Home
TVA2020 - User Manual
Contents
1. 7 Install the five new screws using a Phillips head screw driver 8 Repeat steps 5 7 for opposite head Figure 7 19 Exploded View of Pump Head 7 22 TVA2020 Instruction Manual Thermo Fisher Scientific Replacing the Sample Line in Enhanced Probe Thermo Fisher Scientific Servicing Replacing the Sample Line in Enhanced Probe Use the following procedure to replace the sample line Figure 7 20 1 Loosen the probe nut and remove the sampling assembly by pulling it free from the probe assembly 2 Insert a thin rod through the hole in the side of the probe fitting adapter You will use this rod to pull the adapter and tubing from the probe assembly 3 Remove the tubing from the quick connect fitting by carefully cutting the tubing 4 At the base of the probe handle gently push the tubing into the probe At the same time use the rod to pull the probe fitting adapter and tubing forward from the probe assembly When the adapter fitting is fully exposed cut the tubing from the fitting and pull the remainder of the tubing back through the probe assembly until it is completely out of the probe 5 Cut a new piece of tubing about three inches longer than the old tubing 6 Insert new tubing through the bottom protective covering where you connect to the analyzer and slowly feed the tubing through toward readout handle 7 Once the tubing exits the top protective covering insert the new tubing into the pr
2. 7 Grab the lamp by the cylindrical glass body and clean the window by rubbing it against the polishing pad Use a circular motion and try to keep the window surface flat relative to the pad Five seconds of rubbing will be enough in most cases Another indication of cleaning completeness is that you have used about 1 6 inch of the pads surface during the procedure Figure 7 9 Figure 7 9 Cleaning the PID Lamp TVA2020 Instruction Manual 7 11 Servicing Cleaning the PID Lamp 7 12 TVA2020 Instruction Manual 8 To reassembly install the lamp into the sensor making sure that the lamps metalized pads are aligned with the corresponding excitation springs inside the lamp cavity Figure 7 10 Figure 7 10 Installing the Lamp into the Sensor 9 With the end of the clean tweezers or the clean blade of a screwdriver press down firmly Be careful not to scratch the surface of the lamp Figure 7 11 Figure 7 11 Pressing Down the PID Lamp 10 Using fine tipped tweezers install the cell assembly Align the pins with the corresponding sockets on the sensor and push down on the end with the pins Make sure the cell assembly is flush with the lamp window Figure 7 12 Figure 7 12 Installing the Cell Assembly Thermo Fisher Scientific Thermo Fisher Scientific Servicing Cleaning the PID Lamp 11 Place the spacer around the cell assembly Figure 7 13 Figure 7 13 Placing the Spacer 12
3. Use the following procedure to replace the o rings Figure 5 5 O Rings 1 Unscrew the bottom housing from the top housing 2 Remove filter 3 Remove o ring from either top housing groove or bottom housing groove or both as necessary 4 Place new o ring in groove of housing Make sure o ring is seated properly in groove 5 Repeat for other housing if necessary 6 Place filter into top housing 7 Screw bottom housing into top housing Nozzle Water trap O ring Nozzle Base Figure 5 5 Water Trap Probe Assembly Filter and O Ring Replacing 5 12 TVA2020 Instruction Manual Thermo Fisher Scientific Preventive Maintenance Cleaning or Replacing a Sintered Metal Filter Cup Cleaning OF The filter cup is located behind the sample line quick connect at the sidepack assembly To remove unscrew the quick connect to expose the Replacing a filter Figure 5 6 Pull the filter cup from behind the fitting and replace Sintered Metal and clean as noted below Filter Cup You may clean the filter by immersing it in isopropyl alcohol or equivalent solvent and or swabbing the surface with a Q tip or cotton swab After thoroughly cleaning the surface of the filter place the filter in an oven and dry it at 40 C to 55 C for one hour You can then re insert the filter into the sampling assembly and screw in the cap Nut Filter Holder O ring Spring Filter Cup O rin Sample Inlet ne O ring
4. Bluetooth Communications in the Options section of this manual 2 SETUP 3 INFO This menu contains configuration procedures and menu structure for performing calibrations entering ID numbers setting alarm levels selecting log modes entering response multipliers and setting time date This is a view only menu structure that displays various instrument 4 MEMORY diagnostic information Use this menu to put instrument into USB memory device mode to 3 2 TVA2020 Instruction Manual download upload information to from a personal computer to perform a remote setup from a personal computer to clear memory or to update firmware Thermo Fisher Scientific Thermo Fisher Scientific Display Menus Main Menu Structure Once you are familiar with the various menus and know where to enter specific information you may want to use a short cut method of tracking the menu structure The following figure shows the complete menu structure for the TVA2020 Analyzer RUN SETUP INFO 1 CALIBRATION 1 ZERO 2 SPAN 3 BACKGROUND 4 RF response factors 5 FLOW 1 FLOW 1 2 FLOW 2 3 FLOW 3 4 CALC COEF 6 CONFIGURE 1 NUMBER OF SPAN POINTS 2 SPAN CONCENTRATION down arrow 1 BACKGROUND CORRECT 2 RF CALC MODE down arrow 1 ACCEPT MODE 2 SAVE MODE 2 ALARM 1 STEL 2 LOW CEILING 3 HIGH CEILING 3 LOG 1 NONE 2 AUTO 3 VOC FE 4 CUSTOM 4 PASSCODE 1 NEW PASSCODE 2 ENABLE 3 DISABLE 5
5. GPS POWER ON POWER PID ON POWER FID ON END 9OC0O0OC0CCCOCCOCOCCCCOCCOCOCCOCOC fo Thermo Fisher Scientific Thermo Fisher Scientific Text Data Formats Calibration Ca l ibration Successful download of the calibration file resets the background concentration to zero The calibration is stored as CAL TXT in the following format Format Comments CALIBRATION DATA lt CR gt lt LF gt Title not processed VER XX XX xx xx ver right just lead spaces lt CR gt lt LF gt Blank line not processed lt CR gt lt LF gt Blank line not processed lt detector calibration gt lt detector calibration gt PID if available see below FID if available see below END lt CR gt lt LF gt Marks end of text The detector calibration format is as follows note from 1 to 9 span gas points are allowed ddd CALIBRATION lt CR gt lt LF gt lt CR gt lt LF gt CALIB KNOW CALIB TYPE CONC DATE TIME ZERO aaaaaa uuu dd mmm yy hh mm ss SPAN1 aaaaaa uuu dd mmm yy hh mm ss SPANN aaaaaa uuu dd mmm yy hh mm ss lt CR gt lt LF gt lt CR gt lt LF gt where ddd detector type ETD DID CALIB VALUE lt CR gt lt LF gt lt CR gt lt LF gt UNTS FFFFFFFFF lt CR gt lt LF gt UNTS FFFFFFFFF lt CR gt lt LF gt UNTS FEFFFFFFE lt CR gt lt LF gt aaaaaa gas concentration floating point right justified leading spaces uuu concentration units PPM parts per million percent
6. Press 2 FE to display this message 2 Press the Up Dn arrow keys to choose the type of sampling to be used Max Avg Smp a Max the maximum reading obtained during the log interval b Avg the average reading obtained during the log interval c Smp the last reading obtained during the log interval 3 Type the appropriate time interval desired for F E sampling 4 Press Enter and after a brief message of acceptance the F E method of logging is now ready to be used in the RUN mode Note The allowable Sample Time range is 2 30 seconds If a Sample Time of less than 2 seconds or greater than 30 seconds is selected a warning message will appear and you will be required to re enter the Sample Time A Note If dual option the user must use the enhanced probe to change detector type A 3 20 TVA2020 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Log Custom Display Menus Log Methods e Approximately 2900 data samples may be taken The TVA2020 has several pre designed logging methods However if you find that you would prefer to log data in a different fashion you can use Custom logging to create your own log type The TVA2020 will prompt you through a series of questions which will assist you in setting the various logging options The end result is a logging method which collects data the way you want it to The following variable settings can be configured when designing a Cu
7. Refill tank Replace tank Call Thermo Fisher Scientific Service Department Flameout problems Sample hydrogen content too high Insufficient oxygen in the sample lt 14 Use dilutor kit to achieve concentration within the dynamic range Use dilutor kit to dilute sample with air contaminating sufficient oxygen TVA2020 Instruction Manual 6 3 Troubleshooting Troubleshooting Guide 6 4 TVA2020 Instruction Manual Problem Possible Reason FID capsule contamination Insufficient sample flow min 800 ml min Solution Action Use PID for measurements Clean replace FID capsule Clean replace filter cups Clean replace flame arrestor Service pump Moisture at FID flame arrestor Insufficient sample flow min 800 ml min Clean replace filter cups NOTE Normal operation produces some moisture If performance is affected attempt these solutions Insufficient warmup time Clean replace flame arrestor Service pump Allow 15 20 minutes warmup PID lamp not operating Bad PID lamp Replace PID capsule PID noisy Bad PID lamp Dirty PID window Erratic pump flow Bad calibration Replace PID capsule Clean PID window Clean replace FID capsule Clean replace filter cups Service pump Ensure proper calibration Unable to calibrate PID Pump not on PID lamp not on Span concentration not properly set Cal gas contaminated Sample line filter cup
8. THE OBLIGATIONS CREATED BY THIS WARRANTY STATEMENT TO REPAIR OR REPLACE A DEFECTIVE PRODUCT SHALL BE THE SOLE REMEDY OF BUYER IN THE EVENT OF A DEFECTIVE PRODUCT EXCEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY STATEMENT SELLER DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED ORAL OR WRITTEN WITH RESPECT TO THE PRODUCTS INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE SELLER DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT Thermo Fisher Scientific Thermo Fisher Scientific Appendix B Text Data Formats The TVA2020 supports a USB memory device mode for communications to a computer In this mode the instrument will show up as a USB memory device and the files are directly accessible via this interface The data shown in the USB memory device mode is formatted as printable readable ASCII text All files are read to and written from the root directory of the file system for the memory device The TVA2020 generates files with both lt CR gt and lt LF gt at the end of each line However the TVA accepts files with lt CR gt lt LF gt just lt CR gt or just lt LF gt At startup and after memory device mode is exited an evaluation of all config cal factory files is performed and a warning is displayed on the sidepack if any of the files are in error In addition for any files with errors a file is created with the
9. TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014 2013 Thermo Fisher Scientific Inc All rights reserved Specifications terms and pricing are subject to change Not all products are available in all countries Please consult your local sales representative for details Thermo Fisher Scientific Air Quality Instruments 27 Forge Parkway Franklin MA 02038 1 508 520 0430 www thermoscientific com aqi Thermo Fisher Scientific WEEE Compliance This product is required to comply with the European Union s Waste Electrical amp Electronic Equipment WEEE Directive 2002 96 EC It is marked with the following symbol Thermo Fisher Scientific has contracted with one or more recycling disposal companies in each EU Member State and this product should be disposed of or recycled through them Further information on Thermo Fisher Scientific s compliance with these Directives the recyclers in your country and information on Thermo Fisher Scientific products which may assist the detection of substances subject to the RoHS Directive are available at www thermoscientific com WEEERoHS WEEE Compliance Thermo Fisher Scientific WEEE Compliance Thermo Fisher Scientific About This Manual This manual provides information about installing maintaining and servicing the TVA2020 Toxic Vapor Analyzer TVA It also contains important alerts to ensure safe operation and prevent equipment damage The
10. dd mmm yy hh mm ss iiiiiiiiiiiiiiii tttttttttttttttt ddd xxxxxx uuu fffffffff xxxxxx uuu fffFfFFFFF LEAK REPAIR LEAK SOURCE METHOD chchch cscscs 1111111 ELL RRRRR cccccc GPS DATA lt CR gt lt LF gt lt CR gt lt LF gt aaaaaaaaa oooo00000 hhhh rr nn lt CR gt lt LF gt where dd day 01 31 mmm month JAN January FEB February MAR March APR April MAY May JUN June JUL July AUG August SEP September OCT October NOV November DEC December yy Year 00 99 Thermo Fisher Scientific TVA2020 Instruction Manual B 19 Text Data Formats Logged Data hh Hours 0 23 mm Minutes 0 59 ss Seconds 0 59 tttttttttttttttt tag ID left justified space padded NOTE The following DBID fields are only displayed if a hidden database ID is defined in the route database ih dbid header 1 16 characters centered space padded is dbid separator 1 16 iii dbid 1 16 characters user defined in route database th tag header 1 16 characters centered space padded ts tag separator 1 16 ttt tag 1 16 characters user defined in route database NOTE The following comment fields are only displayed if a Comment is defined in the route database file ch comment header 0 16 characters centered space padded cS comment
11. vi TVA2020 Instruction Manual Thermo Fisher Scientific Contents Battery Warnings and Precautions jcstcicesnessusens sh ccusends taped cepeteceat stead 5 2 Safety Test Criteria senna an ivan anand savateeaeescadeuavelevedens N 5 4 Hydrogen Gras Tank sosta Sie tel aid eoit ledsS canted daa en EE E andes 5 4 Precautions in Handling and Storage c tcccscetisscesacerstiesineeceieeaseietins 5 5 Fuel Refilling Procedwret c csac cee ted ite scat aoa tg eat 5 6 Cleaning an PUD CG arerideers 25s es Bio i uke eres 5 7 Cleaning the FID Detector Cap i664 tee tin Olen ta Cae 5 9 Cleaning the FID Detector Cavities sivecisfeaateeceseieresnesnare ine 5 10 Replacing Water Trap Probe Filter and O Rings eeeeeeeeeereeee 5 11 Replacing the O Rings acs AseceensaciceSaltvhcuacguehevatanyet ledue dd Sanabeteatonee 5 12 Cleaning or Replacing a Sintered Metal Filter Cup oo eee 5 13 Replacing the Charcoal Filter 03 sac sout edt cesta ede 5 14 Chapter 6 Troubleshooting occas steel vecee a 6 1 Troubleshooting Gnide tc siesta ett te ie late hr tae ges 6 2 Warning Mesa ES ccc toate hus tak cassettes cd asaataeeced stars ASE vn aaa aes 6 6 Service Locations iire iaiia E a A iR 6 7 Chapter 7 Servicing iaaeao nena names area NA 7 1 Replacement Parts Listi isiro Ge E Aa aa tte 7 2 Opening the Instrument asiiinsrsen eiiaoe aaia aR 7 5 Replacing the Battery sessar seiiet inea ART AKES 7 6 Hydrogen Gas art kept 2b ioe oset tke eo aaee asui andes 7 6 Precautions in Hand
12. 0 ccee 7 28 Figure 8 1 Sample Probe Option 222 4ecsee eect ees 8 1 Figure 8 2 Enhanced Probe Option cecccccccccescscesestesestesescetesesseseseseeseees 8 2 Figure 8 3 Enhanced Probe Option Display oo cccceccsceccsceestesesteseeteseeeees 8 3 Figure 8 4 Telescoping Extension Option ou cccccccccccsceseesesteteseeteeeeeeee 8 11 Thermo Fisher Scientific Thermo Fisher Scientific Tables Table 1 1 TVA2020 Product Specifications c ccccccccecccceccsceesteseetesesteseeeees 1 9 Table 1 2 TVA2020 External Influences oo cece cecesesesceteesesceteseseetenesens 1 11 Table 1 3 TVA2020 Product Safety Specifications 0 c ccccceeceeeeseeeeeees 1 12 Table 2 1 Keys and Functions 0 ccc ce cececececescseseseetesesesestesesesesteeseseateenees 2 8 Table 3 1 Main Menu Selection and Functions c ccccccccceccseeesteeeteseeeees 3 2 Table 3 2 Alarm Configuiatians 2 2 3 ies ace ee ne a ae 3 11 Table 3 3 Available Log Data Types oo cccccccccetesesescetesesestetesesessenenees 3 15 Table 4 1 Gas Concentration Units cccccccccccccscetesestesesesesteseeteseeteseeneees 4 6 Table 4 2 Detector Counts se ee ee es 4 10 Table 6 1 Troubleshooting oo ccccecesesesceneseseetesesesesteeseseeteessseeteenees 6 2 Table 6 2 Warning M SSaQeS c cccccecececesescscesesescecesesesesteseseseaceseseseateenees 6 6 Table 7 1 TVA2020 Replacement Parts cccccccecesseecsescetesesesteteseseetetenees 7 2 Table 8
13. 1 Enhanced Probe Keys and FUNCTIONS ccc cececeescececeteseseetenenens 8 4 Table 8 2 Enhanced Probe Menu ceccccccecscscsceseseseseseeteseseseeceseseseesenenees 8 5 Table 8 3 Bluetooth Command ou c ccccccceccscsesceseseesestesestesesesestesesteseeeeees 8 13 TVA2020 Instruction Manual xi Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific Chapter 1 Introduction The Thermo Scientific TVA2020 Toxic Vapor Analyzer is an advanced design portable organic inorganic vapor monitor for the gas survey industry The analyzer uses either a flame ionization detector FID or both an FID and a photoionization detector PID For details of the instrument s theory of operation and product specifications see the following topics e Flame Ionization Detection FID on page 1 2 e Photoionization Detection on page 1 4 e Dual Detectors on page 1 6 e Concentration Calculation and Calibration on page 1 8 e Standard Specifications on page 1 9 e External Influences on page 1 11 e Product Safety Specifications on page 1 12 Thermo Fisher Scientific is pleased to supply this portable survey instrument We are committed to the manufacture of instruments exhibiting high standards of quality performance and workmanship Service personnel are available for assistance with any questions or problems that may arise in the use of this instrument For more information on servicing see
14. Dual Detectors Dual Detectors 1 6 TVA2020 Instruction Manual The benefits of each individual detector are very clear both the FID and the PID have their advantages and disadvantages However with either detector alone the number of organic and inorganic vapors that one can detect is limited by the measurement capabilities of that detector With the TVA2020 users can obtain more complete information about organic and inorganic vapors more quickly and easily than with single detector technology alone Since both detectors may be displayed and logged simultaneously the relative response of the two detectors may give some clues about the identity of the compound being measured see Figure 1 3 For instance the PID does not respond to methane at all but the FID responds very well A high FID reading with virtually no PID response might indicate the presence of methane Consequently PIDs respond very well to some inorganic gases that FIDs cannot detect A high PID reading with no FID reading might suggest the presence of an inorganic compound With readings from both detectors readily available the TVA2020 can help a user make decisions about the type of compound present and which detector reading to use Thermo Fisher Scientific Introduction Dual Detectors PID 00000 PPM FID 00000 PPM AMPLIFIER METER EXHAUST FLAME ARRESTOR a TVA2020 COLLECTOR ELECTRODE IGNITION FILAMENT COLLECTOR ELECT
15. Sample Inlet Antenna Connector Instrument Probe Cable Assy O A PID Assy Sample Pump FID Assy Connector Housing Cover Figure 7 1 TVA2020 Component Layout 7 4 TVA2020 Instruction Manual Thermo Fisher Scientific Servicing Opening the Instrument Insert Sleeve 2 1 4 inch 3 4 inch Cap Screw Cap Screw and Washer and Washers 3 Opening the Thermo Fisher Scientific Instrument Barrel Nut 4 Figure 7 2 Hardware Use the following procedure to gain access to the internal components 1 Stand the instrument on end While holding the barrel nut loosen the socket head screws 2 Remove the four 10 24 screws to be reused later Note that one of the screws is much larger than the other three 3 Position the instrument resting on its back with the keypad facing up 4 Grasp the instrument cover near the probe connections and carefully open the cover to the right TVA2020 Instruction Manual 7 5 Servicing Replacing the Battery Replacing the Battery gt gt D Hydrogen Gas Tank 7 6 TVA2020 Instruction Manual Use the following procedure to replace the battery WARNING Do not replace battery in an area classified as hazardous due to P y presence of flammable gases or vapors A WARNING Do not operate battery charger in an area classified as hazardous due to presence of flammable gases or vapors A WARNING Remove the hydrogen tank from the instrument before servicing A
16. Using the special extractor tool provided with the tool kit screw the extractor into the cartridge 4 Remove the cartridge by pulling on the extractor Unscrew the extractor from the cartridge 5 Carefully clean the inside of the detector cavity using a cotton swab and isopropyl alcohol Be sure to clean the high voltage contacts along the side of the cavity Be especially careful around the detector signal collector probe at the rear of the cavity and the thermocouple probe in the FID 6 Dry the inside of the cavity using a low heat gun 7 Insert the cartridges into their respective cavities by reversing the procedure Note that the cartridges must be rotated to properly locate the key tabs Thermo Fisher Scientific Preventive Maintenance Replacing Water Trap Probe Filter and O Rings Replacing Water Use the following procedure to replace the water trap probe filter Figure Trap Probe Filter and O Rings Unscrew the bottom housing from the top housing 2 Remove filter 3 Place new filter into cavity of top housing 4 Filter can be inserted in either direction into top housing 5 Screw bottom housing into top housing Probe Nut Nozzle O ring o Water Trap O ring Nozzle Base Figure 5 4 Water Trap Probe Assembly Filter and O Ring Replacing Thermo Fisher Scientific TVA2020 Instruction Manual 5 11 Preventive Maintenance Replacing Water Trap Probe Filter and O Rings Replacing the
17. dd day of calibration 01 31 mmm month of calibration Jan Dec yy year of calibration 00 99 hh hour of calibration 00 23 mm minute of calibration 00 59 ss second of calibration 00 59 bbbbbbbb detector reading for calibration gas integer right justified leading spaces fffFfFFFFF count status OK no errors DET OFF detector reading invalid TVA2020 Instruction Manual B 11 Text Data Formats Calibration B 12 TVA2020 Instruction Manual DET FAIL detector failed not invalid OVERFLOW A D overflow not invalid UNDERFLOW A D underflow not invalid or A D overflow BAD CALIB invalid calibration parameters CAL_SLOPE calibration sensitivity out of range CLAMPED an intermediate concentration exceeded 100 AVG_OVFLW during log point averaging an overflow occured n Span gas number 2 9 Sample dual detector CALIBRATION DATA VER 1 00 PID CALIBRATION CALIB KNOWN TYPE CONC ZERO o PPM SPAN1 10 23 PPM SPAN2 50 23 PPM SPAN3 100 23 PPM FID CALIBRATION CALIB KNOWN TYPE CONC ZERO o PPM SPAN1 10 23 PPM END CALIB DATE TIME 12 DEC 99 20 20 12 DEC 99 20 20 12 DEC 99 20 20 12 DEC 99 20 20 CALIB DATE TIME 12 DEC 99 20 20 12 DEC 99 20 20 MEASURED CALIB VALUE O COUNTS OK 123456 COUNTS OK 1234567 COUNTS OK 12345678 COUNTS OVERFLOW MEASURED CALIB VALUE O COUNTS OK 123456 COUNTS DET FATL Thermo Fisher Scientific T
18. monitoring different compounds and provides a list of the various gases TVA2020 Instruction Manual About This Manual Safety Safety Review the following information carefully before using the instrument This manual provides specific information on how to operate the instrument however if used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Safety and Equipment This manual contains important information to alert you to potential safety Damage Alerts hazards and risks of equipment damage Refer to the following types of alerts you may see in this manual Safety and Equipment Damage Alert Descriptions Alert PEPE Ee Equipment Damage Description A hazard is present that will result in death or serious personal injury if the warning is ignored A A hazard is present or an unsafe practice can result in serious personal injury if the warning is ignored A The hazard or unsafe practice could result in minor to moderate personal injury if the warning is ignored A The hazard or unsafe practice could result in property damage if the warning is ignored A Safety and Equipment Damage Alerts in this Manual Alert f WARNING ii TVA2020 Instruction Manual Description Do not replace battery in an area classified as hazardous due to presence of flammable gases or vapors A Do not operate battery charger in an area classified as hazardous due to presence of
19. sensitivity In order to adjust the analyzer reading from ppm of Methane or ppm of Isobutylene to ppm of the compound of interest a correction factor must be applied to the reading This correction factor is also known as a Response Factor You can choose from up to nine 9 user defined response factors or use the default response factor of 1 00 Each response factor can be assigned a 9 character alphanumeric name The TVA2020 uses one of two different response factor formats a Multiplier or a Curve Thermo Fisher Scientific Response Factor Multiplier Response Curve Thermo Fisher Scientific Calibration Defining the Response Factor A response factor multiplier is defined as follows Actual Concentration Response Factor ________ Measured Concentration Example If 100 ppm of a compound produces an FID response of 50 ppm on an analyzer calibrated with Methane then the FID response factor would be 2 00 100 50 When using a response factor multiplier to correct a TVA2020 reading the analyzer multiplies the reading by the response factor and displays the corrected reading As in the example give above if a 50 ppm reading is obtained by the TVA2020 the analyzer would automatically multiply 50 times 2 00 to get the actual concentration of 100 ppm Response factors can change as concentration changes The response factor for a compound determined at 500 ppm may not be the same as the
20. A e A 3 10 Accessing the Alarm Menu icsccacetet og etched acter eect cae 3 12 STEE Level Alarmar a a A SN 3 12 Low Ceiling ALAP otii sn eerecieiss antena ias 3 13 High Ceiling Aerie tec 5 oe cathe cence aasan eiia ai 3 14 Los Methods aani as eta iki Ste ed ede han eee dase 3 15 TVA2020 Instruction Manual v Contents Accessing the LOG Ment wisscaneccsstenedeaseneeencnivsins iaden ia 3 16 Log ANT E E decsiousvese 3 16 Tot AUO cas en irren teia ei A a apaid nen eeeontgs 3 17 Lor VOC rreienuieonanonne a a a a a an 3 18 Log E Bvscutaala pnei anann eaa ia T NN 3 19 Log C stom a na n a R A aa a a aa Aiat 3 21 Passcode Protectionissnecisiciciceisrenen iien e gattueane 3 23 New Passcode niscenire si E a E E ASES 3 23 Enableiiss iieiea na a a AKA AA TNR TEESE NKE 3 23 DICE To AS A E E A EEN 3 24 Other Settings enn R ARR R R aN 3 24 User Identification Number ssssssssssssssseesesesesrsrteteererrereeersesrsreres 3 24 DETE A EE 27 as ahs sects E E E 3 25 Time of Dayre an E E A T A AE 3 26 User Options aay scccatsetecansvode recta e A a batty 3 27 Key Click reenn na a E N 3 28 Display Delay aeir aene era a T E alee A 3 28 Infotmaton Men cledecioc ite 5 deceteacceturs aE eonia utie iraa e EiS 3 29 Memory ea eis Bote deco tess ied e EEE EA E ENEN 3 30 Chapter 4 Galibrationochcas tetanic adainankatabnciabenied ahaa 4 1 Calibration SCENAEIOS aeni en dren inci nets te ah 4 1 Scenario Leio Rane A A i a AA Aa Ai 4 1 Seenatio 2 posisi
21. DEFAULT 1 00 0 00 lt CR gt lt LF gt RF1 FID nnnnnnnnnn xx xx yy yy lt CR gt lt LF gt RFQ FID nnnnnnnnnn xx xx yy yy lt CR gt lt LF gt lt CR gt lt LF gt ALARMS lt CR gt lt LF gt Scar lt CR gt lt LF gt STEL PID ffffff PPM LOW PID ffffff PPM HIGH PID fffff PPM lt CR gt lt LF gt lt CR gt lt LF gt lt CR gt lt LF gt Thermo Fisher Scientific Text Data Formats Configuration STEL FID ffffff PPM lt CR gt lt LF gt LOW FID ffffff PPM lt CR gt lt LF gt HIGH FID ffffff PPM lt CR gt lt LF gt lt CR gt lt LF gt USER OPTIONS lt CR gt lt LF gt lt CR gt lt LF gt PROBE ENHANCED lt CR gt lt LF gt KEY CLICK 0FF ON lt CR gt lt LF gt DISPLAY DELAY SHORT MEDIUM LONG lt CR gt lt LF gt DIGITAL DATA LOGGING NONE FID PID PID amp FID SELDET lt CR gt lt LF gt BLUETOOTH POWER OFF ON lt CR gt lt LF gt GPS POWER OFF ON lt CR gt lt LF gt POWER PID OFF ON lt CR gt lt LF gt POWER FID OFF ON lt CR gt lt LF gt lt CR gt lt LF gt END lt CR gt lt LF gt Note Lines containing information on uninstalled detector PID are not output A Note Spaces are allowed in names and other text fields entered by the user A Note Number fields are right justified leading spaces A Note The second detector repeats the RF names During download the last RF name encountered is used A Note Log parameters must be correct for the specified log mod
22. OTHER 1 USER ID 2 DATE 3 TIME 4 USER OPTIONS 1 KEY CLICK 2 DISPLAY DELAY Figure 3 1 TVA2020 Menu Structure MEMORY 1 USB MODE 2 CLEAR LOGGING MEMORY 3 UPDATE FIRMWARE 1 CONNECT USB 2 COPY FILE 3 WAIT REBOOT TVA2020 Instruction Manual 3 3 Display Menus Run Mode Run Mode 3 4 TVA2020 Instruction Manual How the instrument functions in the RUN mode is governed by the selection made in the SETUP Log menu The LOG menu selects how data is stored in the instrument memory There are four possible choices in the LOG mode as defined in Log Methods on page 3 15 In addition if the optional Bluetooth module is active data automatically streams out continuously from the instrument Data will only stream out of the instrument while in the RUN mode The data will always stream at in the same format independent of the log configuration See Bluetooth Communications in the Options section for more information The TVA2020 performs accurately only after it has been properly set up configured The RUN mode displays as governed by the LOG Selection are FID NONE PID FID NONE FID AUTO VOC or F E PID FID AUTO VOC or F E You should install the hydrogen tank and wait 2 3 minutes before entering the RUN mode Upon entering the RUN mode the pump will automatically turn ON and the FID will ignite Note For best performance the instrument must be ON and warmed up for approximately 30
23. Press EXIT to return to MAIN menu A This menu allows you to link the TVA2020 to a personal computer PC through the USB port to copy files from and into the instrument clear data memory and update instrument firmware Note The Flash memory device is erased prior to the file creation process to ensure sufficient memory exists to create the data files Do not store any files on the instrument Flash drive as they will be lost A WARNING Charger and USB connections are not allowed in an area classified as hazardous due to presence of flammable gases or vapors A To enter the MEMORY menu from the MAIN menu press 4 MEMORY The following display will appear Choosing 1 After choosing 1 the display will show Creating Files Please wait as shown below The file creation can take up to one minute depending on the size of the log file Do not connect the USB cable to the instrument until the instrument displays the screen below Thermo Fisher Scientific Display Menus Memory The USB barrier device must be connected in line between the computer and the instrument Two standard USB cables are provided to make this connection If the barrier device in not used the following message will be displayed When first connected to a PC the TVA2020 will send the correct driver files to the computer Administrator privileges may be needed to allow the proper files to be transferred to the PC Once connection has been esta
24. Probe Option e sesesssssesesesssesesesssssssssssssssrsrsrsrsesesesesesesesesrses 8 2 Thermo Fisher Scientific TVA2020 Instruction Manual vii Contents viii Appendix A Appendix B Appendix C TVA2020 Instruction Manual Enhanced Probe lis play xc st ccctandeten testeacessetu tes of cactacee tupehcaceteesaiahe 8 3 Enhanced Probe Keys sciiscasteecesscssssivsovenasdecaheese svedesaivn NN 8 4 The Enhanced Probe and the RUN Mode eeeeeceeeeeereeeeeeeneeeeees 8 4 Auto Logging with the Enhanced Probe eeseeeecseseeceeseeeeeeeeeeens 8 5 VOC Logging with the Enhanced Probesi 0 coi enienedeeat 8 6 FE Logging with the Enhanced Probe iss lt icesissscertvastsnsabivervcrdcaseraserios 8 7 Route Entry Response Factor siciess ccs iveviertsscevaesas cesveunetderdocediens 8 7 Navigating the Route File sscintvenswncesicty aussie necnissevdeuencouesuadyaaicans 8 8 eS ite Data os cost siacesite un enc ugeeataba R A aa teak ashe 8 8 R pa Menusi ests eoprint gun tab iai 8 9 Additional Component Information ssssssssssssesssseessrerererreerersrereres 8 10 Telescoping Extension Option ssssssssssssssssseersersestsesreststerererrsrerreesesess 8 11 Activated Charcoal Filter Adapter ssseeeeeeeeesesrreresesrressse 8 12 Bluetooth Communications sssesesesesesssesesesesesesesesisisesestsesesesesesesees 8 13 Bluetooth Commands sssesesesesesesesesesesrsrsesrsrsesestsrteteerreeeerererereres 8 13 Digital Data Logging Bluetooth s
25. Selecting Pg at the bottom of the display will step you through other INFO pages containing information such as Run ID Record number number of points logged number of leakers found number of repairs performed pump status free memory etc Selecting Exit will return you back to the normal component record Thermo Fisher Scientific Optional Accessories Telescoping Extension Option Telescoping To use an optional telescoping extension loosen the probe nut and unplug Extension Opti on the sampling assembly Next insert the telescoping wand and tighten the probe nut Then insert the appropriate sampling assembly into the other end of the extension unit and tighten the retaining nut PROBE ASSEMBLY PROBE NUT CLOSE AREA SAMPLER FILTER ADAPTER TELESCOPING EXTENSION UNIT Figure 8 4 Telescoping Extension Option Thermo Fisher Scientific TVA2020 Instruction Manual 8 11 Optional Accessories Activated Charcoal Filter Adapter Activated Charcoal Filter Adapter 8 12 TVA2020 Instruction Manual The Activated Charcoal Filter Adapter is an accessory that can be installed or attached to the end of the standard probe or to the end of a telescoping extension The filter is typically filled with activated charcoal which acts as an adsorbent and effectively filters out organic vapors other than methane or ethane A screw cap on the probe end may be removed for refilling the filter with activated charcoal or other fi
26. TEFFFFTFF xxxxxx uuu ffttttttt GPS DATA lt CR gt lt LF gt aaa att deat atat ata ate laat al aatal etait at ataiatat ate lt CR gt lt LF gt aaaaaaaaa oooo00000 hhhh rr nn lt CR gt lt LF gt VOC logged header and data single detector GPS off lt CR gt lt LF gt VOC DATA lt CR gt lt LF gt DATE TIME TAG ddd BACKGROUND ddd CONCENTRATION lt CR gt lt LF gt manner re renee enn nnn nnn nnn nnn ne lt CR gt lt LF gt dd mmm yy hh mm ss tttttttttttttttt xxxxxx uuu fffffffff xxxxxx uuu fffffftff lt CR gt lt LF gt VOC logged header and data dual detector GPS off lt CR gt lt LF gt VOC DATA lt CR gt lt LF gt DATE TIME TAG PID BACKGROUND PID CONCENTRATION T E E E ib tees dd mmm yy hh mm ss tttttttttttttttt xxxxxx uuu ffffffffF xxxxxx uuu ffffFFFFF FID BACKGROUND FID CONCENTRATION lt CR gt lt LF gt lt CR gt lt LF gt XXXXXX uuu FFFFFFFFF xxxxxx uuu fffffffFF lt CR gt lt LF gt VOC logged header and data dual detector and GPS on lt CR gt lt LF gt VOC DATA lt CR gt lt LF gt DATE TIME TAG PID BACKGROUND PID CONCENTRATION E AE TA A E E e oe dd mmm yy hh mm ss tttttttttttttttt xxxxxx uuu ffffffffF xxxxxx uuu ffffFFFFF FID BACKGROUND FID CONCENTRATION BE ewer meena ye ITE EA E S A XXXxxx uuu ffffFfFffff xxxxxx uuu fffFFfFfFFFF GPS DATA lt CR gt lt LF gt lt CR gt lt LF
27. To maximize standard accuracy it is highly recommended that you calibrate with methane for the flame ionization detector and isobutylene for the photoionization detector Almost all published response factors for FIDs and PIDs are based upon methane and isobutylene respectively By employing a multipoint calibration for these compounds you will improve the accuracy of each detector over the entire dynamic range Response factors curves can then be employed for correcting the detector s response to different compounds However once a multipoint calibration has been employed any response curve must characterize only the relative response at each concentration excluding curvature of the calibrated compound Thus use of both multipoint calibration and response curves at the same time is difficult and is not recommended If for example you want to measure several different compounds over wide concentration ranges it is best to use a single point calibration and then enter response curves for each specific compound up to 9 response factors curves can be entered into the analyzer TVA2020 Instruction Manual 4 1 Calibration Calibration Scenarios Scenario 3 4 2 TVA2020 Instruction Manual A If instead you want to measure in direct readings response factor 1 for one specific compound with maximum accuracy over a wide range of concentrations perform a multipoint calibration with the specific compound Up to 9 span points plus
28. a number of different forms 1 Relative to the calibration gas If the compound with which the TVA2020 was calibrated is an acceptable standard then results may be expressed in terms of that compound For example an FID reading of 100 ppm could be Thermo Fisher Scientific Thermo Fisher Scientific Response Factors Response Factors expressed as 100 ppm of Methane if the calibration gas used was Methane Please note that the TVA2020 detectors can be calibrated with gases other than the factory standards as long as the detector has adequate response to that compound 2 Response factor corrected If the TVA is used to measure a single known compound a response factor from this set of response curves can be applied to correct the TVA reading for that compound For example if a survey of Acetone vapors yields a PID reading of 100 ppm the response factor multiplier from the Acetone curve 1 417 could be applied to the reading to obtain the result of 141 7 ppm Acetone 3 Relative to a standard other than the calibration gas If a regulatory agency requires that data be referenced to a standard other than the calibration gas the response factor for that standard can be applied to the reading For example suppose that a survey of total volatile organics must be measured but the results must be reported in equivalents of benzene The FID calibrated with methane can be used and the benzene response factor applied to correct
29. do not necessarily need to have the same number of span points When Both option is chosen a single value will be applied to both PID and FID The span gas concentration is the known concentration of the gas standards used to calibrate your TVA2020 Methane in air is the recommended calibration standard for the FID and Isobutylene in air is the recommended calibration standard for the PID Other gases may be used if desired Thermo Fisher Scientific Thermo Fisher Scientific Calibration Calibration Configuration Note If your instrument is equipped with dual detectors you may choose to calibrate the FID separately or PID FID together A Note If your instrument is configured for multiple span points be sure to set the concentration for ALL span points A 1 From the CAL CONFIG MENU Page 1 display press 2 Span concs The upper display or two displays if the unit is a dual detector version will display the concentration value of your span gas expressed as ppm or as of the last calibration FID PID FID If the TVA2020 is configured for multiple calibration points the span gas concentration values for Point 3 will be displayed The Up and Down arrow keys can be used to scroll through the span gas concentration values for other points FID PID FID 2 To change a span gas concentration value choose the detector s of interest if using a dual detector analyzer or press ENTER if using a single detector analyzer
30. expected when a span gas of known concentration is applied to the detector Finally the detector sensitivity can be calculated by subtracting the zero counts from the span counts and dividing by the span gas concentration Use the following general observations as a guideline Table 4 2 Detector Counts Detector Zero Counts Detector Sensitivity FID lt 6000 160 260 counts ppm Methane PID 10 6 eV lamp 1000 2000 10 35 counts ppm Isobutylene Example A TVA2020 FID is calibrated with zero air and a 100 ppm Methane in air span gas The counts observed for the zero are 2895 and the counts observed for the span are 27395 The span sensitivity is thus 245 counts ppm 27395 2750 100 ppm Since both of these values 2895 zero counts and 245 detector sensitivity are within the acceptable range the calibration is a good calibration Examples of a bad calibration include unusually high zero counts or unusually low detector sensitivity These problems can often be attributed to poor calibration gases contaminated sampling accessories a faulty detector capsule or failure to follow the proper calibration procedure For more information consult the Troubleshooting chapter in this manual or contact Thermo Fisher Scientific for assistance Although the FID and PID are calibrated with span gases of known concentration usually Methane and Isobutylene respectively both detectors respond to many different compounds with differing levels of
31. gt aaaaaaaaa oo0000000 hhhh rr nn lt CR gt lt LF gt B 18 TVA2020 Instruction Manual Thermo Fisher Scientific Text Data Formats Logged Data FE logged header and data GPS off lt CR gt lt LF gt FE DATA LEAK REPAIR lt CR gt lt LF gt DATE TIME ihihihihihih thththththth DET BACKGROUND CONCENTRATION LEAK SOURCE METHOD chchch lt CR gt lt LF gt isisisisisis tstststststs CSCSCS lt CR gt lt LF gt dd mmm yy hh mm ss iiiiiiiiiiiiiiii tttttttttttttttt ddd xxxxxx uuu fffffffff xxxxxx uuu ffffffffFf 1111111 LLL RRRRR ccecccc lt CR gt lt LF gt FE logged header and data with comment GPS off lt CR gt lt LF gt FE DATA i DATE TIME ihihihihihih thththththth DET BACKGROUND CONCENTRATION isisisisisis tstststststs dd mmm yy hh mm ss iiiiiiiiiiiiiiii tttttttttttttttt ddd xxxxxx uuu fffffffff xxxxxx uuu fffFFFFFF LEAK REPAIR lt CR gt lt LF gt LEAK SOURCE METHOD chchch lt CR gt lt LF gt cscscs lt CR gt lt LF gt 1111111 LLL RRRRR cccccc lt CR gt lt LF gt FE logged header and data with GPS data lt CR gt lt LF gt FE DATA DATE TIME ihihihihihih thththththth DET BACKGROUND CONCENTRATION isisisisisis tstststststs 0
32. instrument memory If the file format is incorrect the following message will appear If you choose Exit to clear the following displays If you choose to fix the files re establish the PC USB connection and open the error file that was created to identify the line s of text in error If you choose to revert to the existing files your changes will be lost unless they were saved on your PC The following screen appears Text File Formats Messages to be uploaded and downloaded must be written in straight ASCII text and must contain blank spaces between text not tabs For more information See Appendix B Text Data Formats Thermo Fisher Scientific Display Menus Memory Choosing 2 Clear Logging Memory will erase any loaded logged data stored in TVA2020 memory A warning that You are about to clear all memory will appear and must be acknowledged before the TVA2020 will clear its memory Be absolutely sure that you wish to clear memory before choosing this action Once memory is cleared it cannot be retrieved Choosing 3 Update Firmware allows the user to update instrument firmware upon new releases 1 Confirm that the new firmware image is stored locally on your hard drive 2 Before uploading the new firmware back up all the instrument files as detailed in the USB mode section These files will remain unchanged in the instrument during the upload but a back up is recommended 3 Press 3 UpdateFi
33. lt LF gt MODEL xXxxXXXXXXXX lt CR gt lt LF gt SERIAL NUMBER idididididid lt CR gt lt LF gt DETECTOR FID PID amp FID lt CR gt lt LF gt OPERATOR ID opopopopop lt CR gt lt LF gt lt CR gt lt LF gt LOGGING lt CR gt lt LF gt lt CR gt lt LF gt LOG MODE NONE AUTO VOC FE CUSTOM lt CR gt lt LF gt LOG STORAGE FORMAT NONE AUTO VOC FEVOC lt CR gt lt LF gt LOG SAMPLE TIME nnn lt CR gt lt LF gt LOG SAMPLE TIME UNIT SEC MIN lt CR gt lt LF gt LOG DATA STORED SAMPLE AVERAGE MAXIMUM lt CR gt lt LF gt LOG UNIT LOCK OFF PPM PERCENT lt CR gt lt LF gt LOG AUTO REPEAT OFF ON lt CR gt lt LF gt LOG SAVE MODE MANUAL AUTO lt CR gt lt LF gt TAG ACTIVE OFF ON lt CR gt lt LF gt ROUTE ACTIVE OFF ON lt CR gt lt LF gt lt CR gt lt LF gt CALIBRATION lt CR gt lt LF gt lt CR gt lt LF gt CAL AUTO ACCEPT OFF ON lt CR gt lt LF gt CAL AUTO SAVE OFF ON lt CR gt lt LF gt CAL UNIT LOCK OFF PPM PERCENT lt CR gt lt LF gt BACKGROUND CORRECTION PID OFF ON lt CR gt lt LF gt BACKGROUND CORRECTION FID OFF ON lt CR gt lt LF gt lt CR gt lt LF gt RESPONSE FACTORS lt CR gt lt LF gt lt CR gt lt LF gt SELECTED n lt CR gt lt LF gt MODE FACTOR CURVE lt CR gt lt LF gt RFO PID DEFAULT 1 00 0 00 lt CR gt lt LF gt RF1 PID nnnnnnnnnn xx xx yy yy lt CR gt lt LF gt RFQ PID nnnnnnnnnn xx xx yy yy lt CR gt lt LF gt RFO FID
34. procedure to access the auto log 1 From the LOG MODE display press 2 Auto The previous log selection is displayed on the second line of the screen 2 Use the up and down arrow to select minutes min or seconds sec Then type the appropriate time interval desired for the automatic sampling 3 Press ENTER and after a brief message of acceptance the automatic sampling is ready to be used in the Run mode Note To allow for STEL calculations the log rate value must be less than 90 seconds If a log rate greater than 90 seconds is selected a warning message will appear Press EXIT to bypass this message and proceed A In the LOG Auto mode data is logged in the format described below e Every time logging is initiated in the Run mode a new header is created e Data is logged periodically at the user entered log rate in the following format Thermo Fisher Scientific TVA2020 Instruction Manual 3 17 Display Menus Log Methods 4 e Approximately 9500 data samples may be taken with a single header Log VOC Use the following procedure to access the VOC Log 1 From the LOG MODE display press 3 VOC FE The previous log selection is displayed on the second line of the screen 2 Press 1 VOC The previous log selection is displayed on the second line of the screen 3 Press the Up Dn arrow keys to choose the type of sampling to be used Max Avg Smp a Max the maximum reading obtained durin
35. recording data The TVA2020 will display the reading from the selected detector and begin counting down for the time interval selected in the SETUP menu The countdown is shown on the last line of the display Once the countdown is complete the TVA2020 display will freeze on the reading to be logged and indicate if it is a leak i e if the reading of the active detector exceeds the indicated leak rate You will be given several options 8 8 TVA2020 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Repair Menus Optional Accessories Enhanced Probe Option If you choose Sav the TVA2020 will store the reading in memory replacing any previously logged readings for that Tag and proceed to the next component record in the route If you choose Again the TVA 1000B will repeat the countdown and obtain a new reading If you choose Exit the TVA2020 will return to the previous screen If the reading obtained during the logging interval exceeds the leak definition the TVA2020 will indicate that it is a leak by printing Leaker If you wish you may record what part of the component is leaking and what action was taken as a first attempt at repair Choose Repair instead of Sav and the TVA2020 will proceed to the Repair menus The two lines below the reading show the currently selected leak source and repair method Choosing Select leak source will allow you to record wha
36. same base name and a ERR extension with the following format Error line xxxx mmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmm where xxx is the line number and mmmm is an error description TVA2020 Instruction Manual B 1 Text Data Formats Route Database Route Data base The route database file is stored as ROUTE TXT with the following B 2 TVA2020 Instruction Manual format Format ROUTE DATABASE lt CR gt lt LF gt VER XxXxXxXx lt CR gt lt LF gt RUN ID aaaa lt CR gt lt LF gt lt CR gt lt LF gt ROUTE FIELDS lt CR gt lt LF gt lt CR gt lt LF gt lt route entry field 1 gt lt route entry field 2 gt lt route entry field 15 gt lt CR gt lt LF gt PICKLISTS lt CR gt lt LF gt ERE lt CR gt lt LF gt lt picklist definition 1 gt lt picklist definition 2 gt lt picklist definition 6 gt Comments Title not processed xxxxx ver right just lead spaces a 0 9 A Z a z Blank line not processed Section title not processed Section title not processed See below Blank line not processed Section title not processed Section title not processed See below lt picklist definition LeakSource gt lt picklist definition RepairMethod gt lt CR gt lt LF gt DISPLAY lt CR gt lt LF gt TE lt CR gt lt LF gt lt display entry 1 gt lt display entry 2 gt lt display entry 6 gt lt CR gt lt LF gt lt route entry header line 1 gt lt route entry header line 2 gt lt route e
37. sample log Once configured for VOC the RUN mode prompts you to type a tag identifier After you enter the tag the instrument stands ready until you initiate the logging mode with with the enhanced probe or press the ENTER key on the instrument Once LOG is initiated the instrument counts down Samples are taken over that count down and the highest reading achieved or average or last reading is logged in memory and displayed on both displays The stored values are cleared by pressing either the SELECT key on the enhanced probe or the ENTER key on the instrument Each VOC log must be done manually and individually 3B F E The F E method of sampling is a manual means of triggering a sample log using a pre configured monitor route file which has been downloaded and stored in instrument memory The F E method requires use of the enhanced probe You choose a sample time of between 2 and 30 seconds and you choose to log either the highest reading or average or last reading achieved during the sample time Once configured for F E and the RUN mode entered the FE probe menu screen guides you through the monitor route displaying each pre configured tag identifier in sequential order A SEARCH function is available at the probe menu to allow you to find tags that are out of order The probe display prompts you to confirm the tag identifier after which the instrument stands ready until you either select LOG from the probe or press the ENTER k
38. sisses penou oE irnad o a susi S aK aSa Aras nA TUENA S Rata 4 1 DCO AEC Oe ee aa e aE E E E AE E AEN E SNR 4 2 Calibration Configuration sssssssssessesisssisisisisisisisesesisesesrsesesrses 4 3 N m ber Span Bi gestae oth ope casa etre Ot ort eter 4 4 Defining the Span Gas Concentration s 0 42 scnciettiedaictachee 4 4 B ckoround Correct enaar oi eates tac a N 4 6 Defining the Background Correction ssssesesesesssesesesesereeeeseeeseressses 4 7 RE Gale M ode ins ass seteipiccedapiwervanctsioda E R eee emnene 4 8 Call Accept Modernin nirearen a eatis cele A E emt eel e 4 8 CANS AVS Modenan thane A cect cate E R wis cat snseee 4 9 Detector Counts extern ctr eet rere c ete ai eel cok nasa 4 9 Defining the Response Factor sccecsccstercdssactceadesetsaccdeeareacastunadiccatesdess 4 10 Response Factor Multiplier ssssssssesesesesesesesessssssssssrsrsrsrsrsesssesesessses 4 11 PRESB ONSET VE saan jucnsteaneavyoassobuses cabin den danuncbadest esiosast 4 11 Zero Reference Point Calibration csiccsfesissieendtatiecien taecsboase 4 14 Spat Reference Ponte oiie on tans bt E eT 4 16 Flow Calibra oS sd sete aed R a E RAN GRI 4 18 Chapter 5 Preventive Maintenance s sssssssssssnessnsuunnsununnnnnnunnnnnunnnununnnnnnnnnnnnnunnnnnnnnnnns 5 1 Safety Precautions 94 seh secs a sate R RE RGN 5 2 Replacement Parts deutsena ai aE E E E R a 5 2 Battery and Battery Charger sssesseseseeesesesesisisisisrsiseseresisesessrsesrses 5 2
39. tag may be left blank if desired FID PID FID Once the tag information is entered press the ENTER key The screen then changes to the next display The instrument is now ready to start logging The screen displays live readings and the tag name but no logging takes place until you either enter the logging mode from the enhanced probe or press the ENTER key again on the instrument FID PID FID TVA2020 Instruction Manual 3 7 Display Menus Run Mode Run Log F E 3 8 TVA2020 Instruction Manual FID PID FID Press ENTER or initiate the logging mode from the enhanced probe to start the VOC logging sequence After the instrument counts down for the time selected in the SETUP Log menu the word Save appears When confirmed the instrument logs into memory either the highest average or last reading achieved during the countdown sample After the information is stored into memory the display returns to live measurements and prompts for a new tag Each VOC log must be done manually and individually FID PID FID The operation of the Log F E mode is menu driven from the hand held probe display and requires the use of the optional TVA2020 Enhanced Probe Whenever the TVA2020 is in the RUN mode with Log F E as the logging selection the instrument logs its readings based on a pre configured route file Use the following procedure for use with Fugitive Emissions monitoring 1 Create a monitoring route using a Fugitive E
40. take 2 to 3 minutes because of flow restrictors in the tank and fill adapter Close fill adapter valve Remove TVA2020 tank Close supply tank valve Remove tank fill adapter Always remove the tank from the instrument before storing in the instrument carrying case 5 6 TVA2020 Instruction Manual Thermo Fisher Scientific Preventive Maintenance Cleaning an FID Cartridge Cleaning an FID Use the following procedure to remove and clean the FID cartridge Ca rtridge frequently Figure 5 1 Replace when needed 1 Remove the hydrogen tank Turn the instrument off 2 Using the special spanner wrench provided with the instrument tool kit unscrew the black cap holding the FID cartridge in place Figure 5 1 Replacing the FID Cartridge 3 Screw the special extractor tool provided with the accessory kit into the off center hole in the cartridge Figure 5 2 4 Remove the cartridge by pulling on the extractor Unscrew the extractor from the cartridge Clean or replace the cartridge 5 To insert a new cartridge reverse the procedure Thermo Fisher Scientific TVA2020 Instruction Manual 5 7 Preventive Maintenance Cleaning an FID Cartridge Figure 5 2 Removing the FID Cartridge To clean an FID cartridge you will need a cotton swab and some isopropyl alcohol Dip the swab into the isopropyl alcohol and insert it into the center of the cartridge Swab the surface until clean and discard the swab Take care not
41. to make repairs replacements or corrections required in whole or in part as the result of i normal wear and tear ii accident disaster or event of force majeure iii misuse fault or negligence of or by Buyer iv use of the Products in a manner for which TVA2020 Instruction Manual A 1 Warranty Warranty A 2 TVA2020 Instruction Manual they were not designed v causes external to the Products such as but not limited to power failure or electrical power surges vi improper storage and handling of the Products or vii use of the Products in combination with equipment or software not supplied by Seller If Seller determines that Products for which Buyer has requested warranty services are not covered by the warranty hereunder Buyer shall pay or reimburse Seller for all costs of investigating and responding to such request at Seller s then prevailing time and materials rates If Seller provides repair services or replacement parts that are not covered by the warranty provided in this warranty Buyer shall pay Seller therefor at Seller s then prevailing time and materials rates ANY INSTALLATION MAINTENANCE REPAIR SERVICE RELOCATION OR ALTERATION TO OR OF OR OTHER TAMPERING WITH THE PRODUCTS PERFORMED BY ANY PERSON OR ENTITY OTHER THAN SELLER WITHOUT SELLER S PRIOR WRITTEN APPROVAL OR ANY USE OF REPLACEMENT PARTS NOT SUPPLIED BY SELLER SHALL IMMEDIATELY VOID AND CANCEL ALL WARRANTIES WITH RESPECT TO THE AFFECTED PRODUCTS
42. to touch the igniter coil which is located close to the end of the capsule opposite the threaded hole used to remove capsule Then dry the cartridge in an oven at 45 C to 55 C for one hour When dry re insert the cartridge into the instrument reversing the removal procedure Do not overtighten cap cleaning A CAUTION Do not allow any contact with the igniter coil during 5 8 TVA2020 Instruction Manual Thermo Fisher Scientific Preventive Maintenance Cleaning the FID Detector Cap Cleaning the FID Use the following procedure to clean the FID detector end cap Figure 5 Detector Cap 1 Remove the hydrogen tank from the top of the instrument and turn off the power 2 Using the special spanner wrench provided with the tool kit unscrew and remove the FID detector cap 3 Clean the cap using the isopropyl alcohol followed with a de ionized or distilled water rinse Blow out carefully with compressed dry air 4 Replace the cap Figure 5 3 Cleaning the FID Detector Cap Thermo Fisher Scientific TVA2020 Instruction Manual 5 9 Preventive Maintenance Cleaning the FID Detector Cavities Cleaning the FID Detector Cavities 5 10 TVA2020 Instruction Manual Use the following procedure to clean the FID detector cavity 1 Remove the hydrogen tank from the top of the instrument Turn the instrument off 2 Using the special spanner wrench provided with the tool kit unscrew the cap holding the detector cap 3
43. two most common response factor formats 1 Response Factor Multiplier The response factor multiplier is the most commonly used and most widely recognized response factor format This type of response factor is calculated by dividing the actual concentration of a compound introduced to the analyzer by the detector response TNN Actual Concentration Response Factor Multiplier _ _ Measured Response For example if 100 ppm of a compound produces a 50 ppm FID reading on an analyzer calibrated with methane then the FID response factor for that compound would be 2 00 100 divided by 50 When a using a response factor multiplier to correct a TVA reading simply multiply the analyzer measured response by the response factor As in the example given above if a 50 ppm reading is achieved multiply 50 times 2 00 to get the actual concentration 100 ppm Response Response Factor Mulitplier is what the TVA2020 uses TVA2020 Instruction Manual C 1 Response Factors Response Factors C 2 TVA2020 Instruction Manual 2 Relative Response Factor In some cases a response factor may be expressed as a percent or a ratio of a compound s response to that of the calibration gas This type of response factor is calculated by dividing the detector response by the actual concentration of a compound of interest Measured Response Relative Response Factor Actual Concentration Multiply by 100 to express as a perce
44. usage and environmental conditions vary greatly the components should be inspected frequently until an appropriate maintenance schedule is determined Thermo Fisher Scientific suggests that you return the entire instrument to the factory or an authorized service center once a year for cleaning testing and calibration This chapter includes the following maintenance information and replacement procedures e Safety Precautions on page 5 2 e Replacement Parts on page 5 2 e Battery and Battery Charger on page 5 2 e Hydrogen Gas Tank on page 5 4 e Cleaning the PID Lamp on page 7 9 e Cleaning an FID Cartridge on page 5 7 e Cleaning the FID Detector Cap on page 5 9 e Cleaning the FID Detector Cavities on page 5 10 e Replacing Water Trap Probe Filter and O Rings on page 5 11 e Cleaning or Replacing a Sintered Metal Filter Cup on page 5 13 e Replacing the Charcoal Filter on page 5 14 Thermo Fisher Scientific TVA2020 Instruction Manual 5 1 Preventive Maintenance Safety Precautions Safety Precautions gt gt Replacement Parts Battery and Battery Charger gt gt D Battery Warnings and Precautions gt 5 2 TVA2020 Instruction Manual Read the safety precautions before beginning any procedures in this chapter WARNING Never change an electrical component in an area classified as hazardous due to presence of flammable gases or vapors A WARNING
45. zero can be entered for each detector The use of a response curve is thus unnecessary as the detector is already reading the direct ppm for that specific compound CAUTION If you use multipoint calibration or a gas other than methane or isobutylene and then apply response factors curves that have been generated with reference to a single point methane isobutylene calibration the resulting measurements will probably be incorrect A To provide the specified accuracy the instrument must be calibrated at the beginning of each workday e From the Main Menu choose 2 Setup gt 1 Calib When you reach the CALIBRATION menu you will see the following selections The steps involved in calibrating the TVA2020 are as follows 1 Configure the calibration variables 6 Cfg e Number or span points 1 9 e Span concentration e Background correction None PID FID Both e Response factor calculation mode Multiplication factor or Curve e Accept mode Manual or Auto e Save mode Manual or Auto 2 Zero the instrument using either a zero gas or clean ambient air 1 Zero Thermo Fisher Scientific Calibration Calibration Configuration 3 Calibrate the reference point s using known span gases The TVA2020 can be configured for as many as nine different span gas values 2 Span 4 Optional Set instrument response factors if necessary 4 RF 5 Optional Take background reading 3 Backgnd 6 Optional Calibrate the
46. 1 Identify the four self tapping screws securing the internal frame to the bottom enclosure 2 Remove the four screws Save for reuse 3 Unscrew the two captive screws securing the d sub connector to the battery 4 Remove and replace the battery Reverse the above procedure for reassembly FID instruments are supplied with an 85 cc hydrogen gas tank This tank which may be pressurized to 2200 psi maximum at 25xC will provide 10 hours operation when fully charged The tank has an integrally mounted high pressure gauge that can be easily read when the tank is in or out of the instrument Install the tank in the instrument by inserting it into the receptacle on the top and tightening left hand threads tighten counterclockwise until the rubber tank boot is flush with the instrument sidepack and a slight resistance is felt Do not overtighten Thermo Fisher Scientific Servicing Hydrogen Gas Tank Precautions in The major hazard associated with the handling of hydrogen is flammability Handling and The following specific rules apply when handling hydrogen Never use cylinders of hydrogen in areas where flames excessive heat Storage or sparks may occur e Use only explosion proof equipment and spark proof tools in areas where hydrogen is handled e Ground all equipment and lines used with hydrogen e Never use a flame to detect hydrogen leaks use soapy water e Do not store reserve stocks of hydrogen with cylinders co
47. 10 to 45 C 10 and 45 C 20 to 60 C 4 and 140 F Ambient pressure 860 to 1060 mbar 70 to 108 kPa 20 to 108 kPa Relative humidity 50 10 FID 20 to 95 15 and 95 0 to 100 PID 20 to 70 noncondensing noncondensing Radiated Susceptibility None 3V m 80 to 2700 MHz Conducted susceptibility Not applicable battery operated Conducted emission Not applicable battery operated Radiated emission See Part 15 FCC Bluetooth requirement ESD Sensitivity gt 6000 Volts Battery charger supply 120 or 230 1 Vac 100 to 240 Vac 100 to 240 Vac Not applicable voltage Battery Charger supply 50 60 Hz 50 60 Hz 50 60 Hz Not applicable frequency 0 5 Hz Thermo Fisher Scientific TVA2020 Instruction Manual 1 11 Introduction Product Safety Specifications Product Safety Table 1 3 lists the specifications for the TVA2020 Refer to Figure 1 4 for Sp ec ifi c ati ons Interconnection Control Drawing Table 1 3 TVA2020 Product Safety Specifications Testing Laboratory Types of Protection Conditions of Classification and Area Classification FM intrinsically safe for Class Division 1 Temperature Class T4 Groups A B C and D ASSY TVA2020 57 011966 XXXX_ 57 011966 XXXX 1 SSY BARRIER IS USB TVA2020 114295 00 1142500 3 JASSY POWER SUPPLY TVA 111401 00 11140100 DESCRIPTION DWG NO PART NO ABLE CHARGER 12693 00 11269300 105639 00 2 REF ADDED IS SAFETY AND FM NOTES 2 10 S INSTAL RELEASE 8 N SCRIPTION DATE
48. 16 oxygen is required to support the flame If underground gases or samples in gas bags are to be measured by an FID it is advised that a dilutor be used to combat the problem Thermo Fisher Scientific Introduction Flame lonization Detection FID EXHAUST METER COLLECTOR ELECTRODE IGNITION FILAMENT AIR SAMPLE FUEL H2 Figure 1 1 Typical Flame lonization Detector Benefits of Flame Benefits of Flame Ionization Detection include lonization Detection wide dynamic and linear range e High sensitivity to hydrocarbon vapors including methane e Very stable and repeatable response e Virtually unaffected by ambient levels of CO CO2 and water vapor Thermo Fisher Scientific TVA2020 Instruction Manual 1 3 Introduction Photoionization Detection Photoionization Detection 1 4 TVA2020 Instruction Manual A Photoionization Detector PID consists of an ultraviolet UV lamp of a specific energy and an ionization chamber see Figure 1 2 Compounds passing through the chamber are excited by photons of UV energy and ionized according to the following equation R ho gt R e where R most organic inorganic compounds ese ions are attracted to a collecting electrode producing a curren Th ttracted t llecting electrod ducing t proportional to the concentration of the compound Whether or not a compound can be detected by a PID depends upon the energy required to remove an electron from the compound
49. 2020 Instruction Manual 6 1 Troubleshooting Troubleshooting Guide Troublesh ooting Table 6 1 provides general troubleshooting information and indicates the Guide checks that you should perform if you experience an instrument problem Table 6 1 Troubleshooting Problem Unit will not turn on Possible Reason Battery charge low Bad battery connection Bad battery Blown fuse or faulty keypad Solution Action Charge battery Ensure proper battery connection Replace battery Call Thermo Fisher Scientific Service Department Pump won t turn on Defective pump Faulty keypad Call Thermo Fisher Scientific Service Department Call Thermo Fisher Scientific Service Department Low pump flow Clogged sample intake Clean replace filter cups Clean replace sample line Clean replace flame arrestor Service pump Pump fault Call Thermo Fisher Scientific Service Department Keypad will not respond Faulty keypad Call Thermo Fisher Scientific Service Department Probe display blank or probe buttons will not respond Faulty probe display assembly Replace probe display assembly Call Thermo Fisher Scientific Service Department FID won t ignite 6 2 TVA2020 Instruction Manual Insufficient sample flow Insufficient hydrogen supply lt 500 psi Hydrogen leak Turn pump on Clean replace filter cups Clean replace flame arrestor Service pump Check hydrogen gauge o
50. 8 Standard Specifications vices iiudoiiaicesapceniaupiendel aheenieaecetke 1 9 External TMM CES sh sehtuniacieatansasueaesdl vanbyaasaniuas comin aieiaeo iiia 1 11 Product Safety Specificationsis 0 iia heidi eae 1 12 Hardware and Startup sssscsscssssssssssssssssssssssssssssesseeesesessssneeseesesateaneaees 2 1 OVERVIEW e sn svedisncoena dian sentuadeds A R E chelate 2 1 Instrument PUTICTIO NS crit cia dee ester dace sesraceaveueiecader Serato ie anias 2 3 Using the Instrumenter ca cs eee eae iat pce aes uated 2 4 Stat poani enot iat a E EEE EE EEE Wins EEE RE irae 2 4 Instrument Connections ssesssssssssessssseesseesreseesseeseeseessesseserssressessesees 2 4 Instrument Sidepack Displayis ii cciaiaieepneicdine tae ncaenatecen 2 6 Instrument Sidepack Key pad 30s winced aide wnt esd dake 2 7 Conto MENU hase a ti ates E TTE O ARE 2 9 Quick Stare Procedure nsona e e E A nas 2 10 Display DM NWS aissis egresada aniani 3 1 Main Menu Structur eenei ient aa Mia anit eaaa 3 1 R n Modedern aea a PD A PINE TERT CED N A 3 4 Accessing the Run Ment cns so Olle eee a t 3 5 R n Log IN Gi sac ates seaen g i aah aaa Aa A A AAN 3 5 RUNS Log Ate siescssensiaresdiieudeatt iuasepetsawercaanbapeved a aa 3 5 R n Log VOC aneen oas eie k abep i enro iTia Sei Eosi 3 6 Bile Log 7B Ernenguineon e e A A ai 3 8 Set p Meit aenda an a a a s 3 9 Calibration Settings a accede lad Diets ca E R aes hea 3 9 Alarm Tee ited ol tase are aus nest nai ath Gah Te cae
51. APPROVED ThermoFisher 27 foge Pamay 2 ca TesrmoFiter rarua N a Zar INTERCONNECTION CONTROL NOTES Ik DRAWING TVA2020 CONNECTIONS ALLOWED ONLY IN UNCLASSIFIED LOCATION CECA TO MAINTAIN IS SAFETY RATING LISTED USB BARRIER REQUIRED FOR CONNECTIONS TO TVA2020 USB PORT Pee fe pon ETH ao ME TO MAINTAIN IS SAFETY RATING LISTED CHARGER REQUIRED FOR CHARGING TVA2020 BATTERY once eae coum NO CHANGES ALLOWED WITHOUT PRIOR FM APPROVAL fa bo oe aoe a Toft Figure 1 4 Interconnection Control Drawing 1 12 TVA2020 Instruction Manual Thermo Fisher Scientific Chapter 2 Hardware and Startup For a description of the instrument functions instrument connections and how to start the instrument see the following topics e Overview on page 2 1 e Instrument Functions on page 2 3 e Using the Instrument on page 2 4 e Instrument Connections on page 2 4 e Instrument Sidepack Display on page 2 6 e Instrument Sidepack Keypad on page 2 7 e Control Menu on page 2 9 e Quick Start Procedure on page 2 10 Overview The vapor concentration may be read immediately on either of two displays one mounted directly on the optional enhanced sample probe and the other on the instrument sidepack itself see Figure 2 1 Vapor concentration can be displayed on both displays in parts per million ppm or percent concentration The data displayed may also be collected and saved in analy
52. CENTRATION 0500 10 50 PPM OK 0 03 PPM OK 5 30 PPM OK 0501 103 PPM HIGH ALRM 0 03 PPM OK 99 30 PPM OK 0502 25 67 PPM OK 0 04 PPM OK 5 30 PPM OK TVA2020 Instruction Manual B 21 Text Data Formats Logged Data LOGGED DATA VER 2 00 FE DATA DATE TIME TAG B 22 TVA2020 Instruction Manual DET PID PID PID PID PID PID BACKGROUND Sample FE data LEAK CONCENTRATION PPM DET FAIL PPM DET FAIL PPM DET FAIL PPM DET FAIL PPM DET FAIL PPM DET FAIL REPAIR LEAK SOURCE METHOD COMMENT N A THIS IS MY FIRST COMMENT 1234560 N A GUNP N A ALL FIXED N A A N A Thermo Fisher Scientific Thermo Fisher Scientific Appendix C Response Factors The TVA2020 is a portable vapor analyzer for gas survey monitoring Each unit is factory calibrated with Methane in the case of the FID and Isobutylene in the case of the PID However both detectors respond to many different compounds with differing levels of sensitivity In order to adjust the analyzer reading from ppm of Methane or ppm of Isobutylene to ppm of the compound of interest a correction factor must be applied to the reading This correction factor is also known as a Response Factor Response factors can be expressed in several different formats When you acquire a response factor from any source it is important that you understand the format in which the response factor is expressed so that it can be applied properly Following are the
53. Chapter 7 Servicing Thermo Fisher Scientific TVA2020 Instruction Manual 1 1 Introduction Flame lonization Detection FID Flame lonization Detection FID 1 2 TVA2020 Instruction Manual A Flame Ionization Detector FID measures organic compounds by utilizing a flame produced by the combustion of hydrogen and air see Figure 1 1 When hydrocarbons in the sample are introduced to the detection zone ions are produced by the following reaction RH O gt RHO e gt HO CO where R carbon compound A collector electrode with a polarizing voltage is also located within the detector chamber and the ions produced by this reaction are attracted to it As the ions migrate towards the collector a current is produced which is directly proportional to the concentration of hydrocarbons introduced to the flame This current is then amplified and sent to a microprocessor The FID has a wide dynamic range The effective dynamic range can be further expanded by use of a dilutor kit which reduces very high volatile organic compound VOC concentrations to within the dynamic range or even linear range of the analyzer The dilutor kit can also be used to enrich oxygen deficient samples by adding ambient air that is rich in oxygen 20 9 usually Low oxygen can affect the characteristics of the hydrogen flame causing readings to be artificially elevated and possibly extinguishing the flame As a general rule of thumb greater than
54. It indicates that approximately 15 minutes of operating time remain Press EXIT to return to the operating display The lower right corner of all displays now show blinking If you continue to operate after the warning is displayed the unit will eventually shut off automatically To restart after an auto shut down you must first recharge or replace the battery Note that collected data is not lost Logger memory is already full Log memory is full You cannot continue to log any entries until you have transferred the log memory to a PC or have cleared log memory The instrument will however continue to operate FID Flameout Ignite again This display appears whenever the FID flame is extinguished Press EXIT to clear then CONTROL key and 2 to ignite This warning is active only in RUN and CALIBRATE modes In the RUN mode once the alarm is acknowledged a flashing F appears on the sidepack display and the probe display flashes Low Flow detected This message will follow the FID Flameout message if a low flow condition was measured prior to the flameout High Concentration detected This message will follow the FID Flameout message if the instrument sees a spike in the concentration prior to a flameout condition Log rate too slow for PC STEL calc This message appears whenever you enter a log rate value greater than 90 seconds The data sampling rate is too low for valid STEL calculations in the PC P
55. Keypad ccs ube as ks ee a A eh 2 7 Figure 3 1 TVA2020 Menu Structure 0 c ceccccccsccsesessescesestesesteesteseseeeeseeees 3 3 Figure 5 1 Replacing the FID Cartridge ooo ccc cececescetescetestetesteeeteneees 5 7 Figure 5 2 Removing the FID Cartridge oo c cceccccecescecesceteseetestetesteteeeeseees 5 8 Figure 5 3 Cleaning the FID Detector Cap 001sesiisesiisesiiseeriereiresrirennenn 5 9 Figure 5 4 Water Trap Probe Assembly Filter and O Ring Replacing 9 11 Figure 5 5 Water Trap Probe Assembly Filter and O Ring Replacing 5 12 Figure 5 6 Sintered Metal Filter Cleaning or Replacing cccccccee 5 13 Figure 5 7 Activated Charcoal Filter Adapter 0 c cccccccecceceestesesteeeeeeee 5 14 Figure 7 1 TVA2020 Component Layout 0 c ccccccccccsscesesceteststetesteteeeeneeees 7 4 Figure 7 2 Hardware o ccccccccceccscscscesesescssesesesceeseseseaseseseseetesesessssseateneseseees 7 5 Figure 7 3 Removing the PID Sensor uo ccceccceccsceeseeestetestesesteesteseeeeneeess 7 9 Figure 7 4 Removing the Filter Cap oo ccc cccccescscescetesceteseeesteesteeseeseaees 7 10 Figure 7 5 Removing the Filter Media ccc cceceesceteesestetesesesteeseseees 7 10 Figure 7 6 Removing the Spacer oo ccccccccccccscesessesesteseetesesesteeeeeeeeeseeees 7 10 Figure 7 7 Removing the Cell ASS MDIY ccc ccccecececeseeteseeteseeteeeeeeees 7 11 Figure 7 8 Removing the PID Lam
56. Main Menu Structure on page 3 1 e Run Mode on page 3 4 e Setup Menu on page 3 9 e Calibration Settings on page 3 9 e Alarm Levels on page 3 10 e Log Methods on page 3 15 e Passcode Protection on page 3 23 e Other Settings on page 3 24 e Information Menu on page 3 29 e Memory on page 3 30 Main Menu The display on the TVA2020 analyzer is a menu driven device The various Structure meres Prompt you to select or enter information With various key strokes you can accomplish all necessary setup configuration and operational tasks For more detail see the following topics Thermo Fisher Scientific TVA2020 Instruction Manual 3 1 Display Menus Main Menu Structure Whenever you turn on the TVA2020 a power up self test is performed and warning screens are displayed if any errors are found Please refer to Warning Messages on page 6 6 If errors are detected use the EXIT key to clear error message then MAIN MENU is displayed When you press the corresponding number key as described in Table 3 1 the desired menu and the associated display or menu appear automatically Table 3 1 Main Menu Selection and Functions Key 1 RUN Function Use this selection to assign tags to specific surveys and to view log analysis of organic inorganic compounds If the optional Bluetooth module is active the instrument must be in RUN mode to send commands and capture data See
57. PPPPPP e fEFFFFFTFE e SSSSSSSSS OK u DET OFF DET FAIL OVERFLOW UNDERFLOW BAD CALIB CAL SLOPE CLAMPED AVG_OVFLW HIGH ALRM LOW ALARM STEL ALRM HIGH amp STEL LOWSSTEL PID concentration floating point right justified leading spaces or when PID detector not selected FID concentration floating point right justified leading spaces or when FID detector not selected data status detector not selected no errors or alarms detector not valid detector failed not DET OFF A D overflow not DET FAIL A D underflow not OVERFLOW invalid calibration parameters calibration sensitivity out of range an intermediate concentration exceeded 100 during log point averaging an overflow occured high alarm no STEL not UNDERFLOW low alarm no high alarm or STEL not UNDERFLOW STEL alarm no low or high alarm not UNDERFLOW high alarm and STEL not UNDERFLOW low alarm and STEL no high alarm not UNDERFLOW Digital data logging is initially disabled and the logging type is set according to what detectors are installed so e g if only an FID is installed the logging type will be set to FID data only logging Thermo Fisher Scientific TVA2020 Instruction Manual 8 15 Optional Accessories GPS 8 16 TVA2020 Instruction Manual GPS An optional Global Positioning System GPS module may be installed inside the sidepack to add locati
58. Parts replacement and maintenance should not be performed in areas classified as hazardous due to presence of flammable gases or vapors Use only Thermo Fisher Scientific replacement parts Extended Service Plans are available contact Thermo Authorized Service for further information A See the Servicing chapter for a list of replacement parts You may charge the battery in or out of the instrument or replace it with a charged battery For information on replacing the battery see Replacing the Battery on page 7 6 WARNING Do not replace battery in an area classified as hazardous due to presence of flammable gases or vapors A WARNING Do not operate battery charger in an area classified as hazardous due to presence of flammable gases or vapors A WARNING Charger and USB connections are not allowed in an area classified as hazardous due to presence of flammable gases or vapors A Various safety features have been included in the design of the lithium ion battery packs However misusing or mishandling the battery packs can lead to fluid leakage heat generation fire or an explosion WARNING Misusing or mishandling the battery packs can lead to fluid leakage heat generation fire or an explosion A Thermo Fisher Scientific Thermo Fisher Scientific Preventive Maintenance Battery and Battery Charger To prevent these situations from occurring and to ensure safe use of the battery packs observe the following precaut
59. Place the filter media over the cell assembly centered on top of the sensor Make sure the filters are installed in the correct order Filter media 2 first then filter media 1 on top with the shiny side up Figure 7 14 Figure 7 14 Placing the Filter Media 13 Align the cap key with the notch on the housing Starting at the side opposite the notch press down until the filter cap snaps on to the housing If the cap key is incorrectly aligned there will be a noticeable bulge on the side of the cap Figure 7 15 Figure 7 15 Aligning the Cap Key TVA2020 Instruction Manual 7 13 Servicing Replacing the FID End Cap Replacing the FID End Cap Flow Tests 7 14 TVA2020 Instruction Manual The flame arrestor located in the center of the FID end cap is a critical safety feature to prevent flame propagation into a potentially hazardous area It has been precision fit into the end cap and must be replaced as an assembly 1 Remove the hydrogen tank from the top of the instrument and turn off the power 2 Using the special spanner wrench provided with the tool kit unscrew and remove the FID detector cap 3 Replace the end cap as a full assembly Most flow problems can be traced to a loose connection or dirty filters Based on that premise this section is organized from the least to the most invasive repair procedures Reference the Sample Flow Troubleshooting Flowchart Figure 7 16 to diagnose flow problems Sta
60. RODE AIR SAMPLE IN HYDROGEN SUPPLY Figure 1 3 TVA2020 Dual Detector Configuration Benefits of Dual Benefits of Dual Detectors Detectors Cost effective packaging e Detector response ratios can help characterize compounds e Enhanced analytical capability derived from simultaneous detection Thermo Fisher Scientific TVA2020 Instruction Manual 1 7 Introduction Concentration Calculation and Calibration Concentration Calculation and Calibration 1 8 TVA2020 Instruction Manual The concentration calculation is defined as the process that transforms a detector count into a displayed final concentration The process is as follows DETECTOR COUNT TRANSLATED INTO A STANDARD DET ECT OR CONCENTRATION USING A DETECTOR COUNT VS DETECTOR CONCENTRATION CALIBRATION CURVE STANDARD DETECTOR CONCENTRATION IS CORRECTED TO A SPECIFIC GAS USING THE INSTRUMENT RESPON SE FACTOR SPECIFIC GAS CONCENTRAT ION IS BACKGROUN D C ORRECTED IF ENABLED INSTRUMENT RESPONSE FACTOR AND BACKGROUND CORRECTED C ONCENTRA TION IS RF CORRECTED USING THE CURRENT ROUTE ENTRY RF IF ANY FINAL CONC ENTRATION READING Thermo Fisher Scientific Specifications Thermo Fisher Scientific Introduction Standard Specifications Standard Table 1 1 lists the standard specifications for the TVA2020 P Item Accuracy Table 1 1 TVA2020 Product Specifications Description PID Instrument 20 of reading or 0 5 ppm whichever is great
61. SCODE MENU display press 1 New Passcode Enable allows you to activate the passcode protection feature Once this feature is activated you must enter your passcode whenever you request access to the SETUP MNEU Once passcode protection has been enabled and you exit the SETUP MENU the option can only be disabled by entering the passcode to gain access to the SETUP MENU again REMEMBER YOUR PASSCODE TVA2020 Instruction Manual 3 23 Display Menus Other Settings Disable Other Settings User Identification Number 3 24 TVA2020 Instruction Manual Disable allows you to deactivate the passcode protection feature Once this feature is deactivated you have unlimited access to the SETUP MENU without entering a passcode Note In the event you inadvertently forget your passcode contact Thermo Fisher Scientific for assistance A Other Settings include e 1 User Identification Number e 2 Date e 3 Time e 4 User Options Note Time may be set only if the date is within the valid range of 1980 to 2037 A Use the following procedure to set the user identification number Note To set the User ID number the instrument must be ON but does not have to be warmed up A 1 From the MAIN MENU display press 2 Setup 2 From the SETUP MENU display press 5 Other 3 From the OTHER SETTINGS menu press 1 User ID Thermo Fisher Scientific Display Menus Other Settings 4 From the User ID prompt press ENTER 5 Use th
62. SY Figure 5 6 Sintered Metal Filter Cleaning or Replacing Thermo Fisher Scientific TVA2020 Instruction Manual 5 13 Preventive Maintenance Replacing the Charcoal Filter Replacing the Charcoal Filter 5 14 TVA2020 Instruction Manual Use the following procedure to replace the optional charcoal filter 1 Remove the sampling assembly from the probe by loosening the probe nut and pulling the sampling assembly free from the probe 2 Unscrew the cap from the sampling assembly This will expose the end of the charcoal filter cartridge Figure 5 7 3 Tip the cartridge up and empty the charcoal from the cartridge 4 Clean the inside of the cartridge with a cotton swab dipped in isopropyl alcohol 5 Dry the cartridge thoroughly 6 Re fill the cartridge with new dust free charcoal of the type specified for the filter 7 Apply PTFE tape to the male threads of the charcoal cartridge and screw on the cap 8 Re insert the sampling assembly into the probe and tighten the probe nut THREAD ED EN D WITH CARTRIDGE TEFLON CAP FILLED WITH TAPE ACTIVATED CHARCOAL Figure 5 7 Activated Charcoal Filter Adapter Thermo Fisher Scientific Thermo Fisher Scientific Chapter 6 Troubleshooting This chapter provides the following troubleshooting and service support information e Troubleshooting Guide on page 6 2 e Warning Messages on page 6 6 e Service Locations on page 6 7 TVA
63. The following display will appear TVA2020 Instruction Manual 4 5 Calibration Calibration Configuration Background Correct 4 6 TVA2020 Instruction Manual FID PID FID Use the up and down arrow keys to select the measurement units ppm or and the decimal point position There are 3 selections to choose from Table 4 1 Gas Concentration Units Range Display Units 3 DD DD 2 DDDDDD ppm 1 DDDD D ppm The range of the instrument is determined by your selection of measurement units and decimal point placement In the run mode the instrument will auto range upward only to select the optimum range for displaying measurement information For example if you entered a value in Range 1 and the measured value changes to a high value the instrument will automatically switch to range 2 If it increases further it will switch to Range 3 After you have selected the measurement units and decimal point placement use the keypad to enter the concentration value Press ENTER to accept this value and store it in the TVA s memory 3 Ifyour TVA2020 is a dual detector analyzer you may repeat the procedure for the second detector If your TVA2020 is configured for more than one span point you may use the Up Dn arrow keys to select the next span point and repeat the procedure 4 When the last gas concentration value has been entered you can return to the CALIBRATION menu by pressing the EXIT key This selection allow
64. a Logging detector pid Log PID data only detector fid detector pid_fid Log FID data only Log PID and FID data ignite Reignite FID pump on Turn pump on pump off Turn pump off screen Enter screen mode In screen mode the sidepack screen is shown and is updated when the sidepack screen is changed or every two seconds push co control push do down push en enter push ex exit push le left push ri right push up push 1 push 2 push 3 push 4 push 5 push 6 push 7 Send key to sidepack screen mode only TVA2020 Instruction Manual 8 13 Optional Accessories Bluetooth Communications Command Result push 8 push 9 quit Exit screen mode Note The commands are case insensitive A All commands are terminated by lt CR gt or lt CR gt lt LF gt If an invalid command is received by the TVA it will respond with bad cmd This response is terminated by lt CR gt lt LF gt 8 14 TVA2020 Instruction Manual Thermo Fisher Scientific Digital Data Logging Bluetooth Optional Accessories Bluetooth Communications If digital data logging is enabled streaming concentration data is sent out via Bluetooth at the end of each measurement cycle once per second when the analyzer is in RUN mode The digital data format is as follows Digital data log record pppppppppp sssssssss ffffffffff sssssssss aaaaaaaaa 000000000 hhhh rr nn lt CR gt lt LF gt where PPPP
65. ading spaces No Comms Module installed but no communications No Data Module installed but no data No Power Module installed but power off Comm Err Module installed but messages not recognized oo00000000 GPS longitude ddd ddddd floating point right justified leading spaces z Module installed but no comms or data hhhh GPS altitude 999 to 9999 meters right justified leading spaces Module installed but no comms or data rr GPS Resolution 5 to 99 meters right justified leading spaces Module installed but no comms or data nn Number of GPS satellites O to 20 right justified leading spaces Module installed but no comms or data Sample auto and VOC data low alarm 75ppm high alarm 100ppm PID CONCENTRATION PID BACKGROUND FID CONCENTRATION FID BACKGROUND 10 02 PPM OK 0 02 PPM O 12 03 PPM O 0 43 PPM O 35 02 PPM OK 0 02 PPM O 40 23 PPM O 0 43 PPM O 75 02 PPM LOW ALARM 0 02 PPM O 50 11 PPM O 0 43 PPM O 105 PPM HIGH_ALRM 0 02 PPM O 80 03 PPM LOW ALARM 0 43 PPM O 110 PPM HIGH_ALRM 0 02 PPM O 101 PPM HIGH_ALRM 0 43 PPM O 80 12 PPM LOW ALARM 0 02 PPM O 91 23 PPM LOW ALARM 0 43 PPM O 60 09 PPM OK 0 02 PPM O 74 98 PPM O 0 43 PPM O 50 04 PPM OK 0 03 PPM O 62 03 PPM O 0 39 PPM O 20 03 PPM OK 0 03 PPM O 41 23 PPM O 0 39 PPM O 15 08 PPM OK 0 03 PPM O 25 98 PPM O 0 39 PPM O 9 28 PPM OK 0 03 PPM O 17 03 PPM O 0 39 PPM O TAG PID CONCENTRATION PID BACKGROUND FID CON
66. al B 13 Text Data Formats Factory SUM OF SQUARES 0 934123 END B 14 TVA2020 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Text Data Formats Calibration History Calibration The calibration history data is stored as HISTORY TXT in a read only History format This file is generated from data stored in MRAM each time the instrument is put into USB transfer mode thus not allowing a user to change the data in the HISTORY TXT file If there are fewer history records than the maximum allowed 240 the empty records are not output Format Comments CAL HISTORY lt CR gt lt LF gt Title not processed VER xXx Xx lt CR gt lt LF gt XXxxx ver right just lead spaces lt CR gt lt LF gt lt cal history header line 1 gt See below lt cal history header line 2 gt See below lt cal history header line 3 gt See below lt cal history data record 1 gt See below lt cal history data record 240 gt lt CR gt lt LF gt Blank line not processed END lt CR gt lt LF gt Marks end of text Cal history header CALIB KNOWN CALIB MEASURED lt CR gt lt LF gt DET TYPE CONC DATE TIME CALIB VALUE lt CR gt lt LF gt Pict esse state ee set oa See ee tes a ee eee lt CR gt lt LF gt Cal history data record ddd ttttt aaaaaa uuu dd mmm yy hh mm ss bbbbbbbb COUNTS fffffffff lt CR gt lt LF gt See calibration file description for field details except for CALIB TYPE which is note the indicates that the cali
67. al point position then type the numeric value for the alarm level desired 3 Press ENTER to store new values into instrument memory 4 Press EXIT to return to the ALARM LEVELS without making a change Thermo Fisher Scientific Display Menus Alarm Levels Low Ceiling Alarm Use the following procedure to access the low ceiling alarm 1 From the ALARM LEVELS display press 2 Low Ceiling The previous alarm settings for the Low Ceiling alarm are displayed FID PID FID 2 To change the alarm level to a new value press ENTER on a single detector instrument or 1 2 or 3 on a dual detector instrument FID PID FID Use the up and down arrow keys to select ppm or and decimal point position and then type the numeric value for the alarm level desired 3 Press ENTER to store the new values into instrument memory 4 Press EXIT to return to the ALARM LEVELS without making a change Thermo Fisher Scientific TVA2020 Instruction Manual 3 13 Display Menus Alarm Levels High Ceiling Alarm Use the following procedure to access the high ceiling alarm 1 From the ALARM LEVELS display press 3 High Ceiling The previous alarm settings for the High Ceiling alarm are displayed FID PID FID 2 To change the alarm level to a new value press ENTER on a single detector instrument or 1 2 or 3 on a dual detector instrument FID PID FID Use the up and down arrow keys to select ppm or and decimal point position
68. and then type the numeric value for the alarm level desired 3 Press ENTER to store the new values into instrument memory 4 Press EXIT to return to the ALARM LEVELS display without making a change Note The high ceiling alarm cannot be set higher then 30 000 ppm or 3 A 3 14 TVA2020 Instruction Manual Thermo Fisher Scientific Display Menus Log Methods Log Methods Selecting the LOG mode not only identifies the way in which data is stored in instrument memory but also governs the way the RUN mode works Four possible choices in the LOG mode are available None Auto VOC FE and Custom Table 3 3 Available Log Data Types Option 1 None Description Selecting None means that no logging action takes place at all When the instrument is in the RUN mode you can only display readings 2 Auto When you select and configure the AUTO method the RUN mode functions automatically once an action is initiated You choose an automatic logging rate between 1 second and 999 minutes type a tag identifier and start the instrument You initiate automatic sampling by either pressing the LOG key on the probe or the ENTER key on the instrument From that point on the instrument automatically logs the reading of the sample at the end of each countdown and continues to do so until the battery goes low you press the EXIT key or the logging memory is full 3A VOC The VOC method of sampling is a manual means of triggering a
69. arts may adversely affect the sensors operation Latex gloves are preferred but if they are not used your hands must be clean and free of oils lotions etc It is acceptable to hold the lamp by its glass body or by the edges of the window A 1 Turn the instrument OFF and remove the sensor from the instrument Figure 7 3 Figure 7 3 Removing the PID Sensor 2 Remove the filter cap by applying slight upward pressure with the tip of a screwdriver or X Acto blade just below the hole in the cap and between the cap and housing it will pop off Figure 7 4 TVA2020 Instruction Manual 7 9 Servicing Cleaning the PID Lamp Figure 7 4 Removing the Filter Cap 3 With fine tipped tweezers remove both filter media and set aside Figure 7 5 Figure 7 5 Removing the Filter Media 4 Using the X Acto blade remove the spacer and set aside Figure 7 6 Figure 7 6 Removing the Spacer 5 With fine tipped tweezers carefully remove the cell assembly by prying under the cell s edge where connector pins are located Figure 7 7 7 10 TVA2020 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Servicing Cleaning the PID Lamp Figure 7 7 Removing the Cell Assembly 6 With fine tipped tweezers grasp the lamp by placing the tips in the housing notch and gently pull it out Be careful not to scratch the lamp lens or chip edges Figure 7 8 Figure 7 8 Removing the PID Lamp
70. be inlet From the CONTROL MENU press 1 Pump Off Record the counts associated with the pump off From the CONTROL MENU press 1 Pump On Plug the inlet of the probe The counts should drop to within 200 of the Pump Off recorded value Thermo Fisher Scientific Flow Check Flow Calibration Thermo Fisher Scientific Servicing Flow Check Use the following procedure to check flow 1 Attach a flow meter capable of reading a least 2 LPM to the probe inlet 2 Place the instrument in the RUN mode From the MAIN MENU press 1 Run Note The dynamic flow control is only active in the Run Span or Zero modes A 3 The flow should read between 1 050 LPM and 0 950 LPM 4 Attach one of the red flow restrictors between the flow meter and the probe inlet The flow should drop to 0 80 to 0 85 LPM and slowly climb to 1 05 to 0 950 LPM within 1 2 minutes 5 Attach a second flow restrictor in line with the first between the flow meter and the probe inlet The flow should drop to 0 80 to 0 85 LPM and slowly climb to 1 05 to 0 950 LPM within 1 2 minutes Note If the pump has been serviced the actual flows may be different than above In this case a flow calibration should be performed A Use the following procedure to perform a flow calibration 1 Confirm that the instrument passes the Leak Check on page 7 16 prior to performing this calibration A new water trap filter and a new or clean sintered metal filter should be installe
71. blished the files listed below will be displayed in a newly opened TVA2020 window These files can be dragged and dropped into any user folder or directory If the TVA2020 window does not appear try disconnecting and reconnecting the USB cable from the PC or open Explorer and look for the TVA2020 device The Flash files can also be deleted without any adverse affect on the instrument These files are backed up inside the instrument memory The instrument keypad is disabled while the instrument is connected to the PC via the USB cable To resume keypad control close the TVA2020 dialog window and eject the TVA2020 memory device using the PC controls The following information can be saved from the TVA2020 to the PC a ROUTE TXT Route List b CONFIG TXT TVA2020 Configuration c CAL TXT TVA2020 Calibration Parameters d FACTORY TXT TVA2020 Factory Calibration e LOG TXT Logged Data f CAL_HIST TXT Calibration History The following information can be saved from a PC to the TVA2020 Thermo Fisher Scientific TVA2020 Instruction Manual 3 31 Display Menus Memory 3 32 TVA2020 Instruction Manual a ROUTE TXT Route List b CONFIG TXT TVA2020 Configuration c CAL TXT TVA2020 Calibration Parameters d FACTORY TXT TVA2020 Factory Calibration Note TVA2020 only recognizes the above naming format otherwise the file will be deleted a Upon exiting the USB MODE screen any files that have changed will be loaded into the
72. bration was loaded from a file instead of being performed on the instrument tttttt ZERO SPA SPA SPA SPA SPA SPA SPA SPA SPANQ ZERO SPAN1 SPAN2 SPAN3 SPAN4 SPANS SPAN6 SPAN7 SPANS SPANQ CONDUPWNPR TVA2020 Instruction Manual B 15 Text Data Formats Calibration History Sample calibration history file CAL HISTORY VER 1 00 CALIB KNOWN CALIB MEASURED DET TYPE CONC DATE TIME CALIB VALUE FID ZERO O PPM 12 DEC 99 20 20 10 450 COUNTS OK PID ZERO O PPM 12 DEC 99 20 20 10 400 COUNTS OK FID SPAN1 10 23 PPM 12 DEC 99 20 20 30 123456 COUNTS OK PID SPAN1 10 23 PPM 12 DEC 99 20 20 30 123456 COUNTS DET FAIL FID SPAN2 50 23 PPM 12 DEC 99 20 20 40 1234567 COUNTS OK FID SPAN3 100 23 PPM 12 DEC 99 20 20 50 12345678 COUNTS OVERFLOW B 16 TVA2020 Instruction Manual Thermo Fisher Scientific Text Data Formats Logged Data Logged Data The logged data is stored in binary format in MRAM when the user enters the USB transfer mode the software generates the LOG TXT file The logged data is stored as LOG TXT in the following format Format Comments LOGGED DATA lt CR gt lt LF gt Title not processed VER XX XX XXxXxx ver right just lead spaces lt logged data record 1 gt See below lt logged data record n gt lt CR gt lt LF gt Blank line not processed END lt CR gt lt LF gt Marks end of text The logged data record is different for each data storage type A head
73. c Information Menu Display Menus Information Menu 3 From the OTHER SETTINGS menu press 4 User Options 4 From the USER OPTIONS menu press 2 Display Delay 5 From the Display delay menu press 1 Short 2 Medium or 3 Long The screen displays the previous selection When a new selection is made the display returns to USER OPTIONS Note Short is approximately 0 5 sec Medium is approximately 1 5 seconds and Long is approximately 3 5 seconds A The information menu is a view only list of 14 items parameters existing in the instrument No changes may be made in this menu Enter the INFO Menu from the MAIN menu by pressing 3 Info Use the up down keys to page through the list The parameters items you may view are e Model TVA2020 e S NDDDDDDDDDDDD e Date e Time e Det PID FID or both e Pump On or Off and e Bat Battery voltage listed OK or low e Memory how much is free to use e Ver Current Software Version No e Optional Bluetooth status e Optional GPS status e PID zero calibration Date and time of last calibration e FID zero calibration Date and time of last calibration TVA2020 Instruction Manual 3 29 Display Menus Memory Memory 3 30 TVA2020 Instruction Manual e PID span calibration Date and time of last calibration e FID span calibration Date and time of last calibration Note To view calibration information you must return to the calibration menu in SETUP
74. calibration value or repeat the calibration 1 Yes or 2 Again If Auto is chosen the instrument will automatically store the accepted calibration value without prompting the user Detector counts are the raw unscaled detector output values associated with a gas measurement performed by the FID or the PID Before a detector reading is displayed or recorded the detector signal is converted from analog to digital The result is a raw number or A D counts When a detector is calibrated the detector counts for the zero gas and each of the span gases are saved in memory These detector counts are then used as reference points for calculating the concentration values to be displayed or stored When calibrating the TVA2020 in the Manual accept mode the counts from the last calibration Zero or Span are displayed before the calibration process is initiated Once the calibration process is initiated the live detector counts are then displayed You can refer to these counts as an TVA2020 Instruction Manual 4 9 Calibration Defining the Response Factor Defining the Response Factor 4 10 TVA2020 Instruction Manual indication of when the reading has stabilized or as a means of tracking the repeatability of your calibrations You can also use these counts as an indication of the success of a calibration The zero counts are the counts expected when a zero gas is applied to the detector The span counts are the counts
75. d prior to performing this calibration 2 From the SETUP MENU choose 1 Calib gt 5 Flow 3 Establish the 1 01 LPM flow at the probe inlet a Typical values are 50 60 power with 28K 30K counts 4 Press Enter Accept to save the Flow 1 values TVA2020 Instruction Manual 7 17 Servicing Flow Calibration 5 Attach one of the red flow restrictors between the probe inlet and the flow meter 6 Press 2 for the second flow calibration point 7 Adjust the pump power with the up and down arrows for a reading of 1 01 LPM at the probe inlet a Typical values are 65 75 power with 1K 3K higher counts 8 Press Enter Accept to save the Flow 2 values 9 Attach a second red flow restrictor in line with the first between the probe inlet and flow meter 10 Press 3 for the third flow calibration point 11 Adjust the pump power with the up and down arrows for a reading of 1 01 LPM at the probe inlet a Typical values are 85 95 power with 1K 3K higher counts 12 Press Enter Accept to save the Flow 3 values 13 Press 4 calc coef to calculate the flow calibration values Note The flow calibration values are stored in a file called FACTORY TXT This file can be saved to a computer and reloaded to the instrument if the flow calibration were to become corrupted See the section on flash file transfers A 7 18 TVA2020 Instruction Manual Thermo Fisher Scientific Servicing Servicing the Pump Se rvicing th
76. d the instrument will automatically reboot to the main menu 11 If the instrument does not reboot to the main menu perform the following a With the USB cable still connected to the computer disconnect and reconnect at the instrument This will turn the instrument off b Press the power key to turn the instrument on c The instrument will display two black lines however the bootloader is still active d Return to step 7 in this section and continue from there e Ifthe problem persists after several attempts contact the Thermo Fisher Scientific Service Department 12 Confirm that the calibrations gas and flow are still valid If needed they can be transferred back into the instrument using the Memory USB Mode functions Thermo Fisher Scientific TVA2020 Instruction Manual 3 35 Thermo Fisher Scientific Calibration Scenarios Scenario 1 Scenario 2 Chapter 4 Calibration This chapter describes procedures for performing a calibration on the TVA2020 For details see the following e Calibration Scenarios on page 4 1 e Calibration Configuration on page 4 3 e Detector Counts on page 4 9 e Defining the Response Factor on page 4 10 e Flow Calibration on page 4 18 The use of multipoint calibration and multiple response factors curves with the TVA2020 must be fully understood before employing these features To help explain these TVA2020 capabilities three scenarios follow
77. d the tag name However no logging takes place until you enter the logging mode from the enhanced probe or press the ENTER key on the instrument FID PID FID Press ENTER or initiate the logging mode from the enhanced probe to start the automatic logging The instrument counts down from the time selected in the SETUP Log menu and stores the reading at the end of the count down For instance if 12 seconds was selected as the logging time the instrument will count down from 12 and display 11 10 9 8 7 6 5 4 3 2 1 STORE The instrument logs into memory either the highest average or last reading during the countdown as configured in the SETUP Log menu FID PID FID This action continues until the instrument shuts down due to low battery or the EXIT key is pressed or the log memory is full Whenever the TVA2020 is in the RUN mode with Log VOC as the logging selection the instrument operates as a manual survey tool that logs its readings into memory whenever you initiate logging When the 1 Run Thermo Fisher Scientific Thermo Fisher Scientific Display Menus Run Mode key is pressed the first screen viewed is the one shown below The tag information should be typed The tag may be left blank if desired Note Tag information can only be entered using the sidepack keypad A FID PID FID The tag information should be typed in before starting Enter the tag information in the space provided up to 10 characters The
78. e FID Instrument Less than 4 5 seconds to return to 10 of base line using 10 000 ppm of methane Response time using charcoal filter adapter PID Instrument Less than 15 seconds for 90 of final value using 500 ppm of isobutylene FID Instrument Less than 15 seconds for 90 of final value from 10 000 ppm of methane Recovery time using charcoal filter adapter PID Instrument Less than 20 seconds to return to 10 of original value using 500 ppm of isobutylene FID Instrument Less than 20 seconds to return to 10 of original value using 10 000 ppm of methane TVA2020 Instruction Manual 1 9 Introduction Standard Specifications 1 10 TVA2020 Instruction Manual Data storage interval Auto Mode 1 per second to 1 per 999 minutes user selectable VOC or FE Mode 2 to 30 seconds user selectable Item Description Sample flow rate 1 liter minute nominal at sample probe inlet Battery The battery can be fully charged in less than 10 hours The battery operating time is 10 hours minimum at 0 C 32 F Use of the backlight on the probe display shortens battery life The battery is replaceable by accessing the inside of the instrument WARNING Do not replace battery in an area classified as hazardous due to presence of flammable gases or vapors Battery charger PID lamp life FID life The battery charger is a separate unit capable of operating the analyzer while simultaneousl
79. e A Sample CONFIGURATION VER 1 00 MODEL TVA2020 SERIAL NUMBER 123456789012 DETECTOR PID amp FID OPERATOR ID USERID LOGGING LOG MODE NONE LOG STORAGE FORMAT NONE LOG SAMPLE TIME 0 LOG SAMPLE TIME UNIT SEC LOG DATA STORED SAMPLE LOG UNIT LOCK OFF LOG AUTO REPEAT OFF LOG SAVE MODE MANUAL Thermo Fisher Scientific TVA2020 Instruction Manual B 9 Text Data Formats Configuration B 10 TVA2020 Instruction Manual TAG ACTIVE OFF ROUTE ACTIVE OFF CALIBRATION CAL AUTO ACCEPT OFF CAL AUTO SAVE OFF CAL UNIT LOCK OFF BACKGROUND CORRECTION PID ON BACKGROUND CORRECTION FID ON RESPONSE FACTORS SELECTED 0 MODE FACTOR RFO PID DEFAULT 1 00 RF1 PID 1 00 RF2 PID 1 00 RF3 PID 1 00 RF4 PID 1 00 RF5 PID 1 00 RF6 PID 1 00 RF7 PID 1 00 RF8 PID 1 00 RFQ PID 1 00 RFO FID DEFAULT 1 00 RF1 FID 1 00 RF2 FID 1 00 RF3 FID 1 00 RF4 FID 1 00 RF5 FID 1 00 RF6 FID 1 00 RF7 FID 1 00 RF8 FID 1 00 RFQ FID 1 00 ALARMS STEL PID 0 00 PPM LOW PID 0 00 PPM HIGH PID 0 00 PPM STEL FID 0 00 PPM LOW FID 0 00 PPM HIGH FID 0 00 PPM PROBE ENHANCED KEY CLICK OFF DISPLAY DELAY MEDIUM PC LINK TYPE TEXT AUTO DIGITAL DATA LOGGING PID amp FID BLUETOOTH POWER ON
80. e 7 24 1 Locate the probe connector connection 2 Disconnect the probe cable from the main board 3 Loosen the cross recess screw holding the connector assembly in the housing 4 Install the new probe connector cable assembly and secure in the housing 5 Reconnect the cable to the main board Screw Figure 7 24 Replacing the Internal Probe Connector Cable Thermo Fisher Scientific Sample Probe Thermo Fisher Scientific Option Chapter 8 Optional Accessories The TVA2020 is available with the following options e Sample Probe Option on page 8 1 e Enhanced Probe Option on page 8 2 e Telescoping Extension Option on page 8 11 e Activated Charcoal Filter Adapter on page 8 12 e Bluetooth Communications on page 8 13 e GPS on page 8 16 For information on additional accessories such as calibration kits enhanced probes or dilutor kits please contact Thermo Fisher Scientific The sample probe option is a hand held device that enables you to take vapor samples at precise locations It connects to the instrument by means of an umbilical The umbilical has a quick disconnect fastener The probe comes with both a straight tip and a bent tip The probe ends can be quickly changed depending on user preference Figure 8 1 Sample Probe Option TVA2020 Instruction Manual 8 1 Optional Accessories Enhanced Probe Option Enhanced Probe This probe display option is used to pe
81. e Servicing the pump involves removing the pump assembly accessing the Pump orifice and replacing the pump Removing the Pump Use the following procedure to remove the pump assembly Assembly 1 Disconnect the 8 pin connector from the pump board Figure 7 17 2 Using the correct wrenches remove the hex elbow attached to the valve assembly Figure 7 17 3 Remove the Tygon tubing from the barb fitting on the sample inlet assembly Figure 7 17 Tygon Tubing Lg FANN Hex Elbow 8 pin Connector Figure 7 17 Removing the Pump 4 Lift the pump up and out of the instrument 5 Disconnect the pump motor from the pump board Thermo Fisher Scientific TVA2020 Instruction Manual 7 19 Servicing Servicing the Pump 6 Gently remove the tubing from the pressure transducer both ports Figure 7 18 7 Loosen the screw on the backside of the board Figure 7 18 8 Grasp the pump housing with one hand and the orifice block with the other 9 Pull the pump board assembly away from the orifice block assembly Pressure Transducer Cross Recess Screw Screws 2 Figure 7 18 Replacing the Pump Accessing the Use the following procedure to access the orifice Orifice 1 Remove the hex head screw 2 Pull the orifice block apart 3 Remove the orifice for cleaning 7 20 TVA2020 Instruction Manual Thermo Fisher Scientific Replacing the Pump Replacing the Pump Thermo Fisher Scienti
82. e display is larger 8 line x 20 character display and is capable of providing several menu driven functions These Menu selections allow the user to start a datalogging application switch detectors ignite the FID and perform a background scan In Fugitive Emissions monitoring applications the MENU selections also allow the user to step through a tag route search for a selected tag start a logging event log repair data etc The display which may be backlit under low light conditions is active only in the Run mode In all other modes the display is OFF MEASURED VALUE OVERRANGE INDICATOR MEASUREMENT UNITS 9 DETECTOR PPM PPB BEING DISPLAYED SEGMENTED BARGRAPH PROBE MENU DISPLAY BUTTON FOR BUT TON MENU SELECTOR DISPLAY BACKLIGHTING BUTTON Figure 8 3 Enhanced Probe Option Display Thermo Fisher Scientific TVA2020 Instruction Manual 8 3 Optional Accessories Enhanced Probe Option Enhanced Probe Keys The enhanced probe has three keys labeled MENU SELECT and a lamp Symbol MENU and SELECT are multi functional Table 8 1 Enhanced Probe Keys and Functions Key Function MENU Press and release this key to show the probe display menu or to move to the next menu option Press and hold this key approx 2 seconds to toggle the info screen on off SELECT Press and release this key to select the highlighted menu option Lamp Symbol Press this key to turn the backlight for the probe on or off As it dra
83. e keypad to type your user ID 6 Press ENTER to store the user ID into memory Date Use the following procedure to set the date Note To set the correct date the instrument must be ON but does not have to be warmed up Date entries earlier than Jan 1 1980 or later than 2037 are invalid A 1 From the MAIN menu display press 2 Setup 2 From the SETUP menu display press 5 Other 3 From the OTHER SETTINGS menu press 2 Date Thermo Fisher Scientific TVA2020 Instruction Manual 3 25 Display Menus Other Settings 4 The next screen reads the current date If OK press EXIT or ENTER to change EXIT returns to the OTHER SETTINGS menu ENTER prompts you to type the correct date Do so by typing month day year 5 Press ENTER to store the date in memory The display then returns to OTHER SETTINGS Time of Day Use the following procedure to set the time of day Note To set the correct time the instrument must be ON but does not have to be warmed up The date must be within the valid range of 1980 to 2037 A 1 From the MAIN MENU display press 2 Setup 2 From the SETUP MENU display press 5 Other 3 From the OTHER SETTINGS menu press 3 Time 3 26 TVA2020 Instruction Manual Thermo Fisher Scientific Display Menus Other Settings 4 The next screen reads the current time If it is OK press EXIT or ENTER to change EXIT returns to the OTHER SETTINGS Menu ENTER prompts you to type the correct time Do so by
84. ed as ending in trailing spaces Partial number entries are processed as if complete The default value for fields not included in the entry are Thermo Fisher Scientific Field Type String Number Tag Hidden Leak rate Dead band Response factor Default all spaces all spaces all spaces all spaces 0 00 0 00 1 00 Text Data Formats Route Database Sample download beginning characters only various lengths ROUTE DATABASE VER 1 00 RUN ID 0001 ROUTE FIELDS T 16 Tag H 16 DBID P 16 Plant S 8 Area SX 8 Subarea Max RF EQP EQPick 6 Size 48 Description C 32 Comment WTMDWE ww ko n PICKLISTS LABEL Plant HOUSTON 1 HOUSTON 2 LEFT OF TRACKS LABEL EQPick Subarea Max EQP EQP Size VLV FLG PMP DISPLAY Tag 11 16 EQP 2 1 3 Size 2 10 6 Max 3 1 5 RF 3 12 4 Tag DBID Plant Area Description 1 1000000001 HOUSTON 1 AREA1 2 1000000002 HOUSTON 1 AREA2 3 1000000003 HOUSTON 2 AREA3 4 1000000004 HOUSTON 2 AREA4 5 1000000005 LEFT OF TRACKS AREAS END Thermo Fisher Scientific SUBAREA1 SUBAREA2 SUBAREA3 SUBAREA4 SUBAREA5 O O Go G6 Peep PR N A BON CSL DFL PKG TVA2020 Instruction Manual 00 50 aa he 00 50 DESCRIPTION DESCRIPTION DESCRIPTION DESCRIPTION DESCRIPTION Oa SFWNE B 5 Text Data Formats Route List Route List This is a legacy file format for ROUTE TXT to support older third party applications The route list format is as foll
85. election from the CAL CONFIG MENU produces the following display The second line of the display shows what option is currently selected If Factor is chosen the TVA2020 will use a single constant response factor which is multiplied by the reading If Curve is chosen the TVA2020 will use a two constant equation For more information refer to the section of the manual on Response Factors This selection allows you to choose whether or not calibrations will be automatically accepted Choosing this selection from the CAL CONFIG MENU produces the following display Thermo Fisher Scientific Cal Save Mode Detector Counts Thermo Fisher Scientific Calibration Detector Counts The second line of the display shows what option is currently selected If Manual is chosen the instrument will display the detector count during calibration and prompt the user to decide when to accept the calibration value If Auto is chosen the instrument will automatically determine the value to be stored and when to do it This selection allows you to choose whether or not the TVA2020 will automatically save an accepted calibration Choosing this selection from the CAL CONFIG MENU produces the following display The second line of the display shows what option is currently selected If Manual is chosen after a calibration value has been accepted the instrument will prompt the user to decide whether to save the
86. en the tank is in or out of the instrument Install the tank in the instrument by inserting it into the receptacle on the top and tightening left hand threads tighten counter Thermo Fisher Scientific A A Precautions in Thermo Fisher Scientific Handling and Storage Preventive Maintenance Hydrogen Gas Tank clockwise until the rubber tank boot is flush with the instrument sidepack and a slight resistance is felt Do not overtighten WARNING A safe refill operation means there are no hydrogen leaks Before any valves are opened use a wrench to firmly tighten connections to the hydrogen supply tanks and the tank fill adapter If escaping hydrogen is heard during the filling operation close all valves and correct the leak before proceeding Leak test with soapy water or equivalent A CAUTION Do not fill hydrogen tank to a pressure greater than 15 2 MPa 2200 psig 4 Note Use prepurified ZERO grade hydrogen certified total hydrocarbons as methane lt 0 5 ppm recommended A Note The hydrogen filling assembly contains a flow limiting safety device Approximately two minutes are required to fill the tank A The major hazard associated with the handling of hydrogen is flammability The following specific rules apply when handling hydrogen Never use cylinders of hydrogen in areas where flames excessive heat or sparks may occur e Use only explosion proof equipment and spark proof tools in areas where hydrogen is hand
87. entific Thermo Fisher Scientific Text Data Formats Route List 40 1 delimiter Space 41 4 Response factor floating point 45 1 delimiter Space 46 3 Equipment type text 49 1 delimiter space 50 6 Equipment size text 56 1 delimiter Space 57 48 Description text 105 1 lt CR gt ODH 106 1 lt LF gt OAH Downloaded route entries may be terminated any time after the tag by the lt CR gt lt LF gt Before the first entry is accepted the instrument s route and log memory is cleared Partial text entries are processed as ending in trailing spaces Partial number entries are processed as if complete The default value for fields not included in the entry are Field Default Area all spaces Subarea all spaces Leak rate 0 00 Response factor 1 00 Equipment type all spaces Equipment size all spaces Description all spaces Sample download beginning characters only various lengths ROUTE LIST VER 1 00 RUN ID 0486 LEAK RESP EQP EQUIP TAG AREA SUBAREA RATE FACT TYP SIZE DESCRIPTION 0123 PLANT 1 SECTION2 10 3 2 00 VLV 5 125 OUTLET 0124 PLANT 14 SECTION2 100 1 00 PMP 5 125 0124 PLANT 14 SECTION2 100 1 00 0125 END TVA2020 Instruction Manual B 7 Text Data Formats Configuration Co nfiguration The configuration is stored as CONFIG TXT with the following format B 8 TVA2020 Instruction Manual options are separated by a Format CONF IGURATION lt CR gt lt LF gt VER XX XX lt CR gt lt LF gt lt CR gt
88. er from 0 5 to 2 000 ppm Accuracy listed is achieved using isobutylene with a 1 point calibration at 1 000 ppm including drift at the temperature and humidity of the calibration FID Instrument 10 of reading or 1 0 ppm whichever is greater from 1 0 to 10 000 ppm Accuracy listed is achieved using methane with a 1 point calibration in the range from 100 to 500 ppm including drift at the temperature and humidity of the calibration 10 of reading from 3 000 to 30 000 ppm with a 1 point methane calibration at 15 000 ppm Repeatability PID Instrument 1 at 100 ppm of isobutylene FID Instrument 2 at 500 ppm of methane Dynamic range PID Instrument 0 5 to 2 000 ppm of isobutylene FID Instrument 1 0 to 30 000 ppm of methane Linear range Minimum detectable level Response time using close area sampler PID Instrument 0 5 to 2 000 ppm of isobutylene FID Instrument 1 0 to 30 000 ppm of methane The minimum detectable level is defined as seven times the standard deviation of the peak to peak noise PID Instrument 0 5 ppm of isobutylene FID Instrument 0 5 ppm of methane PID Instrument Less than 3 5 seconds for 90 of final value using 500 ppm of isobutylene FID Instrument Less than 3 5 seconds for 90 of final value using 10 000 ppm of methane Recovery time using close area sampler PID Instrument Less than 4 5 seconds to return to 10 of base line from 500 ppm of isobutylen
89. er is printed whenever the data storage type changes a new auto run starts or the configuration excluding background changes The data type records are lt auto logging record gt lt voc logging record gt lt fe logging record gt Auto logged header and data single detector GPS off lt CR gt lt LF gt AUTO DATA tttttttttttttttt lt CR gt lt LF gt DATE TIME ddd BACKGROUND ddd CONCENTRATION lt CR gt lt LF gt lt CR gt lt LF gt dd mmm yy hh mm ss xxxxxx uuu ffffffFffFf xxxxxx uuu ffFfFFFFF lt CR gt lt LF gt Auto logged header and data dual detector GPS off lt CR gt lt LF gt AUTO DATA tttttttttttttttt lt CR gt lt LF gt DATE TIME PID BACKGROUND PID CONCENTRATION dd mmm yy hh mm ss xxxxxx uuu fffffffff xxxxxx uuu fffFFFFFF FID BACKGROUND FID CONCENTRATION lt CR gt lt LF gt lt CR gt lt LF gt XXXXxx uuu fffFfFffffFf xxxxxx uuu fffffFfFFF lt CR gt lt LF gt Thermo Fisher Scientific TVA2020 Instruction Manual B 17 Text Data Formats Logged Data Auto logged header and data dual detector and GPS on lt CR gt lt LF gt AUTO DATA tttttttttttttttt lt CR gt lt LF gt DATE TIME PID BACKGROUND PID CONCENTRATION Siete ahs iy oe ute Oe Re oan De ae kOe dd mmm yy hh mm ss xxxxxx uuu ffffffffF xxxxxx uuu fffFFFFFF FID BACKGROUND FID CONCENTRATION gemegeeeavoteecusens S set sec XXXXxx uuu
90. ey on the instrument Once initiated the instrument begins a count down sampling over the configured sample time The reading sampled during the countdown is displayed on both displays and you are given the option to save the reading in memory or to select LOG again The stored values are cleared by selecting LOG on the probe display or by pressing ENTER on the instrument Each F E log must be done individually and manually 4 Custom Thermo Fisher Scientific The Custom mode allows the user to configure all the logging parameters Refer to the Log Custom section for a more complete description TVA2020 Instruction Manual 3 15 Display Menus Log Methods Accessing the LOG Menu Log None 3 16 TVA2020 Instruction Manual The RUN mode displays as governed by the LOG selection are FID NONE PID FID NONE FID AUTO VOC or F E PID FID AUTO VOC or F E Note To display and change the log menu the instrument must be ON but does not have to be warmed up A Use the following procedure to access the log menu 1 From the MAIN MENU display press 2 Setup 2 From the SETUP MENU display press 3 Log The previous log selection is displayed on the second line Follow the procedure below to set the log mode From the LOG MODE menu press 1 None to display this message briefly The screen then returns to the SETUP MENU Thermo Fisher Scientific Display Menus Log Methods Log Auto Use the following
91. fic Thermo Fisher Scientific Hardware and Startup Quick Start Procedure Note If FID only enter concentration of methane If dual enter concentration of both gases A 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 Press 1 Zero Press 1 Both Press ENTER start Challenge analyzer with zero gas sample Wait to stabilize Press ENTER to accept Press 2 Span PID 1st Press 2 PID Press ENTER start Challenge analyzer with isobutylene span gas and wait for readings to stabilize Press ENTER to accept Press 2 Span Press 3 FID Press ENTER Start Challenge analyzer with methane span gas and wait for readings to stabilize Press ENTER Accept TVA2020 Instruction Manual 2 11 Hardware and Startup Quick Start Procedure 28 Press 4 Response Factor 29 Confirm that Response Factor says RFO DEFAULT 30 Press EXIT 3 times to main menu 31 Press 1 Run You are now in the survey mode Note To perform more sophisticated operations you will need to read the rest of the manual A To power down this instrument simply press the OFF key You must also unscrew the hydrogen tank to avoid depleting the tank supply 2 12 TVA2020 Instruction Manual Thermo Fisher Scientific Chapter 3 Display Menus This chapter describes the main menu display and menu driven software For more detail see the following topics e
92. fic Valve Servicing Servicing the Pump 4 Apply a very small amount of non hydrocarbon based grease to the o rings to facilitate reassembly 5 Reinstall the orifice into the L shaped block first 6 Reverse the procedure to reassemble the pump assembly Note Be sure to push the pump firmly into the orifice block while tightening the cross recess screw If the gap is too large the pump will not fit correctly in the bottom housing recess A Use the following procedure to replace service the pump 1 Remove the two cross recess screws holding the board to the pump Figure 7 18 2 The pump can be further disassembled to replace the diaphragm and valves 3 Reverse the above procedures to reassemble the pump assembly Use the following procedure to replace the pump valve 1 Disconnect the pump from electrical power Make a sketch of the position of any tubes and fittings for ease of re assembly later 2 Remove the five screws 010 with a Phillips head screw driver and remove the headplate 020 Check that all parts are free from dirt and clean as necessary Do not scratch the parts 3 Remove the old valveplate 030 from the Compressor housing 040 4 Repeat steps 2 and 3 for opposite pump head 5 Place the new valveplate 030 in the contour on the compressor housing 040 Do not use tools TVA2020 Instruction Manual 7 21 Servicing Servicing the Pump 6 Install headplate 020 on to compressor housing 040
93. filling assembly contains a flow limiting safety device Approximately two minutes are required to fill the tank A Note All hydrogen fittings are left hand thread do not overtighten a 9 Turn supply tank valve OFF Attach tank fill adapter to supply tank with valve OFF and with manifold valve on OFF position Attach TVA2020 hydrogen tank to tank fill adapter Note left hand thread do not overtighten Open supply tank valve Move fill adapter valve to FILL position Wait for TVA2020 tank to fill This may take 2 to 3 minutes because of flow restrictors in the tank and fill adapter Close fill adapter valve Remove TVA2020 tank Close supply tank valve Remove tank fill adapter Always remove the tank from the instrument before storing in the instrument carrying case Thermo Fisher Scientific Cleaning the PID Lamp gt be Thermo Fisher Scientific Servicing Cleaning the PID Lamp Use the following procedure to remove and clean the PID lamp Replace when needed WARNING Power down the instrument before performing any service procedures A WARNING To prevent a potential explosion do not operate the instrument with the PID compartment open or FID detector cap removed A WARNING All maintenance procedures must be performed on a clean surface using clean tools Avoid touching the lamp s window as well as the metalized portion of the cell assembly with your bare fingers Fingerprints left on these p
94. flammable gases or vapors A Misusing or mishandling the battery packs can lead to fluid leakage heat generation fire or an explosion A Charger and USB connections are not allowed in an area classified as hazardous due to presence of flammable gases or vapors A To prevent a potential explosion do not operate the instrument with the PID compartment open or FID detector cap removed A Power down the instrument before performing any service procedures A Thermo Fisher Scientific About This Manual Safety and Equipment Damage Alerts Alert Description Never change an electrical component in an area classified as hazardous due to presence of flammable gases or vapors A Parts replacement and maintenance should not be performed in areas classified as hazardous due to presence of flammable gases or vapors Use only Thermo Fisher Scientific replacement parts Extended Service Plans are available contact Thermo Fisher Scientific for further information A A safe refill operation means there are no hydrogen leaks Before any valves are opened use a wrench to firmly tighten connections to the hydrogen supply tanks and the tank fill adapter If escaping hydrogen is heard during the filling operation close all valves and correct the leak before proceeding Leak test with soapy water or equivalent A Remove the hydrogen tank from the instrument before servicing amp All maintenance procedures must be performed on a clean
95. flow sensor 5 Flow Note Prior to performing calibration the instrument must be on and warmed up for approximately 30 minutes The pump must be ON the PID lamp must be ON and the FID must be ignited throughout the warm up period A Calibration Before you calibrate the TVA2020 for the first time you may want to customize certain calibration settings Once you have configured these Configuration settings you don t need to set them again every time you calibrate unless you want to change one In order to set your calibration options from the CALIBRATION menu from the CALIB MENU display choose menu selection 6 Cfg This will produce the following CAL CONFIG MENUs which can be scrolled through by using the Up and Down arrows Thermo Fisher Scientific TVA2020 Instruction Manual 4 3 Calibration Calibration Configuration Number Span Pt Defining the Span Gas Concentration s 4 4 JVA2020 Instruction Manual The Number Span Pt screen is used to determine the number of span points that will be used to calibrate your TVA2020 Choosing this selection will produce the following screen FID PID FID From this screen you may choose which detector you wish to set the number of span points for Choosing one of these selections produces the following screens FID PID FID Both option From these screens press the number of span points that you wish to use Both detectors can have up to 9 span points each and they
96. from the optional Bluetooth device if installed and turned on WARNING Charger and USB connections are not allowed in an area classified as hazardous due to presence of flammable gases or vapors A WARNING To prevent a potential explosion do not operate the instrument with the PID compartment open or FID detector cap removed A To place the TVA2020 in Startup mode press and hold the ON key for 2 seconds until you hear a beep The instrument will display a Please Wait message shortly after the MAIN MENU screen is displayed In this screen you may scroll to the various different menu selections There are four external instrument connections on the TVA2020 as shown in Error Reference source not found The umbilical attached to the probe consists of two connections an electrical cable with locking connector and a sample line with locking fitting Two additional connections battery charger input and USB interface are located on the bottom of the unit below the keypad All mechanical connections are keyed for easy orientation Thermo Fisher Scientific Hardware and Startup Instrument Connections Carrying Strap Connection Press probe lines into groove to _ Q secure after making connections to instrument _ te 7 mi Ei Probe Sample Connection M pe 3 iN ee IS Charger Connection FID Cap Connection Figure 2 2 TVA2020 Instrument Connections WARNING Charger a
97. g Replacement Parts List Replacement Table 7 1 lists the replacement parts for the TVA2020 major subassemblies Refer to Figure 7 1 to identify the component location Table 7 1 TVA2020 Replacement Parts Parts List 7 2 TVA2020 Instruction Manual Part Number Description 56 011811 Battery Pack 114012 00 Hydrogen Tank CRO12WJ Hydrogen Refill Assembly 111256 00 PID Capsule 114971 00 PID Maintenance Kit 111177 00 FID Capsule 111420 00 FID End Cap FM 620090 Filter Cups 5 pk 56 011950 Enhanced Probe complete assembly 510033 1 Split Ferrule Probe for enhanced probe CRO12EG Probe Buttons for enhanced probe CRO15AC Water Trap Probe for enhanced probe CRO12EK Probe Nut for enhanced probe 55 011879 Sample Probe complete assembly 34 011877 0001 Straight Tip 34 011877 0002 Angled Tip 34 011878 Probe Nut CRO15DK Water Trap Filter pk 10 both probes 112387 00 Sample Line Tubing 80 both probes 112415 00 Sample Line Fitting both Probes CRO12YH Sample Inlet Connection BO164KB Sample Side Tubing 12 section of TPFE lined Tygon tubing 111246 00 Pump Assembly 113305 00 O Ring Kit 113306 00 Hardware Kit 113413 00 Pump Rebuild Kit 111401 00 Battery Charger 112461 00 Sidepack Display 111409 00 Sidepack Probe Connection Thermo Fisher Scientific Part Number Servicing Replacement Parts Lis
98. g the Use the following procedure to replace the connector access door Figure Connector Access Door Cut off the retaining tail from the door if needed 2 Tilt the case top half up so that the cut off piece slips into the gap between the board and the case 3 Take the new door and press the retaining tail into the hole Use a small blade screwdriver to work the tail into the hole Figure 7 22 Replacing the Connector Access Door 7 26 TVA2020 Instruction Manual Thermo Fisher Scientific Servicing Replacing the Internal Battery Charger Connector Cable Replacing the Use the following procedure to replace the internal battery charger Internal Batte ry connector cable Figure 7 23 Charger Connector Cable Locate the battery charger connection 2 Disconnect the charger cable from the main power cable 3 Remove the two screws holding the connector in the housing 4 Re use the screws to mount the new connector cable 5 Reconnect the main power cable to the new charger cable Battery Charger Connector from Main Power Cable Screws Figure 7 23 Replacing the Internal Battery Charger Connector Cable Thermo Fisher Scientific TVA2020 Instruction Manual 7 27 Servicing Replacing the Internal Probe Connector Cable Replacing the Internal Probe Connector Cable Connector on Main Board 7 28 TVA2020 Instruction Manual Use the following procedure to replace the internal probe connector cable Figur
99. g the log interval b Avg the average reading obtained during the log interval 3 18 TVA2020 Instruction Manual Thermo Fisher Scientific Display Menus Log Methods c Smp the last reading obtained during the log interval 4 Type the appropriate time interval desired for VOC sampling 5 Press ENTER and after a brief message of acceptance the VOC method of logging is now ready to be used in the Run mode Note The allowable Sample Time range is 2 30 seconds If a Sample Time of less than 2 seconds or greater than 30 seconds is selected a warning message will appear and you will be required to re enter the Sample Time A In the LOG VOC mode data is logged in the format described below e Every time the user initiates logging the instrument will countdown by a defined interval and will prompt the user to save the captured value The user will need to re initialize the logging sequence including tag for successive samples Data will fall under the same heading e Data is logged in the following format whenever you press the LOG button e Approximately 4500 data samples may be taken Log F E Use the following procedure to access the F E Log Note The F E method requires use of the enhanced probe A Thermo Fisher Scientific TVA2020 Instruction Manual 3 19 Display Menus Log Methods 1 From the LOG MODE menu press 3 VOC F E The previous log selection is displayed on the second line of the screen
100. h the 1 01 LPM flow at the probe inlet a Typical values are 50 60 power with 28K 30K counts 4 Press Enter Accept to save the Flow 1 values 5 Attach one of the red flow restrictors between the probe inlet and the flow meter 6 Press 2 for the second flow calibration point 7 Adjust the pump power with the up and down arrows for a reading of 1 01 LPM at the probe inlet Thermo Fisher Scientific Calibration Flow Calibration a Typical values are 65 75 power with 29K 31K counts 8 Press Enter Accept to save the Flow 2 values 9 Attach a second red flow restrictor in line with the first between the probe inlet and flow meter 10 Press 3 for the third flow calibration point 11 Adjust the pump power with the up and down arrows for a reading of 1 01 LPM at the probe inlet a Typical values are 85 95 power with 31K 33K counts 12 Press Enter Accept to save the Flow 3 values 13 Press 4 calc coef to calculate the flow calibration values Note The flow calibration values are stored in a file called FACTORY TXT This file can be saved to a computer and reloaded to the instrument if the flow calibration were to become corrupted See the section on flash file transfers A Thermo Fisher Scientific TVA2020 Instruction Manual 4 19 Chapter 5 Preventive Maintenance This chapter describes the periodic maintenance procedures that should be performed on the instrument to ensure proper operation Since
101. hermo Fisher Scientific Text Data Formats Factory Facto ry The factory calibration is stored as FACTORY TXT with the following format options are separated by a Format FACTORY lt CR gt lt LF gt VER XX XX lt CR gt lt LF gt lt CR gt lt LF gt lt CR gt lt LF gt POI POI POI POI T 1 TIME XXXXXXXX lt CR gt lt LF gt T 1 FLAGS xxxxxxxx lt CR gt lt LF gt T 1 COUNT xxxxxxxx lt CR gt lt LF gt T 1 DUTY XxXxXxXxxxx lt CR gt lt LF gt lt CR gt lt LF gt POI POI POI POI T 2 TIME XXXXXXXX lt CR gt lt LF gt T 2 FLAGS xxxxxxxx lt CR gt lt LF gt T 2 COUNT xxxxxxxx lt CR gt lt LF gt T 2 DUTY XxXxXxXxXxxx lt CR gt lt LF gt lt CR gt lt LF gt POI POI POI POI T 3 TIME XxXxXxXxXxxx lt CR gt lt LF gt T 3 FLAGS xxxxxxxx lt CR gt lt LF gt T 3 COUNT xxxxxxxx lt CR gt lt LF gt T 3 DUTY XxXxXxXxxxx lt CR gt lt LF gt lt CR gt lt LF gt SLOPE XXXXXXXX lt CR gt lt LF gt INT SUM ERCEPT XXXXXXXX lt CR gt lt LF gt OF SQUARES xxxxxxxx lt CR gt lt LF gt lt CR gt lt LF gt lt CR gt lt LF gt END lt CR gt lt LF gt Sample FACTORY VER 1 14 HH RPRPRRB ri ri 1 1 1 1 NNN N O t a Ww w w w SLOPE INTERCE 00 TIME 1334746787 FLAGS 0 COUNT 26601 DUTY 98 TIME 1324747069 FLAGS 0 COUNT 28642 DUTY 34 TIME 1334747227 FLAGS 0 COUNT 29690 DUTY 180 36 6918 PT 23264 TVA2020 Instruction Manu
102. ing This chapter includes the following parts information and component replacement procedures Thermo Fisher Scientific suggests that you return the entire instrument to the factory or an authorized service center once a year for cleaning testing and calibration From time to time you must remove and renew several components of the TVA2020 analyzer Some components may be replaced as normal maintenance functions performed by operating personnel Other components however should be replaced only by personnel thoroughly trained and familiar with the analyzer instrument and its applications This chapter includes the following parts information and component replacement procedures e Replacement Parts List on page 7 2 e Replacing the Battery on page 7 6 e Hydrogen Gas Tank on page 7 6 e Cleaning the PID Lamp on page 7 9 e Replacing the FID End Cap on page 7 14 e Flow Tests on page 7 14 e Leak Check on page 7 16 e Flow Check on page 7 17 e Flow Calibration on page 7 17 e Servicing the Pump on page 7 19 e Replacing the Sample Line in Enhanced Probe on page 7 23 e Replacing Sample Line in Sample Probe on page 7 25 e Replacing the Connector Access Door on page 7 26 e Replacing the Internal Battery Charger Connector Cable on page 7 28 e Replacing the Internal Probe Connector Cable on page 7 28 TVA2020 Instruction Manual 7 1 Servicin
103. ions e Do not use or store the battery packs in environments with high temperatures such as in strong direct sunlight in vehicles during hot weather or directly in front of a heater 80 C or higher e Do not immerse the battery packs in water Do not allow them to get wet e Do not break open or damage the battery packs of the casing of the lithium ion batteries Do not pierce the battery packs with a nail or other sharp object Do not strike it with a hammer or step on it e Do not put the battery packs into a microwave oven or pressurized container e Do not strike drop or throw the battery packs e Do not assemble or modify the battery packs or the lithium ion batteries located inside the battery packs e Do not install the battery packs into to the TVA2020 if the battery packs leak electrolyte change color change shape or become deformed in any other way If leaked electrolyte comes in contact with your eyes flush your eyes immediately Wash them thoroughly with clean water and consult physician If skin or clothing comes in contact with leaked electrolyte wash the area immediately with clean water e Do not charge the TVA2020 unit in an environment below 0 C or above 40 C e Do not charge the TVA2020 unit near fire or in an extremely hot environment e Always properly dispose of the TVA2020 battery packs recycling is preferable contact Thermo Fisher Scientific for further information Do not dispose of
104. ions provided that a Buyer shall promptly notify Seller in writing upon the discovery of any defect which notice shall include the product model and serial number if applicable and details of the warranty claim b after Seller s review Seller will provide Buyer with service data and or a Return Material Authorization RMA which may include biohazard decontamination procedures and other product specific handling instructions and c then if applicable Buyer may return the defective Products to Seller with all costs prepaid by Buyer Replacement parts may be new or refurbished at the election of Seller All replaced parts shall become the property of Seller Shipment to Buyer of repaired or replacement Products shall be made in accordance with the Delivery provisions of the Seller s Terms and Conditions of Sale Consumables including but not limited to lamps fuses batteries bulbs and other such expendable items are expressly excluded from the warranty under this warranty Notwithstanding the foregoing Products supplied by Seller that are obtained by Seller from an original manufacturer or third party supplier are not warranted by Seller but Seller agrees to assign to Buyer any warranty rights in such Product that Seller may have from the original manufacturer or third party supplier to the extent such assignment is allowed by such original manufacturer or third party supplier In no event shall Seller have any obligation
105. its ionization potential If the lamp energy is greater than the compound s ionization potential the PID will detect it The lamp in the TVA2020 is 10 6 eV Because of its smaller dynamic range 0 2000 ppm the PID is not the detector of choice for measuring high concentrations of vapors A PID is also more susceptible to interference from water vapor than a FID However as a PID does not require hydrogen or oxygen it is the detector of choice when fuel is limited or unavailable or when ambient oxygen concentrations are low The PID is also very sensitive to aromatic and chlorinated compounds and can even measure some inorganic compounds that the FID does not detect at all ammonia carbon disulfide carbon tetrachloride chloroform ethylamine formaldehyde and hydrogen sulfide to name a few Thermo Fisher Scientific Introduction Photoionization Detection UV SOURCE LAMP IONIZATION CHAMBER SAMPLE IN es ye NDOW METER gt a COLLECTING ELECTRODE Figure 1 2 Typical Photoionization Detector Benefits of Benefits of Photoionization Detection Photoionization High sensitivity to aromatics unsaturated hydrocarbons and Detection chlorinated hydrocarbons e Ability to measure some inorganic gases e Very simple operation e No support gases required e Non destructive detector allows sample to be recovered Thermo Fisher Scientific TVA2020 Instruction Manual 1 5 Introduction
106. led e Ground all equipment and lines used with hydrogen e Never use a flame to detect hydrogen leaks use soapy water e Do not store reserve stocks of hydrogen with cylinders containing oxygen or other highly oxidizing or combustible materials e Store hydrogen tanks in a well ventilated area e Follow all regulatory safety and labeling precautions when shipping hydrogen in the TVA2020 Note A bleeder assembly is provided in the tool kit to allow the hydrogen tank to be emptied for common carrier shipment To use the bleeder manually screw the bleeder onto the tank valve and allow the hydrogen to vent This process will take about two minutes to complete A TVA2020 Instruction Manual 5 5 Preventive Maintenance Hydrogen Gas Tank 1 CAUTION Observe all hydrogen handling procedures listed below a Note When transporting the instrument remove the hydrogen tank and P 8 yarog place it in its normal location in the carrying case A Fuel Refilling Use the following procedure to refill the hydrogen tank Procedure Note All hydrogen fittings are left hand thread do not overtighten a 9 Turn supply tank valve OFF Attach tank fill adapter to supply tank with valve OFF and with manifold valve on OFF position Attach TVA2020 hydrogen tank to tank fill adapter Note left hand thread do not overtighten Open supply tank valve Move fill adapter valve to FILL position Wait for TVA2020 tank to fill This may
107. ling and Storage s scecvsteaticdetincdencin 7 7 Fuel Refilling Procedutei inanan aE asa 7 7 Cleaning the PID Lampir aa ee eG a ee 7 9 Replacing the FID End Cap sssssssssssssesesesssesesesesrsrsrsereseressessssssssssseses 7 14 Fl w AR SS 2c erase tac bic can eer cate R REA a EA RES iai 7 14 Weak Cheek is ccsizs vides 2tecapstvencstesitigepdar sides iiancusuticaaveaschacapiaguiesniaueeies 7 16 Fl w Check ccs cle tea sae kc teen ce vane oad hc RA eaea E Rsi iss 7 17 Flow Calibrations aes Jonstatets ncaa acs tun aR a a RR 7 17 Servicing the Pump aaesinennneas tite a Mahe eee aca asec acta les Ae 7 19 Removing the Pump Assembly spss 2ic5 acrsecsssartede Gente ete 7 19 Accessing the Orifice mrs ected ete 7 20 Replacing the Pumps s itizccat cd each iiei e RE 7 21 Replacing the Pump Valve sssssssssssssssssssssssssssssssesetereerreeersrersrsreres 7 21 Replacing the Sample Line in Enhanced Probe sssr 7 23 Replacing Sample Line in Sample Probe ssssssssssssssseseisserersrersrersrsrses 7 25 Replacing the Connector Access Dooft sssssssesseseserererereeereeersrersrees 7 26 Replacing the Internal Battery Charger Connector Cable 0 4 7 27 Replacing the Internal Probe Connector Cable ee eeeeeseeeeeereeee 7 28 Chapter 8 Optional Accessories ssssssssssssesssunnnessseunuunnenssuunnunnnnnoounnunnnnneoennnnnnnennnnnnne 8 1 Sample Probe Option sssssssssssssssssrsisrsrsrstsrsrsrsrsrsrsrstsrsrststseseseseseseses 8 1 Enhanced
108. ltering medium Applications of the filter include e Obtaining a clean air sample for zero baseline check and adjustment e Rapid screening of methane and non methane organic vapors e Selective screening for natural gas surveys The charcoal filter adapter fits directly into the telescoping wand The life of the filter depends on the time in use the types of compounds and concentrations of the compounds being filtered Under typical industrial air monitoring conditions the filter will last for many days of continuous sampling For information on replacing the charcoal filter see Replacing the Charcoal Filter on page 5 14 Thermo Fisher Scientific Bluetooth Communications Thermo Fisher Scientific Bluetooth Commands Optional Accessories Bluetooth Communications An optional Bluetooth module may be installed inside the sidepack to provide a wireless data interface via Bluetooth Bluetooth is used for Digital Data Logging to turn the pump on or off or to reignite the FID The TVA 2020 will advertise the Bluetooth friendly name TVA2020 lt last six hex digits of Bluetooth address gt A PC or PDA can then connect to it using a passkey of 0000 After connecting to the Bluetooth module from a computer or PDA the Serial Port Profile SPP is used to issue the following commands according to Table 8 3 Table 8 3 Bluetooth Commands Command Result log start Start Digital Data Logging log stop Stop Digital Dat
109. manual is organized into the following chapters and appendices to provide direct access to specific operation and service information e Chapter 1 Introduction provides an overview of the product describes the theory of operations and product specifications e Chapter 2 Hardware and Startup describes the setup and different instrument configurations and provides a quick start procedure e Chapter 3 Display Menus describes the functions of the instrument keypad and the enhanced probe display e Chapter 4 Calibration describes the calibration process and provides procedures for calibrating the instrument e Chapter 5 Preventive Maintenance provides maintenance procedures to ensure reliable and consistent instrument operation e Chapter 6 Troubleshooting provides guidelines for diagnosing analyzer problems isolating faults and includes recommended actions for restoring proper operation e Chapter 7 Servicing provides guidelines for diagnosing analyzer problems isolating faults and includes recommended actions for restoring proper operation e Chapter 8 Optional Accessories describes the optional equipment that can be used with this analyzer e Appendix A Warranty is a copy of the warranty statement e Appendix B Text Data Formats is a description of the files available for download from the instrument e Appendix C Response Factors describes response factors for
110. ment that has been put on the market after 13 August 2005 A Service is available from exclusive distributors worldwide Contact one of the phone numbers below for product support and technical information or visit us on the web at www thermoscientific com aqi 1 866 282 0430 Toll Free 1 508 520 0430 International We continue to support our customers with advanced online resources Our Air Quality Instruments Online Library allows our customer s access to product documents and information on a constant basis Available 24 hours a day and seven days a week the online library provides quick access to information regardless of time zone or office hours To register for an account or log in please visit www thermoscientific com aqilibrary Thermo Fisher Scientific Thermo Fisher Scientific Chapter 1 Chapter 2 Chapter 3 Contents lntroductiom nii a a a 1 1 Flame Ionization Detection FID esssssesssssssessssessssreessereesssseessssrses 1 2 Benefits of Flame Ionization Detection 0csssscssssossssssesssesssees 1 3 Photoionization ete CIO Ter sais tetas tack casica heads adauniuepastanoutaeaatees 1 4 Benefits of Photoionization Detection cccccssscesseseeesseesssceneeeeees 1 5 Wile Detector 2 6 ii A eNO ened Fehler es oe eat 1 6 Benefits of Dual Detectors 0csseecssscossessotvsssecssssestcsensssssseeases 1 7 Concentration Calculation and Calibration ccccccccecseesseeeseceseeeeees 1
111. minutes with the FID flame ignited throughout the warmup period If the instrument is PID equipped and the PID is to be used the lamp must be ON throughout the warmup period A Thermo Fisher Scientific Display Menus Run Mode Accessing the Run From the MAIN MENU display press 1 Run Menu Run Log None Whenever the TVA2020 is in the RUN mode with Log None as the logging selection the instrument operates as a survey tool only The readings on the probe display and instrument display show the live reading for the samples at that time No logging action is taking place Any alarms that are set will function normally FID PID FID Run Log Auto Whenever the TVA2020 is in the RUN Mode with Log Auto as the logging selection the instrument operates as an automatic survey tool that logs its readings into memory at an interval selected in the Setup Menu After the 1 Run key is pressed the first screen viewed is the one shown FID PID FID The tag information should be typed in before starting Enter the tag information in the space provided up to 16 characters The tag may be left blank if desired Thermo Fisher Scientific TVA2020 Instruction Manual 3 5 Display Menus Run Mode Run Log VOC 3 6 TVA2020 Instruction Manual FID PID FID Once the tag information is entered press the ENTER key The screen then changes to the next display The instrument is now ready to start logging The screen displays live readings an
112. missions personal computer software program as described in Appendix B Text Data Formats 2 Copy the route file to the TVA2020 by overwriting the ROUTE TXT file in the instrument flash with the one created It must be named ROUTE TXT for the instrument to recognize it After exiting the USB mode the instrument reads the properly named files from flash into the processor memory Thermo Fisher Scientific Setup Menu Thermo Fisher Scientific Calibration Settings Display Menus Setup Menu 3 Log readings by following the monitoring route shown on the TVA2020 display 4 At the end of the monitoring route copy the logged readings to the personal computer The Setup Menu configuration of the TVA2020 is the most important step in obtaining accurate gas measurements The Setup Menu contains configuration procedures and menu structure for performing calibrations entering ID numbers setting alarm levels selecting log modes entering response multipliers and setting time date From the MAIN MENU display press 1 Run During instrument setup you must set parameters as follows 5 Calibration Settings 6 Alarm Levels 7 Datalogging Setup 8 Passcode 9 Other The Calibration Menu is used to calibrate the instrument For more detailed information on calibration see Chapter 4 Calibration From the MAIN MENU display press 2 Setup From the SETUP MENU display press 1 Calib TVA2020 Instruction Man
113. n supply tank Refill tank if necessary Check low pressure hydrogen output gauge If lt 8 psi call Thermo Fisher Scientific Service Department Thermo Fisher Scientific Thermo Fisher Scientific Troubleshooting Troubleshooting Guide Problem Possible Reason Solution Action Broken ignitor Inspect ignitor coil on FID capsule for breakage If broken replace capsule Dirty ignitor contacts Inspect clean ignitor contacts not ignitor coils Battery charge low Charge battery pack FID noisy Water contamination in the Clean replace FID capsule detector chamber Erratic pump flow Bad calibration and flame detector Clean replace cups Clean replace flame arrestor Service pump Ensure proper calibration Unable to calibrate FID FID flame out Span concentration not properly set Cal gases contaminated Sample line filter cups contaminated FID capsule contaminated or faulty Contaminated hydrogen tank Internal detector fault or contamination Ignite FID Input correct span gas concentration at CAL menu Use clean cal gases and sampling equipment Clean replace sample line Clean replace filter cups Service pump Clean replace FID capsule Replace hydrogen tank Call Thermo Fisher Scientific Service Department Excessive hydrogen consumption lt 8 hours of run time for 2200 psi hydrogen Insufficient hydrogen pressure Leaking hydrogen tank Internal hydrogen leak
114. n auto logging mode 1 Inthe RUN mode from the Opening menu press the select key to Enter logging mode 2 Use the sidepack to enter a tag or simply select Accept to enter a blank tag 3 Choose Start log sampling to begin Auto logging 4 The Enhanced probe will display the countdown and continue to log until EXIT is selected at the probe or the sidepack keypad TVA2020 Instruction Manual 8 5 Optional Accessories Enhanced Probe Option 5 Once EXIT is selected you may enter another tag and begin logging again or exit from the RUN mode voc Logging with Use the following procedure to use the enhanced probe in VOC logging the Enhanced Probe mode 1 Inthe RUN mode from the Opening menu press the select key to Enter logging mode 2 Use the sidepack to enter a tag or simply select Accept to enter a blank tag 3 Choose Start log sampling to begin Auto logging 4 The Enhanced probe will display the count down for the interval selected in the SETUP menu and freeze on the reading to be saved Maximum Average or Last Reading 5 You may choose to save the reading SAV repeat the reading for the same tag AGAIN or return to the previous menu EXIT 8 6 TVA2020 Instruction Manual Thermo Fisher Scientific FE Logging with the Enhanced Probe Route Entry Response Thermo Fisher Scientific Factor Optional Accessories Enhanced Probe Option 6 Once you have
115. nd USB connections are not allowed in an area classified as hazardous due to presence of flammable gases or vapors A WARNING To prevent a potential explosion do not operate the instrument with the PID compartment open or FID detector cap gt gt removed A Thermo Fisher Scientific TVA2020 Instruction Manual 2 5 Hardware and Startup Instrument Sidepack Display Instrument The liquid crystal display LCD in the instrument sidepack has four 16 Sid epa ck character lines for three types of displays MENU ENTRY and RUN In menu displays the whole screen is normally dedicated to the menu In entry Display displays the screen provides prompts and instructions for inputting new data The normal run display consists of the live measurement data on lines 1 and 2 and menu items on lines 3 and 4 Other display information appears as you page through various menus e In Menu Display the whole screen is normally dedicated to the menu e In Entry Display the screen provides prompts and instructions for entering new data e The normal Run Display consists of the live measurement data in Lines 1 and 2 and menu items on lines 3 and 4 Other display information appears as you page through various menus 2 6 TVA2020 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Instrument Sidepack Keypad Hardware and Startup Instrument Sidepack Keypad The keypad as shown in Figure 2 3 has 19 key
116. nt Note that the relative response is the reciprocal of the response factor multiplier i e 1 RR RFM As in our example above if 100 ppm of a compound of interest produces a response of 50 ppm the relative response would be 50 divided by 100 0 50 or 50 When using a relative response factor to correct a TVA2020 reading simply divide the analyzer measured response by the relative response factor expressed as a decimal As in the example given above if a 50 ppm reading is achieved divide 50 by 0 50 to get the actual concentration 100 ppm Like any FID or PID the TVA2020 will respond to many different compounds producing a survey reading of all detectable compounds present If a single compound is present a response factor can be applied to correct for the response of that compound If a mixture of compounds is present the TVA2020 will respond to all components of the mixture and will not differentiate between them In general response factors are not available for mixtures of compounds However if the composition of the mixture is known a response factor can be calculated by adding weighted fractions of the response factors of the individual compounds For instance ifa 75 Benzene 25 Toluene mixture is to be measured multiply the Benzene response factor by 0 75 the Toluene response factor by 0 25 and add them together to achieve the weighted response factor The data from a survey analysis may be expressed in
117. ntaining oxygen or other highly oxidizing or combustible materials e Store hydrogen tanks in a well ventilated area e Follow all regulatory safety and labeling precautions when shipping hydrogen in the TVA2020 Note A bleeder assembly is provided in the tool kit to allow the hydrogen tank to be emptied for common carrier shipment To use the bleeder manually screw the bleeder onto the tank valve and allow the hydrogen to vent This process will take about two minutes to complete A f CAUTION Observe all hydrogen handling procedures listed above A When transporting the instrument remove the hydrogen tank and place it in its normal location in the carrying case Fuel Refilling Use the following procedure to fill the tank Procedure f WARNING A safe refill operation means there are no hydrogen leaks Before any valves are opened use a wrench to firmly tighten connections to the hydrogen supply tanks and the tank fill adapter If escaping hydrogen is heard during the filling operation close all valves and correct the leak before proceeding Leak test with soapy water or equivalent A Thermo Fisher Scientific TVA2020 Instruction Manual 7 7 Servicing Hydrogen Gas Tank 7 8 TVA2020 Instruction Manual A CAUTION Do not fill hydrogen tank to a pressure greater than 15 2 MPa 2200 psig amp Note Use prepurified ZERO grade hydrogen certified total hydrocarbons as methane lt 0 5 ppm recommended A Note The hydrogen
118. ntry 1 gt lt route entry 2 gt lt route entry 3 gt lt route entry n gt lt CR gt lt LF gt END lt CR gt lt LF gt Blank line not processed Section title not processed Section title not processed See below Blank line not processed See below not processed See below not processed See below Blank line not processed Marks end of text There may be up to 15 route entry fields The TAG field MUST be included as the first entry in the database so the minimum number of route entry fields is one Each route entry field consists of three columns of data the data type field width and title Thermo Fisher Scientific Thermo Fisher Scientific Text Data Formats Route Database The first column in the route entry field defines the type of data as follows String alphanumeric Pick List alphanumeric Number numeric Tag alphanumeric Hidden Database Key alphanumeric Leak Rate numeric Dead Band Rate numeric Response Factor numeric Comment alphanumeric DNA VUrTFtaA2zuYyN The String Pick List and Number field types may have more than one entry in the route fields All other types may be used only once in the route fields An optional X character may be added after the data type to indicate that that entry field may not be edited it is only shown on the TAG screen The second column of the route entry field defines the field width which is used to define data storage size displa
119. obe handle and carefully push it through to the other end When the tubing is in position cut the end square and push it over the barbed adapter fitting 8 Grasp the tubing below the probe assembly handle and pull it back through the probe assembly until the adapter fitting is properly seated in the probe assembly 9 Cut the other end of the tubing to the proper length and insert it into the end of the quick connect fitting Allow 5 8 inch of tubing inside the fitting TVA2020 Instruction Manual 7 23 Servicing Replacing the Sample Line in Enhanced Probe Figure 7 20 Replacing Sample Line in Enhanced Probe 7 24 TVA2020 Instruction Manual Thermo Fisher Scientific Servicing Replacing Sample Line in Sample Probe Replacing Use the following procedure to replace the sample line in sample probe Sample Line in 7 Sample Probe Remove the three screws holding the two halves together 2 Separate the halves to access the tubing 3 Remove the water trap assembly 4 Carefully cut the tubing lengthwise to free it from the barbs Do not cut too deeply or the barbs could be damaged 5 Attach the new tubing and return the water trap to either handle half Make sure the tubing is routed correctly in the grooves 6 Reinstall the three screws Figure 7 21 Replacing the Sample Line in Simple Probe Thermo Fisher Scientific TVA2020 Instruction Manual 7 25 Servicing Replacing the Connector Access Door Replacin
120. olumn number 1 20 to start displaying this field The fourth column is the number of characters to display If the number of characters to be displayed exceeds the number of characters available on the line it will be wrapped to the next line If the number of characters to display is less than the number of characters defined in the ROUTE FIELDS section the first n characters of the field are displayed If the number of characters to display is a negative number then the last n characters of the field are displayed An error is reported when the file is transferred if any of the display fields overlap The route header line is composed of route entry field titles line 1 centered over an underline of dashes line 2 for each field The fields are delimited with a single space Each column is of width defined in the route entry field section These two lines are ignored when receiving a route database file so formatting is not critical The route entry format depends on the data in the Route Fields section Each field goes from left to right in the order they were defined in that section Each field is delimited by a single space Numbers may be either left or right justified Each database entry is ended with lt CR gt lt LF gt Downloaded route entries may be terminated any time after the tag by the lt CR gt lt LF gt Before the first entry is accepted the instrument s route and log memory is cleared Partial text entries are process
121. ommand before logging auto repeat OFF 7 Log save accept This setting determines whether the TVA2020 prompts the user before saving a logged reading Manual or automatically saves the reading without a user prompt Auto 8 Tag active This setting determines whether or not you want to assign tags to logged readings If you wish to assign tags choose ON If you wish to skip tagging choose OFF 9 Route active This setting determines whether the logging should follow a pre loaded route list of tags If you wish to upload a list of tags to the TVA2020 and follow that list while logging choose ON If you do not wish to use routing choose OFF Thermo Fisher Scientific Passcode Protection New Passcode Thermo Fisher Scientific Enable Display Menus Passcode Protection The TVA2020 Setup Menu can be passcode protected This option allows you to protect the setup parameters from anyone who is not familiar with your 6 digit passcode If this feature is enabled you need to enter your passcode each time you choose to display the SETUP menu 1 From the MAIN MENU display press 2 Setup 2 From the SETUP MENU display press 4 Passcode 3 Press ENTER to accept passcode New Passcode allows you to enter and or change the 6 digit calibration passcode You may use any characters from the alphanumeric keypad in your passcode You must then verify the passcode to make sure you have entered it properly From the PAS
122. on to the logged data The GPS module is based on the SiRF Star II chipset which provides exceptional sensitivity even in dense refineries The module can acquire and track up to 20 satellites simultaneously and provide down to 5 meter resolution The following data is provided by the GPS module for inclusion with logged data and real time digital data logging Latitude Degrees with 5 digits to the right of the decimal Longitude Degrees with 5 digits to the right of the decimal Altitude Meters Resolution Meters Number of Satellites Thermo Fisher Scientific Thermo Fisher Scientific Appendix A Warranty Seller warrants that the Products will operate or perform substantially in conformance with Seller s published specifications and be free from defects in material and workmanship when subjected to normal proper and intended usage by properly trained personnel for the period of time set forth in the product documentation published specifications or package inserts Ifa period of time is not specified in Seller s product documentation published specifications or package inserts the warranty period shall be one 1 year from the date of shipment to Buyer for equipment and ninety 90 days for all other products the Warranty Period Seller agrees during the Warranty Period to repair or replace at Seller s option defective Products so as to cause the same to operate in substantial conformance with said published specificat
123. opane The addition of a secondary response factor allows each entry to be corrected for detector sensitivity Thus if the secondary response factor is 1 0 for methane and 2 0 for propane both route samples will read correctly as 100 ppm TVA2020 Instruction Manual 8 7 Optional Accessories Enhanced Probe Option Only the linear correction mode is used for the secondary route response correction Linear FinalConc RouteRF x BackConc The secondary RouteRF is downloaded to the instrument as part of each route entry This factor is applied to both detectors as it is expected that when operating in this mode only one detector will be used Entries inserted by the instrument default to a RouteRF of 1 0 Navigating the Route File Selecting Nx from the first logging mode screen will allow you to step to the next record in the route Selecting Pr will allow you to step to the previous record in the route Selecting List will show 10 records per page for faster searching through the route You can also enter a tag via the keypad and the TVA2020 will automatically skip to that tag if it is contained within the route If the tag is not contained within the route a TAG NOT FOUND message will appear and you will be given the option to Insert that tag into the route or Exit Logging Data Once you have reached the correct tag for the component you wish to monitor you may select Start log sampling to begin
124. otely by using the USB connection to a personal computer PC In INFO mode you can review operational parameters entered or selected in SETUP mode as well as instrument serial number Bluetooth friendly name battery status GPS coordinates etc In MEMORY mode you can download data stored within the TVA2020 to a PC for analysis and printing upload route list calibration and configuration parameters from a PC to the TVA2020 clear data memory and update instrument firmware Each of the four modes is explained in detail later in Chapter 3 Display Menus TVA2020 Instruction Manual 2 3 Hardware and Startup Using the Instrument Using the Instrument A A instrument Connections 2 4 TVA2020 Instruction Manual While operating this instrument in the field you normally carry the TVA2020 at your side using the shoulder strap or in a backpack not supplied With the pump on detector s on and the unit warmed up you monitor the area of concern As soon as the instrument analyzes a sample the sidepack displays concentration of the vapor The display on the optional enhanced probe duplicates the vapor concentration on the sidepack display Using the enhanced probe you can toggle the display between detector types by pressing the SELECT button on the probe configure to log the survey data by pressing the MENU button and backlight the LCD display by pressing the lamp button The concentration data is also streamed out
125. ows B 6 TVA2020 Instruction Manual Format ROUTE LIST lt CR gt lt LF gt VER XXXXX RUN ID aaaa lt CR gt lt LF gt lt CR gt lt LF gt lt route lt route lt route lt route lt route lt route lt route entry header line 1 gt See below entry header line 2 gt See below entry header line 3 gt See below Comments Title not processed XXXXX VEI a 0 9 A Z a z left just trail spaces Blank line not processed entry 1 gt See below entry 2 gt entry 3 gt entry n gt lt CR gt lt LF gt END lt CR gt lt LF gt not processed not processed not processed Blank line not processed Marks end of text The route header line is composed of text headers lines 1 amp 2 and an underline of dashes line 3 The route header is specified below by specification of header text for lines 1 amp 2 and starting position all other characters blank Line 3 is a dash character for all route entry characters except the delimiter space instead Route header Start 0 17 26 35 41 46 50 57 The route entry format is Byte 0 16 17 25 26 34 35 Length 16 ve OP OR Item Tag delimiter Area delimiter Subarea delimiter Leak rate Header Line 1 Header Line 2 TAG AREA SUBAREA LEAK RATE RESP FACT EOP TYP EQUIP lt CR gt lt LF gt SIZE DESCRIPTION lt CR gt lt LF gt Format text space text space text space floating point Thermo Fisher Sci
126. p oo ccc ccecescecetesesestetesesesteteseseees 7 11 Figure 7 9 Cleaning the PID Lamp cesscserh cient were nate ace uence 7 11 Figure 7 10 Installing the Lamp into the Sensor ccccccccecetecsteeeteeee 7 12 Figure 7 11 Pressing Down the PID Lamp 7 12 Figure 7 12 Installing the Cell ASS MDIY cece cccccecceteseeteseeteeeeteeeeseeees 7 12 Figure 7 13 Placing the Spacer cccccccccccccssestesestesesteseeseseseseseeeseeeeseeees 7 13 Figure 7 14 Placing the Filter Media oo ccc cececesescetesesesteteseeesesesteteseseaes 7 13 Figure 7 15 Aligning the Cap Key 2 ccccccsscceesecestecseteseeseceeseseeeesereeee 7 13 Figure 7 16 Sample Flow Troubleshooting Flowchart cccccceeceeceeee 7 15 Figure 7 17 Removing the PUMP cece ceceesescetesesesteesesesteesesesteseseseaes 7 19 Figure 7 18 Replacing the PUMP ccc cece cscetesesesteeseseetetesesestenessseaes 7 20 Figure 7 19 Exploded View of Pump Head 7 22 Figure 7 20 Replacing Sample Line in Enhanced Probe ccccccccee 7 24 TVA2020 Instruction Manual ix Figures x TVA2020 Instruction Manual Figure 7 21 Replacing the Sample Line in Simple Probe cccecceee 7 25 Figure 7 22 Replacing the Connector Access Door ccccccceccscecesteeeteeee 7 26 Figure 7 23 Replacing the Internal Battery Charger Connector Cable 7 27 Figure 7 24 Replacing the Internal Probe Connector Cable
127. pick list entry The pick list is terminated by TVA2020 Instruction Manual B 3 Text Data Formats Route Database B 4 1VA2020 Instruction Manual the start of a new pick list label or by the blank line at the end of the PICKLISTS section There are two special picklists These picklists are defined using the labels LeakSource and RepairMethod If these picklists are present in the route database file they are used for the Leak Source and Repair Method fields in the logged data If they are not included in the file default pick lists are loaded Note that the LeakSource picklist has a hard coded size of 3 characters and the RepairMethod picklist has a hard coded size of 5 characters The first entry in each of these picklists must be something equivalent to N A as these are the default settings indicating that nothing has been selected from the list The next section of the file is the DISPLAY section This section defines what database fields are displayed on the Enhanced probe while monitoring Since there are only three lines of 20 characters each available on the display during monitoring the user may select which fields are displayed The first column of the display field is the label for the route entry field to be displayed This must match a label previously defined in the ROUTE FIELDS section of the file The second column is the line number 1 3 to display this field The third column is the c
128. response factor determined at 10 000 ppm By using a response curve you can characterize a compounds response over a broader range of concentrations If the actual concentration is plotted as Y vs X measured concentration the resulting curve can be represented by the rational equation Y AX Pipe 10000 ma When using the RF Calc Mode curve the terms A and B are entered into the TVA2020 A represents the response factor multiplier at very low concentrations and B represents a change to the response for every percent concentration increase The curve can only be defined by collecting response data at several different concentrations and using a xy math spreadsheet to calculate the best fit polynomial expression Note Once the response curve has been fit the calibration gas concentrations and number of span points should not be changed without readjusting the curve fit A TVA2020 Instruction Manual 4 11 Calibration Defining the Response Factor 4 12 TVA2020 Instruction Manual To define and or choose a response factor choose 4 RF from the CALIB MENU The following display will appear The top line shows the currently active response factor If no response factor is applied the currently active response factor will be the factory DEFAULT 1 00 for FID methane and 1 00 for PID isobutylene The name and value of the default response factor cannot be changed You can scroll through the other response factors by u
129. ress EXIT to bypass the warning Log rate invalid This message appears whenever you enter a log rate of zero To correct the problem enter a log rate other than zero Entry is invalid This message appears whenever you enter a value equal to zero except for alarm levels To correct the problem enter numbers other than zero Bad Calibration parameters This warning display appears during calibration if the current zero calibration value equals or exceeds the span calibration value To correct the problem perform a re zero a re span or both Hardware Failure This message appears as a result of a self test failure Return the instrument to the factory for service Thermo Fisher Scientific Troubleshooting Service Locations Message Description FID not operating Displays when the instrument detects a problem with the FID module Most common cause is the FID is not plugged into the main board Measurement range This message appears if the PID reading exceeds 2 000 ppm exceeded or the FID reading exceeds 30 000 ppm Se rvice Service is available from exclusive distributors worldwide Contact one of Locations the phone numbers below for product support and technical information or visit us on the web at www thermoscientific com aqi 1 866 282 0430 Toll Free 1 508 520 0430 International Thermo Fisher Scientific TVA2020 Instruction Manual 6 7 Thermo Fisher Scientific Chapter 7 Servic
130. rform menu driven Option operational datalogging functions from the handheld unit and reduces the p need to access the sidepack keypad There are three buttons on the Enhanced Probe which allow you to interact with the TVA2020 e MENU Used to cause the menu to appear on the bottom of the probe display and to step through the available selections e SELECT Used to choose the currently highlighted menu selection e g LOG e Light Bulb Icon Used to turn the display backlight on and off The probe accepts a single display control message The message is tested for validity and if valid the display is updated The probe displays the contents of the last valid message received since power on The display provides two display modes Both display modes display the current concentration for one detector 1 Bar graph value gt 0 Display bar graph 2 Bar graph value 0 Displays 6 line x 20 character text PROBE NUT AND SLEEVE CLOSE AREA SAMPLNG ASSEMBLY PROBE KEYPAD AND DISPLAY THE CLOSE AREA SAMPLER CHARC OAL FILTER OR WATER TRAP CAN BE PLACED DIRECTLY INTO THE READOUT UNIT OR EXTENDED BY USING THE OPTIONAL TELESCOPING WAND OPTIONAL Figure 8 2 Enhanced Probe Option 8 2 TVA2020 Instruction Manual Thermo Fisher Scientific Optional Accessories Enhanced Probe Option Enhanced Probe The enhanced probe also has an LCD screen which displays the basic Display measurement information The enhanced prob
131. rmware 4 Press ENTER at the first warning screen 5 Press ENTER at the second warning screen Please note that pressing Enter will delete the current firmware installed in the instrument 6 Follow the steps listed on the display Thermo Fisher Scientific TVA2020 Instruction Manual 3 33 Display Menus Memory 7 After connecting the USB cable the following window will open Depending on your computer configuration you may need to cancel one or two other message windows before this window appears Removable Disk E File Edit View Favorites Tools Help Qe O FB Powe Cras 3 amp X 19 Address E 8 Drag and Drop or Copy and Paste the new firmware file into this Removable Disk window Removable Disk E SE Gy File Edit View Favorites Tools Help a O B A Psh f Foes Q B A P serh E Folders Address E x E co Address c Tvaz020 BE READY TAT VA2020_app_O100375 519 ext Document 9 File H OKB F j 3 34 TVA2020 Instruction Manual Thermo Fisher Scientific Display Menus Memory 9 The copying window should appear while the file transfers Removable Disk E Se Snr f ox ar ine ik Q B A Psh 5 Folders Address c Tvaz0z0 po r Gy TVA2020_app_0100375 519 519 File TVA2020_app_0100375 519 916 KB From T A2020 to E fone cance 10 After transfer the Removable Disk window will close an
132. rt with the Leak Check procedure with the probe connected to the instrument The leak and flow check procedures can be performed on the sub assemblies to identify the source of the problem Thermo Fisher Scientific Servicing Flow Tests Aiquessy dwng y9 y49 4297 ye j Joy qw ssy 4u jdwe zooys jqno L yea 104 qw ssy dwnd mu jqu ssy J949 Mo 4 dwnd woad Si uogou s J Joy Aiquassy ul a dwes zooys jqno L qw ssy dwnd y9 49 MO 4 aqold JooYyse qnol ssedq ssed y Bd1Mas A quwessy A quwessy dwng Aju u wnasu aqoJg pue zu wn su dwnd Wioped teq yoayo J297 yes yoaYO J27 Wes yOaYO J227 UONOLN SOI JO a qold Jooyse qnol Aju u wna su Y90940 MOJA qOJq pue zu wnysu ssed J949 Mo 4 ssed uoq 2019S Jey peyomoj4 Bunooysejqnos Moj4 ajdwes Wes ayesqueo ssed Figure 7 16 Sample Flow Troubleshooting Flowchart 7 15 TVA2020 Instruction Manual Thermo Fisher Scientific Servicing Leak Check Leak Check 7 16 TVA2020 Instruction Manual Use the following procedure to verify no leaks l Attach a flow meter capable of reading a least 2 LPM to the probe inlet From the SETUP MENU choose 1 Calib gt 5 Flow Press 1 for the first flow calibration point Adjust the pump power with the up and down arrows for a reading of 1 LPM at the pro
133. s contaminated PID window dirty or capsule contaminated or faulty Internal detector fault or contamination Turn pump on Turn PID lamp on Input correct span gas concentration at CAL menu Use clean cal gases and sampling equipment Clean replace sample line Clean replace filter cups Clean PID window Replace PID capsule Call Thermo Fisher Scientific Service Department Slow response time Insufficient sample flow Clean replace filter cups Clean replace flame arrestor Clean replace sample line Thermo Fisher Scientific Thermo Fisher Scientific Troubleshooting Troubleshooting Guide Problem Possible Reason Solution Action Service pump Sampling non volatile Call Thermo Fisher Scientific compounds Service Department High background readings High ambient concentration N A Zero drift improper Ensure proper zero span calibration calibration Sample line contamination Detector capsule contamination Clean replace sample line Clean replace filter cups Clean replace FID capsule Clean PID window Replace PID capsule TVA2020 Instruction Manual 6 5 Troubleshooting Warning Messages Wa rning The following is a list of error and warning messages for the TVA2020 Messages 6 6 TVA2020 Instruction Manual Table 6 2 Warning Messages Message Battery Power is low Description This display occurs whenever the battery has reached a low level approximately 6 5 V
134. s some of which are dual function When you press a key the screen displays the selection When you make a selection that creates or changes a parameter you must then press the ENTER key The left right arrow keys move the character entry position The up down keys make page selections or switch from ppm or to another reading Table 2 1 shows the functions of all keys SVS US hor Ua te TVA2020 Toxic Vapor Analyzer Figure 2 3 Keypad Note To turn the instrument ON press and hold the on ON key until the speaker sounds and display flashes A Note To activate OFF CONTROL EXIT and ENTER functions press and HOLD the key for approximately 1 2 second A TVA2020 Instruction Manual 2 7 Hardware and Startup Instrument Sidepack Keypad Table 2 1 Keys and Functions Key Function ON The ON key enables power from the battery to the instrument OFF The OFF key disables power from the battery to the instrument CONTROL The CONTROL key is multi function and is used to turn the pump PID and FID Bluetooth and GPS on or off and to ignite the FID Note Value displayed is opposite of existing status If pump displays OFF you would press the 1 key to turn it off EXIT The EXIT key clears any entry made in error or bypasses information that you do not want to change and clears error or warning screens ENTER The ENTER key has three functions 1 Press ENTER if you have typed one or more characters and wish
135. s you to choose whether or not to apply background correction to the detector readings displayed and logged in memory If you choose to apply background correction the last background reading stored in memory will be subtracted from the measured reading This corrected Thermo Fisher Scientific Defining the Background Thermo Fisher Scientific Correction Calibration Calibration Configuration reading will be the value displayed and the value logged Choosing this selection from the CAL CONFIG MENU produces the following display FID PID FID The second line of the display shows what option is currently selected You can change this option by choosing any one of the menu selections in dual detector instruments You may choose to apply correction to FID PID or both detectors From the CALIB MENU display press 3 Backgnd FID PID FID Press 1 Update to read live values of ambient air FID PID FID FID PID FID Press ENTER TVA2020 Instruction Manual 4 7 Calibration Calibration Configuration RF Calc Mode Cal Accept Mode 4 8 TVA2020 Instruction Manual FID PID FID Note A default background value of 0 is stored in memory until a background reading is taken A Note The background correction is reset to zero when the unit is powered up and when the user exits the USB Mode screen A This selection allows you to choose how response factor correction will be applied to the reading Choosing this s
136. saved a reading you may enter another tag and begin logging again or exit from the RUN mode In FE logging you follow a preconfigured route list consisting of component records for equipment to be monitored You should first download a route to the TVA2020 s memory before entering the RUN mode For more information on downloading routes see the Memory section in Chapter 3 In the RUN mode from the Opening menu press the select key to Enter logging mode In addition to the concentration display at the top the FE mode shows several fields which have been filled in by the downloaded route The TVA2020 starts at the first record in the route and displays the component tag number Tag equipment type Eqp equipment size Size and leak definition Leak If a reading has already been logged into memory for this record it will also be displayed followed by the words LastLog A secondary response factor correction is based on the response factor associated with the current route entry This secondary correction adjusts the concentration reading for different component measurements in the same route For example let a route contain two entries one for propane and one for methane Assume that propane has only half the detector response of methane If each route location has a leak of 100 ppm a single response factor based on methane would produce a concentration reading of 100 ppm for methane but only 50 ppm for pr
137. separator 0 80 ccc comment 0 80 characters user input during log XXXXXX concentration floating point right justified leading spaces or when DET_OFF uuu concentration unit determined by the current unit_lock at the time of the upload FEFEFFFFF OK DET OFF 3 DET FAIL BAD CALIB OVERFLOW UNDERFLOW CAL_SLOPE CLAMPED gt AVG_OVFLW LOW FLOW HIGH ALRM LOW ALARM STEL_ALRM HIGH amp STEL LOW amp STEL data status no errors or alarms detector not valid detector failed not DET OFF invalid calibration parameters A D overflow not DET FAIL A D underflow not OVERFLOW calibration sensitivity out of range an intermediate concentration exceeded 100 during log point averaging an overflow occured low sample flow detected high alarm no STEL not UNDERFLOW low alarm no high alarm or STEL not UNDERFLOW STEL alarm no low or high alarm not UNDERFLOW high alarm and STEL not UNDERFLOW low alarm and STEL no high alarm not UNDERFLOW ddd detector 1111111 OK leak indicator no leak LEAKER leak B 20 1VA2020 Instruction Manual Thermo Fisher Scientific LOGGE VER 2 AUTO DAT Thermo Fisher Scientific D DATA 00 DATA 01234 TIME 08 05 10 08 15 25 08 25 37 Text Data Formats Logged Data aaaaaaaaa GPS latitude dd ddddd floating point right justified le
138. sing the Up and Down arrow keys The other response factor names will show as you scroll through them If you would like to activate a different response factor press ENTER Accept The TVA2020 will show an ACCEPTED message If you would like to modify the response factor name or numbers press 1 Change FID PID FID If you would like to change the name assigned to the response factor press 1 Name If you would like to change the response factor value press 2 Numbers If you choose 1 Name the following display will appear Thermo Fisher Scientific Calibration Defining the Response Factor Use the alphanumeric keypad to enter a compound name for the response factor Press enter when you are finished If you choose 2 Numbers the following display will appear FID PID FID Use the keypad to enter a new response factor and press ENTER to store it in instrument memory Note The TVA2020 will not accept a response factor of 00 00 A Note If a response curve is used the TVA2020 will prompt you to enter two numbers A FID PID FID Press 2 Numbers Thermo Fisher Scientific TVA2020 Instruction Manual 4 13 Calibration Defining the Response Factor Zero Reference Point Calibration 4 14 JVA2020 Instruction Manual FID PID FID Press ENTER FID PID FID Use the following procedure to zero the instrument 1 From the CALIB MENU display press 1 Zero Note The following sequence shows the proced
139. sssssssssesesessssssrsreesrsrresesessses 8 15 REED cuddle dh sdb ent iR PE EE castes dcieendoan teu cacnladuedauedse 8 16 Warranty Donen en A 1 Text D ta Formats sisisssccssscnssctesctesscansnntisanstendsonsectancnoussdaasdensdacscusncdascdteniavcaeaues B 1 Route Databases miressa nisreen ieena eE RNE REE ETKEN B 2 Ro te Listi esiet ng ut teoei aironi inair iaa earr naan kiaii B 6 Configurati N siisisrsiseri asis esanen REKATE EEEE ASE ETES EN EATE na T LAEE NES B 8 Cali Dration nna lente a A A ihe A ba a Saha Lace asia B 11 PAGCTOLY Gis Cissus ested eS an ae A a E E AEA B 13 Calibration History s sesessssesseseseseesssessssesesttsesestsseststesrsestsseseseeeesese B 15 Logged DATA ears torei oieee ise alte a EEEa AA RO REIES B 17 Response Factors s sessssssssusueoseusunoosusunouunnoooununoosunnnoosunnnoosunnnnoounnnoosunnnnosesnne C 1 Thermo Fisher Scientific Thermo Fisher Scientific Figures Figure 1 1 Typical Flame lonization Detector ccc cceceescetetesesteteseseeee 1 3 Figure 1 2 Typical Photoionization Detector oo ccc ceceseesceteeseeteteseseees 1 5 Figure 1 3 TVA2020 Dual Detector Configuration 0 0 ccccccccecceceteeeteeee 1 7 Figure 1 4 Interconnection Control Drawing s sessisessieeeiieiieriierireninna 1 12 Figure 2 1 Instrument Sidepack Display ccccccccececescetestetestetesteteeeeeees 2 2 Figure 2 2 TVA2020 Instrument Connections cccccecccesceceseseeteseeteeeeeeee 2 5 Figure 2 3
140. stom log method 1 Log data type This setting allows you to choose the format in which data will be saved The choices are as follows a None b Auto Data stored in a fashion similar to Auto logging tag calibration information Detector Counts time date alarms c VOC Data stored in a fashion similar to VOC logging Tag Calibration Information Detector Counts time date alarms d FE Data stored in a fashion similar to FE logging 2 Sample time This setting allows you to adjust the sampling interval 3 Log time units TVA2020 Instruction Manual 3 21 Display Menus Log Methods 3 22 TVA2020 Instruction Manual This setting allows you to choose the units for the sample interval seconds or minutes 4 Log sample type This setting allows you to choose the way in which sample readings are collected over the sample rate or interval The choices are as follows a Last sample the last reading obtained within the sample interval is stored b Max the maximum reading obtained within the sample interval is stored c Avg the average reading obtained over the sample interval is stored 5 Log unit lock This setting allows you to decide whether to enable auto ranging or lock the TVA2020 units to either ppm or 6 Log auto repeat This setting allows you to determine whether the TVA2020 should automatically cycle through logging events auto repeat ON or pause for a user c
141. surface using clean tools Avoid touching the lamp s window as well as the metalized portion of the cell assembly with your bare fingers Fingerprints left on these parts may adversely affect the sensors operation Latex gloves are preferred but if they are not used your hands must be clean and free of oils lotions etc It is acceptable to hold the lamp by its glass body or by the edges of the window A CAUTION If you use multipoint calibration or a gas other than A N methane or isobutylene and then apply response factors curves that have been generated with reference to a single point methane isobutylene calibration the resulting measurements will probably be incorrect A Do not fill hydrogen tank to a pressure greater than 15 2 MPa 2200 psig A Observe all hydrogen handling procedures listed below A Do not allow any contact with the igniter coil during cleaning amp Equipment Damage Do not operate the TVA2020 if its case is damaged or otherwise compromised A Thermo Fisher Scientific TVA2020 Instruction Manual iii About This Manual WEEE Symbol WEEE Symbol Where to Get Help iv TVA2020 Instruction Manual The following symbol and description identify the WEEE marking used on the instrument and in the associated documentation Symbol Description Marking of electrical and electronic equipment which applies to electrical and electronic equipment falling under the Directive 2002 96 EC WEEE and the equip
142. t Description 111328 00 Sidepack Charger Connection 34 01 1690 0060 Rubber Connector Door 34 01 1690 0030 Internal Frame goes over battery and H2 cylinder 111782 00 Strap Rods 112694 00 Tool Kit parts indicated by an asterisk below CR012PT Spanner Wrench part of tool kit 112697 00 Capsule Extraction Tool part of tool kit 112696 00 Flow Restrictors part of tool kit CRO12TA Hydrogen Bleeder part of tool kit 105639 00 USB Cable 114295 00 USB Barrier Device 112693 00 Charger Cable N0249XS Charger Line Cord 120V 8926 Charger Line Cord 220V 107315 00 Charger Line Cord 230V China 112451 00 Carrying Case CR012DY Carrying Strap procedures A gt gt WARNING Power down the instrument before performing any service WARNING Parts replacement and maintenance should not be performed in areas classified as hazardous due to presence of flammable gases or vapors Use only Thermo Fisher Scientific replacement parts Extended Service Plans are available contact Thermo Fisher Scientific Department for further information A gt Thermo Fisher Scientific WARNING Never change an electrical component in an area classified as hazardous due to presence of flammable gases or vapors A TVA2020 Instruction Manual 7 3 Servicing Replacement Parts List Hydrogen Tank Assy Battery f Internal Housing Frame Bluetooth
143. t part of the component is leaking by choosing from a list of leak sources Valve Bonnet Compressor Seal Downstream Flange Valve Packing Pump Seal Upstream Flange Choosing Select repair method will allow you to record what method was used for the first attempt at repair Place Cap Gun Pack Valve Place Plug Repack Valve Replace Gasket Replace Seal Job Steam Seal Tighten Bonnet Tighten Cap Tighten Flange Tighten Packing Tighten Plug Wash Seal Choosing Exit will return you to the previous display TVA2020 Instruction Manual 8 9 Optional Accessories Enhanced Probe Option Additional Component Information 8 10 TVA2020 Instruction Manual If you choose a leak source and repair menu be sure to select Sav when you return to this menu so the reading and repair information will be saved in memory before proceeding to the next component record Displays containing additional information about the component and the TVA2020 status are available from the handheld unit If you press and hold the MENU key for approximately 2 seconds the Enhanced display will present an INFO page The INFO page overwrites the bottom four lines of the record with other fields such as Location Loc and Description the two lines below the location The Location field contains the Area and the Subarea where the component is located The Description field displays up to 40 characters of additional information about the component
144. t is provided with each instrument The kit contains special tools for accessing the FID and PID compartments and special flow restrictors for calibrating the pump flow Approximate mass Analyzer FID only 9 2 Ib Dual 9 4 Ib Probe Assembly Standard Probe 0 5 Ib Enhanced Probe 1 5 Ib Thermo Fisher Scientific External Influences Introduction External Influences Bluetooth Wireless communications module certified to Bluetooth ver2 0 Support for SPP Serial Port Profile Transmit Class 2 typical range of 6 This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received above the radiated susceptibility limits including interference that may cause undesired operation This device contains FCC ID POOWML C40 GPS Frequency 1575 MHz Channels up to 12 simultaneously Update rate 1 Hz This product is intended for use in indoor and outdoor environments as a portable instrument carried by a user as specified in Table 1 2 The same environmental conditions also apply to the sample stream being monitored External Influences Table 1 2 TVA2020 External Influences Reference Operating Normal Operating Operative Limits Transportation and Conditions Conditions Storage Limits Ambient temperature 23 2 C
145. th and GPS Selecting Option 1 toggles the pump on or off Selecting Option 2 toggles the FID on or off In the RUN mode when the FID is off dashes will appear instead of a reading and all FID alarms are overridden Selecting Option 3 initiates the FID flame ignition sequence which momentarily turns on the ignite coil and simultaneously reduces the pump flow Selecting Option 4 toggles the PID on or off In the RUN mode when the PID is off dashes will appear instead of a reading and all PID alarms are overridden Selecting Option 5 toggles the next screen if Bluetooth or GPS are installed If neither is installed option 5 does not display TVA2020 Instruction Manual 2 9 Hardware and Startup Quick Start Procedure Quick Start Procedure 2 10 TVA2020 Instruction Manual After running the initial startup diagnostic the Control menu can be accessed at any time Before starting the unit perform the following steps 1 Charge battery 2 Connect sample probe 3 Fill install hydrogen tank which starts hydrogen flowing To start the unit execute the following procedure 4 Press ON until the speaker sounds and the display flashes approximately 1 second 5 Press CONTROL 6 Press 3 to ignite 7 Press 2 Setup 8 Press 1 Calibrate 9 Press 6 Configure 10 Press 2 Span Concentration 11 Enter Span Concentration for calibration gas being used and press ENTER to accept Thermo Fisher Scienti
146. the battery packs in fire or heat Equipment Damage Do not operate the TVA2020 if its case is damaged or otherwise compromised A The lithium ion battery supplied with the unit lasts for a minimum of ten hours of continuous use at 20 C Extreme heat or cold and or use of the backlight however will shorten that time A battery charger with cable is shipped with the instrument You do not have to remove the battery for charging Simply plug the output of the charger into the mating connector marked with the battery symbol in the instrument The cable and connectors are keyed Orient the cable so that TVA2020 Instruction Manual 5 3 Preventive Maintenance Hydrogen Gas Tank Safety Test Criteria Hydrogen Gas 5 4 TVA2020 Instruction Manual Tank the flat side is facing up Then insert the charger plug into the appropriate wall outlet The charger has three LEDs The left most LED indicates power status When green the charger is ready to charge a battery Orange or Red indicates a problem with the charger and should not be used Once the instrument is connected to the charger the center LED will light The light will initially turn orange for 5 to 10 seconds while measuring the temperature of the battery If the battery temperature is within the charging limits then the light will turn green If the light turns red the battery may be too hot or too cold and will not charge until the proper temperature is reached If you are s
147. the reading TVA2020 Instruction Manual C 3 Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific
148. the right of the live measurement for the appropriate detector Low Ceiling This alarm is a warning that a lowest level threshold has been exceeded Once acknowledged the letter L appears to the right of the live measurement for the appropriate detector High Ceiling This alarm is a warning that a second higher level threshold has been exceeded Once this alarm is acknowledged the letter H appears to the right of the live measurement for the appropriate detector The FID maximum setting is 30000 ppm The three alarm messages appear on the instrument display as follows Note To display and change the alarm menus the instrument must be ON but does not have to be warmed up A TVA2020 Instruction Manual 3 11 Display Menus Alarm Levels Accessing the Alarm Menu STEL Level Alarm 3 12 TVA2020 Instruction Manual Use the following procedure to access the alarm menu 1 From the MAIN MENU display press 2 Setup 2 From the SETUP MENU display press 2 Alarm and follow the procedures for the appropriate alarm settings Use the following procedure to access the STEL level alarm 1 From the ALARM LEVELS display press 1 STEL The previous alarm settings for STEL alarm are displayed FID PID FID 2 To change the alarm level to a new value press ENTER on a single detector instrument or 1 2 or 3 on a dual detector instrument FID PID FID Use the up and down arrow keys to select ppm or and decim
149. to keep that information 2 Press ENTER to respond to a menu question 3 Press ENTER instead of entering the logging mode from the enhanced probe to initiate logging Left Right Arrows The left and right arrow keys move character entry positions Up Down Arrows The up and down arrow keys make page selections or scroll through options in SETUP entry screens and info screen Alphanumeric The alphanumeric keys enable you to type letters or numbers into various menus If a display asks for a number only simply press the desired key Two steps are required to type an alphanumeric character First press the key with the desired letter or number The screen then displays a selection prompt at the bottom in which 1 first letter 2 second letter 3 third letter and 0 number Press the appropriate key to execute the selection Three uses e Select menu options e Enter numbers 0 9 using single keystroke e Enter alphanumeric data A Z 0 9 SPACE using 2 keystrokes per character 2 8 TVA2020 Instruction Manual Thermo Fisher Scientific Hardware and Startup Control Menu Co ntrol Menu The Control Menu is used for turning the sampling pump on and off Thermo Fisher Scientific turning the PID lamp and FID on and off for initiating gas ignition of FID and Bluetooth and GPS options The CONTROL MENU screen has five options 1 Turn Pump ON OFF 2 Turn FID ON OFF 3 Ignite FID 4 Turn PID OFF ON 5 More Bluetoo
150. typing hour minute second 5 Then press ENTER to store the time in memory The display then returns to OTHER SETTINGS User Options This menu contains user settable options to customize the features described below 1 From the MAIN MENU display press 2 Setup 2 From the SETUP MENU display press 5 Other 3 From the OTHER MENU display press 4 User Options Thermo Fisher Scientific TVA2020 Instruction Manual 3 27 Display Menus Other Settings Key Click Display Delay 3 28 TVA2020 Instruction Manual User Options include e 1 Key Click e 2 Display Delay If the key click is on a chirp is heard every time a key is pressed Note To select key click on off the instrument must be ON but does not have to be warmed up A 1 From the MAIN MENU display press 2 Setup 2 From the SETUP MENU display press 5 Other 3 From the OTHER SETTINGS menu press 4 User Options 4 From the USER OPTIONS menu press 1 Key Click 5 From the Key click Menu press 1 On or 2 Off The screen displays the previous selection When a new selection is made the display returns to USER OPTIONS This function determines the length of time that temporary messages remain on the screen Note To select the display delay the instrument must be ON but does not have to be warmed up A 1 From the MAIN MENU display press 2 Setup 2 From the SETUP MENU display press 5 Other Thermo Fisher Scientific Thermo Fisher Scientifi
151. ual 3 9 Display Menus Alarm Levels 1 Zero the instrument 2 Calibrate the reference point s using known span gas 3 Take background reading Note For most monitoring background is set to zero Use if regulations or permits require allow background elimination A 4 Set instrument response factors if necessary 5 Calibrate the flow sensor 6 Configure a Calibrate the reference point s using known span gases The TVA2020 can be configured for as many as nine different span gas values 2 Span Alarm Levels The TVA2020 is supplied with three user configurable alarms STEL short term exposure limit Low Ceiling and High Ceiling When any of these alarms is exceeded an alarm message appears on the sidepack display and probe display and an alarm tone is generated Press EXIT to acknowledge the alarm message and sounder Once acknowledged the display returns to the live measurement with an upper case letter representing the alarm or alarm combination appearing to the right of the display 3 10 TVA2020 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Display Menus Alarm Levels The three types of user configurable alarms are Table 3 2 Alarm Configurations Alarm Type STEL Description The Short Term Exposure Limit alarm indicates that measurements averaged over a running 15 minute interval have exceeded the set alarm limit Once this alarm is acknowledged the letter S appears to
152. ure the temperature is correct then either the battery or charger is defective After the center LED turns green the charger tests the battery to determine the correct charge state The right most LED will illuminate as defined below to indicate the charge state e Orange Constant current charge state e Orange Green alternating Constant voltage 90 full charge e Steady Green Fully charged e Red Indicates the charger or battery is defective and should not be used Normal charge time for a fully discharged battery is approximately eight hours or approximately one hour of charge for every hour of use The battery can be charged while the instrument is ON and running It takes approximately 14 hours to fully charge a battery with the instrument running Do not leave the battery on charge for extended periods greater than 96 hours If you wish to remove the battery from the TVA2020 for charging or swapping with a spare battery turn the instrument off The lithium ion battery packs are designed to meet all U S Department of Transportation DOT criteria per UN Manual of Tests and Criteria Part III sub section 38 3 T1 T2 T3 T4 T5 T7 This is currently pending FID instruments are supplied with an 85cc hydrogen gas tank This tank which may be pressurized to 2200 psi maximum at 25xC will provide ten hours operation when fully charged The tank has an integrally mounted high pressure gauge that can be easily read wh
153. ure when the TVA2020 is configured with both Cal Accept mode and Cal Save mode Auto If either mode is Manual an extra confirmation is required at the appropriate step A Note If your instrument is a dual detector type PID and FID you can zero the PID and FID separately both together or one in zero gas and one in clean ambient air To do so follow the same procedure However it is suggested that you zero both detectors together A Thermo Fisher Scientific Thermo Fisher Scientific Calibration Defining the Response Factor If you press 1 Zero the display will show FID PID FID 2 To perform the actual ZERO procedure e FID only Press ENTER e PID FID Press 1 2 or 3 FID PID FID 3 Apply the zero gas to the probe at ambient pressure using a clean and labeled gas sampling bag and then press ENTER FID PID FID The instrument analyzes the zero sample TVA2020 Instruction Manual 4 15 Calibration Defining the Response Factor Span Reference Point s 4 16 TVA2020 Instruction Manual FID PID FID The ACCEPTED message appears for a short time and is then replaced by the normal CALIBRATION menu When the ACCEPTED message disappears and the CALIB MENU appears the ZERO reference value is stored This value is stored in non volatile memory until the next calibration is performed The date and time of this calibration are stored and can be accessed through the INFO menu Note For optimum accurac
154. utylene e FID only Press ENTER e PID FID Press 1 2 or 3 FID PID FID 3 Apply the appropriate span gas to the probe at ambient pressure using a clean and labeled gas sampling bag and then press ENTER FID PID FID The instrument analyzes the span sample When only one reference span gas concentration is used the ACCEPTED message appears for a short time and is then replaced by the normal CALIBRATION menu TVA2020 Instruction Manual 4 17 Calibration Flow Calibration Flow Calibration 4 18 TVA2020 Instruction Manual FID PID FID When the ACCEPTED message disappears and the CALIBRATION menu appears the SPAN reference value is stored This value is stored in non volatile memory until the next calibration is performed The date and time of this calibration are stored and can be accessed through the INFO menu 4 If more than one reference span gas concentration has been defined the span calibration display returns Move to the next concentration point Up Dn keys and repeat steps 2 3 above until each reference gas concentration point has been calibrated Use the following procedure to perform a flow calibration 1 Confirm that the instrument passes the Leak Check on page 7 16 prior to performing this calibration A new water trap filter and a new or clean sintered metal filter should be installed prior to performing this calibration 2 From the SETUP MENU choose 1 Calib gt 5 Flow 3 Establis
155. ws G additional power from the battery and thus shortens instrument run time use It only during low light conditions The Enhanced Probe Use the following procedure to use the enhanced probe in Run mode and the RUN Mode 1 From the MAIN MENU choose 1 Run 2 The enhanced probe display will display the detector reading FID or PID at the top and an analog bargraph on the bottom 3 Press the MENU key on the Enhanced probe display to show the Opening menu 8 4 TVA2020 Instruction Manual Thermo Fisher Scientific Auto Logging with the Enhanced Probe Thermo Fisher Scientific Optional Accessories Enhanced Probe Option The highlighted menu item is the item surrounded by brackets Pressing the MENU key will cause the brackets to move from menu item to menu item When you reach the menu item you wish to choose press SELECT The Opening menu items are as follows Table 8 2 Enhanced Probe Menu Menu Item Function Enter logging mode Enters FE VOC or Auto logging mode allowing you to enter tags toggle through route files log data etc If no logging is selected in the SETUP menu then this selection does not appear Ignite Ignites FID Other det Toggles between FID and PID reading This selection does not appear in single detector models Background Records a new background reading Exit Clears the menu and returns to the bar graph display Use the following procedure to use the enhanced probe i
156. y re zero the FID every time the hydrogen supply is turned on A To set the span reference point execute the procedure described below Note that the procedure is the same as that for setting the zero reference except that a span gas is used instead of a zero gas The procedure is 1 From the CALIB MENU display press 2 Span Note The following sequence shows the procedure when the TVA2020 is configured with both Cal mode and Cal Save mode Auto If either mode is Manual an extra confirmation is required at the appropriate step A Note If your instrument is a dual detector type PID and FID you can set the span reference for the PID and FID separately both together or one in one type of calibration gas and one in another type of gas To do so follow the same procedure A Note The FID and PID span concentrations must be the same for both span reference point selections otherwise an error screen will be displayed A Note If multiple span points are used repeat the following procedure for each and every span point A Thermo Fisher Scientific Thermo Fisher Scientific Calibration Defining the Response Factor When you press 2 Span the display will show FID PID FID If more than one reference span gas concentration has been configured pressing the Up Dn keys will allow you to scroll through all defined reference gas concentration points 2 To perform the actual SPAN calibration for example at 100 ppm isob
157. y charging the internal battery An adapter cable is provided to charge the battery separate from the analyzer The charger can charge a fully discharged battery in a maximum of 10 hours Charging takes 14 hours if performed while the instrument is operating Charger capable of operation from 90 Vac to 260 Vac 50 60Hz with regional plugs WARNING Do not operate battery charger in an area classified as hazardous due to presence of flammable gases or vapors Greater than 2000 hours with normal cleaning Greater than 5000 hours Audio output level Greater than 90 dB at 3 feet Gas cylinder capacity Pressure 15 3 MPa at 25 C 2200 psi at 77 F maximum Empty 85 cc 5 19in3 Hydrogen supply operating time 10 hours of continuous operation starting from a cylinder charged up to 15 3 MPa 2200 psi Instrument Enclosure Description The analyzer enclosure and front panel are made from a chemically resistant thermoplastic material or equivalent Can be decontaminated 10 chlorine solution Electrical communication connections Probe connections are weather tight WARNING Charger and USB connections are not allowed in an area classified as hazardous due to presence of flammable gases or vapors Mechanical connections A quick connect fitting is used for the sample line connection Portability A removable shoulder strap is provided with the instrument Tool kit An accessory tool ki
158. y data on the probe and during file transfers The sum of all defined fields may not exceed 365 characters Each field may not be larger than 60 characters with the exception of the Tag field and Pick lists 16 characters The third column of the route entry field is the label for the route entry field which is displayed as the column header in the route database file and on the probes as a field prompt Each label may be no more than 12 characters Each route entry field is ended with lt CR gt lt LF gt Note that the leak source repair method and comment fields are not actually stored as part of the route entry Instead these items are stored with each logged data point They are included in this section so the user may define the pick lists and comment field size The next section of the file defines the PICKLISTS There may be up to 8 pick lists included in the database definition this includes the LeakSource and RepairMethod lists but none are required Each pick list may have up to 16 items consisting of no more than 16 characters each The pick list definition consists of multiple lines of data The first line is the pick list label as follows LABEL 1111111111111111 Where MHHHHHHI is the pick list label which is left justified and terminated by lt CR gt lt LF gt at any time does not require 16 characters The label must match the label previously defined in the ROUTE FIELDS section of the file Each additional line is a
159. zer memory and downloaded to a personal computer for analysis The optional Bluetooth device will continuously output the concentration data for use with a third party hand held device Through the sidepack keyboard choose the mode of operation select concentration units for the display set alarm thresholds select data collection mode and change setup configuration parameters This unit is shipped with the battery installed The unit is ready for operation upon completion of setup calibration and charging Thermo Fisher Scientific TVA2020 Instruction Manual 2 1 Hardware and Startup Overview Figure 2 1 Instrument Sidepack Display 2 2 TVA2020 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Instrument Functions Hardware and Startup Instrument Functions This analyzer functions in any of four modes e RUN e SETUP e INFO e MEMORY In the RUN mode the instrument automatically displays its measured values in units of ppm or The RUN mode may be operated either as survey only in which the instrument displays measured values but does not store any data or survey and log in which the instrument displays measured values and also stores the information in memory In SETUP mode you can enter or select operational parameters such as calibration values alarm levels operator ID datalogging method and interval date and time This may be performed locally by using the analyzer keypad or rem
Download Pdf Manuals
Related Search
Related Contents
HP ProLiant 598132-B21 server Max Zoom TM - Sears PartsDirect owners maual.indd - Dumoulin Bicyclettes User Manual - Turner Designs Copyright © All rights reserved.
Failed to retrieve file