Home

Multirall Instruction Guide EN

image

Contents

1. 1 Install Multirall with the 2 Connect the desired lifting accessory A Multirall must not be equipped universal connector directly to the Q link see Operation p 5 with a Strap Gripper when the lift into the carriage hook The lift strap and lifting accessory are strap is mounted under the lift unit then suspended under the lift unit This may result in the Strap Gripper getting caught in the lift strap After assembly make sure that e the lifts functions correspond to the markings of the e the battery charger works properly and the hand control indicator lamps are illuminated during charging e the mechanical and electrical emergency lowering e the batteries are charged both work Multirall 200 7EN125103 03 4 www liko com Operation Operating Change of lifting accessories Multirall is operated by lightly pressing the buttons on Multirall is adapted to Liko s Quick Release the hand control The arrows indicate the direction of System for safe and easy changing of lifting movement The movement stops when the buttons accessories corresponding buttons on the end cover of the lift unit Quick Release Hook can easily be updated with A The arrow corresponds to the direction of travel when a Quick Release Hook so that they can be used the lift is mounted with the lift strap above the lift unit with Multirall See Quick Release Hook p 9 A Operate the lift only when tension is
2. See the Guide to Liko s Quick Release Hook System which can be downloaded from our website www liko com or contact Liko Hill Rom gt for more information Multirall 200 7EN125103 03 9 www liko com Stretcher Most of the stretchers in Liko s product range can be used with Multirall Contact Liko Hill Rom for further information Scale For weighing patients in combination with Multirall we recommend LikoScale 350 The LikoScale adapter kit is required LikoScale 350 is certified in accordance with the European Directive NAWI 90 384 Non Automatic Weighing Instruments Contact Liko Hill Rom for further information Liko Support Springs for Bounce and Gait Training Elastic springs are available as an accessory in order to produce a softer springy motion for example during gait training Short 22 cm 8 6 in max 70 kg 154 Ibs weight pair Short 22 cm 8 6 in max 100 kg 220 Ibs weight pair Long 28 cm 11 in max 70 kg 154 Ibs weight pair See the instruction guides for Liko MasterVest models 60 and 64 or Liko LiftPants model 92 for more information Extension Arm Multirall Facilitates the connection disconnection of the lift strap to from the carriage Fits Carriage S65 with single hook prod no 3136011 Multirall Helping hand Facilitates the connection disconnection of the lift strap to from the carriage Fits Carriage S50 with single hook prod no 3136010 Multirall Trolle
3. between max lifting height and min lifting height measured in CSP Multirall 200 7EN125103 03 3 www liko com Assembly Before the first use the lifts battery should be charged for at least 8 hours For detailed instructions see Charging the Batteries p 6 Multirall can be mounted to a carriage in two different ways See alternatives below Alt A with the lift strap above the lift unit This alternative is recommended in cases where the lift is to be moved frequently between different rail systems or when the lift is to be used for room to room transfers AA lift assembled with the lift strap above the lift unit must not be used in wet areas Multa PW Ss 1 Feed out desired length of lift 2 Connect the Q link 3 Connect the NOTE Remove the strap by applying tension to the to the rail carriage desired lifting Hang Up HandControl lift strap and simultaneously and raise the lift to an accessory to Hanger when the lift is pressing the electrical appropriate height the universal installed with the lift strap emergency lowering button connector of above the lift unit A Operate the lift only when the lift unit see tension is applied to the lift strap Operation p 5 Alt B with the lift strap under the lift unit This alternative is recommended in cases where the lift is to be installed in the same rail system for a longer period of time without being moved a D 0 Ww Bg D
4. aise the lift until the weight of the patient has been fully transferred to the carriage in room 2 A While the lift is raised it is very important to ensure that the lift strap is twisted correctly in order for it not to fold when it enters the lift unit Then disconnect the Strap Gripper Multirall 200 7EN125103 03 8 www liko com Maximum Load Different maximum loads may apply to different products on the assembled lift system rail lift sling bar sling and any other accessories used For the assembled lift system the maximum load is always the lowest maximum load rating of any of the components For example a Multirall that is approved for 200 kg 440 Ibs can be equipped with a sling bar that is approved for 300 kg 660 Ibs In this case the maximum load of 200 kg 440 Ibs applies to the assembled lift system Check the markings on the lift and lifting accessory or contact your Liko Hill Rom representative if you have any questions Recommended Lifting Accessories A Using lifting accessories other than those approved can entail a risk Recommended sling bars and accessories for Multirall are described below To select suitable slings and other lifting accessories please see the Lifting Accessories brochure For additional guidance in selecting a sling study the instruction guide for the respective sling models Here you will also find guidance for combining Liko s sling bars with Liko s slings Contact y
5. applied to the lift strap te OW Co A Electrical Emergency Lowering Mechanical Emergency Installation of Latches In the event that the hand control or Lowering After installation check that the electronics malfunction the lift can be Press the red button marked latch locks and moves freely in lowered by pressing the button on the end Emergency on the end the sling bar hook cover of the lift unit cover of the lift unit Note that A The arrow corresponds to the direction 0ad must be applied to the of travel when the lift is mounted with the _ lift in order for the mechanical lift strap above the lift unit emergency lowering to work Ensure the patient is lowered into a bed Ensure the patient is lowered wheelchair or other suitable place into a bed wheelchair or other suitable place Activate Reset J Emergency Stop Hang up HandControl Hanger Enable the emergency stop press the red button When not in use the hand control Reset the emergency stop turn the button in the direction indicated can be hung on the Hang up by the arrows on the button HandControl Hanger The red button on the lift unit end cover is intended for use in an emergency situation When the button is pressed contact between the motor and power source is broken which stops the lifting motion Multirall 200 7EN125103 03 5 www liko com friction brake a SSP Limit Swit
6. ary Bracket for Charger Contact Liko Hill Rom for further information a Simple Troubleshooting The lift doesn t work gt Check that the Emergency Stop button is not activated Check that the battery is charged Check that the hand control is correctly connected 4 If the lift still does not work satisfactorily please contact Liko Hill Rom oO N N The lift emits a repeated sound gt or light signal Charge the battery immediately If the lift still does not work satisfactorily please contact Liko Hill Rom N N The lift is stuck in the gt high position Check that the Emergency Stop button is not activated 2 Check that the battery is charged 3 Check that the hand control is correctly connected 4 Use indicated electrical or mechanical emergency lowering to lower the patient onto a firm surface 5 If the lift still does not work satisfactorily please contact Liko Hill Rom NS You hear unusual sounds gt Contact Liko Hill Rom Multirall 200 7EN125103 03 11 www liko com Inspection and Maintenance Care and Maintenance For trouble free use certain details should be checked each day the lift is used e Inspect the lift and check to make sure that there is no external damage e Check the sling bar attachment e Check the lift strap for wear and to ensure the strap is not twisted e Check the functionality of t
7. ch It is important that the lifting motion of the lift strap is performed as vertically as possible to ensure safe operation The SSP Limit Switch is intended to stop the lifting motion if the lift strap is subjected to harmful strain for example if it is pulled sideways or folded over during the lifting motion The SSP Limit Switch also provides protection against pinching If the SSP Limit Switch has been activated and the lifting motion stopped the lift strap must first be operated downwards before it can be operated upwards again Charge Indicator Multirall has two indicators that warn when the battery has a low charge e buzzer that sounds when lifting e LED that is lit during lifting When either of these signals sounds or illuminates the unit should be charged as soon as possible See Charging the Batteries below Charging the Batteries Green Yellow White no function a Charging in during charging Adjustable Friction Brake The amount of drag on the lift can be adjusted with the friction brake on the carriage Turn the brake clockwise for increased resistance and counter clockwise for reduced resistance In order to ensure maximum battery life it is important to charge the batteries regularly We recommend that you charge the batteries after use or every night The batteries are fully charged after approximately 8 hours When fully charged batteries in good condi
8. d Liko A D Company al nes pute yo 0 ae Cg IDo Multirall 200 Instruction Guide English 7EN125103 03 2012 01 04 Applies to the following model Multirall 200 Prod No 3130001 Multirall Yoyo Product Description Multirall 200 is a flexible overhead lift with a versatile range of applications It is easily movable without compromising on safety and functionality Multirall also works perfectly as a room to room lift Multirall can be mounted to the rail system in two different ways depending on intended use When mounted with the lift strap under the lift unit left picture it works as a normal overhead lift and is easily movable between different rail systems without the need of tools When mounted with the lift strap above the lift unit right illustration Multirall becomes a flexible room to room lift that can be used for transfers of patients between different rail systems in different rooms without requiring openings over doorways Multirall is appropriate for all common lift and transfer situations for example between bed wheelchair to from floor toilet visits gait training and for horizontal lifts with stretchers Accessories A large and comprehensive range of accessories is available for Multirall Liko s product range includes several sling models in many sizes and different designs Also available as an accessory is a practical trolley for transfer and
9. ent applicable assembly instructions adjustment see applicable forwards as close as possible to the assembly instructions doorway Lower the patient as far down as is comfortably possible but not so far that the caregiver cannot reach the Room to Room Strap with the Strap Gripper a Z Strap Gainer N Y ee Pa w a 2 Connect the Strap Gripper with both hooks to the Room to Room Strap see illustration to the right gt Multirall 200 7EN125103 03 7 www liko com Se S A Lower the Multirall until the Disconnect the lift strap from the carriage extension belt in room 1 Room to Room Strap carries the and connect it to the carriage extension belt in room 2 To prevent the full weight of the patient Then lift strap from twisting which in turn may cause the SSP limit switch to run out an extra 10 cm 4 in of lift activate see p 6 the lift unit must be turned manually so that the lift strap to provide enough slack for strap is straight when it enters the lift unit see small illustration disconnection A Check carefully to ensure that the Q link is correctly applied A Keep hand tension on the lift to the carriage hook extension belt strap as you run it out of the lift motor KA k m Only when the lift strap runs straight into the lift unit can the Move the patient further on in room 2 transfer to room 2 proceed see small illustration R
10. he latches e Check the control function of the lift movement e Check to make sure that the emergency lowering works e Charge the batteries each day the lift is used and check to ensure the charger works When necessary clean the stretcher with a moist cloth using warm water or disinfectant NOTE Do not use cleaning agents that contain phenol or chlorine since these can damage aluminium and plastic materials A The lift should not be exposed to running water Service Multirall should be periodically inspected at least once a year A Periodic inspection repair and maintenance should be performed only in accordance with the Liko Service Manual and by personnel authorized by Liko and using original Liko spare parts Service Agreement Liko Hill Rom offers the opportunity to enter into service contracts for the maintenance and regular inspection of your Liko product Contact your Liko Hill Rom representative for more information Expected Life Time The product has an expected life time of 10 years when correctly handled serviced and inspected in accordance with Liko s instructions Transport and Storage During transportation or when the lift is not to be used for a long time the emergency stops should be engaged The environment where the lift is transported and stored should have a temperature of 10 40 C 50 104 F and a relative humidity of 30 to 75 The air pressure should be 700 1060 hPa Recycling For instruct
11. ions on how to recycle your Liko product please visit our website www liko com Product Changes Liko s products undergo continuous development which is why we reserve the right to make product changes without prior notice Contact your Liko Hill Rom representative for advice and information about product upgrades Design and Quality by Liko in Sweden Liko is quality certified in accordance with ISO 9001 and its equivalent for the medical device industry ISO 13485 Liko is also certified in accordance with environmental standard ISO 14001 A Liko a CEID company www liko com Liko AB Nedre v gen 100 S 975 92 Lule Tel 46 0 920 474 700 Fax 46 0 920 474 701 info liko com Copyright Liko AB 2012 01
12. irall has been tested by an accredited testing institute and fulfils the requirements specified in the Medical Device Directive 93 42 EEC for Class products Multirall fulfils the requirements specified under EN ISO 10535 2006 IEC 60601 1 EN 60601 1 2 ANSI AAMI ES60601 1 and CAN CSA C22 2 no 60601 1 A The lift must not be modified under any circumstances If you have any questions please contact Liko Hill Rom Particular care must be taken when using powerful sources of electromagnetic disturbance such as diathermy etc so that for example diathermy cables are not positioned on or near the lift If you have any questions please consult the responsible assistive device technician or the supplier The lift is not suitable for use in the presence of flammable mixtures for example in areas where flammable goods are stored Multirall 200 7EN125103 03 2 www liko com Definitions N 1 Carriage S65 48 2 Q link quick fastener 3 SSP Limit Switch 4 4 Emergency stop ne 5 End cover ee 6 Mechanical Emergency Lowering 416 7 Electrical Emergency Lowering Raising 8 Contact for hand control 9 Hand control 13 10 Hang up HandControl Hanger 11 Sling bar 12 Latches 14 13 Quick Release Hook 43 14 Universal connector qa 15 Lift unit 16 Handle aD 17 Extension Belt connector E 18 Lift strap J Technical Data Maximum load 200 kg 440 Ibs Intermittent Int Op 10 90 active ope
13. n guide Remember to keep the instruction guide in a place where it is always available to those using the product Safety Instructions Before using the first time make sure that the lift is assembled in accordance with the assembly instructions e the lifting accessories are properly attached to the lift e the batteries have been charged for at least 8 hours e you have read the instruction guide for the lift and lifting accessories personnel using the lift are informed of the correct use of the lift and lifting accessories e the lifting accessory is selected appropriately in terms of type size material and design with regard to the patient s needs Before lifting always ensure that e the lift strap is not twisted or worn and can move in and out of the lift freely e the lifting accessories are not damaged e the lifting accessory is correctly and securely applied to the patient in order to prevent injuries e the lifting accessories are properly attached to the lift e the lifting accessories hang vertically and can move freely e the sling bar latches are intact missing or damaged latches must always be replaced e the sling s strap loops are correctly connected to the sling bar hooks when the sling straps are extended but before the patient is lifted from the underlying surface A Never leave a patient unattended during a lifting situation ETL CLASSIFIED GHEE C a ve uy Intertek Mult
14. olation is therefore retained between the rooms supported by the system In order to perform room to room transfers with Multirall the following conditions are required Both rooms have separate rail systems each with a carriage mounted in the rail e Between the rooms suspended under the doorframe a Room to Room Strap is mounted acting as a bridge between the rooms e Multirall is equipped with a Strap Gripper which is installed to either of the Extension Belt connectors on the lift e An adjustable extension belt is mounted to the carriage in order to obtain an appropriate lifting height Alternatively if the lifting height becomes too low when using the Extension Belt the Extension Arm Multirall Multirall Helping Hand can be used for connecting Q link directly to the carriage hook providing a maximum lifting height Room to Room Transfer Below is a description of an R2R transfer with Multirall Strap Gripper Room to Room Strap and adjustable Extension belt A Check carefully at each stage that the Q link is correctly applied to the carriage hook extension belt Room 2 Extension Belt R2R strap XS E af Install the lift according to assembly Install the Room to Room Multirall is pending in the carriage alternative A p 4 Install the Strap between the two rooms alternatively in an extension belt in Strap Gripper in accordance with For installation and length room 1 Move the lift with the pati
15. our Liko Hill Rom representative or visit www liko com for advice and information on Liko s product range Universal SlingBar 350 with Quick release Hook Prod No 3156084 amp Max 300 kg 660 Ibs Cnm n 5 Universal SlingBar 450 with Quick release Hook Prod No 3156085 eC Max 300 kg 660 Ibs Universal TwinBar 670 with Quick release Hook Prod No 3156087 Max 300 kg 660 Ibs Liko Sling Cross bar 450 with Quick release Hook Prod No 3156022 Max 300 kg 660 Ibs Liko Sling Cross bar 670 with Quick release Hook Prod No 3156019 Max 300 kg 660 Ibs Liko SlingBar Mini 220 Prod No 3156005 fel GE Max 205 kg 450 Ibs Universal SlingBar 600 Prod No 3156076 Max 300 kg 660 Ibs must be updated with Quick release Hook to use with Multirall see Quick release Hook below Quick Release Hook Quick release Hooks form a system providing safe and easy changing of lifting accessories Liko s Quick release Hooks protect against unintentional detachment Update existing Liko sling bars not equipped with a Quick release Hook The Quick Release Hook Universal fits the Universal Bars 350 Quick Release Hook Quick Release 450 and 600 prod no 3156074 3156076 The Quick Release Hook Universal Hook TDM TDM fits the SlingBar Mini 220 prod no 3156005 Sling Cross bar Prod No 3156508 Prod No 3156502 450 and 670 prod no 3156021 and 3156018 and Universal TwinBar 670 prod no 3156077
16. ration Batteries 2 X 12 V DC 2 6 Ah Valve regulated lead Paws maA MIN acid gel type batteries New batteries are Sound level 62 2 dB A provided by Liko Protection IP 30 lift unit Battery charger SMP CC 10 43 24 100 240 V AC class IP 54 hand control 40 60 Hz max 600 m A Operating Lifting speed 60 mm s 2 3 in s forces of Button on hand control 4 5 N controls Button on end cover 4 N Lifting interval 1600 mm 63 in Electrical data 24 V DC 8 5A fa Intended for indoor use Lift motor Type B in accordance with the electrical shock weight 8 7 kg 19 Ibs protection class Emergency E Class I1 equipment lowering device Mechanical and electrical Multirall is equipped with an SFS Single Fault Safety safety drum This safety design provides protection against uncontrolled lowering The lift strap has a tenfold safety Dimensions F Gilg NOY i Lateral view A F m m pi CSP Central Suspension Point _j 7 i L Overhead C view y B Measurements in mm A B C D H pes 230 295 210 387 263 1600 Measurements in inch B E D H L 9 12 8 15 10 4 63 Min distance from ceiling to CSP at max lifting height Installation dimensions the distance between the attachment point for the lift unit on the carriage and the CSP at max lifting height Lifting interval the distance
17. storage of the lift as well as an Extension Arm which facilitates the connection of the lift to the rail system Multirall is adapted to Liko s Quick Release System for safe and easy changing of lifting accessories In this document the person being lifted is referred to as the patient and the person helping them is referred to as the caregiver Aisa warning triangle used for situations which require extra care and attention Cli IMPORTANT Read the instruction guide for both the patient lift and lifting accessories before use Lifting and transferring a person always involves a certain level of risk It is important to completely understand the contents of the instruction guide The equipment should be used only by trained personnel Please contact Liko Hill Rom in the event of any uncertainties or questions Table of Contents Safety Instructions eee Definitions cccceceeeeeeeeeeeereteees Technical Data cceeeeeeeeeees DIMENSIONS osoa e ASSOMDIY oo eee eeeeeeeeeteteeeeeeeetteeeeeeeee OPE TATION penino Charging the Batteries Room to room with Multirall Maximum Load s Recommended Lifting Accessories Simple Troubleshooting 008 Inspection and Maintenance A NOTE This instruction guide contains important information for users of the product All those who use the product should review and fully understand the contents of the instructio
18. tion are sufficient for approximately 60 lift cycles The connector cable of the battery charger cannot be repaired If the cable is damaged it should be replaced by a new one 1 Check to ensure that the emergency stops are not pressed 2 Place the hand control in its intended place on the charger 3 Connect the charger to an electrical outlet 100 240 V AC The charging starts automatically The green LED indicates that the charger is connected to a power supply The yellow LED indicates that the battery is charging When the battery is fully charged the charger is switched off automatically and the yellow LED turns off NOTE If the lift is not to be used for a longer period of time the hand control should be placed in the charger If the charger is not connected to a power supply the emergency stop P button should be pressed in order to prevent the battery from discharging Multirall complies with the requirements in the 2002 96 EC directive for waste electronic and electrical equipment Used batteries should be deposited at the nearest recycling station in accordance with local regulations or given to personnel authorized by Liko Hill Rom Multirall 200 7EN125103 03 www liko com Room to Room with Multirall Liko s R2R Room to Room system is an effective solution for safe and easy transfer of patients between two or more rooms The R2R system is mounted without making openings in walls over doors and full is
19. y Convenient trolley for transfer and storage of Multirall and accessories Holder for Accessories For attaching Extension Arm Multirall Multirall Helping Hand to the Multirall Trolley Easy Switch Fits Multirall hand control with membrane key Hang Up HandControl Hanger 10 pcs set SlingBar Cover Paddy 30 Fits Universal SlingBars 350 450 and 600 as well as SlingBar Slim 350 Strap Gripper For room to room transfers Extension Belt adjustable length 30 40 cm 12 16 in 40 60 cm 16 24 in 60 100 cm 24 40 in and 100 140 cm 40 55 in Room to Room Strap length 117 cm 46 in 1770 cm 70 in and 237 cm 93 in Multirall 200 7EN125103 03 Prod Prod Prod Prod Prod Prod Prod Prod Prod Prod Prod Prod Prod No No No No No No No No No No No No No LikoScale 350 Prod No 3156228 3156512 3156513 AN 3156511 3136004 3136006 3136300 3136310 3107010 3156100 3607001 3136250 3136226 29 3136251 53 Adapter Kit LikoScale Prod No 3156232 NLA Pas ce 9 Le gidip Bl A O T x www liko com Parking panel Parking Panel 600 LR MR Prod No 3126075 Parking Panel 1500 LR MR Prod No 3126080 Complete the Parking Panel with appropriate Quick Reference Guide Hook for SlingBar Hook for Accessories charger and if necess