Home
        Alaris ® GP Volumetric Pump - Directions for Use
         Contents
1.                                                                          WAN R KI   A  AN d IN     i   AN WEE A  A Ve    WWGANA y  gd Wad  Kwa   NN   d d S    A V               31 40    1000DF00459 Issue 1    Maintenance    Routine Maintenance Procedures       To ensure that this pump remains in good operating condition  it is important to keep it clean and carry out the routine maintenance  procedures described below     Interval Routine Maintenance Procedure   As per Hospital Policy Thoroughly clean external surfaces of the pump before and after prolonged period of  storage    Each usage 1  Inspect AC power supply plug and cable for damage     2  Inspect case  keypad and mechanism for damage   3  Check Start up self test operation is correct     Before the transfer of the pump toa Clean the pump by wiping over with a lint free cloth lightly dampened with warm water and a  new patient and as required standard disinfectant   detergent solution     If the pump is dropped  damaged  subjected to excessive moisture or high temperature  immediately take it out of  service for examination by a qualified service engineer     All preventative and corrective maintenance and all such activities shall be performed at a compliant work place in  accordance with the information supplied  CareFusion will not be responsible should any of these actions be performed  outside the instructions or information supplied by CareFusion  For Preventative and Corrective Maintenance  instructions please
2.      Alaris   DS Docking Station and Alaris   Gateway Workstation     Pump can only be mounted on the horizontal section of the docking stations listed  above        1000DF00459 Issue 1 10 40    The Alaris   Safety Clamp    ALARIS   SAFETY CLAMP       Safety Clamp    Safety Clamp  Slider    SAFETY CLAMP IN OCCLUDED POSITION     After infusion set is loaded into the pump  opening the door  activates door hooks which will pull the Safety Clamp slider out   as shown     Clamp is in OCCLUDED  POSITION    SAFETY CLAMP IN NON OCCLUDED POSITION     When a new infusion set is removed from packaging the Safety  Clamp will be in this position      Clamp in NON  OCCLUDED POSITION       FLOW ENABLED      This is necessary to avoid tube damage during storage and to ensure  correct sterilisation and allows immediate priming     MANUALLY OPERATING THE SAFETY CLAMP    To move the slider into the non occluded position manually   push up Safety Clamp Tab and push Safety Clamp Slider  completely into Frame     Pushing on the Safety Clamp Slider enables full set flow to the patient  Therefore it is recommended to always close the    roller clamp as well     However  if gravity infusion is required  push up Safety Clamp Tab and push orange Safety Clamp Slider completely into  Frame to enable flow  The gravity infusion can be regulated using the roller clamp on the set     1000DF00459 Issue 1            Hereinafter referred to as to as  Safety Clamp       11 40    Loading an Infusion Set       Ens
3.      e EEN  In line Clamp 1  Check Valve  Prevents secondary infusions  from flowing back up the    primary Infusion set instead of  to the patient      1  Ensure primary set has a backcheck  valve upstream from the Y site     In line Clamp  e g  roller clamp        Upper Y Site  SmartSite   Needle Free  Valve  on Primary Infusion Set         1  The secondary set connects to the upper  Y connection on the primary Infusion set                 1000DF00459 Issue 1 20 40    Service Configuration Mode    This section comprises of a list of options which can be configured  Some can be entered via the pump SERVICE CONFIGURATION menu   available in Technician Mode  and others through the Alaris   GP Editor Software     Enter the access code on Alaris   GP Volumetric Pump for SERVICE mode  then select SERVICE CONFIGURATION  see the Technical Service  Manual for details     Use Alaris   GP Editor to configure the pump configuration  drug list and units enabled for each data set     11  Access codes should only be entered by qualified technical personnel        Date  amp  Time    Select DATE  amp  TIME from the SERVICE CONFIGURATION menu using the K2C keys and press the OK softkey   Press the OK softkey to confirm    Use the RDC keys to adjust the date displayed  pressing the NEXT softkey to access the next field    When the correct date and time are displayed press the OK softkey to return to the SERVICE CONFIGURATION menu   Press the QUIT softkey to return to the SERVICE menu and pres
4.     Pole Clamp Installation    A pole clamp is fitted to the rear of the pump and will provide secure fixing to vertical I V  poles of a diameter between 15 and 40 mm     Recessed         1  Pull the folded pole clamp towards you and unscrew the clamp to leave enough  room for the size of the pole     2  Place pump around pole and tighten screw until the clamp is secured to the pole     Never mount the pump such that the infusion stand becomes top heavy or  unstable     Ensure pole clamp is folded away and stored within recessed area at the  rear of the pump before connecting to a Docking Station Workstation  or  when not in use     Docking Station Workstation  or Equipment Rail Installation    The rotating cam can be fitted to the rectangular bar on the  Docking Station Workstation  or equipment rails measuring 10mm  by 25mm     Rotating  Rectangular bar cam    Release lever  push 1  Align the rotating cam on the rear of the pump with the  to release  rectangular bar on the Docking Station Workstation  or the  equipment rail     2  Push the pump firmly onto the rectangular bar or equipment  rail        Ensure that the pump  clicks  securely into position onto  the rail or bar     3  To release  push the release lever and pull the pump  forwards     It is recommended that infusion bags be located ona  hanger directly above the pump with which they are          being used  This minimises the potential for confusion  of infusion sets when multiple volumetric pumps are  used    
5.    BOLUS          RATE Bolus  Administering a controlled volume of fluid or drug at an increased rate for diagnostic or therapeutic  SETWITH ANVY purposes  The pump should always be infusing and always attached to the patient   Drugs given by  an IV bolus could achieve immediate and high drug concentration levels    80 ml h The bolus feature can be configured via the Alaris  GP Editor to   VOLUME a  BOLUS Disabled  0 0 mt b  BOLUS Enabled       Y HOLD KEY BELOW BOLUS Disabled    If configured to Disabled  pressing the ED button will have no effect and the pump will continue to infuse  at the set rate           A Bolus cannot be administered if the feature is disabled for the selected data set or  specific drug  During BOLUS the pressure limit alarm is temporarily increased to the    maximum level  L8         BOLUS Enabled    Press and hold the  flashing  BOLUS softkey to deliver the required bolus  The bolus rate can be adjusted   The bolus volume is limited in the configuration via Alaris   GP Editor     1  During infusion press the EY button once to display the bolus screen    2  Use the keys to adjust the bolus rate if required    3  To deliver the bolus press and hold the BOLUS softkey  During the bolus  the volume being infused  is displayed  When the desired bolus volume has been delivered or the bolus volume limit is reached   release the softkey  The bolus volume is added to the total volume infused displayed     If the volume to be infused  VTBI  is reached during a 
6.    NOTE  On completion of VTBI pump will continue to infuse at KVO rate     KVO       INFUSING KVO    AAA     CANCEL                At the end of VTBI  the pump will first display VTBI DONE INFUSING  0 ml h KVO  Press CANCEL to display KVO screen        The pump continues to infuse at a very low rate is used to keep the  VTBI Joni patients vein open  in order to prevent blood clots and catheter    occlusions   VOLUME 0 mi NOTE  If the KVO rate  5ml h  is greater than the set infusion  O Oh 00m00s parameters then the pump will continue to infuse at the set  VOLUME infusion rate  The KVO rate will flash on screen to indicate this is  not the usual infusion rate     The pump will beep every 5 seconds while in KVO mode     To check and adjust the pressure level  press the button  The display will change to show the current  pumping pressure level and the pressure alarm limit  The pressure alarm limit can be set via the Alaris    GP Editor           1  Press the  amp  DCO keys to increase or decrease the alarm limit  L2 L5 or L8   The new limit will be  indicated on the display     2  Press OK to exit the screen     The pressure alarm limit is auto adjusted and is fixed at level 8  L8  for rates  above 200mI h     The interpretation of pressure readings and occlusion alarms are the    responsibility of the clinician depending on the specific application     Occlusion levels for the Alaris   GP Volumetric Pump are configured in the Data Set  Editor        1000DF00459 Issue 1 15 40 
7.   10    5   E         5   As       i i i d d  0 5 10 15 20 25 30 35  Observation Window  Minutes   oo Minimum Rate Error  RM Maximum Rate Error     A    Overall Mean Error    0 4   Trumpet Graph at 25ml h  After 72 Hours   15 0 7  10 0    5 0    0 0     p aa  50      10 0     15 0     t t t t    0 5 10 15 20 25 30 35    Observation Window  Minutes        Minimum Rate Error      Maximum Rate Error     A    Overall Mean Error    0 7     Note  The typical flow rate and trumpet curves are achieved using a recommended infusion set   The plot range has been increased to   150   to allow visualization of the graph     1000DF00459 Issue 1    38 40    Products and Spare Parts    Alaris    Infusion System  Range of products in the Alaris   Infusion System product family are     80013UN01 Alaris   GS Syringe Pump  80023UN01 Alaris   GH Syringe Pump      For Docking Stations and Workstation contact local customer services representative to obtain configurations availability and part numbers        Spare Parts   A comprehensive list of spare parts for this pump is included within the Technical Service Manual    The Technical Service Manual  1000SM00013  is now available in electronic format on the World Wide Web at     www carefusion com alaris intl     A username and password are required to access our manuals  Please contact a local customer services representative to obtain login  details     10005P00487 Internal Battery Pack  1000SP01183 RS232 Cable    AC Power Lead   UK  AC Power Lea
8.   16mmHg at 5   C and  41  18mmHg  at 40   C respectively  High Pressure nominally 4  14mmHg at 5   C and  38  21mmHg  at 40   C respectively    The specified accuracy may not be maintained if the above    conditions are not met  see notes 1 to 4        36 40    Trumpet and Flow Rate Curves       In this pump  as with all infusion systems  the action of the pumping mechanism and variations cause short term fluctuations in rate  accuracy     The following curves show typical performance of the system in two ways  1  the accuracy of fluid delivery over various time periods is  measured  trumpet curves   and 2  the delay in onset of fluid flow when infusion commences  start up curves      Trumpet curves are named for their characteristic shape  They display discrete data averaged over particular time periods or    observation  windows  not continuous data versus operating time  Over long observation windows  short term fluctuations have little effect on  accuracy as represented by the flat part of the curve  As the observation window is reduced  short term fluctuations have greater effects as  represented by the  mouth  of the trumpet     Knowledge of system accuracy over various observation windows may be of interest when certain drugs are being administered  Short term  fluctuations in rate accuracy may have clinical impact depending on the half life of the drug being infused and the degree of inter vascular  integration  the clinical effect cannot be determined from the trump
9.   1A rating  in the clinical environment to control receive data from  the pump lies with the user of the equipment  This  software should include detection of the disconnection Typical Connection Data    or other failure of the RS232 cable    Any connected analogue and digital components are  required to meet IEC EN60950 for data processing and  2 Transmit Data  TXD  Output  IEC EN60601 for medical devices  Anyone connecting 3 Received Data  RXD  Input  additional devices to the signal input or output is a   system configurator and responsible for meeting the 4 DTR  gt  DSR  6   requirements of the system standard IEC EN60601 1 1  5 Ground  GND    To connect to the RS232 port use spare part 1000SP01183    RS232 cable     1 Nursecall  Relay  Normally Closed    6 DSR     DTR  4   7 Not used       8 Nursecall  Relay  Normally Open    9 Nursecall  Relay  Common    Ulli  ed awo DIR mo mD ncc  321    5 4    Baud Rate 115k Baud  Start Bits 1 Start Bit  Data Bits 8 Data Bits  Parity No Parity  Stop Bits 1 Stop Bit       1000DF00459 Issue 1 35 40       Infusion Specifications    System Accuracy   Rate Accuracy is  5   achieved under nominal conditions 2 2       Rate Accuracy is  10   achieved under low flow conditions       Occlusion Alarm Limits  Achieved under nominal conditions  4     lt   200 ml h    gt  200 ml h    Pressure   mmHg   approx     Maximum Occlusion Alarm Pressure  1250 mmHg   Post Occlusion Bolus    Bolus volume generated at 25 ml h when the minimum occlusion  alarm
10.   30             1  oot 0 5 10 15 20 25 30 35  02       i   d   d H d d d Observation Window  Minutes   O 10 20 30 40 50 60 70 80 90 100 110 120      Minimum Rate Error    Time  minutes  BR  Maximum Rate Error     A    Overall Mean Error   0 7     Note  The typical flow rate and trumpet curves are achieved using a recommended infusion set     1000DF00459 Issue 1 37 40       Error        Flow Rate  ml h     Trumpet and Flow Rate Curves  Continued     Trumpet Graph at 1 0ml h  Initial Period     30 7   20     104 Ka  hk he    10               20         30 d           d  o 5 10 15 20 25 30 35  Observation Window  Minutes           Minimum Rate Error  RB  Maximum Rate Error     4    Overall Mean Error   0 7        Start up Graph at 25 0ml h  Initial Period           5             d    0 10 20 30 40 50 60 70 80 90 100 110 120       Time  minutes     Trumpet Graph at 25 0ml h  Initial Period     15    10          Error              15                  0 5 10 15 20 25 30 35       Observation Window  Minutes        Minimum Rate Error     EE Maximum Rate Error     A    Overall Mean Error   0 6     Error        Error        Error        Trumpet Graph at 1 0ml h  After 72 Hours                                         30 0 y  20 0    10 0    za a  0 0   d   zk he   10 0     20 0     30 0   4           f i  0 5 10 15 20 25 30 35  Observation Window  Minutes    4  Minimum Rate Error  EE Maximum Rate Error     A    Overall Mean Error   1 0   Trumpet Graph at 25 0ml h  After 24 Hours   15 E
11.   automatically switch off in  the immediate future     SAFETY CLAMP Pump on hold Safety clamp broken or e Clamp infusion set using roller clamp   missing    Replace infusion set     Investigate and correct set loading   SET MISLOAD Pump on hold Set loaded incorrectly    Clamp infusion set using roller clamp      Investigate and correct set loading     FLOW SENSOR Infusion stopped Flow sensor unplugged e Check   replace flow sensor or set VTBI   during infusion   DISCONNECT       1000DF00459 Issue 1 24 40       Alarms  Continued     Infusion Status   Cause Action    WRONG SET Pump on hold Safety clamp not detected  e Clamp infusion set using roller clamp   e Check set and close door     Replace infusion set   If necessary     DOOR CLOSE Pump on hold Safety clamp in non occluded e Clamp infusion set using roller clamp   position with door open or   Investigate and correct set loading   INCOMPLETE obstructed    Close door   DO NOT USE Pump on hold   infusion Internal error has occurred    Remove pump from use   stopped    LEVER OPEN Infusion stopped Door lever is open e Check door lever   e Check lever hooks   e Check lever is not obstructed  if so  free  obstruction        Warnings alert the user but may not stop the infusion and are indicated by an audible sound  a flashing amber warning indicator and a message on the  display or both     1  Check the display for a warning message  Press to silence the sound for 2 minutes  CANCEL to clear the message   2  Rectify the cause of
12.  Close door and open roller clamp  Ensure no drops are   Air in line sensor falling in the drip chamber    Tubing guide 9  Ensure all air is removed from the set  Connect the  infusion set to the patient access device     1  Follow steps 1 to 4 as above where necessary   2  Ensure roller clamp is closed     3  Open door and load infusion set as follows   e Fit blue adaptor on infusion set into blue top set  retainer   e Insert orange safety clamp  leaving slider extended   in the occluded position into orange retainer     Alaris   Safety Pushing on the Safety Clamp Slider may lead to  Clamp  Orange  uncontrolled flow to the patient  Therefore  always  in occluded close the roller clamp before pushing on the safety  position  See clamp slider     previous page   4  Ensure infusion set is fully inserted into tubing  guide   Tubing guide 5  Close door and open roller clamp  Ensure no drops are  falling in the drip chamber   6  Ensure all air is removed from the set  Connect the  infusion set to the patient access device     1000DF00459 Issue 1 12 40       Starting the Infusion    1  PRIME AND LOAD THE SET  Refer to    Loading an Infusion Ser        1  Ensure the pump is connected to an AC power supply  also operates from battery    Connect flow sensor  if reguired   See  Flow Sensor Operation    Press the key   The pump will run a short self test  Check two beeps are activated during this test   Check the displayed date and time are correct   Check display shows the data set nam
13.  DFU  manual carefully        Initial Set Up    1    Check that the pump is complete  undamaged and that the voltage rating specified on the label is compatible with your AC power    supply   2  Items supplied are   e   Alaris   GP Volumetric Pump    Directions For Use  CD   e AC Power Cable  as requested   o Protective Packaging     Alaris   GP Editor Software  including the Alaris   GP Transfer Tool    per hospital    3  Connect the pump to the AC power supply for at least 22 hours to ensure that the internal battery is charged  verify that the ACE is  lit     4  On initial start up the pump will display the Select Language screen  Select the required language from the list displayed using the  aDC keys     5  Press the OK softkey to confirm your selection     The Alaris   GP Editor Software can be used to create an approved data set that can be uploaded into the pump  However  a  default data set is already installed in the pump  See details below      The pump will automatically operate from its internal battery if the pump is switched on without being connected to the    power supply   Should the pump fail to perform correctly  replace in its original protective packaging  where possible and contact a  qualified service engineer for investigation        Factory Default Data Set  The Alaris   GP Volumetric Pump is supplied with the following factory default data set    Bolus Rate Max 1200ml h ug kg h  5ml       1000DF00459 Issue 1 9 40       Getting Started  Continued    
14.  Volumetric Pump  hereinafter referred to as  Pump   is a small lightweight volumetric infusion pump that provides accurate  and reliable infusions over a range of rates     The Alaris   GP Editor Software  is a medical device accessory  which allows the hospital to develop a best practice data set of IV medication  dosing guidelines for patient specific care areas  Each data set contains a specific library of drugs  as well as a pump configuration  appropriate for the care area     The hospital defined data set is developed and approved through pharmacy and clinical input  and then transferred into the Alaris   GP  Volumetric Pump by qualified technical personnel     INTENDED USE     The pump is designed to meet the infusion requirements within the operating environment specified in this Directions For Use  DFU   including general wards  critical and intensive care  operating rooms and accident and emergency rooms     The clinician is responsible for evaluating whether the pump is appropriate for use in a specified patient care area     This pump is suitable for use by appropriately trained clinicians or nurses  This pump can be used for Intravenous modes  supporting fluid  therapy  drug therapy  blood transfusions and parenteral nutrition     The Asena    brand name has been recently changed to the Alaris   brand name  This change in brand name has no effect on the intended use  or functionality of the product  Recommended disposable products for use with this product may r
15.  a secondary infusion  Piggyback  in the same pump channel   The maximum permissible setting for the Volume To Be Infused for secondary infusions     The maximum permissible infusion rate for secondary infusions       These settings may be overriden by drug list settings     1000DF00459 Issue 1 22 40    Drug List available via the Alaris   GP Editor Software       The following drug parameters are only configurable via the Alaris   GP Editor Software  PC based   see Alaris   GP Editor help files for  further details     CONCENTRATION SETTINGS     Concentration Specifies the drug concentration   Concentration Min The weakest permissible concentration for this drug  amount of drug per ml    Concentration Max The strongest permissible concentration for this drug  amount of drug per ml      DOSE RATE SETTINGS     Weight Based Units Selects weight based or non weight based units    Dose Rate Default The default dose rate for infusing this drug    Dose Rate Units The unit for dose rate parameters    Dose Rate Max The maximum permissible dose rate for infusing this drug     BOLUS SETTINGS     Bolus Mode  Allows the use of the bolus feature for this drug   Bolus Rate Default  The default value for bolus rate for this drug   Bolus Volume Max  The maximum permissible bolus volume per bolus session  for this drug       These settings override pump configuration settings     1000DF00459 Issue 1 23 40       Alarms stop the infusion and are indicated by a combination of an audible sound  f
16.  burette infusion sets    60103E   2 SmartSite    Needle Free Valve Port   e 1 Burette  150ml     Length  270cm  63441E   4 SmartSite    Needle Free Valve Port   e 1 Burette  150ml     Length  330cm    Alaris   GP low sorbing infusion sets    60953 e 15 Micron Filter     Polyethylene lined PVC tubing  e Length  260cm                                                          63260NY e Polyethylene lined PVC tubing    No Filter  e Length  295cm    Alaris   GP syringe adapter infusion sets    63280NY e Length  270cm   1  Restricted to maximum infusion rate of 150ml h          Alaris   GP secondary infusion set    72213N 0006 e Male luer and hanger  e Length  76cm                            72951NE  For use e 1 SmartSite    Needle Free Valve Port   with 60950E  e Male luer with Backcheck Valve   e Length  71cm    Do not use with pump in secondary infusion  mode when infusing critical drugs     Check infusion set materials and drug compatibility before selecting an infusion set   It is recommended that infusion sets are changed according to the instructions in the  Changing the Infusion Set  section   Carefully read the Directions For Use supplied with the infusion set prior to use        Please note these drawings are not to scale    1000DF00459 Issue 1 30 40    Associated Products    The Alaris   Gateway Workstation    The Alaris   DS Docking Station       x  wl EE ee  do    ll ls lili   OA      ES                                                     Y                             
17.  including domestic establishments and those directly  connected to the public low voltage power supply network that supplies buildings used for domestic  purposes      This pump is not intended to be used in the presence of a flammable anaesthetic mixture with air or  oxygen or nitrous oxide    Operating Pressure    The pumping pressure alarm system is not designed to provide protection against  or detection of  extravasation or tissuing  complications which can occur   Alarm Conditions  SE        Several alarm conditions detected by this pump will stop the infusion and generate visual and audible    alarms  Users must perform regular checks to ensure that the infusion is progressing correctly and no  alarms are operating        1000DF00459 Issue 1 7 40    Operating Precautions  Continued     Electromagnetic Compatibility and Interference      This pump is protected against the effects of external interference  including high energy radio frequency  emissions  magnetic fields and electrostatic discharge  for example  as generated by electrosurgical and  cauterising equipment  large motors  portable radios  cellular telephones etc   and is designed to remain safe  when unreasonable levels of interference are encountered        e Therapeutic Radiation Equipment  Do not use the pump in the vicinity of any Therapeutic Radiation  Equipment  Levels of radiation generated by the radiation therapy equipment such as Linear Accelerator  may  severely affect functioning of the pump  Pl
18.  infusion set  by pulling back the handles  Refer to the illustration  above     3  Proceed with load  priming  and set up instructions as described in section    Getting Started        NOTE  Ensure drip chamber is half full and upright     Always attach the flow sensor before you start an infusion    Avoid using the flow sensor in direct sunlight   Always ensure lens is clean     1  Always replace the flow sensor interface cover when the flow sensor is disconnected     Flow  Sensor  Interface  Cover       1000DF00459 Issue 1 27 40       infusion sets  The user is responsible for verifying the suitability of a customers  Please contact your local CareFusion    The Alaris  GP Volumetric Pump uses standard  single use  disposable 1  New sets are continuously being developed for our  product used  if it is not recommended by CareFusion  representative for availability        Alaris   GP standard infusion sets    60073 e 2 Injection Ports   No Filter   Length  260cm  60093E   2 SmartSite    Needle Free Valve Ports    15 Micron Filter  e 1 Backcheck Valve   Length  260cm  60123E    2 SmartSite    Needle Free Valve Ports     1 2 8  15 Micron Filter   Length  265cm  60293E   2 SmartSite    Needle Free Valve Ports    1 Backcheck Valve  e No Filter   Length  260cm  60693 e 1 Injection Port   15 Micron Filter   Length  255cm  60693E e 1 SmartSite    Needle Free Valve Port   e 15 Micron Filter   Length  255cm                                                             60793 e 2 Injectio
19.  rate as prompted on the display in the next screen   GHIJ KLM  NOPQRS  TUVWXYZ    wY    ml h       SELECTWITH A A VY    SELECT  ml h 1  Select DOSING ONLY from the list using the ASYM keys   DOSING ONLY 2  Press OK to confirm   DRUGS  A EEN 3  Select the dosing units from the list using the aDC keys  press OK to confirm   NOPQRS 4  Enter WEIGHT  using the     SY V keys  press OK to confirm   TUVWXYZ 5  Use the A DY Y keys to select the TOTAL VOLUME   press OK to confirm        6  Enter DRUG AMOUNT using the ASD keys and if units need to be changed  select UNITS which  will scroll through the units available  Press OK to confirm selection    7 A summary of the DOSING ONLY information is displayed  to CONFIRM  all details shown press OK  The  BACK softkey may be used at any time to return to the previous screen     SELECTWITH AN VY       1   Only displayed if weight based units are used     N      Total Volume   Drug Volume   Diluent Volume i e  Total Volume of fluid in the fluid container after a drug is added        SELECT  ml h    DOSING ONLY  DRUGS  ABCDEF      Select the required DRUGS alphabetical row from the list using the ARE keys     Press OK to confirm       Select the drug from the displayed list using the ASD keys  press OK to confirm     Enter WEIGHT  using the ASD keys  press OK to confirm     Use the ADC D keys to enter the TOTAL VOLUME  press OK to confirm       Enter DRUG AMOUNT using the aDC keys  press OK to confirm selection     A summary of the DRUG infor
20.  refer to the Technical Service Manual  TSM      All servicing should only be performed by a qualified service engineer with reference to the TSM     Please refer to Technical Service Manual for calibration procedures  The units of measurement used in the calibration  procedure are standard SI  The International System of Units  units           Battery Operation    The internal rechargeable battery allows continued operation when the AC power is unavailable  for example during patient transfer or  AC power failure  Mean Time To Battery Empty from fully charged is a minimum of 6 hours  When connected to the AC power supply for  4 hours   whether the pump is in use or not  a new battery pack will be fully charged     The battery is maintenance free  sealed Nickel Metal Hydride and requires no routine servicing  However  to achieve optimum operation   ensure that the battery is fully recharged after full discharge  before storage  and at regular 3 month intervals during storage     Charge retention will eventually degrade  Where retention is critical the internal battery should be replaced every 3 years     It is recommended that only a qualified service engineer replaces the battery  For further information regarding the replacement of  batteries refer to the Technical Service Manual     The battery pack used in this Alaris   Volumetric Pump is manufactured by CareFusion and includes a proprietary PCB  printed circuit board   designed specifically for the Alaris   Volumetric Pu
21.  the warning or proceed with caution     Display Infusion Cause Action  Status  BATTERY LOW Infusion continues Less than 30 minutes of battery     Connect to power supply   life remaining  e Check power cable   AC POWER FAIL AC power disconnected or failed      Reconnect to power supply   VTBI DONE Infusing KVO Intended VTBI completed    Set new VTBI or clear VTBI     AIR IN LINE Pump on hold Air detected in infusion set at   Ensure set is fitted correctly in the air in line  the start of infusion  detector   Set not fitted correctly into air in     Assess air in infusion set   line detector  e Check fluid level in drip chamber   e Check level of fluid in container   SET CLOCK Pump on hold Date   time not set    A qualified service engineer must set date    time     Press cancel softkey to continue   TITRATION Infusion continues Rate titration not confirmed  e Confirm or cancel new rate   RATE LOCK Infusion continues Rate lock not confirmed  e Select YES or NO as required          If pump was on hold the alarm will still be activated but this message will not be displayed     1000DF00459 Issue 1 25 40       Prompts alert the user but may not stop the infusion and are indicated by an audible sound  a flashing amber warning indicator and a message on the  display or both     1  Check the display for a prompt message  Press to silence the sound for 2 minutes  CANCEL to clear the message   2  Rectify the cause of the prompt or proceed with caution     Display Infusion Cause Actio
22.  threshold is reached  lt 0 45 ml    Bolus volume generated at 25 ml h when the maximum occlusion  alarm threshold is reached  lt 0 95 ml    Bolus Volume Accuracy   Typical   4 1   Max   3 2   Min   5 5  1ml E 10ml h    Typical   1 3   Max   0 9   Min   1 6  100ml e 1200ml h    Administering a Bolus    10   1200ml h in steps of 10ml h       Bolus Volume 0 0ml   100 0ml in steps of 0 1ml  Displayed    Starting the Infusion   Set up    Infusion Rate 0 1   99 9ml h in steps of 0 1ml h 8   100   999ml h in steps of 1ml h    1000   1200ml h in steps of 10ml h    VTBI Primary  0   OFF   1   9999ml  VI  Total  0 1   9999ml    Maximum time for activation of occlusion alarm   At Maximum Pressure  time to alarm at 0 1ml h is nominally 735   450  minutes  Maximum  lt 883 min     At Minimum Pressure  time to alarm at 0 1ml h is nominally 234   425  minutes  Maximum  lt 309 min     At Maximum Pressure  time to alarm at 1 0ml h is nominally 65   4   minutes  Maximum  lt 95 min     At Minimum Pressure  time to alarm at 1 0ml h is nominally 16   2   minutes  Maximum  lt 28 min     At Maximum Pressure  time to alarm at 25ml h is nominally 119   7   seconds  Maximum  lt 3 min     At Minimum Pressure  time to alarm at 25ml h is nominally 29   3   seconds  Maximum  lt 50 sec     1000DF00459 Issue 1          Air Sensor    Integral Ultrasonic Sensor    Air in line detection    Single Bubble  configurable   Soul  100ul  250ul  amp  500ul     Bubble accumulation  1ml over a 15 minute window     Crit
23.  to hospital protocol   Insert spike into new container    Squeeze the drip chamber approximately half full or up to fill line  if the drip chamber is marked  with fluid    Restart infusion  see  Getting Started         OF an ie    It is recommended that infusion sets are changed in accordance with the Directions For Use     1  When changing the infusion set or the fluid container use aseptic technique according to hospital protocol   Carefully read the Directions For Use supplied with the infusion set prior to use        SmartSite   Needle Free System Instructions    SmartSite   Needle Free Valve is designed to permit safe gravity flow and automated flow  injection and aspiration of fluids without the use  of needles by utilising Luer lock and luer slip connectors     Precautions    Discard if packaging is not intact or protector caps are unattached    If Needle Free Valve is accessed by a needle in an emergency the valve will be damaged causing leakage  Replace Needle   Free Valve immediately    Needle Free Valve contraindicated for blunt cannula system    DO NOT leave slip luer syringes unattended        DIRECTIONS   Use Aseptic Technique   1  Prior to every access  swab top of Needle Free Valve port with 70  Isopropyl alcohol  1 2  seconds  and allow to dry  approximately 30 seconds     NOTE  Dry time is dependent on temperature  humidity  ventilation of the area    2  Prime valve port  If applicable  attach syringe to Needle Free Valve port and aspirate minuscule  air bu
24. 295cm    Alaris   GP blood infusion sets    60393 e 2 Injection Ports   200 Micron Filter   e Length  270cm   60393E   2 SmartSite    Needle Free Valve Ports     200 Micron Filter     Length  270cm                                                                60893 e 1 Injection Port   200 Micron Filter   e Length  255cm   60894 e 1 Injection Port   200 Micron Filter   e Length  255cm   60980 e Twin Spike  1 Injection Port    200 Micron Filter   e Length  250cm    63477E e 2 Non  Vented Spikes   e 180 Micron Filter   Length  305cm  e 1 SmartSite    Needle Free Valve Port     1  Check infusion set materials and drug compatibility before selecting an infusion set                                                     Itis recommended that infusion sets are changed according to the instructions in the  Changing the Infusion Set  section   Carefully read the Directions For Use supplied with the infusion set prior to use        Please note these drawings are not to scale  1000DF00459 Issue 1 29 40       Infusion Sets  continued     New sets are continuously being developed for our  customers  Please contact your local CareFusion  representative for availability     infusion sets  The user is responsible for verifying the suitability of a    The Alaris  GP Volumetric Pump uses standard  single use  disposable  product used  if it is not recommended by CareFusion  11        Alaris   GP light resistant infusion sets    e 15 Micron Filter    PVC tubing  e Length  250cm       Alaris   GP
25. Alaris   GP Volumetric Pump Directions For Use    EN       Page  UU EE 2  AboutthisManWal occasion a a di   2  Creatingia Data iii A IA 2  Features of the Alaris   GP Volumetric PUMP         6 cece cece cnn cece nnn cnn rr 3  Controls  ane Indicators t   NEE NN Reha Ne EN Reg EEN deg TN Se Greed A  Symbol DEM iIi EN EEN REENEN REENEN REENEN 5  Main Display Features iia 6  Operating Precautions  iii A aa ia wa 7  E len EE 9  TAASI DEE 11  LoadingianmINfUSION Aaa 12  Starting the INfUSION sad AARE IA EAR EAA A AEE AA REE 13  ee EE 14  Secondary  Piggyback  Infusions  0 0 0    cece ee cnc rr 20  Service Configuration MAA Shwe dade dE EERSTEN AER 21  Pump Configuration available via the Alaris   GP Editor Software                                            22  Drug List available via the Alaris   GP Editor Software    23  CHE IEN 24  Warning EE 25  PROM EIA WI ene Bl E E NE E EEEE E OREN RE 26  Restarting an Infusion following an Air in Line Alarm    26  Flow  Sensor Operation  Optional   ssi siers Ua 27  Interes 28  Associated Product  II IA 31  E Wee EE 32  lee Kee Storage EE 33  N  I le CEET 33  EE AI aa 34  IrDA  RS232 and Nursecall Specification Wa aaa 35  InfUSION SPECINCALIONS  mi IAA Piles eae TEE SEs kaa das 36  Trumpet and Flow Rate Curves 1 0       cece cc cece II nen enn rr Ene een EEE EEE EE 37  Products and Spare Parts viii ii AANEREN sa Meee ea aoe ada ANE CTEM E AE 39  Servic COMA EE 40  Po MA AA aa 40    1000DF00459 Issue 1 1 40    Introduction    The Alaris   GP
26. Time To Battery Empty of 6 hours     Battery Charging   2 5 hours to 95      Alarm Conditions      AC POWER FAIL  VTBI DONE  BATTERY LOW  AIR IN LINE  TITRATION   SET CLOCK  RATE LOCK    AIR IN LINE  SINGLE BUBBLE   AIR IN LINE  ACCUMULATED   DOOR OPEN   DOWNSTREAM OCCLUSION  UPSTREAM OCCLUSION   NO FLOW   FLOW ERROR   FREE FLOW   BATTERY EMPTY   SAFETY CLAMP   SET MISLOAD   FLOW SENSOR  DISCONNECTED   WRONG SET   DOOR CLOSE INCOMPLETE  DO NOT USE   LEVER OPEN    ATTENTION  SETVTBI   SET NOT FITTED  LOCKED    Memory Retention    The electronic memory of the pump will be retained for more  than 2 years with normal use     34 40    IrDA  RS232 and Nursecall Specification  IrDA   RS232   Nursecall Feature RS232   Nursecall Connection Data    The IrDA  or RS232   Nursecall optional feature  is a feature on Nursecall Specification    Alaris   GP Volumetric Pump that allows the pump to be connected    8 Sa Connector D Type   9 Pin  to an external device for the purpose of data communication   TXD RXD EIA RS232 C Standard  Baud Rate 115k Baud   The nursecall interface provides a remote backup  Start Bits 1 Start Bit   to the internal audible alarm  It should not be relied  Data Bits 8 Data Bits   upon to replace monitoring of the internal alarm    ji   Refer to the Technical Service Manual for further Parity No Parity    information regarding the RS232_ interface  Stop Bits 1 Stop Bit   The An tf or the suitability of any software used Nurse Call Relay Contacts Pins 1  8   9  30V dc
27. areFusion    Edificio Veganova    Avenida de La Vega  n  1   Bloque 1   Planta 1    28108 Alcobendas  Madrid   Espa  a     Tel   34  902 555 660    Fax   34  902 555 661  FR    CareFusion    Parc d affaire le Val Saint Quentin  2  rue Ren   Caudron   78960 Voisins le Bretonneux  France    T  l   33  1 30 05 34 00  Fax   33  1 30 05 34 43    November 2009    1000DF00459 Issue 1    GB    CareFusion    The Crescent  Jays Close   Basingstoke    Hampshire  RG22 4BS   United Kingdom     Tel   44  0800 917 8776    Fax   44  1256 330860  HU    CareFusion   D  brentei t  r 1   H 1013 Budapest   Magyarorsz  g     Tel   36  14 88 0232  Tel   36  14 88 0233    Fax   36  12 01 5987  IT    CareFusion    Via Ticino 4    50019 Sesto Fiorentino   Firenze  Italia     T  l   39  055 30 33 93 00  Fax   39  055 34 00 24  NL    CareFusion    De Molen 8 10   3994 DB Houten   Nederland     Tel   31  30 228 97 11    Fax   31  30 225 86 58  NO    CareFusion   Solbr  veien 10 A   1383 ASKER   Norge     Tel   47  66 98 76 00  Fax   47  66 98 76 01    40 40    NZ    CareFusion    14B George Bourke Drive   Mt Wellington 1060    PO Box 14 518    Panmure 1741  Auckland   New Zealand    Tel  09 270 2420  Freephone  0508 422734    Fax  09 270 6285  SE    CareFusion   Hammarbacken 4B   191 46 Sollentuna   Sverige     Tel   46  8 544 43 200    Fax   46  8 544 43 225  US    CareFusion    10020 Pacific Mesa Blvd    San Diego  CA 92121   USA     Tel   1  800 854 7128  Fax   1  858 458 6179  ZA    CareFusion    Uni
28. attery     e An explosion hazard exists if the pump is used in the presence of flammable anaesthetics  Exercise care to locate  the pump away from any such hazardous sources     A   Dangerous Voltage  An electrical shock hazard exists H the pump s casing is opened or removed  Refer all servicing  to qualified service personnel       Do not open the RS232 Nurse Call protective covering when not in use  Electrostatic discharge  ESD  precautions  are required when connecting RS232 Nurse Call  Touching the pins of the connectors may result in ESD protection  failure  It is recommended that all actions must be taken by appropriately trained personnel           If this pump is dropped  subjected to excessive moisture  fluid spillage  humidity or high temperature  or otherwise  suspected to have been damaged  remove it from service for inspection by a qualified service engineer  When  transporting or storing the pump  use original packaging where possible  and adhere to temperature  humidity  and pressure ranges stated in the Specifications section and on the outer packaging        e     If this pump behaves abnormally  remove from service and contact a qualified service engineer     Care should be taken to ensure power leads and RS232 cables do not present a trip hazard       Care should be taken in the placement of power leads and RS232 cables to prevent accidental tugging     1000DF00459 Issue 1 8 40    Getting Started    1  Before operating the pump read this Directions For Use 
29. bbles    3  When used with administration sets always refer to individual set directions for use as change  interval may vary according to clinical application  e g  infusions of blood  blood products  and  lipid emulsions     NOTE  During use of Needle Free Valve port  fluid may be observed between the housing and   blue piston  This fluid does not enter the fluid path and requires no action    NOTE  For product questions or needle free valve educational materials  contact your   CareFusion representative  Consult facility protocols  Consult other organizations that  publish guidelines useful in developing facility protocols        1000DF00459 Issue 1 19 40    Secondary  Piggyback  Infusions    Secondary  or piggyback  Infusion mode is only available if it has been configured   The application of secondary infusions should be limited to the intermittent therapy of medications which are not sensitive to the total time  reguired to complete an infusion     Typically antibiotics may be infused using a secondary infusion  where the primary infusion is limited to maintenance  fluid  If intending to use the secondary infusion facility  the primary infusion should be a maintenance fluid only and is  not indicated for drug therapy     The application of secondary infusions for delivery of critical drugs  particularly those with a short half life  is NOT  indicated for use  These drugs should be administered through a dedicated pump channel     Dependent upon factors such as fluid vi
30. bolus  the VTBI complete alarm  will sound  Press      to silence the alarm or CANCEL to acknowledge the alarm  See    VTBI section for more details on VTBI operation   When using infusion set 63280NY the maximum infusion rate is 150ml h        TITRATE  PRESS  lt  gt  TO CONFIRM    If Rate Titration is enabled  via the Alaris   GP Editor  the infusion rate or dose rate  if available  can be adjusted  25 0 mih while infusing     16 7 ug kg 24h    RATE    1  Select the new rate using the keys     The message  lt TITRATE PRESS TO CONFIRM  gt  will flash on screen and the pump continues to  infuse at the original rate     2  Press the button to confirm the new infusion rate and start infusing at the new rate        VTBI 45 0 ml      VOLUME 50 0 ml 34    O 1h 48m00s If Rate Titration is disabled the rate can only be adjusted whilst ON HOLD   Q 1  Press the Y button to put the pump ON HOLD    2  Select the new rate using the E2 lt 2 keys    3  Press the button to start infusing at the new rate              1000DF00459 Issue 1 16 40    Basic Features  continued   Rate Lock  If Enabled   V1 7 X onwards     If Rate Lock is enabled  when the infusion rate has been set and the infusion started  or following a bolus infusion  the rate lock prompt will  appear on the main display     To select the rate lock function press the YES softkey  Press the NO softkey if the rate lock is not required   When rate lock is enabled  the following are unavailable   e Changing the infusion rate   titrat
31. d   European       Alaris   GP Editor Software    The following item may be useful when using the Alaris   GP Volumetric Pump     1000SP01310 Alaris   GP Editor PC Software Kit    1000DF00459 Issue 1 39 40    Service Contacts    For service contact your local Affiliate Office or Distributor     AE    CareFusion   PO Box 5527   Dubai  United Arab Emirates     Tel   971  4 28 22 842    Fax   971  4 28 22 914  AU    CareFusion    3 167 Prospect Highway   PO Box 355   Seven Hills  NSW 2147   Australia     Tel   61  2 9838 0255    Fax   61  2 9674 4444  BE    CareFusion   Leuvensesteenweg 248 D   1800 Vilvoorde    Belgium     Tel   32  2 267 38 99  Fax   32  2 267 99 21  CA    CareFusion    235 Shields Court   Markham   Ontario L3R 8V2   Canada     Tel   1  905 752 3333    Fax   1  905 752 3343  CH    CareFusion Switzerland 221  Sarl   Critical Care   A One Business Centre   Zone d   activiti  s Vers la Pi  ce  n   10   1180 Rolle   Switzerland   Ph   0848 244 433    Fax  0848 244 100    Document History    CO Number Date    Revision    1 9685    CN    CareFusion    Shanghai Representative Office  Suite  A  Floor 24    Shanghai Times Square Office Building   No 500 Zhangyang Road    Shanghai 200122  China     Tel   86  21 58368028    Fax   86  21 58368017  DE    CareFusion   Pascalstr  2   52499 Baesweiler   Deutschland     Tel   49  2401 604 0    Fax   49  2401 604 121  DK    CareFusion   Firskovvej 25 B   2800 Lyngby   Danmark    TI  45 70 20 30 74  Fax   45 70 20 30 98  ES    C
32. e  alarm will resound after this time     BOLUS button   Press to access BOLUS softkey  Press and hold down softkey  to operate    BOLUS   fluid or drug delivered at an accelerated rate    e Pump is infusing   e Infusion set is connected to patient    e Volume infused  VI  is added to the total volume infused displayed     OPTION button   Press to access optional features     PRESSURE button   Use this button to display the pumping pressure and  adjust the alarm limit     CHEVRON keys   Double or single for faster   slower increase   decrease of  values shown on display     BLANK SOFTKEYS   Use in conjunction with the prompts shown on the  display     AC POWER indicator   When illuminated the pump is connected to an AC  power supply and the battery is being charged     BATTERY indicator   When illuminated the pump is running on the internal  battery  When flashing the battery power is low with less than 30 minutes of  use remaining        1000DF00459 Issue 1 4 40       Symbol Definitions    Labelling Symbols     Symbol    Attention  Consult accompanying document     Potential Equalisation  PE  Connector    RS232 Nursecall Connector     Defibrillation proof type CF applied part   Degree of protection against electrical shock     Protected against spraying water    Alternating Current    Device complies with the requirements of Council Directive 93 42 EEC as amended by      s    2007 47 EC     O  O      OH    Date of Manufacture    Manufacturer    Connector for Flow Sensor  Imp
33. e and version number   NOTE  The pump starts up and displays previous settings     e No Drug Name e Drug Protocol e Primary Secondary    eml h e Dosing Only e Drug Names   e Drug Name       CLEAR SETUP           CLEAR SETUP  CLEAR SETUP     RATE 150   vi  1 50 mon    0 01mo mi  1 4 ml                   4  CLEAR SETUP    Selecting KEEP will retain all previous rate and volume settings  go to step 6   Selecting CLEAR will automatically reset the rate and volume settings to zero and the SELECT screen will be  displayed  if configured      SELECT    ml h    5  If CLEAR was selected  choose from either ml h  DOSING ONLY or DRUGS  A Z   and press OK to confirm  Then follow the prompts as required   Refer to    Basic  Features  Drugs and Dosing  section          DOSING ONLY       i        DRUGS  ABCDEF 6  Clear VOLUME infused   if required   Refer to  Clear Volume Infused  section  this  GHIJKLM is recommended for a new patient or when a new infusion is set up    NOPQRS       7  Enter VTBI  if required  by selecting VTBI softkey on main display   Refer to  Setting a VTBI  or  Setting VTBI over Time section     TUVWXYZ          SELECT WTH AN VY  OK       Set VTBI by using the BAGS option and or    S      keys and press OK to          confirm   8  Enter or adjust the RATE  if necessary  using the ASV V keys   9  Press key to start the infusion     INFUSING will be displayed     NOTE  The green run LED will flash to show that the pump is infusing     If the infusion requires to be stop
34. e sets are not  kinked as this may occlude the tubing    Using Collapsible bags  Glass Bottles  amp  Semi Rigid containers         It is recommended that the air vent be opened on the Alaris  GP Volumetric Pump set if using glass bottles  or semi rigid containers  to reduce the partial vacuum formed as the fluid is infused from the container   This action will ensure the pump can maintain volumetric accuracy whilst the container empties  The  action of opening the air vent for semi rigid containers should take place after the spiking of the container  and priming of the drip chamber    Steps for the Collapsible bags Steps for Semi rigid containers   Follow steps 1 to 3 as shown for the   semi rigid containers  however do not   A e 3  4   open vent as in step 4  but prime the ae  set as per step 5  Ensure the bag outlet Fildrip Open theairyent  chamber to to allow pressure  is fully pierced before filling the drip equalisation   ready  chamber  for infusion  Prime the set by  opening   closing the  Close the roller clamp  roller clamp  Operating Environment     When using any infusion pump in conjunction with other pumps or devices requiring vascular access   extra care is advised  Adverse delivery of medication or fluids can be caused by the substantial variation  in pressures created within the fluid channels of such pumps  Typical examples of those pumps are used  during dialysis  bypass or cardiac assist applications      The pump is suitable for use in all establishments 
35. eaning of flow sensors which have been heavily soiled  contaminated or if the handle operation is not free moving  then the flow  sensor may be immersed and soaked in clean soapy water  see      The inside of the spring mechanism can be cleaned by activating it whilst  submerged in the water    After cleaning  the sensor should be allowed to dry fully prior to use     The plug of the flow sensor must not be immersed in water as damage will occur        Information on Disposal for Users of Waste Electrical  amp  Electronic Equipment    This X symbol on the product and or accompanying documents means that used electrical and electronic products should not be mixed  with household waste     If you wish to discard electrical and electronic equipment  please contact your CareFusion affiliate office or distributor for further  information     Disposing of this product correctly will help to save valuable resources and prevent any potential negative effects on human health and  the environment which could otherwise arise from inappropriate waste handling     Information on Disposal in Countries outside the European Union    This symbol is only valid in the European Union  The product should be disposed of taking environmental factors into consideration   To ensure no risk or hazard  remove the internal rechargeable battery and the Nickel Metal Hydride battery from the control board and  dispose of as outlined by the local country regulations  All other components can be safely dispo
36. ease consult manufacturer   s recommendations for safe distance and  other precautionary requirements  For further information  please contact your local CareFusion representative       Magnetic Resonance Imaging  MRI   The pump contains ferromagnetic materials which are susceptible to  interference with magnetic field generated by the MRI devices  Therefore  the pump is not considered an MRI  compatible pump as such  If use of the pump within an MRI environment is unavoidable  then CareFusion highly  recommends securing the pump at a safe distance from the magnetic field outside the identified    Controlled  Access Area    in order to evade any magnetic interference to the pump  or MRI image distortion  This safe distance  should be established in accordance with the manufacturers    recommendations regarding electromagnetic  interference  EMI   For further information  please refer to the product technical service manual  TSM    Alternatively  contact your local CareFusion representative for further guidance     e Accessories  Do not use any non recommended accessory with the pump  The pump is tested and compliant  with the relevant EMC claims only with the recommended accessories  Use of any accessory  transducer or cable  other than those specified by CareFusion may result in increased emissions or decreased pump immunity       In some circumstances the pump may be affected by an electrostatic discharge through air at levels close to or  above 15kv  or by radio frequency 
37. efer to either the Asena   brand name or  Alaris   brand name and both types are suitable for use with this infusion pump       Only some parts of the Alaris   GP Editor software are classified as a medical device accessory    About this Manual    The user must be thoroughly familiar with the pump described in this manual prior to use   All illustrations used in this manual show typical settings and values which may be used in setting up the functions of the pump   These settings and values are for illustrative use only  The complete range of settings and values are detailed in the specifications section     Creating a Data Set    To create a data set for the Alaris   GP Volumetric Pump  first the hospital will need to develop  review  approve  upload according to the  following process  Refer to the Alaris   GP Editor help file for further details and operating precautions     1  Create Care Area data set  Using Alaris   GP Editor     Drug List Drug names and concentrations for a data set with default value and maximum limits     Up to 100 unique drug names drug protocol set ups     Pump Configuration Pump configuration settings and units for dosing only     2  Review  approve and export data set  Using Alaris   GP Editor   Review and Approve Entire data set report to be printed  reviewed and signed as proof of approval by an  authorised person  according to hospital protocol  Signed printout to be kept safe by    hospital  Data set status to be set to Approved  Password is 
38. et curves alone     The start up curves represent continuous flow versus operating time for two hours from the start of the infusion  They exhibit the delay  in onset of delivery due to mechanical compliance and provide a visual representation of uniformity  Trumpet curves are derived from the  second hour of this data  Tests performed per IEC EN60601 2 24 standard     Start up Graph at 0 1ml h  Initial Period  Trumpet Graph at 0 1ml h  After 24 Hours     150 00                               En A     Mn l m   n i   lh JA    00  4 4   S   a  sf ea ea wr a   SE    6    Time  minutes  Observation Window  Minutes          Minimum Rate Error  RB  Maximum Rate Error     A amp     Overall Mean Error    5     Trumpet Graph at 0 1ml h  Initial Period  Trumpet Graph at 0 1ml h  After 72 Hours     150                                                                         1500 5  100  g 50          amp  e  LI Di D D D a   0 0  J kk te te ts      2         LI  50  100  150  1500 r 1 r 1 r 1 1     ji S e     l 7 S o 5 10 15 20 25 30 35  Observation Window  Minutes  S n E  Observation Window  Minutes        Minimum Rate Error         Minimum Rate Error      Maximum Rate Error BR  Maximum Rate Error     A    Overall Mean Error    6     A    Overall Mean Error   4 0   Start up Graph at 1 0ml h  Initial Period  Trumpet Graph at 1 0ml h  After 24 Hours   207 30 7  18  F  20    16  l  144 10   KM  12  L D  E A     04i A          ae      amp     ost 5 10   y e  E 06 o    a 20    z 044 d  9  m 02
39. eys  as required    3  Press OK softkey to confirm or QUIT to exit screen     Pump Details    To review pump information    1  Press the   button to access the options menu    2  Select PUMP DETAILS    3  Review the information and then press the QUIT softkey     1000DF00459 Issue 1 18 40    Basic Features  continued     Changing the Infusion Set    Press to put the pump ON HOLD    Close in line clamp and ensure the access to the patient is isolated    Disconnect the infusion set from the patient    Open pump door and remove infusion set from the pump and discard the set and fluid container according to hospital protocol   Prepare the new infusion set  load infusion set into pump and close the door  see  Loading the Infusion Set     Restart infusion  see  Getting Started      Sr St Wie zs    When changing the infusion set or the fluid container use aseptic technigue according to hospital protocol   It is recommended that infusion sets are changed in accordance with the Directions For Use  Carefully read the  Directions For Use supplied with the infusion set prior to use  The set change interval is up to 72 hours with the  following exceptions     e Transfusion  Blood  Sets   e 60953 Alaris   GP Low Sorbing Infusion Set  e 60033E Alaris   GP Low Sorbing Infusion Set  e 60950E Alaris   GP Oncology Infusion Set    Changing the Fluid Container    Press to put the pump ON HOLD    Remove bag spike on infusion set from empty   used container  Discard empty   used container according
40. follows     1  Press Dto put the pump ON HOLD   Press the   button to access the options menu     9 y      Select PRIMARY SETUP and press the OK softkey to confirm  Make changes to the primary setup as necessary     Setting VTBI over Time  V1 7 K onwards     This option allows a specific VTBI and delivery time to be set  The rate necessary to deliver the reguired volume within the specified time is  calculated and displayed     1  Stop the infusion  Press   button to access the options menu   Select the SET VTBI OVER TIME option using the ASAD keys and press the OK softkey     Adjust the volume to be infused using the ASD keys   Or select BAGS softkey to set the VTBI  When the desired volume has  been reached press the OK softkey     EN Zei     gt     Enter the time over which the volume is to be infused using the RACD keys   The infusion rate will automatically be calculated   Press OK softkey to enter the value or BACK to return to the VTBI     an    Adjust Alarm Volume  V1 7 K onwards     This option allows adjustment of the volume    Press the   button to access the options menu    Select ADJUST ALARM VOLUME    Select HIGH  MEDIUM or LOW using the ANY Y keys   Press OK softkey to confirm or QUIT to exit screen     Pp  Nie    Enable   Disable Rate Lock  V1 7 X onwards     This option allows configuration of the Rate Lock feature to be enabled or disabled   1  Press the   button to access the options menu     2  Select ENABLE RATE LOCK or DISABLE RATE LOCK using the ASV Y k
41. get back to the SERVICE menu and press to exit and  power down     1000DF00459 Issue 1 21 40    Pump Configuration available via the Alaris   GP Editor Software       The following options are only configurable via the Alaris   GP Editor Software  PC based   see Alaris   GP Editor help files for further    details   GENERAL SETTINGS   AC Fail Warning    Alarm Volume    PRESSURE SETTINGS   Pressure Default    Pressure Max    RATE SETTINGS     Rate Titration    Infusion Rate Max    BOLUS SETTINGS   Bolus Mode    Bolus Rate Max  Bolus Rate Default     Bolus Volume Max     PATIENT SETTINGS   Weight Default    AIR IN LINE SETTINGS   AIL Limit Max    VTBI SETTINGS   VTBI Max    SECONDARY INFUSION SETTINGS     Secondary Infusion  Sec  VTBI Max    Sec  Infusion Rate Max    Warning to indicate that the AC Power has been disconnected and the pump is operating on  battery power     The audio volume of the pump used for alarms and warnings     The default occlusion alarm limit     The maximum occlusion pressure alarm limit     Allows the adjustment of the infusion rate while the pump is infusing  without putting the pump  on hold     The maximum permissible infusion rate     Allows use of the bolus feature   The maximum permissible bolus rate   The default values for bolus rates     The maximum permissible bolus volume in a session     The default patient weight     The single bubble AIL setting     The maximum permissible setting for the Volume To Be Infused  VTBI      Allows the use of
42. ical Volume  The maximum volume infused following a single fault condition  is for rates  lt  10ml h      025 ml  rates  lt  100ml h      0 5ml  rates   gt 100ml h      2 ml    Set based  pump activated Safety Clamp Device to prevent free  flow    la  Nominal conditions are defined as   Set Rate  1 to 1200ml1 h   Recommended disposable  60593   Needle  18 gauge x 40mm   Solution Type  De ionized  amp  Degassed Water   Temperature  23  C   2  C  Fluid Head Height   300   30mm   Back Pressure  0   10mmHg     1b  Low flow conditions are defined as   Set Rate  less than 1 0ml h  Recommended disposable  60593   Needle  18 gauge x 40mm   Solution Type  De ionized  amp  Degassed Water   Temperature  23  C   2  C  Fluid Head Height   300   30mm   Back Pressure  0   10mmHg     2  The system accuracy will change by the following percentages 3  Temperature nominally 5 7   1 5   at 5  C and nominally   0 3   1 7   at 40   C  Fluid Head Height  nominally  3 4   1 3   at  0 5m and  0 0   1 1   at  0 5m  Duration  nominally  1 1  0 2  over 24 hours of continuous use  Back Pressure  nominally  2 0  1 3  at  100mmHg   13 4  1 8   at  800mmHg respectively  Atmospheric pressure    5  at 25ml h at 700hPa    3  Tested using Distilled water  20  lipid  50  glucose  0 9   Normal Saline and 5  Alcohol solutions     4  The occlusion pressure accuracy will change by the following   Temperature  Low setting nominally 7  12mmHg at 5   C and   24  17mmHg at 40   C respectively  Normal setting nominally 4
43. ion   e Bolus   e Switching the pump off      VTBI over time infusions      Secondary infusions  if enabled    To turn rate lock off    1  Press the   button to access the options menu    2  Select UNLOCK RATE and press the OK softkey    To turn rate lock on    1  Press the   button to access the options menu    2  Select RATE LOCK and press the OK softkey        Adjusting Existing Dosing or Protocol Infusions   Set By ml h   Set By Doserate    To set doserate or flowrate in precise increments it may be necessary to switch between the rate adjust options SET BY DOSERATE and    SET BY ml h  An arrow to the left of the rate display shows the rate changed when the E2 lt 2 keys are used to increase decrease the  infusion rate     To set a doserate precisely the arrow must be pointing to the doserate  for example  mg kg h   the flowrate will be calculated from the  doserate     To precisely set a flowrate the arrow must be pointing to flowrate  ml h   the doserate will be calculated from the flowrate   Selecting the SET BY ml h Option    1  Press the   button to access the options menu     2  Select the SET BY ml h option using the    gt    YW keys and press the OK softkey indicated on the screen  This will select the set by  flowrate option  the arrow on the display will automatically select the flowrate  the flowrate can be adjusted if necessary     Selecting the SET BY DOSERATE Option  1  Press the   button to access the options menu     2  Select the SET BY DOSERATE option usi
44. lashing red alarm indicator and a message on the display    1  Check the display for an alarm message and review table below for cause and action  Press to silence the sound for 2 minutes  CANCEL to clear the  message    2  When the cause of the alarm has been rectified  press the QI key to resume the infusion   Exceptions are DO NOT USE  amp  BATTERY EMPTY     Status    AIRIN LINE Infusion stopped Single air bubble exceeds     Assess the amount of air detected by air in line   alarm limit  detector   Set not fitted correctly into     Opening the door may cause an air bubble to rise  air in line detector  in the set  Check set for air      Remove air according to hospital policy      Ensure set is fitted correctly in the air in line   detector    e Check level of fluid in container    e Check enough fluid left in drip chamber       Restart infusion     AIR IN LINE Infusion stopped Accumulated air bubbles     Review infusion set for air bubbles and take  exceeds alarm limit  appropriate action    Multiple bubbles smaller      Check level of fluid in container   than the single bubble e Check enough fluid left in drip chamber   alarm limit  which has   Restart infusion   been detected over a 15  min  window and  gt 1ml      DOOR OPEN Infusion stopped Door was opened during an     Close door or clamp infusion set using roller clamp   infusion    Restart infusion     DOWNSTREAM OCCLUSION    nfusion stopped A blockage has occurred   Check fluid path between pump and patient for  d
45. mation is displayed  to CONFIRM  all details shown press OK  The  BACK softkey may be used at any time to return to the previous screen     TUVWXYZ       SELECT WTH AN VY    NO UO A U N gt         1   Only displayed if weight based units are used     2   Total Volume   Drug Volume   Diluent Volume i e  Total Volume of fluid in the fluid container after a drug is added     1000DF00459 Issue 1 14 40    Basic Features  continued     VOLUME       VOLUME INFUSED       Om   OFF           SEL ECT WITH ANNY    VTBI DONE    Clear Volume Infused    This option enables the volume infused to be cleared   1  Press the VOLUME softkey on main display to show the clear VOLUME INFUSED option   2  Press the CLEAR softkey to clear the volume infused  Press the QUIT softkey to retain the volume     has not been cleared  then the message DOSE INFUSED HAS BEEN CLEARED will be  displayed     1  When a new drug or a new concentration has been setup and the previous volume infused       Setting a VTBI    1  Using the RACD keys     a  Press the VTBI softkey on main display to enter the volume to be infused screen     b  Enter the volume to be infused using the AED keys and press OK to confirm   OR  2  Using the BAGS softkey    a  Press the VTBI softkey on main display to enter the volume to be infused screen     b  Select the BAGS softkey  select the required bag volume using the ASAD keys and  press OK to confirm the selection     c  Press OK to confirm again  or adjust the VTBI using the ASAD keys  
46. mp  and in conjunction with Alaris   Volumetric Pump software  controls battery use   charge and temperature  Any use of battery packs that are not manufactured by CareFusion in the Alaris   Volumetric Pump is at your sole  risk  and CareFusion does not provide any warranty for or endorsement on any battery packs that are not manufactured by CareFusion   CareFusion s product warranty shall not apply in the event the Alaris   Volumetric Pump has suffered damage or premature wear  or  malfunctions or otherwise operates incorrectly  as a result of use with a battery pack that is not manufactured by CareFusion     1000DF00459 Issue 1 32 40       Cleaning and Storage    Cleaning the pump     Before the transfer of the pump to a new patient and periodically during the use  clean the pump by wiping over with a lint free cloth lightly  dampened with warm water and a standard disinfectant   detergent solution     Before cleaning always switch off and disconnect Recommended cleaners are   from the AC power supply  Do not allow liquid to  Brand Concentration  enter the casing and avoid excess fluid build up on   Hibiscrub 20   v v   the pump  Virkon 1   w v   y p Do not use the following disinfectant types   Do not use aggressive cleaning agents as these   NaDcc  such as PRESEPT   may damage the exterior surface of the pump    Hypochlorites  such as CHLORASOL   Do not steam autoclave  ethylene oxide sterilise or   Aldehydes  such as CIDEX     immerse this pump in any fluid    Cationic S
47. n  Status  ATTENTION Pump on hold Pump left on hold for 2 minutes     Review pump setup   without starting the infusion    Start infusion or turn off pump     SET VTBI Pump on hold No VTBI   flow sensor    Set VTBI or fit flow sensor     LOCKED Rate change attempted whilst e Unlock rate to adjust infusion settings   locked     Restarting an Infusion following an Air in Line Alarm    The pump may be restarted by opening the door  assessing and removing any air from the tubing guide area and in the  infusion set on the patient side of the pump  if required  according to hospital policy  Close the door and cancel the air   in line alarm  Restarting the infusion will reactivate the air in line system and will alarm if the preset air in line limit is  exceeded        1000DF00459 Issue 1 26 40    Flow Sensor Operation  Optional        The flow sensor automatically monitors the infusion flow rate through the drip chamber  The flow sensor will  cause the pump to alarm ifa significant deviation from the infusion rate occurs  The flow sensor will also be able to  detect empty containers  For this reason we recommend use ofa flow sensor wherever possible excluding secondary  infusions     IVAC   Model 180 Flow Sensor        ee    Pull back handles  gt     _ 1        chamber    Flow sensor  interface                      1  Plug the flow sensor into the flow sensor interface located on the top rear part of the pump     2  Attach the IVAC   Model 180 Flow Sensor to the drip chamber of the
48. n Ports   15 Micron Filter   Length  255cm    60793E   2 SmartSite    Needle Free Valve Ports    15 Micron Filter  e Length  255cm   60903 e 15 Micron Filter   Length  250cm   60593 e 15 Micron Filter    Length  260cm    60173E e 1 SmartSite   Needle Free Valve Port    No Filter  e Length  260cm   63120V e 1 Split Septum Injection Port  1 Backcheck Valve  e No Filter   Length  305cm                            Check infusion set materials and drug compatibility before selecting an infusion set   Itis recommended that infusion sets are changed according to the instructions in the  Changing the Infusion Set  section   Carefully read the Directions For Use supplied with the infusion set prior to use        Please note these drawings are not to scale  1000DF00459 Issue 1 28 40       Infusion Sets  continued             WA EE customers  Please contact your local CareFusion  infusion sets  The user is responsible for verifying the suitability of a y    The Alaris    GP Volumetric Pump uses standard  single use  disposable 11  New sets are continuously being developed for our  product used  if it is not recommended by CareFusion  representative for availability        Alaris   GP standard infusion sets    63200NY e No Filter   Length  260cm   63110V   2 Split Septum Injection Ports    No Filter    Length  290cm   63401E    1 SmartSite    Needle Free Valve Port    No Filter    Length  275cm   63420E   2 SmartSite    Needle Free Valve Ports     1 Backcheck Valve   No Filter    Length  
49. ng the   2 lt  keys and press the OK softkey indicated on the screen  This will select the set  by doserate option  the arrow on the display will automatically select the doserate  the doserate can be adjusted if necessary     Dosing Summary    To review currently selected dosing information     1  Press the   button to first access the options menu   2  Select DOSING SUMMARY   3  Review the information and then press the QUIT softkey     Infusion Setup    To change the Infusion Setup  refer to  Basic Features   Drugs and Dosing  Selecting the INFUSION SETUP  section     Drug Name Only    This feature adds a drug name to an existing infusion  when infusing using ml h or dosing only options   1  Press the    button to access the options menu    2  Select DRUG NAME ONLY    3  Press the OK softkey to confirm the drug name or press the QUIT softkey to exit the option     1000DF00459 Issue 1 17 40    Basic Features  continued     Clear Drug Name    Clearing the drug name is only available if drug name only has been selected   1  Press Dto put the pump ON HOLD    Press the      button to access the options menu    Select DRUG NAME ONLY using the 42  keys  press OK to confirm     Select CLEAR DRUG NAME  displayed if a name only is selected  using the   2 lt 2 keys  Press the OK softkey to confirm the  selection     wai 5    Primary Setup    Ifa secondary infusion has already been setup  see  Secondary  Piggyback  Infusions  section   then access to the primary infusion setup is  as 
50. ortant Information    Not for Municipal Waste    Fuse rating  Authorised representative in the European Community    iad    m  D       1000DF00459 Issue 1 5 40    Main Display Features       Main Display   If VTBI has not been set  flow sensor must be used      Infusion Status     Drug Name Primary or 0 N HO L D ONHOLD  Secondary     a   SETRATEWIHA A Y Y   Only if secondary is enabled in the RATE  data set              E RATE If the rate has not been set  Infusion Rate        gt  and is showing at 0 0ml h   25 0 then message a  will be  D ml h displayed      34  Volume Infused 50 0 ml  Softkey Identifiers  VOLUME   ONHOLD  ten b  RATE TOO LON  RATE   Clear Volume If programmed rate is    Inf d between 0 0ml h and 1 0ml h  nruse exclus ve n drug p otocol     Set VTBI Option message b  will be shown     Main Display   If VTBI is set     Infusion Status     Drug Name Primary or    gt  A D R E NA L IN E z           Secondary   Only if secondary is enabled in the  ONHOLD  Infusion Rate     25 0 mih d RATE TOO HIGH  RATE  Dose Rate    gt       16 7   g kg 24h  Volume to be Infused 45   If programmed rate is greater  0 m than the Infusion Rate Max  Volume Infused    gt  50 0 mi EH Get    Clear Volume  Infused    TIME REMAINING DISPLAY icon   Indicates time remaining before VTBI will be completed  If  the time is greater than 24 hours then 24  will be displayed     PRESSURE INFORMATION icon   Shows the pressure from level 0 being the first bar to level 8   Alarm limits  level 2  5 o
51. ownstream  clamps  connectors  kinks or blockages       Examine access site for signs of  complications  redness  swelling  pain  heat      UPSTREAM OCCLUSION Infusion stopped A blockage has occurred   Check set above the pump     upstream  Possible   Check all clamps above pump   container empty  e Check fluid level in container     Ensure drip chamber is half filled     Ensure that the bag spike is inserted correctly     Ensure air vent on drip chamber is open on all  glass and semi rigid containers     NO FLOW Infusion stopped Flow sensor detects no flow      Check flow sensor   e Check fluid level in container   e Ensure all clamps above pump are open   e Ensure drip chamber is half filled   e Ensure that the bag spike is inserted correctly   e Check flow sensor is clean     FLOW ERROR Infusion stopped Gross difference between   Clamp infusion set using roller clamp   detected drops and   Check flow sensor   expected amount of drops      Check fluid level in drip chamber     FLOW ERROR  In secondary Infusion stopped Unexpected drops   Hang secondary container above primary   detected  e Check drops are from secondary container when    infusion mode only  infusing     e Flow sensor disconnection is recommended     Infusion stopped Uncontrolled flow e Clamp infusion set using roller clamp   FREE FLOW  possible  e Remove pump from use     BATTERY EMPTY Infusion stopped The internal battery is   Connect to power supply  exhausted  The pump will immediately or switch pump off 
52. ped immediately  the following actions may be applied   e by pressing the key  recommended action     e by closing the roller clamp  e by opening the door         If a drug name is selected  then CLEAR SETUP  will alternate with the drug name     If secondary infusions have been enabled in the data set  then PRIMARY may also alternate     1000DF00459 Issue 1 13 40    Basic Features    Drugs and Dosing    The following options enable the pump to be set up for use with a specific drug name and or drug protocol  Drugs are pre configured in the  Alaris   GP Editor to enable rapid selection of the drug name  dosing units and default rate  For increased security using a configured drug   maximum and minimum safety limits are programmable for concentration and dose rates   Using the Alaris   GP Editor software     infusion rate in ml h  This does not affect the accuracy of the infusion     1  When adjusting an infusion using the dose rate  the display may not show any corresponding changes to the       Selecting the INFUSION SETUP    1  Press the    button to first access the options menu   Drugs and dosing set up options are available by selecting INFUSION SETUP from the list using the K2C keys     Select from the list of the options  ml h  DOSING ONLY or DRUGS  as detailed below and press the OK softkey to confirm the  selection     SELECT    ml h 1  Select ml h from the list using the GC ei keys  if necessary    DOSING ONLY 2  Press OK to confirm   DRUGS  ABCDEF 3  Enter the ml h
53. r 8        1000DF00459 Issue 1 6 40    Operating Precautions      To ensure correct and accurate operation  only use CareFusion single use infusion sets described in this     gt  Directions For Use   It is recommended that infusion sets are changed according to the instructions in the  Changing the Infusion Ser   LD  section  Carefully read the Directions For Use supplied with the infusion set prior to use   YA e Use of non specified infusion sets may impair the operation of the pump and the accuracy of the infusion     pal i    Qa               When combining several apparatus and or instruments with infusion sets and other tubing  for example  via a 3 way tap or multiple infusion  the performance of the pump may be affected and should be  monitored closely                                        Uncontrolled flow may result if the infusion set is not properly isolated from the patient i e  closing a tap in    the set or activating an in line clamp   roller clamp      The infusion set may be fitted with an in line clamp  which can be used to occlude tubing in case it is  required to stop fluid flow      The Alaris   GP Volumetric Pump is a positive pressure pump  which should use infusion sets fitted with  Luer lock fittings or equivalent locking connectors      To infuse from a burette  close the roller clamp above the burette and open the clamp on the vent on top  of the burette    e Discard infusion set if the packaging is not intact or the protector cap is detached  Ensur
54. radiation close to or above 10v m  If the pump is affected by this external  interference the pump will remain in a safe mode  the pump will duly stop the infusion and alert the user by  generating a combination of visual and audible alarms  Should any encountered alarm condition persist even  after user intervention  it is recommended to replace that particular pump and quarantine the pump for the  attention of appropriately trained technical personnel          This pump is a CISPR 11 Group 1 Class B device and uses RF energy only for its internal function in the normal  product offering  Therefore  its RF emissions are very low and are not likely to cause any interference with the  nearby electronic equipment  However  this pump emits a certain level of electromagnetic radiation which  is within the levels specified by IEC EN60601 2 24 and IEC EN60601 1 2  If the pump interacts with other  equipment  measures should be taken to minimise the effects  for instance by repositioning or relocation     Earth Conductor    The Alaris   GP Volumetric Pump is a Class   device  therefore must be earthed when connected to an AC power  supply       This pump also has an internal power source      When connected to an external power source  a three wire  Live  Neutral  Earth  supply must be used  If the  integrity of the external protective conductor on the AC power cable has been compromised  the pump should  be disconnected from the AC power source and operated utilising the internal b
55. required      Export Export data set for use by the Alaris   GP Transfer Tool  or to back up a data set  or to  move the data set to another PC     3  Upload data set to Alaris   GP Volumetric Pump  Using Alaris   GP Transfer Tool     Drug parameters have to be in accordance with local regulation and prescribed information     11  Data set transfers should only be performed by qualified technical personnel        1000DF00459 Issue 1 2 40    Features of the Alaris   GP Volumetric Pump       Alarm indicator                    Softkeys          Chevrons    Mute    Pressure       Battery indicator    AC power indicator On Off       Door Lever          Handle    Release lever for rotating cam    Flow sensor  connector    RS232 Nursecall  connector  cover  removed for clarity        Rotating cam to lock onto    Folded pole clam  horizontal rectangular bars  S p       Medical device  interface  MDI     Mains fuses cover    Mains inlet             IR communications port          Potential Equalisation  PE   Connector                   1000DF00459 Issue 1 3 40    Controls and Indicators       Controls   Symbol    ON OFF button   Press once to switch the pump ON  Press and hold down  for approximately 3 seconds to switch the pump OFF     RUN button   Press to start the infusion  The green LED will flash during  infusion     HOLD button   Press to put the infusion on hold  The amber LED will be lit  while on hold     MUTE button   Press to silence alarm for  approximately  2 minutes  Th
56. s    to exit and power down     ER EA e    Pump Reference Text    This option is used to add reference text to be shown on the pump start up display    1  Select PUMP REFERENCE from the SERVICE CONFIGURATION menu using the  amp  C   keys and press the OK softkey   Use the RDC    keys to enter the text and NEXT to move to the next character    When the desired text has been selected press OK softkey to return to the SERVICE CONFIGURATION menu      gt  wm ba    Press QUIT to exit back to the main SERVICE menu and press to exit and power down     Language    This option is used to set the language of messages shown on the pump display    1  Select LANGUAGE from the SERVICE CONFIGURATION menu using the          keys and press the OK softkey   Use the  amp  SC keys to select the language    When the desired language has been selected press OK softkey to return to the SERVICE CONFIGURATION menu   Press QUIT to exit back to the main SERVICE menu and press to exit and power down     d  r ac ad    Backlight  amp  Contrast    This option is used to set the backlight and contrast on the pump display   1  Select BACKLIGHT  amp  CONTRAST from the SERVICE CONFIGURATION menu using the    S   keys and press the OK softkey     2  Use the    gt YY keys to adjust BACKLIGHT  CONTRAST  amp  DIMMING  The contrast of the display will change when scrolling through  the numbers   Use PARAM to scroll between each option     3  When the desired value has been reached press the OK softkey  then QUIT to 
57. s   to access the OPTIONS screen     4  Select SECONDARY SETUP  press OK to confirm  13  Press OK to start infusing at the displayed rate   5  Select either NO DRUG NAME or DRUGS A Z  Press OK to  confirm either selection  Setting up a subsequent secondary infusion   6  Enter the secondary RATE using the keys  On completion of the secondary VTBI  the pump will  7  Press OK to confirm     automatically transition to the primary infusion   An audible    8  Set VTBI using the AED keys   Refer to  Setting  BEEP  will be heard     a VTBI  section   9  Press OK to confirm  1  Press to place the primary infusion ON HOLD   10  Review PRIMARY SECONDARY setup summary  2  Follow instructions 3 to 13 of  Setting up a secondary infusion      11  If correct  press OK to continue  or BACK to adjust VTBI or RATE   of the SECONDARY mode               Typical Secondary infusions   IN Pole 1      Extension Hook  approx 26cm   Normally included with the secondary Infusion set   Primary fluid container must hang lower to allow the  secondary infusion to run and primary infusion to restart  on completion of the secondary infusion        Secondary Fluid Container  Usually a smaller container e g   50ml  100ml  200ml or 250ml     Secondary Infusion Set  e g  72213N 0006  Usually shorter  tubing to reach the Y site on the  primary Infusion set                 Primary Fluid Container                Primary Infusion set  e g  63420E with an upper Y Site  SmartSite   Needle Free Valve           gt  gt  
58. scosity  the secondary infusion rate  head height between the secondary and  primary fluid containers and the use of clamps  flow may occur from the primary fluid container during a secondary  infusion  This could result in drug remaining in the container at the end of the secondary infusion  delaying its delivery    for a period of time which is dependent upon the primary infusion rate  For example  a secondary infusion of 250ml  at 300ml h could result in approximately 33ml remaining  requiring up to 25 minutes additional time to complete the  delivery  assuming a primary infusion rate of 80ml h  and the use of a 72213N 0006 secondary infusion set and its  supplied extension hook   Therefore it is recommended that flow sensors  if used  are disconnected from the pump during  secondary infusions     Regular monitoring for unexpected primary flow is recommended  If flow from the primary fluid container is not desired  during secondary infusion and or the patient is sensitive to fluid balance  the clamp on the primary infusion set should be  closed  Check that no drops fall in the primary drip chamber        On completion of the primary infusion the pump will continue at Keep Vein Open rate  KVO  rate     Setting up a secondary infusion     12  Press to start the infusion in secondary mode     1  Ensure Primary infusion has been setup in ml h  rate  gt  Oml h   An ADVISORY screen will be displayed   ENSURE SECONDARY  2  Press to put the pump ON HOLD  INFUSION SET OPEN  3  Pres
59. sed of as per local regulations     1000DF00459 Issue 1 33 40    Electrical Protection   Class 1  Type CF  Defibrillation proof   Electrical Mechanical Safety   Complies with IEC EN60601 1 and IEC EN60601 2 24     Electro Magnetic Compatibility  EMC   Complies with IEC EN60601 1 2 and IEC EN60601 2 24     Electrical Safety  Typical earth leakage current 78uA     Typical Enclosure Leakage Current  Normal Condition    OyA  Typical Protective Earth Resistance   32mOhms    The above measurements are for guidance only  IEC EN60601 1  limits are defined below     Earth Leakage Current  Normal Condition   lt   500yA  Enclosure Leakage Current  Normal Condition   lt   100uA  Protective Earth Resistance  lt   200mOhms  Classification   Continuous mode of operation  Portable Equipment   AC Power Supply     100   230 VAC  50   60Hz  60VA  Maximum     Fuse Type     2 XT 1 25 A  slow blowing    Dimensions      148mm  w  x 225mm  h  x 148mm  d   Weight  approx  2 5kg   excluding power cable      Protection against fluid ingress    IPX3   Protected against spraying water     Environmental Specifications     5  C    40  C  20  C    50  C    20    9090  15    95      Atmospheric 700hPa   1060hPa 500hPa   1060hPa  Pressure     Non condensing        1000DF00459 Issue 1       BATTERY SPECIFICATIONS      Rechargeable NiMH  Nickel Metal Hydride   Automatically charges  when the pump is connected to AC power     Battery Life   For a 24 hour battery charge time  the pump at  25ml h will have a Mean 
60. t 2 Oude Molen Business Park   Oude Molen Road  Ndabeni   Cape Town 7405  South Africa     Tel   27   0  860 597 572  Tel   27  21 510 7562    Fax   27  21 5107567    Alaris  Asena  IVAC and SmartSite are registered  trademarks of CareFusion Corporation or one of its  subsidiaries  All rights reserved    All other trademarks are property of their  respective owners       2005 2010 CareFusion Corporation or one of its  subsidiaries  All rights reserved     This document contains proprietary information of  CareFusion Corporation or one of its subsidiaries   and its receipt or possession does not convey any  rights to reproduce its contents  or to manufacture  or sell any product described  Reproduction   disclosure  or use other than for the intended  purpose without specific written authorization of  CareFusion Corporation or one of its subsidiaries is  strictly forbidden     ul CareFusion Switzerland 317 Sarl   CH 1180  Rolle    CareFusion U K  305 Ltd  RG22 4BS  UK    1000DF00459 Issue 1    carefusion com       
61. ure the appropriate infusion set for the fluid drug to be infused has been selected   Follow the instructions supplied with the individual infusion set   Only use Alaris   GP Volumetric Pump infusion sets   Refer to  Infusion Sets  section of the DFU     Position the fluid container to avoid spillage onto the pump   Ensure that the tubing is inserted completely into the top set retainer through to the tubing guide avoiding any slack     Loading an Infusion Set  Alaris   Safety Clamp in the NON OCCLUDED position   FLOW ENABLED    Adaptor on infusion 1  set  Blue    Top set retainer    Blue       Remove infusion set from package and close roller  clamp     2  Insert the bag spike into the fluid container and hang  appropriately   Pressure sensor Ata minimum height of 300 mm above the pump     UPSTREAM    3  Fill the drip chamber to the fill line if shown    Approximately half full  Refer to operating precaution  Pumping section   Using Collapsable bags  Glass Bottles  amp  Semi Rigid  mechanism containers       4  Open roller clamp and prime set slowly  to prevent air  bubbles  ensuring all air is removed   Pressure sensor    DOWNSTREAM 5  Close roller clamp   6  Switch the pump on  Open door and load infusion set  as follows   e Fit blue adaptor of infusion set into blue top set  retainer   ji   Insert orange safety clamp into orange retainer   Alaris   Safety    Clamp  Orange  7  Ensure infusion set is fully inserted into tubing guide     Safety Clamp   retainer  Orange    8 
62. urfactants  such as Benzalkonium Chloride     Please ensure the membrane covering the pumping   lodine  such as Betadine    mechanism is intact prior to cleaning  If faulty    Concentrated Isopropyl alcohol based cleaners will  remove from use and contact a qualified service degrade plastic parts    engineer        Storing the pump      If the pump is to be stored for an extended period it should be first cleaned and the internal battery fully charged  Store in a clean  dry  atmosphere at room temperature and  if available  employ the original packaging for protection    Once every 3 months during storage  carry out functional tests as described in the technical service manual and ensure that the internal  battery is fully charged     Cleaning and storing the infusion set     The infusion set is a disposable single use item and should be discarded after use according to hospital protocol     Cleaning the door     Refer to the Technical Service Manual for information for removing the door to facilitate cleaning of the fluid path  the use of a screwdriver   torx  is required and should only be carried out by a qualified service engineer     Cleaning the Flow Sensor      Before the transfer of the flow sensor to a new infusion set and periodically during use  clean the flow sensor by wiping over with a lint free  cloth lightly dampened with warm water and a standard disinfectant   detergent solution  Ensure the connector does not get wet  Dry flow  sensor before use    To aid cl
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Lenovo 6281 Computer Hardware User Manual  FW DUAL STYLE  Manual del usuario administrador del sistema web para  取扱説明書    Hisense AS-18UR4SFATD5 air conditioner  interpolation dans un maillage et visualisation par - Infoterre  ZTE Director User Manual  Leica TS11/TS15 User Manual  istruzioni per l'uso e la manutenzione operation and service manual    Copyright © All rights reserved. 
   Failed to retrieve file