Technical Procedure for Scanning Electron Microscope/ Energy

image

Contents

1. between 4000 and 8000 counts using SEM spot size Do not follow the on screen suggestion of 1000 counts that specific criterion is only for detectors cooled by liquid nitrogen and is not the criterion for the SDD detector Follow the prompts Pass Criteria The extrapolated strobe resolution at process time 5 or 6 shall be no greater than the detector resolution of 127 eV If resolution requirement is not met maintenance shall be performed or a service engineer called Once maintenance is performed and this criterion is met the instrument may be returned to service When test is complete select REPORT icon in Navigator Pane and Print Page 3 of 9 All copies of this document are uncontrolled when printed Technical Procedure for Scanning Electron Microscope Energy Version 4 Dispersive X Ray System SEM EDX for non GSR Casework Effective Date 10 18 2013 Trace Evidence Section Issued by Trace Evidence Forensic Scientist Manager 5 1 4 5 1 5 5 1 6 5 1 3 2 9 Load and analyze the Stainless steel 316 Standard Print report Pass Criteria The following peaks should be present Fe Ni Si Mo Cr and Mn If these peaks are not present maintenance shall be performed or a service engineer called Once maintenance is performed and this criterion is met the instrument may be returned to service 5 1 3 2 10 Fill out LEO Oxford SEM EDX Performance Check Log and place all printed reports in LEO Oxford Notebook Perform
2. When the pulse processor is calibrated zero and gain will be displayed in Green Record the date value of the Gain 10 000 5 and initials on the Aspex Monthly Calibration Log Pass Criteria The value of the Gain shall be 10 000 5 If this value is not obtained maintenance shall be performed or a service engineer called Once maintenance is performed and this criterion is met the instrument may be returned to service Remove Copper Standard from stage and load the Stainless Steel 316 Standard Saturate the beam and align the filament at 20 kV Obtain an image of the standard and focus on it Identify the peaks and save the spectra Print out the spectra and place in the Aspex Monthly Performance Check Manual Page 6 of 9 All copies of this document are uncontrolled when printed Technical Procedure for Scanning Electron Microscope Energy Version 4 Dispersive X Ray System SEM EDX for non GSR Casework Effective Date 10 18 2013 Trace Evidence Section Issued by Trace Evidence Forensic Scientist Manager 5 2 3 5 2 4 5 2 5 5 2 2 8 Pass Criteria The following peaks should be present Fe Ni Si Mo Cr and Mn If these peaks are not present maintenance shall be performed or a service engineer called Once maintenance is performed and this criterion is met the instrument may be returned to service Performance Verification for New Instrument Set Up 5 2 3 1 5 2 3 2 5 2 3 3 A new SEM with EDX detec
3. EO Oxford SEM EDX System 5 1 1 LEO Microscope Startup 5 1 1 1 Turn instrument ON The computer for the SEM starts automatically Enter user name and password when prompted 5 1 1 2 Open the SmartSEM icon Enter user name and password SEM control screen will appear All commands for the SEM are accessed through this screen Page 1 of 9 All copies of this document are uncontrolled when printed Technical Procedure for Scanning Electron Microscope Energy Version 4 Dispersive X Ray System SEM EDX for non GSR Casework Effective Date 10 18 2013 Trace Evidence Section Issued by Trace Evidence Forensic Scientist Manager 5 1 2 5 1 1 3 5 1 1 4 5 1 1 5 5 1 1 6 5 1 1 7 5 1 1 8 5 1 1 9 5 1 1 10 5 1 1 11 Open Fantasia Software Open RemCon32 Software Click on the Vacuum option on the Status Toolbar and select VENT Select YES at the prompt Unlatch both chamber door latches and adjust nitrogen pressure to 3 4 psi Turn off nitrogen supply after chamber is finished venting Pull out stage chamber door to access stage holder Using gloves remove the stage holder and place it on a clean surface Using mounting tweezers place samples in the holder and note position of each sample Tighten the screws on the holder for each mount position Place the holder back on the stage ensuring that the flat edge of the mount is against the flat area of the stage Roll the stage back into the chamber and securely latch t
4. In addition the Stainless Steel 316 Standard shall be used for performance verifications These standards have no special storage requirements 5 1 9 Instrument Maintenance Routine maintenance shall be performed such as changing pump oil checking liquid chiller status and replacing a filament Any maintenance performed shall be documented in the maintenance logbook for that particular instrument ASPEX Variable Pressure SEM EDX 5 2 1 Start Up Procedure amp Loading of Samples 5 2 1 1 5 2 1 2 5 2 1 3 Turn instrument ON Enter user name and password when prompted The ASPEX engine will start automatically After ASPEX engine loads double click Perception icon Move black lever on the stage from the downward position towards the upward position This will begin the process of venting the chamber The black lever will automatically move to the full upward position once the chamber is at atmospheric pressure Page 5 of 9 All copies of this document are uncontrolled when printed Technical Procedure for Scanning Electron Microscope Energy Version 4 Dispersive X Ray System SEM EDX for non GSR Casework Effective Date 10 18 2013 Trace Evidence Section Issued by Trace Evidence Forensic Scientist Manager 5 2 2 5 2 1 4 5 2 1 5 5 2 1 6 Pull the stage from the microscope column and load the sample s into the specimen holder using mounting tweezers Insert specimen holder into stage and tighten the thum
5. Technical Procedure for Scanning Electron Microscope Energy Version 4 Dispersive X Ray System SEM EDX for non GSR Casework Effective Date 10 18 2013 Trace Evidence Section Issued by Trace Evidence Forensic Scientist Manager Technical Procedure for Scanning Electron Microscope Energy Dispersive X Ray System SEM EDX for non GSR Casework 1 0 Purpose This technical procedure shall be followed for the operation of the Scanning Electron Microscope Energy Dispersive X Ray System SEM EDX including ASPEX and the LEO Oxford systems This procedure shall be used for non GSR casework 2 0 Scope This procedure applies to the ASPEX and the LEO Oxford systems These instruments are used for high resolution and magnification imaging with enhanced depth of field for trace evidence and non destructive elemental analysis paint metals powders and other trace particulate material 3 0 Definitions N A 4 0 Equipment Materials and Reagents 4 1 Equipment e ASPEX Variable Pressure Scanning Electron Microscope Energy Dispersive X Ray System e LEO 1450 Variable Pressure Scanning Electron Microscope e Oxford Energy Dispersive X ray System SDD Detector 4 2 Materials Mounting tweezers for SEM lifts Kimwipes Gloves Adhesive lifts with carbon backed tape Copper Cu Standard Stainless Steel 316 Standard Manganese Mn Rhodium Rh 5492 Standard Nitrogen gas compressed Purity Grade 5 0 4 3 Reagents N A 5 0 Procedures 5 1 L
6. ance Verification for New Instrument Set Up 5 1 4 1 5 1 4 2 A new SEM with EDX detector shall be installed by a certified engineer according to the manufacturer s guidelines Spectra shall be obtained from a Manganese Rhodium 5492 Standard Copper Standard and a Stainless Steel 316 Standard Collection and Storage of Data Non GSR on LEO SEM 5 1 5 1 5 1 5 2 5 1 5 3 5 1 5 4 5 1 5 5 Set the parameters of the LEO SEM instrument e g different accelerating voltage current setting signal collector or display variable based on training and experience to enhance the image on the display monitor Perform sizing of particles if needed Label pertinent information on the monitor using Annotation drop down menu Save images and or annotation to the hard drive Print image and import into Forensic Advantage FA Case Record Collection and Storage of Data Non GSR on OXFORD EDS 5 1 6 1 5 1 6 2 5 1 6 3 5 1 6 4 5 1 6 5 Use Acquisition Rate 3 for most analysis If performing quantification a higher process time of 4 5 6 may be needed depending on sample type Analysis time will vary with 100 seconds being an average time Click on Point and ID icon in the Navigator Pane To set up analysis follow the sequential steps in the navigator pane starting with PROJECT Save the project with a unique name before starting analysis During and after collection you may expand different region
7. arl Zeiss SMT Nano Technology Division SmartSEM Operating Software for Felid Emission Scanning Electron Microscopes Manual V05 04 Printed 2009 Hearle J W J T Sparrow and P M Cross The Use of the Scanning Electron Microscope Oxford Pergamon Press Ltd 1972 Oxford Instruments Analytical INCA Energy Operator Manual Issue 1 Printed 2004 Located on Desktop of X ray computer Oxford Instruments Analytical INCA Feature Instruction Manual Located on Desktop of X ray computer 2007 Records e LEO Oxford SEM EDX Performance Check Log e Maintenance Log e ASPEX SEM EDX Performance Check Log e Request for Instrumental Examination of Evidence Page 8 of 9 All copies of this document are uncontrolled when printed Technical Procedure for Scanning Electron Microscope Energy Version 4 Dispersive X Ray System SEM EDX for non GSR Casework Effective Date 10 18 2013 Trace Evidence Section Issued by Trace Evidence Forensic Scientist Manager 10 0 Attachments N A Revision History A Version Effective Date Reason Number 09 17 2012 Original ISO Document 09 30 2013 3 5 1 3 2 Removed from performance check that it shall be performed by first person to use the instrument 10 18 2013 Added issuing authority to header 1 02 01 2013 2 Request for Instrumental Examination of Evidence was added as a record 4 Page 9 of 9 All copies of this document are uncontrolled when printed
8. b screw on the right side of the stage Gently slide stage back inside the microscope column Push black lever downward to begin the vacuum process Choose Variable Pressure checkbox in the Perception software Default variable pressure setting is 0 1 Torr Click Power Button turns bright green when power is ON and SEM EDX is at proper vacuum Set stage point to copper standard Saturate the beam and align filament at 20 kV by adjusting the filament drive up to around 68 or until an image appears Turn the filament drive up slowly until the image is at maximum brightness Select Source Image Mode from the drop down menu to view the emission image Adjust filament drive until the emission image is slightly under saturation Select Line Scan option from the drop down menu Adjust the X and Y axis controls to adjust the line scan to its maximum height Close the Line Scan mode Performance Check Required once a month by the first person to use that instrument in the month while in use 5 2 2 1 5 2 2 2 5 2 2 3 5 2 2 4 5 2 2 5 5 2 2 6 5 2 2 7 Load a Copper standard saturate the beam and align the filament at 20 kV Obtain an image of the standard and focus on it Select Control Right Mouse click and click on Calibrate from the drop down menu Collect X Ray spectra of the Copper Standard The zero offset and gain of the pulse processor are being extracted while the spectrum is being collected
9. he chamber door latches Click the Vacuum option on the Status Toolbar and select PUMP Wait for system to pump down Turn on the Extra High Tension EHT and Filament by clicking on GUN in the right bottom corner of screen and select BEAM ON The last saved working conditions shall be automatically brought up for the filament In emission mode check to determine if the filament is properly aligned and saturated Adjust emission image and filament by going to User Toolbar Beam and Gun Alignment Use the electronic positioning adjustments to center the fully saturated emission image Adjust saturation of the filament slightly below or at the second crossover Adjust the working parameters of the instrument as necessary by accessing Tools through the User Toolbar and Go to Control Panel Oxford EDX Start Up 5 1 2 1 5 1 2 2 5 1 2 3 5 1 2 4 5 1 2 5 Turn computer on by pressing the power button located on the HP computer in lower right area of the EDS table Login to INCA user by clicking on INCA Enter password Double click on the INCA icon on the desktop Select Detector Control in the Options menu In Thermal Tab the Detector State shall say operate and COLD and in the Slide tab the Slide Position shall say in Close window Select Microscope Control in the Options menu Check the box for HV Close window Page 2 of 9 All copies of this document are uncontrolled when p
10. rinted Technical Procedure for Scanning Electron Microscope Energy Version 4 Dispersive X Ray System SEM EDX for non GSR Casework Effective Date 10 18 2013 Trace Evidence Section Issued by Trace Evidence Forensic Scientist Manager 5 1 3 Performance Check for the LEO OXFORD System 5 1 3 1 A quantitative performance check is required whenever quantitative results are needed 5 1 3 1 1 5 1 3 1 2 5 1 3 1 3 5 1 3 1 4 Using either a Copper or Manganese Standard in INCA select the Quant Optimization icon in the navigation pane Obtain an optimal image by adjusting the microscope settings Use a minimum Acquisition Rate of 4 for optimum resolution Select the element of choice Once the element selection is made this program runs automatically The system will state the quantification was successful when finished 5 1 3 2 Resolution vs Process Time Check Performed monthly if in use 5 1 3 2 1 5 1 3 2 2 5 1 3 2 3 5 1 3 2 4 5 1 3 2 5 5 1 3 2 6 5 1 3 2 7 5 1 3 2 8 User must exit INCA software to access the program for this check Click INCA Monitor icon in task bar in lower right corner Select Options Monitor Options and Engineer login Enter the password In Test Details tab Highlight Resolution vs Process Time then click add Select next Follow prompts until it asks for user to adjust counts on a standard Load a standard such as Manganese then adjust count rate
11. s of the spectra and identify spectral peaks either automatically or manually Once spectral peaks are identified and confirmed save project to the hard drive Page 4 of 9 All copies of this document are uncontrolled when printed Technical Procedure for Scanning Electron Microscope Energy Version 4 Dispersive X Ray System SEM EDX for non GSR Casework Effective Date 10 18 2013 Trace Evidence Section Issued by Trace Evidence Forensic Scientist Manager 5 2 5 1 6 6 Print spectra in pdf format and save in FA Case Record 5 1 7 Shut Down Procedures 5 1 7 1 5 1 7 2 5 1 7 3 5 1 7 4 5 1 7 5 Turn off the filament by clicking ALL in the lower right corner of SmartSEM software Close the SEM operating window and answer the dialog questions Close the SEM user software Close the EMServer Close the RemCon32 software Shut down the computer Once computer is shutdown press the yellow STANDBY button on the front panel of the instrument to keep the electronics on but in a standby mode Pressing the red OFF button will remove all the voltage to the instrument and will lose stage settings Therefore never press the red OFF button except for major resetting of the instrument or for servicing of the system Close all windows in INCA software and shut down the computer 5 1 8 Standards and Controls This instrument requires the use of Manganese Rhodium 5492 standard for performance checks
12. tor shall be installed by a certified engineer according to the manufacturer s guidelines Spectra shall be obtained from a Copper standard and a Stainless Steel 316 standard If the spectra are acceptable and all major peaks identified the instrument is approved for use in casework Collection and Storage of Data Non GSR on ASPEX 5 2 4 1 5 2 4 2 5 2 4 3 5 2 4 4 5 2 4 5 5 2 4 6 5 2 4 7 5 2 4 8 Set the parameters of the ASPEX instrument based on training and experience to enhance the image on the display monitor Perform sizing of particles if needed Label pertinent information on the monitor using Annotation Save images and or annotation to the hard drive Collected x ray data User may have to adjust spot size and restart to achieve optimal dead time collection Analysis time will vary with 100 seconds being an average time During and after collection you may expand different regions of the spectra and identify spectral peaks either automatically or manually Once spectral peaks are identified and confirmed save spectra to the hard drive Print image and spectra and save in FA Case Record Shut Down Procedure 5 2 5 1 5 2 5 2 5 2 5 3 5 2 5 4 The SEM vacuum always remains on Ensure that the X Ray system is not actively collecting spectra Turn filament drive to 0 and close Perception software Select Start icon then Shut Down Page 7 of 9 All copies of this document are
13. uncontrolled when printed Technical Procedure for Scanning Electron Microscope Energy Version 4 Dispersive X Ray System SEM EDX for non GSR Casework Effective Date 10 18 2013 Trace Evidence Section Issued by Trace Evidence Forensic Scientist Manager 6 0 7 0 8 0 9 0 5 2 6 Standards and Controls This instrument requires the use of a Copper Standard for performance check In addition the Stainless Steel 316 standard is used for performance verifications These standards have no special storage requirements 5 2 7 Instrument Maintenance Routine maintenance shall be performed such as changing pump oil changing out a filament etc Any maintenance performed shall be documented in the maintenance logbook for that particular instrument 5 3 Sampling and Sample Selection No sampling is performed When sample selection occurs it shall be based on the Forensic Scientist s training and experience 5 4 Calculations N A 5 5 Uncertainty of measurement N A Limitations N A Safety Concerns 7 1 The greatest safety concern is radiation from the X ray tube The x ray system is monitored for leaks on a regular basis 7 2 There is high voltage current which can cause electrocution Avoid contact with any live circuitry components Potentially lethal voltages exist with the high voltage x ray supply References ASPEX LLC Personal SEM Console User Manual Perception Suite User Documentation Version 1 37 2002 C