Home
5991-0419EN
Contents
1. For example if the return loss of a damaged antenna is actually 6 dB but a jumper cable with 3 dB Insertion loss is connected between the antenna and the analyzer then the measured return loss of the combination of the antenna and cable would be 12 dB 3 OB eners 0 UB eae eo CBee loss IN this case it would be necessary to attach the analyzer directly to the antenna in order to properly measure the true performance of the antenna or perform a user calibration at the end of the jumper cable When making measurements with FieldFox attached to an antenna it may be possible that undesired interference from outside the system may be captured by the antenna and measured by FieldFox The interfer ence may appear as a spike or lack of stability in the measured trace FieldFox includes an Interference Re jection setting that may minimize the effects of any interfering signal The Interference Rejection setting is found under the Measurement Setup menu Once enabled up to six measurement sweeps may be required to reduce or eliminate the interference 08 Keysight Techniques for Precise Cable and Antenna Measurements in the Field Using FieldFox Handheld Analyzers Application Note Antenna to Antenna Isolation In many wireless applications dif ferent systems will be co located to share tower and shelter structures For example figure 6a shows a typ ical repeater where antennas from multiple systems are mounted onto the co
2. 18 00 GHz Swp 197 ms Figure 3 Measurement of coaxial cable insertion loss using a novel one port technique yellow and a traditional two port technique blue both available on FieldFox 06 Keysight Techniques for Precise Cable and Antenna Measurements in the Field Using FieldFox Handheld Analyzers Application Note Cable Insertion Loss Measurements continued There is another configuration for measuring cable insertion loss that does not require a long coaxial cable in the return path back to FieldFox In this configuration FieldFox is config ured as a power meter connected to a Keysight U2000 Series USB power sensor As shown in figure 4 one end of the cable under test is directly con nected to the RF OUT port of FieldFox The other end of the installed cable is connected to the USB power sensor FieldFox is configured to generate a CW signal at the RF OUT port This CW signal is the test signal that will be transmitted along the cable and measured by the USB power sensor If the cable ends are physically separat ed by a large distance the USB sensor can be connected back to FieldFox through a USB cable extender USB cable extenders are commercially available with ranges of 500 meters or more This configuration overcomes the limitation of running an expensive coaxial cable between one end of the installed cable back to the instrument The main constraint using the power meter solution is that swept frequency measur
3. antenna and any free space loss are removed from the transmission measurement resulting in the desired antenna gain The measurement of antenna gain in the field is rather difficult as environmental reflections and inaccessibility around the anten na prevent this type of testing When an antenna is installed in a system it will therefore be required to only measure the reflection properties of the antenna in order to determine If an antenna is faulty or damaged Return loss and the associated VSWR will be used to characterize the performance of an antenna These 1 port measure ments are typically performed over the intended frequency band of interest A typical antenna with a return loss of 10 dB or higher would represent a reasonably well matched antenna A 10 dB return loss is equivalent to having 90 of the incident energy radiated by the antenna with 10 of the energy being reflected back to the transmitter For some applications where high efficiency is required such as base station antennas the return loss may be specified at 15 dB or higher A 10 dB return loss is approx imately equal to a VSWR of 2 1 which is also quoted in the specifications Appendix A includes a conversion table relating values of return loss to VSWR Fortunately FieldFox handles this conversion and can be configured to display the measured response as return loss dB or VSWR Figure 5 shows the measured reflec tion characteristics of a 2
4. calculated from the range resolution multiplied by the number of points minus one The third DTF instrument setting is the DTF Window In general when troubleshooting faults in transmis sion systems the default setting with a Maximum Window provides the highest dynamic range for the DTF measurements Under conditions when attempting to resolve two close ly spaced discontinuities a Minimum Window would be preferred Addition al information regarding Frequency Mode Resolution and Windowing can be found in the FieldFox User s Manual Calibrating FieldFox s Test Ports FieldFox offers a set of user calibra tions that can remove the effects of cables and adapters that may be placed between the instrument and the device under test These types of user calibrations are performed by the operator on a regular basis and are independent of the general instrument calibration that may be required ona yearly cycle The yearly instrument cal ibration is a traceable process that is performed by acertified test lab such as Keysight and confirms that the instrument is performing to its stated specifications The yearly instrument calibration should be traceable to International System SI units through a national metrology institute such as NIST NPL or BIPM With a fully trace able instrument calibration FieldFox is capable of making highly accurate measurements at the instrument test ports In practice different test require
5. measure cable insertion loss from only one end of the cable under test This eliminates the need to carry an extra long high quality test cable as part of the equipment requirements Figure 2b shows the simple 1 port configuration for measuring cable insertion loss by attaching one end of the cable to FieldFox and leaving the other end open or terminated in a Short It is preferred at microwave frequencies to use the short circuit to eliminate fringing fields found in an open ended cable which could alter the measured results In this configuration FieldFox injects a test signal into the cable from the RF OUT port same as before The test signal passes through the cable is com pletely reflected from the open end or shorted end passes through the cable a second time and lastly measured by FieldFox using the same port FieldFox is configured with an internal factory calibrated reflectometer 1 for measuring reflected signals at the RF OUT port This internal calibra tion called CalReady is available at instrument turn on and instrument preset and will be discussed later in this application note Once the reflected measurement is complete FieldFox using a built in model for coaxial cable dispersion and knowing that the measured signal contains twice the cable insertion loss due to the round trip path can now report the cable insertion loss as a function of frequency Figure 3 shows a comparison of the measured i
6. required Reducing the IF bandwidth setting on FieldFox will also reduce displayed noise The IF bandwidth can be adjusted over the range of 100 Hz to 300 kHz It should be noted a lower IF bandwidth will also increase the overall measurement time as lower bandwidths increase the instrument sweep time extender Figure 4 Configuration for measuring cable loss using FieldFox and a USB power sensor Return Loss dB CAT Ref 0 00 dB Start 2 200 GHz Points 201 Output Power HIGH CAT Ref 20 00 Stop 2 700 GHz Swp 165 0 ms Start 2 200 GHz Points 201 07 Keysight Techniques for Precise Cable and Antenna Measurements in the Field Using FieldFox Handheld Analyzers Application Note Stop 2 700 GHz Output Power HIGH Swp 738 0 ms Figure 5a and 5b Frequency response measurement of a panel antenna showing a return loss dB and b VSWR Antenna Return Loss and VSWR Measurements Antennas are specified by their gain factor and return loss or voltage standing wave ratio VSWR The return loss and VSWR are standard measurements reported for most RF and microwave components and sys tems and can easily be measured with FieldFox in the field or in the lab An tenna gain measurements are typically performed in a special test facility such as an anechoic chamber During antenna gain testing a transmission measurement is made from the anten na under test to a second calibrated antenna The effects of the calibrated
7. that coaxial cable having an air di electric has the lowest insertion loss at 75 ohms and the highest power han dling capability at 30 ohms The Cable TV CATV industry has standardized on 75 ohm cable in order to maximize signal transmission over very long cable distances required by the sys tem On the other hand the majority of the RF and microwave industry has Standardized on 50 ohm impedance as a compromise between the lowest loss and the highest power handling Most RF and microwave analyzers such as FieldFox are configured with 50 ohm test port impedances When measur ing 7 o ohm cables and components 90 ohm to 75 ohm adapters such as the Keysight Technologies Inc N9910X 846 should be connected to the analyzer Once the transmission line cables are installed into a system It is often difficult and costly to remove them in order to verify their operation and troubleshoot cable failures Also with very long cable runs access to both ends of the cable at the same time is typically impossible especially when attempting to connect the cable to the test instrumentation Under these conditions techniques that allow cable insertion loss measurements to be made from only one end of the cable are preferred Figure 2 shows two configurations for measuring the insertion loss of a cable using Field Fox with CAT mode Figure 2a shows the traditional method for measuring the insertion loss of a cable using a two port te
8. these two peaks marker 4 is measured to be located at 23 7 meters and marker 5 is located at 27 6 meters These measurement peaks are the re sult of signals that travel through sec tions of the cables multiple times and interact with open circuit and adapter discontinuities along the way Locating Faults Along Transmission Lines continued Figure 11 shows a simplified diagram on the various propagation paths back to FieldFox for the cable con figuration shown in figure 8 Figure 11a shows the main reflected signal from the open ended cable resulting in the signal at 13 8 meters marker 3 in figure 10 Figure 11b shows the re reflections between the adapter and the open resulting in a peak at approximately 23 7 meters marker 4 The last peak not shown in figure 11 represents the re reflections between FieldFox and the open resulting in a peak at approximately 27 6 meters marker 5 4m 9 8m 9 8m 4m 4m 9 8m 9 8m 4m 2 Under some instrument settings it is possible that a measured peak could be an artifact or alias of the calculat ed DIF response An alias is not a true cable response and may appear when FieldFox reports an Alias free Range Off as shown under the DTF Settings page Aliases will not appear when FieldFox reports an Alias free Range On These undesired peaks are more likely to occur with FieldFox settings at short stop distances less than 10 meters and or high resolution 1001 p
9. to enter into FieldFox in order for the instrument to correctly display the distance to each cable discontinuity Velocity factor is the signal s propagation velocity in the transmission line relative to signal velocity travelling in a vacuum speed of light When a signal travels in a coaxial cable the dielectric between the inner conductor and outer conduc tor of the cable will reduce the signal s velocity resulting in a velocity factor less than 1 0 For the cable specified in figure 8 the signal s velocity factor in this cable is 0 66 which implies that the signal travels at a rate of 66 of the speed of the light in free space The velocity factor is entered into FieldFox under the DTF Cable Speci fications located in the Measurement setup menu Cable specifications Cable loss dB m 50 MHz Cable loss dB m 100 MHz 0 06 Cable loss dB m 500 MHz Cable loss 4B m 1000 MHz FieldFox DTF cable spec menu Figure 8 Configuration for measuring the distance to fault dB in a system having two coaxial cables con nected to a load also shown are the specifications for the cables used in this example Locating Faults Along Transmission Lnes continued Figure 9 shows the DTF measurement for the simple example shown in figure 8 The cables are initially terminated in a 50 ohm coaxial load Markers are placed at the three peaks in the DIF response The first marker shown on the far left reports a d
10. watch videos and learn more www keysight com find FieldFox 18 Keysight Techniques for Precise Cable and Antenna Measurements in the Field Using FieldFox Handheld Analyzers Application Note myKeysight P g m y w WARRANTY WARRANTY myKeysight www keysight com find mykeysight A personalized view into the information most relevant to you Three Year Warranty www keysight com find ThreeYearWarranty Keysight s commitment to superior product quality and lower total cost of ownership The only test and measurement company with three year warranty standard on all instruments worldwide Keysight Assurance Plans www keysight com find AssurancePlans Up to five years of protection and no budgetary surprises to ensure your instruments are operating to specification so you can rely on accurate measurements www keysight com go quality Keysight Technologies Inc DEKRA Certified ISO 9001 2008 Quality Management System Keysight Channel Partners www keysight com find channelpartners Get the best of both worlds Keysight s measurement expertise and product breadth combined with channel partner convenience www keysight com find FieldFox KEYSIGHT TECHNOLOGIES Unlocking Measurement Insights For more information on Keysight Technologies products applications or services please contact your local Keysight office The complete list is available at www keysight com find contactus Americas Can
11. 1 30 24D 1 09 37 0 1 03 47 0 1 01 136 1 91 17 8 1 30 27 5 1 09 oo 1 03 47 5 1 01 VSWR is sometimes stated as a ratio For example 1 2 1 one point two to one VSWR The first number is the VSWR value taken from the table or calculated using the formula The second number is always 1 17 Keysight Techniques for Precise Cable and Antenna Measurements in the FieldUsing FieldFox Handheld Analyzers Application Note Carry Precision With You Every piece of gear in your field kit had to prove its worth Measuring up and earning a spot is the driving idea behind Keysight s FieldFox analyzers They re equipped to handle routine maintenance in depth troubleshooting and anything in between Better yet FieldFox delivers Keysight quality measurements wher ever you need to go Add FieldFox to your kit and carry precision with you Related literature Number FieldFox Combination Analyzers Technical Overview 9990 9780EN FieldFox Spectrum Analyzers Technical Overview 5990 9782EN FieldFox Vector Network Analyzers Technical Overview 5990 9781EN FieldFox Handheld Analyzers Data Sheet 5990 9783EN FieldFox Handheld Analyzer Configuration Guide 5990 9836EN FieldFox N9912A RF Analyzer Technical Overview 5989 8618EN FieldFox N9912A RF Analyzer Data Sheet N9912 90006 FieldFox N9923A RF Vector Network Analyzer Technical Overview 9990 5087EN FieldFox N9923A RF Vector Network Analyzer Data Sheet 5990 5363EN Download additional application notes
12. 4 GHz patch panel antenna This antenna included a short 6 inch length of 50 ohm coaxial cable ending with a Type N female connector at the end A standard Type N male to male adapter was re quired to interface the antenna to the Type N female connector of FieldFox When interfacing any component to an instrument it is very important to use high quality adapters to improve the accuracy and repeatability of the measurement For this measurement example the adapter and cable was included as part of the measured results The adapter characteristics can be removed from the measured results by using the QuickCal or O S L calibration available on FieldFox Calibration options will be discussed later in this application note Figure 5a shows the measured return loss over the range of 2 2 to 2 7 GHz Two mark ers are placed at the points of 10 dB return loss signifying an appropriate bandwidth for this antenna A marker table lists the amplitude and associ ated frequency for each active marker up to a total of six Figure 5b shows the equivalent VSWR measurement for the same antenna When measuring the reflection characteristics of an antenna It is important to attach the analyzer as close to the antenna as possible If a long length of cable is inserted between the antenna and the analyzer or if the cable has become excessively lossy due to damage then the high insertion loss of the cable will mask the true return loss of the antenna
13. Keysight Tecnnologies Techniques for Precise Cable and Antenna Measurements in the Field Using FieldFox handheld analyzers Application Note This application note introduces the practical aspects of cable and antenna testing interpreting mea surement results and instrument operation including calibration options such as CalReady and QuickCal using a Keysight Tech nologies Inc FieldFox handheld analyzer configured as a cable and antenna analyzer Measurement examples are provided showing techniques for measuring insertion loss return loss and locating faults in a transmission system Carry precision with you KEYSIGHT TECHNOLOGIES Unlocking Measurement Insights Introduction Cable and antenna measurements are often required to verify and troubleshoot the electrical performance of RF and microwave transmission systems and antennas In RF and microwave communications industries measurements are often made along the coaxial cable connecting a transmitter to its antenna and or between an antenna and its receiver This process is sometimes referred to as Line Sweeping and results in the measurement of the signal attenuation and return loss as a function of frequency Line sweeping may also be used to estimate the physical location of a fault or damage along the transmission line using the Distance to Fault DTF measurement available on many RF and microwave signal analyzers such as FieldFox Testing the performance of transmis
14. acy and repeatability in the measured data CalReady The first and easiest calibration to implement is CalReady also known as Preset Cal CalReady is immediately available upon instrument power on or instrument preset and requires no ad ditional steps to calibrate This built in calibration is performed at the factory over the full frequency range of Field Fox and applies directly to the RF OUT test port of FieldFox CalReady as sumes that the device under test DUT will be connected directly to FieldFox s Type N female test port connector or the 3 5 mm male connector on 26 5 GHz models As shown in figure 14a CalReady has a calibration plane at the connector of FieldFox and is available for all reflection measurements includ ing return loss VSWR 1 port cable loss and DTF When using CalReady any jumper cables and or adapters connected between FieldFox and the DUT will be included in the measured response 13 Keysight Techniques for Precise Cable and Antenna Measurements in the FieldUsing FieldFox Handheld Analyzers Application Note QuickCal The second cal type available for reflection measurements is QuickCal QuickCal is useful when jumper cables and adapters are placed between FieldFox and the DUT and the user does not want to include their ef fects in the measured data QuickCal corrects for phase shift and loss of cables and adapters and electrically moves the calibration plane to the end of these compo
15. ada Brazil Mexico United States Asia Pacific Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Other AP Countries Europe amp Middle East Austria Belgium Finland France Germany Ireland Israel Italy Luxembourg Netherlands Russia Spain Sweden Switzerland United Kingdom 877 894 4414 oo J 3351 7010 001 800 254 2440 800 829 4444 1 800 629 485 800 810 0189 800 938 693 1 800 11 2626 0120 421 345 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 65 6375 8100 0800 001122 0800 58580 0800 523252 0805 980333 0800 6270999 1800 832700 1 809 343051 800 599100 32 600 58580 0800 0233200 8800 5009286 800 000154 0200 882255 0800 805353 Opt 1 DE Opt 2 FR Opt 3 IT 0800 0260637 For other unlisted countries www keysight com find contactus BP 04 23 15 This information is subject to change without notice Keysight Technologies 2012 2015 Published in USA May 15 2015 5991 0419EN www keysight com
16. al cable waveguide and twisted pair line and antennas requires fast and accurate measurements of the insertion loss and the return loss as a function of frequency Also important to field operations is the capability to quickly troubleshoot and locate faults and damage in cables and connectors A modern handheld analyzer such as FieldFox configured for cable and antenna testing CAT is an extremely useful tool for characterizing the entire transmission system as well as the performance of individual components in the system In the next few sections of this application note measurement examples will be provided showing techniques for measuring insertion loss return loss and locating faults in a transmission system The measure ments will be made using a FieldFox N991XA analyzer with frequency range up to 26 5 GHz Cable Insertion Loss Measurements The insertion loss of transmission line or coaxial cable is often measured as a function of the intended operating fre quency band The insertion loss also referred to as the cable attenuation is the amount of energy dissipated in the cable The insertion loss also includes energy lost due to mismatch reflection between the source and load It is expected that the source transmitter the transmission line coaxial cable and the load antenna are all designed for the same characteristic impedance Zo usually 50 ohms or 75 ohms when using coaxial cable It has been shown 1
17. ame cable would have 7 4 dB loss at 1 5 GHz Those familiar with vector network analyzers VNA will recognize this insertion loss measurement as the absolute value of the S21 transmission scattering parameter S parameter Additional information regarding VNA options on FieldFox and the basics of network analysis can be found at the following references 1 and 2 It should be noted that accurate inser tion loss measurements require an initial calibration of FieldFox Open or short Cable loss 1 port b Figure 2 Configurations for measuring the insertion loss of a cable using FieldFox with CAT mode Cable Insertion Loss Measurements continued The calibration removes the insertion loss effects of the equipment adapt ers and jumper cables as a function of frequency The calibration of FieldFox for making one and two port measure ments will be discussed later in this application note Figure 3 shows the insertion loss measurement of a co axial cable using the 2 port technique over the frequency range of 30 kHz to 18 GHz This figure also shows a mea surement of the insertion loss using a 1 port cable loss technique to be discussed next As previously mentioned when a cable is installed into a system it is often difficult to physically connect FieldFox to both ends of a very long cable with out Introducing an equally long jumper cable into the test setup Fortunately FieldFox has a convenient technique to
18. and reflection testing While these cal types result in the highest level of measurement accuracy the user will be required to carry a high quality calibration kit into the field Jumper cable Jumper cable Figure 15 Configuration for the Normalization Cal Type using a jumper cable s required for 2 port insertion loss testing of a DUT 16 Keysight Techniques for Precise Cable and Antenna Measurements in the FieldUsing FieldFox Handheld Analyzers Application Note Conclusion References This application note has introduced 1 Keysight Application Note Network Analyzer Basics Literature Number measurement and calibration tech 5965 7917E August 2004 niques for cable and antenna testing 2 www keysight com find FieldFox CAT using FieldFox Measurement 3 Copyright J M Briscoe and licensed for reuse under the Creative examples were provided for cable Commons Licence insertion loss return loss VSWR 4 Keysight Application Note 1287 12 Time Domain Analysis Using a Net and Distance to Fault DTF testing work Analyzer Literature Number 5989 5723EN May 2012 Antenna measurements were also re viewed and several examples of return loss and antenna to antenna isolation were provided Additional informa tion regarding CAT and line sweeping can be found at the Keysight web CalReady allows users to make cable and antenna measurements at test site Keysight also provides local and ports without calibrati
19. antenna to antenna isola tion is above 44 dB across the 902 to 928 MHz band but is only 35 dB in the 2 4 to 2 485 GHz band This level of isolation implies that energy transmit ted from the 2 4 GHz system will enter the 9OO MHz receiver front end ata Isolation VA a y Antenna 1 Antenna 2 AE T N Insertion loss 2 port Figure 6b Configuration for measuring antenna to antenna isolation Figure 6a Microwave repeater showing multiple co located antennas Ref 0 00 dB Insertion Loss 2 Port Start 800 0 MHz Points 201 Stop 3 000 GHz Output Power HIGH Swp 304 0 ms Figure 7 Measured antenna to antenna isolation between a 900 MHz panel antenna to a 2 4 GHz panel antenna mounted on the same structure system will need to have adequate filtering to provide additional rejection of the 2 4 GHz signal otherwise the 900 MHz receiver will be desensitized by this high level signal 09 Keysight Techniques for Precise Cable and Antenna Measurements in the Field Using FieldFox Handheld Analyzers Application Note Locating Faults Along Transmission Lines Once It has been determined that the cable insertion loss is higher than expected or the return loss and associated VSWR is out of spec it becomes necessary to find the possi ble locations for the fault s along the transmission system The location of discrete and periodic cable faults can be determined using the distance to faul
20. bra tion now includes the loss and phase response of the jumper cable this cable should also be included when making measurements of the DUT as shown in figure 15a If the jumper cable is removed after the calibration the measurements of the DUT will be off set by the value of the insertion loss of the jumper cable resulting in a slightly lower measured value of insertion loss for the DUT It is also possible to use two jumper cables and or adapters during the instrument calibration with one connected to each port of FieldFox as shown in figure 15b Figure 15b also shows the DUT measure ment configuration with connection between the two jumper cables While the Normalization calibration is one of the easiest to implement is the least accurate of all the cal types available on FieldFox It is often advisable to use the one of the other cal types such as CalReady or QuickCal for improved measurement accuracy It should also be noted that the Normalization cal type is also available for reflection measurements where an open or short is used to terminate the test port during the calibration QuickCal QuickCal allows users to extend the reference plane to the end of the jumper or adapter without using a cal kit and supports many commonly used connector types QuickCal for insertion loss measure ments begins with the same calibra tion procedure as that for reflection testing but the measurement sweep of the open ended jump
21. chnique Figure 2b shows a novel technique available on the Field Fox for measuring the insertion loss of a cable using measurements obtained from only one end of the cable Both of these techniques will be discussed In the traditional two port setup the cable under test is connected to two Separate ports on FieldFox FieldFox injects a test signal into the cable from the RF OUT port conveniently posi tioned along the top of the analyzer As the test signal passes through the cable a small portion of the energy Is absorbed by resistive and dielectric losses In the cable Discontinuities from cable connectors cable splices damage and other factors will reflect a portion of the energy back to the source resulting in an additional in crease in the measured insertion loss The remaining signal exiting the cable is then measured by FieldFox at the RF Insertion loss 2 port a 04 Keysight Techniques for Precise Cable and Antenna Measurements in the Field Using FieldFox Handheld Analyzers Application Note The ratio between the input and output signals represents the total insertion loss of the cable usually expressed in decibels dB An ideal lossless cable would have O dB insertion loss The cable manufacturer usually provides a table of the insertion loss as a func tion of frequency For example an LMR9OO coaxial cable would have 2 2 dB insertion loss for a 100 meter length of cable measured at 150 MHz The s
22. ements are not available and therefore changing test frequencies will require manual tuning of FieldFox s settings When measuring cables with high insertion loss the displayed measure ment trace may exhibit a relatively high level of noise as the measured signal amplitude approaches the noise floor of the analyzer One way to improve the relative signal to noise and im proving the associated measurement accuracy is to set the output power on FieldFox to High Power mode High Power mode is the default setting for FieldFox and is typically used for most passive device measurements The Low Power mode Is typically used when measuring high gain or low noise amplifiers to prevent saturation and potential damage to the amplifier The High Power setting results in a port power of approximately 5 dBm the Low Power setting is approximately 25 dBm Other techniques to reduce the displayed noise level include increas ing the number of trace averages or decreasing the IF bandwidth setting Trace averaging will reduce the effects of random noise on the measurement by averaging the results from multiple sweeps FieldFox trace averaging can be set over a range from 1 no averag ing to 1000 Power sensor Long cable under test USB cable Knowing that trace averaging requires the measurement of multiple traces it should be expected that there will be an increase in the total measure ment time as multiple sweeps are
23. er cable s and the optional LOADs occurs at both RF OUT port 1 and RF IN port 2 of FieldFox QuickCal for insertion loss testing also requires an additional measurement sweep of a jumper ca ble s connecting RF OUT port 1 to RF IN port 2 This additional measure ment of the jumper cable is similar to the calibration step shown in figure 15 After calibration FieldFox will display insertion loss measurements without including the effects of any jumper cables and adapters used as part of the calibration Enhanced response and full 2 Port calibrations The Enhanced Response and Full 2 Port cal types provide the highest level of measurement accuracy of all the cal types available on FieldFox Calibration Jumper cable Jumper cable or Testing DUT Jumper cable a A Full 2 Port calibration requires measurement sweeps of the OSL standards placed at both test ports of FieldFox This cal type also requires measurement sweeps of the jumper cable s connecting RF OUT port 1 to RF IN port 2 The Enhanced Response follows the same calibration procedure as the Full 2 Port but does not require measurement sweeps of the OSL standards placed at port 2 of FieldFox Without measuring the OSL standards at port 2 the Enhanced Response calibration is slightly less accurate than the Full 2 Port calibra tion The Enhanced Response and Full 2 Port cal types can be used for both insertion loss testing
24. er on or instrument preset CalReady is factory calibrated at FieldFox test ports and is very useful to characterize the Insertion loss of jumper cables In this case the jumper cable will be connected directly to RF OUT port 1 and RF IN port 2 of FieldFox When measuring the inser tion loss of a DUT jumper cables and or adapters connected between Field Fox and the DUT will be included as part of the DUT s measured response CalReady QuickCal 0 S L Open Load Open Short Load alibration i a na Me nes i joes O eaumecins 7 ENEA plane Adapter Adapter Jumper cable Jumper cable Optional a b c Figure 14 Calibration types and calibration standards required for accurately measuring reflection properties using FieldFox 15 Keysight Techniques for Precise Cable and Antenna Measurements in the FieldUsing FieldFox Handheld Analyzers Application Note Calibrating FieldFox s Test Ports continued Normalization A very simple user calibration for removing the effects of the jumper cable and the frequency response of the instrument is the Normalization cal type During this calibration process a short jumper cable is connected be tween the RF OUT port and the RF IN port on FieldFox see figure 15a Once the jumper cable is connected a single measurement sweep is performed during the calibration All DUT mea surements are then normalized or ref erenced to this cable As the cali
25. guide components As resolution in distance is inversely proportional to measure ment bandwidth the Bandpass mode will not provide the same high reso lution ina DTF measurement as the Lowpass mode In Lowpass mode the measured frequency range is automat ically determined after the start and stop distances are entered for the DIF display Ref 0 00 dB Start 0 000 m Points 1001 Frequency Mode Lowpass Output Power HIGH 12 Keysight Techniques for Precise Cable and Antenna Measurements in the FieldUsing FieldFox Handheld Analyzers Application Note Distance To Fault dB Stop 5 000 m Swp 683 0 ms velocity Factor Cable Loss Cable Corr Edit Save Recall TT Figure 13 Measurement of a Category 6 patch cable with adjustment to FieldFox s velocity factor correction in order to match the measured distance to the known physical distance In Bandpass mode the measured start and stop frequencies are manually entered based on the approximate frequency range for the components under test The second DTF instrument setting is related to the Resolution or number of measurement points The Resolu tion can be selected from 101 201 401 601 801 and 1001 points A larger number of points increases the measurement sweep time but also increases the maximum distance that can be measured on the DIF display The range resolution is calculated from the measured frequency span The maximum distance that can be displayed is
26. he cable is known then the DIF mea surement may be used to estimate the velocity factor In this case the cable is attached to FieldFox and the VF is adjusted until the measured distance matches the known physical length For example figure 13 shows the mea surement of a 3 meter length of Cate gory 6 twisted pair cable terminated in an open end The cable was attached to a 100 ohm balanced to unbalanced transformer to match the 100 ohm impedance of the twisted pair cable to the 50 ohm impedance of FieldFox s test port connector The VF setting was adjusted until the measured length was 3 meters and for this exam ple the velocity factor was 0 70 70 This velocity factor may now be used for DIF measurements of other Cate gory 6 cables from this manufacturer There are three additional instrument settings that control how FieldFox calculates and displays DIF mea surements The first setting is related to the Frequency Mode of operation found under the Measurement Setup menu on FieldFox There are two fre quency modes available when making DTF measurements Lowpass and Bandpass The Lowpass frequency mode will provide the optimum resolu tion in measured distance but requires test components that have a very wide frequency bandwidth such as trans mission lines including coaxial cables and adapters The Bandpass frequency mode is required for components that have a narrow operating band such as filters antennas and wave
27. ipboard aircraft and certain automotive applications At an outdoor installation of a wireless cellu lar system the antenna and portions of the coaxial cabling are often exposed to extreme weather conditions includ ing rain snow ice wind and lightning severe environmental conditions can potentially create physical damage to the system components including failed waterproof sealing at interfaces between cables and connectors failed sealing at cable splices and weath er induced cracks in the insulation resulting in an ingress of moisture into the cables and interfaces Sheltered installations can expose components to mishandling stress heat vibra tion and oils and other containments that could leak into the system Cable fault can also occur at any Interface between the cables and connectors where solder joints and cable crimps become fatigued and break It is also possible that cables can be dented crushed or improperly routed during or after installation Conditions may also be created where the minimum bend radius is exceeded beyond the value specified by the cable manufacturer For example low loss coaxial cable may have a specified minimum bend radius of 1 inch to over 10 inches and bending the cable below the specifi cation will cause damage to the cable and substantially degrade its electrical performance Verifying and maintaining the opera tion of the RF and microwave transmis sion system Including coaxi
28. istance of O meters This marker represents the interface between FieldFox and the first coaxial cable The second marker reports a distance of 4 meters This marker is located at the adapter be tween the two cables It also implies that the length of the first cable is 4 meters The third marker is located at the 50 ohm load and is reported at 13 8 meters This measurement can be used to calculate the length of the second cable which is 9 8 meters 13 8 meters 4 meters There is a notice able drop in the measured amplitude after the 50 ohm load signifying the location of the end of the cable It should be noted that if the cable s ve locity factor was not correctly entered the distance measurements would not be correct making it difficult to accurately locate the physical dis tance to a discontinuity in the cable It should also be noted when two cable types with different velocity factors are included in the measurement such as a short jumper cable connected to a long Helix cable the velocity factor of the longer cable should be entered into FieldFox Ideally the short jumper cable should be included as part of QuickCal and therefore its effects will be calibrated out from the measure ment Figure 10 shows the measurement for the same configuration as in figure 8 but the 50 ohm load was removed and the cable was left open ended As a comparison the yellow trace is the current measurement having the cable with the
29. le type selection and cable parameter menus on FieldFox the measured amplitude will not be cor rect For this measurement the cable loss was entered as 0 dB m default setting so the insertion loss of this transmission system is mask ing the true return loss from the open circuit If the application requires an accurate amplitude measurement of the return loss from a particular reflection on the DIF display the cable loss should be entered either manually into FieldFox or selected from a list of Stored cable types Figure 12 shows the cable selection menu left menu on FieldFox In this case the RG8U cable type is highlighted and selected The cable parameters can be viewed and edited as shown in the table on the right of figure 12 For this RG8U type the velocity factor is 0 848 and the cable loss dB m is stored as a function of frequency Cable loss values are inter polated between the entered frequen cy points In general DIF measure ments of transmission systems are primarily concerned with locating the physical position of faults along the line and therefore precise amplitude measurements of the return loss and associated VSWR may not be required Accordingly most applications only require that the cable s velocity factor be entered into the FieldFox cable specification table Locating Faults Along Transmission Lines continued If the velocity factor VF of a cable is unknown but the physical distance of t
30. ments may result in a variety of jumper cables and adapters being connected to FieldFox Often it is pre ferred that the effects of these cables and adapters are not included in the measurement In this case FieldFox allows a user calibration to be per formed with or without a calibration kit in order to extend the calibration plane beyond the jumper cables and adapters Operators familiar with the operation of vector network analyzers VNAs will be familiar with these types of user calibrations The remaining sections of this application note will introduce the different types of user calibrations available for accurately measuring the reflection characteristics and insertion loss of components and systems Calibrating for reflection measurements There are three types of user calibra tions available for accurately measur ing the reflection characteristics of transmission systems and antennas User calibration will improve the results for all previously discussed measurements of return loss VSWR 1 port cable loss and DIF The cali bration types cal types for reflection measurements are CalReady QuickCal and 1 port OSL Figure 14 shows the configuration diagrams for these three cal types including any jumper cables adapters and calibration standards that may be required In general for all cal types jumper cables and adapt ers should be high quality items and maintained in very good condition to improve the accur
31. mmon structure It is important that antennas either operating within the same system or between different systems are physically spaced far enough apart to maintain an adequate level of isolation in order to prevent the creation of intermodulation distor tion in the transmit amplifiers and or noise desensitization in the receivers Intermodulation distortion and receiver desensitization may be created when energy from a high power transmitter radiated from one antenna couples into a nearby antenna and enters the am plifiers of the coupled system Systems are often designed with duplexing and other filters to reject signals from nearby transmitters and other interfer ence but system performance may re quire an antenna to antenna isolation of 60 dB or more It is therefore neces sary to sweep the antenna to antenna isolation across all frequency bands of interest Figure 6b shows a configura tion for measuring the antenna to an tenna isolation using FieldFox FieldFox is set to measure 2 port insertion loss and calibrated using short jumper cables with a normalize calibration type The measurement of isolation is reciprocal so it does not matter which antenna is attached to the RF OUT port of FieldFox Figure 7 shows an example of the measured isolation between two panel antennas mounted to a common structure One antenna was designed to operate in the 9OO MHz band and the other at 2 4 GHz As shown in this figure the
32. nents see figure 14b QuickCal requires a single measure ment sweep of an open at the end of the jumper cables and adapters no additional calibration standards are required Once the calibration sweep is complete the DUT may be mea sured without including the effects of these jumper cables and adapters Jumper cables should be high quality ophase stable test cables It should be noted that if the DUT has a good return loss greater than 20 dB to 25 dB an optional LOAD standard may be measured during QuickCal to improve the accuracy of the DUT measure ments Ideally the LOAD standard should have the same connector type as the DUT The Keysight N9910X 800 801 tee calibration kit contains a Type N male female 50 ohm termi nation 14 Keysight Techniques for Precise Cable and Antenna Measurements in the FieldUsing FieldFox Handheld Analyzers Application Note Calibrating FieldFox s Test Ports continued 1 port OSL The third cal type available for re flection measurements is 1 port OSL This calibration requires a calibration kit where the O is the OPEN standard the S is the SHORT and the L is the LOAD standard Calibration kits are available in Type N 50 ohm and 75 ohm 7 16 3 5 mm Type F 7 mm TNC and several waveguide kits including X band P band and K band When the 1 port OSL cal type is selected on FieldFox the connector type must be chosen including the connector gender male or female The gende
33. nser tion loss using the 2 port and 1 port techniques In this figure the blue line represents the 2 port insertion loss measurement and the yellow line represents the 1 port cable loss mea surement As observed in figure 3 the 1 port technique contains additional amplitude ripple across the measured frequency range due to mismatch effects between the open short and the input connector Using FieldFox the amplitude ripple may be reduced with an additional measurement of a 50 ohm load placed at the end of the cable under test In this case the measurement of the 50 ohm load is saved to memory using the Trace Data gt Mem CAT Ref 1 80 dB Start 30 00 kHz Points 201 Output Power High 05 Keysight Techniques for Precise Cable and Antenna Measurements in the Field Using FieldFox Handheld Analyzers Application Note The load is removed and the cable is terminated in an open or short same as before The measurement of the open short is then subtracted from the memory using the Data Math Da ta Mem This additional measurement step may improve the observed ripple in the 1 port cable measurement In general the traditional 2 port insertion loss measurement will be more accu rate than a 1 port cable measurement but having a measurement process that does not require an instrument connection to both ends of the cable is a great benefit when characterizing installed cabled systems Cable Loss 1 Port Stop
34. oints When an alias is suspected a simple measurement procedure will help determine which peaks are true responses and which are aliases Place a marker on the response in question and note the measured distance Change the start or stop distance setting on FieldFox The displayed distance of a true response will not change On the other hand the dis played distance of an alias will appear to move and therefore this peak should not be recorded as an actual DIF response It should be noted that for the mea surement example shown in figure 10 the expected return loss from an open circuit should be O dB but is mea sured at 9 9 dB marker 3 Unless the specified cable loss dB m is properly entered into FieldFox I Cable 1 11 Keysight Techniques for Precise Cable and Antenna Measurements in the FieldUsing FieldFox Handheld Analyzers Application Note Cable 2 Open 4 meter t 9 8 meter Adapter a 4 9 8 9 8 4 9 9 distance m amh k j en 13 8 meter b distance m 23 6 meter Figure 11 Signal propagation paths between FieldFox and an open ended cable a signal reflection from the open end b double reflection from the open end and the adapter Cable type selection menu Select a cable file Then press Recall File amp Interna jserData Cables RG8U cable parameters nob Press Edit to make changes Press Done to exit G8U Coaxialcable RG Sf 0 8480 Figure 12 Cab
35. on when a good quality phase stable cable is used on site training of CAT and FieldFox to connect FieldFox to the DUT QuickCal allows users to extend the refer applications including the Line Sweep ence plane to the end of the jumper or adapter without the need of a cal kit and Antenna Test Training Course it supports many commonly used connector types H7215x 121 Appendix A Return Loss and VSWR Conversion Table Return loss can be expressed as a voltage standing wave ratio VSWR value using the following table a y VSWR sek a VSWR a i VSWR art 1 VSWR ae ui VSWR 4 0 4 42 14 0 1 50 18 0 1 29 28 0 1 08 38 0 1 03 6 0 3 01 14 2 1 48 16 5 1 27 28 5 1 08 30 9 1 02 8 0 2 32 14 4 1 47 19 0 1 25 29 0 1 07 300 1 02 10 0 1 92 14 6 1 46 19 5 1 24 29 5 1 07 395 1 02 105 1 85 14 8 1 44 20 0 122 30 0 1 07 40 0 1 02 11 0 1 78 15 0 1 43 20 5 1 21 305 1 06 40 5 1 02 11 2 1 76 15 2 1 42 21 0 1 20 31 0 1 06 41 0 1 02 11 4 1 74 15 4 1 41 pales 1 18 31 0 1 05 41 5 1 02 11 6 1 71 15 6 1 40 22 0 L 32 0 1 05 42 0 1 02 11 8 1 69 15 8 aoe 2 1 16 32 5 1 05 42 5 1 02 12 0 1 67 16 0 kag 430 115 33 0 1 05 43 0 1 01 12 2 1 65 16 2 1 37 235 1 14 500 1 04 43 5 1 01 12 4 1 63 16 4 1 36 24 0 1 13 34 0 1 04 44 0 1 01 126 1 61 16 6 1 35 24 5 1413 J45 1 04 44 5 1 01 12 8 1 59 16 8 1 34 25 0 112 50 0 1 04 45 0 1 01 13 0 1 58 17 0 1 33 23 0 1 11 50 0 1 03 45 5 1 01 Tee 1 56 TZ 1 32 26 0 1 11 36 0 1 03 46 0 1 01 13 4 1 54 17 4 1 31 26 5 1 10 30 5 1 03 46 5 1 01 13 6 1 53 17 6
36. open end and the blue trace stored to memory is the measurement using the 50 ohm load As expected the start the adapter and the cable end are measured at the same location using the open but there is a notice CAT Ref 0 00 dB gt Adapter Enter velocity factor f Select distance to fault dB Start 0 000 m Points 601 10 Keysight Techniques for Precise Cable and Antenna Measurements in the FieldUsing FieldFox Handheld Analyzers Application Note Distance To Fault cB 50 ohm Ta oY e End of k Frequency Mode Lowpass Output Power HIGH Stop 30 00 m Swp 710 0 ms Q0 f 4 fee jz f 44444 b e H aMem iti if Avg 1 Start 0 000 m Points 601 Frequency Mode Lowpass Output Power HIGH Stop 30 00 m Swp 526 0 ms Figure 10 DTF measurement comparison between cable terminated in open circuit yel low and 50 ohm load blue able increase in the amplitude of the reflected signal from the open As FieldFox is measuring the two way reflection properties from the cables under test the displayed DIF mea surements are calculated by taking one half of the actual measured distance From figure 10 it is also observed that two additional peaks ap pear to the right of the open As these measured cables are not that long these peaks are created by re reflec tions from the multiple discontinuities throughout the cable set For
37. r is defined as that of the cal kit connec tor As each calibration kit is different proper selection of the connector type and gender will ensure that FieldFox is using the correct cal kit definitions For the highest measurement accuracy the cal kit connector type should be the same as the DUT connector In this way the calibration plane is located directly at the DUT input see figure 14c The 1 port OSL cal type is the most accurate of the three available cal types but the calibration process requires three separate measurement sweeps one for each of the three cal Standards For the highest accuracy when using 1 port OSL it is import ant that the OPEN standard from the calibration kit is used instead of just leaving the cable open ended Calibrating for insertion less measurements There are several types of user calibra tions available for accurately measur ing the insertion loss of cables and systems The cal types are CalReady QuickCal Normalization Enhanced Response and Full 2 port CalReady CalReady allows users to make cable and antenna measurements at the test port without calibration if a good quality phase stable cable is used 25 to 30 dB return loss and much better than system return loss that is under test to connect FieldFox to the DUT This is the easiest method to make a cable and antenna sweep in the field CalReady for insertion loss measure ments is immediately available upon instrument pow
38. sion lines is not limited to only coaxial cable types but systems using waveguide and twisted pair cables may also be characterized once the appropriate adapter is installed between the transmission line and the coaxial interface on the analyzer In addition antenna measurements In the form of signal reflection or namely return loss and VSWR may be used to verify the performance of an antenna at the installation site When multiple antennas are required at an installation site whether the antennas are associated with the same system or different systems the antenna to antenna isolation may also be verified using FieldFox Figure 1 shows a typical return loss measurement upper trace and DTF measurement lower trace for an X band antenna connected to a 3 meter length of coaxial cable The antenna was mounted on a vertical mast at a roof top installation and measured using FieldFox configured with a preset calibration called CalReady The return loss measurement includes two frequency markers placed across the specified operating band at 8 GHz and 12 4 GHz The DIF measurement includes a distance marker placed at the location of the antenna at 3 23 meters This application note will introduce the practical aspects of cable and antenna testing interpreting measurement results and instrument operation including calibration options such as CalReady and QuickCal using FieldFox Ref 0 00 dB Start 6 000 GHz DTF Coupled Freq OFF Stop 15 50 GHz Poin
39. t DTF measurement available on FieldFox DIF uses reflection measure ments taken from the transmission line and calculates the individual amplitude response of any discontinuities as a function of distance DIF measure ments using FieldFox are similar to time domain measurements found on a vector network analyzer 4 but DIF reports distance to the discontinuity instead of time delay Figure 1 shows a dual display of the return loss as a function of frequency upper trace and the associated DIF measurement lower trace from a section of coax ial cable connected to an antenna When examining the DTF display high amplitude level signals are located at the points where discontinuities exist along the transmission line and at the load The location of these high ampli tude signals will provide an important troubleshooting tool when attempting to determined faults in the cabled transmission system Distance to fault dB Figure 8 shows a typical configuration for measuring the DTF on FieldFox In this example it is assumed that some type of the load is connected to Field Fox through two sections of coaxial cables The load can be the system antenna a 50 ohm termination or just an open ended cable In this example the two cable sections are connected together with an adapter Figure 8 also shows a table of important cable specifications as supplied by the man ufacturer The specification for velocity factor is very important
40. ts 401 f Output Power High Swp 334 ms Ref 0 00 dB Distance To Fault dB Hoar i Aiaran K Pe Start 0 000 m Frequency Mode Bandpass Stop 5 000 m Points 401 Output Power High Resolution 12 5 mm Cable Loss 0 00 dB rn Max distance 6 38 m Figure 1 Measured return loss upper trace and distance to fault lower trace for an X band antenna and coaxial feed cable 03 Keysight Techniques for Precise Cable and Antenna Measurements in the Field Using FieldFox Handheld Analyzers Application Note Why Is Cable and Antenna Testing Important During the installation operation and maintenance of a telecommunica tion system the connecting cables adapters and antennas may become damaged or show reduced perfor mance over time It is known that fifty to sixty percent of cellular base station problems are caused by faulty cables connectors and antennas Failure in these components causes poor cover age and unnecessary handovers in the cellular system In any communication system It becomes important to have a process and the appropriate instru mentation to verify the performance of the cables and antennas when issues are discovered at the system level Cabling and antennas may be installed in a variety of environments includ ing outdoor and indoor installations Outdoor installations may include roof top tower mounted and under ground cabling Indoor installations may include equipment shelters office buildings sh
Download Pdf Manuals
Related Search
5991 0419EN
Related Contents
Installer manual User Manual CC2400DK Development Kit HPRcv™ - Datamed Contato magnético RADION RM User Manual - X SMART Notebook 10.8 Amana UXT2030AYW Owner's Manual iDream ADSL2/2+ Modem Router ADSL Document d`accompagnement User Manual REALQUALITY RS Copyright © All rights reserved.
Failed to retrieve file