1. Tare Sequence In Progress Please Wait Begin Tare e The motorized head will begin to lower the Contact Sensor towards the exposed surface Experiment Parameters 02710206 1 2 e This process will proceed slowly to assure that the correct distance to the exposed region 1s calculated by the control box e tis recommended that the investigator watch to be sure that the contact sensor touches the exposed brain surface and not bone during this process IC AHCEL e Once the surface has been contacted the motorized head will automatically raise the contact sensor so it can be removed Speed 0 00m s e Remove the Contact Sensor by unscrewing ee ee the thumb screw and disconnecting the wire Confirm Experiment Parameters to the black Connection box pas Speed 0 Omis e When finished with this task press Done Co Omsec Tissue Depth e The software will again ask to confirm that the injury parameters are correct If not they aha can be altered again at this time CONFIRM e If the parameters are correct press the CONFRIM button Page 22 of 26 User Manual TBI 0310 e Once the experiment parameters have been set and confirmed the instrument will position the motorized head and impact sensor for the injury Experiment Parameters 0710206 12 29 Speed e Press the Confirm button to indicate that this process can be initiated Positioning Tip for Impact Please Wait
2. Confirm e During the positioning movement the control panel will display a warning sign e 15 important to inspect the surgical field to be sure that nothing can obstruct the impactor rod and tip from performing a complete impact Experiment Pararneters 0 10 06 12 29 e If something does not look correct at this Tissue Depth 1 00mm Speed 0 00m s 4 HI Dwell Time msec time the impact sequence can be aborted by pressing the CANCEL button on the Stage is Set to Impact control panel DOWN e If everything appears to be okay then press the red IMPACT button IMPACT In Progress stand Clear Experiment Parameters e There will be a short delay followed by two Tissue Depth 1 0mm Speed 0 00mis audible beeps indicating that the impactor rod ee SE and tip will perform an impact Experiment Results Actual Impact Speed 0 00m s e After the impact cycle has been completed the _ Requested Parameters __ control panel will display results from that Speed 0 00m s particular impact Dwell Time Omsec Tissue Depth 1 00 e The actual impact speed will be displayed PREVIOUS Exit to Main Menu e To perform another experiment press the Exit to Main Menu button Page 23 of 26 User Manual TBI 0310 Appendix A Compressor maintenance Refer to the pictures The receiver tank needs to be purged of condensation occasionally If the auto drain is not installed the receiver should be
3. Tighten just enough to squeeze the plastic nub on the set screw into the thread of the impactor tip Over tightening will wear out the plastic nub and can damage the threads on the impactor tip Page 6 of 26 User Manual TBI 0310 Calibrating the CONTACT SENSOR e Prior to using the TBI Head Impactor for the first experiment it is extremely important to calibrate the CONTACT SENSOR e This procedure informs the control unit what the distance is between the bottom of the impactor tip and the bottom of the CONTACT SENSOR e This information is extremely important in order that the correct cortical deformation is carried out as desired by the investigator e the main menu screen the control box press the Calibration button Starts the Calibration routines necessary for proper Traumatic Brain Injury Device functioning of the head impactor Experiment Calibration Maintenance e The Calibration Menu will appear on the control box e Press the Calibrate Contact Sensor button on the screen 02710706 12 25 Starts the Contact Sensor Calibration routine e This will activate the Contact Sensor Routine e Anew screen will appear asking for confirmation that it is safe to lower the impactor cylinder rod e Remove all objects below the cylinder before pressing the Confirm button Page 7 of 26 User Manual TBI 0310 Contact Sensor Calibration e Ifall objects
4. Control box to raise the cylinder rod until there is enough room Life 0i 15 45 Contact Sensor Calibration CANCEL i e Make sure there is approximately 1 cm space between the bottom of the Contact Sensor and the Calibration Block e If more room is need use the UP arrow on the Control Box e Press the Done button when everything is in place Page 9 of 26 User Manual TBI 0310 e Anew screen will appear on the Control box asking for confirmation that all objects are in place e Confirm that it is okay to Tare the Contact Sensor on the Calibration Box e The motorized head stage will slowly move the Contact Sensor towards the Calibration Block 16 04 Contact Sensor Calibration Begin Tare e Once the two meet the motorized EE head stage will move the Contact Contact Sensor Calibration Sensor away from the Calibration Block e new screen will request that the Contact Sensor be removed from the head of the cylinder rod e Also unplug the connecting wire from the Black Connection box Contact Sensor Calibration e Connect the wire attached to the eee Calibration Block to the Black Connection box on the Impactor frame e Lift the Calibration Block so that it touches the bottom of the Impactor tip e If the connection is good you will hear an audible beep Page 10 of 26 User Manual TBI 0310 e Immediately after the audible beep the screen on the control box will change e Pla
5. Manual TBI 0310 e Mount the contact sensor onto the head of the cylinder rod as indicated in the diagram Don t over tighten the thumb screw Turn it enough that it just starts to get tight Then go less than another 1 8 of a turn The Contact sensor should be firmly attached at this point Experiment Parameters 08710706 12 27 Speed 0 00m s Dwell Time Omsec Press this button only after the Contact Sensor has been fully mounted This requires that the Contact Sensor cable also E 7 be connected to the Contact Sensor and to the Black Connection Box on the Impactor 1 frame CANCEL PREVIOUS Thumb screw for attaching the Contact Sensor Contact sensor with cable connected to the Black Connection Box on the Impactor frame It is important that the connecting plugs be inserted completely e 15 very important that the head of the Impactor tip and mounted Contact Sensor do not rotate during this procedure Page 21 of 26 User Manual TBI 0310 e If the Contact Sensor is not mounted correctly or the wire to the black Connection box on the impactor frame is not attached the Begin Tare prompt will not appear e If this happens be sure that the cable attaches to both the Connection box and the Contact Sensor e The control panel will prompt to begin the Tare process Press the Begin Tare button Experiment Parameters derided dere Speed 0 00m s
6. cleared occasionally by opening the drain cock on the top of the tank at the opposite end from the pressure regulator Before opening the drain cock cover the open end with a disposable towel and hold onto to it to absorb the exhaust otherwise it will blow a mix of oil and water into the room If the auto drain is installed then the receiver will be cleared regularly and only the collection bottle needs to be emptied Check local ordinances for proper disposal The MSDS is enclosed Also the filter bowl needs to be relieved of condensation and oil At the bottom of the bowl is a pin which when pushed in will relieve the bowl of contents Cover the bottom of the bowl with a disposable towel before pressing for the same reason indicated above Pressure regulator adjustment Compressed air outlet port Receiver tank Pressure release Filter drain Page 24 of 26 User Manual TBI 0310 Appendix A Compressor maintenance cont Maintain the oil level between the max min lines visible on the side of the compressor Remove the fill plug at the top center and add oil as necessary Additional oil can be purchased through PSI Stock Number 1256 500 1 bottle is supplied with the original delivery Oil fill Oillevel s sight glass Page 25 of 26 User Manual TBI 0310 Appendix B Contact sensor maintenance The contact sensor is delicate and should be handled with care Accidently impacting the sensor tip can do damage
7. hose Page 5 of 26 User Manual TBI 0310 Attaching the impactor tips on the TBI 0310 There are two different size tips that accompany the impactor unit The larger one has an external diameter of 5 0 mm and is used for large rodents such as adult rats The small tip has an external diameter of 3 0 mm and can be used for smaller rodents such as very young rats and adult mice Additional impactor tips can be custom made by contacting PSI Each tip has a standard size male threaded region that can be connected to the node at the bottom of the impactor rod that extends from the motorized head stage of the TBI 0310 On one side of the node there is a small set screw When first delivered the set screw will not be installed Extra set screws are provided with the accessories Go ahead and install one Using a hex wrench partially unscrew this set screw which will enable the impactor tip to be connected to the bottom of the node It is very important that the impactor tip be securely hand tightened When tightening the impactor tip firmly hold the node Be extremely careful not to scratch the impactor rod portion that enters the impactor cylinder Once the impactor tip has been attached use the hex wrench to tighten the set screw This is important to assure that the tip does not become loose with repeated use of the impactor Check the tip for tightness often It can become loose after many hits It is important not to OVER TIGHTEN the set screw
8. It should be inspected before use to insure it 1s okay Occasionally dirt particles blood or other contaminates may cause the circuit between the stationary and movable contact to not complete If that happens the tare process will not start an audible indication will be given and the begin tare prompt will not appear The area indicated as the contact surfaces can be cleaned and or burnished Place a strip of bond paper between the contacts and gently pull it while gently squeezing the contact closed Don t squeeze so hard as to tare the paper and leave particles of paper attached This area can also be flushed with alcohol The movable contact is a screw The extra length with the multiple nuts forms a counter balance which assists in achieving the light contact load required to detect the dura The screw also adjusts the axis of the tip relative to the axis of the impactor rod Do not adjust this It is set at the factory and should not need any adjustment The set screw below is used to set the contact load following the procedure defined this manual 6 i lt Stationary 4 5 contact Contact Sensor cable Movable contact amp counter balance Contact surfaces p
9. User Manual TBI 0310 Precision systems and Instrumentation pee aes 6 Fi ee 7 he 8 in 1 ice USER MANUAL TBI 0310 6 2012 TRAUMATIC BRAIN INJURY DEVICE Precision Systems and Instrumentation LLC Page 1 of 26 User Manual TBI 0310 Setting up the TBI 0310 Head Impactor The TBI 0310 Head Impactor when fully assembled has the following components 1 Control box 2 Impactor head stage assembly 3 Base plate for stereotaxic frame with stand post 4 Jun Air compressed air pump 5 Flexible air hose with couplers 1 6 Contact sensor 7 Calibration block 8 Rat impactor tip 5 0mm 9 Mouse impactor tip 3 0mm 10 Connection wire for contact sensor 11 AC power cord for control box 12 Quick release air coupler female 1 13 Hex wrench set 14 Spare impactor tip set screws 5 The TBI 0310 stand has been disassembled for shipping Assembly 15 fairly simple However it may require 2 people due to the weight of the components The bolts are socket head cap screws fitting a 3 16 hex wrench included For ease of assembly start all bolts before tightening any Screw them in nearly all the way but leave them slightly loose Refer to the photographs of the finished assembly for guidance The 2 upright frame pieces are identical The top of the uprights have an offset step cut into it for the cross member to rest on This face
10. ce the Calibration Block immediately below the impactor tip in the position it was during the Tare procedure for the Contact Sensor e When everything is in place press the green Proceed button on the control box screen S 11 03 09 14 27 Contact Sensor Calibration Stage is Set to Calibrate The motorized stage will slowly lower Wem Contact Sensor calibration procedure has been es completed e During this final step a new screen will appear on the control box the impactor tip towards the Calibration Block Proceed e Once the tip touches the block the Calibrating e The calibration of the Co ntact Contact Sensor will Please Wait automatically be saved to the control box e This calibration procedure should be followed every time the Impactor tip is removed from the cylinder rod or when Impactor tips are changed 1 changing from rat to mouse tip Page 11 of 26 User Manual TBI 0310 Calibrating the SPEED SENSOR e This procedure is necessary prior to the first time use of the instrument o Itis NOT necessary to calibrate the Speed Sensor before each experiment or when the tip is changed e This calibration procedure should PS only be carried out if the instrument a ee Precision Systems and Instrumentation has not been used for an extended period of 3 months or if Traumatic Brain Injury Device maintenance that affects the position n of the sensors has been prefo
11. have been cleared from the beneath the cylinder press the Confirm button on the screen e There may be a slight delay before the rod is lowered and a new screen appears on the control box Confirm e Mount the contact sensor onto the head of the cylinder rod as indicated in the diagram 02010206 12 25 Contact Sensor Calibration Pressing this button will automatically lower the rod within the cylinder E T Press this button only after the Contact Sensor has been fully mounted This requires that the Contact Sensor cable also be connected to the Contact Sensor and to the Black Connection Box on the Impactor frame Thumb screw for attaching the Contact Sensor Contact sensor with cable connected to the Black Connection Box on the Impactor frame It is important that the connecting plugs be inserted completely Page 8 of 26 User Manual TBI 0310 e There are small guide posts on the side of the cylinder rod head that match the notched indentations on the Contact Sensor as shown in the schematic diagram above e Be sure that the Contact Sensor is screwed tightly into the Cylinder rod head using the thumb screw e When the connections have been completed press the Done button on the control box screen e Anew screen the control box will ask you to place the Calibration Block underneath the Contact Sensor If there is not enough room to fit the Calibration Block use the UP button on the
12. ke a little while for the top of the cylinder rod to go past both sets of sensors e During this time you will see a red screen on the control box e Once the calibration process 15 finished the cylinder will return to its previous position and the new calibration will be saved to the control unit e The TBI 0310 Impactor is now calibrated and ready to use Page 13 of 26 User Manual TBI 0310 Calibrating the Manual Switch of the Contact Sensor e Jt extremely important that the amount of force used to trip the manual switch on the contact sensor is the same for all experiments in order for the injury to be consistent from experiment to experiment This calibration procedure should be carried out every couple of months to assure consistency In order to perform this calibration it is necessary to have a small digital scale supplied by user with a reading sensitivity of 0 10 271070 12 05 PSI Maintenance Menu Precision Systems and Instrumentation Impact Stage Functions _ Traumatic Brain Injury Device Tare and Hold Sequence Control Box Functions Flash Hew Firmware Serial Maintanence Mode e From the main menu press the Maintenance button e On the Maintenance screen press the Tare and Hold Sequence button 02710208 12 20 02710206 12 20 Tare and Hold Sequence Tare and Hold Sequence ia 1 E 1 Confirm e Before pressing the confirm button be sure
13. onnectors each with a different type of plug configuration e Match connectors with the same type of plug at the end of the wires from the TBI 0310 unit and tighten Align the keying slot insert the cable connector and turn the finger nut approximately 34 turn Attach the AC power cord to back of the control box Cooling fan AC power outlet for compressor Computer interface connectors For maintenance only AC Power Connector Stage interface cables Page 4 of 26 User Manual TBI 0310 Connect the Jun Air compressor to the TBI 0310 After unpacking the Jun Air compressor and assembling according to manufacturers instructions confirm that the compressor is functioning properly Tank gage should indicate 90 to 110 PSI Adjust the regulator until outflow gauge shows 80 psi of pressure 1 Attach a female quick release air connector to the outflow connector of the air pump 2 A fully assembled flexible extension hose 15 supplied 3 Connect one end of the air hose to the outflow connector on the air pump and the other end to the male quick release air connector on the TBI 0310 4 When properly assembled the units should appear as in figure below 5 Refer to the appendices and or the Jun Air user manual for proper pump maintenance draining tank moisture and maintaining proper oil level is very important Outflow on air pump with flexible air hose attached TBI 0310 properly connected to air compressor flexible
14. rmed e the Main Menu press the Experiment Calibration button Calibration Maintenance e Remove the Contact Sensor and the Calibration Block arte e Confirm that it is okay to lower the speed Sensor Calibration cylinder rod e The cylinder rod will extend downward from the cylinder Confirm e Place the Calibration Block beneath the impactor tip e If there is not enough room use the UP button on the control box to raise the impactor tip 11703703 16 12 speed Sensor Calibration DOWN E 2 CANCEL Done PREVIOUS Page 12 of 26 User Manual TBI 0310 e Attach the wire connected to the Calibration Block to the Black Connection Box on the Impactor frame e Raise the Calibration Block so that it speed Sensor Calibration touches the bottom of the Impactor tip at e Ifthe contact is good there will be an DOWN audible beep and the screen on the control box will change e Confirm that everything is in position for the Speed Sensor Calibration by pressing the green Proceed button on the control box screen Q210 08 12 25 e During this procedure the top of the Speed Sensor Calibration cylinder rod will move through a set of sensors on the top portion of the frame e 15 important that both lights on these sensors are glowing Maybe red or orange stage is Set to Calibrate Speed Sensor Calibrating Speed Sensor Please Wait e Jt will ta
15. s the front The uprights are attached to the base first using the 4 longer cap screws and washers Place the base on its back end to expose the bottom Line up one of BOTTOM VIEW BOLTS TO LEFT UPRIGHT the uprights and start the bolts through the appropriate holes in the base into the bottom of the upright Use the rear most hole and the 3 from the back If later you discover you need to have the impactor more to the front you can reposition to another set of holes Finger tighten the screws Repeat this process with the other upright Page 2 of 26 User Manual TBI 0310 d 2 m 2 a gt l r 4 lt 4 CROSS MEMEES GHT LER UPRIGHT s RIGHT ax Set the base down The uprights will be vertical with the offset step facing the front The cross member contains the valve regulator and impactor stage assembly with all the cabling The stage assembly will face the front with the valve and regulator facing toward the back Place it onto the offset step and start a bolt on each side to hold it Start the remaining bolts and finger tighten Once everything is in position tighten the bolts completely Tilt the assembly on its back and tighten the bottom bolts Place the assembly back on the base Page 3 of 26 User Manual TBI 0310 Connecting the interface cables between control box and TBI 0310 e Qn the back of the control box there are four 4 black c
16. se of the impactor e tis important that that base plate fit flat on the impactor base To assist in the mounting of the base plate one of the holes has an oblong shape e the main screen of the control box press the Experiment button Experiment Parameters 2 10 06 12 27 Tissue Depth 1 00mm Speed 0 00m s Dwell Time Omsec Precision Systems and Instrumentation Traumatic Brain Injury Device E Experiment o E Confirm e Confirm that it is safe to lower the impactor tip We recommend that for the first time when lowering the tip that the animal and stereotaxic frame be moved to the side to make sure that there is enough room to lower the tip Calibration Maintenance e Using the X amp Y adjustment wheels on the base of the impactor move the animal into position so that the impactor tip is directly above the area to be impacted e tis important that the impactor tip be perpendicular to the area to be hit This adjustment can be made by altering the horizontal angle of the stereotaxic frame It is important to have the horizontal plane of Experiment Parameters terin 42 27 Tissue Depth 4 00mm Speed 0 00m s the stereotaxic frame adjustable but not IT dices extremely loose e The tip of the impactor can be adjusted with the red up and down arrows on the control box e This allows for a more precise determination that the area to be injured has been centered Page 19 of 26 U
17. ser Manual TBI 0310 e Continue to adjust the base plate with the X amp Y adjustment wheels to align the impactor tip e The head of the impactor rod has a small threaded hole in the front that is used to attached the Contact sensor e Make sure that the tip will fit completely into the craniotomy and not hit the sides e Press the Done button e tis time to enter the experiment parameters such as the Speed Dwell Time and Tissue Depth Experiment Pararmneters 081006 12 27 Tissue Depth 4 00mm Speed 0 00 Owell Time Omsec Enter Experiment Parameters Parameters can be changed at ti me Buy pressing the parameter display button button at the top of the screen Speed 0 00 Omsec Tissue Depth 1 00 Done e The default values always show the first time the control box has been turned on After the new experimental parameters have been entered they will remain that way until they are changed by the experimenter or the control box 1s turned off e To enter new parameters press twice on the highlighted field e g 0 00m s that needs to be adjusted e A small number pad will appear e Press the numbers for the values to be entered e g 3 50 and press enter e Continue to change each of the parameters as necessary e Dwell Time refers to the amount of time the impactor tip will remain in contact with the brain surface The default value is 500 msec Page 20 of 26 User
18. t onto the base of the Impactor Hole 1 m e Mount the base plate onto the impactor base with the stereotaxic frame attached to the stand post e Lower the impactor rod and tip and check to see if you can adjust it with the X and Y control wheels to travel in the full useful range for performing an impact on the animal of choice rat or mouse e When the instrument is shipped the Impactor Base 15 set at the factory to minimize the X axis for shipping purposes and may need to be adjusted especially for small rodents such as mice e There are a series of four screws on the Impactor Base that can be loosened with a hex wrench for this adjustment Page 17 of 26 User Manual TBI 0310 Loosen each of the four adjust screws just enough to easily slide the Impactor Base on the frame e Slide the Impactor Base up or back to and tighten the adjustment screws e Mount the Base plate with the stereotaxic frame and use the adjustment wheels to check the Pawel Adjustment movement the plate Scns e necessary repeat the adjustment procedures above until you are satisfied with the X amp Y movements e Tighten the adjustment screws on the Base plate Y amp X adjustment wheels Page 18 of 26 User Manual TBI 0310 Performing an impact with TBI 0310 e Place the animal in the stereotaxic frame on the base of the impactor and be sure that the holes in the base plate line up with the holes on the ba
19. t screw Remount the Contact Sensor and run the Tare and Hold Sequence again Check and see if the grams necessary to trip the mechanical switch is between 0 6 0 9g If not then raise the sensor and repeat the process The injury device has now been completely calibrated for use Page 15 of 26 User Manual TBI 0310 Using the TBI 0310 Injury Device In order to use the injury device a stereotaxic frame needs to be attached to the base plate i On the base plate is a stand Rigid Stand post that will accommodate a Post and universal swivel that attaches Base Plate to the front portion of the stereotaxic frame While different stereotaxic frames can be used with the device they must be free to rotate in a horizontal plane Rotating in this plane is necessary in order to position the animal so that the piston tip will contact the cortical surface in a perpendicular fashion David Kopf Instruments Tujunga CA sells a universal clamp for some models of their stereotaxic frames We recommend the universal clamp that is part of the 1725 clamp and stand post David Kopf 1725 universal clamp designed to attach a 1400 or 1700 stereotaxic frame to the TBI Impactor Smaller stereotaxic frames such as the Kopf 900 series can be adapted by purchasing the Kopf 925 swivel mount alone Page 16 of 26 User Manual TBI 0310 e The base plate used for mounting the stereotaxic frame has two holes that are used to mount i
20. that the area beneath the cylinder is clear e Once the cylinder rod has been extended mount the Contact Sensor It is necessary to connect the Contact Sensor to the black connection box on the frame e Place the digital scale beneath the mounted Contact Sensor If there is not enough room use the UP arrow on the control box to raise the motorized stage Page 14 of 26 User Manual TBI 0310 e To begin the Tare and Hold Sequence use the red arrow to lower the contact sensor to approximately 1 cm above the middle of the scale s Tare and Hold Sequence Tare and Hold Sequence l 3 EF SCALE Press the Done when the height has been properly adjusted With the digital scale perpendicular to the Contact Sensor and level press the Begin Tare button The motorized stage will lower the contact sensor onto the top of the scale and stop when enough grams of force trip the mechanical switch on the sensor Read the amount of grams on the scale It should be in the range of 0 6 and 0 9 If it is not you can adjust the mechanical switch by turning a small screw on the contact sensor Raise the contact sensor from the top of the scale with the arrows and unmount the sensor A black set screw can be turned counter clockwise to increase the sensitivity lt 0 9g and clockwise to decrease the sensitivity gt 0 6g Mechanical switch adjustment se