ProXenon 350 Surgical Illuminator Service Manual
Contents
1. Figure 1 3 Replace the blown fuse s with the same size and rating 6 3A time delay T6 3A 250V AC fuses size 5 x 20mm 4 Re install the fuse cover Re connect the power cord 6 Press the Power button to apply power until a click is heard and power symbol illuminates Note If the ProXenon 350 Surgical Illuminator fails to operate properly again contact Welch Allyn for repair listed on page 5 PROXENON LIGHTSOURCE Welch Ally Figure 1 Part 1201281 15 Door Latch amp Interlock Switch Spring To Install the Interlock Switch Spring 1 Turn the ProXenon 350 Surgical Illuminator lamp off insuring continuous fan operation 2 Wait 10 minutes for the lamp to cool 3 Turn the ProXenon 350 Surgical Illuminator off and unplug the power cord 4 Unlatch and open the access door located on the side of the ProXenon 350 Surgical Illuminator Figure 2 5 Loosen M4 screw and install Interlock Switch Spring on bottom edge of latch cutout Slide spring against door support Figure 3 6 Reattach latch bracket add 1 drop of Loctite 242 to end of screw and tighten screw Torque screw to 5 1 in Ib Figure 2 Part H 1201281 16 IN Door Spring Edge Spring Door 2 Figure 3 Lamp Care and Replacement WARNING The lamp module and nearby structures become VERY HOT A even after brief operation To prevent burns turn off the ProXenon 350 Surgical Illuminator and allow it to cool for 10 minut
2. Caution Since the ProXenon 350 Surgical Illuminator uses a custom lamp module always have a spare lamp module available for replacement Use only Welch Allyn replacement lamp module REF 90209 Caution To prevent damage to the ProXenon 350 Surgical Illuminator prevent overheating and maintain the warranty replace lamp only with the Welch Allyn REF 90209 lamp module Read instructions before replacing lamp module Caution Grounding reliability is achieved only when connected to hospital use or hospital grade receptacles Inspect electrical plug and cord routinely Do not use if damaged Caution PROVIDE VENTILATION TO PREVENT OVERHEATING Keep cooling vents free from obstructions Do not cover or drape the ProXenon 350 Surgical Illuminator Provide a 6 inch 15 24 cm distance between the ProXenon 350 Surgical llluminator and any solid objects Use the ProXenon 350 Surgical Illuminator only when it is in the horizontal position 11 Caution Do not use the turret as a light attenuator or operate with the turret misaligned with the light port Caution The metal end of the fiberoptic cable gets hot during use Allow to cool before touching Caution IPXQ Equipment not protected against the ingress of water Do not use or store liquids above or on the surgical illuminator Caution The ProXenon 350 Surgical Illuminator produces very high levels of visible light and is optimized to power the Welch Allyn ProXenon Headlight and Fiber
3. When using endoscopic or surgical headlight fibers without canes lenses or fusing at the input end irreparable fiber damage will result when the ProXenon 350 Surgical llluminator intensity level is set above 70 and total accumulated lamp hours are 100 hours or less Caution When using fibers other than the 902 Series ProXenon Headlight fiber do not exceed 70 intensity setting when total accumulated lamp hours are less than 100 Caution Do not operate the ProXenon 350 Surgical Illuminator without a lamp module installed Service Cleaning Instructions AN Part 1201281 Caution Do not expose the ProXenon 350 Surgical Illuminator to autoclaving or any cleaning sterilization process involving excessive heat or moisture This could lead to damage and void the warranty Caution To reduce risk of electric shock do not remove cover Refer servicing to qualified personnel No cleaning of any interior components Is required Caution Use hospital approved disinfectants e g 10 Clorox 90 water solution Caution Never introduce any liquid directly to the surface of the ProXenon 350 Surgical Illuminator Wipe the surface of the ProXenon 350 Surgical Illuminator periodically with a damp soft cloth using isopropyl alcohol or a mild detergent solution to remove surface contamination Allow the instrument to dry before use 12 Basic Troubleshooting No light ouptut Part H 1201281 Power cord is not plugged
4. 0 3 3219 0074 Part 1201281 Canada Customers Welch Allyn Ltd Canada 160 Matheson Blvd E Unit 2 Mississauga Ontario Canada L4Z 1V4 Tel 1 905 890 0004 or 1 800 561 8797 Fax 1 905 890 0008 Welch Allyn Ltd Singapore 10 Hoa Chiang Road 19 03 04 Keppel Towers Singapore 089315 Tel 6419 8100 Fax 6291 5780 Welch Allvn China Unit 1101 Central Plaza No 227 HuangPi North Rd Shanghai China 200003 Tel 0 21 6327 9631 Fax 0 21 6327 9632 Welch Allyn Latin American Customer 11300 N W 41 Street Miami FL 33178 USA Tel 1 305 669 9003 Fax 1 305 669 1971 Warrantv Welch Allyn warrants the ProXenon 350 Surgical llluminator when new to be free of defects in material and workmanship and to perform in accordance with manufacturer s specifications when subject to normal use and service for a period of two years from date of purchase from Welch Allyn or an authorized agent Welch Allyn will either repair or replace any components found to be defective or at variance from manufacturer s specifications within this time at no cost to the customer lt shall be the purchaser s responsibility to return the instrument to the authorized distributor agent or service representative Welch Allyn warrants the ProXenon 350 lamp to be free of defects in materials and workmanship This warranty does not cover the ProXenon 350 lamp for breakage or failure due to tampering misuse neglect accidents improper installation m
5. 21 5 Ibs 6Cto 40 C 43 Fto 104 F 20 C to 85 C 4 F to 185 F 10 85 relative humidity non condensing 1013 hPa 170 hPa lt 45 dB per ISO 7779 bystander measurement standard per ISTA 1A CAN CSA C22 2 No 601 1 CAN CSA C22 2 No 601 1 2 UL 60601 1 IEC EN 60601 1 IEC EN 60601 1 2 Relies upon the connection to the protective earth conductor of the installation to prevent shock hazards Applied part is floating from earth No protection against the ingress of liquids Suitable for continuous operation Not suitable for use in the presence of a Flammable Anesthetic Mixture 41 Electromagnetic Compatibilitv WARNING AC cables other than those specified by manufacturer may result in increased EMISSIONS or declared IMMUNITY after brief operation To prevent burns turn off the ProXenon 350 Surgical Illuminator and allow it to cool for 10 minutes prior to removing lamp module Electromagnetic Emissions The Welch Allyn ProXenon 350 Surgical Illuminator is intended for use in the electromagnetic environment specified below The customer or user of the Welch Allyn ProXenon 350 Surgical Illuminator should assure that it is used in such an environment RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emissions IEC 61000 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Part H 1201281 Group 1 Class A Class A Complies The Welch Allyn ProXenon 350 Surgical Illumi
6. While holding the Full Dim Decrease buttons press and hold the Lamp ON OFF button for three seconds The the Lamp Hour window flashes three times and resets to zero Note To cancel reset release any button during the three second period Cover Removal Part 1201281 Remove the eight 6 32 screws four on the back and four on the bottom of the ProXenon 350 Surgical Illuminator Figure 6 Unlatch and open the access door located on the side 19 3 Slide the cover out WARNING Failure to open the lamp door prior to cover removal installation would result in damage to the microswitch PCC Lat g0oubtCR dh fe So lut i KEN sane CE AS d D H i 151 Un D a El Figure 6 Plunger Ball Removal 1 Remove plunger ball from the turret housing 2 When reinstalling apply Loctite 242 to the threads of the set screw Screw the set screw in all the way and back out half a turn no i md PLUNGERBALL 4 PS x 8 i eg de PAD n A gt PISTA Figure 7 Part H 1201281 20 Turret Cover Removal Remove the 10 32 screw securing the turret cover to the front panel Figs 8 amp 9 m len Praxkonon 2350 LA om mum Turret Knob 5 Front Panel 5 AC Switch Cap 17 Membrane Figure 8 Part 1201281 21 Figure 9 Part H 1201281 22 Turret Housing Figure 10 Turret Knob Removal 1 2 Remove the four 6 32 Phillips fl
7. can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer Part 1201281 45 REPLACEMENT PARTS Bubble f Part Description Number 2 0 0 713407 SPRING DOOR _ 3 5 4 711760 JTURRETCOVER 8 5 713433 WRETKNOB 8 6 713434 TURRETHOUSING___________ 10 8 713432 CONTROLLOGICBOARDASSEMBLY 16 9 713430 HOT MIRROR ASSEMBLY 7 15 713424 2 FRONTPANEL TI 8 16 1208 MEMBRANE A7 714488 JACSWITCHCAPUI 8 181703491 AC SWITCHOVERLAY 8 19 714481 PAN HEAD SLOTTED 6 32 X 3 8 LG 18 NYLON 1 714484 NYLON STANDOFF 6 32 X 3 8 LG X 1 4 HEX 1 713429 TOP COVER 714772 WIRE HARNESS ASSY TURRET INTERLOCK HARNESS POWER SUPPLY INTERLOCK HARNESS SHUTTER HARNESS AC HARNESS 12 PINS INTERFACE HARNESS A a 713408 PLASTIC RIVET 1 713409 BLACK FOOT PLASTIC 1 Part H 1201281 46 Final Inspection Verifv the following 1 2 3 4 5 D Turret rotation check AC Switch LED lights when the button is pressed Both fans operate properly Lamp ON OFF functionality amp LED indicator with and without fiber installed Light Output New Lamp 1800 Lumens minimum initial input through 3mm glass rod at full rated power No light leakage when the stutter is at close position 0 Display goes from 0 100 and 100 0 with 10 increment
8. 0 dip for 25 cycle gt 95 dip in 5 seconds 8 kV air 2 kV for power supply lines Not applicable 1kV differential mode 2 kV common mode gt 95 dip in 0 5 cycle 60 dip in 5 cycles 30 dip for 25 cycle gt 95 dip in 5 seconds should be at least 30 Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of the Welch Allyn ProXenon 350 Surgical Illuminator requires continued operation during power mains interruptions it is recommended that the Welch Allyn ProXenon 350 Surgical Illuminator be powered from an uninterruptible power supply or battery Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment 43 Electromagnetic Immunitv The Welch Allyn ProXenon 350 Surgical Illuminator is intended for use in the electromagnetic environment specified below The customer or user ofthe Welch Allyn Eu SE ISO GUI UG Compliance Electromagnetic Environment Guidance Test Level Level Portable and mobile RF communications equipment should be used no closer to any part of the Welch Allyn ProXenon 350 Surgical Illuminator including cables than the recommended separation distance calculated from the
9. 1281 33 Figure 23 Part H 1201281 34 Figure 24 Part H 1201281 35 Figure 25 36 Part H 1201281 i ny fl Lage u e S WAN i i cr ha p vd d m SC r rn h ee Zar E Cu Power Supply A Figure 26 Power Supply Removal gt Disconnect the 12 pin connector Con2 using a flat head screwdriver Fig 26 Disconnect live neutral amp ground connections Fig 27 Remove the four M3 mounting screws from the bottom of the base Part 1201281 37 Figure 27 Line Filter Removal 32 nut Fig 28 Remove the two 6 32 screws and one 6 Disconnect live neutral amp ground connections Fig 28 Remove the line filter shield Fig 28 Fig 29 Remove the 6 32 screw and two 4 40 nuts 38 Part 1201281 ski AAA LANA A tiddi I T 7 S Tresen Line Filter Shield in Di d killa l IN Wing i d k NW MAR l ji ey l N a gt die fe ai ie Jk Ii Li y ki hu LE l an A ye i AMO A Kar on g ANER N ad WM jil 4 U Saat i L N jihu ii i HAU i i eu NG d ei G ui A ALL ui M WA II EN Y wd MRI A a EN ja Il VA vy tk rita em V _ ur Figure 28 ali i eee ri i A A UA A A A 1 KE Li ie MANGIO Gamm ym Kl AA A A A A A A Y ee BEEBE Line Filter Assembly Di Figure 29 Part H 1201281 39 Electrical Input Input Voltage AC Power Connector Line Cord Fuse Perform
10. O 100 Maximum Light intensity setting 0 Minimum Light Intensity Setting Light Intensity Setting Increases light intensity in 10 increments Decreases light intensity in 10 increments Lamp Hours Agency Symbols OD UL 60601 1 IEC 60601 1 CAN CSA STD C22 2 NO 601 1 M90 N344 EMC Framework of Australia Part 1201281 CE The CE mark on this product indicates that it has been tested to and conforms with the provisions noted within the 93 42 EEC Medical Device Directive Authorized European Representative Address European Regulatory Manager Welch Allyn Ltd Navan Business Park Dublin Road Navan County Meath Republic of Ireland Telephone 4353 46 90 67700 Fax 353 46 90 67755 Safetv Warnings and Cautions Familiarize all operating personnel with the general safetv information in this summarv Specific warnings and cautions are also found throughout this manual Such specific warnings and cautions mav not appear in this summarv Warnings A warning statement in this manual identifies a condition or practice which if not corrected or discontinued immediately could lead to injury illness or death Part 1201281 WARNING Only qualified personnel should use the ProXenon 350 Surgical Illuminator WARNING Before operating the ProXenon 350 Surgical Illuminator read the Directions for Use The ProXenon 350 Surgical Illuminator is a source of high electrical voltage intense light an
11. ProXenon 350 Surgical Illuminator FC ee ee IL u pp ProXenon 350 Service Manual WelchAllyn Advancing Frontline Care Part H 1201281 Contents Contents INLFOGUELIO Dci nee acne dan 4 General Information NEE 4 Technical Help Information AA anno TAKE nennen anno nnn sure LYN nennen 4 Customer Service EEF at ee ed AER EFO NC NO Y FERN FYN 4 service EE d i re RF YRFA FE A AE GY Ee 5 ee O RNA NF Y en OE E E EI 6 le S A IA 6 Safety Warnings and CAUTIONS semanas idad 9 o o b A A 12 cleaning Ic UL tele p L A ehe tds 12 Basic TOUDIESNOOUNO sn rindas 13 TOOS REU riot ops 14 Component REMOVA leerte 14 Fase FeO lac CF WEN een A ee 15 Lamp Care and Replacement 22ss20400080008000n0n0nonnn nun nun nun nun eure Niana dder LED NN LE nennen 17 COVEN RENOVA en A a a rear 19 Plunger Ball NNO Vall are ee ers ja i A 20 Turret Cover Removal a ee dara eda 21 Turrel neie ROMOV ee EYN ropas 23 T rret Disk Removal i e a postas 23 Reinig STEEN te NEE 25 Display Board REMOV al au nennen 26 G ntrol Board an 9 E 26 Hot Mirror Assembly Removal ii 29 Fan ASSEMDIY RemnoV Adler aca 29 Power Supply Removal iii 37 LINEFINEOF Le Vlad a ern 38 SPECON EE 40 EICCHICAIINPU rra aid 40 Wein Le 40 Mechanical and ENVITON Mental tikek ae 41 Part H 1201281 2 Classiticalio A A A eg 41 Electromagnetic Compatible re 42 Electromagnetic ENUSSIONS in i A ra 42 Electro
12. RNING Use only non electrically conductive fiberoptic cables Use of conductive fiber optic cables will compromise the safety and effectiveness of this instrument WARNING High intensity light can cause burns to tissue even if the tip of the fiber optic cable or attached terminal device is cool Burns are caused by the absorption of light by tissue and the subsequent conversion of that light energy into thermal energy Full thickness burns can be caused by prolonged exposure to concentrated light energy Anesthesized patients and poorly perfused tissues are particularly susceptible to burn injuries via this mechanism There is serious risk of igniting fires if energized fiberoptic cables are placed on flammable materials IT IS THE USER S RESPONSIBILITY TO DETERMINE AND MAINTAIN THE MINIMUM SAFE DISTANCE BETWEEN THE END OF THE ENERGIZED FIBEROPTIC CABLE AND ANY LIVING TISSUE OR FLAMMABLE MATERIAL WARNING Intense light emitted from the ProXenon 350 Surgical Illuminator can cause permanent eye damage if viewed directly with unprotected eye The risk of injury is reduced by using the minimum level of illumination necessary by minimizing exposure time and by avolding close stationary viewing WARNING There is a risk of injury to tissue from exposure to the intense illumination Using light intensity settings in excess of that required for good visualization contributes to tissue warming and should be avolded WARNING Always have back up light so
13. ance Part H 1201281 Light Output Over temperature Protection Over Heat Recoverv AutoCool Fiberoptic Connection Safety Feature Lamp Power Supply Lamp Module Lamp Module Replacement Lamp Life Specifications 100 240 VAC 50 60 Hz universal 6 0A input Located on rear panel dual fuses IEC60320 250 VAC T6 3A 5x20 mm 1800 Lumens minimum initial input through 3 mm glass rod at full rated power Spectral output 380 750 nm nominal Automatic shutdown to prevent overheating Fans will remain on if thermal shutdown occurs when Auto Cool power is Auto Cool ON Use Lamp ON OFF button to re ignite lamp Output shutter will not open unless a fiberoptic cable is fully inserted into the active port on the turret Output shutter will fully close if fiberoptic cable is removed from active port Intensity setting flashes during this status PS300 12 type Welch Allyn 300 Watt Xenon REF 90209 Access via latched hinged door 650 hours to 50 of initial output specification measured through 3mm glass rod 40 Mechanical and Environmental Dimensions Weight Operating Temperature Storage Temperature Operating and Storage Humiditv Operating Pressure Audible Noise Shipping Shock and Vibration Safetv EMC and Regulatorv Compliance Classification Class 1 Tvpe BF IPXO Mode of Operation Flammable Height 12 98 cm 5 1 x Width 43 cm 16 9 x Depth 38 15 cm 15 0 approx 9 75 kg
14. at head screws securing the turret knob to the turret disk Figs 8 amp 10 Rotate the turret knob to gain access to the 4 screws through the inch diameter hole in the turret housing Turret Disk Removal 1 2 3 Part 1201281 Remove the 10 32 screw securing the turret disk to the turret housing Figure 10 Pull the turret disk away from the unit Figure 12 When reinstalling torque 10 32 screw to 4 in Ib rotate turret disk and ensure disk play or movement at each of the four port locations is minimal Install turret knob and test rotation If more adjustment is needed increase torque and test rotation Torque range may be set between 4 7 in Ibs 23 Figure 11 FUN Bright Full Dim aol Part 1201281 Figure 12 24 Front Panel Removal 1 Remove the six 6 32 screws securing the front panel to the base Fig 13 2 Disconnect the 9 pin J2 and 2 pin J4 connectors from the Control Board 3 Disconnect the 20 pin J12 connector from the Displav Board Fig 14 hi a SE F Tt Cw 6 t T Figure 13 Part 1201281 25 Connector 20 pin A z A ATA a eg e Di wt jq P 4 f pi M W e Figure 14 Displav Board Removal Remove the five 4 40 screws securing the display board to the front panel Fig 15 Figure 15 Control Board Removal 1 Remove the four 4 40 screws securing the control board to the base Fi
15. d heat When used properly and with normal precautions the ProXenon 350 Surgical Illuminator is a safe and effective light source WARNING RISK OF FIRE Not suitable for use in the presence of a Flammable Anesthetic Mixture WARNING RISK OF SHOCK The cover of the ProXenon 350 Surgical Illuminator should never be removed Electrical shock hazard due to high internal voltage There are no user serviceable parts inside this surgical illuminator except for the lamp and fuse which are accessible without removing the cover Refer all service to Welch Allyn WARNING RISK OF SHOCK Disconnect power cord before servicing Part H 1201281 WARNING This product comes with a power cord that is intended for use only with this product The power cord has not been tested and approved for use with other products that may have the same power connectors If you cannot locate the original power cord please contact Welch Allyn to obtain replacement parts WARNING Before connecting the ProXenon 350 Surgical Illuminator to AC power verify that the voltage to be applied is within the range specified on the identification label WARNING Extremely high energy light It is the user s responsibility to check the specifications of any device attached to the ProXenon 350 Surgical Illuminator to verify that it can transmit the intense light output without generating high temperatures and heat build up in the device that can cause serious burns WA
16. equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms 150 kHz to IEC 61000 4 6 80 MHz RadiatedRF 3V m d 1 17 VP 80 MHz to 800 MHz IEC 61000 4 3 80 MHz to 2 5 GHz d 2 33 VP 800 MHz to 2 5 GHz where P is the d 1 17 WP maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol A ProXenon 350 Surgical Illuminator should assure that it is used in such an environment Note 1 At 80 MHz and 800 MHz the higher frequency applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Welch Allyn ProXe
17. er to the model number shown on the data label found on the back of the system Technical assistance is contained in Basic Troubleshooting on page 13 of this manual Part 1201281 Service Center If you have an equipment problem that you cannot resolve you may call the Welch Allyn Service Center nearest you for assistance Technical service support is available by telephone on normal business days at the phone numbers listed below If you are advised to return a product to Welch Allyn for repair or routine maintenance schedule the repair with the service center nearest you Before returning a product for repair you must obtain authorization from Welch Allyn An RMA Return Merchandise Authorization number will be given to you by our service personnel Be sure to note this number on the outside of your shipping box Returns without an RMA number will not be accepted for delivery USA Customers Welch Allyn Inc U S A 4619 Jordan Road Skaneateles Falls NY 13153 0187 Telephone 1 866 801 8428 Fax 1 315 685 2854 International Customers Welch Allyn GmbH amp Co KG Zollerstrasse 2 4 72417 Jungingen Germany Telephone 0 7477 927170 Fax 0 7477 927190 Welch Allyn Ltd Australia Unit 5 38 46 South Street Rydalmere NSW 2116 Australia Telephone 0 2 9638 3000 Fax 0 2 9638 3500 Welch Allyn Japan K K 6F Kinseisha Bldg 3 15 Kandanishikicho Chivoda ku Tokvo 101 0054 Japan Telephone 0 3 3219 071 Fax
18. es prior to removing lamp module Caution Do not operate the ProXenon 350 Surgical Illuminator without a lamp module installed Replace the lamp under any of the following conditions accumulated usage of more than 650 hours an increasing number of ignition attempts required to ignite lamp failure of lamp to start or low light output Part 1201281 17 To remove the lamp module 1 Part H 1201281 Turn the ProXenon 350 Surgical Illuminator lamp off insuring continuous fan operation Wait 10 minutes for the lamp to cool Turn the ProXenon 350 Surgical Illuminator off and unplug the power cord Unlatch and open the access door located on the side of the ProXenon 350 Surgical Illuminator Figure 4 Grasp the handle and pull the entire lamp module out of the ProXenon 350 Surgical Illuminator Figure 5 Figure 4 18 Figure b To insert a new lamp module AN Caution Do not touch the face of the lamp glass window Orient the new lamp module so that the window of the Welch Allyn lamp REF 90209 is facing the front of the ProXenon 350 Surgical Illuminator Align the sides of the lamp module and place it in the receiver channel Firmly push in the lamp module to ensure proper connection Close and latch the door Ifthe door does not close completely the lamp module is not fully seated To Reset the Lamp Hours Meter Verify that the lamp is off Press and hold the Full Dim Decrease buttons
19. g 15 2 Disconnect the 3 pin J10 connector Part H 1201281 26 3 Disconnect the 4 pin J7 connector 4 Disconnect the 12 pin J8 connector 5 Disconnect the 6 pin J9 connector Control Logic Board Assembiv Connector 4 pin Connector 6 pin Connector 12 pin Figure 16 Part 1201281 27 Hot Mirror Assembly Removal Remove the two 6 32 screws securing the hot mirror assembiv to the base Fig 17 Figure 17 Part 1201281 28 Figure 18 Fan Assembly Removal D PLA En a Part H 1201281 Remove the four 6 32 nvlon screws on the interface board Fig 18 Remove the interface board by pulling it up Fig 19 Disconnect the 2 pin fan connector Fig 19 Remove the two lower 6 32 nylon standoffs used for the interface board Fig 20 Remove the four 6 32 screws on the interlock board assembly Figs 21 amp 22 Remove the two 6 32 screws securing the nomex duct to the power supply Fig 23 Remove the nomex duct Disconnect the 2 pin J1 connector from the interlock board Fig 23 Remove the interlock board assembly Remove the four plastic mounting rivets for each fan cut cable tie and remove fan assembiv Note orientation of fans when reinstalling Figs 24 amp 25 29 Figure 19 30 Part H 1201281 Part H 1201281 Figure 20 31 yry LA pii ty Pi bu Figure 21 Part H 1201281 32 Figure 22 Part H 120
20. into receptacle at the instrument or into the hospital grade outlet ON OFF button is off Fuse is blown Lamp door is not closed tightly Lamp module is installed incorrectiv Internal power supplv not operating Lamp is at the end of service Lamp module is installed incorrectiv Lamp door is not closed tightly Fiberoptic cable not connected Fiberoptic cable is not fully seated Fiberoptic cable Is worn or damaged Turret misaligned Plug in power cord Press the Main ON OFF button Replace fuse See Fuse Replacement page 14 Close lamp door completely Re install lamp module correctly See Lamp Care and Replacement page 15 Return the ProXenon 350 Surgical Illuminator to Welch Allyn Contact your local Welch Allyn Technical Service Center listed on page 5 Replace lamp module Re install lamp module correctly See Lamp Care and Replacement page 15 Close lamp door completely Shutter is not attenuated Connect fiberoptic cable Push the fiberoptic cable in completely Replace fiberoptic cable Align Turret correctly 13 Xenon lamp flickers or dims Field of view is dim Turns off after a few minutes of operation Lamp is nearing end of service Controls set incorrectiv Lamp module is installed incorrectiv Fiberoptic cable is worn or damaged Turret is misaligned Obstructed air intake overheating causes thermal switch to trip Fan not r
21. magnetic IMMUNI V ia ia 43 Electromagnetic mn mun EE 44 Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Welch Allyn ProXenon 350 Surgical Illuminator ss seenenenzennnzznnnnzenzz 45 Final InSspectio hke 47 Part 1201281 3 Introduction General Information The Welch Allyn ProXenon 350 Surgical llluminator is designed for use with the Welch Allyn ProXenon Headlight and Fiber system It will accept fiberoptic light guides for Wolf Storz Olympus and ACMI instrumentation The ProXenon 350 Surgical Illuminator uses a Welch Allyn high intensity narrow beam rugged xenon short arc lamp with a fixed internal reflector to produce a uniform profile beam Technical Help Information Only fully trained and properly equipped personnel should perform all service and repairs using genuine replacement parts and correct procedures Failure to do so will invalidate the product warranty and could compromise instrument safety and performance Read and understand all safety warnings and service notes printed in this Service Manual and the Operating Manual If there is any doubt about any precaution or procedure for phone help or to order additional copies of ProXenon 350 Surgical Illuminator Manuals contact Customer Service Welch Allyn Inc 4619 Jordan Road PO Box 187 Skaneateles Falls NY 13153 0187 USA Telephone 866 801 8428 or 315 685 2993 Fax 315 685 2854 When calling ref
22. nator uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment The Welch Allyn ProXenon 350 Surgical Illuminator is suitable for use in all establishments other than domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes See above See above 42 Electromagnetic Immunitv The Welch Allyn ProXenon 350 Surgical Illuminator is intended for use in the electromagnetic environment specified below The customer or user of the Welch Allyn ProXenon 350 Surgical Illuminator should assure that it is used in such an environment Dar IEC 60601 Compliance 1 i Test Level Electromagnetic Environment Guidance Electrostatic 6 kV contact 6 kV contact Floors should be wood concrete or ceramic tile If floors discharge ESD are covered with synthetic material the relative humidity IEC 61000 4 2 Electrical fast transient burst IEC 61000 4 4 Surge IEC 61000 4 5 Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 Part 1201281 8 kV air 2 kV for power supply lines 1 kV for input output lines 1kV differential mode 2 kV common mode gt 95 dip in 0 5 cycle 60 dip in 5 cycles 3
23. non 350 Surgical Illuminator is used exceeds the applicable RF compliance level above the Welch Allyn ProXenon 350 Surgical Illuminator should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the Welch Allyn ProXenon 350 Surgical Part 1201281 44 Illuminator Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Welch Allyn ProXenon 350 Surgical llluminator The Welch Allyn ProXenon 350 Surgical llluminator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The Separation Distance According to Freguency of Transmitter m Y a Rated Max Output 150kHzto80 MHzd 80 MHz to 800 MHz d 800 MHz to 2 5 GHz d Power a 1 17 P 1 17 P 2 33 P 0 117 0 117 0 233 customer or user of the Welch Allyn ProXenon 350 Surgical Illuminator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Welch Allyn ProXenon 350 Surgical llluminator as recommended below according to the maximum output power of the communications equipment For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m
24. odification shipping or from improper maintenance service or cleaning procedures This warranty is void if the instrument is not used in accordance with manufacturer s recommendations or if required service is performed by other than Welch Allyn or an authorized agent Purchase date determines warranty requirements No other express or implied warranty is given Symbols The following symbols are associated with the ProXenon 350 Surgical Illuminator Safety Symbols A Identifies information within the manual to avoid injury N Caution consult accompanving documents Alternating current Type BF Equipment _ e Recvcling Svmbol Do not dispose of this product as unsorted municipal waste Prepare this product for reuse or separate collection as specified bv Directive 2002 96 EC of the European _ Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment WEEE If this product is contaminated this directive does not apply See Part 1201281 www welchallvn com weee or contact Welch Allyn Customer Service Identifies information within the manual to avoid equipment failure Storage Humidity Fuse Dangerous Voltage Equipotentiality Attention Hot Surface IPX Equipment is not protected against the ingress of liquid Transport Temperature Consult accompanving documents o Lamp Part H 1201281 Button Svmbols o Power ON OFF Zen Lamp ON OFF
25. s and the light output is changing accordingly Lamp door interlock is functioning properly Ground continuity check input to equipment stud Hi pot test input chassis 1500 VAC Leakage test at 120 VAC 60 Hz less than 300 u A PLEASE NOTE Light output intensity is best evaluated by comparing light output with a known good light source headlight and fiber Part 1201281 47
26. unning overheating causes thermal switch to trip Tools Required o Screwdriver flat blade medium o Screwdriver Phillips Head e Ratcheting wrench and Phillips bit set e Torque wrench 4 7 in Ib Allen wrench set English e Nut driver set English e Nut driver set Metric e Isopropyl Alcohol o Loctite 242 Component Removal Note The removal sections beginning with the Lamp Care and Replacement are in Replace lamp module See Lamp Care and Replacement page 15 Adjust Full Bright or Full Dim controls appropriately Re install lamp module correctly See Lamp Care and Replacement page 15 Replace fiberoptic cable Align Turret correctly Allow instrument to cool 10 minutes Remove obstruction s Contact your local Welch Allyn Technical Service Center listed on page 5 order from start to finish Actions mentioned in previous sections are not repeated only additional steps are listed It is recommended that the lamp be removed prior to removing the cover and servicing internal components Note Component replacement is accomplished by reversing the process Part H 1201281 described by the removal 14 Fuse Replacement WARNING RISK OF SHOCK Disconnect power cord before servicing 1 Turn the ProXenon 350 Surgical Illuminator off and unplug the power cord 2 Remove the fuse cover located below the three prong power connector Carefully pull out the cover using a flat blade screwdriver
27. urce available in case of failure during a procedure WARNING The lamp module and nearby structures become VERY HOT even after brief operation To prevent burns turn off the ProXenon 350 Surgical Illuminator and allow it to cool for 10 minutes prior to removing lamp module 10 Cautions WARNING When the ProXenon 350 Surgical Illuminator is on never disconnect the fiberoptic cable from the terminal device High intensitv light from the end of the fiberoptic cable can ignite flammable materials for example drapes or cause burns to tissue WARNING Replace fuses as marked See Fuse Replacement Section WARNING Do not use for neonate transillumination Ervthema mav result WARNING Ensure that active port of the Four Port Turret is positioned correctiv and the fiberoptic cable is fullv seated WARNING The ProXenon 350 Light Source has been evaluated for use as a Headlight System light source only The safety and effectiveness of this light source for endoscopic use has not vet been evaluated A caution statement in this manual identifies a condition or practice which if not corrected or discontinued immediately could lead to equipment failure equipment damage or data loss AN Part 1201281 Caution Rx Only United States Federal Law restricts this device to sale by or on the order of a health care practitioner Caution Do not touch or change lamp module immediately after operation Allow lamp to cool 10 minutes