Home
        operation and service manual - Advanced Test Equipment Rentals
         Contents
1.                                     n         Ds   RECALL     LOCAL         am pn c STORAGE   ADDRESS  POWER REMOTE 1 2 8 CURRENT RETURN  i mm um um RUN MILLIOHN        SC ODEL   655454  13 24 26 25 232119 18 17 16 1415  24   20  22    19  MILLIOHM OFFSET  This key is used when performing a ground bond test  It  offsets the resistance of the test leads and fixtures to more accurately read the true  resistance during the ground bond test  The operator must first connect the test  connections together at the point they contact the DUT  Pressing the Milliohm Offset  button will calibrate the 6554SA to disregard connection resistance from all further  tests  Re setting the 6554SA to disregard only it   s test lead resistance        be done by  connecting the leads together and pressing the Milliohm Offset key again     20  RECALL   FUNCTION KEY  Depress this key when you wish to recall from  memory a previously stored test setup  You then simply select the memory number  you wish to recall and press enter     21  REVIEW  Use this key to review the results of a test after it has been performed or  to review the test set up parameters you have set and which memory number you are    1n     23               ASSOCIATED  RESEARCH  INC                                                                                                                                                                             ed  QUADCHEK    Electrical Safety Compliance Analyzer     5                        
2.                               5                        010    SECTION 1  OPERATORS MANUAL    e        SAFETY PRECATIONS REQUIRED FOR HIGH  VOLTAGE TESTING     GENERAL   This product and its related documentation must be reviewed for familiarization with  safety markings and instructions before operation     This product is a Safety Class I instrument  provided with a protective earth terminal    Before applying power verify that the instrument is set to the correct line voltage  110 or  220  and the correct fuse is installed     SAFETY SYMBOLS     INSTRUCTION MANUAL SYMBOL  PLEASE REFER TO THE  INSTRUCTION MANUAL FOR SPECIFIC WARNING OR  CAUTION INFORMATION TO AVOID PERSONAL INJURY OR  DAMAGE TO THE PRODUCT    INDICATES HAZARDOUS VOLTAGES MAY BE PRESENT     CHASSIS GROUND SYMBOL     WARNING    CALLS ATTENTION TO A PROCEDURE  PRACTICE  OR  CONDITION  THAT COULD POSSIBLY CAUSE BODILY INJURY  OR DEATH     d CALLS ATTENTION TO A PROCEDURE  PRACTICE  OR    CONDITION  THAT COULD POSSIBLY CAUSE DAMAGE TO  EQUIPMENT OR PERMANENT LOSS OF DATA          ASSOCIATED  RESEARCH  INC  SAFETY  WARNING  A Hipot produces voltages and currents which can cause harmful or fatal    electric shock  To prevent accidental injury or death  these safety procedures must be  strictly observed when handling and using the test instrument     SERVICE AND MAINTENANCE    User Service   To prevent electric shock do not remove the instrument cover  There are no user  serviceable parts inside  Refer servicing to 
3.                         7 620 meters  25 000 feet   The instrument should also be protected against temperature extremes which may cause  condensation within the instrument     Packaging    Original Packaging  Please retain all original packaging materials that you originally  received  If you are returning your instrument to us for servicing please repackage the  instrument in its original container  Contact our customer support department  1 800   858 8378  for a RGA  return goods authorization  number  Please enclose the instrument  with all options  accessories and test leads  Indicate the nature of the problem or type of  service needed  Also  please mark the container  FRAGILE  to insure proper handling   Upon receipt your instrument will be issued an AR service number  Please refer to this  number in all correspondence     28         ASSOCIATED  RESEARCH  INC  INSTALLATION    Other Packaging  If you do not have the original packaging materials please follow these  guidelines     1   Wrap the instrument in a bubble pack or similar foam  Enclose the same information  as above     2   Use a strong double wall container that is made for shipping instrumentation  350 Ib   test material is adequate     3   Use a layer of shock absorbing material 70 to 100 mm  3 to 4 inch  thick around all  sides of the instrument  Protect the control panel with cardboard     4   Seal the container securely   5   Mark the container  FRAGILE  to insure proper handling     6   Please refer in a
4.            ASSOCIATED  RESEARCH  INC     INSTALLATION    INSTALLATION  Introduction    This section contains information for the unpacking  inspection  preparation for use and  storage of your Associated Research  Inc   product     Unpacking and Inspection    Your instrument was shipped in a custom foam insulated container that complies with  ASTM D4169 92a Assurance Level II Distribution Cycle 13 Performance Test Sequence     If the shipping carton is damaged  inspect the contents for visible damage such as dents   scratches or broken meters  If the instrument is damaged  notify the carrier and the  Associated Research customer support department immediately  Please save the shipping  carton and packing material for the carriers inspection  Our customer support department  will assist you in the repair or replacement of your instrument  Please do not return your  product without first notifying us and receiving an RGA  return goods authorization   number     Preparation for Use  Power Requirements and Line Voltage Selection    This instrument requires a power source of either 115 volts AC   15   47 63 Hz single  phase or 230 volts AC  15   47 63 Hz single phase  Please check the rear panel to be  sure the proper switch setting is selected for your line voltage requirements before turning  your instrument on  In addition please be sure the correct fuse is selected and installed  while the instrument is in the off position  see page 26 for fuse changing instructions    CAUTION
5.        ASSOCIATED  RESEARCH  INC     02 Limited Output    OPTIONS    The Limited Output option limits the maximum output current levels of the Dielectric  Withstand Test  This option effects only the Dielectric Withstand Test mode  The high and  low current trip settings are also decreased accordingly  The revised Dielectric Withstand  Test mode specifications are as follows     DIELECTRIC WITHSTAND TEST MODE SPECIFICATIONS     Limited Output option  02     FUSE  OUTPUT RATING  OUTPUT ADJUSTMENT    HIGH TRIP RANGE    LOW TRIP RANGE    FAILURE DETECTOR    VOLTAGE DISPLAY    CURRENT DISPLAY    DC OUTPUT RIPPLE  AC OUTPUT WAVE FORM  AC OUTPUT FREQUENCY    OUTPUT REGULATION    115 VAC  415    47 63 Hz  Single Phase  230 VAC  415    47 63 Hz  Single Phase  User Selectable   115 VAC   10 Amp  230 VAC   5 Amp  SkV      20 mAAC  amp  10mADC   0   5      AC  amp         10 volt step    0 1   20 0mA AC  amp 10 0mA DC  Accuracy    2  of setting   0 02 mA     0 0   19 9mA AC  amp  9 99mA DC  Accuracy    2  of setting   0 02 mA     Audible  amp  Visual  0 00 to 5 00kV Full Scale  LED Display  Accuracy   Reading     2  of reading   1 count     Setting    2  of setting   5 volts     0 05 to 9 99mA  10 0 to 20 0mA  LED Display  Accuracy  Reading    2  of reading   0 02mA      lt  4  ripple RMS  SkVDC  40mA     Sine wave  Distortion  lt  1   50 or 60Hz   100 PPM  User Selectable    1  of setting from no load to full load       55         ASSOCIATED  RESEARCH  INC  OPTIONS     Limited Output o
6.      REMOTE INTERFACE CONFIGURATION    OPERATION    REMOTE INTERFACES FRONT  amp  REAR PANEL     Front Panel Remote Interface Rear Panel Remote Interface          RESET IES REMOTE  e    ld          STATUS OUTPUT  COM       PROCESSING  PASS   7   FAIL   7   ALARM                      A             20             Cr PONI  72  1                                                  PRGM 1        im PRGM 5   cS PRGM 11                 O  ajja                   FRONT PANEL REMOTE INTERFACE    The 6554SA front panel remote connector  2  allows remote operation of the  TEST and RESET functions  In order to activate the remote connector a jumper  must first be wired across pins 1 and 5  The matching female connector to attach  to the front panel connector is provided with the 6554SA  Any type of  momentary switch can then be wired across pins 2 and 3 to allow remote  operation of the TEST function  When the remote connector is activated the front  panel TEST switch  4  is deactivated  A momentary switch can also be wired  across pins 3 and 4 which allows remote operation of the RESET function  For  safety the front panel RESET switch  3  remains active even when a remote reset  switch is connected so that high voltage can be shut down from either location     REAR PANEL REMOTE INTERFACE    The 6554SA can operate in a basic remote control function even when GPIB control is  not utilized  The rear panel connector provides output signals to remotely monitor PASS   FAIL  PROCESSING and ALARM 
7.   ASSOCIATED  RESEARCH  INC                                         EH                                   5       CONTROLS  FRONT PANEL CONTROLS  5 6 7 8 9 10 11 12 13  QUADCHEK    Electrical Safety Compliance Analyzer  W crest             5                OUTPUT  NH OCPIB RMT      LOCK                         5    EH w test    w Pass     NH W FAIL    EN                     MILLIAMPS     EH  esT       rass             STATUS DATA    TEST PARAMETER    TEST VALUE             RESET                   TEST          FUNCTION       DATA ENTRY  8                   HIGH VOLTAGE                                                                               n          2   RECALL     LOCAL         an    an STORAGE   ADDRESS  POWER REMOTE 1 2 3 CURRENT RETURN      mm um um RUN MILLIOHN                ee MODEL  65545    RESEARCH  INC 9  13 24 26 25 232119 18 17 16 1415  24   20  22    markings       POWER SWITCH  Rocker style switch with international ON       and OFF  0       REMOTE CONNECTOR  The remote interface provides a convenient way to    connect the instrument to a remote system for limited remote control capability  See  page 42 for a complete description of the remote interface connector signals and    guidelines for connection to a test system       RESET SWITCH  This is a momentary contact switch  If an out of range leakage    current condition  an arc breakdown  a Hi limit or LO Limit failure or ground bond  failure occur you will need to reset the system before you can proceed
8.   Do not switch the line voltage selector switch located on the rear panel  while the instrument is on or operating  This may cause internal damage and represents a  safety risk to the operator    NOTE    For operation at 115 Volts AC use a 10A fast blow fuse   For operation at 230 Volts AC use a 5A fast blow fuse     27         ASSOCIATED  RESEARCH  INC  INSTALLATION    Power Cable  WARNING    BEFORE CONNECTING POWER TO THIS INSTRUMENT  THE PROTECTIVE GROUND   EARTH  TERMINALS OF THIS INSTRUMENT MUST BE CONNECTED TO THE  PROTECTIVE CONDUCTOR OF THE LINE  MAINS  POWER CORD  THE MAIN PLUG  SHALL ONLY BE INSERTED IN A SOCKET OUTLET  RECEPTACLE  PROVIDED WITH A  PROTECTIVE GROUND  EARTH  CONTACT  THIS PROTECTIVE GROUND  EARTH   MUST NOT BE DEFEATED BY THE USE OF AN EXTENSION CORD  POWER CABLE   WITHOUT A PROTECTIVE CONDUCTOR  GROUNDING      This instrument is shipped with a three wire power cable  When this cable is connected to  an appropriate AC power source  this cable connects the chassis to earth ground  The  type of power cable shipped with each instruments depends on the country of destination     Operating Environment    This instrument may be operated in temperatures from 32     113   F   0     45         Relative humidity of 0 to 95     Altitude up to 15 000 feet  4 600 meters      STORAGE AND SHIPMENT  Environment    This instrument may be stored or shipped in environments with the following limits                                                      40   to  75      
9.   Most  specifications call for test currents of between 10 and 30 amps  Test voltages at these  currents are typically required to be less than 12 volts  Maximum allowable resistance  readings of the safety ground circuit are normally between 100 and 200 milliohms     The 6554SA provides up to 30 amps output current at any voltage between 5 and 12 volts  through the safety ground of the product under test  Simultaneously the instrument  measures the induced voltage across the safety ground circuit to determine the impedance  of the ground connection  The meter displays the resistance reading of the ground circuit  in milliohms     The measured values are typically very low so it is extremely important to avoid reading  the resistance of the test leads that are used to connect the test instrument to the product  under test  If this is not done a device may be tagged as having a safety ground failure  when it is actually the combined resistance of the DUT and the test leads that has caused  the maximum resistance level to be exceeded  The 6554SA milliohm offset feature can be  adjusted to disregard the resistance of the test leads     IF YOU SHOULD HAVE ANY QUESTIONS RELATING TO THE OPERATION       OF YOUR INSTRUMENT CALL 1 800 858 TEST 8378  IN THE U S A     11         ASSOCIATED  RESEARCH  INC     SPECIFICATIONS    Model 6554SA Functional Specifications    DIELECTRIC WITHSTAND TEST MODE    FUSE  OUTPUT RATING  OUTPUT ADJUSTMENT    HIGH TRIP RANGE    LOW TRIP RANGE    FAILURE
10.   Use the    down arrow and review the parameters you wish to change  Make your  changes using the numeric keypad entry for    numeric    values or the    ENTER    key for  toggling a test ON OFF     3   After you have made all your changes press the  RUN CLEAR  key to return to the  ready state for your test  You should then see    SEL FUNC    in the  STATUS DATA   LED window     4   If you wish to keep this changed test or new test for reuse at a later time then you  must store it in memory  If you only wish to use these parameters until power is shut off  you may then bypass the storage and go right into your testing  To store the new  parameters press the  STORAGE  key  You will then see displayed in the   STATUS DATA  window STO     The number shown in place of the underline will  be whatever stored program you are currently in  Please enter the number of the memory  you wish to store this program under from the numeric keypad  Then press    ENTER     key  The test parameters you entered or changed will now be stored in the selected  memory position for reuse at a later time     36         ASSOCIATED  RESEARCH  INC  OPERATION  C   Recalling an Existing Test to Use     If you wish to use the test you just stored then you may proceed by simply pressing the     TEST    button  To recall and use a different test then press    RECALL     Enter the  number for the test you desire using the numeric keypad  Press    ENTER    key  You will  be returned to the Ready Test State  Yo
11.   ac voltage or dc voltage     11  TEST VALUE LED DISPLAY WINDOW  An led numeric display of the test  values are displayed here  These numeric display will vary depending upon which of  the three test modes you are in  milliohms  milliamps  or megohms     12  TEST VALUE LED   S  There are three led   s to the right of the    TEST VALUE       window  Depending on the mode you are in an led will light up for Milliohms   Milliamps  or Megohms     13  HIGH VOLTAGE ON INDICATOR  This indicator lights to warn the operator  that high voltage is present at the high voltage output terminal     21               ASSOCIATED  RESEARCH  INC                                            CONTROLS  QUADCHEK  Electrical              Compliance Analyzer    EH       OUTPUT       GPIB RMT m                                  5     EH     5        w Pass EH         Sas                                              NH  rass M ra       STATUS DATA TEST PARAMETER TEST VALUE          CAUTION  4 HIGH VOLTAGE         5        HIGH VOLTAGE  RESET TEST DATA ENTRY FUNCTION  RECALL                                                                                                                                            umm      umm                      cnan aD  POWER REMOTE 1 2 3 CURRENT RETURN  i                MN        SM  MODEL 65545A  13 24 26 25 232119 18 17 16 14 15  24   20  22    14  HIGH VOLTAGE OUTPUT JACK  For the connection of the detachable 5 foot   1 52 m  high voltage test lead  The silicone rubber
12.   c rass      G ralL OUTPUT       GPiB RMT      LOCK EH                    1     5       w test                 5   E w ra fF a      kac       vans       voc                             51 EH  rass M ra STATUS DATA TEST PARAMETER TEST VALUE CAUTION  HIGH VOLTAGE  4        SKV  HIGH VOLTAGE  RESET TEST DATA ENTRY FUNCTION  7 8 9 M    RECALL            ee     4 5 5 M    STORAGE   ADDRESS          E     POWER REMOTE 1 2 3 CURRENT RETURN  E  6       GEB       MIULIOHN  4                 2   ASSOCIATED MODEL  6554SA    RESEARCH  ING     13 24 26 25 232119 18 17 16 1415    22     store    test program setup  Adjust or enter test parameters then press   RUN CLEAR   key then press    STORAGE    and select a memory  1 15  then press     ENTER  key     23     STORAGE   FUNCTION KEY  Depress this key when in the memory function to    RUN CLEAR  FUNCTION KEY   Use this key when you desire to enter the Run    Mode to initiate a test or use this key when you enter numeric data during a setup  routine and you wish to clear the information     24     UP ARROW        FUNCTION KEY  Use this key to advance through the setup    screens  DOWN ARROW      FUNCTION KEY  Use this key to advance  through the setup screens     25     KEYPAD ENTER  Use the enter button after entering new data through the keypad    to save it or to advance the program to its next routine  Enter will also toggle  between certain selections     26     DATA ENTRY NUMERIC KEYPAD  For numeric entry or change of testing    par
13.  132V  10     196 265V   5A  50 60  2                                           REMOTE 1 0          STATUS OUTPUT                PROCESSING                         SCAN BUS    A caution AY    REMOTE PROGRAM RISK OF ELECTRICAL SHOCK  DO NOT REMOVE COVER   REFER SERVICE TO MANUFACTURER S SERVICE CENTHR         FOR CONTINUED PROTECTION AGAINST FIRE  REPLACE       ONLY WITH 250V FUSE OF THE SPECIFIED CURRENT   spp         RATING                                                          1 2 3 4 5 6    8   FUSE RECEPTACLE  To change the fuse unplug the power  mains  cord and turn  the fuse receptacle counter clockwise  The fuse compartment will be exposed  Please  replace the fuse with one of the proper rating     9   RETURN OUTPUT JACK  For the connection of the detachable 5 foot  1 52 m   return test lead  This lead is always used when performing a test     10   CURRENT OUTPUT JACK  For the connection of the detachable 5 foot  1 52 m   current output lead used for the ground bond test  This lead is only used for the  ground bond test     11   HIGH VOLTAGE OUTPUT JACK  For the connection of the detachable 5 foot   1 52 m  high voltage test lead  The silicone rubber insulation is flexible for easy  handling and is rated at 30KVDC  The jack is recessed for safety when this lead is  not being used     12   ARC SENSITIVITY ADJUSTMENT  When the arc detection in the setup menu    is on  this adjustment will increase or decrease its sensitivity to an arcing condition   See page 40     26 
14.  DETECTOR    VOLTAGE DISPLAY    CURRENT DISPLAY    DC OUTPUT RIPPLE   AC OUTPUT WAVE FORM  AC OUTPUT FREQUENCY  OUTPUT REGULATION    DWELL TIMER    RAMP TIMER    115 VAC  415    47 63 Hz  Single Phase  230 VAC  415    47 63 Hz  Single Phase  User Selectable   115 VAC   10 Amp  230 VAC   5 Amp  5       40           amp  40mADC   0   5      AC  amp         10 volt step    0 1   40 0mA       amp  DC  Accuracy    2  of setting   0 02 mA     0 0   39 9mA AC  amp  DC  Accuracy    2  of setting   0 02 mA     Audible  amp  Visual    0 00 to 5 00kV Full Scale  LED Display    Accuracy   Reading     2  of reading   1 count   Setting    2  of setting   5 volts     0 05 to 9 99mA  10 0 to 40 0mA  LED Display  Accuracy  Reading    2  of reading   0 02mA      lt  4  ripple RMS  SkVDC  40mA    Sine wave  Distortion  lt  1    50 or 60Hz   100 PPM  User Selectable  1  of setting from no load to full load    1   999 seconds in   second increments or  continuous  Accuracy   0 1 seconds    0   99 seconds in 1 second increments   Accuracy   0 1 seconds            ASSOCIATED  RESEARCH  INC     Model 6554SA Functional Specifications Cont d    SPECIFICATIONS    INSULATION RESISTANCE TEST MODE    OUTPUT VOLTAGE RANGE 500   1000 Volts  1 volt step    VOLTAGE METERING 3 Digits  0 00   1 00kV  Accuracy     2  of reading   1 count     RESISTANCE METERING 3 Digits  9 99G    Full Scale    Accuracy  500   1000VDC   0 00        1 99 GO    3  of reading   3 MQ   2 00 GO   9 99        15  of reading    LIMIT RA
15.  RESEARCH  INC  OPTIONS    05 Lockout with Memory Recall    The Lockout with Recall Memory option allows users to enter different parameters in  various memory locations for testing multiple products that might require separate set ups   Once the memories are set up and stored   the user can lock out the front panel set up  mode  See page 40 of this manual   With this option  while in the Lockout mode  the user  can select different memories but the parameters within the memory can not be changed     06 Dual Remote Test Switch    The Dual Remote Test Switch option allows the user to configure dual palm switches for  safe production line operation  The front panel remote interface is reconfigured to allow  two Test switches instead of the standard Reset and Test inputs  The two Test switches  have to be pressed within 0 5 seconds to activate the test process  The Two Test Switch  must remain closed to continue the test  If either of the Test switches are released  the  process will be shut down immediately  The functions of Test and Reset switches on the  front panel will be disabled while the dual Test switches are connected to the Control  Ports  Ifthe dual Test switches are not connected to the Control Ports  the functions of  Test and Reset switches will remained the same as the standard instrument     07 Remote Reset GPIB status    The Remote Reset GPIB status option allows the user to monitor the remote reset input   see page 42 of this manual for remote operation  throu
16.  accuracy of  lt  0 5      1   Digital Multimeter with the following minimum ranges   ac voltage  1 000 volts  dc voltage  1 050 volts  ac current  30 amps  dc resistance  201A  200uA         30 mA ranges     2   125KO 10Watt resistor   3   10KO  4Watt resistor   4   110MCO  A4Watt resistor   TO ENTER CALIBRATION MODE     In order to enter the calibration of this instrument you must depress the 0 and the 1 on the  numeric keypad and at the same time you power up the instrument  You do not need to  continue holding these keys  You may release them when you see the    LED    panel lights  go       Upon power up you will briefly see the              name in the    STATUS DATA     window  It will then change to    CAL  MODE   You are then in the    Calibration Mode    The displays should look similar to those below  There should be no    LED   S    lit at this  time  Ifthere is press the    Run Clear    button then the    Reset    button  This will return  you to the status as shown below     CAL  MODE    STATUS DATA  TEST PARAMETERS TEST VALUE    60         ASSOCIATED  RESEARCH  INC  CALIBRATION  There are essentially three main areas that are addressed in the calibration of the 6554SA     D  Calibration of the Withstand Voltage Measuring Circuit  high voltage and leakage  current  All adjustments are done in DC mode  The AC calibration is automatically  adjusted through the microprocessor by using the DC value     II   Calibration of the Insulation Voltage Measuring Circuit  hi
17.  clean and uncluttered  All test equipment and test leads not absolutely  necessary for the test should be removed from the test bench and put away  It should be  clear to both the operator and to any observers which product is being tested  and which  ones are waiting to be tested or have already been tested     Do not perform Hipot tests in a combustible atmosphere or in any area where combustible  materials are present     TEST OPERATOR    Qualifications    This instrument generates voltages and currents which can cause harmful or fatal electric  shock and must only be operated by a skilled worker trained in its use     The operator should understand the electrical fundamentals of voltage  current  and  resistance  They should recognize that the test instrument is a variable high voltage power  supply with the return lead directly connected to earth ground and therefore  current from  the high voltage output will flow through any available ground path     Safety Procedures   Operators should be thoroughly trained to follow these and all other applicable safety rules  and procedures before they begin a test  Defeating any safety system should be treated as  a serious offense and should result in severe penalties  such as removal from the Hipot  testing job  Allowing unauthorized personnel in the area during a test should also be dealt  with as a serious offense        Dress  Operators should not wear jewelry which could accidentally complete a circuit          ASSOCIATED  R
18.  insulation is flexible for easy    handling and is rated at 30KVDC  The jack is recessed for safety when this lead is  not being used     15  RETURN OUTPUT JACK  For the connection of the detachable 5 foot  1 52 m   return test lead  This lead is always used when performing a test     16  CURRENT OUTPUT JACK  For the connection of the detachable 5 foot  1 52 m     current output lead used for the ground bond test  This lead is only used for the  ground bond test     17  LOCAL   FUNCTION KEY  Use this key when you wish to go from the Remote  operation of the instrument to the Local mode     18  ADDRESS   FUNCTION KEY  Use this key when you desire to change the address  location for your GPIB  IEEE 488  remote control activities  or if you wish to set the  instrument into lockout mode using a special code number see 32                ASSOCIATED  RESEARCH  INC     5    6 7    11 12    CONTROLS    13                   QUADCHIEEK        m    ectrical Safety Compliance Analyzer                                        55        G FAIL                     w rass    w ra    EH  esT       Pass    ra                                  LOCK    STATUS DATA             TEST PARAMETER                     kac   EH                    TEST VALUE         MILLIOHMS        wavs                   5          RESET TEST                         FUNCTION          DATA ENTRY  8                      QUTPUT       CAUTION  4 HIGH VOLTAGE         5KV    HIGH VOLTAGE                                           
19.  is the last memory in this    group     PROGRAM THREE   Momentarily connecting terminals 6  7 and 8 together signals the  instrument to immediately begin the test program that is stored in memory eleven  If this  memory is linked to the next memory it will also move on to perform the next test setup  until it reaches a memory that is not linked or until it reaches the last memory of the  instrument     This section provides information on the proper use and configuration of the remote  interface  GPIB  IEEE 488     A Brief History of IEEE 488        Hewlett Packard designed in 1965 the Hewlett Packard Interface Bus  HP IB  to connect  their line of programmable instruments to computers  This bus had high transfer rates   nominally 1 Mbytes s   and thus quickly gained acceptance  Later  it was accepted as the  IEEE Standard 488 1975 and has since evolved into ANSI IEEE Standard 488 1 1987     IEEE 488 has expanded over the years and is used with many more types of computers  and instruments than just HP  Because of this it is usually referred to as the General  Purpose Interface Bus   GPIB      GPIB Messages    There are typically two types of messages that GPIB devices use to communicate with  other interconnected GPIB devices     Interface messages  often called commands or command messages and Device  dependent messages often called data or data messages     Data Messages  contain information such as programming instructions or measurement  results    Command Messages perf
20.  proper wiring before connecting the Model 655484 to it     CAUTION   9  Check to be sure the correct input line voltage has been selected on    the rear panel  Either 115 volts AC or 230 volts AC  Connect the power  input plug into its socket on the rear panel of the instrument  Connect the male end of the  plug to the outlet receptacle  Please be sure that the safety ground on the power line cord  is not defeated and that you are connecting to a grounded power source     d   Turn on the POWER switch located on the lower left hand side of the front panel   Upon powering the instrument up a POWER ON SELF TEST  POST  will be  automatically be performed  This test will check for the condition of the led indicators   All of the front panel led s will light temporarily then go out  In addition you will see the  Associated Research name  ARI appear in the    STATUS DATA    window and the main  LED S will return to a ready state 0 00 as depicted below     A AC  MILLIOHMS  SEL FUNC B KV AC  B MILLIAMPS    KV DC  MEGOHMS    32         ASSOCIATED   RESEARCH  INC  QUICK START  If you wish to not use        one of these parameters you must overwrite the             1  position or change your parameters and save them in a different memory such as memory    position 2  For detailed instructions on setting up testing parameters and saving them to  memory refer to the OPERATION section    Types of Tests        e   Ifthe instrument defaults are acceptable then be sure to connect the appropri
21.  to the next test   Press and release the red Reset switch  See PASS FAIL LED PANEL  5     19               ASSOCIATED  RESEARCH  INC                                                                                                                                                                             a    QUADCHIEK  Electrical Safety Compliance Analyzer     42   15                         55      G FAIL OUTPUT                          Lock              EH viuos i  EN ws    w Pass NE oral                     EN waes                                                            STATUS DATA TEST PARAMETER TEST VALUE CAUTION  HIGH VOLTAGE  2 MAX         HIGH VOLTAGE  RESET TEST DATA ENTRY FUNCTION  7 8 9     RECALL                                      aD     5     gt     storace                               EP          NND  POWER REMOTE    2    CURRENT   RETURN       uU GNE  ay  RUN MLLO             ED GEN                2   ASSOCIATED MODEL  6554SA   amp  RESEARCH  INC 9    4  TEST SWITCH  This is a momentary contact switch  Press the blue switch to turn  on the high voltage output     5  PASS FAIL LED PANEL  This panel gives an indication of which mode  G Ground  Bond  W Withstand Voltage   or R Resistance mode you are in  In addition it shows  via a    RED    led a failure for each mode or a    PASS    for each mode     6  GPIB REMOTE LED  This indicator will light when the 6554SA is in the  Remote   connection vs  the  local  connection  When the GPIB Remote is on the 6554S
22. A is  able to send and receive signals across the GPIB IEEE 488 bus     7  KEY LOCKOUT LED  When the Key Lockout light is on the  password  software  lockout has been enabled  This means that the users will be unable to access the   program  mode of the instrument to change any settings     8  STATUS DATA LED WINDOW  Test abbreviations as well as    DATA    that you    enter in the instrument via the numeric keypad are displayed here  See page 30 for a  complete list of abbreviations     20               ASSOCIATED  RESEARCH  INC                                            CONTROLS   6 7 8 9 10 11 12 13  QUADCHEK    Electrical Safety Compliance Analyzer         c rass m        GPIB RMT EH Lock EH AAC     s        eee EN      EE          B                 m EE  rass         STATUS DATA    TEST PARAMETER    TEST VALUE          RESET                                                          CAUTION  4 HIGH VOLTAGE  MAX 5KV    HIGH VOLTAGE                                                                            DATA ENTRY FUNCTION  7    9   RECALL  POWER REMOTE 1 2 3 CURRENT RETURN  i        em EJ RUN MILLIO HNM        Ss  MODEL  65545  3 4 26 25 232119 1817 16 1415  24 20  22    9  TEST PARAMETER LED WINDOW  This is a numeric display of test parameters  or results for ac current  ac voltage or dc voltage     10  TEST PARAMETER LED   S  There are three led   s to the right of the    TEST    PARAMETER    window  Depending on the mode you are in an led will light up for  ac current
23. Advanced Test Equipment Rentals    A ished 1981 www atecorp com 800 404 ATEC  2832           2  ASSOCIATED  RESEARCH  INC     OPERATION AND SERVICE MANUAL                                                                            QUADCHEK       Electrical Safety Compliance Analyzer  MAX             Mmm OUTPUT  NH                    LOC NE            vuos  m NHw  ss NH EE             NH            NH              ECONS                            PARAMETER        CAUTION                            HIGH VOLTAGE  DATA ENTRY  UNCTION     AN RECALL LOCAL                                5 M  STORAGE ADI       NE             ND              CURRENT  RETURN              LEAR REVIEW OFFSET                             2   ASSOCIATED MODEL 6554SA    WE RESEARCH  Ince                   ASSOCIATED RESEARCH  INC   905 CARRIAGE PARK AVENUE  LAKE BLUFF  IL   60044  2248 U S A    PHONE  1  847  295 3312  FAX  1  847  295 9165  E MAIL  info asresearch com  http   www asresearch com    FOR TECHNICAL ASSISTANCE    PHONE  1  800  858 TEST  8378     Item   37484    5 YEAR WARRANTY POLICY    Associated Research  Inc   certifies that the instrument listed in this manual meets  or exceeds published manufacturing specifications  This instrument was calibrated using  standards that are traceable to the National Institute of Standards and Technology  NIST      Your new instrument is warranted to be free from defects in workmanship and  material for a period of  1  year from date of shipment  You must retu
24. C  MILLIAMPS       KV DC     MEGOHMS    G TEST G PASS    W TEST W PASS       R TEST mR PASS    A AC  MILLIOHMS  W REVIEW KV AC  MILLIAMPS    E KV DC     MEGOHMS    38    e ASSOCIATED  RESEARCH  INC     Flow Chart of QuadChek Menu Selections     Power On    Use  Defaults       Enter           1 15        use existing  test      change test    Change  Parameters    Press  Run Clear    OPERATION                      don t store             store    Press  Storage Key          Press  Enter             Press    Run Clear       Review  Results    Press  Down Arrow    Press  Run Clear    39         ASSOCIATED  RESEARCH  INC  OPERATION    SETUP MODE LOCKOUT CODE    The Model 6554SA is supplied with a lockout code feature that is intended to prevent an  unauthorized user from entering the set up mode  The lockout is active when the front  panel LED    7  is lit  To enter the  lockout mode  press the  address  front panel key    18  Use the numeric keypad and enter the number 6473  Then press front panel key   25   enter   This will turn on the keylock and also turn it off     KEYS ACTION REQUIRED    ADDRESS Press ADDRESS key    6 4  7 3 Press numeric keys 6 4 7  3    ENTER Press ENTER  key    CALIBRATION MODE LOCKOUT CODE    To enter the calibration mode you must hold down the 0 and the 1 on the numeric keypad  at the same time while you power up the instrument  The screen will display    ARI    in the   STATUS DATA  LED display area then will go through a front panel LED test  When  
25. ESEARCH  INC     Medical Restrictions  This instrument should not be operated by personnel with heart ailments or devices such  as pacemakers     SAFETY       TEST PROCEDURES  INEVER PERFORM A HIPOT TEST ON ENERGIZED CIRCUITRY OR EQUIPMENT     If the instrument has an external safety ground connection be sure that this is connected   Then Connect the return lead first for any test regardless of whether the item under test is  a sample of insulating material tested with electrodes  a component tested with the high  voltage test lead  or a cord connected device with a two or three prong plug     Plug in the high voltage test lead only when it is being used  Handle its clip only by the  insulator   never touch the clip directly  Be certain that the operator has control over  any remote test switches connected to the Hipot  Double check the return and high  voltage connections to be certain that they are proper and secure     On Models 6550DT and 6554SA the return side of the instrument is not grounded   earthed   This allows for the monitoring of very low leakage levels of current  It is  therefore important that the device under test is never grounded  earthed  or the  current meter will essentially be bypassed and you will get incorrect current meter readings        WARNING    NEVER TOUCH THE ITEM UNDER TEST OR ANYTHING CONNECTED TO IT WHILE  HIGH VOLTAGE IS PRESENT DURING THE HIPOT TEST     When testing with DC  always discharge the capacitance of the item under test and  any
26. ILLIAMPS    KV DC  MEGOHMS    34         ASSOCIATED  RESEARCH  INC  OPERATION    PASS FAIL LED INDICATION PANEL   The pass fail indication panel will display the test mode you are currently in such as  ground bond  withstand or insulation resistance  It will also display the status of your test    whether you have a    PASS    condition or a    fail    condition  This LED sequence changes  automatically when you select one or more tests to be sequenced              OF TEST PASS INDICATION FAIL INDICATION TEST DESC   GROUND BOND  WITHSTAND    RESISTANCE       TYPES OF TESTS     A   Perform a test using the    DEFAULTS     NOTE  Your instrument will always return  to the last test used upon power up     B   Enter a NEW TEST or CHANGE an existing test     C   Recall an existing test to use     A   Perform a test using the    DEFAULTS        If the instrument defaults that are setup in    memory 1    are correct for your testing  situation you may then proceed right to performing your test  After the proper connection  to the DUT  device under test  you simply power on the instrument until it comes into the     ready for test state    as shown on the previous page  Then depress the blue    TEST     button  Your test  s  will the automatically be performed  If you wish to review the     DEFAULT    test parameters before initiating your test you can review the defaults as  listed on page 31  or do the following       35         ASSOCIATED  RESEARCH  INC     How to Review Test Param
27. MS  RAH  1 01 KV AC  0 2 MILLIAMPS     KV DC  m MEGOHMS    Key in the reading on the DC Ammeter display using units of 0 1 pA  For example  If the  reading is 99 7 uA then you would enter 9 9 7 ENTER J  Then the low range of the  resistance measuring circuit is calibrated and you are returned to the main calibration  menu                                   b   High Resistance Range    Connect a resistor with a value of about 110     in series with a standard microammeter  between the  H V   and the  RETURN  terminals as in the diagram above  Press the  numeric keypad 5  The B R  TEST led will come on as well as the HV indicator  You will  then be in the high resistance range calibration mode and the display will show as below     A AC  MILLIOHMS  BAI 1 01 KV AC  114 MILLIAMPS     KV DC     MEGOHMS    Please enter the reading on the micoammeter into the instrument in units of 0 01 pA  For  example  If the reading is 9 92 uA then enter 9 9 2 ENTER J  The high range of the  insulation measuring circuit is calibrated and you are returned to the main calibration  screen     63         ASSOCIATED  RESEARCH  INC          Calibration of the Ground Bond Measuring Circuit     CALIBRATION    a   AC Voltage    Connect a standard AC Voltmeter between the    RETURN    terminal and the     CURRENT    terminal per the drawing below                                                        The instrument will automatically provide 10 volts AC across the voltmeter upon pressing  the 0 key on the n
28. NGES HI Limit range   0   9999MO  LO Limit range   0   9999  0    DWELL TIMER 1   999 seconds in 1 second increments  Accuracy   0 1 seconds       GROUND BOND TEST MODE    OUTPUT VOLTAGE Adjustable  5 0 to 12 0 volts AC  0 1 volt step   Setting Accuracy      1  of setting   0 1 volt     OUTPUT FREQUENCY 50 or 60Hz  User Selectable  OUTPUT CURRENT Adjustable  10 0 to 31 0 amps  0 1 amp step     CURRENT DISPLAY 3 digits  31 0A Full Scale  LED Display  Accuracy     1  of reading   1 count     RESISTANCE DISPLAY 3 digits  600mQ Full Scale  LED Display  Accuracy     1  of reading   1 count   Auto Offset function to disregard lead resistance    0   600mQ      15 amps or less  0   200mO   15   31 amps    DWELL TIMER 1   999 seconds in 1 second increments  or continuous  Accuracy   0 1 seconds       13         ASSOCIATED  RESEARCH  INC     SPECIFICATIONS    Model 6554SA Functional Specifications Cont d    GENERAL SPECIFICATIONS    INTERFACE CAPABILITY    TEST SET UP DISPLAY    GPIB  IEEE 488  Control of all  parameters AC  amp  DC test voltages  HI  amp   LO trip current  50 60 Hz mode  Dwell  Timer  Ramp time  HI  amp  LO Resistance  trip  Storage  amp  Recall of memorized setups   Test  amp  Reset  Ground Bond current  Ground  Bond resistance  Ground Bond voltage  Arc  on off  Milliohm Offset     Basic Remote control Inputs  Test  Reset   Memory FunctionsOutputs  Pass  Fail   Remote Alarm  Test in Process    Special port for connection to optional  Scanning system to test up to 8 i
29. ONS  current  This option effects only the Dielectric Withstand Test mode and allows an  increased 50 mA of output current in both AC and DC mode  The high and low current    trip settings are also increased accordingly  The revised Dielectric Withstand Test mode  specifications are as follows     DIELECTRIC WITHSTAND TEST MODE SPECIFICATIONS   High Current Output option  01     FUSE  OUTPUT RATING  OUTPUT ADJUSTMENT    HIGH TRIP RANGE    LOW TRIP RANGE    FAILURE DETECTOR    VOLTAGE DISPLAY    CURRENT DISPLAY    DC OUTPUT RIPPLE   AC OUTPUT WAVE FORM  AC OUTPUT FREQUENCY  OUTPUT REGULATION    DWELL TIMER    RAMP TIMER    54    115 VAC  415    47 63 Hz  Single Phase  230 VAC  415    47 63 Hz  Single Phase    115 VAC   10 Amp  230 VAC   5 Amp  SkV      50 mAAC  amp  50mADC  0   5kV  AC  amp  DC   10 volt step    0 1   50 0mA       amp  DC  Accuracy    2  of setting   0 02 mA     0 0   49 9mA AC  amp  DC  Accuracy    2  of setting   0 02 mA     Audible  amp  Visual    0 00 to 5 00kV Full Scale  LED Display    Accuracy   Reading    2  of reading   1 count   Setting    2  of setting   5 volts     0 05 to 9 99mA  10 0 to 50 0mA  LED Display  Accuracy  Reading    2  of reading   0 02mA      lt  4  ripple RMS  SkVDC  40mA    Sine wave  Distortion  lt  1    50 or 60Hz   100 PPM  User Selectable  1  of setting from no load to full load    1   999 seconds in 1 second increments or  continuous  Accuracy   0 1 seconds    0   99 seconds in 1 second increments   Accuracy   0 1 seconds     
30. This is a key reason that some safety agencies do not accept DC testing as an  alternative to AC     4  When performing AC hipot tests the product under test is actually tested with peak  voltages that the hipot meter does not display  This is not the case with DC testing  since a sinewave is not generated when testing with direct current  In order to  compensate for this most safety agencies require that the equivalent DC test be  performed at higher voltages than the AC test  The multiplying factor is somewhat  inconsistent between agencies which can cause confusion concerning exactly what  equivalent DC test voltage is appropriate     The Insulation Resistance Test       Some  dielectric analyzers today come with a built in insulation resistance tester    Typically the IR function provides test voltages from 500 to 1 000 volts DC and resistance  ranges from kilohms to gigaohms  This function allows manufacturers to comply with  special compliance regulations  BABT and VDE are two agencies that may under certain  conditions require an IR test on the product before a Hipot test is performed  This  typically is not a production line test but a performance design test     The insulation resistance test is very similar to the hipot test  Instead ofthe go no go  indication that you get with a hipot test the IR test gives you an insulation value usually in  Megohms  Typically the higher the insulation resistance value the better the condition of  the insulation  The connections t
31. Voltage  Withstand Test Equipment must be connected to a good ground  Be  certain that the power wiring to the test bench is properly polarized and that the proper  low resistance bonding to ground is in place     Power to the test station should be arranged so that it can be shut off by one prominently  marked switch located at the entrance to the test area  In the event of an emergency   anyone can cut off the power before entering the test area to offer assistance     ess     Work Area   Perform the tests on a nonconducting table or workbench  if possible  If you cannot avoid  using a conductive surface  be certain that it is securely grounded to a good earth ground  and insulate the high voltage connection from the grounded surface     SAFETY    There should not be any metal in the work area between the operator and the location  where products being tested will be positioned  Any other metal in the work area should  be connected to a good ground  never left  floating      Position the tester so the operator does not have to reach over the product under test to  activate or adjust the tester  If the product or component being tested is small  it may be  possible to construct guards or an enclosure  made of a non conducting material such as  clear acrylic  such that the item being tested is within the guards or enclosure during the  test  and fit them with switches so that the tester will not operate unless the guards are in  place or the enclosure closed     Keep the area
32. ameters during the  setup  mode  Keypad entry is unavailable if the Key Lockout  indicator on the front panel is on     24                  ASSOCIATED  RESEARCH  INC     REAR PANEL CONTROLS    CONTROLS    12 11 7 10 9 8                   Fuse Rating  98  132V  10A  196 265V   5A  50 60HZ                                                    STATUS OUTPUT                   E     caution AY    PROGRAM RISK OF ELECTRICAL SHOCK  DO NOT REMOVE COVER   REFER SERVICE TO MANUFACTURER S SERVICE CENTRR  FOR CONTINUED PROTECTION AGAINST FIRE  REPLACH                                                PRON 1  jv          6 ONLY WITH 250V FUSE OF THE SPECIFIED CURREN  4   RATING    Ni                            1 2 3 4 5 6  1   THERMAL COOLING FAN  Automatically cycles on off   2            488  GPIB  CONNECTOR  See page 46     3   SCAN BUS CONNECTOR  For connection of optional 8 channel Scanning Bus  Matrix     4   REMOTE I O  See page 42     5   INPUT POWER RECEPTACLE  Standard IEC 320 connector for connection to  a standard NEMA style line power  mains  cord     6   CHASSIS GROUND  EARTH  TERMINAL  This terminal should be connected  to a good earth ground before operation     7   INPUT POWER SWITCH  Line voltage selection is set by the position of the    switch  In the up position it is set for 110 120 volt operation  in the down position it  is set for 220 240 volt operation     25            ASSOCIATED  RESEARCH  INC  CONTROLS                   ARC  SENSITIVITY        A Fuse Rating      98 
33. an Associated Research  Inc  authorized  service center  Schematics  when provided  are for reference only     Service Interval   The instrument and its power cord  test leads  and accessories must be returned at least  once a year to an Associated Research authorized service center for calibration and  inspection of safety related components  Associated Research will not be held liable for  injuries suffered if the instrument is not returned for its annual safety check and maintained    properly        User Modifications   Unauthorized user modifications will void your warranty  Associated Research will not be  responsible for any injuries sustained due to unauthorized equipment modifications or use  of parts not specified by Associated Research  Instruments returned to Associated  Research with unsafe modifications will be returned to their original operating condition at  your expense        TEST STATION    Location   Select an area away from the main stream of activity which employees do not walk  through in performing their normal duties  If this is not practical because of production  line flow  then the area should be roped off and marked for HIGH VOLTAGE  TESTING  No employees other than the test operators should be allowed inside     If benches are placed back to back  be especially careful about the use of the bench  opposite the test station  Signs should be posted   DANGER   HIGH VOLTAGE  TEST IN PROGRESS   UNAUTHORIZED PERSONNEL KEEP AWAY      Power   Dielectric 
34. ate test  leads to the device under test  DUT  or test fixture  Be sure to connect this safety ground  to a suitable known good ground before energizing this instrument  Then connect the  return lead first  black  to the test fixture or item followed by the high voltage output lead   red      WARNING     Please check your connections to be sure they are making good   contact and that the test station or area is clear of debris and other  personnel  DO NOT TOUCH THE DEVICE UNDER TEST ONCE THE TEST HAS  BEEN STARTED  To initiate the test press the BLUE test button on the front panel   This is a momentary button and does not need to be held in the pressed position during the  test  The instrument will then cycle ON and begin the automated test sequence using the  defaults       failure occurs you will HEAR an audible alarm go off  To stop the alarm  you must depress the RED button marked RESET  This will silence the alarm and reset  the instrument to begin another test     This RESET button may also be used as a safety button to quickly ABORT a test and cut  off the HIGH VOLTAGE  When HIGH VOLTAGE is present a RED indicator located in  the upper right side of the front panel will glow and remain ON until the HIGH  VOLTAGE is OFF  Ifthe device under test PASSED the test then no audible alarm will  sound  You will hear a brief BEEP to let you know the item was successfully tested and it  PASSED  In the case of a FAIL condition the instrument will provide a memory of the  test cond
35. atic test program  TEST     Stop automatic test program  RESET     Enable GROUND BOND test mode  Disable GROUND BOND test mode  Set the output of GROUND BOND test to 60 Hz  Set the output of GROUND BOND test to 50 Hz  Enable WITHSTANDING VOLTAGE test mode  Disable WITHSTANDING VOLTAGE test mode    Set the output of WITHSTANDING VOLTAGE AC  mode    Set the output of WITHSTANDING VOLTAGE DC  mode    Set the output of WITHSTANDING VOLTAGE test  to 60 Hz    Set the output of WITHSTANDING VOLTAGE test  to 50 Hz    Enable INSULATION RESISTANCE test  Disable INSULATION RESISTANCE test    Enable CONNECT mode to run next program    Disable CONNECT mode to run single program only    Enable ARC DETECT mode    Disable ARC DETECT mode       Actuate Milliohm Offset         ASSOCIATED  RESEARCH  INC     Parameter Set  Function    OPERATION    Set GROUND BOND maximum test voltage   5 0   12 0V  0 1 V step     Set GROUND BOND test current   10 0   31 0A  0 1A step     Set GROUND BOND high resistance limit   0   199m  Im step     Set GROUND BOND test time   0   999Sec  1Sec step     Set voltage of WITHSTANDING VOLTAGE test   0 5000v  10v STEP     Set upper limit of leakage current of  WITHSTANDING VOLTAGE test current   0 0    40 0mA  0 1mA STEP     Set lower limit of leakage current of  WITHSTANDING VOLTAGE test current  0 0   39 9mA  0 1mA STEP     Set time of WITHSTANDING VOLTAGE test   0 999Sec   1 Sec STEP     Set ramp time of WITHSTANDING VOLTAGE test   0 99 Sec 1 Sec STEP     Set voltag
36. by devices     Interface Management Lines  Five lines are used to manage the flow of information  across the interface     ATN  attention  ATN is driven true by the controller when it uses the data lines to send  commands  and drivers ATN false when a Talker can send data messages     IFC  interface clear  IFC is driven by the system controller to initialize the bus and  become CIC     REN  remote enable  The REN line is driven by the controller which is used to place  devices in remote or local program mode     45         ASSOCIATED   RESEARCH  INC  OPERATION  SRQ  service request  The SRQ line can be driven by any device to asynchronously  request service from the Controller     EOI  end or identify  This line has two purposes  the Talker uses this line to mark the  end of a message string  and the Controller uses it to tell devices to identify their response  in a parallel poll     GPIB Connector   Connection is usually accomplished with a 24 conductor cable with a plug on one end and     connector at the other end  Devices may be connected in a linear  star      a combination    configuration     The standard connector is the Amphenol or Cinch Series 57 Microribbon or AMP  CHAMP type  The GPIB uses negative logic with standard transistor transistor login   TTL  levels  When DAV is true  for example  it is a TTL low level   lt  0 8 V   and when  DAV is false  it is a TTL high level    2 0 V      Restrictions and Limitations on the GPIB    A maximum separation of 4 m betwee
37. cannot charge a capacitive load the current reading remains  consistent from initial application of the voltage to the end of the test  Therefore   there is no need to gradually bring up the voltage since there is no stabilization  required to monitor the current reading  This means that unless the product is  sensitive to a sudden application of voltage the operator can immediately apply full  voltage and read current without any wait time     Another advantage of AC testing is that since AC voltage cannot charge a load there  is no need to discharge the item under test after the test        AC testing disadvantages    1     A key disadvantage of AC testing surfaces when testing capacitive products  Again   since AC cannot charge the item under test  reactive current is constantly flowing  In  many cases the reactive component of the current can be much greater than the real  component due to actual leakage  This can make it very difficult to detect products  that have excessively high leakage current     Another disadvantage of AC testing is that the hipot has to have the capability of  supplying reactive and leakage current continuously  This may require a current  output that is actually much higher than is really required to monitor leakage current  and in most cases is usually much higher than would be needed with DC testing   This can present increased safety risks as operators are exposed to higher currents     DC testing characteristics       During DC hipot testing 
38. ceptable leakage current  The 6554SA  has the reading resolution to monitor and set trip points at these low levels     17    18     19     20          ASSOCIATED  RESEARCH  INC  SPECIFICATIONS    17     Output voltage fine adjustment    To make the 6554SA usable in all types of applications  a feature was added to allow  the operator to manually bring the voltage up or down in 10 volt increments by  simply pressing the up and down arrow keys  This makes it very easy to adjust the  output voltage even while the 6554SA is in the dwell mode so you can analyze test  results at different voltages     Heavy duty color coded switches    The 6554SA uses the same rugged switches that AR has used on other models of  hipots for over 10 years which have proven to withstand even the roughest  manufacturing environment  The switches are also color coded so that the operator  can quickly distinguish between the TEST and RESET switch at a glance     Milliohm offset capability in the Ground Bond mode    The 6554SA includes an offset function so that the resistance of the test leads can be  easily eliminated from the test results during a Ground Bond test  This system allows  the use of longer test leads and test fixtures without compromising the test results     Adjustable output current  output voltage and trip current on the Ground Bond test  mode    This capability makes the 6554SA versatile enough to meet various safety agency  specifications for the ground bond test requirements        
39. completed you will see the normal    READY    state as depicted below  You are then in  the calibration mode  No LED S in the    PASS FAIL    LED PANEL should be on  See  the SERVICE section of this manual for detailed calibration information     A AC  MILLIOHMS  CAL  MODE KV AC  MILLIAMPS    KV DC  MEGOHMS    ARC SENSITIVITY ADJUSTMENT  The arc sensitivity control is located on the rear panel control   11  The software allows    you to turn the arc detection circuit on off by toggling the  enter  key  You can  when the  arc detector is set to  ON  also adjust its sensitivity  With the arc detector set to OFF you    40    CT     OPERATION  will be able to draw arcing conditions as long as the current you draw is lower than the  HIGH TRIP LEAKAGE setting in your setups  Example  Set the HIGH TRIP  LEAKAGE to the maximum of the instrument  40mA  You will then be able to  with the  addition of a resistor value in series with the high voltage test lead  create an arcing  condition without the instrument going into failure  You should determine the acceptable  leakage values of your devices and then set the arc sensitivity to the proper threshold for a  failure condition     CONNECT PROGRAM    This instrument has the capability of linking memorized setups together so that they can  automatically be performed in sequence  While configuring each setup the operator can  elect to turn CONNECT            in the setup they are currently programming to have the  instrument proceed after 
40. conditions  These signals are provided by four  normally open internal relays that switch on to indicate the current condition of the tester   A common terminal is provided so that an external voltage can be applied to the remote  connector to operate external devices  The maximum contact rating of the internal relays    42         ASSOCIATED  RESEARCH  INC  OPERATION  are 220VAC at 3 amps  Below is a listing that indicates what conditions activate each pin     When a terminal becomes active the relay closes and connects that terminal to the  common terminal thereby allowing the external voltage to operate an external device     PIN ONE   COMMON  PIN TWO   PROCESSING  PIN THREE   PASS   PIN FOUR   FAIL   PIN FIVE   ALARM    The following describes how the relays operate for each test condition     PROCESSING   The relay contact closes the connection between the common pin  1  and  pin  2  while the instrument is performing a test  The connection is opened at the end of  the test     PASS   The relay contact momentarily closes the connection between the common pin  1   and pin  3  after detecting that the item under test passed all tests     FAIL   The relay contact closes the connection between the common pin  1  and pin  4   after detecting that the item under test failed any test  The connection is opened when the  next test is initiated     ALARM   The relay contact closes the connection between the common pin  1  and pin     5  after detecting that the failure alarm has s
41. e  Underwriters Laboratories   Inc   UL   the Canadian Standards Association  CSA   the International Electrotechnical  Commission  IEC   the British Standards Institution  BSI   the Association of German  Electrical Engineers  VDE   the Japanese Standards Association  JSI   These same  agencies may also require that an insulation resistance test and high current ground bond  test be performed     The Dielectric Withstand  Hipot  Test       The principle behind a dielectric voltage   withstand test is simple  If a product will  function when exposed to extremely adverse conditions  it can be assumed that the  product will function in normal operating circumstances     The most common applications of the dielectric withstand test are        Design  performance  Testing     determining design adequacy to meet service  conditions         Production Line Testing     detecting defects in material or workmanship during  processing         Acceptance Testing     proving minimum insulation requirements of purchased parts       Repair Service Testing     determine reliability and safety of equipment repairs     The specific technique varies with each product  but basically  during a dielectric voltage    withstand test  an electrical devise is exposed to a voltage significantly higher than it  normally encounters  The high voltage is continued for a given period of time     During the test  all  stray  current flow to ground is measured  If  during the time the  component is tested  s
42. e of INSULATION RESISTANCE test   200   1000v  1v STEP     Set upper limit of RESISTANCE ON INSULATION  test  0 2000 Megohms  1 Megohm STEP     Set lower limit of RESISTANCE ON INSULATION  test  0 2000 Megohms  1 Megohm STEP     Set judgment delay time of INSULATION  RESISTANCE  0 999Sec  1Sec STEP        51         ASSOCIATED  RESEARCH  INC     Storage  amp  Recall Store all status and parameters to memory bank   1 15     OPERATION    Recall a memory from bank to controller  1 15     Scanning Unit Set GROUND BOND test channel of scanning unit  Control  1 4     Set WITHSTAND VOLTAGE test channel of  scanning unit  H  High  L  Low  O  Off     Set INSULATION RESISTANCE test channel of  scanning unit  H  High  L  Low  O  Off        EXAMPLE OF SETTING VOLTAGE OVER THE IEEE BUS    To write commands over the IEEE bus you must enter a code that is specific to the  software you are using  Then follow the example below     To set the output voltage across the IEEE bus at 1240 volts do the following    Type in the following string      S4 1240     ENTER   This tells the instrument you are  setting the voltage at 1240 volts  A string is a list of ASCII characters  octal or hex bytes  or special symbols  enclosed in double quotes     To set the ramp time of the Withstand Voltage test across the IEEE bus at 10 seconds do  the following  Type in the following string     S8 10     ENTER   This tells the  instrument you are setting the ramp time at 10 seconds     To set outputs 1  amp  2 of 
43. e of about 125KQ  10 watts in series with a standard DC  ammeter with its range set to at least 30mA between the          terminals and the   RETURN   Press the numeric keypad number 3  The     W TEST led will come on as  will the HV indicator light  The instrument will provide aprox   1000 volts automatically  across the resistor in series with the ammeter  See connection diagram below                   cL TURN 9      After proper connection entering number 3 on the numeric keypad the display will look  similar to that below     A AC  MILLIOHMS  WAO  1 00 KV AC  0 03 m MILLIAMPS  a KV DC  MEGOHMS    Current will then flow through the ammeter  Please enter the reading of the ammeter into  the instrument in units of 0 01 mA  For example  If the reading of the ammeter is 8 27  mA  then you would enter 8 2 7 ENTER  J  The leakage current is then calibrated and  you will be returned to the main calibration screen                                        Calibration of the Insulation Voltage Measuring Circuit  a   Low Resistance    Connect a 10K    standard resistor in between the  H V   terminal and  RETURN  and a  DC Ammeter in series with the          output as shown in the next diagram     62         ASSOCIATED  RESEARCH  INC  CALIBRATION                      ETURN    Press the numeric keypad number 4  This will enter you into the low resistance calibration  mode  The B R TEST led will come on as will the HV ON Light  The main LED displays  will look as follows     A AC  MILLIOH
44. eters     OPERATION    From the    ready to Test State  depicted above press the    down arrow key  This will  move you through each of the test parameters that you can accept or change  Some  entries require a numeric value to be keyed into the instrument  others require you to  toggle the    ENTER    key  For a complete list of possible entries and their values see page  30     In addition some menus will turn off if the test is not chosen in the setup routine  For  example  Upon entering the setup mode you have a choice of performing an G test   ground bond  and W test  withstand  or an    test  insulation resistance   If R OFF in  the    STATUS DATA    window you will not see the parameters of this test menu  To see  them the Insulation Resistance Test must be set to on  RON  You can change this  parameter by using the    ENTER    key to toggle between ON OFF     After you have reviewed the test parameters of the    MEMORY 1    position you can then  press  RUN CLEAR  key and perform your test     B   To enter a NEW test     You may elect to leave the defaults as they are and stored in memory position 1  or you  may change memory 1 parameters or any other memory position 1 15  To enter a new  test sequence do the following     1   Press the recall and enter the memory position for the memory you wish to review   change or enter  Use the numeric keypad and enter the memory position 1 15  Press the   ENTER  key  You will then be returned to the normal test ready state    2 
45. gh range and low range   III  Calibration of the Ground Bond Measuring Circuit  ac voltage and ac current   Note  You may selectively calibrate a single function or all functions    D   Calibration of the Withstand Voltage measuring circuit    a   High Voltage    Connect a standard DC voltmeter between the          and the  RETURN  terminal   Please be sure that the measuring range of the DC voltmeter is set to a range of at least  1050 volts and the polarity of  H V   and  RETURN  is     and      Press the numeric  keypad number 2  The instrument will then automatically provide aprox   1000 volts  across the voltmeter  Upon entering the number    2    from the numeric keypad the B W   TEST led will activate and the HV warning light will come on indicating you are  calibrating the high voltage of the dielectric withstand circuit  Please enter the reading of  the DC Voltmeter into the instrument using units of 1 volt  For example  Ifthe reading  on the voltmeter is 985 volts you would key in 9 8 5 and press the ENTER  key  The  high voltage is then calibrated and you are returned the main calibration screen            Fi                                        TURN                    After proper connection to the DC voltmeter per the previous diagram the LED display  will show the following     A AC  MILLIOHMS  HVO  1 00 KV AC  0 03    MILLIAMPS       KV DC  MEGOHMS    61         ASSOCIATED  RESEARCH  INC     b   Leakage Current    CALIBRATION    Connect a resistor with a valu
46. gh the GPIB interface bus  This  allows the remote reset to be configured as a safety interlock for test fixturing  The status  command     0    will respond with the state of the Remote Reset at byte 3  bit 6 of the  status bytes that are read back after the     0     Binary 1  high  indicates Reset active and  Binary 0  low  indicates Reset inactive see page 48 for other status bit information     08 High Resolution Current Meter  1uA   The High Resolution Current Meter option allows the user to monitor leakage current  with 1uA resolution  The range however is limited to 4mA maximum  The specifications    that are changed by this option are listed as follows  All other specifications remain  unchanged  Please see page 12 for detail on other specifications     57         ASSOCIATED  RESEARCH  INC  OPTIONS    High Resolution Current Meter  DIELECTRIC WITHSTAND TEST MODE    OUTPUT RATING 5       4 mAAC  amp  4mADC    OUTPUT ADJUSTMENT 0   5      AC  amp  DC   10 volt step    HIGH TRIP RANGE 1   4000uA AC  amp  DC  Accuracy    2  of setting   2 uA     LOW TRIP RANGE 0   3999uA AC  amp  DC  Accuracy    2  of setting   2 uA     CURRENT DISPLAY 005 to 999uA  1 00 to 4 00mA  LED Display  Accuracy  Reading    2  of reading   2 counts        09 DC Charge Low Limit    The DC Charge Low Limit option allows the user to use the low limit function in DC  mode when the load is almost completely capacitive  This condition yields very little  steady state leakage current but substantial cha
47. h as UL  CSA  IEC  VDE and others with a single instrument which takes up less  rack space and makes the connections much simpler     Full IEEE programmability comes as a standard feature in the 6554SA    All functions of the instrument can be programmed over the IEEE bus which makes  the instrument adaptable to an automated system which can control the instrument  and retrieve all test results     Up to 40mA of current available in the AC and DC hipot modes   This makes this instrument a true hipot tester with enough output current to test even  highly capacitive loads     All parameters for the setups can be adjusted through a simple menu driven program  by using a front panel keypad     This provides the operator with an easy and safe way to set trip currents and output  voltages since all parameters are set without the high voltage activated  The easy to  follow menu makes sure that the operator properly sets up each test mode     Front panel LED s display test parameters and results    3 easy to view front panel LED s allows the operator to monitor the test  A review  mode allows the operator to quickly check the test results of each mode after the test  has been completed     Electronic ramp and dwell settings    This electronic ramp control helps keep test results consistent as well as reduce  damage to sensitive products by providing a method to gradually bring up the test  voltage and eliminate any high voltage spikes  The dwell timer also has a count down  feature s
48. ing  resistance s in excess of 200 megohms  Usually employs a higher voltage power supply than used in  ohmmeters measuring up to 200 megohms     Leakage  Ac or DC current flow through insulation and over its surfaces  and AC current flow through  a capacitance  Current flow is directly proportional to voltage  The insulation and or capacitance is  thought of as a constant impedance  unless breakdown occurs     Resistance  That property of a substance which impedes current and results in the dissipation of power  in the form of heat  The practical unit of resistance is the ohm  Symbol   R    Trip Point  The minimum current flow required to cause an indication of unacceptable performance  during a dielectric voltage withstand test     Voltage  Electrical pressure  the force which causes current through an electrical conductor   Symbol   V         ASSOCIATED  RESEARCH  INC     INTRODUCTION    INTRODUCTION    The importance of testing    User safety   In an era of soaring liability costs  original manufacturers of electrical and electronic  products must make sure every item is as safe as possible  All products must be designed  and built to prevent electric shock  even when users abuse the equipment or by pass built  in safety features     To meet recognized safety standards  one common test is the  dielectric voltage withstand  test   Safety agencies which require compliance safety testing at both the initial product  design stage and for routine production line testing includ
49. ition results on the LED display that will remain until the next test is initiated   Depressing the reset button will not prepare the instrument for the next test but will not  clear the meter memory until the next test is started     33         ASSOCIATED  RESEARCH  INC  OPERATION  MAIN MENU SELECTIONS    When you need to access the program set up mode you must be sure that the  lockout   LED on the front panel is not lit  If it is lit you must first take the instrument out of   lockout  before beginning to change or enter new test set up information  To exit or  enter  lockout  mode see the instructions for the  lockout  on page 40     The keys that are used when navigating around in the setup mode are    up arrow moves  you BACKWARD by one selection or Y down arrow moves you FORWARD by one  selection  An over range condition will be displayed on any  numeric  selection as an   ERROR  in the    STATUS DATA    LED display area  The instrument will then  revert back to the previously saved value  Another condition is    OFL     This occurs  during an actual test a stands for OUTSIDE FULL LIMIT  This means the measured  value was outside the full limit of the instruments range     1   INITIAL POWER ON STATE        G TEST B G PASS B G FAIL    B W TEST B W PASS B W FAIL    B R TEST B R PASS B R FAIL          A AC     MILLIOHMS  LED TEST    KV AC     MILLIAMPS     KV DC     MEGOHMS    2   READY FOR TEST STATE     G PASS    W PASS    R PASS       A AC  MILLIOHMS  SEL FUNC KV AC  M
50. ll correspondence to your AR service number     Field Installation Of Options    There are no field installable options on this instrument     29         ASSOCIATED  RESEARCH  INC     ABBREVIATIONS     INSTALLATION    The following abbreviations are used in the    STATUS DATA    window      Abbreviation Defaults                 ear           Scanning Unit G  Test Channels  1 4       0    ee ee    WLA   WT low current trip point selection   0 0 39 9mA O ImA STEPS   0    ee es     R     Insulation Resistance Test          ONOFF    OFF     RYS   IR test output voltage               200 1000   IV STEPS             RHR  R test high trip point selection 10 9999      1 MO STEPS         RLR   IR test low trip point selection 10 9999      I MO STEPS       RTS   IR test dwell time selection   0 999 SEC ISEC STEPS          CON To connect memories 1 15 in ON OFF OFF  sequence for multiple tests    All entries above that require a numeric value use the    NUMERIC KEYPAD    for entry  and then the    ENTER    key        Other entries with only two choices such as AC or DC mode or 50 and 60Hz use the     ENTER    key to toggle between choices and the Wdown arrow key to advance or     RUN CLEAR       key to return to the Run Mode     30         ASSOCIATED  RESEARCH  INC     QUICK START    QUICK START    This quick start guide assumes the operator has some familiarity with automated testing  and desires to use the  default  settings on the instrument  The default settings shown  will remain i
51. n any two devices and an average separation of 2 m  over the entire bus     A maximum total cable length of 20 m     No more than 15 device loads connected to each bus  with no less than two thirds  powered on     Note  A bus extender which is available from numerous manufacturers is available to  overcome these limitations     46         ASSOCIATED  RESEARCH  INC  OPERATION    Interface Functions     The capability of a device connected to the bus is specified by its interface functions   These functions provide the means for a device to receive  process  and send messages  over the bus  The interface functions of the Model 6554SA are listed in the chart below   All functions may be controlled over the bus except ARC sensitivity and input voltage  which are Selectable on the rear panel see rear panel drawing on page 25 _    GPIB INTERFACE FUNCTIONS           488 INTERFACE SH1  Complete source handshake   capability   AHI  Complete acceptor handshake  capability   T6  Talker function   L4  Listener function   SRO  No service request capability           Complete remote local capability   PPO  No parallel poll capability   DCO  No device clear capability     DTO  No device trigger capability   CO  No controller capability   E2  3 state driver     CONTROLLABLE ITEMS Test Reset control   Setting of test status parameters for test     DATA CODES ASCII       DELIMITER CR   LF    EOI     GPIB ADDRESS    Each device on the GPIB  IEEE 488  interface must have a unique address  You ca
52. n memory unless you choose to override them with your own test program   The instrument default settings are as follows     DEFAULTS    Withstand Test     Input Voltage   Communications   Mode    Voltage Output   Voltage Type   Ramp Time   Dwell Timer     Arc detector     Current Trip     AC Output Frequency     Withstand Test Scanner     Insulation Resistance Test     Insulation Resistance Test     Insulation Resistance Scanner     Ground Bond Test     Ground Bond Test     115 volts AC  rear panel switch selectable   local  front panel key selectable    set up   1 240 volts AC   AC   1 second   1 second    on    High Trip  10 mA  LO Trip  off    60Hz    all off  00000000     off    all off  00000000     on    31         ASSOCIATED  RESEARCH  INC     QUICK START  e Ground Bond Test Current  25 amps  e Ground Bond Test Timer  1 second  e Ground Bond Test Frequency  60 Hz       Ground Bond Test output voltage  6 volts    e Ground Bond Test resistance  100 mQ  e Ground Bond Scanner  off  0      CONNECT  off    Quick Start Instructions Con t     a   Unpack the QuadChek Model 6554SA from its special shipping container  Be sure to  save all packaging materials in case you need to return it to the factory for service     WARNING   0   Locate a suitable testing area and be sure you have read all safety  instructions for the operation of the instrument and suggestions on the    test area set up in SECTION I  Locate a three prong grounded outlet  Be sure the outlet  has been tested for
53. n set  the 6554SA   S address to any value between 1 and 29  The address is set to   8   when  the instrument is shipped from the factory  The 6554SA address is displayed when you  depress the  Address  button on the front panel keypad 718     The address can only be set from the front panel  The address is stored in non volatile    memory and does not change when the power has been off or after a remote interface  reset     47         ASSOCIATED  RESEARCH  INC     OPERATION    6554SA IEEE COMMAND LIST    TYPE   COMMAND EXPLANATION    Binary 20  Status    48    Binary status  If the 6554SA talks after reception of the   20  command  it will output four bytes which indicates the  current programmed stats  the first two bytes are empty   The meaning of the last two bytes is     Byte 3  Status Bits    Bit 7   1 an ERROR COMMAND occurred   Bit 6   1   Bit 5   1 the INSULATION RESISTANCE test FAILED  Bit 4   1 the WITHSTANDING VOLTAGE test FAILED  Bit 3   1 the GROUND BOND test FAILED   Bit 2  1 ALL TESTS were PASSED   Bit 1   1 an OVER LOAD failure occurred   Bit 0   1 an ARC failure occurred    Byte 4  Status Bits    Bit 7   1 the WITHSTANDING VOLTAGE test is in DC  mode   Bit 6   1 the voltage of WITHSTANDING VOLTAGE  TEST is RAMPING   Bit 5   1 the INSULATION RESISTANCE test PASSED   Bit 4   1 the WITHSTANDING VOLTAGE test PASSED   Bit 3   1 the GROUND BOND test PASSED   Bit 2   1 the INSULATION RESISTANCE is  PROCESSING   Bit 1   1 the WITHSTANDING VOLTAGE is  PROCESSING   Bi
54. nd Disadvantages of AC Testing and DC Testing        Please check with the Compliance Agency you are working with to see which of the two  type of voltages you are authorized to use  In some cases a Compliance Agency will  allow either AC or DC testing to be done  However in other cases the Compliance  Agency only allows for an AC test  If you are unsure which specification you must  comply with please contact our CUSTOMER SUPPORT GROUP at 1 800 858 TEST   8378      Many safety agency specifications allow either AC or DC voltages to be used during the  hipot test  When this is the case the manufacturer must make the decision on which type  of voltage to utilize  In order to do this it is important to understand the advantages and  the disadvantages of both AC and DC testing     AC testing characteristics       Most items that are hipot tested have some amount of distributed capacitance  An AC  voltage cannot charge this capacitance so it continually reads the reactive current that  flows when AC is applied to a capacitive load     AC testing advantages       1       testing is generally much more accepted by safety agencies than DC testing  The  main reason for this is that most items being hipot tested will operate at AC voltages  and AC hipot testing offers the advantage of stressing the insulation alternately in  both polarities which more closely simulates stresses the product will see in real use          ASSOCIATED  RESEARCH  INC  INTRODUCTION    2     Since AC testing 
55. o perform the IR test are the same as the hipot test  The  measured value represents the equivalent resistance of all the insulation which exists  between the two points and any component resistance which might also be connected  between the two points     Although the IR test can be a predictor of insulation condition it does not replace the need  to perform a dielectric withstand test          ASSOCIATED  RESEARCH  INC  INTRODUCTION    TYPES OF FAILURES DETECTABLE ONLY WITH A HIPOT TEST    Weak Insulating Materials   Pinholes in Insulation   Inadequate Spacing of Components  Pinched Insulation    THE GROUND BOND TEST    The Ground Bonding test determines whether the safety ground circuit of the product  under test can adequately handle fault current if the product should ever become defective   A low impedance ground system is critical in ensuring that in the event of a product failure  a circuit breaker on the input line will act quickly to protect the user from any serious  electrical shock     International compliance agencies such as CSA  IEC  VDE  BABT and others  have  requirements calling out this test  This test should not be confused with simple low  current continuity tests that are also commonly called out in some safety agency  specifications  A low current test merely indicates that there is a safety ground  connection  it does not completely test the integrity of that connection     Compliance agency requirements vary on how different products are to be tested
56. o the operator can clearly see how much time is left on the test     Hi and Low limits on both the hipot and insulation test modes    This capability makes it possible to ensure that a test item was properly connected  since the 6554SA can be set to look for minimum and maximum levels of current and  resistance during the hipot and insulation resistance tests     Line and load regulation    This system maintains the setting of the output voltage to within 1  even if the load  or the line voltage vary to ensure that test results remain consistent and within safety  agency requirements           ASSOCIATED  RESEARCH  INC       Built in basic remote control     10     11     12     13     14     15     16     SPECIFICATIONS    This makes the 6554SA versatile enough to allow for remote control operation of the  test even when it is not used in the IEEE interface mode     Storage of up to 15 different test programs    A real benefit for manufacturers that test different products  This makes it possible to  store all the various test parameters required and quickly recall them for each of the  different products that needs to be tested  Each program can store all the parameters  of either the hipot or the insulation resistance test so you can quickly switch between  different types of tests  Program memories can also be accessed through the remote  control port so that a manufacturer can quickly toggle through the various programs  without even going into the set up menu     Securi
57. orm functions such      initializing the bus and addressing and  unaddressing devices     44         ASSOCIATED  RESEARCH  INC     Functions     OPERATION    A GPIB device can be a Listener  Talker and or Controller  A Talker sends data messages  to one or more Listeners  which receive data  A Controller manages the information flow   on the GPIB by sending commands to all devices  The GPIB bus is much like a computer   bus except a computer has circuit cards connected via a backplane and the GPIB has stand  alone devices connected via a cable     Signals and Lines     The GPIB consists of 16 signal lines and 8 ground return or shield drain lines  The 16  signal lines are grouped into 8 data lines  3 handshake lines and 5 interface management  lines     Data Lines  The eight data lines  DIO1 through DI08 carry data and command messages   The 7 bit ASCII or ISO code set is used and the eighth bit DIO8 is unused     Handshake Lines  The transfer of message bytes between devices is done via three  asynchronously control lines  Referred to as three wire interlocked handshake  This  guarantees that message bytes on the data lines are sent and received without transmission  error     NRFD  not ready for data  indicates when a device is ready or not ready to receive a  message byte     NDAC  not data accepted  indicates when a device has or has not accepted a message    byte     DAV  data valid  tells when the signals on the data lines are stable  valid  and can be  accepted safely 
58. ounded  The connection is opened when the  instrument is reset and the alarm is deactivated     REMOTE INPUTS       The 6554SA also allows access to its setup programs through a remote control  connection  This gives the user the capability to quickly change parameters remotely even  when the GPIB mode is not utilized  The 6554SA basically operates in a PLC mode by  responding to simple relay contact closures  The built in memory programs of the  instrument are used to accomplish this  The internal memory programs are arranged into  three groups  By connecting terminal 6 7 and 8 in different combinations the user can  select which memorized test setups to activate     The following describes how each program is activated and what memorized test setups  are run     43    OTT                         PROGRAM ONE   Momentarily connecting terminal 6 to 7 signals the instrument to  immediately begin the test program that is stored in memory         If this memory is linked  to the next memory it will also move on to perform the next test setup until it reaches a  memory that is not linked or until it reaches memory five which is the last memory in this  group     PROGRAM TWO   Momentarily connecting terminal 6 to 8 signals the instrument to  immediately begin the test program that is stored in memory six  If this memory is linked  to the next memory it will also move on to perform the next test setup until it reaches a  memory that is not linked or until it reaches memory ten which
59. ption  02  cont   d    DWELL TIMER 1   999 seconds in 1 second increments or  continuous  Accuracy   0 1 seconds    RAMP TIMER 0   99 seconds in 1 second increments   Accuracy   0 1 seconds       03 8 Port Ground Bond Scanner Menu   The 8 Port Ground Bond Scanner Menu option enables the Quadchek to interface with an  HS 8 Scanner that has been modified to include four additional Ground Bond Ports  This  menu allows inputs from 1 through 8 for the Scanning Unit G Test Channel parameter   GCH   See page 30 for standard menu parameters    04 1 3A Continuity Test   The 1 3A Continuity Test option enables the Quadchek to make measurments in a higher  range of resistance but with a limited output Current  All other test specifications are    unchanged  See page 12 for standard specifications  The modified Ground Bond test  specifications are as follows     Continuity Test    OUTPUT VOLTAGE Adjustable  5 0 to 12 0 volts AC  0 1 volt step   Setting Accuracy      1  of setting   0 1 volt     OUTPUT FREQUENCY 50 or 60Hz  User Selectable  OUTPUT CURRENT Adjustable  1 0 to 3 1 amps  0 1 amp step     CURRENT DISPLAY 3 digits  03 1A Full Scale  LED Display  Accuracy     1  of reading   2 count     RESISTANCE DISPLAY 3 digits  1 200 Full Scale  LED Display  Accuracy     1  of reading   1 count   Auto Offset function to disregard lead resistance    TRIP RANGE 0 00   1 200    DWELL TIMER 1   999 seconds in 1 second increments  or continuous  Accuracy   0 1 seconds       56         ASSOCIATED 
60. rging current during test voltage ramping   This option modifies the DC Lo Limit function only WLA   the AC operation remains  unchanged  See pages 12 and 30 for specifications and menu settings     The Lo Limit setting will set the low trip level to a value that is checked at approximately  100mS after the test has started  If the charging current has exceeded the set level the test  continues and no further low limit checking is performed  This allows the steady state  current to fall below the WLA setting without causing a failure condition but still verifies  that all test connections were completed and a test was performed     58    SECTION 2  SERVICE MANUAL         ASSOCIATED  RESEARCH  INC  CALIBRATION    CALIBRATION PROCEDURES    This instrument has been fully calibrated at the factory in accordance to our published  specifications  It has been calibrated to NIST  You will find in this manual a copy of the   Certificate of Calibration   It is recommended that you have this instrument recalibrated  and a safety check done at least once per year  AR recommends you use  Calibration  Standards  with an accuracy of  lt  0 5  to keep this instrument within published  specifications  Calibration adjustments can only be made in the Calibration mode   calibration checks can only be made in a Test mode of operation     Calibration Equipment Required     The following equipment will be needed to properly calibrate your instrument  Please be  sure that you use instruments with an
61. rn the    Owners  Registration Card    provided within  15  days from receipt of your instrument     AR recommends that your instrument be calibrated on a twelve month cycle  This  instrument may have its warranty extended in one year increments to a maximum of  5   years provided it is returned to AR at least annually for calibration and inspection  The  annual calibration and inspection must be performed annually each and every year  following receipt of instrument  Any instrument not calibrated and inspected annually will  not be eligible for extended warranty status  This extended warranty is non transferable  and is offered only to the original purchaser     A return goods authorization  RGA  must be obtained from AR before returning  this instrument for warranty service  Please contact our Customer Support Center at  1 800 858 TEST  8378  to obtain an RGA number  It is important that the instrument is  packed in its original contained for safe transport  Ifthe original container in not available  please contact our customer support center for proper instructions on packaging   Damages sustained as a result of improper packaging will not be honored  Transportation  costs for the return of instrument for warranty service must be prepaid by the customer   AR will assume the return freight costs when returning the instrument to the customer   The return method will be at the discretion of Associated Research     Except as provided herein  Associated Research makes no warran
62. successful completion of the first test program and automatically  run the next setup program  If CONNECT is left           the instrument will only run the  current setup program once the test is initiated     Once the operator has linked all the setups they wish to perform as a single test they must  RECALL the first setup of the sequence in order to begin the sequence that has been  setup  The instrument will automatically stop when it reaches the last setup that has  CONNECT turned    off     Ifthe operator initiates the test again the instrument will once  again begin the program from the first setup that the test was originally initiated from     Several multiple setup programs can be stored into the instrument  As an example the  operator could configure the setups as follows     setup  1 CON      on     setup  2 CON      on     setup  3 CON      off     setup  4 CON      on     setup  5 CON      on     setup  6 CON      off       If the test is started while in setup  1 the instrument would proceed up to setup  3 before  stopping if all tests passed  If started while in memory  4 the instrument would proceed  to setup  6 before stopping if all tests passed  In the event that any test would fail the  instrument will indicate failure and stop the test sequence  If the reset button is pushed to  clear the failure and the test is again initiated the instrument will begin testing from the  first test setup in the sequence once again     41         ASSOCIATED  RESEARCH  INC
63. t 0   1 the GROUND BOND is PROCESSING            ASSOCIATED  RESEARCH  INC     Feedback Value    OPERATION    GROUND BOND TEST current  If the 6554SA talks after  reception of the   1  command  it will output four bytes  which indicates its currently measured value   unit  Amps     GROUND BOND TEST resistance  If the 6554SA talks  after reception of the   2  command  it will output four  bytes which indicates its currently measured value   unit   milliohms     WITHSTANDING VOLTAGE TEST voltage  If the  65545    talks after reception of the   3  command  it will  output four bytes which indicates the current measured  value   unit  KILOVOLTS     WITHSTANDING VOLTAGE TEST current If the  65545    talks after reception of the   4  command  it will  output four bytes which indicates the current measured  value   unit  MILLIAMPS     INSULATION RESISTANCE TEST voltage If the  65545    talks after reception of the   5  command  it will  output four bytes which indicates the current measured  value   unit  KILOVOLTS     INSULATION RESISTANCE TEST value  If the 6554SA  talks after reception of the  26  command  it will output  four bytes which indicates the current measured value    unit  MEGOHMS     TIMER counter  If the 6554SA talks after reception of the   27  command  it will output four bytes which indicates the  current counter value   unit  Seconds        49         ASSOCIATED  RESEARCH  INC     Operation  Function    Operation Status  Function    50    OPERATION    Start autom
64. t flow is not directly proportional to voltage  Once breakdown current has flown  especially for a  period of time  the next gradual application of voltage will often show breakdown beginning at a lower  voltage than initially     Conductive  Having a volume resistivity of no more than 103 ohm cm or a surface resistivity of no  more than 10  ohms per square     Conductor  A solid or liquid material which has the ability to let current pass through it  and which  has a volume resistivity of no more than 103 ohm cm     Current  The movement of electrons through a conductor  Current is measured in amperes   milliamperes  microamperes  nanoamperes  or picoamperes  Symbol   I    Dielectric  An insulating material which is positioned between two conductive materials in such a way  that a charge or voltage may appear across the two conductive materials     Direct Current  DC  Current which flows in one direction only  The source of direct current is said  to be polarized and has one terminal which is always at a higher potential than the other     Hipot Tester  Common term for dielectric withstand test equipment   Hypot    Registered trademark of Associated Research  Inc   for its dielectric withstand test equipment     Insulation  Gas  liquid or solid material which has a volume resistivity of at least 1012 ohm cm and is  used for the purpose of resisting current flow between conductors     Insulation Resistance Tester  An instrument or a function of an instrument capable of measur
65. tems  simultaneously     Allows storage of up to 15 different test  programs     A separate LED displays the test set up  parameters     All readings in all modes can be reviewed  after the test to monitor test results     Password lockout capability to avoid  unauthorized access to test set up program     Detachable 7 ft   2 13m  power cable  terminated in a three prong grounding plug             ASSOCIATED  RESEARCH  INC     Model 6554SA Functional Specifications Cont d    SPECIFICATIONS    TERMINATION   S 5 ft   1 52m  high voltage  return and current  output leads with clips     MECHANICAL Bench or rack mount with tilt up front feet  Dimensions   w x hx d   17 x 8 75 x 20in   432 x 222 x 508mm   Weight  70 Ibs  31 75kg  net    ENVIRONMENTAL Operating Temperature   32     113  F  0     45  C   Relative Humidity  0 to 95     CALIBRATION Traceable to National Institute of Standards and  Technology  NIST   Calibration controlled by  software  Adjustments are made through front panel  keypad in a restricted access calibration mode   Calibration information stored in non volatile  memory        15         ASSOCIATED  RESEARCH  INC     SPECIFICATIONS  KEY FEATURES  amp  BENEFITS SUMMARY COVERING MODEL 6554SA    The only complete 4 in 1 system that includes an AC hipot  DC hipot  Insulation  Resistance Tester  Ground Bond Tester and IEEE interface in a single rack mount  style cabinet    This allows the user to perform most of the electrical safety tests required by agencies  suc
66. the item under test is charged  The same test item capacitance  that causes reactive current in AC testing results in initial charging current which  exponentially drops to zero in DC testing     DC testing advantages       1     Once the item under test is fully charged the only current flowing is true leakage  current  This allows a DC hipot tester to clearly display only the true leakage of the  product under test     The other advantage to DC testing is that since the charging current only needs to be  applied momentarily the output power requirements of the DC hipot tester can  typically be much less than what would be required in an AC tester to test the same  product          ASSOCIATED  RESEARCH  INC     DC testing disadvantages    INTRODUCTION       1  Unless the item being tested has virtually no capacitance it is necessary to raise the  voltage gradually from zero to the full test voltage  The more capacitive the item the  more slowly the voltage must be raised  This is important since most DC hipots have  failure shut off circuitry which will indicate failure almost immediately if the total  current reaches the leakage threshold during the initial charging of the product under  test     2  Since a DC hipot does charge the item under test it becomes necessary to discharge  the item after the test     3  DC testing unlike AC testing only charges the insulation in one polarity  This  becomes a concern when testing products that will actually be used at AC voltages   
67. the scanner to High  outputs 3  amp  4 to Low and outputs 4 8 to Off  type in the following string   SGHHLL0000      FOR MORE INFORMATION ON IEEE  GPIB  PLEASE CONTACT   The Institute of Electrical and Electronic Engineers  Inc    345 East 47th Street    New York  New York 10017    T 1 212 705 7018  Communications Society of IEEE     52         ASSOCIATED  RESEARCH  INC     OPTIONS  Model 6554SA OPTIONS    Introduction   This section contains a list and descriptions of available factory installed options at the  time of this printing  The list of options contains an option code number which can be  referenced on the data plate on the rear panel of the unit     Data Plate    The option code is contained in the OPT field  The complete model number will include  the standard base model number and end with the letter          when options are present     MODEL  6554SAC_ OPT  L__     Standard base  model number Options    For example your data plate would appear as follows     standard model  no options              MODEL  65545      fitted with option           MODEL  6554SAC OPT  02  fitted with option 02 and 03              MODEL  65545     OPT  0203  6554SA Options    High Current Output  50 mA      1 3   Continuity Test         Description  01 High Current Output    The High Current Output option gives the user the capability to perform Dielectric  Withstand Test on devices that may require greater than the standard 40 mA of test    53         ASSOCIATED   RESEARCH  INC  OPTI
68. thing the high voltage may have contacted  such as test fixtures  before handling it or  disconnecting the test leads     HOT STICK probes can be used to discharge any capacitance in the item under test as a  further safety precaution  A hot stick is a nonconducting rod about two feet long with a  metal probe at the end which is connected to a wire  To discharge the device under test   two hot sticks are required  First connect both probe wires to a good earth ground  Then  touch one probe tip to the same place the return lead was connected  While holding the  first probe in place  touch the second probe tip to the same place where the high voltage  lead was connected     KEY SAFETY POINTS TO REMEMBER     Keep unqualified and unauthorized personnel away from the test area   Arrange the test station in a safe and orderly manner    Never touch the product or connections during a test    In case of any problem  turn off the high voltage first    Properly discharge any item tested with DC before touching connections          ASSOCIATED  RESEARCH  INC     INTRODUCTION  GLOSSARY OF TERMS     as used in this manual     Alternating Current  AC  Current which reverses direction on a regular basis  commonly in the  U S A  60 per second  in other countries 50 times per second     Breakdown  The failure of insulation to effectively prevent the flow of current  sometimes evidenced  by arcing  If voltage is gradually raised  breakdown will begin suddenly at a certain voltage level   Curren
69. ties to the  purchaser of this instrument and all other warranties  express or implied  including   without limitation  merchantability or fitness for a particular purpose  are hereby excluded   disclaimed and waived     Any non authorized modifications  tampering or physical damage will void your  warranty  Elimination of any connections in the earth grounding system or by passing any  safety systems will void this warranty  This warranty does not cover batteries or  accessories not of Associated Research manufacture  Parts used must be parts that are  recommended by AR as an acceptable specified part  Use of non authorized parts in the  repair of this instrument will void the warranty     OPERATION AND SERVICE MANUAL    MODEL 6554SA QUADCHEK       Electrical Safety Compliance Analyzer  AC DC HIPOT WITH INSULATION RESISTANCE TESTER  GROUND BOND  TESTER AND IEEE 488  GPIB  INTERFACE    SERIAL NUMBER                       Associated Research  Inc   1997  905 Carriage Park Avenue  Lake Bluff  Illinois  60044 2248 U S A     Item   37484 Printed Jan 9  1997  Ver 3 06    iii    TABLE OF CONTENTS    SECTION I OPERATORS MANUAL    SAFETY                                                                                      2  02                                 8     2 2  220 22222 1 2      CONTROLS niin    INSTALLATION       QUICK STAR LI                                          ettet terti    OPTIONS eerte    SECTION II SERVICE MANUAL          CALIBRATION oinin i                      
70. tion  GVO Ground Bond Voltage Calibration  GAO Ground Bond Current Calibration         The RUN CLEAR key may be used for clearing an error after entry and to exit froma  specific calibration set up screen to the main calibration menu  Then press RESET  key to exit out of the calibration         The instrument must be turned off after calibration in order for the calibration to be  stored into the EPROM  or to return to the testing or test set up mode     65         ASSOCIATED  RESEARCH  INC     PARTS LIST  MODEL 6554SA QuadCHEK    PARTS LIST    FOR SERVICE INFORMATION IN THE U S A  PLEASE CALL 1 800 858 8378  OUTSIDE U S A  CALL 1 847 295 3312    When ordering replacement parts please provide us with the following information     e Model Number   e Serial Number   e Item Number   e Quantity Required    R C 5 0 2  AND TR 3773       66    em ASSOCIATED    SCHEMATIC INDEX       DrawingNumber   Description    Pages    S06554SA Wiring Diagram  537509 PCB Main Control    537508 PCB Mother Board  537515 Input Protection Board   37510 PCB Display Assy       67    
71. tray current flow remains within specified limits  the device is  assumed to be safe under normal conditions  The basic product design and use of the  insulating material will protect the user against electrical shock          ASSOCIATED  RESEARCH  INC  INTRODUCTION    The equipment used for this test  a dielectric withstand tester  is often called a  hipot   for  high potential tester   The  rule of thumb  for testing is to subject the product to twice its  normal operating voltage  plus 1 000 volts     However  specific products may be tested at much higher voltages than 2X operating  voltages   1 000 volts  For example  a product designed to operate in the range between  100 to 240 volts  can be tested between 1 000 to 4 000 volts or higher  Most  double  insulated  products are tested at voltages much higher than the  rule of thumb      Testing during development and prototype stages is more stringent than production run  tests because the basic design of the product is being evaluated  Design tests usually are  performed on only a few samples of the product  Production tests are performed on each  and every item as it comes off the production line     The hipot tester must also maintain an output voltage between 100  and 120  of  specification  The output voltage of the hipot must have a sinusoidal waveform with a  frequency between 40 to 70 Hz and has a peak waveform value that is not less than 1 3  and not more than 1 5 times the root mean square value     Advantages a
72. ty password system    This makes it possible to limit user access to the setup screens so that only authorized  personnel with a security password can change test parameters  This ensures that the  required test parameters can not be tampered with     Optional scanning system available for use with the model 6554SA   The optional scanning system can cycle through up to 8 high voltage tests and 4  ground bond tests to help manufacturers increase throughput in the final test area     Software calibration control    The 6554SA is calibrated through the front panel keypad  All calibration information  is stored in non volatile memory  This allows the 6554SA to be completely calibrated  without removing any covers and exposing the technician to hazardous voltages     User activated arc detection system    Many tests require the monitoring of arcing levels even if they do not exceed the  maximum trip current level  The 6554SA allows the operator to select whether low  level arcs should be detected which makes this instrument flexible enough to test any  product     User selectable output voltage frequencies of 50 or 60 hertz    The 6554SA was designed for the global market  This feature makes it simple for the  user to select the output frequency on the AC hipot test and the Ground Bond test so  that products can be tested at the same frequency they will be used at     The LED display allows monitoring of current down to 10 microamps     Many tests only allow a very low level of ac
73. u are now ready to perform your test     D   To connect test setups    If you wish to link setups so that they execute in sequence you must activate the  CONNECT mode within the setup you wish to connect to the next by using the enter key  to toggle between off and on    The REVIEW key    The REVIEW key may be used after you have completed a test sequence  Here you can  see the results of any    FAIL    or    PASS    condition  When you depress the    REVIEW     key each time it will move from     R REVIEW  review ofthe Insulation Resistance Test Results    G REVIEW  review ofthe Ground Bond Test Results    W REVIEW  review of the Withstand Voltage Test Results    The TEST PARAMETER LED display and the TEST VALUE LED display will either  show a value for the appropriate test ora       in the display if that particular test was not    selected  In addition as you press the    REVIEW    key you will see the LED S to the right  of each display move to their appropriate settings  See examples below     G PASS    W PASS    B R PASS       A AC  MILLIOHMS  R REVIEW 1 00 KV AC  MILLIAMPS    E KV DC     MEGOHMS    37         ASSOCIATED  RESEARCH  INC  OPERATION    This example depicts a    PASS    condition on a Insulation Resistance Test that is in the     REVIEW    mode  If this was the only test being performed than the G TEST review  display and the W TEST review display would look as follows     G TEST G PASS    W TEST W PASS       R TEST mR PASS    A AC  MILLIOHMS  G REVIEW KV A
74. umeric keypad  The B G TEST led will come on  The displays will  then be similar as below        A AC  E MILLIOHMS  GVO  0 1 KV AC  10 5 MILLIAMPS    KV DC  MEGOHMS    Please key in the reading on the Voltmeter into the instrument via the numeric keypad in  units of 10mV  For example  Ifthe reading on the Voltmeter 1s 10 12 volts you would  press 1 0 1 22  ENTER   The AC Voltage is then calibrated  You will be returned  to the main calibration display     b   AC Current  Connect a standard AC Ammeter between the    CURRENT    terminal and the       RETURN    terminal  Place an Ammeter in series with the    CURRENT    lead per the  following drawing        H V                                   ETURN                64         ASSOCIATED  RESEARCH  INC  CALIBRATION    Please press from the numeric keypad the number 1  The B G  TEST led will come on and  the main led displays will look similar to those below        A AC     MILLIOHMS  GAO  30 1 KV AC  1 3 MILLIAMPS  KV DC  MEGOHMS    Please key in through the numeric keypad the reading on the Ammeter using units of  10      For example  If the Ammeter reads 29 35A  please key in 2 9 3 5  ENTER   The current is then calibrated and you are returned to the main calibration screen     Note  Using    Ohms Law  R V I  it is not necessary to calibrate the resistance because  the voltage and current are calibrated     These abbreviations will appear in the STATUS DATA window during calibration     HVO  WAO    Resistance Low Range Calibra
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
BioPAT® Trace  Dell PowerEdge R720 Reference Configuration  Einbau- und Bedienungsanleitung - Kleinkläranlage INNO  Ultimate Products 1000 User's Manual  Configuración - Área de Seguridad y Comunicaciones  2013 1 NO. 300  Sanyo DP50749 Flat Panel Television User Manual  Unitary Controller (UNT)  SPSD/M - Data Library  Verwenden von Adobe Premiere Elements 10    Copyright © All rights reserved. 
   Failed to retrieve file