Home

Rührwerke R100C, R100CL und R100 CT

image

Contents

1. Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 42 3 4 Assembling the Shaft Protector Fastening Rod for Stirring Shaft Protector H Shaft Protector M 4 Set up and Operation 4 4 Intended Use Use a stirring shaft protector to provide protection against injury when working with the unit Attach the fastening rod H of the shaft protector M in front of the chuck to the stand e g use cross over clamp part no 60492 00 to fix the stirring shaft protector Now bring the pockets K to the screws J and turn the stirring shaft protector to the left to the limit stop The screw L is used to adjust the length of the stirring shaft protector Check that the stirring shaft protector is held in position securely prior to each use and also at regular intervals The position of the stirring shaft protector must only be adjusted when the unit is stationary and the power supply is disconnected The R100C R100CL R100CT is designed for use in chemical and biological laboratories of industrial enterprises universities and pharmacies It is suitable for stirring and mixing of liquids from medium to high viscosity For correct use the device must be fixed to an appropriate stand To ensure maximum service life observe the specified ambient conditions temperature and humidity and ensure that the instrument is not exposed to a corrosive atmosphere Betriebsanleitung Instruction
2. PA Falsche Parameternummer keine zu viele oder zu wenige spezifische Parameters Betriebszustand des Bereichs Mindestens ein Parameter ist zu lang Unbekanntes Datenformat Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 19 6 4 RS232 Befehle des R100C R100CL R100CT Tabelle 2 Aktuelle Einstellung 1 Dummy Parameter um bertragung des R100C R100CL einzuleiten R100CT auslesen Controller sendet handshake A aktuelle Motordrehzahl rpm B aktuelles Drehmoment relativ C aktuelles Drehmoment absolut D aktuelle Leistung relativ E aktuelle Leistung 1 0 2592000 1 1 Dummy parameter um Ubertragung einzuleiten Controller sendet handshake 0 2592000 A Zeitwert Sekunden gt 0 Timer ausgeschaltet Setzen der A Motordrehzahl in 1 min 0 2000 Motordrehzahl in 1 min 0 gt Motor halt Drehzahl 0 wenn programmierte Drehzahl ber dem erlaubtem Grenzwert liegt wird der Wert auf die erlaubte Grenze reduziert gt Set up Menu max Motor 1 min Auslesen des Motor 1 Dummy Parameter um Ubertragung 1 Sollwerts SET einzuleiten Controller sendet handshake 0 2000 A Motor Sollwert in rom Auslesen des 1 Dummy Parameter um Ubertragung 1 Systemstatus einzuleiten Controller sendet handshake 1 1 aktueller Systemstatus Code 1 gt normaler Betrieb i 0 1 2 Uberlast 4 0 gt keine Uberlast 1 gt Motor berlast Auslesen der 1 Dummy Parameter
3. 1 Controller sends in handshake 1 actual systemstatus code 1 1 gt normal operation 2 Overload condition 0 1 0 2 no overload condition 1 gt Motor is under overload operation 1 Read Type 1 Dummy parameter to initiate transfer Information of unit Controller sends in handshake A name of connected device B firmware version number C ON OFF counts D total operation time in minutes Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 53 Keyboard or RS232 A Control parameter speed control Write serial mode 0 gt frontpanel control of motor speed 1 gt RS232 control of motor speed via Select if setting for WSE command motor speed is accepted from frontpanel speed control knob or via RS232 command WSE Lock Unlock 1 lock unlock parameter frontpanel keyboard 0 gt frontpanel keyboard is disabled except On Off button 1 gt frontpanel keyboard is enabled WSA Set RS232 slave 1 New slave address of unit address 4 renumber slaves Switch unit off 1 Security parameter 1 WEE Save set up settings 1 Security parameter to EEPROM 6 5 Programming examples The following application examples show how RS232 commands can be sent to the stirrer and how actual parameters can be read out from the stirrer 6 5 1 Writing parameters to the stirrer If for example the motor speed should be set to 500 rpm s the following two command strings have to
4. 3 OVERTEMPERATURE PROTECTION R100C 48 54 OVERTEMPERATURE PROTECTION RIOOCL 48 3 5 OVERTEMPERATURE PROTECTION R100CT eese nennen nennen nere 49 5 6 VO CONNECTOR 9 PIN SUBZ D e UR fe un ea ln 49 6 THE RS232 INTERFACE OF THE R100C R100CL R100CT 50 6 1 RS232 CONNECTION CONFIGURATION BETWEEN PC AND THE RIOOC RIOOCL RIOOCT eese 51 6 2 FORMAT OF AN RS232 COMMAND 5 2552 Get etate eo nee 51 6 3 FORMAT OF THE CONTROLLER HANDSHAKE 52 6 4 RS232 COMMANDS OF THE R100C R100CL R10OOCT eese nennen rna 53 6 5 PROGRAMMING EXAMPEES ic eiae sa Testa gran eoa uaa cem Dres e ue sao ceo Aaaa 54 6 5 1 Writing parameters to the Stifter ia ob tret ae RE ret ise tea ra eati a a eee vns 54 6 5 2 Reading parameters from the stirrer essei 55 6 6 CONNECTION OF MORE THAN ONE UNIT TO A 55 7 THE SETUP MENU eio Fo e Serva bee ee PRA RE EE FERE eR PARU eS ae 57 7 1 EXAMPLE HOW TO CHANGE SETUP SETTINGS 59 ERROR MESSA G S a r aaraa 61 9 CLEANING AND MAINTENANCE sssssseesssssosssssses
5. Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 43 4 2 The Front Panel Controls The innovative key pad makes data entry easy and incorporates a 2 line alpha numeric LCD display 1 which shows all relevant data The entry and changing of timer or setup settings is made via the encoder wheel 5 the motor speed is set with the right speed control knob 7 CAT RI00C Timer button 3 Incremental encoder wheel 5 Power O Speed control knob 7 Change view on the R100C R100CL R100CT front panel LCD Display 1 Timer LED 2 Display select button 4 Load LED 6 Power LED 8 Power on off button 9 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 44 4 3 Description of the front panel functions LCD display 1 The LCD display shows all relevant system data Timer LED 2 indicates that timer is active and unit will be automatically switched off when timer is expired Timer button 3 If required pressing this key initiates the timer function which will shut down the R100C R100CL R100CT To change the timer setting use the left encoder wheel The timer is calibrated to show days hours minutes Select Display button 4 Press this key once shortly to step from one display option to the next Press and hold this key for 3 seconds to zero the relative torque measurement Mr Press and hold this key fo
6. cous v ero ipie bostes o ep Eo De v etg odi 41 3 ASE LeU eS 42 3 1 MOUNTING OF SUPPORT RODS eL Breve vr ER ELE Cove ge 42 3 2 ATTACH THE OVERHEAD STIRRER TO A STAND esee 42 3 3 ATTACHMENT OF STIRRING PADDEE 6 tre 283 Ceu etd te Fono vi Ce ves eI eene E Co vau eite aUo UE Ca 42 3 4 ASSEMBLING THE SHAFT PROTEET R iue t ia rri vas od ce a LER ELE oe v LR Ee EE Dore said 43 4 SET UPAND OPERATION eo ooo 885524258485 ESPERE Re aa TERR e FERE eoa MARRE Henne NER 43 4 1 INTENDED USE E cece EL we eaten ae wa Sage 43 4 2 THE FRONT PANEL CONTROLS beoe caesa ere 44 4 3 DESCRIPTION OF THE FRONT PANEL FUNCTIONS 45 4 4 DESCRIPTION OF THE REAR PANEL ON THE R100C R100CL RI10OOCT 46 4 5 STIRRING OPERATION nk eere te eec be rar La we ED LE v ve Led rev ED SL vw e 47 4 6 TIMER CONTROLLED OPERATION adanada rennen 47 5 SAFETY FUNCTIONS a e a o 48 5 1 ROTOR STUCK DETECTION n ue c ad le e E Hed can eed aed 48 3 2 OVERLOAD PROTECTION 5 e ent elles te oe Ear eet eet eode eee qr ae Salle e Fo euet e Eva ve Se eue coe 48 5
7. disconnecting it from its AC power outlet Important Note Make sure that all safety information of the instrument respectively on the device itself are clearly visible during operation Attention Do not open the instrument Repairs are only to be carried out by trained service technicians Always connect the instrument to an earthed AC power outlet Ensure that the unit is firmly attached to a stand gt W o qil eepe Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 41 3 Set up 3 1 Mounting of Support Rods Chuck key B Screw the two support rods A into the threaded holes at the P N back of the overhead stirrer 4 gt Support rods Use the chuck B to firmly tighten them 3 2 Attach the Overhead Stirrer to a Stand special clamp C Rod stand D Ira Mount the special clamp to the rod C of the stand D Insert the support rods A of the over head stirrer into the openings of the special clamp C and turn the knurled screws E until the drive unit is firmly attached to the stand knurled screw 3 3 Attachment of Stirring Paddle Insert the stirring paddle G into the chuck F as shown in the picture and firmly tighten it in the desired position Use the chuck key B to do so The immersion depths of the stirring tool may be adjusted by loosening the chuck F and axially moving the stirring tool chuck F Stirring paddle G
8. ltlich Eingangsbereich 0 10V Betriebsbereich programmierbar Erm glicht einfache Drehzahlsteuerung ber ein externes Potentiometer Abgabebereich 0 4V Betriebsbereich von Abgabe und Uberwachungsfunktion Drehzahl Drehmoment oder Leistungsabgabe ist programmierbar 5 40 C innerhalb der Umgebung z B Abzug 80 RH IP42 77 5 x 190 x 170 mm 230 mm inklusive Bohrfutter Page Seite 32 13 3 Technische Daten R100CT Type R100CT Netzspannung LCD Display Handrad 5 Timer Funktion Drehmoment Anzeige Anzeige der Leistungsabgabe Motorspezifikationen Multistep Sicherheitssystem Hilfsfunktionen RS232 Schnittstelle Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Spezifikationen hintergrundbeleuchtetes Display zeigt alle relevanten Systemdaten Gleichzeitige Anzeige der eingestellten und tats chlichen Werte von Drehzahl Drehmoment und Leistungsabgabe Programmierbarer switch off timer programmierbar von 1 Minute bis 99 Tage Aufl sung 1 Minute Zeigt tats chliches Drehmoment am Schaft in Ncm Anzeige des relativen oder absoluten Wertes m glich Genauigkeit 10 Zeigt die tatsachliche Leistungsabgabe am Schaft in Watt Anzeige des relativen oder absoluten Wertes m glich Genauigkeit 1096 Mikroprozessorgesteuert f r einfache Bedienung und Genauigkeit Softstart und Softstop logic verhindert Spritzen Pr zise Einstellung der Motordrehzahl von 10 bis 500 1 min
9. um Ubertragung 1 Typeninformation des einzuleiten Ger tes Controller sendet handshake A Name des angeschlossenen Ger tes B Firmenversionsnummer C ON OFF Z hler D Gesamtbetriebszeit in Minuten Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 20 0 1 Auswahl ob A Control Parameter Drehzahlvorgabe ber Frontpanel 0 Vorgabe der Drehzahl am Drehzahlknopf oder Bedienpanel ber RS232 erfolgen Einschaltzustand soll 1 gt Vorgabe der Drehzahl durch RS232 Befehle gt ber WSE Befehl Verriegeln Freigeben 1 ffnen schlie en der Tastatur 0 gt Frontpanel Tastatur ist deaktiviert auBer On Off Knopf 1 gt Frontpanel Tastatur ist aktiviert Einstellen der RS232 1 Neue Slave Adresse des Ger tes Slave Adresse umnummerieren der Slaves Ger t ausschalten 1 Sicherheitsparameter 1 WEE Set up Parameter 1 Sicherheitsparameter abspeichern 6 5 Programmierbeispiele Die folgenden Anwendungsbeispiele zeigen wie RS232 Befehle an den R hrmotor geschickt werden und wie die Parameter des R hrger tes ausgelesen werden k nnen 6 5 1 Schreiben von Parametern zum R hrmotor Wenn z B die Motordrehzahl auf 500 1 min eingestellt werden soll sind die folgenden zwei Befehlsfolgen an das Ger t mit Slave address 1 zu senden Slave address Befehlscode 1 Parameter hier 1 um den RS232 Modus zu aktivieren und um die Drehzahlvorgabe ber das Bedienpanel
10. vaa ee boe ha Deest dich 21 6 5 1 Schreiben von Parametern zum 21 6 5 2 Auslesen von Parametern aus dem R hrmotor eese eese essen theses as 22 6 6 ANSCHLUSS VON MEHR ALS EINEM GER T AN DEN COMPUTER eere rene 22 7 lt DAS SETUP MEN er etse tenere eee hie eeu ve oett ecl i ede 24 7 1 BEISPIEL NDERUNG DER SETUP EINSTELLUNGEN cerent tnnt tentent renean neret 25 8 FEHLERMELDUNGEN sores ooo aa eee sa o oe Pasar aas ESSERE PASE E stern ERN SERRE FERRO e ERR NR Tennesse vanta 27 9 REINIGUNG UND 27 10 ABBAU TRANSPORT UND LAGERUNG ccccccsssssssssssscccccssssscescccscccsssssessccecscssssssesececscessessesecssscesssees 28 IOT 28 10 2 TRANSPORT UND LAGERUNG evo e NE DR EET RN 28 11 concacvcsvessececcusnnsasevssteecavonessscebcenscevsveccstaasaveseveverdsedsvanssovavencssveosvesecevanesessesvesovecnness 28 12 GEW HRLEISTUNG UND HAFTUNGSAUSSCHLUSS cccssssssssssecsssesscscecsssssccsccececsssssccsseesesssaesceaeeess 29 13 TECHNISCHE DATEN PEU 30 13 1 TECHNISCHE DATEN 00 30 132 TECHNISCHE D
11. ziehen Sie den Netzstecker aus der Steckdose Damit Sie das Ger t problemlos abbauen k nnen entfernen Sie bitte Kolben Beh lter oder andere Ger tschaften um das R hrwerk Spannfutter F Nehmen Sie zuerst das R hrwerkzeug aus dem Ger t Hierzu l sen Sie das Spannfutter Spannfutter F mit Hilfe des Spannfutterschl s schl ssel B sels B Lockern Sie dann die Spezialhalteklammer C und entfernen das Ger t aus dieser Jetzt k nnen Sie die Haltestangen A entfernen R hrwerkzeug G 10 2 Transport und Lagerung Verwenden Sie zum Verpacken des Ger tes wenn m glich die Original Verpackung Sch tzen Sie jedes einzelne Modul mit Luftpolsterfolie gegen St Be von auBen Wenn Sie die einzeln verpackten Module in einen Karton geben achten Sie auf ausreichende Zwischenr ume welche abschlieBend mit D mmmaterial ausgef llt werden sollten Sollten Sie die Original Verpackung nicht verwenden kennzeichnen Sie das Paket von auBen mit folgenden Hinweisen Glassymbol Vorsicht zerbrechlich Regenschirm Vor Nasse sch tzen Verpackungsinhalt Angabe ber den Inhalt F r eine sachgem e Lagerung beachten Sie bitte die folgenden Hinweise 1 F r die Lagerung des Ger tes separieren Sie die einzelnen Module voneinander 2 Verpacken Sie jedes Modul in eine separate luftdichte Plastiktasche 3 Stellen Sie folgende Umgebungsbedingungen sicher Max Umgebungstemperatur bis 40 C Max Luft
12. zu deaktivieren 1 WSM 1 1 Befehl Deaktivieren der Steuerung Uber das Bedienpanel und Aktivierung der Steuerung ber RS232 1 WMS 500 2 Befehl Einstellen der Drehzahl auf 500 1 min Der R100C Controller sendet dann folgende Befehle zum n chsten Ger t oder zur ck an den PC wenn nur ein Ger t angeschlossen ist 1 WSM 1 Echo des 1 Befehls 1 HS OK Handshake Antwort zum 1 Befehl akzeptiert kein Fehler 1 WMS 500 Echo des 2 Befehls 1 HS OK Handshake Antwort zum 2 Befehl akzeptiert kein Fehler Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 21 6 5 2 Auslesen von Parametern aus dem R hrmotor Um die tats chlichen Werte der Drehzahl des Drehmoments und der Leistungsabgabe auszulesen senden Sie folgenden Befehl zum Ger t mit Slave Adresse 1 Slave address 1 Befehlscode RAC Auslesen der Istwerte 1 Parameter hier nur dummy Parameter zur Protokollerf llung mindestens ein Parameter muss gesendet werden 1 RAC 1 siehe auch Tabelle 2 RAC Befehl Das Ger t sendet dann die folgenden zwei Befehlsfolgen an das n chste Ger t oder zur ck an den PC 1 RAC 1 Echo des erhaltenen Befehls 1 HS OK 300 0 50 0 56 0 15 7 17 59 Handshake Antwort mit R ckgabeparametern L absolute Leistungsabgabe Pa in Watt relative Leistungsabgabe Pr in Watt absolutes Drehmoment Ma in Ncm relatives Drehmoment Mr in Ncm 1 Parameter ist aktuelle Drehzahl in
13. 00CT a PC mit 9 pin RS232 Stecker 2 RxD 2 TxD PC R100C 9 pin Sub D 9 pol Sub D plug socket 9 pol Sub D 9 pol Sub D Socket plug b PC mit einem 25 Pin RS232 Stecker PC R100C 25 pin Sub D 9 pin Sub D plug socket 25 pin Sub D 9 pin Sub D socket plug 6 2 Format eines RS232 Befehls Jeder Befehl der an das Ger t geschickt wird muss folgendes Format haben ADR CMDCODE PARAMETERLISTE CR Beschreibung ADR Slave Adresse des Ger ts welche den Befehl ausf hren soll CMDCODE Befehlscode PARAMETERLIST 1 bis 6 Parameter getrennt durch Kommata CR Der gesendete Befehl muss durch Carriage Return ASCII 13 beendet werden Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 18 6 3 Format des Controller Handshake Nach Empfang eines RS232 Befehls wird das Ger t 1 den erhaltenen Befehl an das n chste Ger t oder zur ck zum PC senden ECHO 2 mit einem Handshake antworten welcher wie folgt definiert ist ADR HS RETCODE PARAMETERLIST CR Erkl rung ADR Slave Address des Ger tes das den Handshake sendet RETCODE Fehlercode siehe Tabelle 1 unten PARAMETERLIST 1 bis 6 Parameters jeder Parameter ist durch ein Komma getrennt siehe auch Tabelle 2 Befehl Codes CR der Handshake wird wie jeder Befehl mit ASCII code 13 CR beendet Tabelle Return Codes Tabelle1 Befehl ausgef hrt kein Fehler siehe Tabelle 1 Unbekannter Befehl keine
14. 096 Motor specification torque microprocessor controlled for easy operation and accuracy soft start and soft stop logic prevents splashing of liquid precise adjustable motor speed from 10 to 500 rpm max 220 watts shaft power output extra long life high performance brush motor silent belt drive power transmission max torque 600 Ncm 20 min continuous torque output at 250 rpm 400 Ncm maximum limitation of torque programmable Multistep safety system detects and protects against the following dangerous situations rotor stuck block detection motor overtemperature monitoring self test of all safety functions after switching on Auxiliary functions User programmable limits for maximum motor speed 10 500 rpm maximum allowed torque at shaft 10 600 Ncm Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 67 RS232 interface programmable baudrate 1200 2400 4800 8 N 1 Full remote access of all stirrer functions read and write daisy chain feature by sub addressing allows to connect up to 255 units to only one serial interface on the computer side Windows control software optional available RTI temperature DIN EN 60529 77 5 x195 x 195 mm 5 5 Kg 14 Repairs When returning instruments for repair that have come into contact with hazardous substances please A Fill in attached Repair Return Form Provide precise information on the relevant medium Tak
15. 0C R100CL R100CT wurde gem DIN EN 61010 gefertigt und gepr ft Das Ger t hat das Werk in sicherheitstechnisch einwandfreiem Zustand verlassen Um diesen Zustand zu erhalten und einen gefahrlosen Betrieb sicherzustellen muss der Anwender die Hinweise und Warnvermerke beachten die in dieser Gebrauchsanleitung enthalten sind 1 4 Auspacken des Ger tes Untersuchen Sie den u eren Karton der das Instrument umgibt Achten Sie auf Besch digungen wie z B eingedr ckte W nde chemische Beeintrachtigungen des Kartons Wassermarken oder andere physikalische Einfl sse die den Inhalt besch digt haben k nnten Bei Besch digungen informieren Sie bitte sofort den Spediteur und fragen Sie nach den notwendigen Ma nahmen Wenn es keine sichtbare Besch digung des u eren Kartons gibt dann ffnen Sie diesen vorsichtig 1 2 Lieferumfang und Zubeh r Eine vollst ndige Sendung hat folgenden Umfang 1 R hrwerk R100C Art Nr 60218 00 oder 1 R hrwerk R100CT Art Nr 60314 00 oder 1 R hrwerk R100CL Art Nr 60315 00 2 Haltestangen Art Nr 10402 64 1 Betriebsanleitung Zubeh r und R hrwerkzeuge optional U Stativ Art Nr 60494 00 1 Halteklammer zur Befestigung am Stativ Art Nr 60495 00 1 R hrwellenschutz mit Vorrichtung zur Befestigung am Stativ Art Nr 20 Weiteres Zubeh r wie z B geeignete R hrfl gel finden Sie in unserer Preisliste yee 4 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Versi
16. 0CT RS232 Address Defines the address under which the stirrer can be accessed using RS232 commands factory setting 1 range 1 255 Baudrate Baudrate of the RS232 interface factory setting 1200 Baud options 1200 2400 4800 Baud Max MotorRpm User selectable maximum motor speed in rpm factory setting R100C 2000 rom range 40 2000 rpm factory setting R100CL 1100 rpm range 30 1100 rpm factory setting R100CT 500 rpm range 10 500 rpm Min MotorRpm User selectable minimum motor speed in rpm factory setting R100C 40 rpm range 40 2000 rpm factory setting R100CL 30 rpm range 30 1100 rpm factory setting R100CT 10 rpm range 10 500 rpm Max torque Maximum allowed torque on shaft in Ncm factory setting R100C 150 Ncm Range 10 150Ncm Maximum allowed torque on shaft in Ncm factory setting R100CL 250 Ncm Range 10 250Ncm Maximum allowed torque on shaft in Ncm factory setting R100CT 600 Ncm factory setting 600 Ncm Range 10 600Ncm Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 57 CutOff limit Cut off limit sensitivity in percent rotor stuck detection If the torque increase versus time exceeds a certain limit the unit will assume that the rotor was suddenly blocked If this happens the unit will shut down immediately factory setting 6096 range 10 100 10096 disables this safety check The lower the entered value the more sensitive the unit will react on sudden
17. 1 min Return code s Tabelle 1 OK kein Fehler Beschreibt dass dies ein handshake ist und kein Befehl handshake Bezeichner Slave Adresse des Ger tes das den den handshake sendete HINWEIS Jede Befehls oder Handshake Folge wird mit ASCII Code 13 CR beendet 6 6 Anschluss von mehr als einem Gerat an den Computer Das leistungsstarke Protokoll der RS232 Schnittstelle l sst auch zu dass andere CAT Laborger te wie z B Magnetr hrer M26 MCS66 Mikrodosierpumpen HPLH Serie Dispenser Liquid Processor an nur eine RS232 Schnittstelle auf Computerseite angeschlossen werden k nnen Durch Zuweisung unterschiedlicher Slave Adressen der angeschlossenen Ger te hat der Computer individuellen Zugriff F r den Anschluss von mehr als einem Ger t an den Computer wird ein spezielles Y Kabel f r jedes zus tzliche Ger t benutzt RS232 Y Kabel Art Nr 60709 00 Zur Anwendungsanpassung ist ein Windows Programm optional erh ltlich Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 22 Anwendungsbeispiel 11117777777 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 23 7 Das Setup Men Generelle Einstellungen werden in einem nichtfl chtigen Speicher EEPROM gesichert Zur Anderung der Einstellung dr cken und halten Sie die Select Display Taste 4 w hrend das Ger t mit der
18. 2 VDE 0839 6 2 2006 03 DIN EN 61000 6 4 VDE 0839 6 4 2007 09 DIN EN 61000 4 2 VDE 0847 4 2 2009 12 DIN EN 61000 4 3 VDE 0847 4 3 2008 06 DIN EN 61000 4 4 VDE 0847 4 4 2005 07 DIN EN 61000 4 5 VDE 0847 4 5 2007 06 DIN EN 61000 4 6 VDE 0847 4 6 2009 12 DIN EN 55011 VDE 0875 11 2010 05 DIN EN 55014 1 VDE 0875 14 1 2010 02 DIN EN 55014 2 VDE 0875 14 2 2009 06 2006 42 EC 93 68 EEC DIN EN ISO 12100 1 2004 04 DIN EN ISO 12100 2 2004 04 In the case of a modification of the unit which has not been agreed on with us this declaration becomes null and void and the warranty expires Ingenieurb ro CAT M Zipperer GmbH Ellen Gebel CE Dokumentenbeauftragte 79219 Staufen den 12 Juli 2010 Type R100C R100CL R100CT 230V 10 50 60Hz 300W Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 70 16 Appendix A 16 1 Displaying the system info For additional user information the R100C tracks the total operation time and the number of 8 press and hold the Timer key whilst switching on the R100C The unit initialises and then displays the System Info screen Press any key to display the next info The system Info is shown on the LCD display and contains the following information 1 total operation time of the unit days hours minutes 2 number of times the R100C R100CL R100CT has been turned on Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Ve
19. 3 4 5 Ger t einschalten Achtung Achten Sie vor der Verbindung des Ger tes mit dem Stromnetz darauf dass die Netzspannung mit den Angaben auf dem Typenschild bereinstimmt Achtung Achten Sie darauf dass der Drehzahlregler 7 auf Linksanschlag gedreht ist Achtung Das Ger t immer nur mit der ein aus Taste 9 abstellen Das Ger t ist nur dann vollstandig von der Stromquelle getrennt wenn Sie den Netzstecker ziehen gt gt e Dr cken Sie die Power Taste 9 Drehzahl am Drehzahlregler 7 einstellen Der digitale Wert der eingestellten Motordrehzahl kann in der oberen Zeile der Anzeige 1 abgelesen werden HINWEIS Die maximal zul ssige Motordrehzahl kann im Setup Men eingestellt begrenzt werden Das R hrwerk R100C R100CL R100CT ist mit soft start stop zur Vermeidung von Fl ssigkeitsspritzern versehen Achtung Sorgen Sie daf r dass die Fl ssigkeit nicht aus dem Beh lter geschleudert A werden kann Benutzen Sie ein geeignetes Gef bei dem die Gef w nde noch gen gend Uber den Fl ssigkeitsstand hochragen Manchmal ist ein geeigneter Deckel als Spritzschutz sinnvoll Wenn Sie mit Spritzern rechnen m ssen ziehen Sie Schutzkleidung an und tragen Sie eine Schutzbrille 4 6 Zeitgesteuerter Betrieb des R hrwerks Ein interner Zeitgeber kann durch Dr cken der Timer Taste 3 aktiviert werden Dieser Timer erm glicht das automatische Abschalten des gesamten Ger tes Diese Abschaltzeit
20. ATEN R1O0GQLE iine ener ara cre rie ve te ce APVT TREE ihnen ng 31 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 2 13 3 TECHNISCHE DATEN reete ea feoda nennen 33 14 34 14 1 jR CKSENDEEFORMULAR e aia A V rie C ERE NA V Er cR EU 35 15 KONFORMITATE 36 16 APPENDIX qr 37 16 1 ANZEIGE DER SYSTEMINFORMATION ccccccccsccecesescseseseseseseseeeseueseseseseseueseueueseseeeueeeueueseeeueueseeeueueueeeseueueueneneners 37 1 GENERATE INFORMATION 6 cssssssssccecccscassssissececsescesscsdesscovsddessseseesessudacetnascesossvedecaansesedasegocscacasesedesesoancaseabiese 39 1 1 UNPACKING THE INSTRUMENT T rU deter perte vto oe Ee v P de Poetas 39 1 2 SCOPE OF DELIVER Y u 4 cid teu d Gee d Sod b nnt dbi n Nok cad ad e ciet a 39 2 SAFETY INFORMATION scscccisesssvestecsceutscscesssdesscoucacvossessevecwucoevssessuauessvonsv esssaauscesbvsscssedsanesevesescsecasseecvvoosvens 40 2 1 EXPLANATION OF THE SAFETY INFORMATION IN THE INSTRUCTION 40 2 2 EXPLANATION OF THE SAFETY INFORMATION ON THE 41 2 3 GENERAL SAFETY INFORMATION deer repente dg
21. CAT Ingenieurburo M Zipperer GmbH Etzenbach 16 D 79219 Staufen Betriebsanleitung Ruhrwerke R100C R100CL und R100 CT CAT R100C e Select Timer Display x t Load Change O 6 9 NN ur R100C R100CL R100CL R100CT mikroprozessorgesteuertes R hrwerk hohe Leistungsabgabe hinterleuchtetes alphanumerisches LCD Drehmoment und Leistungsanzeige gleichzeitige Anzeige von Drehzahl und Drehmoment Fuzzylogic Regelung Soft motor start stop einstellbare Sicherheitsabschaltfunktion programmierbare max Motordrehzahl programmierbare Drehmomentbegrenzung switch off Timer 1 Min 99 Tage bertemperaturschutz RS232 Schnittstelle zur Computersteuerung analoger Spannungseingang f r einfache externe Drehzahlsteuerung analoger Ausgang f r Analogwertschreiber Inhalt 1 ALLGEMEINE INFORMATION EN de Res RR A 5 1 1 AUSPACKEN DES GER TES te i Der ere decer e a E e vea eds debis reos oe Eoo oe o LE Dex a OS 5 1 2 LIEFERUMFANG UND ZUBEH R a 5 2 SICHERHEFTSHEN WEISE i ciicccscostsosisccoceceossecssdecvaceseosvecoodecsevessssvecedaeesiccoodcactsodssoacsasacasanesscesccaasseaesosecescc
22. Power On Off Taste 9 eingeschaltet wird Die folgenden Einstellungen k nnen im Setup Men des R100C R100CL R100CT justiert werden RS232 Address Max MotorRpm CutOff limit Analog input Definiert die Adresse unter welcher der R hrer ber die RS232 angesprochen werden kann Werkseinstellung 1 Bereich 1 255 Baudrate der RS232 Schnittstelle Werkseinstellung 1200 Baud Optionen 1200 2400 4800 Baud Benutzer w hlbare max Motordrehzahl in 1 min Werkseinstellung R100C 2000 rpm Bereich 40 2000 rpm Werkseinstellung R100CL 1100 rpm Bereich 30 1100 rpm Werkseinstellung R100CT 500 rpm Bereich 10 500 rpm Maximales Drehmoment am Schaft in Ncm Werkseinstellung R100C 100 Ncm Bereich 10 150Ncm Werkseinstellung R100CL 250 Ncm Bereich 10 250Ncm Werkseinstellung R100CT 600 Ncm Bereich 10 600Ncm Abschaltbegrenzungssensibilit t in Prozent gt Rotorblockade Erkennung Uberschreitet der Drehmomentanstieg Uber die Zeit eine bestimmte Grenze nimmt das Ger t an dass der Rotor pl tzlich blockiert wurde und schaltet sofort ab Werkseinstellung 5096 Bereich 10 100 10096 deaktiviert diesen Sicherheitstest HINWEIS Je niederer der eingegebene Wert desto sensibler reagiert das Ger t auf pl tzliche Drehmomenterh hungen Zulassen oder Sperren der Drehzahleinstellung durch externe analoge Volt Steuerung Werkseinstellung deaktiviert M gliche Optionen aktiviert deaktiviert Betriebsanleit
23. R100CL R100CT 07 2010 Version 1 07 Page Seite 48 5 5 Overtemperature protection R100CT If the temperature inside the unit exceeds the maximum operating temperature the unit will automatically reduce power output speed and torque in order to get back to a safe operating point fold back characteristic This prevents the unit to be destroyed by constant overload operation This overload condition is also signalled by the red Load LED 6 Furthermore a cooling fan 11 on the rear side of the instrument is automatically switched on if required If the temperature inside the R100CT exceeds 70 by any means the unit will shut down immediately This safety feature is fully redundant 5 6 Connector 9 pin SUB D Pinout of the I O connector The following signals are led out on the 9 pin SUB D connector on the rear side of the instrument RS232 control signals Analogue input 0 10V for speed override Analogue output 0 4V for recording of process parameters speed torque power Pin Number Description 1 5V max 10mA 2 TxD RS232 transmit data 3 RxD RS232 receive data 4 n c 5 GND signal ground 6 analogue output 0 4V 7 n C 8 n C 9 analogue output 0 10V 1 6 5 9 e socket of the I O connector view on socket 9 pin Sub D socket Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 49 6 The RS232
24. R100CL R100CT you will find 2 threads a to attach the support rods A a 9 pin SUB D connector 10 a fan and the mains power supply cable 12 support rod A Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 46 4 5 Stirring operation CAUTION The units are not to be used in environments with danger of explosion CAUTION The unit is not to be used without supervision CAUTION When connecting the instrument to an AC power outlet ensure that your local supply voltage matches that indicated on the instrument s rating plate gt gt amp Press the power on off button 9 set the desired motor speed using the speed setting knob 7 The digital value of the set motor speed can be read out in the top line of the display 1 The maximum allowed motor speed can be adjusted limited in the setup menu The R100C R100CL R100CT has a soft start feature which helps to prevent the liquid from splashing Attention Avoid splashing of liquids by using a suitable vessel lower liquid level Make sure A that suitable protective clothing and eye wear is used This instruction sheet does not purport to address all of the safety problems which might result from the use of this instrument chemicals reagents apparatus or equipment employed in any specific test or protocols It is the responsibility of the user to consult their authorised safety advisors and establish appropr
25. ahrleistungserklarung enthaltenen verpflichtet Insbesondere sind Personen und Sachsch den die als Folge von fehlerhaften Teilen oder sonstigen Funktionsfehlern auftreten von der Haftung ausgeschlossen Die Verpflichtung des Herstellers ist im Umfang begrenzt auf die Reparatur oder den Austausch von fehlerhaften Teilen Der Hersteller ist unter keinen Umstanden zu Schadensersatz verpflichtet f r Folgesch den jeder Art die aus der Verwendung und dem Besitz dieses Produktes sich ergeben k nnen Achtung Der Benutzer hat die Aufgabe selbst festzustellen ob dieses Ger t f r seinen speziellen Anwendungsfall geeignet ist Bei Fragen wenden Sie sich bitte an Ihren Handler oder direkt an den Hersteller Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 29 13 Technische Daten 13 1 Technische Daten R100C Type R100C Netzspannung LCD Display Handrad 5 Timer Funktion Drehmoment Anzeige Anzeige der Leistungsabgabe Motorspezifikationen Multistep Sicherheitssystem Hilfsfunktionen RS232 Schnittstelle Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Spezifikationen hintergrundbeleuchtetes Display zeigt alle relevanten Systemdaten Gleichzeitige Anzeige der eingestellten und tats chlichen Werte von Drehzahl Drehmoment und Leistungsabgabe Programmierbarer switch off timer programmierbar von 1 Minute bis 99 Tage Aufl sung 1 Minute Zeigt tat
26. ay key 4 Pressing the Power On Off key 9 from within the setup menu will prevent the unit to store eventually changed settings to non volatile memory After pressing the Select Display key 4 the unit will ask if the settings should be stored to non volatile memory To confirm this question press Select Display 4 again If the parameters should not be stored turn the handwheel 5 until NO is displayed and then press the Select Display button 4 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 59 CAN RIOOC LCD Display 1 Timer LED 2 Display select Timer button 4 button 3 a Load LED 6 Incremental FA encoder wheel 5 Load Speed control knob 7 Power LED 8 Power on off button 9 view on the R100C H100CL H100CT front panel Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 60 8 Error messages The R100C R100CL R100CT overhead stirrer has an on board self diagnostic program that detects possible failures and should one occur the R100C R100CL R100CT shuts itself down To enable the user to prevent the problem re occurring the R100C R100CL R100CT will display any of the following error messages as it is switched on again Please read carefully the following table of error messages Error Reason for shut down message Time out the programmable timer has expired timer ke
27. be sent to the unit with slave address 1 Slave address Command code 1 parameter here 1 to enable serial mode and to disable front panel control of motor speed 1 WSM 1 1 command disable front panel control and enable RS232 control 1 WMS 500 2 command set motor speed to 500 rpm The R100 Controller then sends the following strings to the next unit or back to the PC if only one device is connected 1 WSM 1 Echo of the 1 Command 1 HS OK Handshake response to the 1 Command accepted no error 1 WMS 500 Echo of the 2 command 1 HS OK Handshake response to the 2 Command accepted no error Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 54 6 5 2 Reading parameters from the stirrer To read the actual values of motor speed torque and power output send the following command to the unit with SlaveAdress 1 Slave address 1 Command code RAC Read actual settings 1 parameter here only dummy parameter to fulfil protocol at least one parameter must be sent 1 RAC 1 see also table 2 RAC command The Controller then sends the following two strings to the next unit or back to the PC 1 RAC 1 Echo of the received command 1 HS OK 300 0 50 0 56 0 15 7 17 59 Handshake response with return parameters L absolute power output Pa in Watt relative power output Pr in Watt absolute torque Ma in Ncm relative torque Mr in Ncm 1 parameter i
28. des Ger tes am Stativ Jetzt f hren Sie die Aus sparungen K an die Befestigungsschrauben J und drehen den R hrwellenschutz bis zum An schlag der Aussparungen nach links Die H he der beiden Halbschalen k nnen sie mit Hilfe der Schraube L verstellen berpr fen Sie vor jeder Inbetriebnahme und in regelm Bigen Abst nden den festen Sitz des R hrwellenschutzes Die Position des R hrwellenschutzes darf nur im Stillstand und mit gezogenem Netzstecker verandert werden 4 1 Bestimmungsgem er Gebrauch Der R hrmotor R100C R100CL R100CT ist f r den qualifizierten Gebrauch im Labor bestimmt Er eignet sich zum R hren und Mischen von Fl ssigkeiten mittlerer bis hoher Viskosit t Zum bestimmungsgem en Gebrauch muss das Ger t an einem Stativ befestigt werden Die Grundvoraussetzung f r den dauerhaften Einsatz ist die Einhaltung der maximal zulassigen Umgebungstemperaturen korrosionsfreie Atmosphare Temperatur und Feuchtigkeit sowie eine Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 10 4 2 Das Bedienfeld Das Bedienfeld des R100C R100CL R100CT erleichtert die Dateneingabe und dient zur visuellen Darstellung aller relevanten Daten auf einem zweizeiligen alphanumerischen LCD Display 1 Die Eingabe und Anderung von Einstellungen erfolgt ber das linke Handrad 5 in Verbindung mit Funktions und Hotkeytasten Die Drehzahl des R hrmotors wird ber den rechten Drehk
29. dress Zum daisy chaining ist die Sendeleitung TxD des PC an die Empfangsleitung RxD des ersten Ger tes angeschlossen Die Sendeleitung dieses Ger tes wird dann an die Empfangsleitung des nachsten Ger tes in der Kette angeschlossen Die Sendeleitung des letzten Ger tes in der Kette wird wieder zur Empfangsleitung des PC zur ckgef hrt wodurch die Kette geschlossen wird Ring Um ein individuelles Ger t in der Kette zu adressieren tr gt jedes Ger t eine so genannte Slave address Wertebereich 1 255 Als Standard ist die Adresse 1 eingestellt Die slave address 0 ist als allgemeine Adresse definiert auf die alle Ger te antworten Die slave address kann im Setup Menu eingestellt verandert werden Alle Ger te in einer Kette m ssen auf die gleiche Baud Rate eingestellt sein Ein Windows Software Paket zur Steuerung eines Ger tes ist optional erh ltlich Ge 1 6 5 9 SSS Pinbelegung der seriellen Schnittstelle Draufsicht 9 pin Sub D Buchse Parameter der seriellen Schnittstelle 1200 4800 8 N 1 Baudrate 1200 4800 Baud 1200 Standard siehe auch Set up Men Databits 8 Bit Parity none Stopbits 1 Pinausgang des RS232 Anschlusses TxD Sendeleitung RxD Empfangsleit ung 2 GND 5 Masse SUB D Buchse Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 17 6 1 RS232 Anschlusskonfiguration zwischen PC und R100C R100CL R1
30. e Seite 36 16 Appendix A 16 1 Anzeige der Systeminformation Zur zus tzlichen Benutzerinformation z hlt der R100C R100CL R100CT die Gesamtbedienzeit und die Haufigkeit der Ein Aus Schaltungen e Drucken und halten Sie die Timer Taste 3 wahrend des Einschaltens des R100C R100CL R100CT Das Ger t initialisiert und zeigt dann das System Info screen Dr cken Sie eine beliebige Taste um die nachste Information zu sehen Die System Info wird auf dem Display 1 angezeigt und beinhaltet folgende Information 1 Gesamtbetriebszeit des Ger tes Tage Stunden Minuten 2 H ufigkeit der Einschaltungen des R100C R100CL R100CT Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 37 Instruction Manual Stirring motor R100C R100CL and R100CT R100C R100CL and R100CT microprocessor controlled overhead stirrers e high power output CAT R100C e back lit alphanumeric LC Display HERR e Torque and power readout e Timer x t e Fuzzylogic control ed e Soft motor start stop Change O e adjustable safety cut off function e programmable max motor speed Simultaneous display of speed and torque e Power 5 e programmable min motor speed e programmable torque limitation MEME e switch off Timer 1 min 99 days e Overtemperature protection e RS232 interface for computer control e analogue output for recorder Be
31. e protective measures to ensure the safety of our receiving and maintenance personnel Mark the package as appropriate for hazardous materials Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 68 14 1 Repair Return Form CONTACT USER INFORMATION Contact Phone No Fax No Email Billing Shipping Company Company Address Address INSTRUMENT INFORMATION Model Serial No Please describe all problems malfunctions Operating Conditions please fill in if applicable Ambient Temp Humidity Speed Load Volume Viscosity Temperature in Sample Temperature Operating Time Sample Description NOTE If the instrument was exposed to hazardous material it must be decontaminated BEFORE returning itto CAT and an MSDS for hazardous material must be included with the instrument RETURN SHIPPING UPS Air Parcel Post U Collect Other Your account number is required for UPS collect respectively the address and contact of your preferred forwarder if you choose any other transport means PACKAGING INSTRUCTIONS TO RETURN A CAT INSTRUMENT FOR REPAIR Y Remove all accessories e g homogniser tools stirring paddles from the instrument Y Clean excess testing material off the instrument accessory Y Include MSDS sheets for all hazardous materials used with this instr
32. ececeeeceseresereresenens 14 5 SICHERHEITSEUNKTIONUBEN o esas osoo oen aerea assa su 0080 80800008000 15 5 1 ROTORBLOCKADE 15 52 JUBBRLASTSSCHUTZ exin Re Un ciem e a Torsone m DR AM RUD ae 15 5 3 BERTEMPERATURSCHUTZ R 100C esses thee tenista tree tenter 15 5 4 BERTEMPERATURSCHUTZ R 100CT cccccsscsececscceceseccscessscscessecscessscscessacscessacseessacscessacsesetecseeseacsceasarseeseacseessers 15 5 5 BERTEMPERATURSCHUTZ R1OOCL c ccccesscsecessececescecscccecsceevscscessscscsevscscessacsesevacsceaacseeavacaeestacseeevaraeestarseeevers 16 5 6 I O ANSCHEUSS 9 PIN SUB D u anna sinne en neh sn ini 16 6 DIE RS232 SCHNITTSTELLE DES R100C R100CL R100CT essere neenon 17 6 1 RS232 ANSCHLUSSKONFIGURATION ZWISCHEN PC UND R100C R100CL R100CT eer 18 6 2 FORMAT EINES RS232 BEFEHLS cerea e Peta bo eee habere LUCERE EE Va ec ke e LUE ee 18 6 3 FORMAT DES CONTROLLER HANDSHAKE eere rti itr 1 1 1 1 1 nene 19 6 4 RS232 BEFEHLE DES R100C R100CL R10OOCT eese nennen 20 6 5 PROGRAMMIERBEISPIELE u sa eder reported ipe exe
33. eed in the setup menu press and hold the Select Display key 4 whilst switching on the R100C R100CT The unit initialises and then switches into the Setup menu The Select Display key 4 must be pressed down until the Setup menu appears on the display turn the handwheel 5 until MotorRpm occurs on the display press the Timer key 3 to edit the max motor speed gt Edit arrow appears turn the handwheel 5 to adjust the value press the Select Display key 4 to leave the setup menu press the Select Display key 4 once again when the units asks to save the setup parameters Important Note The different options parameters in the setup menu can be displayed selected by simply turning the handwheel 5 To change a displayed value press the Timer key 3 and an edit arrow gt will appear in front of the parameter to be changed As long as the edit arrow is on the display turning the handwheel 5 will change the displayed value If the handwheel 5 is not touched for approx 3 seconds the edit arrow will disappear and an other menu option can be selected by turning the handwheel 5 When editing the setup parameters the Select Display button 4 is used to leave the setup menu or to confirm a selection The Timer button 3 is used to edit change a displayed value The Setup menu can only be left by pressing the Power On Off key 9 or the Select Displ
34. emove the clamp from the support rods Unscrew the support rods A Now you may remove the instrument from the working area huck key B chuck F Stirring tool G 10 2 Transport and Storage Prior to transport Switch the instrument off and unplug the power supply Place the instrument and its parts in its original packaging or another suitable container to protect it during transport Close the packaging with adhesive tape Store the instrument in a dry environment Please observe the specified conditions of the ambient temperature and humidity Do not subject the instrument to mechanical shocks or vibration during transporting it In case you do not use the original packaging please mark the box with the following notes e Glass symbol handle with care fragile e Umbrella keep dry e Content list of content Storage ambient Max ambient temperature RT to 40 C Max humidity 80 11 Disposal recycling collection point Prior to disposal sort according to materials metal glass plastic etc Also be sure to dispose of the packing material in an Please dispose of used instruments and defective components at your local environmental friendly manner Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 62 12 Warranty and Liability The manufacturer agrees to either repair or replace at the manufacturer s discretion any defects in materials or workma
35. en Sie der R cklieferung die entsprechenden Sicherheitsdaten bl tter bei R CKSENDUNG UPS Paketpost Andere 0 Bitte nennen Sie uns Ihren Spediteur Anschrift Sachbearbeiter in sowie dessen deren Telefonnummer VERPACKUNGSHINWEISE v Bitte entfernen Sie s mtliches Zubeh r Z B Dispergierwerkzeuge R hrpaddel etc Y Reinigen Sie das Ger t und Zubeh r von Probematerial R ckst nden Y Legen Sie dem Ger t die entsprechenden Sicherheitsdatenbl tter f r eventuell benutzte gef hrliche Stoffe bei Y Verpacken Sie das Ger t in die Originalverpackung Sollte diese nicht mehr vorhanden sein achten Sie darauf dass das Ger t sowie Zubeh r ausreichend gesch tzt verpackt wird Y Plattenstativ und Halteklammer bitte nicht einschicken Diese senden Sie nur zur ck wenn sie besch digt sind Um Transportsch den zu vermeiden achten Sie bitte darauf dass Ger t und Zubeh r individuell verpackt werden Y Verpacken Sie das Ger t mit dem Zubeh r in einen ausreichend starken Versandkarton Kennzeichnen Sie dessen Au enseite mit Beispiel Vorsicht oder Vorsicht zerbrechlich und schicken Sie die Waren an Ingenieurb ro CAT M Zipperer GmbH Service Department Etzenbach 16 D 79219 Staufen Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 35 Fax 49 0 7636 780345 Ingenieurb ro E Mail info cat ing de M Zipperer GmbH Internet http
36. ers achten Sie bitte unbedingt darauf dass Schutzleiter am neuen Stecker angeschlossen wird 2 Sicherheitshinweise 2 1 Erl uterung der Sicherheitshinweise in der Gebrauchsanleitung Pob eR Dieses Symbol hebt folgende Hinweise hervor welche durch den Anwender unbedingt beachtet werden m ssen Jegliche Missachtung der nebenstehenden Hinweise kann die sichere Funktion wie die Sicherheit des Anwenders selbst beintr chtigen Dieses Symbol hebt Verbote hervor welche durch den Anwender unbedingt beachtet werden m ssen Jegliche Missachtung der nebenstehenden Verbote kann die Funktionst chtigkeit sowie die Sicherheit des Anwenders erheblich beintr chtigen Dieses Symbol hebt Hinweise hervor welche durch den Anwender unbedingt beachtet werden sollten um einen sicheren Betrieb des Ger tes gew hrleisten zu k nnen Achtung Brand oder Explosionsgefahr Hinweis zur Reparatur Wartung Hinweis zum Netzanschluss 2 2 Erl uterung der Sicherheitshinweise auf dem Ger t gt Warnung vor allg Gefahren Dieses Symbol weist darauf hin dass die Betriebsanleitung unbedingt vor Inbetriebnahme des Ger tes vollst ndig gelesen und verstanden werden sollte Eine unsachgem e Handhabung kann den sicheren Betrieb des Ger tes und die Gesundheit des Anwenders gef hrden 2 3 Allgemeine Sicherheitshinweise gt gt Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 7 Achtung Beac
37. esses 7 2 1 ERL UTERUNG DER SICHERHEITSHINWEISE IN DER GEBRAUCHSANLEITUNG 7 2 2 ERL UTERUNG DER SICHERHEITSHINWEISE AUF DEM GER T 7 2 3 ALLGEMEINE SICHERHEITSHINWEISE coset ereinen iiia i eta tereti a 7 SE VD URS 9 3 1 HALTESTANGEN MONTIEREN scence 2008 E d e Der erp E GE ere reU ELE ae eee vex aUos age Dre c Ecce ve e es o Dev 9 3 2 R HRWERK AM STATIV BEFESTIGEN eere itt ttt 1 1 1 1 1 1 1 1 1 1 1 1 1 1 E E E E E E rere 9 3 3 R HRWERKZEUG MONTIEREN sssssesesssssssssessscsssssssessscsssesesesesesesssesesssesesseesesesesesesessenetetsneeeneesueeseseeeerereneeenenenens 9 3 4 MONTAGE DES R HRWELLENSCHUTZES S e a aosida aosi iori iaaiaee iaaa 10 4 BEDIENUNG UND BETRIEB eso so eoe sea sop ena tn eoa saets esq aa tut ensei 10 4 1 BESTIMMUNGSGEMABER GEBRAUCH eren tia oeir ino ei iieii i nnn 10 4 2 lt lt T MM 11 4 3 BESCHREIBUNG DES BEDIENFELDS eere rennen rere ne neret nnn 12 4 4 BESCHREIBUNG DER R CKSEITE DES R100 C R100CL R100 13 4 5 GERAT EINSCHALTEN DEE i A du eee uc ea ae E E Ie os OO A Eve Ee ERE 14 4 6 ZEITGESTEUERTER BETRIEB DES R HRWERKS ccccsesesecesecececececececececececececececececececececececececececec
38. feuchtigkeit 80 11 Entsorgung den daf r vorgesehenen Recycling Stellen abgegeben werden Auch das Verpackungsmaterial sollte umweltgerecht Materialtrennung entsorgt werden Schadhafte und oder entsorgte elektrische oder elektronische Ger te m ssen an Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 28 12 Gew hrleistung und Haftungsausschluss Der Hersteller verpflichtet sich dem Kaufer gegen ber dieses Produkt entweder durch Reparatur oder nach Wahl des Herstellers durch Austausch jeden Fehler in Material oder Verarbeitung zu beheben der sich innerhalb von 24 Monaten nach Auslieferung dieses Produktes herausstellt F r den Fall des Austausches wird die Gew hrleistung weiterhin bis zum Ende der 24monatigen Laufzeit ab Kaufdatum laufen F r allfalligen Kundendienst ist der Handler oder der Hersteller einzuschalten Diese Gew hrleistung wird nicht gew hrt wenn der Defekt oder die Fehlfunktion verursacht wird durch Unfall Nachl ssigkeit unsachgem en Gebrauch unsachgemaBen Service und andere Gr nde die nicht auf fehlerhafte Materialien oder Verarbeitung zur ckzuf hren sind S mtliche Gew hrleistung und Gew hrleistungsanspr che sind beschr nkt auf die fachgerechte und sachgem e Anwendung des Ger tes unter Beachtung der allgemeinen Sicherheitsvorschriften und dieser Betriebsanleitung Der Hersteller ist zu keinerlei anderen Leistungen als den in dieser Gew
39. g dangerous situations rotor stuck block detection motor overtemperature monitoring self test of all safety functions after switching on Auxiliary functions User programmable limits for maximum motor speed 30 1100 rpm maximum allowed torque at shaft 10 150 Ncm RS232 interface programmable baudrate 1200 2400 4800 8 N 1 Full remote access of all stirrer functions read and write daisy chain feature by sub addressing allows to connect up to 255 units to only one serial interface on the computer side Windows control software optional available Permissible ambient 5 40 within operating area e g fume cupboard temperature Permissible humidity 80 RH Safety class acc to IP42 DIN EN 60529 Case dimensions Length including chuck 230 mm 77 5 x 190 x 170 mm Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 66 13 3 Technical Data R100CT Electrical power 110 230V 300W 50 60 Hz see R100CT rear requirements panel LCD Display bright back lit LCD display showing all relevant System data Simultaneous display of set and real values of motor speed torque or power output system settings 1 minute to 99 days Resolution 1 minute Torque output display displays true torque at shaft in Ncm relative or absolute readout possible Accuracy 1096 Power output display displays true power output at shaft in watts relative or absolute readout possible Accuracy 1
40. hten Sie ale im Labor geltenden Sicherheits und Unfallverh tungsvorschriften Achtung Verwenden Sie dieses Ger t nicht in unsicherer Umgebung und speziell nicht in explosionsgef hrdender Umgebung Lebensgefahr gt gt gt Hw be gt Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 8 Achtung Ausschlie lich unterwiesene Anwender d rfen das Ger t in Betrieb nehmen Achtung Achten Sie vor der Verbindung des Ger tes mit dem Stromnetz darauf dass die Netzspannung mit den Angaben auf dem Typenschild bereinstimmt Achtung AuBerste Vorsicht beim Umgang mit leicht entz ndlichen Medien Beachten Sie die Sicherheitsdatenbl tter Das Ger t ist nicht explosionsgesch tzt Der Hauptschalter trennt das Ger t nicht vollstandig von der Stromquelle Um das Gerat vollstandig vom Netz der Stromversorgung zu trennen ziehen Sie bitte den Netzstecker Hinweis Stellen Sie sicher dass alle Sicherheitshinweise des Gerates bzw auf dem Ger t selbst wahrend des Betriebs deutlich zu sehen sind Achtung Offnen Sie das Ger t nicht Reparaturen sind eingewiesenen Service Technikern vorbehalten Bitte schlieBen Sie das Ger t nur an eine geerdete Steckdose an Bitte achten Sie auf sichere Befestigung des Ger tes 3 Aufbau 3 4 Haltestangen montieren Spannfutterschl ssel B Haltestangen A Schrauben Sie die beiden Haltestangen A in den daf r vorgesehenen Gewi
41. iate health and safety practices and then determine the application of regulatory limitations prior to use 4 6 Timer controlled operation The Timer function can be activated by pressing the Timer key 3 The Edit arrow appears in the display 1 and the time can be changed in increments of one minute by turning the encoder wheel 5 If the time is set to zero the timer function is disabled If for example a time of ten minutes is selected 00 10 00 the unit will shut down after ten minutes The readout of the timer is in days hours minutes press Timer 3 and then turn the encoder wheel 5 to set the desired shut down time The Timer LED signals if the timer is running Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 47 5 Safety Functions 5 1 Rotor stuck detection If the torque increase versus time exceeds a certain limit the unit will assume that the rotor was suddenly blocked and will switch off immediately Cut off limit sensitivity can be adjusted in the Set up menu gt Cut off limit 5 2 Overload protection If the torque or power drawn from the stirring motor exceeds the allowed range the unit will automatically reduce the motor speed until the torque is reduced to the allowed range If this happens the LED Load 6 is switched on As long as the Load LED 6 is on the motor is not able to hold the set motor speed 5 3 Overtemperature protection R100C If the tem
42. igten Wert zu ndern dr cken Sie die Timer Taste 3 und ein Editierpfeil gt erscheint vor dem zu andernden Parameter Solange der Editierpfeil auf dem Display 1 erscheint wird durch Drehen des Handrads 5 der angezeigte Wert verandert Wird das Handrad 5 ca 3 Sekunden nicht ber hrt verschwindet der Editierpfeil und eine andere Men option kann durch Drehen des Handrads 5 ausgew hlt werden Beim Editieren des Setup Parameter wird die Select Display Taste 4 zum Verlassen des Setup Men s benutzt oder um eine Auswahl zu best tigen Die Timer Taste 3 wird benutzt um einen angezeigten Wert zu editieren ndern Das Setup Men kann nur durch Dr cken der Power On Off Taste 9 oder der Select Display Taste 4 verlassen werden Dr cken der Power On Off Taste 9 innerhalb des Setup Men s verhindert dass das Ger t eventuell geanderte Einstellungen im nichtfl chtigen Speicher sichert Nach Dr cken der Select Display Taste 4 fragt das Ger t ob die Einstellungen im nichtfl chtigen Speicher gesichert werden sollen Um dies zu bestatigen dr cken Sie Select Display Taste 4 nochmals Sollen die Parameter nicht gespeichert werden drehen Sie das Handrad 5 bis angezeigt wird und dr cken erst dann die Select Display Taste 4 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 26 8 Fehlermeldungen Der R100C R100CL R100CT R hrmotor besitzt ei
43. interface of the R100C R100CL R100CT The RS232 interface enables remote control of all stirrer functions In this configuration a computer can access each connected unit via a sub address daisy chaining This address is programmable for every unit gt Setup menu slave address For daisy chaining the transmit line TxD of the PC is connected to the receive line RxD of the first unit The transmit line of this unit is then connected to the receive line of the next unit in the chain The transmit line of the last unit in the chain is returned to the receive line of the PC which closes the link ring For addressing a specific unit a daisy chain each controller carries a so called Slave address which can be any number from 1 to 255 The default slave address is 1 The slave address 0 is defined as the general call address on which all controllers will respond The slave address can be set changed in the Setup menu A WindowsQO Software package for control of a connected unit is optional available 1 e 6 5 9 socket of the serial interface view on socket 9 pin Sub D socket Parameters of the serial interface 1200 4800 8 N 1 Baudrate 1200 4800 Baud 1200 default see also Set up menu Databits 8 Bit Parity none Stopbits 1 Pinout of the RS232 connector Mb SE ER TxD transmit data RxD receive data 2 GROUND female 9 pin SUB D socket Betriebsanleitung Instructi
44. it Important Note Read this instruction manual carefully before operating the instrument Should there be any additional questions after reading these instructions concerning the set up operation or warranty please contact either your distributor or the manufacturer at the following address CAT Ingenieurb ro CAT M Zipperer GmbH Etzenbach 16 D 79219 Staufen 49 7636 7803 0 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 39 eee 4 Important Note Please store the instruction manual in a place easily accessible to every user A EURO AC plug DIN 49441 CEE 7 VII10 16 A 250 V is standard on all instruments For North America instruments feature standard US plugs NEMA Pub No WDI1961 ASA C 73 1 1961 page 8 15A 125V When operating the instruments in countries with different AC plug systems use an approved adapter or have a qualified electrician replace the AC plug with an approved model suitable fort the country of operation The instrument is earthed as supplied When replacing the original AC plug ensure that the earth conductor is connected to the new plug 2 Safety Information 2 1 Explanation of the safety information in the instruction manual IRBE This symbol emphasizes the following notices which have to be strictly observed by the user Disregarding of warnings may result in impairment of serviceability as well as impairment of the user This symbol empha
45. kann durch Drehen des Handrades 5 in Minuteninkrementen ver ndert werden Eine eingestellte Zeit von 10 Minuten bewirkt das Ausschalten des Ger tes nach 10 Minuten Timer 3 dr cken und durch Drehen des Handrades 5 die gew nschte Ausschaltzeit einstellen dd hh mm Tage Stunden Minuten minimale Ausschaltzeit 1 Minute maximale Ausschaltzeit 99 Tage 23 Stunden 59 Minuten Die Timer LED 2 leuchtet auf wenn die Timer Funktion aktiviert ist Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 14 5 Sicherheitsfunktionen 5 1 Rotorblockade Erkennung Sollte die Drehmomentzunahme ber der Zeit einen bestimmten Grenzwert berschreiten nimmt das Ger t an dass der R hrfl gel Rotor blockiert wurde und stellt sofort ab Die Abschalt Sensitivitat kann im Setup Men eingestellt werden gt Parameter Cut off limit 5 2 berlast Schutz Sollte Drehmoment oder Leistung den erlaubten Bereich berschreiten verringert das Ger t automatisch die Motordrehzahl bis das Drehmoment bzw die Abgabeleistung in den zulassigen Bereich zur ckkehrt Tritt dieser Fall ein leuchtet die rote Load LED 6 auf Hinweis Solange die Load LED 6 leuchtet kann der Motor die eingestellte Solldrehzahl nicht halten 5 3 bertemperaturschutz R100C Sollte die Motortemperatur den maximal zul ssigen Grenzwert berschreiten so reduziert das Ger t automatisch die Abgabeleistung d
46. max 220 Watt Leistungsabgabe am Schaft robuster leistungsstarker B rstenmotor ger uscharmer Riemenantrieb max Drehmoment 600 Ncm 20 min Dauerdrehmoment bei 250 1 min 400 Ncm max Drehmomentbegrenzung programmierbar erkennt und sch tzt vor folgenden gef hrlichen Situationen Rotorblockade Erkennung Motor berhitzung Selbsttest aller Sicherheitsfunktionen nach Einschalten Benutzerprogrammierbare Begrenzungen f r maximal zul ssige Motordrehzahl 10 500 rpm maximal zul ssiges Drehmoment am Schaft 10 600 Ncm programmierbare Baudrate 1200 2400 4800 8 N 1 Vollst ndiger Zugang zu allen R hrfunktionen lesen und schreiben daisy chaining erm glicht es bis 255 Ger te an nur eine serielle Schnittstelle am Computer anzuschlie en Windows control software optional erh ltlich Page Seite 33 Type R100CT Spezifikationen Analoger Spannungseingang zur Uberschreibung der eingestellten Drehzahl mit einem externen Spannungssteuerungs signal Analoger Spannungsausgang zur Dokumentation von Drehzahl Drehmoment oder Leistungsabgabe Zul ssige Umgebungstemperatur Max relative Feuchtigkeit Sicherheitsklasse gem DIN EN 60529 Geh useabmessungen Gewicht 14 Reparaturen Bitte sorgen Sie bei gefahrlichen Arbeitsstoffen in Ber hrung gekommen sind daf r M glichst genaue Stoffangabe des Mediums SchutzmaBnahmen zum sicheren Umgang f r Eingangsbereich 0 10V Betriebsbereich pr
47. n internes Selbstdiagnosesystem Dieses System erm glicht dem Ger t Fehler selbst zu erkennen Tritt ein Fehler auf so schaltet sich das Ger t selbst aus Der letzte Fehler bzw Abschaltgrund wird nachsten Wiedereinschalten des Ger tes f r ca 3 Sekunden im Display angezeigt Fehlermeldung Abschaltgrund der programmierbare Abschalt Timer ist abgelaufen Timer Taste rotor stuck Eine zu schnelle Erh hung des Rotordrehmoments wurde festgestellt siehe auch Setup Men Motortemperatur zu hoch RS232 Off Das Ger t wurde durch einen RS232 Befehl ausgeschaltet Temperaturf hler oder Schaltkreis defekt 9 Reinigung und Wartung Das Geh use des Ger tes besteht aus lackiertem bzw eloxiertem Aluminium Das Bedienfeld ist chemisch sehr best ndig und spritzwassergesch tzt Verwenden sie zur Reinigung ein feuchtes Tuch und eine milde Seifenlauge Verwenden Sie auf keinen Fall Stahlwolle oder aggressive Reinigungsmittel um das Ger t zu reinigen Die Oberflache des Ger tes kann dadurch beschadigt werden Versuchen Sie auf keinen Fall das Ger t bei Fehlfunktion zu reparieren Es gibt keine Komponenten innerhalb des Ger tes welche durch den Benutzer gewartet werden m ssen Die Gew hrleistung erlischt bei jedem Versuch das Ger t zu ffnen oder gar zu reparieren Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 27 10 Abbau Transport und Lagerung 10 1 Abbau Schalten Sie das Gerat ab und
48. ndebohrungen auf der R ckseite des Ger tes fest Ziehen Sie diese mit Hilfe des Spannfutterschl ssels B kr ftig fest 3 2 R hrwerk am Stativ befestigen Randelschraube E Befestigen Sie jetzt die Spezialhalteklammer C an der Stativstange D F hren Sie die Haltestangen A des R hrwerks in die daf r Vorgesehenen Aussparungen ex der Spezialhalteklammer C und drehen die Randelschrauben E so lange bis der Antrieb sicher befestigt ist Spezialhalteklammer C Stativstange D 3 3 R hrwerkzeug montieren Schieben Sie das R hrwerkzeug G wie dargestellt in das Spannfutter F und ziehen Sie es in der gew nschten Position mit dem Spann futterschl ssel B kraftig fest Durch eine Langsverschiebung des R hrwerkzeugs bei gel stem Spannfutter F k nnen Sie die Eintauchtiefe des R hrwerkzeugs einstellen Spannfutter F R hrwerkzeug G Spannfutterschl ssel B Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 9 3 4 Montage des R hrwellenschutzes Befestigungsstange f r R hrwellenschutz H 5 Befestigungsschrauben J mar Aussparungen K Schraube L R hrwellenschutz M 4 Bedienung und Betrieb Verwenden Sie zum Schutz vor Verletzungen beim Arbeiten mit dem Ger t einen R hrwellen schutz Montieren Sie die Befestigungsstange H des R hrwellenschutzes mit Hilfe einer Kreuzklemme z B Art Nr 60492 00 vor dem Bohrfutter
49. nopf 7 eingestellt lt CAF RIOOC LCD Anzeige 1 Timer LED 2 Display select Taste 4 Timer Taste 3 Load LED 6 Handrad 5 Drehzahlregler 7 Power LED 8 Power ein aus Taste 9 Sicht auf das R100C R100CL R100CT Bedienfeld Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 11 4 3 Beschreibung des Bedienfelds LCD Anzeige 1 Die LCD Anzeige zeigt alle relevanten Systemdaten Timer LED 2 gibt an dass der Timer aktiviert ist und das Ger t automatisch ab schaltet wenn die eingestellte Zeit abgelaufen ist Timer Taste 3 Einstellung der Ausschaltverz gerung durch das Hanarad in der Reihenfolge Tage Stunden Minuten Select Display Taste 4 Dr cken Sie diesen Knopf kurz um von einer Anzeigeoption zur n chsten zu gelangen Dr cken Sie diesen Knopf und halten ihn f r min 3 sec um die relative Drehmomentmessung Mr auf Null zu stellen Dr cken und halten Sie diesen Knopf min 5 sec um von der relativen Drehmomentmessung zur absoluten Drehmomentmessung Ma zur ckzugelangen Das gleiche gilt f r die Leistungsabgabe Messfunktion Handrad 5 Load LED 6 gibt an dass das Ger t die eingestellte Drehzahl nicht halten kann aufgrund eines zu hohen Nenndrehmoments an der Motorwelle Unter solchen Bedingungen reduziert das Ger t die Motordrehzahl so dass das max zul ssige Nenndrehmoment am Motor nicht berschri
50. nship which develop within 24 months of the delivery of this product to the original user In the event of replacement the replacement unit will be guaranteed for the remainder of the original twenty four 24 month period or ninety 90 days whichever is longer If this product should require service contact your local distributor or manufacturer for necessary instructions This guarantee will not apply if the defect or malfunction was caused by accident neglect unreasonable use or fitness for a particular purpose which extend beyond the description and period set forth herein The manufacturers sole obligation under this guarantee is limited to the repair or replacement of a defective product and the manufacturer shall not in any event be liable for any incidental or consequential damages of any kind resulting from use or possession of the product Attention The user has to determine if the instrument is suitable for his specific application If there are any further queries contact your local dealer or the manufacturer direct Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 63 13 Technical Data 13 1 Technical Data R100C Type R100C Specifications Electrical power 110 230V 300W 50 60 Hz see R100C rear panel requirements LCD Display bright back lit LCD display showing all relevant system data Simultaneous display of set and real values of motor speed torque or power o
51. ntrol software optional available EHRE ER Ze Permissible ambient 5 40 C within operating area e g fume cupboard temperature Permissible humidity 80 RH Safety class acc to IP42 DIN EN 60529 Case dimensions Length including chuck 230 mm 77 5 x 190 x 170 mm ERSS 13 2 Technical Data R100CL Electrical power 110 230V 300W 50 60 Hz see R100CL rear panel requirements LCD Display bright back lit LCD display showing all relevant system data Simultaneous display of set and real values of motor speed torque or power output system settings 1 minute to 99 days Resolution 1 minute Torque output display displays true torque at shaft in Ncm relative or absolute readout possible Accuracy 596 Power output display displays true power output at shaft in watts relative or absolute readout possible Accuracy 596 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 65 Motor specification torque microprocessor controlled for easy operation and accuracy soft start and soft stop logic prevents splashing of liquid precise adjustable motor speed from 50 to 2000 rpm max 230 watts shaft power output extra long life high performance brush motor silent belt drive power transmission max torque 250 Ncm 20 min continuous torque output at 700 rpm 200 Ncm maximum limitation of torque programmable Multistep safety system detects and protects against the followin
52. ogrammierbar Erm glicht einfache Drehzahlsteuerung ber ein externes Potentiometer Abgabebereich 0 4V Betriebsbereich von Abgabe und Uberwachungsfunktion Drehzahl Drehmoment oder Leistungsabgabe ist programmierbar 5 40 C innerhalb der Umgebung z B Abzug 80 RH IP42 77 5 195 x 195 mm 250 mm inklusive Bohrfutter 5 5 Kg der Anlieferung von Reparaturgeraten die mit unser Annahme und Wartungspersonal Kennzeichnung der Verpackung der Gefahrenstoffverordnung Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 34 14 1 R cksendeformular Wichtig 1 Kalibrieren des Instruments wird automatisch durchgef hrt 2 Sieerhalten einen Kostenvoranschlag f r die zu erwartenden Reparaturkosten 3 Reparaturen werden bei Freigabe durchgef hrt KONTAKT Ansprechpartner Tel Fax Nr Email Rechnungsanschrift Lieferanschrift Firma Firma Adresse Adresse GERATEINFORMATION Modell Serien Nr Problem Fehlerbeschreibung Betriebsbedingungen bitte ausf llen wenn zutreffend Umgebungstemp Rel Feuchtigkeit Drehzahl Zuladegewicht Volumen Viskositat Temperatur in C Probentemperatur Laufzeit Probenbeschreibung HINWEIS Sollte das Gerat mit gefahrlichen Stoffen in Ber hrung gekommen sein muss es vor dem Ver sand an CAT dekontaminiert werden F g
53. on 1 07 Page Seite 6 Wichtiger Hinweis Ist das Ger t unbesch digt und alle Teile vollstandig so k nnen Sie nach dem Studium der Betriebsanleitung mit dem Betrieb des Gerates beginnen Wichtiger Hinweis Bitte lesen Sie die Betriebsanleitung gr ndlich durch und stellen Sie sicher dass Sie alles verstanden haben bevor Sie das Ger t in Betrieb nehmen Sollten Sie nach dem Durchlesen der Anleitung noch weitere Fragen zur Installation zum Betrieb oder zur Wartung haben wenden Sie sich bitte an den Lieferanten oder den Hersteller des Ger tes unter folgender Adresse CAT Ingenieurb ro M Zipperer GmbH Etzenbach 16 D 79219 Staufen 49 0 7636 7803 0 Wichtiger Hinweis Diese Betriebsanleitung sollte jederzeit zur Verf gung stehen insbesondere demjenigen der dieses Gerat benutzen m chte Deshalb sollte diese Betriebsanleitung in der Nahe des Ger tes aufbewahrt werden Die Ger te sind standardm ig mit einem EURO Stecker DIN 49441 CEE 7 NI10 16 A 250 V versehen F r Nordamerika mit einem US Norm Stecker NEMA Pub No WDI1961 ASA C 73 1 1961 Seite 8 15A 125V Falls Sie das Ger t in einem Land mit anderem Stecker System betreiben m chten m ssen Sie einen zugelassenen Adapter verwenden oder der mitgelieferte Stecker muss durch einen Fachmann ausgewechselt und durch einen f r dieses Netz passenden und zugelassenen Stecker ersetzt werden Bei Lieferung ist das Ger t geerdet Beim Auswechseln des Originalsteck
54. on Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 50 6 1 RS232 Connection configuration between PC and the R100C R100CL R100CT a PC with 9 pin RS232 plug 2 RxD 2 TxD PC R100C 9 pin Sub D 9 pol Sub D plug socket 9 pol Sub D 9 pol Sub D Socket plug b PC with a 25 pin RS232 plug PC R100C 25 pin Sub D 9 pin Sub D plug socket 25 pin Sub D 9 pin Sub D socket plug 6 2 Format of an RS232 Command Each command which is sent to the device must have the format ADR CMDCODE PARAMETERLIST CR Description ADR Slave Address of the unit to execute the command CMDCODE Command code PARAMETERLIST 1 to 6 parameters separated by commas CR The command string must be terminated by Carriage Return ASCII 13 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 51 6 3 Format of the Controller Handshake After receiving a RS232 command the unit will 1 Send the received command to the next unit or back to the PC daisy chaining 2 Answer with a handshake string which is defined as follows ADR HS RETCODE PARAMETERLIST CR Explanation ADR Slave Address of the unit sending the handshake RETCODE Errorcode see table 1 below PARAMETERLIST 1 to 6 parameters each parameter is separated by a comma see also table 2 Command codes CR the handshake as any command is terminated by ASCII code 13 CR Table of return codes Table1 Retu
55. perature inside the unit exceeds the maximum operating temperature the unit will automatically reduce power output speed and torque in order to get back to a safe operating point fold back characteristic This prevents the unit to be destroyed by constant overload operation This overload condition is also signalled by the red Load LED 6 Furthermore a cooling fan 11 on the rear side of the instrument is automatically switched on if required If the temperature inside the R100C exceeds 70 C by any means the unit will shut down immediately This safety feature is fully redundant The unit either shuts down if the microprocessor measures a temperature higher than 70 C or if the thermo switch inside the R100C detects a temperature higher than 70 C Important Note Before re starting the unit after cooling down switch it off and then on again by using the on off button 5 4 Overtemperature protection R100CL If the temperature inside the unit exceeds the maximum operating temperature the unit will automatically reduce power output speed and torque in order to get back to a safe operating point fold back characteristic This prevents the unit to be destroyed by constant overload operation This overload condition is also signalled by the red Load LED 6 Furthermore a cooling fan 11 on the rear side of the instrument is automatically switched on if required Betriebsanleitung Instruction Manual R100C
56. r at least 5 seconds to switch back from relative torque measurement Mr to absolute torque measurement Ma The same applies also to the Power output measurement function Encoder Wheel 5 Load LED 6 Indicates when the unit is not able to hold set motor speed due to too high torque taken from motor shaft Under this condition the unit will reduce the real motor speed so that the maximum allowed motor torque is not exceeded Speed control knob 7 setting the speed manually Power LED 8 indicates that unit is powered up Power on off button 9 Press this key to turn the unit on or off Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 45 4 4 Description of the rear panel on the R100C R100CL R100CT Support Rod A 9 pole SUB D connector 10 Tm Thread a O Thread a N Fan 11 Mains input 12 Te rear view of the R100C R100CL R100CT 9 pole SUB D connector In Output for RS232 interface full duplex analogue input for speed setting via voltage signal input range 0 10V operation range programmable analogue output for speed torque or power monitoring output range 0 4V operation range programmable mains power supply make sure that the voltage printed on the R100C corresponds to the voltage from your mains On the rear panel of the R100C
57. rehzahl Drehmoment oder Leistungsabgabe Zul ssige Umgebungstemperatur Max relative Feuchtigkeit Sicherheitsklasse gem DIN EN 60529 Geh useabmessungen Gewicht Eingangsbereich 0 10V Betriebsbereich programmierbar Erm glicht einfache Drehzahlsteuerung ber ein externes Potentiometer Abgabebereich 0 4V Betriebsbereich von Abgabe und Uberwachungsfunktion Drehzahl Drehmoment oder Leistungsabgabe ist programmierbar 5 40 C innerhalb der Umgebung z B Abzug 80 RH IP42 77 5 x 190 x 170 mm 230 mm inklusive Bohrfutter 13 2 Technische Daten R100CL Type R100CL Netzspannung LCD Display Handrad 5 Timer Funktion Drehmoment Anzeige Anzeige der Leistungsabgabe Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Spezifikationen hintergrundbeleuchtetes Display zeigt alle relevanten Systemdaten Gleichzeitige Anzeige der eingestellten und tats chlichen Werte von Drehzahl Drehmoment und Leistungsabgabe Programmierbarer switch off timer programmierbar von 1 Minute bis 99 Tage Aufl sung 1 Minute Zeigt tats chliches Drehmoment am Schaft in Ncm Anzeige des relativen oder absoluten Wertes m glich Genauigkeit 596 Zeigt die tats chliche Leistungsabgabe am Schaft in Watt Anzeige des relativen oder absoluten Wertes m glich Genauigkeit 596 Page Seite 31 Type R100CL Spezifikationen Motorspezifikationen Multistep Sicherheitssystem Hilf
58. rn Code Explanation Parameterlist command executed no error see table 1 unknown command none wrong parameter number none too few or too many parameters specified command is not allowed in actual actual operation operation mode mode at least one parameter is out of none range at least one parameter is too long none unknown data format none Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 52 6 4 RS232 commands of the R100C R100CL R100CT Table 2 CMD Explanation Parameter list Range CODE Read actual settings 1 Dummy parameter to initiate transfer 1 from the unit gt Controller sends in handshake A actual motor speed rpm B actual torque relative C actual torque absolute D actual power relative E actual power absolute wo 1 WTR A Timer value in sec 0 2592000 Read Timer 1 Dummy parameter to initiate transfer gt Controller sends in handshake A Timer value seconds 0 2592000 gt 0 Timer disabled Write Motor SET A Motor speed in rpm s 0 2000 define motor speed 0 gt Motor is switched stopped speed 0 setpoint in rmp s if selected speed is above allowed limit value will be reduced to allowed limit gt Set up menu Max MotorRpm Read motor setpoint 1 Dummy parameter to initiate transfer 1 SET Controller sends in handshake A Motor setpoint in rpm 0 2000 Read system status 1 Dummy parameter to initiate transfer
59. rsion 1 07 Page Seite 71
60. s chliches Drehmoment am Schaft in Ncm Anzeige des relativen oder absoluten Wertes m glich Genauigkeit 596 Zeigt die tatsachliche Leistungsabgabe am Schaft in Watt Anzeige des relativen oder absoluten Wertes m glich Genauigkeit 596 Mikroprozessorgesteuert f r einfache Bedienung und Genauigkeit Softstart und Softstop logic verhindert Spritzen Pr zise Einstellung der Motordrehzahl von 40 bis 2000 1 min max 220 Watt Leistungsabgabe am Schaft robuster leistungsstarker B rstenmotor ger uscharmer Riemenantrieb max Drehmoment 150 Ncm 20 min Dauerdrehmoment bei 1100 1 min 100 Ncm max Drehmomentbegrenzung programmierbar erkennt und sch tzt vor folgenden gefahrlichen Situationen Rotorblockade Erkennung Motor berhitzung Selbsttest aller Sicherheitsfunktionen nach Einschalten Benutzerprogrammierbare Begrenzungen f r maximal zul ssige Motordrehzahl 40 2000 rpm maximal zul ssiges Drehmoment am Schaft 10 150 Ncm programmierbare Baudrate 1200 2400 4800 8 N 1 Vollst ndiger Zugang zu allen R hrfunktionen lesen und schreiben daisy chaining erm glicht es bis 255 Ger te an nur eine serielle Schnittstelle am Computer anzuschlieBen Windows control software optional erh ltlich Page Seite 30 Type R100C Spezifikationen Analoger Spannungseingang zur Uberschreibung der eingestellten Drehzahl mit einem externen Spannungssteuerungs signal Analoger Spannungsausgang zur Dokumentation von D
61. s actual motor speed in rpm Return code see table 1 OK no error Describes that this is a handshake and not a command handshake designator Slave address of the unit which sent the handshake Every command or handshake sequence is terminated by ASCII Code 13 CR 6 6 Connection of more than one unit to a computer The powerful protocol of the RS232 interface allows also for connection of other CAT Laboratory instruments like hotplate stirrers M26 MCS66 micrometering pumps HPLH series dispensers Liquid Processor to only one RS232 interface on the computer side By assigning different slave addresses to each connected unit the computer can access each unit individually For connection of more than one unit to a computer a special Y cable is used for every additional unit RS232 Y cable part no 60709 00 For application adaptation a Windows program is optionally available Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 55 Application example Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 56 7 The Setup menu Global settings are stored in a non volatile memory EEPROM To modify these global settings press and hold the Select Display key while switching the unit on with the Power On Off key The following settings could be adjusted in the setup menu of the R100C R100CL R10
62. sfunktionen RS232 Schnittstelle Analoger Spannungseingang zur Uberschreibung der eingestellten Drehzahl mit einem externen Spannungssteuerungs signal Analoger Spannungsausgang zur Dokumentation von Drehzahl Drehmoment oder Leistungsabgabe Zulassige Umgebungstemperatur Max relative Feuchtigkeit Sicherheitsklasse gem DIN EN 60529 Gehauseabmessungen Gewicht Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Mikroprozessorgesteuert f r einfache Bedienung und Genauigkeit Softstart und Softstop logic verhindert Spritzen Pr zise Einstellung der Motordrehzahl von 40 bis 2000 1 min max 230 Watt Leistungsabgabe am Schaft robuster leistungsstarker B rstenmotor ger uscharmer Riemenantrieb max Drehmoment 250 Ncm 20 min Dauerdrehmoment bei 700 1 min 200 Ncm max Drehmomentbegrenzung programmierbar erkennt und sch tzt vor folgenden gefahrlichen Situationen Rotorblockade Erkennung Motor berhitzung Selbsttest aller Sicherheitsfunktionen nach Einschalten Benutzerprogrammierbare Begrenzungen f r maximal zul ssige Motordrehzahl 30 1100 rpm maximal zul ssiges Drehmoment am Schaft 250 Ncm programmierbare Baudrate 1200 2400 4800 8 N 1 Vollst ndiger Zugang zu allen R hrfunktionen lesen und schreiben daisy chaining erm glicht es bis 255 Ger te an nur eine serielle Schnittstelle am Computer anzuschlie en Windows control software optional erh
63. sigen Grenzwert berschreiten so reduziert das Ger t automatisch die Abgabeleistung durch Reduzieren der Drehzahl gt Foldback Charakteristik Dies sch tzt das Ger t vor Besch digung durch st ndigen Betrieb unter berlast Dieser berlastzustand wird auch durch die rote Load LED 6 angezeigt Das K hlergebl se 11 auf der R ckseite des Ger tes wird bei Bedarf automatisch zugeschaltet 5 6 Anschluss 9 pin SUB D Pinbelegung des 9 pin SUB D Anschlusses 10 an der R ckseite des Ger tes RS232 Steuersignale Analoger Eingang 0 10V zur externen Drehzahlsteuerung Analoger Ausgang 0 4V zum Aufnehmen der Prozessparameter Drehzahl Drehmoment oder Abgabeleistung Pinausgang des I O Anschlusses Beschreibung 5V max 10mA TxD Sendeleitung Empfangsleitung Nicht belegt analoger Spannungsausgang 0 4V Nicht nicht belegt analoger Spannungseingang 0 10V GND Signal Masse Pinbelegung der I O Buchse Draufsicht 9 pin Sub D Buchse Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 16 6 Die RS232 Schnittstelle des R100C R100CL R100CT Die RS232 Schnittstelle erlaubt die Fernsteuerung aller R hrfunktionen In dieser Konfiguration kann ein Computer auf jedes angeschlossene Ger t ber eine Adresse daisy chaining zugreifen Diese Adresse ist f r jedes einzelne Gerat programmierbar gt Setup Men slave ad
64. sizes prohibitions which have to be strictly observed Any disregard of the prohibitions may result in severe impairment of serviceability as well as severe impairment of the user This symbol emphasizes notices which have to be observed by the user to assure a safe operation of the unit CAUTION Hazard of fire or explosion Advise for repair maintenance Advise for connection to the mains Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 40 2 2 Explanation of the safety information on the device CAUTION This symbol indicates to read the instruction manual carefully prior to operation of the instrument Please mark points which require special attention in your field of application so they are not overlooked Disregarding of warnings may result in impairment of serviceability as well as impairment of the user gt 2 3 General Safety Information Attention Please comply with all safety and accident prevention regulations applicable to laboratory work Attention Use great caution when working in the vicinity of highly flammable or explosive substances The instrument is not explosion proof Attention Only instructed users may operate the instrument Attention When connecting the instrument to an AC power outlet ensure that your local supply voltage matches that indicated on the instrument s rating plate Turn the power switch off when not using the instrument and before
65. ssscsssesssscsssseessecosssesusseocossseeseecosssseseseosssssesseeossseeeeseesseseese 61 10 DISMANTLING TRANSPORT AND STORAGE 62 ARER B 3 ES WASIN K E PA BN E dicen en EE E NE A EAT EANA 62 102 TRANSPORT AND STORAGE 0 0 0 Rn ah san PL OEC Y CEU a a a TRECE b a a aaa ai a 62 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 3 11 DISPOSAL 2 TE iei eneus seed Moi TT 62 12 WARRANTY AND LIABILITY 63 13 TECHNICAL DATA CR 64 13 11 TECHNICAL iii conca ura gilded ates c s tc sug La Bota gU od 64 13 24 TECHNICAL DATA R100GIE a 65 13 3 TECHNICAL DATA 5 n e ti ee oti ira bie ERR ROT e 67 14 Wc 68 141 REPAIR RETURN FORM inhibet eere Rer ia AR ERIT ARRA RE GR 69 15 DECLARATION OF CONFORMITY icscssscccesescevsscsssssconsssevesctvassssensessecessoscesnasestesevencvasetvessesevascessstsestctesess 70 16 D OT 71 16 1 DISPLAYING THE SYSTEM INFO iniii a a a ei DIRT iia e i a dee 71 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 4 1 Allgemeine Informationen Das R hrwerk R10
66. torque increases Analogue input Allow block speed setting via external analogue voltage control An Inp F 0 Defines the control voltage analogue input when the motor has to be stopped speed 0 The voltage at the analogue input functions as multiplication factor onto the frontally set speed speed control knob This value reflects the input voltage at which the multiplication factor is zero This setting is only relevant if the analogue input is enabled see above Factory setting 0 0 V Range 20 0 30 0 V An Inp F 1 This value reflects the input voltage at which the multiplication factor of the speed is 1 0 An Output Selects which function value has to be monitored on the analogue output pin Factory setting DISABLED Available options Torque R relative torque Torque A absolute torque Power R relative power Power A absolute power DISABLED output de activated OV output An Out U 0V This value defines the operation value speed torque or power which is related to a voltage outlet of 0 0V at the analogue pin output pin no 6 An Out U 4V This value indicates which operation value speed torque or power is related to the output voltage of 4 0V at the analogue pin output pin no 6 Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 58 7 1 Example How to change setup settings Example of the necessary steps to set define the maximum allowed motor sp
67. triebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 38 1 General Information The CAT overhead stirrer R100C R100CL R100CT is designed in accordance with Safety Class 1 and built and tested in accordance with DIN EN 61010 According to these regulations the unit is designed to meet the requirements for safe and correct operations To maintain the proper safety and operational functions of the instrument the user should follow the instructions and safety guidelines in this manual 1 4 Unpacking the instrument Unpack the instrument carefully and check to see that it is not damaged It is important that any damage incurred in transport be recognized at the time of unpacking Notify your carrier or forwarding agent immediately in case of such damage 1 2 Scope of Delivery Please check that the package contains the following 1 Overhead Stirrer R100C Part No 60218 00 or 1 Overhead Stirrer R100CT Part No 60314 00 or 1 Overhead Stirrer R100CL Part No 60315 00 2 Support Rods Part No 10402 64 1 Instruction Manual Accessories and stirring tools optional Special Clamp to fix the unit to a stand Part No 60495 00 Stirring shaft protector to Part No 20 be fixed to the stand U Stand Part No 60494 00 Please see price list for further accessories such as suitable stirring paddles Important Note After reading and understanding the instruction manual may now start operating the un
68. tten wird Drehzahlregler 7 zur manuellen Einstellung der Drehzahl Power LED 8 gibt an dass das Ger t eingeschaltet ist Power ein aus Taste 9 Dr cken Sie diesen Knopf zum Ein Ausschalten des Ger tes Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 12 4 4 Beschreibung der R ckseite des R100 C R100CL R100 CT mis 9 pin SUB D Anschluss 10 i Gewindebohrung a Gewindebohrung a Te ee N Ventilator 11 Netzanschluss 12 m Ansicht R ckseite R100C R100CL R100CT 9 pin SUB D Anschluss 10 Ein Ausgang f r HS232 Schnittstelle Duplex analoger Eingang f r Drehzahleinstellung durch Spannungssignal Eingangsbereich 0 10V Arbeitsbereich ist programmierbar analoger Ausgang f r Drehzahl Drehmoment oder Leistungs berwachung Ausgangsbereich 0 4V Arbeitsbereich programmierbar Netzanschluss 12 Achten Sie darauf dass die Spannung auf dem Typenschild des R100C R100CT mit der Spannung Ihrer Netzstromversorgung bereinstimmt An der R ckseite des R100C R100CT finden Sie 2 Gewindebohrungen a zum Anbringen der Haltestangen A einen 9 pin SUB D Anschluss das Netzanschlusskabel 12 und Hal A den Ventilator 11 altestange A Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 1
69. ument Y Pack the instrument in its original box If the box is not available take care to wrap the instrument and accessories with enough material to support them Y DO NOT send pedestal stand unless there is a problem with the upright rod clamp or base If there is a problem with the stand remove the upright rod from the base and individually wrap each item to avoid contact with the instrument Applicable for overhead stirrers and homogenisers Y Pack the instrument and related items in a strong box for shipping Mark the outside of the box with handling instructions Example Handle with care or Fragile Delicate Instrument and send to Ingenieurb ro CAT M Zipperer GmbH Service Department Etzenbach 16 D 79219 Staufen Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 69 Ingenieurb ro M Zipperer GmbH Etzenbach 16 D 79219 Staufen Tel 49 0 7636 78030 Fax 49 0 7636 780345 E Mail info cat ing de Internet http www cat ing de 15 Declaration of conformity Ce We declare under our sole responsibility that this product see table corresponds to the regulations and conforms with the standards or standardized documents 2006 95 EC DIN EN 61010 1 VDE 041 1 1 2002 08 Amendment 1 VDE 0411 1 Amendment 1 2002 11 Amendment 2 VDE 0411 1 Amendment 2 2004 01 DIN EN 61010 2 051 VDE 0411 2 051 2004 07 2004 108 EC DIN EN 61326 1 VDE 0843 20 1 2006 10 DIN EN 61000 6
70. ung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 24 An Inp F 0 Definiert die Steuerungsspannung gt analoger Eingang bei welcher der Motor gestoppt werden soll Drehzahl 0 Die Spannung am analogen Eingang wirkt als Multiplikationsfaktor auf die frontseitig eingestellte Drehzahl gt Drehzahleinstellknopf Dieser Wert gibt die Eingangsspannung wieder bei der der Multiplikationsfaktor Null ist Diese Einstellung ist nur relevant wenn der analoge Eingang aktiviert ist 8 0 Werkseinstellung 0 0 V Bereich 20 0 30 0 V An Inp F 1 Dieser Wert reflektiert die Eingangsspannung bei der der Drehzahlmultiplikationsfaktor 1 0 ist Diese Einstellung ist nur relevant wenn der analoge Eingang aktiviert ist Werkseinstellung 5 0 V Bereich 20 0 30 0 V An Output W hlt aus welche Funktion Wert auf dem analogen Ausgangsstecker berwacht werden soll Werkseinstellung DISABLED Deaktiviert Erh ltliche Optionen Torque R relatives Drehmoment Torque A absolutes Drehmoment Power R relative Leistung Power A absolute Leistung DISABLED Ausgang deaktiviert OV ouput An Out U 0V Dieser Wert definiert den Vorgangswert Drehzahl Drehmoment oder Leistung der zu einem Spannungsausgang von 0 0V am analogen S An Out U 4V Dieser Wert gibt an welcher Vorgangswert Drehzahl oder Leistung zur Ausgangspannung von 4 0V analogen Spannungsausgangsstecker pin No 6 geh rt 7 4 Beispiel Ander
71. ung der Setup Einstellungen Beispiele f r die notwendigen Schritte um die maximal zul ssige Motordrehzahl im Setup Men einzustellen definieren Dr cken und halten Sie die Select Display Taste 4 w hrend der R100C eingeschaltet wird Das Ger t initialisiert und schaltet dann in das Setup Men Hinweis Die Select Display Taste 4 muss gedr ckt gehalten werden bis das Setup Men auf dem Display 1 erscheint Drehen Sie das Handrad 5 bis MotorRpm auf dem Display 1 erscheint Dr cken Sie die Timer Taste 3 um die max Motordrehzahl zu ndern gt Edit Pfeil erscheint Drehen Sie das Handrad 5 um den Wert einzustellen Dr cken Sie die Select Display Taste 4 um das Setup Men zu verlassen Dr cken Sie die Select Display Taste 4 noch einmal wenn das Ger t um die Best tigung der Speicherung der Setup Parameter bittet Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 25 CAN RIOOC LCD Anzeige 1 Timer Timer LED 2 Taste 3 eo Display select Taste 4 Load LED 6 Drehzahlregler 7 Change Hand rad 5 Power LED 8 Power Power ein aus Taste 9 Sicht auf das R100C R100CL R1 00CT Bedienfeld HINWEIS Die verschiedenen Optionen Parameters im Setup Men k nnen einfach durch Drehen des Handrads 5 angezeigt ausgew hlt werden Um einen angeze
72. urch Reduzieren der Drehzahl gt Foldback Charakteristik Dies sch tzt das Gerat vor Besch digung durch st ndigen Betrieb unter berlast Dieser Uberlastzustand wird auch durch die rote Load LED 6 angezeigt Das K hlergebl se 11 auf der R ckseite des Ger tes wird bei Bedarf automatisch zugeschaltet Sollte die Motortemperatur 70 C berschreiten so schaltet sich das Ger t sofort aus Dieser Sicherheitskreis arbeitet unabh ngig zur Mikroprozessorregelung Hinweis Zur Wiederinbetriebnahme nach der notwendigen Abk hlung das Ger t erst ausschalten dann mittels Power ein aus Taste 9 wieder einschalten 5 4 bertemperaturschutz R100CT Sollte die Motortemperatur den maximal zul ssigen Grenzwert berschreiten so reduziert das Ger t automatisch die Abgabeleistung durch Reduzieren der Drehzahl gt Foldback Charakteristik Dies sch tzt das Gerat vor Besch digung durch st ndigen Betrieb unter berlast Dieser Uberlastzustand wird auch durch die rote Load LED 6 angezeigt Das K hlergebl se 11 auf der R ckseite des Ger tes wird bei Bedarf automatisch zugeschaltet Sollte die Motortemperatur 70 C berschreiten so schaltet sich das Ger t sofort aus Dieser Sicherheitskreis arbeitet unabhangig zur Mikroprozessorregelung Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 15 5 5 bertemperaturschutz R100CL Sollte die Motortemperatur den maximal zul s
73. utput system settings 1 minute to 99 days Resolution 1 minute Torque output display displays true torque at shaft in Ncm relative or absolute readout possible Accuracy 5 Power output display displays true power output at shaft in watts relative or absolute readout possible Accuracy 5 Motor specification torque microprocessor controlled for easy operation and accuracy soft start and soft stop logic prevents splashing of liquid precise adjustable motor speed from 50 to 2000 rpm max 220 watts shaft power output extra long life high performance brush motor silent belt drive power transmission max torque 150 Ncm 20 min continuous torque output at 1100 rpm 100 Ncm maximum limitation of torque programmable Multistep safety system detects and protects against the following dangerous situations rotor stuck block detection motor overtemperature monitoring self test of all safety functions after switching on Auxiliary functions User programmable limits for maximum motor speed 40 2000 rpm maximum allowed torque at shaft 10 150 Ncm Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 64 Type R100C Specifications RS232 interface programmable baudrate 1200 2400 4800 8 N 1 Full remote access of all stirrer functions read and write daisy chain feature by sub addressing allows to connect up to 255 units to only one serial interface on the computer side Windows co
74. www cat ing de 15 Konformitat Konformitatserklarung Declaration of conformity D claration de conformit Wir erklaren in alleiniger Verantwortung dass dieses Produkt siehe Tabelle den Bestimmungen der Richtlinien entspricht und mit den folgenden Normen oder normativen Dokumenten bereinstimmt 2006 95 EG DIN EN 61010 1 VDE 0411 1 2002 08 Berichtigung 1 VDE 0411 1 Berichtigung 1 2002 11 Berichtigung 2 VDE 0411 1 Berichtigung 2 2004 01 DIN EN 61010 2 051 VDE 0411 2 051 2004 07 2004 108 EG DIN EN 61326 1 VDE 0843 20 1 2006 10 DIN EN 61000 6 2 VDE 0839 6 2 2006 03 DIN EN 61000 6 4 VDE 0839 6 4 2007 09 DIN EN 61000 4 2 VDE 0847 4 2 2009 12 DIN EN 61000 4 3 VDE 0847 4 3 2008 06 DIN EN 61000 4 4 VDE 0847 4 4 2005 07 DIN EN 61000 4 5 VDE 0847 4 5 2007 06 DIN EN 61000 4 6 VDE 0847 4 6 2009 12 DIN EN 55011 VDE 0875 11 2010 05 DIN EN 55014 1 VDE 0875 14 1 2010 02 DIN EN 55014 2 VDE 0875 14 2 2009 06 2006 42 EG DIN EN ISO 12100 1 2004 04 DIN EN ISO 12100 2 2004 04 Bei einer nicht mit uns abgestimmten nderung des Ger tes verliert diese Erkl rung ihre G ltigkeit und die Gew hrleistung erlischt Ingenieurb ro CAT M Zipperer GmbH 1Ellen Gebel CE Dokumentenbeauftragte 79219 Staufen den 12 Juli 2010 R100C R100CL R100CT 230V 10 50 60Hz 300W Text wurde maschinell erstellt und ist auch ohne Unterschrift g ltig Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Pag
75. y rotor stuck A too fast increase of rotor torque was detected see also Setup menu Overtemp the motor temperature was too high RS232 Off The unit was switched off by an RS232 command Sensor Err The temperature measurement sensor or circuit is damaged 9 Cleaning and Maintenance The outer casing consists of painted or anodised aluminium and a chemically resistant splash proof membrane key pad and therefore easily cleaned with warm water and any suitable liquid laboratory detergent Do not use steel wool or any similar plastic wool sponge to clean the unit In case of malfunction do not attempt to repair the unit There are no user serviceable parts in this instrument The CAT R100C R100CL R100CT should only be opened and repaired by authorised service personnel Any work on the electronics inside the unit should only be carried out by knowledgeable trained personnel Any attempt by the user to repair the unit will immediately render the guarantee null and void Please contact your local distributor in the event of a problem Betriebsanleitung Instruction Manual R100C R100CL R100CT 07 2010 Version 1 07 Page Seite 61 10 Dismantling Transport and Storage 10 1 Dismantling Switch the unit off Disconnect the unit from the mains Remove any glass beakers and any other equipment around the overhead stirrer Remove the stirring tool G by loosening the chuck F with the chuck key B Loosen the special clamp C and r