1. OF Cancel ee The raw data from any single profiled object or series of objects can be exported as a txt file to be used for generating a profile chart for statistical analysis 1 Select export as txt option from the FILE pull down menu A dialogue box will be opened 2 Choose a source file from which data will be exported the default is set to the currently open DAT file 3 If you check Create separate file for each well the software will use the summary text file to access wells information and split data into separate files per well By default the summary text file is in the same directory as the dat file and has the same name This is the way how Copas software creates it If 1t is located in different directory browse to it using Browse button at the bottom of the dialog 4 Option export data for different channels in separate files means that each exported file will contain data only for one channel if you have acquired more profiling channels NOTE if only one channel is collected for profiling exporting different channels or selected channels only is not meaningful 5 Option export data for selected channels only allows exporting not all channels contained in dat file Only channels selected in Setup Channels channels displayed on the graph will be exported 6 All options can be checked independently from one another 7 Select a profile or group of profiles to export by entering the
2. green detector bandpass filter yellow block yellow detector bandpass filter red block red detector bandpass filter The COPAS instrument uses either of the two excitation filters 488nm and 514 nm to get the different fluorochromes emissions that are absorbed by the PMT s Despite the fact that bandpass filters are used to best capture individual color fluorescence there is still a considerable amount of fluorescence overlap between the color parameters The figure above shows the color spectrum overlap of the color channels the instrument uses Clearly there is a broad area of overlap of the emission spectra of the three colors detected especially in emission of green and yellow fluorescence The result is that parameter signals often contain some portion of another color s emitted light For the purposes of displaying an object s fluorescent characteristics it may be necessary to adjust for this emission overlap L Open the compensation dialogue box from the SET UP pull down menu The dialogue box lists each color parameter as an independent function and the remaining two colors as a variable amount for compensation rg 2 Choose a parameter channel to be modified If subtracting green emission from yellow detector choose the yellow signal bar and move K1 cursor green signal right to desired percentage to be subtracted This percentage value appears in the box to the right of the sliding bar 3 Click APPLY then OK The pro
3. a plate without a Plate Handler this column will always contain 1 If dispensing in a receptacle Manual Sort or scanning without dispensing this column will contain 1 ROW If dispensing into a plate a well s numerical row index A B etc as seen on the plate s template It is necessary to align the plate correctly to utilize this 0 display if not using plate format COL If dispensing into a plate a well s alphabetical column index 1 2 etc as seen on the plate s template It is necessary to align the plate correctly to utilize this 0 display if not using plate format CLOG If software detects a clog while sampling file will contain a Y otherwise an N will be displayed to indicate no problem during sampling 11 SCAN RATE A single scan rate must be used to generate profile data This rate is displayed as 156 156 KHz 312 312 KHz 625 625 KHz 1250 1 25 MHz 2500 2 5 MHz 5000 5 MHz STATUS SORT A number representing the sorting status of an event returned by COPAS board Refer to the following key 0 scanning without dispensing 1 unknown sort status cannot be retrieved 2 out of region 3 in enhanced mode coincidence with following un sortable event 4 coincidence in pure mode 5 lost synchronization during acquisition not sorted 6 sorted 7 sorted in superdrop 8 in enhanced mode coincidence with previous un sortable event 9 sorting decision took too long for event to b
4. count below Or set number of peak to ACCEPT ALL and location requirement to EVERYTHING which effectively removes peak count from sorting decision and skip ahead to step 10 Configuring Sorting criteria based on number of peaks Before making a sorting decision the software calculates the number of peaks in the acquired profile or part of a profile based on user defined peak count criteria found in Peaks count setup under the PROFILER pull down menu After the entire Profile is scanned the total number of peaks within defined location is compared to the limits set by user Click Change peak count setup button located beneath the location settings to open dialogue box shown below to setup peak count Set appropriate settings for determining peaks See using profile reader section 7 1 using peak count setup to optimize these settings Peaks count setup x Channel Green hi Average on Peak s relative 3000 height greater than Peak s absolute 10000 height greater than Single peak s width from 1 Tite to 50 Always count peaks Cancel i VERY IMPORTANT Itis crucial to carefully setup peak count criteria accurately Therefore it is recommended that the user collects a number of representative profiles which s he can manipulate in ProfReader software to determine acceptable settings for peak count setup prior to sorting objects using number of peaks Under number of peaks enter the range of numbers of p
5. height greater than Peak s relative height greater than Once the program identifies an area of the profile that meets peak absolute height it locates each individual peak maximum value and analyzes these points for peak relative height and peak width At each maximum the program moves down the peak the distance of the user defined relative height to determine the width of the peak at this point Basically the program extends a line across the peak at this height peak height relative height Wherever the line intersects with the graph twice it creates endpoints that it will measure to determine the width of this peak This distance is the single peak s width 19 The value set for relative peak height affects which peaks are counted See the figure B and C below where an increase of 100 in the relative peak height excludes four peaks from peaks counted Figure 3A part of original profile Figure 3B relative height 200 peaks counted 11 IY Show peaks Channel Average on Peak s relative height greater than Peak s absolute height greater than Single peak s width from e Show peaks Channel Average on Peak s absolute height greater than Single peaks width from LLL Red E Peak s relative 300 height greater than o Peak Count Setup X points 00 500 T to 50 x Peak Count Setup xX NOTE Keep in mind that neighboring peaks may interfere with each other if limi
6. i e 0 4uS Correct scan rate is essential for the conversion If this option is not checked then all values along axis X are presented as number of scanned points It is important to use this option if you use ProfReader to configure peak counting or sorting criteria The template will automatically fill in scan rate when file is opened Scan rate can be found listed in the original TXT file should the user want to use this option in the absence of having a viable template 3 Scale the Y axis parameters by entering integers or selecting adjustable setting for each parameter The scale range can be any integer from flat line on the x axis to 65536 the highest signal level detectable by the filter Again each profiled parameter is independently scaled allowing you to see different parameter features of the profiles Selecting adjustable will change the scale so that the parameter range of the profile fills the entire viewer along the y axis 4 Click OK to implement scale setup and return to viewer display window 6 2 3 CHANNELS Channels allows the user to choose which parameter channels to display in the viewer and where to set the level of width determination Channes x Show profiles Calculate width on level Extinction 4 Green e Yellow i S W Red Apply k 1 Open the dialogue box by selecting CHANNELS from the SETUP pull down menu See dialogue box above 2 Click on the profiled parameters you w
7. sd ls la a ls a da o cr aid Ge A eee 16 6 2 4 LADA dd ll a OT LRT TE eee te TNE REA RIE Vip me eee 16 7 PROFILE SETTINGS FOR SORT CRITERIA 4 scssccscccccccccccccccccccccccccccccccccccccccccccccccccccccccccccccccecces 18 7 1 WS SINGPEAKCOUNT SEU UP a tae a cda 18 7 1 1 OPErat OO VEGI COM A E O O iO 18 7 2 USING COMPENSATION a a aa 21 ZA An example FOP usne COMP CTS CLONE io seca inas eE ANA n 22 S OT THEREFEATURES a n 23 8 1 ARO Ja Buus 2 AAA A SEO eC A AN ROTO EEA NTR TV es 23 8 2 IMPORTA TEMPLATE en COC a NN a Cn ee ER te ee Teen en 23 8 3 AE UNS E S Te easels tec toric A A E sera ios uan acca cabeamenen E A E A E E E EE 23 8 4 EXPORTA S E TEE eaa a a A a a a A 24 1 Introduction 1 1 Profiler An Overview The standard COPAS system measures the value of extinction and fluorescence signals by integrating each signal over the time that the threshold signal is above the threshold value The result for each signal is a single value that has obscured any details of the changing intensity of the signal while the signal is being integrated Therefore an object containing a small intense fluorescent spot and another object with a low diffuse level of florescence throughout would appear the same despite dramatically different spatial organization of the fluorescence signal Instead of making a single integrated measurement of a signal the Profiler option digitizes the instantaneous signal level The result is a list
8. should appear on the display Additionally user should see individual peaks labeled with a p and light blue lines across each peak denoting where the width measurement was made At the bottom left corner there will appear a new data display peaks counted the number of peaks labeled with p that fulfill peak count criteria 9 Using NEXT and PREVIOUS buttons scroll across other profiles in the same DAT file to ensure settings are sufficient for other object profiles It is possible you will need to adjust some of the settings a few times to correctly count important peaks from different objects At this point the values determined for correct peak analysis and count can be input into the COPAS software program and sorting can be performed using new peak count criteria See software peak count feature for more information about this 7 2 Using Compensation Compensation allows the user to manipulate the display of fluorescence overlap by removing a percentage of one color emission from the emission of another fluorochrome x 550 30 r Green k1 Yellow k2 Red i E e E 80 60 Yellow k1 Green k2 Red ss jo 40 Se a o Hx 20 Red k1 Green k2 Yellow e aop 44x 0 468 518 618 Wavelength nm F Compensation enabled Figure 6 Emission spectra green line green emission blue line Apply Cancel yellow emission red orange line red emission green block
9. sort criteria for a detailed description on how to set criteria for profiles to be used as a filter Here the user can select any parameter in which to use as a filter simply by setting limits and applying the set criteria for this parameter box This can be particularly useful when the user needs to locate individual significant profiles from a file containing many other profiled objects 1 Open the filter feature under the SET UP pull down menu Choose a profile channel or multiple channels to set filter limits Extinction Green Yellow or Red 3 Set appropriate limits to include profiles with attributes you are interested in viewing or exclude profiles with attributes you do not wish to view For example it is possible to accept all or reject all profiles with a peak height above a user defined low high range 4 Choose to set rules for a parameter by unselecting the Do not set sorting rules for this channel found at the top left of the dialogue box 5 Click Apply to implement the limits you have set for a parameter channel The right half of the dialogue box Summary changes to reflect the sorting rules applied to this parameter channel 6 Continue setting filter limits for other parameter channels removing the Do not set sorting rules for this channel and Appling the limits set 23 7 Click CLOSE button to implement the filter criteria you have set Once limits are set and applied scrolling through pro
10. useful to set the calculate width on value minimum By doing this the absolute height value will be drawn across the graph offering visual display of which peaks rise above this height Choosing different values will allow you to see which peaks are above considered for peak count analysis and which peaks on the graph are below the absolute height will totally be excluded from analysis See figure below that shows how different absolute height values affect the peaks counted Note that changing the value of absolute peak height changes the number of peaks counted Figure 2A absolute height 500 number of peaks 28 El MV Show peaks Channel Red y Average on E 4 points Peak s relati e O Peak s absolut 500 Meira thon Single peak s width from jo to 50 P pp ES D Y pr P D y P omp PpP in PP ene oS Pp a PpP D Apply O oK fat JIN P P ee IRA N AeA a Number of peaks 28 Figure 2B absolute height 2000 number of peaks 18 z MV Show peaks a Channel ed Average on points Peak s relative height greater than Peak s absolute 2000 height greater than ld Single peak s width from 0 Choose a value for Peak s relative height greater than by inputting a value lower than the Peak s absolute height greater than value Peak s relative height greater than will determine where each potential peak s width is calculated See discussion below to determine peak s relative
11. 4 If desired select other parameter channels to apply sorting rules for peak height and peak width NOTE User can apply profiler rules for peak height peak width and number of peaks to none one or any combination of parameters profiled Keep in mind that if many profiling rules apply there may be no events that will pass all of them Click CLOSE to close the dialogue box and return to COPAS program screen Once settings have been chosen initiate a sort manually or fill plate making sure that you use Data Collection mode to acquire the sample See section 4 sorting in COPAS user s manual Y ou can store your data after acquisition 10 4 Data Storage Once data is acquired data can be stored by clicking the STORE button at the right edge of the COPAS software screen When data is stored four files are created that correspond to the data storage event 4 1 COPAS System template file nameXXX csv The csv file contains settings for the COPAS system preserved at the time of data acquisition All settings such as PMT gains and gate and sort regions are stored as a template that the user can open later if they wish to run additional samples under previous conditions NOTE Sample may not fall within previous gate and sort regions for various reasons 1 Slight differences in sample conditions may result in minute physical changes that the instrument can detect Starvation for example may cause changes in the object s internal s
12. TABLE OF CONTENTS TAP OR CONTENTS waco a as 1 E INTRODUC TION ecan aa ar aaee a ua cuseseeuareceleuceaeelens 2 1 1 PPO P AAO VV dd idad 2 2 USING PROFU ER iii bio 4 2 1 Fe HAVA lO Sl PROFE and ea da de al od de aca da e a di NTE OT 4 22 SFr gl 22 Bt 0 YE OAA tai 4 23 SET SCALING DISPEAY OF PROPIISES cd ccs cece estes cand eucrsce slow cis ol cveedtuic dead 5 3 SORTING WITH PROFILER 0 ica dieses 6 3 1 SS A O EO TET TE aT RTC 6 3 2 CONFIGURING PROFILER SORTING CRITERIA FOR PROFILES USING PEAK HEIGHT AND WIDTH 0cccceeccessceeeseceeeeees 7 3 3 CONFIGURING SORTING CRITERIA BASED ON NUMBER OF PEAKG cscesceccecceccecesccceccscescescnccecescescescscescscesescescescescess 8 a DAMA SIDRA a On E A E NEE E ae RE EN Rene 11 4 1 COPAS SYSTEM TEMPLATE FILE NAMEX XX CSV jn aia its 11 4 2 PCS FORMAT FILE FILENAME XX MD lia 11 4 3 SUMMARY DATA FILE CHILE NAME XOX Das 11 4 4 PROFILE READER COMPATIBLE FILE FILE NAMEX XX DAT cccsscccssecccsscccsscccensceensccescecnscsesccesceeensceeuecsensceeeeseeenss 13 S PROFF READER lt coca 14 5 1 PROFILE READER AN OVER VIE W wij becivativiwedanaoscavabdedsaduatviaauah ddr asii 14 5 2 TPB OTS PIGAY OB AUP ROPING Bierce al alice 14 6 OPERATING PROFILE READER cout ii a iii 15 6 1 OPLENINO A AE e ls a e e al A EN e ll e o ld a al 15 6 2 SETUP THE PROFE DISPLAY a da e ee da de dd T e e da tie dd rn e e de la o 15 6 2 2 SCADE ARA A A A A A A A A IA 15 6 2 3 ECNANNE ES sa ne da ea a
13. ant displayed in the viewer A check mark in the window indicates the parameter will be displayed A ghosted parameter indicates that the parameter channel was not selected for data storage during data acquisition 3 Choose a value to calculate width on by typing in an integer or clicking the up or down arrows next to the appropriate window Appropriate range is integer 1 to 65536 maximum y value The calculate width on level is used to determine where the width of a peak is calculated This can be particularly useful for two reasons 1 to help determine length of the object that rises above the calculate width on that specific level 2 visualizing where an absolute peak height value lies on the graph For instance if EXT was used as the trigger channel the user can determine the TOF by entering 1 as the EXT calculate width on level Under these conditions the entire profile will be contained begin and end within the peak of EXT of at least 1 Therefore the width above level for EXT signal displayed at the top of each profile will equal the object s TOF Additionally drawing a height value on the graph is useful in the peak count mode described in using peak count section 7 1 4 Click Apply to implement the changes 5 Click OK to return to the viewer window Upon returning to the viewer window you should see that peak width calculations have changed and that a horizontal line was drawn across the graph at the calculate width on l
14. corresponding index number s you wish to export in the From and To boxes as shown above left The maximum number of events per file can be selected in the same dialog NOTE if a data file from a ReF 1x acquisition contains more than 100 events per well you have to set this number high enough to export all profiles from one well in the same file 8 Click OK At this point a new dialogue box will be opened shown above right 9 In the File name window type in a name for the new file s 10 Click the Save button 24 File names In general exported files are named name_channel_Px_Wy_z txt where name name selected by user channel name of a channel e g green red etc if exporting data separately per channel otherwise omitted x number or name of a plate from Plate column in summary file if exporting data per well y well e g Al if exporting data per well z sequential numbers starting from 1 in case data should be split between several files NOTE If you check the option Create separate file for each well the selection of a group of profiles to export default 1 through 100 does not apply note that it gets grayed out With this option checked all profiles contained in dat file will be exported If you apply this option to the data collected not during sampling or dispensing then the exported file s will look as if all data belongs to one well W00 because in the summary text fi
15. ct Conversely the slowest scan rate collects data over 6 4 microseconds as the object passes the laser path Effectively each data point contains information over a longer slice of the total length of the object One way to think about this is to consider that each data channel generated using the slowest scan rate may contain information that Spans up to 32 channels generated using the highest scan rate Therefore averaging on 5 channels of 156 KHz scan rate is akin to averaging 160 channels under 5MHz scan rate The averaging effect can be quite large so keep this in mind when choosing a value to average on Once the user inputs the number of channels to average on the smoothed profile is drawn in black on top of the actual profile and this becomes the profile line graph that the program uses to determine and count peaks 18 Determine a threshold value for determining a peak This is the peak s absolute height greater than value in the peak count dialogue box See the discussion of peak s absolute height below Enter an integer into the window Peak s absolute height greater than The absolute height of the peak is a signal level that the graph needs to rise above in order to be analyzed as containing peaks Keep in mind that any area of the profile that does not meet this value will be excluded from analysis as a peak so it is important to determine a height sufficient to contain all peaks of interest For this purpose it is
16. e dispensed 3n Sort status equal 30 n where n is one of the above statuses indicates that the object was very close to the previous object and the sorting drop Gif n 6 started at least 30 microseconds later than it normally should have STATUS SELECT A Selected status is assigned to each object with respect to its parameter s relative to the gate and sort regions and with respect to its profiler characteristics relative to the chosen Profiler sorting configuration 0 outside TOF histogram gate l outside gating 2 dimensional region dot plot 2 outside sorting 2 dimensional region if sorting based on 2d regions or does not satisfy any histogram condition if sorting based on histograms 3 10 does not satisfy a condition for peak s height or width 40 satisfies all sorting conditions NOTE Display Store options affect which data points are stored Option EVERYTHING stores events from every status select code all data points Option STORE GATED excludes events with status select 0 and 1 outside gate regions STORE SELECTED excludes events with status select 0 1 and 2 outside gate and sort regions TOF Time of flight value EXT Extinction integral value Green Yellow Red Integral values for fluorescence on a corresponding channel PH Ext PH Green PH Yellow PH Red Maximum Peak Height of an individual profile In non Profiler configuration 0 PW Ext PW Green PW Yellow PW Red Peak Width of an i
17. e was not found the viewer cannot revert to a template and all settings simply maintain the values currently in use 8 2 Import a template It is possible to open a DAT file and import template scale settings used during a different sample s acquisition 1 Select import a template in the FILE pull down menu Select the CSV file from the template you wish to use 3 Select OK Imported template settings will temporarily replace the existing ones Template settings will only be in effect until the file is closed If file is opened again later data will be displayed using original settings from data acquisition 8 3 Filter Filter allows the user to set criteria so that only significant object profiles contained within the larger stored data are displayed while scrolling through the individual profiles Criteria for profiles x Summary Extincion Mo rules for this channel Green Extinction Green Yellows Red Do not set sorting rules for this channel Low High Max Height fico E 096 E H Accept All mo ae Width above level fioo A 612 A Mo rules for this channel Yellow Lo High No rules for this channel W Mumber of peaks 5 fico A Accept Al From To Location fo 3 ico z Everything gt Change peak count setup Red Mo rules for this channel Choosing this feature opens the criteria for profiles dialogue box from the COPAS software See Profiler manual profiling
18. eaks you wish to sort Acceptable range is 0 250 peaks 7 Select sorting to Accept all include or exclude objects based on number of peaks ACCEPT ALL Ignoring the specified ranges for the number of peaks every object passing the flow cell will be dispensed to a microtiter plate or sorted to a receptacle This may be useful if user wants to count peaks as data is acquired but doesn t need to sort based on peak count number INCLUDE To be dispensed to a microtiter plate or sorted to a receptacle an object must satisfy the following condition The profile has a number of peaks satisfying the Number of Peaks limits and these peaks are within the predefined Location limits along the length of the profile EXCLUDE This condition is the compliment of INCLUDE In other words only objects not satisfying the condition for INCLUDE will be dispensed to a microtiter plate or sorted to a receptacle If a sample usually exhibits 1 to 2 peaks of localized fluorescence it may be possible to sort variable expression patterns by EXCLUDING objects with 1 2 peaks By doing this the user can collect objects with no fluorescence or objects with more than the usual two peaks of fluorescence 8 Set peak location ranges by entering an appropriate range of the entire length of the object to count peaks in Location of peaks is the relative position of peaks expressed as a of length of the whole object in a profile that can be chosen This allows
19. evel for each parameter profiled 6 2 4 LABELS Labels allows the user to change the title and color display of a profiled parameter channel x Channel Label Color 1 Extinction MM Change color gt Green EE Change color 3 yellow Change color 4 Red MA Change color 1 Open the LABELS dialogue box under the SETUP pull down menu See dialogue box above 16 Se O Choose a parameter channel to modify Change the title by typing in the new title under the corresponding Label window Change the color of the display by clicking on the Channel s Change color button Select a color you wish it changed to clicking OK when done to close the color change dialogue box When all changes have been made to Channels click OK to implement the changes The profile viewer will reflect the title and color changes Paige 7 Profile Settings for Sort Criteria 7 1 Using Peak count setup Peak count identifies and counts peaks on any one profiled parameter channel based on user defined criteria The criteria can then be used in COPAS software to make peak count sorting decision Peak Count Setup x Show peaks Channel Average on Peak s relative height greater than Peak s absolute height greater than Single peak s width from fo to 50 bro When activated the COPAS profiler uses peak count as final criteria for sorting a sampled object The ProfReader software serves as a tool to determine appropriate se
20. f the profiles using SETUP menu features 6 2 Setup the Profile Display 6 2 2 SCALE Scale lets the user choose the scale in which each parameter is independently displayed EA ll x Scale Range fi 500 or adjustable E MV Scale X axis to instrument units 0 4microsec for scan rate 1 25 MB y Y Scale Extinction 65536 or adjustable Green 65536 or adjustable Yellow 65536 or adjustable j Red 30000 or adjustable J 1 Open the scale feature from the set up pull down menu Dialogue box shown above 2 Scale the X axis Time of Flight by entering an integer selecting adjustable range or range based on scan rate If a value is entered only the left channels up to that number will be displayed in the viewer window for every profile in that DAT file If profiles are longer than the scale range set only the left part of the object contained within the viewer limits will be displayed If the user chooses to select adjustable each profile will be scaled to fit within the window NOTE Choosing adjustable scale will make it very useful to scan through profiles of objects with varying lengths within a DAT file However it can also make these profiles harder to compare to each other Peak widths for example may appear very different between objects of various lengths Choosing Scale X axis to instrument units causes converting profile s total length and peak s width to the TOF units
21. file display will change to reflect the compensated values 1 2 1 AN EXAMPLE FOR USING COMPENSATION To better understand fluorescence compensation it is useful to look at a profile where this is occurring Shown below is part of a profile where the green and yellow scale settings are set equal to each other You 1l notice that there are two yellow peaks about midway along the graph The green graph at this place does not show peaks However at the right end of the profile there is a large peak displayed for both the green and yellow parameters but phenotypically there should be only one color fluorescence green at this point It is quite likely that the overlapping peaks in both parameter channels is due to overlap of spectra and absorption of green signal by the yellow detector Compensating for green by removing 50 of the green signal from the yellow signal we can clearly see the difference between graphs of the same object Notice that the yellow peaks at the middle of the object have not been altered much by removing a portion of the green signal from the yellow data However the peak at the far right is greatly reduced This suggests that the yellow peak at that end of the object 1s comprised mostly of green signal This may be very useful knowledge in determining how to conduct your sampling keeping two things in mind 1 all compensation values used in ProfReader software can be duplicated in the COPAS software for determination of com
22. files by clicking next and previous will result in skipping over profiles that fail to meet the limits set by user NOTE The specific limits set by the user are unique to individual samples Therefore the user should have a good understanding of the characteristics in the profiles they wish displayed before setting filter limits As an additional note once acceptable limits are established to VIEW profiles these limits can be duplicated in the Criteria for profiles box in the COPAS software to effectively display sort and or store object events fulfilling these criteria limits However before applying these rules make sure you are very confident in your determined settings It is always advisable to test your settings with a small sample using an existing template to ensure consistent PMT gain threshold scan rate settings etc before collecting large data acquisitions 8 4 Export as TXT file xj ea A 2 x Save ire E COPAS fies y COPAS files ce E ES Select source file CAD ocuments and SettingssAico BongaartssMy Documents UBL Create separate file for each well T Export data for different channels in separate files T Export data for selected channels only Select group of profiles to esport From jo To 10 File name Save basimar number of events per file 200 save as ppe gt Files txt Cancel E Summary test fle contamina well intormatior E Documents and Settings Aico BongaartssMy Docum Browse
23. fluorescent markers within an object allowing another dimension for resolving differences between the analyzed individuals in a collection or population of objects 70 60 50 40 30 20 Figure 1 C elegans worm with str 1 GFP mab 5 dsRed unc 17 zs Yellow Profile of transgenic nematode clearly shows green expression in head yellow expression in the animal s vulva and red expression in multiple specific cells along the body Variable scan rates for resolution control Additionally Profiler gives the user the option of setting the scan rate the clock rate of data storage of the object itself as it passes through the flow cell There are six scanning rates available to the user ranging from 156 KHz to 5 MHz A scan rate of 156 Hz takes single integral measurements every 6 4 microseconds while at a rate of 5 MHz single integral measurements are collected every 0 2 microseconds This has an obvious effect on the resolution of the profile itself Because higher scan rates take integral parameter information slices more frequently than slower scan rates the result is better detection of slight differences along the length of the object Using higher scan rates the profile of the object is one of sharper peaks and valleys accurately depicting the small variations along the length of the object Lower scan rates will in essence average the highs and lows of variance to have an overall smoothing affect on the profile NOTE For many users t
24. g the box next to each parameter title 2 At this time you can choose to change the color of a parameter s display by clicking the Change Color button next to the parameter you wish to modify 3 Select one parameter to display its peak features during acquisition mode 4 Implement the changes by clicking OK The profile window will display the values of the chosen parameter channel s HIGHEST peak height and peak width width above level defined in criteria for profiles at the right side of the profile window NOTE The peak width determination is made based on user input in the criteria for profiling function See section 3 2 criteria for profiling 2 3 Set Scaling Display of Profiles The profile of each object will be displayed in the profile window so that a line graph of each parameter profiled correspondingly colored will be graphed along the TOF measurement x axis Because each parameter s information will be unique to each sample it is necessary to set the display scale of each parameter independently This allows the user to adjust how each parameter is displayed in the profile window as the data is acquired L Open the Scale dialogue box under the PROFILER pull down menu Two dialogue box tabs will be visible profile scaling and dot plot scaling Dot plot scaling serves as a summary of scales set for each parameter for their display in dot plot mode Choose profile scaling tab to adjust the scaling of prof
25. h above level 100 Mo rules for this channel Yellow High Number of peaks of e lt T Accept All No rules for this channel A Location jo H TE Everything Change peak count setup Red No rules for this channel 1 Open the criteria for profiles feature under the PROFILER pull down menu The criteria for profiles dialog box shown above will appear allowing the user to enter sorting limits rules for each profiled parameter 2 Choose a parameter to set limits from the tabs at the top left corner of dialogue box 3 Set limits for peak Max Height and peak Width above level by typing in values in the appropriate windows or use up and down arrows at the right side of each window See discussion on determining profiler criteria and ranges for Peak Height and Peak Width determination Profiler Criteria and Ranges for Peak Height and Peak Width determination Yellow peak height Red peak height Red peak low limit Red peak width PEAK HEIGHT The highest measured value along the length of the object Peak height measurement is compared with lower and upper limits specified by the user The limits can be set between 1 and 65536 Keep in mind that peak width sort criteria depend on peak height low limit as value to determine peak width 3 3 PEAK WIDTH The program identifies all areas of the profile that exceed the value of the lower peak height limit The system determines a width f
26. he definition of multiple single peaks is advantageous in distinguishing minute variations in the sample However it is important to note that higher scan rates require more computer processing and can result in very large data files It is important to consider testing different scan rates before conducting a profiler experiment More is not always better Multiple parameter data acquisition Profiler allows for sorting based on profile features of multiple parameters as a secondary sorting mechanism for objects fulfilling sort regions in dot plot In the absence of profiler most settings can be manipulated so that slightly differing subpopulations are separated on the gate and sort dot plots However with activation of profiler sortable objects fulfilling dot plot criteria can be further analyzed for presence absence of a peak of interest in one or more profiled parameters This means that a sort decision can be made based on any user defined profile features contained in one or more of the profiled parameters Moreover specific sorting rules can be established for each profiled parameter independently and only events that fulfill ALL the sorting criteria will be dispensed Number of peaks Finally software can count one parameter s peaks over the entire length of the object Using user defined criteria the system will determine peaks within an object then compare to count limits set by user before a sorting decision is made 2 Using Prof
27. iled parameter display Identify the X axis scaling at the bottom half of the dialogue box Use up or down arrows to select the maximum TOF to set profile window display If you are using TOF on either the Gate or Sort dot plot displays it is best to set a maximum equivalent to the TOF in dot plot scale If profiler plots an object with a TOF larger than the maximum the profile may appear truncated within the profile window 4 Identify the y axis scaling Set scaling for each profiled parameter by using up or down arrows at the right end of the window Keep in mind if a profiled parameter exceeds the limits set in scaling the profile graph will contain a flattened peak at the maximum y axis value upper scale limit It may be necessary to change the scaling However if you find that a peak exceeds system maximum value of 65536 signal is saturating the channel s storage limits you may consider changing scan rates PMT or gain settings for this parameter See sections 7 and 9 in COPAS manual for more information on changing these variables Select OK to implement scale settings NOTE Scaling only affects how the profile is displayed not the raw data generated Scaling can be changed at any time during data acquisition 3 Sorting With Profiler 3 1 Summary A sorting decision for every object passing through the flow cell consists of evaluating three different sets of conditions a HISTOGRAM MARKERS OR GATE AND SORT REGIONS Jus
28. iler 2 1 Activate Profiler After normal startup of the COPAS instrument and COPAS software activate the Profiler by checking the Profiler check box on the screen The area in the upper right hand corner displaying the sorting limits for TOF EXT and three fluorescence colors will change to the Profiler display Figure 2 Screen capture of COPAS software main screen E LLLI a AA File Setup Tools Plate Formats Regions Configuration Profiler Mode About Total 4 40 0 Selected Last Sorted 0 Sort o eel q Delay Bo Width Enrich Mo s drops height 2346 width 31 Full Scale MU Tot ACOUITE Ext Erase Green Store dul Red 4 Pressure OK Plate Mover W ReFlex 4 Profiler PMT Control maare one Green 755 Yellow 550 Red jean _Load Piate Fressure Status 4 63 6 76 Je roz ae ez saz ee W Sheath M sample W Sorter Clean Clean Tot READY Scale The message READY should appear on the display in the lower area 2 2 Set Profiler Channels Using profiler will allow you to view profiles in real time for any combination of the parameters profiled User can choose any or all of the four parameter s available Extinction Green Yellow and Red fluorescence 1 Choose Profiling Parameters under the PROFILER pull down menu A dialogue box will appear that allows you to activate any combination of parameters to be displayed in the profiler window by checkin
29. ings for viewing and or sorting based on the unique characteristics of a particular sample in profiler mode The following contains information for an in depth look at how to use profile reader mo ProfReader generic one File Edit View Window Setup Help D W i S S Next Prewious Profile index 14 how Count peaks Highest peak 65534 65534 11112 Width above level 1452 1250 1202 Figure 1 Screen print of profiles viewed through profile reader Notice the display features of the profiles include overlaying graphs of each profiled parameter green yellow and red In this case EXT was not profiled Highest peak and width above level are displayed for each parameter as well as the scale value for TOF 5 2 The Display of a Profile Profiles are plotted with the time of flight of the object along the x axis The value at bottom right of the screen displays the TOF at that point The y axis contains successive profile channels that contain integral signal data for each parameter profiled The display can be thought of as overlaying line graphs of each parameter correspondingly colored plotted along the objects length Specific to each profile are the parameter displays demonstrated at the top of the screen Profile index This is the profile identifying number that corresponds to its location in the txt file of the stored data NEXT PREVIOUS These activity buttons allow the user to scroll through succe
30. ion will be evaluated independently For an object to be dispensed both sets of conditions must be satisfied simultaneously c PEAK COUNT see configuring sorting based on peak count The user must specify a profiler channel under which to count peaks The user can specify the desirable number of peaks within the segment of a profile defined by Location limits Setting the range to the default values of 0 and 250 removes the Number of Peaks from the sorting decision NOTE Sorting based on peak count is a valuable sorting tool employ however it requires very rigid conditional limits to be useful To configure peak counting successfully the anticipated shape of a profile should be understood It is advisable to use ProfReader software on a test sample to optimize settings for peak count implementation See section 7 1 using peak count setup function in ProfReader software All the above conditions are evaluated independently and must be satisfied simultaneously for an object to be dispensed Any or all of these conditions can be configured so that they will be disregarded by setting limits to a maximally wide range 3 2 Configuring Profiler sorting criteria for profiles using peak height and width Criteria for profiles x Summary Extinction Mo rules for this channel Green Extinction Green Yellow Red f Do not set sorting rules for this channel Low High Max Height fico os J E ma E Accept All widt
31. le column and row are Os if only acquiring The txt file s generated will contain depending the export finction unlabeled data column s for profiled parameter s EXT green yellow and or red The data from export of multiple profiles will also be organized in columns and grouped by object The object s index number can be found in the very last row of each data column At this point the user can select the data from the txt file and import it into whichever statistical or graphing software program they wish to use 25
32. ndividual profile SEE DEFINITION ENTER REFERENCE In non profiler configuration 0 PC Ext PC Green PC Yellow PC Red Number of peaks within selected location on an individual profile ENTER REFERNCE LKOCATION Otherwise displays n a In non profiler configuration 0 12 At the bottom left corner of the txt file is a summary of the conditions under which the sample was run These include the displays of Threshold source This is the trigger on channel and displays the value of this threshold setting Min TOF TOF threshold value Signal gains for EXT GREEN YELLOW and RED color channels PMT voltage of each color channel Data string indicating the location and name of the CSV file corresponding to this data acquisition 4 4 Profile Reader compatible file file nameXXX dat The DAT file is a unique system file only compatible with ProfReader software included in the Profiler package This file allows profile specific portions of the txt file to be displayed for visual analysis Object profiles displayed will appear as they did in the profiler window during data acquisition 18 5 Profile reader 5 1 Profile Reader An Overview Profile reader is a feature provided with the new profiler II package that converts the stored DAT file to composite line graphs displaying the profiled parameters for a single analyzed object Profile reader can be used simply to display stored profiles or as a tool to optimize sett
33. of successive point measurements made while the object passes through the flow cell An object containing a small bright fluorescent spot will produce a fluorescence signal with a corresponding narrow peak and the Profiler will digitize the peak into a succession of numbers that directly trace the fluorescence peak as it passed through the flow cell The computer can now perform tests that will detect the presence of different colored fluorescent peaks for the short time they were present in the original signal and note how much each individually rises above its own background level This ability to detect short duration signal peaks that would be swamped out by a single integrated measurement is one way in which the Profiler can enhance detection sensitivity Positional information An additional advantage of the Profiler is that all digitized points have been recorded proximally along with the peak signal so that the position of the peak can be located proportionally in the total list of points This permits extracting positional information from the complete profile rather than simply the presence or absence of peaks Notice how the fluorescence profile signal aligns with the microphotograph in Figure 1 Profiles can be collected for the changes in optical density which we refer to as extinction or EXT and three fluorescence channels along the length of the object simultaneously This correspondence permits testing for the relative positions of
34. or every area of the profile contained above this value See the figure above The sum of all width measurements is the peak width value It is compared with lower and upper limits specified by the user The limits can be set between 0 and 4096 Choose to accept all include or exclude limits you ve set when sorting the sample by choosing the appropriate setting from the pull down tabs next to peak height and width range settings ACCEPT ALL Ignoring the specified ranges for the Peak Height and Peak Width every object passing the flow cell will be dispensed to a microtiter plate or sorted to a receptacle INCLUDE Only those objects that are within the specified ranges for the Peak Height AND Peak Width will be dispensed to a microtiter plate or sorted to a receptacle The conditions for both parameters must be satisfied simultaneously EXCLUDE Only those objects that are outside the specified ranges for the Peak Height OR Peak Width will be dispensed to a microtiter plate or sorted to a receptacle Either parameter falling outside its range will satisfy the EXCLUDE condition NOTE Remember objects can only be sorted if they have also met gate and sort region criteria as displayed in dot plots Set rules for sorting based on number of peaks by entering limits into the appropriate fields or using up and down arrows next to each field See discussion of setting appropriate limits for sorting based on peak
35. pensated parameter values 2 conducting compensation function during profiling is time consuming and may result in too much processing time for accurate sort decision Note compensation should be set using dual parameter Dot Plots in the acquisition software and using approprioate controls for correct values xl Green k1 Yellow k2 Red e f E Figure 7A Original profile Yellow k1 Green k2 Red gt 2 A A AL jo e ax Red k1 Green k2 Yellow ES E HR o 4x Highest peak 655532 6 Width above level 1252 O 1202 MV Compensation enabled Apply Cancel x r Green k1 Yellow k2 Red dl ES E A d Figure 7B Compensated profile yellow signal reduced by 50 green signal Note reduced yellow peak Yellow k1 Green k2 Red HH ka fo Hz 7 k2 jo m Red k1 Green k2 Yellow Highest peak 63532 32766 11112 o ap az Width above level 1252 1222 1202 HA jo ax J Compensation enabled OK Cancel AA 8 Other Features 8 1 Reset to template The Reset to template feature allows the user to revert any changes back to original template settings Only pertains to a file where original matching template was found To do this simply click on the Revert to template under SET UP pull down menu and all modified values will be reverted to values contained within original template If a templat
36. ssive profiles ordered by index number SHOW This activity button that allows the user to advance to that specific profile when user inputs the number in the index window then clicks this button Highest peak This is a numerical value of the parameter s highest peak within the currently displayed profile Width above level This is the summation of all peaks widths as determined at the user defined calculate width on level under channels feature for the currently displayed profile COUNT PEAKS This activity button opens up peak count dialogue box 14 6 Operating Profile Reader 6 1 Opening a File Select Open under the FILE drop down menu and find the appropriate DAT file to open Upon opening a DAT file in Profiler reader the software automatically searches for a matching CSV file for use as a template When a matching template is found the profile reader is automatically initialized with the profile criteria settings used during data acquisition This gives the user immediate display of profiles as they appeared during data acquisition as well as the settings used for COPAS display and or sorting of the sample If a matching template is not found a message appears stating this When this happens user must click OK button on the message before the file will be opened using default settings At this point the user can choose to import a template see Import Template section 4 4 or set up the program for visual analysis o
37. t as with the standard COPAS system an object must lie inside any histogram markers or gate and sort polygonal regions if it is to be dispensed Non profiling systems can have up to 5 parameters that can be used as parameters for histogram markers or gating and sorting dot plots TOF EXT and three colors of fluorescence The Profiler provides additional parameters Peak Height and Peak Width for EXT and each of the three fluorescence colors green yellow and red Any combination of these thirteen parameters can be selected for dot plot displays See section 7 12 Sort criteria in COPAS manual for more information about sorting based on dot plot NOTE The Selected at the top of the screen takes into account only the non Profiler constraints defined by dot plots and is not affected by the Profiler constraints Specifically the following Profiler specific conditions do not affect this number b PEAK HEIGHT AND PEAK WIDTH LIMITS This set of sorting conditions is available only on Profiler systems By setting limits for these two parameters and selecting INCLUDE or EXCLUDE additional constraints for an object s being dispensed are specified If the user keeps these limits wide open and or selects the ACCEPT ALL option then these parameters do not participate in the sorting decision NOTE if these Profile parameters are selected for dot plots the conditions for the dot plots polygons and conditions set in the profile configurat
38. the user to specify a segment over which the number of peaks is evaluated Notice that user MUST set limits to EVERYTHING MIDDLE or ENDS to evaluate peaks over any length of the object EVERYTHING ignores limits set Peaks along the whole profile are included in the Number of Peaks count the Location limits are disregarded MIDDLE will check for peaks appearing within the limits set Only peaks between the From and To Location limits are included in the Number of Peaks count ENDS will check for peaks satisfying criteria outside the limits set by user Only peaks outside the Location limits are included in the Number of Peaks count 1 e only the heads and tails outside the middle are taken into account 9 Set sorting rules by unchecking the do not set sorting rules for this channel located at the top left corner of dialogue box Summary on right half of dialogue box should reflect the rules you have implemented NOTE Unchecking the do not set sorting rules for this channel automatically activates the system to check ALL profiling criteria peak height width AND number of peaks Be sure to set appropriate limits for all profiling parameters or wide enough limits that allow the system to ignore peak height width or peak count analysis if necessary 10 Click APPLY button at the bottom of the dialogue box to implement all the rules you have set 11 12 13 1
39. tructures the EXT optical density measurement may reflect this difference 2 System pressure changes between samples may cause similar objects to be analyzed differently For example if sheath pressure is set too high resulting in a faster sheath flow rate objects carried in that flow will move past the laser faster than under normal flow rates resulting in a smaller TOF measurement Comparing samples acquired under different sheath pressure conditions may look slightly different for this reason 4 2 FCS format file file nameXXX Imd This format is compatible with most flow cytometry software such as FCSExpress The file contains all measured integral parameters extinction fluorescence TOF and peaks parameters height and width for every channel See COPAS system manual section 7 10 data storage 4 3 Summary data file file nameXXX txt Summary data file contains all data points stored after data acquisition Data is organized into columns as described next DESCRIPTION OF COLUMNS IN THE SUMMARY DATA FILE also described in the COPAS system manual ID Sampled objects are given a chronological identifying number starting with 0 This identifier is preserved in txt and DAT files profiler reader so the same object event can be identified in both files PLATE Plates are indexed in chronological order when dispensing into or aspirating from a microtiter plate using a Plate Handler Twister If dispensing into or aspirating from
40. ts are too large For example it is possible that a series of neighboring small peaks will be labeled as a single peak with a width extending over all of them In this case the tallest peak is labeled as a peak and the program will draw the width over the graph of multiple small peaks as in figure D To avoid this tighten the relative peak height and width limits Figure 3D relative height 700 peaks counted 6 width range changed 20 e Show peaks Channel Average on Peak s relative height greater than Peak s absolute height greater than wn mo a pa cL 4 F points 00 Single peak s width from 0 to ro Apply 0 l 6 Set limits for single peak s width Set upper and lower limits for the peak width for peak analysis by entering values in the Single peak s width from boxes Determined width of peak must lie within these limits to be counted as a peak NOTE a generous set of limits will favor counting many peaks while a smaller range may exclude peak widths just outside either end of the limits It is useful to determine the best set of limits by applying different value sets to see which works best for your sample 7 Click Apply to implement all the settings 8 Click OK to close the dialogue box and return to viewer screen Profile viewer will have additional items displayed A black line average on graph closely laid over the parameter channel used to count peaks
41. ttings specific to individual samples for using this sorting feature of the software Peak count is a feature that as the name suggests counts the number of peaks that fulfill user defined criteria See above figure of the dialogue box 7 1 1 OPERATION OF PEAK COUNT L Open the Peak Count Setup feature by selecting it from the SETUP pull down menu See the dialogue box above Za Choose a profiled parameter to count peaks over Do this by locating the channel window and pull down menu towards the top of the box options available are any of the parameters profiled during data acquisition such as Ext green yellow or red fluorescence 3 Choose the number of data points to average the graph over by using up and down arrows next to the average on feature tool See the discussion below to determine a value to average on Average on value The program looks at every data point of the profile averaging data values from surrounding channels to construct a smoothed graph to determine peaks over For example 1f the user selects 5 channels to average on the value displayed in channel 12 will be the average value of data from channels 10 15 This is designed to remove the little variations in fluorescence along the profile in essence reducing the fluctuations based on noise Consider the scan rate Higher scan rates generate a data point as often as every 0 2 microseconds The data reflects a very small slice of the total length of the obje
Download Pdf Manuals
Related Search
Related Contents
User`s Manual Equipment Bedienungsanleitung, Gold- und Karatwaagen (JEWELRY), JB-L ServerView Operations Manager 取扱説明書改訂履歴:PDF形式、105KB Commission éducation du 8 juillet 2013 Philips AquaTouch Moisturizing shaving system GW 90 740A Attuatore 4 canali 16 AX KNX - da guida DIN Manual del usuario Medidor digital de nivel de Copyright © All rights reserved.