Contents
1. Power Supply SP Battery module Staff call lead Drop sensor Infusomat Space line standard type M686 70 01 05 F04 3891 8838 GB Infusomat Space 1 1 gb Bi BRAUN SHARING EXPERTISE 5 3 Sheet 3 of 4 Technical Safety Check TSC Index d Master to be added to the documentation B BRAUN M686 70 01 05 F04 3891 8838 GB 5 4 SHARING EXPERTISE Infusomat Space 1 1 gb Sheet 4 of 4 Technical Safety Check TSC Index c Master to be added to the documentation Checklist for Technical Safety Check Every 24 Months Unit Power Supply SP Manufacturer B Braun Melsungen AG Z N Observe the Service Manual and the instructions for use of the respective device All measured val ues are to be documented Make exclusive use of calibrated measuring equipment Visual Inspection Electrical Safety Functional Inspection according to IEC EN 60601 1 or VDE 0750 and VDE 0751 O Power Supply SP O Mains voltage ____V AC O Connector locking Cleanliness completeness damage and O Leakage current lt 7 HA _ pA O After connecting the power supply the faults affecting safety damage and Measurement is to be carried out with respective unit displays that it is oper readability of the label service adapter SP between primary ated in mains operation O Connecting line and secondary circuit Cleanliness damage and faults affect O Check of power sup
2. Legend to fig 3 27 ItemDesignation 1 Ribbon cable 2 LC display connection cable lock Fig 3 28 Legend to fig 3 28 ItemDesignation 1 LC display 2 Axle mounting plug 3 Keyboard screws 4 Front side operating unit with keyboard 3 28 5 Open the connector lock Fig 3 27 Item 2 on the key board PCB 6 Pull the LC display ribbon cable Fig 3 27 Item 1 out of the connector 7 Liftthe LC display Fig 3 28 Item 1 out of the operating unit front panel Fig 3 28 Item 4 Note The keyboard screws Fig 3 28 Item 3 must not be released A special procedure is required to install the keyboard so that a uniform pressure point is guaranteed for all keys Infusomat Space 1 0 gb Disassembly Assembly 8 Pull the left Fig 3 29 Item 1 and right hinge cover Fig 3 29 Item 4 off the axle Fig 3 29 Item 6 2 3 4 i ee 9 Press the axle out of the operating unit rear panel Fig 3 Y L 29 Item 7 the bottom inner frame Fig 3 29 Item 5 from the hinge cover spring Fig 3 29 Item 2 the hinge cover cable Fig 3 29 Item 3 and the hinge spring Fig 3 29 Item 8 Fig 3 29 Legend to fig 3 29 ItemDesignation 1 Hinge cover left Hinge cover spring Hinge cover cable Hinge cover right Bottom inner frame Axle Operating unit rear panel 0 N O 0 A UU N Hinge spring
3. Important Preliminary Remarks References to item numbers in figures are shown as follows Fig 1 1 Item 1 In this case Fig 1 1 is the figure number and Item 1 the item number within the figure When the Service Manual is stored as pdf file these references are displayed green Click with the mouse button on a reference to jump to the corresponding source Menu commands are described as Menu File Infusomat Space 1 0 gb List of Abbreviations Infusomat Space 1 0 gb Important Preliminary Remarks Abbreviations which are not generally known but are used in this manual are listed below CAN Controller Area Network CE Communaut Europ enne cs Calibration Step DIN German Industrial Norm EN European Norm ESD Electrostatic Discharge FuP Function Microprocessor IEC International Electrotechnical Commission ISO International Organization for Standardization ISP Infusomat Space ISPS Infusomat Space Silicon ISPP Infusomat Space PVC cs Calibration Step KuP Monitoring Microprocessor LCD Liquid Crystal Display MOS Shortname of the following company MOS Technology Inc Commodore Semiconductor Group PCA Patient Controlled Analgesia PSP Perfusor Space SP Space System SPC SpaceCover SPCC SpaceCover comfort SPCS SpaceCover standard SPCO SpaceCom SPCT SpaceControl SPS SpaceStation TEMP Temperature TS Trouble Shooting Step TSC Technical Safety Check UTS Unit T
4. Power supply Euro HW KK KK KK KK KK KK KK KS 0871 3110A Power supply UK 0 0 0 KK KK KK KK eee 0871 3111A Power supply USA Japan ooooooooooo 0871 3112A Power supply Australia KK KK 0871 3113A Power supply South Africa hh kk 0871 3115A Designation Part No A A specie dare qt 0871 3170 battery charging station Connection cable staff call SP o o 0871 3232 Power supply cable 12V LLuuuuluuuuuuu 0871 3231 for ambulance cars CombiLead SP 12 V ss kalik aa cee eee eee 0871 3133 connection cable pump pump InterfaceLead SP WWW KK KK KK KK KK 0871 3230 interface cable CAN SP Drop sensor S a saxa ad kl aa aka Ga kya ka sd Ala 08713175 PoleClamp SP sa2 syn ki s k kab Ann 0871 3130 The PoleClamp is a holder for one or several Space system pumps Short stand SP sass dis 0871 3135 Infusomat Space 1 1 gb General Infusomat Space 1 0 gb Unit Diagnosis Calibration 2 WARNING WHILE TESTING THE UNIT AND TROUBLE SHOOTING THE OPERA TOR SERVICE TECHNICIAN MUST WORK WITH VOLTAGES OF UP TO 115 230 V AC THESE VOLTAGES MAY CAUSE INJURIES WHICH ARE DANGEROUS TO LIFE AND LIMB THE NATIONAL AND INTERNATIONAL SAFETY REGULATIONS ARE TO BE ADHERED TO Before each disassembly and assembly of a unit subsystem check the connectors plug contacts and connections for corrosion and tight fit These fault types are not described again in the following troubl
5. After eye contact Rinse opened eye for several minutes under running water If symptoms persist consult a doctor After swallowing If symptoms persist consult doctor Suitable extinguishing agents Water haze Foam Fire extinguishing powder Carbon dioxide For safety reasons unsuitable extinguishing agents Water with full jet Special hazards caused by the substance its products of combustion or resulting gases In case of fire the following can be released Carbon monoxide CO Hydrocarbons Protective equipment Do not inhale explosion gases or combustion gases Standard procedure for chemical fires Additional information Cool endangered receptacles with water spray Dispose of fire debris and contaminated fire fighting water in accordance with official regulations Person related safety precautions Not required Measures for environmental protection Do not allow to enter sewers surface or ground water Measures for cleaning collecting Pick up mechanically Dispose of the material collected according to regulations Handling Information for safe handling No special measures required Information about fire and explosion protection No special measures required Storage Requirements to be met by storerooms and receptacles Store in cool dry conditions in well sealed receptacles Information about storage in one common storage facility
6. 3452 1429 Linear poti ISP sa nu kn lA ka l k 3452 1056 Pressure adjustment unit ISPS 3452 1445 spring holder pressure springs adjusting screw adjusting nut bracket bracket locking Pump frame ISP epl W K 3452 1402 incl pump drive motor ISP Pump drive motor ISP 00008 3452 1410 Door bolt ISP si caciones ratita bd 3452 1496 Inner frame ISP sana sun ned eee 3452 1437 Membrane ISP 020000eee eee K 3452 1356 Pressure sensor ISP XAW KRA KS 3452 1372 Air inline sensor ISP 000000 000 3452 1380 Safety clamp ISPS WWW KK KK KK 3452 1399 Front frame ISPS cpl h3 3452 1364 Operating unit ISPS cpl 3452 1470 LC display SP ka i sas ax sk oniinn nala d ka aa 3452 0988 Metal front sheet ISPS 2205 3452 1488 Axle ISP ais Eee Dak 3452 1461 incl hinge covers and hinge spring Processor PCB ISP 000000200 eee 3452 1348 incl connectors Bottom inner frame ISP ooo 3452 1453 9 3 Spare Parts List Infusomat Space 1 0 gb Safety Data Sheets 10 Kl ber POLYLUB GLY 501 y LUBRICATION Page 1 4 Material Safety Data Sheet According to 91 155 EEC Printing date 26 01 2005 Reviewed on 29 03 2004 Product details Trade name POLYLUB GLY 501 Article number 020285 Application of the substance the preparation Grease Manufacturer Supplier KL BER LUBRICATION M NCHEN KG Gei
7. 3477 4386 Emergency release plug ISP 10 St ck 3477 3101 Connector seal P2 50 St ck 3477 3102 Housing foot SP 20 St ck On request Screw kit Infusomat Space 3452 1585 with screw DELTA PT 22x8 WN 5451 TORX plus 6IP 5 pieces screw DELTA PT 25x7 WN 5451 TORX plus 8IP 5 pieces screw DELTA PT 30x8 WN 5452 TORX plus 10IP 10 pieces screw DELTA PT 30x12 WN 5452 A2 TORX plus 10IP 10 pieces countersunk screw DELTA PT 20x9 WN 5454 TORX plus GIP 5 pieces fillister head screw M3x6 TORX fillister head screw M3x10 countersunk screw M3x5 TORX countersunk screw M3x12 A2 TORX hexagon nut M3 A2 toothed locked washer M3 A2 distance sleeve Infusomat Space 1 1 gb 9 1 Spare Parts List 7 l 2 Fig 9 1 Exploded drawing Infusomat Space Infusomat Space 1 1 gb Infusomat Space 1 1 gb 99 N DO 0 A 12 13 14 15 16 17 18 19 20 21 22 23 24 25 Spare Parts List Slide guide compl 0 kk 3452 1330 Cover drop sensor connector 3452 1577 Battery compartment cover ISP cpl 3452 1321 incl emergency release crank Battery pack SP NIMH 008 0871 3180 Housing upper part ISP 00 3452 1313 Loudspeaker SP cece eee eee eee 3452 0937 Housing bottom part ISP 0 3452 1305 Rating plate ISP LAL KK KK Upon request Door bolt drive ISP J
8. Infusomat Space 1 0 gb 3 29 Disassembly Assembly 3 14 Processor PCB Designation Ord No z 5 Processor PCB ISP JAVA KK KK KK KK KK KK KK 3452 1348 incl connectors Bottom inner frame ISP 222000000 3452 1453 Screws Fig 3 30 Legend to fig 3 30 ItemDesignation 1 o Oo A W N Hexagon nut M3 Toothed locked washer M3 Processor PCB Bottom inner frame Fillister head screw M3x10 EMC shield Countersunk screw M3x12 A2 TORX not depicted Distance sleeve 3 30 see Service Parts and Screw Kit p 3 8 Note Please pay attention to the corresponding notes during assembly and installation see Processor PCB amp p 3 32 Disassembly 1 Loosen two hexagon nuts Fig 3 30 Item 1 and remove the nuts with the toothed locked washers Fig 3 30 Item 2 the distance sleeve Fig 3 30 Item 8 and the countersunk screw Remove the processor PCB Fig 3 30 Item 3 from the bottom inner frame Loosen one hexagon nut and remove the nut together with the EMC shield Fig 3 30 Item 6 and the screw Fig 3 30 Item 5 from the bottom inner frame Fig 3 30 Item 4 Infusomat Space 1 1 gb 3 15 Assembly Installation Infusomat Space 1 0 gb Disassembly Assembly Assembly or installation of the modules and subsystems is done in reverse order of disassembly Special steps to be observed are de scribed hereafter
9. Infusomat Space 1 1 gb Unit Diagnosis Calibration The Most Important Error Modes The following list specifies the most important error modes and their clearance Note The device must be checked after every repair or service see De vice Check p 2 9 1 The battery module discharges too fast The device was not used for a longer time O Discharge and charge battery module The battery module was not discharged several times and charged at regular intervals O Replace battery module 2 Pump drive runs continuously Light barrier defective O Exchange processor PCB 3 Noise during opening or closing the oper Toothed wheels in door bolt drive defec O Replace door bolt drive ating unit tive 4 Device alarm 2027 Air inline sensor defective 1 Replace air inline sensor air inline sensor optimized from se rial number 3050 on 5 Device alarms 2021 2022 Upstream pressure sensor defective O Replace upstream pressure sensor 6 Device alarms 2023 2024 Downstream sensor defective O Replace downstream pressure sensor 7 Device alarm 2119 Manufacturing tolerance of the LC dis IJ Exchange ISP processor PCB play too high optimized from serial number 3050 on Table 2 4 2 8 Infusomat Space 1 0 gb Device Check Unit Diagnosis Calibration 1 The device is inserted in a SpaceStation or connect UTS 2
10. Infusomat Space S Pressure Levels Tests v Disconnect a Calibration Cal update Serial number Disposable articles Modification data Operating data Delivery Constant 2006 10 09 00 00 00 1823 Name Addition Space Line v 1 00 Space Line Neut Y 1 00 Intrafix PVC v 1 00 State on active off Inactive ok From device To device Save as Fig 3 42 12 Press the load from button The window Open is displayed 7 on screen nen Suchenin C Disposables 13 Select the desired file with the mouse pointer and press the 2 ER Open button The data loaded is displayed on screen Recent 14 Actuate the to device button The data is saved in the Infu bes somat Space 15 Exit the Service Program see Quit the Service Program Eigene Dateien p 1 _ 15 Arbeitsplatz Netzwerkumgeb Dateiname ISPS 1823 180107 TUT ffnen ung Dateityp Tube Table Y Abbrechen Fig 3 43 Infusomat Space 1 1 gb 3 41 Disassembly Assembly HiBaSeD International Service Program File Extras Tools Update Help Extras Tools Update Help ue aie a ss com3 686E030002 2006 10 09 00 00 00 Fig 3 44 3 42 Infusomat Space 1 1 gb 3 16 Checks after Repair Infusomat Space 1 0 gb Disassembly Assembly Procedure 1 Remove a slide clamp which may be inserted in the safety clamp and open safety clamp press
11. Store away from foodstuffs Store away from oxidizing agents Further information about storage conditions None GB Contd on page 3 Safety data sheet Kl ber POLYLUB GLY 501 Part 2 of 5 10 2 Infusomat Space 1 1 gb Safety Data Sheets Page 3 4 Material Safety Data Sheet According to 91 155 EEC Printing date 26 01 2005 Reviewed on 29 03 2004 Trade name POLYLUB GLY 501 Contd of page 2 Additional information about design of technical facilities No further data see item 7 Ingredients with limit values that require monitoring at the workplace The product does not contain any relevant quantities of materials with critical values that have to be monitored at the workplace Additional information The lists valid during the making were used as basis Personal protective equipment General protective and hygienic measures Immediately remove all soiled and contaminated clothing Avoid close or long term contact with the skin Be sure to clean skin thoroughly after work and before breaks Respiratory protection Not required Protection of hands Preventive skin protection by use of skin protecting agents is recommended Eye protection Not required General Information Form Pasty Colour Beige Odour Product specific Change in condition Drip point gt 250 C DIN ISO 2176 Flash point Not applicable Danger of explosion Product does not pr
12. played for a short moment on the screen The work window of the Service Program appears on the screen All devices recognized are listed in the left column 1 11 System Overview HiBaSeD International Service Program File Extras Tools Update Help eed infusomat Space S Fig 1 11 Select Devices To handle a specific device select one from the device tree on the left side doubleklick Note HiBaSeD must not service several devices at the same time Thus you have to disconnect from the device and select a different one T 12 5 Activate the desired device from the list on the left in the work window with a double click The device data is then dis played below the device name Infusomat Space 1 1 gb Device compatibility Version HiBaSeD 1 1 4 1 21 1 3 5 137 15 0 2 0 1 3 1 0 Fig 1 12 Infusomat Space S Ver 686E030002 Please use HiBaSeD Ver current Perfusor Infusomat Cover Infusomat A 6876030001 6876030002 6876030003 6878030004 6870030001 68 D 030001 6878030007 687803000 687803000 687803000 B8 CO3000 68 D03000 687E030001 EN Infusomat type selection Infusomat type Please select the desired Infusomat type for the current calibration procedure Infusomat Space P Article number 8713070 Fig 1 13 DET Infusomat Space S Article number 8713050 Cancel Inf
13. 1 1 gb Fig 3 34 Infusomat Space 1 0 gb Disassembly Assembly Note Always use a new membrane for assembly Use Meliseptol to fa cilitate installation of the membrane over the pushers 5 Engage the front frame lugs arrow Fig 3 33 exactly in the inner frame openings Note When mounting the inner frame and the housing front panel make sure that the operating unit is opened 6 During installation of the complete housing front panel on the bottom inner frame the positioning hemispheres arrow Fig 3 34 of the inner frame must be adjusted in the slots of the bottom inner frame 3 35 Disassembly Assembly Fig 3 35 3 36 7 The housing front panel is pushed on to the bottom inner frame until stop and aligned in parallel arrow Fig 3 35 Snap in Hook 1 The snap in hook is to be inserted before mounting the hous ing bottom part Housing Bottom Part 1 Check the correct position of the seal relative to the safety clamp before mounting the housing bottom part 2 Make sure that the hinge spring with the housing bottom part is pressed in the spring holder of the housing front panel CAUTION Please pay attention to the various screw lengths when you mount the bottom part Housing Upper Part 1 If you have carried out any activities on the pressure adjust ment unit or dismounted or replaced the pressure adjust ment unit r
14. 3 HH HiBaSeD display DER B Braun Melsungen AG Infusomat S Adjust Data 3 Program Name HiBaSeD Program Version 3 1 0 15 01 2007 14 51 02 0001 Serial Number 1823 3 Recent Ei Desktop Ge Eigene Dateien Arbeitsplatz Fig 3 4 Speichern unter Save in Adjust C ISPS 1823 241106 INI B ISPS 1823 241106 INI CHK Netzwerkumgeb File name isPs_1 823 180107 ut Save as type 7 5 Press the Print button to transmit the device data to a printer 6 Actuate the Display button to display the device data on screen With Close the window is closed again 7 Select the storage position in the Save file as window and input a unique file name 8 Pressthe Save button The data of the pump is saved on the PC hard disk 9 Select the tab Disposable article Infusomat Space 1 1 gb Disassembly Assembly File Extras Tools Update Help El Meu Infusomat Space S com3 686E030002 2006 10 09 00 00 00 1823 HiBaSeD International Service Program Pressure Levels Tests s Disconnect Calibration Cal update Serial number Disposable articles Modification data Operating data Delivery Constant status Nam Action State active off Inactive Load from E Fig 3 5 Infusomat Space 1 1 gb 10 Actuate the from device button The data is read from the device The data of the d
15. 3 ria Version 1 1 Page 3 42 ccccccccecceeceececeaceucenceces Version 1 1 KU D osi ir a ad Version 1 0 Page 3 43 cria kuna Rd Version 1 0 Chapter 10 Safety Data Sheets Page EL LLL brinco VESIA Page 10 A nehn manini d4 A Version 1 1 IAS E MEON O Page 10 92 id n Version 1 1 PROP AO IA version Ton Page 10 3 said ERE Sd ard Version 1 1 Chapter 4 Servicing the Unit Page TOA yen LR ec 4 4240 ads Version 1 1 Page 4 1 RN rr PIP az versions ee ee eee es ar eer ere ima AAA a ed A Version TA Chapter 11 Revision Documentation O Version 1 1 Page iaa Version 1 1 Page G22 sla wa Pe Da rn k b ha an ideis emi TAOS TES 2i Wak n EEE nere n ra kebab KER Version 1 1 UE PNEU HH H HH Version 1 1 Page 11 Z3 ii daa E a k al d wth k Version 1 1 11 4 Infusomat Space 1 1 gb ES i xwan alba nalan ose cidade geet Version 1 1 Page ll SD una rr id Version 1 1 Chapter 12 Index Page IIAG Db mm pr r EE cado Page 12 22 a ER E E E Version 1 1 Infusomat Space 1 1 gb 11 5 Infusomat Space 1 0 gb A ACCESSONES airada heed needs x eee ex Dk 1 16 PAQUIN seein ck ian decide ad ede 2 4 Assembly kk kk kk kk kk kK KK KK nne 3 1 3 31 B Battery module WWW kK kk KK KK KK KK KK KK KK KK KK 3 12 BlOCK dia dado Whe Wila Ded b 1 8 C Calibration cu nae en gr zz 2 13 Procedure nearen raa de BEA ania ID ER 2 13 Checks after repair kk kK KK KK KK KK KK KK eee 3 43 Cleanilgizs sies may n eed boda id ve
16. Error Codes The Most Important Error Modes Device Check Procedural Instructions for the Device Check Calibration Procedural Instructions for Calibration Further Trouble Shooting General Slide Guide Cover for Drop Sensor Connector Emergency Release Plug Unit Foot Battery Module Housing Upper Part Loudspeaker Housing Bottom Part Inner Frame and Housing Front Panel Inner Frame Housing Front Panel Operating Unit Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page 0 5 0 5 0 5 0 5 0 6 0 6 0 6 0 7 0 7 0 11 0 11 0 11 0 11 0 11 0 11 0 11 1 1 1 1 1 2 1 4 1 9 1 10 1 16 1 16 1 16 2 1 2 3 2 8 2 9 2 12 2 13 2 13 2 19 3 1 3 9 3 9 3 10 3 11 3 12 3 14 3 17 3 17 3 19 3 21 3 25 3 26 0 3 Table of Contents Processor PCB Page 3 30 Assembly Installation Page 3 31 Checks after Repair Page 3 43 Servicing the Unit Cleaning Page 4 1 Servicing the Battery Page 4 1 Technical Safety Check TSC Infusomat Space Page 5 1 Technical Safety Check TSC Power Supply SP Page 6 1 Procedural Instructions on the TSC Visual Inspection Page 7 1 Electrical Safety according to IEC EN 60601 1 or VDE 0750 and VDE 0751 Page 7 2 Functional Inspection Infusomat Space Page 7 3
17. Tools Update Help f Infusomat Space S com3 686E030002 2006 10 09 00 00 00 1823 Calibration Tests y Disconnect Cal update Serial number Disposable articles Modification data Operating data Delivery Constant Data selection OK Cancel Fig 2 10 Connect e Connect Please press the blue connect button on the attached device Fig 2 11 2 16 9 Start calibration with the OK button You are prompted to press the blue connection key on the device 10 Press the blue connection key All device data is read out and stored in the PC Infusomat Space 1 1 gb Close clamp Close clamp Close locking clamp Please ensure that the Infusomat locking clamp is closed You may close it manually Fig 2 12 Status of lock bolt calibration 2 Is lock bolt open 1 Fig 2 13 Status of lock bolt calibration Is lock bolt closed Fig 2 14 HiBaSeD Process completed switch machine off Fig 2 15 Infusomat Space 1 1 gb 11 12 13 14 15 16 17 Unit Diagnosis Calibration If the locking clamp clamping lever was opened you are asked to close the locking clamp Close the locking clamp with the slide clamp The Calibrate locking clamp frame is activated and the locking clamp is moved to the open posi tion Check the posi
18. Uncleaned packaging Recommendation Empty contaminated packagings thoroughly They may be recycled after thorough and proper cleaning Land transport ADR RID cross border ADR RID class Maritime transport IMDG IMDG Class Air transport ICAO TI and IATA DGR ICAO IATA Class Transport Additional information Not classified as dangerous according to the above specifications Labelling according to EU guidelines The product is not subject to classification according to the calculation methods of the General Classification Guideline for Preparations of the EU as issued in the latest valid version This information is based on our present knowledge However this shall not constitute a guarantee for any specific product features and shall not establish a legally valid contractual relationship Relevant R phrases 53 May cause long term adverse effects in the aquatic environment Department issuing MSDS Material Compliance Management Contact Tel 49 0 897876 564 GI o Safety data sheet Kliiber POLYLUB GLY 501 Part 4 of 5 10 4 Infusomat Space 1 1 gb Safety Data Sheets a Printing date 26 01 2005 Reviewed on 29 03 2004 Product details Trade name POLYLUB GLY 501 Article number 020285 Application of the substance the preparation Grease Manufacturer Supplier KL BER LUBRICATION M NCHEN KG Geisenhausenerstrasse 7 D 8
19. and engage operating le ver When activities according to chapter 3 2 to 3 9 were carried out a visual inspection and a self test are necessary Check the device chapter 3 10 to 3 14 to ensure safe func tionality of the unit see Device Check s p 2 9 Check the flow accuracy according to the TSC if activities of chapter 3 11 were carried out see Technical Safety Check TSC p 5 1 Depending on the work carried out the specific steps of the TSC must be performed see Technical Safety Check TSC p 5 1 3 43 Disassembly Assembly 3 44 Infusomat Space 1 0 gb Disassembly Assembly Check List for Checks after Repair Visual Inspection Electrical Safety according to IEC EN 60601 1 or VDE 0750 and VDE 0751 Functional Inspection O Cleanliness C Completeness O Damage and faults affecting safety C Damage to and readability of the label O Screw covers O Connectors P2 and P3 O Pressure sensor downstream membrane O Pressure sensor upstream membrane O Pump membrane O Pump slide all the 12 slides are engaged O Coding of pump slide guide O Pump connecting rod O Airinline sensor free from grease or ultrasonic gel free from cracks O Drop sensor lens The patient and housing leakage current of the Infusomat Space is caused exclusively by the operating voltage supply Power Sup ply SP or SpaceStation The Technical Safety Checks of
20. change depending on the software version The following screen illustrations are only examples and represent the state when the manual was printed 1 Start the Service Program see Starting the Service Program p 1 10 2 Select the register tab Modification data 3 37 Disassembly Assembly HiBaSeD International Service Program File Extras Tools Update Help Ve Infusomat Space S Pressure Levels Tests Disconnect Calibration Cal update Serial number Disposable articles Modification data Operating data Delivery Constant com3 686E030002 2006 10 09 00 00 00 10 1823 Select a data source to modify From device Default Accept or reject modified data Eo zou Accepted data must be saved To device To file Show Print Fig 3 36 3 Press the from file button The window Open is displayed on screen 3 38 Infusomat Space 1 1 gb Offnen Look in Adjust J amp ei El BE JISPS_1823_180107 INI MO 1SPS_1823_241106 INI Recent Eigene Dateien Arbeitsplatz isPs 1623 1801071NI z INI File z Netzwerkumgeb File name ung Ce EEEN Files of type Disassembly Assembly 4 Select the desired file with the mouse pointer and press the Open button 5 Press the to device button HiBaSeD International Service Program File Extras Tools Update Help f Infusomat Space S Pressure Leve
21. compartment and tilting the upper part forward 4 Remove the housing upper part vertically to the top Note Pay attention to the length of the connection cable to the battery compartment and to the connectors P2 and P3 when dismounting the housing Do not pull the connectors together with the upper part out of the bottom part Pay special attention to the loud speaker and the length of the connection cable to the battery compartment 3 15 Disassembly Assembly Fig 3 16 Legend to fig 3 16 ItemDesignation 1 Housing upper part 2 Screw DELTA PT 25x7 WN 5451 TORX plus 8IP 3 Contact strip 3 16 5 Unscrew two screws and remove the contact strip Fig 3 16 Item 3 to the battery compartment together with the connection cable out of the housing upper part Fig 3 16 Item 1 Note If necessary the spring mounted contact pins must be carefully pressed in when the contact strip is dismounted Infusomat Space 1 0 gb 3 8 Loudspeaker Fig 3 17 Legend to fig 3 17 ItemDesignation 1 Loudspeaker 3 9 Housing Bottom Part Infusomat Space 1 0 gb Disassembly Assembly Designation Ord No Loudspeaker SP i225 cio cidade 3452 0937 Disassembly 1 Pull the loudspeaker connector off Fig 3 17 Item 1 the processor PCB and remove the loudspeaker Designation Ord No Housing bottom part ISP A K KK 3452 1305 Rating plate ISP 2 KK KK KK
22. data sheets are to be observed see Safety Data Sheets m p 10 1 Plastic guides Fig 3 32 Item 1 of the door bolt Spring bearings and dovetail of the spring holder Fig 3 32 Item 2 Guide 6 mm straight pin of the pump frame at the bear ings Fig 3 32 Item 3 Toothed wheels Fig 3 32 Item 4 of the pump drive Movement areas Fig 3 32 Item 5 of the pump frame Toothed wheels Fig 3 32 Item 6 of the door bolt drive Guides Fig 3 32 Item 7 and supporting area of the door bolt Pay attention to the operating pin of the linear potentiometer when installing the door bolt drive The pin must engage in the fork for the potentiometer when the spindle tappet is clipped in Check all toothed wheels for damage after installation of the door bolt drive The spring holder of the pressure adjustment unit must run smoothly with its dovetail guide in the guide of the pump frame Check smooth running without springs inserted before final assembly Inner Frame and Housing Front Panel 1 Move the door bolt by hand to the left until stop open posi tion before mounting the housing front panel Close the safety clamp with a slide clamp after installation Push the membrane carefully over the pump drive pusher Route the cable of the door bolt drive in the cable guide on the door bolt drive and between the two cable guide pins of the front frame Infusomat Space
23. gb System Overview 1 The Infusomat Space ISP is according to IEC EN 60601 1 or IEC EN 60601 2 24 a portable volumetric infusion pump for infusion of small to high volumes with ultimate precision and is suitable for intravenous applications blood transfusion and enteral nutri tion The medical specialist must decide on suitability for application on the basis of the warranted properties and the technical data The Space system is a modular design of modern infusion technol ogy for stationary mobile or private use The key modules and their connection to the peripheral devices are shown in Fig 1 1 All the pump types Perfusor Space Infusomat Space and Infu somat Space P as well as the other devices of the system are of modular design Up to three pumps can be connected together us ing L rails on the bottom of the unit and grooves on the top They can then be fastened to a drip stand or appropriate rail using the pole clamp The SpaceControl module can be used to extend operation One single pump can be inserted onto this module The pump is then connected via connectors to the module The SpaceStation module allows the set up of a complete pump system with up to 24 pumps Up to four pumps can be installed in every SpaceStation The pumps are supplied with power via the integrated power supply and the built in connectors The pumps are connected to the optional SpaceCom via these connectors SpaceControl can also b
24. in detail Only new cover caps are to be used Special Screws Special screws for plastic housings are used in this unit The screws are not self cutting but produce a thread in the plastic of the housing through deformation when fitted in for the first time If the beginning of the thread is not engaged when the screw is fitted a new thread is produced and the old thread is destroyed so that the security of the fixing can no longer be guaranteed Proceed as follows to fit the special screws 1 Put the screw on the thread 2 Rotate screw anti clockwise loosen until a faint click can be heard This click is produced when the screw thread drops into the existing thread 3 Screw in screw and tighten carefully Zero Force Insertion Connector Note Make sure that the ribbon cable is centered between the plug guides when the zero force insertion connector is locked Check the lock and the ribbon cable for correct fit before any further in stallation 3 31 Disassembly Assembly Processor PCB 1 The adjusting lug of the bottom inner frame must be posi tioned in the processor PCB boring When the processor PCB is replaced all data of the pump except for the calibration data was probably saved on a PC see Prepa rations for Exchanging the Processor PCB sm p 3 2 Carry out the following steps to transmit the data back to the device see Rework when Replacing the Processor PCB sm p 3 37 Op
25. in the manuals are ob served the manuals are a reference for measurements depending on the unit type the Service Program must be called which may lead to a dangerous unit condition in case of inappropriate operation Furthermore a special service connector may be necessary This manual version corresponds to the state when the manual was written B Braun reserves the right to make technical modifi cations The state of the revision is indicated by the index number in the footer of every page The possession of this manual does not automatically mean inclu sion in the revision service You will be included in the revision service after technical training by B Braun Melsungen or a written order placed with the sales department of B Braun fee required Important Preliminary Remarks Responsibility of the Manufacturer Quality Management Checks and Repair Notes on ESD The manufacturer person who assembles installs or imports the device can only be held responsible for safety reliability and per formance if mounting enhancements new settings changes or repairs are carried out by duly authorized persons the electrical installation in the corresponding room meets the requirements of the VDE 0107 VDE 0100 part 710 or IEC 60364 7 710 and the national standards the device is used in accordance with the instructions for use and the Service Manual the Technical Safety Checks are performed at reg
26. the device settings the red LED on service connector SP lights up permanently or for one second af ter an alarm was triggered Status display Battery capacity mains operation Lighting LC display lighting keyboard lighting 2 Operation a b c Open operating unit Yellow warning light beside the safety clamp flashes Insert Infusomat Space line Slide clamp disengages release lever of the safety clamp The Infusomat Space line is squeezed Close operating unit Carry out infusion with intermediate bolus and press all buttons at least once Procedural Instructions on the TSC 8 N N 9 Fig 7 2 Legend to fig 7 2 ItemDesignation 1 ao 4 CO N Electronic pressure gauge 10 ml syringe piston fixed Infusion bottle 1000ml Infusomat Space line standard type Distance between liquid level and middle of the Infusomat Space line Three way valve 1 Three way valve 2 Graduated cylinder Infusion and bolus are performed and all buttons trigger the function desired e Staff call Plug service connector Space on connector P2 Lock roller clamp of the Infusomat Space line The red LED on the service connector Space lights up because of a pressure alarm in the supply line f Open operating unit Slide clamp of the Infusomat Space line is completely pulled on to the line g Operating lever of the safety clamp can be pressed down easily against spring te
27. the internal voltages required through a voltage transforming and monitoring circuit on the processor PCB An independent circuit in the battery module monitors the battery cells and controls their charge condition The keyboard and the display are lighted The Infusomat Space is connected to a SpaceControl by connec tor P3 The function processor controls all the functions of the Infuso mat Space Data is stored in a non volatile memory which also controls the external data transfer The monitoring processor monitors all important responses of the function processor to incoming information If a response does not correspond with that expected by the monitoring processor an error message is generated and the device is switched to a safe stop state The Infusomat Space line and its contents is monitored by the sensors listed below optional drop sensor pressure sensor upstream container side monitors the upstream pressure negative pressure airinline sensor monitors the line contents liquid air temperature sensor monitors the temperature of the medium in the line inte grated in the air inline sensor pressure sensor downstream patient side monitors the downstream pressure overpressure The pump drive motor is monitored by a detector for speed and direction of rotation The motor of the door bolt drive is controlled via the processor PCB The different positions of the door bolt during operation are mo
28. the power supply SP drawing No MOO1 32 10 05 F04 or of the SpaceStation drawing No M690 00 00 46 F04 serve to check whether both limit values are met O Locking with second unit O Battery compartment cover Switch on unit with power supply Indicator lamps LEDs Self test Audible alarm Visual alarm g g g g O Status display g Lighting Operation O Opening and closing mechanism of the operating unit O Slide clamp releases release lever of the safety clamp The Infusomat Space line is squeezed Buttons on the operating unit Infusion Staff call Bolus Slide clamp is completely pulled on to the Infusomat Space line when the operating unit is opened The operating lever of the safety clamp can be pressed down uniformly against spring tension and engages The yellow caution sign flashes Part 1 of 2 Infusomat Space 1 0 gb 3 45 Disassembly Assembly Visual Inspection Electrical Safety according to IEC EN 60601 1 or VDE 0750 and VDE 0751 Functional Inspection Electronic pressure cut off according to TSC Infusomat Space line standard type Delivery rate 250 ml h Test without drop sensor O Upstream O Alarm O Downstream O Pressure stage 1 O Pressure stage 5 O Pressure stage 9 Mechanical pressure limitation according to TSC g Pmax g Pmin Safety clamp O Occlusion of Infusomat Space line Dr
29. unlocking crank to the right clockwise according to arrow direction until the operating unit is re leased and opens System Overview LEDs Display MCP memory 16Mbit Dual Bank Flash 4mbit SRAM Keyboard Operating unit Loudspeaker Drop sensor Motor final stage Recogition of direction of rotation Pump drive Motor final stage Position sensor Door bolt drive Pressure sensor Up Stream Down Stream Safety clamp Air sensor Temperature sensor Fig 1 7 Block diagram Infusomat Space Function processor CAN RX TX CAN 9 transceiver 9 Level lt gt gt EA converter EIE 32 0kHz z 12 0 Mhz lt zz EAE Change over RTC iid logic A gt gt Control microprocessor gt 10 0 Mhz gt RX TX Battery module 12VDC Ub_KM hi reis 4 12VDC Akku OUT 12VDC P1 Infusomat Space 1 0 gb Unit Software Infusomat Space 1 1 gb System Overview Approved Software Versions 686B030001 Basic software 686B030002 Improved functions Move door bolt in pumping position just before occlusion test safety clam
30. 1379 M nchen Tel 0049 0 897876 0 Fax 0049 0 897876 333 Notfallauskunft 0049 0 897876 700 Further information obtainable from Material Compliance Management Chemical characterization Description mineral oil Synthetic hydrocarbon oil special lithium soap UV indicator Dangerous components CAS 597 82 0 0 0 O Triphenylthiophosphate R 53 lt 2 5 EINECS 209 909 9 Additional information For the wording of the listed risk phrases refer to section 16 Hazard description Not applicable Information concerning particular hazards for human and environment The product does not have to be labelled due to the calculation procedure of the General Classification guideline for preparations of the EU in the latest valid version Classification system The classification is according to the latest editions of the EU lists and extended by company and literature data Fig 10 1 Safety data sheet Kl ber POLYLUB GLY 501 Part 5 of 5 Infusomat Space 1 1 gb 10 5 Safety Data Sheets For your notes 10 6 Infusomat Space 1 0 gb Revision Documentation 11 Description of Version Version 1 0 Base Version First version of this Service Manual Release date 31 03 06 Version 1 1 AC adapters changed TSC of the AC adapters changed Spare part number of the ISP processor PCB changed Release date 22 03 07 Version List o
31. 7 3 start infusion and deliver first of all in to an open system without pressure gauge Upstream 5 Push roller clamp of the Infusomat Space line up to appr 30 cm before the pump and close the roller clamp or squeeze the Infusomat Space line between the container and 30 cm be fore the pump An alarm is activated 6 Open roller clamp Downstream 7 Switch over three way valve please see Fig 7 4 and deliv er towards the pressure gauge Read off maximum value on the pressure gauge upon an alarm and before an automatic pressure reduction and com pare with the specifications in the TSC 8 Check all pressure stages listed in the TSC and document val ues Pressure Limitation Mechanical The optional drop sensor must not be connected Pmax and Pmin 1 Plug service connector Space on connector P2 so that the electronic pressure cut off is deactivated 2 Input a delivery rate according to the TSC start infusion and deliver first of all in to an open system without pressure gauge position of the three way cocks please see Fig 7 3 3 Switch over three way valve please see Fig 7 4 and deliv er towards the pressure gauge Read off the corresponding value on the pressure gauge and compare with the specifications of the TSC for Pmax and Prin 4 Document values Do not reduce pressure Procedural Instructions on the TSC Safety Clamp 1 Stop infusion mechanically from the pressure li
32. AN in Switch device off and on ure terface 23 2035 Light barrier or motor defect turn Light barrier defective Exchange processor PCB backwards 24 2036 FUP Language Flash Memory Error Language file faulty Update software 25 2037 FUP Language Version Error Language file faulty Update software 26 2038 FUP CRC Task inactive Communication error on CAN in Switch device off and on terface Table 2 2 Device alarms of the function processor Part 1 of 2 Infusomat Space 1 0 gb Unit Diagnosis Calibration 27 2039 Two pumps with the same serial number are installed in one SpaceStation Check and correct the serial num bers Table 2 2 Device alarms of the function processor Part 2 of 2 Device Alarms of the Control Microprocessor 1 2100 Timebase too fast Quartz of the processor PCB Exchange processor PCB 2 2101 Timebase too slow Quartz of the processor PCB Exchange processor PCB 3 2102 Timebase fail Quartz of the processor PCB Exchange processor PCB 4 2103 Keybord High Keyboard defective Carry out device check 5 2104 EA KEY defect 25sec Keyboard defective Carry out device check 6 2105 No keydecode Keyboard defective Carry out device check 7 2106 ROM Romtest defect Software Update unit software 8 2107 ROM Program defect 9 2109 MPU Test failed 10 2110 RAM Test failed 11 2111 KUP active reset Voltage supply during operatio
33. F04 serve to check whether both limit values are met Power Supply Leakage Current Note The values to be measured are indicated in the TSC see Technical Safety Check TSC p 6 1 The leakage current is to be measured between primary and sec ondary circuit using the service adapter SP Designation Ord No Service adapter SP WW kK KK KK KK KK KK KK KK 0770 5174 Check 1 Select leakage current on the testing equipment 2 Bridge pin 6 7 12 and 13 with service adapter SP at the con nector to the unit 3 Measure value 4 Measure value with reverse polarity of the mains connection 5 Document the largest value Infusomat Space 1 0 gb Functional Inspection Infusomat Space Infusomat Space 1 0 gb Procedural Instructions on the TSC Mechanical Inspection 1 Fit the unit to be tested on top of another Space device and check the proper functioning of the lock 2 Fit the unit to be tested under another Space device and check the proper functioning of the lock 3 Check the battery compartment cover lock for proper opera tion Functional Check Note Carry out the check with power supply connected 1 Switch on unit and check the following details Indicator lamps LEDs yellow green blue light up for a short moment Audible alarm A deep and then a high sound Visual alarm Red LED lights up for a short moment Self test Display on the LC display Staff call Depending on
34. Functional Inspection Power Supply SP Page 7 8 Test Equipment and Special Tools Test Equipment Page 8 1 Special Tools Page 8 2 Spare Parts List Page 9 1 Safety Data Sheets Kl ber POLYLUB GLY 501 Page 10 1 Revision Documentation Description of Version Page 11 1 Version List of the Individual Pages Page 11 1 Index Page 12 1 0 4 Infusomat Space 1 0 gb Service Work Technical Safety Checks Current Versions Revision Service Infusomat Space 1 1 gb Important Preliminary Remarks The present manual is for your information only The possession of this manual does not authorize the performance of service work Service tasks may only be executed by persons who have received appropriate training on the system from B Braun are included in the revision service possess the necessary test equipment and mechanical aids and fulfill the personal requirements training and knowledge The user is obliged to perform or to have performed the Technical Safety Checks on those medial products for which these checks have been prescribed by the manufacturer and to carry them out according to the indications of the manufacturer as well as the generally approved technical standards while adhering to the pe riods stated S 6 MP BetreibV B Braun also recommends training on the Technical Safety Checks or to perform at least the steps indicated in the current version of the manual as the TSC requires that the instructions
35. IT Operate the device with battery until the message Battery discharged is dis played and the unit is switched off Then connect the unit to the mains for at least 6 hours Battery module defective or too old Replace battery module 2 Battery discharged type operating alarm The device was not connected to the mains long enough Connect the unit to the mains for at least 6 hours Battery module defective Replace battery module 3 Battery cover open type operating alarm CJ The battery compartment cover is not correctly closed Insert the battery compartment cover correctly O The magnet in the battery compart ment cover is missing Exchange the battery compartment cover J The battery compartment cover is not recognized by the battery mod ule Replace battery module 4 Device alarm type device alarm A serious internal fault was detected in the system Carry out a device check see Device Check p 2 9 Table 2 1 Alarms Infusomat Space 1 0 gb Unit Diagnosis Calibration Device Alarms of the Function Processor 1 2001 2013 Internal Error 2 2014 Loudspeaker not off Loudspeaker connector Check the loudspeaker connector Loudspeaker Check the loudspeaker 3 2015 Loudspeaker lost Loudspeaker connector Check the loudspeaker connect
36. Infusomat Space Service Manual n w gt UN e Q nQ Automated Version 1 1 English BIBRAUN SHARING EXPERTISE This Service Manual is valid for This Service Manual is available under the following part number Languages of this Manual Designation Part No Infusion Pump Infusomat Space 0871 3050 Designation Part No Service Manual Infusomat Space English 8713 9120 The Service Manual for this unit can be supplied in the following languages Designation Part No Service Manual Infusomat Space German 8713 9110 Service Manual Infusomat Space English us 8713 9120U Service Manual Infusomat Space French 8713 9130 Infusomat Space 1 1 gb Important Preliminary Remarks Contact Persons System Overview Unit Diagnosis Calibration Disassembly Assembly Infusomat Space 1 0 gb Table of Contents Service Work Technical Safety Checks Current Versions Revision Service Quality Management Checks and Repair Notes on ESD Spare Parts and Test Equipment Setting Off Technical Training Entry for Technical Training Ordering of Spare Parts and Test Equipment Service Hotline Return of Spare Parts and Test Equipment Safety Officer 8 30 MPG Translation Description System Overview Physical Construction Function Unit Software Service Program Technical Data Options Accessories General Alarms and
37. K KK KS 3452 1364 Screws see Service Parts and Screw Kit p 3 8 Disassembly 1 Press the membrane out of the housing front panel 2 Unscrew two screws and remove the pressure sensor up stream from the housing front panel 3 Unscrew two screws and remove the pressure sensor down stream from the housing front panel 4 Unlock two lugs of the air inline sensor carefully and remove the air inline sensor out of the housing front panel 5 Open the safety clamp with the operating lever and engage 6 Unscrew two screws and remove the safety clamp out of the housing front panel WARNING THE SAFETY CLAMP MUST NOT BE DISASSEMBLED FURTHER Note Pay attention to the dummy slide ISPS when dismantling the safe ty clamp This slide closes one housing opening on the left hand Fig 3 24 side of the safety clamp Legend to fig 3 24 ItemDesignation 1 Safety clamp Screw DELTA PT 22x8 WN 5451 TORX plus 6IP Pressure sensor downstream Screw DELTA PT 25x7 WN 5451 TORX plus 8IP Pressure sensor upstream Front frame Membrane 0 N O o A WwW N Air inline sensor Infusomat Space 1 0 gb 3 25 Disassembly Assembly 3 13 Operating Unit Fig 3 25 Legend to fig 3 25 ItemDesignation 1 Operating unit ribbon cable 2 Connector lock 3 26 Designation Ord No Operating unit ISPS cpl WAAR KK 3452 1470 LC display SP paingan thee rtm 3452 0988 Metal front she
38. KK KK KK KK KK KK KK RR 1 2 Pressure cut off a ass a y la beta k ala lk a is 7 5 Pressure limitation aeter IE 7 5 Processor PCB ss d cede tees day arm toe dandus 3 30 Assembly ix yal Ra ayak di daa 3 32 Preparations ironia be kk e n kan 3 2 P sher OU T raras arar taa dias 3 9 R Responsibilities 0 0 KK KK KK KK KK eee 0 6 REVISION SEMVICE secca ae ek neuem date te Bm ad ena b a 0 5 S Service Program A Lr 1 15 Service program eases t anien eects 1 10 Starting miro iia e da ds 1 10 12 1 Index Service Program version 0 kk eee eee 1 14 Servicing the battery WW kk kk KK KK KK KK KK KK KK 4 1 Servicing the unit LAWA kk kK kK KK KK KK KK KK KK KK KK KK KK 4 1 Software update XW W kk kk kk kk kK KK KK KK KK KK KK KK KK KK 1 9 Spare parts list ss wa sme ceed k aa aka aha 9 1 Special tools 5 aa aa a a z l bade conte ame see an 8 1 System overview a kd ka klan kay a eee eee ka eee 1 1 T Technical data ss si asan ade a a ke e Wae aa lk e EE 1 16 Technical Safety Checks hk kk kk 0 5 5 1 Power supply SP xax s eee cece dl kala dala an Av a 6 1 Test equipment coria daras aliada 8 1 Trouble shooting J kk kk kk KK KK KK KK KK KK 2 19 TSG Kawan Aa lb ad tia odie a kan DK WA BA A 0 5 5 1 Power supply SP ec cece eee eee ed 6 1 Procedural instructions c c cee KEK 7 1 TeStequipment gt s Asya armada amar
39. KK KK KK Upon request Snap in hook SP with leaf spring see Service Parts and Screw Kit p 3 8 Screws and cover caps see Service Parts and Screw Kit p 3 8 Note Please pay attention to the corresponding notes during assembly and installation see Housing Bottom Part sm p 3 36 3 17 Disassembly Assembly Disassembly 1 Remove the drop sensor connector not depicted out of the housing bottom part Fig 3 18 Item 2 Note The drop sensor connector is integrated in the processor PCB 2 Pierce four cover caps Fig 3 18 Item 4 with a small screwdriver and pry the cover caps out 3 Unscrew three screws Fig 3 18 Item 5 and one screw Fig 3 18 Item 3 4 Take off the housing bottom part 5 Pull the snap in hook Fig 3 18 Item 1 out of the housing bottom part Fig 3 18 Legend to fig 3 18 ItemDesignation 1 Snap in hook Housing bottom part Screw DELTA PT 30x12 WN 5452 A2 TORX plus 10IP Cover cap Screw DELTA PT 30x8 WN 5452 A2 TORX plus 10IP ao A W N 3 18 Infusomat Space 1 0 gb 3 10 Inner Frame and Housing Front Panel Disassembly Assembly Designation Ord No Screws see Service Parts and Screw Kit p 3 8 Note Please pay attention to the corresponding notes during assembly and installation see Inner Frame and Housing Front Panel p 3 34 Disassembly Note When dismount
40. The maximum value read on the pressure gauge must correspond with the specification of the TSC for the respective pressure stage 35 Confirm alarm UTS 36 36 Plug service connector Space on connector P2 UTS 37 37 Open operating unit for a short moment and close UTS 38 it again 38 Press the gt button on the LC display in the Line The following Service info is displayed incl UTS 39 TS 27 selection menu air inline sensor The value displayed water value is within the spec ifications of the TSC 39 Dismantle the test setup UTS 40 40 Switch device off The message Pump is switched off in 1 2 3 sec UTS 41 is displayed 41 The device switches off UTS 42 42 Pull off the power supply The blue LED lights up for a short moment and the UTS 43 LC display is getting dark 43 Switch on unit UTS 44 44 Remove battery A permanent alarm high beep sounds UTS 45 45 Insert battery and switch on the device The message Devicealarm 2111 is displayed UTS 46 46 Switch device off This step terminates the device check Table 2 5 Device check Part 3 of 3 Infusomat Space 1 0 gb 2 11 Unit Diagnosis Calibration Procedural Instructions for the Device Check 8 Fig 2 2 7 6 Legend to fig 2 2 ItemDesignation 1 Electronic pressure gauge 10 ml syringe piston fixed Infusion bottle min 100 ml Infusomat Space line standard type ao A CO N Space line 6 Three way
41. UTS 17 TS 16 low 16 Exchange processor PCB The caution sign at the front of the unit flashes yel UTS 17 15 17 low 17 Replace safety clamp UE 18 Replace door bolt drive The operating unit is tightened by the door bolt and UTS 18 TS 19 then fixed 19 Exchange processor PCB UTS 18 20 Replace air inline sensor The Service info appears on the LC display UTS 20 TS 21 Table 2 7 Trouble shooting Part 1 of 2 Infusomat Space 1 0 gb 2 19 Unit Diagnosis Calibration 21 Exchange operating unit The following Service info is displayed incl UTS 20 TS 22 air inline sensor The value displayed is within the specifications of the TSC Templ Temp2 The difference of the values displayed is within the specifications of the TSC 22 Exchange processor PCB UTS 20 23 Exchange operating unit Vent line 16 ml is displayed UTS 24 TS24 24 Exchange processor PCB UTS 24 25 Exchange pump drive You can hear the pump as it is delivering and the UTS 25 TS 26 line is vented 26 Exchange processor PCB UTS 25 27 Replace air inline sensor The following Service info is displayed incl UIS 39 TS 28 air inline sensor The value displayed water value is within the spec ifications of the TSC 28 Exchange processor PCB UTS 39 29 Charge battery module for about 16 hours in the device with power supply connected 30 Replace battery module Table 2 7 T
42. UTS 3 ed to a SpaceControl 2 Remove the device UTS 3 3 Loosen all connections from the device UTS 4 4 Plug service connector Space on connector P2 UTS 5 5 Connect power supply to the device via service con All LEDs light up for a short moment UTS 6 1S1 nector Space 6 The battery charge state and the mains connection UTS 7 TS 5 are displayed at the top left of the LC display with out lighting 7 Switch on unit All LEDs light up from left yellow green blue UTS 8 TS 5 8 A short deep and then a short high beep sound UTS 9 TS 7 9 The colour of the middle LED changes from green to UTS 10 TS 8 red then the LED goes out The yellow and the blue LED remain on for a short moment 10 The message Self test active and the current soft UTS 11 TS 8 ware version are displayed 11 The keyboard and the LC display are lighted UTS 12 159 12 Open door and insert line or press C to input pa UTS 13 TS 10 rameters is displayed 13 Press button to open the operating unit and con The door bolt unlocks the operating unit which then UTS 14 TS 11 firm opens 14 The safety clamp is opened and the caution sign at UTS 17 UTS 15 the front of the unit flashes yellow 15 Open safety clamp The operating lever of the safety clamp can be UTS 16 TS 14 pressed down uniformly against spring tension and engages 16 The caution sign at the front of the unit flashes yel UTS 17 TS 15 low 17 Insert an air filled Infusomat Space lin
43. Version 1 Open the HiBaSeD window via Help Info The current l version of the Service Program is shown in this window lse 2 Close the window by clicking OK BIBRAUN zh Version 3 1 0 Fig 1 16 1 14 Infusomat Space 1 1 gb Device compatibility Infusomat Space S Ver 686E 030002 Please use HiBaSeD Ver current Perfusor Infusomat Cover Version HiBaSeD 1 1 4 1 21 1 3 5 137 150 2 0 1 3 1 0 Fig 1 17 Infusomat 6878030001 68 C030001 68 D 030001 68 030001 687603000 68 B030002 58 B030003 68 B030004 68203000 68 D03000 687803000 687603000 68 B03000 h Infusomat Space 1 1 gb System Overview Compatibility List 1 Open the Unit Compatibility window via Help Com patibility This window displays the compatibility of the Hi BaSeD version and the unit software version 2 Close the window by clicking OK Quit the Service Program 1 Exit the Service Program via File Exit 2 Disconnect a power supply which might be connected from the unit 3 Switch off the unit 4 Remove the battery module 5 The device can be restarted after appr 10 seconds 1 15 System Overview Technical Data All technical data is indicated in the instructions for use Options Accessories 1 16 The functions of the individual options are detailed in the instruc tions for use Infusomat Space Designation Part No
44. X hexagon nut M3 A2 toothed locked washer M3 A2 distance sleeve Infusomat Space 1 1 gb 3 2 Slide Guide You gt Fig 3 8 Legend to fig 3 8 ItemDesignation 1 Notch 2 Slide guide 3 3 Cover for Drop Sensor Connector Fig 3 9 Legend to fig 3 9 ItemDesignation 1 Cover for drop sensor connector Infusomat Space 1 0 gb Disassembly Assembly Designation Ord No Slide guide compl 2 0 0 0 kalak kak awek ak kla 3452 1330 Disassembly 1 Remove slide guide Fig 3 8 Item 2 together with the 12 pump slides carefully out of the housing front panel Note The slide guide can be pryed out with a pointed object through a notch Fig 3 8 Item 1 in the housing front panel on the right bottom side of the slide guide Designation Ord No Cover drop sensor connector Kh kk 3452 1577 Disassembly 1 Carefully remove the cover Fig 3 9 Item 1 for the drop sensor connector from the rear to the front and out of the housing Note Be careful not to damage the lugs in the housing bottom and up per part Disassembly Assembly 3 4 Emergency Release Plug Fig 3 10 Legend to fig 3 10 ItemDesignation 1 Emergency release plug 3 10 Designation Ord No Emergency release plug ISP incl O ring see Service Parts and Screw Kit m p 3 8 Emergency release plug ISP 10 Stiick 3477 3101 Disas
45. am O Pressure stage 1 0 1 0 6 bar bar O Pressure stage 5 0 4 1 0 bar bar O Pressure stage 9 0 8 1 4 bar bar Pressure limitation mechanical O Pmax 1 9 2 1 bar bar J Pmin gt 1 5 bar gt bar Safety clamp Phin gt 1 bar bar Delivery accuracy Ambient temperature 20 25 C Delivery rate 250 ml h Measured volume 25 ml O Divergence 5 Drop sensor optional Delivery rate 250 ml h O Drop sensor alarm flow O Drop sensor alarm occlusion Part 2 of 3 5 2 Sheet 2 of 4 BIBRAUN SHARING EXPERTISE M686 70 01 05 F04 3891 8838 GB Infusomat Space 1 1 gb Technical Safety Check TSC Index d Master to be added to the documentation Visual Inspection Electrical Safety according to IEC EN 60601 1 or VDE 0750 and VDE 0751 Functional Inspection Air inline sensor O Water value gt 600 mV mV O Difference of temperature sensors lt 3 C sC CJ Air alarm O Air value lt 100 mV mV Note If the water or air value is not reached im mediately these values must be read again after 120 seconds Switch on unit without power supply O Self test O Magnetic function of the battery com partment cover Remove battery module O Alarm for min 3 minutes Part 3 of 3 Mechanical Aids and Measuring Equip ment Used Accessories Used O Service connector SP O O
46. ayed on the LC dis play 11 Read off value after approximately 10 seconds and compare with the specifications of the TSC Note If the water or air value is not reached immediately these values must be read again after 120 seconds Battery Check 1 Switch device off 2 Pull off the power supply 3 Switch on unit Self test is carried out 4 Open battery compartment cover during operation An alarm is activated Mechanical Inspection 1 Connect the power supply to a Space system device and check the lock for proper function Functional Check 1 Thedevice connected operates correctly after being switched on Infusomat Space 1 1 gb Test Equipment and Special Tools 8 Test Equipment Fig 8 1 Slide clamp Infusomat Space 1 0 gb Designation Ord No For Device Check Service connector Space KK KK KK KK KK 3452 1062 HiBaSeD Service CD WW KK KK 0871 3301 Infusomat Space line standard type new not used Infusion bag or bottle min 100 ml Three way valve 2 pieces 10 ml syringe air buffer for measurement with electronic pressure gauge syringe drawn up to 10 ml and piston fixed mechanically 1 ml syringe for air bubble injection Electronic pressure gauge with peak value recognition Pressure cell alternative of electronic pressure gauge Slide clamp of the Infusomat Space line For Repairs HiBaSeD Service CD 0 000 KK KK KK KK 0871 3301 For t
47. d bew RR 4 1 Compatibility list 0 cee eee eee eee 1 15 Cover for drop sensor connector 000 eee eee 3 9 Cutrent versions cenar ree bela 0 5 D Delivery ACCUTACY iii a bates 7 6 DESCrIPtION orar e a 1 1 Device Check i s a Avka a a e ra 2 9 Disassembly start iniii a WA a n b ka Z 3 1 Drop sensor connector kk kk kk kK KK KK KK KK KK 3 9 E Electrical safety ariadna tx ase nak dn 7 2 Emergency release plug AWK kK KK KK KK KK KK KK KK KK 3 10 Error GO ES ak s n xalke a dai aa ad avana i ala ie 2 3 Error modes anan na yad a ek da aia E 2 8 F FUNCION RT 1 4 Functional inspection WWW kk KK KK KK KK KK KK KK KK 7 3 H Housing bottom part AWK kk kk KK KK KK KK KK KK KK KK 3 17 Araya O DD ge aden hails 3 36 Infusomat Space 1 1 gb Index 12 Housing front panel J L kk 3 19 3 25 Assembly east dirt dene ae 3 34 Housing upper part JJA kk kK KK KK KK KK KK KK KK KK KK 3 14 Asset ds 3 36 I lnn r frame s aeo La c eer tds a 3 19 3 21 Assembly novere det rb rd 3 33 3 34 Intended use ooo 1 1 L Linear potentiometer xi kasa a k a kal eee ee 3 24 List of abbreviations MAWA KK KK KK KK KK eee eee eee 0 9 loudspeaket air 3 17 0 Operating Unit escrita kk ci 3 26 Assembly ers eG ai RU Ub RERO RE RAE 3 32 Open manually eo sa l n n d k a kanal 1 6 DIDI xl ka Kat yad ewna na haat nas kak en een aed 1 16 P Physical construction WWW kK KK
48. e and close The operating unit is tightened by the door bolt and UTS 18 TS 18 operating unit then fixed 18 The line selection menu appears on the LC display UTS 19 TS 18 Table 2 5 Device check Part 1 of 3 Infusomat Space 1 0 gb 2 9 Unit Diagnosis Calibration 19 Press the gt key The following Service info is displayed incl UTS 20 TS 20 air inline sensor The value displayed air value is within the specifi cations of the TSC Temp1 _____ Temp2 The difference of the values displayed is within the specifications of the TSC 20 Remove the air filled Infusomat Space line out of UTS 21 the device 21 Prepare test setup see Test Setup amp p 2 12 UTS 22 and set three way valves as shown in Fig 2 3 22 Insert the Infusomat Space line of the test setup The line selection menu appears on the LC display UTS 23 and close operating unit 23 Select line type Vent line 16 ml is displayed UTS 24 TS23 24 Confirm venting by pressing Yes You can hear the pump as it is delivering and the UTS 25 TS 25 line is vented 25 Pull out the service connector Space UTS 26 26 Inputa delivery rate for upstream as specified in the UTS 27 TSC and any volume select the first pressure stage according to the TSC and start infusion 27 Close the roller clamp of the Infusomat Space line The message Alarm Check supply is displayed UTS 28 or squeeze t
49. e cing we x 8 1 U UI DE DO gt ala naa di annua c 3 11 U filit SOL DA GE J oa wu ha da hlna a Rb rd 1 9 V Visual inspection 0 0 eee KK KK KK KK KK eee ees 7 1 Z Zero force insertion connector 00 eee eee 3 31 Infusomat Space 1 1 gb
50. e components HN j SS concor onsec ed and the initial window of the Service Program is displayed The connection to hardware components e g Perfusor or Infusomat may only be made using the prescribed interfaces and leads These products may not be simultaneously used on a 2 Read the notes careful y patient If pre set values are taken over by the user the use of these and all other entries the user s 3 Mark the field I accept all conditions and then the field ibility A TRE Yes to confirm that you have read the notes This software may only be used by trained and specially instructed personnel Furthermore the devices are only to be serviced by personnel trained in the repair of this special equipment By pressing YES the user accepts all these conditions Note Click the field English to switch the language of the notes over to English IV accept all conditions 1 10 Infusomat Space 1 1 gb EH Program selection Program selection History Viewer C Clone Data Service Program xx Password Scan interfaces serial r Detect devices Device detected on com3 Device detection active coml IV all ports Fig 1 10 Infusomat Space 1 1 gb 4 System Overview Enter the password and confirm it by clicking the field Start The Service Program checks the PC interfaces for connected devices of the Space system Units that were found are dis
51. e handled Each workstation must be equipped with a conductive table sur face The conductive surface the soldering iron or the soldering stations must be grounded via protective resistors Chairs must be of antistatic design The floor or floor mats should be of electrically conductive material Personnel must wear conductive wristbands which are connected to a central ground potential via protective resistors e g the ground contact of a wall outlet Furthermore it is recommended that personnel wear cotton clothing and electrically conductive shoes to prevent electrostatic charge Only use original spare parts from the manufacturer Do not tamper with assembly groups which can only be exchanged com pletely The spare parts required are listed in Section 9 Service personnel are responsible for the calibration of their test equipment Original test equipment can be calibrated at the works of B Braun Further information is available upon request Additional notes and warnings are set off as follows Note Is used for additional or special notes concerning information and working steps CAUTION Is used for working steps which may result in damage to the unit system or to a connected device WARNING IS USED FOR WORKING STEPS WHICH MAY RESULT IN PERSONAL INJURY References to chapters are shown as follows see Setting Off pg O 8 References to figures and tables are shown as follows Fig 2 3 or Table 2 1
52. e integrated into the system Up to six SpaceStations can be set up as a column with a total of 24 pumps SpaceStation placed next to each other can be con nected via special connection cables if the maximum number of 24 pumps in maximum three columns is not exceeded SpaceCover Standard or SpaceCover Comfort forms the top of each column Alarms are signalled by a row of LEDs and a loud speaker in the SpaceCover Comfort System Overview Physical Construction The Infusomat Space housing mainly consists of the bottom part the upper part the front part and the operating device The battery module is inserted in the rear of the housing upper part The opening is covered by the battery compartment cover The operating unit is attached to the front of the bottom part The thrust bearing of the tube pump peristaltic pump the spring mounted pressure elements for the two pressure sensors and the air inline sensor as well as the shackle for the slide clamp of the Infusomat Space line are located at the rear of the operating unit The operating unit is mechanically locked in its closed posi tion via three metal pins A motor driven door bolt is used to lock the operating unit In case of an emergency the operating device can be opened through an opening on the left side of the housing top During normal operation this opening is closed with a plug According to the line run the following subsystems are installed in the housing fron
53. e make sure that the toothed wheels do not rub along the metal edges of the inner frame as the toothed wheels can easily be damaged 7 Liftthe door bolt on the right side and push it to the left until the guide rivet is exactly positioned in the mounting hole of the door bolt 8 Lift the door bolt over the guide rivet of the inner frame 3 23 Disassembly Assembly p E 3 Fig 3 23 2 Legend to fig 3 23 Pos Bezeichnung 1 Locking opening for snap in pins 2 Connection leads to the pump drive motor 3 Pump drive motor 3 24 WARNING BE VERY CAREFUL WHEN SOLDERING AS THE BIT OF THE SOL DERING IRON CAN GET VERY HOT WHICH CAN LEAD TO INJURIES IN ADDITION DANGEROUS VAPOURS MAY RELEASE DURING THE SOLDERING PROCESS SOLDERING WORK SHOULD ONLY BE EXECUTED BY TRAINED AND DULY AUTHORIZED PERSONNEL IN A SUITABLE ENVIRONMENT 9 Unsolder the two connection leads at the pump drive motor 10 Pull the linear potentiometer carefully out of the door bolt drive Note The linear potentiometer is held in the door bolt drive by two top snap in pins Infusomat Space 1 1 gb Disassembly Assembly 3 12 Housing Front Panel Designation Ord No Membrane ISP WA KK KK KK KK KK KK KK KK KK 3452 1356 Pressure sensor ISP 220020c cece e eee 3452 1372 Air inline sensor ISP AWA KK KK KK eee 3452 1380 Safety clamp ISPS WW KK KK KK KK eee KK 3452 1399 Front frame ISPS cpl Ah KK K
54. e sensor downstream 00 00 46 F04 serve to check whether both CJ Visual alarm membrane limit values are met J Status display O Pressure sensor upstream O Lighting membrane O Pump membrane Operation 1 Pump slide O Opening and closing mechanism of the O Coding of pump slide guide operating unit O Pump connecting rod O Slide clamp releases release lever of the O Air inline sensor safety clamp The Infusomat Space free from grease or ultrasonic gel line is squeezed free from cracks O Buttons on the operating unit J Drop sensor Infusion lens CJ Staff call O Accessories O Bolus Cleanliness completeness damage and C Slide clamp is completely pulled on to faults affecting safety damage and the Infusomat Space line when the readability of the label operating unit is opened O Check the unit and the accessories for O The operating lever of the safety clamp compatibility can be pressed down uniformly against spring tension and engages The yellow caution sign flashes Part 1 of 3 M686 70 01 05 F04 3891 8838 GB Infusomat Space 1 1 gb BIBRAUN SHARING EXPERTISE 5 1 Sheet 1 of 4 Technical Safety Check TSC Index d Master to be added to the documentation Visual Inspection Electrical Safety according to IEC EN 60601 1 or VDE 0750 and VDE 0751 Functional Inspection Pressure cut off electronic Delivery rate 250 ml h Test without drop sensor O Upstream O Alarm C Downstre
55. e shooting list The following equipment and gauges are necessary for testing the unit and or performing troubleshooting PC Service connector Space Service Program HiBaSeD nterface cable CAUTION Take special care when carrying out measurements on an open and switched on unit Short circuits and wrong measuring meth ods can cause serious damage to or destroy the subsystems of the device Unit Diagnosis Calibration The unit check calibration and trouble shooting are subdivided into numbered working steps Unit Test Step UTS Calibration Step CS Trouble Shooting TS and are based on each other Beginning with UTS 1 the operation described here has to be ex ecuted The consequences of the steps performed are listed in the Function column If the result corresponds to the consequence the working step must be carried out to which reference is made in the column If yes If the result does not correspond with the function described the working step in column If no is to be ex ecuted One example is given in Fig 2 1 Model table 1 1 2 3 4 5 Model table 2 Fig 2 1 Model tables Steps for which additional information is required are described after the table in detail Infusomat Space 1 0 gb Alarms and Error Codes Infusomat Space 1 0 gb Unit Diagnosis Calibration The alarms of the In
56. ed Ros Version 1 0 PAGES 21 cita aene in Version 1 1 Chapter 6 Technical Safety Check TSC Page 3 lt ON Version 1 1 Pag 6 lt 1 rss taa e u a At Aa l van Version 1 1 MM TRENT M MEEREO Version 1 0 Page EZ ii an kak a k er mis Version 1 0 Page 3 S2420 ice yaran a Ak a DR W Ali de av i Version 1 1 Chapter 7 Procedural Instructions on the TSC PALE C99 pS A Version 1 0 A e ne QU P RENE Ea a Version 1 0 i M EE Merion 10 Page 72 zm a m Rint Version 1 0 900 9 OL Peres aa yeke alye ya ceke e eye Wenn Yekl IA l age 7 23 ane mi edo Q A REPE ed a l Version 1 0 LL MM Merion Page Fe uue lel aren nale TTC ae n2 Version 1 0 Ld oe e gl Version TA POE r mm Version 1 0 Rage StS k eyes a Radius Version 1 PUEDE asias circa o cade Version 1 0 _ DM VM Q iris kul PAGETEZ suada duet elan iach a esce Een i Version 1 0 cx Version Tu KE T N pXXSU RE GR eee Version 1 1 Page 3 233 cssc br ke Radix snp PEE Version 1 1 lage 334 on cede ieee theta edie heeds Version 1 1 Chapter 8 Test Equipment and Special Tools PAGE 3 35 aii Version 1 0 Page8 1 000occccccccncrcccrr e Version 1 0 lage 3 9 30 arar ee Rc ec n a B N Version 1 0 Page 8 52 xa kala nane eee eee eee a Version 1 1 Page 3 Dir ed ids Version 1 0 P ge 3 538 hisori EA Ea Rr acu ag Meanie Version 1 1 PA aS List DIE MMC Version 1 1 A oo Version 1 1 E NC NERONE RR TRE Version 1 1 P ge 9 iii eba Version 1 1 e EE T E Version 1 1 lage 9
57. emove the rectangular cover cap from the housing top and keep it in a safe place until calibration be fore mounting the housing upper part Note Take care that no cables are squeezed and check that the connec tor seals at the back are correctly placed in the housing the snap in hook spring is located outside the housing wall and that the loudspeaker is correctly placed when the housing is fitted The two wires must point downwards so that the loudspeaker wires Infusomat Space 1 0 gb Infusomat Space 1 0 gb Disassembly Assembly are placed in the insulation to the coil in the cut out of the hous ing bottom part 2 Check beforehand the position of the cable cover at the top and bottom of the housing upper part and the bottom part 3 Lower the upper part of the housing vertically and carefully on to the lower part 4 Engage the snap in hooks between housing top and front frame Rework when Replacing the Processor PCB When the processor PCB is replaced all data of the pump except for the calibration data was probably saved on a PC see Prepa rations for Exchanging the Processor PCB p 3 2 Carry out the following steps to transmit the data back to the device WARNING WHEN DATA PREVIOUSLY SAVED IS TRANSFERRED BACK TO THE DEVICE CALIBRATION DATA IS NOT TRANSFERRED THEREFORE A COMPLETE NEW CALIBRATION OF THE UNIT IS REQUIRED Note Please note that text and or functions of the Service Program may
58. erating Unit 1 Insert the hinge spring with its long side in the hinge cover opening spring and with its angled side in the rear panel bore hole of the operating unit 2 Please pay attention to the bonding on the keyboard PCB when mounting the rear panel of the operating unit 3 The grooves of the hinge covers are pointing to the operating unit rear panel 4 Push the operating unit connection cable into the processor PCB connector 5 Lock the connector 6 The ribbon cable cover of the operating unit must be layed between the bottom inner frame and the operating unit and over the connector 7 Checkthe seal washers for correct position and completeness once again after assembly 3 32 Infusomat Space 1 0 gb Disassembly Assembly Inner Frame NE m Fig 3 32 Legend to fig 3 32 ItemDesignation 1 Door bolt plastic guides 5 Pump drive movement areas 2 Inner frame spring holder 6 Door bolt drive toothed wheels 3 Pump frame bearing of the 6 mm straight pin 7 Door bolt guides and supporting area 4 Pump drive toothed wheels Infusomat Space 1 1 gb 3 33 3 34 Disassembly Assembly Make sure that the following components are sufficiently lu bricated when installing the inner frame If the grease film is not sufficient stiff or jerky movements grease these areas slightly with Polylub GLY 501 Note If chemical substances such as lubricating grease are used the safety
59. esent an explosion hazard Density at 20 C 7 0 88 g cm Solubility in Miscibility with water Insoluble Thermal decomposition conditions to be avoided No decomposition if used and stored according to specifications Materials to be avoided oxidizing agents Dangerous reactions No dangerous reactions known Dangerous decomposition products none under normal use Additional toxicological information Prolonged skin contact may cause skin irritation and or dermatitis GB Contd on page 4 Fig 10 1 Safety data sheet Kl ber POLYLUB GLY 501 Part 3 of 5 Infusomat Space 1 1 gb 10 3 Safety Data Sheets Fig 10 1 Page 4 4 Material Safety Data Sheet According to 91 155 EEC Printing date 26 01 2005 Reviewed on 29 03 2004 Trade name POLYLUB GLY 501 Contd of page 3 Ecotoxical effects Behaviour in sewage processing plants The product can be mechanically separated General notes Do not allow product to reach ground water water course or sewage system Product Recommendation Can be incinerated in accordance with local and national regulations Waste disposal key For this product no waste disposal key according the European Waste Catalogue EWC can be determined as only the purpose of application defined by the user enables an allocation The waste code number has to be determined in accordance with the local waste disposer
60. essor PCB via the ribbon cable 6 Pull the door bolt drive connector Fig 3 19 Item 2 off the processor PCB 7 Pull the pressure sensor connector upstream Fig 3 19 Item 7 off the processor PCB 8 Pull the pressure sensor connector downstream Fig 3 19 Item 3 off the processor PCB Disassembly 1 Carefully remove the inner frame Fig 3 20 Item 3 from the housing front panel Fig 3 20 Item 1 Note The membrane in the housing front panel must not be used again Always use a new membrane for assembly Infusomat Space 1 0 gb 3 11 Inner Frame Infusomat Space 1 1 gb Disassembly Assembly Designation Ord No Door bolt drive ISP WWW KK cc eee eee KK 3452 1429 linear poti ISP mico aaa 3452 1056 Pressure adjustment unit ISPS 3452 1445 spring holder pressure springs adjusting screw adjusting nut bracket bracket locking Pump frame ISP cpl 2 KK KK KK eee 3452 1402 incl pump drive motor ISP Pump drive motor ISP 2 eee KK RR KAK 3452 1410 Door bolt ISP lt s k ci eee 3452 1496 Inner frame ISP WA KK KK KK KK KK KK KK KK 3452 1437 Bracket locking and screws see Service Parts and Screw Kit p 3 8 Note Please pay attention to the corresponding notes during assembly and installation see Inner Frame p 3 33 Disassembly 1 Carefully press two bracket lockings Fig 3 21 Item 10 at the front and rear out of the i
61. est Step Important Preliminary Remarks VDE Verband der Elektrotechnik Elektronik und Informationstechnik e V german association of engineers 0 10 Infusomat Space 1 0 gb Technical Training Entry for Technical Training Ordering of Spare Parts and Test Equipment Service Hotline Return of Spare Parts and Test Equipment Safety Officer S 30 MPG Translation Infusomat Space 1 1 gb Contact Persons Via local representative Application for a technical training course must be made via the responsible representative Please contact your local B Braun subsidary International Technicians Intercompany Nadja Machal Fax 49 5661 75 47 89 e mail nadja machal 2bbraun com Karl Tippel Tanja K rdel Phone 49 5661 71 35 25 Fax 49 5661 71 35 26 e mail X karl tippel bbraun com e mail tanja koerdel gbbraun com B Braun Melsungen AG Schwarzenberger Weg 73 79 Wareneingang Werk C 34 212 Melsungen Germany Dr Ludwig Sch tz e mail ludwig schuetz bbraun com Br ckner GmbH Germany 0 11 0 Contact Persons For your notes 0 12 Infusomat Space 1 1 gb Description System Overview Fig 1 1 Space system Legend to fig 1 1 ItemDesignation 1 a A CO N SpaceCover Infusion pump Infusomat Space Infusion syringe pump Perfusor Space SpaceControl SpaceStation Infusomat Space 1 0
62. et The drop sensor triggers an alarm immediately 4 Disconnect the Infusomat Space line from the container Alarm by drop sensor within 5 seconds Air Inline Sensor 1 Plug service connector Space on connector P2 2 Insert an Infusomat Space line filled with water in the de vice and close operating unit 3 Press the gt button on the device in the Line change menu The value of the air inline sensor is displayed on the LC dis play 4 Read off value after approximately 10 seconds and compare with the specifications of the TSC Infusomat Space 1 0 gb 7 7 Procedural Instructions on the TSC Functional Inspection Power Supply SP Note If the water or air value is not reached immediately these values must be read again after 120 seconds 5 Read off the values of the temperature sensors on the LC dis play and compare the difference with the specifications of the TSC 6 Enter a delivery rate of 250 ml h and a volume of 250 ml 7 Start infusion 8 Generate an air bubble of appr 0 4 ml appr 56 mm length of line with air in the supply line to the pump An alarm is trig gered when the air bubble is detected 9 Insert an Infusomat Space line filled with air in the device and close operating unit or disconnect the line at the con tainer and vent in order to remove the water out of the line 10 Press the gt button on the device in the Line change menu The value of the air inline sensor is displ
63. et ISPS 000000 K8 3452 1488 Axle ISP ais aa 3452 1461 incl hinge covers and hinge spring Screws see Service Parts and Screw Kit p 3 8 Note Please pay attention to the corresponding notes during assembly and installation see Operating Unit p 3 32 Assembly Disassembly 1 Push the right and left connector lock Fig 3 25 Item 2 on the processor PCB carefully rearward 2 Disconnect the operating unit ribbon cable Fig 3 25 Item 1 from the processor PCB Infusomat Space 1 0 gb 5 Fig 3 26 Legend to fig 3 26 ItemDesignation 1 Countersunk screw EJOT 20x9 WN 5454 TORX plus 6IP Seal washer PA 2 2x0 3 Operating unit rear panel metal front sheet Bonding ao A CO N Operating unit with LC display Infusomat Space 1 0 gb Disassembly Assembly 3 Unscrew 6 screws Fig 3 26 Item 1 and remove the seal washers Fig 3 26 Item 2 from the operating unit rear panel metal front sheet Fig 3 26 Item 3 Note The screws of the keyboard must not be loosened A uniform pres sure point of all keys is only guaranteed when a special tool is used for assembly 4 Remove the operating unit Fig 3 26 Item 5 from the rear panel metal front sheet Note Please pay attention to the bonding Fig 3 26 Item 4 on the keyboard PCB during disassembly 3 27 Disassembly Assembly Fig 3 27
64. f the Individual Pages Infusomat Space 1 1 gb 11 1 Revision Documentation 1 2 Infusomat Space 1 0 gb Title and Leading Pages Page O 10 Page O 11 Page O 12 Chapter 1 System Overview Infusomat Space 1 1 gb Version 1 1 Version 1 1 Version 1 0 Version 1 0 Version 1 1 Version 1 1 Version 1 1 Version 1 0 Version 1 0 Version 1 0 Version 1 1 Version 1 1 Version 1 0 Version 1 0 Version 1 0 Version 1 0 Version 1 0 Version 1 1 Version 1 1 Version 1 0 Version 1 1 Version 1 1 Version 1 1 Version 1 1 Version 1 1 Version 1 1 Version 1 1 Version 1 1 Version 1 0 Version 1 0 Version 1 0 Version 1 0 Page 2 13 Page 2 14 Page 2 15 Page 2 16 Page 2 17 Page 2 18 Page 2 19 Page 2 20 Chapter 3 Disassembly Assembly Page 3 12 Page 3 13 Page 3 14 Page 3 15 Page 3 16 Page 3 17 Page 3 18 Version 1 0 Version 1 1 Version 1 1 Version 1 0 Version 1 0 Version 1 0 Version 1 0 Version 1 0 Version 1 0 Version 1 1 Version 1 1 Version 1 1 Version 1 1 Version 1 1 Version 1 0 Version 1 0 Version 1 0 Version 1 1 Version 1 1 Version 1 1 Version 1 1 Version 1 1 Version 1 1 Version 1 1 Version 1 0 Version 1 1 Version 1 0 Version 1 0 Version 1 0 Version 1 0 Version 1 0 Version 1 0 Version 1 0 Version 1 0 11 3 Page 3 19 ui aa is Version 1 0 Page5 4 paa la i eee eee eee eens Version 1 1 lage 3 920 ici ne cine dh phe edie te
65. fusomat Space are classified in 5 categories These categories are listed hereafter according to their impor tance Alarm advice In case of unacceptable inputs corresponding messages are displayed e g Caution Rate out of range The parameter cannot be changed and a beep sounds Pre alarm Pre alarms are triggered several minutes depending on the service settings before the operating alarms Reminder alarm A reminder alarm is triggered if the device is not operated for two minutes when input or operation was not finished Operating alarm In case of an operating alarm the infusion is stopped An au dible signal is released the red LED flashes and a staff call is triggered The message Alarm and the cause of the alarm appear on the display Device alarm The most important alarms and error codes as well as their mean ing and possible fault clearance are specified in the following lists Note The device should be checked after every repair or service see Device Check p 2 9 Unit Diagnosis Calibration 1 Battery nearly discharged type pre alarm Alarms The device was not connected to the mains long enough Connect device to the supply voltage via the AC adaptor plug Start battery maintenance by pressing OK if this message is displayed when the device is switched off WARNING THE FOLLOWING ACTIVITY MUST NOT BE CARRIED OUT WHEN A PATIENT IS CON NECTED TO THE UN
66. he Infusomat Space line between con tainer and pump 28 Acknowledge alarm and open roller clamp UTS 29 29 Input a delivery rate for downstream according to UTS 30 the TSC and start infusion 30 Set three way valves as shown in Fig 2 4 When the maximum pressure of this pressure stage UTS 31 is reached the delivery is stopped the red LED on the operating unit flashes and the message Alarm Pressure too high is displayed The maximum value read on the pressure gauge must correspond with the specification of the TSC for the respective pressure stage 31 Confirm alarm UIS 32 Table 2 5 Device check Part 2 of 3 2 10 Infusomat Space 1 0 gb Unit Diagnosis Calibration 32 Select the second pressure stage according to the When the maximum pressure of this pressure stage UTS 33 TSC and start infusion is reached the delivery is stopped the red LED on the operating unit flashes and the message Alarm Pressure too high is displayed The maximum value read on the pressure gauge must correspond with the specifications of the TSC for the respective pressure stage 33 Confirm alarm UTS 34 34 Select the third pressure stage according to the TSC When the maximum pressure of this pressure stage UTS 35 and start infusion is reached the delivery is stopped the red LED on the operating unit flashes and the message Alarm Pressure too high is displayed
67. he TSC Measuring instrument for electrical safety Service adapter SP AWK KK KK KK KK KK KK KK 0770 5174 for measuring the electrical safety Infusomat Space line standard type new not used Infusion bag or bottle min 100 ml Three way valve 2 pieces 10 ml syringe air buffer for measurement with electronic pressure gauge syringe drawn up to 10 ml and piston fixed mechanically 1 ml syringe for air bubble injection Electronic pressure gauge with peak value recognition Pressure cell alternative of electronic pressure gauge Graduated cylinder 25 ml 0 4 ml Stop watch Test Equipment and Special Tools Special Tools Designation Ord No TORX screwdriver kit 5 10 25 TORX plus screwdriver kit 5 10 25 Screwdriver 6IPX60 TORX plus 220000 4002 4806 Screwdriver 8IPx60 TORX plus A kk 4002 4814 Screwdriver 10IPx60 TORX plus 4002 4822 Infusomat Space 1 1 gb Spare Parts List 9 Item Designation Ord No Infusomat Space Service part kit Infusomat Space 3452 1593 with housing cover cap 100 pieces operating unit cap 5 pieces housing foot SP 10 pieces snap in hook SP with leaf spring 2 pieces emergency release plug ISP incl O ring 2 pieces emergency release crank ISP 1 piece bracket locking 20 pieces seal washer PA 2 2x0 3 200 St ck safety clamp seal housing bottom part 10 pieces Cover cap kit for housing SP 50 pieces
68. ing the inner frame and the housing front panel please pay attention to the length of the connection cables to the processor PCB Fig 3 19 Legend to fig 3 19 ItemDesignation 1 N O 0d A CO N Connector for ribbon cable operating unit Door bolt drive connector Pressure sensor connector downstream Air inline sensor connector Safety clamp connector Pump drive connector Pressure sensor connector upstream Infusomat Space 1 0 gb Fold the locking Fig 3 19 Item 4 of the ribbon cable connector to the air sensor up and pull the ribbon cable off the connector Pull the safety clamp connector Fig 3 19 Item 5 off the processor PCB Pull the pump drive connector Fig 3 19 Item 6 off the processor PCB 3 19 Disassembly Assembly Fig 3 20 Legend to fig 3 20 ItemDesignation 1 Housing front panel cpl 2 Fillister head screw M3x6 TORX 3 Inner frame cpl 4 PCB of direction of rotation sensor 5 Screw DELTA PT 25x7 WN 5451 TORX plus 8IP 6 Bottom inner frame 3 20 4 Loosen four screws Fig 3 20 Item 2 and lift the inner frame together with the housing front panel Fig 3 20 Item 6 from the bottom inner frame Unscrew one screw Fig 3 20 Item 5 and remove the PCB of the direction of rotation sensor for the pump drive from the pump frame Note The PCB of the direction of rotation sensor is connected to the proc
69. isposable articles read out is displayed on screen Disassembly Assembly HiBaSeD International Service Program File Extras Tools Update Help Ve Infusomat Space S Pressure Levels Tests s Disconnect O Calibration Cal update Serial number Disposable articles modification data Operatingdata Delivery Constant 2006 10 09 00 00 00 1823 status Name addtin on S Space Line v 1 00 on S Space Line Neut V 1 00 on P Intrafix PVC v 1 00 State on active off Inactive From device To device Load from Fig 3 6 11 Press the Save as button In the window which opens now you are asked for the storage position of the file on the PC hard disk and the file name 3 6 Infusomat Space 1 1 gb Speichern unter Save in Disposables d S IsPS 1823 241106 TUT 2 Eigene Dateien Arbeitsplatz Netzwerkumgeb File name isPs 1823 180101 Save ung Save as type Tube Table X Cancel Infusomat Space 1 1 gb Disassembly Assembly 12 Select the storage position in the Save file as window and input a unique file name Note The storage position which is proposed should not be changed 13 Press the Save button The data of the pump is saved on the PC hard disk 14 Exit the Service Program see Quit the Service Program p 1 15 Note When the new processor PCB has been installed the saved data must be transfer
70. le current 32 2208 stepmotor not high current 33 2209 stepmotor PH1A PH1B PH2A PH2B not 0 34 2210 K_PKM_UB not off Door bolt drive defective Replace door bolt drive or ISP 35 2211 K_PKM_UB not on processor PCB 36 2215 no V_MOT Processor PCB defective Exchange ISP processor PCB 37 2216 overvoltage test fail 38 2217 no V_MOT 39 2218 undervoltage test fail 40 2220 door poti defect Door bolt drive potentiometer Replace door bolt drive defective 41 2221 tube change timeout Defective or old software Update software change ISP proc essor PCB if necessary 42 2222 temperatur sensor difference Air inline sensor defective Replace air inline sensor 43 2223 airsensor defect 44 2224 rate compensation difference Dose calculation error Update unit software 45 2225 rate compensation max value 46 2226 rate compensation min value 47 2227 received Airdata defect Air inline sensor defective Replace air inline sensor 48 2228 defect temperature sensor switch 49 2229 FuP not stopped by KuP alarm Software defective or obsolete Update software change ISP proc essor PCB if necessary 50 2230 drive totphase is defect Software defective or obsolete Update software change ISP proc essor PCB if necessary 51 2231 tube data set is defect Software defective or obsolete Update software change ISP proc essor PCB if necessary 52 2237 2255 Internal Error Table 2 3 Device alarms of the control microprocessor Part 2 of 2
71. ls 686E030002 2006 10 09 00 00 00 Name Rates Kvo 1823 Service Select a data source to modify From device From file Default Accept or reject modified data Accepted data must be saved com3 Calibration Cal update Serial number Disposable articles Other Status Options Infusomat Language Time Tests s Disconnect Modification data Operating data Delivery Constant Alarms Therapy lo Fig 3 38 Infusomat Space 1 1 gb 3 39 Disassembly Assembly Bestatigung Write data to device Abbrechen Fig 3 39 Worker ID Please insert data Worker ID Fig 3 40 Information Writing task successfully completed Modification of a device includes the possibility of changing therapy relevant data 1 You are responsible for verification of the device settings Fig 3 41 3 40 10 The data is transferred to the device Select the tab Disposable article Press the OK button when the window Confirmation is Input your user number in the window User number as well as the six digit serial number of the device if necessary Confirm the input with OK A message is displayed that a change of the device data may affect the patients Confirm by pressing the OK button Infusomat Space 1 1 gb Disassembly Assembly HiBaSeD International Service Program File Extras Tools Update Help f
72. mitation check and unlock operating unit but do not open by hand Read off value on the pressure gauge after min 10 seconds and compare with the specifications in the TSC Document value Pull out the service connector Space Delivery Accuracy Requirements 1 Moog B5 Test setup according to Fig 7 2 Ambient temperature according to the TSC Insert the Infusomat Space line of the test setup Fig 7 2 in the device Close three way cock 1 to the syringe please see Fig 7 5 Vent test setup Bottom part of drop chamber must be 2 3 filled Insert cannula in empty graduated cylinder Set delivery rate according to the TSC Simultaneously start infusion and stop watch Stop the stop watch when the measured volume defined in the TSC is reached Evaluate deviation in percent according to Table 7 1 and document Infusomat Space 1 0 gb Procedural Instructions on the TSC Measuring Time Deviation 6 min 40 0 sec 10 6 min 18 9 sec 5 6 min 15 0 sec 4 6 min 11 1 sec 3 6 min 7 3 sec 2 6 min 3 6 sec 1 6 min 0 0 sec 0 5 min 56 4 sec 1 5 min 52 9 sec 2 5 min 49 5 sec 3 5 min 46 2 sec 4 5 min 42 9 sec 5 5 min 27 3 sec 10 Table 7 1 Drop Sensor Option 1 Connect the optional drop sensor to the device 2 Start infusion according to the specifications of the TSC 3 Squeeze the drop chamber bottom part to generate a j
73. n interrupted 12 2112 2114 Internal Error 13 2115 Drive too fast Motor drive Exchange processor PCB Recognition of direction of rota tion 14 2116 Drive too slow Motor drive Exchange processor PCB Recognition of direction of rota tion 15 2117 2118 Internal Error 16 2119 Icd backlight on defect LC display defective Exchange operating unit 17 2120 led backlight off defect LC display defective Exchange operating unit 18 2121 red led on defect LC display defective Exchange operating unit 19 2122 red led off defect LC display defective Exchange operating unit 20 2123 key pressed too long without EA Keyboard defective Carry out device check Key 60sec 21 2124 2127 Internal Error 22 2128 Drive motion rightless forward Motor drive Exchange ISP processor PCB 23 2129 Drive motion rightless backward Recognition of direction of rota tion 24 2130 2200 Internal Error Table 2 3 Device alarms of the control microprocessor Part 1 of 2 Infusomat Space 1 1 gb Unit Diagnosis Calibration 25 2201 different version FuP to KuP Software Update unit software 26 2202 stepmotor PH1A not 0 Stepper motor defective Replace pump drive motor or ISP 27 2203 stepmotor not middle current processor PCB 28 2204 stepmotor K_SM_CLK 29 2205 stepmotor PH1A not 1 30 2206 stepmotor not zero current 31 2207 stepmotor not midd
74. n process is terminated Infusomat Space 1 1 gb Unit Diagnosis Calibration Further Trouble Shooting Note The following trouble shooting cannot be carried out independ ently It is based on the precise observance of the steps for the de vice check see Device Check p 2 9 From there reference is made to the corresponding trouble shooting steps 1 Replace power supply All LEDs light up for a short moment UTS 6 152 2 At least one LED lights up for a short moment 15 9 TS4 3 Replace the LC display PZP USs5 4 A message is displayed on the LC display TS 6 TSS 5 Exchange processor PCB All LEDs light up for a short moment and a message UTS 6 TS 6 is displayed on the LC display 6 Exchange operating unit UTS 7 7 Exchange loudspeaker and switch on unit A short deep and then a short high beep sound UTS 9 TS 8 8 Exchange processor PCB UTS 9 9 Exchange operating unit The keyboard and the LC display are lighted UTS 12 TS 10 10 Exchange processor PCB UTS 12 11 Replace door bolt drive The door bolt unlocks the operating unit which then UTS 14 TS 12 opens 12 Exchange processor PCB The door bolt unlocks the operating unit which then UTS 14 15 13 opens 13 Exchange operating unit UTS 14 14 Replace safety clamp UTS 16 15 Exchange pressure sensor downstream The caution sign at the front of the unit flashes yel
75. nitored by a linear potentiometer Infusomat Space 1 0 gb Infusomat Space 1 0 gb System Overview The safety clamp squeezes the Infusomat Space line when the operating unit is opened or internal calibration is carried out to protect the patient and prevent an uncontrolled drug flow to the patient Beside the position monitoring via the door bolt the open closed state of the safety clamp is monitored by a light barrier The safety clamp is closed and the Infusomat Space line is squeezed mechanically when the slide clamp of the Infusomat Space line is inserted When the operating device is opened the slide clamp is pulled out of the safety clamp until the Infusomat Space line is squeezed The safety clamp is only opened by manual operation of the safety clamp lever and the Infusomat Space line can be removed To ensure patient safety the operating unit is automatically locked For this purpose the operating unit is manually pressed in the door bolt with the three locking pins The resistance of the lin ear potentiometer which is monitored by the controller is changed when the position of the door bolt changes The locking process is started and can take up to 10 seconds During the lock ing process several calibrations are carried out automatically e g occlusion test to check for example the function of the safety clamp pressure calibration of the two pressure sensors The operating unit is also unlocked aut
76. nner frame Fig 3 21 Item 3 2 Push the bracket Fig 3 21 Item 6 forward against spring tension unhook from the inner frame and remove it together with the pressure adjustment unit Fig 3 21 Item 7 3 Press the pump frame guide 6 mm straight pin Fig 3 21 Item 8 in longitudinal direction out of the linear bearings of the pump frame and the inner frame holders and remove the pump frame from the inner frame 3 21 Disassembly Assembly T ios Fig 3 21 Legend to fig 3 21 ItemDesignation 1 ao A CO N Door bolt 6 Bracket Countersunk screw M3x5 TORX 7 Pressure adjustment unit Inner frame 8 Pump frame Door bolt drive 9 Guide 6 mm straight pin Linearpotentiometer mit Kabelbaum 10 Bracket holder 3 22 Infusomat Space 1 1 gb Fig 3 22 Legend to fig 3 22 ItemDesignation 1 Screw DELTA PT 30x8 WN 5452 TORX plus 10IP 2 Pump drive motor Infusomat Space 1 0 gb Disassembly Assembly 4 Unscrew one screw Fig 3 22 Item 1 turn the pump drive motor Fig 3 22 Item 2 clockwise out of the holder and remove it from the pump frame 5 Move the door bolt drive by hand in the middle position 6 Unscrew two screws Fig 3 21 Item 2 and pull the door bolt drive Fig 3 21 Item 4 off the door bolt tappet Fig 3 21 Item 1 Note The door bolt drive is slightly engaged when installed When dis mantling the driv
77. nsion engages and the caution sign flashes yellow Test Setup Perform test setup with the subassemblies listed below please see also Fig 7 2 Note The liquid level in the container must be appr 50 cm above the middle of the unit opening for the Infusomat Space line Infusomat Space line standard type new can be used for the complete TSC incl functional check 1 piece Infusion bag or bottle min 100 ml 1 piece Three way valve 2 pieces for measurement with electronic pressure gauge 10 ml syringe air buffer for measurement with electronic pressure gauge syringe drawn up to 10 ml and piston fixed mechanically 1 piece 1 ml syringe for air bubble injection 1 piece Electronic pressure gauge with peak value recognition 1 piece Pressure cell alternative of electronic pressure gauge Graduated cylinder 25 ml 0 4 ml 1 piece Infusomat Space 1 0 gb 2 1 Fig 7 3 yu 2 1 Fig 7 4 Infusomat Space 1 0 gb Procedural Instructions on the TSC Pressure Cut Off Electronic The service connector Space must not be connected with the de vice The optional drop sensor must not be connected 1 Insert the Infusomat Space line of the test setup Fig 7 2 in the device 2 Enter a delivery rate according to the TSC and a volume of 250 ml 3 Select pressure stage according to the TSC 4 Vent test setup position of the three way cocks please see Fig
78. omatically via the door bolt In case of a power failure or malfunction the operating unit can be unlocked manually by a small emergency unlocking crank To do so move the door bolt manually with the crank to the open position as described hereafter System Overview Opening the Operating Unit Manually 1 Press the locking of the battery compartment cover using a pointed tool and remove the battery compartment cover from the housing 2 Remove the emergency unlocking crank out of the battery compartment cover Fig 1 3 3 Press the emergency unlocking crank carefully in the emer gency unlocking plug on the left side of the housing until stop Note The hexagon socket of the emergency unlocking plug is tapered to the inside If the emergency unlocking crank is not inserted until stop the hexagon socket may be damaged Fig 1 4 1 6 Infusomat Space 1 1 gb Fig 1 5 Fig 1 6 Infusomat Space 1 1 gb System Overview 4 Turn the emergency unlocking plug carefully with the emer gency unlocking crank through 90 to the right clockwise in arrow direction until stop and remove the plug out of the housing 5 Insert the emergency unlocking crank in the housing opening The crank must be pushed in the hexagon socket of the door bolt drive CAUTION Make sure that the emergency unlocking crank is not turned be yond the stop by force 6 Turn the emergency
79. op sensor optional Delivery rate 250 ml h O Drop sensor alarm occlusion O Drop sensor alarm flow Air inline sensor according to TSC Infusomat Space line standard type Water value Air alarm O O O Difference of temperature sensors O Air value Switch on unit without power supply O Self test O Magnetic function of the battery com partment cover C Charge state of the battery module Remove battery module O Alarm for min 3 minutes Part 2 of 2 3 46 Infusomat Space 1 0 gb Cleaning Servicing the Battery Infusomat Space 1 0 gb Servicing the Unit 4 Clean and disinfect the Infusomat Space with a humid cloth at regular intervals To clean the system we recommend mild soap suds WARNING WHILE CLEANING AND DISINFECTING THE INFUSOMAT SPACE DISCONNECT THE UNIT FROM THE MAINS SUPPLY CAUTION Hold the device horizontally and make sure that no fluids pene trate in the device through the safety clamp or the locking open ings The pushers of the pressure sensors and the air inline sensor in the operating unit must not be pressed during cleaning For disinfection by wiping you should use for example Melisep tol from B Braun Allow the unit to dry for at least one minute When you disinfect the device by spraying make sure not to spray in the system openings such as interface sockets and connectors loudspeaker opening The instructions for use contain a detailed de
80. or Loudspeaker Check the loudspeaker 4 2016 Loudspeaker shorted Loudspeaker connector Check the loudspeaker connector Loudspeaker Check the loudspeaker 5 2017 KUP no work Software defective or obsolete Update software change ISP proc 6 2018 FUP KUP OS Cycle Timeout Software defective or obsolete essor PCB if necessary 7 2019 FUP Flash Memory Error Software defective or obsolete 8 2020 FUP different version KuP to FuP Software defective or obsolete 9 2021 FUP UPSTREAM TOO HI Upstream pressure sensor defec Replace sensor tive 10 2022 FUP UPSTREAM TOO LOW Upstream pressure sensor defec tive 11 2023 FUP DOWNSTREAM TOO HI Downstream sensor defective 12 2024 FUP DOWNSTREAM TOO LOW Downstream sensor defective 13 2025 FUP EA key closed to long Keyboard defective Exchange operating unit 14 2026 FUP Task inactive 15 2027 FUP Airsensor CALDATA CRC Air inline sensor defective Replace sensor 16 2028 FUP Airsensor PROGDATA CRC Air inline sensor defective Replace sensor 17 2029 FUP Airsensor Calibration Error Air inline sensor defective Replace sensor 18 2030 No Airtest Kup Air inline sensor defective Replace sensor 19 2031 FUP Temperature Sensors differ Air inline sensor defective Replace sensor ence 20 2032 FUP Temperature Raw Values dif Air inline sensor defective Replace sensor ference 21 2033 FUP Airsensor Signature Error Air inline sensor defective Replace sensor 22 2034 FUP CAN paket scheduler list fail Communication error on C
81. ot damaged Pump slide 12 pieces smooth running and engaged Coding of pump slide guide present Pump connecting rod 12 pieces not damaged Air inline sensor free from oil and ultrasonic gel free from cracks Pressure webs for air inline sensor smooth running movable without jamming Drop sensor lens Screw covers Connectors P2 and P3 Completeness and correct position of the insulating washers on the operating unit back panel Completeness and readability of both pictograms on the inside of the operating unit Procedural Instructions on the TSC Electrical Safety according to IEC EN 60601 1 or VDE 0750 and VDE 0751 N wo Fig 7 1 Legend to fig 7 1 ItemDesignation 1 Mains connection 2 Primary of the powper supply 3 Secondary of the power supply 4 Service adapter SP with probe 5 Leakage current measuring device 7 2 Power Supply SP 1 Check the Power Supply SP and connecting line for cleanli ness 2 Check the Power Supply SP incl connecting line for damage and the labels for readability Infusomat Space The patient and housing leakage current of the Infusomat Space is caused exclusively by the operating voltage supply Power Sup ply SP or SpaceStation The Technical Safety Checks of the poer supply SP drawing No M001 32 10 05 F04 or of the SpaceStation drawing No M690 00 00 46
82. p opens 686B030003 Improved functions CAN bus functioning 686B030004 Improved functions 686C030001 Dose calculation Changed CAN log 686D030001 Drug list data base Changed user language 686E030002 Improved functions Piggyback Softlimits Software Update of the Unit The instructions for updating the software are supplied with the software itself CAUTION If the device is disconnected while the software is being updated or the device or PC is switched off a component of the software may be seriously damaged so that repairs are no longer possible In such a case the software cannot be updated via the PC and the device must be returned to B Braun System Overview Service Program Approved Versions Note Please note that text and or functions of the Service Program may change depending on the software version The following screen illustrations are only examples and represent the state when the manual was printed 1 1 4 1 2 1 1 3 5 137 150 2 0 1 3 1 0 Note Installation and further operation of the Service Program is de scribed in its separate instructions for use HiBaSeD security advice Pi ther devi t not b ted t tient whil A ns motes 1 Start the HiBaSeD exe program History Barcode Service Drug list on the PC The Service Program is loaded and start Please observe the instructions for use of the software and the connected hardwar
83. ply ord no ing safety Note 3310 2708 This measurement ensures that the admissi Check lock of primary adapter with ble limit values regarding the patient and power supply The adapter must engage housing leakage current of the Perfusor on both sides Space or the Infusomat Space are met Mechanical Aids and Measuring Equip ment Used J Perfusor Space serial No O Infusomat Space serial No O M001 32 10 05 F04 3891 6142 GB B BRAUN Infusomat Space 1 1 gb SHARING EXPERTISE 6 1 Sheet 1 of 2 Technical Safety Check TSC Index c Master to be added to the documentation For your notes B BRAUN M001 32 10 05 F04 3891 6142 GB 6 2 SHARING EXPERTISE Infusomat Space 1 0 gb Sheet 2 of 2 Procedural Instructions on the TSC 7 Visual Inspection Infusomat Space 1 Check the Infusomat Space and accessories for cleanliness 2 Check the Infusomat Space and accessories for complete ness and check configuration 3 Check the Infusomat Space and its accessories for damage and the labels for readability Pay special attention to the fol lowing parts a Safety clamp b Pressure sensor membrane downstream c Infusomat Space 1 0 gb not damaged Pressure sensor membrane upstream not damaged Pressure webs for pressure sensors smooth running movable without jamming Pump membrane n
84. red back to the device see Processor PCB gt p 3 32 Disassembly Assembly Service Parts and Screw Kit All small parts such as cover caps are contained in an Infusomat Space service part kit Designation Ord No Service part kit Infusomat Space 3452 1593 with housing cover cap 100 pieces operating unit cap 5 pieces housing foot SP 10 pieces snap in hook SP with leaf spring 2 pieces emergency release plug ISP incl O ring 2 pieces emergency release crank ISP 1 piece bracket locking 20 pieces seal washer PA 2 2x0 3 200 Stiick safety clamp seal housing bottom part 10 pieces Cover cap kit for housing SP 50 pieces 3477 4386 Emergency release plug ISP 10 pieces 3477 3101 Connector seal P2 50 pieces oooo oo o 3477 3102 Housing foot SP 20 pieces oocoocooo On request All screws used in the device are included in an Infusomat Space screw kit Designation Ord No Screw kit Infusomat Space hh kk k 3452 1585 with screw DELTA PT 22x8 WN 5451 TORX plus GIP 5 pieces screw DELTA PT 25x7 WN 5451 TORX plus 8IP 5 pieces screw DELTA PT 30x8 WN 5452 TORX plus 10IP 10 pieces screw DELTA PT 30x12 WN 5452 A2 TORX plus 101P 10 pieces countersunk screw DELTA PT 20x9 WN 5454 TORX plus GIP 5 pieces fillister head screw M3x6 TORX fillister head screw M3x10 countersunk screw M3x5 TORX countersunk screw M3x12 A2 TOR
85. rnummer Bitte geben Sie die Daten ein Werkernummer Seriennummer OK Infusomat type Please select the desired Infusomat type for the current calibration procedure Tap ami Infusomat Space P Fig 2 8 EN Infusomat type selection UD Mer i A VENIR Inu E peli H ni O Saa TI Infusomat Space S Article number 8713070 Article number 8713050 Infusomat type selection Fig 2 9 The type of the Infusomat will be set Does the type definitely correspond with the device See type plate on base of device Abbrechen Infusomat Space 1 1 gb Unit Diagnosis Calibration 4 Input your user number in the window User number as well as the six digit serial number of the device if necessary 5 Confirm the input with OK The frame Data selection is now activated Note If HiBaSeD could not clearly read the device serial number the number must be entered according to the rating plate 6 If the device type was not clearly recognized the user is prompted to select a device type Select the device type by clicking the flag with the mouse Note The device type can be defined via the article number This article number is indicated on the rating plate 7 Click the OK button You are prompted to confirm the se lection again 8 Click the OK button 2 15 Unit Diagnosis Calibration HiBaSeD International Service Program File Extras
86. rouble shooting Part 2 of 2 2 20 Infusomat Space 1 0 gb 3 1 General Infusomat Space 1 0 gb Disassembly Assembly 3 Remarks on Disassembly Assembly Before disassembling the unit the system must be checked see Device Check 9 p 2 9 to isolate the part to be exchanged The necessary steps to disassemble the complete unit all its sub systems and spare parts are detailed in the following description Steps that are not necessary can be skipped Before disassembling the unit make sure that all Infusomat Space lines are removed from the device and that the operating unit is in the open position CAUTION Remove the battery module before opening the device to avoid damage to the system or the battery module through a short cir cuit Note Special screws for plastic housings are used in this unit Pay at tention to the corresponding notes when you fit the screws Disassembly Assembly Preparations for Exchanging the Processor PCB If the processor PCB is to be replaced a back up of the pump set tings is to be carried out if this is still possible Note Please note that text and or functions of the Service Program may change depending on the software version The following screen illustrations are only examples and represent the state when the manual was printed 1 Start the Service Program see Starting the Service Program p 1 10 File Extras Tools Update Help V ie
87. scription on how to service the battery If a battery module is not discharged completely for more than 28 days a servicing program for the battery module can be started on the unit Servicing the Unit For your notes 4 2 Infusomat Space 1 0 gb Technical Safety Check TSC Index d Master to be added to the documentation A Observe the Service Manual and the instructions for use All measured values are to be documented Checklist for Technical Safety Check Every 24 Months Unit Infusomat Space Manufacturer B Braun Melsungen AG Accessories used should be included in testing Make exclusive use of calibrated measuring equip ment Visual Inspection Electrical Safety Functional Inspection according to IEC EN 60601 1 or VDE 0750 and VDE 0751 O Infusomat Space The patient and housing leakage current of Locking with second unit Cleanliness completeness damage and the Infusomat Space is caused exclusively fJ Battery compartment cover faults affecting safety damage and by the operating voltage supply Power Sup readability of the label ply SP or SpaceStation Switch on unit with power supply Particularly The Technical Safety Checks of the power Indicator lamps LEDs J Connectors P2 and P3 supply SP drawing No MOO1 32 1005 F04 CJ Self test O Safety clamp or of the SpaceStation drawing No M690 0 Audible alarm O Pressur
88. sembly 1 Turn the emergency release plug Fig 3 10 Item 1 using a spanner e g the emergency release crank in the battery compartment cover through 90 to the right and pull the plug out of the housing Note The spanner must be pressed carefully until stop in the opening of the emergency release plug The hexagon socket of the emergency release plug tapers to the inside to guarantee a certain bond when the plug is removed Infusomat Space 1 1 gb Disassembly Assembly 3 5 Unit Foot Designation Ord No Unit foot see Service Parts and Screw Kit p 3 8 Disassembly 1 Pull the unit foot Fig 3 11 Item 1 out of the housing Fig 3 11 Legend to fig 3 11 ItemDesignation 1 Unit foot Infusomat Space 1 0 gb 3 11 Disassembly Assembly 3 6 Battery Module Fig 3 12 Legend to fig 3 12 ItemDesignation 1 Battery compartment cover 2 Battery compartment cover lock 3 12 Designation Ord No Battery compartment cover ISP cpl 3452 1321 incl emergency release crank Battery pack SP NIMH KAK KK eee 0871 3180 Disassembly 1 Press the lock Fig 3 12 Item 2 on the battery compart ment cover Fig 3 12 Item 1 with a pointed tool and re move the battery compartment cover Infusomat Space 1 0 gb Legend to fig 3 13 ItemDesignation 1 Battery pack 2 Battery compartmen
89. senhausenerstrasse 7 D 81379 M nchen Tel 0049 0 897876 0 Fax 0049 0 897876 333 Notfallauskunft 0049 0 897876 700 Further information obtainable from Material Compliance Management Chemical characterization Description mineral oil Synthetic hydrocarbon oil special lithium soap UV indicator Dangerous components CAS 597 82 0 O 0 O Triphenylthiophosphate R53 lt 2 5 EINECS 209 909 9 Additional information For the wording of the listed risk phrases refer to section 16 Hazard description Not applicable Information concerning particular hazards for human and environment The product does not have to be labelled due to the calculation procedure of the General Classification guideline for preparations of the EU in the latest valid version Classification system The classification is according to the latest editions of the EU lists and extended by company and literature data After inhalation Supply fresh air consult doctor in case of complaints After skin contact Wash off with soap and plenty of water Contd on page 2 GB Fig 10 1 Safety data sheet Kl ber POLYLUB GLY 501 Part 1 of 5 Infusomat Space 1 1 gb 10 1 Safety Data Sheets Fig 10 1 Page 2 4 Material Safety Data Sheet According to 91 155 EEC Printing date 26 01 2005 Reviewed on 29 03 2004 Trade name POLYLUB GLY 501 Contd of page 1
90. t cover 3 Battery pack lock Infusomat Space 1 0 gb 2 Disassembly Assembly Lift the lock Fig 3 13 Item 3 on the battery pack Fig 3 13 Item 1 and remove the battery pack out of the device 3 13 Disassembly Assembly 3 7 Housing Upper Part Designation Ord No Housing upper part ISP WAAR KK eee 3452 1313 Screws and cover caps see Service Parts and Screw Kit p 3 8 Note Please pay attention to the corresponding notes during assembly and installation see Housing Upper Part p 3 36 Disassembly 1 Pierce six cover caps Fig 3 14 Item 1 with a small screwdriver and pry the cover caps out 1 2 Unscrew six screws Fig 3 14 Item 2 a 2 Note e The other 4 cover caps on the housing bottom side must not be L2 T removed during this disassembly state Fig 3 14 Legend to fig 3 14 ItemDesignation 1 Cover cap 2 Screw DELTA PT 30x12 WN 5452 A2 TORX plus 10IP 3 14 Infusomat Space 1 0 gb Fig 3 15 Legend to fig 3 15 ItemDesignation 1 Snap on openings Infusomat Space 1 0 gb Disassembly Assembly 3 Loosen the snap on mechanism of the housing upper part Fig 3 15 Item 1 carefully To do so a screwdriver can be inserted in the two openings between the housing front panel and the housing top Note It is also possible to remove the housing upper part by gripping into the battery
91. t panel from right to left pressure sensor upstream container side Slide guide with 12 slides mechanically coded can be dis mantled without tool coding for ISPS top left coding for ISPP middle right airinline sensor pressure sensor downstream patient side safety clamp ISPS The slides are moved in the slide guide by pump connecting rods The connecting rods are mounted on an eccentric shaft and are led outside through a seal diaphragm The complete pump is flex ibly mounted in the inner frame of the device and is moved in combination with the door bolt The door bolt drive is also in stalled in the inner frame The processor PCB with the external connectors P2 and P3 is located at the bottom of the housing bottom part Infusomat Space 1 0 gb System Overview Fig 1 2 Infusomat Space Legend to fig 1 2 ItemDesignation 1 Infusomat Space 2 Operating unit 3 Cover for drop sensor connector Infusomat Space 1 0 gb Connector P2 for SpaceStation module external 12 V DC and accessories Connector P3 connection to SpaceControl module Battery compartment cover Function System Overview There are two power options for the Infusomat Space via the inserted battery module via an external 12 V DC power supply e g SpaceStation SpaceControl an external power supply or from an ambu lance car connected to connector P2 The voltage supplied is converted to
92. tion Procedural Instructions for Calibration Infusomat Space 1 0 gb Test Equipment Designation Slide clamp of the Infusomat Space line Procedure Note Calibration must be carried out with power supply connected since the calibration can be interrupted suddenly if the unit is battery operated and the battery gets discharged so that the de vice is switched off Note Please note that text and or functions of the Service Program may change depending on the software version The following screen illustrations are only examples and represent the state when the manual was printed 2 13 Unit Diagnosis Calibration 1 Start the Service Program see Starting the Service Program p 1 10 2 Select the register tab Calibration HiBaSeD International Service Program File Extras Tools Update Help El i ieu Infusomat Space S Pressure Levels Tests y Disconnect om Calibration Cal update Serial number Disposable articles Modification data Operating data Delivery Constant 2006 10 09 00 00 00 Calibration procedure a 1823 Infusomat Space 5 com3 Fig 2 5 3 Press the New device button in the frame Calibration pro cedure The window User number is opened Note If you do not have an allocated user number enter 0001 2 14 Infusomat Space 1 1 gb Fig 2 6 Worker ID Please insert data Worker ID 0001 Fig 2 7 Werke
93. tion of the door bolt in the device The door bolt lugs must not be visible in the three openings It must be possible to tilt the operating unit nearly in the closed position gap between upper edge of the operating unit and upper edge of the housing top is about 5 mm Check the gap and confirm the door bolt position by pressing the Yes button Check the position of the door bolt in the device The door bolt lugs are visible in the three openings It must not be pos sible to tilt the operating unit in the closed position gap be tween upper edge of the operating unit and upper edge of the housing top is about 10 mm Check the gap and confirm the door bolt position by pressing the Yes button The door bolt is moved in the open position After calibration you are prompted to switch the device off Disconnect all connectors from the device Switch the device off without connection to the power sup ply and press the OK button 2 17 Unit Diagnosis Calibration BR Report Report Print Software Version 686E030002 Software Date 2006 10 09 00 00 00 Calibration Calibration process completed successfully Fig 2 17 2 18 18 At the end of calibration the result with all the values is dis played in the frame Calibration process terminated success fully and data is stored in the unit This report can be printed out by pressing the Print button 19 With OK the calibratio
94. ular inter vals a current manual which corresponds to the revision state is used when carrying out maintenance repair and service the service technician takes part in the revision service the technician has participated in a technical training course for the specific B Braun unit B Braun is certified in accordance with DIN EN ISO 9001 and ISO 13485 This certification also includes maintenance and serv ice The unit has the CE label The CE label confirms that the device corresponds to the Directive of the Council for Medical Products 93 42 EC of June 14 1993 Training may only be performed by B Braun The possession of the manual does not authorize the performance of repairs The in structions on electrostatic sensitive components ESD standards must be observed After repair a device check or diagnosis is to be carried out Semiconductors can be destroyed by electrostatic discharge Es pecially MOS components can be damaged by interference from electrostatic fields even without discharge via contact This type of damage is not immediately recognizable Unit malfunctions can even occur after a longer period of operation Infusomat Space 1 1 gb Spare Parts and Test Equipment Setting Off Infusomat Space 1 1 gb Important Preliminary Remarks Each workstation must be equipped according to the recommen dations with the necessary static protective measures if ESD components or boards ar
95. usomat type selection Fig 1 14 The type of the Infusomat will be set Does the type definitely correspond with the device See type plate on base of device Abbrechen Infusomat Space 1 1 gb System Overview If the unit software version is not compatible with the Service Program version a window opens prompting the operator to change the Service Program version This window displays a compatibility list of the Service Program and unit software versions If Service Program and unit software versions are compati ble all the Service Program functions are activated If the software was updated or the processor PCB was ex changed you are prompted to enter the device type You can select either the Infusomat Space P or the Infusomat Sace S Note The designation Infusomat Space S corresponds to the Infuso mat Space 6 The device type must be confirmed again If the Service Program and the unit software are compatible all the Service Program functions are activated 1 13 System Overview HiBaSeD International Service Program File Extras Tools Update Help E M ieu Infusomat Space S com3 Calibration 686E030002 2006 10 09 00 00 00 1823 Calibration procedure Tests e Disconnect Serial number Disposable articles Modification data Operating data Delivery Constant Modify New device Infusomat Space 5 com3 Fig 1 15 Service Program
96. valve 1 7 Three way valve 2 8 Graduated cylinder Distance between liquid level and middle of the Infusomat T 2 1 Fig 2 3 2 12 Test Equipment Designation Infusomat Space line standard type new unused 1 piece Infusion bag or bottle min 100 ml 1 piece Three way valve 2 pieces 10 ml syringe air buffer for measurement with electronic pressure gauge 1 piece syringe drawn up to 10 ml and piston fastened mechanically e g with aspiration lever for Sonocan part No 0480 3000 Electronic pressure gauge with peak value recognition or baro metric cell 1 piece Graduated cylinder 25 ml 0 4 ml 1 piece Test Setup 1 Mount test setup as shown in Fig 2 2 Note The liquid level in the container must be appr 50 cm above the middle of the unit opening for the Infusomat Space line 2 Vent Infusomat Space 1 0 gb Fig 2 4 Calibration 1 Connect unit to PC with interface cable Unit Diagnosis Calibration 3 Measure pressure C52 2 Start Service Program on the PC see Starting the Service Program p 1 10 The desired device is found by the Service Program CS 3 and then displayed 3 Start calibrating the unit see Procedural Instruc tions for Calibration p 2 13 CS 4 4 Close the Service Program see Quit the Service Program p 1 15 Table 2 6 Calibra
97. x Infusomat Space S com3 686E030002 2006 10 09 00 00 00 1823 HiBaSeD International Service Program Pressure Levels o Select a data source to modify From file Default Accept or reject modified data HA ee Accepted data must be saved odeve tome coca Lew mu Tests s Disconnect Calibration Cal update Serial number Disposable articles Modification data Operating data Delivery Constant Fig 3 1 Infusomat Space 1 1 gb Disassembly Assembly 2 Select the register tab Modification data 3 Press the From device button The data is read from the de vice and shown on the display HiBaSeD International Service Program File Extras Tools Update Help f Infusomat Space S Pressure Levels Tests Disconnect as Calibration Cal update Serial number Disposable articles Modification data Operating data Delivery Constant T 40 09 00 00 00 Service Therapy lo Name Rates KYO Other Status Options Infusomat Language Time Alarms e N identification gt Data Lock PIN Fig 3 2 4 Select the tab 10 and actuate the To device button In the window which opens now you are asked for the storage position of the file on the PC hard disk and the file name Note The storage position which is proposed should not be changed Infusomat Space 1 1 gb Disassembly Assembly Fig 3