1.     ON     If the  battery voltage is below 6 8Volt  display indicates     LOP     Turn  unit off and replace the battery     Connect the ECG leads of the Defibrillator Monitor to the ECG  output jack  ECG snap banana adapter is provided     Select the ECG source of the Monitor Defibrillator to ECG  leads  Set the MODE switch to Cardio  The display flashes    000  and ECG waveform   60 BPM will appear on the  monitor  Place the paddles on the paddle contacts  ECG  waveform should appear when source selected to paddle     Calibration    DELTA 2000 is factory calibrated and thoroughly tested and  found to meet the Netech s manufacturing specifications before  being shipped  Calibration measurement standards are traceable  to the National Institute of Standards and Technology  NIST    Any attempt to tamper with the calibration seal will void the  warranty     
2.    Energy  Range  Low  0  99 99 Joules  High  0      700 Joules  Resolution    Low Range  0 01 Joule  High Range  1 Joule  Accuracy  Low Range     2   Rdg       1Joule  High Range     2  Rdg     2 Joules    Cardioversion Delay      250 Milli Seconds from Peak R  Accuracy      1  of full scale     1 ms    Waveform Simulation     Internal  ECG   60 BPM  VTAC  VFIB  amp  NSR   90 BPM  External  Waveform input through Input Jack     Discharge Waveform   Real time  Voltage attenuation 1000  1  Expanded time  1 200 repeated every 8 seconds     Display  4 Digit Large LCD disply    Power  9 V alkaline battery or AC adapter   Weight  3 lb  1 6 Kg    Dimensions  10 4 x 7 2 x 4 0   26 4x18 2x10 2cm     Optional Adaptors      Paddle adapter sets are available for  most AEDs  Call Netech for part       TEST AND CALIBRATION  Introduction    Before attempting to test the unit please observe the necessary  safety precautions for safe handling of the unit and also to avoid  potential shock hazard  The operator must be familiar with the  unit and follow the safety precautions listed in the  Safety  precautions  in the General Information section of this manual     Performance Check    This check is recommended upon the initial receipt of the unit   Before proceeding with the test inspect the instrument  thoroughly to ensure that no visual damage caused during  transportation shipping     The DELTA 2000 is shipped with a 9 Volt Alkaline battery   Turn the unit on by flipping the Power ON OFF to
3. INSTRUCTION MANUAL    DELTA 2000       Defibrillator Analyzer    2 m   NETECH  60 Bethpage Drive  Hicksville  NY 11801    www GoNetech com    TABLE OF CONTENTS    General Information 2  Product Description 3  Operating Instructions 5  Theory of Operation 9  Specifications 11  Test and Calibration  12    Safety Considerations    Defibrillators deliver high voltage shocks to a patient in order to    stop a heart attack  The Defibrillator discharge pulse is potentially  dangerous  Before attempting to use Delta2000 get thoroughly  familiar with the defibrillator under test and follow the  manufacturer s safety protocols        GENERAL INFORMATON    Introduction    Delta 2000 1s intended to operate only by qualified  technical personnel  Before attempting to use DELTA 2000  please observe the safety procedures for safe handling of  the unit and also to avoid potential shock hazard  The  operator must be familiar with the unit and follow the  safety precautions listed below  Refer to the Defibrillator  manufactures operating manual for additional safety  considerations     Defibrillators deliver high voltage shocks to a patient in order  to stop heart attack  Defibrillator pulse is potentially  dangerous     Safety Considerations    If any fluid 1s spilled on the unit or the unit 15 dropped  do  not use until thoroughly tested     Do not touch the metal surface of the paddle     Place DELTA 2000 onto a flat firm surface before  discharging the defibrillator  Apply both paddles fir
4. arged energy is greater than 99 99  Joules  then the display flashes HI continuously  In this case  change the range selector to HI and discharge the Defibrillator  a second time  In Hi range if the discharge energy exceeds 700  the display flashes    HELP    continuously     CAUTION  DO NOT USE ELECTRODE PASTE ON PADDLE  CONTACTS     DO NOT DISCHARGE DEFIBRILLATOR IN THE PRESENCE  OF FLAMMABLE ANESTHETICS     Waveform Recording    While the Defibrillator 15 discharged onto DELTA 2000 the  captured waveform will be continuously output through the  ECG leads  paddle contacts  and high level output jack  marked    ECG ARTH OUT IN  located on the left side of  the rear panel of DELTA 2000  This waveform is  amplitude scaled to provide maximum output and time  scaled to provide a time expansion of 1 200  The same  waveform is repeated at 8 seconds intervals  If the  Monitor Recorder associated with the Defibrillator can  sense through the paddles  the waveform from DELTA  2000 will be automatically captured by the Monitor  Recorder  as long as the Defibrillator paddles are in contact  with the DELTA circular plates  If there 1s no paddle  sensing capability  older units   then ECG leads must be  connected to the ECG input leads on the Monitor Recorder     Real time Scope Output    The attenuated Defibrillator discharge waveform can be  Observed on an oscilloscope  Connect an oscilloscope to the  phone jack marked Scope Output  on the right side of the  rear panel of DELTA 2000  S
5. ed and fed to a 13 Bit  12 Bit   sign  bit  A D converter interfaced with an 8 bit Microcontroller   Philips P80C52  The waveform is sensed by the A D and    sampled at a rate of 1000 times per second and saved in the  memory  The captured waveform is time scaled to 1 200  and dumped into a 12 bit D A converter  The stored  waveform can be observed through the paddle and the ECG  Leads  In ENERGY mode the micro controller computes  the energy content of the captured waveform and displays  the result on the display  During cardioversion the micro  controller produces an ECG waveform   60 BPM through  the D A converter and wait for the discharge waveform   The time between the rising edge of the discharge  waveform and the peak R wave of the simulated ECG 1s  measured and displays in milli seconds on the display  If  the discharge occurs on the negative slope a negative sign  also displayed     The waveform is also stored in the memory and when the  MODE switch is changed to energy the micro controller  calculated the energy and displays     The accuracy of the DELTA 2000 is maintained by a  precision calibrated voltage source  a crystal oscillator  running at 11 0592 MHz and a set of 0 1  load resistors  and attenuation resistors  The unit 1s factory calibrated with  N  I  S  T traceable standards  In order to ensure the  specified accuracy  factory recalibration 1s recommended  on an annual basis     SPECIFICATIONS    Internal Test load  50 Ohms     1   Pulse Width  1  65 ms 
6. et the sweep to Ims div and  vertical gain to 1 Volt  Discharge the Defibrillator onto the  paddle contacts on DELTA 2000 and observe the  waveform  The amplitude reduction is 1000 1 Volt   The  output is 1V 1000V       Cardioversion Testing    Set the MODE switch to CARDIO  The display flashes    000  and a simulated ECG waveform   60 BPM is  output through the ECG leads and the paddle contacts   Connect the ECG leads of the Defibrillator monitor  Set the  Defibrillator to synchronize mode  Select the source of the  ECG to leads  Discharge the Defibrillator into the paddle    contacts  DELTA 2000 measures the time delay between  the peak of the R wave of the ECG and the rising edge of  the Defibrillator discharge waveform  The reading is  displayed in Milliseconds  Measured time is positive if the  leading edge of the discharge waveform occurs after the  peak of the R wave  on the negative slope of the QRS of  the ECG   If the discharge occurs on the positive slope of  QRS      sign will be displayed  indicating negative  reading  If the reading exceeds the range     250   the  display flashes    OOPS        Testing Semi Fully Automatic Defibrillators    Refer to the Defibrillator manufactures service manual for  specific device inspection requirements     Observe the safety precautions  For Fully automated  Defibrillators use the correct paddle adapter for the  Defibrillator under test  The paddle contact plate must not  be exposed during discharge     Set the MODE switch o
7. mly  centered on the discharge plates of DELTA 2000  If the  paddles are not firm arcing may occur  Arcing can cause  injury to the operator and damage the unit     Do not touch the contact plates when defibrillator paddles  are placed on the plates     Do not use any gel  pads or conductive material on the  paddles when testing on the DELTA 2000     Use only the recommended paddle adapters for testing the  Automated External Defibrillators     PRODUCT DESCRIPTION    Introduction    DELTA 200 is a compact  state of the art defibrillator  analyzer designed for testing the performance of  Defibrillators including the AEDs  Automated external  Defibrillators      Features    The four performance characteristics of a defibrillator are  discharge energy  synchronized mode operation  automated  external defibrillator  AED  function and ECG monitoring   The DELTA 2000 is specifically intended to test all four  parameters  It measures the output energy and cardio   version delay time in synchronized mode     The internal 50 Ohms load simulates the human body   The DELTA 2000 simulates several test waveforms     The simulated waveforms  ECG  VTAC  VFIB and NSR   Normal Sinus Rhythm   90BPM  are output through the  paddles for testing the Automated External Defibrillators   AED   Additional arrhythmia waveforms can be input into  DELTA 2000 through the phone jack input for interactive  training purposes     Discharge waveform is stored in the memory  The time  expanded discharge wavefo
8. n the DELTA 2000 to AED  Select  the desired waveform for the advisory shock feature of the  Defibrillator  Select the ENERGY range switch to the  desired range  Connect the paddle adapter to the circular  contact plate  for fully automated Defibrillator  or place the  paddle firmly on to the paddle contact  semi automatic  Defibrillator  and discharge  The DELTA 2000 displays the  measured energy     A unique feature of the DELTA 2000 is the capability to  input any desired waveform required to activate the  automated Defibrillator into the unit by connecting an  arrhythmia simulator   Netech s MiniSim 1000   to the   ECG ARTH OUT IN jack  Set the MODE switch to  ENERGY  Select the manufactures recommended  waveform from the arrhythmia simulator     THEORY OF OPERATION    Introduction    The DELTA 2000 is a precision micro controller based  device that performs tests on automated and semi  automated Defibrillators  This instrument utilizes modern  analog to digital converter techniques to capture the  Defibrillator discharge waveform     Switches Output    Micro    Converter  controller    p  5  LE  gt   a rm   o O  g co  a E D  B tA c  c o 2    O  O 5 J  m  z  Un       DELTA 2000 Circuit Block Diagram    Description    DELTA 2000 circuitry is divided into eight functional  blocks for the purpose of simplicity  The input load and  attenuator block consists of the precision 50 Ohm high watt  resistor and attenuation resistors  The discharged waveform  is attenuated and buffer
9. orm is output to through the paddle and ECG leads   DELTA 2000 is now ready to measure the cardio version  delay time     In ENERGY Mode the display flashes  0000  or  00 007  depending on the  ENERGY Range selector switch in  HIGH or LOW  The unit is now ready to measure the  energy discharged into the circular metal plates marked  STERNUM and APEX     The AED MODE selector simplifies the testing of the  AEDs  In this MODE the DELTA 2000 simulates VFIB    NSR or VTAC depending on the waveform selector     Energy Measurement    Defibrillator discharge voltages are potentially dangerous   Before attempting to test the Defibrillator please observe  the necessary safety procedures for the safe handling of  the unit and also to avoid potential shock hazard     Place DELTA 2000 on a flat non skid flat surface close to the  Defibrillator under test  Turn the unit by flipping the ON OFF  switch to ON     Select the MODE switch to ENERGY  middle Position    The display flashes  0000  or  00 00  depending on the position  of the energy range selector     Select the range corresponds to the Defibrillator energy to be  measured  For energy values below 100 select LOW     Place the Defibrillator paddles firmly against the circular plates  on DELTA 2000 and charge the Defibrillator to the desired  energy level  Ensure all the safety precautions  Discharge the  defibrillator  The delivered energy reading will be displayed  until the next discharge test     Note    In LOW range if the disch
10. rm is played back through the  paddles and ECG leads     Real time discharge waveform is output through the output  phone jack for observing on a scope     Product Familiarity    DE    ga  x  XL       TE      arts UC  o ecet  mi  a       2000             E    aa ym         z E  E         E  m e             FERENT    Display   Scope Output Jack  Energy range Selector  On OFF Switch   DC Jack   Defibrillator paddle  Contacts    Mode Selector   ECG High Level Output   Arrhythmia Input Jack  AED waveform selector  Battery Compartment  ECG Leads Output       a    OPERATING INSTRUCTIONS    Setup    Although DELTA 2000 1s simple to operate and requires  minimum setup  the user must read the manual thoroughly  for safe and proper operation     The DELTA 2000 operates with a 9 Volt Alkaline Battery  or optional AC adapter  Netech Part   302   The unit is  shipped with a fresh battery  To replace the battery open  the battery compartment located below the ECG leads in  the rear of the unit  If using the AC adapter  plug the unit  into the Jack on the right side of the unit  Use only the  recommended AC adapter     DELTA 2000 uses a large four digit LCD display and four  slide switches located on both side of the display  Two of  the slide switches  left  are three position and the two   right  are two positions     DELTA 2000 provides three modes of operation  selectable  by the    MODE    switch  When the MODE  switch is in  CARDO position the display flashes   000  and the ECG  wavef
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
DC2060 - Farnell  *49005360* 49005360 KS 1100  P360/460 Memory Scanners  "user manual"  Descargar  VOACAP Propagation Planner User`s Manual  CLUB3D CGAX-57724I 1GB graphics card  Modelo ProFloSXT de co-corriente  Untitled - NRG fitness systems  Guía de Incidencias del Cliente de Firma 5    Copyright © All rights reserved. 