Home
        Notice MW1008P en
         Contents
1.       100       100uH  TD     10 15 Nf K  1mF X   Tour    1uH        300m        d   ApH   10        100         D K        gt  100nH      100mF         100   10m N N  100Hz 1kHz 10KHZ 25kHz    MWI1OOSP  user manual  32    OPERATION MODE    Le MW1008P utilise la technique du pont auto   quilibr   pour d  terminer une imp  dance  inconnue en mesurant la tension aux bornes du composant et le courant le traversant  La  figure A montre le schema bloc du MW1008P et montre comment l   instrument mesure des  impedances inconnues  Un signal sinusoidal est applique sur l   imp  dance inconnue ZX a  travers une r  sistance source Rs  qui change suivant la gamme d   imp  dance  Le r  le de  l   amplificateur    transimp  dance A  est de faire circuler le courant Jr qui circule dans  l   imp  dance inconnue dans la r  sistance de r  f  rence Rr  tout en maintenant le point de  Jonction de Zx et Rr    zero volt  masse virtuelle   Aux bornes de      nous trouvons une tension  V2   Ir    Rr  Les tensions VI et V2 respectivement pr  sentes aux bornes de l   imp  dance  inconnue et aux bornes de la r  sistance de r  f  rence  sont aiguill  es vers un amplificateur  diff  rentiel   2 par le commutateur A L   utilisation du m  me amplificateur pour la mesure de  la tension et du courant permet de s   affranchir des facteurs d     chelle et de ses d  rives  Les  parties r  elles et imaginaire de la tension et du courant sont obtenus en multipliant ces  tensions par un signal carr   coh  rent ave
2.   8  Pressez la touche pour sauvegarder les param  tres d     talonnage en m  moire  non volatile  L   instrument doit afficher fugitivement    Passed    pour signaler que la  m  morisation s   est bien d  roul  e     MWIOOSP  user manual  36    Procedure de modification des valeurs des r  sistances    L   afficheur indique    RCALX  XXXXXX        Pressez la touche  choix EDIT  pour  modifier la valeur  Un curseur appara  t sous le premier chiffre    gauche  Le chiffre peut   tre  incr  ment   par la touche ou d  cr  ment   par la touche  MENU    La position du curseur peut   tre modifi   par la touche  d  placement vers la droite  ou    par la touche  MODEL   d  placement vers la gauche    Validez la valeur en pressant la touche sous OK     La premi  re   tape consistant    param  trer l     talonnage est maintenant termin  e     ne nous  reste plus qu      proc  der    la calibration proprement dite     CALIBRATION    Avant de pr  c  der    la calibration la premi  re   tape d  crite pr  c  demment doit   tre effectu  e  dans son int  gralit    La proc  dure de calibration comprend plusieurs   tapes qu   il faut r  aliser  dans l   ordre  A partir du menu    PARAM CAL EXIT         1  Pressez la touche  n Vs M   choix    CAL   L   instrument affiche l     cran  ci contre  L   instrument va proc  der     une calibration en circuit  ouvert   L   instrument se servira de cette calibration pour compenser les   l  ments  parasites lors des calibrations des r  sistances  Il est pa
3.   de phase  ici   Vi   0 79745     0 23967   0 55778    of il       Vous pouvez effectuer la m  me chose avec le courant en pressant la touche L C R  Les  valeurs affich  e sont des valeurs normalis  es issues directement du convertisseur  allant de 0     1     MWIOOSP  user manual  42    ORMELABS  1  All  e des rochers  94045 CRETEIL   FRANCE    Web   www ormelabs com Email   infos ormelabs com  Tel   33  0  9 51 23 74 80    MWI1OOSP  user manual  43    
4.   pressing of this key changes the measurement impedance range  1     6   Pressing this key for  longer than two seconds returns the unit to auto ranging or normal mode  The range is  displayed in the top right corner of the display  The    A    character behind the Range number   indicates that the instrument is in Auto Range mode  In Manual or Hold ranging mode  this  letter becomes a blinking    H    character  Pressing this key for two seconds or longer returns to    MWI1OOSP  user manual  9    the Auto mode  See also chapter    Impedance ranges    page 24 for more details about  impedance ranges     MODEL    The MODEL key selects between a series or parallel equivalent circuit model for the device  under test  Pressing this key for two seconds or longer places the Model selection in Auto  mode  In this case the instrument selects the most appropriate model     MENU    The  MENU  key allows access to a series of special configurable parameters  Pressing this  key displays the programmable options  The current state of each option is displayed on the  first line of the LCD as that option is selected  To move through the Menu  press the key  under the forward or back arrows displayed on the LCD     Default value  Backlight ON The backlight can be set to ON or OFF  To turn the  backlight off  press the key under OFF   This option turns the audible alert function ON or OFF     Set this feature to ON to compensate for random noise   eg that 1s apparent when measuring some comp
5.  5   C  Base precision  0 2    150  lt   Z   lt  300kQ et f  lt  1kHz     See the accuracy section for detailed accuracy specifications    Various   Fixture  4 wires Kelvin on CB connector  Protection Protected up to 1 Joule of stored energy  100 VDC max  for  charged capacitors   Zeroing open short circuit compensation    Compensation limits Short      lt  20 Q  7   lt  50 Q  Open   Z   gt  10         Sorting Selection of tolerance  1 2 5 10 or 20    Averaging 2 to 8 measurements  General     Operating Conditions  0     50   C   lt  80  relative humidity    Power supply 7 5     15    150 mA with light   lt 100 mA without  Dimensions 6 125 in x 1 5 in x 3 875 in   WxHxL   Options   MW108 Varactor adapter  MWI0 Universal 4 wires Kelvin cable  mini pincers     MW KELV Kelvin pincers cable  MW SMT SMT cable    TERMS AND SYMBOLS    Rs or ESR   Serial resistance            Real part of impedance ene    A Reactance    ohm  Q   imaginary part of impedance  G Conductance  See     Real part of admittance  Y    Susceptance  siemens  S    MWI1OOSP  user manual  5          D   Dissipation factor        Phase angle of  Z     Primary voltage     transformer    transformer     Imaginary par of admittance    PREFIXES OF MEASUREMENT UNITS    Multiplier    1000000   1000   0 001   0 000001  0 000000001  0 000000000001    GLOSSARY    Coil      Capacity      Capacitor     Test frequency      Range     Impedance      Scientific Ingineer Symbol  10   Mega M   10   Kilo k   10   milli m   10   micr
6.  Key  LICIR X  Increments N1 of 10 units  Key   Increments      of one unit  Key  MODEL   Decrements   1 of 10 units  Key  MENU    Decrements   1 of one unit  Key  Frequ    Multiply N2 by 2   Key  ISP     Divide N2 by 2    MWIOOSP  user manual  24    IMPEDANCE RANGES    The MW1008P has 6 ranges of impedance       RI R6   The impedance range can be   Range 2 Resi    Inductance   Capaci   source   stance  H  tance  selected manually or the instrument can be           m  i 0 01 uH   99 9 mF   10Q 2 4 f 10 6 f mF  10 Q   2 4 f   10 6 f mF   1009   3300 52 5 f 482 f uF         330 Q    52 5 f    482 f uF   3 3      525 f 48 2 f uF       13 3 kQ    525 f    48 2 f uF   33 KQ 5252 f 4 82 f uF  kQ    1 1  11   111    configured in auto range mode  which is the  normal mode of the instrument  The current  range is indicated on the right of the first line  of the display  The letter behind the range  number indicates is the instrument is in auto  mode  A  or manual mode  blinking H   A  long press on key       switches         instrument to auto mode  A regular press    33kQ    5252 f    4 82 f uF   800kQ   52520 f 0 48 f          to f th       ro   800K2    52520    0 48 f pF   allows to freeze the current range  Successive      100 MO 9999 0 001 pF   presses allows to choose the desired range     The table below gives the limits of impedance for each range  F corresponds to the test  frequency in kHz  Ranges      and R6 are extended ranges since they use reference resistances  from r
7.  appropriate range  The manual mode allows to get a faster  response since the instrument doesn t have to search for the correct range for each measure  It  is besides one of the few cases where manual mode should be used     SERIES AND PARALLEL MODELS    The key allows to choose the model to use  Please refer also to    PASSIVE  COMPONENTS     paragraph above  In case of doubt use the auto mode by a long press on this  key or use series model which 15 most generally appropriate     DISPLAY MODES    The key  DISP  Allows to select the display mode among three different ones  Successive  presses on the key DISP  allow to get the following display modes      e The measured value    e The difference between value to the current value  The symbol A next to the parameter  indicates that this function is active  If the sorting mode 1s active  the message PASS   FAIL will be displayed on the second line according to the value of the component   the nominal value and the tolerance percentage defined by the user     e The difference in percentage to the current value  The symbol A and unit in    indicates that this function 1s activated     MWI1OOSP  user manual  23    TEST CONDITIONS    TEST FREQUENCIES    There are ten possible test frequencies   100 Hz  120 Hz  250 Hz  500 Hz  1 0 kHz  2 5 kHz  5  kHz  10 kHz  12 5 kHz  15 625 kHz and 25 kHz  An additionnal frequency can be defined by  the user among more than 500 possible frequencies between 100 Hz and 25 kHz  The  precision of th
8.  au primaire et ainsi  d   obtenir une tension secondaire appropri  e lorsque le rapport de transformation est    lev       En mode de mesure d   inductance mutuelle  l   instrument mesure le courant primaire et la    tension secondaire pour d  terminer une imp  dance Zu   vs   ip puis l   inductance  mutuelle est calcul  e        M          o    Source    Resistor       100       MWI1OOSP  user manual  35    CALIBRATION    Le MW1008P peut   tre calibre a l aide d un kit de six r  sistances   talon  10     100     1        10       100      et 1 MQ  ou d un calibrateur externe  La valeur de chaque r  sistance est not  e  sur le l   emballage de ces r  sistances     La calibration n  cessite une connexion Kelvin    quatre fils  Cette connexion peut   tre r  alis  e     l   aide d   un des c  bles disponibles  MW10 ou MW11   La calibration doit   tre effectu  e une  seule fois  Toutefois nous recommandons de refaire la calibration tous les ans afin de se  pr  munir contre les d  rives a longs termes  Les coefficients calcul  s par l   instrument durant le  processus de calibration sont sauvegard  s en m  moire non volatile  Un checksum de la  m  moire est effectu      la mise sous tension et un message d   erreur appara  tra en cas de  probl  me     Mettre en marche le MW1008P et attendre un minimum de 15 minutes avant de commencer  la proc  dure de calibration  La proc  dure comporte deux   tapes diff  rentes  La premi  re   tape  consiste d   une    entrer les valeurs des r  
9.  circuit below  valid for impedance ranges 1 and 2      Point LD corresponds to a virtual    ground  the capacitor 1s thus polarized HS LS  by external voltage VBIAS  This R1 C2  voltage must be well filtered  one will Si Lo  avoid the use of switching power Capacitor  L    under test    supply  For safety reasons one will  limit to 40V the maximum bias    voltage  The value of the Cl capacitor  will have to be such as its impedance 15  low compared to the source resistance   100    for ranges 1 and 2   A value of  470 uF 63V will be sufficient in most VBIAS  lt  100V  cases  The value of resistor R2 will be  of 390     A short zeroing should be  performed replacing the capacitor by a    D       MWI1OOSP  user manual  15    short circuit  One has to wait until displayed value is stabilized because of the circuit time   constants  It 1s possible to use this circuit for low capacities  in this case it 1s necessary to  adapt R2 and C1 values to resistance source     One will choose R2 so that its value is large compared to source resistance       capacitor must  be large enough not to increase resistance source too much  One will take for R2 a value 4 or  5 times the source resistance  and for      a value such that its reactance is negligible compared  to source resistance for the 100 Hz frequency     100    470 uF 390     100    470 uF 390       10      4 7 F  100 kQ 470 nF 390         6   100KQ 470 nF 390 kQ    3 9       3       Inductors      Inductors are measured in Henrie
10.  key  and step backward or forward using the appropriate arrow  key to the AL display option  Select OFF and press the  Menu key again  Save the current configuration        MWI1OOSP  user manual  11    DISP     The ISP   key selects the manner_that the value of a component will be displayed  If the  Sorting mode is disabled  pressing  DISP   cycles through the following display types       The value being measured      The deviation of the value from the current value     The D symbol next to the measured  parameter indicates that this function 15 active      The percent of deviation from the current value   The D symbol next to the measured  parameter indicates that this function 1s active    In Sorting Mode pressing cycles through the following display types       The value being measured      The percent of deviation from a stored value   The D symbol next to the measured parameter  indicates that this function is active  In this mode  a PASS FAIL message 15 shown in the  second line of the display according to the measured deviation and the selected tolerance     CAL    The key allows access to open short compensation  Pressing the CAL key  displays the zeroing options on the second line of the LCD     Open Short Exit  This option  will zero the LCR meter for the currently selected test frequency  Pressing the CAL key  for more than two seconds displays the zeroing options   OPEN SHORT Exit  note that the  options are in all capital letters   This option performs an open
11.  pour la tension  Chacun de  ces deux vecteurs n   a aucune signification propre  par contre le rapport de ces deux vecteurs   division des modules et diff  rence des phase  donnent l   imp  dance  Veuillez aussi vous  reporter au chapitre DIAGNOSTIC pour de plus amples d  tails    Le signal issu du d  tecteur de phase est int  gr   sur une dur  e multiple de la fr  quence de test   et la plus proche possible de 20 ms ou 16 67 ms selon que l   on est en pr  sence d   un secteur     50 Hz ou    60 Hz  La valeur r  sultante est obtenue en mesurant la d  charge du condensateur  d int  gration    l   aide d   un compteur     MWIOOSP  user manual  34    MESURE DES PARAMETRES DES TRANSFORMATEURS    Cette fonctionnalit   permet la mesure des param  tres de transformateurs grace    un    En mode    commutateur analogique  qui permet la mesure de la tension High Sense pour le  secondaire ou de la tension High Drive pour le primaire  Les points Low Drive et  Low Source sont connect  s a la masse via une masse virtuelle    de mesure du rapport de transformation l   instrument mesure successivement la  tension primaire et la tension secondaire  puis effectue le rapport vectoriel de ces  deux grandeurs  Le primaire du transformateur est aliment   par le g  n  rateur du  signal de test via la r  sistance source  La r  sistance source est de 100Q pour les  gammes 2 et 3  et 1      pour la premi  re gamme  Cette derni  re valeur de  r  sistance source permet de r  duire la tension appliqu  e
12.  short calibration through the  entire range of test frequencies     MWI1OOSP  user manual  12    OPERATING    POWER ON THE INSTRUMENT    The instrument can be powered either by a 9V battery or by an external power supply  In this  last case the power supply must be well filtered and its voltage must lie between 8 and 15     Consumption is 150mA approximately when the backlight is activated  In the case of battery  use a warning message 15 displayed when the voltage is below 7V  In this case the backlight is  automatically deactivated  Besides we advise you not to use lighting when the instrument is  supplied by battery  which makes it possible to divide by two the consumption of the  instrument  The battery replacement requires the opening of the battery compartment     PASSIVE COMPONENTS    All non ideal passive component  resistor   capacitor or inductor  can be represented by  a real part  the resistance  in series or in  parallel with an imaginary part  reactance   inductance if positive or capacitance if  negative   The impedance varies with  frequency  The series or parallel model are  equivalent at a given frequency and one can  goes from one model to the other one thanks  to the equations below    Generally one model is more appropriate  than the other one in a given measurement  conditions  that is to say when the series or  parallel resistance represent a physical  property of the part    Temperature can have a large impact on the  DUT impedance  Usually  capacitor
13.  shown on the bottom line  The  number of displayed digits and  the location of the decimal  point are automatically    Indicateur Param  tre d  onction auto principal mode automatique    mod  le Param  tre fr  quence  automatique secondaire de test       adjusted according to the range and resolution  The A symbol in front of the major parameter  indicates that the measurement is displayed as a relative or absolute deviation from a nominal  value  A dark arrow present in the top left of the display indicates the unit is in the Auto  Parameter mode  If a dark arrow is displayed in the bottom left of the LCD  the unit is in the  Auto Model mode  The Range 15 indicated at the top right of the display  The    A    character  behind the Range number indicates that the instrument is in auto range mode  In Manual or  Hold ranging mode  this letter becomes a blinking    H    character  The selected test frequency  is displayed on the bottom right of the display  The word    USER    15 displayed when the    user frequency 15 selected     MWI1OOSP  user manual  8    KEYPAD    The keypad is used to select  measurement conditions and to    enter values  All keys have two            functions  depending      LICIR N Vs M Freq   Hold Range  whether the key is pressed AUT AUTO    momentarily or for two d     E    seconds or longer   MODEL a DISP CAL  L C R AUTO                 This key selects the parameter being measured  Pressing the L C R key steps through the  major parameters to manua
14.  thickness than the  component to measure  and proceed to open circuit calibration     Case of measurements on transformers    The OPEN SHORT calibration must not be used     MWI1OOSP  user manual  27    CONNECTING TO A PASSIVE COMPONENT    The MWI1008P uses a Kelvin type  connection with 4 wires to measure the  component  This allows the instrument to  let a current pass through the component  using wires LD  Low Drive  and HD   High Drive   and to measure the voltage  at its terminals using wires HS  High  Sense  and LW  Low Sense   This  configuration increases the precision for  low impedances    The MWI10 cable included in the  package has 4 mini pincers  This cable  can be used as well for for standard 2  terminals passive components as for  transformers  The 2 red mini pincers  correspond to signals HD and HS  while  the black ones correspond to signals LD  et HS  The red sleeve one one of the  wires correspond to the current signals    The impedance is measured  between those two points       LD LS HS HD     HD and LD   The use of Kelvin pincers cable  MW KELV  is easier because the signals  LD LS and HS HD are internally wired inside each pincer     When you measure weak or strong impedances  it is necessary to perform a calibration of the  measurement cable used  see paragraph before     MWI1OOSP  user manual    28    PINOUT OF THE CB CONNECTOR    The pinout of the CB type DUT connector is given below  side view   This allows to make  your own test cables  For a cla
15.  user definable     The LCR meter is controlled by a high speed micro controller with embedded logic that  controls the display and keypad  as well as setting measurement conditions and performing  calculations  Please refer to chapter    Operation    for more details about instrument operation     SPECIFICATIONS   Functions Auto  L Q  C D  R Q   Z  0  R X  G B  N 0  N  0  Vs Vp   M  L AL  C Vr  varactor option   Equivalent circuit Series or parallel    Displayed parameters Value  Deviation    deviation  Measurement display    L Q  L 0 01 uH     99 99 H  Q 0 0001     100  C D  C 0 001 pF     99999 uF  D 0 0001     10  R Q  R 1          99 9       Q 0 001     100  IZ  0  L   1 mO     99 9 MQ      180 00      180 00    R X  R 1          99 9       A 1          99 9       G B          N 0  N 1     9999      180 00      180 00      N  0 N   0 0001     1      180 00      180 00     Vs Vp Vs 230V N      LI5V N  resolution 0 01 V  Vp 115V or 230V   M     O01uH 9999H   LAL L 0 01 uH     99 99 H  Au _L N     N user definable from 1 to 999     with varactor adapter MW 108      C Vr    0 001 pF     99999 uF  Vr 0 00 5 0V or 0 0     28 0V    Test conditions      Test frequencies 100 Hz  120 Hz  250 Hz  500 Hz  1 kHz  2 5 kHz   5 kHz  7 8125 kHz  12 5 kHz  15 625 kHz  25 kHz  Test voltage 0 5 Vrms  10   no load     MWI1OOSP  user manual  4    Measurement rate 2 measures per second  without averaging     Ranging Auto or manual   Precision    Conditions At least 15 minutes warm up  23   C  
16.  value ix Q   500  is displayed in the first line  the  reactance in the second one  These  two values are respectively the real  part and the imaginary part of the  DUT  Units are Q  kQ or MQ     Conductance and Susceptance  These two  values are respectively the real and  imaginary part of the admittance Y    1 Z  Units        S  mS or LS     D   500  Q   0 002          500  switches the instrument in ee  automatic mode  The instrument  determines the function the most  appropriate function according to  some criterias resumed in the  graphic below    When       impedance is below 10 mQ the  resistance function is selected        MWI1OOSP  user manual  19    MEASUREMENTS ON TRANSFORMERS    The MW1008P has a special function for measuring a transformer ratio between primary and  secondary windings  calculation of the equivalent secondary voltage for a supply transformer       and for measuring mutual inductance between primary and secondary windings   press key n Vs M     The primary winding of the     transformaer shall be connected on VA   wires HD and LD  with a red     sleeve   The secondary winding   inversion  the instrument displays LS HS   an overflow message      OVERFLOW     on ranges 2 and 3  PRIM AIRE SECONDAIRE  The instrument has 3 measurement ranges according to the value   of the turns ratio  The instrument places itself on the first range   voltage is attenuated by a higher source impedance  which   allows a wider secondary voltage measurement range  The rea
17.  with a Q less than 1 will not be automatically detected  The LCR meter will  default to the Rs mode  Increasing the test frequency to where the inductor Q is greater than 1 will  then switch the major parameter to L  If  at the highest test frequency  the    does not raise above 1     manually selecting the L mode will give the value of the inductor     Inductors biasing    Inductor  under test       For measurement in range 1 or 2 one will take the following values      R2   R3   390 Q 1W  C1   C4   470 uF 160V    MWIOOSP  user manual  17    Bias current is I    VBIAS  R2 R3 r     When using this circuit one will avoid to disconnect the insductor while a current is flowing  into it  VBIAS voltage should be decresed down to OV in order to discharge the coil     Opening the circuit can introduce a high voltage with high energy which can be dangerous  id the inductance or the current is high        Resistors    The unit of measurement for resistance 1s the Ohm  Of the three basic circuit components   resistors cause the fewest measurement problems  This is true because it is practical to measure  resistors by applying a dc signal or relatively low ac frequencies  Resistors are usually measured at  dc or low frequency ac where Ohm s Law gives the true value under the assumption that loss  factors are accounted for     Model    For low values of resistors  below 1       the choice usually becomes a low frequency  measurement in a series equivalent mode  Series because the reacti
18. 008P allows comparison of components parameters to a pre defined value  The  instrument will calculate the difference to the pre programmed value  which could be  displayed in absolute difference or percentage formats    To access sorting function  press key Menu  For the other  options of menu   please refer to description of the Menu key   described in page 9        Press the key below right arrow  n Vs M  until the option  Sorting appears  Validate this function by pressing key below   ON      To define the tolerance percentage  press the key below option  Next until the desired tolerance is displayed  Possible values  are 1   2   5   10   and 20     Press the key below right arrow    Enter nominal value of the component  For this  press the key  below EDIT     A cursor appears below the figure to be modified  To change  the figure press the key below option CHANGE  The figure  will increment by one unit at each press  To go to next figure  press key NEXT  The cursor will move below the next figure   Modify this way the value until getting the desired value  The  unit at the end of the number can also be modified  The  available options            pico  10            nano 10        micro 10       m  milli 10     white  as  displayed   k  kilo 10     and M  Mega 10     When the value is entered  press the key below  option OK    The entered value appears now on the first line    Press the key below right arrow        Configure the alert notification when the measured compon
19. AL    C D    R Q     Z  0    R X    G B    AUTO Pressing more than 2s the L C R key    Inductance and quality factor  Inductance is displayed in the first line of the LCD   Units are     mH ou       Q is the ratio of the imaginary part to the real part an  dis unit less  If Q 1s positive the component 15 inductive  capacitive in the other  case     Inductance and     parameter  This values replace L Q values when the AL mode is  selected       value is computed from the inductance and a given turn number    AL   L  n     Capacity and dissipation factor  The capacity is displayed in the first line  The capacity  is either the series equivalent capacity  Cs  either the parallel equivalent  capacity  Cp   Units are pF  nF  uF or mF  D is the ratio of the real part to the  imaginary part  If the value displayed is negative the component is inductive     The resistance is either the series resistance  Rs or ESR  or the parallel resistance  Rp   of the DUT  Units are Q  KQ or MQ     Absolute value and phase angle of the DUT  The absolute value is displayed in the  first line  The phase angle corresponds to the phase difference between the  current and the voltage  The instrument can measure phase angles between   180 00   and  180 00    A negative value indicates a capacitive component  a  positive value indicates an inductive component  High quality capacitors have  a phase angle closed to  90 00    Resistors have a phase angle closed to 0       Resistance et Reactance  Resistance
20. MIRE  LABS          RLC bridge MW1008P    User Manual       Copyright Ormelabs    2010  http   www ormelabs com    MWIOOSP  user manual  1    Table of contents    General Information  Specifications  Terms and symbols  Prefixes of measurement units  Glossary    Generalities  Display  Keypad    Operation  Passive Components  Displayed parameters  Polarisation of electrochemical capacitors  Polarisation of coils  Measurement of batteries impedance  Components sorting  Serial and Parallel models  Displays types  Test conditions  Test frequencies  Impedances ranges  Open short calibration  Connecting to a passive component  Connecting to a transformer  Pinout of DUT connector    Precision  Precision of resistances impedances  Precision of inductances  Precision of capacitors    Theory of operation  Signal generator  Phase detector and analog digital converter  Transformers    Calibration  Parameterizing the calibration  Calibration  Error messages    Diagnostics  Keyboard test  Buzzer test  1    complex  A D results    MWI1OOSP  user manual  2    14  15  17  18  18  20  21  21  23  23  25  25  24  27  28    29  30  30    29  30  30    32  33  34    35  36  36  37    MWIOOSP  user manual  3    GENERAL INFORMATIONS    The MW1008P LCR Meter is a multi frequency impedance measuring instrument capable of  measuring resistance  capacitance  inductance or transformer parameters from 1 mQ to 100  MQ  The MW1008P LCR meter has a basic accuracy of 0 2  and has 11 test frequencies plus  one
21. WI1OOSP  user manual  39    Lorsqu   on appuie sur une touche le r  sultat  a  de la conversion est affich   que la position a     de la touche  Ici la touche L C R est SCENE  Le CAL  appuy  e  donnant la valeur 1023  La croix en           ts  haut    gauche symbolise la touche appuy  e    elle reste affich  e lorsque la touche est rel  ch  e        L     cran ci contre montre      que l   on devrait  obtenir lorsque la touche n Vs M est    appuy  e        La tension issue du clavier peut   tre mesur  e avec un multimetre  imp  dance d   entr  e 10M   aux bornes de R70  Vous pouvez comparer les r  sultats avec ceux donn  s dans le tableau ci   dessous  Les valeurs peuvent diff  rer de quelques pourcents    cause de la tol  rance des  composants   quipant la carte clavier     Tension Valeur  Touche  R70  affich  e   V     Test du buzzer       Ce test est utile lorsque l   on rencontre un     probl  me avec le buzzer  Appuyez sur la a ECS  touche sous ON pour activer le buzzer  ou      sur touche OFF pour d  sactive le buzzer   Lorsque le buzzer est actif on devrait  mesurer une tension d   environ 3 2V aux bornes de celui ci     Pik     a ft 5       i v complexes    Ce mode donne acc  s aux composantes en phase et en quadrature qui sont utilis  es pour  calculer la valeur finale  L   instrument d  termine deux vecteurs dans le domaine complexe  le  premier pour la tension et le second pour le courant  Chacun de ces deux vecteurs comporte  une composante en phase  r  elle  eue 
22. an depend on the quality of the device  but primarily the capacitor s measured value most  closely represents its effective value when the more suitable equivalent circuit  series or  parallel  is used  To determine which model is best  consider the impedance magnitudes of the  capacitive reactance and Rs and Rp  Remember that reactance 15 inversely proportional to      so a small capacitor yields a large reactance  This implies that the effect of parallel resistance   Rp  has a more significant effect than that of Rs  Since Rs has little significance m this case   the parallel circuit model should be used to more closely represent the effective value  The  opposite is true when C has a large value  In this case the Series Resistance  Rs  is more  significant than Rp  thus the series circuit model becomes appropriate  Mid range values of C  require a more precise reactance to resistance comparison but the logic remains the same  The  rule of thumb for selecting the most appropriate model should be based on the impedance   Z    of the capacitor    Above approximately 10 kQ   use parallel model   Below approximately          use series model   Between these values   follow the manufacturer   s recommendation    Polarisation of electrolytics capacitors  The MW1008 generates a sinusoidal signal of 0 5Veff centered on OV  In certain cases it can    be necessary to polarize the capacitor to avoid the inversion of polarity across the capacitor   This can be carried out simply by the
23. anges R2 and RS           R2  R3  R4  RS   6       Note   The ranges are impedance ranges and not capacitance or inductance ranges  The  manual choice of a range implies to know the impedance of the component at the current test  frequency  The manual test mode should be reserved to measure similar components which  impedance is known and doesn t vary too much  In all other cases the instrument should be  configured to auto mode  The choice of a wrong impedance range will give an unstable  measurement or the display of message    OVERFLOW           The source resistance of the generator 1s linked to the range as indicated in the table below  R  SOURCE column   The real voltage at the component terminals will then depend on the range  and on its impedance as shown on the graphics below         Voltage  V        1 E 01 1    00 1 E 01 1 E 02 1 E 03 1 E 04 1 E 05 1 E 06 1 E 07 1 E 08  Component impedance    MWI1OOSP  user manual  25    OPEN   SHORT CALIBRATION    This function allows to compensate for the parasitics elements of the test cable in order to  take them into account during value display  The model used 15 the one shown on the figure  below      Ym   Virtual admittance  Zy    Measured Impedance  Z595   Residual impedance       Ras   Residual resistance  Lss   Residual inductance     pe   Stray admitiance  Cpp   Stray conductance  Cop   Stray copacitance        oi Nermol meosurement mode       os e      ti             ql            D   55   hi OPEN zero mode fc  SHORT z
24. c le g  n  rateur sinusoidal  D  tecteur de Phase   Une  r  ponse proportionnelle a la partie en phase ou en quadrature est obtenue  La mesure finale est  confi  e    un convertisseur double rampe qui est lu par le micro contr  leur  Le rapport de ces  deux vecteurs donne l   imp  dance inconnue  Celle ci est corrig  e par des facteurs de  calibration  pour le module de l   imp  dance comme pour la phase  Les param  tres mesures tels  que L  C  R  Q  D sont math  matiquement d  duits de l   imp  dance mesur  e  selon le mod  le  choisi  s  rie ou parall  le  et la fr  quence de test choisie par l   utilisateur     Phase  Delector    Differential  Amplifier       SIGNAL GENERATOR    Le g  n  rateur de signal a pour r  le de g  n  rer un signal sinusoidal    faible distorsion et dont  la fr  quence est param  trable  La fr  quence peut   tre choisie parmi les choix  suivants   100  120  250  500  1 k  2 5     5     7 8     12 5     15 6 kHz et 25 kHz  La  niveau est fix      0 5V eff    Le signal sinusoidal est g  n  r   a partir d   un signal carr   en utilisant un filtre passe bas  elliptique programmable du 5  ordre    capacit  s commut  es  La fr  quence de coupure est  d  termin  e par la fr  quence du signal d   horloge et vaut 1 28 fois la fr  quence du signal de  test  La sortie du filtre est un signal sinusoidal sur   chantillonn     lt 100   un simple filtre  passe bas du premier ordre  LPF2  supprime une grande partie des harmoniques restantes  La  distorsion harmoni
25. composante en quadrature  imaginaire   Les  coordonn  es absolues de ces deux vecteurs sont relatives    la r  sistance de r  f  rence de la  gamme en cours  Pour donner un sens physique    ces vecteurs il faut les multiplier par la  valeur de la r  sistance de r  f  rence en cours  qui d  pend de la gamme     Dans ce la gamme et la fr  quence peuvent   tre modifi  es  La touche L C R permet de choisir  entre les coordonn  es de la tension ou du courant     MWI1OOSP  user manual  40    Le   crans ci contre monte le r  sultat lorsqu   un condensateur de IuF est connect   a  l   instrument avec une fr  quence test de 1 kHz  L   imp  dance th  orique du condensateur est  d   environ 159Q at 1kHz  S  lectionnez donc la gamme 2 pour obtenir un rapport signal bruit  maximum     La r  sistance de r  ference typique est de 100  Q pour la gamme 2  Les coordonn  es du  vecteur tension sont 55 7Q pour la partie  r  elle et  34 7Q pour la partie imaginaire        Pressez la touche L C R pour acc  der aux coordonn  es du vecteur courant     Les coordonn  es de ce second vecteur sont  21 9 Q pour sa partie r  elle et 32 1Q pour sa  partie imaginaire           est possible de calculer le module de     l   imp  dance du condensateur  CN E  Z    100 x          100    65 62 41 36     A    158 60  En r  alit   cette valeur pourra   tre 3  l  g  rement diff  rente car nous avons suppose   55 70  que la r  sistance de r  f  rence vaut 100 ohms 3  exactement  Les coefficients de calibration ne  so
26. e frequency is  0 01    100 ppm   The current test frequency        be changed  by pressing key Frequ    The frequency is shown on the right of the second line    The test frequency most currently employed is 1 kHz  It s the frequency for which instrument  precision is at its best  The large capacitors and inductors should be measured at low  frequency  while low values of capacitors and inductors should be measured at high  frequency  The large choice of frequency allows for example to test audio transformers on the  whole range of audio frequencies   The 120 Hz frequency must not be used when the  instrument is configured for a mains frequency of 50  7  same as the frequency of 100 Hz  should not be used for a mains frequency of 60 Hz     The user frequency allows to choose with precision a particular frequency  To select the  frequency  press the key  Frequ   Until getting the message       USER    in place of the  frequency  To know of change this frequency  press the key  MENU    Select the option    Usr  f   with keys below the arrows  lt     and      To modify the frequency press the key below     Edit       The frequency 15 determined using the 2 parameters      and N2 according to the following  formula      f  125 kHz    N1    N2     The maximum admissible limit is 25 kHz obtained for N1 5 and N2 1  The lower  functionnal limit is around 90 Hz  N1 can be adjusted between 2 and 255  N2 can take the  values 1 2 4 or 8        Modification of coefficients N1 and N2     
27. en the measured value and the real value of a  component  Accuracy is expressed as a percentage for the main  parameter  The accuracy depends on the impedance and the test  frequency  Generally the secondary parameter accuracy is an absolute  value     The actual accuracy of the instruments depends of some parameters  like test frequency and impedance  The basic accuracy is the best case  accuracy that can be expected  this means 1 kHz test frequency and  impedance between 10    and 100          Output resistance of the test signal generator  The DUT 1s connected to  the signal generator trough this resistor  The source Resistance depends  on the impedance range     Resolution 15 the smallest quantity that the instrument can display  Do  not confuse this value with accuracy     This is the rms value delivered from the instrument with no load for  creating a current through the device under test  The source resistor as  well as the actual DUT impedance imply that actual voltage across the  DUT 15 always below this value     MWI1OOSP  user manual  7    GENERALITIES    This chapter gives an overview of the MW1008O characteristics  For further informations  please refer to the operation section        DISPLAY    The two lines of the LCD show  measured values  selected  parameters  instrument status  and various messages  When  making normal measurements   the major parameter  L  C  R     Z   G  is shown on the top line  and the appropriate minor  parameter      D  X           is 
28. ent  satisfy the sorting conditions  The default notification is a  short beep  To change this option  press the key below option  Next  The possible options are SHORT  short beep   LONG   long beep   NONE  no beep   Press the key below right  arrow     The    backlight    option allows to configure the display lighting  as a visual indicator when the component is within specified  tolerance  Do not confuse with backlighting configuration that  is in main configuration menu  Please note that the back lighting must be activated for this  function     MWI1OOSP  user manual  22    The default value is YES  which means that the back lighting will bright only if the  component value is inside permitted tolerance  To invalidate this function press the key below  option  NO   Press the key below right arrow     To save the current configuration  press the key below YES   The configuration will be memorised in non volatile memory Sags consis      and will be kept even after turning off the instrument  Choose    NO if you don t want to keep the parameters after turning off the instrument        To use the sorting mode  the instrument should be configured in relative measurement by  pressing key   DISP    In sorting mode  the instrument places itself automatically in manual  mode and keeps the current impedance range  The user can change manually the range  but it  is simpler to measure a device in normal mode  auto  before activating sorting mode  the  instrument then keeping the most
29. ero mode    The instrument finds out the values of the serial elements Rss and Lss during a short circuit  measurement  The values of Gpp and Cpp being measured during an open circuit  measurement  The MW1008P takes these elements into account by applying the correction  given on the schematics  a  above  The calibration should be done at each time one proceed to  measurements in the measurement range boundaries  meaning for weak impedance values   high capacitors values of low inductor values  and for high impedances  small capacitors  values and strong inductors values   The open short calibration 15 besides mandatory to  measure with a small error the losses in a components  like the serial equivalent resistance of  a capacitor    Some care has to be taken during open short calibration  The position of the cables must not  change between calibration procedure and component measure  Any change of spacing or  any other change in cables position could change the parasitic elements that could bring an  error in the measurement  During the open short calibration  one can use a small wire piece on  which could be connected the 4 mini pincers  MW10  or the 2 Kelvin clamps  MW KELV    For the measurement of very low impedances  one should place the pincers at right angle  90      to minimise coupling between wires as shown on figure below  The same arrangement  should be used during measurement itself     MWIOOSP  user manual  26    The calibration can       performed at a given t
30. est  frequency or for       frequencies  As a matter of  fact  the parasitics elements  may not be constant over  frequency  and the instrument  keeps in memory the 4  parasitic elements for each  frequency    The calibration is performed  by pressing key CAL  A  menu appears and you can  choose an open circuit  calibration  OPEN  or a short   circuit calibration  SHORT    The correction limits are the  following ones         Figure X   short circuit calibration for the measurement of low  impedances     e Short circuit calibration   the maximum series resistance of cables should be 20 Q  while total impedance should be  lt  50     e Open circuit calibration   the parallel impedance of the cable should be smaller than    10          The instrument displays the message    PASS    if the calibration succeeded     FAILED    in    the other case     To perform a calibration at all frequencies press key CAL long enough so that the instrument  displays the menu             SHORT    in capital letters  The instrument then starts the  calibration procedure at all frequencies  what takes a longer time  The calibration should be  performed at each measurement cable change  or for critical measurements  see above      SMT Pincers cable     The zeroing of the instrument can be performed by shorting the tweezers extremities  The  open circuit calibration is more critical since the spacing between contacts may vary  One  should use for this a small piece of isolating material with the same
31. l   range depends on several factors  such as transformer primary impedance at the test  frequency  During the test of an elevator transformer  one should choose a frequency such that  the primary impedance will be smaller than 100Q in order to get a measurement of N  up to    The use of this function needs use of the 4 wires mini pincers cable MW10  To use this mode  shall be connected to the 2 other  wires  The open short calibration  must not be used  The primary  should be the winding with the  greatest number of turns  In case of  HD LD  allowing the widest measurement  In this range  the primary  10  A smaller primary impedance allows to measure a higher transformer ratio        The following parameters are displayed by successive press on key  n Vs M  for the  characterisation of transformers     N 6 Main parameter N gives the turns ration  between primary and secondary windings   Primary secondary phase shift and test    frequency are displayed on second line     N  0 While pressing twice key  n Vs M  the  instrument displays invert ratio n     1 n     Vs Vp Secondary and primary voltages  In the case of  measurements on a mains supply transformer  the instrument can calculate the    secondary voltage for a primary voltage of  230V or 115V_ starting from measured value   Press key Q Vs M  until parameters Vs and Vp    appear           MWI1OOSP  user manual  20    Default primary voltage is 230V  To change to 115V  press key  Menu    The second line presents the 2 avai
32. lable options  Press  the key below 115 to select virtual primary voltage of  115V  The brackets will move around 115  Press key  EXIT to go back to parameters Vs Vp and to display the  secondary voltage for 115V this time  Spot 1 88   MO Mutual inductance and primary secondary  phase shift The mutual inductance is obtained by measuring primary current  and secondary voltage of the transformer  This direct method can lack of  precision at highest frequencies because to parasitics elements of the  transformer  In case of capacity overflow  the message OVERFLOW 1  displayed  If so  change impedance range by pressing key  old Range   It is  better to start with the lowest range and then increase range until getting a    result  The automatic mode should not be used for this function   The calculation of the mutual inductance             can also be done by measuring the        inductance resulting from serializing 2    WW        windings  by connecting them in the 2     gt   gt   gt   possible configurations as shown on the         gt  CS  figure      The    value is then                              L  Lt  4  The results from the 2  methods can be compared     About mutual inductance      When 2 winding are close to each other  the flux created by one can establish in the  second  The mutual inductance coefficient 1s the ratio of the flux produced by coil   in  coil 2 to the current that flows through coil 1     D       MWI1OOSP  user manual  21    SORTING COMPONENTS    The MW1
33. les quatre possibles     Pressez la touche sous    OK    pour activer le      dr nee per  ei  mode diagnostic clavier  Pressez la touche carne Net     den t A  sous EXIT pour revenir au menu principal  Er      Pressez la touche sous    OK    pour activer le ben CA      e  mode test du buzzer  Pressez la touche sous    te  gee  EXIT pour revenir au menu principal     i fes KN    Pressez la touche sous    OK    pour activer le  mode test des valeurs complexes  courant tension  Pressez la touche sous EXIT  pour revenir au menu principal        Pressez la touche sous    OK    pour activer le  mode test des valeurs du convertisseur A D   Pressez la touche sous EXIT pour revenir au  menu principal      a          a  em  w   jm       Test du clavier    Ce test est utile lorsque l   on rencontre un probl  me avec le clavier  Remarquez qu   un  probl  me de clavier peut emp  cher d   entrer dans ce mode  Il est dans ce cas n  cessaire de  d  panner       la main    les circuits du clavier  On pourra se reporter au tableau donn   plus loin  afin de v  rifier les tensions  Chaque touche correspond    une tension particuli  re qui est lue  par un convertisseur analogique digital    10 bits  L   instrument affiche la valeur lue par  l instrument  de 0    1023  ainsi que la position de la touche appuy  e  La sortie du mode test  du clavier s   effectue lorsque toutes les touches ont   t   test  es     Premier   cran lorsque l   on entre dans le    mode test du clavier  4    i  a Ke       M
34. lly select the desired   L Q  or  L Ax    C D    R Q   121   0     R X    G B    When this key is pressed for more than two seconds the instrument goes in  Auto Parameter mode  In this mode  the instrument will select the most appropriate parameter  pair according to the phase angle and the absolute impedance  This mode is indicated by the  presence of a dark arrow at the top left of the display  The different modes are described  below  The  L A   mode replaces       L Q  mode when the instrument is configured in Ar  mode  see key       n Vs M    This key selects the transformer measurement mode  Pressing this key steps through and  permits the selection of the desired function    N     1 N      Vs Vp    M 0         3 0  Turns ratio and phase angle    1 N 0  Reciprocal turns ratio and phase angle   Vs Vp  Secondary and Primary Voltage   M 0  Mutual Inductance and phase angle  Frequ     e key selects one of the following test frequencies  100 Hz  120 Hz  250 Hz  500  Hz  1 kHz  2 5 kHz  5 kHz  7 8125 kHz  12 5 kHz  15 625 kHz and 25 kHz  The selected  frequency 15 indicated just above this key  When the user frequency 15 selected the word  USER is displayed  Pressing this key for longer than two seconds sets the LCR meter to use  the default parameters  The default frequency 15 1KHz     Hold Range    The Hold Range  key selects the impedance range of measurement appropriate for the device  under test  Pressing this key holds the unit in its current measurement range  Repeated
35. mQ 1kQ 10     Ky  1kQ 1GQ  10     Precision of the resistance     0 2   0 0002       0 2     Precision of      AQ    0 2 100  x 1 0 00057    0 002    MWI1OOSP  user manual  30    Precision of inductances  Precision of L    Ai   A     Ka  K    x 100      Ae  Pr  cision de base  valeur donn  e dans le graphe ci apr  s  Pour les valeurs de Q  gt  0 1  multiplier la valeur de A         OI  Ku      facteurs correspondant a l   erreur aux extr  mit  s des gammes    K  0 16uH  Lmxf  Ky    Lm x f     160         f   fr  quence de test en kHz  Precision of Q         Ai   100   1 Q   Avec A  pr  cision de l   inductance calcul  e ci dessus  Nota   La pr  cision est Q est une grandeur absolue et non pas un pourcentage   Precision of capacitors  Precision of C    A    A  Se  Ka   K    x 100      Ac  Pr  cision de base  valeur donn  e dans le graphe ci apr  s  Pour les valeurs de     gt  0 1  multiplier la valeur de A par  OI  Ku        facteurs correspondant a l   erreur aux extr  mit  s des gammes    K     0 16 pF    Cm x f   Ku    Cm x f  160 mF    f   fr  quence de test en kHz  Precision of Q         Ac   100   1 Q       Avec A  pr  cision de la capacit   calcul  e ci dessus    Nota   La pr  cision est Q est une grandeur absolue et non pas un pourcentage     MWIOOSP  user manual  31    Base precision                10pF 1    _ 10     100 1      100n 10M  lt   10H  3M            1M    lt   300k X    1H     ou      100k y    100mH  100     10k K       x lt    10mH   gt  N   1mH  10m    
36. nt pas pris en compte dans ce mode        34 70    R  sultats A D    Ce mode permet d   acc  der aux mesures primaires qui sont utilis  es pour calculer les  composantes en phase et en quadrature  Les valeurs affich  es correspondent    la valeur  moyenne des signaux pr  sent en sortie du d  tecteur de phase mesur  s par le convertisseur  A D double rampe  Le d  tecteur de phase est un multiplicateur     1  L   entr  e du  multiplicateur est soit le signal de tension soit le signal de courant  Il est possible de passer de  l   un    l   autre en pressant la touche L C R  Le gain    1 est command   par un signal carr   de  m  me fr  quence que le signal de test  La phase de ce signal peut   tre modifi  e de 0      270    par pas de 90   gr  ce    la touche n Vs M     MWIOOSP  user manual  41    Pressez la touche L C R pour s  lectionner la  mesure de tension  puis la touche n Vs M  pour s  lectionner un angle de 0    Notez que  cet angle est un angle qui sert de r  ference et  ne correspond pas a un angle absolu  Pressez  la touche Hold Range et s  lectionnez la gamme 2  L   afficheur montre le r  sultat de la  conversion        Pressez la touche n Vs M pour acc  der aux  valeurs avec un angle de phase de 90    180    et 270       La valeur finale en quadrature est obtenue en  soustrayant les composantes en opposition  de phase  ici   Vq   0 344     0 6909    0 34690       La valeur finale en phase est obtenue en   E             soustrayant les composantes en opposition       
37. o the series resistance is more  significant and the series model is the appropriate choice  For very small inductance values a  higher measurement frequency will yield better accuracy     MWI1OOSP  user manual  16    All inductors have a maximum allowable current  Above this value the core saturates  the magnetic  field remains constant  and the inductance decreases to near zero  The maximum current 1s  dependent on the core material  A core material with high permeability gives a higher inductance  for the same number of turns as a core of low permeability  The drawback is that the core  saturates at a much lower current    Futhermore the inductance varies according to the current level  If you measure the same inductor  with different instrument et the same frequency you can get different velue because the test  current is not the same for both instruments  It is possible to know the current flowing into the  inductor by measuring the resistance and the reactance of the inductor  R and X  and knowing  the source resistance  The source resistor varies according to the impedance range  its value is  100Q for ranges 1 and 2 for example     i rms    Vs V  Rs R    X      with Vs   0 5 Vrms  test voltage at no load      Example   When measuring a coil one measures R   2    and X   43 Q  The range determined  by the instrument is range  2  which has a source resistance arround 100     The current value  flowing into the coil 1s     i   0 5 4 100 2 2 432    4 5 mA  Note  Inductors
38. o u   10   nano n   10   pico       A coil is made of several turns of insulated wire  The property of a coil  is to oppose itself to current variations and 15 characterized by its  inductance     The property of a capacitor  The capacity of a capacitor 1s expressed in  Farad  F      Passive component made of two plates separated by a dielectric  The  property of a capacitor 15 to pass ac current while stopping dc current   See also capacity and reactance     The frequency in which the parameters of a component are measured   The parameter values depends generally upon frequency     Impedance ranges that the instrument uses to perform measurements     Complex value characterizing a passive component  The impedance has  a real component  r  sistance  in series with an imaginary component   reactance   A pure resistance doesn   t have reactance while pure coils  and capacitors doesn   t have resistance     MWI1OOSP  user manual  6    Inductance     Parameter      Accuracy      Basic Accuracy      Source Resistance     Resolution      Test Voltage      A property of a coil by which an electromotive force is induced in it by  a variation of  The inductance of a coil is expressed in Henry  H      Electrical property measured  The main parameter is the most  important parameter of the component  capacity  inductance   resistance   The secondary parameter has less importance and  characterizes component losses  quality factor  dissipation factor or  phase angle      Difference betwe
39. onents   ON There are seven selectable step rates from 2 to 8  Each   step adds approximately  25 seconds to the sampling    refresh rate of the LCD display     Numb  Avrg  Number of measurements used for avaeraging     Varactor OFF This option requires the MW108 Varactor Test Fixture    varicap  available from MW Instruments  Turning the Varactor  option to ON will place the LCR meter automatically in  the Varactor measurement mode upon saving and exiting  the Menu  To return to normal features  select the Menu    key and turn the Varactor feature OFF     Usr f   user frequency  1 25 kHz Set the user frequency from 100 Hz up to 25 kHz    Sorting OFF Use this feature to measure and sort like valued  components    e Turning the Sorting function ON will allow the user to  set the Tolerance between the value of the  benchmark component and like components to be  measures  This Tolerance is selected by pressing the  NEXT key    e When the appropriate tolerance has been selected   pressing the key under the left arrow enables the  value of the benchmark component to be entered b    MWI1OOSP  user manual  10       pressing the    Edit    key  A cursor will appear under  the first digit of the value  Pressing th NEXT Key  will move the cursor to the next digit  Pressing the     Change    key will step the value of the selected digit   When the desired value has been entered  pressing    th   OK _  key will record this value     Pressing the key under the right arrow will allow the  
40. que totale est d   environ 0 05   Le filtre est suivi d   un   tage tampon et  d   un s  lecteur pour le choix de la r  sistance source     MWI1OOSP  user manual  33    Reference Signal Buffer Source Resistor    Clock Signal       DETECTEUR DE PHASE ET CONVERTISSEUR A D    La phase et le module de l   imp  dance inconnue sont obtenus en multipliant la tension de  mesure alternative  courant et tension  par un signal carr   dont la fr  quence est strictement  identique a la fr  quence du signal sinusoidal  et dont la phase peut   tre modifi  e  La valeur  moyenne du signal ainsi produit est proportionnel a la partie en phase ou en quadrature du  signal  Un total de quatre mesures  pour lesquelles la phase du signal carr   est avanc  e de 90    tr  s pr  cis  ment permet la d  termination vectorielle du signal  Ce processus est r  alis   aussi  bien pour la tension que pour le courant  Ainsi huit mesures   l  mentaires sont n  cessaires  pour le calcul final  Les composantes en opposition de phase sont ajout  es  0   avec 180   et  90   avec 270       liminant tous les offsets communs et r  duisant le bruit  Chacune de ces  mesures n   a pas de signification en elle m  me  car la phase absolue entre le signal sinusoidal  de test et le signal carr   n   est pas connu et car le courant parcourant l   imp  dance inconnue  n   est pas contr  l    On peut repr  senter le fonctionnement sous forme vectorielle  ou le  syst  me d  termine un vecteur pour le courant et un second vecteur
41. rtant n  cessaire de conserver la  m  me position des c  bles pour les   tapes suivantes afin de ne pas changer la valeur de  ces   l  ments parasites  Si vous utiliser le cable MW10 les grippe fils LD et LS doivent    tre reli  s ensemble de m  me que les grippe fils HD et HS  Pressez la touche n Vs M    choix    Ok     L   instrument affiche    Opening 100 Hz   ainsi qu   une barre de  progression sur la seconde ligne  L   instrument proc  de a une mesure a chaque  fr  quence  Lorsque l   instrument a termin   cette   tape 11 passe automatiquement     l     tape suivante     est possible de choisir une   tape particuli  re par les touches    choix  lt      et  choix       Il est possible d   effectuer les   tapes STP3 a STP8  dans un ordre quelconque  cependant les calibrations    OPEN    et    SHORT    doivent  toujours   tre effectu  es en premier        2  L   instrument affiche l     cran ci contre  pour la calibration en court circuit   Court circuitez les fils    l   aide d   un d K       morceau de fil   pais  Choisir l   option  Ok  De la m  me fa  on que pour la calibration    OPEN    l   instrument proc  de    une  calibration    toutes les fr  quences puis passe    l     tape suivante  S1 l   instrument affiche     FAIL       la fin de la premi  re calibration    100 Hz v  rifiez que les pinces de test  sont bien en court circuit puis recommencez l   op  ration     ae                                   3    71  5             va maintenant r  aliser les calibration
42. s  An inductor is a device for storing energy in a magnetic  field  which 15 the opposite of a capacitor that 15 a device for storing energy in an electric  field   An inductor consists of wire wound around a core material  Air is the simplest core  material for inductors because it is constant  but for physical efficiency  magnetic materials  such as iron and ferrites are commonly used  The core material of the inductor  its length  and  number of turns directly affect the inductor s value     Model of a real inductor    The series resistance  Rs  represents the resistive losses m the  windings  The parallel capacitance  Cp  is the equivalent  capacitive effect between the turns of the coil  and the  parallel resistance  Rp  is the sum of all losses in the core   Open flux inductors are more sensitive to metallic materials  that are in close proximity  because such materials modify  the magnetic field  Toroidal inductors keep the flux inside the  core and are less sensitive to external conductors in close  proximity                       Inductor measurements can be made in either the series or parallel model  Where the inductance is  large  the reactance at a given frequency 1s relatively large so the parallel resistance becomes more  significant than any series resistance  therefore the parallel model should be used  For very large  inductance values a lower measurement frequency will yield better accuracy    For low value inductors  the reactance becomes relatively low  s
43. s d   imp  dance pour chacune des  gammes et    toutes les fr  quences     y a donc six   tapes STP3 a STP8 pour lesquelles  vous devrez connecter les r  sistances de calibration fournies avec le kit et dont les  valeurs ont   t   pr  c  demment programm  es dans l   instrument lors de la procedure     PARAM     Pour chacune de ces   tapes il est important de garder la position des    MWI1OOSP  user manual  37    c  bles telle qu   elle   tait lors de la calibration    OPEN    et    SHORT    des   tapes  STP1 et STP2  en particulier pour les   tapes STP7  calibration 100 kQ  et STP8   calibration 1 MQ      4  La derni  re   tape consiste    sauvegarder le r  sultat des   tapes de calibration en  m  moire non volatile  L   instrument affiche    SAVE    sur la premi  re ligne  Pressez la  touche sous    OK    pour valider  L   instrument proc  de    une v  rification de l     criture  en m  moire puis affiche    PASSED    si tout s   est d  roul   comme pr  vu  L   instrument  bascule automatiquement en mode normal  Faire un arr  t marche de l   instrument  Il ne  devrait pas y avoir de message d   erreur    la mise sous tension  Si un message d   erreur  appara  t c   est qu   une ou plusieurs   tapes n   ont pas   t   r  alis  e     A l issue de la calibration on pourra v  rifier la valeur des r  sistances dans le mode Z 0   Les valeurs doivent correspondre aux valeurs r  elles avec un angle de phase nul et ce a  toutes les fr  quences     Messages d   erreurs a la mise sous 
44. s have  large temperature coefficients except for  ceramic COG NPO capacitors  which can  exhibit only a 0 003    C variation  Inductors   especially those with non air cores  may vary  largely with temperature  Ambient and DUT  temperature drifts may introduce error into the  measurement  Control ambient temperature  changes to reduce errors     FORME GENERALE       Rp  Rs XS NN   _e e  Xp  aS Rp  Rs    Xp  Kis Xp  Xs    Rp  CONDENSATEUR  R  Rs Cs     Sf           Cp  INDUCTANCE  Rp             E 4     000       Lp  Rp  Rs       Rp   Rs 1 Q     1 Q  S Cs  Cs        1  D     Cp      1 D  Lp 1  it  Q  Q    MWI1OOSP  user manual    13    Capacitors      Capacitors are measured in Farads  The basic construction of a capacitor 1s a dielectric material  between two electrodes  The many different types of capacitors available are classed according to  their dielectric types  The figure below shows the range of generic capacitance values for standard    types     Aluminum Electrolytic       Tantalum Electrolytic    1pF inf LuF imF    A capacitor can be modeled as a pure  capacitor C with some parasitic elements   see the figure below  RS 1s the actual series  resistance  comprised of the lead resistance  and the foil resistance  RS is generally very  low  a few mil   RD symbolizes the  dielectric loss  Its value changes with  frequency    Dissipation factor  also known as loss  tangent  is the ratio of the series  resistance to the reactance  It indicates the capacitor quality  A lo
45. sistances   talon qui vont servir    la calibration et  d autre part    configurer l   instrument selon la fr  quence secteur du lieu o   sera utilis    l   instrument     PARAMETRAGE DE LA CALIBRATION    Commencez par connecter le c  ble de test sur la prise DIN de l   instrument  Mettre  l instrument hors tension    1  Pressez une touche et mettez l   instrument en marche  Relachez la touche lorsque  l   afficheur indique    DIAG CAL EXIT a   2  Pressez la touche  n Vs M sous option CAL  Le menu    PARAM CAL EXIT     apparait    3  Pressez la touche sous l option PARAM  L    instrument affiche     RCAL1  10 0000      qui est la valeur nominale de la premi  re r  sistance de  calibration  Cette valeur doit   tre modifi  e afin qu   elle corresponde    une r  sistance de  pr  cision      0 1   de valeur connue  La proc  dure est d  crite plus bas  Lorsque la  r  sistances est modifi  e l   instrument revient au menu    PARAM CAL EXIT       4  Validez le choix PARAM en pressant la touche   L   instrument affiche     RCAL2  100 000 Q     Modifiez la valeur de la m  me fa  on que pour la premi  re  r  sistance    5  R  p  tez le processus jusqu      la derni  re r  sistance  RCAL6     6  Pressez la touche sous l   option PARAM  L   instrument affiche le menu    Mains Frq   60 Hz    Pressez la touche pour valider le choix 50Hz   L   instrument doit afficher    Mains Frq   50 Hz      7  Pressez la touche sous l   option NEXT  L   instrument affiche le menu    Save  Parameters     
46. ssical use  you just need to use a standard 5 pins male CB plug     The cables should be shielded type  the shields being connected to ground  pin 3       HS  High Sense   2 HD  High Drive   3   OV  GND    Ae LD  Low Drive    S   LS  Low Sense        MWI1OOSP  user manual  29    PRECISION    The measurement precision depends on the base precision of the instrument  which itself  depends on the values of impedance and test frequency  This value is increased by a factor  corresponding to the additional error added when impedance is near the min      max values of  its range  The precision values are valid with the use of cables MW KELV or MW SMT   when an open short calibration has been performed before the measurement  and when the  instrument is in auto ranging mode     Precision of resistances and impedances    Precision of R    A      A  T  Ky K_  x 100      Ae  Base precision  value given in the following table  For Q values  gt  0 1 multiply the  value of A by  1 0    Ku      factors corresponding to range limits error  KL   1 mQ         Ky   Rm  1 GO         is negligible for resistances smaller than 50       K   is negligible for resistances greater than 20 Q    Precision of Q                100   1 0 2    With A  precision of the resistance calculated above  Calculation example      Calculation of a resistance measurement precision of 1      at 1kHz  The instrument gives     value of            0 0005    From the graphics  you find a base precision of 0 2    A   Ki   1
47. tension    Real Error La calibration n   a pas   te effectu  e  Des valeurs par d  faut sont  utilis  es pour la mesure  Les mesures seront toutefois pr  cises     0 5   pour des fr  quences de test jusqu      1 kHz et pour les gammes  d   imp  dance 2    5     Open Cal Error  L     tape de calibration    OPEN    n   a pas   t   effectu  e  Les autres    tapes de calibration doivent   tre effectu  es apr  s avoir r  aliser la  calibration    OPEN       Short Cal Error  L     tape de calibration    SHORT    n   a pas   t   effectu  e  Les autres    tapes de calibration doivent   tre effectu  es apr  s avoir r  aliser la  calibration    SHORT       Cal Error RngXY L     tape de calibration de la gamme X    la fr  quence Y n   a pas   t    effectu  e ou a   chou  e   X prend les valeurs de      6 suivant la gamme concern  e  Y prend les valeurs de A    J selon la fr  quence concern  e  A   100 Hz  J   15 625  kHz     Checksum Error Erreur de checksum  Consulter le service apr  s vente de Ormelabs     MWI1OOSP  user manual  38    DIAGNOSTICS    Le mode diagnostic permet d   effectuer des  tests particuliers    des fins de d  pannage ou  de v  rification  L   acc  s    ce mode s   effectue   i ar pont A gop epe  par le menu sp  cial de calibration diagnostic    Fi 2 FRG die    n   Appuyez sur une touche a la mise sous  tension jusqu      l   apparition du menu    DIAG CAL EXIT    Pressez la touche   choix DIAG         Vous pouvez maintenant s  lectionner un mode particulier parmi 
48. user to set the audible indication when a DUT meets  the programmed parameters  Default_value for the  Pass Beep is SHORT  Pressing th NEXTkey will  change the value to LONG or NONE     Pressing the key under the right arrow will present the  option of turning the backlight on or off when a valid  component is measured  The default value 15 YES     Pressing the Menu key now will allow you to save the  selected configuration and place the LCR meter in  the Relative measurement mode     To turn the Sorting mode OFF  press the Menu key  and step backward or forward using the appropriate  arrow key until the Sorting option is displayed  Press  the key under OFF  Press the Menu key and save this  configuration     Set this feature to ON to measure the AL value of an  unknown toroid core  Press the    forward arrow    key to  enter the number of turns on the inductor  To enter the  number of turns  select EDIT  A cursor will be displayed  under the last digit on the LCD  Pressing the Change key  will step the value of the digit over the cursor  Pressing  the NEXT key will move the cursor to the far left  position  When the number of turns has been entered   press the OK key  For accuracy  a minimum of 10 evenly  distributed turns on the core is recommended     1  Pressing the Menu key and saving the current  configuration will place the LCR meter in L  AL   mode  The AL value will be displayed on the second line  of the display    2  To return to normal measurements  press the Menu
49. ve component most likely to be  present in a low value resistor is series inductance  which has no effect on the measurement of  series R  For high values of resistors  greater than several MQ  the choice usually becomes a low  frequency measurement in a parallel equivalent mode  Parallel because the reactive component  most likely to be present in a high value resistor is shunt capacitance  which has no effect on the  measurement of parallel R  Some high precision resistors are winding resistors  and the number of  turns can be quite large for high value resistors  In this case the series inductance can be quite  large and measurement should be done at low frequency to avoid errors caused by the inter   winding capacitance     MESURE DE LA RESISTANCE  INTERNE DE BATTERIES                under test    La r  sistance interne des batteries peut   tre mesur  e       HE          Es       par le circuit ci contre  La tension continue est  isol  e par le condensateur CI   47uF  La r  sistance  interne sera g  n  ralement mesur  e    1 kHz  Avant  d effectuer la mesure proprement dite on r  alisera  un calibration en court circuit en rempla  ant la  batterie par un court circuit  La mesure est effectu  e     vide  Il est possible de connecter momentan  ment  une r  sistance en parall  le sur la batterie afin de la  faire d  biter  Il suffira de faire un petit calcul afin  de retrouver la r  sistance interne r  elle        MWIOOSP  user manual  18    DISPLAYED PARAMETERS    L Q    L 
50. w D indicates a nearly pure  capacitor  In order to achieve reliable measurement a short zero must be performed before any  ESR or D measurement because in this case the series resistor can be very small  Like most  everything else about capacitors  it changes with time  frequency  and temperature  ESR is a  single resistive value of a capacitor representing all real losses  It includes effects of the  capacitor s dielectric loss        Electrolytic Capacitors     The accurate measurement of electrolytic capacitors  particularly large value caps  can present  unique requirements  The MW1008 LCR meter applies an AC signal to the DUT  To test some  polarized components  such as electrolytic and tantalum capacitors  it may be preferable to use  only positive voltages  During normal operation  the AC current source swings negative 50   of the time  which results in an inverse polarization of the capacitor under test  To prevent this  inverse polarization  a DC bias can be applied to prevent the voltage across the part from    MWI1OOSP  user manual  14    becoming negative  The schematic for a simple test fixture to apply DC bias is provided at  Appendix 1     Test frequency    Generally high value capacitors should be measured at lower test frequencies because the  impedance of the component will be very low  Low value capacitors should be measured at  higher frequencies     Model    Measuring a capacitor in series or parallel mode can provide different results  The difference  c
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
CopyRight Notice  English - Broncolor  取扱説明書  iLEAD controller user manual - COMPLETE  Creative TD020D User's Manual  1 GBBR REAL ESTATE DOCUMENT PREPARATION  Samsung L100 Kasutusjuhend  Istruzioni di funzionamento e montaggio Etanorm  取扱説明書 OPJ-2400シリーズ - LED照明・LED蛍光灯のOPTILED  Intelligent IP-Camera - EnGenius Technologies, Inc.    Copyright © All rights reserved. 
   Failed to retrieve file