        Digitales Vakuum Anzeige- und Regelgerät Digital Vacuum Display
         Contents
1.     At the VD9 pressure readings of certain transducers can be adjusted on  atmosphere and or zero pressure     CAUTION    The connected transducers need an adequate warm up  time of at least 10 minutes before adjustment     For adjustment on atmosphere pressure   Switch the VD9 into configuration mode as described in chapter 5  so  that the display shows  unit      Press the menu key  several times  until the display shows  Adu             Hier        968     Press the  Up  key to select adjustment on atmosphere and confirm by  the  set  key  The value for atmosphere pressure is displayed        OO For transducer types 1  2 and 5  see chapter 5 3  the value of  atmosphere pressure can now be adjusted     Confirm adjustment with  Set  key   The display then shows the next configuration point  In order to return to    measurement mode quit all further configuration points by pressing the  menu key     5 THYRACONT    For adjustment on zero pressure        CAUTION ae  Vor einem Nullpunktabgleich ist sicherzustellen  dass    der Nulldruck im Rezipienten mindestens eine Dekade  unterhalb der Messbereichsgrenze liegt    For transducer type 4 adjustment on zero pressure is not possible    For transducer type 5 a zero pressure less than 0 1mbar is sufficient     Switch the VD9 into configuration mode as described in chapter 5  so  that the display shows  unit      Press the menu key  several times  until the display shows  AdJ              Lofer      Press the  Down  key to select ad
2.    der this configuration point     To select the start mode switch the VD9 into configuration mode as  described in chapter 5  so that the display shows  unit      Press the menu key  several times  until the display shows the currently  selected start configuration     OO By means of the  Up  and  Down  keys you can now toggle be   tween  Stop  and  run      Confirm with  Set  key     The display then shows the next configuration point  In order to return to  measurement mode quit all further configuration points by pressing the  menu key     5 THYRACONT    5 8 Scaling the Analog Output    When the analog output is set to type  outP   you can scale which  pressure range is to be mapped to the 0 10 V output signal     To scale the analog recorder output switch the VD9 into configuration  mode as described in chapter 5  so that the display shows  unit           Sm    Press the menu key  several times  until the display shows  SCAL         gt   5t     Select by means of the  Up  and  Down  keys if the upper   S Hi   or  lower   S Lo   scale end should be adjusted         gt   1 3      Confirm with  Set  key  The current value for this scale end is displayed     OO Adjust the scale end by means of the  Up  and  Down  keys          SCA      Confirm with  Set  key  the display again shows  SCAL   Now you can  adjust the second scale end        gt  5h      Select the second scale end by means of the  Up  and  Down  keys   S  Hi  or  S Lo       THYRACONT 3         gt     Confirm wi
3.   31      66  56 0       4 mm       7 Technische Daten    fe     1284 mn       Dra    Anzeige Numerisches LCD  4 stellig  mit Hintergrundbeleuchtung  45 x 20 mm  Display Refresh Rate 2 Hz  0 5 s   on  Spannungsversorgung 95     265 VAC  50 60 Hz  Leistungsaufnahme max  15 W  inkl  Transmitter     Sicherung 0 8 A T    Umgebungstemperatur   5   50   C    Messeingange 2x 4 20 mA  SubD 9polig  Buchse  1x 0 10 V  Amphenol 6polig  Buchse  Schaltausgange 2 x Relaiskontakt  potentialfreie Wechsel Schalter   Phoenix Klemmleiste 6polig   4 A  250 VAC  2 A 45 VDC  Analogausgang 0 10 V   gt 10 kQ  Serielle Schnittstelle RS232  SubD Stecker  9olig    Schutzart IP 20       2 THYRACONT    8 Konformitatserklarung    C       Erklarung Uber die Konformitat    Diese Erkl  rung gilt f  r folgend bezeichnete Erzeugnisse     Ger  tsart  Vakuum Mess  und Regelger  t  Typenbezeichnung  VD9 52  VDI CV  Zertifikat Nr   261 368    Hiermit wird best  tigt  dass die Produkte den wesentlichen Schutzanforderungen  entsprechen  die in den Richt  nien des Rates zur Angleichung der Rechtsvorschriften  der Mitgliedstaaten   ber die elektromagnetische Vertr  glichkeit  BW336 EWG  ge  ndert  durch 91 26 VEWG  S2 31 EWG   3 68 EWG sowie 93 87 EWG und der  Nrederspannungsrichtiinie TA 2WEWG ge  ndert durch 93 68 EWG festgelegt sind     Diese Erkl  rung wird abgegeban durch    Thyracont Elektronic GmbH  Max Emanuel Str  10  94036 Passau    CE a ON EOS se hinsichtlich elektromagnetischer Vertr  glichkeit sowie  der 
4.  2 1 Zur Orientierung    Diese Betriebsanleitung ist g  ltig f  r Produkte mit den Artikelnummern  VD9S2     Sie finden die Artikelnummer auf dem Typenschild  Technische Anderungen  ohne vorherige Anzeige sind vorbehalten     2 2 Lieferumfang    Zum Lieferumfang geh  ren      VD9 Anzeige  und Regelger  t    Netzkabel      Gegenstecker f  Relaisausgang  Schreiberausgang     Befestigungsschrauben zur Schalttafelmontage  19  System     Betriebsanleitung    Lieferbares Zubeh  r     Transmitter VSC42MA4  1400   1 mbar    Transmitter VSP52MA4  100     1x10   mbar    Transmitter VSP62MV  1000     1x10  mbar    Transmitter VSM72MV  1000   5x1 0   mbar    Transmitter VSH82MV  1000     1x10   mbar     Messkabel f  VSC42 VSP52  2m  W0900002      Messkabel f  VSC42 VSP52 6m  W0900006      Messkabel f  VSP62MV  2m  W0606002     Messkabel f  VSP62MV  6m  W0606006      Messkabel f  VSM72 VSH82 2m  W1506002      Messkabel f  VSM72 VSH82 6m  W1506006    RS232 Kabel zum PC Anschluss  2m  W0800002      Windows Software VacuGraph  VGR    THYRACONT 5    2 3 Produktbeschreibung    Das VD9S2 dient zum Anzeigen und Regeln von Absolutdruck in Verbindung mit  Thyracont Vakuum Transmittern mit Signalausgang 4 20mA oder 0 10V       ber die serielle Schnittstelle kann der Regler vom PC aus gesteuert werden        LCD Anzeige   Taste Menu   Pfeiltasten auf ab   Taste Set   Signaleingang 4 20mA f  VSC42MA4  Signaleingang 4 20mA f  VSP52MA4  Ger  tesicherung 0 8AT   Netzanschluss 95   265 VAC  50 60 Hz  Ana
5.  4 THYRACONT    3 5 Relay Outputs     Use enclosed counter plug for electrical connection  Connect   only when power is off    Maximum ohmic load for the relays is 2 A   45 VDC or   4 A   250 VAC     Phoenix Combicon  6pol  SP 1   The contacts are shown in   state of rest position  i e   one switching function  off     DILI    For process control this output of the VD9 provides switching functions for 2  switch over relays SP1 and SP2     3 6 Analog Output    CAUTION This output can be configured as recorder output or as steady  control output signal  see chapter 5 6    Minimum load resistance for this output is 10 kQ     Jack  female  1  Analog Out  0 10V max 10kQ  2 2  AGND    The recorder output provides an analog voltage signal representing the actual  absolut pressure  The measuring range is hereby mapped to a voltage range of  0 10 V    If required the recorder output can be scaled as described in chapter 5 6  In this  case the scaled pressure range is mapped to 0 10 V     THYRACONT 41    Analog Recorder Output   Transducer Type 1  see chapter 5 3     linear  1   1400 mbar corresponds to 0   10 V    Other Transducer Types  logarithmic according to the following formula  up to max  1000mbar    Vout    10V   Decades  x   log p mbar     expo       Decades  Number of decades that are covered by the measuring range   p  Actual pressure in mbar    exp0  Exponent off the lower measuring range limit in mbar    Example    Transducer Type 2  1400 0 001 mbar    see chapter 5 3    
6.  Gas Correction Factors         uurssuunnseennnnnennnnnnnnnnnnnnnnnnnnnnnnnn nn  4 6 Communication Protocol                 ursnersnneennnennnnnnnnnnnnnnnnnnnnenn  4 7 Survey of COMMANAG            ccceccceceeeeeeeeeeeeeeeeeeeeeeeaeeeseaeeeeaeeeeaes  5 Eonfiguration          u u  en  5 1 Display Unit   4 e ee ee  5 2  Adjustment             ur 444404ennnnnnnennnnnnnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnannnn  5 3 Transducer Type        2nnuenessnnnennennnnnnnnnnnnnnnnnnnnennnnnnnnnnnnnnennn nn  5 4 Switching Mode    snennnnnnnnnnennnnnnnnnnnnnnnnnnnnnnnnnn  5 5 Control Characteristic              nnnnnnennnnnnnnnnnnnnnnnnnnnnnnnnnnn  5 6 Analog Output Type        nnersnsesssssnnsnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn  5 7 Otan MOC veaas cxtcastnencecadinceedsanndetvadsnedecsamn devvsdsnecedsast bebedish ctedsans es  5 8 Scaling the Analog Output                 u0snneensnnnnnnnnnnnnnnnnennnnnnn  6 Maintenance and Service           ccsccccsecesssceseeeeeeeeeeeeeeesseesenseeeeeees  7 Technical   D  ta          12   2222222200  ee  Declaration of CONfOrmity            ccccecececeeeeeeeeeeeeeeeeeeeeeseeeeesaeeesaaeseeneeees    Hersteller Manufacturer Constructeur        Thyracont Vacuum Instruments GmbH  Max Emanuel StraBe 10   D 94036 Passau   Tel     49 851 95986 0   Fax     49 851 95986 40   email  info thyracont de   Internet  http   www thyracont com    33    34    VV    VV WV    THYRACONT    Safety Instructions    Read and follow the instructions of this manual   Inform yoursel
7.  related parameter for Pl feedback is set    Range of Pl parameter  1   hard   to 8   soft       4 4 Start or Stop Control    When the VD9 display shows the actual pressure reading  measurement  mode   control can be started and stopped by means of the  set  key     Control on   Active switching outputs are displayed  with symbols S1 and S2     Control off        THYRACONT 47    4 5 Gas Correction Factors    For vacuum transducers whose pressure reading depends on type and  composition of the gas a correction factor can be entered into the VD9 in  order to adjust the pressure reading  With this method a correct pressure  reading is again achieved then over a wide range    The value for the gas correction factor can be found in the operation  instructions of the transducer or in suitable technical literature    The range of the gas correction factor is 0 20 to 8 00     In order to adjust the gas correction factor     Press the menu key  several times   until the gas correction factor  c  is shown in the display        Adjust the factor by means of the  Up  and  Down  keys     Confirm with  Set  key     If no key is pressed for 5s or if changes are quitted by the menu key   changes are lost and the primary state is restored  undo function      4 THYRACONT    4 6 Communication Protocol    Communication is carried out according to the Thyracont protocol  The com   mands are sent as ASCII code in the following command frame                             Address Code Data cks CR   Ad
8. 10 4 mbar    Unlock setpoint 2  001s2v R   001s2v R    Set setpoint 2 to  0015420016q  r   0015420016q  r    4 2x10 4 mbar   Unlock gas correction factor 1  001c1e R   001c1e R    Set gas correction factor 1  001c000120W  R   001c000120W       to 1 20                   Before writing the parameters  s   setpoint    c   correction factor    h   hystere   sis  and  p   Pl parameter  they must be activated by sending the related unlock  command     For further information please refer to the separate description of the Thyracont  communication protocol     5 THYRACONT    5 Configuration    CAUTION    Changes of the VD9 configuration should be carried out  AN only by qualified personnel     In order to change settings of the VD9 in the configuration mode       m Hold the  set  key pressed when the      IL O VD9 is connected to mains supply until         5s   the display shows  unit      5 1 Display Unit    The unit for pressure display can be set to mbar  bar  hPa  Pa  mTorr or  Torr    To select the pressure unit switch the VD9 into configuration mode as  described in chapter 5  so that the display shows  unit        mbar    I OL  At the right border of the display the currently selected unit is shown     OO Switch the unit by means of the  Up  and  Down  keys     Confirm with  Set  key     The display then shows the next configuration point  In order to return to  measurement mode quit all further configuration points by pressing the  menu key     THYRACONT 51    5 2 Adjustment
9. Actual pressure  0 5 mbar   Decades  6   expo   3  1 0e 3mbar     Vout    10V   6  x   log 0 5      3      4 50 V      Steady Control Output     Actuating Variable  0 V corresponds to 0   10 V corresponds to 100     3 7 PC Connection RS232    SubD 9pin  male     1 Oo 6   2 6   7 Pin2  RxD  4l oS Pin3  TxD  se Pin5  GND    The RS232 interface can be connected to a serial PC port  e g  in order to  document measurements with the Windows Software VacuGraph       2 THYRACONT    4 Operation    4 1 Startup    First connect the required transducers to the corresponding signal inputs  4 20mA or 0 10V    The control lines of the relay outputs or the continuous control signal are  to be wired according to the particular requirements    At last the mains cable has to be connected     After connection to mains supply the VD9 display shows the following  information       Display test  all display segments are shown for 2s     Display of software version     Display of transducer type     Display of actual pressure    The instrument is now operating in the measurement mode    If the controller is configured to start mode  run     as described in  chapter 5 7  the VD9 actuates the two relay outputs simultaneously with  the display of actual pressure     Description of keys        Change between measurement mode and the adjustment of set   points  hysteresis or Pl parameters and gas correction factor     Measurement Mode  Start Stop function for controlling  see chap   ter 5 7   Adjustment 
10. Anderung  des Sollwertes wird die zugeh  rige Hysterese zun  chst automatisch    ber einen prozentualen Zusammenhang an den neuen Sollwert adap   tiert  Eine anschlie  ende manuelle Anderung der Hysterese ist ebenfalls  m  glich  Werte werden dabei auf Plausibilit  t zum Sollwert   berpr  ft      Beispiel  Sollwert 1   10mbar   H1 1mbar  10  vom Sollwert   Schaltausgang 1 wird bei 9mbar deaktiviert  bei 11mbar aktiviert       nderung Sollwert 1   1 0mbar   H1 0 1mbar  automatisch      Schaltausgang 1 wird bei 0 9mbar deaktiviert  bei 1 1mbar aktiviert       Hysteresis H1    Actual Pressur    Setpoint S1      Hysteresis H1    switching function  on        Time    THYRACONT ss    Zum Einstellen der Schaltpunkte von S1 oder S2     M Z mbar   si m L  Taste Menu  mehrmals  dr  cken  bis im Display das Symbol  S1  bzw    S2  erscheint      OLO  Mithilfe der Pfeiltasten den gew  nschten Sollwert einstellen     A HE      gt  BE  S1  Best  tigen mit der Set Taste  Im Display erscheint die momentan    eingestellte Hysterese     OO Mithilfe der Pfeiltasten die gew  nschte Hysterese einstellen   Diese darf max  90  des eingestellten Sollwertes betragen     Best  tigen mit der Set Taste     Erfolgt 5s lang kein Tastendruck oder wird eine   nderung mit der Taste   Menu  quittiert  werden die Anderungen verworfen und der urspr  ngli   che Zustand wieder angezeigt  Undo Funktion      Schaltverhalten  2P   2 Punkt Regler  Regelcharakteristik  Pi   Pl Regler       F  r die Schaltausg  nge S1 u
11. ED INT  6 Bytes mit f  hrenden Nullen   FLOAT  6 Bytes  Exponentialformat    4 Bytes Mantisse  entspricht Mantissenwert x 1000   2 Bytes Exponent  Offset 20    FLOAT Werte werden in hPa  mbar    bertragen   Beispiel  Der Wert  460016  in einem Float Datenfeld steht f  r 4 6x1 0    mbar     Schnittstellen Parameter        9600 baud  8 Datenbits  1 Stopbit  keine Paritat    8 THYRACONT    4 7 Befehls  bersicht      ber die serielle Schnittstelle des VD9 sind folgende Funktionen verf  gbar                                                        Ger  tetyp T STRING lesen  Messwert M FLOAT lesen  Schaltpunkt S s FLOAT lesen S  schreiben s   SP1 und SP2   Hysterese H  h FLOAT lesen H  schreiben h   H1 und H2   Gasart Korrekturfaktor   C  c UNSIGNED INT   lesen C  schreiben c  Pl Parameter P  p UNSIGNED INT   lesen P  schreiben p  Regelung start stop A  a BOOLEAN lesen A  schreiben a  Tastaturverriegelung K  k BOOLEAN lesen K  schreiben k  Beispiele   vom VD9  Messwert lesen  001M   R   001M260014K       gt  2 6x10 6 mbar   Regelung starten  001a1c   R   001a1c   R   Schaltpunkt 2 lesen  001S2V  R   001S4000160  R     gt  4 0x10 4 mbar   Unlock Schaltpunkt 2  001s2v R   001s2v R   Schaltpunkt 2 auf  0015420016q r   0015420016q r   4 2x10 4 mbar setzen  Unlock Gasart Korrekturfaktor 1    001c1e  R     001c1e R   Gasart Korrekturfaktor 1  001c000120W      001c000120W      auf 1 20 setzen                   Vor dem Schreiben der Parameter  s   Schaltpunkt    h   Hysterese    p   PI   Par
12. Mode  Confirmation of adjustments    Increment values in adjustment mode    Q0   G    Decrement values in adjustment mode    THYRACONT 43  4 2 Measuring Absolute Pressure    In measurement mode the VD9 display shows the actual absolute pressure    Pressure reading is numerical above 1 mbar  Torr       and exponential below    The state of the switching outputs is indicated by the symbols S1 and S2  which are  shown at the bottom of the display as soon as the corresponding relay is switched on     4 3 Switching Output    The VD9 can be configured as two  or three state controller with or  without Pl feedback  as described in chapter 5 4 5 5     Switching Mode  2P   2 state control    Control Characteristic   _0   on off     The relay outputs S1 and S2 are adjusted by a setpoint and the related  hysteresis  H1 and H2   Both switching outputs work independently of  each other    The hysteresis lies symmetrically around the setpoint  When the setpoint  is changed the related hysteresis is automatically adjusted by means of  a percentage relation  Afterwards the hysteresis can be manually  adjusted  value is checked for plausibility with regard to the setpoint      Example  setpoint 1   10mbar   H1 1mbar  10  from setpoint   Switching output 1 is deactivated at 9mbar  activated at 11mbar     Change  setpoint 1   1 0mbar   H1 0 1mbar  automatically      Switching output 1 is deactivated at 0 9mbar  activated at 1 1mbar       Hysteresis H1    Actual Pressure    Setpoint  1      Hyst
13. Niederspannungsrichilinie wurden folgende Normen herangezogen     EN 55 022  1994     nderung  1997   A1  1995   A2  1997 Klasse B  EN 50 082 2  1995    EN 61 000 4 2  1995   At  1998  EN 61 000 4 3  1996   A1  1998   ENV 50 204  1995  EN 61 000 4 4  1995  EN 61 000 4 6  1896     EN 61 010 1  1993   A2  1995    Passau  16 06 2002    Datum   Ort       THYRACONT  Content  1 Safety Instructions         usnsessnnannnnannnnnnnnnnnnnnnnnnnnnnnnnnnannnnnnnnnnannnnnn  2 The VD9 Display and Control Unit           uueesnssannnnnnnnnannnnnnnnnnnnnn  2 1 F  r Orienlalion an aeiei ee rennen  2 2 Delivery Content            nerseerssennnnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn  2 3  Product Descripti  n        u   en ea  3 MSU eA AU ON as ooo coset veda ccc ect nennen ran a eteusaseestcaneecsuccureesecanee  3 1 Notes for Installation                     444444 44H nnnnnnnn nn nnnn nn ennn nen  3 2 Mains Connection           useersnennnnennnnnnennnnnnnnnnnnnnnnnnnnnnnnnn nenn  3 3 Transducer Connection 4 20 MA          uunseesssennnnennnnnennnnnnnnnnnnn  3 4 Transducer Connection 0 10 V       unenssseessnennnnennnnnnnnnnnnnnnnenn  8 5     Relay Outputs icense eaaa  3 6  Analog Output  een nee  3 7 PC Connection RS232 riiai anaandaa  4  SOOT AU ON ocr aoe awa cence wesc ee cca epa a ae ae E  4 1 St  lle    cede aaa aaea aaa aaa anaE ei aaea  4 2 Measuring Absolute Pressure     sssssseeessreseerssrrrrssrrrresrens  4 3  Switching  Output    es  4 4 Start or Stop Control sssrinin seian inna  4 5
14. SH82MV  1000     1x10   mbar    Amphenol C91E  6pol  weibl   1   2 Pini  Identifikation    6 3  Pin 2    Analog Out  0 10V max 10kQ  siehe Kapitel 3 6     5 4 Pin 3  Signaleingang  0 10V  Pin 4  GND  Pin 5  24 VDC  max  8W    Pin 6  AGND    THYRACONT 3    3 5 Schalt Ausg  nge    ACHTUNG Beigelegten Gegenstecker zum Verdrahten verwenden   Stecker nur in spannungsfreiem Zustand anschlie  en  anste     cken oder abziehen   Ausg  nge mit max  2 A   45 VDC bzw  4 A  250 VAC belasten     e    0 Phoenix Combicon  6pol  a o SP1   E Darstellung der Kontakte in  ia   Ruhelage  d h  Schaltfunk   L z SP2 tion  aus     Zur externen Steuerung stehen an diesem Ausgang die Schaltfunktionen des  VD9 in Form von 2 Relais Schaltausg  ngen SP1 und SP2 zur Verf  gung     3 6 Analogausgang 0 10 V    ACHTUNG Der Ausgang kann als Schreiberausgang oder als stetiger  Steuerausgang konfiguriert werden   Vgl  Kapitel 5 6   Der Mindest Lastwiderstand f  r den Schreiberausgang  betr  gt 10 kQ     Klinkenbuchse  1  Analog Out  0 10V max 10kQ  2  AGND  2    Am Schreiberausgang steht der gemessene Absolutdruck als analoges Span   nungssignal zur Verf  gung    Hierbei wird standardm    ig der Messbereich des angeschlossenen Vakuum   Transmitters auf den Spannungsbereich 0 10 V abgebildet    Bei Bedarf kann der Analogausgang    wie in Abschnitt 5 6 beschrieben  skaliert  werden  In diesem Fall wird der skalierte Druckbereich auf 0 10 V abgebildet     10 THYRACONT    Analoger Schreiberausgang     Messumformer T
15. THYRACONT    Vacuum Instruments       VD9  Digitales Vakuum Anzeige   und Regelgerat    Digital Vacuum Display  and Control Unit          Betriebsanleitung  Operating Instructions    2 THYRACONT    Inhalt  1 Hinweise f  r Ihre Sicherheit             nessnnsnnsnnnnnnannnnnnannnnnnnnnnnnnnnnnnn 3  2 Das Anzeige  und Regelger  t VD9        unsunssnannnnnonnnnnnnnnnnnnnnnnnnnnnnn 4  2 1 ZufOrlentierung   u  r n  ne nennen 4  2 2  Cielernimmang ap a euer 4  23  Produktbeschreibung             ccccccccescccceeeeeceeeeeeeceeeseeeceeeeeeeaeenenees 5  3 Installation ssia i eE AAEN 7  3 1 Hinweise Zur Installation                     usurnnn en nnnnnnnnnnnnnnnnnnnnnnnn 7  83 2 Netzanschluss netten ee 7  3 3  Transmitter Anschluss 4 20 MA       uueesnseeesssnnenenssnnnnnnnnnnnnn ann 7  3 4 Transmitter Anschluss 0 10 V        nnesessnseenessnnnnnnnnnnnnnnnn nenne 8  3 5  SChAlt AUSGANGE aseccisteccvecceacedcusdane vs gis eadccuvsncevustt secetastarvveettuncciaes 9  3 6  Analogausgang 0 10 V     uunennseeeessnnnnnnssnnnnnnnnnnnnnnnnnnnnnn ann nnn ann 9  3 7  PCG Anschl  ss RS232     une 10  4 Betrieb   2      20202  steele clack lahn 11  4 1 Inbetriebnahme    ae ea kanns era 11  4 2 Absolutdruckanzeige               24uuennnnnenennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nn 12  4 3  Schaltausg  nge          usnersnsensseennnnnnnnnennnnnnnnnnnnnnnnnnnnnnnnnnnnn nen 12  4 4 Regelung starten und stoppen          uuuursnnesnensnnennennnnnnnennnnnnn nn 15  4 5  Gasartkorrektur           erseensnsennnenn
16. alog Output Type    If  Pi  is set as control characteristic  the analog 0 10V output  see  chapter 3 6  can be configured as recorder or control output     outP  Output of actual pressure  recorder output    out S1  Output of a continuous control signal f  switching output S1   2 state control  or evacuation  3 state control     out S2  Output of a continuous control signal f  switching output S2   2 state control  or venting  3 state control      To select the analog output signal switch the VD9 into configuration  mode as described in chapter 5  so that the display shows  unit      Press the menu key  several times  until the display shows the currently  selected output type  for control a  Pi  only       OO By means of the  Up  and  Down  keys you can now toggle be   tween  outP    out S1  und  out S2      Confirm with  Set  key     The display then shows the next configuration point  In order to return to  measurement mode quit all further configuration points by pressing the  menu key     THYRACONT 357    5 7 Start Mode    You can select how the VD9 controller should behave when powered up     Stop  Control is inactive after the VD9 is powered on and has to be  started and stopped manually by pressing the  set  key or by  software command    run  Control is started automatically after the VD9 is powered on  and cannot be stopped with the  set  key     CAUTION  It is possible to start and stop control by software com   mand via RS232 independently of the settings made un
17. ameter  und  c   Korrekturfaktor  m  ssen diese zun  chst mit einem Unlock   Befehl freigeschaltet werden     F  r weitere Informationen beachten Sie bitte die gesonderte Beschreibung des  Thyracont Kommunikationsprotokolls     THYRACONT 19    5 Konfiguration    ACHTUNG    Anderungen der Gerate Konfiguration nur von geschul   tem Fachpersonal durchf  hren lassen   Um im Konfigurationsmodus Grundeinstellungen des VD9 zu   ndern       me beim Einschalten des Ger  tes die  set        He Taste gedr  ckt halten  bis die Anzeige  O5s      unit  erscheint     5 1 Anzeigeeinheit    Die Einheit der Druckanzeige kann zwischen mbar  bar  hPa  Pa  mTorr  und Torr umgeschaltet werden    Zum Umschalten der Anzeige Einheit das VD9 wie unter Abschnitt 5  beschrieben in den Konfigurationsmodus schalten  so dass die Anzeige   unit  erscheint       mbar  ro    LT  IL  Am rechten Displayrand wird die aktuell eingestellte Einheit angezeigt       amp    Mithilfe der Pfeiltasten die Einheit umschalten     Best  tigen mit der Set Taste     Im Display erscheint der n  chste Konfigurationspunkt  Um zum Mess   modus zur  ckzukehren  alle weiteren Konfigurationspunkte mithilfe der   Menu  Taste   berspringen     2 THYRACONT    5 2 Nachjustieren    Die Messwertanzeige bestimmter Transmitter kann am VD9 auf Atmo   sphare und oder Nulldruck abgeglichen werden     ACHTUNG  Die angeschlossenen Transmitter ben  tigen vor dem  A Abgleich eine ausreichende Warmlaufphase von min     destens 10 Minuten     Zum Ab
18. ationspunkte mithilfe der   Menu  Taste   berspringen     2 THYRACONT    5 3 Transmitter Typ    Das VD9 kann f  r den Betrieb mit folgenden Vakuum Transmittern bzw   Transmitter Kombinationen konfiguriert werden     Typ1  VSC42MA4  1400 1 mbar  Typ2  VSC42MAA4 und VSP52MA4  1400 1x10   mbar  Typ3  VSP52MA4  100 1x10   mbar    Typ4    VSH82MV  1000 1x10   mbar oder  VSM72MV 1000 5x10   mbar    Typ5  VSC42MA4 und VSH82MV  1000 1x10   mbar oder  VSC42MA4 und VSM72MV  1000 5x10   mbar    Typ6  VSP62MV1  1000 1x10  mbar    Zum Einstellen des Transmittertyps das VD9 wie unter Abschnitt 5  beschrieben in den Konfigurationsmodus schalten  so dass die Anzeige   unit  erscheint     Taste Menu  mehrmals  dr  cken  bis im Display die Anzeige des mo   mentan ne ne erscheint     OO Mithilfe der Pfeiltasten kann nun zwischen den verschiedenen  Typen gewechselt werden     Die Auswahl mit der Set Taste best  tigen     Im Display erscheint der n  chste Konfigurationspunkt  Um zum Mess   modus zur  ckzukehren  alle weiteren Konfigurationspunkte mithilfe der   Menu  Taste   berspringen     THYRACONT z    5 4 Schaltverhalten    Zweipunktregelung  2P    Die beiden Schaltausgange arbeiten unabhangig mit jeweils einstellba   rem Sollwert und symmetrischer Hysterese bzw  PI Parametersatz     Dreipunktregelung  3P     Die beiden Schaltausg  nge arbeiten als Evakuierungs  bzw  Bel  ftungs   kontakt  Einstellbar sind ein gemeinsamer Sollwert und jeweils eine  asymmetrische Hysterese bzw  je ein Pl Param
19. dress  3 Bytes  decimal  address RS232  001   Code  1 Byte  command parameter  upper case character for read com   mand  lower case character for write command   Data  data field  max  6 Bytes  can be absent depending on code   cks  1 Byte  checksum  defined as sum over all Bytes of the fields  address  code and data  modulo 64 plus 64    CR  Carriage Return  ODh  13d     Data Formats     BOOLEAN 1 Byte   STRING  max  6 Bytes   UNSIGNED INT  6 Bytes with leading zeros  FLOAT  6 Bytes  exponential format    4 Bytes mantissa  means mantissa value x 1000   2 Bytes exponent  offset 20    FLOAT values are transmitted in hPa  mbar    Example  Value  460016  in a float type data field means 4 6x1 0    mbar     THYRACONT 49    4 7 Survey of Commands    By the serial interface of the VD9 the following functions are available              Type T STRING read   Measurement Value M FLOAT read   Setpoint S s FLOAT read S  write s   SP1 and SP2    Hysteresis H  h FLOAT read H  write h   H1 and H2        UNSIGNED INT   read C  write c  UNSIGNED INT   read P  write p    Gas Correction Factor  Pl Parameter                            Al gt   VO  iv olo                   Control start stop   BOOLEAN read A  write a  Keylock   BOOLEAN read K  write k  Examples   Action Telegram to Answer telegram  transducer from transducer   Read pressure measurement  001M4     001M260014K p     gt  2 6x10 6 mbar    Start Control  001a1c   R   001aic   R    Read setpoint 2  001S2V R   001S4000160        gt  4 0x
20. e vom Gerat ausgehen und Gefahren  die von  Ihrer Anlage ausgehen   Beachten Sie die Sicherheits  und Unfall Verh  tungsvorschriften   Pr  fen Sie regelm    ig die Einhaltung aller Schutzma  nahmen   Installieren Sie das VD9 unter Einhaltung der entsprechenden Umgebungsbedingun   gen  die Schutzart ist IP20  d h  die Ger  te sind gesch  tzt gegen Eindringen von  Fremdk  rpern   Beachten Sie beim Umgang mit den verwendeten Prozessmedien die einschl  gigen  Vorschriften und Schutzma  nahmen   Ber  cksichtigen Sie m  gliche Reaktionen zwischen Werkstoffen und Prozessmedien  Ber  cksichtigen Sie m  gliche Reaktionen der Prozessmedien infolge der Eigener   w  rmung des Produkts   Ger  t nicht eigenm  chtig umbauen oder ver  ndern   Informieren Sie sich vor Aufnahme der Arbeiten   ber eine eventuelle Kontamination  Beachten Sie im Umgang mit kontaminierten Teilen die einschl  gigen Vorschriften  und Schutzma  nahmen   Legen Sie beim Einsenden des Ger  tes eine Kontaminationsbescheinigung bei    Geben Sie die Sicherheitsvermerke an andere Benutzer weiter    Piktogramm Definitionen    _ WARNUNG  i Gefahr eines elektrischen Schlages beim Ber  hren der  Kontakte  _ WARNUNG  Gefahr von Personensch  den  ACHTUNG    Gefahr von Sch  den an Ger  t oder Anlage  HINWEIS    Wichtige Information Uber das Produkt  dessen Handhabung  oder den jeweiligen Teil der Betriebsanleitung  auf den beson           ders aufmerksam gemacht werden soll    3    4 THYRACONT    2 Das Anzeige  und Regelgerat VD9 
21. efindet sich nun im Messmodus    Falls der Regler  wie in Abschnitt 5 7 beschrieben  startaktiv konfiguriert  wurde  Startmodus  run    steuert das VD9 simultan mit der Anzeige des  Druck Istwertes bereits die beiden Relais Schaltausg  nge     Tastenbeschreibung        Umschaltung zwischen Messmodus und dem Einstellmodus f   Sollwert  Hysterese bzw  Pl Parameter und Gasart Korrektur   faktor     Messmodus  Start Stop Funktion f  r Regelung  s  Abschnitt 5 7   Einstellmodus  Best  tigung der eingestellten Werte    Inkrementieren der Werte im Einstellmodus    Dekrementieren der Werte im Einstellmodus    O00        2 THYRACONT  4 2 Absolutdruckanzeige    Im Messmodus zeigt das Display den momentan gemessenen Absolutdruck   Die Druckanzeige erfolgt oberhalb 1 mbar  Torr     numerisch  unterhalb in  Exponentialdarstellung    Der Status der Schaltausgange wird Uber die Symbole S1 und S2 angezeigt   wobei die Symbole am unteren Display Rand eingeblendet werden  sobald  das zugeh  rige Relais eingeschaltet ist     4 3 Schaltausg  nge    Das VD9 kann  wie unter Abschnitt 5 4 5 5 beschrieben  als Zwei  oder  Dreipunkt Regler mit oder ohne PI R  ckf  hrung konfiguriert werden     Schaltverhalten  2P   2 Punkt Regler    Regelcharakteristik    0   ein aus     Die Relais Schaltausgange S1 und S2 werden mit Sollwert und zugeh     riger Hysterese  H1 bzw  H2  eingestellt  Beide Schaltausgange arbeiten  voneinander unabh  ngig      Die Hysterese liegt jeweils symmetrisch um den Sollwert  Bei 
22. en Sollwert einstellen     Best  tigen mit der Set Taste  Im Display erscheint die momentan  eingestellte Hysterese H1 f  r Evakuierung     OO Mithilfe der Pfeiltasten Hysterese H1 einstellen  diese darf max   90  des eingestellten Sollwertes betragen     THYRACONT 15    Bestatigen mit der Set Taste  Im Display erscheint die momentan  eingestellte Hysterese H2 f  r Bel  ftung     OO Mithilfe der Pfeiltasten Hysterese H2 einstellen  diese darf max   90  des eingestellten Sollwertes betragen     Bestatigen mit der Set Taste     Erfolgt 5s lang kein Tastendruck oder wird eine Anderung mit der Taste   Menu  quittiert  werden die   nderungen verworfen und der urspr  ngli   che Zustand wieder angezeigt  Undo Funktion      Schaltverhalten  3P   3 Punkt Regler  Regelcharakteristik  Pi   Pl Regler       Die Schaltpunkteinstellung erfolgt zunachst wie oben beschrieben Uber  die Eingabe des Sollwertes und der Schaltpunktabstande f  r Evakuie   rung und Bel  ftung  Zus  tzlich wird jeweils der zugeh  rige PI   Parametersatz gew  hlt    Wertebereich des Regelparameters  1   hart   bis 8   weich       4 4 Regelung starten und stoppen    Wird im Display des VD9 der momentan gemessene Absolutdruck ange   zeigt  Messmodus   so kann mit der  set  Taste die Regelung gestartet  und gestoppt werden     Regelung an   Aktive Schaltausg  nge werden mit S1  bzw  S2 symbolisiert     Regelung aus        6 THYRACONT    4 5 Gasartkorrektur    Bei Totaldruck Transmittern  die ein gasartabh  ngiges Messpri
23. eresis H1    switching function  on        Time    4 THYRACONT  In order to adjust the switchpoints of S1 or S2     Ee  HE  tht    Press the menu key  several times    until the symbol  S1  or  S2  is shown in the display      OLO  Adjust the setpoint by means of the  Up  and  Down  keys            mbar  KE  1    Confirm with  Set  key  The display shows the current hysteresis     OO Adjust the hysteresis by means of the  Up  and  Down  keys   Maximum permissible hysteresis is 90  of setpoint     Confirm with  Set  key     If no key is pressed for 5s or if changes are quitted by the menu key   changes are lost and the primary state is restored  undo function      Switching Mode  2P   2 state control  Control Characteristic  Pi   Pl control          The setpoints for switching outputs S1 und S2 are adjusted as described  above  Instead of a hysteresis the related parameter for Pl feedback is  set    Range of Pl parameter  1   hard   to 8   soft      Both switching outputs work independently of each other     Switching Mode  3P   3 state control    Control Characteristic  I 0   on off     The switchpoints are set by a common setpoint with positive hysteresis  H1 for evacuation and negative hysteresis H2 for venting    Hysteresis H1 and H2 can lie asymmetrically around the setpoint  When  the setpoint is changed hysteresis H1 and H2 are automatically adjusted  by means of a percentage relation  Afterwards the hysteresis can be  manually adjusted  value is checked for plausibilit
24. etersatz  eingestellt werden     THYRACONT 3    5 6 Analogausgang    Falls als Regelcharakteristik  Pi  eingestellt wurde  ist der analoge 0 10V  Ausgang  siehe Abschnitt 3 6  als Schreiber  oder Steuerausgang  konfigurierbar     outP  Ausgabe des Druckistwertes  Schreiberausgang    out S1  Ausgabe der stetigen Stellgr    e f  Schaltausgang S1 bei  Zweipunkt  bzw  f  Evakuierung bei Dreipunktregelung    out S2  Ausgabe der stetigen Stellgr    e f  Schaltausgang S2 bei  Zweipunkt  bzw  f  Bel  ftung bei Dreipunktregelung     Zum Konfigurieren des Analogausgangs das VD9 wie unter Abschnitt 5  beschrieben in den Konfigurationsmodus schalten  so dass die Anzeige   unit  erscheint     Taste Menu  mehrmals  dr  cken  bis im Display die momentan gew  hlte  oo des ee erscheint  nur bei Regelcharakteristik   Pi l      OO Mithilfe der Pfeiltasten kann nun zwischen den Einstellungen   outP    out S1  und  out S2  gewechselt werden     Die Auswahl mit der Set Taste best  tigen     Im Display erscheint der n  chste Konfigurationspunkt  Um zum Mess   modus zur  ckzukehren  alle weiteren Konfigurationspunkte mithilfe der   Menu  Taste   berspringen     2 THYRACONT    5 7 Startmodus    Unter diesem Konfigurationspunkt kann eingestellt werden  wie sich der  Regler beim Einschalten der Spannungsversorgung verhalten soll     Stop  Die Regelung ist beim Einschalten des Ger  tes zun  chst  inaktiv und kann mit der Taste  set  gestartet und wieder ge   stoppt werden    run  Die Regelung ist beim E
25. etersatz mit Schaltpunkt   abstand     Zum Einstellen des Schaltverhaltens das VD9 wie unter Abschnitt 5  beschrieben in den Konfigurationsmodus schalten  so dass die Anzeige   unit  erscheint     Taste Menu  mehrmals  dr  cken  bis im Display die Anzeige des mo   mentan an erscheint     Io  Mithilfe der Pfeiltasten kann nun zwischen den Modi  2P  und   3P  gewechselt werden     Die Auswahl mit der Set Taste bestatigen     Im Display erscheint der n  chste Konfigurationspunkt  Um zum Mess   modus zur  ckzukehren  alle weiteren Konfigurationspunkte mithilfe der   Menu  Taste   berspringen     24 THYRACONT    5 5 Regelcharakteristik    Fur die Regelcharakteristik des VD9 gibt es folgende Optionen        0   Regelung ohne PI R  ckf  hrung  Standard    Pi   Regelung mit PI R  ckf  hrung    Zum Einstellen der Regelcharakteristik das VD9 wie unter Abschnitt 5  beschrieben in den Konfigurationsmodus schalten  so dass die Anzeige   unit  erscheint     Taste Menu  mehrmals  dr  cken  bis im Display die Anzeige der mo   mentan a eas tne erscheint     OO Mithilfe der Pfeiltasten kann nun zwischen den Modi  I 0  und   Pi  gewechselt werden     Die Auswahl mit der Set Taste best  tigen     Im Display erscheint der n  chste Konfigurationspunkt  Um zum Mess   modus zur  ckzukehren  alle weiteren Konfigurationspunkte mithilfe der   Menu  Taste   berspringen     Falls Reglertyp  Pi  gew  hlt wird  kann sp  ter im Eingabemenu als  zus  tzlicher Parameter je Schaltkontakt ein eigener Regelparam
26. etpoint with asymmetric hysteresis or with Pl parameter plus  hysteresis is adjustable     To select the switching mode put the VD9 into configuration mode as  described in chapter 5  so that the display shows  unit      Press the menu key  several times  until the display shows the currently  selected switching mode     OO You can now toggle between mode  2P  and  3P  by means of  the  Up  and  Down  keys     Confirm with  Set  key     The display then shows the next configuration point  In order to return to  measurement mode quit all further configuration points by pressing the  menu key     THYRACONT 5    5 5 Control Characteristic    For the control characteristic of the VD9 there are the following options        0   Control without Pl feedback  standard    Pi   Control with Pl feedback    To select the control characteristic switch the VD9 into configuration  mode as described in chapter 5  so that the display shows  unit      Press the menu key  several times  until the display shows the currently  selected control characteristic     OO You can now select the required characteristic by means of the   Up  and  Down  keys     Confirm with  Set  key     The display then shows the next configuration point  In order to return to  measurement mode quit all further configuration points by pressing the  menu key     If characteristic  Pi  is chosen  you can additionally enter the related PI   parameter for each switching output in the adjustment mode     56 THYRACONT    5 6 An
27. f regarding hazards  which can be caused by the product or arise in  your system   Comply with all safety instructions and regulations for accident prevention   Check regularly that all safety requirements are being complied with   Take account of ambient conditions when installing your VD9  The protection class is  IP 20  which means the unit is protected against penetration of foreign bodies   Adhere to the applicable regulations and take the necessary precautions for the  process media used   Consider possible reactions between materials and process media   Consider possible reactions of the process media due to the heat generated by the  product   Do not carry out any unauthorized conversions or modifications on the unit   Before you start working  find out whether any of the vacuum components are  contaminated   Adhere to the relevant regulations and take the necessary precautions when handling  contaminated parts   When returning the unit to us  please enclose a declaration of contamination    Communicate the safety instructions to other users    Pictogram Definition    WARNING  Danger of an electric shock when touching    WARNING  Danger of personal injury    CAUTION  Danger of damage to the unit or system  PLEASE NOTE    Important information about the product  it s handling or about    a particular part of the documentation  which requires special  attention    THYRACONT 3    2 The VD9 Display and Control Unit    2 1 For Orientation    These operating instructions de
28. g Output 0 10 V   gt 10 kQ  Serial Interface RS232  SubD 9pole  male  Protection Class IP 20       THYRACONT 63    C       Erklarung Uber die Konformitat    Declaration of Conformity    Diese Erkl  rung gilt f  r folgend bezeichnete Erzeugnisse     Ger  toart  Vakuum Mess  und Regelger  t  Typenbezeichnung  VD9 52  VDI CV  Zertifikat Nr   261 368    Hiermit wird best  tigt  dass die Produkte den wesentlichen Schutzanforderungen  entsprechen  die in den Richtlinien des Rates zur Angleichung der Rechtsvorschriften  der Mitgkedstasten Uber die elektromagnetische Vertr  gbchkeit  89 336 EWG  ge  ndert  durch 91 26 VYEWG  S2 I31 EWG  83 68 EWG sowie 93 87 EWG und der  Nrederspannungsrichtinie TA 2WEWG ge  ndert durch 93 68 EWG festgelegt sind     Diese Erkl  rung wird abgegeben durch    Thyracont Elektronic GmbH  Max Emanuel Str  10   94036 Passau   Germany    Zur Beurteilung der Erzeugnisse hinsichtlich elektromagnetischer Vertr  glichkeit sowie  der Niederspannungsrichtlinie wurden folgende Normen herangezogen     EN 55 022  1994     nderung  1997   A1  1995   A2  1997 Klasse B  EN 50 082 2  1995    EN 61 000 4 2  1995   At  1998  EN 61 000 4 3  1996   A1  1998   ENV 50 204  1995  EN 61 000 4 4  1995  EN 61 000 4 6  1896     EN 61 010 1  1993   A2  1995    Passau  16 06 2002    Datum   Ort       
29. gleich auf Atmosph  rendruck   Das VD9 wie unter Abschnitt 5 beschrieben in den Konfigurationsmodus  schalten  so dass die Anzeige  unit  erscheint     Taste Menu  mehrmals  dr  cken  bis im Display  AdJ  erscheint     O  Hier          368     Mit der  Up  Taste Abgleich f  r Atmosph  rendruck w  hlen und per  set   Taste best  tigen  Der Wert f  r Atmosph  rendruck wird angezeigt        CO  Bei Transmittern des Typs 1  2 und 5  siehe Abschnitt 5 3  kann  nun mithilfe der Pfeiltasten der Tagesdruck eingestellt werden     Den Abgleich mit der Set Taste bestatigen     Im Display erscheint der nachste Konfigurationspunkt  Um zum Mess   modus zur  ckzukehren  alle weiteren Konfigurationspunkte mithilfe der   Menu  Taste   berspringen     THYRACONT 21    Zum Nullpunkt Abgleich     se na Vor einem Nullpunktabgleich ist sicherzustellen  dass    der Nulldruck im Rezipienten mindestens eine Dekade  unterhalb der Messbereichsgrenze liegt    Bei Transmitter Typ 4 ist kein Nullpunktabgleich m  glich    Bei Transmitter Typ 5 ist ein Enddruck unterhalb 0 1mbar ausreichend     Das VD9 wie unter Abschnitt 5 beschrieben in den Konfigurationsmodus  schalten  so dass die Anzeige  unit  erscheint     Taste Menu  mehrmals  dr  cken  bis im Display  Adu  erscheint            blob      Mit der  Down  Taste Abgleich fur Nulldruck wahlen     Den Abgleich mit der Set Taste best  tigen     Im Display erscheint der n  chste Konfigurationspunkt  Um zum Mess   modus zur  ckzukehren  alle weiteren Konfigur
30. in eine Netzsteckdose mit Schutzkontakt  eingesteckt werden  Hierzu sind 3 polige Netzkabel mit fachge   rechtem Schutzleiteranschluss zu verwenden    7 Netzanschlussbuchse              7  Ger  tesicherung 0 8 AT                                                    Li J  8  Netzanschluss    3 3 Transmitter Anschluss 4 20 mA  WARNUNG Das VD9 muss sich im ausgeschalteten Zustand befinden  wenn   Transmitter angeschlossen werden    Nichtbeachten dieser Anweisung kann zu Sch  den am Ger  t f  h    ren     An diesen Anschluss des Ger  ts k  nnen Thyracont Transmitter f  r Absolutdruck  mit 4 20 mA Signalausgang angeschlossen werden  Die Transmitter werden vom  VD9 mit Strom versorgt  Zul  ssige Typen hierf  r sind     8 THYRACONT    VSC42MA4  1400   1 mbar   SubD  9pol  weibl     Pin 4  4 20 mA  Pin 7  24 VDC  max  5W    O  O  O  O  O    akon  o000  OONO    VSP52MA4  20   0 001 mbar     1 0   SubD  9pol  weibl    Pr 9 o  7 Pin3  4 20 mA   4  o9 2 Pin 7  15 VDC  max  5W   gt   0   Pin 9  AGND    3 4 Transmitter Anschluss 0 10 V    WARNUNG Das VD9 muss sich im ausgeschalteten Zustand befinden  wenn  Transmitter angeschlossen werden   Nichtbeachten dieser Anweisung kann zu Sch  den am Ger  t f  h     ren     An diesen Anschluss des Ger  ts k  nnen Thyracont Transmitter f  r Absolutdruck  mit 0 10 V Signalausgang angeschlossen werden  Die Transmitter werden vom  VD9 mit Strom versorgt  Zul  ssige Typen hierf  r sind      VSP62MV  1000   1x10    mbar     VSM72MV  1000   5x10   mbar     V
31. inschalten des Ger  tes automatisch  eingeschaltet  Sie kann nicht per Tastendruck gestartet oder  gestoppt werden     ACHTUNG  Ein Starten oder Stoppen der Regelung per Softwarebe   fehl   ber RS232 ist unabh  ngig von den hier gemachten  Einstellungen in jedem Falle m  glich     Zum   ndern des Startmodus das VD9 wie unter Abschnitt 5 beschrieben  in den Konfigurationsmodus schalten  so dass die Anzeige  unit  er   scheint     Taste Menu  mehrmals  dr  cken  bis im Display die momentan gew  hlte  Start  ae a    OO Mithilfe der Pfeiltasten kann nun zwischen den Einstellungen   Stop   und  run  gew  hlt werden     Die Auswahl mit der Set Taste best  tigen     Im Display erscheint der n  chste Konfigurationspunkt  Um zum Mess   modus zur  ckzukehren  alle weiteren Konfigurationspunkte mithilfe der   Menu  Taste   berspringen     THYRACONT 2    5 8 Skalierung Analogausgang    Ist der Analogausgang auf Typ  outP  eingestellt  kann skaliert werden   welcher Druckbereich auf das Ausgangssignal 0 10V abgebildet werden  soll    Zum Skalieren des Analogausgangs das VD9 wie unter Abschnitt 5  beschrieben in den Konfigurationsmodus schalten  so dass die Anzeige   unit  erscheint     Taste Menu  mehrmals  dr  cken  bis im Display die Anzeige  SCAL   erscheint           Sin    Mit den Pfeiltasten wird zun  chst gew  hlt  ob der obere   S Hi   oder  untere   S Lo   Skalenendwert eingestellt werden soll         gt   1 3      Die Auswahl mit der Set Taste best  tigen  Es erscheint der aktue
32. justment on zero pressure     Confirm adjustment with  Set  key     The display then shows the next configuration point  In order to return to  measurement mode quit all further configuration points by pressing the  menu key     THYRACONT 53    5 3 Transducer Type    The VD9 is configurable for operation with following vacuum transducers  and transducer combinations     Type1  VSC42MA4  1400 1 mbar  Type2  VSC42MA4 and VSP52MA4  1400 1x1 0   mbar  Type3  VSP52MA4  100 1x10   mbar    Type4  VSH82MV  1000 1x10   mbar or  VSM72MV 1000 5x10   mbar    Type5  VSC42MA4 and VSH82MV  1000 1x1 0   mbar or  VSC42MA4 and VSM72MV  1000 5x10   mbar    Type6  VSP62MV1  1000 1x10  mbar    To select the transducer type switch the VD9 into configuration mode as  described in chapter 5  so that the display shows  unit           EFT     Press the menu key  several times  until the display shows the currently  selected transducer type     lo  You can now select the required type by means of the  Up  and   Down  keys     Confirm with  Set  key     The display then shows the next configuration point  In order to return to  measurement mode quit all further configuration points by pressing the  menu key     5 THYRACONT    5 4 Switching Mode    2 State Control  2P    Both switching outputs work independently of each other  each with  adjustable setpoint  symmetric hysteresis or Pl parameter     3 State Control  3P     The two switching outputs operate as relays for evacuation and venting   A common s
33. led out correctly and must be  attached to all vacuum gauges and components  which are sent back to us for repair or service   Otherwise delays will be the consequence  This declaration must be filled out and signed by  authorized and qualified staff only        2 Reason for Return                3 Used Machinery Materials          4 Harmful Contamination of the Product    toxic noo  corrosive non  microbiological non  explosive non  radioactive non  other substances noO       5 Harmful substances and dangerous products of reaction which were in  contact with the product     Chemical Steps in case of First ald in case  Identification tego escape of the harm  of an accident    6 Legally Binding Declaration    guarantee that all statements in this form are are correct and complete  The dispatch of the  contaminated products will be arranged according to legal regulations     Name          62    7    THYRACONT    660    oY mm  r    Technical Data    106 0 mm    Je     1284 mm       Display Numeric LCD  4 digits  with background illumination  45 x 20 mm    Display Refresh Rate 2 Hz  0 5 s   en  Voltage Supply 95     265 VAC  50 60 Hz    Fuse    Ambient Temperature 5   50   C    0 8 A T    Power Consumption max  15 W  incl  transducer     Storage Temperature  20    60   C  Measuring Input 2x 4 20 mA  SubD 9pole  female  1x 0 10 V  Amphenol 6pole  female  Switching Output 2 x relay  potential free switch over contacts   Phoenix strip terminal 6pin   4 A   250 VAC  2 A 45 VDC  Analo
34. lle  Skalenendwert     OGO Mit den Pfeiltasten nun den Skalenendwert einstellen     Die Auswahl mit der Set Taste best  tigen  das Display zeigt wieder   SCAL   Nun kann die Einstellung f  r den zweiten Skalenendwert  erfolgen     SH     Mit den Pfeiltasten zun  chst den zweiten einzustellenden Skalenendwert  ausw  hlen   S Hi  oder  S Lo       2 THYRACONT            Uc      Die Auswahl mit der Set Taste bestatigen  Es erscheint der aktuelle  Skalenendwert     OGO Mit den Pfeiltasten nun den Skalenendwert einstellen     Fro    Die Auswahl mit der Set Taste bestatigen  das Display zeigt wieder   SCAL       set  Die Einstellungen mit der Set Taste bestatigen und den Konfigu   rationsmodus beenden  Das VD9 schaltet in den Messmodus     THYRACONT 29    6 Wartung und Service    WARNUNG    Vorsicht bei kontaminierten Teilen    Es kann zu Gesundheitssch  den kommen  Informieren Sie  sich vor Aufnahme der Arbeiten Uber eine eventuelle Kontami   nation  Beachten Sie beim Umgang mit kontaminierten Teilen  die einschlagigen Vorschriften und SchutzmaBnahmen     Das Ger  t ist wartungsfrei    u  erliche Verschmutzungen k  nnen mit einem  feuchten Tuch beseitigt werden   Sollte wider Erwarten ein Schaden an Ihrem VD9 auftreten  senden Sie das  Ger  t bitte mit einer Kontaminationserkl  rung zur Reparatur an uns     ACHTUNG    Fehlermeldungen    Das Ger  t ist nicht zur kundenseitigen Reparatur vorgesehen        Anzeige    M  gliche Ursache    Behebung        ur    Messbereich unterschritte
35. logausgang 0 10V   Signaleingang 0 10V f  VSP62 VSM72 VSH82  Serielle Schnittstelle RS232  Relaiskontakte    ANOoaRWD         3A     N  oO    6 THYRACONT    Bestimmungsgem    e Verwendung   Das VD9 dient in Verbindung mit Transmittern der Firma Thyracont zur Messung  und Regelung von Absolutdruck  Es darf nur an geeignete und hierf  r vorgese   hene Komponenten angeschlossen werden     Nicht bestimmungsgem    e Verwendung   Als nicht bestimmungsgem     gilt der Einsatz zu Zwecken  die von oben genann    ten abweichen  insbesondere      der Anschluss an Ger  te oder Komponenten  die laut ihrer Betriebs   anleitung hierf  r nicht vorgesehen sind     der Anschluss an Ger  te  die ber  hrbare  spannungsf  hrende Teile  aufweisen     Bei nicht bestimmungsgem    en Einsatz erlischt jeglicher Haftungs  und  Gew  hrleistungsanspruch    Die Verantwortung im Zusammenhang mit den verwendeten Prozessmedien  liegt beim Betreiber     THYRACONT 7    3 Installation  3 1 Hinweise zur Installation    WARNUNG Keine eigenm  chtigen Umbauten oder Ver  nderungen am Ger  t  vornehmen  Vor dem Anschlie  en der Spannungsversorgung da   rauf achten  dass der auf dem Typenschild des VD9 angegebene  Spannungsbereich mit der   rtlichen Netzspannung   bereinstimmt     Aufstellungsort  Innenr  ume  F  r nicht vollklimatisierte Betriebsr  ume gilt     Temperatur   50      506  Rel  Luftfeuchte  5   85   nicht betauend  Luftdruck  860   1060 hPa    3 2 Netzanschluss    WARNUNG       Der Netzstecker darf nur 
36. n       or    Messbereich   berschritten                    Erri  Druck Transmitter nicht Steckverbindungen und  angeschlossen Leitung   berpr  fen  Falscher Ger  tetyp Richtigen Ger  tetyp  eingestellt einstellen  Kapitel 5 3   Druck Transmitter defekt Transmitter zur Reparatur  einschicken   Err4  EPROM Lesefehler VD9 einschicken          3 THYRACONT    Kontaminierungserklarung THYRACON T    ACHTUNG  Diese Kontaminierungserkl  rung muss korrekt und vollst  ndig ausgef  n allen  Vakuumger  ten und  komponenten beigelegt werden  die Sie zur Reparatur oder Wartung an  uns zur  cksenden  Ansonsten kommt es zu einer Verz  gerung der Arbeiten  Diese Erkl  rung  darf nur von autorisiertem Fachpersonal ausgef  llt und unterschrieben werden        2 Grund f  r die Einsendung                   3 Verwendete s  Betriebsmittel          4 Einsatzbedingte  use women Kontaminierung des Produkts  toxisch nein         tzend nein  Im  rticogich nein  explosiv nein     radioaktiv nein       sonst  Schadstoffe nein            5 Schadstoffe und prozessbedingte  gef  hrliche Reaktionsprodukte  mit denen das Produkt in Kontakt kam           6 Rechtsverbindliche Erkl  rung     Hiermit versichere n  ich wir  dass die Angaben in diesem Vordruck korrekt und vollst  ndig sind   Der Versand des kontaminierten Produkts erfolgt gem     den gesetzlichen Bestimmungen     Firma institut Name _   Stra  e   PLZ  Ort   Telefon   Telefax 5   Email Firmenstempel  rechtsverbindiche Unterschrift          THYRACONT 
37. nd S2 werden jeweils  wie oben beschrie   ben  die Sollwerte eingegeben  Statt der Hysterese wird jeweils der  zugeh  rige Regelparametersatz f  r die Pl R  ckkopplung eingestellt   Wertebereich des Regelparameters  1   hart   bis 8   weich      Beide Schaltausg  nge arbeiten voneinander unabh  ngig     Schaltverhalten  3P   3 Punkt Regler    Regelcharakteristik   _ 0   Ein Aus     Die Schaltpunkte werden in Form eines gemeinsamen Sollwertes mit  positiver Hysterese H1 f  Evakuierung bzw  negativer Hysterese H2 f   Bel  ftung eingestellt      Die Hysteresen k  nnen asymmetrisch zum Sollwert liegen  Bei Ande   rung des Sollwertes werden die Hysteresen zun  chst automatisch   ber  einen prozentualen Zusammenhang an den neu eingegebenen Sollwert    14 THYRACONT    adaptiert  Eine anschlie  ende manuelle   nderung der Hysteresen ist  ebenfalls m  glich  Werte werden dabei auf Plausibilitat zum Sollwert    berpr  ft      Beispiel  Sollwert   100mbar   Hysterese H1   10mbar  H2   20mbar  Schaltausgang f  r Evakuierung S1 wird bei 100mbar deaktiviert  bei  110mbar aktiviert    Schaltausgang f  r Bel  ftung S2 wird bei 100mbar deaktiviert  bei  80mbar aktiviert       Hysteresis H1    Actual Pressure    Setpoint      Hysteresis H2    switch SP1    on       switch SP2    off          Time    Zum Einstellen der Schaltpunkte     ur  E mbar    gt     S1 S2    Taste Menu  mehrmals  dr  cken  bis im Display die Symbole  S1  und   S2  erscheinen     OO Mithilfe der Pfeiltasten den gew  nscht
38. nnnnnnnnnennnnnnnnnnnnnnnnnnnnennnnnnn nn 16  46 Kommunikationsprotokoll                  usuerseennnnnnnnnnnnnnnennnnnnnnnnnn 17  4 7 Befehls  bersicht               u  ennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nn 18  5 Konfiguration    ee een 19  5 1  Anzeigeeinheit              eenenenennnennnnnnnnnnnnnnnnnnnnnnnnnn nn nnn nen 19  5 2  Nachjustieren          uuuessnsesersnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nn nnn nen 20  5 3 Transmitter Typ      neeessnneeeersnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nn 22  5 4  Schaltverhalten               nennnneennnnnennnnnnnnnnnnnnnnnnennnnnnnnenn 23  5 5  Regelcharakteristik                     44 nenn nnnnnnnnnnnnnnnnnnnnnnnn nen 24  5 6 Analogausgang seseris irainak anaia earna aA NE 25  S7    St  rtmodUus     ue een 26  5 8 Skalierung Analogausgang                440ersnneennnnnnnnennnnnennnnnnn nenn 27  6 Wartung und Service      uuunnssusnnnnnnnannnnnnnnnnnnnnnnnnnnnunnnnnannnnnnnnnannnnnnn 29  7 Technische Daten         nununsennnennnennnnnnnnnnnnnnnnnnnnn nennen anne 31  Konformit  tserkl  rung sssrinin nasa aana E nn ennn nn 31    Hersteller Manufacturer Constructeur     Thyracont Vacuum Instruments GmbH  Max Emanuel Stra  e 10   D 94036 Passau   Tel     49 851 95986 0   Fax     49 851 95986 40   email  info thyracont de   Internet  http   www thyracont com    VVV VV      VV WV    VV V    THYRACONT    Hinweise fur Ihre Sicherheit    Lesen und befolgen Sie alle Punkte dieser Anleitung   Informieren Sie sich Uber Gefahren  di
39. nzip  verwenden  kann zur Anpassung der Druckanzeige ein Gasart   Korrekturfaktor eingegeben werden  Dadurch wird in weiten Bereichen  wieder eine korrekte Druckanzeige erzielt    Der Wert des einzustellenden Faktors ist der Betriebsanleitung des  verwendeten Transmitters oder geeigneter Fachliteratur zu entnehmen   Der Wertebereich des Korrekturfaktors reicht von 0 20 bis 8 00     Zum Einstellen des Gasart Korrekturfaktors     o        00     Taste Menu  mehrmals  dr  cken  bis im Display der Gasart   Korrekturfaktor  c     erscheint        Per Pfeiltasten den gew  nschten Korrekturfaktor einstellen     Best  tigen mit der Set Taste     Erfolgt 5s lang kein Tastendruck oder wird eine   nderung mit der Taste   Menu  quittiert  werden die Anderungen verworfen und der urspr  ngli   che Zustand wieder angezeigt  Undo Funktion      THYRACONT 17    4 6 Kommunikationsprotokoll    Die Kommunikation erfolgt gemaB Thyracont Protokoll  Die Befehle werden in  folgendem Rahmen als Zeichenfolge im ASCII Code   bertragen                             Address Code Data cks CR   Address  3 Bytes  dezimal  Adresse RS232  001   Code  1 Byte  Befehlsparameter  Gro  buchstaben f  r Lesen  Kleinbuch   staben f  r Schreiben   Data  Datenfeld  max  6 Bytes  kann je nach Code auch fehlen   cks  1 Byte  Checksumme  definiert als Summe   ber alle Bytes der  Felder Addresse  Code und Data  modulo 64 plus 64    CR  Carriage Return  ODh  13d    Datenformate    BOOLEAN 1 Byte   STRING  max  6 Bytes   UNSIGN
40. rly wired earth conductor        Mains Connector           7  Fuse 0 8 AT                                                       8  Socket for mains cable    3 3 Transducer Connection 4 20 mA      WARNING    The instrument must be switched off before any transducers are  connected  Disregarding this instruction may lead to damage of  the instrument     THYRACONT 3  To this port Thyracont transducers for absolute pressure with 4 20 mA signal  output can be connected  The VD9 provides voltage supply for the transducer   Admissible transducer types   VSC42MA4  1400   1 mbar   SubD  9 pole  female    Pin 4  4 20 mA  Pin 7  24 VDC  max  5W    akon   o0000  o000  OONO    VSP52MA4  20     0 001 mbar     1     o      SubD  9 pole female   3  90 7 Pin 3  4 20 mA   4  09  gt  Pin 7  15 VDC  max  5W  a Pin 9  AGND    3 4 Transducer Connection 0 10 V  WARNING  The instrument must be switched off before any transducers are  connected  Disregarding this instruction may lead to damage of  the instrument     To this port Thyracont transducers for absolute pressure with 0 10 V signal  output can be connected  The VD9 provides voltage supply for the transducer   Admissible transducer types      _ VSP62MV  1000   1x10    mbar      VSM72MV  1000   5x10   mbar     VSH82MV  1000   1x10   mbar    Amphenol C91E  6 pole female  1   2 Pini  Identification  6 4 3  Pin 2    Analog Out  0 10V max 10kQ  see chapter 3 6     5 4 Pin 3  Signal input  0 10V  Pin 4  GND  Pin 5  24 VDC  max  8W    Pin 6  AGND   
41. scribe installation and operation of products with  article number    VD9S2     The article number can be found on the product s type label  Technical  modifications are reserved without prior notification     2 2 Delivery Content    Included in the delivery consignment are     VD9 display and control unit   mains cable   counter plug for relay outputs and recorder output  fasteners for panel installation  19  system   operating instructions    Available accessories     Vacuum transducer VSC42MA4  1400     1 mbar  Vacuum transducer VSP52MA4  100     1x10   mbar  Vacuum transducer VSP62MV  1000     1x10    mbar  Vacuum transducer VSM72MV  1000   5x10    mbar  Vacuum transducer VSH82MV  1000     1x10   mbar  Measuring cable f  VSC42 VSP52  2m  W0900002  Measuring cable f  VSC42 VSP52 6m  W0900006  Measuring cable f  VSP62MV  2m  W0606002  Measuring cable f  VSP62MV  6m  W0606006  Measuring cable f  VSM72 VSH82 2m  W1506002  Measuring cable f  VSM72 VSH82 6m  W1506006  RS232 cable for PC connection  2m  W0800002  Windows Software VacuGraph  VGR    3 THYRACONT    2 3 Product Description    The VD9 is designed to display and control absolute pressure  You can connect  Thyracont vacuum transducers with 4 20mA signal output     Via serial interface RS232 the instrument can be remote controlled by a PC              9  THYR CONT   O fer    se 1  8 J ic  1 11  2   j  amo 12   3 lo zZ  1 LCD display  2 Menu key  3 Up Down keys  4 Set key  5 Signal input 4 20mA f  VSC42MA4  6 Signal inpu
42. t 4 20mA f  VSP52MA4  7 Fuse 0 8AT  8 Mains connector 95     265 VAC  50 60 Hz  9 Analog output 0 10V  10 Signal input 0 10V f  VSP62 VSM72 VSH82  11 Serial interface RS232  12 Relay outputs    THYRACONT 37    Proper Use   The VD9 serves exclusively to display and control absolute pressure in combina   tion with Thyracont vacuum transducers  It may only be connected to compo   nents specifically provided for such purpose     Improper Use   The use for purposes not covered above is regarded as improper  in particular     the connection to components not allowed for in their operating instructions    the connection to components containing touchable  voltage carrying parts     No liability or warranty will be accepted for claims arising from improper use     The user bears the responsibility with respect to the used process media     3 THYRACONT    3 Installation    3 1 Notes for Installation  WARNING Unauthorized modifications or conversions of the instrument are  not allowed  Before connecting to mains power make sure that the  supply voltage range stated on the type label complies with your   local mains voltage     Installation location  Indoor  For not fully air conditioned open buildings and operation rooms     Temperature   50      506  Rel  Humidity  5   85   not condensing  Air pressure  860   1060 hPa    3 2 Mains Connection    WARNING The mains connector must be plugged into a mains socket with    protective earth conductor  Use three pole cables  only  with prop   e
43. th  Set  key  The current value for this scale end is displayed      OO  Adjust the scale end by means of the  Up  and  Down  keys         SER      Confirm with  Set  key  the display again shows  SCAL       set  Confirm the adjustments made by means of the  set  key  The  configuration mode is terminated and the VD9 switches into the  measurement mode     6 THYRACONT    6 Maintenance and Service    WARNING Danger of possibly contaminated parts   Contaminated parts can cause personal injuries  Inform your   self regarding possible contamination before you start working   Be sure to follow the relevant instructions and take care of  necessary protective measures     The unit requires no maintenance  External dirt and soiling can be removed by a  damp cloth    Should a defect or damage occur on the VD9  please send the instrument for  repair  enclosing a contamination declaration     CAUTION    The unit is not planned for customer repair     Error messages and malfunction                wrong instrument type  selected    defective transducer    Problem Possible Cause Correction    ur  measurement under range    or  measurement over range    Erri  no transducer connected check connectors and    cables    select correct instrument  type  see chapter 5 3     send transducer for repair        Err4           EPROM reading error       send VD9 for repair          THYRACONT  amp     Declaration of Contamination THYRACONT    ATTENTION  This declaration about contamination has to be fil
44. y with regard to the  setpoint      THYRACONT 45    Example setpoint   100mbar   hysteresis H1   10mbar  H2   20mbar  Switching output for evacuation S1 is deactivated at 100mbar  activated  at 110mbar   Switching output for venting S2 is deactivated at 100mbar  activated at  80mbar       Hysteresis H1    Actual Pressure    Setpoint      Hysteresis H2    switch SP1  on     switch SP2    off          Time    In order to adjust the switchpoints     je mbar    Yams Va  S1 S2    Press the menu key  several times   until the symbols  S1  and  S2  are  shown in the display      OO Adjust the setpoint by means of the  Up  and  Down  keys     mbar    Confirm with  Set  key  The display shows the current hysteresis H1 for  evacuation     OO Adjust hysteresis H1 by means of the  Up  and  Down  keys   Maximum permissible hysteresis is 90  of setpoint     44 THYRACONT    ee mbar   mm   Confirm with  Set  key  The display shows the current hysteresis H2 for  venting     lo  Adjust hysteresis H2 by means of the  Up  and  Down  keys   Maximum permissible hysteresis is 90  of setpoint    Confirm with  Set  key     If no key is pressed for 5s or if changes are quitted by the menu key   changes are lost and the primary state is restored  undo function      Switching Mode  3P   3 state control  Control Characteristic  Pi   Pl control          The switchpoints are adjusted as described above with setpoint  positive  hysteresis for evacuation and negative hysteresis for venting   Additionally the
45. yp 1  s  Abschn  5 3   linear  1   1400 mbar entspricht 0   10 V    andere Messumformertypen  logarithmisch gem     Formel  bis max  1000mbar    Vout    10V   Dekaden  x   log p mbar     exp0       Dekaden  Anzahl der vom Messbereich   berstrichenen Dekaden   p  Druckistwert in mbar   exp0  Exponent der unteren Messbereichsgrenze in mbar     Beispiel    Messumformer Typ 2  1400 0 001 mbar   vgl  Abschnitt 5 3    Istwert  0 5 mbar   Dekaden  6   exp0   3  f  1 0e 3mbar     Vout    10V   6  x   log 0 5      3      4 50 V      Stetiger Steuerausgang     Stellgr    e  0 V entspricht 0   10 V entspricht 100     3 7 PC Anschluss RS232    1     o g SubD 9pol  m  nnl   3 6   7 Pin2  RxD  4 05  5 Pin3  TxD  as Pin5  GND    Dieser RS232 Anschluss kann mit einer seriellen PC Schnittstelle verbunden  werden  um in z B  in Verbindung mit der Windows Software VacuGraph  Messungen zu dokumentieren     THYRACONT 141    4 Betrieb    4 1 Inbetriebnahme    Zun  chst den oder die ben  tigten Messumformer an die entsprechenden  Signaleing  nge 4 20mA bzw  0 10V anschlie  en    Die ben  tigten Steuerleitungen der Schaltausg  nge bzw  des stetigen  Steuerausganges sind entsprechend den Anforderungen zu verkabeln   Zuletzt das Netzkabel anstecken     Nach Anschluss des Netzkabels erscheinen am VD9 folgende Anzeigen        Displaytest  alle Anzeigesegmente werden 2s eingeschaltet     Anzeige der Softwareversion     Anzeige des Transmitter Typs     Anzeige des momentanen Druck Istwerts    Das Ger  t b
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
CPM User`s Manual  Digimerge DNS1010 User's Manual  FERSYSTEM 25 C - Certificazione Energetica  Bedienungsanleitung Infrarotthermometer PCE  ASUS J7824 User's Manual  User Manual — Tobii Studio  ピオニー・ブリオニー取扱説明書  MANUEL DU PROPRIÉTAIRE  Sahel Dimanche - Nigerdiaspora    Copyright © All rights reserved. 