E1401-90009
Contents
1. 25 Installing the Power Line Fuse 15 L Line Voltage 9 Mainframe description 9 size 25 using 9 weight 25 Maximum Power Output 26 P Parts List 20 Power Cords 16 Power Supply Outputs 26 R Rear Panel Auxiliary Connector 12 13 Remote Standby Switch 14 Replaceable Parts List 20 S Safety Grounding Information 17 Safety Warnings 4 Specifications 25 27 U Using the Mainframe 9 V Variable Speed Fan 11 VXI Ripple Noise 26 VXIplug play Example Programs See VXIplug play Online Help VXIplug play Function Reference See VXIplug play Online Help VXIplug amp play Programming See VXIplug play Online Help VXIplug play Soft Front Panel See VXIplug play Online Help W WARNINGS 4 Index 29 Notes 30 Index
2. interrupteur avant de cet appariel ne coupe pas la totalit de l alimentation des circuits internes En cas d urgence l utilisateur peut devoir debrancher cet appareil Ne pas bloquer l acces au cordon d alimentation ou si l appareil est install sur un chassis au disjoncteur alimentant l appareil Si un interrupteur de secours a distance est branche a cet appariel au moyen d un connecteur Sub D a 25 broches du panneau arriere il est possible que l utilisateur n arrive pas a mettre l appareil hors tension au moyen de l interrupteur frontal Avisez l utilisateur de emplacement du dispositif de sectionnement cordon ou disjoncteur de l appareil Chapter 1 Agilent E1401B Safety Grounding Information WARNING AVERTISSMENT Grounding Procedure Chapter 1 For protection from electrical shock when operating at mains frequencies greater than 66 Hz connect the chassis ground terminal to permanent earth ground Risque de Choch lectrique Si la fr quence du secteur est sup rieure 66 Hz relier la borne de masse du chassis une prise de terre fixe Connect a 16 AWG 1 3 mm or larger wire to the PEM nut shown in Figure 1 6 The wire must be green with a yellow stripe or bare no insulation Use an M4 x 10 screw grounding lug and toothed washers or toothed lug as shown in Figure 1 7 Securely attach the other end of the wire to a permanent earth ground using toothed washers or a toothed lug M4 X
3. 0 240 Vac 10 Nominal Frequency Range 50 60 Hz 10 400 Hz Input Power Nominal Voltage Range 100 120 VAC 10 Nominal Frequency 400 Hz 10 DC Input Power Input Voltage Range 100 VDC minimum 370 VDC maximum General Power Supply Power Factor Corrected Power Switch On Standby with lighted indicator in front Rear panel connection allows remote operation Inrush Current 40A max Designed to meet EN 60555 2 Socket for detachable line cord IEC 320 HOT Chassis ground tap on rear panel M4 x 0 7 threaded insert Fuse Voltage Maximum Current Fuse 100Vac 15A 15AF 120Vac 12A 15AF 220 240Vac 8A 15AF Appendix A Agilent E1401B Specifications 25 Power Supply Outputs DC Output Peak DC Output P P Dynamic CD Current Limit Voltage Current IMP Current IMD supply shuts down 40 C 55 C 5V 65A 60A 9 0A 70 80A 12V 14A 12A 2 5A 14 18A 12V 14A 12A 2 5A 14 18A 24V 14A 12A 5 0A 15 19A 24V 12A 10A 5 0A 12 16A 5 2V 65A 60A 8 5A 70 80A 2V 35A 30A 4 5A 35 41A IMP Rated mainframe peak DC output current as defined by the VXIbus Specification IMD Rated mainframe peak to peak dynamic current as defined in the VXlbus Specification by a current vs frequency curve VXI Ripple Noise DC Output Allowed Variation Max DC Load Max Induced Voltage Ripple Noise Ripple Noise 45V 0 25 0 125 50 mV 50 mV 12V 0 06 0 36 50 mV
4. 0 7 PEM Nut for Permanent Earth Ground Connection x Figure 1 6 Grounding Connection Wire Lug Toothed Washers 7 power Supply Case Figure 1 7 Grounding Screw Toothed Washers Using the Mainframe 17 Notes 18 Using the Mainframe Chapter 1 Chapter 2 Replaceable Parts and Assemblies Module Replacement Component Locator WARNING AVERTISSEMENT Cleaning Fan Filters Chapter 2 There are no user sevicable parts in the Agilent E1401B Repair is limited to replacing the power supply module the backplane module or other mechanical parts only The power supply is available on an exchange basis The backplane assembly should be replaced and not repaired no exchange on the backplane Backplane E1401 66511 Power Supply E1401 69202 Refer to Figures 2 1 through 2 3 for component locator information There are no user servicable parts inside Refer repair and servicing to trained service personnel Ne contient pas d l ment que l utilisateur puisse r parer Confier les r parations et la maintenance un technicien qualifi Remove and clean the mainframe s fan filter regularly see Figure 2 3 Replaceable Parts and Assemblies 19 Agilent E1401B Replaceable Parts List Reference Agilent Part Total Designator Number Qty Description Al E1401 66511 1 PC Backplane Assembly A2 E1401 6920
5. 2 1 Power Supply Power 650 Watts 8 Outputs B1 B4 3160 0864 4 Fan Tubeaxial 12 VDC BZL1 E1401 04306 1 Front Bezel Bottom BZL2 E1401 04304 1 Bezel Top BZL3 E1401 04301 1 Front Bezel Left BZL4 E1401 04302 1 Front Bezel Right CBL1 E1401 61611 1 Cable Back Plane Interface CBL2 E1401 61612 1 Cable Power Switch LED CBL3 E1401 61610 1 Cable Fan CR1 1990 1027 1 LED Lamp CVR1 E1401 04110 1 Cover Top F1 2110 0054 1 Fuse 15A 250V NTD FE UL MP2 E1401 04603 1 Air Filter MP94 1400 1502 1 Clip LED Mounting Polypropylene MP95 MP97 E1401 81206 6 Horizontal Rail MP100 MP102 MP98 E1401 00102 1 Chassis MP99 E1401 04602 1 Air Plenum MP103 MP128 E1401 41202 26 Guide PC Extractor MP129 E1401 01211 1 Bracket Fan MP130 E1401 00203 1 Fan Carrier Blank Panel MP131 E1401 00204 1 Rear Panel MP132 E1401 01212 1 Bracket Power Supply Bottom MP133 E1401 01213 1 Bracket Power Supply Top MP134 E1401 01215 1 Bracket Air Filter MP135 MP136 5001 0543 2 Trim Side Front Frame MP137 5021 8411 1 Frame Front MP138 MP141 5041 8801 4 Foot MP142 5041 8802 1 Trim Strip Top MP143 MP146 5041 8821 4 Stand Off Rear Panel MP149 MP150 E1401 81205 2 Mounting Rail Front uses E1401 21200 1 per front rail 1494 0075 2 Tapped Strip mounts inside MP97 and MP102 E1401 21200 2 Tapped Strip Non symmetric mounts inside MP149 MP150 PNL2 PNL4 E1400 00209 3 Filler Panel 1 Slot PNL5 E1400 00203 1 Filler Panel 3 Sl
6. 50 mV 12V 0 60 0 36 50 mV 50 mV 24V 1 2 0 72 150 mV 150 mV 24V 1 2 40 72 150 mV 150 mV 5 2V 0 26 0 156 50 mV 50 mV 2V 0 10 0 10 50 mV 50 mV Per VXI Specification rev 1 4 Maximum Power Output 1100 F Thermal Shutdown 1000 FH May Occur TOTAL 900F OUTPUT POWER 800 w 700r 600 F Thermal Shutdown Will Not Occur 500 F 0 L L 0 25 40 55 AMBIENT TEMPERATURE C Uncertainty in shutdown is due to variable loading Heavy loading on 5V 2V and 5 2V will cause shutdown at the lower edge of the band Heavy loading of the 12V 12V 24V or 24V will cause shutdown at the higher edge of the band Low line voltage 90Vac will cause shutdown at the lower edge of the band 26 Agilent E1401B Specifications Appendix A Humidity Temperature Range Shock and Vibration Cooling Provided Acoustical Noise Classification EMC Testing Module Weight Safety Appendix A Exceeds MIL T 28800D test limit 3 7 2 40 C and 95 RH Exceeds MIL T 28800D Test Limits 3 7 2 Non operating 40 C to 75 C Operating 0 C to 55 C Meets the following MIL T 28800D Test Limits Vibration test limits 3 7 4 1 and 4 5 5 3 1 Shock Test Limits 3 7 5 1 and 4 5 5 4 1 Bench handling test limits 3 7 5 3 and 4 5 5 4 3 o Minimum Airflow per Slot AP Across Module mm 40 F 0 5 High Fan Speed Airflow liters sec Fan set to Var
7. Angle Connector Denmark 8120 5398 250 VAC 10A Right Angle Connector Japan 8120 5400 125 VAC 15A Right Angle Connector Agilent supplied Power Cords have polarities matched to the power input socket on the instrument L Line or Active Conductor also called live or hot N Neutral or identified conductor E Earth or safety Ground CSA Certification includes only this power cord Note WARNING AVERTISSMENT 16 Using the Mainframe These are special high power cords Make certain that any replacement cord is capable of handling the indicated loads For protection from electric shock hazard power cord ground must not be defeated The front power switch on this instrument does not disconnect all power from internal circuits In case of emergency the operator may need to disconnect the power to this mainframe Do not block the operator s access to the power cord or if installed ina rack to the circuit breaker which supplies power to the mainframe If a remote standby switch is connected to the mainframe by way of the rear panel 25 pin Sub D connector then the operator may not be able to turn this unit off by using the front panel switch Notify the operator of the location of the disconnect device either the power cord or a circuit breaker for this mainframe Pour ne pas compromettre la protection contre le choc lectrique ne pas couper le circuit de continuit des masses de la fiche L
8. FY AND ARE IN ADDITION TO THE MANDATORY STATUTORY RIGHTS APPLICABLE TO THE SALE OF THIS PRODUCT TO YOU U S Government Restricted Rights The Software and Documentation have been developed entirely at private expense They are delivered and licensed as commercial computer software as defined in DFARS 252 227 7013 Oct 1988 DFARS 252 211 7015 May 1991 or DFARS 252 227 7014 Jun 1995 as a commercial item as defined in FAR 2 101 a or as Restricted computer software as defined in FAR 52 227 19 Jun 1987 or any equivalent agency regulation or contract clause whichever is applicable You have only those rights provided for such Software and Documentation by the applicable FAR or DFARS clause or the Agilent standard software agreement for the product involved IEC Measurement Category Il Overvoltage Protection This is a measurement Category II product designed for measurements at voltages up to 300V from earth including measurements of voltages at typical mains socket outlets The product should not be used to make voltage measurements on a fixed electrical installation including building wiring circuit breakers or service panels E1401B High Power Mainframe User s Guide and Service Manual Er Agilent Technologies Edition 3 Rev 2 Copyright 1997 2006 Agilent Technologies Inc All Rights Reserved Documentation History All Editions and Updates of this manual and their creation date are listed below The first E
9. Mainframe 11 Rear Panel Auxiliary Connector A 25 pin D connector is provided on the rear panel of the mainframe to allow monitoring of power supply voltages and other functions The pinouts for this connector are shown in the following table Description REM STDBY 5 D en E amp E ET EN 5 EA ELA E 18 2 5 25 SYSRESET O D 2 AC FAIL N C 12 Using the Mainframe 5V STDBY s emer 5V STDBY 24V for voltage monitoring 2 4mA max 2V for voltage monitoring 2 4mA max Remote Power Switch 5V output for charging batteries running external TTL circuits etc 1A maximum 12V output for charging batteries running external TTL circuits relays etc 1A max Input for 5V standby current e g from external battery Max 1A combined total pins 8 and 21 Chassis Ground Backplane SYSRESET signal input or output If you use this pin be careful not to violate backplane electrical specifications i e keep extender cable as short as possible An analog output signal proportional to the power supply exhaust temperature Not Used Not Used 12V for voltage monitoring 2 4mA max 24V for voltage monitoring 2 4mA max 5 2V for voltage monitoring 2 4mA max Chassis Ground Remote Power Switch Return Chassis Ground Chassis Ground Input for 5V standby current e g from external battery Max 1A combined total pi
10. ary Connector ene rennen entree 12 Remote Standby Switch essere nnne 14 Installing the Power Line Fuse esee enne nene ren nnne 15 Power Cords dae egeo et ein ceto arie diens 16 Agilent E1401B Safety Grounding Information eee 17 Grounding Procedure muii e este rete e S 17 Chapter 2 Replaceable Parts and Assemblies ee eee ee eee eene ee ee en see etta sese tease s etta sese teens 19 Module Replacement cuado energie 19 Component SA rib Lor paii Ee bete 19 Cleaning Fan Filters tent Reb C C a ee 19 Agilent E1401B Replaceable Parts List esses 20 Appendix A Agilent E1401B Specifications ocmmoccrmscrcssss ta asse eaa 25 j M 29 Contents 1 Notes 2 Contents Certification Agilent Technologies Inc certifies that this product met its published specifications at the time of shipment from the factory Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology formerly National Bureau of Standards to the extent allowed by that organization s calibration facility and to the calibration facilities of other International Standards Organization members AGILENT TECHNOLOGIES WARRANTY STATEMENT PRODUCT El401B DURATION OF WARRANTY 1 year 1 Agilent warrants Agilent hardware accessories and su
11. dition of the manual is Edition 1 The Edition number increments by 1 whenever the manual is revised Updates which are issued between Editions contain replacement pages to correct or add additional information to the current Edition of the manual Whenever a new Edition is created it will contain all of the Update information for the previous Edition Each new Edition or Update also includes a revised copy of this documentation history page Edition 1 is is January 1995 Edition Vie Mee a res li M Pes August 1995 Edition 3 E1401 90009 0 March 1997 Edition 3 Rev 2 E1401 90009 May 2006 Trademarks Microsoft is a U S registered trademark of Microsoft Corporation Windows NT is a U S registered trademark of Microsoft Corporation Windows and MS Windows are U S registered trademarks of Microsoft Corporation Safety Symbols Instruction manual symbol affixed to product Indicates that the user must refer to Ny Alternating current AC the manual for specific WARNING or CAUTION information to avoid personal A Di D injury or damage to the product u nes irect current DC Indicates hazardous voltages Indicates the field wiring terminal that must ES be connected to earth ground before operating the equipment protects against electrical shock in case of fault Calls attention to a procedure practice or condition that could cause bodily injury or death WARNING Calls attenti
12. e VXI Hi Power Mainframe 13 Slot Model Number E1401B Product Options This declaration covers all options of the above product s complies with the essential requirements of the following applicable European Directives and carries the CE marking accordingly Low Voltage Directive 73 23 EEC amended by 93 68 EEC EMC Directive 89 336 EEC amended by 93 68 EEC and conforms with the following product standards EMC Standard Limit IEC 61326 1 1997 A1 1998 EN 61326 1 1997 A1 1998 CISPR 11 1990 EN 55011 1991 Group 1 Class A IEC 61000 4 2 1995 A1 1998 EN 61000 4 2 1995 4kV CD 8kV AD IEC 61000 4 3 1995 EN 61000 4 3 1995 3 V m 80 1000 MHz IEC 61000 4 4 1995 EN 61000 4 4 1995 0 5 kV signal lines 1 kV power lines IEC 61000 4 5 1995 EN 61000 4 5 1995 0 5 kV line line 1 kV line ground IEC 61000 4 6 1996 EN 61000 4 6 1996 3 V 0 15 80 MHz IEC 61000 4 11 1994 EN 61000 4 11 1994 Dips 30 10 ms 60 100 ms Interrupt gt 95 5000 ms Canada ICES 001 1998 Australia New Zealand AS NZS 2064 1 The product was tested in a typical configuration with Agilent Technologies test systems Safety IEC 61010 1 2001 EN 61010 1 2001 Canada CSA C22 2 No 1010 1 1992 USA UL 61010B 1 2003 Supplementary Information This DoC applies to above listed products placed on the EU market after December 31 2003 A 3 Date Tan Boon Juan Quality Manager For further information please contact your local Agilent Technologies sales o
13. f the 13 slots even if one of the three mainframe circulation fans fails to operate The air passes through the installed modules and escapes out the sides and top of the mainframe If any plug in module exceeds 40 Watts dissipation you must provide a minimum of 20 mm 0 8 in of clearance above the mainframe A fourth fan provides cooling for the power supply This fan is mounted on the side of the mainframe When installing the mainframe in your system ensure that the air inlets and outlets are not obstructed This prevents damage to the mainframe and installed modules due to overheating Clean the air filter regularly Chapter 1 Variable Speed Fan Normally the speed of the four internal circulation fans is determined by the ambient temperature of the air drawn into the mainframe One fan cools the power supply module and three fans cool the VXI modules inserted into the mainframe You can override the temperature controlled speed of the fans by changing the position of the Fan Speed Control Switch on the rear of the mainframe to HIGH This forces all four fans to remain at full speed all of the time The two fan modes are Variable Speed Mode factory default Quiet operation 75 airflow at lt 30 C ambient temperature High speed operation at gt 40 C ambient temperature High Speed Mode switch selectable Full airflow at all times Figure 1 2 Forcing the Fan to High Speed Chapter 1 Using the
14. ffice agent or distributor or Agilent Technologies Deutschland GmbH Herrenberger Stra e 130 D 71034 B blingen Germany Notes Notes Notes Chapter 1 Using the Mainframe Mainframe Description The Agilent E1401B mainframe is designed in full compliance with the VXIbus specification Rev 1 4 and VMEbus system specification Rev C 1 The mainframe contains 13 slots for plug in modules and can be rack mounted in either the forward or reverse position Improved Power Supply Reliability Solid state automatic bus grant sensing to bypass empty slots Variable speed fan cooling Rear mounted 25 pin D connector for auxiliary power supply monitoring Easy replacement of major components most can be replaced from the rear of the mainframe even while the mainframe is mounted in a rack Seven standard VXIbus power supply voltages all overvoltage overcurrent and temperature protected Maximum Usable Power total power supply output power before thermal protection shutdown 900 W 25 C 800W 40 C 650 W 55 C Line Voltage and Input Power Chapter 1 WARNING The Agilent E1401B mainframe includes a power supply which automatically adjusts for standard nominal line voltages of 100 120 VAC and 220 240 VAC and nominal power line frequencies of 50 Hz and 60 Hz The power supply has a power factor corrected input section and thus can be powered from any nominal line input over the ran
15. ge of 100 VAC to 240 VAC The power supply can also be powered from a 400 Hz AC line frequency or DC power line voltages Refer to Appendix A for complete input power specifications Refer to Agilent E1401B Safety Grounding Information on page 17 for complete grounding information when power line frequencies exceed 66Hz Using the Mainframe 9 Air Flow Caution 10 Using the Mainframe The air flow into the plug in module slots in the Agilent E1401B mainframe is significantly more than previous Agilent mainframes The mainframe supplies each of the 13 slots with enough air to cool 60 Watts while maintaining only a 10 C rise across a typically dense module Up to 75 Watts per slot may be dissipated if a 15 C rise is acceptable across a typically dense module The graph of Figure 1 1 shows the air flow volume liters second versus change in pressure across the module mm H O for the mainframe Minimum Airflow per Slot 1 0 ae High Fan Speed mm H O L Qu r oa 2 3 4 d Airflow liters sec Figure 1 1 Minimum Airflow Available Per Slot The mainframe has a fan chamber with three circulation fans The speed of the fans is controlled by the ambient temperature of the air drawn into the mainframe Air from the fan chamber is forced into the compartmentalized plenum at the bottom of the mainframe Although the plenum does contain a compartmentalized structure some air flow is always present in each o
16. iable Quiet operation approximately 75 airflow at lt 30 C ambient temperature and high speed operation at gt 40 C ambient temperature Fan set to HIGH Full airflow all the time For modules with typical density 10 C rise at 60W per slot 15 C rise at 75W per slot Low fan speed 48 dBA sound pressure at bystander position Lpa 48 dB fiktiver Arbeitsplatz normalier Betrieb nach DIN 45635 T 1 High fan speed 56 dBA sound pressure at bystander position Lpa 56 dB fiktiver Arbeitsplatz normalier Betrieb nach DIN 45635 T 1 Classified under MIL T 28800 as Type III Class 5 Style E Color R Within the requirements of MIL STD 461 methods CE01 CE02 relaxed 30 dB below 2 MHz relaxed 10 dB between 2 and 15 MHz CS01 CS02 CS06 REOI relaxed 10 dB REO2 and RSO2 RSO3 Meets requirements of CISPR 11 Level A Maximum 3 5 Kg 7 7 lbs per slot to comply with vibration and shock specifications Heavier modules may be installed if vibration and shock environment is less severe IEC 348 UL 1244 and CSA C222 231 Agilent E1401B Specifications 27 Notes 28 Agilent E1401B Specifications Appendix A Index A Air Flow 10 C Certification 3 Cleaning Fan Filters 19 Cooling Provided 27 D Declaration of Conformity 5 Documentation History 4 F Filters cleaning fan 19 Fuse 25 installing 15 G General Power Supply 25 Grounding Information 17 Grounding Procedure 17 Input Power 9
17. ion about technical and professional services product support and equipment repair and service on the web www keysight com Select your country from the drop down menu at the top Under Zlectronic Test and Measurement click on Services The web page that appears next has contact information specific to your country For more detailed product information go to www keysight com find lt product model gt i e for the M9514A use www keysight com find M9514A Hypertext links to documents on agilent com are no longer active Use this substitution to access PDF files Broken links have the form http cp literature agilent com litweb pdf lt ferature part number Substitute links with this form http literature cdn keysight com litweb pdf terature part number Where ferature part number has the form M9300 90001 pdf For service notes use www keysight com find servicenotes KEYSIGHT TECHNOLOGIES Contents Walk Vir iia in ta el el isn in q dead adidas 3 Safety SyMbolS cita NR amp nee 4 WARNINGS 525 d peo ee A a ae ec EN Gat 4 Declaration of Conformity ootasin rrisnin nennen nennen eren nnne 5 IET dV A RE A A Aa 6 Chapter 1 Using the Mainframe c s0s0sssosssssssnnsnsssnnsnsssssnnsssnnnnsssnsnnsssnsnsnssssnsnssssssnssssssssssssssnnene 9 Mainframe Description cdi 9 Line Voltage and Input POWT sitene n aai 9 At EON tau A A bie Cte Res 10 Narable Speed Fano eret eee ee e e t e as 11 Rear Panel Auxili
18. l shock DO NOT perform procedures involving cover or shield removal unless you are qualified to do so DO NOT operate damaged equipment Whenever it is possible that the safety protection features built into this product have been impaired either through physical damage excessive moisture or any other reason REMOVE POWER and do not use the product until safe operation can be verified by service trained personnel If necessary return the product to an Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained DO NOT service or adjust alone Do not attempt internal service or adjustment unless another person capable of rendering first aid and resuscitation is present DO NOT substitute parts or modify equipment Because of the danger of introducing additional hazards do not install substitute parts or perform any unauthorized modification to the product Return the product to an Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained T i i DECLARATION OF CONFORMITY e Agilent Technologies According to ISO IEC Guide 22 and CEN CENELEC EN 45014 C Manufacturer s Name Agilent Technologies Malaysia Sdn Bhd Manufacturer s Address Bayan Lepas Free Industrial Zone Supplier s Address 11900 Penang Malaysia Declares under sole responsibility that the product as originally delivered Product Name C Siz
19. ns 8 and 21 Chassis Ground Backplane ACFAIL signal input or output If you use this pin be careful not to violate backplane electrical specifications i e keep extender cable as short as possible Chassis Ground Not Used Chapter 1 Chapter 1 5V 12V 5V STDBY SYSRESET ACFAIL EXHAUST_TEMP The 5 volt output lets you power external TTL circuitry if required The maximum current allowed from this supply is 1 0A The 12 volt output lets you power a battery charging circuit to provide a source for the 5V STDB Y input The maximum current allowed from this supply is 1 0A This provides an input to the mainframe backplane for a 5 volt standby power source This may be from a separate battery or power supply The maximum current input allowed is 1 0A This means that all of the standby circuitry in all of the modules mounted in the mainframe can draw a total of 1 0A when active This pin provides an extension of the backplane SYSRESET signal It can be used to monitor for SYSRESET or to send SYSRESET to the backplane Shorting this line to ground inputs a SYSRESET signal to the system If you use any extension cable out of the 25 pin D connector on the mainframe rear panel make certain that you do not violate VXI backplane electrical specifications i e keep the cable as short as possible or buffer the signal etc This pin lets you monitor the ACFAIL signal from the backplane Since thi
20. on to a procedure practice or or Frame s id ground T DRY CAUTION condition that could possibly cause damage to connects Lo ME equipments metal tame equipment or permanent loss of data WARNINGS The following general safety precautions must be observed during all phases of operation service and repair of this product Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design manufacture and intended use of the product Agilent Technologies Inc assumes no liability for the customer s failure to comply with these requirements Ground the equipment For Safety Class 1 equipment equipment having a protective earth terminal an uninterruptible safety earth ground must be provided from the mains power source to the product input wiring terminals or supplied power cable DO NOT operate the product in an explosive atmosphere or in the presence of flammable gases or fumes For continued protection against fire replace the line fuse s only with fuse s of the same voltage and current rating and type DO NOT use repaired fuses or short circuited fuse holders Keep away from live circuits Operating personnel must not remove equipment covers or shields Procedures involving the removal of covers or shields are for use by service trained personnel only Under certain conditions dangerous voltages may exist even with the equipment switched off To avoid dangerous electrica
21. ot PNL6 E1400 00204 1 Filler Panel 7 Slot x1 2110 0565 1 Fuseholder Cap 15A Max for UL E1401 00210 1 EMI Spring Clip 20 Replaceable Parts and Assemblies Chapter 2 DETAIL A MP95 MP97 MP94 MP128 MP103 MP137 O llo i PNL4 PNL2 PNL5 PNL6 1 Mainframe Replaceable Parts Figure 2 21 Replaceable Parts and Assemblies Chapter 2 MP143 MP146 Figure 2 2 Power Supply and Rear Panel Replaceable Parts 22 Replaceable Parts and Assemblies Chapter 2 MP146 Figure 2 3 Fan Assembly and Rear Panel Replaceable Parts 23 Replaceable Parts and Assemblies Chapter 2 Notes 24 Replaceable Parts and Assemblies Chapter 2 Appendix A Agilent E1401B Specifications The Agilent E1401B High Power mainframes are 100 compatible with VXIbus specification revision 1 4 Overall Mainframe Size 426mm 16 8 E 602mm 23 7 EST T D 310mm 12 2 Mainframe Weight Agilent E1401B 22 Kg 48 5 lbs with no modules installed Module Size Thirteen 13 C Size slots The mainframes also accept A or B Size modules using the optional Agilent E1403 Module Carrier Input Power 50 60 Hz Input Power Automatic Voltage Ranging nominal range 10
22. ove the power supply Al PC Backplane Assembly Located Under the Connector Figure 1 4 Location of R93 Surface Mount Resistor 14 Using the Mainframe Chapter 1 Installing the Power Line Fuse The mainframe is shipped with a 15A fast blow 15AF fuse Agilent Part Number 2110 0054 already installed This fuse is suitable for all line voltages Figure 1 5 shows how to replace the fuse if necessary ne ae 9 0 ato on PEO de b Q ofb 0 029 00500 000 CO QUES 090200 Tao 090 0008 0 0500 Fuseholder 0000 050 970 050 Cap quo gt prd S 2 Pd lt gt a Bees P quio Figure 1 5 Replacing the Power Line Fuse WARNING For continued protection against fire hazard replace fuse only with fuse of specified type and current rating AVERTISSMENT Pour ne pas compromettre la protection contra le risque d incendie remplacer le fusible par un fusible de m mes type et courant nominal Chapter 1 Using the Mainframe 15 Power Cords Table 1 1 shows the power cords and their Agilent part numbers Table 1 1 Power Cords for the Agilent E1401B Country Part Number Volts Rated Amps Type U K 8120 5392 250 VAC 10A Right Angle Connector Australia 8120 5393 250 VAC 10A Right Angle Connector Europe 8120 5394 250 VAC 10A Right Angle Connector U S Canada 8120 5395 125 VAC 13A Right Angle Connector Switzerland 8120 5397 250 VAC 10A Right
23. pplies against defects in materials and workmanship for the period specified above If Agilent receives notice of such defects during the warranty period Agilent will at its option either repair or replace products which prove to be defective Replacement products may be either new or like new 2 Agilent warrants that Agilent software will not fail to execute its programming instructions for the period specified above due to defects in material and workmanship when properly installed and used If Agilent receives notice of such defects during the warranty period Agilent will replace software media which does not execute its programming instructions due to such defects 3 Agilent does not warrant that the operation of Agilent products will be interrupted or error free If Agilent is unable within a reasonable time to repair or replace any product to a condition as warranted customer will be entitled to a refund of the purchase price upon prompt return of the product 4 Agilent products may contain remanufactured parts equivalent to new in performance or may have been subject to incidental use 5 The warranty period begins on the date of delivery or on the date of installation if installed by Agilent If customer schedules or delays Agilent installation more than 30 days after delivery warranty begins on the 31st day from delivery 6 Warranty does not apply to defects resulting from a improper or inadequate maintenance or calibration b
24. s pin is connected directly to the backplane do not violate VXI backplane electrical specifications i e keep the cable as short as possible or buffer the signal etc This pin provides an analog output that is proportional to the power supply exhaust temperature Using this circuit you can monitor the actual temperature of the power supply exhaust temperature to receive a warning signal that the mainframe is running under abnormal conditions This output provides OVdc at 0 C with a rise of 10mV per degree centigrade when referenced to chassis ground Using the Mainframe 13 Remote Standby The remote standby switch connection is provided to let you turn the Switch mainframe ON or put it in STANDBY mode remotely These two connections are wired in parallel with the front panel ON STANDB Y switch This means that the front panel switch must be in STANDBY position for the remote switch to function correctly POWER SUPPLY FRONT PANEL m Y Yo x ia PES SUB D CONNECTOR o E1401 F_3_1 Figure 1 3 Remote Standby Switch Wiring If you wish to completely disable the front panel switch you can remove R93 a surface mount zero Ohm resistor on the back of the backplane PCA Figure 1 4 shows the location of this resistor Once R93 is removed the only way to control power is through a switch connected to pins 5 and 18 of the 25 pin D connector on the rear panel of the mainframe To access the interface board rem
25. software interfacing parts or supplies not supplied by Agilent Technologies c unauthorized modification or misuse d operation outside of the published environmental specifications for the product or e improper site preparation or maintenance 7 TO THE EXTENT ALLOWED BY LOCAL LAW THE ABOVE WARRANTIES ARE EXCLUSIVE AND NO OTHER WARRANTY OR CONDITION WHETHER WRITTEN OR ORAL IS EXPRESSED OR IMPLIED AND AGILENT SPECIFICALLY DISCLAIMS ANY IMPLIED WARRANTY OR CONDITIONS OF MERCHANTABILITY SATISFACTORY QUALITY AND FITNESS FOR A PARTICULAR PURPOSE 8 Agilent will be liable for damage to tangible property per incident up to the greater of 300 000 or the actual amount paid for the product that is the subject of the claim and for damages for bodily injury or death to the extent that all such damages are determined by a court of competent jurisdiction to have been directly caused by a defective Agilent product 9 TO THE EXTENT ALLOWED BY LOCAL LAW THE REMEDIES IN THIS WARRANTY STATEMENT ARE CUSTOMER S SOLE AND EXLUSIVE REMEDIES EXCEPT AS INDICATED ABOVE IN NO EVENT WILL AGILENT OR ITS SUPPLIERS BE LIABLE FOR LOSS OF DATA OR FOR DIRECT SPECIAL INCIDENTAL CONSEQUENTIAL INCLUDING LOST PROFIT OR DATA OR OTHER DAMAGE WHETHER BASED IN CONTRACT TORT OR OTHERWISE FOR CONSUMER TRANSACTIONS IN AUSTRALIA AND NEW ZEALAND THE WARRANTY TERMS CONTAINED IN THIS STATEMENT EXCEPT TO THE EXTENT LAWFULLY PERMITTED DO NOT EXCLUDE RESTRICT OR MODI
26. uw Amm mmm Agilent 75000 Series C wu BEN MA BEN Y Ae i bus Agilent E1401B High Power Mainframe AEREA User s Guide and Service Manual Where to Find it Online and Printed Information System installation hardware software VXIbus Configuration Guide Agilent VIC VXI installation software Module configuration and wiring This Manual VXIplug amp play programming VXIplug amp play Online Help VXI plug amp play example programs VXI plug amp play Online Help VXI plug amp play function reference VXIplug amp play Online Help Soft Front Panel information VXIplug amp play Online Help VISA language information Agilent VISA User s Guide Agilent VEE programming information Agilent VEE User s Manual Supplied with Agilent Command Modules Embedded Controllers and VXLink CO Y ea t Agilent Technologies Manual Part Number E1401 90009 Printed in Malaysia E0506 Errata Agilent References in this manual NOTICE This document contains references to Agilent Technologies Agilent s former Test and Measurement business has become Keysight Technologies For more information go to www keysight com About this manual We ve added this manual to the Keysight website in an effort to help you support your product This manual provides the best information we could find It may be incomplete or contain dated information Support for your product You can find informat