1. 17 Section IA eR Rem 18 SectionX X Electromagnetic Conformity Declaration 20 This product contains software solely owned by Covidien Covidien grants the user a non exclusive limited license to use the software according to the operating instructions A copy of the license can be obtained from Covidien Kangaroo 924 Enteral Feeding Pump with Pole Clamp Section General Information The Covidien Kangaroo 924 Enteral Feeding Pump with Pole Clamp is a rotary peristaltic pump designed to regulate the flow rate of enteral feedings Among the features provided by this pump are Portable operation built in rechargable battery Wide flow rate range 1 300 mL hr in 1 mL hr increments Small and lightweight for easy transport Language Selection 13 languages available Touch panel and knob operation Shut off and activation of audible and visible alarms when Feeding container is empty Feeding tube becomes occluded Drip sensors blocked Battery low alarms 15 minutes before pump shut off Unitisleft on Hold longer than 5 minutes Feeding set is improperly loaded Self diagnosed problem which requires technical servicing Pump motor malfunction Ambient light error Automatic systems check after pump has been turned ON Large LED display Low occlusion pressure Pump will not overcome back pressure greater than approximately 15 psi 103 4 kPa
2. Integral pole clamp Use of Optix feeding sets to prevent free flow from inadvertent misload of feeding set onto pump Pump running indicator lights DC motor with intermittent operation Adjustable alarm volume Memory of flow rate setting is retained for 16 or 24 hrs after pump shut off depending on AC connection status AN Warning This Kangaroo 924 Enteral Feeding Pump with Pole Clamp requires the use of a Kangaroo 924 Feeding Set The pump will not operate if sets other than Kangaroo 924 Feeding Sets are loaded into the pump Kangaroo 924 Enteral Feeding Pump with Pole Clamp 1 Table of Contents Figure 2 Handle Upper Drip Chamber Guide including drip sensor Tubing Guide Lower Drip Chamber Guide Optix Lock Rotor Assembly Buzzer Hole LED Display and Pump Running Indicator Lights Kangaroo 924 Enteral Feeding Pump with Pole Clamp 2 Pole Clamp Pole Clamp Knob AC Light VOL CLR Light Adjustment Knob Sensor Lens Tubing Latch Touch Panel Table of Contents Section Il Safety and Warnings Caution U S federal law restricts the sale of this device to physicians or to their direct representatives 1 Read this booklet thoroughly before using the Kangaroo 924 Enteral Feeding Pump 2 Do not use this device near flammable anesthetics 3 Disconnect power supply before cleaning or servicing Inspect power cord for damage and send in for service if damaged 4 Use only Kangaroo 924 Enteral Feeding Pump po
3. b has been subjected to unauthorized or non Covidien repair or use of non Covidien supplied parts has been altered misused abused or neglected e hasbeen subjected to fire casualty or accident seas e suffers damage caused by Customer s negligent acts or omissions or f suffers damage beyond normal wear and tear 5 For purposes of this Limited Warranty damage beyond normal wear and tear includes without limitation a Damage to housing LCD display overlay or power supply b PCBA damage due to fluid ingress Use of non qualified power supply or battery or d Use of unauthorized cleaning fluids 6 Ifa Pump does not operate as warranted during the applicable warranty period Covidien may at its option and expense a repair or replace the defective part or Pump or b refund to Customer the purchase price for the defective part or Pump 7 Dated proof of original purchase is required to process warranty claims Removal defacement or alteration of serial lot number voids this Limited Warranty 8 Shipping costs for Pumps being returned to Covidien shall be borne by Customer Customer is responsible for proper packaging for return shipment Loss or damage in return shipment to Covidien shall be at Customer s risk Kangaroo 924 Enteral Feeding Pump with Pole Clamp 18 Table of Contents 9 Covidien disclaims all other warranties expressed or implied including any implied
4. AU Power Cord Part Number F090475 AC Light On whenever pump is plugged in VOL CLR Light On when volume delivered has been cleared ON Powers the pump See Figure 3 for location OFF Shuts pump off Note To disconnect AC power it is necessary to unplug power cord from the AC wall outlet See Figure 3 for location A Increases rate or VTBD desired W Decreases rate or VTBD desired Note Numbers displayed will change more rapidly if either arrow button is depressed and held Figure 3 Adjustment Knob Increases or decreases rate or VTBD desired VOL Displays volume delivered for 5 seconds when pressed When the pump has been off for 24 hours 16 hours battery the volume delivered is automatically reset to zero VOL CLR Clears volume delivered when pressed after VOL has been pressed The VOL CLR light will go on when volume delivered has been reset to zero Volume delivered cannot be cleared until the preset VTBD has been delivered or cleared VTBD Volume To Be Delivered When pressed allows user to preset a volume to be delivered by pressing A or or rotating adjustment knob Holding for 5 seconds will clear VTBD setting and volume VOL CLR light flashes 3times to signal that VTBD has been cleared START HOLD Starts pump rotor to deliver feeding formula to patient and alternately puts unit on Hold Hold is used to stop fluid flow to 1 correct alarm condition alarm is silenced 2 change the flow rate 3 change administrative
5. Clamp 6 Table of Contents Warning Do not stretch the silicone tube beyond what is required to load the set Over stretching the silicone tube can result in an over delivery of formula to the patient Warning Verify that the feeding set is properly loaded into the pump before proceeding An improperly loaded feeding set could result in uncontrolled flow and possible patient injury 13 Press ON 14 Wait while pump performs a system check Displays B B B B fora moment 15 Check VOL CLR light or press VOL to insure that volume delivered is clear if desired 16 Setthe delivery rate The rate may be changed when the delivery rate is shown on the display by using the adjustment knob or by pressing the A or W buttons Depressing and holding the button will cause the numbers to increase decrease more rapidly 17 Set Volume To Be Delivered VTBD if desired Press VTBD then rotate adjustment knob or press A W buttons to set VTBD desired 18 Open the control clamp 19 Press START HOLD 20 Proceed with feeding 21 When feed complete or after 24 hour use dispose of feeding sets set not to exceed 24 hours of use Used feeding sets shall be handled as follows To prevent employee exposure risk the materials must be handled in accordance UJ with bloodborne pathogen standards before being placed in an approved disposal container Disposal must be performed in accordance with current medical practices or local regulations in r
6. Community Certification abel i Assure that the white cap and red stepped 0123 Catheter connector are securely tightened Kangaroo 924 Enteral Feeding Pump with Pole Clamp 17 Table of Contents SectionIX Warranty Limited Warranty 1 Covidien warrants to the original purchaser Customer that this newly manufactured enteral feeding pump Pump or Pumps will be free of defects in materials and workmanship under normal use for three 3 years from the date of shipment from Covidien This Limited Warranty as applied to Pump batteries and power cords is limited to one 1 year from the date the date of shipment from Covidien for all Pumps 2 This Limited Warranty does not extend to routine maintenance of Pumps such as cleaning and all recommended Performance Tests set forth in this Pump Operation and Service Manual which remain the sole responsibility of Customer Failure of Customer to perform cleaning routine maintenance and recommended performance testing on any Pump as outlined in this Pump Operation and Service Manual may void this Limited Warranty 3 Customer agrees that with the exception of customer serviceable parts and troubleshooting steps outlined in this Pump Operation and Service Manual Covidien or its authorized dealer must perform Pump repairs 4 This Limited Warranty does not cover any Pump product or part that has been operated in an unsuitable environment or used for purposes other than intended
7. Directions before starting e Clean lens the Optix lock area with a cotton swab dampened with isopropyl alcohol e Besure sensor areas of the lens are clear e Avoid using abrasive cleaners to prevent scratching damaging the lens Kangaroo 924 Enteral Feeding Pump with Pole Clamp 10 Table of Contents Directions For Cleaning Rotor Assembly e Referto General Cleaning Directions before starting e Loosen rotor set screw with 5 64 2 mm Allen wrench and gently pull rotor forward off shaft After removing rotor avoid getting any moisture in the rotor shaft opening e Clean rollers thoroughly with warm soapy water or isopropyl alcohol if necessary e Be sure all parts of rotor are completely dry before putting it back onto shaft e To replace rotor align set screw on hub of rotor with the flattened section of the output shaft e Push rotor into place and tighten set screw do not over tighten Preventive Maintenance This pump must be periodically serviced to assure proper functioning and safety The recommended service interval is six 6 months but in no case less than once each twelve 12 months Servicing must be performed by personnel trained in servicing Kangaroo Enteral Feeding Pumps and may be done at the user s Biomedical Engineering department an outside service or by Covidien Factory Service To arrange for Covidien Factory Service call 1 800 448 0190 in the US or contact your local customer service cente
8. Voltage fluctuations Complies flicker emissions IEC 61000 3 3 Radiated Disturbance Complies Immunity EN60601 1 2 IEC 61000 4 3 2002 Conducted Disturbance Complies Immunity EN60601 1 2 IEC 61000 4 6 2001 Voltage dips andsags Complies Immunity EN60601 1 2 IEC 61000 4 11 2001 Electrical Fast Transient Complies Bursts Immunity EN60601 1 2 IEC 61000 4 4 2001 Electrostatic Discharge Complies Immunity EN60601 1 2 IEC 61000 4 2 2001 Surge Immunity Complies EN60601 1 2 IEC 61000 4 5 2001 Power Frequency Complies Magnetic Field Immunity EN60601 1 2 IEC 61000 4 8 2001 Kangaroo 924 Enteral Feeding Pump with Pole Clamp 20 Table of Contents Recommended separation distances between portable and mobile RF communications equipment and the Kangaroo 924 Enteral Feeding Pump The Kangaroo 924 Enteral Feeding Pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Kangaroo 924 Enteral Feeding Pump can help prevent electromagnetic interference by maintaining the minimum distance between portable and mobile RF communications equipment transmitters and the Kangaroo 924 Enteral Feeding Pump recommeneded below according to the maximum output power of the communication equipment Rated maximum output Separation distance according to frequency of transmitter power of transmitter m 150 kHz to 80 MHz 80 MHz to 8
9. set When on Hold the LCD display flashes on and off and the total delivery is retained Note If the pump has been placed on Hold and not restarted within 5 minutes an alarm will sound Normal Operation Insert power cord into AC wall outlet Kangaroo 924 Enteral Feeding Pump with Pole Clamp 4 Table of Contents Battery Operation Unplugging the pump from the AC line outlet will automatically put the pump on battery power new battery when fully charged will operate the instrument for more than 16 hours at 125 milliliters per hour Approximately 15 minutes prior to battery discharge a low battery alarm will occur see alarm section When complete discharge occurs the pump will automatically turn itself off Note As the batteries get older the time from low battery alarm to complete discharge may be less than 15 minutes To recharge battery insert power cord into AC wall outlet Battery will automatically begin charging The recharge time for a fully discharged battery is approximately 12 hours The battery will charge whenever the pump is plugged into a wall outlet even when the OFF button has been pressed To be certain that power to all circuits has been disconnected the power cord must be unplugged from the wall outlet Battery Replacement battery packs are available from Covidien See Section VII Factory Service AN Warning Replace the battery pack only with the same type and rating of battery Used battery packs m
10. until the START HOLD button is pressed again Kangaroo 924 Enteral Feeding Pump with Pole Clamp 8 Table of Contents VTBD Delivered Vtbd dEL Preset VTBD has been delivered to patient When preset VTBD has been delivered perform one of the following Repeat the VTBD amount by pressing START HOLD after clearing the alarm Cancel the VTBD feature by pressing and holding VTBD for 5 seconds Tofeed additional volume increase VTBD amount by first pressing VTBD Then rotate adjustment knob or press A until new VTBD is displayed Tum pump off Note For all alarm conditions pressing the START HOLD button will silence the audible alarm The visible display will continue to flash until the START HOLD button is pressed again The alarm volume is set using the front panel in one of two ways 1 Using the Biotech Mode as described See Section VI 2 While the alarm is sounding press to increase the volume press V to decrease the volume or rotate adjustment knob The volume is changed in steps from 1 to 10 with 1 the quietest and 10 the loudest AN Warning If alarm does not sound during system check or when any alarm condition exists see No Audible Alarm symptom in Troubleshooting guide Section V Cleaning Cleaning is to be performed as required The user will define the cleaning intervals based on their knowledge of the environment in which the pump is used Cleaning may become necessary between inte
11. 00 Mhz 800 MHz to 2 5 GHz 1 2VP d 1 2VP d223JP For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer Note 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Kangaroo 924 Enteral Feeding Pump with Pole Clamp 21 Table of Contents Guidance and Manufacturer s Declaration Electromagnetic Immunity The Kangaroo 924 Enteral Feeding Pump with Pole Clamp is intended for use in the electromagnetic environment specified below The customer or the user of the Kangaroo 924 Enteral Feeding Pump with Pole Clamp should assure that it is used in such an environment Immunity Test IEC60601 test level Compliancelevel Electromagnetic environment guidance Conducted RF 3Vrms Portable and mobile RF communications equipment IEC 61000 4 6 150 kHz to 80 MHz should be used no closer to any part of the Kangaroo 924 Enteral Feeding Pump with Pole Clamp including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Radi
12. 3 4 kPa Nominal Alarms e Flow Error Low Battery Hold Error LoadSet System Error VTBD Delivered Memory of Settings When Pump is Off 16 hours for battery 24 hours for AC Battery Rechargeable NiMH Automatically charges when pump is plugged into wall outlet Recharge time for fully discharged battery is approximately 12 hours A new battery when fully charged will operate for approximately 16 hours at 125 milliliters per hour Dimensions Approximately 7 25 High x 6 61 Wide x 4 33 Deep 18 4 cm High x 16 8 cm Wide x 11 cm Deep Kangaroo 924 Enteral Feeding Pump with Pole Clamp 15 Table of Contents Weight Approximately 3 89 Ibs 1 76kg Case Material ABS PC Plastic Fire Resistant Power Requirements 120V 60 Hz 1 Amp North America 230V 50Hz 0 5 Amp UK Australia If the integrity of the protective earth conductor arrangement is in doubt operate unit from its internal electrical power source Ambient Operating Conditions 10 to 40 C 50 to 105 F Transport and Storage not to exceed 30 days 0 to 50 C 32 to 122 F 10 95 RH non condensing In the event that the environmental conditions for transport and storage are exceeded return the unit for factory service In the event that the specified storage time limit is exceeded within the environmental limits specified perform the performance tests prior to use Refer to the Kangaroo 924 Enteral Feeding Pump with Pole Clamp manual for s
13. E Operating Manual Kangaroo 924 Enteral Feeding Pump with Pole Clamp Table of Contents Section I General Information sss 1 Section 1 Safety and 3 Section 11 Operating 4 Batte 5 D c Feeding a RU Uva UR n 6 IDEE C mc 8 Section V 9 Important User 9 General Cleaning 10 Directions for Cleaning Pump 10 Directions for Cleaning Pump Power Cord 10 Directions for Cleaning Drip 10 Directions for Cleaning Optix 10 Directions for Cleaning Rotor 11 Preventive Maintenance 11 Section VI Troubleshooting 11 Biotech MDUB REA Rcs d 13 Section VII Factory Service eise re nre Pocos 14 Section VIII Spedficationis isdosco dup D ip RR b nb ERR 15 Pp m
14. ance including replacement of battery pack and line fuses is to be performed only by the user s appropriately qualified technical person C To order a replacement for the power cord or tubing latch use the following part number U S Power Cord Not User Serviceable U K Power Cord F090444 AU Power Cord F090475 Tubing Latch F31975 Kangaroo 924 Enteral Feeding Pump with Pole Clamp 14 Table of Contents Section VIII Specifications ES60601 1 2005 UL 60601 1 CAN CSA C22 2 No 60601 1 08 Type Infusion Device Volumetric Pumping Mechanism Rotary Peristaltic Feeding Set Required Covidien Kangaroo 924 Enteral Feeding Pump with Pole Clamp Feeding Sets see www covidien com for details on compatible administration sets Operating Range Rate 1 300 mL hr in 1 mL hr increments Dose 1 2000 mL in 1 mL increments Volume Delivered 1 9999 mL rollover is accounted for Accuracy 1096 or 0 5 mL hr whichever is larger with top of the fluid column at 71 12 cm 28 0 76 cm 0 3 above the top of the pump at a room temperature of 22 C 2 C 72 F 3 F using a new 924 feeding set for no longer than the recommended 24 hours of maximum usage Confidence limits for accuracy are based upon those included in the ANSI AAMI 1026 1992 American National Standard for Infusion Devices Reported accuracy is based upon the time based accuracy calculation in this standard as applicable for enteral feeding Occlusion Pressure 15 psi 10
15. ated RF 3V m IEC 61000 4 3 80 MHz to 2 5 GHz Recommended separation distance 12 4 1 2 80 800 MHz d 2 3VP 800 2 5 GHz where is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol a Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflec tion from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Kangaroo 924 Enteral Feeding Pump with Pole Clamp is used exceeds the applicable RF compliance level above the Kangaroo 924 Enteral Feeding Pump with Pole Clamp should be observed to verify normal operation If abnormal p
16. beyond the pump s capability to pump Alarm will activate at approximately 15 psi 103 4 kPa To correct this condition determine where in the system the occlusion has occurred Correct occlusion condition Press START HOLD Continue with feeding C The drip sensors are blocked and will not be able to detect drops as they come through the drip chamber To correct this condition check drip sensors on inside of upper drip chamber guide and clean off any material that may be blocking sensors D Tubing is pinched or kinked Low Battery Lo bAt This alarm is activated when the battery begins to run down and flow rate accuracy can no longer be maintained To correct this condition plug the pump into an AC wall outlet When the pump is connected to a wall outlet it will continue to operate and charge the battery at the same time Once the low battery alarm has sounded it will take approximately 12 hours to fully recharge Hold Error HLd Err If the pump is left in the hold mode for approximately 5 minutes Hold Error will alarm System Error SYS Err A Rotor is not turning while pump is in start mode Note Dots will flash across top of LED display while pump is running B Rotor is turning but is not detected by rotor sensors Do not try to repair pump Return pump for technical service Note For all alarm conditions pressing the START HOLD button will silence the audible alarm The visible display will continue to flash
17. d permanently Kangaroo 924 Enteral Feeding Pump with Pole Clamp 13 Table of Contents Section VII Factory Service Inthe event that it is necessary to return a unit for repair please observe the following 1 Call Customer Service for an Authorized Return Number and shipping instructions using the appropriate phone number below 2 Packthe instrument carefully and ship the insured parcel to the following locations or your local service center United States Canada U K Service Center Covidien Covidien Outside of U S and Canada Attention Service 7300 Trans Canada Highway Covidien 5920 Longbow Drive Pointe Claire QC H9R 1C7 Unit 2 Talisman Business Center Boulder CO 80301 1 877 664 8926 London Road 1 800 448 0190 Bicester OX266HR UK 44 1869 328065 B place an order for repair or if technical assistance is required call customer service or where designated by your local customer service center Technical information will be made available on request to allow the user s appropriately qualified technical personnel to repair the parts of the equipment designated by the manufacturer as repairable Refer to the Kangaroo 924 Enteral Feeding Pump with Pole Clamp manual for more service information The Kangaroo 924 Enteral Feeding Pump with Pole Clamp contains no user serviceable parts User maintenance consists only of cleaning the exterior of the pump as stated in the cleaning section of this manual All other mainten
18. egards to disposal of infectious biological or medical waste Example of proper placement of Feeding Set in Kangaroo 924 Enteral Feeding Pump with Pole Clamp See Figure 6 Note Assure feeding bag is set so that the top of the fluid column is 28 71 1 cm above the top of the pump Figure 6 Kangaroo 924 Enteral Feeding Pump with Pole Clamp 7 Table of Contents SectionIV Alarms The Covidien Kangaroo 924 Enteral Feeding Pump with Pole Clamp has been designed to provide audible and visible alarms under several conditions where proper performance cannot be maintained When alarm occurs press the START HOLD button on the pump and perform the necessary procedures to correct the alarm condition Note The visible alarm is in the form of a message shown on the display Load Set LoAd SEt A The feeding set is not installed or is installed improperly See Operating Procedure for installation instructions B Excessive Ambient Light dots displayed with LoAd SEt message indicates an excessive ambient light condition To correct this situation move pump away from bright light condition Flow Error FLo Err A feeding container is empty To correct this situation if more feeding is required refill the container with the desired amount of formula and follow the steps in the Operating Procedures of this manual B The feeding container the feeding set tubing or the feeding tube itself has become occluded
19. ents Low Battery Alarm Lo bAt Battery has been run down below point of maintaining accuracy of pump Plug pump power cord into wall outlet If required replace the power cord only with Covidien approved part as listed below U S Power Cord Part Number Not User Serviceable U Power Cord Part Number F090444 AU Power Cord Part Number F090475 Hold Erro Alarm HLd Err Pump has been left in the hold mode for over 5 minutes Press START HOLD to silence alarm then press START HOLD again to clear the visual alarm message Press START HOLD again to restart Load Set LoAd SEt Feeding set has been improperly placed onto pump Check to make sure only Kangaroo 924 Feeding Set has been placed onto the pump Check to make sure Optix connector on feeding set has been properly positioned into the Optix lock on pump High Ambient Light Condition If dots appear with the LoAd SEt message on the display this indicates a high ambient light condition Remove from bright light area until error condition clears Feeding solution or other foreign substances have accumulated in the Optix lock and blocked or obscurred the optical set detection mechanism Follow the instructions in Section V Cleaning under Directions for Cleaning Optix Lock to remove the material blocking proper detection of the feeding set System Error Alarm SYS Err Rotor is not turning while pump is running Return pump for technical ser
20. erformance is observed additional measures may be necessary such as re orienting or relocating the Kangaroo 924 Enteral Feeding Pump with Pole Clamp Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Kangaroo 924 Enteral Feeding Pump with Pole Clamp 22 Table of Contents Rx li Consult 01 23 instructions for use Eg covipiEN Manual No SMF0211 100 COVIDIEN COVIDIEN with logo and Covidien logo are U S and internationally registered trademarks of Covidien AG Spray Nine is a trademark of Spray Nine Corp bHisoHex is a trademark of The Mentholatum Co Hibiclens is a trademark of Regent Medical Ltd Vesta Syde is a trademark of Steris Inc Other brands are trademarks of a Covidien company O 2012 Covidien wl Covidien llc 15 Hampshire Street Mansfield 02048 USA Covidien Ireland Limited IDA Business amp Technology Park Tullamore www covidien com 1 800 962 9888 T REV 04 2012
21. ervicing information Type of Protection Against Electric Shock Class North America UK and Australia Degree of Protection Against Electric Shock Type BF Mode of Operation Continuous Degree of Protection Against Ingress of Fluids Drip Proof IPX1 Per IEC 529 Degree of Protection Against Ingress of Flammable Anaesthetic Mixtures Not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide Kangaroo 924 Enteral Feeding Pump with Pole Clamp 16 Table of Contents Symbols European Union s Waste Electrical Equipment WEEE Directive in accordance with European Standard EN 50419 Alternating Current Store between these temperatures Set Loading Diagram Caution For indoor use only Consult Instructions for Use Type BF Protection Degree of protection against electrical shock there is no conductive connection to the patient Do not use Kangaroo 924 Feeding Sets for greater than 24 hours Do not use if package is opened or damaged Manufacturing Date Code Feeding Formula Serial Number SQR Feed See Kangaroo 924 Enteral Feeding Pump pia with Pole Clamp manual for Feeding 1 Set loading instructions Maximum operating pressure 103 kPA p A 5 White Cap Reorder number for the device located on the carton label A B RB We x PX1 Drip Proof Degree of protection against ingress of fluids Protective Earth 12 Red Stepped Connector CE Mark European
22. fied for this enteral feeding pump For optimal accuracy level of formula should be 28 inches 71 1 cm above the pump do not reuse feeding sets and avoid overstretching the silicone tube that wraps around the pump rotor 13 The Kangaroo 924 Enteral Feeding Pump has been programmed to optimize accuracy by taking into account the viscosity of formula Formula that has been diluted and has a low viscosity will tend to deliver like water and thus deliver to the high side of the accuracy specification 14 Caution Ensure buzzer hole is unobstructed during normal operation so as to allow clear recognition of alarm 15 Caution The battery cells used in this device may present a fire or chemical hazard if mistreated Do not disassemble heat above 100 C 212 F or incinerate 16 Caution Avoid getting black Optix connector wet Wetting of the connector may cause a false Load Set alarm 17 Warning Always disconnect the feeding set from the patient before priming the feeding set 18 Warning For safe operation use only Covidien brand replacements battery power cord or tubing latch Kangaroo 924 Enteral Feeding Pump with Pole Clamp 3 Table of Contents Section Ill Operating Procedures Note To disconnect AC power it is necessary to unplug power cord from the AC wall outlet Replace the power cord only with Covidien approved part as listed below U S Power Cord Part Number Not User Serviceable U K Power Cord Part Number F090444
23. h this solution can damage the plastic pump case Isopropyl alcohol applied with a damp not wet cotton swab may be used for cleaning difficult to reach areas such as the drip sensor however it should be used sparingly because repeat cleaning may damage the case e not use strong cleaners such as Spray Nine pHisoHex Hibiclens or Vesta Syde because damage to the pump case housing can result Directions For Cleaning Pump Housing e Refer to General Cleaning Directions before starting e Clean outside surface with a damp not wet cloth or sponge and keep pump in upright position as much as possible e Avoid excess moisture near handle hinges e Avoid excess moisture near pole clamp area Directions For Cleaning Pump Power Cord e Referto General Cleaning Directions before starting e Unless soiling is observed the power cord should not be cleaned Ifcleaning ofthe power cord is necessary wipe the exterior surfaces of the wall plug of the power cord with a cloth dampened with isopropyl alcohol Avoid exposing power cord to excess moisture as this can lead to an electrical shock or fire hazard Directions For Cleaning Drip Detector e Referto General Cleaning Directions before starting e Clean the drip sensors in the upper drip chamber guide with a cotton swab dampened with isopropyl alcohol Besure sensor areas are clear Directions For Cleaning Optix Lock e Referto General Cleaning
24. r The Kangaroo 924 Enteral Feeding Pump with Pole Clamp contains no user serviceable parts User maintenance consists of cleaning the exterior of the pump as stated in the cleaning section of this manual All other maintenance is to be performed by the user s appropriately qualified technical personnel Section VI Troubleshooting Guide Listed below are some of the probable causes of alarm conditions and their corrections Flow Error Alarm FLo Err Feeding container is empty Refill feeding container If bag has been in use for 24 hours replace Feeding tube or feeding set tubing is occluded Locate point of occlusion and correct Note occlusion may occur upstream or downstream from pump e Feeding Container e Feeding Set control clamp closed e Feeding Tube patency of feeding tube should be checked Drip chamber is improperly placed in pump Check to make sure drip chamber is properly locked in position in drip chamber guide Drip chamber walls are coated with feeding formula make sure formula is not preventing detectors from proper operation If formula cannot be removed from inside walls of chamber via chamber manipulation replace set Sensors in upper drip chamber guide are blocked Check to make sure detectors are free of dried formula Remove any deposits by using a cotton swab dampened with warm soapy water Kangaroo 924 Enteral Feeding Pump with Pole Clamp 11 Table of Cont
25. rvals due to unforeseen soiling Only personnel trained in the cleaning of medical devices shall perform cleanings Important User Information AN Caution Do not immerse pump or power cord in water or other cleaning solution clean using a damp not wet cloth or sponge Failure to follow the cleaning procedures described herein could result in hazards to users As with any A C powered electrical device care must be taken to prevent liquid from entering the pump or the power cord receptacle to avoid electrical shock hazard fire hazard or damage to electrical components Should any of the following events occur do not use the pump until it has been properly cleaned and serviced by personnel trained in servicing Kangaroo pumps wetting of the pump s power cord or leakage into the pump interior during cleaning spillage of large amounts of formula onto the pump exterior or any spillage on the power cord Kangaroo 924 Enteral Feeding Pump with Pole Clamp 9 Table of Contents A A Caution Caution General Cleaning Directions Cleaning of Kangaroo 924 Enteral Feeding Pumps must be performed as follows Disconnect pump from A C power source before cleaning Upon completion of cleaning do not connect to power source until pump and power cord are thoroughly dry Amild detergent should be used for general cleaning If necessary the pump may be cleaned with a 10 1 water and hypochlorite mixture however repeated cleaning wit
26. ust be disposed of in accordance with local environmental and institutional policies Kangaroo 924 Enteral Feeding Pump with Pole Clamp 5 Table of Contents Beginning Feeding Below are general instructions for inserting all Covidien Kangaroo 924 Feeding Sets into the Kangaroo 924 Enteral Feeding Pump with Pole Clamp Close the control dlamp on the Kangaroo 924 feeding set completely Fill feeding set bag with desired amount of formula Close and hand bag Ensure the formula fluid column hands 28 71 1 cm above the top of the feeding pump at the start of feeding Figure 4A Figure 4B Slide drip chamber into upper drip chamber guide See Figure 4A Seat and secure chamber in lower drip chamber guide A Remove distal connector protective cover from the feeding set Slowly open control clamp and fill entire line with fluid leaving a few drops of fluid in the drip chamber Note Avoid filling the drip chamber more than half full 9 Close control clamp 10 Attach distal connector to feeding tube 11 Inspect lens in Optix lock area Clean any contaminants from lens before proceeding to the next step Figure 4B Figure 5A Figure 5B 12 Grasp the clear tubing below the Optix connector and guide the silicone tubing around rollers see Figure 5A Insert Optix connector into Optix lock B Thread tubing into tubing guide C Secure latch D over tubing see Figure 5B Kangaroo 924 Enteral Feeding Pump with Pole
27. vice Rotor is turning but is not detected by rotor sensors Return pump for technical service VTBD Delivered Alarm Vtbd dEL Preset VTBD has been delivered to patient No Audible Alarm Battery is fully discharged Charge battery for a minimum of 12 hours Alarm circuitry failure Return pump for technical service Kangaroo 924 Enteral Feeding Pump with Pole Clamp 12 Table of Contents The Kangaroo 924 Enteral Feeding Pump with Pole Clamp is designed to minimize the effects of uncontrolled electromagnetic interference and other types of interference When using the Kangaroo 924 Enteral Feeding Pump with Pole Clamp avoid the use of equipment that causes erratic operation or degradation in performance Biotech Mode To enter Biotech mode 1 Turn pump off 2 Pressand hold the ON button for 6 7 seconds 3 Pump alternately displays biotECH for 5 seconds Pump then displays SEt LAng for 5 seconds To select the language press A or W buttons or rotate adjustment knob to cycle through the list When the desired language is displayed press the ON button to set 4 Pump now displays bEEP LVL and volume of speaker Select desired volume 1 to 10 by rotating adjustment knob or pressing or W buttons 5 Pumpreturns to Hold mode Note Pressing the ON button will set the value displayed and continue to the next item in the sequence This will also happen automatically by waiting 5 seconds Note Changes are save
28. warranty of merchantability or fitness for a particular purpose or application other than as expressly set forth in the product labeling In no event shall Covidien be liable for any incidental indirect or consequential damages in conjunction with the purchase or use of the Pump even if advised of the possibility of the same Kangaroo 924 Enteral Feeding Pump with Pole Clamp 19 Table of Contents Section X Electromagnetic Conformity Declaration The Kangaroo 924 Enteral Feeding Pump with Pole Clamp has been built and tested according to UL60601 1 560601 1 2005 CAN CSA C22 2 No 60601 1 08 EN60601 1 2006 and EN60601 1 2 Standards The Kangaroo 924 Enteral Feeding Pump with Pole Clamp is intended for use in the electromagnetic environment specified below The user of the pump should assure that it is used in such an environment Electromagnetic Environment Guidance RF emissions Group 1 The Kangaroo 924 Enteral Feeding Pump with Pole Clamp uses RF energy only CISPR 11 for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions Class B The Kangaroo 924 Enteral Feeding Pump with Pole Clamp is suitable for use CISPR 11 in all establishments including domestic establishments and those directly Harmonic emissions Class A connected to the public low voltage power supply network that supplies IEC 61000 3 2 buildings used for domestic purposes
29. wer cord with built in A C adapter with pump See Section VII Factory Service for replacement of power cord 5 Use only Kangaroo 924 Enteral Feeding sets with this device Pump is not to be used with other sets Sets are a single patient use device 6 Used feeding sets should be disposed of in accordance with current hospital procedure or local disposal guidelines 7 For service or for technical information please contact Factory Service Section VII 8 Do not open the main housing as there are no user serviceable parts inside Opening may affect function of device and voids the warranty To replace battery see Section VII 9 Dispose of old battery powered electronic equipment in a manner consistent with institutional policy for expired equipment disposal 10 Cleaning frequency and practices must be consistent with institutional policy for cleaning of non sterile devices See Section V Cleaning for instructions on cleaning the Kangaroo 924 Enteral Feeding Pump 11 This device is designed to minimize the effects of uncontrolled electromagnetic interference and other types of interference from external sources Avoid use of other equipment that may cause erratic operation or degradation in the performance of this device 12 Caution This enteral feeding pump should only be used for patients who can tolerate the flow rates and accuracy levels delivered by the pump Premature infants may require higher accuracy rates than speci