1. 2010 Teddington Valid Until 18 July 2020 Middlesex TW11 0JZ Reference No T1147 0006 United Kingdom Descriptive Annex 1 INTRODUCTION This pattern of an automatic continuous totaliser belt weigher designed to weigh large quantities of loose material from bulk to bulk is designated the IPC 14 It comprises awelghing platform aspeed sensor an integrator type IPC 14 2 FUNCTIONAL DESCRIPTION 2 1 Mechanical 2 1 1 Weighing platform The weighing platform is designed to react only to forces perpendicular to the line of the conveyor belt It consists of a fixed support frame static and a live frame dynamic and may include a ramp section bridging the static and dynamic parts The static frame is the main support between the conveyor stringers which in turn supports the dynamic frame including the pair s of load cells The dynamic frame supports the scale idlers and transfers the weight of the material to the load cells As the material travels along the conveyor belt a force is exerted through the suspended idlers to the dynamic frame The movement in the load cell is sensed by its strain gauges when excited by voltage from the electronic integrator and produces a signal proportional to weight which is returned to the integrator 2 1 2 Load cells Any compatible load cell s may be used providing the following conditions are met There is a respective OIML Certificate of Conformity R60 or a test certificate E
2. UK 0126 0084 NWMI MI 006 NATIONAL WEIGHTS AND MEASURES LABORATORY United Kingdom of Great Britain and Northern Ireland Certificate of EC type examination of a measuring instrument Number UK 0126 0084 issued by the Secretary of State for Business Innovation amp Skills Notified Body Number 0126 In accordance with the requirements of the Measuring Instruments Belt Weighers Regulations 2006 SI 2006 1259 which implement in the United Kingdom Council Directive 2004 22 EC this certificate of EC type examination has been issued to Control Systems Technology Pty Ltd Unit 9 41 45 Lorraine Street Peakhurst NSW 2210 Australia in respect of continuous totalising automatic weighing instrument belt weigher designated the IPC 14 as described in the descriptive annex and having the following characteristics Qmax lt dependent upon application Qmin gt 20 Qmax dt gt 0 001 t Imin gt 800 d for class 0 5 gt 400 d for class 1 gt 200 d for class 2 Belt speed 0 1 m s to 15 0 m s Accuracy class gt 05 The necessary data principal characteristics alterations securing functioning etc for identification purposes and conditions when applicable are set out in the descriptive annex to this certificate P Signatory P R Dixon for Chief Executive National Weights amp Measures Laboratory Part of National Measurement office Department for Business Innovation amp Skills Stanton Avenue Date 19 July
3. Document Drawing No IPC 14 Installation Manual 02500470 User Service Manual 02500760 Dimensional drawing 88500571 Assembly drawing 88500571 Parts list 88500571 Electrical schematics 88500580 Block diagram 8850905 Main PCB circuit diagram 88500600 3 3 Software 3 3 1 The software is held in the EPROM and its version number is 092316xx tsk where 092316 is fixed for the legally relevant part and xx is for minor updates xx must be greater than 75 The software version number can be accessed from the 1 Info menu then 2 Equipment 3 3 2 Software download and configuration parameters are protected by a calibration amp set up switch located inside the controller enclosure which can only be accessed with breaking the tamper evident seal see 6 2 4 PERIPHERAL DEVICES AND INTERFACES 4 1 Interfaces Load cell connection Speed sensor RS232 485 Digital VO Analogue I O 5 9 4 2 Peripheral devices The instrument may be connected to any peripheral device that has been issued with a parts or test certificate by a Notified Body responsible for Annex B MI 006 under Directive 2004 22 EC in any Member State and bears the CE marking of conformity to the relevant directives or A peripheral device without a parts or test certificate may be connected under the following conditions it bears the CE marking for conformity to the EMC Directive it is not capable of transmittin
4. N45501 issued for the load cell by a Notified Body responsible for type examination under Directive 2009 23 EC The certificate contains the load cell types and the necessary load cell data required for the manufacturer s declaration of compatibility of modules WELMEC 2 Issue 5 May 2009 No 11 and any particular installation requirements A load cell marked NH is allowed only if humidity testing to EN45501 has been conducted on this load cell The compatibility of the load cells and indicator is established by the manufacturer by means of the compatibility of modules calculation contained in the above WELMEC 2 document at the time of verification or declaration of EC conformity of type The load cell transmission must conform to one of the examples shown in the WELMEC Guide 2 4 Guide for Load cells 2 9 2 1 2 Belt Displacement Transducer The conveyor belt displacement transducer may be shaft driven from a suitable pulley or may involve tags affixed to the end of a pulley instrumented to send pulses each representing a precise amount of belt travel to the integrator A fixed number of pulses are generated for each rotation of the pulley The displacement transducer must be applied in a manner which ensures that slipping does not occur It shall be driven from the clean side of the conveyor belt It shall not be attached to a driven pulley of the conveyor belt A dual redundant displacement transducer with automat
5. estroyed The markings and inscriptions shall fulfil the requirements of Paragraph 9 of Annex I of the Directive 2004 22 EC 6 2 Components that may not be dismantled or adjusted by the user calibration and setup switch load cell junction box EPROM must be secured by either a wire and seal or a tamper evident label and securing mark The securing mark may be either a mark of the manufacturer and or manufacturer s representative or _ an official mark of a verification officer 7 ALTERNATIVES There are currently no alternatives 8 ILLUSTRATIONS Figure 1 IPC 14 integrator Figure 2 Enclosure Figure 3 Rating plate typical example 9 CERTIFICATE HISTORY ISSUE NO DATE DESCRIPTION UK 0126 0084 19 July 2010 Type examination certificate first issued No revisions have been issued 7 9 Figure 1 IPC 14 integrator Figure 2 Enclosure 8 9 Figure 3 Rating plate typical example OCrown Copyright 2010 NATIONAL WEIGHTS AND MEASURES LABORATORY Part of National Measurement Office Department for Business Innovation amp Skills 9 9
6. g any data or instruction into the weighing instrument other than to release a printout checking for correct data transmission or validation it prints weighing results and other data as received from the weighing instrument without any modification or further processing and it complies with the applicable requirements of Directive 2004 22 EC Paragraph 8 1 of Annex I 5 APPROVAL CONDITIONS The certificate is issued subject to the following conditions 5 1 Legends and inscriptions 5 1 1 The following legends are legibly and durably marked on the rating plate securely fixed to the integrator console Figure 3 Manufacturer Serial number Zero setting will have a duration of at least belt revolutions Voltage frequency Certification number Accuracy class Totalisation scale interval Belt speed range Maximum flow rate Qmax Minimum flow rate Qmin Minimum totalised load Zmin Designation of the type s of product to be weighed Max capacity Weigh length Supplementary metrological marking CE mark 6 9 6 LOCATION OF SEALS AND VERIFICATION MARKS 6 1 The CE marking supplementary metrology marking and certificate number are located on the rating plate bracket which is fixed to the side of the integrator console The CE mark shall be impossible to remove without damaging it The rating plate shall be impossible to remove without it being d
7. ic fail over may be used 2 2 Electrical 2 2 1 IPC 14 integrator The IPC 14 integrator Figure 1 is the user interface and data handling terminal It comprises a keypad with 16 keys a back lit LCD display with 4 lines of 40 characters and a circuit board with CPU The load and belt displacement travel signals from the weighing platform and speed sensor are processed to derive rate of material flow and totalisation The primary values of speed and load are available on a diagnostic display while primary derived values of rate and total are available on the main display The display is checked at power on however characters are a 5x7 dot matrix and any display faults are self evident The integrator is protected by a metallic enclosure with lockable front door according to drawing 0979902 rev B Figure 2 The integrator is mains operated and supplies the nominal 10 VDC needed to excite the load cells The displacement transducer tachometer speed sensor is powered from the integrator The integrator is fitted with a non volatile memory FRAM in which all parameters affecting the metrological characteristics of the instrument are stored the FRAM memory retains the data indefinitely without the need for battery backup The integrator is also fitted with a 3V Lithium battery to backup up volatile dynamic data such as the totaliser figures The Integrator is powered from a built in float charged Battery Backed power supply which allows the integrat
8. ler The totaliser can continue to operate and can be read when the alarm is cleared the alarm should be recorded along with Totaliser logging records have been lost when memory space is freed and recording resumes 2 6 5 An error message is displayed and must be acknowledged whenever a significant fault occurs with this information recorded in the fault log The measurement data also shows Significant fault weighing result may not be accurate 3 TECHNICAL DATA 3 1 The instrument has the following technical characteristics Power supply 110 Vac 50 60 Hz 240 Vac 50 60 Hz Totalisation scale interval gt 0 001 t Qmax Dependent upon application Qmin gt 20 Qmax 2min gt 800 d for class 0 5 gt 400 d for class 1 gt 200 d for class 2 4 9 Operating speed 0 1 to 15 m s Weigh length Dependent upon application Load cells excitation voltage 10 Vde Minimum load cell impedance 43 75 O Maximum load cell impedance 1100 Q Minimum dead load voltage 2mV Measuring range minimum voltage at max capacity 3 mV Measuring range maximum voltage at max capacity 50 mV Climatic environment 20 C to 55 C Condensing open Electromagnetic environment El and E2 Accuracy class 0 5 6 wire and screen controller to Load cell cable junction box then 4 wire and screen 3 2 Documentation and drawings Description
9. n only be accessed via the calibration setup switch by authorised personnel They allow viewing and editing the parameters to suit the application The Access Level 0 is the only mode accessible to the end user and allows running totalisations while displaying rate and total weight The keypad also allows the end user to view but not edit the system s parameters 2 6 3 The totaliser is continuously updated and displayed and is also logged in a built in history record space along with the information needed to identify the measurement The system is fitted with alarms out of range flow rate or speed power failure to ensure accurate totalisation results Error messages are displayed when the alarms are set and must be acknowledged by the user the error message the totalisation may not be accurate also appear in the datalog when a critical alarm is detected Each time there is a power outage this is considered to be a critical alarm and the totalisation may not be accurate alarm is displayed 2 6 4 When the memory is full measurement data is automatically overwritten provided it is older than x days with x fixed by parameter 658 which is protected by the calibration and setup switch If the memory is full and all records are not old enough to be overwritten parameter 658 and memory size should be chosen for each application so that this situation does not occur the conveyor should stop and an error message is displayed on the control
10. or to be powered up by the push of a button for the purpose of viewing the totaliser figure in the event of a power outage 2 5 Devices 2 5 1 The instrument is provided with the following devices General totalisation device non resettable 10 digit Main totalisation device resettable 8 digit Semi automatic zero setting device lt 2 Max over a whole number of belt revolution Automatic zero setting device also known as Zero Tracking which operates over a whole number of belt revolution lt 2 Max fitted with interlock preventing zero setting while it is possible for material to feed onto the belt Indication of the flow rate Q Indication of the belt speed Error messages and alarms 3 9 2 6 Operation 2 6 1 Zero and Span calibrations are only available to authorised personnel requiring password protected Access Level greater than 0 and access to the sealed calibration setup switch inside the console These calibrations as well as NTEP zero Semi automatic zero executed in level 3 access and Initial span Menu 3 BeltScale Calibrate Span may need to be performed depending upon operational needs A semi auto zero 3 BeltScale Calibrate Zero can be performed by an end user using the keypad and is to be used to maintain the accuracy of the system as required 2 6 2 Access levels greater than 0 are password protected and used to define the operation of the system They ca