Laboratory practice of high performance liquid chromatography

image

Contents

1. 71 Pelagia Research Library Ufere Emele N and Ajanwobu O Innocent Der Pharmacia Sinica 2015 6 5 75 80 2 Flush with tetrahydrofuran 2 5 5 Protein size exclusion media There are two wash regeneration procedures associated with the removal of contaminants from protein size exclusion media Weakly retained proteins Flush with 30 mL 0 1M pH 3 phosphate buffer Strongly retained proteins Flush for 60 minutes using 100 water to 100 acetonitrile gradient 2 5 6 Porous graphitic carbon There are four wash or regeneration procedures associated with porous graphitic carbon The one s used will depend on the analytes and solvents that have been used with the column 2 5 7 Acid Base Cleaning Suitable for ionized species analyzed in strongly aqueous mobile phases 1 Invert the column Flush at 1 mL min with 50 mL tetrahydrofuran water 1 1 containing 0 1 trifluoroacetic acid Flush at 1 mL min with 50 mL tetrahydrofuran water 1 1 containing 0 1 triethylamine or sodium hydroxide Flush at 1 mL min with 50 mL tetrahydrofuran water 1 1 containing 0 1 trifluoroacetic acid Flush with methanol water 95 5 to re equilibrate Re invert the column ee ae 2 5 8 Strong Organic Cleaning Suitable for applications involving polar and or ionized species analyzed in aqueous mobile phases 1 Flush at 1 mL min with 50 mL acetone 2 Flush at 1 mL min with 120 mL dibutylether 3 Flush at 1 mL min with 50 mL acetone 4 Flush with a
2. Perform a ten fold dilution of the column efficiency test solution Inject 5 uL of this diluted mix Adjust the detector sensitivity until the peak height is approximately 75 of the full scale readout Measure the peak width at 4 4 peak height 5 sigma column efficiency method Convert the peak width to mL using the following conversion and spread uL peak width x 1 20 x 1 x 1000 DANK Where peak width is measured at 4 4 peak height and expressed in cm 1 20 represents the chart speed min cm 1 represents the flow rate mL min and 1000 represents the volume correction factor uL min RFERENCES 1 Jerry Demenna workshop in spectroscopy and chromatography Buck Scientific Incorporated USA pp 9 27 127 2 Gurdeep R Chatwal Instrumental methods of chemical analysis 5 ed Mumbai Himalaya publishing house 2002 pp 2 566 2 568 2 640 3 Chmwiki eucdavis edu api deki pages 2928 pdf 4 www fda gov ucm 5 www hplctraining blogspot com 6 User manual for Liquid chromatography buck scientific incorporated BLC Series pp 172 186 7 http www kobisi si material HPLCTroubleshooting pdf 8 www kobis s material hplctroubleshooting 9 http www knauer net fileadmin user_upload product files document applica tion_notes si material HPLC_troubleshooting_guide pdf 10 www sigma aldrich com 11 http www pharmatutor org 12 www unine ch files content sites s 80 Pelagia Research Library
3. Available online at www pelagiaresearchlibrary com Pelagia Research Library Der Pharmacia Sinica 2015 6 5 75 80 PP ed Research SEIE Der Pharmacia Sinica Library ISSN 0976 8688 CODEN USA PSHIBD Laboratory practice of high performance liquid chromatography HPLC Ufere Emele N and Ajanwobu O Innocent Department of Pharmaceutical and Medicinal Chemistry Faculty of Pharmaceutical Sciences Nnamdi Azikiwe University Awka ABSTRACT This compilation is intended to guide for efficient and smooth operation of HPLC system especially those that are problems free Its primary aims are to maximize system operation time and to promote good chromatographic practices It has also been extended to cover some aspect of troubleshooting Wherever possible diagrams or schematics are used for clarity This enables the guide to be used quickly and efficiently by operators with varying experience Correct solvent preparation is very important It can save vast amounts of time spend troubleshooting spurious peaks baseline noise etc INTRODUCTION High performance liquid chromatography HPLC is a form of liquid chromatography to separate compounds that are dissolved in solution HPLC instruments consist of a reservoir of mobile phase a pump an injector a separation column and a detector Compounds are separated by injecting a plug of the sample mixture onto the column The different components in the mixture pass through the column at di
4. fferent rates due to differences in their partitioning behavior between the mobile liquid phase and the stationary phase All reagents and solvents should be of the highest quality HPLC grade reagents may cost slightly more than lower grade reagents but the difference in purity is marked HPLC grade reagents contain no impurities to produce spurious peaks in a chromatogram baseline Water used in buffer preparation is of highest purity demonized water often contains trace levels of organic compounds therefore is not recommended for HPLC use Ultra pure HPLC water 18M_ resistivity is generated by passing demonized water through an ion exchange bed Modern water purification instruments use this mechanism to produce water of suitable quality in high volumes Alternately HPLC grade water can be purchased from solvent suppliers All buffers should be prepared freshly on the day required This practice ensures that the buffer pH is unaffected by prolonged storage and that there is no microbial growth present Changes in pH and microbial growth will affect chromatography If buffer solutions are stored be aware that they have a finite lifetime This can be found in pharmacopeia or monographs on buffer shelf life Buffer reagents can contain a stabilizing agent chemicals that inhibit reaction of other chemicals eg Sodium metabisulphite These stabilizing agents often affect the optical and chromatographic behaviour of buffer solutions so it is w
5. l that is compatible with any solvents that may be flushed through it 78 Pelagia Research Library Ufere Emele N and Ajanwobu O Innocent Der Pharmacia Sinica 2015 6 5 75 80 2 7 Cutting tubing Tubing should always be cut using instruments that are designed for the purpose The use of scissors and wire cutters is not recommended as they can deform the tubing bore and almost always produce angled cuts that are one of the biggest causes of leaks and increased dead volume Note Where possible pre cut tubing should always be used 2 7 1 Stainless Steel Stainless steel can be effectively cut using the procedure outlined below 1 Estimate the length required Remember to allow extra length if the tubing is to go around corners as sharp bends in the tubing will distort the inside bore and hence the solvent flow through it 2 Using a stainless steel cutter score cleanly round the tubing If a cutter is unavailable use a knife file to score the tubing Care should be taken whichever method is used not to distort or damage the tubing 3 Using two pairs of smooth jawed pliers one above the score and one below it gently bend the tubing back and forth until it snaps Excessive bending should be avoided as this will damage the tubing and not give a clean break 4 There will be one or two burrs present on the cut surface These can be carefully filed away to give a smooth flat cut surface Care should be taken not to allow any filed material
6. orth buying reagents that contain no stabilizer Containers of solid reagent are easily contaminated by repeated use For this reason we recommend that regents be purchased in low container weights 75 Pelagia Research Library Ufere Emele N and Ajanwobu O Innocent Der Pharmacia Sinica 2015 6 5 75 80 MATERIALS AND METHODS 2 1 Mobile Phase Properties Do not use highly acidic or basic solvents unless your HPLC system and column have been engineered to accommodate them Seals plungers etc can be damaged by extreme pH conditions If in doubt please contact your column or instrument supplier before using an aggressive solvent The use of highly aqueous mobile phases is becoming more popular as safety guidelines demand less exposure to organic solvents Care should again be taken that the HPLC column has been engineered to accommodate highly aqueous solvents Highly aqueous mobile phases are ideal breeding grounds for microbes Ensure that an organic solvent is flushed through the HPLC system and column at least once every 48 hours to kill unwanted microbial growth 2 2 Solvent Use Each solvent line should be fitted with an inlet filter This is the first line of system defense against particulate contamination from solvents The filters should be kept clean to prevent cross contamination When they are not being used it is recommended that they are stored in a solution of 50 acetonitrile 50 water This will inhibit microbial growth and
7. queous mobile phase until equilibrated 2 5 9 Normal Phase Cleaning Suitable for applications running predominantly in normal phase mobile phases 1 Flush at 1 mL min with 50 mL dichloromethane Flush at 1 mL min with 50 mL methanol Flush at 1 mL min with 50 mL water Flush at 1 mL min with 50 mL 0 1 M hydrochloric acid Flush at 1 mL min with 50 mL water Flush at 1 mL min with 50 mL methanol Flush at 1 mL min with 50 mL dichloromethane Flush with mobile phase until equilibrated ot Se ee ee 2 5 10 Removal of Trifluoroacetic acid Suitable for applications running mobile phases containing trifluoroacetic acid Flush the column with acetonitrile that has been heated to 75 C the column should also be maintained at this temperature 2 6 System plumbing and fittings The purpose of a well plumbed HPLC system is to minimize dead volume between its components and to eliminate leaks System tubing errors show themselves in many ways for example as band broadening baseline noise etc detection of incorrect diameter tubing is often very difficult The internal diameter of tubing used in a HPLC system varies with the position in the instrument refer to your system maintenance manuals The type of tubing used is determined by the application that is being performed The two most common types of tubing are steel and PEEK although others are also available when changing tubing make sure that the replacement is manufactured from a materia
8. stop dust and dirt from embedding in the filter pores The solvent lines should be clean growth free and should have no sharp bends or creases in them Solvent reservoirs should be placed as high as possible on or in the instrument always higher than the pump inlet manifold The solvent lines and filters should be of sufficient length to reach the bottom of the solvent reservoir Note Before starting an analysis calculate the total volume of mobile phase required Prepare all the phases at the same time and place in reservoirs that are large enough to accommodate them Insufficient mobile phase may cause the system to pump dry This is undesirable because it will fill the system with air If this does happen purge the solvent lines and pump with fresh solvent then allow the system to pump solvent until it is equilibrated 2 3 Changing Solvents Ensure that the buffer is soluble in the proposed wash or storage phase If it is not first flush the system with a solvent mix that is highly aqueous to remove the buffer from the system and column and then change to the proposed wash or storage solvent mixed There are few occasions where these solvent changes will be necessary Such occasions include columns that can be used with both solvent types or systems where both normal and reversed phase analyses are performed Where possible systems should be dedicated to normal or reversed phase chromatography To convert a normal phase system column
9. th system instability spurious baseline peaks etc on 4 J Nm O Pa 2 i 4 5 a a u m Time Minutes ime Minules Mobile Phase Not Degassed Mobile Phase Degassed Fig 2 1 Degassed and non Degassed mobile phase The most efficient form of degassing is bubbling with helium or another low solubility gas If this method is available I recommend that the mobile phase is continually degassed at very low levels throughout the analysis This will inhibit the re adsorption of gases over the analysis time 2 6 Column Cleaning In all instances the volume of solvent used is 40 60 column volumes unless otherwise stated Ensure that no buffers samples are present on the column and that the solvent used prior to the clean up is miscible with the first wash solvent After the clean up ensure that the test mobile phase is compatible with the last solvent in the column 2 5 1 Normal phase media 1 Flush with tetra hydro furan 2 Flush with methanol 3 Flush with methylene chloride 4 Flush with benzene free n hexane 2 5 2 Reversed phase media 1 Flush with HPLC grade water inject 200 uL DMSO during this flush 2 Flush with methanol 3 Flush with chloroform 4 Flush with methanol 2 6 3 Anion exchange media 1 Flush with HPLC grade water 2 Flush with methanol 3 Flush with chloroform 2 5 4 Cation exchange media 1 Flush with HPLC grade water inject 200 uL DMSO during this flush
10. to a reversed phase system column flush with a solvent that is miscible with both the current normal phase solvents and ideally the proposed reversed phase solvents If the final reversed phase solvents include a buffer then it is advisable to move from the 100 methanol flush to a 50 aqueous methanol flush For example Normal Phase Hexane Ethy1 Acetate Flush IPA then Methanol Finally 50 50 Methanol Water Reversed Phase Buffered Aqueous Methanol Flush 50 50 Methanol Water Normal Phase Hexane Ethyl1 Acetate 2 4 Filtration Ideally all HPLC solvents should be filtered through a 0 45 um filter before use This removes any particulate matter that may cause blockages After filtration the solvents should be stored in a covered reservoir to prevent contamination with dust etc Filtering HPLC solvents will benefit both your chromatography and the wear and tear of the HPLC system Pump plungers seals and check valves will perform better and lifetimes will be maximized 76 Pelagia Research Library Ufere Emele N and Ajanwobu O Innocent Der Pharmacia Sinica 2015 6 5 75 80 2 5 Degassing Before the freshly prepared mobile phase is pumped around the HPLC system it should be thoroughly degassed to remove all dissolved gasses Three methods of degassing are listed below e Bubbling with helium e Sonication e Vacuum filtration If the mobile phase is not degassed air bubbles can form in the high pressure system resulting in problems wi
11. to block the tubing bore 2 7 2 PEEK or Polymeric Tubing The simplest way to cut polymeric tubing is using a blade for example a razor blade or craft knife The polymeric tubing cutters that are supplied by many manufacturers are essentially a razor blade in safety housing The advantage of using one of these cutters is that they hold the tubing at 90 to the blade so a clean flat cut is assured Polymeric tubing can be effectively cut using the procedure outlined below 1 Estimate the length required Remember to allow extra length if the tubing is to go around corners as sharp bends in the tubing will distort the inside bore and hence the solvent flow through it 2 Use a sharp blade or specialist cutter to cut the tubing Do not use a sawing action This will give an uneven cut surface Make the cut in a single action 3 Inspect the cut surface for burrs These can be carefully filed away to give a smooth flat cut surface Care should be taken not to allow any filed material to block the tubing bore As a final action after cutting the tubing and before connecting it to the HPLC flush it with solvent to remove any filed material dust or other debris that could be in the tubing bore The tubing is now ready to have fittings attached to it 2 7 Fittings Fittings are available in stainless steel and a range of polymeric compounds In general you should use stainless steel fittings with stainless steel tubing PEEK fittings with PEEK
12. tubing etc I recommend that compression fittings and ferrules from different manufacturers are never interchanged The dimensions of the fitting and ferrule vary between suppliers so mixing them often leads to leaks stripped threads and damage to female ports such as column inlet and outlets detector inlets etc The amount of tubing that extends past the ferrule also varies with manufacturer If the tubing does not seat properly within the female fitting then leaks and increases in dead volume will occur Laboratories that contain instruments from more than one manufacturer often find it useful to purchase universal fittings These fittings are compatible with all instruments RESULTS 3 1 Band spreading or Band broadening Broad peaks often accompanied by a change in retention time indicate band spreading It can occur within the HPLC column but is more often due to incorrect system plumbing The following procedure describes a method for measuring band spreading due to the HPLC system Column effects can be measured using efficiency calculations 1 Remove the HPLC column from the system and replace with a zero dead volume union 2 Configure the HPLC system with the following parameters Flow rate 1 mL min Detector sensitivity 0 5 to 1 0 AUFS Detector time constant 0 2 or less Chart speed if required 20 cm min 79 Pelagia Research Library Ufere Emele N and Ajanwobu O Innocent Der Pharmacia Sinica 2015 6 5 75 80

Download Pdf Manuals

image

Related Search

Related Contents

  feuillet technique eco2 - Laboratoire de biologie médicale du Quai  IC50/IC46 用チップリセッター YXD968 取扱説明書  Groundwater Protection Code: Septic Tanks  PDFを閲覧する  Serien DPO2000 und MSO2000 Oszilloskope Benutzerhandbuch  Notice d`installation - La Boutique du Sauna  intext:Installationsanleitung filetype:pdf  Samsung LE19A656A1D Наръчник за потребителя  BA 10 FAG Motion Guard CONCEPT6 Sistema Automático de  