1. BE RETURNED TO WALLACH FOR INSPECTION AND SERVICE AT NO TIME SHOULD CHLORINE MIXTURES BLEACH OR IODINE BASED PRODUCTS BE USED FOR SOAKING OR CLEANING WALLACH CRYOTIPS CRYOTIPS WILL DEGRADE IF EXPOSED TO SUCH SOLUTIONS MOISTURE IN LINES AND ORIFICES OF THE CRYOPROBE CRYOTIP AND CRYO CONSOLE CAN CAUSE SPUTTERING POOR FREEZE AND OTHER MALFUNCTIONS WHENEVER USING FLUIDS OR STEAM TO CLEAN OR STERILIZE ALL GAS LINES AND CRYOTIP ORIFICES MUST BE PLUGGED OR CORROSION MAY RESULT PLUGS ARE AVAILABLE FROM WALLACH FOR THE CRYOTIPS OVEREXPOSURE OF TISSUE MAY CAUSE CRYOGENIC BURNS SIMILAR TO SEVERE FROSTBITE AS A RESULT OF THE RAPID EVAPORATION OF THE LIQUIFIED GAS CAUTIONS U S FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN IN THE EVENT OF A DEFROST FAILURE TURN TANK OFF LEAVE TIP IN PLACE REMOVE TIP WHEN DEFROST IS EVIDENT AND USE STERILE WATER OR SALINE TO EXPEDITE DO NOT DEPRESS TRIGGERS WHILE CHANGING TIPS IF TRIGGERS ARE DEPRESSED WHILE CHANGING TIPS INJURY MAY OCCUR DUE TO TIP SEPARATION FROM CRYOSURGICAL FREEZER WEIGH THE CYLINDER TO DETERMINE IF ENOUGH GAS IS AVAILABLE FOR THE PROCEDURE THE EMPTY OR TARE WEIGHT IS STAMPED ON THE CYLINDER VALVE OR SHOULDER SUBTRACT THIS TARE WEIGHT FROM TOTAL WEIGHT TO DETERMINE WEIGHT OF GAS IN CYLINDER SEE INSTRUCTIONS ON VENTING EXCESS WEIGHT CYLINDER SHOULD CONTAIN A MINIMUM OF 1 36 KILOS 3 LBS OF GAS IN ORDER TO BEGIN THE PROCEDURE IMCRY
2. LA BOMBONA A UN LUGAR BIEN VENTILADO PARA DEJAR SALIR EL N20 UNA CONCENTRACI N EXCESIVA DE XIDO NITROSO PUEDE SER PELIGROSA PARA LA SALUD ES ESENCIAL REALIZAR UNA INSPECCI N VISUAL PERI DICA DE LA UNIDAD Y DE LAS PUNTAS DE CRIOCIRUG A INTERRUMPA EL USO DE CUALQUIER PRODUCTO QUE PAREZCA DETERIORADO EL CUAL DEBERA DEVOLVERSE A WALLACH SURGICAL DEVICES PARA SU INSPECCI N Y REPARACI N EN NING N CASO DEBEN USARSE MEZCLAS DE CLORO LEJ A NI PRODUCTOS A BASE DE YODO PARA LIMPIAR O DEJAR EN REMOJO LAS PUNTAS DE CRIOCIRUG A WALLACH YA QUE SE DEGRADAR AN EN CONTACTO CON DICHAS SOLUCIONES LA HUMEDAD EN LOS CONDUCTOS Y ORIFICIOS DE LA SONDA LAS PUNTAS Y LA CONSOLA DE CRIOCIRUG A PUEDEN PROVOCAR CHISPORROTEO MALA CONGELACI N Y OTROS FALLOS DE FUNCIONAMIENTO SIEMPRE QUE SE UTILICEN L QUIDOS O VAPOR PARA LA LIMPIEZA O ESTERILIZACI N DEBER N TAPARSE TODOS LOS CONDUCTOS DEL GAS Y ORIFICIOS DE LAS PUNTAS DE CRIOCIRUG A LOS TAPONES PARA LAS CRIOPUNTAS PUEDEN SOLICITARSE A WALLACH LA SOBREEXPOSICI N DE LOS TEJIDO PUEDE CAUSAR QUEMADURAS CRIOG NICAS SIMILARES A UNA CONGELACI N GRAVE COMO RESULTADO DE LA EVAPORACI N R PIDA DEL GAS LICUADO ADVERTENCIAS LAS LEYES FEDERALES ESTADOUNIDENSES PERMITEN S LO LA VENTA DE ESTE DISPOSITIVO A M DICOS O POR PRESCRIPCI N FACULTATIVA EN CASO DE FALLO EN LA DESCONGELACI N CIERRE LA BOMBONA DEJE LA PUNTA DE CRIOCIRUG A EN SU SITIO Y RET RELA UNA VEZ COMPROBADA LA DESCONGELACI N O U
3. 008 Rev G 5 13 1 Wallach Surgical Devices INSTRUCTIONS 1 2 3 4 5 6 7 8 9 IMPORTANT SAFETY FEATURE Wallach has incorporated a rupture disk into the LL100 This is a safety device that prevents damage to the instrument from overheated cylinders Note If tank is overheated and pressure reaches 1500 psi gas will exhaust through the rupture disk instead of pressurizing the whole unit thereby avoiding damage to the unit The rupture disk is easily replaced by removing it with a 14 2 mm wrench Additional rupture disks may be purchased from Wallach Surgical Devices The rupture disk is not covered under the Remove protective sleeve which covers the tip of the instrument Save the sleeve for protection if you need to transport the instrument at any time in the future Secure the pressure gauge block to the cylinder The hand wheel should be hand tightened only Open gas cylinder valve and observe gas pressure Gauge needle should move to the green area Note Green area denotes proper pressure for operation Yellow area denotes that cylinder pressure is too low for operation and cylinder needs to be changed Red area denotes excess pressure Vent tank as described in section titled HOW TO VENT CYLINDER Choose appropriate sterilized cryotip with clear shield and install Hand tighten only Do not use tools to tighten To see if apparatus is working properly it should be tested prior to use Pointing to th
4. TILICE AGUA CALIENTE PARA ACELERAR EL PROCESO NO APRIETE LOS GATILLOS MIENTRAS CAMBIA LAS CRIOPUNTAS PUES EN CASO DE MANTENERSE APRETADOS PODR A PRODUCIRSE UNA LESI N SI LA PUNTA SE SEPARA DE LA PISTOLA DE CRIOCIRUG A PESE LA BOMBONA PARA DETERMINAR SI HA Y SUFICIENTE GAS DISPONIBLE PARA EL PROCEDIMENTO EL PESO EN VAC O O TARA APARECE IMPRESO EN LA V LVULA DE LA BOMBONA O EN LA PARTE SUPERIOR RESTE LA TARA DEL PESO TOTAL PARA DETERMINAR EL PESO DEL GAS QUE QUEDA EN LA BOMBONA LEA LAS INSTRUCCIONES PARA ELIMINAR EL EXCESO DE PESO LA BOMBONA DEBE CONTENER UN M NIMO DE 1 36 KILOS 3 LBS DE GAS PAPA PODER INICIAR EL PROCEDIMIENTO IMCRY008 Rev G 5 13 5 Wallach Surgical Devices INSTRUCCIONES 1 Retire la funda protectora que cubre el extremo del instrumento y cons rvela para protegerlo cuando lo tenga que trasladar 2 Asegure el bloque del man metro en la bombona La rueda manual debe apretarse a mano solamente 3 Abra la v lvula de la bombona de gas y observe la presi n La aguja del man metro debe moverse hacia la zona verde Nota La zona verde indica que la presi n es correcta para el funcionamiento La zona amarilla indica que la presi n de la bombona es insuficiente y que por lo tanto ser necesario cambiarla La zona roja indica un exceso de presi n Vac e la bombona como se describe en la secci n titulada C MO VACIAR LA BOMBONA 4 Escoja una criopunta esterilizada apropiada con un protector t
5. TOLA DE CRIOCIRURG A DE DOBLE GATILLO CON N O LL100 REF 900001 900019 70 900019 72 900020 900629 900630 Instrucciones de uso Espanol Spanish C ooss MA LEA TODA LA INFORMACI N DE SEGURIDAD E INSTRUCCIONES ANTES DE USAR ESTE PRODUCTO DESCRIPCI N Y USO DEL PRODUCTO La LL100 es una unidad o pistola de criocirug a utilizada para t cnicas quir rgicas de tipo ablativo La popularidad de este procedimiento se atribuye a varios factores su pS aplicabilidad a diversos sistemas org nicos una amplia gama de indicaciones de enfermedad p ej tejido premaligno o con afectaci n v rica la rapidez en los procedimientos su coste relativamente bajo su facilidad de uso en el consultorio y sus m nimas complicaciones operatorias y postoperatorias La referencia 900001 es la principal LL100 para N O La referencia 900019 70 es la LL100 con estribo de seguridad brit nico La referencia 900019 72 es la LL100 con una conexi n de rosca La referencia 900020 es la Pistola de congelaci n con N O de tija corta de LL100 La referencia 900629 es la consola con unidad de congelaci n de varias puntas con termopar Y la referencia 900630 es la consola con unidad de congelaci n de varias puntas AVISOS NO UTILICE LAS BOMBONAS DEMASIADO LLENAS YA QUE LA PRESI N DEL GAS PODR A SER EXCESIVA Y SUPONER UN PELIGRO LAS BOMBONAS DE 9 KILOS 19 8 LBS NO DEBEN CONTENER M S DE 9 KILOS 19 8 LBS DE N 0 L QUIDO LLEVE
6. WAILACH SURGICAL DEVICES LL100 Two Trigger Cryosurgical Unit for N20 REF 900001 900019 70 900019 72 900020 900629 900630 Directions for Use English C oose MA READ ALL SAFETY INFORMATION AND INSTRUCTIONS BEFORE USING THIS PRODUCT PRODUCT DESCRIPTION USAGE The LL100 is a cryosurgical unit used for ablative type surgical technique The popularity of this procedure can be attributed to several factors applicability to multiple organ systems a wide range of disease indications i e viral and premalignant tissue rapid procedural times relatively low cost office based suitability and minimal operative and post operative complications The 900001 is the main LL100 for NO The 900019 70 is the LL100 with the pinned yoke British The 900019 72 is the LL100 with a threaded connector The 900020 is the LL100 short shaft Cryosurgical Freezer N20 The 900629 is the console freezer multi tip with thermocouple The 900630 is the console freezer multi tip WARNINGS DO NOT USE OVERFILLED CYLINDER GAS PRESSURE MAY BE EXCESSIVE AND PRESENT A HAZARD 9 KILO 19 8 LB CYLINDERS SHOULD NOT CONTAIN OVER 9 KILOS 19 8 LBS OF LIQUID N O ONLY MOVE CYLINDER TO A WELL VENTILATED AREA WHEN VENTING N O NITROUS OXIDE IN EXCESSIVE CONCENTRATION MAY BE HAZARDOUS TO HEALTH PERIODIC VISUAL INSPECTION OF CRYO INSTRUMENT AND TIPS IS ESSENTIAL DISCONTINUE USE OF ANY PRODUCT SUSPECTED OF BEING DAMAGED SUCH PRODUCT SHOULD
7. completely off Detach instrument from gas cylinder Call Wallach Surgical Devices at 203 799 2000 and ask the Service Department for assistance If a repair is needed carefully sanitize and package the unit in a protective carton Include a note stating the problem and ask for a repair estimate Do not attempt in house repairs as they will void your warranty WARRANTY The LL100 system carries a five year warranty One year unconditional guarantee plus additional four year warranty agreement against manufacturer s defects on parts and labor WALLACH SURGICAL DEVICES 95 Corporate Drive Trumbull CT 06611 USA Phone 800 243 2463 203 799 2000 Fax 203 799 2002 customerservice wallachsurgical com www wallachsurgical com EXPLANATION OF SYMBOLS Reorder Number For Professional Use Only ATTENTION See instructions for use I Over pressure Relief Device Serial Number sims le Air Flow Gas Exhaust single use rupture disk Port F Activate Defrost Activate Freeze R Z Product contorns tothe Authorized European Representative C oose6 Medical Device Directive EC REP Leisegang Feinmechanik GmbH 93 42 EEC LeibnizstraBe 32 D 10625 Berlin GERMANY re Latex Free LL100 and Wallach are trademarks of Wallach Surgical Devices 2013 Wallach Surgical Devices IMCRY008 Rev G 5 13 4 Wallach Surgical Devices WAILACH SURGICAL DEVICES UNIDAD O PIS
8. e floor depress freeze trigger for several seconds Tip should frost and smoke Release freeze trigger and depress defrost trigger Frost should disappear within 3 to 4 seconds NOTE The freeze trigger will automatically remain in the depressed position for your comfort Depressing the defrost trigger will release the freeze trigger If you wish to release the freeze trigger without defrosting momentarily depress and release the defrost trigger Apply cryotip to moistened treatment area It is recommended that a lubricant be applied to the treatment site Depress freeze trigger to freeze tissue Depress and hold down defrost trigger to defrost the tip Remove instrument from treatment area When finished with treatment set instrument in holder and shut off gas cylinder valve NOTE If treatment requires more than one tip 1t is not necessary to shut off the gas cylinder valve while changing tips Simply unscrew the tip presently in use and replace with tip desired To reduce concentration of N20 in the immediate cryosurgical area you may vent the exhausted gas by attaching one end of the scavenger hose to the exhaust port at bottom of the pressure gauge block The other end of the hose goes to a discharge location Hose must have minimum of 6 mm internal diameter We recommend Wallach Scavenger Hose Part 901015 approximately 6 meters warranty HOW TO VENT CYLINDER CAUTION MOVE CYLINDER TO A WELL VENTILATED AREA A Turn gas cy
9. e them as soon as possible with fully immersible probes INSTRUCTIONS FOR CLEANING STERILIZABLE CRYOTIPS 900200AA 900217AA 900300AA 900306AA 1 Remove disposable shield from cryotip and discard 2 Plug cryotip before cleaning and sterilizing Plug Orifice does not require sterilization 3 Scrub tip to remove blood and particles 4 Rinse with water 5 Cryotips can be steam sterilized by Gravity or Pre vacuum using the following procedures Stage Process Parameter Set Points Set up Cycle Profile Gravity Pre vacuum PTEE ponu de Chamber Purge 1 minute 1 minute Conditioning Sterilization Temperature 132 C 270 F 132 C 270 F Exposure Steam Injection Pare f Sterilization Time 15 minutes 4 minutes Post Exposure i Dry T 1 t 20 t Exhaust Dry ry Time 5 minutes minutes 6 Attach new disposable plastic shield to the cryotip ensuring that the tabs on the shields are correctly fitted into the grooves on the metal tip 7 Cryotip is now ready to use WARNING DO NOT USE ETO E BEAM OR GAMMA STERILIZATION Rutala William A Ph D M P H David J Weber M D M P H and the Healthcare Infection Control Practices Advisory Committee HICPAC CDC Guideline for Disinfection and Sterilization in Healthcare Facilities pg 19 2008 IMCRY008 Rev G e 5 13 3 Wallach Surgical Devices SERVICE AND REPAIR In case of malfunction immediately turn gas cylinder master valve
10. linder master valve completely off B Slowly loosen hand wheel nut on attachment to gas cylinder This releases residual gas in the feeder hose When hissing sound stops remove pressure gauge block C Point cylinder opening away from you and slowly open the cylinder master valve just enough to hear gas steadily released After approximately 15 seconds close the valve D If cylinder is being vented to reduce overfilled condition re weigh cylinder before proceeding Reconnect pressure gauge block open gas cylinder valve and observe gas pressure Gauge needle should point to the green area Repeat the procedure if it still points to the red area HOW TO STORE LL100 BETWEEN PROCEDURES A B C Close cylinder master valve fully It is not necessary to disconnect pressure gauge and tips from cylinder between procedures Depressurize the unit by engaging the defrost trigger until pressure gauge drops to 0 IMCRY008 Rev G 5 13 2 Wallach Surgical Devices IMPORTANT REMINDERS A Keep unit including cylinder at room temperature 68 C 74 F 18 C 22 C Cold will decrease the system s efficiency and significantly increase defrost time Heat causes excess gas consumption Allow time for temperature recovery if extremes have been experienced B Never store or use cylinders near radiators furnaces or other heat supplies C Do not mix or use gas other than N O D Never siphon tanks of liquid N20 NOTE For complete informa
11. mente la llave principal hasta que empiece a o r c mo sale el gas de manera constante Tras unos 15 segundos cierre la llave D Si est vaciando la bombona porque esta demasiado llena vuelva a pesarla antes de seguir adelante Conecte de nuevo el bloque del man metro abra la v lvula de la bombona y compruebe la presi n del gas La aguja del man metro debe estar se alando hacia la zona verde Repita la operaci n si sigue se alando hacia la zona roja C MO ALMACENAR LA LL100 ENTRE PROCEDIMIENTOS A Cierre por completo la v lvula principal de la bombona B Noes necesario desconectar el man metro y las criopuntas de la bombona entre procedimientos C Despresurice la unidad apretando el gatillo de descongelaci n hasta que man metro baje a 0 ADVERTENCIAS IMPORTANTES A Mantenga la unidad incluida la bombona a temperatura ambiente 18 C 22 C 68 F 74 F ya que el fr o reducir la eficacia del sistema y aumentar considerablemente el tiempo de descongelaci n mientras que el calor provoca un consumo de gas excesivo En caso de alcanzar temperaturas extremas deje pasar un tiempo para que se recupere la temperatura ambiente B Nunca almacene o use las bombonas cerca de radiadores calderas u otras fuentes de calor C No mezcle ni use gases que no sean N30 D No utilice nunca el sistema de sif n para vaciar bombonas de N O l quido NOTA Para una completa informaci n sobre el manejo de las bombonas de gas c
12. nta met lica 7 La criosonda ya est lista para utilizarse PRECAUCI N NO ESTERILIZAR CON XIDO DE ETILENO NI CON IRRADIACI N DE ELECTRONES NI CON RAYOS GAMMA SERVICIO Y REPARACI N En caso de mal funcionamiento cierre OFF de inmediato la v lvula principal de la bombona de gas Desconecte el instrumento de la bombona de gas Llame a Wallach Surgical Devices al 203 799 2000 y solicite la ayuda del Servicio t cnico Si se precisa reparaci n desinfecte cuidadosamente la unidad y emb lela en una caja de cart n Incluya una nota mencionando el problema y pida un presupuesto de reparaci n No intente hacer ninguna reparaci n por su cuenta ya que sta anular a su garant a GARANT A El sistema LL 100 tiene cinco a os de garant a Un a o de garant a incondicional m s un contrato de garant a adicional de cuatro a os cubriendo defectos de fabricaci n y gastos de reparaci n Rutala William A Ph D M P H David J Weber M D M P H and the Healthcare Infection Control Practices Advisory Committee HICPAC CDC Guideline for Disinfection and Sterilization in Healthcare Facilities pg 19 2008 IMCRY008 Rev G e 5 13 7 Wallach Surgical Devices WAILACH SURGICAL DEVICES 95 Corporate Drive Trumbull CT 06611 USA Tel fono 800 243 2463 203 799 2000 Fax 203 799 2002 customerservice wallachsurgical com www wallachsurgical com EXPLICACI N DE LOS S MBOLOS Referencia para nuev
13. omprimido consulte la Compressed Gas Association www cganet com email cga cganet com IMCRY008 Rev G 5 13 6 Wallach Surgical Devices DONDE ABASTECERSE DE GAS PARA GARANTIZAR UN RENDIMIENTO OPTIMO SOLAMENTE DEBERA UTILIZARSE GAS DE CALIDAD MEDICA Y SIN CONTAMINANTES El gas N20 de calidad m dica est disponible en todo el mundo Consultar la gu a telef nica local en cualquiera de los siguientes apartados Suministros para soldadura Gas medicinal Gas carb nico Suministro y Equipos de distribuci n de bebidas con gas Muchos de estos proveedores sirven a los consultorios INSTRUCCIONES DE LIMPIEZA DE LA LL100 La unidad de criocirug a LL100 de Wallach puede limpiarse con un pa o empapado en un desinfectante apropiado Este procedimiento de limpieza puede aplicarse tambi n al mango de pl stico negro a las mangueras blancas de silicona al man metro y al conector de la bombona de gas No se recomiendan otros m todos que impliquen el uso de calor l quidos remojo en fr o o vapor RECOMENDACIONES PARA LA DESINFECCI N DE LA SONDA DE AISLAMIENTO SEG N LA GU A DE LOS CDC PARA LA DESINFECCI N Y ESTERILIZACI N EN CENTROS SANITARIOS Algunas sondas de criocirug a no son totalmente sumergibles Durante el reprocesado la punta de la sonda debe sumergirse en un desinfectante de alto nivel durante el tiempo que sea apropiado cualquier otra parte de la sonda que pueda entrar en contacto con una membrana mucosa se puede desinfec
14. os pedidos 0 Uso exclusivo profesional N mero de serie A ATENCI N Consulte las instrucciones de uso mire lt 23 Flujo de aire Puerto de salida del gas Activar descongelaci n NA Dispositivo de despresurizaci n Disco de ruptura de un solo uso Activar congelaci n C oose Producto conforme a la Directiva 93 42 CEE de Productos Sanitarios Representante europeo autorizado Leisegang Feinmechanik GmbH Leibnizstrafe 32 D 10625 Berlin GERMANY Es No contiene l tex LL100 y Wallach son marcas comerciales de Wallach Surgical Devices O 2013 Wallach Surgical Devices IMCRY008 Rev G e 5 13 Wallach Surgical Devices
15. ransparente y m ntelo Apri tela a mano sin usar herramientas 5 Para comprobar que la unidad funciona correctamente deber probarla antes de utilizada Para ello apriete el gatillo de congelaci n durante varios segundos apuntando hacia el suelo La criopunta se helar y soltar vapor Suelte el gatillo de congelaci n y apriete el gatillo de descongelaci n La escarcha desparecer en 3 4 segundos NOTA El gatillo de congelaci n permanecer autom ticamente apretado para su comodidad Al pulsar el gatillo de descongelaci n el gatillo de congelaci n se desbloquear Si desea soltar el gatillo de congelaci n sin descongelar apriete y suelte moment neamente el gatillo de descongelaci n 6 Aplique la criopunta al rea de tratamiento humedecida Se recomienda aplicar un producto lubricante al rea de tratamiento Apriete el gatillo de congelaci n para congelar el tejido 7 Apriete el gatillo de descongelaci n y mant ngalo apretado para descongelar la punta Aleje el instrumento del rea de tratamiento 8 Una vez terminado el tratamiento coloque el instrumento en su base y cierre la v lvula de la bombona de gas NOTA si el tratamiento requiere m s de una punta de crioterapia no es necesario cerrar la v lvula de la bombona mientras cambia de punta Limitese a destornillar la criopunta en uso y c mbiela por la punta adecuada 9 Para reducir la concentraci n de N20 en el rea crioquir rgica inmediata puede desp
16. resurizar el gas de salida conectando un extremo de la manguera de vaciado al orificio de salida del gas situado en la parte inferior del bloque de man metro El otro extremo de la manguera deber dirigirse a una zona de vaciado La manguera debe tener un di metro interior m nimo de 6 mm Se recomienda la manguera de vaciado Wallach N de pieza 901015 de aproximadamente 6 metros de longitud DISPOSITIVO DE SEGURIDAD IMPORTANTE Wallach ha incorporado un disco de ruptura a la unidad LL100 Este dispositivo de seguridad impide que el instrumento sufra da os a causa del sobrecalentamiento de las bombonas Nota Si la bombona se sobrecalienta y la presi n alcanza 1500 psi el gas se liberar a trav s del disco de ruptura en lugar de ejercer presi n sobre toda la unidad evit ndose as da os al equipo El disco de ruptura puede cambiarse f cilmente con una llave inglesa de 14 2 mm Pueden solicitarse discos de ruptura adicionales a Wallach Surgical Devices El disco de ruptura no est cubierto por la garant a C MO VACIAR LA BOMBONA ADVERTENCIA TRASLADE EL CILINDRO A UN LUGAR BIEN VENTILADO A Cierre por completo la v lvula principal de la bombona de gas OFF B Afloje lentamente la tuerca de la rueda manual de la bombona para expulsar el gas residual en la manguera de alimentaci n Cuando cese el silbido desconecte el bloque del man metro C Mantenga alejada de su cuerpo la abertura de la bombona y abra lenta
17. tar por inmersi n o envolvi ndola con un pa o empapado de un desinfectante de alto nivel para asegurar el periodo de contacto recomendado Tras la desinfecci n la sonda debe enjuagarse con agua corriente y secarse antes de volver a usarla Los centros sanitarios que utilicen sondas no sumergibles deben sustituirlas lo antes posible por sondas totalmente sumergibles INSTRUCCIONES DE LIMPIEZA DE LAS PUNTAS DE CRIOCIRUG A ESTERILIZABLES 900200AA 900217AA 900300AA 900306A A 1 Retire el protector desechable de la punta y t relo 2 Tape la punta antes de limpiarla y esterilizarla El orificio del tap n no necesita esterilizarse 3 Frote la punta para eliminar la sangre y las part culas 4 Enjuague con agua 5 Las criosondas pueden esterilizarse con vapor por sistema de Prevac o o Gravedad mediante los siguientes procedimientos Fase Proceso Par metro de ajuste Configuraci n Perfil del ciclo Gravedad Prevac o Preexposici n C mara de purgado 1 minuto 1 minuto Preparaci n a Temperatura de 132 C 270 F 132 C 270 F Exposici n esterilizaci n I i n d ae f Diea sane Tiempo de esterilizaci n 15 minutos 4 minutos cis ida Tiempo de secado 15 minutos 20 minutos Expulsi n secado 6 Coloque un nuevo protector de pl stico desechable en la punta de criocirug a asegur ndose de que las lengiietas de los protectores se insertan correctamente en las ranuras de la pu
18. tion on handling compressed gas cylinders refer to Compressed Gas Association www cganet com email cga cganet com WHERE TO REFILL GAS TO ENSURE OPTIMAL PERFORMANCE ONLY MEDICAL GRADE GAS FREE OF CONTAMINANTS SHOULD BE USED Medical grade NO gas is available throughout the world See your local telephone listing under any of these headings Welding Supplies Gas Medical Carbonic Gas Soda Dispensing Equipment and Supplies Many of these suppliers deliver to the office INSTRUCTIONS FOR CLEANING LL100 The Wallach LL100 Cryosurgical unit can be wiped down with an appropriate disinfectant This wipe down procedure can include the black plastic handle white silicone gas hoses gauge block and gas cylinder connector Other methods involving heat fluid cold soaking or steam are not recommended RECOMMENDATION FOR DISINFECTION OF INSULATION PROBE BASED ON CDC GUIDELINE FOR DISINFECTION AND STERILIZATION IN HEALTH CARE FACILITIES Some cryosurgical probes are not fully immersible During reprocessing the tip of the probe should be immersed in a high level disinfectant for the appropriate time any other portion of the probe that could have mucous membrane contact can be disinfected by immersion or by wrapping with a cloth soaked in a high level disinfectant to allow the recommended contact time After disinfection the probe should be rinsed with tap water and dried before use Health care facilities that use nonimmersible probes should replac