10x doc www dascor com Page 3 11 28 2007 kborsum dascor com of 6 Woy Wie a SOLID STATE pH Probes We have recently encountered ruminal systems that use a solid state FET pH probe Advantages include storing the probe dry an extended life and small size However the sensor manufacturer s literature clearly states the following 23 What are the applications to avoid Avoid long term immersion The system is designed for momentary tests in liquids or semi solids Do not use the probe for continuous measurement or in line use Do not keep the probe stored in solution Always store the probe dry The key word here is momentary Issues include low stability 0 2 pH or more drift o

Download Pdf Manuals

image

Related Search

Related Contents

クラッチディスク (スポーツフェーシング) 取付・取扱説明書  Mohawk Home 395742 Use and Care Manual    Dahle 50044  BBM-0001 AIVEE BOITIER DE PEDALIER  800 Series 803S 804S 805S SCMS HTM3S HTM4S DS8S Owner's  MultiSync P750 User`s Manual - NEC Display Solutions Europe  9373 B-25 Mitchell DE_FR_EN  UNDER COOLER CART - Lincoln Electric    