Home
        Cobas Integra 400 Troubleshooting
         Contents
1.   2 Call Roche Service  if the status of the system is Fatal Error     This section lists those problems that may occur when results are obtained from the  instrument     Quality control results out of range    Quality control results are obtained but they are outside the expected range for the test        gt  Control or a diluent in the wrong position    E 22    Use Status   Missing  amp  Blocked to check which items are missing     2 Use Status   Samples to check the current positions of all on board controls and  diluents     3 Replace the controls or diluents in their correct positions as required     User Manual   Version 2 0    COBAS 16 Troubleshooting  INTEGRA 400 Results       gt  Control material has deteriorated or was incorrectly reconstituted    1 Use Status   Samples to check the current positions of all on board controls     2 Replace the controls with fresh or freshly prepared material in the correct rack  positions        gt  Lot specific assigned values are not correct    e Check the values in Configuration   Controls   Lot        gt  Pipetting error  The volume of fluid delivered by the dosage pipettes is incorrect        Leakage    on page E 25      Air bubbles in one or more pipettes    on page E 26     Dripping pipettes    on page E 26      Dripping probes    on page E 27       gt  Reagent in the cassette has deteriorated    1 Check the cassette expiration date in Status   Cassettes     If the cassette has reached its expiration date  replace it with a ne
2.   Water supply or cleaner not properly connected    1 In Status   Analyzer  check whether the cleaner is running low  If necessary  replace  the cleaner     Check the connections to the cleaner and water supply     Perform the Prime the fluid system service action        Replacing the cleaner    on page B 41     Waste  water  and power connectors    on page E 29     Prime the fluid system    on page D 33    Dripping pipettes    If a visual inspection shows that either of the pipettes is leaking  you must take immediate  action        gt  Pipette not correctly installed  e Reinstall the pipette        Replace plunger tip dosage pipette  B or C     on page D 41     amp  Piston worn out    1 Replace piston or pipette     2 Ifthe piston only is replaced and drips reappear after replacement  replace the  complete pipette        Replace plunger tip dosage pipette  B or C     on page D 41    E 26 User Manual   Version 2 0    INTEGRA 400    Dripping probes    User Manual   Version 2 0    16 Troubleshooting    Hardware    Ifa visual inspection shows that either of the probes is leaking  you must take immediate  action     Loose tubing connections    1 Check that the tubing connectors are firmly screwed into place and tighten  if  necessary  do not over tighten any connections      2 Replace the complete probe and tubing assembly  if necessary        Replace probe B or C    on page D 44    Damaged tubing or probe  e Replace the complete probe and tubing assembly  if necessary        Re
3.  B Click to change the size of the message log     Working with messages    You can confirm messages  display or print details on a message  review confirmed  messages  or display Help on a message     Chapter 4     Online Help       Confirming messages    A shortened form of each message is stored on the New Messages tab on the View  Message dialog box  When you confirm a message  it is automatically removed from  the New Messages tab and transferred to the Message Log tab  The system can carry on  running even if there are unconfirmed messages     Recommendation Do not confirm a message until you are sure that you have resolved the problem that  caused it to appear in the first place     The running man on the Messages button remains until you confirm   1  The message that caused it to be displayed     2  All subsequent messages that could have caused a running man to be displayed     E 8 User Manual   Version 2 0    INTEGRA 400          15 Messages    Working with messages       gt  To confirm a message    1 Click Messages on the navigation bar   The View Message dialog box is displayed   2 Click the New Messages tab  if necessary   3 When you have fixed the problem  select the message and click Confirm   The selected message is automatically transferred to the message log   Tips    Double click a message to get more information     Follow any Help or instructions contained in the message to remove the cause or  causes of the message  if you have not already fixed the prob
4.  Stop  Tests in process are completed     C Stop  Terminate run and discard testin process        A Stop pipetting samples but finish processing tests that are already in progress     B Stop all system operations immediately  The system aborts all current activities  The system may need  special attention before it can resume its normal operation     Tips  e Ifthe system is already in Standby when you click Stop  it goes into Sleeping     e Ifyou stop the run immediately  all the partially run tests are stopped  In practice   this means you lose all the reagent  control  calibrator  and sample material for any  tests that have been started but not yet completed     e The background color of the Status button tells you what the current system state  is  green means Operating  unless there is a resource alert      e You can also use F12 to stop the processing        The Start button serves to start or restart a run     When the system is in the Stopped  AutoService  or Operating status  the text of the  Start button changes to Restart     Tips  e Clicking Restart causes the system to perform a re scheduling of all patient orders   For example  clicking Restart after inserting a new rack into the system while in    Operating status  causes the system to start pipetting on the sample rack that is  positioned furthest to the left     e You can also use F11 to restart the processing     E 11    15 Messages COBAS  Stopping and restarting the system INTEGRA 400  Re initializing When 
5.  To replace the fuse  you need the following tools and materials   O Replacement fuses    O 4mm screwdriver       gt  To replace a mains fuse    1 Switch off the instrument at the main switch  A         Wait for the power down to complete  display will be black    Switch off the mains supply manually     Isolate the instrument from the mains power supply by removing the power  connector from the back of the instrument     If you do not know how to do this  do not attempt this procedure     5 When you are sure that the instrument is isolated from the mains power supply   remove the left side panel  B      E 32 User Manual   Version 2 0    COBAS 16 Troubleshooting     INTEGRA 40000 Checking and replacing fuses                            7 Using a small screwdriver  turn the fuse holder counter clockwise   The fuse is spring loaded and pops out as you turn the fuse holder    8 Place the new fuse into the fuse holder and reinsert    9 Replace the left side panel  B     10 Reconnect the instrument to the mains power supply    11 Switch on at the main switch  A      12 If the fuse fails again  contact Roche Service     User Manual   Version 2 0 E 33    16 Troubleshooting  Checking and replacing fuses    INTEGRA 400    Replacing controller board fuses    The controller board fuses protect the electronic controller circuit boards that are  located under the main front cover  To see if a fuse has failed  check the LEDs  there is  one for each fuse   a green LED means that the fuse is 
6.  dialog is displayed     Create  C  Roche Instrument Report Version  txt             E 35    16 Troubleshooting COBAS  Contacting Roche Service INTEGRA 400    5 When the report has been generated  the Problem Report dialog box is displayed    again        The report has been completed  Do you want to copy the report to  floppy disk s             You will need 1 disk s      Warning  All existing files on the disks  will be ereased     Message ID 28 1230       6 Insert the disk  any data on the disk will be destroyed    7 Click Yes     Tips  e You can also create a problem report from the message dialog box  for some types  of messages     e You need one or more disk  3 5 inch floppy disk  to store the report     e Contact your Roche contact person  and send the disk to your local Roche Service  organization     e Ifyou turn off the main switch and turn on again  you may get a message  prompting you to create a problem report before the system re initializes  You can  ignore this message     Remote diagnostics    Use the Roche remote diagnostics support service whenever you have a problem that  you cannot resolve     This service allows a Roche Service engineer to log on and control the system remotely   For this level of support  a modem must be connected to the specified serial port on  the instrument        gt  To enable a remote diagnostic session    1 Check that the modem is connected to the serial port COM 2 and to a telephone  line with a direct external connection  that
7.  is  no connection to another telephone  user         Data connectors    on page E 30  2 Ensure that the system and the modem are on  You do not have to be logged on     the Service engineer will automatically log on using the remote access service     3 The Roche Service engineer can now access your system remotely     Customer confidentiality When a service engineer logs on to your system remotely  patient demographics are  automatically switched off so that any patient information is not directly available in  the remote session or on the monitor screen on the instrument     Sample results are  however  available on the remote system   as are records of  calibrations  controls  and service actions     A message will be displayed on your system  informing you that the system is currently  accessed by remote access     E 36 User Manual   Version 2 0    INTEGRA 400    Make a note of the  following information    Problems with chemistry    User Manual   Version 2 0    16 Troubleshooting    Contacting Roche Service    When you report a problem to Roche Service  please have the following information  available     e Analyzer serial number   e Maintenance information  see service notes in the Service work area   e System software version   e TAS version of the test in question   e TAS version installed on the system   e Description of the problem    e Description of the last operations   carried out immediately before the problem  arose    For chemistry troubleshooting  you shoul
8.  log    1 Click Messages on the navigation bar   The View Message dialog box is displayed   Click the Message Log tab    3 Click Print     Tip    The message log contains messages that have been confirmed  and therefore have been  automatically removed from the list on the New Messages tab      Stopping and restarting the system    Stop processing    Stop    E 10    You often need to stop the system when you are troubleshooting  The method you  choose depends not only on what you need to do in terms of corrective action  but also  what the system is doing  the system state  and how your throughput will be affected     There are two levels of stopping   1  Stopping and continuing processing  2  Shutting down and restarting the system     Check the messages and Help to see which method is recommended for a particular  case     If the system is in the Operating state  for example  when it is processing tests  you can  interrupt sampling or to stop sampling once the current actions are finished  by  clicking Stop on the navigation bar     The precise function of the buttons depends on the current system status     User Manual   Version 2 0    INTEGRA 400    n gt     Restarting the system    Start    User Manual   Version 2 0    15 Messages  Stopping and restarting the system    When clicking Stop  the Stop Action dialog box is displayed  which offers the Sampling  Stop and the Stop options        Stop Action  x     AN Stop has been pressed     Indicate action to take     Sampling
9.  mans seen 32  Replacins controller Board USES ri nae nere r ee een 34   Contacting Roche Service anreisen 35  Creating a prod emi Tepo susanne 35  Remote GIA CIO SCICS nassen een 36   E 15    INTEGRA 400    Introduction    16 Troubleshooting    Introduction    Effective troubleshooting on a COBAS INTEGRA 400 system requires a good  understanding of all basic operating procedures for the instrument and the system  software  However  you can avoid most common problems if you observe the  recommended procedures at all times and if the operating environment is effectively  controlled     For new users in particular  there are the usual familiarization problems including   e Connecting various parts of the computer system  if necessary    e Locating the main switch and function keys    e Starting up the program and logging on    e Finding your way around the user interface    All the information you need is available either in the User Manual or in the online    Help     For details on flags  error messages  and recommended actions for ISE related problems  see    Flags and  user actions    on page F 35 and    General ISE troubleshooting    on page F 42     Information about problems    The user interface keeps you informed about the status of the instrument as a whole   and about particular hardware  software  and chemistry problems as they arise  It does  this through     e Messages displayed on the screen    e Status work area    e Buttons on the navigation bar    e Flags on r
10.  started  SUPER  7 16 98 14 42 Diagnostics not started  SUPER    Message 7 of 10           Help On Message   Print    N G       se        Icon showing the severity of the message if any    The date and time when the message was created    Short message text    Click to get Help on the selected message  if available     Name of the user logged on at the time the message was created   The selected message  Double click to display details on this message     Click for Help on this dialog box  this does not give Help on the selected message      Tips    The system stores only the most recent confirmed messages in the message log     The log has a limited capacity and so the system automatically deletes the oldest  messages in the log when it needs more space for new ones     The message log contains only confirmed messages     15 Messages COBAS  Optionstab INTEGRA 400    Options tab    Use the Options tab to specify the maximum number of messages stored in the  message log  Type a number in the Buffer Size text box and click Save     You can increase or decrease the size of message log buffer as required  A buffer size of  100 means that a maximum of 100 messages are stored in the log  The system saves the  most recent messages and  if the log is full  and deletes the oldest messages to make  way for the more recent ones        m  View Message x   New Messages   Message Log Options    Message Log    Buffer Size  I             A Specifies the number messages in the message log    
11. COBAS  INTEGRA 400    Troubleshooting       Part E describes messages generated by the system  what they mean and  how to deal with them  It also describes specific problems and gives    step by step procedures on how to deal with them     INTEGRA 400    Messages    15 Messages  Table of contents    General problem solving techniques    User Manual   Version 2 0    This chapter explains how to identify and interpret messages  and suggests ways of  reacting to these messages     In this chapter Chapter 15    POU EINES ee een meer 4  DIE so Duo ana ee ee ee nro nee 5  NEW NIE 0 0  ee ee ee ee le Dre 6  Message LO TAD ans a 7   DPM OS TAD seadeciess ce AO NUR ORERIEN a EEEHIENEEG SIEH EURUERER CENT DEN EUESURSEN aA 8  Workine Wiih messages  ee ee een ine 8  COMMENTS Message sauna N 8  Display  ne messaseletals  rs Rein 9  Printinsmessase orria e r ENEE E E 10  Stoppie and restarting the Systirin E A N aai 10  DOPP OCES O a E E A EE 10  SEIKE DAA ee E E E S 12    15 Messages COBAS  About messages INTEGRA 400    About messages    Messages inform you about errors or other events that have occurred on the system   The system displays important messages as popup message dialog boxes  for example     Hardware Warning x          IN A hardware error has occurred in the Fluid controller   Message ID  14 3238 03 010 63    q 08 010    Click Help for further information     The icon displayed on the message dialog box reflects the severity of the problem     Symbol Meaning    Information me
12. barcode scanner window  if necessary     Check whether the barcode setup  including the reading of the installation  barcodes  was properly completed     Check whether the printing quality of the barcode labels is sufficient     Check whether there is condensation on the barcode labels     External connectors    The connectors that you may need to check are located either at the back of the  instrument  A  or on the right side near the back  B  as shown below                          A Waste  water  and power connectors  not visible here     B Data connectors    Waste  water  and power connectors    The waste  water  and power connectors are located at the back of instrument        Mains input connector   Mains supply to  optional  printer  Water reservoir level sensor   Not used   Waste reservoir level sensor    Water reservoir connector    A  B  C  D  E  F  G    User Manual   Version 2 0 E 29    Waste reservoir connector    16 Troubleshooting    aah          External connectors INTEGRA 400  Tips    e Do not touch mains cables  if your hands are wet     e Do not attempt to connect or disconnect either of the mains cables  if the  instrument is switched on     e If any power connector becomes worn or frayed  it must be replaced immediately  with an approved cable  Contact your Roche Service representative     e You must connect the waste sensor  if you are using a separate waste container   that is  the instrument is not connected to the main laboratory waste system      e Yo
13. bcomponents has failed  Also  the LEDs for the  fuses for the controller boards are not lit when the fuse has blown  normally  they are  bright green      A complete set of replacement fuses is delivered in the start up kit  The order number  for these fuses is listed in the COBAS INTEGRA 400 Installation Manual     If you use a fuse that has too low a rating  it will almost certainly blow again  This  wastes time  but is not in itself dangerous     If you use a fuse that has too high a rating  your system will not be protected from  excessive currents  This is very dangerous and can lead to a situation where a fire  develops  Even without a fire you can get temperature control problems or other  electrical faults if the underlying problem reoccurs     Fire hazard    For continued protection replace fuses only with fuses of the same type and rating    Danger of electrocution    Always disconnect the instrument from the mains before attempting to replace any fuse  Do not  attempt to work with fuses if your hands are wet or if there is excess moisture on any fittings or  components     E 31    16 Troubleshooting COBAS  Checking and replacing fuses INTEGRA 400    Replacing a mains fuse    There are two mains fuses  both of which must have the same current rating  which  depends on the mains voltage supplied to the instrument     Mains voltage Fuse rating  230 VAC 8A  115 VAC 15 A  I   z Q   The mains input voltage is pre selected by a switch inside the voltage selection unit    
14. ck the printer cable  This is a standard PC printer cable connected from the  LPT1  25 pin  connector on the analyzer to the    Centronics    type connector on  the printer itself     e Check that the printer power connector is connected at both ends  A 3 pin Euro  connector  plug end of the cable  must be connected to the back of the analyzer  and a 3 pin Euro connector  socket end of the cable  must be connected to the  power input on the printer     e Check that the water supply is connected  The analyzer will not initialize properly  if a water supply is not connected     e Check that the water sensor is connected  if you are getting your water supply from  a water container   The water sensor must be inside the water container       e Check that the waste outlet is connected to a suitable waste container or to a waste  drain     e Check that the waste level sensor is connected  if your waste is going to a waste  container   The waste level sensor itself must be inside the waste container       Checking and replacing fuses    How do I know whether a  fuse has blown     Where do I get the fuses  from     Why is the correct fuse  important     Danger    User Manual   Version 2 0    A set of fuses protects the analyzer from excessive electrical currents  If a fuse blows  a  message is displayed to inform you that you need to replace that particular fuse     Typically  you get one or more messages from the appropriate controller informing  you that the power to one or more su
15. d also have the following additional  information available     Item Lot number  Cassette   Calibrator   Control    Diluent    You may also need the following    O The last calibration and control results   O A printout of the cassette status   O A printout of the raw data of the specific test s     E 37    16 Troubleshooting COBAS  Contacting Roche Service INTEGRA 400    E 38 User Manual   Version 2 0    
16. d in Configuration   Result Handling apply to the workflow in    your laboratory and change the result handling if applicable  See the Online Help for  details     Aborted e Test aborted    Test was aborted either by stopping the run and selecting    STOP   Terminate Run and  discard tests in progress    or by switching off the instrument     Recommended action Restart the run     ISE    Not on board e Activator not on board  e Deproteinizer not on board  e Etcher not on board    e No ISE solution available    Recommended action 1 Open the Not on Board folder in Status   Missing  amp  Blocked  to see the solutions  that are not on board and their correct rack positions     Remove the ISE rack from the instrument     Load the required activator  deproteinizer  etcher  or ISE solutions on their correct  ISE rack positions     4 Reinsert the ISE rack into slot O   The ISE Rack Replaced dialog box is displayed   5 On the dialog box select the ISE solutions that you have replaced       Q   Leave the checkboxes cleared for ISE solutions you have not replaced     Empty e Activator is empty  e Deproteinizer is empty  e Etcher is empty    e ISE solution is empty    E 20 User Manual   Version 2 0    INTEGRA 400    Recommended action    Not ready    Recommended action    Disabled    Recommended action    No electrode    Recommended action    Error    Recommended action    User Manual   Version 2 0    16 Troubleshooting    Blocked orders  tests  calibrations  or controls    1 Open the Em
17. d take the necessary action     2 Ifthe initialization is incomplete and no error is shown  contact Roche Service     Blocked orders  tests  calibrations  or controls    User Manual   Version 2 0    The following main conditions must be fulfilled before you can successfully run any  test on the INTEGRA 400     e All required reagents  diluents  cleaners  calibrators and controls are on board and  have enough volume to run the test     e The last calibration and control results are accepted  for calibration and control  tests      e The sample is on board and has enough volume to run the test   e The test and the measurement modules are enabled   e The instrument and the measurement modules are in an error free state     If any of these conditions are not fulfilled  the test will be blocked and the cause must  be eliminated     To check the cause of blocked orders  tests  calibrations  or controls    Click Orders on the navigation bar    Click the Worklist tab    Open the Blocked folder and select a blocked order    Detailed information of the selected order is displayed in the area on the right     4 Make a note of the Blocked Status Text and look up the corresponding  recommended actions in the list below     E 17    16 Troubleshooting    Blocked orders  tests  calibrations  or controls INTEGRA 400    Cassettes and cleaner    Not on board    Recommended action    Empty    Recommended action    Not Enough for  Worklist    Recommended action    Tips    Double click a blocked o
18. e all the slots for cassette racks     6 If this does not fix the problem  contact Roche Service     Computer system and accessories    This section lists problems that may be associated with the computer system including  connection problems  and other basic problems with the computer  monitor  printer   and barcode reader       gt  Cables detached or damaged  1 Ensure that all the required connections to your instrument are made        Connections check list    on page E 30    2 If any cables are damaged  contact Roche Service      amp  Monitor not on    1 Check that the instrument is plugged into the laboratory mains   2 Check the main fuses        Replacing a mains fuse    on page E 32    3 If this does not fix the problem  contact Roche Service      gt  Printer connections and printer indicators OK  but still no printout    Check the printer settings in Windows NT     2 Check that the printer driver  currently used as the default  corresponds to the  printer actually connected  Modify the printer settings  if necessary     3 Switch the printer off and back on again to re initialize     If a printing is still not possible  shut down the system  Restart and wait until the  system is initialized     5 Ifthe problem reoccurs  contact Roche Service     E 28 User Manual   Version 2 0    COBAS 16 Troubleshooting    INTEGRA 40000 External connectors       gt  Barcode scanner cannot read the barcodes    Check that the barcode scanner window is free of grease and dirt   Clean the 
19. ections  Leaking tubing or connections in the fluid system may lead to pipetting errors        Leakage    on page E 25      Air bubbles in one or more pipettes    on page E 26     Dripping pipettes    on page E 26      Dripping probes    on page E 27       gt  Evaporation or deterioration of the sample    1 Use fresh material  if possible     2 Ifthe sample is in a secondary cup  transfer a new aliquot to a new cup from the  original primary sample      gt  System water quality is unsuitable  1 Check the quality of the input water according to the instructions provided by the  manufacturer of the deionized water supply system   You must use reagent grade  type 1 water or equivalent     If the water quality is not within specification  take the necessary action as  recommended by the manufacturer of the deionized water supply system      gt  The wash station is defective    Check for leakage on the bench under the wash station     2 Check that the waste connector is properly connected and that the tubing is not  clogged     3 If any leaks persist  contact Roche Service        Leakage    on page E 25     Air bubbles in one or more pipettes    on page E 26     Dripping pipettes    on page E 26     Dripping probes    on page E 27  Test results low or drifting  Results are obtained  but the values obtained are lower than expected     Results are obtained  but the values show a consistent trend to drift  either up or down   from their expected values      gt  Pipetting incorrect sa
20. er there are messages waiting to be viewed  The background color also indicates  the severity of the message     Button Background Meaning    Gray The messages that are occurring need no action  from the user     ao    should deal with these messages promptly  to    a Yellow Warning messages waiting to be dealt with  You  ensure that a critical situation does not develop              Red flashing Error or critical messages waiting to be dealt  A  Running with  You should deal with the causes of these  messenger  messages immediately     You can clear the red and yellow versions of the Messages button by confirming all  Warning and Critical error messages        Confirming messages    on page E 8    15 Messages  New Messages tab    COBAS  INTEGRA 400    New Messages tab    Use the New Messages tab on the View Message dialog box to review all currently  unconfirmed messages  listed by time      New Messages   Message Log   Options      zZonmo OUV D                                      6 27 96 17 00  6 26 96 14 31  8 26 98 14 31  6 26 96 14 29    Rack position slot position 13 not defined   A hardware problem has been detected in the resource controller    ISE solution standard 1 not on board        Error in file  c  roche instrument Log  Trace tra  occurred     7 16 96 17 01 Diagnostics not started  SUPER  G  Message 3of5      Help On Message     Confirm Print    H    Icon showing the severity of the message  if any     The date and time when the message was created    Short me
21. esults from samples  calibrations  and controls     You can also often see that something is wrong before any results are produced  A  dripping probe for example  will affect the generated results     Preventative maintenance    User Manual   Version 2 0    The most common sources of problems are related to service actions and water purity     e For service actions  the main problem is a failure to keep to the maintenance  schedules as defined by the service actions  Ensuring that the instrument is  correctly maintained is the most effective thing that you can do to reduce  problems     e Poor water quality can cause problems over a long period of time and can be  difficult to troubleshoot especially if the water quality varies from day to day        System water quality is unsuitable    on page E 24    E 16    INTEGRA 400    16 Troubleshooting  Start up problems    Start up problems    This section lists potential problems arising from start up or system initialization     Mains power connector is not plugged in    1 Switch off the instrument   2 Insert the plug into the socket     3 Switch on the instrument     Laboratory power supply has been interrupted    1 Check other electrical power sources and inform your in house service personnel     2 When the power is re established  the system will restart automatically  wait until  the system is initialized     System cannot initialize due to a fault in one or more of the analyzer  modules    1 Check the displayed message s  an
22. lem from the original  message dialog box      If you need to refer to the message again  use the message log     Displaying message details       User Manual   Version 2 0          gt  To display message details    Co    B             elp On Message   Print   Help    cose      Click Messages on the navigation bar   Click either the New Messages or Message Log tab   Double click a message     The View Message   Detail dialog box is displayed     View Message   Detail Xx  Date  amp  Time fi 9 04 2000 14 46 Message ID  19 583    Message        A    Not enough fluid was detected in slot J position 10                     sample volume is too small for the tube being used  transfer  he sample to a smaller tube  if possible     g    Internal Information       Threadname  0x66 DAM_ExpRec Session ID  No Session      0    Message 8 of 158    A Long text of the message   B Click for Help on this message   C Service and diagnostics information     D Click for Help on the dialog box  not message Help      E 9    15 Messages    Stopping and restarting the system    Printing messages                INTEGRA 400    You can print a list of messages from either the New Messages tab or the Message Log    tab     To print a list of messages    1 Click Messages on the navigation bar   The View Message dialog box is displayed   Click the New Messages tab   Click Print     Tip  This procedure prints only non confirmed messages  for problems that have not yet  been resolved      To print the message
23. lts are accepted  you may request a rerun   though      1 Accept the last calibration or control results in the Validate tab in the Results work  area     2 Ifthe last calibration or control result was not accepted because it was flagged   follow the recommended action for the displayed flag     Accepting flagged calibrations without taking appropriate action will cause all subsequent patient  results also to be flagged     e No calibration definition available    e No control definition available    Install the required calibration or control definitions  See Online Help for details     e No fluid in sample cup    No fluid level was detected in a tube or cup  Possible reasons are   o The tube or cup is empty   o The tube or cup is overfilled        Wrong tube definition     Do one of the following   e Refill the sample tube or cup   e Remove some sample from the tube or cup     e Check the parameters Trigger Point and Lower Sample Margin in Configuration    Lab   Sample Tubes     E 19    16 Troubleshooting    Blocked orders  tests  calibrations  or controls INTEGRA 400  Tests  Disabled e Test disabled    Possible reasons are   o Test was manually disabled     o Test was automatically disabled due to a flagged calibration or quality control  result  This is the case if the flag action is set to Stop Test in Configuration    Result Handling     Recommended action 1 Open Status   Test Status   2 Select the required test   3 Select Test Enabled     Tip  Check if the rules define
24. mple or reagent volumes    1 Perform the service action Deproteinize Probes   2 Perform the service action Clean Probes Manually     3 Perform a Check Test     E 24 User Manual   Version 2 0    INTEGRA 400    Hardware    Cuvettes    Leakage    User Manual   Version 2 0    16 Troubleshooting    Hardware    4 If this is unsuccessful   a  Replace the probes   b  Check the pipettes and pistons    c  Check for leaks      Leakage    on page E 25     Air bubbles in one or more pipettes    on page E 26       Dripping pipettes    on page E 26     Dripping probes    on page E 27    Sample has deteriorated    Sample has deteriorated  for example  due to instability of the analyte   1 Use fresh material  if possible     2 Ifthe sample is in a secondary cup  transfer a new aliquot to a new cup from the  original primary sample     This section lists problems associated with hardware functions  particularly regarding  the cuvettes and the fluid system     Cuvettes must meet stringent optical specifications before they can be used on the  instrument  Defective cuvettes are discarded automatically     System discards a large number of unused cuvettes    Remove several cuvettes from the reservoir     2 Check whether they are cloudy  scratched  or dirty  If so  empty the cuvette  reservoir manually and refill with new cuvettes     3 If this is unsuccessful  open the analyzer cover and carefully remove some cuvettes  from the cuvette buffer slide with tweezers  If these cuvettes are scratched o
25. or messages in View Message   New Messages   2 Select the message and click Help on Message to open the Online Help     3 Follow the recommended actions of the selected message     E 21    16 Troubleshooting  Results    Hardware errors    Abs  error    Recommended action    FP error    Recommended action    FP disabled    Recommended action    Fatal error    Recommended action    Results    INTEGRA 400    e Error in Abs  Module    Hardware error in the absorbance photometer module     1 Check for new hardware error messages in View Message   New Messages   2 Select the message and click Help on Message to open the Online Help     3 Follow the recommended actions of the selected message     e Error in FP Module    Hardware error in the fluorescence polarization module     Check for new hardware error messages in View Message   New Messages   Select the message and click Help on Message to open the Online Help     3 Follow the recommended actions of the selected message     e FP module disabled  The FP module was disabled     1 Open Configuration   General   System   2 Select FP Module to enable the FP measurements   Tip    The instrument must be in Standby for enabling the FP module     e Fatal HW error    The system has a fatal hardware error  The robotic transfer  the ISE module and the  analyzer including absorbance and FP module are not working due to a serious  hardware problem  No measurement of tests is possible     1 Check the status of the system in Status   Analyzer 
26. place probe B or C    on page D 44    Valve is defective    A Hardware Warning message may be displayed  but not necessarily  stating that a valve  is defective     e Contact Roche Service     Blocked or partially blocked probes    If you suspect that a probe is blocked or partially blocked   1 Perform a Check Test     2 Ifthe flow checks do not meet the specifications  clean or replace the probes as  necessary        Replace probe B or C    on page D 44    Damaged probes    1 Check the probes visually for damage and replace  if necessary     2 Check whether there is any physical obstruction that might have damaged the  probe  if so  remove the obstruction        Replace probe B or C    on page D 44    Due service actions not performed    Due service actions  related to pipettes and probes  were not performed on schedule     e Use Service   All and perform all due service actions     E 27    16 Troubleshooting COBAS  Computer system and accessories INTEGRA 400    Temperature control    If the ambient temperature exceeds 30   C  the refrigerant pressure becomes too high and  the cooling system generates a hardware error     1 Switch off the instrument at the main switch     2 Ensure that at least one cassette slot is empty  to improve the air circulation inside  the cooling box      Switch on the main switch   Allow the system to start up and complete its initialization     Run the system with at least one spare cassette rack slot for about ten minutes   After this  you can us
27. pty folder in Status   Missing  amp  Blocked  to see the empty activator   deproteinizer  etcher  or ISE solution and their correct rack positions     Remove the ISE rack from the instrument   Refill the empty activator  deproteinizer  etcher  or ISE solution   Reinsert the ISE rack into slot O   The ISE Rack Replaced dialog box is displayed   5 On the dialog box select the ISE solutions that you have replaced     Leave the checkboxes cleared for ISE solutions you have not replaced     ISE module is not ready  The ISE module is not initialized  Possible reasons are      The last ISE initialization was not successfully completed   o Not all required solutions are available  for example Activator or Std 1     o The ISE module was recently enabled but no ISE initialization was performed     e Make sure all solutions are on board     e Perform the service action Initialize ISE module in Service   All     e ISE module disabled  The ISE module was disabled manually     1 Open Status   ISE    2 Select Enable ISE Module to enable the ISE measurements   Tip   The instrument must be in Standby for enabling the ISE module     e No ISE Electrode  The ISE electrode is not defined or disabled     1 Open Configuration   General   ISE   2 Select the required electrode setting   3 Click OK    4 Open Status   ISE    5    Select Lithium  Potassium  Chloride  or Sodium to enable the required electrodes     e Error in ISE module    Hardware error in the ISE module     1 Check for new hardware err
28. r dirty   contact Roche Service     Fluid leaks can sometimes occur  especially when service actions  for example  Replace  Dosage Pipette Plunger Tip  are not properly completed     Leaking pipettes    1 Check that the pipette is firmly screwed into place and tighten  if necessary  do not  over tighten any connections      2 Ifthe leak is from the barrel of the pipette  replace the pipette or piston     E 25    16 Troubleshooting COBAS  Hardware INTEGRA 400     gt  Leaking tubing or tubing connections  1 Check that the tubing connectors are firmly screwed into place and tighten  if  necessary  do not over tighten any connections      2 Check the external water and external waste tubes and ensure they are properly  connected and not damaged     3 Turn off the external water supply and reconnect or replace the water tubes  if  necessary     4 Ifthe leakage is from the internal fluid system  on the instrument   perform a  shutdown and switch off the system  then contact Roche Service     Air bubbles in one or more pipettes  Air bubbles are a sign that the fluid system  most likely a pipette  is leaking or that the    cleaner is running low or the water supply is not properly connected      gt  Pipette not correctly installed    1 Reinstall the pipette or the piston ensuring no air bubbles are trapped   2 Check for air bubbles appearing after reinstallation   3 If air bubbles reappear  call Roche Service        Replace plunger tip dosage pipette  B or C     on page D 41       gt
29. rder in Results   Validate to see the Blocked Status Text   To get help on a blocked item  select the Blocked Status Text and press F1     Reagent not on board    Cleaner not on board    Place the required reagent or cleaner cassette in a cassette rack     Insert the cassette rack into the cassette area     Reagent is empty    Cleaner empty    Locate the rack position of the empty reagent or cleaner cassette in Status    Cassettes     Remove the cassette rack from the cassette area   Replace the empty cassette with a new one     Reinsert the cassette rack into the cassette area     The item is loaded  but there is not enough of it to run the tests in the current worklist     1  2    Locate the rack position of the cassette in Status   Missing  amp  Blocked     Place an additional cassette of the same kind on a rack     Calibrators  controls  and diluents    Not on board    Recommended action    E 18    1  2    Tip    No calibrator on board  Control not on board    Diluent not on board    Load the required calibrators  controls  or diluents on their correct rack positions     Insert the rack into a free slot in the sample or cassette area     To check the correct calibrator  control  or diluent position  do one of the following     Open the Not on Board folder in Status   Missing  amp  Blocked  to see the missing  calibrators  controls  or diluents and their correct rack positions     Click a test folder in Status   Test Status  to see the calibrator  control  or diluent  info
30. ressing  Alt   F4     maintenance functions even after you have switched off  These functions ensure that the system    v7 Because the COBAS INTEGRA 400 has an internal battery  the system performs certain essential  Q  especially the ISE module  can be restarted on request  without manual intervention     E 12 User Manual   Version 2 0    INTEGRA 400    Restarting the system  after shutdown    User Manual   Version 2 0    15 Messages  Stopping and restarting the system    You should perform a restart only after the screen has become dark     To restart the system    1    Switch on the main switch     The system undergoes the complete startup procedure  The COBAS INTEGRA  400 program is automatically restarted  Log on  Shift   F3  when the Log On  dialog box is displayed     Tips    After restarting     o Windows NT is automatically restarted  You do not have to log on to Windows  NT     o The COBAS INTEGRA 400 program is automatically started      The system is re initialized     When you log on  type in your user ID and password as usual  You must rerun tests  that were in progress when the system was stopped     E 13    15 Messages COBAS  Stopping and restarting the system INTEGRA 400    E 14 User Manual   Version 2 0    INTEGRA 400    16 Troubleshooting  Table of contents    Troubleshooting    General problem solving techniques    User Manual   Version 2 0    This chapter explains how to identify and fix problems that can affect the running of  the COBAS INTEGRA 400  It e
31. rmation of the selected test     If you see  for example  a status of Diluent not on board  and you replace the  diluent while the system is in Standby  the values for Status and Next Cal  in the  Test group are not updated until the system is in Operating status  The  information in the Diluents group  however  is updated immediately  You may  have to click  Re Start on the navigation bar to get the system to Operating mode     User Manual   Version 2 0    INTEGRA 400    Empty    Recommended action    Not accepted    Recommended action    No definition    Recommended action    Samples    No fluid    Recommended action    User Manual   Version 2 0    16 Troubleshooting    Blocked orders  tests  calibrations  or controls    e Calibrator empty  e Control empty    e Diluent is empty    1 Open the Empty folder in Status   Missing  amp  Blocked  to see the empty calibrators   controls  diluents  and their correct rack positions     Remove the required rack from the instrument   Refill the empty calibrator  control  diluent  and cups or tubes     Reinsert the rack into a free slot in the sample or cassette area     e Last calibration is not accepted  e Last control not accepted    The last calibration or control result was not accepted and a further calibration or  control request has been made  either by the system  as a result of a cassette change  for  example   or by the user   Note that you can only request a further calibration or  control once the calibration or control resu
32. ssage  Often a prompt to ask you if you really want to do  something     Warning  Alerts you to some condition or consequence  but the system as a    N whole can keep running   Some of these types of messages may mean one or more measurement modules  may not be able to run tests     Critical error  The system cannot continue  You must fix the problem and then  restart the system     All messages are listed on the New Messages tab on the View Message dialog box until  you confirm them        gt  Toreacttoa message    On the message dialog box  click Si to turn off the audible alarm  if necessary   Read the message and take appropriate action     Click Help for additional information  if available     FF Q N      Click Report if you want to create and save a problem report        Creating a problem report    on page E 35    Ol    Click Close when you have read the message     Confirm the message        Confirming messages    on page E 8    E 4 User Manual   Version 2 0    INTEGRA    Messages butt    400    on    Clearing the Messages    User Manual   Version 2 0    button    15 Messages    About messages    Tips  e Ifyou need to refer to the message again after confirmation  use the Message Log  tab        Message Log tab    on page E 7    e All messages that generate a dialog box are listed on the New Messages tab on the  View Message dialog box  Double click a message for more details     The background color of the Messages button shown on the navigation bar indicates  wheth
33. ssage text   Click to get Help on the selected message  if available     Name of the user logged on at the time the message was created   The selected message  Double click to display details on this message   Click to confirm the selected message     Click for Help on this dialog box  this does not give Help on the selected message      Tips    You click Confirm to inform the system that you have fixed the problem in the  message     If you have not fixed the problem  the system will either not restart  or the same  message will reappear again  if the same conditions are met     Confirmed messages are automatically transferred to the message log     User Manual   Version 2 0    INTEGRA 400    Message Log tab    User Manual   Version 2 0    15 Messages  Message Log tab    Use the Message Log tab on the View Message dialog box to review messages that you   or other users  have already confirmed as resolved  Double click any message to  display the Message Detail dialog box     m View Message x     NewMessages Message Log   Options                                8727798 17 01 FP high voltage  W  out of range  300 700   L   E  8 27 98 17 00 Database logon failed  LABTO3   6 27 98 17 00 Rack position slot position 13 not defined  ADMIN   8 26 98 14 31 A hardware problem has been detected in the resource controller         6 26 98 14 31 ISE solution standard 1 not on board  F  6 26 98 14 29 Error in file  c  roche instrument Log  Trace tra  occurred   7 16 98 17 01 Diagnostics not
34. the system is in Standby  you can re initialize  if necessary  Sometimes you will    be asked to do this in the recommended actions for a particular message  You would  also need to do this if your troubleshooting procedure caused you to knock  or move   an initialization post  for example        gt  To re initialize the system    Click Stop on the navigation bar  if necessary    Click Service on the navigation bar    Choose Tools  gt  Initialize System    System initialization takes a few minutes to complete     4 Click Restart  if necessary     Tips  e The system must be in Standby or Fatal Error state when you initialize     e The prompt displayed after you click Restart is not listed here      Stop processing    on page E 10  In all messages and recommended actions in the Help  your reaction to this    prompt is assumed     e Step 4 is not required if Autostart is enabled  in Configuration   General   System      System shutdown    A system shutdown means that the COBAS INTEGRA 400 program is closed down  and the system comes to a complete stop     Shut down the system by pressing the  main switch  A  to the    zero     position     The system informs you that it will be  closing down in 10 seconds     After shutdown     e All tests and procedures in  progress are stopped     e The COBAS INTEGRA 400 main  window is closed        e When the shutdown is complete  the monitor screen becomes dark   Tip    You can also perform a shutdown by choosing File  gt  Shut Down or by p
35. u must connect the water sensor  if you are using an external water reservoir     Data connectors    These connection sockets are for serial data communications  networking  and printer    output  The connectors are located on the right hand side of the instrument near the  back     NVI NVI    xnv    Z WOO        WOO    bid     CE  EE    e      O  L WOO          AUX   Auxiliary connector  not used   COM 1   Serial communications connector  host connection   COM 2   Serial communications connector  modem connection     COM 3   Serial communications connector  not used     mo 0 WB  gt     LPT 1   Parallel port  Printer connection   Tip  e Serial ports  if used  are configured by your Roche Service representative     Connections check list    Use this list if you need to check that your system is properly connected     Mains connection   Check that the main power cable is plugged in at both ends     e Check that the mains switch on the wall socket is in the ON position     E 30    User Manual   Version 2 0    INTEGRA 400    Serial communications   for host computer  installations only     Printer connections   optional     Water supply    Waste outlets    16 Troubleshooting  Checking and replacing fuses    e Check that host computer connection  COM 1  is made  This connection is  required for automatic exchange of orders and results between the host computer  and the analyzer     e Check the modem connection  COM 2  for remote diagnostics  if applicable to  your system     e Che
36. w cassette when  the rack is released     2 Check the results of the last calibration and control     If the results deviate significantly from the previous calibration  order a new  calibration     If the results are still wrong  check the calibrator or the control material  If these  are OK  replace the reagent cassette     3 Check when the next calibration is due  Status   Test Status      If the due time is some time away and the last calibration was performed some  time back  order a new calibration     Test results not reproducible    Results are obtained but are frequently flagged  In particular  when a particular test 1s  rerun  the results are not consistent with earlier values      gt  Air bubbles or foam on the surface of the sample    Carefully remove air bubbles or foam from the surface of the sample     2 When transferring samples to secondary cups  pipette the sample carefully to avoid  the introduction of air bubbles or foam     3 Insert sample racks carefully to avoid disturbing the samples      gt  Fibrin clots or particles in the sample    1 Re centrifuge the sample  if necessary     2 Alternatively  transfer some sample  free of clots or particles  into a secondary cup     User Manual   Version 2 0 E 23    16 Troubleshooting COBAS  Results INTEGRA 400     gt  Probes not clean    1 Perform the service action Deproteinize Probes   2 Perform the service action Clean Probes Manually     3 Ifthis does not work  replace the probes      gt  Leaking tubing or conn
37. w the  recommended procedures     Search the online Help for the information you require  You can perform a full text  search to help you find this information     If this fails  contact your system supervisor for further advice     If you still cannot solve the problem  make a note of any special conditions that  you think might be relevant     Create a problem report   See    Creating a problem report    on page E 35    Contact Roche Service     In the first instance  contact the Roche remote diagnostics support service  This  service allows a Roche service engineer to log on and control the system remotely    See    Remote diagnostics    on page E 36   For this level of support  a modem must  be connected to the specified serial port on the instrument     Creating a problem report    You can automatically create a problem report on a floppy disk  This report collates all  the messages and other system information that the Roche Service engineers will need   if you cannot solve a particular problem on your own  You should create the report  if  possible  before contacting Roche Service        gt  To create a problem report on diskette    Click Messages on the navigation bar    Click the Options tab    Click Report    The Problem Report dialog box is displayed     Problem Report x     Do you want to add DB export to the problem report         Message ID 28 4160       4 Click Yes to add a database export to the report     User Manual   Version 2 0    The Create Problem Report
38. working     The layout of the controller board fuses is as follows       VDOOOOSD    ot ee    zi    b  NG ik E    Lamp Supply FP Lu Pu EN San Lun ZN Fn oa  Fuse 2 AT 20 n BM  AR DM Zn OM GM  so Co    ae Que Lo ze Iov NO  F2 23023230363  gt 3 283   gt  0   x oO   gt L 5u zu L  gt L su yu eu  x ro  3    N  lt   N  lt   cc c       gt     N    Lamp Supply ABS Be F  Fuse 6 3 AT N x  A    Supply     15 V    Fuse 2 AT    Supply 36 V Supply 100 V  Fuse 5 AT Fuse 3 2 AT       gt  To replace a controller board fuse    7  3  9    Wait for the instrument to go into Standby     Lift the main front cover  A  and lock it in the raised position        Identify the fuse s  to replace  LED not lit   Make a note of the fuse type and  rating  as shown next to the fuse     Switch off the instrument at the main switch   Wait for the power down to complete  display will be black      Open the controller rack cover and  using a small screwdriver  turn the fuse holder  counter clockwise     Replace the failed fuse with a fuse of the same type   Reinsert the fuse     Switch on the instrument     10 If the problem occurs again  contact Roche Service     E 34    User Manual   Version 2 0    INTEGRA 400    16 Troubleshooting    Contacting Roche Service    Contacting Roche Service    Roche Service can help you solve many problems that you are unable to solve for  yourself  However  before you contact Roche Service     l     Check this chapter to see if your particular problem is described  If it is  follo
39. xplains how to view and respond to system generated  messages and how to troubleshoot problems that do not generate messages  A  description of how to identify and replace fuses is also included     If you encounter any problems that you cannot resolve  information about contacting  Roche Service is also provided     In this chapter Chapter 16    Introduction    16  Information  about problems    anheben 16  Preventativemanlenantee  kernel 16   LAT UB Problems ne  ne ee see EE EE E E E 17   Blocked orders  tests  calibrations  or controls             sseseessssnesseenesseenesseeseesseseesseseessesee 17  Gassettesand ceanen  o a Gadcasec A AATAS 18  Calibrators  controls  and diluents           cceacaneeeeeeeneseeennnnnssennnnnsnnnnnnnnnnnnnnnnnnnnnnnn nenn 18  STD Ines een ren 19  Test entire t  ten  20  SE aan ae 20  Flat Iware et or ee re 22   Re SS ee 22  Quality control results out of TAN Be wc II 22  Test results notreproducib lenien e T E re asses aa geese e T 23  Testresults low or Ins a RR ATNA 24   Hardware nenne 25  UV CLLES een na ee 25  Eeakase anne A N 25  Air bubbles in One Or more pipetles serie 26  PDE 1S Pipete seele ihn 26  Dippie probs naeh 27  Temperature no eines Nie 28   Computer system and   ccesonesn a nr RI 28   External COMMECIOT Sia  ss  Rasen 29  Waste water  and Power connectersse  c ha Rus 29  Dataconnectors a s ER tee eoneet iuneas ehe 30  Connect ons check TiSi ee eisen 30   Chreckins and PEP ACTS TU Se See EEr NANE e EEEREN ENEE 31  Beplacns a
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Table of Contents  取扱説明書(pdf)  Samsung BF641FST Korisničko uputstvo  Digital Video Surveillance Center Management Software User Manual  Vantec NexStar DX NST-530U2  Sharp FO-475TH User's Manual  暖房便座 脱臭暖房便座  6 - Müller Elektronik GmbH & Co.  GoPro HERO3 User's Manual  Kalorik USK GRB 32231 S User's Manual    Copyright © All rights reserved. 
   Failed to retrieve file