Electrophoresis Power Supply EPS 1001
Contents
1. W T 5 pause continue STOP 4 1 2 Keyboard ENTER Enters a value locks the value if valid and moves programming to the next field Valid values are voltage 5 1000 V current 1 400 mA power 1 100 W time 0 01 500 h volthours 1 500000 Vh SET Puts the instrument into SET its programming and starting mode In RUN mode pressing SET also shows the programmed parameters for the actual run In addition SET allows you to make changes in the program during a run after first pressing PAUSE CONTINUE Change up Change down Changes the parameter value or other choice in the field which is blinking N umerical values are changed in an accelerating manner when a key is held down Clicking a key changes the value in preset increments Parameters or units e g Vh and yes no choices are changed with one key push The keys can also be used to switch between time and volthours in RUN PAUSE and END The values scroll i e they automatically change from maximum to minimum value or vice versa RUN PressingRUN starts the run and puts the program into RUN mode The program number current values for voltage current and power are shown on the display The elapsed time or volthours are also displayed Switch between time and volthours with 4 7 PAUSE Puts the instrument in PAUSE mode and switches off the voltage The display shows the status of the run at the time the key was pressed PAUSE CONTINUE only operates in RUN mode Time and
2. 1 Perdita di messa a terra funzionale Se l alimentatore collegato ad un unit per elettroforesi dotata di dispersione di messa a terra l EPS 1001 rilever il difetto e disattiver la tensione 2 Controllo della corrente di avviamento Per assicurarsi che l unit per elettroforesi sia collegata correttamente l alimentatore controlla che la resistenza non superi un limite specifico a bassa tensione 40 V Sela resistenza troppo elevata viene disattivata l alimentazione Una resistenza troppo elevata pu essere dovuta anche all utilizzo di tamponi con una conduttivit estremamente ridotta L alimentazione viene disattivata anche in questo caso ed appare un messaggio di errore Q uesta funzione pu essere disabilitata per alcune applicazioni vedere Fig 4 disattivazione del controllo della corrente 3 Rilevamento di cambi di carico improvvisi Q uesta funzione previene incidenti durante il funzionamento dovuti ad un guasto nell impianto elettrico come ad esempio un collegamento errato all unit per elettroforesi In tal caso viene disattivata l alta tensione 3 Disimballaggio ed installazione Disimballaggio Accertarsi cheil contenuto della confezione sia conforme all elenco allegato Controllare che lo strumento non abbia subito danni durante il trasporto Qualora vi siano danni rivolgersi immediatamente al rappresentante A mersham Pharmacia Biotech ed alla ditta di trasporti interessata Collegamento alla rete
3. SDS PAGE Polyacrylamide Gel Electrophoresis Native PAGE Agarose electrophoresis Electroblotting DNA pulsed field electrophoresis DNA fragment separation 7 DNA submarine electrophoresis OO 01 bh UN Electrophoresis separations can be controlled by voltage current or power The EPS 1001 automatically switches over the controlling parameter according to programmed limits and conductivity variations in the system Two electrophoresis units can be connected to the EPS 1001 and run with the same programmed method at one time Three programs can be saved 2 Safety information 2 1 Safety precaucions Extreme caution should be exercised in the operation of this instrument as it can develop sufficient voltage and current to produce a lethal shock To avoid any risk of injury the instrument should only be operated by properly trained personnel and always in accordance with the instructions provided Read this entire manual before using this power supply 1 This instrument is designed for indoor use only 2 Theinstrument must always be used with the protective earth lead of the power cord correctly grounded to earth at the mains outlet 3 To permit sufficient cooling ensure that the vents in the rear and sides of the instrument are not covered 4 Do not operate the instrument in extreme humidity above 95 Avoid condensation by letting the unit equilibrate to ambient temperature when taking the power suppl
4. OFF para cambio manual Confirmar con la tecla ENTER 5 En el modo cambio autom tico fijar el valor a utilizar para el punto de cambio 4 7 con las teclas y pulsar ENTER 6 Si se ha elegido el evento seg n el punto 4 el sistema pregunta si se desea que suene la alarma al terminar el proceso Seleccionar s Y ES o no NO con las teclas Y y pulsar ENTER Si se elige alarma ello se indica con una A peque a en supra ndice como en 1A a la derecha del numero de programa en la pantalla 7 Al terminar la programaci n conectar la unidad electrofor tica a la toma y pulsar la tecla RUN 8 La pantalla muestra los valores actuales de tensi n amperaje y potencia as como el tiempo transcurrido o los voltioshora conmutar las teclas 4 1 9 Latecla PAUSE CONTINUE puede usarse para interrumpir y continuar el funcionamiento 10 Los par metros programados pueden controlarse durante el funcionamiento pulsando la tecla SET 11 El proceso programado se para autom ticamente pulsando la tecla STOP EPS 1001 Manual del usuario ES Los caracteres intermitentes se visualizan en negrita 1 Elegir el n mero de SET P1 programa 200 V 40 mA Seleccionar con Confirmar con 2 an de po OFF 400 mA 100 W Seleccionar con Confirmar con 3 Fij l mite d eu SET P2 OFF amperaje 1000 V 400 mA 100 W Seleccionar con 0 U Confirmar con 4 Fijar el l mite de SET P2 potencia 3 1000 V 50 mA Selecc
5. 4 1 3 Ausgangsbuchsen Es sind zwei S tze von Ausgangsbuchsen vorhanden so da zwei Elektrophoreseger te angeschlossen und gleichzeitig betrieben werden k nnen Abb 1 siehe Deckblattinnenseite Der Spannungsausgang ist 0 1000 V Die negative Ausgangsbuchse gibt O bis 500 V die positive O bis 500 V 4 2 R ckwand Die R ckwand ist in Abb 2 siehe Deckblattinnenseite dargestellt An der R ckwand befinden sich 1 Ein Netzschalter Durch Dr cken von wird der Strom zum Stromversorgungsger t eingeschaltet Durch Dr cken von 0 wird der Strom abgeschaltet 2 Eine Buchse f r das N etzkabel 3 Ein Schalter f r den Spannungsbereich Die linke Schalterstellung entspricht dem Bereich 100 120 V die rechte dem Bereich 220 240 V 4 Bel ftungs ffnungen DE EPS 1001 Bedienungsanleitung 5 Betrieb 5 1 berblick Der Betrieb des EPS 1001 wird in der nachfolgenden A blauffolge beschrieben I 2 Programmieren oder Bearbeiten eines V erfahrens siehe Abb 3 und Abb 4 Abarbeiten eines Verfahrens siehe Abb 5 5 2 Kurzanweisungen Dieses Kapitel fa t die H auptpunkte f r den Betrieb des EPS 1001 zusammen Es wird auch auf die separate schematische Betriebsanleitung verwiesen die dem Stromversorgungsgerat beiliegt Es wird empfohlen diese schematische A nleitung stets in der N he des Ger ts aufzubewahren 1 2 ER 8 9 Den N etzstrom einschalten Auf der Anzeige sollte die Programm
6. Choose between manual break or automatic break For automatic break choose breakpoint units in either V olthours Vh or time h Select the correct unit or for manual break choose OFF with 4 9 Confirm with ENTER 5 If volthours or time were chosen set the value to be used for the breakpoint with 4 7 and ENTER 6 If an automatic breakpoint was chosen under 4 you will be asked if you wish an alarm to sound at the end of the run Select YES or NO with Y and press ENTER If Y ES was chosen this is indicated by a small A as in 1A to the right of the program number in the display 7 When programming is completed connect your electrophoresis unit to the outputs and press ENTER 8 The actual voltage current and power are shown on the display together with the elapsed time or volthours switch with 44 7 9 PAUSE CONTINUE can be used to interrupt and then continuethe run 10 The programmed parameters can be checked during the run by pressing SET 11 The programmed run stops automatically or by pressing STOP 11 GB EPS 1001 User Manual 6 Maintenance Wipe the instrument regularly with a damp cloth Let the instrument dry completely before use Otherwise no user maintenance is necessary All servicing should be entrusted to qualified personnel only Please contact your local Amersham Pharmacia Biotech representative for more service information 7 Trouble shooting If an error that can be corrected by
7. Confirm with ENTER Repeat the same procedure for limiting current and limiting power Programmable values for voltage are 5 1000 V current 1 400 mA power 1 100 W time 0 01 500 h volthours 1 500 000 Vh Setting break point Choose between manual break or automatic break For automatic break choose breakpoint units in either Volthours Vh or time h Select the correct unit or for manual break choose OFF with 4 7 Confirm with ENTER If volthours or h time were chosen the display will flash for the break value Set the value with 4 1 and confirm with ENTER Choosing alarm When a breakpoint has been entered the following question is shown Alarm at End Y Select yes or no with Y and confirm with ENTER If yes is selected a small A appears on the right of the program number i e 2 EPS 1001 User Manual GB 1 Choose Program number 2 Set Voltage limit 3 Set Current limit 4 Set Power limit 5 Choose Breakpoint unit h or Vh or no Breakpoint OFF 6 Set Breakpoint 7 Choose Alarm YES or NO 8 Program ready SET P1 200 V 40 mA Select by Confirm by P2 OFF 400 mA 100 W Select by Confirm by SET P2 OFF 1000 V 400 mA 100 W Select by Confirm by SET P2 1000 V 50 mA Select by Confirm by SET 1000 V P2 50 mA Select by Confirm by SET P2 1000 V 50mA Select by Breakpoint off Confirm by Select by Confirm b
8. Manuale dell operatore 1 Introduzione L alimentatore per elettroforesi Amersham Pharmacia Biotech EPS 1001 un unit di alta qualit e precisione per applicazioni di elettroforesi che richiedono un massimo di 1000 volt 100 Watt e 400 mA L EPS 1001 stato concepito principalmente per le seguenti tecniche 1 SDS PAGE Elettroforesi con gel poliacrilammide PAGE nativa Elettroforesi di agarosio Elettroassorbimento Elettroforesi pulsata del DN A Separazione dei frammenti di DNA 7 Elettroforesi sommersa del DNA OU BW NN Le separazioni per elettroforesi possono essere controllate mediante tensione o corrente L EPS 1001 imposta automaticamente i parametri di controllo in conformita ai limiti programmati ed alle variazioni di conduttivita nel sistema Possono essere memorizzati tre programmi 2 Informazioni sulla sicurezza 2 1 Norme di sicurezza Poich questo strumento pu sviluppare una tensione ed una corrente sufficienti per provocare una scossa elettrica letale prestare particolare attenzione durante il suo utilizzo Per evitare il rischio di incidenti l alimentatore deve essere utilizzato esclusivamente da personale esperto ed addestrato ed in conformit alle istruzioni del fabbricante Leggere attentamente l intero contenuto del manuale prima di utilizzare l alimentatore 1 Questo strumento stato progettato per essere utilizzato solamente in ambienti chiusi 2 Assicurarsi che il condut
9. 1 1 Display sense dene 4 4 1 2 Keyboard sort 5 4 1 3 Output sockets 5 4 2 Rear WANG A Es 5 5 Dperalloli ccce pietre 6 5 1 Overview scorre oua rd wes 6 5 2 Programming ix esr en 6 5 3 Running a program 8 5 4 Optional programming 9 5 5 Choosing run parameters 10 5 6 Short instructions 11 6 Maintenance isses n 12 7 Troubleshooting ie as 12 8 Technical specifications 13 9 Orderinginformation 14 Please read this entire manual to fully understand the safe use of EPS 1001 N that must be strictly followed in order to avoid personal injury Be sure not to proceeed until the instructions are clearly understood and all stated conditions are met WARNING The Warning sign highlights an instruction Declaration of conformity Safety Standards This product complies with the requirement of the Low Voltage Directive LVD 73 23 EEC through the harmonized standard EN 61010 1 1993 A1 1992 EMC Standards This product complies with the requirement of the EM C Directive 89 336 EEC through the harmonized standards EN 50081 1 emission and EN 50082 1 immunity The CE symbol and corresponding declaration of conformity is valid for the instrument when it is used as a stand alone unit or connected to other CE marked Amersham Pharmacia Biotech instruments or connected to other products r
10. 50mA 60W programma SET P2 2 00h 1000 V 50mA 60W PAUSE P24 0 30h 500V 50mA 100W Programmazione 4 Interrompere il funzionamento RUN P2A 1 45 h manualmente 1000V 35mA 35W END P2 1 45h 1000 V 35mA 35W 5 Il funzionamento viene interrotto END P22 2 00h automaticamente 1000 V 20 mA 20W Fig 5 Riepilogo di funzionamento e funzioni di pausa passo dopo passo o amersham pharmacia biotech Printed in Sweden by Snits amp design AB TK i Uppsala AB December 1998
11. A switch for voltage range The left position corresponds to 100 120 V and the right to 220 240 V 4 Vent openings GB EPS 1001 User Manual 5 Operation 5 1 Overview The operation of the EPS 1001 is described in the following sequence 1 Programming or editing a method 2 Running a method W hen the method is running it can be interrupted put in PAUSE mode to apply samples and or change the program During a run it is also possible to view the program settings Programming and running are discussed in more detail in the following two sections Blinking characters are shown as bold characters 5 2 Programming Start position When the power supply is switched on the display shows the start position in SET mode The previous program set is shown and that program number is blinking for example SET P1 2 00 h 200V 40mA 10 W If the power supply is switched on for the first time the breakpoint parameter and the alarm are off and the values for voltage current and power are 5 V 400 mA and 100 W respectively Choosing a program Up to 3 programs can be entered Press ENTER to confirm the program shown on the display or use 4 9 keys to choose another Confirm with ENTER The programming procedure can be followed step by step in Fig 3 Setting voltage current and power limits The display will now flash for the set voltage limit Using the Y keys select the maximum voltage desired for the run
12. IT run view STOP STOP Ferma il funzionamento e commuta lo strumento modo End La tensione viene disattivata e vengono visualizzati i parametri finali Dopo aver premuto STOP non possibile continuare il funzionamento MORE E possibile disattivare il controllo della corrente di avviamento il cui compito rilevare se la resistenza troppo elevata e disattivare l alimentazione EXIT Interrompe l esecuzione di un operazione come l inserimento di un valore Vengono conservati solamente i valori unit confermati in precedenza Lo strumento ritorna nel modo nel quale era rimasto o nella posizione di avviamento nel modo SET 4 1 3 Prese di uscita Vi sono due gruppi di prese di uscita per il collegamento ed il funzionamento contemporaneo di due unit per elettroforesi vedere Fig 1 La tensione erogata 0 1000 V La presa di uscita negativa fornisce una tensione compresa tra 0 e 500 V quella positiva tra 0 e 500 V 4 2 Pannello posteriore La Fig 2 vedere prima di copertina mostra il pannello posteriore Sul pannello posteriore si trovano 1 Un interruttore principale Premere per accendere l alimentatore Premere O spegnere l alimentatore 2 Una presa per il cavo di rete 3 Un commutatore di tensione La posizione sinistra corrisponde a 100 120 V quella destra a 220 240 V 4 Bocchette T EPS 1001 Manuale dell operatore 5 Funzionamento 5 1 Vista generale La seguente sequenza i
13. Programmlauf und Pausefunktion EPS 1001 Manual del usuario ES Contenido ao CONS uc ade eR Paes wes 2 Seguridad SERIEN E d 2 2 1 Medidas de seguridad 2 2 2 Funciones de seguridad incorporadas pica rss das 2 Desembalaje e instalaci n 3 Descripci n t cnica ais exo ad baee 4 4 1 Panel delantero 4 Haga el favor de leer el manual completo para comprender perfectamente el uso del EPS 1001 A equil tero en el manual advierte al usuario sobre la presencia de instrucciones importantes de operaci n y mantenimiento del aparato ADVERTENCIA El signo de admiraci n en un tri ngulo Informaci n importante para el usuario Declaraci n de conformidad N ormas de seguridad Este producto cumple las disposiciones de la Directiva relativa a Bajo Voltaje 72 23 EEC a trav s de la norma armonizada NE 61010 1 1993 A1 1992 NormasEMC EMC Electromagnetic Compatibility Este producto cumple las disposiciones de la Directiva EM C 89 336 EEC a trav s de la normas armonizada NE 50081 1 emisiones y NE 50082 1 inmunidad T rminos y condiciones de venta Todas las mercanc as y servicios se venden seg n los t rminos y condiciones de venta de la empresa dentro del Grupo A mersham Pharmacia Biotech que los suministra Se puede obtener un ejemplar de estos t rminos y condiciones solicit ndolo Si tiene alg n comentario sobre estos produ
14. Selezionare con Punto di interruzione Confermare con me disattivato 7 Attivare o disattivare allarme YES o NO Selezionare con Confermare con me 8 Programma pronto SET P2A 2 00 h SET P2 OFF 1000 V 50mA 50 W 1000 V 50mA 50 W Fig 3 Riepilogo passo dopo passo di programmazione AT EPS 1001 Manuale dell operatore Fig 4 Disabilitazione del controllo della corrente 1 Inserire MORE SET P2A 1000 V 50mA Selezionare MORE 1 con 2 Selezionare il programma MORE P2 na Selezionare programma Disabilitando il Selezionare con 6 U controllo della corrente di avviamento l alimentatore pu 3 Attivare o disattivare il controllo fornire alta tensione della corrente anche se non e collegato all attrezzatura per elettroforesi Confermare con Controllo della corrente di avviamento YES no A W Selezionare con Confermare e uscire con 4 Tornare al modo SET SET P2A 2 00 h 1000 V 50 mA 50 W 1 Selezionare il numero del programma SET P1 OFF 200 V 40mA 10W Selezionare con 2 Avviare le operazioni SET P24 2 00h 1000 V 50mA 60W Avviare con RUN P24 0 01h 300 V 50mA 25W RUN P2 00 30h 500V 50mA 25W oppure attendere 5 secondi 3 Durante il funzionamento possibile visualizzare le impostazioni E anche 3 n 1 possibile interrompere il P SET R2 2 00h funzionamento e modificare il 1000 V
15. changent automatiquement du maximum au minimum ou vice versa RUN Ex cution Une pression sur la touche RUN d clenche l ex cution du programme et met le programme en mode RUN Le num ro du programme les valeurs effectives de la tension du courant et de la puissance s affichent sur l cran Le temps coul ou le nombre de volt heures s affichent aussi Pour visualiser le temps ou les volt heures utiliser les touches y PAUSE CONTINUE Une pression sur cette touche met l instrument en mode PAUSE et coupe la tension L cran d affichage indique les diff rents param tres au moment o l op rateur a appuy sur la touche La touche PAUSE CONTINUE n est active qu en mode RUN Le temps et les valeurs int gr es sont maintenues En mode PAUSE on peut utiliser la touche SET pour effectuer des modifications dans le programme Pour repasser en mode RUN appuyer sur la touche PAUSE CONTINUE ou sur RUN EPS 1001 Manuel d Utilisation STOP ST OP Arr t Une pression sur la touche STOP interrompt l ex cution du programme en cours et met l instrument en mode EN D Fin La tension est coup e et les param tres de fin de programme sont affich s L ex amp cution du programme ne peut pas tre reprise son point d interruption apr s une pression sur la touche 4 7 MORE Suite Une pression sur cette touche permet de d sactiver la mise sous tension apr s v rification du courant qui normalement d tectent si la r sistance
16. de red pulsar para conectar la electricidad de la fuente de alimentaci n pulsar O para desconectarla 2 Un enchufe para el cable de red 3 Un conmutador de intervalo de tensi n la posici n izquierda corresponde a 100 120 V y la derecha a 220 240 V 4 Respiraderos ES EPS 1001 Manual del usuario 5 Operaci n 5 1 Resumen La operaci n de la fuente de alimentaci n EPS 1001 queda descrita brevemente en esta secuencia 1 Programaci n o edici n de un m todo Ver las figuras 3 y 4 2 Ejecuci n de un m todo Ver la figura 5 5 2 Instrucciones resumidas En este cap tulo se resumen los puntos m s importantes para el manejo correcto de la EPS 1001 Ver tambi n la gu a de manejo esquem tica entregada con el equipo Se recomienda mantener la gu a junto al mismo 1 Conectar a la red La pantalla debe mostrar el n mero de programa intermitentemente 2 Pulsar la tecla ENTER del n mero de programa actual o seleccionar el ENTER n mero de programa deseado con las teclas Y Pulsar ENTER 3 Para cada uno de los par metros de tensi n amperaje y potencia pulsar Y N hasta alcanzar el valor previsto Confirmar con la tecla ENTER despu s de cada selecci n la pantalla vuelve autom ticamente al par metro siguiente 4 Elegir entre cambio manual o autom tico Para el cambio autom tico seleccionar las unidades del evento V oltioshora Vh o tiempo h Utilizar las teclas 4 7 para la selecci n de unidades
17. dieses Handbuch sorgf ltig lesen unbedingt eingehalten werden mu um Verletzungen zu verhindern Sorgen Sie daf r da Sie erst dann fortsetzen wenn Sie die Anleitungen eindeutig verstanden haben und wenn alle erforderlichen Sicherheitsma nahmen getroffen wurden WARNUNG Das Z eichen WARNUNG deutet auf eine Anweisung hin die Konformit tserkl rung Sicherheitsnormen Dieses Produkt entspricht den Anforderungen gem der N iederspannungsvorschriften LVD 72 23 EEC und zwar im Rahmen der angepa ten Norm EN 61010 1 1993 A1 1992 EM C N ormen EMC Electro magnetic C ompatibility Dieses Produkt entspricht den Anforderungen gem der EM C Vorschriften 89 336 EEC undzwar im Ramhen der angepa ten N orm EN 50081 1 Strahlungsabgabe und EN 50082 1 Unempfindlichkeit Das CE Symbol und die sich dadurch ergebende Erkl rung bez glich Einhaltung der Vorschriften ist nur dann f r das Instrument g ltig wenn es auf folgende W eise verwendet wird Es wird als alleinstehendes Ger t verwendet Es wird an andere Instrumente der Firma Amersham Pharmacia Biotech die mit einem CE Symbol gekennzeichnet sind angeschlossen oder Es wird an andere Produkte angeschlossen die in diesem H andbuch empfohlen oder beschrieben werden und Es wird in jenem Z ustand verwendet in dem es von der Firma Amersham Pharmacia Biotech AB geliefert wurde abgesehen von jenen Anderungen die i
18. est trop lev e et coupent la tension lorsque c est le cas EXIT Sortie Une pression sur cette touche permet d arr ter l ex cution de l op ration en cours comme la saisie d une valeur par exemple A pr s une pression sur cette touche seules les valeurs unites qui ont d j t confirm es par une pression sur la touche ENTER vont tre gard es De plus le g n rateur est replac dans le mode interrompu ou dans le mode D marrage de la touche SET 4 1 3 Prises connecteurs de sortie L instrument comporte deux jeux de connecteurs pour permettre le branchement et l utilisation simultan e de deux unit s d lectrophor se Figure 1 voir verso de la page de couverture La tension de sortie se situe entre 0 1000 V Le connecteur n gatif est un potentiel compris entre 0 et 500 V et le connecteur positif est un potentiel compris entre 0 et 500 V 4 2 Panneau arri re Le panneau arri re est repr sent sur la Figure 2 voir verso de la page de couverture Sur le panneau arri re sont log s 1 Un commutateur d alimentation secteur Basculer ce commutateur en position l pour mettre l instrument sous tension Le basculer en position O pour mettre l instrument hors tension 2 Un r ceptacle pour le c ble d alimentation 3 Un commutateur de tension secteur de l instrument La position gauche correspond 100 120 V et la position droite 220 240 V 4 Des ouvertures d a ration FR EPS 100
19. innesota Street PO Box 77387 San Francisco CA 94107 USA N vert aux Etats Unis 800 227 4750 T l phone local 415 282 2307 Fax 415 821 1081 T lex 470778 Adresses des bureaux Amersham Pharmacia Biotech AB SE 751 84 Uppsala Sweden Amersham Pharmacia Biotech UK Limited Amersham Place Little Chalfont Buckinghamshire England HP7 9NA Amersham Pharmacia Biotech Inc 800 Centennial Avenue P O Box 1327 Piscataway N J 08855 1327 USA FR EPS 1001 Manuel d Utilisation 1 Introduction L unit d alimentation lectrique d lectrophor se EPS 1001 produite par Amersham Pharmacia Biotech est un instrument de haute pr cision de grande qualit et fonctionnant en s curit maximale lorsqu elle est utilis e pour des applications d lectrophor se qui n cessitent un maximum de 1000 volts de 100 Watts et de 400 mA L EPS 1001 est destin e en premier lieu aux techniques d application suivantes SDS PAGE lectrophor se en Gel de Polyacrylamide SDS PAGE Native Electrophor se en Gel de PolyAcrylamide L lectrophor se A garose L lectrotransfert L lectrophor se d ADN en champ puls Electrophor se H aute R solution des produits d amplification ADN L lectrophor se sous marine des acides nucl ides ADN ARN NOU BRB WN HG Les s parations lectrophor tiques peuvent tre contr l es par la tension le courant ou la puissance L EPS 1001 commute automatiqueme
20. principale Selezionare la tensione appropriata 100 120 o 220 240 V Fig 2 vedere prima di copertina Avvertenza Se l alimentatore viene collegato ad una tensione di 220 240 V mentre impostato su 100 120 V lo strumento pu subireseri danni Selezionare un cavo di rete idoneo e collegarne un estremit alla presa di rete dell alimentatore EPS 1001 vedere Fig 2 el altra ad una presa CA con messa a terra Accendere l unit Ad ogni accensione lo strumento effettua un test di autodiagnosi Se durante il test viene rilevato un errore sul display apparir un messaggio e verr emesso un allarme acustico Collegamento alla e unit per elettroforesi Collegare i cavi dell alimentatore per elettroforesi rosso con rosso e nero o blu con blu Fig 1 vedere prima di copertina Il cavo rosso positivo ed il cavo nero o blu negativo Avvertenza Utilizzare esclusivamente cavi elettrici non danneggiati ed attrezzature specificate per la tensione utilizzata Con lo stesso programma possono essere utilizzate due unit per elettroforesi contemporaneamente In questo caso si prega di ricordare di raddoppiarei limiti di corrente e potenza La tensione non varia a seconda del numero di unit utilizzate Ge EPS 1001 Manuale dell operatore VIT pause continue 4 Descrizione tecnica 4 1 Pannello anteriore Sul pannello anteriore si trovano un display alfanumerico una tastiera con 9 tasti a memb
21. showing the elapsed time or volthours by 4 7 The limiting parameter will attain the programmed limit The unit for that parameter is underlined A light emitting diode shows when voltage is applied HV on If no current is displayed please check the electrical connections to the electrophoresis equipment Pausing Y ou can interrupt the electrophoresis for sample loading and or changing parameter values by pressing PAUSE CONTINUE Voltage and current will no longer be supplied the HV on LED goes off and you may safely load your samples The display shows the status of the run when PAUSE CONTINUE was pressed Switch between time and integrated voltage by pressing 4 9 When sample loading is complete press PAUSE CONTINUE again or RUN and the run will continue from where it was interrupted When in the PAUSE mode you can also press SET to make changes in the program This modeis called the P SET mode When the P SET mode is entered you can make changes as described under programming The P SET modeis the same as the SET mode apart from restrictions in setting the breakpoint N aturally it is not possible to enter a time or volthour that is already passed Press EXIT or PAUSE CONTINUE to go back from P SET to PAUSE Press RUN or PAUSE CONTINUE to proceed with the electrophoresis View programmed values It is also possible to view the programmed values during a run by pressing SET N ote that no values can be changed here The display returns auto
22. tes stets korrekt am N etzausgang geerdet sein 3 Um eine ausreichende K hlung zu erm glichen ist zu gew hrleisten da die Bel ftungs ffnungen hinten und an den Seiten des Ger tes nicht verdeckt sind 4 Das Ger t nicht bei hoher Luftfeuchtigkeit ber 95 betreiben Kondensation vermeiden indem man das Ger t nach einem Transport von einer k lteren in eine w rmere Umgebung sich stets auf Umgebungstemperatur erw rmen l t 5 Das Ger t so trocken und sauber wie m glich halten Regelm ig mit einem weichen und feuchten Tuch abwischen Vor dem Gebrauch das Stromversorgungsger t vollst ndig trocknen lassen Vor dem Befeuchten stets den N etzstecker des Ger tes ziehen und ihn erst dann wieder einstecken wenn es trocken ist 6 Stets intakte elektrische Kabel und Ausr stung verwenden die f r diezu verwendenden Spannungen geeignet sind H ochspannungskabel sollten der elektrischen Sicherheitsnorm IEC 1010 2 031 1993 entsprechen S mtliche an die Stromversorgung angeschlossene Elektrophorese A usr stung sollte der elektrischen Sicherheitsnorm IEC 1010 1 1993 entsprechen 7 Man beachte da der Ausgang an Geh use R eferenzerde angeschlossen ist EPS 1001 Bedienungsanleitung DE 2 2 Eingebaute SicherheitsmaBnahmen Das M odell EPS 1001 wurde gem der elektrischen Sicherheitsnorm IEC 61010 1 EN 61010 1 gepr ft Das Stromversorgungsger t verf gt auch ber mehrere eingebaute Schutzfunktionen Die folgende
23. the same time The voltage will be the same regardless of the number of units GB EPS 1001 User Manual Local regulation for Great Britain WARNING IMPORTANT This appliance must be earthed The wires in the mains lead are coloured in accordance with the following code Green and yellow Earth Blue N eutral Brown Live If the plug provided is unsuitable for socket outlets the plug must be cut off and suitable plug fitted The cut off plug should be disposed of and must not be inserted into any 13 amp socket as this can result in electric shock The plug or adapter or the distribution panel should be provided with 13 amps fuse As the colours of the wires in the mains lead of this appliance may not correspond with coloured markings identifying the terminals in your plug proceed as follows The green and yellow wire must be connected to the terminal in the plug which is marked with the letter E or by the earth symbol or coloured green or green and yellow The blue wire must be connected to the terminal which is marked with the letter N or coloured black The brown wire must be connected to the terminal which is marked with the letter L or coloured red Note After replacing or changing a fuse the fuse cover in the plug must be replaced with a fuse cover which corresponds to the colour of the insert in the base of the plug or the word that is embossed on the base of the plug and the appliance must not be used without a fu
24. the user occurs either during a run or when switching on the power supply the program enters the HALT mode and the output is switched off Four different type of errors can cause HALT Thefollowing list shows the error message on the display the cause and the remedy Error message Cause Remedy HALT Low start The current is less than the lower limit 1 Check connections and or buffers This can be due to incorrect I n current connection of the electrophoresis 2 Press MORE Switch off the start equipment or due to use of buffers with current check in the MORE mode extremely low conductivity IMPORTANT Read Section 5 4 first HALT Ground The current to ground leakage in the Check the electrophoresis unit leakage current electrophoresis unit is too high HALT Mains The mains voltage is too low see Check voltage selector Voltage too low Specifications Check mains voltage HALT Mains fail Mains power failure for more than 8 s Press PAUSE CONTINUE to continue Program stopped running a program or STOP to interrupt the run If a serious error occurs the program enters the FAIL mode The output is switched off and an error message is shown in the display FAIL Code No xxx Call service Please read the error code number and contact your Amersham Pharmacia Biotech representative 12 EPS 1001 User Manual GB 8 Technical specifications Regulation Output mode Pr
25. 01 von A mersham Pharmacia Biotech ist ein qualitativ hochwertiges u erst genaues und sicheres Stromversorgungsger t f r Elektrophoreseanwendungen die eine Stromversorgung von maximal 1000 Volt 100 Watt und 400 mA ben tigen Das M odell EPS 1001 wurde vornehmlich f r die folgenden Arbeitsverfahren entwickelt 1 SDS PAGE Polyacrylamidgel Elektrophorese NativePAGE Agarose Elektrophorese Elektroblotting DNA Pulsfeld Elektrophorese DNA Fragmenttrennung 7 DNA Submarine Elektrophorese ou RU NN Elektrophoresetrennungen k nnen mit Spannung Strom oder Leistung gesteuert werden Die Steuerparameter mA V W werden in Abhangigkeit von den einprogrammierten Grenzwerten sowie von der Leitfahigkeit wahrend des Laufes automatisch vom EPS 1001 angepa t Es konnen drei Programme gespeichert werden 2 Sicherheitsangaben 2 1 Sicherheitsma nahmen Da das Gerat gen gend Spannung und Strom entwickeln kann um einen t dlichen elektrischen Schlag zu verursachen wird bei dessen Betrieb um u erste V orsicht gebeten Zur Vermeidung von V erletzungsgefahren sollte das Ger t nur von entsprechend geschultem Personal und stets gem den beiliegenden Anweisungen betrieben werden Das Handbuch vor der Benutzung dieses Stromversorgungsger tes vollst ndig durchlesen 1 Das Ger t ist nur f r den Gebrauch in geschlossenen R umen geeignet 2 Der Schutzerdleiter des Stromversorgungskabels mu beim Gebrauch des Ger
26. 1 Manuel d Utilisation 5 Utilisation 5 1 G n rali s Le fonctionnement de l EPS 1001 est d crit dans l ordre suivant 1 Programmation ou s lection d une proc dure voir Figures 3 et 4 2 Ex cution d une proc dure voir Figure 5 5 2 Instructions simplifi es Cette section est un r sum des principaux points retenir pour l utilisation de l EPS 1001 N ous vous recommandons aussi de vous r f rer au sch ma du guide d utilisation fourni avec l unit d alimentation lectrique Conserver ce guide d utilisation pr s de l instrument 1 Mettre l instrument sous tension ON Le num ro de programme doit clignoter sur l cran d affichage 2 Appuyer sur la touche ENTER pour valider le num ro de programme m moris ou utiliser les touches Y pour s lectionner le num ro de programme souhait A ppuyer sur la touche ENTER 3 Pour chacun des param tres Tension Courant et Puissance appuyer sur une des touches 4 7 jusqu ce que la valeur s lectionn s affiche Confirmer la s lection en appuyant sur la touche EN TER apr s chacune de ces valeurs et l cran affiche automatiquement le param tre suivant 4 S lectionner entre param tres de fin manuel ou automatique Pour le param tre de fin automatique s lectionner les unit s soit en nombre Volt heures Vh soit en dur e h S lectionner l unit appropri e ou si vous pr f rez la version manuelle choisir OFF arr t automatique d sactiv l a
27. 4 Description technique 4 1 Panneau avant Le panneau avant comporte un cran d affichage alphanum rique un clavier avec 9 touches une diode lectroluminescente LED qui s allume lorsque l instrument est mis sous tension Sortie HT active et des connecteurs pour deux unit s d lectrophor se 4 1 1 cran d affichage Un cran d affichage 32 caract res alphanum riques vous permet de suivre les diff rentes phases de programmation affiche les valeurs actuelles des param tres pendant l ex cution d lectrophor se et les valeurs finales des param tres Il affiche aussi des questions et des messages d erreur L cran d affichage comporte une barrette sup rieure et une barrette inf rieure La Figure 1 voir verso de la page de couverture montre l cran initial lorsque l interrupteur d alimentation est mis en position M arche On Le mode dans le cas pr sent SET c d programmation est affich en haut gauche et le num ro du programme au centre de la barrette sup rieure Le num ro du programme indiqu est celui du programme pr c dent Le clignotement du chiffre en l occurrence 1 indique qu il est possible de le modifier en appuyant sur une des touches T En haut droite s affiche le param tre de fin programme pour l op ration d lectrophor se Dans le cas pr sent O ff Rupture M anuelle En bas gauche au centre et droite sont indiqu es respectivement la tension le courant et la pui
28. Electrophoresis Power Supply EPS 1001 User Manual Manuel d Utilization Bedienungsanleitung Manual del usuario Manuale dell operatore 18 1130 20 Edition AB amersham pharmacia biotech amersham pharmacia biotech O O Program 0 0 00 O A 96 000 0 Electrophoresis Power Supply EPS 1001 J L Fig 1 The front panel of the EPS 1001 Figure 1 Le panneau avant de l EPS 1001 Abb 1 Die Fronttafel des EPS 1001 Figura 1 Panel frontal de la EPS 1001 Figura 1 Pannello anteriore dell EPS 1001 Voltage 100 120 220 240 V AC O C 2 Frequency 50 60 Hz Power 140W 2 5AMax C gt C J b C C D 2 indi dd servicing C 2 c 2 BI C0 9 all Olle C di gt 0 C IE gt 100 120 LIO 220 240 C IL 5 C 2 JO y L L Fig 2 The rear panel of the EPS 1001 Figure 2 Le panneau arri re de l EPS 1001 Abb 2 Die R ckwand des EPS 1001 Figura 2 Panel posterior de la EPS 1001 Figura 2 Pannello posteriore dell EPS 1001 EPS 1001 User Manual Contents INErOdUEHON 26256 tage weceeien ens 2 Safety information air 2 2 1 Safety precautions 2 2 2 In built safety features 3 Unpacking and installation 3 Technical description 4 4 1 Front panels 4 4
29. Eomprating 1 MORE eingeben SET P2A 1000 V 50 mA d MORE w hlen mit 2 Programm w hlen MORE Warnung Select e Wenn die Funktion W hlen mit Start Current Check gesperrt ist dann ist es m glich da an dem Best tigen mit Current check at Stromversorgungsger t 3 Strompr fung ja nein w hlen start YES no selbst dann H och spannung anliegt wenn W hlen mit es nicht an das Elektro Best tigen und A O phoreseger t Verlassen mit angeschlossen ist 4 R ckkehr zum SET Modus SET p2A 2 00 h 1000 V 50 mA 50W 1 Programmnummer w hlen SET P1 OFF 200V 40mA 10W A U SET P24 2 00h 1000 V 50mA 60W Starten mit W hlen mit 2 Programmlauf starten RUN 300 V 50 mA 0 01 h 25W RUN P2 20 30h 500V 50mA 25W EJ oder 5 s warten SET pas 1000 V 50 mA 2 00 h 60 W PAUSE 0 30 h 500V 50mA 100W Der Benutzer kann w hrend eines Programmlaufs die Einstellungen ablesen Er kann den Lauf auch P SET P2 2 00h unterbrechen und nderungen im 1000 V 50mA 60W Programm vornehmen Programmierung 4 Programmlauf manuell stoppen RUN pas 1000 V 35 mA 1 45h 35 W END pas 1000 V 35 mA 1 45h 35 W END 1000 V 20 mA 2 00 h 20 W 5 Programmlauf wird automatisch gestoppt Abb 5 Schrittweise Zusammenfassung von
30. LVD Basse tension par la norme harmonis e EN 61010 1 1993 A1 1992 NormesEMS EMC Electromagnetic C omaptibility Ce produit est conforme la directive EM C 89 336 EEC par la norme harmonis e EN 50081 1 mission et EN 50082 1 immunit Le symbole CE et la d claration de conformit correspondante s appliquent l instrument lorsqu il est utilis comme unit seule ou connect d autres instruments Amersham Pharmacia Biotech marqu s CE ou connect d autres produits recommand s ou d rits dans ce manuel et utilis dans le m me tat que lors de sa livraison par Amersham Amersham Pharmacia Biotech AB l exception des modifications d crites dans ce manuel Modalit s et Conditions de Vente Toutes les marchandises et prestations sont vendues sous r serve des modalit s et conditions de vente dela soci t membre du groupe Amersham Pharmacia Biotech qui les fournit Une copie de ces modalit s et conditions est disponible sur demande N ous vous sommes gr d adresser tout commentaire que vous auriez sur ce produit Amersham Pharmacia Biotech Inc 654 M innesota Street San Francisco CA 94107 USA Marques de fabrique Amersham est une marque d pos e de N ycomed A mersham plc Pharmacia et Drop Design sont des marques d pos es de Pharmacia amp Upjohn Inc Copyright 1998 Amersham Pharmacia Biotech AB Tous droits r serv s H oefer Pharmacia Biotech Inc 654 M
31. NTER Introduce un valor lo fija si es correcto y mueve la programaci n al campo siguiente Los valores v lidos son tensi n 5 1000 V amperaje 1 400mA potencia 1 100 W tiempo 0 01 500 h voltioshora 1 500 000 Vh SET Pone el equipo en la modalidad SET modo de programaci n y arranque Pulsando la tecla SET en la modalidad RUN funcionamiento se visualizan los par metros activos A dem s la tecla SET permite efectuar cambios en el programa durante el funcionamiento despu s de pulsar la tecla PAUSE CONTINUE Change up Change down Flecha ascendente y flecha descendente Cambian el valor par metro u otras variables en el campo intermitente Los valores se cambian de forma acelerada manteniendo oprimida una tecla y en incrementos prefijados con pulsaciones sucesivas El tipo de par metro o unidad por ejemplo Vh y las alternativas Y ES N O se cambian con una pulsaci n Las teclas tambi n pueden usarse para el cambio entre tiempo miliamperhoras y voltioshora en las modalidades RUN PAUSE y END Los valores se desplazan en pantalla cambian autom ticamente entre m ximo y m nimo y viceversa RUN Pulsando la tecla RUN seinicia el proceso y se pone el programa en modalidad RUN La pantalla muestra el n mero de programa los valores actuales de tensi n amperaje y potencia as como el tiempo transcurrido voltioshora o miliamperhoras el cambio entre estos tres ltimos par metros se hace con las teclas 4 1 PA
32. P2 0 30h 500V 50mA 25W SET pas 2 00h 1000 V 50mA 60W PAUSE P2 0 30h 500V 50mA 100W o esperar 3 Durante el funcionamiento pueden verse los ajustes El funcionamiento x puede interrumpirse para efectuar P SET P2 2 00h modificaciones en el programa 1000 V 50mA 60W 4 Parar el funcionamiento manualmente END P2A 41 45h 1000 V 35mA 35W RUN P24 1 45h 1000V 35mA 35W 5 Elfuncionamiento se para autom ticamente END P2 2 00 h 1000 V 20mA 20W Figura 5 Sumario paso a paso de un proceso con pausa EPS 1001 Manuale dell operatore IT Indice 1 Introduzione isis kr da 2 4 1 1 DISPO sez rares 4 Informazioni sulla sicurezza 2 4 1 2 Tastiera 4 2 1 Norme di sicurezza 2 4 1 3 Prese di uscita 5 2 2 Funzioni di sicurezza incorporate 3 4 2 Pannello posteriore 5 3 Disimballaggio ed installazione 3 5 Funzionamento 6 4 Descrizionetecnica 4 5 1 Vista generale 6 4 1 Pannello anteriore 4 5 2 Brevi istruzioni 6 Informazioni importanti per l operatore Leggere attentamente il manuale per comprendere completamente l uso dell EPS 1001 triangolo equilatero indica all operatore la presenza di importanti istruzioni di funzionamento e manutenzione nella AVVERTENZA Il punto escl
33. USE CONTINUE Pone el aparato en modalidad de pause y desconecta la tensi n La pantalla muestra el estatus del proceso en el momento de pulsarse la tecla La tecla PAUSE CONTINUE s lo funciona en modalidad de RUN Se retienen los valores de tiempo e integraci n En la modalidad de pausa la tecla SET puede usarse para introducir modificaciones en el programa Para volver a la modalidad RUN pulsar PAUSE CONTINUE o RUN EPS 1001 Manual del usuario ES STOP STOP Interrumpe el funcionamiento y pone el aparato en modalidad EN D Se desconecta la tensi n y se visualizan los par metros finales Un funcionamiento interrumpido con la tecla STOP no puede continuar MORE Inhabilita el control de la corriente de arranque que detecta si la resistencia es demasiado alta y desconecta la electricidad EXIT Para la ejecuci n de una operaci n como la introducci n de un valor S lo se retienen los valores o unidades ya confirmadas con la tecla ENTER Devuelve el equipo a la modalidad abandonada o a la posici n de arranque en SET 4 1 3 Tomas de salida Hay dos tomas de salida para la conexi n y funcionamiento simult neo de dos unidades electrofor ticas la figura 1 ver interior de la cubierta La tensi n saliente es de 0 1000 V El enchufe negativo da entre 0 y 500 V y el positivo entre 0 y 500 V 4 2 Panel posterior En el panel posterior ilustrado en la figura 2 ver interior de la cubierta hay 1 Un interruptor
34. allows the user to set limits for voltage current and power During electrophoresis only one of the parameters is limiting at a time The limiting parameter determines together with the conductivity in the electrophoresis system the values for the other two parameters Voltage current power and conductivity are related by the following equations U 1 L 1 P Uxl 2 Where U Voltage Current P Power and L Conductivity Equation 1 is more familiar if the conductivity is replaced by the reciprocal resistance R U R x Ohm s law The electric field E measured in V cm is the driving force behind electrophoresis E U d where E Electrical field strength U Voltage d distance The electrical field strength is achieved by applying a voltage The higher the voltage the faster the electrophoresis Fast electrophoresis is beneficial since it counteracts diffusion The disadvantage of increasing the voltage too much is that most of the generated electrical energy the product of power and time is transformed to heat Therefore cooling of electrophoresis equipment is recommended Cooling will also reduce smiling effects which are caused by mobility differentials across an electrophoresis gel resulting from poor heat transfer Since the cooling efficiency cannot be increased indefinitely the power should be limited when programming the power supply EPS 1001 User Manual GB The parameter that should be chosen
35. alta tensi n deben satisfacer los requisitos de la norma el ctrica IEC 1010 2 031 1993 Todos los equipos electrofor ticos que se conecten a la EPS 1001 deben satisfacer los requisitos de la norma el ctrica IEC 1010 1 1993 7 T ngase en cuenta que la salida se conecta en el chasis masa de referencia 2 2 Funciones de seguridad incorporadas La fuente de alimentaci n EPS 1001 est homologada seg n la norma de seguridad el ctrica IEC 61010 1 EN 61010 1 Adem s incorpora varias funciones de seguridad y mensajes de error que se indican en la pantalla EPS 1001 Manual del usuario ES 1 Fuga a tierra funcional Si la fuente de alimentaci n se conectara a una unidad electrofor tica que tiene fuga a tierra el fallo es detectado y se desconecta la alta tensi n 2 Control de la corriente de arranque Para asegurar la conexi n correcta de la unidad electrofor tica la fuente de alimentaci n controla que la resistencia no supere un l mite especificado en una tensi n de seguridad baja lt 40V Si esta resistencia es demasiado alta la tensi n se desconecta Una resistencia demasiado alta tambi n puede deberse a la utilizaci n de tampones con conductividad excesivamente baja En este caso la alta tensi n tambi n se desconecta y se visualiza un mensaje de error Esta funci n puede inhabilitarse para ciertas aplicaciones ver la figura 4 Inhabilitaci n del control de corriente 3 Detecci n de camb
36. amativo all interno di un documentazione allegata al prodotto Dichiarazione di conformit Standard di sicurezza Questo prodotto conforme alla Direttiva sulla bassa tensione LVD 72 23 EEC in base alla norma armonizzata EN 61010 1 1993 A1 1992 Standard EMC EMC Electromagnetic Compatibility Il prodotto conforme alla direttiva sulla compatibilit elettromagnetica 89 336 EEC in base alle norme armonizzata EN 50081 1 emission en EN 50082 1 immunit Termini e condizioni di vendita Tutti i prodotti ed i servizi sono soggetti alle condizioni di vendita dell azienda del gruppo Amersham Pharmacia Biotech di provenienza Una copia dei termini e delle condizioni disponibile su richiesta Saremo lieti di ricevere eventuali commenti sul prodotto al seguente indirizzo Amersham Pharmacia Biotech Inc 654 M innesota Street San Francisco CA 94107 USA Marchi registrati Amersham un marchio registrato della N ycomed Amersham plc Pharmacia e Drop Design sono marchi registrati della Pharmacia amp Upjohn Inc Copyright 1998 Amersham Pharmacia Biotech AB Tutti i diritti riservati Indirizzi degli uffici Amersham Pharmacia Biotech AB SE 751 84 Uppsala Sweden Amersham Pharmacia Biotech UK Limited Amersham Place Little Chalfont Buckinghamshire England HP7 9NA Amersham Pharmacia Biotech Inc 800 Centennial Avenue P O Box 1327 Piscataway N J 08855 1327 USA IT EPS 1001
37. as separaciones electrofor ticas pueden controlarse por tensi n amperaje y potencia La EPS 1001 conmuta autom ticamente el par metro de control de acuerdo a los l mites programados y a las variaciones de conductividad del sistema Pueden almacenarse 3 programas 2 Seguridad 2 1 Medidas de seguridad La fuente de alimentaci n debe manejarse con suma precauci n puesto que puede desarrollar tensi n y amperaje suficiente para producir descargas mort feras Para evitar riesgos de lesiones el aparato s lo debe manejarlo personal adecuadamente formado y siempre siguiendo las instrucciones adjuntas 1 El aparato est dise ado para uso exclusivo en interiores 2 El aparato debe utilizarse con el hilo de masa del cable de energ a correctamente conectado a la tierra de la toma de corriente 3 Losrespiraderos laterales y posteriores del aparato deben mantenerse destapados para permitir una ventilaci n suficiente 4 Noutilizar el aparato en condiciones de humedad extrema m s del 95 Evitar la condensaci n dejando que la unidad se equilibre a la temperatura ambiente al trasladarla desde un ambiente m s caliente 5 Mantener el aparato tan seco y limpio como sea posible Limpiarlo regularmente con un trapo suave h medo Dejar que se seque del todo antes de usarlo Si se moja desenchufarlo hasta que se seque 6 Utilizar solamente cables el ctricos intactos y equipos especificados para las tensiones a usar Los hilos de
38. as the constant and thus control the electrophoresis depends on the type of electrophoresis In the case of homogeneous buffers throughout the system same electrode and gel buffer the conductivity is constant during the electrophoresis If the conductivity is constant the voltage will be proportional to the current and the power to the square of the current according to 1 and 2 This means that the result of the electrophoresis will be the same regardless which parameter is chosen for constant For historical and practical reasons voltage is most commonly used for regulation Submarine gel electrophoresis of DNA RNA and pulsed field electrophoresis are usually run at constant voltage SDS PAGE using continuous buffer systems is run at constant voltage or current For discontinuous buffer systems the resistance will increase as the electrophoresis proceeds due to a decrease in conductivity Running at constant voltage will result in decreasing current and power Constant voltage will thus be safe in the respect that the power will not increase and produce more and more heat On the other hand the separation will slow down and impair resolution due to an increased time available for diffusion Running at constant power would give a faster electrophoresis and controlled power while running at constant current would at the first sight seem to be problematic because of increasing voltage and power During discontinuous electrophoresis h
39. azioni Selezionare Y ES o NO con 4 ft e premere ENTER La selezione di Y ES viene indicata sul display da una piccola A sulla destra del numero del programma cio 1A Una volta completata la programmazione collegare l unit per elettroforesi alle prese e premere RUN Sul display vengono visualizzati gli attuali valori di tensione corrente e potenza insieme a tempo trascorso o voltore commutare con 4 1 PAUSE CONTINUE puo essere utilizzato per interrompere e riprendere il funzionamento 10 Durante il funzionamento i parametri programmati possono essere controllati premendo SET 11 Il funzionamento si interrompe automaticamente o premendo STO P EPS 1001 Manuale dell operatore IT caratteri lampeggianti vengono illustrati in neretto 1 Selezionare il numero SET P1 del programma 200 V 40 mA Selezionare con Confermare con 2 Selezionare il limite di tensione P2 OFF 400 mA 100 W Selezionare con Confermare con 3 Selezi il limite di elezionare il limite di SET P2 OFF corrente 1000 V 400 mA 100 W Selezionare con 0 Q Confermare con 4 Selezionare il limite di SET P2 potenza 1000 V 50 mA Selezionare con Confermare con 5 Attivare il punto di interruzione h o Vh o disattivare il punto di interruzione OFF SET 1000 V P2 50 mA Selezionare con Confermare con 6 Selezionare il punto di interruzione SET 1000 V P2 50mA
40. ble L essuyer r guli rement avec un chiffon doux humide Veiller ce que l unit soit compl tement s che avant de l utiliser Lorsque l unit est humide la d brancher jusqu ce qu elle soit s che 6 N employer que des fils lectriques compl tement intacts et des quipements pr vus pour les tensions que vous utilisez Les fils haute tension doivent tre conformes la norme lectrique IEC 1010 2 031 1993 Tout quipement d lectrophor se branch l unit d alimentation lectrique doit tre conforme la norme lectrique IEC 1010 1 1993 7 Noter que la sortie est branch e la terre du ch ssis r f rence EPS 1001 Manuel d Utilisation 2 2 Caract ristiques de s curit intrins ques La conformit de l EPS 1001 la norme de s curit lectrique IEC 61010 1 EN 61010 1 a t certifi e Cette unit d alimentation lectrique est aussi dot e de fonctions de s curit intrins ques Des messages d erreur sont affich s l cran 1 Functional earth leakage Fuite la terre fonctionnelle Lorsque l unit d alimentation lectrique est branch e sur une unit d lectrophor se qui a des fuites la terre l EPS 1001 d tecte cette d faillance et coupe imm diatement la haute tension 2 Start current check V rification du courant initial Pour s assurer que l unit d lectrophor se est branch e correctement l unit d alimentation lectrique v rifie basse tension de s cur
41. chzeitig betrieben werden Die Spannung ist unabh ngig von der Anzahl der angeschlossenen Ger te gleich 4 Technische Beschreibung 4 1 Fronttafel Die Fronttafel besteht aus einer alphanumerischen Anzeige einer Tastatur mit 9 Folientasten einer Leuchtdiode LED die aufleuchtet wenn Spannung anliegt H V on sowie Anschl ssen f r zwei Elektrophoreseger te 4 1 1 Anzeige Eine 32 stellige alphanumerische Anzeige f hrt den Benutzer durch die Programmierung zeigt aktuelle Parameterwerte w hrend der Elektrophorese sowie Endparameterwerte danach Sie stellt auch Fragen und zeigt Fehlermeldungen an Die Anzeige hat eine obere und eine untere Reihe Abb 1 siehe Deckblattinnenseite zeigt die Anzeige in der Startposition wenn der Strom eingeschaltet ist Der M odus in diesem Fall SET wird in der oberen Reihe auf der linken Seite angezeigt Die Programmnummer befindet sich in der M itte der oberen Reihe Die angezeigte N ummer ist die N ummer des Programms in das der Benutzer zuvor gegangen ist Die blinkende Z ahl in diesem Fall 1 zeigt an da sie durch Dr cken der Tasten 4 9 ge ndert werden kann Die obere Reihe auf der rechten Seite zeigt den programmierten Unterbrechungspunkt Diese Funktion ist in diesem Falle abgeschaltet Die Positionen unten links in der M itte und rechts zeigen jeweils die Spannung den Strom und die Leistung an 4 1 2 Tastatur ENTER Durch D r cken dieser Taste wird ein Wert eingegeben de
42. ctos nos complacer recibirlos en Amersham Pharmacia Biotech Inc 654 M innesota Street San Francisco CA 94107 USA Marcas Registradas Amersham es marca registrada de N ycomed Amersham plc Pharmacia y Drop Design son marcas registradas de Pharmacia amp Upjohn Inc 4 1 1 Pantalla 5i sae murs 4 2 1 2 Tela d casa EE as 4 4 1 3 Tomasdesalida 5 4 2 Panel posterior 5 o Stains es imet iiti 6 5 1 Resumen sad criada 6 5 2 Instrucciones resumidas 6 Copyright 1998 Amersham Pharmacia Biotech AB Reservados todos los derechos Direcciones de oficinas Amersham Pharmacia Biotech AB SE 751 84 Uppsala Sweden Amersham Pharmacia Biotech UK Limited Amersham Place Little Chalfont Buckinghamshire England HP7 9NA Amersham Pharmacia Biotech Inc 800 Centennial Avenue P O Box 1327 Piscataway N J 08855 1327 USA ES EPS 1001 Manual del usuario 1 Introducci n La fuente de alimentaci n EPS 1001 de Amersham Pharmacia Biotech es un aparato seguro de gran calidad y precisi n para aplicaciones de electroforesis que requieren un m ximo de 1000 V 100 W y 400 mA La EPS 1001 est dise ada principalmente para aplicaciones de Electroforesis en gel de poliacrilamida SDS PAGE PAGE nativa Electroforesis en gel de agarosa Electroforesis de DNA en campo pulsante Separaci n de fragmentos de DNA Electroforesis submarina de DN A L
43. ecommended or described in this manual and isusedin the same state as it was delivered from Amersham Pharmacia Biotech except for alterations described in this manual Important user information X JykJX Terms and Conditions of Sale All goods and services are sold subject to the terms and conditions of sale of the company within the Amersham Pharmacia Biotech group which supplies them A copy of these terms and conditions is available on request Should you have any comments on this product we will be pleased to receive them at Amersham Pharmacia Biotech Inc 654 M innesota Street San Francisco CA 94107 USA Trademarks Amersham Pharmacia Biotech Inc Amersham is a trademark of 800 Centennial Avenue N ycomed Amersham plc P O Box 1327 Pharmacia and Drop Design are Piscataway N J 08855 1327 trademarks of Pharmacia amp Upjohn USA Inc Copyright 1998 Amersham Pharmacia Biotech AB All rights reserved Office Addresses Amersham Pharmacia Biotech AB SE 751 84 Uppsala Sweden Amersham Pharmacia Biotech UK Limited Amersham Place Little Chalfont Buckinghamshire England H P7 9NA EPS 1001 User Manual 1 Introduction The Amersham Pharmacia Biotech Electrophoresis Power Supply EPS 1001 is a high quality high precision and safe power supply for electrophoresis applications that require a maximum of 1000 volts and 400 mA EPS 1001 is primarily designed for the following techniques
44. er l attuale funzionamento Inoltre il tasto SET permette di cambiare il programma durante il funzionamento dopo aver premuto PAUSE CONTINUE Change up C hange down Consentono di modificare i parametri il valore o un altra opzione nel campo lampeggiante Tenendo premuti i tasti i valori numerici vengono modificati in modo accelerativo Premendo brevemente si cambiano i valori in incrementi preimpostati Le selezioni di parametri unit come Vh e Y ES N O si cambiano con un unica pressione del tasto tasti possono essere utilizzati anche per commutare tra tempo e voltore nei modi RUN PAUSE ed END I valori scorreranno automaticamente cio una volta raggiunto il massimo si passa al minimo o viceversa RUN Premendo RUN si avvia il funzionamento e commuta il programma nel modo RUN Sul display vengono visualizzati il numero del programmi ed i valori attuali di tensione corrente e potenza V engono anche visualizzati il tempo trascorso o le voltore La commutazione tra tempo e voltore avviene tramite i tasti W T PAUSE CONTINUE Commuta lo strumento nel modo PAUSE e disattiva l alimentazione Il display visualizza lo stato di funzionamento alla pressione del tasto PAUSE CONTINUE funziona solamente nel modo RUN Vengono visualizzati i valori di tempo ed integrazione residui N el modo PAUSE SET pu essere utilizzato per modificare il programma Per tornare al modo RUN premere PAUSE CONTINUE o RUN EPS 1001 Manuale dell operatore
45. er manual restart 4 40 C 0 95 84 106 kPa according to GPR 100 120 200 240 V 50 60 Hz Max 140 W 250 x 385 x 95 mm 3 8 kg 13 GB EPS 1001 User Manual 9 Ordering information Designation Code No Electrophoresis Power Supply EPS 1001 18 1130 03 14 EPS 1001 Manuel d Utilisation Table des mati res 1 Introduction Renseignements concernant la s curit 2 4 1 1 cran d affichage 4 1 2 Clavier 2 1 Pr cautions d emploi 2 4 1 3 Prises connecteurs de sortie 5 2 2 Caract ristiques de s curit 4 2 Panneau arri re 5 intrins ques 3 5 Utilisation esso hah ire 6 3 R ception et installation 3 5 1 G n ralit s 6 4 Description technique 4 5 2 Instructions simplifi es 6 4 1 Panneau avant 4 Renseignements importants d utilisation Veuillez lire ce manuel enti rement pour une compr hension totale de l usage de EPS 1001 A scrupuleusement suivie afin d viter des blessures corporelles N e pas poursuivre tant que les instructions n ont pas t enti rement assimil es et que toutes les conditions indiqu es n ont pas t r unies AVERTISSEMENT Le signe d avertissement souligne une instruction qui doit tre D claration de conformit N ormes de securit Ce produit est conforme la directive 73 23 EEC
46. est branch e sur la tension secteur de 220 240 V avec une tension secteur s lectionn e sur 100 120 V Employer le cordon d alimentation secteur appropri et brancher une de ses extr mit s sur le r ceptacle d alimentation secteur de l unit EPS 1001 voir Figure 2 et l autre extr mit la prise d alimentation Secteur avec prise de terre M ettre l unit sous tension Chaque fois que l instrument est mis sous tension un essai automatique est effectu par le syst me Si cet essai d tecte une erreur il affiche un message l cran et d clenche un signal d alarme sonore voir 2 2 Branchement de ou des unit s d lectrophor se Brancher les fils de l unit d lectrophor se rouge sur rouge et noir ou bleu sur bleu comme indiqu dans Figure 1 voir verso dela page couverture Le fil rouge est le positif et le fil noir ou bleu est le n gatif Recommandation importante N utiliser que des fils lectriques parfaitement intacts et des quipements autoris s pour la tension utilis e Il est possible d utiliser deux unit s d lectrophor se simultan ment avec le m me programme Il est indispensable de doubler les limites d intensit mA et de puissance W lorsqu on utilise deux unit s d lectrophor se en m me temps Par contre la tension V reste la m me qu on utilise une ou deux unit s d lectrophor se la fois FR EPS 1001 Manuel d Utilisation enter 4 1 pause continue
47. h extremely low conductivity The high voltage is also turned off in this case and an error message is shown This function can be disabled to perform certain applications see 5 4 Optional programming 3 Sudden load change detection This function prevents accidents under running conditions due to a break in the electrical circuit such as a bad connection to the electrophoresis unit The high voltage is turned off in such an event Error messages are also shown on the display for sudden fails 3 Unpacking and installation Unpacking Select the appropriate voltage range 100 120 or 220 240 V See Fig 2 inside front cover Warning If the power supply is connected to 220 240 V with the range set to 100 120 V the instrument can be severely damaged Select the appropriate mains cable and connect one end to the mains socket on the EPS 1001 power supply see Fig 2 inside front cover and the other end to an AC grounded outlet Connection of the electrophoresis unit s Connect the leads from the electrophoresis unit red to red and black or blue to blue see Fig 1 inside front cover The red lead is the positive and black or blue is the negative Warning Use only undamaged electrical wire and equipment approved for the voltage you will use Two electrophoresis units can be run simultaneously with the same program Please remember to double the limiting current and power if two electrophoresis units are run at
48. ide d une des touches Y Confirmer en appuyant sur ENTER 5 Si vous avez s lectionn comme unit les V olt heures ou la dur e entrez la valeur du param tre de fin l aide de l une des touches 4 1 et ENTER 6 En cas de s lection param tre de fin automatique p 4 le syst me demande si vous souhaitez une alerte sonore en fin d ex cution du programme Entrez Y ES oui ou NO non avec et appuyez sur EN TER La s lection Y ES s affiche sous la forme d un A comme dans 1A droite du num ro de programme l cran 7 Lorsquela programmation est termin e brancher l unit d lectrophor se sur les connecteurs de sortie et appuyer sur la touche RUN 8 La tension effective le courant effectif et la puissance effective sont affich s ainsi la dur e coul e ou le nombre de Volt heures basculer d une valeur l autre en appuyant sur l une ou l autre touche W T 9 Utiliser PAUSE CONTINUE pour interrompre et reprendre ensuite le cours d ex cution du programme 10 Les param tres programm s peuvent tre v rifi s pendant l ex cution en appuyant sur la touche SET 11 L Electrophor se ainsi programm e s arr te automatiquement ou en appuyant sur la touche STOP EPS 1001 Manuel d Utilisation Les caract res clignotants sont repr sent s en caract res gras 1 Choisir un num ro de SET P1 Programme 200 V 40 mA s lectionner avec confirmer avec 2 R gler la tension li
49. integration values are retained In the pause mode SET can be used to make changes in the program Return to RUN mode by pressing PAUSE CONTINUE or RUN STOP Stops the run and puts the instrument in EN D mode The voltage is switched off and the end parameters are displayed Switch between time and integrated voltage by pressing 4 9 A run cannot be continued after pressing STOP Continue by pressing RUN to run the same method again or pressing SET to choose another program program a new method or make changes in an existing method MORE Y ou can disable the start current check that otherwise detects if the resistance is too high and switches off the power See Section 5 4 EXIT Stops the execution of an operation such as the entry of a value O nly values units that have already been confirmed by ENTER are retained Returns the instrument to the mode that was left or the start position in SET 4 1 3 Output sockets There are two sets of output sockets to allow two electrophoresis units to be connected and run at the same time see Fig 1 inside front cover The voltage output is 5 1000 V The negative output socket gives between 0 and 500 V and the positive gives between 0 and 500 V 4 2 Rear panel The rear panel is shown in Fig 2 See inside front cover On the rear panel there is 1 A mains switch Press in I to switch on the power to the power supply Press 0 to switch off the power 2 A socket for the mains cable 3
50. io de carga repentino Esta funci n impide accidentes durante el funcionamiento debidos a la ruptura del circuito el ctrico como una mala conexi n a la unidad electrofor tica La alta tensi n se desconecta 3 Desembalaje e instalaci n Desembalaje Controlar la entrega con la lista de contenido adjunta Ver si se han producido da os durante el transporte De haberlos comunicarlo inmediatamente al representante local de Amersham Pharmacia Biotech y a la empresa transportista Conexi n a la red Seleccionar la tensi n apropiada 100 120 220 240 V la figura 2 ver interior de la cubierta Precauci n El aparato puede sufrir aver as graves si se conecta a 220 240 V habiendo seleccionado la tensi n de 100 120 V Seleccionar el cable de red adecuado Conectar uno de sus extremos a la toma de la fuente de alimentaci n ver la figura 2 y el otro a un enchufe de corriente alterna conectado a tierra Conectar la corriente Cada vez que se conecta el aparato se ejecuta un test de diagn stico autom tico y si se detecta un error se visualiza un mensaje en la pantalla y suena una alarma Conexi n de la s unidad es electrofor tica s Conectar los cables de la unidad rojo con rojo y azul o negro con azul la figura 1 ver interior de la cubierta El cable rojo es el positivo y el azul o negro el negativo Precauci n Utilizar solamente hilos el ctricos intactos y equipos homologados para la tensi n a emp
51. ionar con Confirmar con 5 Elegir la unidad del evento h Vh o anular la ruptura OFF SET 1000 V P2 50 mA Seleccionar con Confirmar con e 6 Fijar el evento SET P2 1000V 50mA Seleccionar con Programauo n de Confirmarcon Fr eeventos desactivada 7 Elegir alarma s YES o no NO Seleccionar con Confirmar con me 8 Programa listo SET p24 2 00 h SET P2 OFF 1000 V 50 mA 50 W 1000 V 50 mA 50 W Figura 3 Sumario paso a paso de la programaci n ES EPS 1001 Manual del usuario Figura 4 Inhabilitaci n del iente PrO AE coments 1 Introducir MORE m s SET p2 1000 V 50 mA Seleccionar MORE A m s con 2 Seleccionar el programa MORE p2 Advertencia Seleccionar programa Si se inhabilita el Salesiano DI control de la corriente de arranque la fuente de alimentaci n puede 3 Elegir inhabilitaci n del control suministrar alta de corriente no tensi n aunque no est conectada al equipo de electroforesis Seleccionar con Confirmar y salir con Confirmar con Control de corriente de arranque YES no 4 4 Vuelta a la modalidad SET SET P2A 2 00 h preparar 1000 V 50 mA 50 W 1 Elegir el n mero de programa SET P1 OFF 200V 40mA 10W Seleccionar con 2 Iniciar el funcionamiento SET PQA 2 00 h 1000V 50mA 60W Iniciar con RUN pos 00 01h 300V 50mA 25W RUN
52. it inf rieure 40 V quela r sistance n est pas sup rieure une limite d termin e Si cette r sistance est trop lev e la tension est coup e imm diatement Une r sistance trop lev e peut aussi avoir pour cause l utilisation de tampons conductivit extr mement faible Dans ce cas l EPS 1001 coupe aussi la haute tension et affiche un message d erreur Cette fonction de V rification du courant peut tre d sactiv e pour certaines applications voir Figure 4 D sactivation de la v rification du courant 3 Sudden load change detection D tection d une variation soudaine de charge Cette fonction emp che tout accident pendant l ex cution provoqu par une rupture du circuit lectrique par exemple une connexion d fectueuse du branchement l unit d lectrophor se L EPS 1001 coupe imm diatement la haute tension dans une telle situation 3 Reception et installation R ception V rifier le contenu l aide de la liste d inventaire V rifier que le mat riel n a pas t endommag pendant le transport Signaler imm diatement toute d t rioration au repr sentant Amersham Pharmacia Biotech local et la soci t de transport en question Branchement sur la tension secteur S lectionner la tension appropri e 100 120 V ou 220 240 V Figure 2 voir verso dela page couverture Recommandation importante Le g n rateur risque d tre s rieusement endommag si l unit d alimentation lectrique
53. lear Con un mismo programa pueden operarse dos unidades electrofor ticas simult neamente Para ello debe doblarse la corriente limitadora y la potencia La tensi n es la misma independientemente del n mero de unidades ES EPS 1001 Manual del usuario VIT E pause continue 4 Descripci n t cnica 4 1 Panel delantero En el panel delantero hay una pantalla alfanum rica un teclado con 9 teclas de membrana un diodo luminiscente que se enciende cuando se aplica la tensi n H V on y tomas para dos unidades electrofor ticas 4 1 1 Pantalla La pantalla alfanum rica de 32 d gitos sirve de gu a en la programaci n visualiza los par metros durante la electroforesis y los par metros finales posteriores A dem s hace preguntas y visualiza mensajes de error La pantalla tiene dos renglones La figura 1 ver interior dela cubierta ilustra una pantalla en posici n de arranque con la corriente conectada La modalidad en este caso SET se indica a la izquierda del rengl n superior En el centro del mismo rengl n se indica el n mero del programa previamente introducido La cifra intermitente en este caso 1 indica que puede cambiarse con las teclas Y P A la derecha del rengl n superior se indica el evento programado en este caso desconectado Las posiciones izquierda central y derecha del rengl n inferior indican la tensi n el amperaje y la potencia respectivamente 4 1 2 Teclado E
54. llustra il funzionamento dell EPS 1001 I 2 Programmazione o editaggio di un metodo V edere Figg 3 e 4 Esecuzione di un metodo Vedere Fig 5 5 2 Brevi istruzioni Questa sezione riassume i punti principali per l utilizzo dell EPS 1001 Fare riferimento anche alla guida schematica di funzionamento allegata all alimentatore Si raccomanda di conservare questa guida schematica vicina allo strumento 1 2 3 8 9 Accendere l unita Il display visualizza il numero del programma Premere ENTER per visualizzare il numero attuale del programma oppure utilizzare 4 7 per selezionare il numero del programma desiderato Premere ENTER Per ognuno dei parametri di tensione corrente e potenza premere 7 fino a raggiungere il valore desiderato Confermare premendo EN TER dopo ogni impostazione il display passer automaticamente al parametro successivo Selezionaretra interruzione del funzionamento manuale o automatica Per l interruzione automatica selezionare il punto di interruzione in Voltore Vh o tempo h Selezionare l unit corretta o per l interruzione manuale selezionare OFF con V f Confermare premendo ENTER Se e stato scelto voltore o tempo impostare il valore da utilizzare come punto di interruzione con T e ENTER Nel caso in cui sia stato selezionato un punto di interruzione inferiore a 4 il display chieder all operatore se desidera l emissione un allarme acustico al termine delle oper
55. matically to show RUN values after 5s Alternatively use EXIT or RUN Stopping the run and viewing end parameter values W hen the programmed time or volthour is attained the program will enter the End mode It is also possible to break the run manually by pressing STO P In both cases the voltage current and power will go to zero as indicated by the HV on LED switching off The end parameter values are displayed Switch between elapsed time and integrated voltage by 4 7 An alarm will sound if selected in the program Y ou can stop the alarm by pressing STOP EPS 1001 User Manual GB 1 Choose Program number SET P1 OFF 200V 40mA 10W Select by 2 Startthe run SET P2A 2 00h 1000V 50mA 60W Start by e RUN P24 0 01h 300 V 50mA 25W au JE tar PAUSE P24 0 30h 500 V 50mA 25W 500 V 50mA 100W or wait 5 s pause u 3 During a run you can view the settings It is also possible to interrupt the run and make changes P SET P24 2 00h in the program 1000 V 50mA 60W SET P24 2 00h 1000 V 60mA 60W Programming 4 Stop the run manually RUN Po 45h 1000V 35mA 35W END P2A 145h 1000V 35mA 35W 5 The run is stopped automatically END P2 20h 1000V 20mA 20W Fig 4 Step by step summary of run and pause function A run cannot be continued after pressing STOP Disconnect the leads and proceed with post electrophoretic techniques Since diffusio
56. mite P2 OFF v 400 mA 100 W s lectionner avec confirmer avec SET P2 OFF 1000 V 400 mA 100 W 3 R gler le courant limite mA s lectionner avec confirmer avec SET 1000 V 4 Regler la puissance limite W P2 50 mA s lectionner avec confirmer avec SET 1000 V 5 Choisir l unit du param tre de fin h ou Vh ou pas OFF D sactiv P2 50 mA s lectionner avec confirmer avec 6 Fixer le param tre de fin SE 1000 V P2 50mA s lectionner avec confirmer avec 7 S lectionner l option signal d alarme YES NO oui non s lectionner avec confirmer avec 8 Programme pr t SET 1000 V SET pas 1000 V 50 mA 50 W Param tre de fin supprim P2 50 mA Figure 3 R sum d taill de la programmation FR EPS 1001 Manuel d Utilisation Figure 4 D sactivation des v rifications du courant initial 1 Passer en mode MORE suite SET P2A 1000 V 50 mA s lectionner MORE avec Avertissement 2 S lectionner le programme MORE P2 Lorsqu on d sactive la S lection programme fonction v rification du s lectionner avec 9 4 courant initial n anmoins fournir une D srificat haute tension m me si 3 S lectionner l option Me addet ication elle n est pas branch e du courant YES no l alimentation peut sur l quipement N s lectionner avec d electro
57. n Fehlermeldungen werden auf der Anzeige angezeigt 1 Functional earth leakage Fehlerstromanzeige Wird das Stromversorgungsger t an ein Elektrophoreseger t angeschlossen das einen Erdschlu aufweist dann erfa t das EPS 1001 diesen Fehler und die Hochspannung wird abgeschaltet 2 Start current check Strompr fung starten Um zu gew hrleisten da ein Elektrophoreseger t richtig angeschlossen ist berpr ft das Stromversorgungsger t ob der Widerstand h her ist als ein vorgegebener Grenzwert bei einer niedrigen Sicherheitsspannung lt 40 V Ist dieser Widerstand zu hoch dann wird die Spannung abgeschaltet Ein zu hoher Widerstand kann auch durch die Verwendung von Puffern mit u erst niedriger Leitf higkeit verursacht werden Auch in diesem Fall wird die Hochspannung abgeschaltet und eine Fehlermeldung angezeigt Diese Funktion kann f r die Durchf hrung bestimmter Anwendungen gesperrt werden siehe Abb 4 Sperren der Strompr fung 3 Sudden load change detection Lastwechsel Pr fung M it dieser Funktion werden Unf lle unter Programmlaufbedingungen aufgrund eines Zusammenbruchs der elektrischen Schaltung wie zum Beispiel aufgrund einer schlechten Verbindung mit dem Elektrophoreseger t verhindert In einem solchen Fall wird dieH ochspannung abgeschaltet 3 Auspacken und Installation Auspacken Den Inhalt anhand der beigef gten Packliste pr fen Sollte das Ger t beim Transport besch digt worden sein sofo
58. n diesem Handbuch beschrieben sind 5 Betrieb 4 1 1 Anzeige 4 4 1 2 Tastatur seta aes 4 4 1 3 Ausgangsbuchsen 5 4 2 R ckwand 5 1 Uberblick 5 2 Kurzanweisungen Verkaufsbedingungen S mtliche Waren und Dienstleistungen werden zu den V erkaufsbedingungen desjenigen Unternehmens innerhalb der Amersham Pharmacia Biotech Gruppe verkauft das die W aren liefert Auf Anfrage ist eine Kopie dieser V erkaufsbedingungen erh ltlich Bei Anmerkungen zu diesen Produkten wollen Sie sich bitte an H oefer Pharmacia Biotech Inc wenden Sollten Siezu dem Produkt irgendwelche Bemerkungen haben w rden wir uns dar ber freuen wenn Sie sie uns an folgende Adresse zukommen lassen k nnten Amersham Pharmacia Biotech Inc 654 M innesota Street PO Box 77387 San Francisco CA 94107 USA Warenzeichen Amersham ist ein Warenzeichen der N ycomed Amersham plc Pharmacia und Drop Design sind Warenzeichen der Pharmacia amp Upjohn Inc Copyright 1998 Pharmacia Biotech AB Alle Rechte vorbehalten Gesch ftsadressen Amersham Pharmacia Biotech AB SE 751 84 Uppsala Sweden Amersham Pharmacia Biotech UK Limited Amersham Place Little Chalfont Buckinghamshire England HP7 9NA Amersham Pharmacia Biotech Inc 800 Centennial Avenue P O Box 1327 Piscataway N J 08855 1327 USA EPS 1001 Bedienungsanleitung 1 Einleitung Das Elektrophorese Stromversorgungsger t EPS 10
59. n will begin as soon as the voltage is turned off you should remove the gel and begin staining or blotting immediately 5 4 Optional programming Y ou can disable the start current check that otherwise checks that the resistance is not higher than a specified limit This disabling function is set in the M ORE mode instead of the SET mode in order not to confuse users who do not need this feature and for safety reasons TheM ORE modeis described in Fig 5 After pressing M ORE the program asks you for the program number and if you want to keep the start current check Choose program number and change betw een Y ES and NO by Y and confirm with EN TER Choosing NO for start current check means that the function is disabled and high voltage can be applied despite a very high resistance Leave the M ORE mode by EXIT N Warning By disabling the start current check the power supply can deliver high voltage even if it is not connected to electrophoresis equipment GB EPS 1001 User Manual 10 1 Enter MORE SET P2A 1000 V 50 mA Select MORE by 2 Select program MORE Select d Select by Confirm by 3 Choose current check no Current check at start YES no Select by Confirm and 8 4 Exit by 4 Back to SET mode SET P2A 2 00 h 1000 V 50 mA 30 W Fig 5 Disabling the current check 5 5 Choosing run parameters EPS 1001 is an automatic cross over power supply that
60. nt sur le param tre de contr le en fonction des valeurs limites programm es et en fonction des variations de conductivit du syst me Il est possible de sauvegarder trois programmes d une tape chacune 2 Renseignements concernant la s curit 2 1 Pr cautions O p rer avec la plus grande prudence lorsque vous utilisez l instrument car il peut g n rer une tension et une intensit suffisantes pour provoquer des chocs lectriques mortels Pour viter tout risque d accident l instrument doit tre utilis uniquement par du personnel qualifi et toujours en suivant strictement les instructions fournies N ous vous recommandons de lire attentivement ce manuel avant d utiliser l unit d alimentation lectrique 1 Cet instrument a t concu pour amp re utilis l int rieur uniquement 2 Cet instrument doit toujours tre utilis avec le fil de terre du cordon d alimentation correctement mis la terre de la prise secteur 3 Pour que la ventilation fonctionne normalement veiller ce que les fentes d a ration l arri re et sur les c t s de l instrument ne soient jamais obstru es 4 Ne pas faire fonctionner l instrument en milieu extr mement humide humidit sup rieure 95 Eviter la condensation en laissant l unit s quilibrer la temp rature ambiante lorsqu elle est transf r e d un environnement froid un environnement plus chaud 5 Conserver l instrument aussi sec et aussi propre que possi
61. nummer blinken ENTER f r die aktuelle Programmnummer dr cken oder mit den Tasten 4 9 die gew nschte Programmnummer w hlen EN TER dr cken F r jeden der Parameter Spannung Strom und Leistung so lange Y dr cken bis der gew nschte W ert erreicht ist Durch Dr cken von ENTER jeden Parameter best tigen wor auf die Anzeige automatisch auf den n chsten Parameter bergeht Manuelle oder automatische Unterbrechung w hlen F r eine automatische Unterbrechung die Zwischenstopparameter in Voltstunden V h oder in Zeit h w hlen Entweder die korrekte Einheit oder f r eine manuelle Unterbrechung OFF mit Y fw hlen Durch Dr cken von EN TER best tigen Wurde Voltstunden oder Zeit gew hlt den f r den Zwischenstopparameter zu verwendenden Wert mit den Tasten 4 9 und ENTER einstellen Wurde unter 4 ein automatischer Z wischenstopparameter gew hlt dann wird der Benutzer gefragt ob er m chte da nach jedem Programmlauf ein Alarm ert nt M it 4 9 Y ES oder NO w hlen und dann ENTER dr cken Wird Y ES gew hlt dann wird dies durch ein kleines A wiein 1A auf der rechten Seite der Programmnummer in der Anzeige angezeigt Nach abgeschlossener Programmierung sollte das Elektrophoreseger t an die Ausg nge angeschlossen und die Taste RUN gedr ckt werden Es erscheinen die aktuellen Werte von Spannung Strom und Leistung auf der Anzeige zusammen mit dem Wert f r die abgelaufene Zeit oder die Voltstunden Um
62. ogramming range Output resolution Programming resolution Accuracy Line regulation Load regulation Ripple Short term stability Long term stability Start current check Sudden load change check Ground leakage check Output protection Recovery after power failure Ambient operating temperature Ambient operating humidity Ambient operating pressure Mains requirements Power consumption Dimensions WxDxH Weight Maximum voltage current and power with automatic cross over at preset limits Constant Voltage 0 1000 V DC Constant Current 0 400 mA Constant Power 0 100 W Voltage 5 1000 V DC Current 1 400 mA Power 1 100 W Time 00 01 500 h Volthour 1 500 000 Vh Voltage 1V Current 100 yA 0 39 9 mA 1 mA 40 400 mA Power 1W Voltage 1V Current 1 mA Power 1W Time 1 min 00 01 99 59 1h 100 500 h Volthour 1 Vh 1 9999 Vh 100 Vh 10 0 99 9 kVh 1 kVh 100 500 kVh Voltage 1 3V Current 1 1 mA Power 2 1 W Timer 0 1 1 min 00 01 99 59 h 0 195 10 min 100 500 h lt 0 2 lt 1 at load change 10 90 of maximum load lt 1 at 1000 V lt 0 2 10 h after warm up lt 1 year Resistance not greater than 0 2 MQ at 40 V current less than 200 uA Can be disabled Shuts off output voltage on extreme load change Leakage not greater than 500 HA Fully protected against any overload conditions Duration lt 8 s The program continues automatically Duration gt 8 s The program continues aft
63. owever the voltage is not distributed evenly across the gel These gels have a region with low onic strength that causes a high electrical field strength This region increases as the electrophoresis proceeds This means that the main part of the voltage is spread over a longer and longer distance and a higher and higher power is tolerated This is the reason why constant current is chosen for such applications It is however recommended to also limit the power as a precaution against overheating the gel The power supply will propably switch over to limiting power at the end of the run dueto increased voltage For detailed information about parameter values always follow the gel supplier s recommendations 5 6 Short instructions This section summarizes the main points covered earlier in this chapter Use it as a check list once you are familiar with the detailed programming and running instructions Refer also to the separate schematic operating guide included with the power supply We recommend you keep this schematic guide close by the instrument 1 Turn mains power ON The display should have the program number flashing 2 Press ENTER for the actual program number or use to V select the program number desired Press ENTER 3 For each of the parameters Voltage Current and Power press 4 9 until the value desired is reached Confirm by pressing ENTER after each one and the display will automatically move to the next parameter 4
64. phorese confirmation et NM O confirmer avec sortie avec v rification du courant 4 Retour au mode SET SET p2 2 00 h Figure 5 R sum d taill programmation d marrage 1000V 50mA 50W des fonctions RUN et PAUSE 1 Choisir le num ro de programme SET P1 OFF s lectionner avec 200V 40mA 10W s lectionner avec 2 D marrer l ex cution du programme Z SET P24 2 00h d marrer avec 1000V 50mA 60W d marrer avec RUN P24 0 01h 300 V 50mA 25W Y RUN pos 0 30 500V 50mA 25W ou attendre 5 sec SET P24 2 00h 1000 V 50mA 60W PAUSE P2 0 30h 500V 50mA 100W 3 Pendant l ex cution du programme vous pouvez ou visualiser les P SET P2 2 00h param tres programm s ll est 1000V 50mA 60W possible d interrompre l ex cution pour effectuer des modifications dans le programme Programmation 4 Arr t manuel de l ex cution RUN poA 1 45 h 1000 V 35mA 35W END P2A 1 45h 1000 V 35mA 35W END P22 72 00h 1000 V 20mA 20W 5 Arr t automatique de l ex cution EPS 1001 Bedienungsanleitung Inhaltsverzeichnis 1 Einleitung Sicherheitsangaben 2 1 SicherheitsmaBnahmen 2 2 2 Eingebaute Sicherheitsmerkmale 3 4 1 Fronttafel Auspacken und Installation Technische Beschreibung Wichtige Benutzerinformationen Zum vollen Verst ndnis vor Gebrauch von EPS 1001
65. r wenn er korrekt ist aufgenommen wird und der Programmiervorgang geht auf das n chste Feld ber G ltige W erte sind f r die Spannung 5 1000 V f r den Strom 1 400 mA f r die Leistung 1 100 W f r die Zeit 0 01 500 h f r die Voltstunden 1 500 000 Vh SET Diese T aste schaltet das Ger t in den SET M odus seinen Programmierungs und Startmodus Im RUN M odus werden nach dem Dr cken der Taste SET die programmierten Parameter f r den aktuellen Programmlauf angezeigt Au erdem kann der Benutzer durch Dr cken der Taste SET Anderungen im Programm w hrend eines Programmlaufs vornehmen wenn zun chst die Taste PAUSE CONTINUE gedr ckt wurde Change up Change down M it dieser Taste werden Parameter Werte oder eine andere O ption in dem blinkenden Feld ge ndert N umerische W erte laufen schneller ab wenn eine T aste gedr ckt gehalten wird durch Anklicken einer Taste ndert sich der Wert in voreingestellten Stufen Parameter oder Einheiten z B Vh und Y ES N O O ptionen werden durch einmaliges D r cken einer Taste ge ndert Die Tasten k nnen auch zum Umschalten zwischen Zeit und Voltstunden in RUN PAUSE und END verwendet werden Die Werte rollen d h sie laufen automatisch von ihrem H chstwert bis zu ihrem M indestwert bzw umgekehrt RUN N ach dem Dr cken der Taste RUN beginnt der Programmlauf und das Programm wird in den RUN M odus geschaltet Auf der Anzeige erscheinen die Programmnummer und die aktuellen W e
66. rana un diodo luminoso LED che si illumina quando viene applicata tensione H V attivata e connettori per due unit per elettroforesi 4 1 1 Display Un display alfanumerico a 32 cifre guida l operatore nella programmazione ed illustra i valori attuali dei parametri durante l elettroforesi nonch i parametri finali ad elettroforesi terminata Esso pone anche domande e visualizza i messaggi di errore Il display dotato di una riga inferiore ed una superiore La Fig 1 vedere prima di copertina mostra il display all attivazione dell unit Il modo in questo caso SET viene visualizzato nella riga superiore sulla sinistra Il numero del programma si trova al centro della riga superiore Il numero indicato quello del programma precedente La cifra lampeggiante in questo caso 1 indica che pu essere cambiato utilizzando i tasti 4 7 La riga superiore destra indica il punto di interruzione programmato In questo caso disattivato Le posizioni inferiore sinistra centrale e destra visualizzano rispettivamente tensione corrente e potenza 4 1 2 Tastiera ENTER Inserisce un valore fissa il valore se corretto e commuta la programmazione al campo successivo valori disponibili sono tensione 5 1000 V corrente 1 400 mA potenza 1 100 W tempo 0 01 500 ore e voltore 1 500 000 Vh SET Commuta lo strumento in SET nel modo di programmazione ed avviamento N el modo RUN premendo SET si visualizzano anche i parametri programmati p
67. rt die rtliche Amersham Pharmacia Biotech N iederlassung und das entsprechende Speditionsunternehmen verst ndigen Netzanschlu Den entsprechenden Spannungsbereich w hlen 100 120 oder 220 240 V Abb 2 siehe Deckblattinnenseite Achtung Wird der Spannungsbereich auf 100 120 V eingestellt und das Ger t an 220 240 V angeschlossen dann wird das Ger t stark besch digt Das entsprechende N etzkabel w hlen und ein Ende an die N etzanschlu buchse des EPS 1001 Stromversorgungsger tes siehe Abb 2 und das andere Ende an eine geerdete N etzsteckdose anschlie en Den Strom einschalten Bei jedem Einschalten des Ger tes durchl uft dieses einen Selbstdiagnosetest Wird bei diesem Test ein Fehler entdeckt so erscheint eine M eldung auf der Anzeige und es ert nt ein Alarm Anschlu des der Elektrophoreseger te s Die von dem Elektrophoresegerat kommenden K abel anschlie en Rot auf Rot Schwarz oder Blau auf Blau Abb 1 siehe Deckblattinnenseite Das rote Kabel ist das positive das blaue oder schwarze das negative Kabel Achtung Stets intakte elektrische Kabel und Ausr stung verwenden die f r diezu verwendenden Spannungen geeignet sind DE EPS 1001 Bedienungsanleitung VIT Es k nnen zwei Elektrophoreseger te gleichzeitig mit demselben Programm betrieben werden Bitte beachten da die Grenzwerte f r Strom und Leistung verdoppelt werden m ssen wenn zwei Elektrophoreseger te glei
68. rte f r EPS 1001 Bedienungsanleitung DE pause continue Spannung Strom und Leistung Auch die abgelaufene Zeit bzw die Voltstunden werden angezeigt Das Umschalten zwischen Zeit und Voltstunden erfolgt mit 4 9 PAUSE CONTINUE M it dieser Taste wird das Ger t in den PAUSE M odus geschaltet und die Spannung wird abgeschaltet Die Anzeige zeigt den Status des Programmlaufs zu dem Zeitpunkt an an dem die Taste gedr ckt wurde PAUSE CONTINUE funktioniert nur im RUN M odus Zeit und Integrationswerte bleiben erhalten Im Pausemodus k nnen durch Dr cken der Taste SET Anderungen am Programm vorgenommen werden Durch Dr cken der Taste PAUSE CONTINUE oder der TasteRUN kann der Benutzer in den RUN M odus zur ckkehren STOP Diese T aste unterbricht den Programmlauf und schaltet das Ger t in den EN D M odus Die Spannung wird abgeschaltet und die Endparameter werden angezeigt Ein Programmlauf kann nach dem Dr cken der Taste STOP nicht fortgesetzt werden MORE M it dieser Taste kann die Funktion Start Current Check Strompr fung starten die sonst berpr ft ob der Widerstand zu hoch ist und den Strom abschaltet gesperrt werden EXIT Diese Taste stoppt die Ausf hrung eines Vorgangs sowie die Eingabe eines Wertes Es bleiben nur W erte Einheiten erhalten die zuvor bereits mit der Taste ENTER best tigt waren Bringt das Ger t in den vorhergehenden M odus oder in die in SET eingestellte Startposition zur ck
69. schalten mit 4 1 DieTaste PAUSE CONTINUE kann zum Unterbrechen und nachfolgenden Fortsetzen des Programmlaufs verwendet werden 10 Die programmierten Parameter k nnen w hrend des Programmlaufs durch Dr cken von SET berpr ft werden 11 Der Programmlauf stoppt automatisch oder nach dem Dr cken von STOP EPS 1001 Bedienungsanleitung DE Blinkende Zeichen sind im Fettdruck dargestellt Programmnummer w hlen Spannungsgrenzwert einstellen Stromgrenzwert einstellen Leistungsgrenzwert einstellen Wahlen des Zwischenstoppara meters h oder Vh oder kein Zwischenstop OFF AUS Zwischenstoppara meter einstellen Alarm YES oder NO wahlen Programm bereit SET P1 200 V 40 mA W hlen mit Best tigen mit P2 OFF 400mA 100W W hlen mit Best tigen mit SET P2 1000 V 400 mA OFF 100 W W hlen mit Best tigen mit SET 1000 V P2 50 mA W hlen mit Best tigen mit SET 1000 V P2 50 mA W hlen mit Best tigen mit SET 1000 V P2 50mA W hlen mit Best tigen mit Alarm at End 2 Y W hlen mit 6 DI Best tigen mit SET pas 1000 V 50 mA Zwischenstoppara meter aus SET P2 1000 V 50 mA OFF Abb 3 Schrittweise Zusammenfassung des Programmiervorgangs DE EPS 1001 Bedienungsanleitung Abb 4 Sperren der pori S
70. se cover Only 13 Amps fuse approved to B S 1362 A S T A should be used Switch on the power Each time the instrument is turned on a self diagnostic test is done If an error is detected during the test a message will appear on the display and an alarm will sound 4 Technical description 4 1 Front panel The front panel consists of an alphanumeric display a keyboard with 9 membrane keys a light emitting diode LED that lights when voltage is applied HV on and connectors for two electrophoresis units 4 1 1 Display A 32 digit alphanumeric display guides you through the programming shows current parameter values during the electrophoresis and final parameter values afterwards It also asks questions and shows error messages The display has an upper and lower row Fig 1 inside front cover shows the display in the start position when power is switched on for the first time The mode in this case SET is shown in the upper row on the left The program number sin the center of the upper row The number shown is that of the previously entered program As no program has yet been entered the first available program number P1 is shown The blinking figure in this case 1 indicates that it can be changed by using the 4 7 keys The upper right shows the programmed breakpoint In this case it is off The lower left middle and right positions show voltage current and power respectively EPS 1001 User Manual GB
71. ssance 4 1 2 Clavier Touche ENTER Entr e Une pression sur cette touche permet successivement d entrer une valeur de la valider si cette valeur est correcte et de faire avancer le programme au champ suivant Les valeurs correctes sont tension 5 1000 V courant 1 400 mA puissance 1 100 W dur e 0 01 500 heures nombre de volt heures 1 500 000 Vh SET Programmation D marrage Une pression sur cette touche met l instrument en mode SET qui est le mode programmation et le mode d marrage En mode RUN ex cution une pression sur SET affiche les param tres programm s pour l ex cution en cours De plus une pression sur la touche SET permet d effectuer des modifications sur le programme en cours d ex cution apr s avoir appuy sur la touche PAUSE CONTINUE Change up Change down M odif Une pression sur l une de ces deux touches permet de modifier le param tre la valeur ou le choix oui non dans le champ qui clignote On modifie les valeurs num riques de mani re acc l r e en maintenant la pression sur la touche concern e Un pression rapide sur une de ces touches modifie la valeur par incr ments pr d termin s On modifie les param tres ou unit s par exemple V h et les choix YES NO oui non avec une seule pression sur la touche en question On peut aussi utiliser ces touches pour basculer entre la dur e et les volt heures en mode RUN ex cution PAUSE et END fin Les valeurs d filent en l occurrence elles
72. tore di messa a terra del cavo di alimentazione dello strumento sia collegato correttamente a terra alla presa di rete 3 Per consentire un sufficiente raffreddamento assicurarsi che le bocchette posteriori e laterali dello strumento non siano coperte 4 Nonutilizzare lo strumento in ambienti estremamente umidi oltre il 95 Evitare la condensazione consentendo all unit di ritornare gradualmente a temperatura ambiente dopo il trasporto da ambienti molto caldi o freddi 5 Conservare lo strumento il pi asciutto e pulito possibile Pulirlo regolarmente con un panno morbido Assicurarsi che lo strumento sia completamente asciutto prima dell utilizzo In caso contrario disinserire la spina dalla rete di alimentazione finch lo strumento non asciutto 6 Utilizzare esclusivamente cavi elettrici non danneggiati ed attrezzature specificate per la tensione utilizzata cavi ad alta tensione devono essere conformi alla norma IEC 1010 2 031 1993 Tutte le attrezzature per elettroforesi collegate all alta tensione devono essere conformi alla norma IEC 1010 1 1993 7 Notare che l uscita collegata al telaio massa di riferimento EPS 1001 Manuale dell operatore IT 2 2 Funzioni di sicurezza incorporate L EPS 1001 omologato in conformit alle norme di sicurezza elettriche IEC 61010 1 EN 61010 1 L alimentatore dotato di diverse funzioni di sicurezza incorporate Sul display vengono visualizzati anche i messaggi di errore
73. y SET 2 2 00 h SET P2 1000 V 50 mA 50W 1000 V 50 mA Fig 3 Step by step summary of programming GB EPS 1001 User Manual Back to start position The program is now back to the start position in SET mode with the program number flashing It is possible to go back to this position at any stage during programming in SET mode by pressing EXIT or STOP Note that the program is saved automatically when EXIT is pressed Disabling the current check See Section 5 4 if you want to use this feature 5 3 Running a program Connect the leads from the electrophoresis unit red to red and black or blue to blue Red is positive and black or blue negative Up to two electrophoresis units can run at the same voltage at one time Please remember to double the maximum current and power conditions if two units are to be run Voltage will be the same regardless of how many units are run The current should also be doubled if two gels are run on the same electrophoresis unit Choosing a program Press SET and select the program you wish to run by pressing Y until the value after the P is correct Confirm with ENTER O mit this step if you have just programmed a method as described insection 5 2 Running a program is described schematically in Fig 4 Running Press RUN to start the electrophoresis The display will show current values for voltage current power and elapsed time or volthours Y ou can switch between
74. y from a colder to a warmer environment 5 Keep the instrument as dry and clean as possible Wipe regularly with a soft damp cloth Let the power supply dry completely before use If wetted unplug the power supply until the instrument is dry 6 Use only undamaged electrical wire and equipment specified for the voltages you will use High voltages wires must meet the requirements of the IEC 1010 2 031 1993 electrical standard Any electrophoresis equipment connected to the power supply should meet the requirements of the IEC 1010 1 1993 electrical standard 7 Note that the output is connected to the chassis reference earth EPS 1001 User Manual GB 2 2 Built in safety features The EPS 1001 has been tested and complies with the IEC 61010 1 EN 61010 1 electrical safety standard The power supply also has several built in safety functions Error messages are also shown on the display see 7 Trouble shooting 1 Functional earth leakage Should the power supply be connected to an electrophoresis unit that has a leakage path to earth the EPS 1001 will detect this fault and the high voltage is turned off 2 Start current check To ensure that an electrophoresis unit is connected correctly the power supply checks that the resistance is not higher than a specified limit at a low safety voltage lt 40V If this resistance is too high the voltage is turned off Too high a resistance can also be caused by using buffers wit
Download Pdf Manuals
Related Search
Related Contents
Manual Usuario SII - Servicio de Impuestos Internos Cet appareil peut être utilisé par des enfants âgés d`au moins 8 ans (MSDS付き):スタンダードタイプ Leica DM4000B User's Manual DEUTSCH DEUTSCH - CONRAD Produktinfo. Quick Start Guide - Honeywell Analytics ティーチングボックス CON Les Cahiers de la Fourmi - extrait de Mélimo CE Hansgrohe 04286800 Instructions / Assembly Copyright © All rights reserved.