PSM-1000 Microscope user`s manual
Contents
1. Adjust the center of the standard point on work piece to the center of the right field of view The second highest objective can be set onto the work piece by turning the revolver until hearing a click sound Adjust the two centering screws of the second highest objective to the position where the standard point on the work piece is centered with centering keys centering key Migri centering sc ey ee objective lt profocality adjustable AS highest objective tae nonadjustable Similarly parcenter the other objectives in order of highest magnification 3 10 Change over tube Lens The tube lens can be changed over by turning the changeover turret Change over to 1X NIR VIS 1X UV VIS or 2X VIS lens by turning the changeover turret of tube lenses Check the kind of lens based on the mark on turret The turret is marked as follows 3 SETUP CONTINUED AA MENS PE ETEA E EEEE 1X Ultraviolet Visible cce cece ea ees 1X Near Infrared Visible 3 11 Changing Objectives Change the objective by turning revolver Note Don t touch the objective face by hand It may cause damage to the objective face 1000 3 12 Change Optical Path Pull or push the optical path changeover knob to change the optical path of the microscope Pull outward to enhance the laser operation by pulling the beam splitter out of the field of view Push inward to change over to the optic
2. 3 screws as shown below and remove the binocular tube and shim block meee shim block 3 screws e 3 SETUP CONTINUED Insert the safety filter the 2 rings and the binocular tube then hold and tighten the 3 screws as mentioned in the previous step 3 6 Mount Objective Insert the objective into the flange of revolver MLL MULL parfocality adjustable Objective Objective 4i5 X 1 36 thread Note 1 The color sticker near the flange and the color line on the objective is the same color 2 If you purchased an objective which has a thread 4 5 X 1 36 insert the optional Objective Adapter for 4 5 X 1 36 thread into revolver first and then insert the objective into the Objective Adapter 3 Ifthe objective is parfocality adjustable take care not to loosen the lock ring 3 SETUP CONTINUED Repeat this procedure for the other objective lens 3 7 Adjusting the eyepiece diopter Adjust the diopter of the eyepiece as follows Step1 Turn the diopter adjustment ring of the eyepiece until the lower edge of the eyepiece tube is aligned with the 0 index line white Align both eyepieces p as spa A Sm 4 Step2 Prepare a sample and set the illumination Set the revolving nosepiece to a high magnification 20X or 50X and focus the sample with the focusing knobs Set the magnification of the tube lens to 1X Step3 Turn the revolving nosepiece to cha
3. Long working ELWD5X SG01S162C4 1 N A 0 14 Working Distance 34 O distance Plan APO ELWD10X SG01S022C4 1 N A 0 28 Working Distance 33 5 O objectives parfocality ELWD20X_ SG01S032C4 1 N A 0 42 Working Distance 20 O adjustable ELWD50X SG01S122C4 1 N A 0 55 Working Distance 13 O ELWD100X SG01S062C4 1 N A 0 70 Working Distance 6 O Ultra Long working ULWD50X SG01S122C7 1 IN A 0 42 Working Distance 20 5 O distance Plan APO objectives parfocality JULWD100X SG01S062C7 1 IN A 0 55 Working Distance 13 O adjustable 1X Camera Adapter SP100386 1 C mount can be attached 2 3 CCD O 0 5X Camera Adapter SP100363 1 C mount can be attached 2 3 CCD O 0 4X Camera Adapter SP100365 1 C mount can be attached 1 2 CCD O 0 3X Camera Adapter SP100367 1 C mount can not be attached 1 3 CCD O Analyzer SWO01991Z 1 O Polarizer SW01992B 1 O Centering keys SW0199BH 2 Shims SW018880 2 J For the laser safety filter Screw M3X10 SS0120594 3 J Safety pin SW010013 1 For the laser adapter J Objective Adapter SW011840 1 4 For 4 5 X1 36 thread O Inspection report SP050002 1 Lock Screw M5X10 SS0120225 4 J Allen key 4mm SP070008 1 J User s manual SP010205 1 Warranty card SP160003 1 J Standard O Optional 1 OVERVIEW CONTINUED 1 3 Nomenclature 1 Eyepiece 10 Tube Lens Change over Turret 2 Binocular Tube 11 Interpupillary Scale 3 Trinocular Head 12 Focus Block 4 Body 13 F
4. performance Don t insert anything into open spaces This may cause the microscope to perform poorly and or invalidate your warranty Laser safety and use optimization If you plan to use your PSM 1000 for laser work you need a laser safety filter To operate the laser safely and properly please see the laser and or laser filter manufacturer s documentation Motic Instruments assumes no responsibility whatsoever for the performance and or safety of the laser and or laser filter used with the PSM 1000 microscope The PSM 1000 microscope is optimized for 355nm 532nm and 1064nm wavelength lasers WARRANTY In the event that the Microscope PSM1000 series should prove defective in workmanship or material within 5 years from date of purchase it will be repaired or replaced at our option free of charge upon its return to us prepaid The following cases will void the warranty 1 2 Defects caused by improper usage or invalid modification Defects caused by moving of installation fall and transfer after original purchase Defects caused by disaster earthquake lighting and other force majeure cases Case of no warranty sheet presented upon return No proof of date of purchase the name of customer and or improper rewriting of the required information Changing of the consumer parts with non registered parts 1 OVERVIEW 1 1 Outline Innovative ergonomic design leading edge optical systems and a host of unique advanced
5. PRECAUTIONS User Manual To obtain the best possible performance and maximum service life from the Motic Microscope PSM1000 hereinafter referred to as the PSM1000 please read this manual thoroughly This manual describes all of the components available for the PSM1000 system including optional components available for purchase Avoid impacts Do not allow any part of the microscope unit to be subjected to stress or blows during installation and operation Carrying the microscope unit Carefully carry the microscope unit by supporting its base Do not touch any moving parts on the microscope unit Installation Place the microscope unit on a sturdy and level desk where it will not be subjected to dust direct sunlight high temperature high humidity or excessive vibration Keep the lenses clean Dirty optical parts reduce image clarity Always keep the lenses mirrors filters and the bulb free from stains fingerprints and dust If the objectives and or ocular lenses become dirty wipe softly in a circular motion from the center towards the outside using lens paper moistened with alcohol Do not apply excessive force to the focusing knobs Do not over rotate the focusing knobs located at side of the main body Do not disassemble Do not attempt to disassemble the microscope unit unless absolutely necessary for replacement of specified consumable parts Disassembly of this microscope unit may result in damage or deterioration in
6. ain the sharpest image Ifthe position of the main body is upper adjust the objective as follows A anticlockwis ath trini Step1 Remove the objective from the revolver inib ee 4 Step2 Grasp the objective body and loosen the lock ring tighten es dey icosen a el Step3 Turn the adjustment ring anticlockwise and tighten the lock ring SE SUTA Step4 Insert the objective back into the flange of revolver inl S PLAN A 20x10 42 anes ae Ifthe position of the main body is lower adjust the objective as follows clockwise A I Step1 Remove the objective from the revolver adjustment oY SS 1 Step2 Grasp the objective body and loosen the lock ring 0 d MWH j Step3 Turn the adjustment ring clockwise and tighten the lock ring f Step4 Insert the objective back into the flange of revolver Repeat these procedures until the second highest objective is parfocal with the highest objective 3 SETUP CONTINUED Similarly adjust the focal point of objective in the order of higher magnification highest objective rau m adjustment ring nonadjustable X y D profocality adjustable 3 9 Parcentration of objectives The highest objective nonadjustable is not required to parcenter Insert a eyepiece with reticle into the right tube or left tube The highest objective nonadjustable can be set onto the work piece by turning the revolver until hearing a click sound
7. al path for visible viewing through the eyepieces 3 SETUP CONTINUED 3 13 Control Light Brightness Rotate the Aperture Open and Close Dial to control the light brightness Rotate towards you The aperture s diameter narrows and darkens Rotate away from you The aperture s diameter widens and lightens Aperture Open and Close Dial 3 14 Polarization Insert the polarizer and analyzer all the way into the mounting slots The analyzer has an adjuster to rotate the plane of polarization by 360 3 15 Remove and mount nosepiece The nosepiece of PSM 1000 is removable Remove and mount the nosepiece as follows With the provided Allen wrench loosen the 4 locking screws and remove the nosepiece 4 locked screws two sides lt 14 3 SETUP CONTINUED Mount the nosepiece and tighten the 4 locking screws with the Allen wrench nosepiece 4 locked screws two sides TROUBLE SHOOTING Phenomenon Causes Measures Visual field is insufficient or invisible Revolver cannot turn to the position where the click sound can be heard Make the revolver turn to the position where the click sound can be heard Image lens changeover turret cannot turn to the position where the click sound can be heard Make the image lens changeover turret turn to the position where the chick sound can be heard Quantity of light of illuminator is insufficient Control the qu
8. antity of light through aperture knob Eyepiece is not mounted Mount the eyepiece Objective lens is not mounted Mount the objective lens Resolution or image quality is considerably bad Eyepiece is dirty Objective lens is dirty Observational objective is dirty Clean the dirty area as instructed in the Precautions section Quantity of light is too much or too little Control the quantity of light through aperture knob Not focused Adjust the focus Observing through any medium of glass Etc Remove the medium One side of image is unclear Observational object is inclined Correct the inclination between microscope and object Objective lens is not screwed in completely Mount the objective lens into the revolver correctly In observation through binocular microscope right and left visual images do not match Interpupillary distance of siedentopf is not matched to observer s eye Adjust the interpupillary distance In observation eyes feel tired Interpupillary distance of Adjust the interpupillary siedentopf is not matched to distance observer s eye Diopter movement of Adjust the diopter eyepieces is not matched to observer s eye Illumination is too bright Control the quantity of light through aperture stop knob
9. features make this Motic metallurgical industrial microscope ideally suited to the needs of manufacturing and research amp development in the fields of metallurgy semiconductors and electronics 1 The PSM 1000 provides a 3 lens turret with 1X NIR tube lens 1X NUV tube lens and 2X visible These lenses cover wavelengths from 355nm to 1064nm and are designed around the new range of Wave Research Lasers A laser safety pin is provided 2 The optical system has adopted a combination of ultra long working distance objectives which are capable of independent correction of lateral chromatic aberration and eyepieces with an ultra wide field of view 3 The quadruple forward facing spring loaded nosepiece with adjustments located on the side along with the turret can be easily centered using the tools provided The turret is placed on a dovetail mount for easy access and removal The turret comes with colored stickers to indicate the position of objectives 4 The body incorporates polarizing filter slots for easy adaptation 5 Various peripherals optional including CCD or CMOS camera adapter Polarizer and Analyzer are available to broaden applications 1 2 System Configuration The PSM 1000 Microscope unit consists of a main body objectives a pair of eyepieces standard accessories and other optional peripheral units Parts list of microscope PSM 1000 Part Name Part No Qty Note IWF10X 24 SG02S0142 2 Field
10. ilter Slot in Light Source Side Attaches 5 Nosepiece an optional filter polarizer 6 Objective 14 Optical Path Change over Knob 7 Filter Slot in Observation Side Attaches 15 Aperture Open and Shut Dial an optional filter analyzer 16 Fiber Insertion Unit 8 Coarse Adjustment Handle 17 CCD Camera Adapter mount 9 Fine Adjustment Handle 2 Check the accessories Before Operation Before operating check that all accessories have been included If any accessory is missing please advise our department immediately regarding the missing item 3 SETUP 3 1 Mount eyepiece When you first use this microscope after unpacking remove the protective cap of the ocular upper tube and mount the eyepiece The eyepieces use a high eye point system and have rubber eye cups attached If you wear eyeglasses turn the rims outward 3 2 Adjust the pupil distance of the eyepiece While looking into the eyepieces and holding the binocular tubes with both bands move the tubes in the direction of the arrows see the diagram below until the two fields of view coincide The pupil adjustment range is between 55 and 75mm or 2 17 and 2 95 inches Interpupillary distance scale 3 SETUP CONTINUED 3 3 Set up Illumination Loosen up the lock screw and remove the protective cap of the fiber insertion unit protective cap Fiber insertion Unit Trinocular Head After inserting the 0 272 or 6 9mm outer diameter illuminati
11. ng fiber into the fiber insertion unit tighten it with the lock screw Illuminating Fiber Se gt He 0 272 or 069MM Fiber Insertion Unit i 3 4 Mount CCD Camera Unit You can select camera adapter for your CCD camera as below 0 5X camera adapter use for 2 3 CCD camera 0 4X camera adapter use for 1 2 CCD camera 0 3X camera adapter use for 1 3 CCD camera C mount adapter can not be Attached Attach a CCD camera to the 0 5X camera adapter option e CCD Camera 0 5X Camera Adapter Option 3 SETUP CONTINUED Loosen up the lock screw and remove the protective cap of the CCD camera adapter mount Attach the assembled camera unit to the CCD camera adapter mount on trinocular head and tighten it with the lock screws a assembled camera unit VE gok screw mss i CCD Camera Adapter mount AND j j M Trinocular Head 3 5 Mount Laser Unit If you plan on using your PSM 1000 for laser work you need a laser safety filter To operate the laser safely and properly please see the laser and or laser filter manufacturer s documentation Motic Instruments assumes no responsibility whatsoever for the performance and or safety of the laser and or laser filter used with the PSM 1000 microscope To install the safety filter and the laser safety pin we have provided you with the shims and a laser safety pin please follow the instructions below With a screwdriver loosen the
12. nge the lens magnification from a high magnification 20X or 50X to a lower magnification 5X or 2X Turn the diopter adjustment ring to obtain the sharpest image for each eye without moving the focusing knobs Step4 Turn the diopter adjustment ring a Clockwise if near sighted b Counterclockwise if far sighted The diopter has a range of 5D to 5D Repeat this procedure twice for even sharper images 3 8 Parfocality of Objectives Be sure to parfocus the objective lens after mounting the objective onto the microscope completely Parfocality adjustable objective only The highest objective nonadjustable can be set onto the work piece by turning the revolver until hearing a click sound 3 SETUP CONTINUED Turn the coarse adjustment handle and move the main body of the microscope to either an upper or lower position to obtain the image coarsely Turn the fine adjustment handle and move the main body of the microscope to either an upper or lower position to obtain the sharpest image Fine Adjustment Scale Tumper scale The second highest objective lens can be set onto the work piece by turning the revolver until hearing a click sound Turn the coarse adjustment handle and move the main body of the microscope to either an upper or lower position to obtain the image coarsely Turn the fine adjustment handle and move the main body of microscope to either an upper or lower position to obt
13. number 624 with eyeguard o Eyepiece WF15X 17 2 Field number 16 O IWEF20X 13 2 Field number 12 O Binocular Siedentopf infinity Erect image Interpupillary Distance 55 75mm Binocular Tube SP100243 1 J Optical Path 50 50 simultaneous observation or 100 0 Optical path system Turret change over system Trinocular Head SW0199BD 1 e 2X for visible 1X for visible and NIR 1X NUV Body SW0199BE 1 Quadruple Nosepiece SW011715 1 Parcenterable for each hole J Stroke 50mm Operation Coaxial system Coarse 4mm per rotation Fine Focus Block SW0199BG 1 e 0 1 per rotation Mountable Weight 45Kg 1 OVERVIEW CONTINUED Part Name Part No Qty Note IELWD2X_ SG01S102C3 1 N A 0 055 Working Distance 34 O ELWD5X SG01S162C3 1 IN A 0 14 Working Distance 34 O Extra Long working ELWD10X SG01S022C3 1 N A 0 28 Working Distance 33 5 O distance Plan APO objectives ELWD20X SG01S032C3 1 N A 0 42 Working Distance 20 O ELWD50X SG01S122C3 1 N A 0 55 Working Distance 13 O ELWD100X SG01S062C3 1 N A 0 70 Working Distance 6 O Ultra Long working ULWD50X SG01S122C5 1 N A 0 42 Working Distance 20 5 O distance Plan APO N A 0 55 Working Distance 13 objectives ULWD100X SG01S062C5 1 O ELWD2X SG01S102C4 1 N A 0 055 Working Distance 34 O Extra
Download Pdf Manuals
Related Search
Related Contents
Samsung GT-E1202 Керівництво користувача Foaming Penetrant & Cutting Oil Aceite Espumoso para Cortar y Leica M 9 Airless Warranties assembly machines - Manuli Hydraulics Operating instructions EXZ1200 - Support SHIMADZU APPLICATION NEWS Dell PowerConnect W-Airwave 7.6 Installation Manual Shuttle™ ULT:25 -86˚C Portable Ultra-Low Copyright © All rights reserved.