1. Locating holes 2 Cell holder base thumbscrew 3 Cell lifter 4 Microcell holder thumbscrew 5 Adjustment screw 6 Hexagonal ball driver To install the microcell holder 1 Place a microcell in the microcell holder if the cell has ground glass sides or sides featuring the Cary logo hold the cell by these sides and check that the cell aperture is vertically centered in the cell holder aperture Cary 50 Spectrophotometer Hardware Manual Installation 2 Ifthe cell aperture is not at the correct height remove the cell and adjust the grub screw in the microcell holder accordingly using the 2 5 millimeter hexagonal ball driver Replace the cell in the microcell holder 3 Slide back the sample compartment lid refer to Figure 2 Figure 2 Cary 50 showing the sample compartment lid 1 and front panel 2 4 Remove the sample compartment front panel refer to Figure 2 5 If not already fitted install the cell holder base in the sample compartment as follows a Place the cell holder base in the sample compartment aligning the two locating holes over the two locating pins in the floor of the sample compartment b Firmly tighten the thumbscrew on the cell holder base item 2 in Figure 1 6 Place the microcell holder on the cell holder base aligning the holes in the microcell holder over the raised black knobs in the cell holder base 7 Tighten the thumbscrew on the microcell holder You should now align the microcell hold
2. hazard they depict is shown below each symbol gt gt gt Broken glass Corrosive liquid Electrical shock gt gt gt Eye hazard Fire hazard Heavy weight danger to feet ee gt gt Heavy weight Hot surface Noxious gas danger to hands The following symbol may be used on warning labels attached to the instrument When you see this symbol refer to the relevant operation or service manual for the correct procedure referred to by that warning label 8 Cary 50 Spectrophotometer Hardware Manual Safety Practices and Hazards The following symbols also appear on the instrument or in the documentation C Indicates that a high voltage xenon flash lamp is present H Oy Indicates viewing hole to check the operation of the Xenon flash NA lamp When attached to the rear of the instrument indicates that the product complies with the requirements of one or more EU directives S p When attached to the rear of the product indicates that the product has been certified evaluated to CSA 1010 1 and UL 3101 1 Color coding The various indicator lights appearing on Agilent instruments and associated accessories are color coded to represent the status of the instrument or accessory The green indicator lamp on the front of the Cary 50 indicates the instrument is powered up that is the computer is switched on and is in normal standby condition When the indicator lamp is flashing this indicates the instrume
3. the Cary WinUV software you see an Access is denied message Solution You must be logged on with administrator rights to install the Cary WinUV software Cary 50 Spectrophotometer Hardware Manual Troubleshooting 1 Click the Windows Start button and choose Log Off 2 Logon as an administrator or ask your Windows XP administrator to log on for you Uninstall the Cary WinUV software see Page 28 Uninstall the Cary 50 driver see Page 19 Ensure the computer is turned off and the power cable has been removed Remove the Cary 50 PCI card from the computer 6 Restart the computer 7 Reinstall the Cary WinUV software and Cary 50 PCI according to the Cary 50 Installation Instructions publication number 8510185100 FCary service failed to start Problem After installing the Cary WinUV software this message is displayed The FCary service failed to start due to the following error The system cannot find the file specified Solution This indicates the Cary 50 device driver was not copied to the correct directory It is possible you may not have access rights to the installation drive To rectify this 1 Logon as an administrator or ask your Windows XP administrator to log on for you Uninstall the Cary WinUV software see Page 28 Uninstall the Cary 50 driver see Page 19 Ensure the computer is turned off and the power cable has been removed Remove the Cary 50 PCI card from the computer 5 Res
4. Cary 50 Spectrophotometer Hardware Manual oh Agilent Technologies Notices Agilent Technologies Inc 1998 1999 2002 2004 2006 2010 and 2011 No part of this manual may be reproduced in any form or by any means including electronic storage and retrieval or translation into a foreign language without prior agreement and written consent from Agilent Technologies Inc as governed by United States and international copyright laws Manual Part Number 8510159600 Edition Ninth edition March 2011 Printed in Australia Agilent Technologies Inc Warranty The material contained in this document is provided as is and is subject to being changed without notice in future editions Further to the maximum extent permitted by applicable law Agilent disclaims all warranties either express or implied with regard to this manual and any information contained herein including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing use or performance of this document or of any information contained herein Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms in the separate agreement shall control Technology Licenses The hardwar
5. ctices and Hazards e 1 Safety Practices and Hazards Lamp module 6 Electrical hazards Modules covers and panels 6 Other precautions 7 Warning and caution messages 7 Information symbols 8 Color coding 9 CE compliance 10 Your Agilent Cary 50 instrument and accessories have been carefully designed so that when used properly you have an accurate fast flexible and safe analytical system If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Information about safety practices appears throughout the documentation both hard copy and online provided with your instrument and accessories to help you safely operate the instrument and accessories Before using the instrument or accessories you must thoroughly read these safety practices ALWAYS operate the instrument and accessories in accordance with these safety practices Cary 50 Spectrophotometer Hardware Manual 5 Safety Practices and Hazards Lamp module The lamp is enclosed in a self contained module This module contains components operating at high voltages To avoid electric shock NEVER disassemble the module When operating the lamp module emits high intensity light which can cause serious damage to eyes To avoid eye damage never operate the lamp outside the instrument Electrical hazards Modules The Cary 50 spectrophotometer is powered by the computer control
6. dures not specifically mentioned in this chapter should be carried out only by Agilent trained Agilent qualified or Agilent authorized customer support representatives M Eye Hazard This instrument contains an intense light source Direct viewing of the light source will cause eye damage Operators and other unauthorized personnel must NEVER remove the main cover This section refers only to maintenance procedures for the instrument You should refer to your computer and printer plotter manuals for their maintenance procedures and to the Cary WinUV online Help for the maintenance procedures for any accessories you ordered Cary 50 Spectrophotometer Hardware Manual 21 Maintenance Cleaning Any spills in the sample compartment should be immediately wiped up The exterior surfaces of the Cary 50 spectrophotometer should be kept clean All cleaning should be done with a soft cloth If necessary this cloth can be dampened with water or a mild detergent Do not use organic solvents or abrasive cleaning agents Lamp module This section describes how to replace the lamp module and realign the light beam Before changing the lamp module ALWAYS disconnect the Cary 50 from the computer These instructions are also provided in the Cary WinUV Help complete with a video demonstration Refer to the Cary WinUV Software Manual part number 8510162500 for details on using the online Help Replacing Shock and Eye Hazard
7. e and or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government prime contract or subcontract Software is delivered and licensed as Commercial computer software as defined in DFAR 252 227 7014 June 1995 or as a commercial item as defined in FAR 2 101 a or as Restricted computer software as defined in FAR 52 227 19 June 1987 or any equivalent agency regulation or contract clause Use duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms and non DOD Departments and Agencies of the U S Government will receive no greater than Restricted Rights as defined in FAR 52 227 19 c 1 2 June 1987 U S Government users will receive no greater than Limited Rights as defined in FAR 52 227 14 June 1987 or DFAR 252 227 7015 b 2 November 1995 as applicable in any technical data Safety Notices A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in damage to the product or loss of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met A WARNING notice denotes a hazard It calls attention to an operating
8. er Cary 50 Spectrophotometer Hardware Manual 17 Installation Alignment To align the microcell holder 1 Start the Align application by clicking the Windows Start button and choosing All Programs Agilent Cary WinUV or Agilent UV UV FO Dissolution then Align Click the Cary tab Under Instrument Parameters set the wavelength to 0 nm white light by selecting Zero Order Click Apply The green power indicator on the instrument should start flashing to indicate that the instrument is active Place a cell in the microcell holder if you have not already done so Place a small piece of white paper in the light path to the right of the cell If the beam appears as though it will strike the cell aperture move the paper to the left of the cell and check that the beam is passing through the cell If the beam does not appear as though it will pass through the cell adjust the height of the cell as described on Page 17 Using the 2 5 millimeter hexagonal ball driver adjust the microcell holder adjustment screw item 5 in Figure 1 and note the intensity of the light striking the paper Continue to adjust the adjustment screw until the beam hitting the paper appears the most intense You may need to dim the room lights to see the light beam Other sample holders Other sample holders are available for use with the Cary 50 such as the Solid Sample Holder Instructions for their use are included in the on
9. erconnecting cable 0110637390 3M disposable grounding wrist strap 7910031300 Mouse mat 7910022300 Cary 50 Spectrophotometer Hardware Manual 37 Spare Parts 38 Figure 8 Cary 50 showing 1 Sample compartment lid 2 Sample compartment front panel 3 Lamp mirror adjustment cover 4 Accessory cable port plate at rear Ordering details for other Cary 50 accessories are available on the Agilent Web site www agilent com Cary 50 Spectrophotometer Hardware Manual
10. he pReinstallation manual for future reference If you have misplaced your copy you can obtain a free replacement from your local Agilent office Alternatively download a PDF from the Agilent Web site www agilent com Documentation You have been provided with the following documentation to help you set up and operate your Agilent Cary 50 spectrophotometer Installation guides one for each supported operating system with information on unpacking the instrument installing the interface card in the computer and setting up the system This operation manual with safety practices and hazards information instructions for installing and maintaining the components of the Cary 50 and troubleshooting information Cary 50 Spectrophotometer Hardware Manual 11 Introduction Cary WinUV Software Manual publication number 8510162500 with instructions for installing the Cary WinUV software an overview of the software and software related troubleshooting information Extensive online Help provided with the Cary WinUV software containing context sensitive Help step by step instructions for frequently performed analyses and instructions for using any accessories you ordered Conventions The following conventions have been used throughout the documentation Menus menu items buttons and check boxes have been typed in bold For example click OK and From the Edit menu choose Copy ALL CAPITALS indicate keyboard co
11. it should change to Start Cary 50 Spectrophotometer Hardware Manual 31 Troubleshooting Hard disk power supply connection Problem There is no spare hard disk power supply connector in the computer to connect to the Cary 50 PCI card Solution 1 With the disposable grounding wrist strap supplied with the Cary 50 fitted plug the appropriate end of the supplied Y connector refer Figure 6 into the connector at the end of the Cary 50 PCI card NI jo y Figure 6 Y connector supplied with the Cary 50 2 Unplug a hard disk power supply connector and connect it to the Y connector 3 Connect the free end of the Y connector to the hard disk Absorbance is 10 and fluctuates wildly during a scan Problem The Cary 50 is reporting 10 Abs and a scan like that shown in Figure 7 is displayed on the screen 32 Cary 50 Spectrophotometer Hardware Manual Troubleshooting 10 T T 700 750 800 Wavelength nm Figure 7 Wild absorbance fluctuation during a scan Solution The detector cable inside the Cary 50 sample compartment is not connected Check that the 8 pin plug on the left side of the sample compartment is firmly plugged in Instrument performance testing Problem The results of your instrument performance tests do not meet specifications the results obtained during factory testing are included in the packing crate with
12. le corrosive toxic or otherwise hazardous Careless improper or unskilled use of such materials solvents and solutions can create explosion hazards fire hazards toxicity and other hazards which can result in death serious personal injury and damage to equipment and property ALWAYS ensure that laboratory safety practices governing the use handling and disposal of such materials are strictly observed These safety practices should include the wearing of appropriate safety clothing and safety glasses Warning and caution messages Carefully read all warnings and cautions and observe them at all times A Warning message is used in the text when failure to observe instructions or precautions could result in death or injury Warnings have the following format Hazard Type Nature of the hazard information on how to avoid the hazard and possible consequences if you don t The triangular symbols that appear in conjunction with warnings are outlined in the next section Cary 50 Spectrophotometer Hardware Manual 7 Safety Practices and Hazards A Caution message is used when failure to observe instructions could result in damage to equipment Agilent supplied and or other associated equipment Cautions have the following format CAUTION Caution information appears here Information symbols The following triangular symbols appear in conjunction with warnings on the spectrometer and associated documentation The
13. line Help provided with the Cary WinUV software Refer to the Cary WinUV Software Manual part number 8510162500 for details on using the online Help Cary 50 Spectrophotometer Hardware Manual Installation Uninstalling the Cary WinUV software For Pharma software you will need to contact your local Agilent office as a service call will be required to uninstall Cary WinUV Pharma software To uninstall the Cary WinUV Software 1 Click the Windows Start button and choose Control Panel gt Add Remove Programs Scroll through the list on the Install Uninstall tabbed page until you find Agilent Cary WinUV Click Add Remove and then Yes Follow the instructions on the screen Repeat Step 3 for the Cary WinUV Help Uninstalling the Cary 50 driver To uninstall the Cary 50 driver 1 Click the Windows Start button and choose Control Panel gt System On the Hardware tab select Device Manager and click the plus icon next to Other devices Right click Cary 50 PCI Rev03 and then click Uninstall Restart the computer Cary 50 Spectrophotometer Hardware Manual 19 Installation 20 This page is intentionally left blank Cary 50 Spectrophotometer Hardware Manual Maintenance e 4 Maintenance Cleaning 22 Lamp module 22 This chapter includes the Agilent Cary 50 spectrophotometer maintenance procedures that may be carried out by an operator Any maintenance proce
14. ling the instrument Safe operation of the instrument depends on the integrity of the switching power supply of the computer Your computer must comply with IEC 60950 covers and panels The ONLY modules of the Cary 50 spectrophotometer that operators and other unauthorized personnel are permitted to access are the lamp module on the underside of the instrument and the sample compartment White circles indicate the screws you need to undo to remove the lamp module ALWAYS switch off the computer before changing a lamp The only panel you are permitted to remove is the snap out panel covering the lamp mirror adjustment screws on the front of the instrument Any other panels or covers that are retained by screws on the spectrophotometer and accessories may be opened ONLY by Agilent trained Agilent qualified or Agilent approved customer support representatives Consult the manuals or product labels supplied with your computer monitor and printer to determine which parts are operator accessible The safety classification is given as Safety Class III EN 61010 1 Cary 50 Spectrophotometer Hardware Manual Safety Practices and Hazards Other precautions Do not block any ventilation grilles present on the computer Consult the manuals supplied with your computer monitor and printer for their specific ventilation requirements Use of the Cary 50 system and accessories may involve materials solvents and solutions that are flammab
15. mmands For example Press ENTER and Press SHIFT F93 Throughout this manual UV Dissolution and UV Fiber Optic Dissolution users should replace Cary WinUV Software with UV Dissolution Software or UV Fiber Optic Dissolution Software respectively Specifications The instrument is suitable for indoor use only and is classified Pollution degree 2 and Installation Category II EN 61010 1 Cary 50 Spectrophotometer Hardware Manual Introduction Environmental Table 1 Suitable conditions during instrument transportation non operation and operation Condition Altitude m ft Temperature Relative humidity C F non condensing Non operating transport 0 3050 0 10 000 5 45 41 113 20 80 Operating within performance 0 3050 0 10 000 10 35 50 95 8 80 specifications For optimum analytical performance it is recommended that the ambient temperature of the laboratory be between 20 and 25 C and be held constant to within 2 C throughout the entire working day Weights and dimensions Weights Packed 23 kg 50 7 1b Unpacked 21 kg 46 Ib Dimensions width x depth x height Packed 615 x 710 x 360 mm 24 x 28 x 14 in Unpacked 590 x 500 x 200 mm 23 x 20 x 8 in Computer power supply To ensure safe operation of the instrument the switching power supply in the computer must comply with standard IEC 950 or equivalent Cary 50 Spectrophotometer Hardware Man
16. nce Aligning the source mirror For optimum performance of the instrument the source mirror should be aligned to suit the new lamp module To align the source mirror 1 Start the Align application by clicking the Windows Start button and choosing All Programs Agilent Cary WinUV or Agilent UV UV FO Dissolution then Align On the Cary tabbed page set the following parameters Beam Mode Single Beam Normal Y Mode T Ave Time 0 5000 Wavelength 500 nm Click Apply Remove the snap out panel refer to Figure 5 from the front of the instrument using a flat blade screwdriver to pry the panel open This exposes the two source mirror adjustment screws Do not remove the plastic bung next to the adjustment screws Cary 50 Spectrophotometer Hardware Manual 25 Maintenance Figure 5 The snap out panel 1 on the front of the Cary 50 On the Lamps tabbed page monitor the aqua colored Current Signal bar as you use the supplied 2 5 millimeter hexagonal ball driver to slowly adjust one of the adjustment screws at the front of the instrument If the length of the Current Signal bar decreases slowly turn the screw in the other direction You can expect to see some fluctuation due to noise If the signal is out of range or excessively weak or strong click Rescale to bring the signal back into range for display Continue to adjust the screw until the length of the Current Signal bar is maximi
17. nt is busy Cary 50 Spectrophotometer Hardware Manual 9 Safety Practices and Hazards CE compliance Agilent Cary 50 instruments have been designed to comply with the requirements of the Electromagnetic Compatibility EMC Directive and the Low Voltage electrical safety Directive commonly referred to as the LVD of the European Union Agilent has confirmed that each product complies with the relevant directives by testing a prototype against the prescribed EN European Norm standards Proof that a product complies with the directives is indicated by The CE marking appearing on the rear of the product The documentation package that accompanies the product containing a copy of the Declaration of Conformity This declaration is the legal declaration by Agilent that the product complies with the directives and also shows the EN standards to which the product was tested to demonstrate compliance Cary 50 Spectrophotometer Hardware Manual Introduction e 2 Introduction J Installation requirements 11 Documentation 11 Specifications 12 Installation requirements Before receiving your instrument you will have been provided with a Cary 50 PReinstallation Manual publication number 8510161800 which describes the environmental and operating requirements of the Cary 50 spectrophotometer You must prepare your laboratory according to these instructions before the Cary 50 can be installed You should keep t
18. procedure practice or the like that if not correctly performed or adhered to could result in personal injury or death Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met Cary 50 Spectrophotometer Hardware Manual Contents Contents 1 Safety Practices and Hazards 5 Lamp module 6 Electrical hazards 6 Modules covers and panels 6 Other precautions 7 Warning and caution messages 7 Information symbols 8 Color coding 9 CE compliance 10 2 Introduction 11 Installation requirements11 Documentation 11 Conventions 12 Specifications 12 Environmental 13 Weights and dimensions 13 Computer power supply 13 Connections 14 3 Installation 15 Cell holder 16 Alignment 18 Other sample holders 18 Cary 50 Spectrophotometer Hardware Manual 3 Contents Uninstalling the Cary WinUV software 19 Uninstalling the Cary 50 driver 19 4 Maintenance 21 Cleaning 22 Lamp module 22 Replacing 22 Aligning the source mirror 25 5 Troubleshooting 27 Windows XP service or driver failure 27 Access denied during installation28 FCary service failed to start 29 No green light indicating that the Cary 50 is powered 30 Start button replaced with Connect button 30 Hard disk power supply connection 32 Absorbance is 10 and fluctuates wildly during a scan 32 Instrument performance testing 33 Wavelength calibration 35 6 Spare Parts 37 4 Cary 50 Spectrophotometer Hardware Manual Safety Pra
19. r application for example Scan before this initialization has finished the application s Start button will be unavailable Wait for the status line at the bottom of the application to display Idle and the Start button will become active 30 Cary 50 Spectrophotometer Hardware Manual Troubleshooting Problem The Start button will not become active or the Connect button will not change to Start Solution If the Cary software cannot locate the Cary 50 PCI card the Start button will not become active or the Connect button will not change to Start This is due to either A memory address conflict with the Cary 50 PCI card and another card in the computer An IRQ conflict with another card An I O address conflict with another card A defective Cary 50 PCI card To fix the problem when using Windows XP 1 Uninstall the Cary WinUV software see Page 28 2 Uninstall the Cary 50 driver see Page 19 3 Ensure the computer is turned off and the power cable has been removed Remove the Cary 50 PCI card from the computer Restart the computer Reinstall the Cary WinUV software and Cary 50 PCI according to the Cary 50 Installation Instructions publication number 8510185100 6 Check that the computer is now communicating with the Cary instrument when you open an application the Start button should become active or when the Connect button is clicked
20. s When operating the lamp module emits high intensity light which can damage eyes To avoid eye damage never operate the lamp module outside the instrument The lamp module contains components operating at high voltages To avoid electric shock NEVER disassemble the lamp module 22 Cary 50 Spectrophotometer Hardware Manual To remove the lamp module 1 Remove the plug from the inst connector on the rear panel of the Cary 50 Turn the Cary 50 onto its side to give access to the base Using a star head screwdriver undo and remove the six screws marked by the white solid circles on the lamp module cover refer to Figure 3 As you remove the last screw use one hand to support the lamp module Figure 3 Underneath the Cary 50 The six screws to be removed are numbered 4 Hold the lamp module out from the instrument and unplug the 3 pin connector from the Cary 50 refer to Figure 4 Discard the lamp module Maintenance 24 Figure 4 Unplugging the 3 pin connector To fit a new lamp module 1 Plug the 3 pin connector of the new lamp module into the instrument 2 Fit the lamp module in the base of the Cary 50 ensuring that the wiring is kept clear of the other parts of the instrument 3 Replace and tighten the six screws Return the instrument to its upright position and replace the cable at the rear 5 Reconnect the computer to the Cary 50 Cary 50 Spectrophotometer Hardware Manual Maintena
21. t which can damage eyes To avoid eye damage never operate the lamp module outside the instrument The lamp module contains components operating at high voltages To avoid electric shock NEVER disassemble the lamp module 34 Cary 50 Spectrophotometer Hardware Manual Troubleshooting Wavelength calibration Problem The position of an absorption transmission or reflectance peak appears to shift over a prolonged period of time Solution Perform a wavelength calibration Restart the computer Agilent recommends that as part of your standard operating procedure you restart the computer attached to the Cary 50 monthly Perform the Instrument Performance Test suite in Validate to ensure the instrument results meet specification Cary 50 Spectrophotometer Hardware Manual 35 Troubleshooting This page is intentionally left blank 36 Cary 50 Spectrophotometer Hardware Manual Spare Parts 7 6 Spare Parts These spare parts are available for use with your Agilent Cary 50 spectrophotometer Always use Agilent supplied spare parts unless otherwise indicated Part Part number Sample compartment lid 1 in Figure 8 0910127200 Sample compartment front panel 2 in Figure 8 0110642990 Lamp mirror adjustment cover 3 in Figure 8 0910127400 Accessory cable port plate 4 in Figure 8 0910124000 Xenon lamp module 0110639690 Cary 50 PCI card 0210155790 Cary 50 computer int
22. tart the computer Reinstall the Cary WinUV software and Cary 50 PCI card according to the Cary 50 Installation Instructions publication number 8510185100 Cary 50 Spectrophotometer Hardware Manual 29 Troubleshooting No green light indicating that the Cary 50 is powered The Cary 50 is powered directly from the computer power supply A green power indicator light on the front of the Cary 50 indicates when the instrument is powered Problem The power indicator on the Cary 50 does not light when the computer is switched on Solution Check the connection of the main instrument cable connecting the computer to the Cary 50 If the power indicator on the front of the Cary 50 is still not on check the connection of the computer power supply connector to the Cary 50 card inside the computer Instructions for the connection of the power supply connector are listed in the installation guide supplied with your Cary 50 and in the Hard disk power supply connection section on Page 31 if you need to use a Y connector Start button replaced with Connect button Problem I have a Connect button instead of a Start button Solution Only one Cary WinUV at a time can communicate with the instrument If you want to change to another application you can click the Connect button to bring this application online When you turn on the computer the Cary 50 calibration program initializes the Cary 50 If you start anothe
23. the instrument Solution Check that The sample compartment is empty The cable connecting the instrument to the computer is correctly connected and the retaining screws are tightened Cary 50 Spectrophotometer Hardware Manual 33 Troubleshooting The lamp is pulsing This is indicated if the green power indicator on the front of the instrument flashes you should also hear the monochromator and the filter wheel moving You should also turn the instrument on its side and look through the small lamp viewing hole in the base of the instrument If the lamp is not pulsing you may have a hardware conflict The lamp is correctly aligned see Page 25 for instructions on aligning the lamp Calibrated appears on the Calibration tab in System Information To open System Information double click the Cary WinUV folder or the UV Dissolution or UV FO Dissolution folder on the Windows desktop and then double click the System Information icon If the text reads UnCalibrated on the Calibration tabbed page select Initialize Wait for the instrument to finish initializing and click Calibrate When the calibration routine finishes the text will change to Calibrated Your instrument is now calibrated and is ready to use Repeat the Instrument Performance Test suite in Validate WARNING Eye and Shock Hazards When operating the lamp module emits high intensity ligh
24. ual 13 Introduction The power requirements of the Cary 50 must be met by the computer power supply and the following capacity in addition to what is required by the computer itself is required 4 5VDC lt 1A 412VDC lt 1L5A 12 VDC lt 0 25 A a total of 26 W for the instrument Allowance must also be made for external accessories powered by 12 V DC Operating motor driven accessories will increase the 12 V current by a further 2 A a maximum of 24 W for accessories Agilent recommends a computer power supply rating of 220 W asa minimum Connections 3 5 mm phono jack socket in the left side of the sample compartment for accessories 8 pin DIN connector in the left side of the sample compartment for the diode detector 25 pin D range connector in the right side of the sample compartment for accessories optional If an external cooling air supply system is required an air inlet duct attachment part number 0110595300 must be ordered with the instrument 14 Cary 50 Spectrophotometer Hardware Manual Installation e 3 Installation Cell holder 16 e r Uninstalling the Cary WinUV software 19 Uninstalling the Cary 50 driver 19 The Agilent Cary 50 spectrophotometer is designed to be completely customer installed Instructions for setting up the system are included in the installation guides supplied with the instrument You should select the one appropriate for the operating s
25. ver will be indicated by a red stop sign on the left side Cary 50 Spectrophotometer Hardware Manual 27 Troubleshooting 3 Double click the failed event to determine the cause of the failed device If it was the Cary 50 that failed to load you will see the following description The FCary service failed to start due to the following error The device is not ready This generally means that the device driver loaded but there was no response from the Cary 50 instrument This indicates that the Cary 50 interface hardware is not responding 4 Ensure that the Cary 50 card is correctly installed and that the interconnecting cable is secure Solution 2 Another cause of this error may be due to a conflict between the Cary 50 driver and another driver on the system The most common cause is a sound card for which a driver has not been installed The computer may have a sound card built into the motherboard as well as a separate sound card that the computer recognizes but for which it does not have a driver installed You will need to 1 Click the Windows Start button and choose Control Panel gt System Click the Hardware tab and select Device Manager 2 Ifyou see an exclamation mark next to one of the devices you need to fix the problem usually by installing the latest driver before installing the Cary WinUV software and Cary 50 driver Access denied during installation 28 Problem During installation of
26. ystem you are using on your computer M Heavy Weight Hazard The Cary 50 weighs over 20 kg 44 Ib To avoid injury to personnel or damage to equipment always use two or more people when lifting or carrying the instrument NEVER attempt to lift the instrument alone Details for unpacking the instrument and what to do in case it has been damaged in transit are outlined in the pReinstallation manual Following the instructions in the Cary 50 Installation Instructions Publication Number 8510185100 you should have Unpacked the spectrophotometer and placed it on the intended workbench Connected the computer to the power supply Installed the Cary WinUV software Installed the instrument interface card in the computer if you supplied your own computer Cary 50 Spectrophotometer Hardware Manual 15 Installation Installed the Accessory Cable kit Gf ordered Connected the instrument to the computer Completed the instrument performance tests This chapter describes how to install the sample holders used with the Cary 50 and how to uninstall the software and driver should you ever need to Instructions for installing replacing the lamp module are included in the Maintenance chapter Cell holder A microcell holder see Figure 1 is supplied as standard with the Cary 50 Follow these instructions to install and align it x o Figure 1 Cell holder base left and microcell holder right 1
27. zed When the length of the Current Signal bar is at its greatest repeat for the other adjustment screw NOTE It does not matter in which order the adjustment screws are aligned 26 Replace the snap out panel on the front of the instrument The source mirror is now aligned and the instrument is ready for use Cary 50 Spectrophotometer Hardware Manual Troubleshooting 5 Troubleshooting Windows XP service or driver failure 27 Access denied during installation 28 FCary service failed to start 29 No green light indicating that the Cary 50 is powered 30 Start button replaced with Connect button 30 Hard disk power supply connection 32 Absorbance is 10 and fluctuates wildly during a scan 32 Instrument performance testing 33 Wavelength calibration 35 This chapter contains troubleshooting information to help you solve various problems you may encounter when setting up or using your Agilent Cary 50 hardware Windows XP service or driver failure Problem When you start the Microsoft Windows XP operation system you see a message Service Control Manager At least one service or driver failed during system startup Use Event Viewer to examine the event log for details Solution 1 1 Click the Windows Start button and choose Control Panel gt Administrative tools gt Event Viewer 2 Inthe Event Viewer find the hardware device driver that failed to load during startup event This dri