Home
        KODAK MIN-R L Film System User Guide - Spectrum Medical X
         Contents
1.       KODAK MIN R L Film System  User Guide    W a en a  A     4 f avra be     4    Ng I     ci as  n    AA VETES       KODAK MIN R L Film System    User Guide    The following information is intended  only as a guide for troubleshooting the  KODAK MIN R L Film System  site  specific conditions may cause different  results  For optimum results  KODAK  MIN R Films  Screens  and Cassettes  should be used as part of a system with  KODAK Processing Chemicals and  Equipment  If you have any questions  concerning the information contained in  this guide  contact your local Kodak  representative or contact Health  Imaging Technical Support at  1 800 336 4722  716 724 9362  or on  Kodak   s Health Imaging Internet site   www  kodak com go mammo     How to Use This Guide    The KODAK MIN R L Film System User  Guide includes the following sections  to provide information and to help  optimize image quality    e Keys for Properly Processing  KODAK MIN R L Film   e Sensitometric Variability   e Optimizing Exposure   e Minus Density Artifacts   Shadow Images Pick Off   Processor Digs    e A Method for Evaluating Film  Performance in Other Manufacturers     Developer Solutions   e Fixer Considerations   e Drying   e Other Technical Considerations   e Troubleshooting Tools   e Uniformity of Screen Speed   e Information Needed by Kodak to  Facilitate Troubleshooting    Description of  KODAK MIN R L Film    e KODAK MIN R L Film for mammography  was designed for radiologists who rate  visualiz
2.   10  in optical  density at the speed step  e  0 15 for average gradient of the films  processed through KODAK Developer     To verify adequate developing  the  optical density of the 21st step should  be measured  If the optical density is  lower than 4 0  for KODAK MIN R L Film  with KODAK RP X OMAT Developer   the  test should be declared invalid  Other  factors could affect the validity of the  test  Please consult Kodak if you   have questions     Typically Acceptable  Results       Results are based on testing in fresh   developer  using the manufacturer s   recommended starter volumes  We   recommend using 6 ounces per     30 gallon of KODAK RP X OMAT  Developer Starter  Actual sensitometric  results may vary with seasoning     15    Fixer Considerations    Assessing the Results of the  Fixer  Hypo  Retention Test    A fixer  hypo  retention test should be  performed regularly  i e   at least  quarterly  to make sure mammography  films are properly washed and retain a  minimal amount of fixer  Excessive fixer  retention by the film may result in image  degradation over time  It would then be  of less or no value when making  comparisons with subsequent  mammographic studies  The procedure is  outlined in the instructions included in  the Hypo Retention test kit     It is important to note the following  points regarding assessing the results of  the test    e The Hypo Test Solution must be   applied to the primary emulsion side of   KODAK MIN R L Film   There should b
3.   Screen Speed    Objective   To assess the uniformity of the  radiographic speed of image receptors  routinely used for mammographic  imaging     Caveats for Success   e Set aside enough film from a single  emulsion lot to perform the test on  all cassettes    e Automatic Exposure Control  AEC  of  the mammography x ray unit must be  performing within standards    e Processor must be in control    e Feed all films through the same side of  the processor    e Measure the optical density of the film  in the same place on each film     Required Test Equipment  e The image receptors normally used for  mammographic imaging   e Film of the type used for  mammography  Note that the use of a  single emulsion lot of film  and most  preferably  film from a single box  is  very important in obtaining useful  results from this test   A 4 0 cm thick cassette sized phantom  made of either acrylic or BR 12  Some  provision should be made on the  phantom to allow density  measurements to be made in the same  place on each of the test radiographs   If density measurements are not taken  from the same location  significant  variation can be obtained due to  variation in beam intensity   e A spot reading densitometer     Test Procedure Steps   1  Identify all of the image receptors   cassettes  of each size and type to be  evaluated with some form of numbering  system for further reference     2  Select one of the cassettes and load it  with a sheet of film from the box set  aside for this tes
4.   current clinical images and images taken  one or more years previously  Such  comparisons may assist in  troubleshooting image quality concerns     The easiest way to retain this  information is to start an emulsion  number log  The log should contain the  following information    e Type of film    e Date the box of film was opened    e Expiration date    o Film size    e Complete emulsion number    e Any other comments  such as which  emulsions were used for processor QC  or phantom images  Refer to the  example above of a emulsion number  log  Every time a box of film with a  different emulsion number is opened     a new entry should be made in the log     KODAK MIN R 2000 Film    DATE EXPIRATION   OPENED DATE FiLM SIZE   12 22 99 02 01 18 x 24 cm  02 01 24 x 30 cm   2 02 00 03 01 18 x 24 cm    Split Phantom Test    A split phantom test should be  performed to radiographically determine  relative speed differences between two  different boxes of film  one of which is  suspected of being much faster or slower  than the film in current use for either  clinical films or for processor quality  control  Speed comparisons made using a  sensitometer may not accurately reflect  the differences in speed between two  films exposed by light from an  intensifying screen        CUT EDGE  gt  CUT EDGE map          S   N          The procedure is as follows     1  Assemble the tools that are needed for   the test    e A phantom used for mammography  quality control testing    e The 18 x
5.   tions  are shown below  For the emulsion  shown in this example  the highest con   trast occurs between optical densities  1 45 to 2 2     viewing conditions and radiologists     preferences may result in a lower  selected optical density than the maxi   mum  2 0 in the above curve   Note   however  that the contrast optical den   sity curve in this example demonstrates  highest contrast over a range of optical  densities  1 45 2 2   A selected optical  density within this range will still pro   duce very high image quality  A lower  preferred optical density may result in  slightly lower contrast  but will provide  more exposure latitude     Characteristic Curve    Density              RP      EI                Selecting Optical Density    Procedure  2  using the same tools as  in Procedure  1  page 7  for assessing  if the optical density or densities  tested above are optimal  i e   provide  the highest image quality     1  Using the kVp of the preferred phan   tom image  Procedure  1   expose a  series of phantom images using photo   timing  automatic exposure control  and  varying the density control setting  For  example  make exposures using the  2    1  0   1  and  2 settings on the den   sity control  Aim for the background  optical density of all phantom images to  range from a minimum of 1 20 to  approximately 1 80  Use the densitome   ter to check the actual optical density in  the center of each image     2  As done for the kVp series in  Procedure  1  prepare 
6.  24 cm mammography cassette  normally used for the phantom test    e A piece of cardboard from the film box  cut in half to use as a guide    e A pair of scissors    e A lead pencil     EMULSION   NuMBER QC Comments   103 016 11 Jv   101 011 15   105 023 10 y Re established QC    The mammography x ray unit and the  processor will also be used for this test   Note that roomlight film handling  systems may not be able to remove the  two pieces of film from the cassette or  the films may not transport properly  If  the roomlight film handling equipment is  installed through the darkroom wall and  has a back door  the two films may be  processed using that route  Alternatively   the two films may be processed in the  processor used as backup for the  mammography processor  or in any   other processor available elsewhere in  the facility     2  In the darkroom  in total darkness to  reduce any additional density added to  the films due to long safelight exposure   cut a sheet of film from the current box  in half by using the cardboard as a  guide   This can be done by lining up the  18 cm edges of the cardboard and film   with the film closest to the countertop  and the card board half on top  Use care  in cutting the film in the dark      3  Place the film primary emulsion side  up in the cover of the opened cassette  with the film on the right side and the  cut edge toward the right edge of the  cassette  use a lead pencil to mark the  corner N for normal     4  Cut a sheet of 
7.  In the recent  past  exposure times of 3 to 5 seconds  were common  especially for previous  generations of mammographic x ray  equipment with three phase  six pulse  and single phase generators  Modern  x ray equipment usually employs high   frequency constant potential generators  which are more efficient  Motion is still a  significant problem even if using modern  equipment if long exposure times are  used  To minimize motion  breast tissue  should be adequately compressed  and  all exposures should ideally last 1 to 1   seconds  with 2 seconds generally  considered the maximum     Note that it may still be necessary for  exposure techniques to last longer than 2  seconds when performing magnification  views or when imaging dense breast  tissue  A higher kVp setting and or a  faster intensifying screen  such as the  KODAK MIN R 2190 Screen  may be used  to reduce long exposure times and motion  during magnification or with very dense  breast tissue  Additionally  any  combination of factors in the imaging  chain which results in an extremely low  dose  e g   less than 1 0 mGy  100  millirads   should be avoided  As total  system speed increases above 200  noise  increases and detail visibility decreases     Reproducibility   The capabilities of the x ray equipment  to provide reproducible exposures   consistent optical density  when using a  high contrast screen film combination  must also be assessed  As the contrast of  a screen film combination increases  its  exposur
8.  control system   technique charts showing the appropriate  program s  should also be posted on  each piece of equipment   Refer to the  manufacturer of the equipment as to  their recommendations for choosing the  correct program for specific screen film  combinations and technical parameters   It is very important that all types of  mammographic x ray equipment be  initially calibrated for a new screen   film combination      All technologists operators should be  trained on the use of the technique  charts and on any differences associated  with the use of a different screen film  combination  It is important to note that  as technology changes occur  different  practices may be not only necessary but  absolutely critical to obtain optimal  image quality  In particular  higher kVp  or different technical parameters than  used previously with lower contrast  screen film combinations may   be appropriate     Other Tips    e Be as precise as possible when  positioning the patient and the  photocell  photoreceptor     e Ensure that adequate compression  is employed    e Installation of new film should be  accompanied by mammographic x ray  equipment calibrated by qualified  personnel    e A medical physicist should be  consulted with any questions  or for  any help with this procedure     Minus Density Artifacts  Shadow Images   Pick Off Processor Digs     Minus density artifacts are those which  appear white or lighter than the  surrounding background  They are of  particular 
9.  equipment  to control extraneous light    e Control extraneous light within all  viewing areas    e Consult the instructions in the  Kodak Viewing Conditions Test Tool   Kodak publication No  M7 207  CAT No   150 1915     kVp    The traditional rules of thumb for  mammographic imaging have been that  the lowest kVp be selected in order to  achieve the highest possible subject  contrast while maintaining an acceptable  dose  e g   less than 3 0 mGy  milligrays    300 millirads   and an exposure time of  less than 2 seconds to reduce motion   The range of 24 to 28 kVp has typically  been used worldwide  However  the use  of higher kVp with KODAK MIN R L Film  may be acceptable     All other things being equal    e As kVp is decreased  the length of  the exposure increases  which in  turn increases the risk of motion on  clinical films    e As kVp is increased  the length of the  exposure decreases     Procedure  1 for assessing the effect  of varying kVp on image quality    The following procedure assumes  satisfactory kVp accuracy and  reproducibility     1  Assemble the tools which will be   needed    e Mammographic phantom  e g   phantom  that adheres to the ACR mammographic  accreditation phantom     e Mammographic cassette of the type  used clinically  ideally  the same  cassette should be used for all  exposures     e Fresh box of mammography film  all  images must be made using the same  emulsion     e Densitometer  7    Optimizing Exposure  continued     2  For each ex
10.  mark the  emulsion orientation  up  U  or down    D    as well as which side of the  processor feed tray  right  R  or left   L   is being used  when feeding the  film into the processor  e g   TUR    Process one film so that the narrowest   dimension of the film is the leading   edge  process a second film so that the  widest dimension of the film is the  leading edge       To thoroughly evaluate the entire  processor  up to eight films should  be exposed and processed as  described above  one pair primary  emulsion up on the right side of the  film feed tray  one pair primary  emulsion up on the left side of the  feed tray  one pair primary emulsion    18    down on the right side of the film  feed tray  and one pair primary  emulsion down on the left side of  the film feed tray    Expose the first two film pairs which   will be processed as described above    primary emulsion up on right  primary   emulsion up on left  using the small   moving grid device  18 x 24 cm bucky   and use to evaluate whether artifacts  are being generated by the x ray unit  in conjunction with the small bucky    Expose the third and fourth film pairs   which will be processed as described   above  primary emulsion down on  right  primary emulsion down on left   using the large moving grid device    24 x 30 cm bucky  and use to   evaluate whether artifacts are being   generated by the x ray unit in  conjunction with the large bucky    Once processed  view the film pairs    or single films  in th
11.  of silver recovery     Drying    The last step in processing an x ray  film is drying  Drying prepares the film  for handling  viewing  and storage   The physical characteristics of   KODAK MIN R L Film should allow for  good drying     A blower in the processor supplies  heated air to the dryer section  Most of  the warm air is recirculated  the rest is  vented to prevent buildup of excessive  humidity in the dryer  Proper processor  ventilation is important to achieve  proper film drying and decrease surface   pattern  or drying  artifacts  Refer to   Service Bulletin No  101  October 1992    Dryer Venting Requirements for All    KODAK X OMAT Processors  publication No     N 923  FaxBack document No  800220     The dryer temperature should always be  set as low as possible while still being  consistent with good drying  Use of a  lower setting will also result in energy  savings for processor operation     The dryer temperature should never  exceed the film manufacturer s  recommendations     A common tendency is to raise dryer  temperature higher than recommended   especially in response to films not  thoroughly dried  or still tacky  exiting  the processor  Over dried films usually  exhibit severe drying artifacts that are  visible on transmitted light as well as on  reflected light     For these circumstances  check the  quality of the chemicals  both developer  and fixer  and that replenishment rates  and processor ventilation are  appropriate   Chemical quality and  a
12.  processor service firm  DD  J slight    Check film expiration date  B F  T slight    Manage film inventory using first in  first out method  MD  J slight    Store unprocessed unexposed film and processed radiographs  B F  T between 50 to 70  F  10 to 21  C  and 30 to 50  relative    humidity    5 Base Plus Fog values for KODAK MIN R L Film are generally stable  T or     gt  may occur  as indicated in the    Comments    column  J does not generally occur     TRENDS IN GRAPH    MD  1  DD  T  B F   gt        POSSIBLE CAUSES    Slight contamination of  developer with fixer    Developer temperature higher  than recommended   within 5  F  3  C      Water temperature higher than  recommended  if developer  temperature above set point     Developing time longer than  recommended    Over replenishment    No starter added to fresh  developer in processor  developer tank    Insufficient amount of  starter added    Developer temperature lower  than recommended   within 5  F  3  C      Water temperature lower  than recommended    if developer temperature  below set point     Developing time shorter  than recommended    COMMENTS    DD  T slight    B  F     slight    B  F     slight    B  F     slight    MD     slight  DD  4 slight  B  F     slight    MD     slight  B  F     slight    MD     slight  DD  J slight  B F   gt     DD   gt     DD   gt     CORRECTIVE ACTIONS               Call service to clean the processor      Drain developer tank      Thoroughly flush tank and developer rack w
13.  room only if the dust particles that  could be stirred up from the vacuum  exhaust have enough time to settle   e g   overnight        Then clean the countertops  floors   and film feed tray    Contain dust from any new construction   in or near the darkroom       Avoid drywall dust accumulation in  vent pipes during construction or  long term dust shadow image  problems may result       Thoroughly clean vents before  access is limited       Avoid installing any new equipment   such as a roomlight processor  in an  area still under construction    Minimize storage in the darkroom    Open cases of film and other   corrugated cardboard boxes outside   the darkroom and carry the contents  inside to reduce introducing additional  cardboard fibers into the darkroom    Open new boxes of film slowly and   carefully to reduce cardboard fibers   from becoming airborne    Remove the flap and box top from the   darkroom after placing the film  in its   envelope and box  in the film bin     e Remove the stiffener cardboard from  film boxes and the darkroom    e If the film is packaged in a sealed    foiled lined paper envelope  use a pair   of scissors to cut the envelope to   reduce airborne fibers       Do not remove the film from the  envelope or store loose film in  the box  film bin  or film safe at  any time    Avoid replacing unexposed film that   has been inside a cassette back into   the film supply in the film bin       This could introduce additional  dust into the film bin a
14.  type of film to be tested     NOTE  This procedure will take   from 2 to 4 hours to perform  Do not  attempt to perform this test unless both  developers will be tested within that time  frame on the same day  This test can best  be performed using components of the  complete KODAK System  i e   processors  and cassettes     Items Needed   1  A calibrated Sensitometer   Densitometer  similar to KODAK Process  Control Sensitometer Densitometer   that accurately reads above 4 0 0 D    optical density      2  Calibrated Digital Thermometer   3  Stopwatch   4  A box of KODAK MIN R L Film     5  Freshly mixed KODAK RP X OMAT  Developer and Replenisher or KODAK  X OMAT EXI Developer Replenisher and  a bottle of KODAK RP X OMAT  Developer Starter     6  Freshly mixed other manufacturer s  developer replenisher and a bottle of  that manufacturer s starter     Setup  Exposure  and  Processing the Film    Information to complete steps 1 4 can  be found in Service Bulletin No  30   revised December 1999   available by  calling Health Imaging Technical Support  at 1 800 336 4722  716 724 9362      1  Fill the processor with KODAK RP  X OMAT Developer or KODAK X OMAT  EXI Developer Replenisher  and  Replenisher  adding Kodak s  recommended amount of starter for  the mammography film to be used     2  Using the calibrated thermometer   verify that the developer temperature is  set as per Kodak recommendations     3  Set the developer and fixer replenish   ment rates to Kodak s recommendat
15. 00 MIN R 2000 Standard 150 3 60 RP 3 80 EXI   gt 4 0  MIN R 2000 MIN R 2190 Standard 190 3 60 RP 3 80 EXI   gt 4 0    1 Relative speed determined from matched density radiographs of a mammography phantom   KODAK MIN R M Film and MIN R Screen arbitrarily assigned a relative speed of 100   2 Contrast   Measured as the average gradient between densities 0 25 and 2 00 above gross fog     Keys for Properly Processing  KODAK MIN R L Film    e KODAK MIN R L Film should be  processed using a standard processing  cycle  developer immersion time and  developer temperature should be as  specified by the processor manufac   turer as equivalent to a standard cycle      Note that decreased  not increased    contrast will result if extended cycle  processing is employed       The table below reflects standard  cycle processing parameters for  KODAK X OMAT Processors  for other  brands of processors  check with the  manufacturer to ensure use of a  standard cycle       The Troubleshooting Tools section of  this guide contains information on  using a time in solution  TIS  test  tool to ensure the proper processing  of KODAK MIN R L Film  page 20      KODAK X OMAT ann MIN R Processor s     M35  M35A  M35A M  M20  and  MIN R Mammography Processor    M7  and M74     M7     M7A     M7B  270 RA   and 3000 RA  XML 300    M6B  M6 N  M6A N  M6AW  460 RA   480 RA  and 5000 RA    M8    e The processing environment must be   properly initiated and the processor   quality control program established  ac
16. Compare MD  DD  and B   F results from  the processing control chart to the  information below to determine possible  causes and corrective actions     COMMENTS CORRECTIVE ACTIONS  MD  4 severe    Check for kinked or air locked developer replenishment line  DD  J severe    Check for improperly mixed  overdiluted  developer replenisher  B F   gt     Check amount of water added     Check developer replenishment rate     Check that rate set as recommended in Service Bulletin No  30     Check chemical usage     Check for change in film volume since rate was set and adjust accordingly  raise      Communicate changes in film volume to processor service firm     If EXI Developer is used  Contrast will increase just before dropping severely  B F  T   Call service to clean the processor     Drain developer tank     Thoroughly flush tank and developer rack with water     Change the developer filter      Use splash guards    Remove and insert the fixer rack carefully    Process film with edges in contact with film feed tray guides to avoid film jam  B F   gt     Always use a floating lid on top of the developer replenisher  inside the replenisher holding tank    Check developer replenishment rate     Call service to check processor  B F   gt    Check developer replenishment rate     Check that rate set as recommended in Service Bulletin No  30      Check chemical usage     Check for change in film volume since rate was set and adjust accordingly     Communicate changes in film volume to
17. No  30  Eastman Kodak Company  December 1999     wo      Recommended Specifications for New Mammography Equipment  Screen Film X Ray Systems  Image Receptors  and Film Processors  Reston  VA  American College of Radiology  1995     10  KODAK MIN R 2000 Film System Mammography Optimization Guide  Kodak Publication No  M3 108  1999     11  KODAK MIN R 2000 Film System Mammography Optimisation Guide  EAMER version   Kodak Publication No  KHI 60  1999     Notes    Notes    www kodak com go mammo    Health Imaging Division  EASTMAN KODAK COMPANY  Rochester  New York 14650    No duplication in whole or part of this publication without eee 6 4722  the written permission of Eastman Kodak Company  1 716 724 9362    KODAK CANADA INC   Toronto  Ontario M6M 1V3  CANADA    HEALTH IMAGING    Kodak  Min R  and X Omat are trademarks of Eastman Kodak Company  A B ETTER VI EW OF L IFE     Printed on recycled paper containing 10  post consumer waste fiber  using soybean based inks     M3 255 Printed in U S A  10 00    Eastman Kodak Company  2000 CAT No  179 6739    
18. a mask from an  exposed and processed 35 x 43 cm  film  mount the films sequentially  and  label each image with the measured  optical density     3  Place the masked phantom images on  the illuminated viewbox in a darkened  room and view  Identify the image which  shows the greatest amount of informa   tion  details   The optical density of the  preferred image may indicate the best  technique on the specific piece of  mammographic x ray equipment  In  studying the images from your kVp and  optical density series  notice that chang   ing the kVp resulted in very little  change to the radiographic contrast of  the images  while changing the optical  density resulted in the perception of  significant contrast changes  Contrary to  widely held beliefs  kVp has a minor  influence on radiographic contrast when  using a combination of molybdenum and  rhodium for anodes and filters in  mammography  however optical density  has a major influence on contrast    Radiographic contrast is a combination  of subject and film contrast  and is  defined as the difference in optical  density  or film blackening  between  areas of interest in a radiograph      The preferred kVp and optical density  selected will also be based upon the  light output and viewing conditions  where the assessment took place  If  viewboxes or viewing conditions are  sub optimal  i e   low light output   as  mentioned previously  the contrast and  optical density preferred may not be  consistent with regulatory requ
19. ality assurance  program  the darkroom should be  checked for any significant factors which  may contribute to shadow images  After  addressing any outstanding issues   properly cleaning intensifying screens  used for mammography as discussed  above will generally control shadow  images satisfactorily  In some  circumstances  it may be necessary to  thoroughly troubleshoot the darkroom  for dirt and dust  Consider the following  suggestions which are intended to help  reduce airborne dust  dirt  and lint from  the environment  and make the darkroom  easier to keep clean     12    Perform initially    e Clean screens and cassettes at   least weekly       Clean screens and cassettes more  frequently  e g   daily   if necessary    Ensure that film magazines and   roomlight film handling equipment are   cleaned periodically  if used    Use an ultraviolet  or black  light to   help identify sources of dust and dirt   in the darkroom       Note that all dust particles do not  fluoresce under ultraviolet  illumination       Limit exposure prudently and  observe appropriate safety  precautions for eyes  face  and  exposed skin  do not look directly  at the light       Do not expose film to ultraviolet  light       Use the light again after cleaning     Perform daily    e Damp mop darkroom floors to remove  any dust that settled overnight    e Clean darkroom countertops using a  lint free wipe and appropriate non   residual cleaner    e Clean the processor film feed tray last      Moi
20. aphic film that is  high in contrast  such as KODAK MIN R L  Film  the window for achieving a properly  exposed mammographic image is smaller   i e   it has reduced exposure latitude and  care must be taken in mammographic  technique  In addition  when film with   a different emulsion number is opened  for clinical use  it may then be necessary  to make a slight adjustment in the  density control setting in order to  achieve the proper optical density on all  clinical images     Technique Charts    It will be necessary to generate technique  charts for each piece of equipment   Technique charts  generally used with  automatic exposure control where the kVp  is selected by the technologist operator   should indicate how the kVp is altered  from that used on the mammographic  phantom as the breast thickness and  density changes  Slightly lower kVp should  be used for breasts compressed thinner  than the phantom  slightly higher kVp  should be used for breasts compressed  thicker than the phantom  Technique  charts should also include reference to    different target and filter combinations if  used for different breast thicknesses  and or density  In all cases  the kVp  selected for any particular exposure  should allow that exposure to be  completed without exceeding the  maximum desirable length of time     In many circumstances  regulatory  requirements stipulate that technique  charts be posted on each piece of  equipment  Note that even if using a  fully automatic exposure
21. ation of the breast periphery  on par with high contrast in the  breast parenchyma    KODAK MIN R L Film has a novel  double emulsion design  for use with  standard  green emitting  single screen  cassettes  Total system speeds are  shown in the table below    KODAK MIN R L Film is identified by   a single V notch  The primary emulsion  is up when the notch is on the right  hand side at the top edge of the   film  The primary emulsion should   be in contact with the screen    e Kodak has added an orientation layer  to the film to aid in the identification  of the primary emulsion side  the dull  side identifies the primary emulsion  side of the film  the side which is  to be placed against the screen  during imaging         not to scale    e In documentation which refers to the  emulsion side of the film  this should    during imaging  be considered a reference to the primary    emulsion of KODAK MIN R L Film           MIN R L Inverse Square Sensi  primary  emulsion  side up                      not to scale       Density  ie D  oO uo                                  Relative Log E             Screen Film Comparison Chart    KODAK Processing  Screen Cycle    Relative  Speed       Contrast       MIN R L MIN R 2000 Standard 150 3 40 RP 3 60 EXI   gt 4 0  MIN R L MIN R 2190 Standard 190 3 40 RP 3 60 EXI   gt 4 0  MIN R L MIN R Standard 100 3 40 RP 3 60 EXI   gt 4 0  MIN R M MIN R Standard 100 2 95 RP 3 15 EXI 3 9  MIN R 2000 MIN R Standard 100 3 60 RP 3 80 EXI   gt 4 0  MIN R 20
22. ccess to the area  above the ceiling       It may also be helpful to seal the  surface and edges of the ceiling  tiles themselves to decrease elimi   nate airborne particles from the  tiles  check with the ceiling tile  manufacturer for an appropriate  sealant  e g   semi gloss paint or  polyurethane  and with local  authorities to ensure that sealing  the tiles in any manner does not  violate local fire codes       If access to the area above the sus   pended ceiling is required  securely  tape a large continuous sheet of  plastic over the ceiling to contain  the dust  it may easily be removed  when access is necessary       Install a revolving  barrel type  door  to eliminate door slamming dust  sifting down from the ceiling and to  improve traffic flow into and out of  the darkroom       Check the felt installed as part of  revolving doors as it may be an  additional source of lint       If the darkroom is old  i e   con   structed many years ago   check all  materials used in the ceiling for  brittleness and replace  if necessary     e Darkroom walls should be smoothly  finished  painted using a light colored  semi gloss paint  and clean       Walls painted with semi gloss paint  are easier to keep clean       The use of a light colored paint  maximizes safelight illumination and  makes it easier to accomplish all  tasks performed in the darkroom       Avoid flat paint because dust can  cling more easily to walls and they  are difficult to wash       Repaint walls with pe
23. ck for other sources of fog   Sometimes fog arises from exposure to  light other than from safelights   Sources to consider are light leaks  around doors  cracks in walls  or cracks  where ceiling partitions join walls   Occasionally  automated processors will  vibrate sufficiently to cause small  openings between the light seals and  the adjoining partition or the  processor cover may not be tightly  closed  Also  afterglow from some  types of fluorescent lights can cause  fogging  Glowing lights or dials on  equipment located in the darkroom can  fog film     Grid Lines    Reciprocating or moving grids are  commonly used for screen film  mammography  The visualization of grid  lines or of the surface texture on top of  the grid device may be more pronounced  because of the improved image detail  visibility and high contrast of KODAK  MIN R L Film  Refer to the table below for  possible causes and corrective actions         longer than maximum time recommended by equipment  manufacturer     Use higher kVp to shorten exposure time     AIl exposure times should meet minimum and  maximum time specifications       High kVp used  resulting in short exposure times   shorter than minimum time recommended by equipment  manufacturer  e g   0 5 seconds     Use lower kVp to lengthen exposure time     AIl exposure times should meet minimum and maximum  time specifications       Mechanical problem with grid     Grid did not move or moved too slowly    Check electrical connection  Make su
24. concern in mammographic  imaging due to the use of single screen     Minus density artifacts may be caused by  handling or processing  i e   processor  digs   The most frequently occurring  minus density artifact  however  is  termed shadow images  Shadow images  are caused by dirt and dust  superimposed inside the cassette  between the intensifying screen and film  emulsion  Shadow images appear as  small  random  minus density spots on  the film  and are frequently confused  with pick off or processor digs  It is  estimated that shadow images may  account for as much as 95 percent of  minus density artifacts seen on  mammographic images     Controlling shadow images is   dependent upon    e Thoroughly cleaning the surface of  intensifying screens and cassette  interiors as recommended    e Periodically cleaning the interiors of  film magazines and roomlight film  handling devices  such as the KODAK  X OMAT Miniloader 2000    e Proper darkroom hygiene     The high contrast of KODAK   MIN R L Film  however  may increase the  visualization of minus density artifacts   necessitating even more care in  maintaining a clean darkroom  environment and clean intensifying  screens and cassettes     Cleaning Mammographic  Intensifying Screens    Maintaining clean mammography  intensifying screens and cassette  interiors will reduce the occurrence of  minus density artifacts caused by dust or  dirt  Intensifying screens and cassettes  should always be cleaned following the  recommendati
25. cording to regulatory guidelines       The addition of three ounces per  U S  gallon  3 8 liters   25 milli   liters per liter  of KODAK RP X OMAT  Developer Starter is required to  achieve appropriate image quality    Contact your local Kodak representa    tive or contact Kodak   s Health Imaging   Technical Support at 1 800 336 4722    716 724 9362  to obtain the follow    ing service bulletins containing impor    tant processing information for KODAK   MIN R L Film via the Health Imaging   FaxBack System       Service Bulletin No  30  December  1999 revision   Processing  Recommendations for KODAK X OMAT  Processors  publication No  N 923   FaxBack document No  800210      Conversion Instructions and  Processing Recommendations for  KODAK MIN R L Film    DEVELOPER IMMERSION TIME  APPROXIMATE     33 seconds    27 seconds    27 seconds    25 seconds    21 seconds    3 KODAK X OMAT M7 and M7A Processors with 140 second total processing time  4 KODAK X OMAT M7 and M7A Processors with 122 second total processing time    DEVELOPER TEMPERATURE    92  F  33 5  C   92  F  33 5  C     94  F  34 4  C     95  F  35  C     96  F  35 6  C     Sensitometric Variability    MD  Mid Density  Speed     Note that only the combinations of MD   DD  and B   F presented in this guide  typically occur due to the decreased  sensitivity of KODAK MIN R L Film to  processing fluctuations  When MD  DD   and B   F differ greatly from expected  values and or indicate an out of control  processing environ
26. ction to minimize costs and  environmental impact  and fixer reuse  have become more prevalent  In some  processing situations  KODAK MIN R L  Film may exhibit dye stain  Dye stain may  be recognized by a general pinkish or  lavender tint in a D min area of the film     Dye stain may be more pronounced in   the following situations    e A combination of electrolytic silver  recovery with reduced fixer  replenishment rates    e The use of fixers exhibiting a high pH  and or high hardener level   Note  It is  normal for the pH of to increases with  seasoning  increasing the fixer  replenishment rate may help control  the pH      It may be possible to reduce dye stain by    e Checking that the fixer has been  properly mixed    e Increasing the fixer replenishment rate    e Verifying that the film is being  processed in the appropriate cycle    e Checking the temperature of the fixer   solution in the processor  and   increasing to approximately 5  F  3  C    above the developer temperature    if possible    Checking the temperature of the wash   water in the processor  and increasing    if possible   Refer to Service Bulletin   No  30  revised December 1999  for   general processor information     If using electrolytic silver recovery    increasing the fixer replenishment rate   and lowering the fixer pH by adding   acetic acid   Contact Health Imaging   Technical Support at 1 800 336 4722    716 724 9362  for additional   information     e Considering use of a different method 
27. d for mammography in a  darkened room and view  Identify the  image which shows the greatest amount  of information  details   The kVp and  optical density of the preferred image  may indicate the best technique for the  phantom on the specific piece of  mammographic x ray equipment used     8    Contrast  amp  Optical Density    Contrast can be determined in several  ways using the information on a sensito   metric strip  This information is usually  plotted as contrast optical density  curve  A contrast optical density curve is  also known as a gamma plot     At the maximum of the contrast optical  density curve  the film is the most  responsive to changes in exposure    In terms of optimum film contrast  the  selected optical density would corre   spond to this point  Taking into  consideration dose  length of exposure     Contrast Optical Density Curve    Gamma                 RP      EXI                T T 1  2 0 3 0 4 0    Density    Contrast optical density curves can be  plotted manually  by using computer  programs  or by using an automatic  scanning densitometer with the capabil   ity of calculating the gamma  or con   trast  anywhere on the characteristic  curve  Such densitometers provide  gamma value at 0 10 intervals  The char   acteristic curve and contrast optical  density curve  gamma plot  of a selected  emulsion of KODAK MIN R L Film  curves  were generated for reference purposes  using inverse square x ray sensitometry    processed according to recommenda 
28. dequate replenishment rates are  important not only for image quality but  also for proper drying and film  transport   Drying artifacts visible with  transmitted light and other drying  problems can usually be eliminated or  significantly improved when these areas  are corrected     Note that drying or surface patterns  that are visible only on reflected light  are not artifacts  they are considered  normal and acceptable  When the film  surface is closely examined in reflected  light  the primary emulsion side of  KODAK MIN R L Film   slight drying  patterns close to the film edges may be  visible when the film has been properly  dried using the lowest setting as  recommended     To troubleshoot drying problems  review  the following causes and implement the  solutions  as necessary     Causes     e Depleted  oxidized  or poor quality  processing chemicals    e Poor squeegee action at the wash rack  exit or entrance to squeegee    e Clogged  missing  or out of position  dryer air tubes  non uniform or  excessive airflow in dryer    e Inadequate processor venting    e Excessively high dryer temperature    e Very cold wash water     Solutions     e Replace processing chemicals with  fresh solutions       Check replenishment rates and  adjust according to the film  manufacturer s recommendations       Use processing chemicals  recommended by the film  manufacturer    e Adjust or replace the wash rack  exit squeegee    e Replace clean air tube s     e Check processor ventilatio
29. e   no hypo retention in the secondary   emulsion     The KODAK Hypo Estimator  the test   tool used to estimate the amount of   retained fixer  or hypo  on processed  radiographs  contains four colored  sections  1 4  which indicate the  estimated grams of thiosulfate ion per   square meter of film  0 01  0 02  0 05    and 0 12  respectively    The KODAK Hypo Estimator is a generic   test tool for use with all types of   photographic papers  films  and  processed radiographs    The color of the test spot on KODAK   MIN R L Film may differ in color from   the patches printed on the estimator    making it difficult to judge which  color patch most closely matches the  test spot       A yellowish  greenish  or brownish  tone may result    To minimize the difficulty in making a   visible or subjective comparison  use a   spot reading densitometer to make   optical density measurements       Make the first measurement after  laying the clear central portion of  the estimator in its plastic sleeve  over the damp spot on the film     16       Make at least two additional  measurements through the color  patches on the estimator  the plastic  sleeve in which the estimator is  enclosed  and a clear portion of  the film       The optical density reading which  comes closest to the reading taken  from the test spot indicates the  best match    e The test spot should not be darker  than the third patch on the estimator   indicating approximately 0 05 grams of  retained fixer per square 
30. e  avoided  as undesired sensitometric  changes may occur     Any film used as cleanup film should be  discarded after one use to avoid  contaminating the developer solution  with fixer and redepositing lint or dirt  back onto the rollers in the processor   Contact your local Kodak representative  for additional information about KODAK  Roller Transport Cleanup Film 4955     20    Time in Solution Test Tool    Films used for medical imaging should be  processed in the processing cycle or  cycles recommended for the particular  film type     The temperature of the developer  solution and the length of time a film  spends in the developer solution are  important considerations in determining  whether the film will be properly  processed  optimal optical density and  contrast   The length of time the film  spends in the developer is known as  developer time  development time  or  developer immersion time  It is defined  as the amount of time from the leading  edge of the film into the developer to the  leading edge of the film into the fixer     A time in solution  TIS  test tool  which  consists of a strip of clear blue film base  from which all film emulsion components  have been chemically removed and two  white tape strips  may be used to verify  the correct developer immersion time as  specified by film and processor  manufacturers  Note that motor speeds  will vary slightly from processor to  processor of the same type  the TIS tool  should be used to check for times  sign
31. e latitude decreases  If the  optical density selected as preferred is  positioned at the point of maximum  contrast on the characteristic curve   exposure latitude is further decreased  In  other words  the tolerance in the amount  of exposure required to produce a desired  density range decreases     Optimizing Exposure  continued     The exposure algorithms of older   x ray equipment when using the  automatic exposure control  AEC  system  may not be able to respond fast enough  so that the proper optical density is  obtained when a high contrast screen   film combination is used and or the  optical density preferred is at the point  of maximum contrast  Achieving the  desired optical density consistently is  important for consistent image quality   If the optical density varies from image  to image  the viewer generally perceives  variable radiographic contrast  Modern  equipment  when properly  is usually  capable of consistently achieving the    desired optical density on clinical images     Practical Considerations    The goal for technologists  operators  for all mammographic imaging should  be to achieve a consistent optical  density to avoid the misperception of  variable image quality  changes in  radiographic contrast      Assume for a moment that processing is  totally invariant and that all mammo   graphic equipment in the facility has  been calibrated using a specific emulsion  of film  e g   emulsion 104  When the  phantom is exposed using the following   e au
32. e same   orientation as processed       Artifacts that are parallel to each  other on a film pair occurred in the  processor  artifacts may be parallel  or perpendicular to film travel        Artifacts that are perpendicular to  each other on a film pair occurred  during exposure from the x ray unit    Careful analysis of all of the films will   indicate whether the primary emulsion   up or down gives the best overall  processing results    The protocol should subsequently be   posted in the darkroom so all films are   processed consistently       AIl clinical images  sensitometric  strips  and phantom images  primary emulsion up or all primary  emulsion down    Films should be processed on both   sides of the processor to prolong   roller life    Quality control films  phantom images   and sensitometric strips  should also   be processed in a specific location  with respect to the film feed tray    right side or left side  with the edge   of the film butted against the guide of   the film feed tray     Processing Nonuniformity  with KODAK MIN R L Film    Processing nonuniformity  characterized  by uneven optical densities on  radiographs  may occur when KODAK  MIN R L Film is processed primary  emulsion side up in the following  processors  KODAK X OMAT M35A M   M35 M  M35  M35A  MIN R Mammog   raphy Processor  270 RA  and 3000 RA  Processors  The nonuniformity occurs due  to the accumulation of processing by   products adjacent to the film emulsion  inside the developer rac
33. ed with  a small amount of the KODAK MIN R  Screen Cleaner     3  Dry the cover  It may be advisable to  use a camel   s hair brush or other soft  lint free brush to remove any dust  accumulated in the corners of the  cassette cover and along the hinge  Such  dust may not be easily reached with the  dampened cloth     4  Moisten a lint free wipe with a small  amount of KODAK MIN R Screen Cleaner  and gently rub the wipe across the screen   e Avoid the use of abrasive wipes such as  surgical gauze pads   e Avoid excessive pressure or rubbing on  the screen   e Avoid pouring the screen cleaner  directly onto the screens or cassettes   e Use the minimum amount of screen  cleaner needed to moisten the wipe   Excess screen cleaner will not improve  screen cleanliness  will prolong the  drying period  and may stain the screen   Clean around any labels on the screen   e g   those used to individually  number the cassettes    DO NOT WIPE THE SCREEN DRY  Allow  the screen surface to air dry so that  the full effects of the screen cleaner  may be realized   A solution of 70  isopropyl alcohol  may be used occasionally to remove  stubborn dirt or dust   After using 70  isopropyl alcohol   KODAK MIN R Screen Cleaner must be  used following the procedure outlined  above  The use of 70  isopropyl  alcohol should be limited to  occasional use only     5  Stand the cassettes on edge to dry    e Allow the screens and cassettes to  completely air dry before returning  them to service    e Load
34. eling paint    Replace any open shelving       No open shelving of any kind should  be used  especially above darkroom  countertops    Install cabinets with doors to reduce   areas that may accumulate dirt and   dust       Use cabinets that reach all the way  to the ceiling for upper storage     13    Minus Density Artifacts  Shadow Images   Pick Off Processor Digs   continued     Processor Digs Pick Off Marks    Appearance    Pick off marks processor digs appear as  very small minus density spots on the  film where the emulsion has been  removed down to the film base  Often  the piece of the emulsion that is  removed may be deposited on the film  near the trailing edge of the minus den   sity mark or may be deposited randomly  on the film  Pick off marks are most  apparent on single screen images  Pick   off marks are most visible in transmitted  light  Unlike shadow images  pick off  marks can be seen in reflected light     Pick Off Marks    t Direction of Travel    14    Causes    e Rough dirty rollers   e Varying transport speed   e Inactive processing chemicals   e Inadequate film emulsion formulation    Remedies    e Do the periodic maintenance  procedures as outlined in the  Service Manual    e Mix new processing solutions    e Check the surface finish of the knurled  rollers for excessively sharp or rough  areas  If necessary  install new rollers    e Check the replenishment rates for the  developer and fixer solutions     Red Grease Pencil Test to  Determine Source o
35. ence and Regulation    The optimal kVp range and optical  density depend upon several factors   including the preference of the  radiologist s  for contrast and optical  density  the capabilities of the mammo   graphic x ray equipment  the light  output of the viewboxes and control of  extraneous light  etc     It is also important to check existing  local and national regulations   guidelines as to whether any direction is  given regarding optimal exposure for  mammography  For example  a minimum  mammographic phantom background  optical density  0 D   may be suggested   such as at least 1 50  and more  important  minimum optical density for  the glandular tissue on mammograms of  dense breast tissue may be required   Radiologist reviewers of clinical films for  the American College of Radiology  Mammography Accreditation Program  consider glandular tissue with an optical  density lower than 1 0 under exposed   It is important that all clinical and  phantom images meet local regulations     Preparation    In order to optimize exposure techniques   it is exceedingly important that a number  of technical factors be understood and  balanced  A medical physicist may be  able to direct this assessment in  consultation with the radiologist s   Note  that a physics survey  including kVp  accuracy and AEC  automatic exposure  control  reproducibility tests  should first  be performed and the results judged  satisfactory before doing any of the  additional procedures discussed in thi
36. f Minus   Density Artifacts    Tools required    e Film with minus density artifacts   e Red grease pencil   e Lint free wipe  used for screen  cleaning     Procedure    1  On the primary emulsion side of the  film  use the red grease pencil color   in all the minus density spots that are  of concern    2  Take the wipe and gently clean away  the red grease pencil marks from all the  spots colored     Analysis    1  The minus density spots that are  cleaned of red grease pencil marks are  shadow images caused by dust dirt    2  The minus density spots that DO NOT  clean and remain red are caused by  processor digs pick off     A Method for Evaluating Film Performance  in Other Manufacturers    Developer Solutions    The United States Food and Drug  Administration  FDA  Mammography  Quality Standards Act  MQSA  Final  Regulations  effective April 28  1999   states that processing be capable of  developing films in a manner equivalent  to the minimum requirements specified  by the film manufacturer  It is likely that  compliance guidance from the FDA will  recommend that facilities have  documentation from the film or chemical  manufacturer to support this     To determine if another manufacturer s  developer is within Kodak   s acceptable  range for KODAK Mammography Films  you  will need to have access to a processor  using fresh KODAK RP X OMAT Developer  and Replenisher or KODAK X OMAT EXI  Developer with the appropriate amount of  KODAK RP X OMAT Developer Starter for  the
37. film from the suspect  box in half by using the cardboard as  a guide     5  Place the film primary emulsion side  up in the cover of the opened cassette  with the film on the left side and the cut  edge toward the left edge of the  cassette  use the lead pencil to mark the  corner S for suspect     21    Troubleshooting Tools  continued     6  Make sure the film edges in the center  of the cassette are directly adjacent to  one another and not overlapping   before closing the cassette     7  Place the cassette with the two film  halves in the bucky  image receptor  of  the mammography x ray unit     8  Place the phantom on top of the grid  in the standard location used for  mammography quality control testing     9  Position the photocell beneath the  center of the phantom  standard  location   assuming the phantom  exposure is always made using   the phototimer     10  Select the same technique factors  usually employed when imaging the  phantom  same kVp  etc       11  Make the exposure and process the  two film halves immediately in the same  manner  e g   primary emulsion side up  and on the right side of the processor    NOTE  Daylight handling system should  not be used to unload the cassette  Film  halves should be manually fed through  the processor     12  Use a densitometer to take two  optical density readings in the center of  the phantom  just to the right and left of  the cut edges  one on the normal and  one on the suspect film      13  Calculate the density d
38. g the center of the phantom  and average them  This will eliminate  errors due to film or processing artifacts     12  The data should be recorded on an  appropriate form  see page 24  and  should be recorded in the order in which  the cassettes were run  Accurate record  keeping is vital to the interpretation of  the test results     Data Interpretation and Analysis  Using the measured densities from the  three films exposed in the control cas   sette  calculate the standard deviation  of the control film optical densities  If  the standard deviation exceeds 0 05  the  variability of the x ray exposures or film  processing is excessive and the screen  speed uniformity test cannot be carried  out adequately under these conditions   Corrective action should be taken to  reduce this variability before assessing  screen speed uniformity  If the standard  deviation of the control films does not  exceed 0 05  then determine the maxi   mum and minimum optical densities  from all cassettes  The difference  between the maximum and minimum  optical densities of a single size and  type should not exceed 0 30  Corrective  action is necessary for any cassette   screen combination that does not fall  within this range     Suggested Performance Criteria and  Corrective Action   Any individual cassette screen s  within  a given speed group of one size or type  that does not meet the above criteria  should be checked to try to determine  the cause of the problem  One obvious  item to check i
39. haracteristic  H amp D  curve    e Optical densities of the normal and  suspect film halves from the split  phantom test    e Phantom optical density and technical  factors used       Conversion Instructions and Processing Recommendations for KODAK MIN R L Film in KODAK MIN R and KODAK X OMAT Processors  Models 460 RA  480 RA  270 RA  3000 RA  5000 RA  Multiloader 300  M6B  M6A N  M6A W       Film Artifact Diagnostics Guide for KODAK X OMAT Automatic Film Processors  Kodak Publication No  1C0948  Eastman Kodak Company  1996   Haus AG  Film Processing Systems and Quality Control  In  Gould RG and Boone JM  eds  A Categorical Course in Physics  Technology Update and Quality Improvement of Diagnostic X Ray Equipment  Oak Brook  IL     Haus AG  Jaskulski SM  The Basics of Film Processing in Medical Imaging  Madison  WI  Medical Physics Publishing  1997    Haus AG  Jaskulski SM  Richards A  Technical and Clinical Considerations for High Contrast Screen Film Mammography  Eastman Kodak Company  1997   Introduction to Medical Radiographic Imaging  Kodak Publication No  M1 18  1993    Mammography Quality Control Manual for Radiologists  Medical Physicists and Technologists  Reston  VA  American College of Radiology  1999   Processing Recommendations for KODAK MIN R and KODAK X OMAT Processors  Models M35  M35A  M35A M  M35 M  M43  M43A  Clinic 1  M7B  M7B E  270 RA  3000 RA  180 LP  180 LPS  Multiloader 300  M6A N  M6AW     M6B  M8  M6RA  460 RA  480 RA  and 5000 RA  Service Bulletin 
40. ifference by  subtracting the optical density value of  the suspect film from the optical density  value of the normal film  If the density  difference is a negative value and the  suspect film is darker than the clinical  film  the suspect film is faster  If the  density difference is a positive value and  the suspect film is lighter than the  clinical film  the suspect film is slower   According to the American College of  Radiology  ACR  in Recommended  Specifications for New Mammography  Equipment  June 1995  pages 31 33      22    e A density difference of 0 30 between  any two films of the same type from  the same manufacturer  exposed and  processed together  is a reasonable  maximum to be expected from  manufacturing variability for films of  roughly the same age and storage  conditions    e If the difference between the two film  densities exceeds 0 30 at a density of  approximately 1 25  as specified by the  test   then the film supplier should be  contacted to determine the source of  the problem     Note that a difference of 0 30 at a  density of approximately 1 25 may  translate into a bigger difference for  clinical films exposed at a greater optical  density  For example  high contrast  mammography films  such as KODAK  MIN R L Film  are frequently exposed at  an optical density between 1 50 to 1 70  in order to maximize contrast  The  density difference at this optical  density level may be greater due to the  increased contrast     Procedure for Uniformity of
41. ificantly more or less than the times  listed by the processor manufacturer     The procedure is as follows     1  Remove the lid of the processor  A  small magnet placed near the  microswitch may be needed to allow the  processor to operate without the lid   in place     2  Locate the approximately    in   6 3  mm  gap between the entrance detector  crossover and the guide shoe  Also locate  the gap between the developer to fix  crossover assembly and the guide shoe     3  Feed the test tool into the processor   placing the tool either along the film  feed tray guide or in the center of the  film feed tray  with the bottom of the  taped T feeding first     4  With a stopwatch in hand  get ready  to begin timing when the black line  drawn across the tape passes through  the space in the entrance detector  crossover     5  Begin timing when the cross of the T  reaches the same space  Either the top or  bottom of the strip of tape may be used     6  Stop timing when the same part of the  cross of the T reaches the space in the  developer to fix crossover assembly     7  Repeat the timing sequence three  times and take an average to determine  the developer time     8  The total processing time should also  be checked and verified by measuring  the amount of time either a  mammography film  18 x 24 cm  with 24  cm of film travel  or a general radiology  film  35 x 43 cm  with 35 or 43 cm of  film travel  depending on the width of  the processor  takes to transport through  the p
42. ing film into a cassette with wet  screens may damage the film and lead  to staining of the screens     6  Inspect the screen and cassette cover   for any stray particles of dust    e An ultraviolet  or black  light is helpful  in determining if screens and cassettes  are clean    e Limit exposure to the ultraviolet light  and observe appropriate safety  precautions for eyes  face  and exposed  skin  do not look directly at the light     11    Minus Density Artifacts  Shadow Images   Pick Off Processor Digs   continued     7  Reload the cassette with film    e If using improved KODAK MIN R 2  Cassettes  wait at least 5 minutes after  loading before using    e If using KODAK MIN R Cassettes or an  older version of KODAK MIN R 2  Cassettes  wait at least 15 minutes  before using    NOTE  Improved KODAK MIN R 2 Cassettes   have a rounded latch lever  The older   version of the cassette has an angular  latch lever     8  Reclean the countertop  if the  cleaning procedure was done in the  darkroom     9  Between regular cleanings  a camel s   hair brush may be used to remove dust  particles from the cassette and screen  surface  Care should be taken when using  any brush to protect the screen surface  from scratches  which will degrade  image quality     10  Each intensifying screen and cassette  should be marked with a unique number  to facilitate locating a specific cassette  suspected of requiring cleaning or  inspection     Troubleshooting the  Darkroom    When initiating a qu
43. ion is needed     24    If standard deviation of control cassette densities is less than 0 05 AND density range exceeds 0 3  then       Information Needed by Kodak to  Facilitate Troubleshooting    When contacting Kodak for assistance  with image quality concerns  having as  much of the following information as  possible will facilitate all  troubleshooting efforts    e Make and model of processor    e Mix of film types processed    e Volume of films  number of sheets per  typical eight hour day     e Brand of chemicals  developer  and fixer     e Type of chemicals  e g   premixed   automixed  or facility mixed        Replenishment rates  milliliters of  developer and fixer  fully defined in  terms of number and size of film  and  length of film travel  e g   30 milliliters  of developer and 30 milliliters of fixer  per one 18 x 24 cm KODAK MIN R L  Film  single film feeding  24 cm of  film travel      References    m    M7B  M35A M  and M35 M  Eastman Kodak Company  1999     w    Radiological Society of North America  1996  49 66     E a ar al    e Dates last and previous preventive  maintenance  PM  of processor  performed    e Developer temperature    e Current and previous emulsion  numbers    e Date the change or problem  was noticed    e Description of change  what was  observed and who noticed it     e Description of any unusual  circumstances  e g   films processed  only 2 days per week  etc      e Optical densities of the 21 steps from  the sensitometric strip    e C
44. ions  for processor type and film     4  Measure the drop time for an  18 x 24 cm film and verify that the  transport speed is set accurately     5  Set the sensitometer for green  sensitivity  For KODAK MIN R L FILM   be sure that the primary emulsion side  faces the light exposing source     6  Expose three films     7  Process the film on the right   hand side of the feed tray   primary emulsion side down     Figure 1   8  Label all films  Indicate  developer used  date  time   and any other relevant  information  Drain the  developer solution  change  the developer filter  and  thoroughly rinse the tank   rack and developer to fixer  crossover assembly  Repeat  steps 1 8 using the other  manufacturer s developer   Use the developer  manufacturer s  recommended starter and  replenishment volumes     Evaluating the Films    1  Read the films using the calibrated   scanning densitometer  Note the speed  value  step closest to 1 0 above gross   fog  and the average gradient     2  Average the speed and average  gradient for the three films processed  through KODAK Developer  Repeat for the  films processed through the other  manufacturer s developer     3  Compare the speed and average  gradient for the film processed in the  other manufacturer s chemistry to that  in KODAK RP X OMAT Developer and  Replenisher or KODAK X OMAT EXI  Developer Replenisher  Typically accept   able results are within the ellipse shown  in Figure 1  This ellipse is defined by   e a maximum range of
45. irements  for image quality  images may be  underexposed      NOTES    e The above procedure will help establish  the optimal kVp for the specific  phantom which will simulate one  specific breast thickness and type only   The facility should establish an optimal  kVp range for different breast  thicknesses and densities  The use of  one kVp for all breast types is not  recommended    e Each mammographic x ray unit should  be evaluated    e Consultation with the medical physicist  or governmental agency regarding  local national dose regulations is  suggested before making any permanent  exposure technique changes     Exposure Time  amp  Dose    Dose  and the length of the exposure in  terms of minimizing motion on clinical  films must also be considered with  respect to the selected kVp and optical  density  All factors should then be used  to create technique charts  refer to  page 10 of this guide for additional  information     After the optimal kVp and optical density  have been determined by performing two  procedures described above  assess  whether local national dose regulations  can be met when that kVp and optical  density are used for clinical images  If  dose regulations cannot be met ora  lower dose is desired  the use of a  slightly higher kVp and or of a slightly  lower optical density may be used   Alternatively  a faster intensifying screen   i e   KODAK MIN R 2190 Screen  may   be used     Furthermore  it is necessary to assess the  length of exposure time 
46. ith water      Change the developer filter   Use splash guards   Remove and insert the fixer rack carefully   Process film with edges in contact with film feed tray guides to avoid  film jam    Measure developer temperature using accurate or calibrated  thermometer   Verify developer temperature set as recommended in Service Bulletin  No  30 for model of processor   Lower thermostat to correct temperature   Check thermostat and recirculation pump for malfunction   Check water flow and temperature    Check temperature of incoming water   Verify water temperature set as recommended in Service  Bulletin No  30 for model of processor   Install mixing valve to regulate water temperature       Check immersion time using time in solution test tool    Check Service Bulletin No  30 for recommended rates  and  adjust    Check Service Bulletin No  30 for the amount recommended  for the model of processor in use   Add the correct amount of KODAK RP X OMAT Developer  Starter    Check Service Bulletin No  30 for the amount recommended  for the model of processor in use  Add the correct amount of KODAK RP X OMAT Developer Starter    Measure developer temperature using accurate or calibrated  thermometer   Verify developer temperature set as recommended in Service  Bulletin No  30 for model of processor   Raise thermostat to correct temperature   Check thermostat and recirculation pump for malfunction  Check water flow and temperature    Check temperature of incoming water   Verify water tem
47. k  Besides  nonuniformity  the characteristic curve  plotted from the information from a  sensitometric strip may also reflect  nonuniformity at approximately 2 20 to  3 00 optical densities     Due to nonuniformity  processing  KODAK MIN R L Film primary emulsion  side down is recommended in the  processors noted above  This allows  developer solution to reach the primary  emulsion more efficiently  reducing the  occurrence of nonuniform development   Additionally  it is necessary to install  smooth guide shoes in the developer  rack to reduce guide shoe marks  accentuated by processing film with   the emulsion side down  Refer to  Conversion Instructions and Processing  Recommendations for KODAK MIN R L Film  for information on guideshoe kits for  specific processors  Contact your local  Kodak representative  visit Kodak s  Health Imaging Internet site   www kodak com go mammo   or contact  Health Imaging Technical Support at  1 800 336 4722  716 724 9362      The Darkroom Fog Test    The darkroom fog  safelight  test should  be performed at least semi annually   Several factors can affect test results     Expose the phantom test film to the  appropriate optical density   The film should be exposed to an    optical density of between 1 2 and 1 5     This density range is appropriate  regardless of the optical density   used for other phantom tests   Adhering to this density range is  especially important for high   contrast mammography films such as  KODAK MIN R L Film    U
48. ment  e g   reaching  or exceeding  0 15 of the operating  levels for MD and DD   always generate  another sensitometric strip  making  sure that   e The primary emulsion side of the film  was toward the light source of the  sensitometer     TRENDS IN GRAPH PossIBLE CAUSES    MD  J Severe under replenishment    DD  J f  B  F     or  gt        Major contamination of  developer with fixer    Severe oxidation of developer    Exhausted developer    Expired film    Film stored above the    recommended temperature    and relative humidity    DD  Density Difference  Contrast     B  F  Base Plus Fog    The temperature of the developer  solution in the processor has stabilized  before exposing and processing the  sensitometric strip    The settings on the sensitometer were  properly set  i e   single sided  exposure  green  recommended DIP  switches if the sensitometer is  exposure adjustable  etc      The film used was taken from the box  set aside for processor quality control   The delay between exposing and  processing the film is as normally  occurs from day to day   Note that  immediate processing after exposure of  the sensitometric strip by the  sensitometer is recommended         T  Increasing    J  Decreasing      No Change    e The same densitometer was used to  measure MD  DD  and B   F values    e The calibration of the densitometer  is correct    e If an automatic scanning densitometer  is used  the correct channel was  used and the programming has not  been changed     
49. meter of film     Other important points include    e Perform the fixer retention test in  normal lighting conditions  avoid  bright light or sunlight     e Perform only on a freshly processed  film  or one that has been processed  no more than two weeks previously       The shelf life of the test chemicals is   approximately two years  assuming the   chemicals have been properly stored  with the cap tightly closed  within the  required environmental temperature  range  etc    Process another film and repeat the   test if failing results are obtained    If failing results are obtained after   repeating the test  consider replacing   the test solution first if the solution is  older than two years  or its age is  unknown   if the conditions under  which it has been stored are unknown   or if it is possible that the cap on the  bottle may not always have been  tightly closed       When purchasing a new bottle of  test solution  write the date it is  first opened on the bottle    If failing results are obtained after   repeating the test  and the test   chemicals are not in question  contact  the processor service company or your  local Kodak representative     Dye Stain   Following the film manufacturer s  recommendations for fixer in terms of  type  replenishment rates  etc   is just    as important as for developer since  mammography films may be more difficult  to clear and image stability storage  requirements are longer     Additionally  systematic replenishment  rate redu
50. n    e Reduce dryer temperature to minimum  required to produce dry films    e Provide wash water within the  recommended temperature range     All actions suggested under    Solutions     should be done by the most appropriate  person  that is  one who has received  the proper training to perform the  necessary task     17    Other Technical Considerations    Evaluating Processors  for Uniformity    Processors in which MIN R L Mammog   raphy Film will be processed should be  evaluated initially to determine whether  films should be processed primary  emulsion side up or primary emulsion  side down which orientation provides the  best uniformity and the fewest artifacts  before processing clinical and quality  control films     Film must be processed primary emulsion  side down in the X OMAT M 35AM   X OMAT M 35M and MIN R Mammography  Processors   Multiloaders may be  excluded as film orientation during  processing is fixed   The following  procedure may be used to make this  evaluation    e Expose films to an optical density of  1 10 to 1 50  using the same cassette  and a 1 inch thick uniform sheet of  acrylic       The exposure time used should be at  least 0 5 seconds or longer       The cassette should be known to  have good screen film contact    e Lay the film on the film feed tray in   the darkroom  and mark an arrow  1    on the corner of the film with a lead   pencil immediately before processing  to indicate the direction of film travel       It is also helpful to
51. nd into the  film supply    Switch from an identification  ID    printer or camera located in the dark    room to a camera located outside or   to an ID capable mammography unit   to eliminate fibers from the thin paper   cards inserted into the printer or   camera in the darkroom    Eliminate non essential items that   contribute to dust and paper fiber       Newspapers  magazines  facial  tissue  notebooks  paper pads  etc    Require all darkroom personnel and   technologists to wear lint free   clothing or to wear smocks or lab   coats over clothing    Avoid hanging or storing articles of   clothing in the darkroom    Avoid bringing materials that   shed fibers into the darkroom    e g   knitting     Avoid washing machines and dryers in   the darkroom    Thoroughly clean the entire darkroom   if it is necessary to clean screens   and cassettes more frequently than   once a day     Evaluate darkroom features     e Darkroom ceilings should preferably be  solid  if possible  smoothly finished  and painted       Suspended ceilings  especially in  darkrooms with a single door  entrance  may be a significant  source of dirt and dust       Slamming the darkroom door can  cause suspended ceiling tiles to  move and dust to sift down from  the tiles and dead space above  the ceiling       It may be helpful to seal tiles to the  suspension frame to prevent ceiling  tile movement when the darkroom  door is closed  slammed   this  should be done only if there is no  periodic need for a
52. ons of the manufacturer   Generally  for effective cleaning  prolonged screen life  only the screen  cleaning solution  e g   KODAK MIN R    Screen Cleaner  and any other products  specified by the manufacturer of the  screens and cassettes should be used     Mammography screens should be cleaned  at least once a week  or more frequently  as needed to reduce shadow images  caused by dust and dirt   If it is  necessary to clean the screens more  often than once a day  attention to the  darkroom environment itself may be  necessary  refer to page 12 of this guide  for information on troubleshooting   the darkroom      For best results  the recommendations  below should be followed when cleaning  KODAK MIN R and MIN R 2 Cassettes in  which KODAK MIN R  MIN R 2000  or  MIN R 2190 Screens are mounted  New  screens and cassettes must be cleaned  prior to first use     1  Chose a clean location to clean   screens and cassettes    e If working on a countertop in the  darkroom used for processing  mammography film  wipe the outside  of the cassettes and clean the  countertop with a damp cloth prior to  cleaning the screens    e Manage cassettes so they are empty at  the time of cleaning  end of the day is  usually best     e Avoid replacing film previously loaded  into cassettes back into the film box or  film bin  since this could introduce  dust into the film supply     2  Clean the inside plastic cover of the  cassette  tube side panel  using a lint   free wipe that has been dampen
53. perature set as recommended in Service  Bulletin No  30 for model of processor   Install mixing valve to regulate water temperature    Check immersion time using time in solution test tool    Sensitometric Variability  continued    TRENDS IN GRAPH POSSIBLE CAUSES COMMENTS CORRECTIVE ACTIONS   MD  J Developer temperature lower MD only affected  4    Measure developer temperature using accurate or calibrated  DD   gt  _ than recommended thermometer   B R     within 5  F  3  C      Verify developer temperature set as recommended in Service    Bulletin No  30 for model of processor             Raise thermostat to correct temperature     Check thermostat and recirculation pump for malfunction     Check water flow and temperature  Water temperature lower    Check temperature of incoming water  han recommended    Verify water temperature set as recommended in Service  if developer temperature Bulletin No  30 for model of processor  below set point     Install mixing valve to regulate water temperature  Developing time shorter    Check immersion time using time in solution test tool  han recommended  MD      gt  Too much starter added DD only affected  T    Check Service Bulletin No  30 for the amount recommended  DD  T i  gt 50 percent  for the model of processor in use  B  F   gt    Add the correct amount of KODAK RP X OMAT Developer Starter  Developer replenishment rate    Check Service Bulletin No  30 for recommended rates and adjust  oo low KODAK EXI Developer  MD      gt  Exhau
54. posure  in the darkroom  under appropriate safelighting  load a  sheet of film into a clean mammographic  cassette  If using an improved KODAK  MIN R 2 Cassette  wait at least 5  minutes prior to each exposure  wait   at least 15 minutes prior to each  exposure if using a KODAK MIN R  Cassette or an older version of the  KODAK MIN R 2 Cassette     3  Prepare to make the first exposure    e Place the cassette into the  moving grid cassette holder of the  mammographic unit    e Position the phantom on top of the grid    e Lower the compression device until it  contacts the top of the phantom    e Position the photoreceptor so it is  centered underneath the phantom    e The maximum exposure time of any  image should not exceed some pre   determined value  such as 2 seconds    NOTE  The target optical density should   be as typically used clinically  Each image   in the following series should be within   0 05 of the target optical density     4  Expose a series of phantom images  varying the kVp by one kVp increments   from the lowest kVp to the highest  i e    24 30 kVp      5  Prepare a mask for the images by  exposing a 35 x 43 cm film to light  pro   cessing the film and cutting the mask  and films to accommodate the images   NOTE  Mask to the edge of wax insert   Tape the phantom images sequentially   Label each image with the kVp used  so  it is either visible or hidden  as desired      6  Place the masked phantom images   one or more series  on an illuminated  viewbox use
55. re grid device is stable    Make sure nothing is interfering with smooth  passage of grid during each exposure    Check with the equipment manufacturer or service       Older grid technology       Check with the equipment manufacturer about  the availability of modifications to the existing  grid or new grid technology          19    Troubleshooting Tools    Some simple but very effective tools are  available for troubleshooting the  processor and KODAK MIN R L Film  System  They include    e Cleanup film    e Time in solution test tool    e Emulsion number log    e Split phantom test    e Screen speed     Cleanup Film    KODAK Roller Transport Cleanup Film  4955  CAT No  122 4310  is a specialized  film designed to be used in conjunction  with the processing environment  Each  14 x 17 in   35 x 43 cm  sheet of film  features a non light sensitive coating on  both sides of the film base     Cleanup film picks up lint  dirt  and  other deposits and helps carry them out  of the processor  For best results  one or  two sheets should be processed     Cleanup film is particularly useful to  control a processing artifact called delay  streaks  Cleanup film may be used in all  processors  except those with area  replenishment  which cannot sense this  clear based film     Fogged or expired single  or double   emulsion film that has not previously  been processed may also be used as  cleanup film  Note that mixing films  from different manufacturers in   the same environment should b
56. rocessor  Time from the leading  edge of the film into the processor to  the trailing edge exiting the processor   Repeat the timing sequence three times  and take an average to determine the  total processing time     Correct developer immersion time at a  specified developer temperature is  particularly important when processing  mammography films  which are generally  processed in standard or in extended  processing cycles     As previously discussed  KODAK MIN R L  Film should be processed only in a  standard processing cycle     The TIS test tool and instructions are  available at a nominal charge by calling  KODAK Parts Services at 1 800 431 7278   716 724 7278   part No  5B6497   instructions are also available separately   part No  8B7020      Emulsion Number Log    Film manufacturers designate each box  of x ray film with a multiple digit  emulsion number  This number provides  important information such as which  particular emulsion batch was used as  well as which roll it came from  the  specific part  slit  of the roll  and  which variation of emulsion was used   variation code      Keeping track of the emulsion number of  the film used for processor quality  control is generally done in every  processor quality control program  It is  also extremely advantageous to keep  track of the complete emulsion numbers  of all film used clinically  especially for  mammography  Doing so allows film  manufacturers to make comparisons in  speed  contrast  D max  etc   between
57. s  guide  The processor used for  mammography must also be in control    i e   verify that the processor and  processing environments are functioning  according to the manufacturers     recommendations  This information is  contained in Service Bulletin No  30   December 1999 revision   Processing  Recommendations for KODAK X OMAT  Processors  publication No  N 923   FaxBack document No  800210  and in  Conversion Instructions and Processing  Recommendations for KODAK MIN R L Film   Both documents are available by visiting  the Health Imaging Internet site   www kodak com go mammo   or by  calling Kodak   s Health Imaging Technical  Support at 1 800 336 4722   716 724 9362  to access the Health  Imaging FaxBack System     Viewing Conditions   The viewbox is an integral part of the   mammographic imaging chain  It is   essential for good viewing practices to  be implemented to ensure optimal  interpretation  The following factors  should be considered    e Evaluate the lightboxes for cleanliness   light output  and consistency  clean  the interiors and surfaces of viewboxes   replace all lightbulbs  etc  NOTE  All  lightbulbs should be replaced whenever  it is necessary to change one lightbulb    e The intensity for all viewing panels  should be matched  both within and  between viewboxes     e It is recommended that the technologist    radiographer viewboxes match the intensity   of interpretation viewboxes to aid in   ensuring adequate exposure    e Use masking materials and
58. s misidentification of a  cassette with the type of screen s  it  contains  Also  if screens of the same  speed are contained with cassettes of  different manufacturers  it is possible  that variations in attenuation of the  cassettes may cause significant varia   tions in film density  Should no identifi   able cause for image density variation  be determined  it is reasonable to  replace the cassette screen s  that  results in optical densities outside the  0 30 0 D  range     23    Uniformity of Screen Speed       Screen type        Film type           Focal spot           Small Cassettes       Size     cm          Film Emulsion             Cassette ID  Artifacts     mAs    Density       Control Cassette        Processor used        kVp setting           AEC density control       Large Cassettes                   Size  cm  Film Emulsion     Cassette ID  Artifacts  mAs   Density          Control Cassette                       Mean of control cassette densities                Standard deviation of  control cassette densities    Mean of control cassette densities          Other Cassettes     Standard deviation of  control cassette densities          Other Cassettes                                                                             Mean Density                Minimum Density    Mean Density          Maximum Density    Minimum Density          Density Range    Maximum Density                   Density Range             Comments             Action Limit   corrective act
59. se the appropriate safelight filter   A KODAK GBX 2 Safelight Filter  or  equivalent  is the correct safelight filter  for all KODAK Mammography Films   Change the safelight filter  periodically    Safelight filters can fade and crack  with age  Heat from incandescent  bulbs can crack safelight filters   over time     Possible Causes of Grid Lines    Low kVp used  resulting in long exposure times    Install the safelight filter correctly   Filters should be installed so that the  identification printing can be read  when looking at the lamp  If the filter  orientation is reversed  excessive heat  buildup inside the lamp housing may  cause the dye layer to crack and thus  leak unsafe light    Use the appropriate wattage bulb   Oversized bulbs can damage the  safelight filter from excessive heat    A 7 5 watt incandescent bulb is  recommended for all KODAK  Mammography Films    The distance of the safelight from the  work surface and the direction of the  light to the work surface can affect  results  Also consider directing the  safelight toward the ceiling    Use the appropriate number of  safelights for the size of your  darkroom    In large rooms with white ceilings   place no more than one lamp for every  64 square feet  6 square meters  of  ceiling area  You can use a number of  safelight fixtures  with the correct  filter and bulb  if they are at the  proper distance from the work surface  and are placed at least 8 feet  2 5  meters  apart     Corrective Actions    e Che
60. sted fixer  if film cloudy    Check fixer replenishment rate  DD     a or exhibits dye stain    Check that rate set as recommended in Service Bulletin No  30  B  F     or  gt     pink or lavender color      Check fixer replenishment line for kinks     Check that fixer was properly mixed  Fixer temperature lower B F  T    Measure fixer temperature using accurate or calibrated  han recommended  if film thermometer  cloudy or exhibits dye stain    Verify fixer temperature set as recommended in Service   pink or lavender color   Bulletin No  30 for model of processor      Raise thermostat to correct temperature  Developer replenishment rate    Check Service Bulletin No  30 for recommended rates and adjust     oo high KODAK EXT Developer    Optimizing Exposure    KODAK MIN R L Film is a novel  mammographic film  extending the  exposure latitude to allow the  simultaneous visualization of peripheral  structures with the parenchymal detail   In order to achieve the full benefits of  this technology  and get the best images  on the viewbox  it is important to  optimize the exposure characteristics   The objective of this section is to review  procedures that can help in the selection  of the optimal optical density and  preferred kVp  This can lead to the  establishment of technique charts which  are an important part of reliably  achieving the optimal image quality in  mammography     There is no current optimal standard for a  single level of contrast and optical density     Prefer
61. sten a lint free wipe with a small   amount of KODAK MIN R Screen  Cleaner used for intensifying  screens       If the feed tray needs to be cleaned  more frequently  use the ultraviolet  light to check the cleanliness of the  incoming air     Perform occasionally    e Wipe darkroom walls  the fronts of  cabinets  safelights  vent surfaces   exposed pipes  etc   with a damp lint   free wipe periodically to eliminate  clinging dust    e Vacuum or use a damp lint free wipe to  remove dirt and dust from the inside of  the film bin occasionally  e g   once a  year  or as needed     e If passboxes are used  check the insides  for paint particles  metal flakes  etc      Clean periodically     Institute good practices    e Maintain relative humidity between  30 and 50 percent to minimize the  attraction of particles onto films and  intensifying screens    e Do not routinely place mammography  cassettes on the floor    e Dirt may be carried into the darkroom  along with the cassettes    e Filter air coming into the darkroom    either at the system level  or by   placing furnace type filters in the air  vents feeding the darkroom       Change air filters according to the  frequency recommendations of the  manufacturer       Electrostatic air cleaning devices  may be helpful if a significant dust  problem exists in the darkroom    Locate ventilation louvers in darkroom   doors near eye level  not near the   floor  and vacuum or wipe periodically       Use a vacuum cleaner in the dark  
62. sures     8  After the first half of the cassettes  have been exposed  unload the film from  the control  first  cassette and process  the film  Reload  wait the appropriate  time for air purge  and expose the  control cassette to prepare a second  control film  Continue testing the  cassettes from the second half of the  testing group  as noted in Step 7  Reload  and expose the control cassette after all  test cassettes have been exposed  This  will result in three control films from the  one cassette  from the beginning   middle  and end of the test sequence  If  possible  additional measurements of the  control cassette should be made  spread  throughout the measurements of the test  cassettes  This will greatly improve the  reliability of the results     9  Process the exposed films in the film  processor routinely used for  mammographic imaging  Process the  films on the same side of the processor     10  If more than one size or type of  cassette is used for mammographic  imaging  repeat Steps 1 through 10 for  each  The same lot of film should be  used across all cassette sizes if the  results are to be grouped together as the  same cassette type     11  Measure the optical density at the  center of the phantom on each of the  processed images and record optical  density on the data form   Identify the  center by diagonals    drawn    from corner  to corner across the phantom image   If  possible  take several density  measurements from the area immediately  surroundin
63. t  Record on the data  form the specifics of the image receptor  being evaluated and the emulsion  number of the film used for the test   This cassette will be called the   control cassette     3  Select the mammographic imaging  mode and kVp most commonly used for  clinical examinations  e g   contact  grid   28 kVp  Mo Filter  using an automatic  exposure control  AEC   if available and  used clinically     4  Position the phantom on the cassette  holder assembly  bucky  at a location  over the image receptor that would be  occupied clinically  breast   Place the  compression device in contact with   the phantom     5  Using the imaging mode and kVp  selected in Step 3  determine the    density control setting  if photo timed   or mAs  if manual  required to obtain an  image optical density  0 D   greater than  1 20  as measured in the center of the  phantom image  Record this technique  on the data form     6  Load all of the mammographic imaging  cassettes to be evaluated with film from  the same box as used in Step 2     7  Sequentially expose each cassette in  the first half of the test group under the  conditions determined in Step 5  It is  useful to use lead numbers to identify  the images  If AEC exposures are used   make sure that the lead numbers are not  placed in the vicinity of the AEC  detectors  Record the actual mAs for  each exposure on the data form  It is  important to ensure that the position of  the phantom and AEC detector is  unchanged during all expo
64. tomatic exposure control   e the designated cassette   e emulsion 104 loaded into the cassette      the photocell  photoreceptor  in the  ideal designated location   e the preferred kVp  and  e the density control is set at    0    the  preferred optical density  e g   1 60   is obtained on the processed film   and the exposure time is just under  2 seconds     In this scenario  all clinical imaging  should be done by varying the kVp as  recommended in the technique chart  while the density control setting is  left at 0     It has been common practice to move  the density control setting  i e    changing from    0    to a plus or minus    10    value   especially when using older  equipment whose phototiming  algorithms could not properly  compensate for reciprocity law failure as  the length of the exposure increased    The reciprocity law states that the  optical density of a radiograph will be  the same if the milliampere seconds   mAs  is constant  regardless of mA  this  law  however  applies to direct exposure  radiography and fails when intensifying  screens are used  necessitating  additional exposure in order to achieve  the proper optical density   If using  modern equipment  significant  improvements to phototiming algorithms  have been implemented  Therefore  it is  no longer necessary to change the  density control setting  and doing so  only results in changes in optical density  and the perception of variable image  quality and contrast     When using a mammogr
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Rapport d`activité 2007  Pamp Catalogus  Samsung YP-T4H Наръчник за потребителя  Sanyo PDG-DHT8000L data projector  Hampton Bay BSG8212-2 Installation Guide  FAXユニット 取扱説明書  Oricom 310 Baby Monitor User Manual  Eclat version 1 07 2011  3-Heights™ PDF to Image Converter Shell, User Manual  NOTICE DE MONTAGE DU KARTING KT8 Réf. T6570F72 / T6570F73    Copyright © All rights reserved. 
   Failed to retrieve file