Home
        Model 2000 Service Manual
         Contents
1.                                                          Field Definition Units or Range of Values  PT1 Patient Temperature  1   C  PT2 Patient Temperature  2   C  FL Water Flow Rate ml min  T1 Monitor Water Temperature    C  T2 Control Water Temperature    C  T3 Return from Pad Water Temperature     C  T4 Chiller Loop Water Temperature   C  IP Inlet Water Pressure 0 to  10 psi  L Reservoir Level 1   Empty  5   Full  VV Vent Valve 1 Open  0   Closed  BV Bypass Valve 1 Open  0   Closed  PWM P   Circulating Pump Speed 0  Stop to 235   Full  PWM M   Mixing Pump Speed 0   Stop to 150   Full  PWM H   Heater Power 0   Off to 32   Full Power       6 1 2 Error Log and Operational Log    The Model 2000 maintains two history logs  The logs may be accessed while in Stop mode  once  Diagnostic Mode is activated  by pressing the Down Arrow key  The Error Log screens are labeled  UD2 and the Operational Log screens are labeled UD3  The Operational Log contains a history of  the last 16 alerts and recoverable alarms  The Error Log contains a history of the last 8 non   recoverable alarms  Note that power failures do not generate a logged alarm  Each screen provides  information on the state of the Control Module when the event occurred including the current system    hours        Error Log Display  Figure 6 3    31    The history logs are interpreted as follows     Table 6 2    Log Definitions                                                       Field Definition Units or Range of Values   
2.    Chiller Capacity Test       Model 2000 02   Water Temp   lt  12  C    Model 2000 03 or  10   Water Temp   lt  13  C        Blank Page     29    CHAPTER6 TROUBLESHOOTING    6 1 Diagnostic Mode    The Model 2000 has a diagnostic mode which is designed to provide assistance with troubleshooting  and provides access to the Calibration software  To enter this mode  power up the Control Module  while holding the Automatic and Purge Mode keys located on the Remote Display  The keys must be  held for at least 5 seconds  The LCD display will appear as shown below  See Figures 6 1 and 6 2   after the normal power up screen sequence  The control module will operate normally  except these  displays will replace the home display     Caution  Do not operate the device in Diagnostic Mode when the device is providing therapy to a  patient unless closely monitored     Access to screens is available through the Up and Down Arrow Keys  The order of screens in the  Diagnostic Mode while in Stop mode are as follows     1  Diagnostic Home Screens  Standard Operation Screens    UD1     System hours  pump hours  and system voltage setting    2  3  4  UD2  Operational Log  5  UDS  Error Log   6    UD4   Calibration Procedure    30    6 1 1 Diagnostic Home Screens    Toggle between these two diagnostic displays by pressing the Home key           Diagnostic Home Page  1    Figure 6 1       Diagnostic Home Page  2    Figure 6 2       The display fields are     Table 6 1  Display Field Definitions    
3.    Federal law  USA  restricts this device to sale  by or on the order of a physician       Medivance will not be responsible for patient safety or equipment performance if the procedures  to operate  maintain  modify or service the Medivance Arctic Sun are other than those specified  by Medivance       This product should be serviced only by trained  qualified personnel     Use only distilled or sterile water  The use of other fluids will damage the Arctic Sun System       Any device connected to the RS232 data port must comply with the applicable IEC standard for  that device       When the system is powered off  all changes will revert to the default unless the new settings  have been established as new defaults       Users should not use cleaning or decontamination methods different from those recommended by  the manufacturer without first checking with the manufacturer that the proposed methods will not  damage the equipment       His advisable not to cancel the alarm or alert until the situation is resolved  If an alarm is  cancelled and the condition has not been cleared  the alarm will recur  If an alert is cancelled and  the alert condition has not been cleared  the alert will not recur unless the Stop Mode is activated       Do not operate the unit in a diagnostic mode while delivering therapy to a patient      Blank Page     CHAPTER 2 THEORY OF OPERATION    2 4 Introduction    The Model 2000 was designed to control patient temperature with a high level of precision
4.    TEMPERATURE MANAGEMENT SYSTEM    Model 2000  Service Manual    C  Medivance        CHAPTER INTRODUCTION  escscccccesesctcchecceacectetsacetacesacecuacsaccescceacbenstiaceecaaesdndeabcesadeaaecdanesstaececasstes 1  1 1 Use of the Model 2000 Service Manual                     sse nnne 1  1 2 Customer Information and Technical Support    1  1 9  System  Description  i nere EE 1  1 4     Symbols and Standards dne da Anal dn        nn 3  1 5 Environmental             4  1 07    General Warnings  ate eno ee enden Dn 4  127          A A 5  CHAPTER2 THEORY OF OPERATION                      sieesiseeiseessnee nenne entr NEEN                                  tnn              nnne nnns 7  2 1   lhntroduction       serene edie edet ENEE ee ee e tte eee 7  2 2 System Components iii 7  2 9  Hydraulic    Syste M sssi 2               Lie eee aate 7  2 4    Electronic Control System   iiaea he tet edi die tice eed 11  CHAPTER 3  INSTALLATION             recrutent tre EEN ER 13  3 1 Installation of Cables and Accessories    14  3 27    Power  Up System ioco Saee  15  3 3 Filling the Control 6  15  CHAPTER 4                    17  4 1 Recommended Maintenance Schedule eene 17  4 2 Clean External            17  4 3 Adjust Remote Display      18  4 4 Inspect Connectors and Cables eene nnne 18  4 5 Clean the            19  4 6    Drain and  Replace             eet RR RR e ER HR     20  4 7 Calibration and Safety      21  CHAPTER5 CALIBRATION AND SAFETY CHECK                        eeeesesseeeese
5.   Low Water Flow        Low water flow and flow  through one pad is different  from other     Restriction in one pad  allows flow to  preferentially shunt  through other pad   This causes significant  efficiency loss     Eliminate kinks in lines or replace  low flow pad        Pads not adhered completely     Movement of pads  relative to patient   Pads drying out  causing loss of  adhesion     Adhere or replace pads to assure  full contact        Small pads used on large  patient    Improper size selection  reduces surface area  available for transfer     Use largest pads that can fit on  patient  Assure pads are not  overlapped in back        Poor water temperature  control  but flow appears to be  good     Main Tank is overfilled     Drain 0 5 liter water from the  upper drain port  Power up in  User Diagnostic Mode and check  the level  Repeat this process  until the level is less than 5        Poor water temperature  control  Main Tank is not full     Heater or chiller circuit  fault     See  Water does not heat  or   Water does not cool        Pads connected     Water does not heat to  target temperature in  Manual Mode     Flow rate less than 800  ml min     Heater does not  provide full power at  low flow rates     See Low Water Flow        CTU connected     Water does not heat to  target temperature in  Manual Mode     Power up in User Diagnostic  Mode  Enter Manual Mode  with a water target of 42  C   After stabilizing for several  minutes  note whether the  display 
6.   Open the snap ring with a small screwdriver and pull  the tube off the reservoir     Reach through the right side of the chassis and slide the lever up  to unlock the fitting from the flow pump     Remove the fitting     Unscrew the bracket screws on the left side of the control unit  chassis with a 5 32    hex key and remove the heater assembly  and the bracket through the opening     After re assembly of the unit  it is recommended to perform a  Dielectric Withstand  Hipot  Chassis Test per Section 7 23     7 18 Chiller Power Relay    Tools and Supplies Required     5 16    nut driver  Remove the rear panel per Section 7 3     Note the position of each wire and remove all connections from  the relay     Remove the nut on the top and on the bottom of the relay with a  5 16    nut driver     Remove the relay   Reverse the process to replace the relay with a new one     After re assembly of the unit  it is recommended to perform a  Dielectric Withstand  Hipot  Chassis Test per Section 7 23     70    7 19 Chiller Control Relay  Tools and Supplies Required     1 4    nut driver  1  Remove the rear panel per Section 7 3     2  Note the position of each wire and remove all connections from  the relay     3  Remove the standoffs on the top and on the bottom of the relay  with a 74  nut driver     4  Remove the relay     5  After re assembly of the unit  it is recommended to perform a  Dielectric Withstand  Hipot  Chassis Test per Section 7 23     7 20 Inlet   Outlet Manifold  Too
7.   half second tone conditions such as a momentary occlusion  air  cool  every 10 seconds  bolus  hot restart  or no load condition   If the  Alarm Alert indicator   alarm is recurrent  use the CTU to check the  flashing  calibration of the Control Module  Calibrate if  Alarm necessary  If the alarm persists  call Medivance  Service   38 Water temperature Water temperature is   Changes to Stop Operate the unit in User Diagnostic Mode and       below safety level   Allow system to  warm     Alarm       below above 3 0  C   37 4  F         Mode  Audio alarm  half second tone  every 10 seconds   Alarm Alert indicator  flashing        Manual Mode with the CTU attached at a setpoint  of 4  C  Observe T1  T2  T3  and T4 for at least 10  minutes after reaching stability  If T4 drops below  3  C for more than a few seconds  check the  connector at J5 of the I O Card  Verify that it is  plugged in securely  the wires emanating from it are  intact  and that the other end of the wires entering  the solid state relay are intact  Repeat the test   Replace cables or the relay  if necessary  and  retest  If the error persists  call Medivance Service        49          Table 6 4    Alarm and Alert Troubleshooting                               Alarm    Message Displayed   Alarm Alert System Status Troubleshooting   Alert Description   40 System unable to In Manual Mode  Changes to Stop Use the CTU to check the calibration the Control  maintain stable system is unable to Mode  Audio alarm Modul
8.  24 and U 26     2  Place the new chip with the odd number extension  number  after the dash  into    24     3  Place the new chip with the even number extension into U 26     Software Chip Location  Figure 7 18    Monitor Software Chips  1  Using the extractor tool remove the chip from U11     2  Install the new chip     Programmable Logic Device  1  Using the extractor tool remove the chip from U 21     2  Install the new chip     7 13 Handle  1  Unscrew the knobs on both sides of the handle   2  Replace with a new handle    67       U11    7 14    Py w qu  ce    Circulating Pump   Tools and Supplies Required      5 32    hex key  Remove the rear panel per Section 7 3  Remove the left side panel per Section 7 4   Drain the control unit per Section 7 1     Locate the cable that comes from the pump and unplug from  the Input Output circuit card     Note  Observe the routing of the wire harness for the installation    7 15    Aog m      of the new pump assembly     With a 5 32  hex key remove the 2 screws on the side panel  that hold the bracket of the pump  The two screws are  located toward the rear of the unit   Disconnect the tube from the pump by pushing the lever on  the pump up   Remove the tube fitting   Repeat the process for the tube fitting in the back of the  pump   At this point the pump assembly can be removed from the  chassis    Mixing Pump    Tools and Supplies Required       5 32    hex key    Remove front grill per Section 7 2   Remove rear panel per Section
9.  7 3   Remove the side panels per Section 7 4     Disconnect the cable that comes from the pump to the  Input Output Circuit Card  see Section 7 10      Note  Observe the routing of the wire harness for installation of  the new pump assembly     5     Remove the two screws that hold the pump bracket using a  5 32    hex key  located to the front of the unit     Push the lever up on the pump ports to disconnect the tube  fitting     Repeat the process for the tube fitting on the back of the  pump     At this point the mixing pump can get removed from the  chassis     68       E Ri    Circulating Pump Location  Figure 7 19       Circulating Pump Mounting Screws  Figure 7 20       Mixing Pump Mounting Screws  Figure 7 21       Mixing Pump Mounting Screws  Figure 7 22    7 16 Chiller Pump  Tools and Supplies Required     10   11     12     harness  Blade screwdriver Fe  5 32    hex key  Remove the rear panel per Section 7 3   Remove the right side panel per Section 7 4   Disconnect the chiller harness at the connector located  directly above the chiller pump assembly  Grad    Remove the insulation from around the tubing on the bottom  of the chiller reservoir     With a screwdriver  loosen the hose clamp ring around the  tubing   Pull the tubing off from under the tank  Chiller Pump Harness Location    Remove the two screws on the metal frame that hold the Figure 7 23  bracket of the chiller pump assembly with a 5 32    hex key     Remove the insulation from around the tube attac
10.  Allow them to dry  if necessary  and retest   Refer to Operator resistance half second tone If the error persists  call Medivance Service   Manual  every 10 seconds   Alarm Alert indicator  Alarm flashing   75 Non recoverable T3 Thermistor signal   Changes to Stop Check the connector at J9  pin 1 and J9  pin 7 of  System error out of range   high Mode  Audio alarm the I O Card  Verify that it is plugged in securely   Refer to Operator resistance half second tone the wires emanating from it are intact  and that the  Manual  every 10 seconds  other end of the wires entering the sensor body is  Alarm Alert indicator   intact  Power up and allow the Control Module to  Alarm flashing  run for 10 minutes  If the error persists  call  Medivance Service   76 Non recoverable T3 Thermistor signal   Changes to Stop Ensure that the I O Card or J9  pin 1 and J9  pin 7  system error out of range   low Mode  Audio alarm connector on it are not wet  Allow them to dry  if  Refer to Operator resistance half second tone necessary  and retest  If the error persists  call  Manual  every 10 seconds  Medivance Service   Alarm Alert indicator  Alarm flashing   77 Non recoverable T4 Thermistor signal   Changes to Stop Check the connector at J10 of the I O Card  Verify  system error out of range   high Mode  Audio alarm that it is plugged in securely  the wires emanating  Refer to Operator resistance half second tone from it are intact  and that the other end of the wires  Manual  every 10 seconds  e
11.  Ensure that the rubber washers are installed on the  temperature sensors of the new harness  Observe the  labeling on the cable to identify the temperature sensors     Install T4 into the chilled water reservoir located toward the  front of the control module     10  Install T1 in the bottom corner of the mixing reservoir     11  Install T2 in the bottom center of the mixing reservoir     12  Cover the temperature sensors with the provided insulation     Note  Don t bend the wires too sharply while applying the  insulation     13  If the temperature sensor harness is replaced  calibrate per    Chapter 5     72       Sensor Locations   tank insulation not shown for clarity     Figure 7 26    7 22 Level Sensor Harness    Be    D  gt     10     11     Tools required     7 16  nut driver    7 16    open end wrench     3 32 hexkey    Silicone RTV Adhesive  Drain the control module per Section 7 1     Remove the front grill per Section 7 2    Remove the rear panel per Section 7 3    Remove the left side panel per Section 7 4    Note  Observe the routing of the Level Sensor Harness     Remove the Input Output Manifold for access per Section 7 20   Pull the block toward you until it disengages from the chassis  but do not remove connectors or lines     With a 7 16  wrench hold the tank insert in place so it does not  rotate  This can be done through the opening of the left side of  the control module     Remove the screw and terminal with a 3 32  hex key from the  tank insert     Re
12.  F  and the flashing  system is unable to  heat in Automatic  Mode  24 Patient Temperature   Patient Temperature   Current Mode Adjust Patient High Temperature Alert in the  2 below user defined   2 reading is equal or   continues  Audio custom menu if desired   limit  less than the user  alarm half second  defined value tone every 2  established in the minutes  Alarm Alert   Alert custom settings indicator flashing   menu    25 System is warming Patient Temperature   Current Mode Operate the system in Manual Mode with a setpoint  and Patient 2 reading is above continues  Audio below 38  C until the water temperature is less than  Temperature 2 high    38 0  C  100 4  F  alarm half second 38  C  Ifthe error persists  see General   and water tone every 2 Troubleshooting   Will Not Cool   temperature is above   minutes  Alarm Alert  Alert 38 0  C  100 4  F  indicator flashing   and the system is  unable to cool in  Manual Mode   26 System is cooling Patient Temperature   Current Mode Operate the system in Manual Mode with a setpoint  and Patient 2 reading is below continues  Audio above 32  C until the water temperature is greater  Temperature 2 low  32   C  89 6  F  and alarm half second than 32  C  If the error persists  see General   water temperature is   tone every 2 Troubleshooting   Will Not Heat   below 32 0 C minutes  Alarm Alert  Alert  89 6  F  and the indicator flashing   system is unable to  heat in Manual  Mode   27 Patient Temperature   Patient Temperature   Chan
13.  If the alarm persists  call Medivance Service     Note  The calibration process can be aborted and restarted at any  time by pressing Enter     6  Power up while holding the Automatic and Purge Keys     7  At the User Diagnostic Home Screen  press the Up Key once   twice if Patient Temperature 2 is enabled   to scroll to the  UD4 Screen  Press the Enter Key to initiate the Test and  Calibration Mode     8  Follow the prompts on the Remote Display Module     Note  The Control Module software will ask for the CTU calibration  data  termed  A  through  G   at various points in the process   This data is printed on the CTU label     25       CTU Connections  Figure 5 2    Medivance   Calibration Test Unit   REF 741 00 115   46  60Hz 4 5A   is Colorado USA 8007   Dg a M ENT    Ltd  Arundel House  1 Liverpool   Tr  CA                    CTU Switch 1  Figure 5 3       CTU Switches 2 and 3  Figure 5 4    CTU S N 7001    A  10 88  B  25 55    C 34 54  D  39 99  E 0 02  F   0 07       CTU Calibration Data  Figure 5 5    5 2 2 Stage 4  Temperature Out Calibration    1     Press the Enter Key if calibration is desired  The sequence  may be aborted at this stage if only verification is desired     The    Calibrate System    menu screen displays the number of     Cal Hours    that have elapsed since the system was last  calibrated     5 2 3 Stage 5  Acceptance    1s    Save the calibration data  if desired  by pressing Enter on the     Save Calibration Values  screen  If only verification
14.  Mode for 5 Water leaking under Replace the leaking component   minutes with pads connected  positive pressure will  Place the pads on the floor indicate the exact leak  and then place the system in site   Stop Mode  Dry all water  from the pads  lines and  manifold then place them on  the floor  Wait 5 minutes   check carefully for water  leaking from the pads  lines or  valves   Pads connected  Check the pads for folds or Pads folded or creased Smooth out and reapply pad to  buckles created during foam to foam will remove buckles or creases   Low water flow  The pump placement  reduce flow  can be heard  significantly   CTU connected  Power up in User Diagnostic If the pressure cannot Debris may be cleared by running  tow wateriow Tow Mode  Enter Manual Mode be maintained  the flow the pump a full speed  To  ue with a water target of 28  C  is low  or the PWM is attempt to dislodge debris   pressure  or high PWM  After stabilizing for several greater than 175  there replace the CTU with a fluid  The pump can be heard  minutes  note whether the may be an external or delivery line and connect two  display indicates  internal leak or debris shunt tubes to the fluid delivery  e 2550  lt   FL  flow   lt   is caught in a pump line manifold  Enter Manual  3150 valve causing Mode  The flow rate should    inefficient pumping  exceed 4 lpm for 5 minutes   e P  6710  7 2  psi  If Remove the shunts and  value is less than 6 7 reconnect the CTU  Repeat the  then the system is unable che
15.  by  rapidly altering water temperature in response to a patient core temperature sensor using a  closed loop algorithm  To achieve rapid response with tight control  the system is designed to  circulate a minimum volume of water  have a high heating and cooling capacity  and include  sensors for real time feedback of energy transfer     2 2 System Components  The Model 2000 system is composed of several components       Control Module     The main unit which includes the hydraulics and major elements of the  electronics       Remote Display Module     The primary control and monitoring interface for the user       Fluid Delivery Line     The reusable dual lumen tubing connecting the Control Module to the  pads       Interconnect Cables and Accessories     The cables that connect the Control Module to mains  power  the Control Module to the Remote Display Module  and optional patient temperature  cables  Accessories include fill tubes  drain bags  Fluid Delivery Line straps and others     2 3 Hydraulic System  The hydraulic system within the Control Module consists of the following major elements       Pumps    Circulating pump    Chiller pump    Mixing pump      Sensors    4 Water temperature sensors    Flow meter    Pressure sensor      Valves    Vent valve    Bypass valve    Fill valve      Heater      Tanks    Circulating Tank    Chiller Tank    Main Tank        nlet Outlet Manifold    Chiller    The hydraulic system is composed of two major fluid circuits  the circulatin
16.  designed and built to have a high degree of reliability  however  occasional failures  do occur  Use the trouble shooting methods in Chapter 6 or consult with Medivance Customer  Service to determine the source component of the failure  Once the source of the failure has been  determined  follow the appropriate procedure for replacement of the component  Not all components  can be replaced in the field  A complete list of spare parts is available from Medivance Customer  Service  If a component fails that is not available as a spare part  contact Medivance Customer  Service     In general  reverse the order of removal to install a component  Otherwise  note any special  instructions     Caution  Observe the electrostatic discharge control procedures  ESD  when working with any circuit  card assembly     56    Fluid Reservoir d      Latch    Fill Line Port Inlet Outlet Manifold    Mains Power Switch  Chiller Control Relay  Locations   Mains Power Inlet  Input Output Circuit    Equipotential Stud Card Connections       Heater Fuse    Circulating Pump  Card Cage    Chiller Unit    Chiller Heat  Exchanger    Reservoir Drain Port    Chill Loop Drain Port       Control Module  Rear Inside View    Figure 7 1    57    Circulating Pump    Side Panel  Keyholes  Mixing Pump  Chiller Pump  Bracket  Chiller Pump  Chiller Pump Bracket   behind    fasteners       Control Module  Right Side    Figure 7 2    58    Level Sensors  6     Heater Flowmeter  Bracket fasteners    Circulating Pump  
17.  half second tone 38  C  Press Automatic and retest  If the error  water temperature is   every 2 minutes  persists  see General Troubleshooting   Will Not  above 38   C Alarm indicator Cool   Alarm  100 4  F  and flashing  System is unable to  cool in Automatic  Mode  09 Patient Temperature   Patient Temperature   Current Mode Adjust Patient High Temperature Alert in the  1 above user defined   1 reading is equal or   Continues  Audio custom menu if desired   limit  greater than the alarm half second  user defined value tone every 2  established in the minutes  Alarm Alert  Alert custom settings indicator flashing  menu   10 System cooling and Patient Temperature   Changes to Stop Operate the system in Manual Mode with a setpoint  Patient Temperature   1 reading is below Mode  Audio alarm above 32  C until the water temperature is greater  1 low  32 0 C  89 6  F  and   half second tone than 32  C  Press Automatic and retest  If the  water temperature is   every 10 seconds  error persists  see General Troubleshooting   Will  below 32 0 C Alarm Alert indicator   Not Heat   Alarm  89 6   F  and the flashing  system is unable to  heat in Automatic  Mode  11 Patient Temperature   Patient Temperature   Current Mode Adjust Patient High Temperature Alert in the  1 below user defined   1 reading is equal or   continues  Audio custom menu if desired   limit less than the user  alarm half second  defined value tone every 2  established in the minutes  Alarm Alert  Alert custom settings
18.  hands at the top sides of the panel  pull it straight  out and away from the control module     3  To replace  align the studs with the holes and push all four  corners        Front Grill Removal  Figure 7 6    7 3 Rear Panel  Tools and Supplies Required     5 32    hex key  Turn off the power and unplug the power cord     2  Detach all cables and lines from the rear panel  fluid delivery  line  fill line  patient temperature cables  remote display cable  and strain relief     3  Remove the four mounting screws at the corners with the hex  key from the rear panel     4  Pullthe panel off horizontally        Rear Panel Removal  Figure 7 7    61    7 4 Side Panels  Tools and Supplies Required     5 32    hex key  1  Remove the rear panel per Section 7 3   Remove the front grill per Section 7 2     Remove the two screws located on the underside of one side  panel with a 5 32    hex key        4  Hold the panel on both sides and move it towards the rear of  the unit  Slide the panel out from front to back  There are four Side Panel Screw Removal  key slots which accept white plastic key nuts  Figure 7 8    5  Repeat this process for the other side panel    6  To replace the side panels  align the white key nuts horizontally   rotated to allow the threaded stud to engage the slots in the  frame    7 5 Card Cage   Tools and Supplies Required     5 64    hex key    5 16  nut driver    Note  Observe electrostatic discharge control procedures when  handling circuit cards     1  Remov
19.  important that the Model 2000 be properly maintained and serviced in accordance with  the instructions described in this manual  Only trained personnel should service this product     1 2 Customer Information and Technical Support    Prior to using the Arctic Sun System  a Medivance representative will provide on site in services  to ensure adequate training has occurred     Additional requests for technical support  information or orders may be placed by mail  fax  or by  calling Medivance Customer Service     Medivance hours of operation are 8 am to 5 pm Mountain Standard Time  Emergency support is  offered after hours in the U S  by calling 303 926 1917 or 877 267 2314  Other customers should  contact their local distributor     Additional educational materials  such as training programs for in house staff  operator manuals   quick reference guides  and a bibliography of relevant materials are available to all Medivance  customers     1 3 System Description    The Arctic Sun Temperature Management System is a thermoregulatory device that monitors and  controls patient temperature within a range of 33  C to 37  C  91 4  F to 98 6 F   The Arctic Sun  System consists of the Model 2000 and disposable Arctic Sun Energy Transfer Pads  The Model  2000 consists of the Control Module  Remote Display  Fluid Delivery Line  cables and  accessories     With the Energy Transfer Pads adhered to the patient s skin  the Model 2000 pulls temperature   controlled water ranging between 4  C 
20.  indicator flashing   menu   12 Patient and water Patient Temperature   Current Mode Operate the system in Manual Mode with a setpoint       temperature high   Check water  temperature     Alert       1 reading is above  38 0  C  100 4  F   and water  temperature is above  38 0  C  100 4  F   and the system is  unable to cool in  Manual Mode       continues  Audio  alarm half second  tone every 2  minutes  Alarm Alert  indicator flashing        below 38  C until the water temperature is less than  38  C  If the error persists  see General  Troubleshooting   Will Not Cool        44          Table 6 4    Alarm and Alert Troubleshooting                      Alarm    Message Displayed   Alarm Alert System Status Troubleshooting   Alert Description   13 Patient and water Patient Temperature   Current Mode Operate the system in Manual Mode with a setpoint  temperature low  1 reading is below continues  Audio above 32  C until the water temperature is greater  Check water 32   C  89 6  F  and alarm half second than 32  C  If the error persists  see General  temperature  outlet water tone every 2 Troubleshooting   Will Not Heat    temperature is below   minutes  Alarm Alert   32 0  C  89 6  F  and   indicator flashing   Alert the system is unable   to heat in Manual   Mode   14 Patient Temperature   Patient Temperature   Changes to Stop Use the CTU to verify the accuracy of the Control  1 probe 1 reading is outside Mode  Audio alarm Module  Check connection of patient temperature  di
21.  is  desired  do not save  If the new data is not saved  the  calibration will revert to the previous set of values     Note  If the Up arrow is pressed at this time a  No Calibration  Values were Saved  prompt displays and pressing the Enter  produces the Exit screen     5 2 4 Stage 6  Water Temperature Alarm Test    1   2     10     Press the Manual Mode Key     Press the Down Key once and change the  Water Target  Temperature  to 42 C     Press the Home Key     Allow the water temperature to stabilize at 42 C for at least 5  minutes     Note  It is important to allow the temperature to fully stabilize  at 42  C  If it does not  it may not be possible to attain the  alarm conditions     Press the S4 switch on the CTU  Monitor the T3 temperature  and release the switch once the temperature reaches 44 5 C     Start the timer when the T1 temperature reaches 43 01  C   Verify that Alarm 37 sounds in less than 18 seconds     Note  If the T1 water temperature drops below 43 01  C   restart the timer when the water again reaches 43 01  C     Turn power off   Turn power on and allow the Control Module to start normally   Press the Manual Mode Key     Press the Down Key once and change the  Water Target  Temperature  to 42  C     26    11   12     13     Press the Home Key     Allow the water temperature to stabilize at 42  C for at least 5  minutes     Note  It is important to allow the temperature to fully stabilize  at 42  C  If it does not  it may not be possible to attain the  
22.  kinking        Look for pad leaks by  inspecting for a continuous  stream of bubbles in the tubes  exiting the pads when no air  is entering     Leak in pad     Replace pad        Enter Stop Mode  then hold  manifold above Control  Module and pads  pad  connector side up  Air fills the  connectors     Leak at seal between  pad connector and the  fluid delivery line  manifold  Air may be  entering the return side  of the fluid delivery line  and is not visible in the  pad lines     Reseat the connector  If leak  continues  replace fluid delivery  line           Disconnect each pad in turn  from the manifold to  determine whether one pad is  contributing to the low flow   Try to prevent air from  entering the lines when  removing the pads and run  long enough to assure  stability before checking the  flow        One pad is not  functioning properly     There is an air leak in  the manifold        Replace pad     Replace fluid delivery line        34       Table 6  3  General Troubleshooting Guide                               Conditions   Symptom Observation Cause Solution  Pads connected  Disconnect all pads from Fluid   There is an external or Replace Fluid Delivery Line     Delivery Line  Enter Manual internal leak in the  baw water lawor alr Mode  The system does not system of fluid line   continuously streaming remain in bypass mode for at This observation rules  from pads  The pump can least 3 minutes or    Alert 1  out leaks in the pad   be heard  Occurs   Run in Manual
23.  lines kinked     Reconnect the line onto the circulating pump  Push the lever  on the pump down to latch  Reinstall the back panel     Check the calibration according to Section 5 1 5 using a  Calibration Test Unit  Recalibrate if required        1 0 Manifold Connector Location  Figure 7 35    77    Parameter    APPENDIX A SPECIFICATIONS    Specification       Control Modes    Automatic  Manual  Purge  Stop       Heater Capability    750 W       Circulating Fluid    Distilled or Sterile Water       Reservoir Capacity    5 liters       Water flow rate  total     0 5   8 0 liter min       Patient Probe Type    YSI 400 Series compatible       Number of Patient Probes    2  1 monitor  controls  alarms   1 monitor   alarms        Patient Temperature Display Range    Resolution    10  C to 42  C  50  F to 107 6  F  in 0 1   C   F       Patient Temperature Measurement Accuracy     0 4  C  10  C to 32  C    0 2  C  32  C to 38  C    0 4  C  38  C to 42  C    includes  0 1  C external probe        Patient Temperature Control Range    Automatic Mode    33  C to 37  C  91 4  F to 98 6   F    in 0 1  C F increments       Water Temperature Display Range    3  C to 45  C  37 4  F to 113 0 F    in 1  C F increments       Water Temperature Control Range    Manual Mode    4  C to 42  C 39 2  F to 107 6   F    in 0 1   C F increments       Maximum Water Temperature   Automatic Mode     32  C to 42  C  89 6  F to 107 6   F    in 1  C F increments       Minimum Water Temperature   Automatic Mod
24.  storing    27    Table 5 1  Model 2000 Verification Data       Test    Criteria       Flow Rate    Factor H   300ml min 6  Flow  lt  Factor H    300ml min       Inlet water pressure     7 2 psi    P     6 8 psi       Patient Temperature 1 PT1  4 checks      0 3   C  lt  CK1  lt  0 3  C   0 3   C  lt  CK2  lt  0 3  C   0 1  C  lt  CK3  lt  0 1 C   0 3  C  lt  CK4  lt  0 3  C       Patient Temperature 2 PT2  4 checks      0 3  C  lt  CK1  lt  0 3  C   0 3  C  lt  CK2  lt  0 3  C   0 1  C  lt  CK3  lt  0 1 C   0 3  C  lt  CK4  lt  0 3  C       Patient Temperature Echo Out  4 checks      0 2  C  lt  CK1  lt  0 2  C   0 2  C  lt  CK2  lt  0 2  C   0 2  C  lt  CK3  lt  0 2  C   0 2  C    CK4  lt  0 2  C       Water Temperature T1  3 checks      0 4  C  lt  CK1  lt  0 4  C   0 4  C  lt  CK2  lt  0 4  C   0 2  C    CK3    0 2  C       Water Temperature T2  3 checks      0 4  C  lt  CK1  lt  0 4  C   0 4  C  lt  CK2  lt  0 4  C   0 2  C  lt  CK3  lt  0 2  C       Water Temperature T3  3 checks      0 4  C  lt  CK1    0 4 C   0 4  C    CK2    0 4 C   0 4  C  lt  CK3  lt  0 4  C       Water Temperature T4  3 checks      0 4  C  lt  CK1  lt  0 4  C   1 0  C  lt  CK2  lt  1 00   1 0  C  lt  CK3  lt  1 0  C       Miscellaneous  Heat  Cool  Flow    Heat  lt  120 seconds  Cool  lt  150 seconds  Factor H   300ml min  lt   Flow  lt  Factor H    300ml min       Secondary Alarm       Alarm 37 in    18 seconds       28       Test    Criteria       Primary Alarm    Alarm 34 in  lt  14 seconds    
25.  thermal breaker mounted on the heater casing  This breaker is  not used for normal temperature control  only to ensure that in the event of control failures the heater  does not over heat     CHAPTER3 INSTALLATION          Bw e  Fluid Reservoir Vent                              gt     ds Handle Release Knob    a Fluid Delivery Line Latch    y   WK       Fluid Delivery Line Connection  Cable Storage to Inlet Outlet Manifold  El e    Remote Display Cable US Panel Mounting Screw  On Off Switch for Mains Data Collection RS 232  Power  Reserved for Future  Remote Display  Connection Patient Temperature  1  Power Cord Mains Inlet Patient Temperature  2    Equipotential Stud Temperature Out    Power Cord Storage Fluid Delivery Line    Fill Line in Retainer Clip    Serial Number Label    Drain Ports  Panel Mounting Screws    Control Module Rear  Figure 3 1    13    3 1 Installation of Cables and Accessories  1  Unpack the Model 2000 and accessories     2  Adjust the Control Module handle to a vertical position and secure  by tightening the knobs on both sides     3  Connect the Remote Display Cable to the back of the Remote  Display  Install the Remote Display clamp onto the handle  Route  the cable through the three clips on the side of the handle   Connect the Remote Display Cable connector at the labeled  location on rear panel of the Control Module     4  Insert the Patient Temperature Cable into the labeled location on  the rear panel of the Control Module     5  Insert the opt
26. 0 seconds and  calibration error  internal temperature   continues  Audio turn power back on  If the error persists  replace  Restart system  calibration from the alarm half second the Isolation Card and retest  If the error persists    Patient Temperature   tone every 2 call Medivance Service  Calibrate upon  1 probe input is not minutes  Alarm Alert   completion   Alert functioning  indicator flashing    19 Patient Temperature   In Automatic Mode  Changes to Stop Turn the Control Module off  Wait 30 seconds and  calibration error  internal temperature   Mode  Audio alarm turn power back on  If the error persists  replace  Restart system  calibration from the half second tone the Isolation Card and retest  If the error persists    Patient Temperature   every 10 seconds  call Medivance Service  Calibrate upon  1 probe input is not Alarm Alert indicator   completion   Alarm functioning  flashing    20 Patient Temperature   In Manual Mode  Current Mode Turn the Control Module off  Wait 30 seconds and  calibration error  internal temperature   continues  Audio turn power back on  If the error persists  replace  Restart system  calibration from the alarm half second the Isolation Card and retest  If the error persists    Patient Temperature   tone every 2 call Medivance Service  Calibrate upon  1 probe input is not minutes  Alarm Alert   completion   Alert functioning  indicator flashing    21 Patient Temperature   Patient Temperature   Changes to Stop Operate the system in Manua
27. 1 0 CCA does not illuminate  during the purge     Electronics fault     Call Medivance Service        Water spills from the Main  Tank Vent in Manual or  Automatic Modes     Control Module has been  switched into service during a  case     Overfilled tank due to  pads being purged into a  full Control Module     Drain some water from the upper  drain port        Water jets from the Main    Tank Vent in Purge Mode     Control Module has been  Switched into service during a  case     Overfilled tank due to  pads being purged into a  full Control Module     Drain some water from the upper  drain port        Water jets from the Main  Tank Vent in Fill Mode     Cleaning Solution has not  been added to the distilled  water being used to fill the  Control Module     Low conductivity of the  fluid     Drain all water from the upper  drain port  Refill after adding  Cleaning Solution to fluid being  used to fill the Control Module        Reservoir overfills    Cleaning Solution has been  added to the fluid being used  to fill the Control Module     Level sensor or cable  fault     Check the sensor connection at  the I O Card  Replace the level  sensor harness        Cleaning Solution has been  added to the fluid being used  to fill the Control Module  Level  sensor cable replacement  unsuccessful     Electronics fault     Call Medivance Service        Control Module will not fill    At power up  the level is read  and indicates  full   The  Control Module will not even  attempt to fil
28. 62  7o Card Cage rh me a A an nn a lt ne nd    62  ZE  Power Circ  it Gard   x  nes arene Gani ee Aan Ati wee 64  Ls  Power MOJU                           64  7 8 Processor Circuit Card iii 65  7 9   Isolation Circuit  Gard  ccu ed t et etd tei deeds 65  7 10 Input   Output Circuit Card Connections    nnns 66  7 11 Input   Output Circuit         66  LL Software CHIPS o aeter E vente atm tuns uter ete ei  67  7 13  Sblaligl8  rdi iae 67  7 14  Circulating PUMP  erica 68  LADA Mixing  PUmp   uin icr ccce veinte aia 68  7416 Ghiller  Pumps  e eerte o et e om LM peers 69  PAT Oi en iiec 70  7 18  Chiller Power Relay    o1 greet p mati att aed 70  7 19  Chiller Control Relay          2 ne e a ge Hd tete c E de ii 71  7 20 Inlet   Outlet Manitold issues 71  7 21 Temperature Sensor      72  7 22  Level Sensor Harness erriari enit em heat n e ee Duden Medals 73  7 23 Dielectric Withstand  Hipot  Chassis Test    74  7 24 Dielectric Withstand  Hipot  Patient Applied Part Test    74    7 25 Inlet   Outlet Manifold Valve Harness iii 76    CHAPTER1 INTRODUCTION    1 1 Use of the Model 2000 Service Manual    The Arctic Sun Temperature Management System provides a safe and effective way of modifying  patient temperature and delivering precise temperature control in a novel way  This Service  Manual provides a detailed description of the Model 2000  its construction  routine maintenance  and calibration  The Model 2000 Operators Manual should be consulted for general operation  detail  It is
29. Bracket fasteners    59       Control Module  Left Side    Figure 7 3    Bracket fasteners    The electronics are housed in the Electronics Card Cage and control all machine processes   Five subassemblies comprise the electronics     Power Module    Power Inlet    Equipotential  Chassis Ground    Stud    Heater Fuse    The Power Module converts AC to DC power and controls the heater   The Power Card provides low voltage power for the electronics    The Processor Card houses the microprocessors and peripherals    The Isolation Card provides interfaces to the patient and external devices     The I O  Input Output  Card interfaces to the internal sensors and actuators     Power Card    150     44822 E     Processor Card    Isolation Card    500M 18  NOW Novi    DZ  118 900M    12 1          2  3  3    Input Output Card       Card Cage  Circuit Card Identification    Figure 7 4    60    7 1 Drain the Control Module  Tools and Supplies Required     Arctic Sun Drain Bag   1  Turn the Control Module off     Caution  Draining the system with power on may damage the  chiller     2  Attach the two fittings on the bag to the two drain ports on the  back of the control module  The device will passively drain all  tubing  reservoirs and pumps within the system  This is an       adequate drain for maintenance procedures  iiis Attached  igure 7 5    7 2 Front Grill    1  Remove the front grill by first pulling the bottom out by reaching  under the center of the bottom edge     2  Using both
30. C Error Code See Table 6 4   M Control Module Operating Mode 0   Initialization  12 Stop  2    Automatic  3   Manual  4   Purge   5   Fill   P Patient Temperature 1   C   OM Outlet Temperature Monitor  T1    C   O Outlet Temperature Control  T2    C     Inlet Temperature  T3    C   F Flow Rate ml min   P Inlet Pressure 0 to  10 psi   L Water Level 1   Empty  5   Full   P Circulating Pump PWM 0  Stop to 235   Full   C Mixing Pump PWM 0   Stop to 150   Full   H Heater PWM 0   Off to 32   Full Power   V Vent Valve State 1 Open  0   Closed   B Bypass Valve State 1 Open  0   Closed   Hr System Hours Hours             32       6 2 Card Cage Indicators and Test Points    LEDs on the edges of the cards indicate the status of major functions and may be helpful in  troubleshooting   Figures 6 4 to 6 7  Access to the power supply voltages is also available at the  card edge     Note  Use care in accessing these test points  An insulated probe is needed to avoid shorting the  test point to the frame  Doing so will cause circuit board fuses to blow  resulting in the need for  component level repair         NOT ACTIVE  VGG VCC 24VU GND 24VU 24VS 24VS  NOT ACTIVE   TEST TEST LED TEST TEST TEST LED  Power Card  Figure 6 4    1  oo oo  UI MON MON CTL CTL  CONN BIT WDOG WDOGBIT    Processor Card    Figure 6 5  RS 232  1 RS 232 PAT  PAT  ECHO   2 TEMP  TEMP    1  2    Isolation Card  Figure 6 6    O O O O O O O    FLOW GAS COOL FILL CHIL VV BV    1 0 Card  Figure 6 7    6 3 General Troubleshoot
31. Control Module off  Wait 30 seconds and  System error start up Mode  Audio alarm turn power back on  If the error persists  replace  Refer to Operator synchronization fault   half second tone the Processor Card  If unsuccessful  call  Manual  every 10 seconds  Medivance Service  Calibrate upon completion   Alarm Alert indicator  Alarm flashing   61 Non recoverable Control Processor Changes to Stop Turn the Control Module off  Wait 30 seconds and  system error parameter memory Mode  Audio alarm turn power back on  If the error persists  replace  Refer to Operator fault half second tone the Processor Card  If unsuccessful  call  Manual  every 10 seconds  Medivance Service  Calibrate upon completion   Alarm Alert indicator  Alarm flashing   62 Non recoverable Monitor Processor Changes to Stop Turn the Control Module off  Wait 30 seconds and  system error parameter memory Mode  Audio alarm turn power back on  If the error persists  replace  Refer to Operator fault half second tone the Processor Card  If unsuccessful  call  Manual  every 10 seconds  Medivance Service  Calibrate upon completion   Alarm Alert indicator  Alarm flashing   64 Non recoverable Unable to enable Changes to Stop Replace the Power Card and attempt to power up   system error pump power Mode  Audio alarm If the error persists  call Medivance Service   Refer to Operator  Control  half second tone  Manual  every 10 seconds   Alarm Alert indicator  Alarm flashing   65 Non recoverable Unable to enable Changes to S
32. Earth Ground     1 5 Environmental Conditions  The Model 2000 should be stored and used in specific operating conditions     Ambient Temperature Range     Operating  6      10  C to 27  C  50  F to 80  F   Note  At higher room operating temperatures  the units cooling capacity is compromised     Storage 1115       30  C to 50  C   20  F to 120  F     Note  If the device is stored in a location where the temperature may fall below the  freezing point  the unit should be completely drained of water to prevent damage  See  Section 4 6     Ambient Humidity Range     Operating  atii ere Ted ipa dade 50  to 70  relative humidity  non condensing  Storag  C  5  to 95  relative humidity  non condensing    1 6 General Warnings      Do not use the Arctic Sun in the presence of flammable agents because an explosion and or  fire may result       There is a risk of electrical shock and hazardous moving parts inside the machine  Care  should be exercised by service personnel and attention should be particularly paid to areas  highlighted with warning symbols       North America only  Power cord has a hospital grade plug  Grounding reliability can only be  achieved when connected to an equivalent receptacle marked  hospital use  or  hospital  grade    4    1 7 Cautions    See Operator s Manual for complete list of Cautions prior to using the Arctic Sun Temperature  Management System to treat a patient  This is not an exhaustive list and is not intended to replace  the Operator   s Manual    
33. alarm conditions     Press the S4 switch on the CTU  Monitor the User Interface  Water temperature and start the timer once the temperature  reaches 42 6   Verify that alarm 34 sounds in less than 14  seconds  Release the switch immediately     Note  If the water temperature drops below 42 6  C  restart the  timer when the water again reaches 42 6 C     Note  If the water temperature exceeds 44 1  C for more than  1 second Alarm 33 will occur  and the test must be repeated    Note  If S4 is held too long or is pressed when no water is  flowing  a safety circuit will turn off power to the heater  The  CTU must be allowed to cool completely if this should occur     5 2 5 Stage 7  Chiller Capacity Test    1   2     Press the Manual Mode Key     Press the Down Key once and change the  Water Target  Temperature  to 6  C     Press the Home Key     Allow the water temperature to stabilize at 6  C for at least 5  minutes     Press and hold S4 for 10 minutes  Note the temperature of  the water  Release the switch     Note  If S4 is held too long or is pressed when no water is  flowing  a safety circuit will turn off power to the heater  The  CTU must be allowed to cool completely if this should occur     Once verification and calibration is complete  press the   Purge  key to remove the water from the CTU     Turn off the Arctic Sun Model 2000  and remove the CTU   Disconnect associated cabling from the Model 2000 and  reconnect operator cabling     Drain the water from the CTU before
34. ance Kit     Caution  Use only sterile or distilled water  The use of other fluids  will damage the system     From Stop Mode  press the down arrow until the fill screen  appears   Water Level Empty  Enter to Fill   Press Enter to  proceed     Note  Pads cannot be attached to the Fluid Delivery Line during  the fill process     Follow directions on the display screen which read     Place the Fill  Tube in the container of water  Press Enter to fill      Press Enter again and the Model 2000 will automatically fill until  the reservoir is full      Blank Page     CHAPTER4 MAINTENANCE    4 1 Recommended Maintenance Schedule                         Procedure Interval   Clean External Surfaces As required   Adjust Remote Display Clamp As required   Inspect connectors and cables 3 months   Clean the Condenser 3 months   Replace Fluid 3 months   Calibration and Operational Check Every 2000 hours or 250 uses  which ever is   less  as indicated by system display                4 2 Clean External Surfaces    Cleaning should include the exterior of the Control Module and the power cord  Use a soft cloth and  mild detergent or disinfectant according to hospital protocol  The display screen can be wiped with a  clean  lint free  moist cloth  Fluid Delivery Lines and Temperature Cables should be cleansed with a  hospital approved disinfectant     4 3 Adjust Remote Display Clamp  Tools and Supplies Required       1 2    hex key     The Remote Display clamp tension may need adjusting over ti
35. and 42  C  39 2 F and 107 6 F  through the pads  This  results in heat exchange between the water and the patient  Unlike conventional water based  Systems  this system operates under negative pressure relative to ambient  Any break or leak in  the fluid path will result in air entering the circuit instead of water leaking out     The system functions in either an Automatic Mode or Manual Mode  In Automatic Mode  a patient  temperature probe  connected to the Control Module  provides feedback to an internal control  algorithm  The temperature of the water increases or decreases automatically to achieve a pre   Set patient target temperature determined by the clinician  In Manual Mode  the feedback control  algorithm is not activated  The operator adjusts the temperature of the water delivered to the pads  directly and is responsible for monitoring patient temperature     Model No          Arctic Sun Temperature Management System Model 2000    Phone               Toll free in the U S  877 267 2314 or 303 926 1917  FAX dun 303 926 1924  Address            321 South Taylor Avenue  Suite 200    Louisville  CO USA 80027  E mail               customerservice medivance com    Internet  www medivance com    European Representative   Bard Limited  Forest House  Brighton Road    Crawley  West Sussex UK  RH 11 9BP  0050    1 4 Symbols and Standards    The Model 2000 bears the following symbols affixed to the system      Ai    C    an    Certified    v  A    STERILE    DISTILLED       Ta
36. ar     n   amp   o    J10  Thermistors  402479 00    J8  Level Sensors    402476 00    J5   Chiller   Hot  Gas Bypass  Control  402475 00    Input Output Circuit Card Connections    Figure 7 16    Input   Output Circuit Card  Tools and Supplies Required     1 16    hex key    Observe electrostatic discharge control procedures when    handling circuit cards   1  Remove the rear panel per Section 7 3     2  Carefully disconnect all connectors from the Input Output circuit  card  releasing each locking tab before pulling    3  Remove the Input Output circuit card by removing the allen  head screw with a 1 16    hex key and carefully prying outward at  the top with a flat head screw driver  Carefully pull outward until  the circuit card is removed     4  Ifthe Input Output circuit card is replaced  calibrate per Chapter    5     66    J2  Flowmeter  402470 00    J4  Mixing Purp  402471 00    J6  Flow Pump    402471 00    J7  Card Cage Fan  402481 00    J9  1 0 Harness  402391 00       Input Output Circuit Card  Figure 7 17    7 12 Software Chips  Tools required     PLCC extractor    Note  Observe electrostatic discharge control procedures when  handling circuit cards     1  Remove the rear panel per Section 7 3   2  Remove the Processor circuit card per Section 7 8     Note  The software upgrade is performed by replacing the firmware  chips Handle the chips as little as possible to avoid static damage     Control Software Chips  1  Using the extractor tool remove the chips from U
37. ble 1 1  Model 2000 Symbols    For the safe and effective use of this device  the operator must consult the  accompanying documents prior to use     This symbol adjacent to the patient connections means that the thermal probe  connection is a  Defibrillator Proof  Type BF Applied Part   per standard IEC  60601 1 and affords the degree of patient protection defined in that standard for  this type of applied part     Models of the Arctic Sun that bear the Entela US C Monogram have been  Certified for Safety by Entela  Inc  against standards C22 2  No  601 1 and UL  2601     Identifies the equipotential terminal on the equipment  which is intended to be  connected with the equipotential terminal s  on one or more other types of  equipment in close proximity  in order to bring all pieces of equipment to the  same potential for safety purposes     Indicates high temperature part or component     Indicates that only sterile or distilled water should be used when filling the Arctic  Sun Control Module     Identifies Patient Temperature 1  the primary patient temperature probe input     Identifies Patient Temperature 2  the secondary patient temperature probe  input     Identifies the drain port     Identifies the heater fuse     Table 1 1  Model 2000 Symbols   continued     50  C     Identifies the storage temperature range   30  C    2 Ss Identifies the storage relative humidity range     Indicates electrical hazard     Indicates sharp object or a moving part             Indicates 
38. ck of pump PWM   to generate full negative  pressure   e Circulating Pump PWM is  greater than 200  indicating pump is  running faster than  normal to maintain flow  and pressure   No leak is found  Operate the The pump head may Replace the Circulating Pump   system as above  Circulating be ineffective due to  Pump PWM is high  debris stuck in the  pump   CTU connected  No leak is found  Operate the   Flowmeter fault  Replace the Heater   Flowmeter  OW E system as above  No flow is assembly   e indicated in User Diagnostic  pressure  or high PWM  Mode although some water  The pump can be heard  appears to be flowing   CTU connected  Check the Circulating Pump Pump connection fault  Reseat the connector  Allow the               connection at the I O Card  card to dry  if necessary   e Verify that the circuit card is  pressure  or high PWM  not wet                 35       Table 6  3  General Troubleshooting Guide       Conditions   Symptom    Observation    Cause    Solution       The pump cannot be heard     Swap the Circulating Pump  and Mixing pump connections  on the I O Card  Attempt to  operate the system  Mixing  pump does not run     Pump electronics  failure     Replace the Circulating Pump        Pressure is more negative  than  8PSI for more than a  minute     Pressure sensor or  cable fault    Replace the I O Manifold  Assembly       Pads connected     Poor temperature control of  patient    Low water flow or air leaving  pads     Air leak in pads or  lines     See
39. e     4  C to 32  C  39 2  F to 89 6   F    in 1  C F increments       Mains Input     115VAC  60 Hz  11 A  nominal   230VAC  50 Hz  5 5 A  nominal        Current Leakage     lt  300uA       Circuit Breaker       12 Amp     Mains supply voltage fluctuations are not to exceed  10  of nominal supply voltage                                Parameter Specification   Relative humidity range  operating  5 70    Relative humidity range  storage  5 95    Temperature range  operating  10  C to 27  C  50  F to 80  F  Temperature range  storage   30  C to 50  C   20  F to 120  F  Height of unit  handle down  66 0cm  26    Length of unit 48 3cm  19    Width of unit 31 8cm  12 5    Weight of unit filled 52 7kg  116 lbs       79    APPENDIX B SHIPPING    Due to the size and weight of the Model 2000 Control Module  it  should be shipped on a pallet using Medivance provided packaging  materials  If the original packaging is not available  a shipping kit  may be ordered from Medivance     1  Drain the control module per Section 4 7  Note the additional  steps to completely drain the unit in order to prevent damage  from freezing     2  Place the Control Module on the foam support which is mounted  to the pallet  Figure B 1      3  Cover the unit with the plastic bag  Figure B 2      4  Place the cardboard tube over the top of the unit so the edges  are outside the wood strips attached to the pallet     5  Place the molded foam insert on top on the unit    6  Place the cardboard cap on top of t
40. e  Calibrate if necessary  If the error  water temperature  control water within half second tone persists  call Medivance Service    1 0  C F of set point every 10 seconds    after 25 minutes in Alarm Alert indicator  Alarm current mode or at flashing    the time of last   change in Water   Target Temperature    41 Low bypass flow  Insufficient internal Current Mode Use the CTU to check the calibration of the Control  Heating not flow during priming Continues  Audio Module  Calibrate if necessary  If the error  available  or bypass mode  alarm half second persists  call Medivance Service    tone every 2  minutes  Alarm Alert  Alert indicator flashing   43 User settings not Saved customer Current Mode Turn the Control Module off  Wait 30 seconds and  saved  System user profile settings Continues  Audio turn power back on  If the error persists  replace  default settings are invalid  alarm half second the Processor Card  If unsuccessful  call  restored  tone every 2 Medivance Service  Calibrate upon completion    minutes  Alarm Alert  indicator flashing  Alert   44 Invalid system log In Diagnostic Mode    Current Mode Turn the Control Module off  Wait 30 seconds and  entry  Service one or more entries Continues  Audio turn power back on  If the error persists  replace  information may be in the internal alarm half second the Processor Card  If unsuccessful  call  corrupt  system event log are   tone every 2 Medivance Service  Calibrate upon completion    invalid  Used by a min
41. e circulating water  In addition  draining is required  for replacement of some components or prior to shipment or  storage     1  Power off the system     Caution  Draining the system with power on may damage the  chiller     2  Attach the two fittings on the bag to the two drain ports on the  back of the control module  The device will passively drain all  tubing  reservoirs and pumps within the system  This is an  adequate drain for maintenance procedures     3  Proceed with a complete drain for shipping or storage when  freezing is a risk     3 1 Perform the partial drain procedure in step 2     3 2 Allow the control module to stand for 15 minutes to allow  the chiller to return to room temperature     3 3 Power on the system  The  Level Empty  alarm should  occur and the chiller should not run  Cancel the alarm  If  the chiler runs  do not continue and power down  immediately  Drain the system again as in step 2 and  power up again     3 4 From the main screen  press the Down key three times to  enter the Fill Mode  Press enter  Do not place the fill tube  in water     3 5 With the Drain Bag still attached  press enter to start a fill  cycle which will empty the Circulating Pump     3 6 Press and hold the Down key for a minimum of 20  seconds to empty the internal Mixing Pump     3 7 Press the Stop key then power off the system  Remove  the Drain Bag and empty it     4  Fill Control Module per Section 3 3 prior to returning to use     20    A  f    Drain Bag Attached  Fi
42. e the rear panel per Section 7 3  2  Remove right side panel Section 7 4     3  Through the right side of the frame  disconnect the wire which  connects to the rear of the card cage        4  Disconnect the  heater  cable from under the card cage     Chiller Power Wire Harness  Figure 7 9       Heater Wire Harness  Figure 7 10    62    Disconnect all cables on the Input Output circuit card on the  right side of the card cage per Section 7 10     Remove the two screws on the right side of the chassis  left of  the card cage  with a 5 64    hex key     Remove the two nuts in the back of the Input Output circuit card  on the right side  on the top and bottom with a 5 16    nut driver     Lift out the card cage     After re assembly of the unit  it is recommended to perform a  Dielectric Withstand  Hipot  Chassis Test per Section 7 23     63       Card Cage Mounting Screws  Figure 7 11       Card Cage Mounting Nuts  Figure 7 12    7 6 Power Circuit Card  Tools and Supplies Required     1 16    hex key    Note  The machine does not have to be drained in order to  complete this procedure     Note  Observe electrostatic discharge control procedures when  handling circuit cards     Remove the rear panel per Section 7 3     2  Remove the Power Circuit Card  second from left  by removing  the bottom allen head screw with a 1 16  allen wrench and  carefully prying the bracket at the top with a flat head screw  driver        3  Carefully slide the circuit card out of the card cage until 
43. ee Figure 7 34   If  necessary  place a blade screwdriver into the slot at the top    Label  of the stem to hold it secure while loosening the nuts  ae    Remove the solenoid coils from the stems  Note the location  of each and the associated label on the wire  The 3 wires  are labeled FV  BV  and PV  The new solenoid coils will  have to be returned to their proper location        Removing Solenoid Nuts  Figure 7 33    With a 7 16    wrench  loosen the temperature probe and  remove it from the manifold block     Unplug the manifold harness from the Input Output Circuit  Card by first compressing the tab on the top of the plug  Do      Valve  not stress the wires  It may be necessary to unplug the Stem  three closest connectors to gain access  The tabs on these Vake    connectors must be compressed in the same manner to Washer       release them     Replace the harness  Install the new temperature probe into  the manifold replacing the orange rubber washer  Tighten  the probe until finger tight then tighten Ye turn        Replace valve washers then each solenoid assuring the  proper location of each  see Figure 7 33 above   Tighten the Removing Temperature Probe  nuts onto the valve stems assuring the solenoid covers are Figure 7 34  properly aligned with the valve washers     Reconnect the harness to the I O board with the connector  tab oriented down  Assure the connector is fully engaged     Replace the manifold in the frame  assuring the wires do not  become pinched or the
44. eeenen nnne rca 22  5 1 The Calibration and Test UNI ss 22  5 1 1 Basic CTU Operation  e dite tisane alerte etx ee dua aee de eene eus 22  A Un EE 22  5 1 3 Electrical une CET 22  5 1 4 ET wu Ile 23  5 1 5     Calibration  edel Ee 23  5 1 6 Environmental Requirements                      essen nennen eene 24  5 2 The Test and Calibration Procedure                    ssssssssssseseeeeneeneeeenn nnne 24  5 2 1 Stage 1  2  and 3  Warm up  Flow Check  Patient Temperature Calibration  Water  Temperature Calibration  and Interim Results Heporting   25  5 2 2 Stage 4  Temperature Out Calibration                                        nennrennrenneen nsen nnn 26  5 2 3   Stage 5  ACCeptarice eec te ein dite Dee ig el deca 26    TABLE OF CONTENTS    REF 000742 01 Rev K    5 2 4 Stage 6  Water Temperature Alarm Test  26    5 2 5 Stage 7  Chiller Capacity Test  27  CHAPTER 6 TROUBLESHOOTING  nn nnnrrrsrsrsnneennnnneenennenneneennenenneneeneeneneneennenennenennes   ge le ele EE EE 20   6 1 1 Diagnostic Home Screens eene nnn enne 31   6 1 2 Error Log and Operational Log ss 31  6 2 Card Cage Indicators and Test Points                     ssssssssssssseeee enne 33  6 3 General Troubleshooting Guide eene 34  6 4 Troubleshooting Alarms and Alerts AA 42   CHAPTER 7 COMPONENT REPLACEMENT           s cccssecssseecesseeeeseeesesesesseeseseeeeeseeesseeseseanenseeeeeees   7 1 Drain the Control Module    is 61  4 2  Front  EU 61  1 93  Rear Patel  ea tuetur eet veu 61  LI    SIMS  Panels  esse      
45. er Module           Control Module is noisy    The noise is present at all  times and emanates from the  bottom of the card cage     A cooling fan may be  failing     Replace the Card Cage   Circuit  cards need not be replaced            The noise is present at all  times and emanates from the  lower frame chiller area        The chiller fan may be  failing        Call Medivance Service        40       Table 6  3  General Troubleshooting Guide       Conditions   Symptom    Observation    Cause    Solution       The noise occurs only when  water is flowing     The Circulating Pump  may have air trapped  in the pump head     Debris may be cleared by running  the pump a full speed  Connect  two shunt tubes to the fluid  delivery line manifold  Enter  Manual Mode  The flow rate  should exceed 4 Ipm for 5  minutes  If the pump is still noisy   call Medivance Service        The noise occurs only when  water is flowing  High flow  flush is unsuccessful     The Circulating Pump  may be failing     Replace Circulating Pump        The noise occurs only during  a control mode when the  temperature is being  regulated at the target  The  noise seems to burst on and  off     The Mixing Pump may  be failing     Replace the Mixing Pump        Control Module will not  power up     When turned on  the Control  Module beeps  possibly  flashes the displays  and goes  silent  beeps again  and  continues     Electronics fault    Call Medivance Service        When turned on  a fan can be  heard  b
46. fault  Test the input with a  cable and probe proven good on another Control  Alarm Module  If the Control Module proves inaccurate   replace the Isolation Card and retest  If the error  persists  call Medivance Service  Calibrate upon  completion   29 Patient Temperature   In Automatic Mode  Current Mode Turn the Control Module off  Wait 30 seconds and  2 calibration error  internal temperature   continues  Audio turn power back on  If the error persists  replace  Restart system  calibration from the alarm half second the Isolation Card and retest  If the error persists   Patient Temperature   tone every 2 call Medivance Service  Calibrate upon  2 probe input is not minutes  Alarm Alert   completion   Alert functioning  indicator flashing   30 Patient Temperature   In Manual Mode  Current Mode Turn the Control Module off  Wait 30 seconds and  2 calibration error internal temperature   continues  Audio turn power back on  If the error persists  replace  Restart system  calibration from the alarm half second the Isolation Card and retest  If the error persists   Patient Temperature   tone every 2 call Medivance Service  Calibrate upon  2 probe input is not minutes  Alarm Alert   completion   Alert functioning  indicator flashing   31 Patient Temperature   In Automatic Mode  Current Mode Turn the Control Module off  Wait 30 seconds and  2 calibration error  internal temperature   continues  Audio turn power back on  If the error persists  replace  Restart system  calibration fr
47. fy  system error out of range   high Mode  Audio alarm that it is plugged in securely  the wires emanating  Refer to Operator resistance half second tone from it are intact  and that the other end of the wires  Manual  every 10 seconds  entering the sensor body is intact  Power up and  Alarm Alert indicator   allow the Control Module to run for 10 minutes  If  Alarm flashing  the error persists  call Medivance Service   72 Non recoverable 12 Thermistor signal   Changes to Stop Ensure that the I O Card or J10 connector on it are  system error out of range   low Mode  Audio alarm not wet  Allow them to dry  if necessary  and retest   Refer to Operator resistance half second tone If the error persists  call Medivance Service   Manual  every 10 seconds   Alarm Alert indicator  Alarm flashing   73 Non recoverable T1 Thermistor signal   Changes to Stop Check the connector at J10 of the I O Card  Verify  system error out of range   high Mode  Audio alarm that it is plugged in securely  the wires emanating  Refer to Operator resistance half second tone from it are intact  and that the other end of the wires  Manual  every 10 seconds  entering the sensor body is intact  Power up and  Alarm Alert indicator   allow the Control Module to run for 10 minutes  If  Alarm flashing  the error persists  call Medivance Service   74 Non recoverable T1 Thermistor signal   Changes to Stop Ensure that the I O Card or J10 connector on it are  system error out of range   low Mode  Audio alarm not wet 
48. g circuit and the chiller  circuit  see Figure 2 1   The circulating circuit controls the flow of water to the pads  The chiller  circuit provides a constant source of cold water that is pumped  as needed  into the circulating  circuit     Water in the circulating circuit is pulled from the Circulating Tank through the Inlet Outlet Manifold  and out to the Energy Transfer Pads  Water then returns from the pads through the Inlet Outlet  Manifold to the negative pressure side of the positive displacement Circulating Pump  Water  leaves the Circulating Pump under positive pressure then passes through a turbine flow meter  and heater before returning to the Circulating Tank  Water in the Main Tank replenishes the  circulating circuit     The chiller circuit operates independent of the circulating circuit  Water in the Chiller Tank drains  into the centrifugal Chiller Pump where it then is pumped under positive pressure through the  Chillers    evaporator  The evaporator is a heat exchanger with cold refrigerant on one side and  water on the other  Water then returns to the Chiller Tank to complete the cycle  The system  controls the temperature of the water in the Chiller Tank to 4  C by activating a valve on the  refrigerant side when additional cooling is no longer needed     When the system determines water must be warmed  the heater is activated in the circulating  circuit  When the system determines water must be cooled  water is drawn out of the Circulating  Tank by the pos
49. ges to Stop Use the CTU to verify the accuracy of the Control       2 probe  disconnected or out  of range     Alarm       2 reading is outside  the lower limits of the  display range in  Automatic Mode        Mode  Audio alarm  half second tone  every 10 seconds   Alarm Alert indicator  flashing        Module  Check connection of patient temperature  probe and cable  If wet  replace cable  Check  temperature cable connection at the rear of the  Control Module  Flex the cable to ensure that there  is not an intermittent fault  Test the input with a  cable and probe proven good on another Control  Module  If the Control Module proves inaccurate   replace the Isolation Card and retest  If the error  persists  call Medivance Service  Calibrate upon  completion        47          Table 6 4    Alarm and Alert Troubleshooting                               Alarm    Message Displayed   Alarm Alert System Status Troubleshooting  Alert Description  28 Temperature probe Patient Temperature   Changes to Stop Use the CTU to verify the accuracy of the Control  2 malfunction  2 reading is outside Mode  Audio alarm Module  Check connection of patient temperature  Replace the upper limits of half second tone probe and cable  If wet  replace cable  Check  temperature probe the display range in every 10 seconds  temperature cable connection at the rear of the  or cable  Automatic Mode  Alarm Alert indicator   Control Module  Flex the cable to ensure that there  flashing  is not an intermittent 
50. gure 4 6    4 7 Calibration and Safety Check  See Chapter 5 for the complete procedure for calibration and safety check     21    CHAPTER 5 CALIBRATION AND SAFETY CHECK    5 1 The Calibration and Test Unit    The Model 2000 Calibration and Test Unit  CTU  is an instrument that is used to calibrate the Model  2000 Control Module  It is a hand sized unit that connects directly to the fluid delivery line port  taking  the place of the Model 2000 fluid delivery lines and pads     5 1 1 Basic CTU Operation    The CTU is used as an external stimulus and measuring device by software embedded in the Control  Module  The CTU contains precision resistors  a sensor for measuring water temperature  and water  heaters that are calibrated at the factory  The factory calibration data is listed on the CTU and is  entered into the Control Module during the calibration process  The resistors simulate several patient  temperatures  After a warm up period  the simulated temperatures are first applied to the patient  temperature inputs to temporarily calibrate each channel  This calibration is temporary in the sense  that it is used to perform the balance of the testing  It is not saved by the Control Module until the  operator accepts it  Once the patient temperature channel s accuracy has been established  the  Temperature Out circuit and the CTU   s temperature sensor are connected to the Control Module   s  Patient Temperature inputs  in turn  The Control Module then circulates water through 
51. hat there  indicator flashing is not an intermittent fault  Test the input with a  Alert cable and probe proven good on another Control  Module  If the Control Module proves inaccurate   replace the Isolation Card and retest  If the error  persists  call Medivance Service  Recalibrate upon  completion   52 Extended period of Circulating water Remains in current Assess patient s skin  Check to determine if water  cold water delivery  temperature has mode  Audio circulating through the pads is at least 2 3 liters per  Assess Patient been below alarms  minute  Determine whether the temperature probe  10  C 50  F for 8 of is placed properly  Check to make sure the  the previous 10 recommended number of pads is in use  Determine  Alert hours  whether the patient requires additional sedation for  shivering  Assess whether environmental factors   such as room temperature and hot lamps are  interfering with the cooling process   53 Prolonged cold The Arctic Sun has Changes to Stop Assess patient s skin  Assess whether the minimum  exposure issued 11 alerts for Mode water temperature should be increased or whether  extended periods of the temperature management therapy should be  cold water delivery  discontinued   Alarm                   51          Table 6 4    Alarm and Alert Troubleshooting                                     Alarm    Message Displayed   Alarm Alert System Status Troubleshooting  Alert Description  60 Non recoverable Control and Monitor   Changes to Stop Turn the 
52. he tube     7  Band the cardboard tube and cap securely with at least two  bands evenly spaced  Figure B 3 and B 4      80       Foam on Pallet    Figure B 1          Bagged Unit on Pallet  Figure B 2           Packaged and Banded to Pallet  Figure B 3    Packaged and Banded to Pallet  Figure B 4    
53. hed to the  chiller pump  which extends to the back of the pump  in the  lower part of the unit      Loosen the hose clamp ring with a small screwdriver    Pull the pump assembly off the tubing    To replace  insulate the pump head and the tubing with the  adhesive coated foam insulation provided     After re assembly of the unit  it is recommended to perform a  Dielectric Withstand  Hipot  Chassis Test per Section 7 23        69    7 17 Heater    OV d qx qe            10   11     12     Tools and Supplies Required     Blade screwdriver    5 32    hex key  Drain the control unit per Section 7 1     Remove front grill per Section 7 2        Remove rear panel per Section 7 3  Snap Ring Removal    i Figure 7 24  Remove the side panels per Section 7 4    Disconnect the cable that comes from the heater and goes to the  cable connection directly behind the circulating pump     Disconnect the flowmeter cable that comes from the heater  assembly and plugs into the Input   Output circuit card  see  Section 7 10      Note  Observe the routing of the cable to the Input Output circuit  card     Disconnect the tubing from the upper drain port  Push the  insulation back where the tubing meets the elbow fitting  Open  the snap ring with a small screwdriver and pull the tubing from  the fitting     Disconnect the tube from the bottom of the mixing reservoir  The  tube is located closest to the bottom of the outside of the  chassis  Pull back the insulation where the tube meets the  reservoir
54. ic cleaning of the exterior surfaces  none is  required  To clean the CTU  spray a mild detergent solution onto a towel and wipe down the unit  Do  not submerge the unit in water     If heater becomes inoperative  return the CTU to Medivance for service  An open fuse is indicative of  a problem that requires immediate service     5 1 5 Calibration    The CTU should be returned to the manufacturer for calibration every two years  The date of last  calibration is indicated by the international symbol for calibration  an upside down solid black triangle     v which can be found on the top of the CTU     CTU S N 7001    A  10 88    25 55     34 54     39 99     0 02      0 07      0 14   i  2850  Calibration Date W 2007 06    Calibration Data       CTU Calibration Data  Figure 5 1    23    5 1 6 Environmental Requirements  The Calibration Test Unit should be stored and used in specific operating conditions     Ambient Temperature Range     Operating              18  C to 24  C  65  F to 75  F   Storage          30  C to 50  C   20  F to 120  F     Ambient Humidity Range   Operating      5  to 95  non condensing    Note  If the CTU is used at a room temperature outside of the specified operating  range  the test system   s accuracy will be degraded     5 2 The Test and Calibration Procedure    The entire test and calibration process takes approximately 75 minutes  Operator assistance is  required only in certain stages       Stage 1  Warm up  Flow Check  and Patient Temperature Ca
55. indicates any rise in  the water temperature T2   Water does not warm at all     Heater fuse is open   caused by fault in  heater circuit     Replace fuse and Heater    Flowmeter assembly           CTU connected     Water does not heat to  target temperature in  Manual Mode        Heater fuse is intact  Power  up in User Diagnostic Mode   Enter Manual Mode with a  water target of 42  C  After  stabilizing for several minutes   note whether the display  indicates any rise in the water  temperature T2  Water does  not warm at all        Power Module fault        Replace the Power Module        36       Table 6  3  General Troubleshooting Guide       Conditions   Symptom    Observation    Cause    Solution       Heater fuse is intact  Power  Module replacement was  unsuccessful  Power up in  User Diagnostic Mode  Enter  Manual Mode with a water  target of 42 C  After  stabilizing for several minutes   note whether the display  indicates any rise in the water  temperature T2  Water does  not warm at all     Heater thermal breaker  tripped  caused by fault  in heater control circuit     Call Medivance Service        Pads connected     Water cools but is reported    to be unable to cool    effectively when used with    a patient     Remove plastic grille and  examine the condenser     Chiller condenser  clogged with lint     Clean condenser        CTU connected     Water does not cool to  target temperature in  Manual Mode     Power up in User Diagnostic  Mode  Operate the system i
56. ing   04 Water reservoir level   Reservoir level is Remains in Stop Fill Reservoir  If the error persists  see General  below minimum  Fill empty or below the Mode  Troubleshooting   Will Not Fill   reservoir  minimum to run the     Audio alarm half  system   Alarm second tone every  10 seconds  Alarm  indicator flashing    05 Water reservoir level   After pads are Changes to Stop If connected to pads  purge pads  Fill Reservoir  If  below minimum  Fill purged  reservoir is Mode  Audio alarm the error persists  see General Troubleshooting    reservoir or purge empty or below half second tone Will Not Fill   pads minimum required to   every 10 seconds    Alarm run the system  Alarm indicator  flashing   07 Purge not complete    A significant amount   Remains in Stop See General Troubleshooting   Fails To Purge       Pads may contain  extra water     Alert       of water was still  being returned from  the pads at the end  of the purge cycle        Mode  at end of  purge   Audio alarm  half second tone  every 2 minutes   Alarm indicator  flashing           43          Table 6 4    Alarm and Alert Troubleshooting                         Alarm    Message Displayed   Alarm Alert System Status Troubleshooting  Alert Description  08 System warming and   Patient Temperature   Changes to Stop Operate the system in Manual Mode with a setpoint  Patient Temperature   1 reading is above Mode  Audio alarm below 38  C until the water temperature is less than  1 high  38   C  100 4  F  and  
57. ing Guide    The following is a list of potential issues and methods of resolution  The suggestions for each issue  are listed from most to least probable  Therefore  each step assumes the previous steps have been  checked  For component replacement instructions  see Chapter 7     Note that this guide assumes that no special equipment  such as the Medivance Calibration Test Unit  is available  If available  the Calibration Test Unit is a powerful diagnostic tool and it should be used  in association with the instruction in its manual     Table 6  3  General Troubleshooting Guide       Conditions   Symptom    Observation    Cause    Solution       An alarm or alert not solved  by recommendations in the  Operator s    See Section 6 4  Troubleshooting  Alarms and Alerts          Pads connected     Low water flow or air  continuously streaming  from pads  The pump can  be heard     The pads have been filled   then purged and then refilled     Trapped air bubbles  may restrict flow  temporarily     Allow the flow to stabilize for over  20 minutes before rechecking the  flow rate  Air can be removed by  lightly tapping on the pads        The pad connectors are not  fully seated in the fluid  delivery line manifolds     Air leak in fluid delivery  line manifold  connection     Reseat connectors        Pad lines are kinked or  occluded at either the  connector and pad ends     Restriction in fluid lines    Remove kinks  Recommend use  of fluid delivery line straps to keep  lines from
58. ional Patient Temperature  2 Cable into the labeled  location on the rear panel of the Control Module     6  Insert the optional Patient Temperature Out Cable into the labeled  location on the rear panel of the Control Module        7  Using the clips and wing nuts from the Strain Relief Kit  anchor the Remote Display Cable and Clamp Rear  cables to the back panel  View    Figure 3 2   8  Move the latch of the fluid In Out Manifold to the left  Connect the  Fluid Delivery Line to the Inlet Outlet Manifold  Move the latch on   the manifold to the right to lock it in place     9  Gently push the fitting of the fill line into the fill port  Secure the line  to the rear panel as shown in the picture        Strain Relief  Figure 3 3       Fill Line  Figure 3 4    14    3 2 Power Up System    1     3 3    Insert the power cord into the power inlet  Using the clips and wing  nuts from the Strain Relief Kit  anchor the cord to the back panel   Plug the power cord into the wall outlet  Locate the switch on the  rear panel and turn the power on     After the software initializes an alarm will sound and the display  will read   Alarm 5  Water level below minimum  Fill reservoir or  purge pads      Clear the alarm by pressing the alarm key     The display reads   Medivance Arctic Sun Temperature  Management System      Filling the Control Module    Add one vial of Arctic Sun Cleaning Solution to 4 liters of sterile or  distilled water  The solution is contained in the Arctic Sun  Mainten
59. ists  replace  calibration data in alarm half second the Processor Card  If unsuccessful  call  non volatile memory   tone every 2 Medivance Service  Calibrate upon completion   Alert invalid  minutes  Alarm Alert  indicator flashing   50 Patient Temperature   Patient temperature   Current Mode Use the CTU to verify the accuracy of the Control  1 erratic  Check discontinuity  Continues  Audio Module  Check connection of patient temperature  temperature probe  Significant change in   alarm half second probe and cable  If wet  replace cable  Check  patient temperature tone every 2 temperature cable connection at the rear of the  reading for more minutes  Alarm Alert   Control Module  Flex the cable to ensure that there  Alert than 8 minutes  indicator flashing is not an intermittent fault  Test the input with a  cable and probe proven good on another Control  Module  If the Control Module proves inaccurate   replace the Isolation Card and retest  If the error  persists  call Medivance Service  Recalibrate upon  completion   51 Patient Temperature   Patient temperature Current Mode Use the CTU to verify the accuracy of the Control  1 below control less than 33  C when   Continues  Audio Module  Check connection of patient temperature  range  Check Automatic mode alarm half second probe and cable  If wet  replace cable  Check  temperature probe  entered  tone every 2 temperature cable connection at the rear of the  minutes  Alarm Alert   Control Module  Flex the cable to ensure t
60. itive displacement Mixing Pump and metered into the Chiller Tank  This causes  water to overflow the Chiller Tank and refill the Circulating Tank with cold water where it mixes   cooling the circulating water     The flow of water to the pads is controlled by regulating the Circulating Pump speed using  pressure sensor  P1   The pressure is regulated to  7 psi  This assures that the flow through  each pad is maintained regardless of how many pads are connected  Since water in the pads  and lines is at a negative pressure relative to ambient  any break causes air to enter the circuit  instead of water leaking out where it may contaminate the patient  This valuable feature can  make troubleshooting leaks more difficult     The temperature of the water going out to the pads is controlled using a temperature sensor  T2   and monitored by a temperature sensor  T1   Both sensors are located in the Circulating Tank   Temperature sensor  T3  measures the temperature of water returning from the pads  The  difference between outlet and inlet water temperature along with the flow rate provides the control  algorithm real time feedback on the current heat load for optimal temperature regulation   Temperature sensor  T4  is used to regulate the cold water within the chiller circuit     There are three valves installed in the Inlet Outlet manifold to permit filling the system with water   pre conditioning the water to a desired temperature  and emptying the pads of water at the end of  aca
61. k the I O Manifold connector  at the I O Card  Replace the I O  Manifold assembly  if necessary   Recalibrate upon completion        At power up  the level is read  and indicates less than full   Upon initiation of fill  one or  no  clicks  are heard  The  Control Module tries to fill  the  pump can be heard   but no  water is ingested  Only one  or neither of the BV and FILL  LEDs on the I O Card  illuminate     Electronics fault     Call Medivance Service        No patient temperature is  displayed even though a  probe is installed     The probe and cable have not  been tested on another  Control Module     An open circuit exists  in either the probe or  cable     Replace the probe and cable in  turn  or try each on another  Control Module  to determine  which is at fault  Flex the cable  and probe  particularly near  connections to determine whether  there is an intermittent  connection        The probe and cable have  been shown to work on  another Control Module     Electronics fault     Call Medivance Service        A patient temperature is  displayed  but no probe is  installed     At power up  no alarm occurs     The patient  temperature channel is  out of calibration     Calibrate the Control Module        At power up  no alarm occurs   Recalibration unsuccessful     Electronics fault    Call Medivance Service        The Control Module is  noisy     The noise is present at all  times and emanates from the  Power Module     A cooling fan is failing     Replace the Pow
62. l  Attempting to  drain the tank yields little or  no water   Note  the tank can  hold about 2 8 liters  The  resistance between the GND  TEST point on the Power  Card and the chassis is  greater than 1 ohm     Low voltage circuitry  ground fault     Replace the Power Module        Indicates  full  when tested as  above  ground resistance low     Electronics fault     Call Medivance Service           Control Module will not fill       At power up  the level is read  and indicates less than full   Upon initiation of fill  two   clicks  are heard  The  Control Module tries to fill  the  pump can be heard   but no  water is ingested        Pads are connected to  the patient lines   preventing suction from  being applied to the fill  tube        Disconnect all pads        39       Table 6  3  General Troubleshooting Guide       Conditions   Symptom    Observation    Cause    Solution       At power up  the level is read  and indicates less than full   Upon initiation of fill  two   clicks  are heard  The  Control Module tries to fill  the  pump can be heard   but no  water is ingested  Pads not  connected     Vacuum leak in a  patient line manifold  valve     See Low Water Flow        At power up  the level is read  and indicates less than full   Upon initiation of fill  one or  no  clicks  are heard  The  Control Module tries to fill  the  pump can be heard   but no  water is ingested  Both the  BV and FILL LEDs on the I O  Card illuminate     Bypass or Fill Valve  fault    Chec
63. l Mode with a setpoint  2 high  Check 2 reading is above Mode  Audio alarm below 38 C until the water temperature is less than  patient temperature    38   C  100 4  F  and   half second tone 38  C  Press Automatic and retest  If the error   water temperature is   every 2 minutes  persists  see General Troubleshooting   Will Not  above 38   C Alarm indicator Cool   Alarm  100 4  F  and flashing  System is unable to  cool in Automatic  Mode  22 Patient Temperature   Patient Temperature   Current Mode Adjust Patient High Temperature Alert in the       2 above user defined  limit     Alert       2 reading is equal or  greater than the  user defined value  established in the  custom settings  menu        Continues  Audio  alarm half second  tone every 2  minutes  Alarm Alert  indicator flashing       custom menu if desired        46          Table 6 4    Alarm and Alert Troubleshooting                            Alarm    Message Displayed   Alarm Alert System Status Troubleshooting   Alert Description   23 Patient Temperature   Patient Temperature   Changes to Stop Operate the system in Manual Mode with a setpoint  2 low  Check patient   2 reading is below Mode  Audio alarm above 32  C until the water temperature is greater  temperature  32 0  C  89 6  F  and   half second tone than 32  C  Press Automatic and retest    If the   water temperature is   every 10 seconds  error persists  see General Troubleshooting   Will  below 32 0  C Alarm Alert indicator   Not Heat   Alarm  89 6 
64. le The Monitor Changes to Stop Replace the I O Card and attempt to power up  If  system error Processor has failed   Mode  Audio alarm the error persists  call Medivance Service   Refer to Operator to detect a simulated   half second tone Calibrate upon completion   Manual  water temperature every 10 seconds   fault  Alarm Alert indicator  Alarm flashing   82 Non recoverable A D converter fault Changes to Stop Turn the Control Module off  Wait 30 seconds and  system error Mode  Audio alarm turn power back on  If the error persists  replace  Refer to Operator half second tone the Processor Card  If unsuccessful  call  Manual  every 10 seconds  Medivance Service  Calibrate upon completion   Alarm Alert indicator  Alarm flashing   83 Non recoverable The Monitor Changes to Stop Turn the Control Module off  Wait 30 seconds and  system error Processor does not Mode  Audio alarm turn power back on  If the error persists  replace  Refer to Operator communicate  half second tone the Processor Card  If unsuccessful  call  Manual  every 10 seconds  Medivance Service  Calibrate upon completion   Alarm Alert indicator  Alarm flashing   84 Non recoverable The Control Changes to Stop Turn the Control Module off  Wait 30 seconds and  system error Processor does not Mode  Audio alarm turn power back on  If the error persists  replace  Refer to Operator communicate  half second tone the Processor Card  If unsuccessful  call  Manual  every 10 seconds  Medivance Service  Calibrate upon completio
65. libration  10 minutes  after a 10  minute warm up   operator assistance required       Stage 2  Water Temperature Calibration  30 minutes  automatic     Stage 3  Interim Results Reporting  1 minute  operator assistance required     Stage 4  Temperature Out Calibration  30 minutes  automatic       Stage 5  Acceptance  1 minute  operator assistance required       Stage 6  Water Temperature Alarm Test  15 minutes  operator assistance  required      Stage 7  Chiller Capacity Test  15 minutes  operator assistance required    24    5 2 1 Stage 1  2  and 3  Warm up  Flow Check  Patient  Temperature Calibration  Water Temperature  Calibration  and Interim Results Reporting    1  Replace the fluid delivery lines with CTU     2  Connect the Blue circular connector labeled    PT1    to the  Model 2000 connector Patient Temperature 1  larger  thermometer and patient symbol      3  Connect the Blue circular connector labeled    PT2    to the  Model 2000 connector Patient Temperature 2  smaller  thermometer and patient symbol      4  Connect the Black circular connector labeled    TO    to the  Model 2000 connector labeled    TEMP OUT        5  Set the CTU switches       Gi B    S2A  s SZA    Note  For best results  act promptly when instructed by the  machine  Do not allow the machine to lie idle waiting for data  input or test unit switching     Note  If an alarm occurs during calibration  check the CTU  connections and switch positions  If the alarm persists  power  down and start over 
66. ls and Supplies Required     5 32  hex key    3 32  hex key  13  Remove the rear panel per Section 7 3     14  Note  Observe the routing of the cable to the Input Output  circuit card  Ensure that wires are not pinched when  reassembling     15  The Inlet   Outlet Manifold is attached to the chassis by two  shoulder screws inserted into keyholes  Pull the Inlet  Outlet    manifold forward then down to slide it off the chassis  Figure 7 25       Inlet   Outlet Manifold    16  Remove the tubing and connectors from the bottom of the  manifold by first removing the screws securing each  connector using a 3 32    hex key     17  Unplug the manifold harness from the Input Output Circuit  Card  see Section 7 10      71    7 21    1   2   3     Temperature Sensor Harness  Tools and Supplies Required      7 16    open end wrench     5 32    hex key    Drain the control unit per Section 7 1   Remove the rear panel per Section 7 3    Remove the right side panel per Section 7 4     Note  Observe the routing of the harness     4     Remove the Inlet  Outlet manifold block for access  Pull the  block toward you until it disengages from the chassis     Remove the harness from the cable clip located behind the  manifold block     Through the opening of the right side of the chassis remove  the insulation from all three temperature sensors located on  the bottom of the mixing and chilled water reservoirs     Remove the temperature sensors from the reservoirs using a  7 16  open end wrench    
67. me     1  Remove the Remote Display from the handle  Disconnect the  Remote Display Cable        2  Loosen the clamp knob until the bottom jaw swings clear of the    top head  Remove the black rubber plug in the top head by dida SE  grabbing an edge and pulling straight out  19656     3  Using a 77  hex key  adjust the hex bolt until proper tension is  achieved  Replace the plug and tighten the clamp knob        Removing Rubber Plug  Figure 4 2       Adjusting Tension  Figure 4 3    4 4 Inspect Connectors and Cables    Inspect all cables for breakage  fraying  excessive twisting or other signs of damage  Replace  damaged cables     18    4 5 Clean the Condenser   The Arctic Sun condenser fins should be inspected for dirt and  debris every three months or more often with heavy use or dusty  environment  The condenser is located underneath the grill on the  front of the control module     1  Remove the front grill by pulling the bottom out  then using  both hands at the top of the panel pull it straight out and away  from the control module  For maximum performance  the  condenser fins should be free of dust and debris     2  Use a brush or vacuum to remove any dust or debris     19       Removing Front Grill  Figure 4 4       Condenser  Figure 4 5    4 6 Drain and Replace Fluid  Tools and Supplies Required       Arctic Sun Drain Bag    It is recommended the Control Module be drained and refilled  every 3 months to assure a satisfactory concentration of Cleaning  Solution in th
68. move the remaining screws and terminals in the same  fashion     Note  It might be necessary to remove the silicone sealant for  the upper tank insert     Install the new harness in the same manner following the color  coding as in Figure 7 27     Note  If a level sensor insert gets rotated  rotate it back into the  position where there is only a very slight compression on the  rubber washer  Seal the top level sensor insert with silicone  RTV if it is removed or broken     Route the harness back through the cable clip and plug it into  the Input Output Circuit Card  see Section 7 10      73    Blue  Black  Brown    Red  Orange    Yellow       Front of Tank  Installation of Level Sensor Terminals  Figure 7 27       Hold Insert While Tightening Screw  Figure 7 28    7 23 Dielectric Withstand  Hipot  Chassis Test    This test is to be performed in accordance with UL2601 1    EN60601 1  Section 20 4  which describes the application of  voltage rising from no more than half the prescribed level to the  peak over the course of ten seconds  The voltage is maintained at  that level for one minute and is then decreased over the course of  ten seconds to less than half the peak  A tester compliant with this  standard should be used     1  Turn off the tester and the Control Module    2  Remove the Control Module power cord from Mains power and  plug it into the tester outlet    3  Connect the tester ground lead to the equipotential stud on the   Control Module located just below the p
69. n   Alarm Alert indicator  Alarm flashing   85 Non recoverable The model of the Changes to Stop Call Medivance Service        system error  Refer to Operator  Manual     Alarm       Control Module  could not be verified        Mode  Audio alarm  half second tone  every 10 seconds   Alarm Alert indicator  flashing           54          Table 6 4    Alarm and Alert Troubleshooting                      Alarm    Message Displayed   Alarm Alert System Status Troubleshooting  Alert Description  86 Non recoverable Power supply Changes to Stop Turn the Control Module off  Wait 30 seconds and  system error voltage out of range    Mode  Audio alarm turn power back on  If the error persists  replace  Refer to Operator half second tone the Power Card and attempt a power up  If the  Manual  every 10 seconds  error persists  call Medivance Service  Calibrate  Alarm Alert indicator   upon completion   Alarm flashing   98 x Non recoverable Processor fault Changes to Stop Replace the Processor and retest  Calibrate upon  system error Mode  Audio alarm completion   Refer to Operator half second tone  Manual  every 10 seconds   Alarm Alert indicator  Alarm flashing   99 Non recoverable Processor Fault Changes to Stop Replace the Processor Card  Calibrate upon       system error  Refer to Operator  Manual     Alarm          Mode  Audio alarm  half second tone  every 10 seconds   Alarm Alert indicator  flashing        completion        55       CHAPTER 7 COMPONENT REPLACEMENT    The Model 2000 is
70. n  Manual Mode with a water  target of 4  C for 5 minutes   No sound emanates from the  chiller  The CHILL LED on  the I O Card is on     Chiller Power Relay or  cable fault     Check the cables between the  relay and the I O Card  Also  check the cables between the  relay and the Power Module and  chiller  Replace cables and or the  relay  if necessary        Power up in User Diagnostic  Mode  Operate the system in  Manual Mode with a water  target of 4  C for 5 minutes   No sound emanates from the  chiller  The CHILL LED on  the I O Card is off  The  reservoir level in the  diagnostic display  Lx  is less  than 2     Water reservoir level  insufficient to run the  chiller     Fill reservoir  If unsuccessful   see  Will Not Fill         Power up in User Diagnostic  Mode  Operate the system in  Manual Mode with a water  target of 4  C for 5 minutes   No sound emanates from the  chiller  The CHILL LED on  the I O Card is off  The  reservoir level in the  diagnostic display  Lx  is  greater than 2     Electronics or chiller  fault     Call Medivance Service           CTU connected     Water does not cool to  target temperature in  Manual Mode        Power up in User Diagnostic  Mode  Operate the system in  Manual Mode with a water  target of 4 C for 5 minutes   The chiller runs but T4  remains at or above room  temperature        Chiller fault        Call Medivance Service        37       Table 6  3  General Troubleshooting Guide       Conditions   Symptom    Observation    Cau
71. nds  bolus  hot restart  or no load condition   If the  Alarm Alert indicator   alarm is recurrent  use the CTU to check the  Alarm flashing  calibration of the Control Module  Calibrate if  necessary  If the alarm persists  call Medivance  Service   35 Water temperature Water temperature is   Changes to Stop Operate the unit in User Diagnostic Mode and  low  Allow system to   below 3 5  C Mode  Audio alarm Manual Mode with the CTU attached at a setpoint  warm   38 3   F   half second tone of 4  C  Observe T1  T2  T3  and T4 for at least 10  every 10 seconds  minutes after reaching stability  If T4 drops below  Alarm Alert indicator   3  C for more than a few seconds  check the  Alarm flashing  connector at J5 of the I O Card  Verify that it is  plugged in securely  the wires emanating from it are  intact  and that the other end of the wires entering  the solid state relay are intact  Repeat the test   Replace cables or the relay  if necessary  and  retest  If the error persists  call Medivance Service   36 Water temperature Water temperature is   Changes to Stop Call Medivance Service   above safety level  above 44  C Mode  Audio alarm  Allow system to  111 2 F   half second tone  cool  every 10 seconds   Alarm Alert indicator  flashing   Alarm  37 Water temperature Water temperature is   Changes to Stop Allow the system to cool  then press Automatic or  above safety level  above 43 C Mode  Audio alarm Manual   This alarm may be caused by transient  Allow system to  109 4 F 
72. ng circuit cards     1  Remove the rear panel per Section 7 3     2  Remove the Processor circuit card by removing the bottom hex  head screw with a 1 16    hex key and carefully prying outward  at the top with a flat head screw driver  Carefully pull outward  until the circuit card is removed     3  If the Processor circuit card is replaced  calibrate per Chapter  5     7 9 Isolation Circuit Card  Tools and Supplies Required     1 16    hex key    Note  Observe electrostatic discharge control procedures when  handling circuit cards     1  Remove the rear panel per Section 7 3     2  Remove the Isolation circuit card by removing the bottom  allenhead screw with a 1 16  allen wrench and carefully prying  outward at the top slot with a flat head screw driver  Then   carefully pull outward until the circuit card is removed     3  If the Isolation circuit card is replaced  calibrate per Chapter 5     After re assembly of the unit  it is recommended to perform a   Dielectric Withstand  Hipot  Patient Applied Part Test per  EE   Section 7 24  Isolation Circuit Card  Figure 7 15       65    7 10 Input   Output Circuit Card Connections    1  Many of the wiring harnesses connect to the Input   Output  circuit card  In the course of troubleshooting or repair  it may be  necessary to remove some or all of the connections  The  following figure describes these connections    7 11    Note           2   e    ke  mm J    T K    ALES    J3  Not Used    ki                     Se  Ji ais da   
73. ntering the sensor body is intact  Power up and  Alarm Alert indicator   allow the Control Module to run for 10 minutes  If  Alarm flashing  the error persists  call Medivance Service   78 Non recoverable T4 Thermistor signal   Changes to Stop Ensure that the I O Card or J10 connector on it are       system error  Refer to Operator  Manual     Alarm       out of range   low  resistance       Mode  Audio alarm  half second tone  every 10 seconds   Alarm Alert indicator  flashing        not wet  Allow them to dry  if necessary  and retest   If the error persists  call Medivance Service        53          Table 6 4    Alarm and Alert Troubleshooting                                  Alarm    Message Displayed   Alarm Alert System Status Troubleshooting  Alert Description  79 Non recoverable T1 and T2 differ by Changes to Stop Turn the Control Module off  Wait 30 seconds and  system error more than 1 0 Mode  Audio alarm turn power back on  If the error persists  call  Refer to Operator degree C half second tone Medivance Service   Manual  every 10 seconds   Alarm Alert indicator  Alarm flashing   80 Non recoverable The Control Changes to Stop Replace the I O Card and attempt to power up  If  system error Processor has failed   Mode  Audio alarm the error persists  call Medivance Service   Refer to Operator to detect a simulated   half second tone Calibrate upon completion   Manual  water temperature every 10 seconds   fault  Alarm Alert indicator  Alarm flashing   81 Non recoverab
74. o be drawn into the pads and the  circulating circuit  emptying the pads     The water level in the Main Tank is sensed by detecting conductivity with an array of stainless  steel studs in the tank  There are five level sensors and a reference sensor     y ELLE LE LS        Chiller  Pump    Chiller Circuit    O nnnnnnunnun A  lt a OA    A Eeer        Vent Line Inlet Outlet  Manifold            Circulating    Pump           Circulating Circuit    Chiller       Air Flow       O Drain Ports    Model 2000 Hydraulic Schematic 10    Figure 2 1    2 4 Electronic Control System    The electronic control system consists of two independent subsystems  the Control and the Monitor   The Control Subsystem performs all of the normal machine operations while the Monitor Subsystem  continually monitors important machine operations with independent sensors  If the Control acts  incorrectly with respect to the safety of the patient  the monitor will disable pump power  thus  preventing inappropriate therapy from being delivered  Each subsystem has a microprocessor   watchdog  keyboard input circuitry  audio alarm  outlet water temperature sensing circuitry  and  patient temperature sensing circuitry     The Control Subsystem performs the following functions      Command  keyboard  interpretation and information display    Outlet water temperature control     Water flow control     Patient temperature measurement  Patient Temperature 1  and control    Temperature Out signal generation     Cold 
75. om the alarm half second the Isolation Card and retest  If the error persists   Patient Temperature   tone every 2 call Medivance Service  Calibrate upon  2 probe input is not minutes  Alarm Alert   completion   Alert functioning  indicator flashing   32 Patient temperature   In Manual Mode  Current Mode Turn the Control Module off  Wait 30 seconds and  2 calibration error  internal temperature   continues  Audio turn power back on  If the error persists  replace  Restart system  calibration from the alarm half second the Isolation Card and retest  If the error persists   Patient Temperature   tone every 2 call Medivance Service  Calibrate upon  2 probe input is not minutes  Alarm Alert   completion   Alert functioning  indicator flashing   33 Water temperature Water temperature is   Changes to Stop Call Medivance Service        high  Allow system  to cool     Alarm       above 44 0  C   111 2  F         Mode  Audio alarm  half second tone  every 10 seconds   Alarm Alert indicator  flashing           48          Table 6 4    Alarm and Alert Troubleshooting                            Alarm    Message Displayed   Alarm Alert System Status Troubleshooting  Alert Description  34 Water temperature Water temperature is   Changes to Stop Allow the system to cool  then press Automatic or  high  Allow system above 42 5  C Mode  Audio alarm Manual   This alarm may be caused by transient  to cool   108 5   F   half second tone conditions such as a momentary occlusion  air  every 10 seco
76. omprise the Isolation Card  One island  is connected to the Backplane and provides an interface to the rest of the card cage  This island  11    is referenced to chassis ground  The other islands are isolated from the Backplane island by  opto isolators or transformers  The RS 232  1  RS 232  2  and Temperature Out  Echo  islands  are isolated from the Backplane and from each other to a breakdown voltage of at least 500V   The patient temperature island is isolated from all others to a level of 1500V  It should be noted  that  in addition to this  the 24V power supply in the Power Module isolates all low voltage  circuitry to a level of 4000V from Mains         Input Output  I O  Circuit Card  The I O Card contains the majority of the actuator and sensor  circuitry  T1  T2  T3  and T4  the flowmeter  pressure sensor  and level sensor are conditioned on  the I O Card  Both pump motors  the valves  and the refrigeration system are controlled by this  circuit card     The remaining circuitry is housed in the Remote Display Module  This module communicates with  the card cage via a cable having 13 twisted pairs using a technology called Low Voltage Differential  Signaling  LVDS  which minimizes both the risk of spurious signals interfering with communication  and the emission of electromagnetic energy  The Remote Display Module houses LED display  drivers  keyboard interpretation circuitry  an audible alarm  primary   and an LCD text module     There is also a manually resettable
77. or  sterile water to fill the system        Pads connected     System fails to purge  completely     Pads do not empty well during  purge cycle  A  click  is  heard when the purge is  initiated     Vent line is kinked    Inspect vent line which is small  diameter line exiting the bottom of  the Inlet Outlet Manifold and  entering top of tank  Replace if  kinked        Pads do not empty well during  purge cycle  A  click  is  heard when the purge is  initiated     Water has  inadvertently entered  purge line or valve     Repeat purge multiple times to  attempt to clear water  If  necessary  detach one end and  drain           Pads connected     System fails to purge  completely        Pads do not empty well during  purge cycle  A  click  is  heard when the purge is  initiated  Multiple purge  cycles are unsuccessful        Tank is overfilled        Drain 0 5 liter water from the  upper drain port  Power up in  User Diagnostic Mode and check  the level  Repeat this process  until the level is less than 5        38       Table 6  3  General Troubleshooting Guide          Conditions   Symptom Observation Cause Solution  Pads do not empty well during   Purge valve Replace the I O Manifold  purge cycle  No  click  is inoperative  assembly  Recalibrate upon    heard when the purge is  initiated  The VV LED on the  UO CCA illuminates during  the purge     completion        Pads unable to empty during  purge cycle  No  click  is  heard when the purge is  initiated  The VV LED on the  
78. ower inlet    Turn on the Control Module power switch    Turn on the tester    Set the tester voltage to 1500VAC    Stand clear of the Control Module and start the test  Do not   touch the Control Module while the test is in progress    Verify that no alarm occurs                   gue              Tester Ground Lead to Equipotential Stud  Figure 7 29    7 24 Dielectric Withstand  Hipot  Patient Applied Part  Test    Call Medivance Service for instructions     74     blank page     75    7 25 Inlet   Outlet Manifold Valve Harness  Tools and Supplies Required       5 32  hex key    9 16    wrench      7 16    wrench       1  Remove the rear panel per Section 7 3        2  Note  Observe the routing of the cable to the Input Output N  2  circuit card  Ensure that wires are not pinched when Inlet   Outlet Manifold  reassembling    Figure 7 30    3  Lift the lever on the Circulating Pump  located below the  manifold  then remove the connector from the pump        Removing Pump Port  Figure 7 31    4  Pull the Inlet  Outlet manifold forward then down to slide it off  the chassis     5  Unplug the pressure transducer plug from the green circuit  board by first compressing the tab on the underside of the  connector  Do not stress the wires        Removing Plug  Figure 7 32    76    10     11    12     13     14     15       Check that the valve stems are tight but do not over tighten     Remove the nut from the top of each solenoid coil being  careful not to unscrew the valve stem  s
79. rol       malfunction  Replace  temperature probe or  cable     Alarm       1 reading is outside  the high limit of the  display range in  Automatic Mode    shorted        Mode  Audio alarm  half second tone  every 10 seconds   Alarm Alert indicator  flashing        Module  Check connection of patient temperature  probe and cable  If wet  replace cable  Check  temperature cable connection at the rear of the  Control Module  Flex the cable to ensure that there  is not an intermittent fault  Test the input with a  cable and probe proven good on another Control  Module  If the Control Module proves inaccurate   replace the Isolation Card and retest  If the error  persists  call Medivance Service  Calibrate upon  completion        45          Table 6 4    Alarm and Alert Troubleshooting                            Alarm    Message Displayed   Alarm Alert System Status Troubleshooting   Alert Description   17 Patient Temperature   In Automatic Mode  Changes to Stop Turn the Control Module off  Wait 30 seconds and  calibration error  internal temperature   Mode  Audio alarm turn power back on  If the error persists  replace  Restart system  calibration from half second tone the Isolation Card and retest  If the error persists    Patient Temperature   every 10 seconds  call Medivance Service  Calibrate upon  1 probe input is not Alarm Alert indicator   completion   Alarm functioning  flashing    18 Patient Temperature   In Manual Mode  Current Mode Turn the Control Module off  Wait 3
80. sconnected or out the lower limits of the   half second tone probe and cable  If wet  replace cable  Check  of range  display range in every 10 seconds  temperature cable connection at the rear of the   Automatic Mode  Alarm Alert indicator   Control Module  Flex the cable to ensure that there  flashing  is not an intermittent fault  Test the input with a  Alarm cable and probe proven good on another Control  Module  If the Control Module proves inaccurate   replace the Isolation Card and retest  If the error  persists  call Medivance Service  Calibrate upon  completion    15 Unable to obtain Patient temperature   Changes to Stop Use the CTU to verify the accuracy of the Control  stable patient discontinuity  Mode  Audio alarm Module  Check connection of patient temperature  temperature for Significant change in   half second tone probe and cable  If wet  replace cable  Check  automatic control  patient temperature every 10 seconds  temperature cable connection at the rear of the   reading for more Alarm Alert indicator   Control Module  Flex the cable to ensure that there  than 10 minutes  flashing  is not an intermittent fault  Test the input with a  Alarm cable and probe proven good on another Control  Module  If the Control Module proves inaccurate   replace the Isolation Card and retest  If the error  persists  call Medivance Service  Calibrate upon  completion   16 Temperature probe 1   Patient Temperature   Changes to Stop Use the CTU to verify the accuracy of the Cont
81. scription of alarm and alert codes and  messages and the condition limits that triggered the alarm or alert     Note  This guide includes non recoverable alarms  These alarms should be address only by  someone trained and qualified to work on this system     42       Table 6 4    Alarm and Alert Troubleshooting                            Alarm    Message Displayed   Alarm Alert System Status Troubleshooting   Alert Description   01 Low fluid pressure in   Fluid delivery line is   Current Mode See General Troubleshooting   Low Flow   lines  Check open to air  leaking Continues  Audio  connections  or has significant air   alarm half second   in line tone every 2  minutes  Alarm Alert  Alert indicator flashing  02 Low water flow to Flow to the pads is Current Mode See General Troubleshooting   Low Flow   pads decreased by  gt 50    Continues  Audio    of the maximum flow   alarm half second  Check lines  detected during the tone every 2  connections and case or since the last   minutes  Alarm Alert  pads  low flow alert  or the indicator flashing  flow is  lt  0 3 liters  per minute  300  Alert ml minute     03 Water reservoir level   Reservoir fluid level Alert occurs after Fill Reservoir  If the error persists  see General   low  almost empty  There   Purge is complete  at   Troubleshooting   is only enough water   start up  or after a fill  left in the reservoir to   cycle   Alert run one procedure  Audio alarm half  second tone every 2  minutes  Alarm Alert  indicator flash
82. se    Solution       Power up in User Diagnostic  Mode  Operate the system in  Manual Mode with a water  target of 4 C for 5 minutes   The cooling command  PWM   Mxx  is nonzero and T4  stabilizes below 6C     Pump cable or  connection fault     Check the Mixing Pump  connection at the I O Card  Verify  that the circuit card is not wet   Reseat the connector  Allow the  card to dry  if necessary        Power up in User Diagnostic  Mode  Operate the system in  Manual Mode with a water  target of 4 C for 5 minutes   The cooling command  PWM   Mxx  is nonzero and T4  stabilizes below 6C  The  Mixing Pump connection at  the I O Card is dry and  secure     The pump head may  be ineffective due to  debris stuck in a valve  or the pump electronics  have failed     Replace the Mixing Pump        Power up in User Diagnostic  Mode  Operate the system in  Manual Mode with a water  target of 4 C for 5 minutes   The cooling command is  nonzero and T4 stabilizes  below 6C  Mixing Pump  replacement ineffective     Electronics fault    Call Medivance Service        Power up in User Diagnostic  Mode  Operate the system in  Manual Mode with a water  target of 4 C for 5 minutes   the cooling command is  nonzero and T4 changes little     Electronics or chiller  fault     Call Medivance Service        Water drained from system  appears green    Highly alkaline tap  water used at some  point to fill system   Causes chelating of  copper ion from chiller  evaporator     Drain and only use distilled 
83. se  When the Fill mode is activated  water is drawn up through the Fill Tube  through the Fill  Valve  FV  by the Circulating pump  The water then enters the Circulating Tank and by gravity  passes to the Main Tank  Air  displaced by the water  exits the Main Tank through a small hole  in the top of the tank  It is important to note that pads cannot be connected during this operation     The user may wish to pre condition the temperature of the water prior to pads being connected    The system will detect the absence of pads connected by feedback from the flow meter and   pressure sensor and will open the Bypass Valve  BV   Water is drawn out of the Main Tank   through the Bypass Line and is shunted past the Fluid Delivery Lines into the circulating circuit  8    where it returns back into the Circulating Tank  Water flows through the leveling line from the  Circulating Tank back into the Main Tank  During this cycle  the temperature of the water is  controlled by activating the Heater or Mixing Pump  Automatic activation of the Bypass Valve  occurs briefly at the beginning of both Automatic Mode and Manual Mode in order to completely  prime the Circulating Pump  If pads are connected  the Bypass Valve will close after one minute     A user can initiate the Purge Mode in order to empty the pads of water prior to disconnecting  them  Purge Mode causes the Vent Valve  VV  to open to atmosphere through the vent line  communicating with the top of the Main Tank  This causes air t
84. the  card protrudes approximately one inch to expose the Piezo Power Circuit Card  Alarm and the DC Power cable connectors  Disconnect the Figure 7 13  Piezo Alarm and DC Power cables     4  Carefully slide the Power circuit card approximately another  inch and disconnect the AC Sync cable     5  Now the Power circuit card may be fully removed     6  After re assembly of the unit  it is recommended to perform a  Dielectric Withstand  Hipot  Chassis Test per Section 7 23     7 7 Power Module  Tools and Supplies Required     1 16    hex key    Note  Observe electrostatic discharge control procedures when  handling circuit cards     1  Remove the rear panel per Section 7 3    2  Remove the right side panel per Section 7 4    3  Remove the Power circuit card per Section 7 6   4    Disconnect the Chiller Power cable from the back of the card  cage per Section 7 5     5  Disconnect the heater power cable at the bottom of the card  cage per Section 7 5        6  Remove the Power Module by removing the 1 16  hex head Power Module  screws at the top and bottom of the bracket and carefully pulling Figure 7 14  outward     7  After re assembly of the unit  it is recommended to perform a  Dielectric Withstand  Hipot  Chassis Test per Section 7 23     64    7 8 Processor Circuit Card  Tools and Supplies Required     1 16    hex key    Note  The machine does not have to be drained in order to  complete this procedure     Note  Observe electrostatic discharge control procedures when  handli
85. the CTU at  various temperatures  In this way the patient temperature channel can then be used to check and  calibrate the accuracy of the Control Module s internal water temperature sensors and circuits     The fluid path in the CTU is designed to allow a specific flow rate at the operating pressure  Since  the Model 2000 controls the inlet pressure at a constant  7 psi  a crosscheck of flow and pressure  measurement accuracy can be obtained  The inlet pressure offset is zeroed during the initial warm   up period     The CTU water heaters are used to slowly raise the system water temperature above its high  temperature limits in order to verify correct operation of the Control Module   s safety systems  The  test unit heaters are thermally and electrically protected  and are only powered while pressing the  Heater Power Switch     5 1 2 Warnings    The CTU must not be used on equipment connected to a patient       Do not use the Calibration and Test Unit in the presence of flammable agents because an  explosion or fire may result       Anyone performing the procedures must be appropriately qualified     Carefully observe the Calibration and Test unit for fluid leaks before and during use  Leakage  may result in lower flow rates or an electrical shock hazard     5 1 3 Electrical Ratings    Voltage  115V 60Hz or 230V 50Hz      Current  4 5A or 2 25A    Fuse  6 3A Slow Blow  2     Power  500W    22    5 1 4 Maintenance    The CTU contains no serviceable parts  Other than period
86. top Replace the Power Card and attempt to power up   system error pump power Mode  Audio alarm If the error persists  call Medivance Service   Refer to Operator  Monitor  half second tone  Manual  every 10 seconds   Alarm Alert indicator  Alarm flashing   66 Non recoverable Unable to disable Changes to Stop Replace the Power Card and attempt to power up   system error pump power Mode  Audio alarm If the error persists  call Medivance Service   Refer to Operator  Control  half second tone  Manual  every 10 seconds   Alarm Alert indicator  Alarm flashing   67 Non recoverable Unable to disable Changes to Stop Replace the Power Card and attempt to power up   system error pump power Mode  Audio alarm If the error persists  call Medivance Service   Refer to Operator  Monitor  half second tone  Manual  every 10 seconds   Alarm Alert indicator  Alarm flashing   70 Non recoverable Mains signal is Changes to Stop Replace the Power Card and attempt to power up        system error  Refer to Operator  Manual     Alarm       missing or  inconsistent       Mode  Audio alarm  half second tone  every 10 seconds     Alarm Alert indicator    flashing        If the error persists  call Medivance Service        52          Table 6 4    Alarm and Alert Troubleshooting                                     Alarm    Message Displayed   Alarm Alert System Status Troubleshooting  Alert Description  71 Non recoverable 12 Thermistor signal   Changes to Stop Check the connector at J10 of the I O Card  Veri
87. ut no lights appear     Power Card fault     Replace the Power Card           The Control Module has  failed Biomed checkout              Call Medivance Service        41       6 4 Troubleshooting Alarms and Alerts    The Arctic Sun is designed with the patient s and operator s safety in mind  The Control Module has  redundant  cross checking systems and a multitude of self tests  It detects inappropriate conditions  and notifies the user  Many of these conditions can be addressed by clinical actions alone  These  are described in the Operator s Manual  If the alarm or alert cannot be resolved clinically or is  recurrent  the matter should be addressed by technical personnel  Table 6 4 below is intended as an  aid in such an investigation     Alerts and Alarms have specific definitions     Alert  Alerts are intended to inform the user about patient and equipment status without interrupting  the procedure     Alarm  When an Alarm condition occurs  there is a chance that a potentially unsafe condition has  occurred  or could occur with respect to the equipment  the patient  or an improperly functioning  component     Alarms are divided into two types     recoverable and non recoverable  Recoverable alarms may be  cleared when a condition is met  Non recoverable alarms require correcting the condition and  powering the system back up to clear the alarm  Non recoverable alarms are identified as such on  the display     Refer to the Alarm  amp  Alert Troubleshooting Table for a de
88. utes  Alarm Alert   designated clinical indicator flashing  Alert engineering or   Medivance service    46 Non recoverable User interface not Changes to Stop Check Remote Display Cable connections on the  System error detected by control Mode  Audio alarm remote display module and Control Module    Refer to Operator System  half second tone Ensure that they are secure  Turn the Control  Manual  every 10 seconds  Module off for 30 seconds and turn power back on   Alarm Alert indicator   If the error persists  replace the Remote Display  If  flashing  unsuccessful  call Medivance Service   Alarm  47 Non recoverable User interface not Changes to Stop Check Remote Display Cable connections on the       system error  Refer to Operator  Manual     Alarm       detected by monitor  system        Mode  Audio alarm  half second tone  every 10 seconds   Alarm Alert indicator  flashing        remote display module and Control Module   Ensure that they are secure  Turn the Control  Module off for 30 seconds and turn power back on   If the error persists  replace the Remote Display  If  unsuccessful  call Medivance Service        50          Table 6 4    Alarm and Alert Troubleshooting                            Alarm    Message Displayed   Alarm Alert System Status Troubleshooting  Alert Description  48 Patient temperature   Patient Temperature   Current Mode Turn the Control Module off  Wait 30 seconds and  out invalid  Output  Echo  Continues  Audio turn power back on  If the error pers
89. water supply temperature control     Valve control   The Monitor Subsystem performs the following functions       Redundant command  keyboard  interpretation     Outlet water temperature monitoring     Patient temperature measurement  Patient Temperature 2     Pump power control     Power supply monitoring     Data logging output to the RS 232 port    The majority of the hardware comprising the electronic control system is located within the card cage   The card cage consists of six major elements  See Figure 7 4        Power Module  The Power Module includes a 24VDC medical universal input switching power  supply  IEC601 1 compliant   a Mains circuit breaker  cooling fan  heater circuit fuse   equipotential stud  and solid state relay  The Power Module is connected to the Power Card by  two pigtail connectors       Backplane  The Backplane is passive  serving only to connect the four circuit cards listed below       Power Circuit Card  The Power Card provides low voltage power regulation to the balance of the  electronics as well as Mains line synchronization  with 4000V isolation from Mains   pump power  control  a Mains loss alarm  and a secondary audible alarm  The Mains loss circuit is powered by  low leakage capacitors and needs no battery       Processor Circuit Card  The Processor Card houses the Control and Monitor microprocessors  and their associated circuitry  including nonvolatile memory for each      Isolation Circuit Card  Five isolated  islands  of circuitry c
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Télécharger la notice  DR-V1HD Owner`s Manual  PS710 Anleitung 010 DE.indd    Mellanox Firmware Tools (MFT) User's Manual    V7 LED215W2  Radio Shack 21-1599 User's Manual  DEMONTAGE DU RETRACTEUR A EVENTAIL MONTAGE DU  Symantec Event Manager for Security Gateways (10142990)    Copyright © All rights reserved. 
   Failed to retrieve file