Home
RIGOL User`s Guide M300 Series Data Acquisition/Switch System
Contents
1. N Lad Figure 1 1 Front View Unit mm Ca re 373 4 gt 6ST ww IN Figure 1 2 Side View Unit mm M300 User s Guide 1 3 RIGOL Chapter 1 Quick Guide Front Panel Data Acquisition Switch System LXI E Gaass AN AA 10 Figure 1 3 Front Panel of M300 1 LCD 4 3 inches high resolution color LCD displaying the menu configuration guide measurement parameters system status and prompt messages 2 Function dd N Configure the scan list and the measurement parameters of each channel Create edit read and save the scan list Set the number of scans trigger mode and scan interval Provide the channel configuration guide used to configure the measurement parameters scaling parameters alarm paramete
2. Chapter 6 Specifications Module Specifications MC3120 MC3132 MC3164 MC3324 MC3416 MC3648 Multiplexer Actuator Matrix General MC3120 MC3132 MC3164 MC3324 MC3416 MC3648 Number of Channels 20 32 64 20 Voltage 4 Current 16 4x8 2 4 wire 2 4 wire 1 wire 2 4 wire SPDT 2 wire Connect to DMM Yes Yes Yes Yes No No Module Scanning Speed 60Ch s 60Ch s 60Ch s 60Ch s Open Close Speed 200Ch s 200Ch s 200Ch s 200Ch s 200Ch s 200Ch s Maximum Input Voltage DC AC rms 300Vrms 300Vrms 300Vrms 300Vrms 300Vrms 300Vrms Current DC AC rms lArms lArms lArms lArms 2Arms lArms Power W VA 50VA 50VA 50VA 50VA 60VA 50VA Isolation ch ch ch earth DC AC 300Vrms 300Vrms 300Vrms 300Vrms 300Vrms 300Vrms rms DC Characteristics Offset Voltage 5uV 5uV 5uV 5uV lt 3uV 5uV Initial Closed 19 19 19 19 0 10 19 Channel Resistance ne gt 10GQ gt 10GQ gt 1060 gt 1060 gt 1060 gt 1060 AC Characteristics Bandwidth 1MHz 1MHz 1MHz 1MHz 1MHz 1MHz Ch Ch Cross Talk dB P 45 45 18 9l 45 15 18 1MHz Capacitance HI LO 100pF 100pF 100pF 100pF lt 500pF 100pF Capacitance 200pF 200pF 200pF 200pF lt 200pF 200pF LO Earth M300 User s Guide Chapter 6 Specifications RIGOL Volt Hertz Limit 108 10 105 10 105 105 Other T C Cold J unction 0 8 C 0 8 C 0 81 0 8 C _ Accuracy Typical uten e 10 100M 100M 100M 100M 100M 100M
3. Function Range Frequency Range 24 Hour 90 Day 1 Year Temperature Coefficient Tcu C 1 C Tea C 5 C Tcu C 5 C OC to Tea C 5 C Tea C 5 C to 50 C Frequency 200mV to 3 Hz 5 Hz 0 07 0 07 0 07 0 005 Period 300V 5 Hz 10 Hz 0 04 0 04 0 04 0 005 10 Hz 40 Hz 0 02 0 02 0 02 0 001 40 Hz 300 kHz 0 005 0 006 0 007 0 001 300 kHz 1 MHz 0 005 0 006 0 007 0 001 Additional Low Frequency Errors of reading Frequency Gate Time Resolution 1s 0 1 ppm 0 1 s 1 ppm 0 01 s 10 ppm 0 001 s 100 ppm 3 Hz 5Hz 0 0 12 0 12 0 12 5 Hz 10Hz 0 0 17 0 17 0 17 10 Hz 40Hz 0 0 20 0 20 0 20 40 Hz 100Hz 0 0 06 0 21 0 21 100 Hz 300Hz 0 0 03 0 21 0 21 300 Hz 1 kHz 0 0 01 0 07 0 07 gt 1kHz 0 0 0 02 0 02 1 Specifications are for 90 minutes warm up and 1 s gate time 2 For frequency lt 300 kHz the specification is for AC input voltage of 10 to 110 of range For frequency gt 300 kHz the specification is for AC input voltage of 20 to 110 of range The maximum input is limited to 750 Vrms or 8 x 107 Volts Hz whichever is less 200 mV range is full range input or input that is larger than the full range For 20 mV to 200 mV inputs multiply of reading error by 10 3 Relative to calibration standards 6 6 M300 User s Guide Chapter 6 Specifications RIGOL Measuring Characteristics Frequency and Period Measurement Method Reciprocal counting technique AC coupled input using the AC v
4. 7 RS 232 Alarms Ext Trig Mixed Interface Male This interface is 25 pin male interface You can convert this interface into two 9 pin interfaces using the Mixed Interface Convert Cable accessory This breakout cable consists of a 9 pin male interface used as a standard RS232 interface and another a 9 pin female interface used for alarm output external trigger signal input etc RS232 via this interface M300 can communicate with the PC and you can control the instrument remotely via the PC Pin Name Definition 1 s 1 6 2 RXD Received Data 2 3 TXD Transmitted Data 3 4 DTR Data Terminal Ready 8 5 SGND Signal Ground 3 9 6 DSR Data Set Ready 5 7 RTS Request To Send 8 CTS Clear To Send RS232 Interface gt e Alarms Ext Trig as shown in the figure below pin 1 to pin 4 are used to output TTL pulse when an alarm is generated in the corresponding alarm channel Pin 6 is used to receive external trigger signal When an external digital multimeter is connected for measurement pin 5 and pin 6 are used together to control the measurement Pin Definition 1 6 1 Alarm 1 Output 2 2 Alarm 2 Output 3 7 3 Alarm 3 Output 8 4 Alarm 4 Output 4 5 Channel Closed Output 5 9 6 Ext Trig Input Channel Advance Input 7 GND 8 Not Used Alarms Ext Trig Interface 9 Not Used Note The Pin 5 outputs a TTL negative pulse with about 7us pulse width each time the measurement is completed The output rate is dependent upon the actual m
5. isses erre fce edente eds 1 17 Alarm Information 2 72 Integration Times 2 24 Alarm EN 2 18 Intervalle 2 3 Alarm Mode nenn 2 18 P Address ae 2 90 Alarm Trigger eessss 2 11 LAN Parameters ssss 2 89 Alarm Value 2 18 L tch een 2 67 all channel monitor 2 46 lock the front panel 1 6 ALPHA a ee tis 2 31 MAC erede een 2 91 Analog Bus Interface 1 8 Manual Trigger 2 8 Any sensor measurement 2 33 ManuallP seeesss 2 90 Aperture Time 2 27 MEIN ee 1 18 arithmetic CUFVe 2 33 MC3120 en 1 18 Xd nie age tataewat 2 90 MC31322 er 1 18 Auto Trigger ans 2 8 MC31064 en 1 18 RUTO ZEO ae 2 25 MC3324 a 1 18 Bank Joined nccc 2 50 MBIT ee 1 19 Bank Separate 2 50 MC3534 EHE 1 19 BriOMtnesS sere een 2 85 MC3 64G e eect ne nennen nenn 1 19 built in help system 1 20 Measurement Configuration 2 74 Channel Configuration 2 12 Measurement Connections 1 23 Channel Copy 2 41 Measurement Data 2 75 Channel delay 2 27 Menu Quick Navigation 1 26 Channel Number 2 13 Mirror Configuration 2 75 Cleaning ie exe IX module control 2 47 current channels 2 56 Module copy ccecce 2 40
6. 4 Advanced Configuration include integration time AC filter etc Press Extended to enter the extended copy interface e Switch switch the cursor between the Source Channel area and Target Module area OK copy the configuration of the source channel selected to the target module selected Return return to the upper level menu 1 Local arget Module Switch Return Figure 2 11 Extended Copy Steps of Extended Copy 1 After entering the extended copy interface the Source Channel area is selected by default Use the direction keys to select the desired source channel 2 Press Switch to move the cursor to the Target Module area and use the up down direction keys to select the desired target module 3 Press OK to execute extended copy Note For extended copy only copies among the same type of modules are allowed This function makes it easy to configure several channels for the same measurement M300 User s Guide 2 43 RIGOL Chapter 2 Front Panel Operations Channel Monitor The channel monitor function of M300 allows users to monitor a single channel multiple channels or all of the channels in the scan list The instrument monitors the channel and takes readings from the channel monitored continuously even during a scan This function is helpful for troubleshooting your system before a test or for watching an important signal The channel monitor function can monitor the following channels mu
7. 4 TrigMode When the channel status is READ use the up down direction keys to select TrigMode and use the left right direction keys to select the desired mode e Rising configure the totalizer to count on the rising edge of the input signal when the input signal passes through the trigger threshold e Falling configure the totalizer to count on the falling edge of the input signal when the input signal passes through the trigger threshold Note The TrigMode here refers to the mode in which to trigger the counting of the totalizer Please pay attention to distinguish it from the TrigMode of the M300 User s Guide 2 37 RIGOL Chapter 2 Front Panel Operations instrument Threshold For TOT channels S07 and S08 after selecting the trigger mode use the up down direction keys to select Threshold and use the numeric keyboard to input the desired value between 12 V and 12 V Note This function is only applicable to TOT channels S07 and S08 For TOT channels S05 and S06 the threshold is fixed at CMOS 3 3 V The terminal block of the multifunction module provides G and G terminals which are used to control the trigger of the TOT channels ATTL high level applied to the G terminal and a TTL low level applied to the G terminal enables counting Only providing a TTL high level applied to the G terminal enables counting Only provide a TTL low level applied to the G terminal enables counting When a
8. MUX 20 20 Channel Multiplexer and its model is MC3120 MUX 32 32 Channel Multiplexer and its model is MC3132 MUX 64 64 Channel Single Ended Multiplexer 64 and its model is MC3164 MUX MIX 24 24 Channel Multiplexer and its model is MC3324 MFC Multifunction Module and its model is MC3534 Function Display the function DCV ACV DCI ACI 2WR 4WR FREQ PERIOD TEMP SENSOR DIN or TOT of the channel currently monitored Range Display the range of the channel currently monitored This is only applicable to multiplexer channels Scaling Value Display the scaling value of the channel currently monitored only when the current channel uses the scaling function Measurement Value Display the measurement value of the channel currently monitored For TOT channel the count value is displayed If the current channel does not use the scaling function the scaling value is not displayed the measurement value is displayed at the center of the screen in relatively larger font size Alarm Channel Alarm Value Alarm State Display the alarm information of the channel currently monitored including the alarm channel alarm value displayed only when alarm is generated and alarm state PASS LO FAIL or HI FAIL For DIO channels the alarm state can be PASS EQU FAIL or NOT EQU FAIL If the channel monitored does not use the alarm function ALARM NO CONFIGURE will be displayed System Time Display the system time M300 User s Gu
9. O Jj No I Figure 2 25 MC3416 Schematic Diagram The control interface of MC3416 is as shown in the figure below Use the up down and left right direction keys to select the desired channel The channel currently selected is highlighted for example You can perform various operations on the channel currently selected such as disconnect and close the channel Card Reset Return Figure 2 26 MC3416 Control Interface M300 User s Guide 2 59 RIGOL Chapter 2 Front Panel Operations 1 Open Disconnect the channel currently selected The channel number of the channel currently disconnected is displayed in gray for example H Pressing loK can also open the channel currently selected 2 Close Close the channel currently selected The channel number of the channel currently closed is displayed in green for example 102 Pressing loK can also close the channel currently selected 3 Card Reset Initialize the status of each channel of the current module 2 60 M300 User s Guide Chapter 2 Front Panel Operations To Control MC3534 RIGOL MC3534 is a multifunction module MFC for short This module combines four 8 bit digital input output DIO channels four 32 bit totalizer TOT channels and four analog output DAC channels ais Input Signals Terminal Block M3TB34 jM O a 78 Pin Interface O O p M Bit 0 Port 1 LSB Channel 01 Bit 7 TTT Bit 0 Port 4
10. RIGOL TECHNOLOGIES INC Communicate test Right click the resource name M300 USBO 0x1AB1 0x0C80 M300123123123 INSTR and select SCPI Panel Control to open the remote command control panel through which users can send commands and read data 4 8 M300 User s Guide Chapter 4 Remote Control RIGOL To Control M300 through RS232 1 Connect the devices 4 Convert the RS 232 Alarms Ext Trig interface at the rear panel to two 9 pin interfaces using the Mixed Interface Convert Cable accessory Wherein one is a 9 pin male interface used as a standard RS232 interface Connect this interface with the computer using RS232 cable Set the RS232 parameters Set proper parameters according to To Set RS232 Parameters Search device resource Start up Ultra Sigma and ji to open the setting panel as shown in the figure below Set the matched paremeters r n must be selected in End Mark and click TEST If the test succeeds the instrument resources searched will be displayed in the right resource list If not please process according to the prompt message BS232 amp GPIB Setting R5232 Setting GPIB Setting ET m Serial Resource ASRL1 INSTR ASRLI INSTR Baud Rate Parity nen W Read 2 times 9600 v Hone bd Data Bits Stop Bit e m a ad TEST Flow Control End Mark View the resource Click OK to return the main interface of Ultra Sigma The resources that have been se
11. or Track 2 66 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Latch In this mode the corresponding pin is latched to the status high level or low level specified in Output when the alarm channel generates an alarm until you clear it by initiating a new scan or cycling power You can also manually clear pressing Clear the output of this pin at any time even during a scan and the alarm data is not cleared The alarm data is cleared when you initiate a new scan e Track In this mode the corresponding pin jumps to the status high level or low level specified in Output when a reading of the channel crosses a limit and remains outside the limit When a reading returns to within the limits the output of this pin is cleared automatically You can also manually clear pressing Clear this pin at any time even during a scan and the alarm data is not cleared The output pin and alarm data are both cleared when you initiate a new scan 2 Ouput Set the output status of the corresponding pin when an alarm is generated Use the up down direction keys to select Output and use the left right direction keys to select High or Low e Low when an alarm is generated the corresponding pin outputs a TTL low level 0 V e High when an alarm is generated the corresponding pin outputs a TTL high level 3 3 V 3 Channel Display the numbers of channels that select this alarm channel By default all the channe
12. 1 Print Use the up down direction keys to select Print and use the left right direction keys to select ON or OFF to enable or disable the measurement data output function of the serial port After connecting the instrument and the computer via RS232 and enable the print function the instrument will output the measurement readings automatically through RS232 interface when the instrument is measuring At this point you can capture the readings using serial port data acquisition tool Parity Use the up down direction keys to select Parity and use the left right direction keys to set the parity to None 8 data bits Odd 7 data bits or Even 7 data bits Please select a parity that matches the PC or DTE When you set the parity you are also indirectly setting the number of data bits The current setting is stored in the non volatile memory Flow Control Use the up down direction keys to select Flow and use the left right direction keys to select NONE XON XOFF DTR DSR RTS CTS or MODEM Please select a flow control mode that matches the PC or DTE e NONE In this mode data is sent and received over the RS232 interface without any flow control used When this mode is selected please select relatively slower baud rate lower than 9600 Baud and please do not send data that is too long longer than 128 characters and does not contain end bit e XON XOFF This mode is software flow control mode This mode uses s
13. 1 Up to 7 channels can be monitored at the same time 2 The monitor interface displays the list of the channels monitored including the channel number function range measurement value and alarm state 3 Add or delete the channel monitored All channel Monitor 1 Monitor all the channels in the scan list at the same time 2 The monitor interface displays the list of the channels monitored including the channel number function range measurement value and alarm state View Switch Ke Scan History Information Read the measurement data stored in the internal or View the information of the last external memory of the scan including the start time scan sweep count channel number and the corresponding internal or external memory of the instrument Channel View all the measurement Data values with time stamps of the channel selected Quickly locate the specified channel Note function as well as the maximum minimum average and standard deviation of the measurement alues The channel status table displays the status of each channel in real time After an alarm is generated the channel is displayed in red and will be displayed in green again if the next Channel Status Table measurement value is judged as PASS 1 Yellow channels that are not configured to the scan list 2 Green channels that are configured to the scan list 3 Red the channel gen
14. 4WR and FREQ sensors With the any sensor measurement function you can easily convert the measured quantity pressure flow temperature etc to quantity that is easy to measure voltage resistance etc After inputting the response curve the instrument converts and amends the data according to the internal arithmetic to obtain the physical quantity measured Users can edit and modify the display unit of the measured physical quantity at will The parameters for the any sensor measurement function are as shown in Table 2 1 Wherein the integration time and AC filter are introduced in the preceding section and this section introduces the configuration methods of the other parameters 1 Select the sensor type Enter the first step of the channel configuration guide select the desired channel number and function SENSOR use the up down direction keys to select Type and use the left right direction keys to select the desired type DCV DCI 2WR 4WR or FREQ 2 Settheunit Use the up down direction keys to select Unit and use the left right direction keys to select the desired unit C F QorK You can also define a unit Select User and press OK to enter the filename input interface Input the desired unit and press OK The length of user defined unit cannot exceed 3 characters 3 Configure the arithmetic curve Use the up down direction keys to select Scaling and Ax 2 Bx C is displayed by default Press OK to open th
15. 600 ms 20 Hz medium 300 ms 200 Hz fast 100 ms e Delay Use the numeric keyboard to enter the desired delay The range is from 0 to 60 s and the resolution is 1 ms M300 User s Guide Chapter 2 Front Panel Operations RIGOL Temperature Measurement Parameters The parameters to be set in temperature measurement are as shown in Table 2 1 Wherein for the integration time offset compensation and channel delay refer to the introduction in the preceding section This section introduces the configuration methods of the other parameters 1 Select the temperature sensor type Enter the first step of the channel configuration guide select the desired channel number and function TEMP use the up down direction keys to select Type and use the left right direction keys to select the desired sensor type e TC thermocouple THER thermistor RTD thermal resistance 2 wire connection mode RTD 4W thermal resistance 4 wire connection mode not applicable to single ended multiplexer 2 To select the unit Use the up down direction keys to select Unit and the left right direction keys to select the desired unit The M300 can report temperature measurements in C Celsius F Fahrenheit and K Kelvins The temperature conversion relations are as shown below F 9 5 x C 32 K x C 273 15 Different channels on the same module can use different temperature units 3 Use Thermocouple to Measure Temperature The thermocou
16. Bonn 2 63 Module Indicators 1 7 Date oi rie ah 2 78 multi channel monitor 2 46 DC osse NER Yep hes 2 14 Offset Compensation 2 26 D citmal 5 rip pus 2 83 Outside view drawing of the module DHCP Leaves d cbe Rmo ZOO re Ur 1 16 Digital Input 2 35 PERIQD iacet iret rt mrt oig 2 14 DIO sce te ren dut 2 14 Power Switch ccecce 2 84 Error Information 2 72 Power off Status 2 81 M300 User s Guide 1 RIGOL Index Power on Settings 2 81 T C Check cren 2 30 Preset Settings 2 81 TG racine range 2 29 BU LS AED eed aho t 2 31 TEMP riis dee 2 14 Bande eO 2 23 terminal blocks 1 21 Reference Source 2 30 THER scies een eere 2 29 RS 232 Alarms Ext Trig Mixed Mu rc T 2 78 lhterfacE areenan k 1 10 To Create a New List 2 2 RTD RETE 2 29 To Edit the LISt css eret 2 6 RTD AW a 2 29 To Read the List 2 7 Scaling Configuration 2 15 To Save the List 2 7 Screen SaVer c 2 83 TOT d ite a a 2 14 Select the temperature sensor type Totallzer ne t ente 2 37 mE 2 29 Track u a ana 2207 SENSOR u PR tr 2 14 Trigger Edge Type 2 5 sensor TYPE rina an tere nenn 2 33 Trigger Mode oor rene ees 2 3 Separator 2 84 trigger Modes 2 8 single channel monitor 2 44 unlock the front panel 1 6 Slots i ord ent 1 8 User Interfac
17. Load Typical Switch Life Rated Load Typical 100K 100K 100K 100K 100K 100K Operating 0C 55 C 0C 55 C 0C 55 C 0C 55 C 0C 55 C 0C 55 C Temperature Storage Temperature 20 C 70 C 20 C 70 C 20 C 70 C 20 C 70 C 20 C 70 C 20 C 70 C Humidity 40 C 80 RH 40 C 80 RH 40 C 80 RH 40 C 80 RH 40 C 80 RH 40 C 80 RH non condensing 1 The 20 channel multiplexer can be used as 20 two wire measurement channels or 10 four wire measurement channels The 32 channel multiplexer can be used as 32 two wire measurement channels or 16 four wire measurement channels 2 The2 groups of channels 32 channels per group of the 64 channel multiplexer use the same LO terminal 3 The24 channel multiplexer can be used as 20 two wire measurement channels or 10 four wire measurement channel and 4 current channels 4 Integration time 0 02PLC channel delay 0 auto zero off alarm off scaling off display off data to internal memory 5 Matching impedance is 509 6 The isolation between banks is greater than 40dB 7 Thethermocouple precision of this module depends on the LO setting rather than the temperature of the cold terminal 8 Only for resistive loads RIGOL MC3534 Chapter 6 Specifications Digital Input Output DI O Port 1 2 3 4 8 bit input or output non isolated Type Vin L Vin H Vout L V
18. Press Config Edit gt Copy to enter the copying interface Y Local Module Channel Extended vEEEE Diagram Module Channel Extended Return Figure 2 8 Configuration Copy M300 User s Guide 2 39 RIGOL Chapter 2 Front Panel Operations Module Copy Module copy can copy the configuration of a module source module to another module target module of the same model After that the configuration of each channel of the source module is copied to the target module automatically Press Module to enter the module copy interface e Switch switch the cursor between the Source Module and Target Module areas OK copy the configuration of the source module selected to the target module selected Return return to the upper level menu 1 Local Target M Oo d ule Slot J 74 TA TA Switch OK Return Figure 2 9 Module Copy Steps of Module Copy 1 After entering the module copy interface the Source Module area is selected by default Use the up down direction keys to select the desired source module 2 Press Switch to move the cursor to the Target Module area and use the up down direction keys to select the desired target module 3 Press OK to execute module copy Note The source module and target module selected must be of the same model and it is forbidden to copy the configuration of a module to itself 2 40 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Ch
19. and Manuall P The three IP configuration modes cannot be all disabled at the same time 1 DHCP Dynamic Host Configuration Protocol In this mode the DHCP server of the current network distributes network parameters such as the IP address to the instrument Use the up down direction keys to select DHCP and use the left right direction keys to enable or disable the DHCP configuration mode By default DHCP is enabled 2 AutolP 3 In this mode the instrument acquires an IP address from 169 254 0 1 to 169 254 255 254 and subnet mask 255 255 0 0 according to the current network configuration Use the up down direction keys to select Autol P and use the left right direction keys to enable or disable the AutolP configuration mode By default AutolP is disabled Note To acquire an IP address using this mode you need to disable the DHCP mode If DHCP is not disabled the instrument will use the DHCP mode automatically Manuall P In this mode users define the network parameters of the instrument including the IP address subnet mask default gateway and DNS Domain Name Service Use the up down direction keys to select Manuall P and use the left right direction keys to enable the ManuallP configuration mode By default ManuallP is disabled At this point you need to use the up down direction keys to select IP Addr SubMask Gateway and DNS respectively and use the numeric keyboard to input the correct values Not
20. closing this channel the instrument opens the channel currently closed Pressing OK again can read the measurement result again 1 Open Disconnect the channel currently selected The channel number of the channel currently disconnected is displayed in gray for example E if one or more channels of the current module are added to the scan list this operation is invalid 2 Close Close the channel currently selected The channel number of the channel currently closed is displayed in green for example BEL it one or more channels of the current module are added to the scan list this operation is invalid 3 Read Read the measurement value of the current channel This operation is only applicable to channels added to the scan list 4 Card Reset Initialize the status of each channel of the current module If one or more channels of the current module are added to the scan list this operation is invalid M300 User s Guide 2 49 RIGOL Chapter 2 Front Panel Operations 5 Bank Separate Bank J oined By default this module is in Bank Joined state namely the bank switch in Figure 2 14 is closed and the control interface is as shown in Figure 2 15 At this point this module can only be used for 2 wire connection Press Page Dn gt Bank Sepa to execute the Bank Separate operation namely disconnect the bank switch in Figure 2 14 and the control interface is as shown in Figure 2 16 At this point this module is divided into tw
21. display English lowercase input mode display In additional in Chinese input mode this area displays Chinese characters for selection Use the numberic keyboard to input the number corresponding to the Chinese character to input the desired Chinese character 2 Filename Input Area Display the letter entered You can use the left right direction keys to select the desired letter to edit 3 Letter Selecting Area Use the direction keys to select desired letter and press OK 1 14 M300 User s Guide 4 Chapter 1 Quick Guide RIGOL Switch the I nput Mode Press Ch A a to switch the input mode to Chinese English uppercase or English lowercase Chinese can input Chinese characters and number from 0 to 9 and is displayed in the Input Mode Icon Area e English uppercase can input English uppercase letters from A to Z and number from 0 to 9 and is displayed in the Input Mode Icon Area e English lowercase can input English lowercase letters from a to z and number from 0 to 9 and is displayed in the Input Mode Icon Area Switch the Cursor Press Cursor to switch the cursor to select Filename Input Area Letter Selecting Area or Chinese Character Selecting Area only available in Chinese input mode e When Filename Input Area is selected you can use left right direction keys to select desired letter to edit When Letter Selecting Area is selected use the direction keys to select des
22. 200 uA 2 mA 2 A and 10 A ranges when frequency is gt 1 kHz Measuring Characteristics True RMS AC Voltage Measurement Method AC coupled True RMS measure the ac component of input with up to 300 V DC bias on any range Crest Factor lt 5 at full range Input Impedance 1 MQ 2 in parallel with 150 pF capacitance on any range Input Protection 300 Vrms on all ranges AC Filter Bandwidth Slow 3 Hz 300 kHz Medium 20 Hz 300 kHz Fast 200 Hz 300 kHz CMRR common mode rejection ratio 70 dB for the 1 KQ unbalance in LO lead 60 Hz common mode signal frequency 300 Vpeak maximum True RMS AC Current Measurement Method Direct coupled to the fuse and shunt AC coupled True RMS measurement measure the AC component Crest Factor lt 3 at full range Max Input DC AC current peak value 30096 of range Current with DC current component 1 Arms Shunt Resistor 100 Q for 200 uA 2 mA 1 Q for 20 mA 200 mA 0 01 O for 1A Settling Time Considerations The default measurement delay of the multimeter can ensure the correctness of the first readings of most of the measurements Make sure the RC circuit of input terminal has been fully settled about 1 s before accurate measurement M300 User s Guide 6 5 RIGOL Chapter 6 Specifications Frequency and Period Characteristics Accuracy Specifications 96 of reading
23. 300 000V 0 0020 0 0006 0 0040 0 0010 0 0055 0 0010 0 0005 0 0001 200 0000LA 0 03V 0 010 0 012 0 040 0 015 0 050 0 015 0 0020 0 0030 2 000000mA 0 25V 0 007 0 003 0 030 0 003 0 050 0 003 0 0020 0 0005 DC Current 20 00000mA lt 0 07V 0 007 0 012 0 030 0 015 0 050 0 015 0 0020 0 0020 200 0000mA 0 7V 0 010 0 002 0 030 0 003 0 050 0 003 0 0020 0 0005 1 000000A 0 12V 0 050 0 020 0 080 0 020 0 100 0 020 0 0050 0 0010 200 00000 1mA 0 0030 0 0030 0 008 0 004 0 010 0 004 0 0006 0 0005 2 000000kO 1mA 0 0020 0 0005 0 008 0 001 0 010 0 001 0 0006 0 0001 Resistance 20 00000kO 100uA 0 0020 0 0005 0 008 0 001 0 010 0 001 0 0006 0 0001 4 200 0000kO 10uA 0 0020 0 0005 0 008 0 001 0 010 0 001 0 0006 0 0001 1 000000MO 2uA 0 002 0 001 0 010 0 001 0 012 0 001 0 0010 0 0002 10 00000MO 200nA 0 015 0 001 0 030 0 001 0 040 0 001 0 0030 0 0004 100 0000MO 200nA 10MQ 0 300 0 010 0 800 0 010 0 800 0 010 0 1500 0 0002 1 Specifications are for 90 minute warm up and 100 PLC integration time 2 1096 overrange on all ranges 3 Relative to calibration standards 4 Specifications are for 4 wire resistance measurement Add 3 Q additional error in 2 wire resistance measurement M300 User s Guide 6 1 RIGOL Chapter 6 Specifications Measuring Characteristics DC Voltage Input Impedance 200mV 2V 20
24. C Thermocouple K 270 C to 1372 C 0 5 C 0 03 C N 270 C to 1300 C 0 5 C 0 04 C R 50 C to 1768 1 C 0 5 C 0 09 C S 50 C to 1768 1 C 0 6 C 0 11 C T 270 C to 400 C 0 5 C 0 03 C 1 Specifications are for 90 minutes warm up Probe error excluded 2 Specification is for 4WR resistance measurement 3 Relative to cold junction temperature accuracy is based on ITS 90 Built in cold junction temperature refers to the temperature inside the banana jack and its accuracy is 2 5 C 6 8 M300 User s Guide Chapter 6 Specifications RIGOL Measuring Characteristics Thermocouple Conversion ITS 90 software compensation Reference J unction Type Internal Fixed or External T Check Selectable per channel When the channel resistance is gt 5kQ it is considered as Open RTD Alpha 0 00385 DIN IEC 751 using ITS 90 software compensation 0 00389 0 00391 or 0 00392 using IPTS 68 software compensation Thermistor 44004 44007 44006 series Measurement Considerations The built in cold junction temperature tracks the temperature inside the terminal block The change of temperature in the terminal block might cause additional error When using the built in cold junction compensation connect the sensor terminal of the thermocouple to the terminal block and warm it up for more than 3 minutes to minimize the error M300 User s Guide 6 9 RIGOL
25. Gefahrenquellen die leibliche Sch den oder den Tod von Personen zur Folge haben k nnen CAUTI ON Die Kennzeichnung Caution Vorsicht beschreibt Gefahrenquellen die Sch den am Ger t hervorrufen k nnen Begriffe auf dem Produkt Diese Bedingungen k nnen auf dem Produkt erscheinen DANGER weist auf eine Verletzung oder Gef hrdung hin die sofort geschehen kann WARNING weist auf eine Verletzung oder Gef hrdung hin die m glicherweise nicht sofort geschehen CAUTION weist auf eine Verletzung oder Gef hrdung hin und bedeutet dass eine m gliche Besch digung des Instruments oder anderer Gegenst nde auftreten kann Symbole auf dem Produkt Diese Symbole k nnen auf dem Produkt erscheinen A A Hh GefahrlicheS Sicherheits Schutz erde Geh usemasse Erde pannung Hinweis VIII M300 User s Guide RIGOL General Care and Cleaning General Care Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time Cleaning Clean the instrument regularly according to its operating conditions To clean the exterior surface perform the following steps 1 Disconnect the instrument from all power sources 2 Clean the loose dust on the outside of the instrument with a lint free cloth with a mild detergent or water When cleaning the LCD take care to avoid scarifying it CAUTI ON To avoid damage to the instrument do not expose it to caustic liquids W
26. Ignore the instrument does not execute any operation and automatically recognize the module Usually the system works more efficiently when Ignore is selected CAUTI ON In any situation hot plug might result in unstability as well as product damage or data loss Note Hot plug hot swap is defined as inserting or removing the module without turning off the instrument Analog Output Select Analog Output using the up down direction key and select OFF or ON using the left right direction key to open or close the analog switch on the back board inside the M300 main frame The analog switch on the back board inside the M300 main frame is used to control the connection between the multiplexer module and DMM module When the switch is open OFF is selected the multiplexer module and DMM module is not connected at this point M300 cannot measure the signal input from the multiplexer channel correctly To acquire correct measurement close the switch select ON Brightness Use the up down direction keys to select Bright and use the left right keys to adjust the brightness between 0 and 15 the default is 8 The current setting is stored in non volatile memory M300 User s Guide 2 85 RIGOL Chapter 2 Front Panel Operations I O Configuration M300 supports RS232 GPIB LAN and USB interfaces After configuring proper parameters for the above interfaces you can control M300 remotely Press Utility gt I O to e
27. LSB Channel 04 M a I DIO 8 ci 8 i CHN O 3 IN OS Channel 05 nn lt G V G 57 INO 4 Channel o8 lt IN G o G O TOT 32 jy 32 Channel 09 Channel 12 DAC1 DAC4 Figure 2 27 MC3534 Schematic Diagram 16 16 i You can enter the channel configuration guide of the MFC module by pressing Control selecting MC3534 and pressing Config The channel number is S01 by default wherein S is the number of the module slot and the range is from 1 to 5 M300 User s Guide 2 61 RIGOL Chapter 2 Front Panel Operations and the function is fixed as DIO For detailed information about the channel configuration please refer to Channel Configuration This section mainly introduces how to control MC3534 The control interface of MC3534 is as shown in the figure below Use the up down and left right direction keys to select the desired channel The channel currently selected is highlighted for example a1 You can perform various operations on the channel currently selected such as the write and read operations 1 Local DIN DIN DIN DIN 01 66666606 D2 666660666 U3 666666006 D4 66660666 0b1111 1111 oT F9 IN art 9 IN at 9 IN at 9 IN 05 Gate 06 ror 07 Gate 08 ror 77 9 Mi DAC gt 42 Write Chan Reset Return Figure 2 28 MC3534 Control Interface 1 DIO Channel Control Digital Input DIN the f
28. NONE Alarm Channel Alarm1 HI 0 LO 0 Alarm Channel Configuration Alarm Queue Not cleared Output Status Cleared Output Mode Latch Alarm Output Low Level Channel Monitor Monitor in Progress Stop Module Control MC3120 MC3132 MC3164 All channels off MC3416 All channels off 4 DIO ports input MC3534 4 TOT counts 0 4 DAC 0 VDC MC3648 All channels off System Related Date No change Time No change Language No change Sound ON DMM No change Screen Saver OFF Decimal Point f Separator None Power on Default 2 80 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Power Key OFF Module Plug No change Brightness 8 Error Queue Not cleared Power on Settings Use the up down direction keys to select Power on and use the left right direction keys to select Default or Last the configuration before the last power off as the initial status at power on If Default is selected the system will be restored to its factory settings when starting up next time The following settings always recall the configuration before the last power off and will not be affected by Default Power Switch Language e Module Plug 1 0 Configuration Power off Status Use the up down direction keys to select Power off and use the left right direction keys to select ON or OFF the default is OFF When ON is selected the prompt messag
29. SENSE LO terminal on the external terminal block 5V 2 Connected to the DGND 2 digital temperature SDA 1 sensor 1 C type on SCL 1 the external terminal block Notel n is an integer from 01 to 32 M300 User s Guide 7 3 RIGOL MC3164 Chapter 7 Appendix Pin Label Pin Qty Definition nl 64 Correspond to the 64 channel input terminal on the external terminal block SE COM Correspond to the COM terminal on the external terminal block NC 11 Connection is forbidden EX HI Correspond to the HI terminal on the external terminal block EX LO Correspond to the LO terminal on the external terminal block Note n is an integer from 01 to 64 7 4 M300 User s Guide Chapter 7 Appendix MC3324 RIGOL Pin Label Pin Qty Definition HI n 28 Correspond to the 24 channel differential positive input terminal on the external terminal block LO n 28 Correspond to the 24 channel differential negative input terminal on the external terminal block NC 10 Connection is forbidden EX HI Correspond to the SOURCE HI terminal on the external terminal block EX LO Correspond to the SOURCE LO terminal on the external terminal block EX HS Correspond to the SENSE HI terminal on the external terminal block EX LS Correspond to the SENSE LO
30. alarm messages of the current scan the first 100 including the reading and time when the alarm is generated and error messages up to 20 Note Up to 100 alarm messages on each alarm channel can be logged in the alarm queue and the other alarm messages would be lost and up to 20 error messages can be logged in the error queue and the other error messages will also be lost Module I ndicators Correspond to the 5 module slots at the rear panel respectively The corresponding module indicator goes on when the module is inserted into the slot at the rear panel M300 User s Guide 1 7 RIGOL Chapter 1 Quick Guide Rear Panel 1 Figure 1 4 Rear Panel of M300 Slots The M300 provides 5 slots for 5 modules The 5 slots from left to right in the figure above correspond to the 5 module indicators at the front panel respectively When a module is inserted in the slot the corresponding module indicator at the front panel goes on Figure 1 4 is the schematic diagram of the rear panel without the modules inserted For information of each module refer to Plug in Module Overview CAUTI ON Only one DMM module is permitted for one mainframe USB Device Via this interface M300 can communicate with the PC as a slave device and you can control it remotely via the PC LAN M300 conforms to the LXI Core 2011 Device standard The instrument can be connected to the network via this i
31. brevet rb utu bm a a eus 1 26 MORIEOE sat ooo a dehet E eda deni a der data 1 27 View Switch Key scie niece esas an x ote enu eaa S Rupees bb UH E Rand 1 27 eu 1 28 Alarm Channel Setting Keys ev ee ont OD GG RU RI RE Sei 1 28 Chapter 2 Front Panel Operations ecce 2 1 Scan Contigurati n user RATA ead al ces th teaver n P Re OG ore re e ERR Re Ng 2 2 se NM p 2 2 Trigger Modes opta erede ee e npe eei dea e a eg 2 8 Channel Configuration sssseseee mmn 2 12 Channel Configuration When Using External DMM aese 2 21 Measurement Parameter Configuration sseeeeene 2 22 Digital Iit err ir i ERR re e o A ERE T Du 2 35 Totallzer na oae e eR ex p e o e es 2 37 M300 User s Guide XIII RIGOL Contents Configuration COPY isse itte eere rra ee Sor cepe or era ndi ds 2 39 Channel Monit fics oreet nere p o d Ene ED Ue DER RN REY 2 44 Single channel Monitor essessseseeeen mnn 2 44 Multi channel Monitor essen 2 46 All Channel Monitor esseeeenereeensneenenennnen A emm nnn 2 46 Module Control penn eene nnne nnn nns 2 47 To Control MC3120 Hrn aan le E e turn 2 48 To Control MC3132 s n EHE retra ER Fr con a dd 2 5 To Control MCS TD4 u er 2 54 To Control MC3324 nenne CORDE 2 56 To Control MCI ED en ede REPE a or e ra mc aat 2 59 To Control MC3534 icr retis eer re Eel Rate Rp t edv 2 61 TO Control MC3648 crier erret Hr een a rH C d d 2 64 Al
32. direction keys The instrument executes a scan when an alarm is generated from the channel selected 8 Trigger Edge Type When the trigger mode is External select Edge using the up down direction keys and select Rising or Falling using the left right direction keys The instrument executes a scan when a TTL trigger pulse with specified edge is received by the external trigger signal input terminal M300 User s Guide 2 5 RIGOL Chapter 2 Front Panel Operations To Edit the List Press Config Edit to enter the scan list editing interface list TENE uf I Copy Delete Modify Page Dn Figure 2 2 Scan List Editing Interface New Create a new channel configuration Press this softkey to enter the channel configuration guide Finish all the configurations according to the introductions in Channel Configuration and press Done to return to the scan list editing interface Copy Copy the channel configuration created to other channel or module For details refer to the introduction in Configuration Copy Modify Modify the channel configuration created Select the desired channel using the direction keys and press Modify to open the configuration guide of the corresponding channel At this point you can modify the configuration of this channel and after the modification is finished press Done to return to the scan list editing interface Note The channel in Chan No can only be modified to the channels in
33. ie Precision ee Retum peau Bar Figure 1 5 User Interface 1 Status Bar Help Displayed when the built in help is enabled Error Displayed when an error occurs Displayed when the keyboard is locked Displayed when an alarm is generated Displayed when the system sound is disabled Displayed when the instrument is connected to a network correctly Displayed when an USB storage device is detected Local Displayed when the instrument is in local mode Rmt Displayed when the instrument is in remote mode 2 Content Area Display the measurement configuration guide measurement parameters settings etc For more detailed information please refer to the User s Guide 3 Menu Bar Display the menus of the current function corresponding to the menu softkeys below respectively Pressing the softkey can activate the corresponding menu M300 User s Guide 1 13 RIGOL Chapter 1 Quick Guide To I nput Filename M300 supports filenames consisting of Chinese characters English letters and numbers The length of the filename can not exceed 12 bytes The filename input interface is as shown in the figure below Input Mode Icon Chin se Character Selecting Area Letter Selecting Area Filename Input Area Cursor Delete Figure 1 6 Filename Input Interface 1 Input Mode I con Chinese Character Selecting Area Display the icon of the input mode currently selected Chinese input mode display English uppercase input mode
34. in the figure below USB TMC RS232 Verify All INSTR ntrol RIGOL TECHNOLOC 6 Communication Test Right click the resource name M300 GPIBO 7 INSTR to select SCPI Panel Control to turn on the remote command control panel through which you can send commands and read data M300 User s Guide 4 7 RIGOL Chapter 4 Remote Control To Control M300 through USB 1 Connect the devices Connect M300 USB Device with your PC USB Host using USB cable I nstall the USB drive As this instrument is a USB TMC device after you connect it to the PC and turn both on the PC will display the hardware update wizard dialog box Please install the USB Test and Measurement Device IVI driver following the directions of the wizard The detailed steps are the same as the instruction in User defined Programming Search device resource Start up Ultra Sigma the software starts searching the EXIT resources currently connected to the PC automatically or click to search manually The status bar of Ultra Sigma at present is as shown in the figure below View the resource The resources that have been searched successfully are shown under the RIGOL Online Resource catalog and the instrument model and the USB interface information are also displayed as shown in the figure below LAN GPIB USB TMC RS232 Verify All RIGOL Online Resource a M3 AB1 0x0C50 M30012 23 INSTR Other Online
35. input range M300 User s Guide 1 23 RIGOL Chapter 1 Quick Guide 3 Connectthe terminal block to the module via interface 1 the interface definition is in Figure 1 7 on the module and screw down the two screw rods onto the nuts on the module 1 24 M300 User s Guide Chapter 1 Quick Guide RIGOL To Replace the Fuse If a new fuse is required please refer to the following steps 1 Turn off the instrument and remove the power cord 2 Insert the small straight screw driver into the slot at the power socket and pry out the fuse seat 3 4 CAUTI ON Please use the specified power fuse and make sure that the fuse seat is not short circuited to avoid electric shock and fire M300 User s Guide 1 25 RIGOL Chapter 1 Quick Guide Menu Quick Navigation This section provides the structures of the main menus for M300 to guide users to quickly get familiar with the operations For more detailed information please refer to Front Panel Operations Config 1 26 M300 User s Guide Chapter 1 Quick Guide RIGOL Monitor Single channel Monitor i5 1 By default the first channel in the scan list is monitored 2 The monitor interface displays the channel number module type function range alarm information and measurement value 3 You can switch the channel monitored by editing the channel number or pressing N Single LAN Multi channel Monitor
36. measurement curve in the coordinate Up to 4 channels can be added 3 Delete To delete a channel press Switch to switch the cursor to select the channel number and use the left right direction keys to select the channel number to be deleted and then press Delete 4 Zoomin Press this softkey to reduce the maximum of the horizontal axis to zoom in the curve 5 Zoom Out Press this softkey to increase the maximum of the horizontal axis to zoom out the curve 60 90 Switch Add Delete Zoom In Figure 2 34 Measurement Curve M300 User s Guide 2 71 RIGOL Chapter 2 Front Panel Operations Channel Information Press at the front panel again and the instrument enters the channel information display interface Figure 2 35 Relay Cycle Alarm Info Error Info 1 Relay Cycle M300 measures relay cycles and stores the result in the non volatile memory of the corresponding module When starting up and detecting the modules currently insterted M300 reads the cycle of each relay Via this interface you can view the statistical results to ensure proper maintenance of the relay For more information please refer to the schematic diagrams of the modules in Module Control 2 Alarm Information Read and clear the first 100 alarm messages generated on 4 alarm channels of the last scan including the alarm reading the time channel and mode of the alarm as well as the alarm channel 3 Error Information Read and clear the first
37. of M300 and computer are identical If the problem remains please refer to the next step Check whether the RS232 cable is a cross line If the problem remains please contact RIGOL 14 The LAN interface does not work normally 1 2 Check whether the cable works normally and the connection is correct and reliable Check the IP address configuration mode e f DHCP is enabled please ensure the network supports DHCP mode and can distributes network parameters such as the IP address to the instrument automatically e f DHCP is disabled and Auto IP is enabled please ensure the IP addresses of M300 and computer is in the same network segment e f Manual IP is enabled please ensure the IP addresses of M300 and computer is in the same network segment If the problem remains please contact RIGOL M300 User s Guide 5 3 Chapter 6 Specifications Chapter 6Specifications DC Characteristics RIGOL Accuracy Specifications of reading of range EET Range Test Current or 24 Hour 90 Day 1 Year ee ee Load Voltage Tca C 1 C Tcu C 5C Ta C 5 C Tan C 50 to 50 C 200 0000mV 0 0020 0 0020 0 0030 0 0025 0 0040 0 0025 0 0005 0 0005 2 000000V 0 0015 0 0005 0 0020 0 0006 0 0035 0 0006 0 0005 0 0001 DC Voltage 20 00000V 0 0020 0 0004 0 0030 0 0005 0 0040 0 0005 0 0005 0 0001 200 0000V 0 0020 0 0006 0 0040 0 0006 0 0050 0 0006 0 0005 0 0001
38. off and insert the modules following the instructions below As shown in Figure 1 7 there is a golden circuit board edge both on the top and bottom of the module Aim the two edges with the sliding chutes as pointed out by the arrows in the figure below into the main frame slot and push the module in until you hear a sound At this point interface 2 is connected to the main frame Note Please refer to Measurement Connections to connect the external Figure 1 8 Insert the Module into the Mainframe Note All the modules of the M300 do not support hot plugging Please turn the mainframe off before inserting the modules into the mainframe M300 User s Guide 1 17 RIGOL Chapter 1 Quick Guide Module Overview MC3065 DMM Module The DMM module is used to measure the signals under test and has 6 digit reading resolution The measurement functions include DCV ACV 2 wire resistance 4 wire resistance frequency period temperature and any sensor After connecting the DMM module make sure that the signal under test connected to the analog bus is no greater than 300 Vdc or 300 Vrms MC3120 20 Channel Multiplexer All 20 channels switch both HI and LO inputs thus providing fully isolated inputs to the DMM module MC3120 is divided into two banks called A and B of 10 two wire channels each When making 4 wire resistance measurement the instrument automatically pairs channels of A bank and B bank All channels in
39. performed could result in damage to the product or other devices connected to the product Symbols Used on the Product These symbols may appear on the product A A a Hazardous Safety Protective Chassis Voltage Warning Earth Ground Terminal 2 Test Ground M300 User s Guide RIGOL Allgemeine Sicherheits Informationen berpr fen Sie diefolgenden Sicherheitshinweise sorgf ltigumPersonensch denoderSch den am Ger tundan damit verbundenen weiteren Ger tenzu vermeiden Zur Vermeidung vonGefahren nutzen Sie bitte das Ger t nur so wiein diesem Handbuchangegeben Um Feuer oder Verletzungen zu vermeiden verwenden Sie ein ordnungsgem es Netzkabel Verwenden Sie f r dieses Ger t nur das f r ihr Land zugelassene und genehmigte Netzkabel Erden des Ger tes Das Ger t ist durch den Schutzleiter im Netzkabel geerdet Um Gefahren durch elektrischen Schlag zu vermeiden ist es unerl sslich die Erdung durchzuf hren Erst dann d rfen weitere Ein oder Ausg nge verbunden werden Anschluss einesTastkopfes Die Erdungsklemmen der Sonden sindauf dem gleichen Spannungspegel des Instruments geerdet Schlie enSie die Erdungsklemmen an keine hohe Spannung an Beachten Sie alle Anschl sse Zur Vermeidung von Feuer oder Stromschlag beachten Sie alle Bemerkungen und Markierungen auf dem Instrument Befolgen Sie die Bedienungsanleitung f r weitere Informationen bevor Sie weitere Anschl sse an das Instrument l
40. series number of M300 gt lt the start time of the scan gt for example G M300 data M300123123123 20131121_140506 The instrument configuration file is usually config csv and the scan data file is usually dat00001 csv If 64K row limit is selected in this scan the files such as dat00002 csv and etc may be created When OFF is selected the scan data will be stored into the internal memory Note When auto storage of data into the USB storage device is selected the measurement speed of the instrument might be affected 2 86 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Export Data Set whether the scan data stored in the internal memory will be exported to the USB storage device when a scan finishes Row Limit If 64K is selected the file stored in the USB storage device can not exceeds 64K rows Each time when the data in the scan data file exceeds 64k rows a new csv file will be created automatically by the instrument to store the following data File Separator Select Tab or as the separator of the scan data file stored in the USB storage device Import If Yes is selected when an USB storage device is inserted the instrument searches and lists the blcfg files in the root directory of the USB storage device Select the desired file and press the Import softkey to import it bicfg file stores the measurement configuration After it is imported the scan list settings including tri
41. the scan list Delete Delete a channel configuration created Select the channel to be deleted using the direction keys and press Delete to delete the corresponding channel 2 6 M300 User s Guide Chapter 2 Front Panel Operations RIGOL 5 Search When a great number of channels are added into the scan list you can press Page Dn Search use the numeric keyboard to input the desired channel number and then press OK to quickly locate this channel You can copy modify or delete the channel configuration located To Save the List The M300 allows users to save the configured scan list in internal or external memory as measurement configuration file Press Config gt Save to enter the store and recall interface the file type is automatically set to Measurement configuration Select the desired directory according to the method introduced in Store and Recall and then press Save to enter the filename input interface Input the specified filename to save the current scan list You can read the scan list stored as well as modify its configuration and save it again To Read the List The M300 allows users to read the scan list stored Press gt Read to enter the store and recall interface the file type is set to Measurement Configuration automatically Select and read the desired file according to the method introduced in Store and Recall You can modify the configuration of the scan list read and save it again
42. the DMM module and is divided into two banks called A and B of 10 two wire channels each When making 4 wire resistance measurement the instrument automatically pairs channels of A bank and B bank All channels in the scan list are break before make You can close multiple channels on this module only if you have not configured any channel to be part of the scan list 1 18 M300 User s Guide Chapter 1 Quick Guide RIGOL 4 current channels are used in combination with the DMM module to measure DC current or AC current All channels in the scan list are break before make In any case you cannot close multiple channels on this module This module can be connected with MC3065 DMM Module if MC3065 is currently inserted MC3416 16 Channel Actuator The MC3416 can connect signals to the device under test or enable external devices Any of the 16 channels can switch to Normally Open NO and Normally Closed NC states MC3534 Multifunction Module This module provides 3 kinds of functions with 4 channels for each function This module can be used to check status or control external devices such as solenoid power relays and microwave switches You can also read digital inputs and the totalizer count during a scan Four 8 bit digital input output DIO ports 4 totalizer TOT input terminals the first two channels are 10 MHz TOT and the other two channels are 100 kHz TOT with 1 Vpp sensitivity 4 analog output term
43. the count of the channel selected Start or stop counting 3 DAC channel edit the output value of the channel selected 4 Module reset MC3534 Alarm Channel Setting Keys Alarm Channel Setting 1 Set the output mode of the corresponding pin to Latch o r Track when an alarm is generated 2 Set the output status of the corresponding pin to High or Low when an alarm is generated 3 Display the numbers of channels that select the current alarm channel 4 Clear the alarm output of the corresponding pin The Alarm data is not cleared Chapter 1 Quick Guide 1 28 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Chapter 2 Front Panel Operations This chapter introduces in details the front panel operation methods and functions of M300 Topics in this chapter Scan Configuration Channel Monitor Module Control Alarm Channel Setting View Switch Store and Recall Utility M300 User s Guide 2 1 RIGOL Chapter 2 Front Panel Operations Scan Configuration Press Config at the front panel to configure M300 to create a scan create and edit the scan list and configure the parameters of each channel During the scan each time the DMM module is connected to one channel in the scan list and executes a measurement according to the channel configuration Scan List Create a scan list add any channels of the multiplexer or the DIO or TOT channels of the multifunction m
44. the range It is not user selectable For thermocouple measurements the instrument internally selects the 200 mV range For thermistor RTD and RTD 4W measurements the instrument will autorange to the correct resistance range M300 User s Guide 2 23 RIGOL Chapter 2 Front Panel Operations I ntegration Time The Integration time is the period of time that the instrument s analog to digital A D converter samples the input signal for a measurement The longer the integration time is the slower the measurement speed and the better the measurement resolution will be The shorter the integration time is the faster the measurement speed and lower the measurement resolution will be For the relationship between intergration time and resolution please refer to Specifications Integration time is applicable to DCV DCI 2WR 4WR TEMP and SENSOR except the FREQ sensor functions You can specify the integration time in the number of Power Line Cycles and aperture time Number of Power Line Cycles the unit is PLC It can be set to 0 02 PLC 0 2 PLC 1 PLC 2 PLC 10 PLC 20 PLC 100 PLC or 200 PLC and the default is 1 PLC The instrument detects the power frequency automatically at power on frequencies within 55 Hz and 66 Hz are regarded as 60 Hz PLC 1 60 s Other frequencies are regarded as 50 Hz PLC 1 50 s Aperture Time user defined integration time The range is from 33 us to 4 s When the current measurement function is
45. 0 15 0 06 0 015 0 006 5kHz 10kHz 0 35 0 70 0 35 0 70 0 35 0 70 0 030 0 006 200 0000mA 3Hz 5Hz 1 00 0 04 1 00 0 04 1 00 0 04 0 100 0 006 5Hz 10Hz 0 30 0 04 0 30 0 04 0 30 0 04 0 035 0 006 10Hz 5kHz 0 10 0 04 0 10 0 04 0 10 0 04 0 015 0 006 5kHz 10kHz 0 20 4 0 25 0 20 0 25 0 20 0 25 0 030 0 006 1 000000A 3Hz 5Hz 1 10 0 06 1 10 0 06 1 10 0 06 0 100 0 006 5Hz 10Hz 0 35 0 06 0 35 0 06 0 35 0 06 0 035 0 006 10Hz 5kHz 0 15 0 06 0 15 0 06 0 15 0 06 0 015 0 006 5kHz 10kHz 0 35 0 70 0 35 0 70 0 35 0 70 0 030 0 006 5Hz 10Hz 0 35 0 08 0 35 0 10 0 35 0 10 0 035 0 008 10Hz 5kHz 0 15 0 08 0 15 0 10 0 15 0 10 0 015 0 008 1 2 3 4 6 4 Specifications are for 90 minute warm up slow ac filter and sine wave input 1096 overrange on all ranges Relative to calibration standards Specifications are for sine wave input gt 5 of range For inputs from 1 to 5 of range and 50 kHz add 0 196 of range additional error For M300 User s Guide Chapter 6 Specifications RIGOL 50 kHz to 100 kHz add 0 1396 of range 5 When the frequency is lower than 100 Hz the specification of slow filter is only for sine wave input 6 Specifications are for sine wave input gt 5 of range For inputs from 1 to 5 of range add 0 196 of range additional error Specifications are typical values for
46. 0 open the first channel and close the next channel At this point the pin 5 of the Alarm Ext Trig interface outputs pulse again to control the external DMM to start measuring Then running down accordingly all channels in the scan list are scanned M300 User s Guide 3 5 Chapter 4 Remote Control RIGOL Chapter 4 Remote Control This chapter introduces the method to control M300 remotely You can bulid communication between M300 and computer via GPIB USB RS232or LAN to realize remote control based on the SCPI Standard Commands for Programmable Instruments commands There are two remote control methods including user defined programming and sending command using PC software Note When the instrument is working in remote mode Rmt is displayed in the status bar in the user interface and the keys except Monitor and NE at the front panel are locked At this point you can press to exit the remote mode Topics in this chapter m User defined Programming m To Use PC Software M300 User s Guide 4 1 RIGOL Chapter 4 Remote Control User defined Programming Users can select desired programming tool such as LabVIEW Matlab Visual C etc program and control M300 by using SCPI commands on the basis of NI VISA National I nstrument Virtual Instrument Software Architecture 1 Install NI VI SA Library You need to install the VISA library of NI download from http www ni com visa on your PC NI VI
47. 1 00 0 02 1 00 0 03 1 00 0 03 0 100 0 003 5Hz 10Hz 0 35 0 02 0 35 0 03 0 35 0 03 0 035 0 003 10Hz 20kHz 0 04 0 02 0 07 0 03 0 08 0 03 0 008 0 003 20kHz 50kHz 0 10 4 0 04 0 12 0 05 0 15 0 05 0 012 0 005 50kHz 100kHz 0 55 0 08 0 60 0 08 0 60 0 08 0 060 0 008 100kHz 300kHz 4 0 0 50 4 0 0 50 4 0 0 50 0 20 0 02 300 000V 3Hz 5Hz 1 00 0 02 1 00 0 03 1 00 4 0 03 0 100 0 003 M300 User s Guide 6 3 RIGOL Chapter 6 Specifications 5Hz 10Hz 0 35 0 02 0 35 0 03 0 35 0 03 0 035 0 003 10Hz 20kHz 0 04 0 02 0 07 0 03 0 08 0 03 0 008 0 003 20kHz 50kHz 0 10 0 04 0 12 0 05 0 15 0 05 0 012 0 005 50kHz 100kHz 0 55 0 08 0 60 0 08 0 60 0 08 0 060 0 008 100kHz 300kHz 4 0 0 50 4 0 0 50 4 0 4 0 50 0 20 0 02 True 200 00001A 3Hz 5Hz 1 10 4 0 06 1 10 0 06 1 10 0 06 0 200 0 006 RMS AC 5Hz 10Hz 0 35 0 06 0 35 0 06 0 35 0 06 0 100 0 006 SHINE 10Hz 5kHz 0 15 0 06 0 15 0 06 0 15 0 06 0 015 0 006 5kHz 10kHz 0 35 0 70 0 35 0 70 0 35 0 70 0 030 0 006 2 000000mA 3Hz 5Hz 1 00 0 04 1 00 0 04 1 00 0 04 0 100 0 006 5Hz 10Hz 0 30 0 04 0 30 0 04 0 30 0 04 0 035 0 006 10Hz 5kHz 0 12 0 04 0 12 0 04 0 12 0 04 0 015 0 006 5kHz 10kHz 0 20 0 25 0 20 0 25 0 20 0 25 0 030 0 006 20 00000mA 3Hz 5Hz 1 10 0 06 1 10 0 06 1 10 0 06 0 200 0 006 5Hz 10Hz 0 35 0 06 0 35 0 06 0 35 4 0 06 0 100 0 006 10Hz 5kHz 0 15 0 06 0 15 0 06
48. 120 MC3132 MC3164 All channels off MC3416 All channels off 4 DIO ports input MC3534 4 TOT counts 0 4 DAC 0 VDC MC3648 All channels off System Related Date No change 2 82 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Time No change Language No change Sound ON DMM No change Screen Saver OFF Decimal Point Separator None Power on Default Power Key OFF Module Plug No change Brightness 8 Error Queue Not cleared Screen Saver Use the up down direction keys to select Saver and use the left right direction keys to select OFF or set the time to wait before entering the screen saver state OFF disable the screen saver function Time you can use the numeric keyboard to set a value between 1 and 60 The unit is minute and the resolution is 1 minute The instrument enters the screen saver state automatically when the instrument has not been operated for the specified period of time Decimal Use the up down direction keys to select Decimal and use the left right direction un keys to set the display form of the decimal point of the screen data to or M300 User s Guide RIGOL Chapter 2 Front Panel Operations Separator Use the up down direction keys to select Separator and use the left right direction keys to set the display form of the separator of the screen data to or None or Space Tip The dis
49. 20 channels of this module are in Bank Joined state namely the bank switch in Figure 2 23 is closed and the control interface is as shown in Figure 2 24 At this point this module can only be used for 2 wire connection Press Page Dn Bank Sepa to execute the Bank Separate operation namely disconnect the bank switch in Figure 2 23 and the control interface is as shown in Figure 2 25 At this point this module is divided into two independent banks and can be used for 4 wire connection When it is used for 4 wire resistance measurement the instrument pairs channel n and channel n4 10 automatically wherein n is an integer between 1 and 10 Channel n is used to connect the source terminal of DMM and channel n 10 is used to connect the sense terminal of DMM Page Up Bank Joined Return Figure 2 24 MC3324 Control Interface Bank Separate 2 58 M300 User s Guide Chapter 2 Front Panel Operations RIGOL To Control MC3416 MC3416 is a 16 channel actuator ACT 16 for short This module contains 16 independent SPDT single pole and double throw switches This module does not connect to the DMM module and only provide access to the NC Normally Closed NO Normally Open and COM Common contacts on each switch It can be used to control external devices Terminal Block M3TB16 HC co ri E ME 01 COM Is O el un 78 Pin NE O e Interface Signals s to b e lt eee witched e NC H Le 16 O Sag COM
50. 20 error messages in the error queue When remote command grammar error or hardware error is detected the system displays Error in the user interface and stores the first 20 error messages in the error queue of the instrument If error is still generated the last error message in the error queue will be replaced by Error queue overflow At this point the instrument will not store the new error messages unless error messages in the error queue are removed Turning off the instrument or using the remote command CLS can clear the error queue 2 72 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Store and Recall M300 allows you to save the system configuration measurement configuration and measurement data into the internal or external memory and recall them when required Press Store Recall to enter the interface as shown in the figure below SysConfig Format Return Directory File File Type Current Directory Figure 2 36 Store and Recall 1 File Type Display the four file types of M300 storage and recall You can select the desired file type using the left right direction key 2 Current Directory Dispay the directory currently selected In the interface displayed by default when is pressed this area is empty To select the specified directory please execute the following operation in the interface as shown in Figure 2 37 1 Select the desired drive letter C or D using the up down directi
51. 300 User s Guide Chapter 4 Remote Control RIGOL Hardware Update Wizard Completing the Hardware Update Wizard The wizard has finished installing the software for USB Test and Measurement Device IVI Click Finish to close the wizard 7 EE mem Using RS232 Interface Convert the RS 232 Alarms Ext Trig interface at the rear panel to two 9 pin interfaces using the Mixed Interface Convert Cable accessory Wherein one is a 9 pin male interface used as a standard RS232 interface Connect this interface with the computer using RS232 cable and set proper parameters according to To Set RS232 Parameters Using LAN Interface Connect M300 to the LAN of the computer and set proper parameters according to To Set LAN Parameters 3 Programming Then you can select a familiar software developing tool such as LabVIEW Matlab Visual C etc for programming For detailed information about SCPI commands and programming method of M300 refer to M300 Programming Guide M300 User s Guide 4 5 RIGOL Chapter 4 Remote Control To Use PC Software You can control the instrument remotely by sending SCPI commands using PC software RIGOL Ultra Sigma This section introduces in details how to use Ultra Sigma to send commands through multiple interfaces to control M300 Please refer to Ultra Sigma Help to install the software and the components required correctly About how to acquire the Ultra Sigma software please con
52. ARNI NG To avoid injury resulting from short circuit make sure the instrument is completely dry before reconnecting to a power source Environmental Considerations The following symbol indicates that this product complies with the applicable WEEE Directive 2002 96 EC K Product End of Life Handling The equipment may contain substances that could be harmful to the environment or human health In order to avoid release of such substances into the environment and harm to human health we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately Please contact your local authorities for disposal or recycling information M300 User s Guide IX RIGOL M300 Overview The M300 series data acquisition switch system provides 5 slots to accept any combination of 8 kinds of plug in modules including a DMM module multiple multiplexers an actuator a matrix switch module and a multifunction module With its modular structure it combines precision measurement capability with flexible signal connections and can provide versatile solutions for the applications that require multiple points or signals to be tested This can be especially useful in product performance testing during the R amp D phase and automatic test during production process It comes with a 4 3 inch high resolution color LCD user friendly interface design easy to use keyboard layout as w
53. DCV DCI 2WR 4WR TEMP or SENSOR except the FREQ sensor enter the fourth step of the channel configuration guide use the up down direction keys to select Integ and use the left right direction keys to select the desired integration time The integration time can be set to User 0 02PLC 0 2PLC IPLC 2PLC LOPLC 20PLC 100PEC and 200PLC If User is selected you need to use the numeric keyboard to input the desired integration time and then press OK The range is from 33 us to 4 s Input I mpedance The DC input impedance setting is applicable to DCV and SENSOR DCV sensor measurement functions The default is 10 MQ For 200 mV 2 V and 20 V ranges you can select gt 10 GQ to reduce the affect caused by the loading error of the measurement object The current setting is stored in the volatile memory When the current measurement function is DCV or SENSOR DCV sensor enter the fourth step of the channel configuration guide use the up down direction keys to select IMP and use the left right direction keys to select the desired impedance e 10M Set the DC input impedance to 10 MQ The input impedances are 10 MQ on all ranges e gt 10G Set the DC input impedance to gt 10 GQ The input impedances are gt 10 GQ on the 200 mV 2 V and 20 V ranges and are 10 MQ on the other ranges 2 24 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Auto Zero Auto Zero AZ is appli
54. IOD or SENSOR FREQ sensor enter the fourth step of the channel configuration guide use the up down direction keys to select Filter and use the left right direction keys to select 3Hz 20Hz or 200Hz The default is 20Hz 2 26 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Gate Time Gate time also called Aperture Time applies to FREQ or PERIOD function It decides the resolution of low frequency measurement The longer the gate time is the higher the resolution of the low frequency measurement is and the slower the measurement is and vice versa When the current measurement function is FREQ or PERIOD enter the fourth step of the channel configuration guide use the up down direction keys to select GateTime and use the left right direction keys to select 1ms 10ms 100ms or 1s The default is 100ms Channel Delay Channel delay is applicable to all the multiplexer channels Channel delay is the time interval between two neighboring channels during a scan as shown in the figure below During the scan the instrument immediately disconnects the previous channel and closes the current channel after finishing measuring the previous channel The instrument starts the next measurement after the specified channel delay time has been reached A Scan EE uu Channel Delay When the current channel is a multiplexer channel enter the fourth step of the channel configuration guide use the up down dir
55. L Unable to do the configurations or operations related to the DMM module when it has been inserted 1 Ensure the setting of gt System gt DMM is ON If the problem remains please refer to the next step 2 Press Utility Test select the DMM module using the left right direction keys and press View to view the information of the DMM module If the information does not been displayed or displayed correctly the DMM module is abnormal due to high current or power load failure probably At this point please turn the mainframe off and restart it a few minutes later 3 If the problem remains please contact RIGOL 5 2 M300 User s Guide Chapter 5 Troubleshooting RIGOL 11 The GPIB interface does not work normally Check whether the cable works normally and the connection is correct and reliable Ensure the GPIB addresses of M300 and computer are identical If the problem remains please contact RIGOL 12 The USB Device interface does not work normally 1 2 3 Check whether the cable works normally and the connection is correct and reliable Check whether the USB Test and Measurement Device drive program has been installed in your computer namely check whether USB Test and Measurement Devices is displayed in the device manager of your computer If the problem remains please contact RIGOL 13 The RS232 interface does not work normally Ensure the RS232 parameters baud rate etc
56. M300 User s Guide 2 7 RIGOL Chapter 2 Front Panel Operations Trigger Mode The M300 provides 5 trigger modes including auto trigger manual trigger external trigger absolute time trigger and alarm trigger Auto Trigger In auto trigger mode the scan rate is determined by the specified interval If the time required for a complete scan of the scan list is greater than the specified interval the instrument will execute a continuous scan at the fastest possible rate no error message will be generated The range of the interval is from 0 ms to 99 59 59 999 hours the time format is Hour Minute Second Millisecond and the resolution is 1 ms If the number of scans is set to Infinite press Run Stop at the front panel and the instrument executes a continuous scan automatically at the specified interval You can press and hold the key to stop the scan If the number of scans is a specific value 1 to 50000 press Run Stop at the front panel and the instrument executes the specified number of scans at the specified interval and then stops automatically During the scan the instrument stores the measurement data scaling data and alarm data in the non volatile memory automatically You can press at the front panel to view the history measurement data including the maximum minimum average and standard deviation Manual Trigger In manual trigger mode after you press Run Stop at the front panel the instrument e
57. MC3132 and MC3164 add it to the scan list and use it for measurement you need to first insert and enable the DMM module Utility gt DMM gt ON or connect an external digital multimeter 2 Function Use the up down direction keys to select Function and the left right direction keys to select the desired measurement function If the current channel is the voltage channel of the multiplexer the measurement functions available include DCV ACV 2WR 4WR only supported by the first 16 channels of MC3132 and the first 10 channels of MC3120 and MC3324 FREQ PERI OD TEMP and SENSOR If the channel is current channel only for MC3324 the measurement functions available include DCI ACV and SENSOR After selecting the desired measurement function please configure the basic measurement parameters for the function selected by referring to Measurement Parameter Configuration and press Next to configure the scaling or alarm parameters If the current channel is the DIO channel of the multifunction module Function is fixed at DIO Please refer to Digital I nput to set the basic parameters and press Next to configure the alarm parameters If the current channel is the TOT channel of the multifunction module Function is fixed at TOT Please refer to Totalizer to set the basic parameters and press Next to configure the alarm parameters 2 14 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Scaling Configurat
58. Panel Operations If the channel using the scaling function is removed from the scan list the scaling function of the channel will be disabled but the scaling coefficients will not be reset When the channel is added into the scan list again without changing the measurement function the scaling function will be enabled and the scaling coefficients remain unchanged For a channel using the scaling function when it is in single channel monitor all the measurement readings are scaling readings 1 Enable Scaling Configuration Enter the scaling configuration interface and Scaling is selected by default You can use the left right direction keys to select ON At this point the fitting function coefficients and unit are displayed on the interface 2 Set the Fitting Function Coefficients Use the up down direction keys to select A B and C respectively and use the numeric keyboard to input the desired values The range is from 1E 15 to 1E 15 and the default is 0 The fitting function defined is displayed in Fitting 3 Setxl Use the up down direction keys to select x1 and press Curr Value to acquire the current reading as x1 You also can set x1 by using numeric keyboard 4 Setthe Unit Set the unit of the measurement result Use the up down direction keys to select Unit and then press OK to enter the filename input interface Input the desired unit and press OK The length of user defined units cannot exceed 3 bytes You can use direction ke
59. RIGOL User s Guide M300 Series Data Acquisition Switch System Dec 2014 RI GOL Technologies Inc RIGOL Guaranty and Declaration Copyright 2013 RIGOL Technologies Inc All Rights Reserved Trademark I nformation RIGOL is a registered trademark of RIGOL Technologies Inc Publication Number UGC07110 1110 Software Version 00 02 00 04 19 Software upgrade might change or add product features Please acquire the latest version of the manual from RIGOL website or contact RIGOL to upgrade the software Notices RIGOL products are covered by P R C and foreign patents issued and pending RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at company s sole decision Information in this publication replaces all previously corresponding material Information in this publication is subject to change without notice RIGOL shall not be liable for either incidental or consequential losses in connection with the furnishing use or performance of this manual as well as any information contained e Any part of this document is forbidden to be copied photocopied or rearranged without prior written approval of RIGOL Product Certification RIGOL guarantees this product conforms to the national and industrial standards in China as well as the 1S09001 2008 standard and the 15014001 2004 standard Other international standard conformance certifications are in prog
60. SA is an application program interface developed by NI according to the VISA standards You can use NI VI SA to realize the communication between the generator and PC through instrument buses such as USB VISA defines a set of software commands with which users can control the instrument without the need to understand how the interface bus works For details please refer to NI VISA Help 2 Build Communication between Instrument and PC You need to build the communication between the generator and PC e Using GPIB Interface Connect M300 with the PC with GPIB card installed using GPIB cable before using the GPIB General Purpose Interface Bus Interface and set proper parameters according to To Set GPIB Address Using USB Interface Use USB cable to connect M300 and PC At this point the hardware update wizard dialog box is displayed Then follow the instructions to install the USB Test and Measurement Device IVI The steps are as following Select Install the software automatically Recommended and click Next Select Don t search will choose the driver to install and click Next Select USB Test and Measurement Device and click Next When the installation is finished click Finish I BWN 4 2 M300 User s Guide Chapter 4 Remote Control RIGOL Found New Hardware Wizard This wizard helps you install software for USB Test and Measurement Device IVI If your hardware came with an
61. V ranges 10MQ or gt 10GQ For these ranges input beyond 26 V are clamped through 106 kQ 200V and 300V ranges 10MQ 1 Input Protection 300V Input Offset Current 50pA at 25 C typical CMRR common mode 140 dB for 1 kQ unbalanced resistance in LO lead 300 VDC peak maximum rejection ratio Resistance Measurement Method 4 wire or 2 wire resistance Current source referenced to LO input Open circuit Voltage Limited to 10 V Max Lead Resistance 10 of range per lead for 200 Q and 2 kQ ranges 1 KQ per lead on all other ranges 4 wire Input Protection 300V on all ranges Offset Compensation Available on 200 Q 2 kQ and 20 kQ ranges DC Current Shunt Resistor 100 Q for 200 uA 2 mA 1 Q for 20 mA 200 mA 0 01 Q for 2 A 10A Auto Zero OFF Operation typical value Following instrument warm up at the environment temperature 1 C and 5 minutes add 0 0001 range 2 uV error for DCV and 2 mQ error for resistance Settling Considerations Reading settling times are affected by source impedance cable dielectric characteristics and input signal changes The default measurement delay can ensure the correctness of the first reading for most measurements Measurement Considerations Teflon or other high impedance low dielectric absorption wire insulation is recommended for these measurements 6 2 M300 User s Guide Chapter 6 Specifications AC Characteristics Accuracy Specifi
62. When used as digital input ports you can read the real time status of each bit of the ports and read the digital signal automatically by configuring the DIO channels into the scan list via the channel configuration guide or press Control and use the left right direction keys to select the multifunction module and the press Config DIO channel configuration guide includes Measurement Configuration Alarm Configuration and Advanced Configuration For alarm configuration refer to Alarm Configuration This section introduces the parameters to be set in Measurement Configuration and Advanced Configuration for the DIO channels 1 Chan No Enter the first step of the channel configuration guide Use the up down direction keys to select Chan No and use the numeric keyboard to input or use the left right direction keys to select the desired channel number Generally the DIO channels are numbered S01 to S04 where S denotes the slot number of the multifunction module 2 Function After selecting the correct channel number Function is fixed to DIO and users cannot modify it 3 Status Use the up down direction keys to select Status and use the left right direction keys to select the desired status READ configure the current DIO channel to the scan list At this point this channel can only be used as a digital input port During a scan the instrument can read the digital signal of this chann
63. alue or the count of the TOT channel is greater than HI an alarm will be generated from the corresponding alarm channel LO when the measurement value is lower than LO an alarm will be generated from the corresponding alarm channel OFF disable the alarm parameters only for DIO channel ON enable the alarm configuration Note that you need to select the alarm channel set the alarm condition and alarm value only for DIO channel Select the Alarm Channel Use the up down direction keys to select Channel and use the left right direction keys to select the desired alarm channel Alarml select Alarml as the alarm channel of the current channel Alarm2 select Alarm2 as the alarm channel of the current channel e Alarm3 select Alarm3 as the alarm channel of the current channel Alarm4 select Alarm4 as the alarm channel of the current channel Set the Alarm Limits only for multiplexer channel and TOT channel After setting the alarm mode to HI LO or HI LO you can use the up down direction keys to select HI and or LO to set the alarm limits HI use the numeric keyboard to input the desired value and press OK For TOT channel the range of HI is from 0 to 27 1 4294967295 LO use the numeric keyboard to input the desired value and press OK Note f the current alarm mode is HI LO the value of HI must not be lower than the current value of LO Set the Alarm Condition only for DI O channe
64. annel Copy Channel copy can copy the configuration of a channel source channel to another channel or multiple channels target channels of the same type After that the following settings of the source channel are copied to the target channels automatically 1 Measurement Configuration include measurement function range etc 2 Scaling Setting on off status of the scaling function and scaling coefficients 3 Alarm Setting on off status of the alarm function alarm channel number and alarm limits 4 Advanced Configuration include integration time AC filter etc Press Channel to enter the channel copy interface This interface is divided into the upper and lower areas Switch switch the cursor to select desired area This menu is available only when the current scan list has 2 or more channels Upper Area use the channel selected as the source channel You can select the target channels Lower Area use the channel selected as the target channel You can select the source channel Done copy the configuration of the source channel selected to the target channel selected Return return to the upper level menu 3an 101 104 Switch Done Return Figure 2 10 Channel Copy Steps of Channel Copy After entering the channel copy interface the channel selected is fixed as the source channel and the upper area is selected by default Use the left right direction keys to select the desired target channels
65. annels of the current module are added to the scan list this operation is invalid 2 52 M300 User s Guide Chapter 2 Front Panel Operations RIGOL 5 Bank Separate Bank J oined By default this module is in Bank J oined state namely the bank switch in Figure 2 17 is closed and the control interface is as shown in Figure 2 18 At this point this module can only be used for 2 wire connection Press Page Dn gt Bank Sepa to execute the Bank Separate operation namely disconnect the bank switch in Figure 2 17 and the control interface is as shown in Figure 2 19 At this point this module is divided into two independent banks and can be used for 4 wire connection When it is used for 4 wire resistance measurement the instrument pairs channel n and channel n 16 automatically wherein n is an integer between 1 and 16 Channel n is used to connect the source terminal of DMM and channel n 16 is used to connect the sense terminal of DMM 25 26 27 28 29 30 31 32 uU u M db t oo t oo rririzrzizrizre Jrzizrzmr zmJ db FTETEFREFEFEFETERE IE L H L H L ia 3 2 9 H L H L H L Pagelp Bank Joined Return Figure 2 19 MC3132 Control Interface Bank Separate M300 User s Guide 2 53 RIGOL Chapter 2 Front Panel Operations To Control MC3164 MC3164 is a 64 channel single ended multiplexer MUX 64 for short MC3164 is divided
66. arched successfully are shown under the RIGOL Online Resource catalog GPIB USB TMC RS232 Verify All RIGOL TECHNOLOGIES INC M300 User s Guide 4 9 RIGOL Chapter 4 Remote Control 5 Communicate test Right click the resource name M300 ASRL1 INSTR and select SCPI Panel Control to open the remote command control panel through which users can send commands and read data 4 10 M300 User s Guide Chapter 4 Remote Control RIGOL To Control M300 through LAN 1 Connect the devices Connect M300 to your computer or the LAN of the computer using network cable 2 Configure LAN parameters 1 DHCP mode If the network supports DHCP the DHCP server in the network assigns network parameters IP Address Subnet Mask Gateway and DNS for M300 automatically 2 Auto IP mode When the network doesn t support DHCP the DHCP mode of M300 is disabled or M300 is connected to the PC directly the instrument selects Auto IP mode and acquires the IP address from 169 254 0 1 to 169 254 255 254 and subnet mask 255 255 0 0 automatically 3 Manual IP mode Enable the Manual IP mode and disable the DHCP mode as well as the Auto IP mode to set the network parameters manually If M300 is connected to the PC directly set the IP Addresses Subnet Masks and Gateways for both of the PC and the instrument The Subnet Masks and Gateways of M300 and PC must be the same and the IP Addresses of them must be within the same netw
67. arm Channel Seti osa Heu 2 66 View Switcheri neuerer 2 68 Scan History Information ss sesso decer oie yo See ui Pes e eru on Ex dues 2 68 Channel Status Table nen ee a oa 2 70 Measurement CUrVe iret ce ea eH P na Hv En ena sun 2 70 Channel informations 2 pte Eee S ba 2 72 Store and Recall z ente e Ro os de which manera des 2 13 File Type serere dene Soba does eu e e esis aida cha Vale ERR sa wh E Be FREE Y Rd 2 74 Internal Storage xir en tern tee ere SR e e e Rt Rud Y PE ran 2 15 External Storage t e ee aa t nein klin 2 15 To Save a Filez e Ih 2 76 VO einer 2 77 System SEN Ge ern se E cava d esrb eat 2 78 I O CORTIQUFSEIOTI 20d ott ot it repeto aoa aote do ro uo Fei cie bdo etur EAD 2 86 System nformatlofi eti a HR t ne C d ida 2 92 acq pe ELI 2 92 System Update eode ae dn Ga a ts 2 93 Chapter 3 To Use External DMM 1 eeeeeee eren nnne nnns 3 1 To Connect Instruments ssssseeeen emm nennen 3 2 To Configure the InstrUMent cc Hem 3 4 Scan Control Method ceat e re En ce ode Rl ore ees 3 5 Chapter 4 Remote Control 1 eeneeeeeee nun ener nnns 4 1 User defined Programming sss 4 2 ToU se PC SOFt Wale dose een nre ert bee nee 4 6 To Control M300 through GPIB 4 6 To Control M300 through USB sssssm Hee 4 8 To Control M300 through RS232 sssssseeme meme 4 9 To Control M300 through LAN eee Hm 4 11 Chapter 5 Trou
68. atus Table Press at the front panel again to enter the channel status table interface as shown in Figure 2 34 In this interface you can judge the channel status Yellow channels not added to the scan list is displayed in yellow Green channels added to the scan list is displayed in green Red during a scan an alarm is generated the corresponding channel is displayed in red Note This status table displays the channel status in real time For example the channel is displayed in red after an alarm is generated If the next measurement value is judged as PASS the channel will be displayed in green again 107 Figure 2 33 Channel Status Table Measurement Curve Press at the front panel again and the instrument enters an interface with a blank coordinate This interface can display the measurement curves of the specified channels 1 Switch Press Switch to switch the cursor to select the channel number or the curve By default the channel number is selected At this point you can delete the channel selected or select other channel number by using the left right direction keys Press Switch to switch the cursor to select the curve and use the left right direction keys to view the curve outside of the range of the horizontal axis 2 70 M300 User s Guide Chapter 2 Front Panel Operations RIGOL 2 Add Press Add and use the numeric keyboard to input a channel number in the scan list The instrument starts drawing the
69. automatically The update will take effect after you restart the instrument CAUTI ON Please update the system advised by the technical support of RIGOL and make sure the power supply of the instrument is stable and the instrument is in non operation state M300 User s Guide 2 93 Chapter 3 To Use External DMM RIGOL Chapter 3To Use External DMM You can connect the M300 with an external DMM and configure them in proper way to realize scan and measurement if the DMM module MC3065 is not ordered Any digital multimeter can be connected to the M300 This chapter takes RIGOL DM3068 as an example to illustrate the connection and configuration method Topics in this chapter m To Connect Instruments m To Configure the Instruments m Scan Control Method M300 User s Guide 3 1 RIGOL Chapter 3 To Use External DMM To Connect Instruments 1 Make sure that the DMM module is not installed in the M300 or or not enabled Split the mixed interface of M300 Convert the RS232 Alarm Ext Trig interface of M300 into two 9 pin interfaces RS232 and Alarm Ext Trig using the Mixed Interface Convert Cable accessory Connect the M300 and DM3068 1 Connect the pin 5 and pin 7 of the Alarm Ext Trig interface to the external trigger signal input interface Ext Trig IN of DM3068 2 Connect the pin 6 and pin 7 of the Alarm Ext Trig interface to the VMC output interface VM Complete OUT of DM3068 Note This step is a typica
70. bleshooting uurnnuunnnnnnnnnnnnnnnnnnnnnnn nun nennen nnn 5 1 Chapter 6 Specifications uu 20n20nn00n00nnunnnnn nun nun ann nun eee ann nun nennen 6 1 DC Characteristics si ER E P ERR ln 6 1 XIV M300 User s Guide Contents RIGOL A ChiaracteriStlCS cocco retener ra sew utto b t rei Red 6 3 Frequency and Period Characteristics sssssseee ene 6 6 Temperature Char amp cteristbS rs segeln 6 8 Mod le Specification S a nn etd Cre ad o De Ene ek RR US FER Vui 6 10 MC3120 MC3132 MC3164 MC3324 MC3416 MC3648 sss 6 10 M53 dites E Rie RNC Bears ud emu dm du teas ctor ds TEEN 6 12 General Specifications essen 6 13 Chapter 7 Appendix 1 eeeeeeeeeeesi esee nun nnnn ann nunnnnnannnnnn nennen 7 1 Appendix A Accessories and Options List accen 7 1 Appendix B Definitions of the Pins of the 78 pin Interface of the Plug in Modules ae ERR RU Gn EF vara ne daveb PR TA 7 2 MEZIO ee ead a tarta Mu d Lnd Od CoU ned iUd 7 2 MO3LI32 Sousse rc A ere ert M EI CI IUE 7 3 MUTA Nc tasse eate atenta dabit ie 7 4 gr y Tr 7 5 MEI 5s terne aq hn eR eb eL ea qi EE Fe Ox Aa Mcd 7 6 MCS BSA 2 ns ee asien 7 7 MESBAB m PE 7 8 Appendix B Warranty mmm mene 7 9 Index are ee A ee ute fegato Sc luu iau da gut seuss 1 M300 User s Guide XV Chapter 1 Quick Guide RIGOL Chapter 1Quick Guide This chapter guides users to quickly get familiar with the front rear panels
71. cable to DCV DCI 2WR 4WR TEMP RTD and RTD 4W and SENSOR except FREQ sensor measurement functions When the current function is DCV DCI 2WR 4WR or SENSOR except FREQ sensor enter the fourth step of the channel configuration guide use the up down direction keys to select AZ and the left right direction keys to select ON or OFF e ON When auto zero is enabled the DMM internally disconnects the input signal following each measurement and takes a zero reading Itthen subtracts the zero reading from the preceding reading namely during the measurement the value displayed is the difference between the measurement value and zero reading This prevents offset voltages on the instrument s input circuitry from affecting measurement accuracy e OFF After pressing OFF the auto zero function is disabled The DMM will still take a zero reading automatically each time the function range or integration time is changed and the instrument subtracts this zero reading from the succeeding measurement value Tip Auto zero and offset compensation are mutually exclusive After enabling auto zero offset compensation if currently enabled will be disabled automatically and vice versa M300 User s Guide 2 25 RIGOL Chapter 2 Front Panel Operations Offset Compensation The Offset compensation setting is applicable to the 2WR 4WR TEMP RTD and RTD 4W and SENSOR 2WR sensor and 4WR sensor measurement functions It remo
72. cations 96 RIGOL of reading of range Function Range Frequency Range 24 Hour 90 Day 1 Year Temperature Coefficient Tea C 1 C Tea C 5 C Ter C 5 C OC to Tea C 5 C Teca C 5 C to 50 C True 200 0000mV 3Hz 5Hz 1 00 0 03 1 00 0 04 1 00 0 04 0 100 0 004 RMS AC 5Hz 10Hz 0 35 0 03 0 35 0 04 0 35 0 04 0 035 0 004 Voltage 10Hz 20kHz 0 04 0 03 0 05 4 0 04 0 06 0 04 0 005 0 004 20kHz 50kHz 0 10 4 0 05 0 11 0 05 0 12 0 05 0 011 0 005 50kHz 100kHz 0 55 0 08 0 60 0 08 0 60 0 08 0 060 0 008 100kHz 300kHz 4 00 0 50 4 00 0 50 4 00 0 50 0 20 0 02 2 000000V 3Hz 5Hz 1 00 0 02 1 00 0 03 1 00 4 0 03 0 100 0 003 5Hz 10Hz 0 35 0 02 0 35 0 03 0 35 0 03 0 035 0 003 10Hz 20kHz 0 04 0 02 0 05 4 0 03 0 06 0 03 0 005 0 003 20kHz 50kHz 0 10 4 0 04 0 11 0 05 0 12 0 05 0 011 0 005 50kHz 100kHz 0 55 0 08 0 60 0 08 0 60 0 08 0 060 0 008 100kHz 300kHz 4 00 0 50 4 00 0 50 4 00 4 0 50 0 20 0 02 20 00000V 3Hz 5Hz 1 00 0 03 1 00 0 04 1 00 0 04 0 100 0 004 5Hz 10Hz 0 35 0 03 0 35 0 04 0 35 0 04 0 035 0 004 10Hz 20kHz 0 04 0 04 0 07 0 04 0 08 0 04 0 008 0 004 20kHz 50kHz 0 10 4 0 05 0 12 4 0 05 0 15 0 05 0 012 0 005 50kHz 100kHz 0 55 0 08 0 60 0 08 0 60 0 08 0 060 0 008 100kHz 300kHz 4 00 0 50 4 00 0 50 4 00 4 0 50 0 20 0 02 200 0000V 3Hz 5Hz
73. ce clean and dry Electrostatic Prevention Operate the instrument in an electrostatic discharge protective environment to avoid damage induced by static discharges Always ground both the internal and external conductors of cables to release static before making connections M300 User s Guide Hl RIGOL Proper Use of Battery If a battery is supplied it must not be exposed to high temperature or in contact with fire Keep it out of the reach of children Improper change of battery note lithium battery may cause explosion Use RIGOL specified battery only Handling Safety Please handle with care during transportation to avoid damage to buttons knob interfaces and other parts on the panels IV M300 User s Guide Safety Terms and Symbols RIGOL Terms Used in this Manual These terms may appear in this manual WARNI NG AN Warning statements indicate conditions or practices that could result in injury or loss of life N CAUTI ON Caution statements indicate conditions or practices that could result in damage to this product or other property Terms Used on the Product These terms may appear on the Product DANGER It calls attention to an operation if not correctly performed could result in injury or hazard immediately WARNING It calls attention to an operation if not correctly performed could result in potential injury or hazard CAUTION It calls attention to an operation if not correctly
74. channels HI and LO each and can be used for 2 wire and 4 wire connections When making 4 wire resistance measurement the instrument automatically pairs channel n with channel n 10 wherein n is an integer between 1 and 10 MC3120 is used with the external terminal block M3TB20 which has a built in thermocouple reference junction Terminal Block M3TB20 Channel Switches Bank Switch atO _ OH pm O a e S E Q Source LO Signals 78 Pin to be e Interface Measured 98 Backplane Switches H O Reference 9 V 20 Junction C Mi LO To H DMM N E oO L Sense Figure 2 14 MC3120 Schematic Diagram The control interface of MC3120 is as shown in Figure 2 15 Use the up down and left right direction keys to select the desired channel The channel currently selected is highlighted for example You can perform various operations on the channel currently selected such as open or close the channel 2 48 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Close Card Reset Page Dn Figure 2 15 MC3120 Control Interface Bank Joined If the channel currently selected is not added to the scan list and other channels of the module are not added to the scan list pressing loK can close the channel At this point pressing OK again can open the channel If the channel is added to the scan list pressing OK can close the channel and read the measurement result before
75. connecting external DMM and configuring scan list to scan Chapter 4 Remote Control This chapter introduces how to control the M300 remotely Chapter 5 Troubleshooting This chapter introduces the commonly encountered failures and their solutions Chapter 6 Specifications This chapter provides the performance specifications and general technical parameters of the M300 Chapter 7 Appendix This chapter provides the accessory list of the M300 as well as service and support information M300 User s Guide XI RIGOL Format Conventions in this Manual l Key The front panel key is denoted by the format of Text Box Button Name Bold For example Utility denotes the Utility key 2 Menu The menu is denoted by the format of Character Shading Menu Word Bold For example System denotes the system menu item of the Utility function key 3 Operation Steps The next step of operation is denoted by an arrow gt For example gt System denotes pressing Utility at the front panel and then pressing System User Manuals of This Product The main user manuals of this product include the Quick Guide User s Guide Programming Guide and Data sheet Users can download the desired manual from www rigol com XII M300 User s Guide Contents RIGOL Contents Guaranty and Declaration 1 eeeeeee seien eene nnns l Safety Requirement uuusnaunnnunnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn
76. dule are not added to the scan list pressing OK can close the channel At this point pressing OK again can open the channel If the channel is added to the scan list pressing can close the channel and read the measurement result before closing this channel the instrument opens the channel currently closed Pressing again can read the measurement result again 1 Open Disconnect the channel currently selected The channel number of the channel currently disconnected is displayed in gray for example GE i one or more channels of the current module are added to the scan list this operation is invalid 2 Close Close the channel currently selected Only one channel can be closed each time The channel number of the channel currently closed is displayed in green for example DE it one or more channels of the current module are added to the scan list this operation is invalid 3 Read Read the measurement value of the current channel This operation is only applicable to channels added to the scan list 4 Card Reset Initialize the status of each channel of the current module If one or more channels of the current module are added to the scan list this operation is invalid M300 User s Guide RIGOL Chapter 2 Front Panel Operations To Control MC3324 MC3324 is a mixed multiplexer with 20 voltage channels and 4 current channels The 20 voltage channels from channel 01 to channel 20 are divided into t
77. dung von Kurzschlu im Ger teinneren und Stromschlag betreiben Sie das Ger t bitte niemals in feuchter Umgebung Nicht in explosiver Atmosph re betreiben Zur Vermeidung von Personen und Sachsch den ist es unumg nglich das Ger t ausschlie lich fernab jedweder explosiven Atmosph re zu betreiben Ger teoberfl chen sauber und trocken halten Um den Einflu von Staub und Feuchtigkeit aus der Luft auszuschlie en halten Sie bitte die Ger teoberfl chen sauber und trocken Schutz gegen elektrostatische Entladung ESD Sorgen Sie f r eine elektrostatisch gesch tzte Umgebung um somit Sch den und Funktionsst rungen durch ESD zu vermeiden Erden Sie vor dem Anschlu immer Innen und Au enleiter der Verbindungsleitung um statische Aufladung zu entladen Die richtige Verwendung desAkku Wenneine Batterieverwendet wird vermeiden Sie hohe Temperaturen bzw Feuer ausgesetzt werden Bewahren Sie es au erhalbder Reichweitevon Kindern auf Unsachgem e nderung derBatterie Anmerkung Lithium Batterie kann zu einer Explosion f hren VerwendenSie nur von RIGOL angegebenenAkkus Sicherer Transport Transportieren Sie das Ger t sorgf ltig Verpackung um Sch den an Bedienelementen Anschl ssen und anderen Teilen zu vermeiden M300 User s Guide VII RIGOL Sicherheits Begriffe und Symbole Begriffe in diesem Guide Diese Begriffe k nnen in diesem Handbuch auftauchen WARNI NG Die Kennzeichnung WARNING beschreibt
78. e 1 13 Subnet Mask sssssss 2 91 View Switch 2 68 Sweep Count 2 3 Voltage Selector 1 9 System Configuration 2 74 2 M300 User s Guide
79. e To use this mode to set the IP address you need to disable the DHCP and AutolP modes Otherwise the instrument will use the DHCP or AutolP mode automatically e IP Address The format of IP address is nnn nnn nnn nnn for example 172 16 2 13 After enabling the ManuallP mode use the up down direction keys to select IP Addr and use the numeric keyboard to set the desired IP M300 User s Guide Chapter 2 Front Panel Operations RIGOL address e Subnet Mask The format of subnet mask is nnn nnn nnn nnn for example 255 255 255 0 After enabling the ManuallP mode use the up down direction keys to select SubMask and use the numeric keyboard to set the desired subnet mask e Gateway The format of gateway is nnn nnn nnn nnn for example 172 16 2 1 After enabling the ManuallP mode use the up down direction keys to select Gateway and use the numeric keyboard to set the desired gateway e DNS The format of DNS address is nnn nnn nnn nnn for example 202 96 199 133 After enabling the ManuallP mode use the up down direction keys to select DNS and use the numeric keyboard to set the desired DNS address Generally users do not need to set the DNS address of the network and this parameter setting can be ignored 3 VISAAddr 6 VISA Virtual Instrument Software Architecture is an advanced application programming interface developed by NI National Instruments for communication with various instrument buses using the same method re
80. e asking whether you want to restore all the settings to the status before the last power off will be displayed Press OK Preset Settings Use the up down direction keys to select Preset and use the left right direction keys to select ON or OFF the default is OFF When ON is selected the prompt message asking whether you want to restore all the settings to the preset status will be displayed Press OK The preset settings are as shown in the table below Table 2 6 Preset Settings Parameter Factory Setting Scan Configuration Scan List No change Reading Memory Cleared Max Min Average and SDEV Cleared Trigger Mode No change Scan Interval No change Number of Scans No change M300 User s Guide 2 81 RIGOL Chapter 2 Front Panel Operations Measurement Configuration Function No change Range No change Integration Time No change Input Impedance No change Channel Delay No change TOT Read Mode No change TOT Trigger Mode No change Scaling Configuration Scaling Configuration No change A No change B No change C No change x1 No change Unit No change Alarm Configuration Alarm Mode NONE Alarm Channel Alarm1 HI 0 LO 0 Alarm Channel Configuration Alarm Queue No change Output Status No change Output Mode No change Alarm Output No change Channel Monitor Monitor in Progress Stop Module Control MC3
81. e configuration list interface The arithmetic between the first start value and the second start value conforms to the first fitting function the arithmetic between the second start value and the third start value conforms to the second fitting function from the last start value on the arithmetic conforms to the last fitting function e Add This menu is used to add the start value and fitting function Press this softkey to open the any sensor configuration interface Start Value use the up down direction keys to select Start and use the numeric keyboard to input the desired value between 1E 15 and 1E 15 Fitting Function Coefficients use the up down direction keys to select A B and C respectively and use the numeric keyboard to input the desired values between 1E 15 and 1E 15 M300 User s Guide 2 33 RIGOL Chapter 2 Front Panel Operations Start Value and Coefficient Setting Explanation When setting the start value and fitting function coefficients A B and C press any key on the numeric keyboard to open the dialog box in the form below e Sign use the left direction key to move the cursor to the sign bit and press on the numeric keyboard to switch the sign to or Unit use the right direction key to move the cursor to the unit bit and use the up down direction keys to select the desired unit The units available are related to the sensor type currently selected The unit coeffici
82. e device Note The file with the suffix csv can only be stored in the USB storage device Internal Storage The internal C disk holds 256 Mbytes storage space and can store four types of files including system configuration measurement configuration and measurement data only dat file Note When executing internal storage for the first time please format C disk External Storage The D disk can store four types of files including system configuration measurement configuration mirror configuration and measurement data dat or csv file according to its capacity Before using the external storage insert the USB storage device into the USB Host interface at the front panel When the instrument detects the USB storage device will be displayed in the screen status bar CAUTI ON When executing operations in the D disk storage area do not remove the USB storage device M300 User s Guide 2 75 RIGOL Chapter 2 Front Panel Operations To Save a File When the current directory can store the specified type of file press Save to enter the filename input interface as shown in Figure 1 6 The length of the filename cannot exceed 12 characters Please refer to the introduction in To Input Filename to input the filename and press OK to finish the save operation 2 76 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Utility Press at the front panel to enter the utility interface in wh
83. e mutliplexer module for measurement make sure that both the amplitude of the signal under measurement and the commom mode voltage are no greater than the maximum input voltage of the module When the multiplexer module is used to measure an object with high power output ability such as battery and power supply to avoid damaging the relay you are recommended to connect a fuse in serial when making connections as shown in the figure below Channel Terminal Switches Box TB32 H e L 78 Pin ui Interface e e M300 DMM Module Mainframe MC3065 1 22 M300 User s Guide Chapter 1 Quick Guide Table 1 2 Measurement Connections RIGOL DCV ACV FREQ PERI OD Sensor DCV FREQ E Signal to be measured DCI ACI Sensor DCI ES Signal to be measured Resistance to be measured Thermocouple Thermocouple under test H L H L CH21 CH1 Note Only available for the channel 21 to channel 24 of MC3324 2WR RTD Thermistor 4WR RTD 4W Sensor 4WR Sensor 2WR Source Resistance to be measured H L CH17 Sense Note Only available for all the channels of MC3120 and MC3132 as well as the channel 1 to channel 20 of MC3324 WARNI NG A To avoid electric shock the connecting wire should meet the maximum curent requirement To avoid damaging the module make sure that all the input signals are within the rated
84. easurement rate The maximum output rate is 1000 times per second In external trigger mode Pin 6 accepts external trigger signal The trigger can have a pulse width equal to or greater than 2us with specified edge To configure press Config Select External trigger mode and select Rising or 1 10 M300 User s Guide Chapter 1 Quick Guide RIGOL 10 Falling in Edge item If the external trigger signal is continuous pulse please make sure the pulse period is larger than 100ps GPI B The M300 conforms to the EEE 488 2 standard Via this interface the M300 can be controlled remotely via a PC with a GPIB interface Power Socket The M300 can accept two kinds of AC power supplies Use the power cord provided in the accessories to connect the AC power supply to the instrument via this socket CAUTI ON Before connecting the AC power supply select the correct voltage scale using the voltage selector Power Fuse The specification of the fuse of M300 is AC 250V T3 15 A If a new fuse is required please refer to To Replace the Fuse CAUTI ON Please use the specified power fuse and make sure that the fuse seat is not short circuited to avoid electric shock and fire M300 User s Guide 1 11 RIGOL Chapter 1 Quick Guide Power On I nspection 1 Connect the power 1 Adjust the power voltage selector at the rear panel according to the power supply voltage M300 supports two types of AC power If your supply v
85. ection keys to select Delay and use the left right direction keys to select Auto or set a specific delay value e Auto In auto mode the instrument calculates the channel delay according to the function range integration time and AC filter of the current channel automatically At this point the channel delay time changes each time the configuration of the current channel is changed M300 User s Guide 2 27 RIGOL Table 2 4 Auto Delay Time Chapter 2 Front Panel Operations a DCV DCI TEMP TC SENSOR DCV DCI all the ranges Integration Time Channel Delay gt 1 PLC 2 0 ms lt 1 PLC 1 0 ms b 2WR 4WR TEMP THER RTD RTD 4W SENSOR 2WR 4WR Channel Delay Channel Delay Range Integration Time gt 1 Range I ntegration Times1 PLC PLC 200 Q 2 0 ms 200 Q 1 0 ms 2 kQ 2 0 ms 2 kQ 1 0 ms 20 kQ 2 0 ms 20 kQ 1 0 ms 200 kQ 25 ms 200 kQ 20 ms 1 MQ 30 ms 1 MQ 25 ms 10 MQ 200 ms 10 MQ 200 ms 100 MQ 200 ms 100 MQ 200 ms Notel For DCV TEMP 2WR 4WR and SENSOR except FREQ sensor if the current integration time is set to User the instrument calculates the channel delay according to the condition Integration Time gt 1 PLC automatically c ACV ACI all the ranges AC Filter Channel Delay 3 Hz slow 7 05 20 Hz medium 1 0s 200 Hz fast 120 ms FREQ PERI OD SENSOR FREQ AC Filter Channel Delay 3 Hz slow
86. egen Verwenden Sie einen geeigneten berspannungsschutz Stellen Sie sicher da keinerlei berspannung wie z B durch Gewitter verursacht das Ger t erreichen kann Andernfallsbestehtf r den Anwender die GefahreinesStromschlages Nicht ohne Abdeckung einschalten Betreiben Sie das Ger t nicht mit entfernten Geh use Abdeckungen Betreiben Sie das Ger t nicht ge ffnet Der Betrieb mit offenen oder entfernten Geh useteilen ist nicht zul ssig Nichts in entsprechende ffnungen stecken L fter z B Passende Sicherung verwenden Setzen Sie nur die spezifikationsgem en Sicherungen ein Vermeiden Sie ungesch tzte Verbindungen Ber hren Sie keine unisolierten Verbindungen oder Baugruppen w hrend das Ger t VI M300 User s Guide RIGOL in Betrieb ist Betreiben Sie das Ger t nicht im Fehlerfall Wenn Sie am Ger t einen Defekt vermuten sorgen Sie daf r bevor Sie das Ger t wieder betreiben dass eine Untersuchung durch RIGOL autorisiertem Personal durchgef hrt wird J edwede Wartung Einstellarbeiten oder Austausch von Teilen am Ger t sowie am Zubeh r d rfen nur von RIGOL autorisiertem Personal durchgef hrt werden Bel ftung sicherstellen Unzureichende Bel ftung kann zu Temperaturanstiegen und somit zu thermischen Sch den am Ger t f hren Stellen Sie deswegen die Bel ftung sicher und kontrollieren regelm ig L fter und Bel ftungs ffnungen Nicht in feuchter Umgebung betreiben Zur Vermei
87. el OFF do not configure the current DIO channel to the scan list At this point this channel can be used as both digital input you can read the digital input signal of this port manually and digital output port you can edit the digital output signal of this port Please refer to the introduction in To Control MC3534 4 Bits Use the up down direction keys to select Bits and use the left right direction keys to select the desired number of bits Channels S01 to S04 are all 8 bit ports You can configure the 4 channels to be independent 8 bit channels You can also combine S01 and S02 as well as S03 and S04 respectively and configure them to 16 bit Finally all 4 channels can be combined and configured to be 32 bit input M300 User s Guide 2 35 RIGOL Chapter 2 Front Panel Operations 8 bit S01 S02 S03 S04 XXXXXXXX XXXXXXXX XXXXXXXX XXXXXXXX 16 bit 02S01 504503 XXXXXXXXXXXXXXXX XXXXXXXXXXXXXXXX 32 bit 504503502501 XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX Note n the 16 bit S02S01 S02 is the upper 8 bit bank In the 16 bit 504503 S04 is the upper 8 bit bank In the 32 bit 504503502501 S04 is the upper 8 bit bank Type When the channel status is READ use the up down direction keys to select Type and use the left right direction keys to select the desired level TTL CMOS5V CMOS3 3V CMOS2 5V or User If User is currently selected use the up down direction keys to s
88. elect Level and Threshold respectively and use the numeric keyboard to input the desired values The range of level is from 2 000 V to 5 000 V and the default is 5 000 V e The range of threshold is from 0 500 V to 3 500 V and the default is 2 500 V When setting the level and threshold they should fulfill the relation equation Level gt Threshold 0 5 V When the input signal Threshold 0 3 V it is regarded as logic 0 when the input signal gt Threshold 0 3 V it is regarded as logic 1 Coding Enter the fourth step of the channel configuration guide By default Coding is selected and use the left right direction keys to select Binary Decimal or Hex The coding mode selected will be used in both the digital input and output of this channel Key Points 1 When the DIO channels are used as digital input channels they do not have to be part of the scan list and the DMM module does not have to be enabled to generate an alarm when the reading matches or dismatches the specific data pattern DIO channels added into the scan list cannot be used as digital output channels M300 User s Guide Chapter 2 Front Panel Operations RIGOL Totalizer The multifunction module provides four 32 bit totalizers TOT which can count TTL pulses at a 10 MHz rate for high speed TOT channels or 100 kHz rate for normal TOT channels You can manually read the TOT count or you can read the TOT count automatically by con
89. ell as clear key backlight and operation prompts The M300 is extremly flexible and easy to use For remote control versatility it comes with RS232 USB LAN and GPIB interfaces standard Main Features Up to 320 switch channels per mainframe with very low cost per channel e Can run independently without PC USB data logging Interval scanning with storage of up to 100 000 time stamped readings e 8 modules supported e 6 digit DMM can be inserted into any slot It also supports multiple measurement functions including DCV DCI ACV ACI 2WR 4WR PERIOD FREQ TEMP Thermocouple Thermistor and RTD and any sensor e Standard multiple communication interface USB Device USB Host GPIB LAN LXI Core 2011 Device and RS232 e Standard SCPI commands for remote programming capability Math statistics function AVG MAX MIN and SDEV Friendly interface Built in help system e Multiple system language e 4 3 inch color LCD PC control and analysis software X M300 User s Guide RIGOL Document Overview Main topics in this manual Chapter 1 Quick Guide This chapter introduces the front panel rear panel user interface and plug in modules of the M300 as well as the precautions when using it for the first time Chapter 2 Front Panel Operations This chapter introduces in details the front panel operation methods and functions of M300 Chapter 3 To Use Extemal DMM This chapter illustrates the method of
90. ents include n 10 u 10 m 10 _ 1 K 10 M 10 G 10 T 10 and P 10 e Number use the left right direction keys to move the cursor to the number bit and use the numeric keyboard to input the desired value Sign 25 V nit Number After configuring a group of start value and fitting function press Done to return to the arithmetic curve configuration list interface Press Add again to configure the next group of start value and fitting function Delete Use the up down direction keys to select a group of start value and fitting function in the arithmetic curve configuration list interface and press Delete to delete the configuration selected e Modify Use the up down direction keys to select a group of start value and fitting function in the arithmetic curve configuration list interface and press Modify to enter the any sensor configuration interface At this point you can modify this group of configuration e Done Finish the arithmetic curve configuration and return to the measurement configuration interface e Return Give up the configuration currently edited and return to the measurement configuration interface 2 34 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Digital Input The multifunction module has four non isolated 8 bit digital input output DI O ports which you can use as digital output ports refer to Module Control and digital input ports
91. erates an alarm during the scan Draw the measurement value curve of the specified channel in the scan list Channel Information 1 The cycles of all the relays of each module 2 Alarm information 3 Error information Note Display the first 100 alarm messages of each alarm channel and the alarm messages generated later will be discarded The alarm messages viewed are cleared Notel Display the first 20 error messages generated and the error messages generated later will be discarded The error messages viewed are cleared M300 User s Guide 1 27 RIGOL Control MC3065 DMM module con not be controlled Multiplexer Module 1 Disconnect the channel selected 2 Close the channel selected 3 Read the measurement value of the channel selected 4 Module reset 5 Bank Joined used for 2 wire measurement 6 Bank Separate used for 4 wire measurement MC3120 MC3132 MC3324 Single ended Multiplexer module 1 Disconnect the channel selected 2 Close the channel selected 3 Read the measurement value of the channel selected 4 Module reset MC3164 Actuator Matrix Switch MC3416 1 Disconnect the channel selected MC3648 2 Close the channel selected 3 Module reset Multifunction Module 1 DIO channel Read the input value of the channel selected Edit the output value of the channel selected 2 TOT channel Read
92. erature It features in high measurement accuracy and stable performance The instrument can measure the resistance using 2 wire RTD or 4 wire RTD 4W method The internal software conversion program converts the resistance measured to the corresponding temperature For the reference table about each kind of RTD please refer to the related standards e RO After setting the temperature sensor type to RTD or RTD 4W use the up down direction keys to select RO and use the numeric keyboard to input the desired resistance The range is from 49 Q to 2100 Q and the default is 100 Q ALPHA Temperature coefficient After setting the temperature sensor type to RTD or RTD 4W use the up down direction keys to select ALPHA and use the left right direction keys to select the desired coefficient 0 00385 0 00389 0 00391 or 0 00392 e Precision After setting the temperature sensor type to RTD or RTD 4W use the up down direction keys to select Precision and use the left right direction keys to select 1 0 1 or 0 01 M300 User s Guide 2 31 RIGOL Chapter 2 Front Panel Operations Tip For the temperature measurement principles of thermocouple RTD and thermister refer to the Temperature Measurement Technique Application Instruction contact RIGOL to acquire the manual 2 32 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Any Sensor Measurement Parameters Any sensor measurement supports DCV 2WR
93. erface as shown in Figure 2 13 You can control the status of each module except the DMM module The 5 column diagrams in the module control interface represent the 5 slots The corresponding number such as slot1 and type of the module inserted into the slot are displayed above and below the column diagram respectively and the name of the module is displayed in the middle of the column diagram Use the left right direction keys to select the desired module slot without module and slot with the DMM module cannot be selected The module currently selected is displayed in blue Press Single to enter the control interface of the module currently selected Note n the module control interface the model and name of the module are related to the module currently inserted into the instrument Module Model Module Name Slot Number Single Figure 2 13 Module Control Interface Explanation When the multiplexer is used as the DMM input channel only one channel can be closed at each time For example use channel 1 of MC3132 to measure temperature and channel 2 to measure DCV The instrument first closes channel 1 to measure temperature and then disconnects channel 1 and closes channel 2 to measure DCV namely the break before make feature M300 User s Guide 2 47 RIGOL Chapter 2 Front Panel Operations To Control MC3120 MC3120 is a 20 channel multiplexer MUX 20 for short MC3120 is divided into two banks of 10 two wire
94. ew the measurement value of the specified channel Note When the count reaches to 100000 the new readings will overwrite the first readings stored Save Chan Data Search Figure 2 32 Scan History Information Read the measurement data stored in the internal or external memory Press Read and the instrument enters the store and recall interface the file type is set to Meas Data automatically Please refer to Store and Recall to select the desired file to read 2 Save Save the measurement data into the internal or external memory Press Save and the instrument enters the store and recall interface the file type is set to 2 68 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Meas Data automatically Please refer to Store and Recall to save the data 3 Channel Data Select a channel and press Chan Data to view the measurement values of this channel These values contain time stamps Press Search and use the numeric keyboard to input the point number to quickly locate the data of the specified point Press Return to return to the scan history information interface 4 Search Press Search and input the desired channel number to display the specified channel in the interface When a larger number of channels are contained in the scan list this softkey can be used to quickly locate the desired channel to view the channel information M300 User s Guide 2 69 RIGOL Chapter 2 Front Panel Operations Channel St
95. figuring the TOT channels to a scan list via the channel configuration guide The TOT channel configuration guide contains Measurement Configuration Alarm Configuration and Advanced Configuration For alarm configuration refer to Alarm Configuration This section introduces the parameters to be set in Measurement Configuration and Advanced Configuration for the TOT channels 1 Chan No Enter the first step of the channel configuration guide Use the up down direction keys to select Chan No and use the numeric keyboard to input or use the left right direction keys to select the desired channel number Generally the TOT channels are numbered S05 to 508 where S denotes the current slot number of the multifunction module S05 and S06 are high speed TOT channels and can count TTL pulses at a 10 MHz rate S07 and 08 are normal TOT channels with 100 kHz count rate 2 Function After selecting the correct channel number Function is fixed at TOT and users cannot modify it 3 Status Use the up down direction keys to select Status and use the left right direction keys to select the desired status READ configure the current TOT channel to the scan list During a scan the instrument can read the current TOT count OFF do not configure the current TOT channel to the scan list You can start and stop the counting or read the count of the TOT channel manually Please refer to the introduction in To Control MC3534
96. from slot 1 to slot 5 automatically Operations on any channels of the module of which the channel is in the scan list are forbidden You can press and hold Run Stop at the front panel to stop the scan but after the scan has stopped it cannot be resumed When a new scan is executed all the readings in the memory will be cleared M300 User s Guide 2 11 RIGOL Chapter 2 Front Panel Operations Channel Configuration After creating a new scan list and configuring proper parameters such as the scan sweep and trigger mode you need to add channels into the scan list and configure the measurement scaling alarm and advanced parameters of the channels The M300 provides special channel configuration guide for you to easily finish the channel configuration The flow diagram of the channel configuration guide is as shown in the figure below Measurement Configuration Scaling Configuration 2 Alarm Configuration Advanced Configuration Figure 2 3 Flow Diagram of Configuration Guide Notel For multiplexer channels using the SENSOR function or the DIO channels and TOT channels of the multifunction module the configuration guide does not contain Scaling Configuration Note For the multiplexer channels if the DMM module is not inserted or it is inserted but is not enabled you can connect an external DMM for scan and measurement At this point the configuration guide does not contain Scaling Configurat
97. gardless of the type of the instrument interface GPI B USB LAN Ethernet or RS232 The instrument communicating with it via the GPIB USB LAN Ethernet or RS232 interface is called Resource The VISA address also called VISA descriptor refers to the resource name and describes the accurate name and position of the VISA resource For example if LAN interface is currently used for communicating with the instrument the VISA descriptor is TCPIP0 172 16 2 13 I NSTR MAC The physical address also called MAC Media Access Control address or hardware address is used to define the position of the network device For an instrument the MAC address is unique When distributing an IP address for the instrument the MAC address is always used to distinguish the instrument The MAC address 48 bits namely 6 bytes is usually denoted by a six section hexadecimal number for example 00 14 0E 42 12 CF Default In the LAN parameter setting interface press Default to restore the network parameters to their defaults Apply In the LAN parameter setting interface press Apply to configure the network parameters using the current settings M300 User s Guide 2 91 RIGOL Chapter 2 Front Panel Operations System I nformation Press gt Info to view the system information of the instrument such as the model serial number hardware version and software version Detect Press Utility gt Detect to enter the interface as shown in the fi
98. gate signal is not detected in both the terminals counting is always enabled Read Enter the fourth step of the channel configuration guide Use the up down direction keys to select Read and use the left right direction keys to select the desired setting READGRST read the count and reset the totalizer without losing any counts READ only read the count Key Points 1 2 The maximum count is 4294967295 2 1 The count rolls over to 0 after reaching the maximum allowed value You can configure the instrument to generate an alarm in the specified alarm channel when a specific count is reached for TOT channel that is not added to the scale list the channel status is OFF Please refer to Alarm Configuration to configure the desired parameters You can execute single channel monitor on a TOT channel even if the channel is not part of the scan list the channel status is OFF S05 and S06 are high speed TOT channels with 10 MHz count rate they only receive standard TTL pulse or 3 3 V to 5 V square signal and their thresholds are decided by the internal circuit 507 and S08 are normal TOT channels with 100 kHz count rate They can receive analog signal or TTL pulse and their thresholds are set by users the range is from 12 V to 12 V M300 User s Guide Chapter 2 Front Panel Operations RIGOL Configuration Copy The configuration copy function supports module copy channel copy and extended copy
99. gative signs Used to input parameter value date and time CI C3 CI Use this key to open the previous or next page of the current interface Note These keys are different from the menu page up down keys Page Up and Page Dn Direction keys and confirmation key If a USB storage device is currently connected to the instrument press this key to store the current content on the screen in USB storage device in bmp format M300 User s Guide 1 5 RIGOL Chapter 1 Quick Guide oN When the instrument is in local mode pressing this key can lock the front panel At this point all the keys at the front panel except this key become invalid Press and hold this key to unlock the front panel When the instrument is in remote mode press this key to switch to local mode Store and recall the system configurarion measurement configuration and measurement data etc e Store the file in the internal non volatile memory or external USB storage device Read the file stored Store NN in Configure the system related parameters Set various parameters such as the time date system language and display brightness Configure the remote interface parameters RS232 LAN GPIB and USB Query the system information such as the model and serial number Execute self test and view the information of each module such as the name and model of the module e System u
100. gers the instrument exits the wait for trigger state During the scan the instrument stores the measurement data scaling data and alarm data in the non volatile memory automatically 2 10 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Alarm Trigger To use the alarm trigger mode configure the alarm by referring to Alarm Configuration and select alarm trigger mode then select an alarm channel The instrument executes a scan when an alarm is generated in the specified alarm channel n alarm trigger mode if the number of scans is set to Infinite after you press at the front panel the instrument enters the wait for trigger state Then the instrument executes a scan when an alarm is generated in the specified alarm channel Press and hold Run Stop and the instrument will exit the wait for trigger state If the number of scans is a specific value after you press Run Stop at the front panel the instrument enters the wait for trigger state Then the instrument executes a scan when an alarm is generated in the specified alarm channel After executing the specified number of alarm triggers the instrument exits the wait for trigger state During the scan the instrument stores the measurement data scaling data and alarm data in the non volatile memory automatically Explanation The instrument triggers a scan when the trigger condition is met and scans the channels in the scan list in ascending order
101. gger count trigger mode interval for auto trigger mode and trigger edge type for external trigger and the channel configurations including measurement configuration scalling configuration alarm configuration and advanced configuration are overwritted To Set RS232 Parameters You can convert the RS 232 Alarms Ext Trig interface at the rear panel to two 9 pin interfaces using the Mixed Interface Convert Cable accessory Wherein one is a 9 pin male interface used as a standard RS232 interface the definitions of the pins are as shown in the table below The 9 pin female interface is used for alarm output external trigger signal input etc nn amp Ww TN H Pin Name Definition 1 DCD Data Carrier Detect 2 RXD Received Data 3 TXD Transmitted Data 4 DTR Data Terminal Ready 5 SGND Signal Ground 6 DSR Data Set Ready 7 RTS Request To Send 8 CTS Clear To Send 9 RI Ring Indicator RS232 Interface Connect the RS232 interface with the PC or data terminal equipment DTE device using a RS232 cable and set the interface parameters such as the baud rate and M300 User s Guide 2 87 RIGOL Chapter 2 Front Panel Operations parity to match the PC or terminal device At this point you can control the instrument remotely The measurement result of the instrument can be output to serial port receiving devices via this interface Press gt 1 0 gt RS232 to enter the interface
102. gure below The 5 column diagrams represent the 5 slots The corresponding slot number such as Slot1 and the model of the module inserted into the slot are displayed below and above the column diagram respectively and the name of the module is displayed in the middle of the column diagram Use the left right direction keys to select the desired module slot without module and slot with the DMM module cannot be selected and the module currently selected is displayed in blue Module Model Module Name Slot Number Slot1 Slot2 Slot3 Slots Self test View Re test Return Figure 2 38 Detect Interface l Self test Select the corresponding slot and press this softkey to test the module 2 View Press this softkey to view the information of the module currently selected including the function model serial number and version 3 Re test If the module is inserted or removed when the instrument is running select the corresponding slot and press this softkey to test the module of this slot again to acquire the latest status 2 92 M300 User s Guide Chapter 2 Front Panel Operations RIGOL System Update M300 can update the system by reading the update file stored in the USB storage device Insert the USB storage device with the update file After the instrument recognizes the USB storage device press gt Update Now you can operate the instrument according to the prompts The instrument will read the update file and finishes the update
103. ht direction keys to select Rising or Falling As the External advanced source input and the external trigger input use the same pin the external trigger mode does not support External advanced source Pin 6 Ext Trig Input Channel Advance Input 00000 0000 Alarms Ext Trig Interface 6 Absolute Time Absolute time is defined as the time when the instrument executes the scan in 2 4 M300 User s Guide Chapter 2 Front Panel Operations RIGOL absolute time scan mode When the trigger mode is AbsTime select AbsTime using the up down direction keys and set the desired time using the numeric keyboard The time format is Month Date Hour Minute Second You need to set a valid time The ranges of the parameters are as follows Month 01 to 12 Date 01 to 31 30 29 or 28 Hour 00 to 23 Minute 00 to 59 Second 00 to 59 The instrument can also ignore some time parameters and trigger when the system time fulfills some of the parameters Use on the numeric keyboard to set the parameters to be ignored and the corresponding position will display For example set the instrument to ignore the Month and Date and trigger at 8 00 every day use the numeric keyboard to set the absolute time to 08 00 00 7 Alarm Channel When the trigger mode is set to Alarm select Channel using the up down direction keys and set the alarm channel to Alarm1 Alarm2 Alarm3 or Alarm4 using the left right
104. ich you can set the system parameters and I O parameters view the system information as well as update the system Time English d ON DMM eee Factory ei c Info Detect Update Figure 2 37 Utility 1 System Press System to set the system parameters such as the date time and language For details refer to System Setting 2 1 0 Press I O to configure the remote interfaces such as RS232 and LAN For details refer to 1 O Configuration 3 Information Press Info to view the instrument serial number software version etc 4 Detect Press Detect to execute self test or view the type model serial number and version of the module in each slot 5 Update Insert the USB storage device with the update file After the instrument recognizes the USB storage device press this softkey the instrument read the update file and finish the update automatically The update takes into effect after restarting the instrument Note Before updating the system make sure the instrument power source is stable and the instrument is in non operation state M300 User s Guide 2 77 RIGOL Chapter 2 Front Panel Operations System Setting Press Utility and the instrument enters the system configuration interface by default or press Utility gt System Date Use the up down direction keys to select Date and use the numeric keyboard to set the system date The format is as follows YYYY MM DD Whe
105. ide 2 45 RIGOL Chapter 2 Front Panel Operations Key Points 1 2 3 For single channel monitor function only one channel can be monitored at a time but you can change the channel being monitored at any time Readings acquired during a monitor are not stored in memory but they are displayed on the screen When the alarm function are applied to the selected channel during monitor all the alarm data is stored in the alarm queue which will be cleared at power off You can monitor a multiplexer channel only if the DMM module is installed and enabled Utility gt DMM gt ON The channel must also be configured to be part of the scan list You can monitor a DIO or TOT channel even if the channel is not part of the scan list Multi channel Monitor Press Monitor gt Multiple to enter the multi channel monitor interface At most 7 channels can be monitored You can add or delete the channels being monitored All Add use the numeric keyboard to input the channel number Delete use the up down direction keys to select the desired channel number and then press Delete Channel Monitor Press IMonitor gt All Moni to enter the all channel monitor interface All channels in the scan list can be monitored M300 User s Guide Chapter 2 Front Panel Operations RIGOL Module Control The M300 provides a user friendly module control interface Press at the front panel to enter the int
106. inals 12 V calibrated voltage MC3648 4x8 Matrix Switch The MC3648 is used to connect multiple devices to multiple points on the device under test You can connect the rows and columns of multiple matrix switches to form relatively larger matrix such as 8x8 and 4x16 the number of crosspoints cannot exceed 160 32 two wire cross points which can connect any combination of inputs and outputs at the same time M300 User s Guide 1 19 RIGOL Chapter 1 Quick Guide To Use the Built in Help System The built in help system of M300 provides the help information about any front panel key and menu To acquire the help information of any front panel key or menu press Help the backlight goes on and Help is displayed in the status bar in the user interface and then press the desired key The built in help interface is as shown in the figure below Press OK to exit gt 1 Utility Auto Set the system information and interface information view the instrument and module information perform module s Figure 1 9 Built in Help System 1 20 M300 User s Guide Chapter 1 Quick Guide RIGOL Measurement Connections Multiple signals to be measured are connected to the plug in modules through the external terminal blocks and then switched to the DMM module or external DMM M300 provides 7 kinds of terminal blocks The models of the terminal blocks and corresponding modules are listed in the table below This section introd
107. installation CD BP or floppy disk insert it now What do you want the wizard to do O Install the software automatically Recommended Install from a list or specific location Advanced Click Next to continue 2 e Found New Hardware Wizard Please choose your search and installation options ey Search for the best driver in these locations Use the check boxes below to limit or expand the default search which includes local paths and removable media The best driver found will be installed Search removable media floppy CD ROM Include this location in the search G MM L4N BROADCOM G Don t search will choose the driver to insta 3 Choose this option to select the device driver from a list Windows does not guarantee that the driver you choose will be the best match for your hardware M300 User s Guide 4 3 RIGOL Chapter 4 Remote Control Found New Hardware Wizard Select the device driver you want to install for this hardware Select the manufacturer and model of your hardware device and then click Next If you co have a disk that contains the driver you want to install click Have Disk Show compatible hardware Model C TE Test and Measurement Device a 5 Es amp This driver is not digitally signed Tell me why driver signing is important 6 Cs eo Found New Hardware Wizard Please wait while the wiid installs the software en e USB Test and Measurement Device IVI EH p 4 4 M
108. into two banks of 32 single ended channels HI only the 32 channels has a common LO terminal each and cannot be used for 4 wire connection MC3164 is used with the external terminal block M3TB64 which has a built in thermocouple reference junction Terminal Block M3TB64 oO 5 Channel Switches Signals M 78 Pin to be 5 e Interface Measured l Backplane Switches H DMM O u az j Source N64 o fi EN 2 eference k unction V Lowe comt O Figure 2 20 MC3164 Schematic Diagram The control interface of MC3164 is as shown in Figure 2 22 Press Page Up and at the front panel to open the previous and next pages of the control interface Use the up down and left right direction keys to select the desired channel The channel currently selected is highlighted for example oi You can perform various operations on the channel currently selected such as disconnect and close the channel 2 54 M300 User s Guide Chapter 2 Front Panel Operations i pos 4 oo H L H L H L H H L H L H L H L 3 9395 BS BS 93 9 BSS BS BS BS 83 83 93 8 958385 05 BS BS 3S co Close oO I D E RIGOL 17 18 19 20 21 22 23 24 8 958385 SS BS BS 92 Se Mog d nano nd ee nd wm C mn c4 papa nd gen ox Card Reset Return Figure 2 21 MC3164 Control Interface If the channel currently selected is not added to the scan list and other channels of the mo
109. ion The second step of the channel configuration guide configure the scaling parameters 2 OFF mon Fitting A boxTP 2 B boxT C 0 000000 F Next Return Figure 2 5 Scaling Configuration Note For multiplexer channels that are using the SENSOR function and the DIO and TOT channels of the multifunction module the configuration guide does not contain scaling configuration Also note that the multiplexer channels that are not using the DMM module the configuration guide also does not contain Scaling Configuration The M300 provides a scaling function This function has three scaling coefficients A B and C for all the readings of the channel selected The formula for the scaling function is as follows Scaling Reading Ax x x1 Bx x x1 C Wherein x is the reading measured x1 is the correction value can be a reading measured at certain time or a value defined by user Substracting x1 from all readings can correct the zero point of the fitting curve You can configure the scaling function for all the multiplexer channels except channels using the SENSOR function or when external DMM is connected Before applying the scaling function you must configure the measurement parameters of the channel After applying the scaling function if the measurement function is modified the scaling function will be disabled and the scaling coefficients will be reset M300 User s Guide 2 15 RIGOL Chapter 2 Front
110. ion and Alarm Configuration and there are some differences in Measurement Configuration and Advanced Configuration Press Config Edit New to enter the channel configuration guide After finishing all the configurations press Done 2 12 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Measurement Configuration The first step in the channel configuration guide select the channel number select the measurement function of the specified channel and configure the basic measurement parameters of the selected function Return Figure 2 4 Measurement Configuration 1 Channel Number Select the desired channel number Generally the channel number is a 3 digit number where the first digit from the left denotes the number of the module slot and the other two digits denote the number of the channel SCC The number of the module slot from 1 to 5 The number of the channel from 01 to 64 The channels available include all the channels of the multiplexer S01 to 32 S64 S24 the DIO S01 to S04 and TOT S05 to S08 channels of the multifunction module After entering the channel configuration guide Chan No is selected by default At this point you can use the left right direction keys to modify the channel number or the numeric keyboard to input the channel number M300 User s Guide 2 13 RIGOL Chapter 2 Front Panel Operations Explanation If you want to configure the channel of the multiplexer
111. ired letter and press OK At this point letters in Filename Input Area or Chinese Character Selecting Area varies e In Chinese input mode when Chinese Character Selecting Area is selected you can use the up down direction keys to open the previous or next page of Chinese Character Selecting Area and use the numberic keyboard to input the number corresponding to the Chinese character to input the desired Chinese character Delete Press this softkey to delete the letter selected by the cursor in Filename Input Area OK Press this softkey to use the letters displayed in Filename Input Area as the filename Return Press this softkey to cancel inputting filename and return to the previous menu M300 User s Guide 1 15 RIGOL Chapter 1 Quick Guide Plug in Module Overview The M300 provides 8 kinds of modules These include a DMM module 20 channel multiplexer 32 channel multiplexer 64 channel multiplexer multi function module etc This section introduces the functions and characteristics of each module Outside View Drawing of the Module Golden Circuit board Edge each on the top and bottom for sliding into the slot Interface 1 connect the signal under test via the terminal block the DMM module does not have this interface Figure 1 7 Outside view drawing of the module 1 16 M300 User s Guide Chapter 1 Quick Guide RIGOL Insert the Module into the Mainframe Please turn the mainframe
112. is ready to receive data The instrument sets the RTS pin True when it is ready to receive data The instrument sets the RTS pin False when the input buffer is almost full approximately 100 characters and set the RTS pin True when space is available again 4 Baud Rate Use the up down direction keys to select Baud and use the left right direction keys to select the desired value 4800 9600 19200 38400 57600 and 115200 The unit is Baud Please select a baud rate that matches the PC or DTE The current setting is stored in the non volatile memory To Set LAN Parameters Connect the instrument to the PC or the network of the PC using network cable before using the LAN interface Press Utility gt I O gt LAN to enter the interface 1 Status Display the current LAN Status Use the up down direction keys to view other information about network connection as shown in the figure below Connected Disconnected LAN Configure IP Conflict I P Mode The IP address configuration mode can be DHCP AutolP or ManuallP By default DHCP is enabled AutolP and ManuallP are disabled In different IP configuration mode the configuration mode of network parameters such as the IP address is different M300 User s Guide 2 89 RIGOL Chapter 2 Front Panel Operations When all the three IP configuration modes are enabled the priority of the parameter configuration from high to low is DHCP Autol P
113. its scan mode and then press Next to enter into the advanced setting interface in which you can set the channel delay time SCAN 2 wire connection mode which is supported by all the multiplexer channels SCAN AW 4 wire connection mode which is supported by the first 16 channels of MC3132 and the first 10 channels of MC3120 and MC3324 Configure other channels into the scan list using the same mothod 3 4 M300 User s Guide Chapter 3 To Use External DMM RIGOL Scan Control Method The typical scan control method based on the connection shown in Figure 3 1 is as follows 1 Afterthe external DMM and M300 are connected and configured properly press the Run Stop key at the front panel of M300 to place M300 into wait for trigger state 2 When the preset trigger condition of the scan list is reached the first channel in the scan list is closed At this point the pin 5 Channel Closed Output of the Alarm Ext Trig interface outputs a negative TTL pulse its pulse width is about 7us and its output speed is related to the measurement speed and can not exceed 1000 times per second to the external trigger input interface and then the external DMM starts measuring 3 When the measurement is finished the VMC Voltage Measurement Complete output interface of the external DMM outputs a negative TTL pulse with about 2us pulse width to the pin 6 Channel Advance Input of the Alarm Ext Trig interface Thus M30
114. l Use the up down direction keys to select Condition and use the left right direction keys to select the desired alarm condition Not Pattern generate an alarm when the reading of the current channel does not match the alarm value set Pattern generate an alarm when the reading of the current channel matches the alarm value set Set the Alarm Value only for DI O channel Use the up down direction keys to select Value and use the numeric keyboard to input the desired alarm value The alarm value is a binary number that is formatted in Bits configuration 8 bit 16 bit or 32 bit You can input a number M300 User s Guide Chapter 2 Front Panel Operations RIGOL outside the range but when you press OK an error prompt message will be displayed The instrument can also ignore some digits and alarm when the values of some digits are satisfied Use on the numeric keyboard to set the digits to be ignored and the corresponding position will display x Key Points 1 You must complete the measurement configuration and scaling configuration before configuring the alarm parameters If you change the measurement configuration or enable scalling function alarms are turned off and the limit values are cleared 2 If you remove a channel from the scan list alarms are no longer evaluated on that channel but the limit values are not cleared If you decide to add that channel back to the scan list the original limit values are re
115. l connection which uses the VMC signal output from DM3068 as the advanced source signal At this point the trigger mode of M300 can not be External and the advanced source type of M300 shoild be set to External and the edge type must be set to Falling Alarm Ext Trig Interface of M300 Pin Definition 5 Channel Closed Output 6 Ext Trig Input Channel Advance Input GND Ext Trig IN VM Complete OUT Figure 3 1 To Connect External DMM Schematic Diagram Connect the signals to be measured To connect the signals to be measured to the modules multiplexer or multifunction module through the external terminal blocks according to Measurement Connections 3 2 M300 User s Guide 5 Chapter 3 To Use External DMM RIGOL Connect the signals to be measured into the external DMM Connect the analog bus interface at the rear panel of M300 and the measurement signal input terminal at the front panel of the external DMM using the analog bus external port option as shown in the figure below Note When connecting the external DMM the 5 terminals of the analog bus external port should be connected to the measurement signal input terminals of the external DMM according to the corresponding relations as shown in the table below between the numbers of the colored rings on the analog bus external port terminals and the mea
116. ls select Alarm1 You can modify the alarm configuration via the third step of the channel configuration guide refer to Alarm Configuration or using Add and Delete Add Use the up down direction keys to select Channel press Add and input the desired channel number After that the specified channel uses the current alarm channel and the number of the channel added is displayed in Channel e Delete Use the up down direction keys to select Channel use the left right direction keys to select the desired channel number and press Delete to delete the specified channel The number of the channel deleted will not be displayed in Channel 4 Clear Press this softkey to clear the alarm output of the corresponding pin of the current alarm channel manually but the alarm data is not cleared M300 User s Guide 2 67 RIGOL Chapter 2 Front Panel Operations View Switch M300 provides view switch function which provides multiple display forms for the scan list and scan results Press at the front panel to display the scan history information channel status table measurement curve and the channel information Scan History Information Press at the front panel to enter the scan history information interface through which you can view the start time scan sweep count channel number and the corresponding function as well as the maximum minimum average and standard deviation measured You can also store read the measurement data and vi
117. ltiplexer channels configured to the scan list and the DMM module is enabled DIO and TOT channels of the multifunction module no matter whether they are configured to the scan list and the DMM module is not required Press at the front panel to configure the channel monitor function Single channel Monitor Press Monitor gt Single to enter the single channel monitor interface as shown in Figure 2 12 and by default the first channel in the scan list is monitored If no scan list is currently configured the first channel of the multifunction module DIO channel is monitored Alarm Channel Alarm Value Alarm State Channel Number Module Type Function Range 201 Auto Alarmd 300 7450V LO FAIL 300 450V 100 2484V Single channel Moriitor Single Multiple All Moni Scaling Value Measurement Value System Time Figure 2 12 Single channel Monitor 2 44 M300 User s Guide 1 Chapter 2 Front Panel Operations RIGOL Channel Number The channel number is in the editable status At this point you can modify the number of the channel monitored by inputting the desired channel number using the numeric keyboard you can also press at the front panel to switch the channel monitored Note f the channel number of the multiplexer channel that is not added into the scan list is input the instrument will generate an error prompt message Module Type Display the module type of the channel currently monitored
118. meric keyboard to set a value The range is from 1 to 50000 After executing the specified number of scans the instrument stops scanning automatically 3 Trigger Mode Select TrigMode using the up down direction keys and select Auto Manual External AbsTime or Alarm using the left right direction keys Auto the instrument executes infinite scan or the specified number of scans at the specified interval automatically e Manual the instrument executes a scan when you press Run Stop e External the instrument executes a scan when the external trigger signal input interface receives a TTL trigger pulse e AbsTime the instrument executes a scan when the system time reaches the specified time e Alarm the instrument executes a scan when an alarm is generated from the specified alarm channel For details of the trigger mode refer to Trigger Mode 4 Interval Interval is defined as the time interval between the start of a scan to the start of the next scan in auto scan mode as shown in the figure below When the trigger mode is Auto select Interval using the up down direction keys and use the numeric keyboard to set a desired interval The range is from 00 00 00 000 to 99 59 59 999 the time format is Hour Minute Second Millisecond and the default is 00 00 00 000 A Scan a Interval M300 User s Guide 2 3 RIGOL Chapter 2 Front Panel Operations 5 Advanced Source Available only when the DMM
119. module is not installed or disabled This setting is only available when an external digital multimeter is used The advanced source provides the advanced channel signal When receiving the advanced channel signal the instrument disconnects the channel currently selected and closes the next channel in the scan list When an external DMM is connected use the up down direction keys to select Adv Src and use the left right direction keys to select the desired advanced source Auto After finishing measuring the current channel the current channel is disconnected and the next channel in the list is closed automatically e Single After finished measuring the current channel the unit will wait for user input before moving to the next channel in the list Pressing Single to disconnect the current channel and close the next channel Because both the Single advanced source and the manual trigger use the same way pressing Single to generate signal the manual trigger mode does not support Single advanced source External The scan will advance after receiving a trigger on the external advanced channel signal This signal typically the VMC signal from the external DMM is input from pin 6 of the Alarm Ext Trig interface converted from the RS 232 Alarms Ext Trig interface After External is selected you need to select the edge type of the input signal use the up down direction keys to select Input Edge and use the left rig
120. nches Power Supply AC 100V 120V 45Hz 440Hz AC 200V 240V 45Hz 66Hz Detect the power frequency automatically at power on 400 Hz defaults to 50 Hz Power Consumption 25 VA Max Working Environment Full accuracy for 0 C to 50 C Full accuracy to 8096 R H at 40 C Non coagulation Storage Temperature 40 C to 70 C Operation Altitude Up to 2000m Safety IEC 61010 1 EN 61010 1 UL 61010 1 CAN CSA C22 2 No 61010 1 Measurement CAT 1000V CAT II 300V Pollution Degree 2 EMC EN 61326 1 Weight About 5 7 kg without package Dimension height x width x length 159 0mm x 239 0mm x 373 4mm Remote Interface GPIB 10 100Mbit LAN USB 2 0 Full Speed Device amp Host support USB storage device RS 232 Programming Language SCPI LXI Compatibility LXI Core 2011 Device Version 1 4 Warm up Time 90 minutes Chapter 7 Appendix Chapter 7 Appendix RIGOL Appendix A Accessories and Options List Peserintion oO M300 Data Acquisition Switch System M300 Data Acquisition Switch System DMM Module M300 Data Acquisition Switch System DMM Module MC3120 20 Channel Multiplexer Power Cord conforming to the standard of the countr USB Cable Mixed interface Separator Line Four Spare Fuses 2 AC 250 V T3 15 A fuses 2 AC 250 V T250 mA fuses Quick Guide Resource CD User s Guide and Ultra Acquire DMM Module 6 digits 20 channel Multiplexer 32 channel Multiplexer Mai
121. nframe Standard Accessories Optional 64 channel Single ended Multiplexer Accessories 20 voltage channel 4 current channel Mixed Module Multiplexer 16 channel Actuator Multifunction Module 4x8 Matrix Switch MC3120 Terminal Block MC3132 Terminal Block Optional MC3164 Terminal Block Pesos MC3324 Terminal Block Terminal j Block MC3648 Terminal Block MC3534 Terminal Block MC3416 Terminal Block RS232 Serial Cable GPIB Invert Port Optional External Port for Analog Bus Accessories Rack Mount Kit Rack Mount Kit for Two Instruments PC Application Software for M300 Series Ordering No M300 M301 M302 CB USBA USBB FF 150 MIX SEPARATOR MC3065 MC3120 z 1 O w rA w N MC3164 MC3324 MC3416 MC3534 MC3648 M3TB20 M3TB32 M3TB64 M3TB24 M3TB48 M3TB34 M3TB16 M3GPIB A BUS EXT PORT RM 1 M300 RM 2 M300 Ultra Acquire M300 User s Guide 7 mn RIGOL Chapter 7 Appendix Appendix B Definitions of the Pins of the 78 pin I nterface of the Plug in Modules MC3120 MC3120 PinLabel Pin Qty Definition HI n 20 Correspond to the 20 channel differential positive input terminal on the NC HI_01 A e e Ed siena terminal T HI 02 DOCK e 2 e Ad LO nl 20 Correspond to the x Quo 20 channel x Goo differential negative e o input terminal on the x Qo external terminal no Hos block 10 09 10 05 NC 28 C
122. nnnnnnnnnnn nun lI General Safety Summary sss eem nnns I Safety Terms and Symibols yes teen ae era V Allgemeine Sicherheits Informationen VI Sicherheits Begriffe und Symbole sseeee n VIII General Care and Cleaning erssssreennenen nenne nennen enennen essen nn nnnn nennen IX Environmental Considerations une rn perm hd PRU dne eh EO De EY Enna IX M300 Overview uauunaunnunnnnnaunnnnnannaunnnnnnunnunnannnunnunnannnunnannaunnunnannnnnnnnnnnnnnen X Document Overview uuunsunnunnunnaunnnnnnnnaunnunnnnnnnnnnnnnunnunnannnunnnnnnunnunnnnnnnnnn XI Chapter 1 Quick Guide 1 e eeeee eere nun nun nn nnn nun nnne 1 1 General fspecHOTI een ea 1 2 Appearance and DIMENSIONS ce Hee 1 3 Front Panel id roter HR ROUES UEBER FREE RR ERE cond er ER ed 1 4 Rear Panel EnA 1 8 Power On nSpectiOn ei e n inc IST Shi 1 12 User Interface a teorie Pis dn db vb d ds 1 13 TO INBURPllename neiise edic e etd e b ee cda AA 1 14 Plug in Module OVErViCW ci mmn 1 16 Outside View Drawing of the Module 1 16 Insert the Module into the Mainframe l enn 1 17 Module OVERVIEW oves ikea 1 18 To Use the Built in Help System sies epo e eie te qx tede qat eee 1 20 Measurement Connections 00 cece eee mmn nnns 1 21 To Replace the F USe iier ntm Rete Y e ee irr Rete ees 1 25 Menu Quick Navigation sseeen mmn 1 26 GOnfIG estre ero edu ded eerte
123. not be recognized 1 Check whether the USB storage device can work normally 2 Make sure the USB storage device used is Flash storage type as this instrument does not support hardware storage type USB storage device 3 Restart the instrument and insert the USB storage device to check it 4 Ifthe USB storage device still cannot work normally please contact RIGOL 3 How to recover the configurations from the last power failure or unexpected shutdown and continue the work before the shutdown automatically 1 Restart the instrument 2 Press gt System select Power off using the up down direction keys and select ON using the left right direction keys 4 The screen is too dark Press gt System select Brightness using the up down direction keys and set the proper brightness using the left right direction keys 5 Theinstrument is locked 1 Check whether the instrument is in remote mode namely check whether L FON Rmt is displayed in the status bar of the interface If yes press the V key to exit the remote mode and unlock the instrument 2 Check whether the front panel of the instrument is locked namely check whether Gi is displayed in the status bar of the interface If yes press and hold the Nad key to unlock the front panel 3 Check whether a progress bar is displayed in the interface If yes please wait for a moment and operate the instrument when the progress bar M300 User s Guide 5 1 RIGOL Chap
124. nter the interface configuration interface To Set GPI B Address Connect the instrument with the PC with GPIB card installed using GPIB cable before using the GPIB General Purpose Interface Bus Interface Each device on the GPIB interface must have a unique address Press Utility gt I O gt GPIB to enter the interface Use the numeric keyboard to set the GPIB address to any integer between 0 and 30 and the default is 7 The address set is stored in the non volatile memory To Set USB Parameters A USB Device interface and a USB Host interface are provided by M300 The former is used to connect computer to realize remote control and the latter is used to connect an USB storage device to store measurement data information in screen ect Before using the USB Device interface please use the USB cable connect the USB Device interface at the rear panel of M300 and the USB Host interface of the computer and press Utility gt I O gt USB to enter the USB Device setting interface in which USB ID is displayed At this point you can control M300 remotly via USB Device interface Press gt 1 O gt USB gt Host to enter the USB Host setting interface Log Data When Auto is selected the instrument configuration and scan data will be stored as csv files in the USB storage device during a scan The path of the files stored is usually lt the drive letter of the USB storage device gt M300 data lt the
125. nterface and you can control the instrument remotely via a PC in the same network Analog Bus Interface Female M300 provides two internal 2 wire analog buses for signal routing as well as an 1 8 M300 User s Guide Chapter 1 Quick Guide RIGOL external interface namely Analog Bus Interface The analog bus interface is a 9 pin female interface Definition Current Input Lo LoSense Qo aon O Hi HiSense Float GND Ui FB wN HE o oo 1 OY U1 B W IN S y Analog Bus Interface 5 DMM Module Power Fuse The AC power supply from the power socket is divided into two paths with one for the DMM module and the other for other circuits except the DMM module M300 is equipped with the DMM module power fuse AC 250V T250 mA To replace the fuse refer to the method below 1 Turn off the instrument and remove the power cord 2 Press down the fuse cover using a straight screw driver and rotate clockwise 3 Select a proper voltage through the voltage selector 4 Disassemble the fuse cap and fuse 5 Replace a specified fuse and install the fuse cap CAUTI ON Please use the specified fuse to avoid electric shock and fire 6 Voltage Selector M300 supports two kinds of AC voltages 115 V and 230 V Please select the proper voltage scale according to the AC power used M300 User s Guide 1 9 RIGOL Chapter 1 Quick Guide
126. o independent banks and can be used for 4 wire connection When it is used for 4 wire resistance measurement the instrument pairs channel n and channel n 10 automatically wherein n is an integer between 1 and 10 Channel n is used to connect the source terminal of DMM and channel n 10 is used to connect the sense terminal of DMM Page Up Bank Joined Retum Figure 2 16 MC3120 Control Interface Bank Separate 2 50 M300 User s Guide Chapter 2 Front Panel Operations RIGOL To Control MC3132 MC3132 is a 32 channel multiplexer MUX 32 for short MC3132 is divided into two banks of 16 two wire channels HI and LO each and can be used for 2 wire and 4 wire connections When making 4 wire resistance measurement the instrument automatically pairs channel n with channel n 16 wherein n is an integer between 1 and 16 MC3132 is used with the external terminal block M3TB32 which has a built in thermocouple reference junction Terminal Block M3TB32 Channel Switches Bank Switch vrai Dot OMM H pli o 0 O at du O H DMM 01 6 LO o e o7 56 D Source L Signals e 78 Pin J to be e Interface P4 Measured ACY r Reference H x O Junction LO N EN 97 o oO H DMM se O L Sense Figure 2 17 MC3132 Schematic Diagram The control interface of MC3132 is as shown in Figure 2 18 Use the up down and left right direction keys to select the desired channel The channel cur
127. odule to the scan list and configure the basic measurement scaling alarm and advanced measurement parameters for each channel The edited scan list can be stored in internal or external memory and be recalled when required To Create a New List The instrument enters the scan list configuration interface when you start the M300 or press Config You can set the scan parameters such as sweep count tigger mode as well as create edit recall and save the scan list TrigMode Interval Edit Read Figure 2 1 Scan List Configuration Interface 1 To Create Scan List Press New to create a new scan list and enter the filename input interface At this point you can input the desired list name by referring to To Input Filename and press OK The scan list name can consist of Chinese characters English uppercase lowercase letters and numbers and can not exceed 12 bytes 2 2 M300 User s Guide Chapter 2 Front Panel Operations RIGOL each English letter or number occupies 1 byte and each Chinese character occupies 2 bytes To modify the name of the scan list created please select Scan List firstly and press OK to enter the filename input interface 2 Sweep Count Select Sweep using the up down direction keys and select Infinite or 1 you need to set a value using the left right direction keys Infinite execute continuous scan You can press and hold Run Stop to stop the scan Number of Scans you can use the nu
128. oltage function Input Impedance 1 MQ x 2 in parallel with 150 pF capacitance on any range Input Protection 300 Vrms on all ranges Measurement Considerations All frequency counters are susceptible to error when measuring low voltage low frequency signals Shielding inputs from external noise pickup is critical for minimizing measurement errors Settling Considerations Errors will occur when attempting to measure the frequency or period of an input following a dc offset voltage change The input blocking RC time constant must be allowed to fully settle about 1 s before the most accurate measurements are possible M300 User s Guide 6 7 RIGOL Chapter 6 Specifications Temperature Characteristics Accuracy Specifications Function Probe Type Type Optimum Range 1 Year Temperature Coefficient Tea C 5 C OC to Tcu C 5 C Teca C 5 C to 50 C Temperature RTDUI a 0 00385 200 C to 660 C 0 16 C 0 01 C Rois within 49 a 0 00389 200 C to 660 C 0 17 C 0 01 C Q and 2 1 kQ a 0 00391 200 C to 660 C 0 14 C 0 01 C a 0 00392 200 C to 60 C 0 15 C 0 01 C 2 2 kQ 40 C to 150 C 0 08 C 0 002 C Thermal 3 KQ 40 C to 150 C 0 08 C 0 002 C Kae ee 5kQ 40 C to 150 C 0 08 C 0 002 C 10 ko 40 C to 150 C 0 08 C 0 002 C 30 kQ 40 C to 150 C 0 08 C 0 002 C B OC to 1820C 0 76 C 0 14 C E 270 C to 1000 C 0 5 C 0 02 C J 210 C to 1200 C 0 5 C 0 02
129. oltage is in the range of 115 10096 1096 V and115 10096 10 V please select 115 and if it is in the range of 230 10096 1096 V and 230 10096 10 V please select 230 Connect the instrument to AC power using the power cord provided in the accessories f e J gO amp Oo ols g eE 2 AC Power Socket o v 2 Start the instrument Press the power key at the front panel and the instrument starts After the instrument starts it checks the modules This process would take several to dozens of seconds according to the number of modules inserted During this process user operation is forbidden 3 Check the instrument following the steps below if the instrument does not start normally 1 2 3 4 5 Check whether the power is correctly connected Check whether the power key at the front panel is really on Check whether the power fuse is burned out If the fuse needs to be changed use the specified fuse Check whether the correct power voltage is selected Please select the proper power voltage according to the power supply voltage Restart the instrument after finishing the above inspections If it still does not work correctly please contact RIGOL 1 12 M300 User s Guide Chapter 1 Quick Guide RIGOL User Interface Status Bar 5n 1 TEMP Content el J Area Unit
130. on key and press OK for example D 2 Select the desired folder under this drive letter using the up down direction key and press OK for example RIGOL At this point the current directory area displays D RIGOL 3 Directory File Display the files and folders under the current directory In the interface displayed by default when Store Recall is pressed this area displays the drive M300 User s Guide 2 73 RIGOL Chapter 2 Front Panel Operations letter C and or D The C disk is the local disk with 256 Mbytes storage space The D disk is the external USB storage device it will only be displayed when an USB storage device is connected Use the up down direction key to select the desired file folder or drive letter Menu Explanations 1 Format Press Format to delete all the data under the corresponding directory This menu is only available when C disk is currently selected 2 Read Press Read to read the file currently selected This menu is valid only when the specified type of file is selected Note Before reading a system configuration file make sure that the module currently inserted and the module used when storing the configuration file is of the same type 3 Save Press Save to enter the filename input area This menu is valid only when the current directory can store the specified type of file 4 Delete Press Delete to remove the file currently selected This menu is only valid the specified
131. onnection is mio ur oc forbidden e p P EX HI 1 Correspond to the K HILO 2 W HI terminal on p e ps the external terminal IC HI 08 block y e Bu EX LO 1 Correspond to the j 2 W LO terminal on e p P s the external terminal Q xis sa block xas SCL EX_HS 1 Correspond to the vcno Por SENSE HI terminal on 5 the external terminal H _ s block x 10 18 EX LS 1 Correspond to the Q c Quz SENSE LO terminal on 9 Q 10 17 the external terminal Q 10 20 Hi block 11 20 Q ois 15V 2 Connected to the 0 19 uis DGND 2 digital temperature s 10 15 SDA 1 sensor 1 C type on x nu SCL 1 the external terminal ex LO 14 block e p o e Notel n is an integer from 01 to 20 NC LO_13 n H2 Q r o e Ni e HI 11 Q Digital interface NC LO 11 O Analog input 7 2 M300 User s Guide Chapter 7 Appendix RIGOL MC3132 Pin Pin Definition Label Qty sd HI nl 32 Correspond to the 32 channel differential positive input terminal on the external terminal block LO ni 32 Correspond to the 32 channel differential negative input terminal on the external terminal block NC 4 Connection is forbidden EX HI 1 Correspond to the 2 W HI terminal on the external terminal block EX LO 1 Correspond to the 2 W LO terminal on the external terminal block EX HS 1 Correspond to the SENSE HI terminal on the external terminal block EX LS 1 Correspond to the
132. ork segment for details refer to the TCP IP rotocol An example is shown in the table below Network Parameters PC Instrument IP Address 192 16 3 3 192 16 3 8 Subnet Mask 255 255 255 0 255 255 255 0 Default Gateway 192 16 3 1 192 16 3 1 If your M300 is connected to the LAN of your PC please acquire a set of valid network parameters IP Address Subnet Mask Gateway and DNS and configure the network parameters of the instrument according to the description in To Set LAN Parameters 3 Search device resource Start up Ultra Sigma click and click EEE in the pop up window Ultra Sigma searches for the instrument resources currently connected to the LAN and the resources found are displayed at the right of the window Select the desired resource name and click en to add it as shown in the figures below M300 User s Guide 4 11 RIGOL Chapter 4 Remote Control Create LAH Instrument Resource SS Remove Manual Input LAN Instrument IP Auto detect of LAN Instrument Search Create LAN Instrument Resource Remove Manual Input LAN Instrument IP TCPIP 1T2 18 Auto detect of LAN Instrument Search b 4 View device resource The resources found will appear under the RIGOL Online Resource directory as shown in the figure below GPIB USB TMC RS232 Verify All nline ce a INSTR Other Online GOL TECHNOLOGIES INC 4 12 M300 User s Guide Chapter 4 Rem
133. ote Control RIGOL 5 Communication test Right click the resource name M300 TCPIP 172 16 3 21 INSTR to select SCPI Panel Control to turn on the remote command control panel through which you can send commands and read data 6 LoadLXI webpage As this instrument conforms to LXI Core 2011 Device standards you can load the LXI webpage through Ultra Sigma right click the resource name and select LXI Web or directly input the IP address in the browser Information about the instrument model manufacturer serial number description MAC address and IP address will be displayed on the webpage M300 User s Guide 4 13 Chapter 5 Troubleshooting RIGOL Chapter 5 Troubleshooting This section lists some questions and failures which might occur when using this instrument Please solve them according to the following method If the problem remains please contact RIGOL and provide the device information of your instrument Utility gt Info 1 The instrument does not start 1 Check the power cable connection 2 Check whether the power switch at the front panel is turned on 3 Remove the power cord and check whether the voltage selector is at the proper scale and whether the fuse specification is correct and the fuse is in good condition To replace the fuse refer to the instructions 4 Restart the instrument after executing the above steps 5 Ifthe problem remains please contact RIGOL 2 The USB storage device can
134. our DIO channels can all be used as digital input ports You can configure the DIO channels to the scan list to read the input value of the specified channel automatically or you can read the input value manually via the MC3534 control interface Digital Output DOUT only DIO channels that are not configured to the scan list can be used as digital output DOUT ports Read read the input value of the channel currently selected The value is displayed on the interface Write edit the output value of the channel currently selected Press Write and use the numeric keyboard to input the desired value between 00000000 and 11111111 If the channel currently selected is added to the scan list this operation is invalid 2 TOT Channel Control You can configure the TOT channels to the scan list to read the count of the specified channel automatically or you can read the count manually via the 2 62 M300 User s Guide Chapter 2 Front Panel Operations RIGOL MC3534 control interface Read read the count of the channel currently selected The count is displayed on the interface Start press this softkey and the current TOT channel starts counting e Stop press this softkey and the current TOT channel stops counting This function is only valid when the current channel is counting Def Start when the instrument starts next time the power on setting is Last this TOT channel starts counting by default You do not need to
135. out H Vin H Max TTL 0 8V 22 0V lt 0 2V I ou 500MA gt 4 8V l oy 1MA lt 42V with external open 5V CMOS lt 1 5V 23 5V lt 0 2V I ou 500mA gt 4 8V l ou 1mA drain pull up 3 3V CMOS lt 1 0V gt 2 3V lt 0 2V lo4 500MA 3 15VGl 21mA 2 5V CMOS lt 0 75V gt 1 75V 0 2V8l 2 500mA gt 2 35V 1 1mA User defined Threshold 0 3V Threshold 0 3V 0 2V8l 2 500mA gt Level 0 2V I u 2 1mA Alarming Match or mismatch maskable Match or mismatch maskable Speed 4ms Max alarm sampling 4ms Max alarm sampling Latency 5ms 5ms Read Write 100 s 100 s Speed Totalizer Input TOT High Speed TOT1 TOT2 Normal Speed TOT3 TOT4 Maximum 32 32 Count dies i Totallzer Input LOMEIZ max rising arfalling edge 100kHz max rising or falling edge programmable programmable Signal Level CMOS 3 3V 5V tolerable 1Vp p min 42Vpk max Vem 12V 12V Threshold Fixed at CMOS 3 3V 12V 12V programmable Gated Input CMOS 3 3V Hi CMOS 3 3V Lo or none CMOS 3 3V Hi CMOS 3 3V Lo or none 5V tolerance 5V tolerance Count Reset Manual or Read Reset Manual or Read Reset Read Speed 100 s 100 s Analog Voltage Output DAC DAC 1 2 3 4 12V non isolated earth referenced Resolution 1mV lout 10mA max Setting Time 1ms to 0 01 of output 6 12 M300 User s Guide Chapter 6 Specifications RIGOL Accuracy output mV 1 year 5 C 0 25 20mV Temp 0 015 1mV C Coefficient General Specifications Display 4 3 i
136. pdate Enable the built in help system Provide Chinese English help information for the front panel function keys and menu items Alarm Channel Setting M300 provides 4 alarm channels When an alarm is generated the corresponding pin of the Alarms Ext Trig interface converted from the RS 232 Alarms Ext Trig interface at the rear panel outputs a pulse You can press the corresponding key to set the output mode of the corresponding pin Power Key Press this key to turn on or off the instrument Via this interface M300 can be as a host device and connected with an USB storage device to store or recall instrument status measurement configuration measurement data etc 1 6 M300 User s Guide Chapter 1 Quick Guide RIGOL 8 9 10 Menu Softkeys Correspond to the menus on the screen Pressing any softkey can enable the corresponding menu View Switch View the scan history information channel status table measurement curve and channel information e Scan history information view the start time scan sweep count channel number as well as the corresponding function maximum minimum average and standard deviation measured and the scan EN readings of each channel of the latest scan j Channel status table display the current status of each channel Measurement curve display the measurement curve of the specified channel e Channel information view the cycle count of each relay of the module the
137. pecial characters embedded in the data stream to control the flow If the instrument is addressed to send data it continues sending data until the XOFF character 13H is received When the XON character 11H is received the instrument resumes sending data e DTR DSR This mode is hardware flow control mode The instrument monitors the state of the DSR pin When the state goes True the instrument sends data over the interface When the state goes False the instrument stops sending data The instrument sets the DTR pin False when the input buffer is almost full approximately 100 characters and sets the pin True M300 User s Guide Chapter 2 Front Panel Operations RIGOL when space is available again e RTS CTS This mode is hardware flow control mode and it operates the same as the DTR DSR mode The instrument monitors the state of the CTS pin When the state goes True the instrument sends data over the interface When the state goes False the instrument stops sending data The instrument sets the RTS pin False when the input buffer is almost full approximately 100 characters and sets the pin True when space is available again e MODEM This mode uses the DTR DSR and RTS CTS pins to control the flow of data between the instrument and a modem When the RS232 interface is selected the instrument sets the DTR pin True The DSR pin is set True when the modem is on line The modem sets the CTS pin True when it
138. pendix B Warranty RIGOL warrants that its products mainframe and accessories will be free from defects in materials and workmanship within the warranty period If a product is proven to be defective within the respective period RIGOL guarantees the free replacement or repair of products which are approved defective To get repair service please contact with your nearest RIGOL sales and service office RIGOL does not provide other warranty items except the one being provided by this warranty statement The warranty items include but not being subjected to the hint guarantee items related to tradable characteristic and any particular purpose RIGOL will not take any responsibility in cases regarding to indirect particular and ensuing damage M300 User s Guide 7 9 Index RIGOL I ndex QW ee 2 14 Extended Copy ic teret on 2 43 AWR iiie Re iE 2 14 External Trigger 2 9 Absolute Time 2 4 Factory Settings ten 2 79 Absolute Time Trigger 2 10 FREO HH 2 14 AC Filter senes 2 26 FUNCION iic 2 14 ACV ipne cette eret e Ph dva 2 14 Gate time eese ces 2 27 Advanced Configuration 2 20 Gateway ce ien nes 2 91 Advanced Source 2 4 GPIB Address 2 86 Alarm Channel 2 5 2 18 Input Impedance 2 24 Alarm Condition 2 18 Insert the Module into the Mainframe Alarm Configuration 2 17
139. play form of the decimal point affects the display form of the separator The two cannot be set to or at the same time There are 6 kinds of data formats Decimal Point Separator Example 10 000 00 Space 10 000 00 None 10 00000 10 000 00 Space 10 000 00 None 10 00000 Power Switch Use the up down direction keys to select PwrSW and use the left right direction keys to select ON or OFF to set whether the front panel power key is valid The default is OFF ON the front panel power key is valid After power on you need to press the front panel power key to start the instrument OFF the front panel power key is invalid After power on the instrument starts automatically But you can still use the front panel power key to turn off the instrument 2 84 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Module Plug None of the current M300 modules are hot swappable To avoid damage you can select the mainframe response below Use the up down direction keys to select Card Plug and use the left right direction keys to select the way in which the instrument responses to the hot plug of the module Restart the instrument restarts automatically If you insert the module during a scan the instrument restarts and resumes the scan Confirm a prompt message is displayed At this point users need to confirm the module connection
140. ple is one of the most commonly used devices for measuring temperature in industry It converts temperature variation information into voltage variation information It provides larger temperature measurement range e Model After setting the temperature sensor type to TC use the up down direction keys to select Model and use the left right direction keys to select the desired model The M300 supports the following thermocouple types B Pt Rh 30 Pt Rh6 E NI CR WRCK J Fe WRCK K NI CR NiSi N NiCrNi NiSi R Pt Rh13 Pt S Pt Rh10 Pt and T Cu WRCK Among them B R and S are precious metal thermocouples E J K N and T are low cost metal thermocouples For the reference table about each thermocouple please refer to the ITS 90 International Temperature Scale of 1990 standards M300 User s Guide 2 29 RIGOL Chapter 2 Front Panel Operations Precision After setting the temperature sensor type to TC use the up down direction keys to select Precision and use the left right direction keys to select 1 or 0 1 T C Check The T C Check Thermocouple Check function checks whether the TC is connected correctly After setting the temperature sensor type to TC enter the fourth step of the channel configuration guide use the up down direction keys to select T C Check and use the left right direction keys to select ON or OFF When ON is selected the instrument measures the channel resistance in the 20 kQ
141. press OK to confirm it press OK again to cancel M300 User s Guide 2 41 RIGOL Chapter 2 Front Panel Operations it and press Done to execute the channel copy You also can press Switch to fix the channel selected as the target channel after entering the channel copy interface At this point the lower area is selected Use the left right direction keys to select the desired source channel and press Done to execute channel copy Tips 1 When the channel selected is fixed as the source channel multiple target channels can be selected and are displayed in the lower area 2 When the channel selected is fixed as the target channel only one source channel can be selected 3 The Switch menu is gray and disabled if the scan list only has one channel Note For channel copy only copies between the same type of channels are allowed 2 42 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Extended Copy Extended copy allows copying the configuration of a channel source channel of a module to all the channels which type is the same to that of the source channel of the target module After that the following settings will be copied to the target module automatically 1 Measurement Configuration include measurement function range etc 2 Scaling Setting on off status of the scaling function and scaling coefficients 3 Alarm Setting on off status of the alarm function alarm channel number and alarm limits
142. press Start Def Stop when the instrument starts next time the power on setting is Last this TOT channel stops counting by default You need to press Start to start the counting 3 DAC Channel Control The multifunction module MC3534 provides 4 low noise analog output DAC channels The 4 analog output channels are capable of outputting calibrated voltages between 12 V and 12 V with 16 bits of resolution Each DAC channel can be used as the programmable voltage source for other analog control device The DAC channels are usually numbered S09 to S12 wherein S denotes the number of the slot of the multifunction module Press Write and use the numeric keyboard to input the desired output value between 12 000 V and 412 000 V and the resolution is 1 mV 4 Channel Reset Reset the status of the channel currently selected M300 User s Guide 2 63 RIGOL Chapter 2 Front Panel Operations To Control MC3648 MC3648 is a 4x8 matrix switch and contains 32 two wire crosspoints Each crosspoint relay has its own unique channel label representing the row and column For example channel 24 represents the crosspoint connection between row 2 and column 4 This module does not connect to DMM You can close multiple channels and connect multiple inputs to multiple outputs at the same time Before making any connection make sure that the connection will not introduce any danger lay ag ag ag ay ae a ae 11 aia Mle Mla Ma ia Kt Ro
143. r Channel Delay 2WR Range Integration Time Offset Compensation Auto Zero AZ 4WR Channel Delay FREQ Range AC Filter Gate Time Channel Delay PERIOD i i TC Model Unit Precision Integration Time T C Check Reference Source Auto Zero AZ Channel Delay Model Unit Precision Integration Time Auto Zero TEMP Type THER AZ Channel Delay RTD RO ALPHA Unit Precision Integration Time Offset RTD 4W Compensation Auto Zero AZ Channel Delay DCV Unit Scaling Integration Time Input Impedance Auto Zero AZ Channel Delay DCI Unit Scaling Integration Time Auto Zero AZ SENSOR Type Channel Delay 2WR Unit Scaling Integration Time Offset 4WR Compensation Auto Zero AZ Channel Delay FREQ Unit Scaling AC Filter Channel Delay Note When the measurement function of the current channel is changed all the parameters will be restored to their defaults M300 User s Guide Chapter 2 Front Panel Operations RIGOL To Set the Range The setting modes of range include Auto and Manual In auto range mode the instrument selects the proper range for each measurement according to the input signal automatically In the manual range mode users can select a fixed range for each channel Auto ranging is convenient and manual ranging provides faster scanning because some additional time is required for the instrument to make a range selection in auto range mode Enter the first step of the channel configu
144. range automatically after each thermocouple measurement to ensure a proper connection If an open connection is detected greater than 5 kQ the instrument displays OPEN T C Reference Source Thermocouple measurements require a reference junction temperature After setting the temperature sensor type to TC enter the fourth step of the channel configuration guide use the up down direction keys to select Ref and use the left right direction keys to select Fixed Internal or External a Fixed select a fixed reference temperature Then use up down direction keys to select Fixed and use the numberic keyboard to input desired temperature The range is from 20 C to 80 C always specify the temperature in C regardless of the temperature units currently selected Using this type of reference source you can obtain the highest measurement precision b Internal select an internal isothermal reference temperature on the module Using this type of reference source you will obtain the lowest measurement precision c External before selecting external reference please find the first module which supports temperature measurement function from Slotl Then configure the first channel of this module to perform a temperature measurement function The selected type can be THER RTD or RTD 4W Otherwise you can not select external reference for other channels This channel is then reserved as the reference source At this poin
145. ration guide select the desired channel number and function use the up down direction keys to select Range and use the left right direction keys to select the desired range The ranges available for each function are as shown in the table below Table 2 2 Range Function Range DCV ACV Auto 200 mV 2 V 20 V 200 V 300 V DCI ACI Auto 200 pA 2 mA 20 mA 200 mA 1A 2WR AWR Auto 200 Q 2 kQ 20 kQ 200 kQ 1 MQ 10 MQ 100 MQ FREQ PERI OD Auto 200 mV 2 V 20 V 200 V 300 V Note n frequency period measurement the range set is the voltage range of the signal under test and is used to calculate the measurement resolution The measurement range of frequency period is from 3 Hz to 1 MHz 1 us to 0 33 s Key Points Auto range threshold For signal under test with voltage resistance that is between 10 Range and 11096 Range the instrument selects Range as the current range automatically For example the signal under test for DCV measurement is 5 V between 1096 20 V and 110 20 V the 20 V range is selected automatically e fthe input signal is greater than the current range the instrument gives an overrange indication Atpower on or after a reset the range is Auto by default Auto range is recommended when it is hard to predict the measurement range to protect the instrument and acquire more accurate data Fortemperature measurements the instrument internally selects
146. rein YYYY denotes Year and the range is from 2001 to 2099 MM denotes Month and the range is from 01 to 12 DD denotes Date and the range is from 01 to 28 29 30 or 31 related to the YYYY and MM currently set Time Use the up down direction keys to select Time and use the numeric keyboard to set the system time The format is as follows hh mm ss Wherein hh denotes Hour and the range is from 00 to 23 mm denotes Minute and the range is from 00 to 59 ss denotes Second and the range is from 00 to 59 Tip The date and time forms the system time The instrument stores the measurement data and alarm data with the time stamp during a scan In absolute trigger mode the instrument triggers when the system time reaches the specified time Language M300 supports Chinese and English menus and built in help systems All the messages and help information are displayed in the language selected Use the up down direction keys to select Language and use the left right direction keys to select the desired language 2 78 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Sound Use the up down direction keys to select Sound and use the left right direction keys to select ON or OFF to enable or disable the sound generated when operating any of the front panel keys The current setting is stored in the no volatile memory When the sound is disabled kal is displayed in the status bar of the user in
147. rently selected is highlighted for example ur You can perform various operations on the channel currently selected such as disconnect and close the channel M300 User s Guide 2 51 RIGOL Chapter 2 Front Panel Operations Figure 2 18 MC3132 Control Interface Bank Joined If the channel currently selected is not added to the scan list and other channels of the module are not added to the scan list pressing OK can close the channel At this point pressing OK again can open the channel If the channel is added to the scan list pressing OK can close the channel and read the measurement result before closing this channel the instrument opens the channel currently closed Pressing again can read the measurement result again 1 Open Disconnect the channel currently selected The channel number of the channel currently disconnected is displayed in gray for example E if one or more channels of the current module are added to the scan list this operation is invalid 2 Close Close the channel currently selected The channel number of the channel currently closed is displayed in green for example E it one or more channels of the current module are added to the scan list this operation is invalid 3 Read Read the measurement value of the current channel This operation is only applicable to channels added to the scan list 4 Card Reset Initialize the status of each channel of the current module If one or more ch
148. ress Contact Us If you have any problem or requirement when using our products or this manual please contact RIGOL E mail service rigol com Website www rigol com M300 User s Guide RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it To prevent potential hazards please use the instrument only specified by this manual Use Proper Power Cord Only the power cord designed for the instrument and authorized for use within the local country could be used Ground the I nstrument The instrument is grounded through the Protective Earth lead of the power cord To avoid electric shock it is essential to connect the earth terminal of the power cord to the Protective Earth terminal before connecting any inputs or outputs Connect the Probe Correctly If a probe is used do not connect the ground lead to high voltage since it has isobaric electric potential as the ground Observe All Terminal Ratings To avoid fire or shock hazard observe all ratings and markers on the instrument and check your manual for more information about ratings before connecting the instrument Use Proper Overvoltage Protection Make sure that no overvoltage such as that caused by a thunderstorm can reach the product or else the operator might be exposed
149. rm channel and set the alarm limits high limit low limit or both limits for channel in the scan list During the scan an alarm will be generated in the alarm channel when readings that exceed the specified limit are detected If an alarm is generated during the scan the alarm data is stored in the reading memory as well as the alarm queue which is separate from the reading memory The non scanned alarms alarms during single channel monitor alarms generated by the DIO channel etc can only be stored in the alarm queue You can assign multiple channels to any of the 4 alarm channels and use the same alarm limit on multiple channels but you cannot assign multiple alarm channels and alarm limits to the specified channel You can configure the alarm parameters for the DIO and TOT channels that are not added into the scan list M300 User s Guide 2 17 1 RIGOL Chapter 2 Front Panel Operations Select the Alarm Mode Enter the alarm configuration interface Mode is selected by default and you can use the left right direction keys to select the desired mode If the DIO channel is currently used the options available are OFF and ON If the TOT channel is currently used the options available are only NONE and HI NONE disable the alarm configuration HI LO when the measurement value is greater than HI or lower than LO an alarm will be generated from the corresponding alarm channel HI when the measurement v
150. rs and advanced measurement parameters of the channel Enable or disable the channel monitor function e Support single channel multi channel and all channel monitor functions In single channel monitor function you can switch the channel monitored at any time e In multi channel monitor function you can monitor up to 7 channels in the scan list 1 4 M300 User s Guide Chapter 1 Quick Guide RIGOL In all channel monitor function you can monitor all the channels in the scan list Control the modules currently inserted Control the status of each channel of the modules Reset the modules Configure the DIO and TOT channels of multifunction module e Can not control the DMM module Auto Trigger Stop Press this key the instrument triggers or enters the wait for trigger state and the backlight goes on Press and hold this key the instrument stops triggering or exits the wait for trigger state and the backlight goes off Note This key is invalid when the scan list is empty Single channel Switch When the instrument is in the single channel monitor interface press this key to switch the channel monitored to the next channel in the scan list When the instrument is in other interfaces press this key to switch the instrument to the single channel monitor interface 3 Control Keys C ER Numeric Keyboard Cs Include numbers 0 9 decimal point as well as A J J positive ne
151. stored and alarms are turned back on M300 User s Guide 2 19 RIGOL Chapter 2 Front Panel Operations Advanced Configuration The fourth step of the channel configuration guide configure the advanced measurement parameters relating to the function selected 1 Local Advanced wm EN Re 10PLC 10M0 ON Auto Return Figure 2 7 Advanced Configuration The advanced parameters to be configured differ when different function is selected in the Measurement Configuration interface for the current channel 1 Multiplexer Channel According to the function currently selected you need to configure the integration time DC impedance auto zero etc For details please refer to Measurement Parameter Configuration 2 DIO Channel Select coding mode for the current DIO channel refer to Digital I nput 3 TOT Channel Set the reading mode for the current TOT channel refer to Totalizer 2 20 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Channel Configuration When Using External DMM For the multiplexer channels if the DMM module is not inserted or enabled you can connect an external DMM to realize scan and measurement function for the connection method please refer to To Connect I nstruments Press Edit New to enter into the channel configuration guide which only contains measurement configuration and advanced configuration 1 Measurement Configuration After entering into the channel config
152. strument is in a scan the instrument will ignore the new trigger and continue with the current scan The absolute time format is Month Date Hour Minute Second You need to set a valid time The ranges of the parameters are as follows Month 01 to 12 Date 01 to 31 30 29 or 28 Hour 00 to 23 Minute 00 to 59 Second 00 to 59 The instrument can also ignore some time parameters and trigger when the system time fulfills some of the parameters Use on the numeric keyboard to set the parameters to be ignored and the corresponding position will display For example to set the instrument to ignore the Month and Date and trigger at 8 00 every day use the numeric keyboard to set the absolute time to 08 00 00 When parameters in Month Date Hour Minute Second are all ignored it is similar to auto trigger mode In absolute time trigger mode if the number of scans is set to Infinite after you press at the front panel the instrument enters the wait for trigger state and then executes a scan each time the system time reaches the specified time Press and hold Run Stop and the instrument will exit the wait for trigger state If the number of scans is a specific value after you press Run Stop at the front panel the instrument enters the wait for trigger state and then executes a scan each time the system time reaches the specified time After executing the specified number of absolute time trig
153. surement signal input terminals of the external DMM No Corresponding Input Terminal 6 HI 2 LO 7 HI SENSE 3 LO SENSE 1 CURRENT Connect to the analog bus interface of M300 Connect to the measurement signal input terminals of the external DMM If only a single multiplexer module is used you can connect the corresponding terminals as shown in the figure below on the external terminal block to the measurement signal input terminal of the external DMM M3TB20 M3TB32 M3TB64 M3TB24 M300 User s Guide 3 3 RIGOL Chapter 3 To Use External DMM To Configure the Instruments 1 Toconfigure the external DMM Set the external DMM to external trigger mode and triggering at the falling edge of the external trigger signal 2 To configure the M300 1 To configure the scan list parameters Press Config to set the scan cycles trigger mode advanced source type etc Based on the connection in Figure 3 1 external trigger mode can not be selected advanced source should be set to external and the edge type must be set to Falling Note External trigger mode and external advanced source can not be selected at the same time e Manual trigger mode and Single advanced source can not be selected at the same time To configure channels to the scan list Press Config gt Edit gt New to enter into the channel configuration guide You can select desired channel number and set
154. t you can select other external eferences for other channels If this channel is deleted or its function is modified the channels taking this channel as reference source will revert to the internal reference automatically 4 Use thermistor to measure temperature A thermistor is made up of material of which the resistance changes with temperature and it converts temperature variation information to resistance M300 User s Guide Chapter 2 Front Panel Operations RIGOL variation information Thermistor is especially sensitive to temperature but its temperature measurement range is small and is more suitable for measurements with relatively small temperature variations Thermistor resistance is usually very high Therefor the 2 wire resistance method is preferred when measuring the resistance of the thermistor The internal software conversion program converts the resistance measured to the corresponding temperature e Model After setting the temperature sensor type to THER use the up down direction keys to select Model and use the left right direction keys to select the desired model 2 2kO 3kQ 5kQ 10kQ or 30kQ e Precision After setting the temperature sensor type to THER use the up down direction keys to select Precision and use the left right direction keys to select 1 0 1 or 0 01 5 Use RTD to measure temperature RTD measures the temperature on the basis that the resistance of the measured substance changes with temp
155. tact with RIGOL To Control M300 through GPI B 1 Connect the device Connect the instrument with the PC with GPIB card installed using GPIB cable before using the GPIB General Purpose Interface Bus Interface 2 Install the driver of GPI B card Install the driver of the GPIB card which has been connected to the PC correctly 3 Set the GPIB address Set the GPIB address of the instrument according to the description in To Set GPIB Address 4 Search device resource Start up the Ultra Sigma and click to open the panel as shown in the figure below Click Search and the software will search the GPIB instrument resources connected to the PC The device resources will be displayed on the right side of the panel RS232 Setting GPIB Setting Remove OK GPIBO 7 INSTR O INSTR GPIBO 0 INSTR Hyperchannel Search TEST If resources cannot be found automatically Select the GPIB card address of the PC from the comboBox of GPIBO and select the GPIB address set in the instrument from the comboBox of 0 I NSTR Click Test to check whether the GPIB communication works normally if not please follow the corresponding prompt messages to solve the problem 4 6 M300 User s Guide Chapter 4 Remote Control RIGOL 5 View device resource Click Ok to return back to the main interface of Ultra Sigma The resources found will appear under the RIGOL Online Resource directory as shown
156. ter 5 Troubleshooting 10 disappears 4 Ifthe problem remains please restart the instrument How to change the system language Press gt System select Language using the up down direction keys and select your desired language using the left right direction keys How to view the information and model number of the plug in modules Press Utility Test select the desired module using the left right direction keys and press View to view the information of the module currently selected including the function model serial number and version The corresponding indicator at the front panel does not go on when the module has been inserted 1 Press Utility Test select the desired module using the left right direction keys and press Re test If the problem remains press View to view the information of the module 2 Ifthe information of the module does not been displayed correctly please refer to the problem 9 3 Ifthe information of the module is displayed correctly the module indicator is possibly damaged please contact RIGOL to consult maintenance information Unable to view the information of the module or unable to use the module when the module has been inserted 1 Turn the mainframe off pluck the module from the mainframe and insert the module into the mainframe again firmly 2 Restart the mainframe wait for about 1 minute and view the module information again 3 If the problem remains please contact RIGO
157. ter 7 Appendix MC3534 RIGOL Pin Label Pin Qty Definition Dn In 24 Correspond to the digital input output terminal on the external terminal block Tm GT 4 Correspond to the Tm GT 4 TOT channel input T3_CNT 2 terminal on the T4_CNT external terminal T3_CNT 2 block T4 CNT DACm 4 Correspond to the DAC channel output terminal on the external terminal block NC 2 Connection is forbidden GND 20 Ground terminal S GND 6 Correspond to the GNDterminal of the DAC channel on the external terminal block Note n is an integer from 01 to 08 Note m is an integer from 1 to 4 M300 User s Guide 7 7 RIGOL MC3648 Chapter 7 Appendix MC3648 5 co 9c CHI cnn e 102 co ec Qc coz 9 c0 c 9c co co e c 0 o r 02 ec Q5 o9 5 0 RHR 5 R103 R104 Q5 ri04 2103 85 RH3 giu e 105 e 1105 cH5 eco e 17 106 oc cs cio Qc cv civ cis cos cis eco Pin Label Pin Qty Definition C Lon 16 Correspond to the signal C Hin 16 input terminal on the R LOml 8 external terminal block R Him 8 Not 30 Not used included in the figure Notel C refers to Column n is an integer from 1 to Note R refers to Row m is an integer from 1 to 4 7 8 M300 User s Guide Chapter 7 Appendix RIGOL Ap
158. terface DMM Use the up down direction keys to select DMM and use the left right direction keys to select ON or OFF to enable or disable the DMM module MC3065 When the DMM module is disabled the system initializes the scan list automatically This operation is only valid when the DMM module is inserted If the DMM module is not inserted this option is fixed at OFF and cannot be modified Factory Settings Use the up down direction keys to select Factory and the left right direction keys to select ON or OFF the default is OFF When ON is selected the prompt message asking whether you want to restore all the settings to the factory setting will be displayed Press OK The factory settings are as shown in Table 2 5 Table 2 5 Factory Settings Parameter Factory Setting Scan Configuration Scan List Empty Reading Memory Cleared Max Min Average and SDEV Cleared Trigger Mode Auto Scan Interval 10 s use TRI Gger SOURce TI Mer Number of Scans 1 Measurement Configuration Function DCV Range Auto Integration Time 1 PLC Input Impedance 10 MQ Channel Delay Auto TOT Read Mode READ TOT Trigger Mode Rising M300 User s Guide RIGOL Chapter 2 Front Panel Operations Scaling Configuration Scaling Configuration OFF A 0 B 1 C 0 x1 0 Unit V Alarm Configuration Alarm Mode
159. terminal on the external terminal block EX ILO Correspond to the CURR LO terminal on the external terminal block EX IHI Correspond to the CURR HI terminal on the external terminal block 5V DGND SDA SCL RI N N Connected to the digital temperature sensor I C type on the external terminal block Note n is an integer from 01 to 24 M300 User s Guide 7 5 RIGOL MC3416 Chapter 7 Appendix MC3416 s 7 5 cQ o CH16 NO m CH15 NO eo ous o e CH14 CM CH14 NO Q 759 oso 5 CH12_NO cis vg 5 Qo CH12 CM rc oo 7 99 Io CH9 NO CH10 CM 7909 CH9 CM x so 7 CH7_NO eo CH7 NC CH8 CM Hs o o o5 x CH5_NO m Hm 9 CH4 NO Hin CHS_NC CH3 CM o 9o CH2_NO 8 CH2_CM n aim Hin go Pin Label Pin Qty Definition CHn_NC 16 Correspond to the 16 channel NC terminal on the external terminal block CHn NOH 16 Correspond to the 16 channel NO terminal on the external terminal block CHn CM 16 Correspond to the 16 channel COM terminal on the external terminal block Not 30 Not used included in the figure Notel n is an integer from 01 to 16 Notel NC refers to Normally Closed Note NO refers to Normally Open Note COM refers to Common 7 6 M300 User s Guide Chap
160. the scan list are break before make You can close multiple channels on this module only if you have not configured any channel to be part of the scan list This module can be connected with MC3065 DMM Module if MC3065 is currently inserted MC3132 32 Channel Multiplexer All 32 channels switch both HI and LO inputs thus providing fully isolated inputs to the DMM module MC3132 is divided into two banks called A and B of 16 two wire channels each When making 4 wire resistance measurement the instrument automatically pairs channels of A bank and B bank All channels in the scan list are break before make You can close multiple channels on this module only if you have not configured any channel to be part of the scan list This module can be connected with MC3065 DMM Module if MC3065 is currently inserted MC3164 64 Channel Single Ended Multiplexer All 64 channels can switch the HI input only MC3164 is divided into two banks A and B of 32 single ended channels each All channels in the scan list are break before make In any case you cannot close multiple channels on this module This module can be connected with MC3065 DMM Module if MC3065 is currently inserted but cannot be used for 4 wire resistance measurement MC3324 24 Channel Multiplexer The MC3324 is a mixed multiplexer with 20 voltage channels and 4 current channels All 20 voltage channels switch both HI and LO inputs thus providing fully isolated inputs to
161. to the danger of electrical shock Do Not Operate Without Covers Do not operate the instrument with covers or panels removed II M300 User s Guide RIGOL Do Not I nsert Anything I nto the Holes of Fan Do not insert anything into the holes of the fan to avoid damaging the instrument Use Proper Fuse Please use the specified fuses Avoid Circuit or Wire Exposure Do not touch exposed junctions and components when the unit is powered Do Not Operate With Suspected Failures If you suspect damage occurs to the instrument have it inspected by RIGOL authorized personnel before further operations Any maintenance adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel Keep Well Ventilation Inadequate ventilation may cause an increase of instrument temperature which would cause damage to the instrument So please keep the instrument well ventilated and inspect the intake and fan regularly Do Not Operate in Wet Conditions In order to avoid short circuiting to the interior of the device or electric shock please do not operate the instrument in a humid environment Do Not Operate in an Explosive Atmosphere In order to avoid damage to the device or personal injuries it is important to operate the device away from an explosive atmosphere Keep Product Surfaces Clean and Dry To avoid the influence of dust and or moisture in the air please keep the surface of the devi
162. trigger signal In external trigger mode the instrument initialize a trigger when a trigger signal met the conditions is received by the Ext Tig Input terminal 5 Pin 6 Ext Trig Input Channel Advance Input 00000 0000 Alarms Ext Trig In external trigger mode if the number of scans is set to Infinite after you press Run Stop at the front panel the instrument enters the wait for trigger state and then executes a scan each time an external trigger signal is received Press and hold Run Stop and the instrument will exit the wait for trigger state If the number of scans is a specific value after you press Run Stop at the front panel the instrument enters the wait for trigger state and then the instrument executes a scan each time an external trigger signal is received After executing the specified number of external triggers the instrument exits the wait for trigger state During the scan the instrument stores the measurement data scaling data and alarm data in the non volatile memory automatically M300 User s Guide 2 9 RIGOL Chapter 2 Front Panel Operations Absolute Time Trigger To use the absolute time trigger mode set an accurate system time for the instrument by referring to Date and Time Then select absolute time trigger mode and set a time The instrument executes a scan when the system time reaches the specified time If the system time reaches the next specified time point when the in
163. trol Interface Bank Joined If the channel currently selected is not added to the scan list and other channels of the module are not added to the scan list pressing OK can close the channel At this point pressing OK again can open the channel If the channel is added to the scan list pressing OK can close the channel and read the measurement result before closing this channel the instrument opens the channel currently closed Pressing 1 OK again can read the measurement result again Open Disconnect the channel currently selected The channel number of the channel currently disconnected is displayed in gray for example E if one or more channels of the current module are added to the scan list this operation is invalid Close Close the channel currently selected Only one channel can be closed each time The channel number of the channel currently closed is displayed in green for example E it one or more channels of the current module are added to the scan list this operation is invalid Read Read the measurement value of the current channel This operation is only applicable to channels added to the scan list Module Reset Initialize the status of each channel of the current module If one or more channels of the current module are added to the scan list this operation is M300 User s Guide 2 57 RIGOL Chapter 2 Front Panel Operations invalid 5 Bank Separate Bank J oined By default the first
164. type of file is selected File Type You can store or read various types of files 1 System Configuration Store the current system configuration as a file with the suffix sfg into the internal memory or external USB storage device The system configuration saves the states of all the modules including all the system information under Utility System except date and time and the configurations of the I O interfaces under Utility gt 1 0 2 Measurement Configuration Store the current measurement configuration as a file with the suffix mfg into the internal memory or external USB storage device The measurement configuration saves the scan list settings including trigger 2 74 M300 User s Guide Chapter 2 Front Panel Operations RIGOL count trigger mode interval for auto trigger mode and trigger edge type for external trigger and the channel configurations including measurement configuration scalling configuration alarm configuration and advanced configuration 3 Mirror Configuration Combine the System Configuration and Measurement Configuration files into one file with the suffix mir and store it in the internal memory or the external USB storage device The mirror configuration is the combination of the system configuration and the measurement configuration 4 Measurement Data Store the current measurement data as a file with the suffix dat or csv into the internal memory or external USB storag
165. uces how to connect the signals to be measured through the external block Table 1 1 External Terminal Blocks Terminal Block Module 1 M3TB20 MC3120 2 M3TB32 MC3132 3 M3TB64 MC3164 4 M3TB24 MC3324 5 M3TB48 MC3648 6 M3TB16 MC3416 7 M3TB34 MC3534 WARNI NG Cut off all the power supplies before opening the terminal block or removing the terminal block from the module l Place the terminal block with the front with the RIGOL caption facing downward and pull the two latches in the arrow direction in the figure below and press them down Place the terminal block with the front facing upward and pull the upper cover of the terminal block upward and remove it 2 Connect the cables for connecting the signal under test according to the labels on the circuit board of the terminal block for the connecting method refer to M300 User s Guide 1 21 RIGOL Chapter 1 Quick Guide Table 1 2 Pass the binding wires through the two groups of holes as pointed out by the arrows in the figure below to fix the cable connected Pay attention to the channel or terminal connected to the cable to ensure that the signal connected is correct Note For the connection method of each module and the corresponding terminal block as well as the specification requirements of the connecting cable please log in RIGOL official website www rigol com to get more related materials When using th
166. uration guide you can select desired channel number and use the up down direction keys to select the Mode item and then use the left right direction keys to select desired connection mode SCAN 2 wire connection mode which is supported by all the multiplexer channels SCAN AW 4 wire connection mode which is supported by the first 16 channels of MC3132 and the first 10 channels of MC3120 and MC3324 2 Advanced Configuration Press Next to enter the advance configuration interface and set the channel delay time according to Channel Delay Explanation In this case M300 only control the connection mode 2 wire or 4 wire and the delay time of the channels in the scan list The function and parameters for each channel are determined by the external DMM M300 User s Guide 2 21 RIGOL Chapter 2 Front Panel Operations Measurement Parameter Configuration This section introduces the measurement parameters such as the range and integration time that can be configured in the first and fourth steps of the configuration guide The parameters available for configuration vary for each measurement function as shown in the table below Table 2 1 Measurement Parameter Function Parameter DCV Range Integration Time Input Impedance Auto Zero AZ Channel Delay ACV Range AC Filter Channel Delay DCI Range Integration Time Auto Zero AZ Channel Delay ACI Range AC Filte
167. user interface as well as the measurement connections of the M300 Topics in this chapter General Inspection Appearance and Dimensions Front Panel Rear Panel Power On Inspection User Interface To Input Filename Plug in Module Overview To Use the Built in Help System Measurement Connections To Replace the Fuse Menu Quick Navigation M300 User s Guide 1 1 RIGOL Chapter 1 Quick Guide General I nspection 1 Inspect the shipping container for damage Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has passed both electrical and mechanical tests The consigner or carrier shall be liable for the damage to instrument resulting from shipment RIGOL would not be responsible for free maintenance rework or replacement of the unit Inspect the instrument In case of any damage or defect or failure notify your RIGOL sales representative Check the Accessories Please check the accessories according to the packing lists If the accessories are incomplete or damaged please contact your RIGOL sales representative 1 2 M300 User s Guide Chapter 1 Quick Guide RIGOL Appearance and Dimensions
168. ves the affects of small DC offsets of the measurement lead on the measurement result When the current measurement function is 2WR 4WR TEMP RTD or RTD 4W or SENSOR 2WR sensor or 4WR sensor enter the fourth step of the channel configuration guide use the up down direction keys to select OCOM and use the left right direction keys to select ON or OFF e ON After selecting ON add two different currents onto the resistance under test and measure the variation of the voltage between two ends of the resistance Calculate the resistance under test by dividing the voltage variation by the current variation e OFF After selecting OFF no offset compensation will be made Tip Offset compensation and auto zero are mutually exclusive After enabling offset compensation auto zero if currently enabled will be disabled automatically and vice versa AC Filter AC filter is applicable to the ACV ACI FREQ PERIOD and SENSOR FREQ sensor measurement functions The AC filter optimizes the low frequency accuracy and minimizes the AC settling time The instrument selects slow medium or fast filter based on the frequency of the input signal of the current channel as shown in the table below Table 2 3 AC Filter Type Input Frequency AC Filter Type 3 Hz to 300 kHz 3 Hz slow 20 Hz to 300 kHz 20 Hz medium 200 Hz to 300 kHz 200 Hz fast When the current measurement function is ACV ACI FREQ PER
169. vides more complicated signal switch You can connect more inputs to more outputs at the same time Please make sure that the connection will not introduce any danger M300 User s Guide 2 65 RIGOL Chapter 2 Front Panel Operations Alarm Channel Setting M300 provides 4 alarm channels Alarm1 Alarm2 Alarm3 and Alarm4 You can assign an alarm channel to a configured channel via the channel configuration guide An alarm will be generated from the corresponding channel when a reading fulfills the specified alarm condition during a scan or monitor When an alarm is generated the corresponding pin of the Alarm Ext Trig interface at the rear panel converted from the RS 232 Alarms Ext Trig interface outputs a pulse with specified edge If an alarm channel is assigned to multiple channels the corresponding pin outputs a pulse when any of the channel generates an alarm Pin Definition 1 Alarm 1 Output Alarm 2 Output Alarm 3 Output Alarm 4 Output UNBE AJWJIN Alarm Ext Trig Interface Press WAS NAN AAY or NAN at the front panel to set the output mode of the corresponding pin of the Alarm Ext Trig interface for the 4 alarm channels when an alarm is generated Figure 2 31 Alarm Channel Setting Interface 1 Mode Set the output mode of the corresponding pin when an alarm is generated Use the up down direction keys to select Mode and use the left right direction keys to select Latch
170. w 2 JA AAA AA S ow AA z a 2 a a a a 48 Col Col2 Col3 Cold Col5 Col amp Col7 Col8 5 J v Row 4 Rod O iub WE Signals bad to b e Switched m Terminal Block M3TB48 Figure 2 29 MC3648 Schematic Diagram 78 Pin Interface The control interface of MC3648 is as shown in Figure 2 31 Use the up down and left right direction keys to select the desired channel The channel currently selected is highlighted You can perform various operations on the channel currently selected such as disconnect and close the channel 2 64 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Col 1 Col Col 3 Col 4 Col 5 Col B Col Open Close Card Reset Return Figure 2 30 MC3648 Control Interface 1 Open Disconnect the channel currently selected The channel number of the channel currently disconnected is displayed in gray Pressing OK can also open the channel currently selected 2 Close Close the channel currently selected The channel number of the channel currently closed is displayed in green Pressing OK can also close the channel currently selected 3 Card Reset Initialize the status of each channel of the current module Matrix Combination For M300 you can combine two or more up to 5 matrixes together to form a larger matrix For example you can connect the rows or columns of two 4x8 matrixes to form a 4x16 8x8 matrix Note The combined matrix pro
171. wo banks of 10 two wire channels each and can be used for 2 wire and 4 wire connections When making 4 wire resistance measurement the instrument automatically pairs channel n with channel n 10 wherein n is an integer between 1 and 10 4 current channels from channel 21 to channel 24 are used to measure DC current or AC current MC3324 is used with the external terminal block M3TB24 which has a built in thermocouple reference junction Channel Switches Backplane Switches o ES oa O E E H DMM f L o o o Source O OL Terminal Block M3TB24 o 10 m N Q O Com 1 170000 O S ot O Com 4W Sense 78 Pin Interface 97 ae o QOH DMM ge O Sense Signals Pi s to be e Measured e Reference 96 Junction 222 I se mO DMM O d O 32 DO L O VA j A r Figure 2 22 MC3324 Schematic Diagram The control interface of MC3324 is as shown in Figure 2 24 Press Page Up and Page Dn at the front panel to open the previous and next pages of the control interface Use the up down and left right direction keys to select the desired channel The channel currently selected is highlighted for example Jl You can perform M300 User s Guide 2 56 Chapter 2 Front Panel Operations RIGOL various operations on the channel currently selected such as disconnect and close the channel 18 17 18 19 20 Close Card Reset Page Dn Figure 2 23 MC3324 Con
172. xecutes a scan and then enters the wait for trigger state After that the instrument triggers each time you press Run Stop at the front panel If the number of scans is set to Infinite after each trigger is finished the instrument will always stay in wait for trigger state You can press and hold the key to stop the scan and the instrument will exit the wait for trigger state If the number of scans is set to a specific value 1 to 50000 after the trigger is finished the instrument enters the wait for trigger state The instrument will exit the wait for trigger state when the number of triggers reaches the preset value During the scan the instrument stores the measurement data scaling data and 2 8 M300 User s Guide Chapter 2 Front Panel Operations RIGOL alarm data in the non volatile memory automatically You can press at the front panel to view the history measurement data including the maximum minimum average and standard deviation External Trigger To use the external trigger mode first use the mixed interface separator line accessory to convert the RS 232 Alarms Ext Trig interface at the rear panel to two 9 pin interfaces Wherein the 9 pin male interface is a standard RS232 interface and the 9 pin female interface is used for alarm output external trigger signal input pin 6 as shown in the figure below and so on Secondly you need to select the edge type rising or falling for the external
173. ys to select C F and O respectively and then press and hold the OK key to enter C F and Coefficient Setting Explanation When setting the fitting function coefficients A B and C press any key on the numeric keyboard to open the dialog box in the form below e Sign use the left direction key to move the cursor to the sign bit and press on the numeric keyboard to switch the sign to or Unit use the right direction key to move the cursor to the unit bit and use the up down direction keys to select the desired unit The units available include n 10 u 109 m 10 1 K 10 M 10 G 10 T 10 and P 107 e Number use the left right direction keys to move the cursor to the number bit and use the numeric keyboard to input the desired value Sign 112 5 K Unit Number 2 16 M300 User s Guide Chapter 2 Front Panel Operations RIGOL Alarm Configuration The third step of channel configuration guide configure the alarm parameters Alarm NONE PAFO Return Figure 2 6 Alarm Configuration Note For multiplexer channel if the DMM module is not currently inserted or it is inserted but is not enabled the configuration guide does not contain Alarm Configuration when you connect an external DMM to execute scan and measurement M300 has 4 alarm channels Alarm1 Alarm2 Alarm3 and Alarm4 refer to Alarm Channel Setting You can configure an ala
Download Pdf Manuals
Related Search
Related Contents
『ストリングチェッカー(ソラメンテ-Z)』のレンタル料金はこちらから Indesit IWE 81481 S GRUPO F - TÉCNICO DE NÍVEL MÉDIO III Operating Instructions Submersible drainage pump Autonomie de la batterie et nombre d`images pouvant être catalogo serie b50 Comment maîtriser l`acidité des vins - Vinnopôle Sud fflBitte Bilderseite Oyster HDTV CI+: Bedienung ROI Villers 2012-2013 Copyright © All rights reserved.