        40-6532/40-6543 Auto-leveling Rotating Laser Service Manual
         Contents
1.                                                                                                                                                                                                                                                                                                                                  Figure 20 Electric Connection Drawing       Prepared By  Tim Wojo Revision  2 042109 Page 15 of 16    a  F  BCL easyPDF  Printer Driver    Click here to purchase a license to remove thisimage         omens al sham  eee fae SE OS ee    LASER LEVELS    4 0 Troubleshooting Guide    40 6532 40 6543 Service Manual                                                                         No  Symptom Cause Corrective Action  Battery case is loose Change battery box    1  control board is defective   Change 1  control board  1 Unit doesn   t turn on  Power switch is defective Change Power Switch  6  circuit board Change 6  circuit board  Remote control does    2 8  control board is broken Change 8  remote control board  not work  Laser point beam does   Hair spring is broken Change hair spring  3  not light Laser source is defective Change laser source  Hair spring is broken Change hair spring  Head does not rotate    4 _ 5  alarm board is defective   Change 5  alarm board  when unit is turned on   1  control board is defective   Change 1  control board  Instrument is good but  indicator light does      5 _   1  control board is defective   Change 1  control board  not c
2.      n ol abh am  i 3   f E          n M   LAREn LEYEL SO PRO    40 6532 40 6543 Auto leveling Rotating  Laser Service Manual                                                                                  Item Description Pages  1 0 Overall Instrument Assembly 2  1 1 Main Housing Assembly 2 3  1 2 Upper Housing Assembly 3 4  1 3 Core Part Assembly 4 5  1 3 1 Main Axis Assembly 5  1 3 2   Top Bracket Assembly 6  1 3 4   Bottom Bracket Assembly 7  1 4 Base assembly 8  2 0 Calibration 8 15  2 1 Horizontal Mode Operation 9  2 1 1 Quantifying Accuracy Error 9  2 1 2   Characterize the type of error 10  2 1 2 1   Oblique Error 11  2 1 2 2   Taper Error 12  2 1 2 3    Wave Error 13   22 Vertical Mode Operation 14   2 2 1   Vial Calibration 14  2 3 Leveling Range Alarm Calibration 14  3 0 Circuit Board Connections 15  4 0 Troubleshooting Guide 16          omens al sham    LAGER teve DRO 40 6532 40 6543 Service Manual    1  Overall Instrument Assembly    This rotating laser is a highly accurate instrument  Out side of a few customer  adjustments  outlined in the owners manual   all adjustments service operations are  internal to the instrument and to be performed only by authorized service personnel   Authorized personnel should adhere to the guidelines described within this service    manual for all repairs and or service work        This service manual specifically discusses two very similar products    e 40 6532   e 40 6543  Physical construction and components of the two product
3.   the instrument is self leveling  where by a major factor of accuracy is how well the leveling  compensator is balanced  Different types of errors require different methods of calibration  These errors are easily  characterized by placing the instrument on a flat surface following the guidelines below  and running the instrument with the    rotating head continuously rotating 360       2 1 1  Quantifying Accuracy Error    Establish the test set up shown in the following illustration and confirm that it meets the following requirements    1  The debugging platform should be horizontal with a error less than 0 1   6      2  The distance L from the center of the debugging clamp to the wall should be L gt 10m    3  The accuracy of the reference marking is no less than 0 05mm m    4  The adjacent light ruler should be put within 0 3m  12     of the instrument  the zero position of the ruler should be    located at the same height as that of the laser line at the emitting hole    Far Target placed on wall at least 10m  33ft  from Test    Stand  where    0    corresponds to them same height as     0    on the target next to the aperature of the instrument     Laser line  Instrument for test  laser spinning 360   rotation              Distance  L   gt 10m  33ft                 Accuracy  A    Difference between laser line          height and reference line height  divided by the       Flat Surface  Test Stand  o distance that the laser is from the wall  L   Near Target placed within 
4.  7 3  by removing 2 retaining screws  7 1     2  Remove the prism base part  7 3  from the main axis assembly  7 7  by loosening Allen Notched Tightening Screw   7 21     3  Remove Hair spring switch over PCB  3  7 5  from Hair spring switch over PCB  4 by de soldering hair springs  not  shown  and removing 4 retaining screws  7 4     4  Remove Bracket  7 20  from top bracket assembly  7 14  by removing 3 retaining screws  7 9    Remove 4  Hairspring Switchover PCB  7 8  from top bracket assembly  7 14  by removing 4 retaining screws  7 1    Remove Switch PCB  7 17  from bottom bracket assembly  7 15  by removing its 2 retaining screws  7 18 and 7 13  respectively     7  Remove top bracket assembly  7 14  from lower bracket assembly  7 15  by removing 4 retaining screws  7 10  and their  respective elastic gaskets  7 19  and 2 retaining screws  7 9  from front semi circle piece  7 16     8  Remove Press board  2  7 11  from top bracket assembly  7 14  by removing 2 retaining screws  7 13  and their  respective flat washers  7 12      9  Remove main axis assembly  7 7  from top bracket assembly by removing 4 retaining screws  7 6      1 3 1 Main Axis Assembly  AP1902          Item   JLT Part   Description Qty  7 7 1 AP1838 M2 x 4 Cross Slot Plate Screw  7 7 2 AP1909 Progressive Motor   7 7 3 AP1910 Main Axis   7 7 4 AP1911 Laser Source Part   7 7 5 AP1491 M2 x 8 Cross Slot Plate Screw                      Rl  e  Re  Ww                             ue     um    1  Remove laser s
5.  direction  X or Y  is necessary for the adjustment   Laser module  Prepared By  Tim Wojo Revision  2 042109 Page 12 of 16    wal            Glick hers          omens al sho    LAGER teve DRO 40 6532 40 6543 Service Manual    2 1 2 3 Wave Error _       ae N  ee  This error occurs when the rotating laser surface is not a plane  but a turn up one as ene S  shown in the figure to the right  The rotating beam projected on the wall is crooked  Du  similar to waviness  which is caused by the unbalance of rotating head  The value A Wave error    checked in the X direction deviates upwards while the one in the Y direction  downwards  or the value A in the Y direction upwards while the one in the X direction downwards  Meanwhile  you can see  the laser point above the head is mobile and its orbit is inclined to be a circle   The error can be removed by adjusting the balance block of the Rotating head as follows   a  Remove the top cover as discussed in section 1 of this document   b  Regulate the adjusting screws on the four balance block as shown in the figure to the right  until the laser point    above the head becomes stable  namely  the laser point projected on the ceiling being a stable one     1  As shown in the figure below  put the table vice on the debugging platform    2  Fix the core part on the adjustment fixture  and Install the adjustment fixture on the table vice  to make the laser spot  towards the wall    3  Turn the core part  to make the laser revolve around the ro
6. 0 3m  12     of  instrument  Note that    0    of the target  corresponds to the height of laser line at the    aperature of the instrument     Place the instrument for test on the test stand  preconfigured from the illustration above  with handle facing the far target    2  Power the laser and function in full rotation mode  head rotates 360      Note the errors in all four quadrants of the instrument as read on the far target  i e  x axis A   0   and B   180    y axis C    90   and 270       e You should end up with 4 numbers    i AQO    ii  B 80          iii  C 90       iv  D 270        4  Since all of the errors are referenced against    0    on the far target  essentially the largest number from the data collected  A      D    the distance should be equal to or less than the published specification for the product  If not  characterize the error    and determine method of calibration as defined by section 4 1 2 of this document        Prepared By  Tim Wojo Revision  2 042109 Page 9 of 16             omens al sho    sasen LEVELS DRC  40 6532 40 6543 Service Manual    2 1 2  Characterize the type of error    Use the following rules to determine what to of accuracy error the instrument for test has  and reference the stated  section to adjust for it  Note that the unit my have a combination of different types of errors to adjust for  in  which case  multiple calibration must be performed             Accuracy is beyond Accuracy meets published           published specificati
7. 15 2     2  Remove Alarm Spring PCB  7 15 17  from Connecting board  7 15 3  by loosening 2 retaining screws  7 15 15   Note  that retaining screws have elastic gaskets on them  7 15 16     3  Remove magnetic Base part  7 15 11  from Bottom bracket  7 15 9  by loosening 4 retaining screws  7 15 13   Note  that retaining screws have elastic gaskets on them  7 15 12     4  Remove Connecting Columns  7 15 14  from Bottom Bracket  7 15 9  by loosening retaining screw  7 15 1   Note that  retaining screws have elastic gaskets on them  7 15 2     5  Remove Shield Ring  7 15 8  from Bottom Bracket  7 15 9  by loosening 3 retaining screw3  7 15 6   Note that retaining    screws have elastic gaskets on them  7 15 7         Prepared By  Tim Wojo Revision  2 042109 Page 7 of 16         omens al sham    LASER LEVELS PRO    1 4 Base Assembly  AP1887    40 6532 40 6543 Service Manual                                                                               Item JLT Part   Description Qty  10 1 AP1930 Connecting Column 2 1  10 2 AP 1843 ST2 2 x 6 5 Cross Slot Plate 2  Tapping Screw  10 3 AP1931 Buzzer 1  10 4 AP1932  2 Switch Part 1  10 5 AP 1933 Base 1  10 6 AP 1493 M2 5 x 8 Cross Slot Plate Screw 1  10 7 AP 1934 Hand Wheel 2 1  10 8 AP1935 Sealed Ring 1  10 9 AP 1936 Screw Bolt 1  10 10 AP1937 Slipper Block 1  10 11 AP 1938 5mm    E    Ring 1                Buzzer  10 3  is epoxied in place Base  10 5  and should not be removed    2  Remove Horizontal Vertical Switch part  10 4  from 
8. Base  10 5  by loosening 2 retaining screws  10 2      Note that    Buzzer  10 3  wires need to be de soldered before switch can be removed     3  Remove connecting column  10 1  from Slipper Block  10 10  by removing e ring  10 11      4  Remove Slipper Block  10 10  from Base  10 4  by removing Hand wheel  10 7     loosening retaining screw  10 6      Hand wheel  10 7  is removed by    5  Remove Screw Bolt  10 9 from Slipper Block  10 10  by rotating it counterclockwise until it comes off     2 0 Calibration    Calibration is a process that is used to correct for accuracy and or functional errors above and beyond those stated    in published specifications     While Manual leveling  Self leveling  and Automatic leveling  motor driven  devices    have different mechanisms that require calibration  there are similarities with optics that is consistent regardless of    the leveling mechanism  This section of the service manual discusses calibrations specific to the 40 6532 40 6543     Each item discussed is shown below     2 1 Horizontal Mode Operation  2 1 1 Quantifying Accuracy Error    2 1 2 Characterize the type of error  2 1 2 1 Oblique Error   2 1 2 2 Taper Error   2 1 2 3 Wave Error    2 2 Vertical Mode Operation  2 2 1 Vial Calibration    2 3 Leveling Range Alarm Calibration       Prepared By  Tim Wojo Revision  2 042109    Page 8 of 16         omens al sho    LAGER L  vete ey 40 6532 40 6543 Service Manual    2 1  Horizontal Mode Operation    In this mode of operation
9. ave been removed the upper housing assembly    4  can be removed  Remove this by loosening the 5 retaining screws  5  located at the base of the unit    e As the upper housing assembly is being removed  note that there are wires that make the electrical connections  between the upper housing and the core module  These wires can break and caution should be taken as the two parts  are being separated    Remove the Core module assembly  7  from the Base Assembly  10  by removing the 6 retaining screws  5  and their 6    elastic gaskets  6     1 2 Upper Housing Assembly  AP1881                                                                                     Item   JLT Part   Description Qty  4 1 AP1891 Top Cover Assembly 1  4 2 AP1892 Window 1  4 3 AP1893 Housing 1  4 4 AP1558 ST2 9 x 9 5 Cross Slot Tapping 4   Screw  4 5 AP1894 8  Remote Control Receiving PCB 1  4 6 AP1843 ST2 2 x 6 5 Cross Slot Plate 4  Tapping Screw  4 7 AP1538 2 5mm Nut 8  4 9 AP1895 1  Key Pad PCB 1  4 10 AP1896 Key Pad Membrane 1  4 11 AP1897 M2 5 x 10 Cross Sunk Screw 4  Prepared By  Tim Wojo Revision  2 042109 Page 3 of 16         omens al sham    sasen teve DRO  40 6532 40 6543 Service Manual    1  The key pad assembly  4 9  can be removed by disconnecting electrical wires from the remote control receiving PCB   4 5   Remove the 4 retaining nuts  4 7  from the retaining screws  4 11  and remove key pad assembly  4 9  from  housing  4 3   The retaining screws can be removed by peeling back the key pad 
10. e  if one adjustment is loosened     A pene ecm   the adjustment to the opposite side of it needs to be   tightened by the same amount   X Y orientation of  housing determines X Y adjustments within the unit   After adjustments have been completed loc tight all 4 adjustments to prevent them from moving through temperature cycling     vibration  normal use  etc        Prepared By  Tim Wojo Revision  2 042109 Page 11 of 16             omens al sham    LAGER teve DRO  40 6532 40 6543 Service Manual    2 1 2 2 Taper Error awe    Taper error results when the prism base module  discussed in section 1 3  item 7 2  does not v       reflect the laser light exactly 90    i e  non perpendicularity between the rotating line and the    rotating axis   The end result is a laser rotating surface that is not a plane  but a tapered one Taper error   as shown in the figure to the right  The value A in the left target shares the same symbol as   the one in the right target during the check  that is  Aj  2 and Ajgo    2  The error is removed by adjusting the optical wedge  of the laser output window to make the rotating laser beam and rotating axis plumb beam  The details are listed as follows     a  Remove the top cover as discussed in section 1 of this document             b  Screw the wedge out of its base with special spanner as shown in the figure Optical wedge  to the right   c  Adjust the laser to the zero position of the target   d  Loc tight adjust after it has been set   Note  Only one
11. he Upper housing assembly as shown in section 1 1    Attach unit to special mounting platform    f  Turn unit on and adjust the height of the laser dot such that it is at the same height on the near and far targets  disregard  reading of vial at the point    g  Now that the laser dot is shooting the same height on both the near and far targets  adjust the 4 adjustment screws  shown    above  until the bubble in the center of the vial     2 3  Alarm Calibration    The alarm function is made up of 3 components  the 2 alarm PCB   s  7 15 17  7 14 13  and the buzzer  10 3   Calibrating it is  a matter of centering the contact ring of 7 14 13 to the spring post of 7 15 17  When the spring post makes contact with the  contact ring of 7 14 13  the buzzer  10 3  sounds  Centering the post between the ring offers equal angular alarming in all    directions  Adjustments are made as follows           a  Remove the Upper housing assembly as shown in section 1 1  Alarm ring    Locate the adjustment screws 7 14 12  reference section 1 3 2   Alarm spring  c  Loosen the adjustment screws and manually center the contact ring of  7 14 13 to the spring post of 7 15 7     d  Tighten the adjustment screws and reassemble the upper housing assembly to    Adjustment screw    the base assembly        Prepared By  Tim Wojo Revision  2 042109          BCL easyPDF  Printer Driver           omens al sho    tasen LEVELS DRC  40 6532 40 6543 Service Manual    3 0 Circuit Board Connection                   
12. membrane  4 10  and removing the 4  retainer nuts  4 8     2  Remove the remote control receiver PCB  4 5  by removing its 4 retaining screws  4 6     3  Remove the top cover   4 1  and window  4 2  by removing the 4 retaining screws located on the under side of the  housing  4 3      1 3 Core Part Assembly  AP1884                                                                                                     Item   JLT Part   Description Qty  7 1 AP1647 M2 x 5 Cross Slot Plate Screw 6  7 2 AP1898 Rotary Head Assembly 1  7 3 AP1899 Prism Base Part 1  7 4 AP1843 ST2 2 x 6 5 Cross Slot Plate 4   Tapping Screw  7 5 AP1900  3 Hairspring Switchover PCB 1  7 6 AP1572 M2 5 x 10 Cross Slot Plate Screw 4  7 7 AP1902 Main Axis Assembly 1  7 8 AP1903  4 Hairspring Switchover PCB 1  7 9 AP1493 M2 5 x 8 Cross Slot Plate Screw 5   7 10 AP1511 M3 x 12 Cross Slot Plate Screw 4   7 11 AP1904 Press Board 2 1   7 12 AP1582 2 5mm Flat Washer 2   7 13 AP1616 M2 5 x 5 Cross Slot Plate Screw 3   7 14 AP1877 Top Bracket Assembly 1   7 15 AP1878 Bottom Bracket Assembly 1   7 16 AP1905 Front Semi Circle Piece 1   7 17 AP1906  6 Sensitive Switch PCB 1   7 18 AP1513 M3 x 8 Cross Slot Plate Screw 1   7 19 AP1883 3mm Elastic Gasket 4   7 20 AP1907 Bracket 1   7 21 AP1908 M3 x 4 Allen Head Tightening 1   Screw  Prepared By  Tim Wojo Revision  2 042109 Page 4 of 16         omens al sham    CARER teve DRO 40 6532 40 6543 Service Manual    1  Remove the rotary head assembly  7 2  from the prism base part 
13. ome on when unit  is turned on   After turn off the  6 instrument  the keypad   1  control board is defective   Change 1  control board  indicator stays on  7 Range scan is Hair spring is broken change hair spring  nonfunctional 1  control board is defective   Change 1  control board  Hair spring is broken change hair spring  8 No alarm indication 5  alarm board and or 2  Change 5  alarm board  2   circuit board is defective circuit board  Unit does not work in f    9   2  board is defective Change 2  board  Vertical g mode does             Prepared By  Tim Wojo    Revision  2 042109    Page 16 of 16         
14. on specification           ada on Section 2 1 2 1  and B  lt  gt  0 but   dD  lt 0 but  Oblique Error oa om        A B 0 C D 0    Aand B  lt 0 but Section 2 1 2 2        Cand D  lt  gt  0 but  A B Taper Error        C D        Aand B  lt  gt  0 and C and D    0 and  Oblique Error        AtB lt  gt 0 C D lt o0    Taper Error    A B lt C D Section 2 1 2 3    Wave Error       Prepared By  Tim Wojo Revision  2 042109 Page 10 of 16     gt    BCL easyPDF  Printer Driver           Click here to purchase a license to remove this image      omens al sham    LAGER teve DRO 40 6532 40 6543 Service Manual    2 1 2 1  Oblique error    This type of error occurs when leveling compensator does not hang straight  as illustrated to the    right   i e  is not properly balanced  When the laser spins  it actually rotates on a plain that is not       level  i e  perfectly horizontal whose rotating axis is not plumb  So value A on the left target does  not has the same sign as that on the right target  like  A   p  2  Aig0   2  A number of factors can  cause this situation to exist   e Normal use     Depending upon how careful the user is with the instrument  finely tuned adjustments can be altered by  very large changes in temperature  vibrations  due to handling and transportation   e Shock     The instrument was dropped and components of the leveling system  gimbal  compensator weight and shaft    screw adjustments  have either been damaged or came out of alignment  If defective components are p
15. ource part  7 7 4  from main axis  7 7 3  by removing 4 retaining screws  7 7 5      2  Remove the progressive motor  7 7 2  from main axis  7 7 3  by removing 3 retaining screws  7 7 1        Prepared By  Tim Wojo Revision  2 042109 Page 5 of 16         omens al sham    LASER LEVELS PRO                                                          1 3 2 Top Bracket Assembly  AP1877  Item JLT Part   Description Qty  7 14 1 AP1912 Adjustment Base 1  7 14 2 AP1913 Gimbal Part 1  7 14 3 AP1914 Top Bracket 1  7 14 4 AP1915 2 5mm Elastic Gasket 3  7 14 5 AP1513 M3 x 8 Cross Slot Plate Screw 3  7 14 6 AP1916 Gimbal Mount 1  7 14 7 AP1883 3mm Elastic Gaskets 3  7 14 8 AP1530 M3 x 25 Cross Slot Plate Screw 3  7 14 9 AP1917 Pendulum Parts il  7 14 10 AP1908 M3 x 4 Allen Head Tightening 1  Screw  7 14 11 AP1918 M2 5 x 25 Cross Slot Plate Screw 4  7 14 12 AP1022 M3 x 6 Cross Slot Plate Screw 2  7 14 13 AP1919  5 Alarm PCB 1  7 14 14 AP1920 Vial Assembly 1                         40 6532 40 6543 Service Manual                1  Remove Vial Assembly  7 7 14  from Base Part  7 14 6  by loosening 4 retaining screws  7 14 11      2  Remove Alarm PCB  7 14 13  by loosening 2 retaining screws  7 14 12   Note that wires running to the alarm PCB will    need to be de soldered before it can be removed    3  Remove Pendulum part  7 14 9  by loosening Allen head tightening screw  then rotating the Pendulum part  7 14 9     counterclockwise to un screw it from the Base part  7 14 6     4  Remove Base par
16. resent  they    will need to be repaired or replaced prior to the calibration procedures defined below being initiated        Depending upon the magnitude of the error   5mm or 0 2      only fine adjustments need to be made  Larger errors   gt 5mm or  0 2     require coarse adjustment to get close  the fine adjustment to bring the unit within specification  Both adjustments are  defined below       Self calibration  aperture in Y  direction    Self calibration  aperture in X  direction    Plastic Rubber    screw       With the instrument on the measuring platform describe in section 2 1 1  remove the battery box and remove off plastic screws   Remove rubber plug to expose the calibration adjustment  Calibration screws are adjusted with a flat head screw driver   Clockwise rotation shifts the weight on the compensator to the opposite side of the adjustment screw causing the laser line to  drop  Conversely  counterclockwise rotation shifts the weight on the compensator to the near side of the adjustment screw  causing the laser line to rise  If the error is too large to calibrate through the self calibration apertures  coarse calibration is    required     Remove housing  reference section 1 1  to expose the 4         inner calibration adjustments  right   Using a Phillips I Fine adjustment    Y screw in X axis    screwdriver  adjust the instrument accuracy by decion        systematically loosening tightening the adjustments in a Fine adjustment screw    alternating fashion  i 
17. s are identical  The main  difference is the laser diode used for each product  40 6543 green vs  40 6532 red                  1 1 Main Housing Assembly  40 6543                                                                                  Item   JLT Part   Description Qty          tie  1 AP 1449 M2 5 x 6 Cross Slot Plate Screw 4 EY oe    2 AP1879   Bracket Plate 1  3 AP1880 Pivot Screw 2  4 AP1881 Upper Housing Assembly 1  5 AP1513 M3 x 8 Cross Slot Plate Screw 11  6 AP1883 3mm Elastic Gasket 6  7 AP1884 Core Part Assembly 1  8 AP1885 Press Board 1 1  9 AP 1886 Pin 1  10 AP1887 Base Assembly 1  11 AP1493 M2 5 x 8 Cross Slot Plate Screw 1  12 AP1888 On Off Switch 1  13 AP 1889 Sealed Ring 1 1  14 AP1890 Battery Housing 1  Prepared By  Tim Wojo Revision  2 042109 Page 2 of 16       BC    omens al sham    CARER teve DRO 40 6532 40 6543 Service Manual    Remove battery housing  14  by loosening the flat head retaining screw    Remove ON OFF switch  12  by first removing retaining screw  11   Once retaining screw  11  has been removed    ON OFF switch  12  and sealing ring 1  13  can be removed    Remove Bracket plate  2  by removing 2 pivot screws  3   The bottom portion of the bracket plate is held in place by the   Press board 1  8   4 retaining screws  1  and retaining pin  9   Loosen the 4 retaining screws  1  to remove press board    8  and slide retaining pin  9  out to remove bracket plate  2    Now that bracket plate  2   battery housing  14   and ON OFF switch  12  h
18. t  7 14 6  and Adjustment Base  7 14 1  from Gimbal part  7 14 2  from loosening 3 retaining screws     7 14 8      Note that retaining screws have elastic gaskets on them  7 14 7     5  Remove Gimbal part  7 14 2  from Top bracket  7 14 3  by loosening 3 retaining screws  7 14 5   Note that retaining    screws have elastic gaskets on them  7 14 4        Prepared By  Tim Wojo    Revision  2 042109    Page 6 of 16         omens al sham    LAGER teve DRO  40 6532 40 6543 Service Manual    1 3 3 Bottom Bracket Assembly  AP1878       Item JLT Part   Description Qty  7 15 1 AP1513 M3 x 8 Cross Slot Plate Screw  7 15 2 AP1883 3mm Elastic Gasket  7 15 3 AP1921 Connecting Board  7 15 4 AP1922 Compression Spring  7 15 5 AP1923 Back Semi circle Press piece  7 15 6 AP1449 M2 5 x 6 Cross Slot Plate Screw  7 15 7 AP1582 2 5mm flat washer  7 15 8 AP 1924 Shield Ring  7 15 9 AP1925 Bottom Bracket   7 15 10 AP 1926 Crankshaft   7 15 11 AP1927 Magnetic Base Part   7 15 12 AP1915 2 5mm Elastic Gasket   7 15 13 AP1616 M2 5 x 5 Cross Slot Plate Screw  7 15 14 AP1928 Connecting Column   7 15 16 AP1516 3mm flat washer   7 15 17 AP1929  7 Alarm Spring PCB                                                        NI NIAI Ble  mele le unju  e N                             1  Remove Connecting board  7 15 3   Compression springs  7 15 4   Back semi circle Press Piece  7 15 5   and Crank  shaft  7 15 10  by loosening 3 retaining screws  7 15 1   Note that retaining screws have elastic gaskets on them   7 
19. tary axis  Adjust the 4pcs of adjustment crews which are used  for fixing the laser indicator  in order to make the diameter of the laser spot circle no more than 1 25cm    4  Spread the screw and laser source part with shellac     Fixture  1 25CH       Main axis part       Adjust Vises   ment  screw       Prepared By  Tim Wojo Revision  2 042109 Page 13 of 16         BCL easyPDF  Printer Driver          nove thisimage      se to remor    omens al sho    LAGER teve DRO 40 6532 40 6543 Service Manual    2 2  Vertical Mode Operation    In this mode of operation  the rotating laser surface is a plumb plane  while the laser line projected from the top is in projecting  horizontal  While in this mode  the instrument is manually level  meaning that accuracy is a function of how accurately the    vial is calibrated     2 2 1 Vial Calibration    Vial accuracy calibration is accomplished by Vial part  performing the following     a  Place the laser level on it   s side  mounting Adjustment    trivot  on the test stand described in section g SCTEW  2 1 1  The test stand is referenced  throughout this procedure     b  Adjust the leveling screw  located directly       below the keypad assembly  until the vial       indicates    level    i e  bubble is directly in  the center of the vial    c  Measure the difference between the near  target and far target  Given this test set up  the error should be 1mm or less  Anything greater  requires calibration  procedure that follows    d  Remove t
    