K-Factor Scaler F5140 and Programming Kit F5141

image

Contents

1. blades cuts through the magnetic field generated by the magnetic pick up which in turn generates a frequency output signal that is directly proportional to the speed of the rotor The signal produced is received by the K Factor Scaler input amplifier which has an input sensitivity of 30 mV p p to 30V p p The signal is then sent to an onboard microcontroller which acts as a divisor with a range of 1 to 999 999 999 The divisor K factor is user adjustable through the optional programming software kit The microcontroller handles the dividing process by counting the input pulses and comparing it to the programmed K factor value Once the count equals this value an output pulse occurs for a selectable time period and the counting process starts over In addition if the K factor is set to 1 the K Factor Scaler can be used as a pre amplifier where the frequency from a low level turbine meter is proportional to the logic level frequency output needed by a PLC or CPU data acquisition card This option allows the end device to control the dividing process of the turbine meter output to a recognizable flow rate SPECIFICATIONS External Power Input Voltage 8 5 to 30 Vdc Diode protected Maximum Current draw 18 mA using internal resistor 30 Vdc input Environmental Operating Temperature 22 F 30 C to 158 F 70 C Inputs Magnetic Pick up Input Frequency Range 0 to 4000 Hz Trigger Sensitivity 30 mV p p to 30 V p p Output
2. Flo tech K Factor Scaler F5140 and Programming Kit F5141 Installation amp Operating Instructions Flo tech K FACTOR SCALER F5140 S N 12345 8635 Washington Ave Racine WI 53406 U S A Tel 262 639 6770 Fax 262 639 2267 wf m 8635 Washington Avenue Racine WI 53406 USA Tel 800 433 5263 or 262 639 6770 Fax 800 245 3569 or 262 639 2267 E Mail flo techsales racinefed com www flo tech com Flo h K Factor Scaler o tech Installation amp Operating Instructions TABLE OF CONTENTS POMERAT CCN ecte es capces ces encasement aeee eeen e ane eene aceea a e renane Cenan seana roan ease saa aea iioa cect aaa 3 Operating Principle esesssooooeeseesssssssseeeseessessssssesseessessssssssseeeseesesssssseesreessesssssesseerseessosssessteserosesssssseeseeeoresssssesseersresssssssss 3 Specifications na ansieae r iaei nae EEE E dened ania ee Es ESASEN ES EEEE EEES Niine 3 Installation iiion nien E AE A A E S NAE E E EASE AES EE A 4 ENnclOSUTe MOUNTING AE EAE A IE EEE T EIE IEA 4 edoan ET A EE EEEE EEEE IEE EEA EEE E ET EEE 4 Turbine I CSR cscs enscseschcasccescxssxcdctescvcrcessencressancaconensaxanesccqevaisendecusntaacacer r TSE ET ELC SOK ASSET SESE ni SEa Kae aisi eaS 4 ea EL ETH O70 ao i AEE EA EEES EE E R A EE E N EN 4 Start up and Configuration Using Optional Programming Software Kit F5141 0 esseesseccneeeneenee 5 STOW AN Coos cc esses states So ne ade etched a a e etd haan tes ten aan atcha AS 5 S
3. Signal Max Voltage 30 Vdc Max Power 0 25 W Pulse Type using internal pull up resistor VH Power Input Voltage 0 7 Vdc VL Less then 0 4 V max input power Pulse Type using external pull up resistor VH Input Voltage to external pull up resistor VL VH selected resistor value 47 x 47 Pulse Length 150us 1ms 25ms 100ms 500ms 1s or Auto mode selectable Output Options Jumper Configurations Internal Pull up 3 6K Q to device power External Pull up Open collector Enclosure ABS with mounting flanges Form FLIT 555 11 10 Page 3 Flo tech a K Factor Scaler Installation amp Operating Instructions INSTALLATION The programmable K Factor Scaler was designed to provide a field adjustable frequency divider that converts the low level frequency output from a turbine meter into a scaled square wave output signal Refer to Figure 1 below for the I O terminal connections The board connections include Power Input Turbine Meter Input and the Pulse Output to a totalizing device PROGRAMMING PORT INTERNAL 3 6K PULL UP RESISTOR JUMPER COS e3 OUTPUT L OUTPUT E VIN E VIN L 8 5 30 VDC L TURBINE E PICK UP Figure 1 Board Connections Enclosure Mounting The enclosure has four 4 mounting holes for easy mounting Power The K Factor Scaler requires 8 5 30 VDC to operateand is protected by a diode Polarity is shown in Figure 1 above Turbine Meter The turbine meter c
4. age lt lt No Response gt gt after trying to program the board All information under the Program Values column must be entered before the software will allow the board to be programmed The PROGRAM button programs the K factor Pulse Width and Pulse Output to the K Factor board After the program function is complete the board is automatically read and verified The K Factor board will retain these values when disconnected from power and can be re programmed at any time Form FLIT 555 11 10 Figure 2 Interface connection Flo tech K Factor Progi 5 x Eile Tools r Program Values Board Values m K Factor m K Factor tz ES aaahh Pulse Width Pulse Width C 150us C 150us C ims C ims C 25ms C 25ms C 100ms C 100m C 500ms C 500ms o ls C Auto C Auto Pulse Output r Pulse Output C High TL C High TL iiaa SES C Low E Program Read Verify Status Idle Com Port 1 Figure 3 Programming software screen Page 5 Flo tech ou K Factor Scaler Installation amp Operating Instructions The READ button reads the current information from the K Factor board and displays it under the Board Values column of the PC software The VERIFY button performs the same function as the READ button but compares the Board Values to the Program Values and displays an error if the two do not match Configuring and programming t
5. alues may be required by different setups The external pull up resistor is connected between the receiving device s input and external power source This power source would be the maximum input voltage of the pulse to the receiving device Refer to the following equation to help determine the resistor value needed Page 4 Form FLIT 555 11 10 K Factor Scaler Installation amp Operating Instructions Flio tech i Supply Voltage Where R Resistor value in Ohms Supply Voltage External supply voltage connected to the external pull up resistor l Input current required by the receiving device in amps After the Resistor value has been calculated make sure in the following equation that P is less than or equal to 0 25 P represents the power capability of the output and should not exceed 0 25 Watts Exceeding this value could damage the K Factor board Raising the Resistor value will decrease the available power output P Supply Voltage Supply Voltage R 47 STARTUP and CONFIGURATION USING OPTIONAL PROGRAMMING SOFTWARE KIT F5141 The programmable K Factor Scaler can be factory or user configured through a serial port of a PC by a Windows compatible software utility A programming adapter is required that interfaces the serial port from the PC to the programming port on the board To program the K Factor Scaler begin with the power turned off and connect the adapter cable to the K Factor Scaler
6. board See Figure 2 Then connect the Serial to TTL converter black box to the adapter cable Next attach the serial extension cable to the Serial to TTL converter and connect the opposite end to an available 9 pin serial port of a PC Lastly make sure power is being supplied Power must be supplied to the K Factor Scaler in order to program The programming software is divided into two columns as shown in Figure 3 The Program Values column is the information that is to be programmed into the K Factor board and can be entered by the user The Board Values column is Read ONLY and indicates the information that the K Factor board currently contains The Board Values column will only display the contents of the board after a program read or verify function was performed The proper computer serial port must be selected within the K Factor Programming software for it to communicate with the board This is done by going to TOOLS COM PORT from the menu bar and selecting the Com Port 1 through 16 that the serial programming cable is connected to on the computer If the serial port selected is invalid the software will show the message ERROR Invalid Com Port when trying to program the board If the serial port selected is the incorrect port or if there is a problem with the cable the software will show the mess
7. etting the K factor nisinsin eas sa iad Ba a a a Nai 6 Setting the output pulse Width aiciicsusniiacsnciniidcionisanwasaniueaal enna ee 6 Setting the output level to normally high or normally low 0 cc escsssssssssssssecsecessecseecsecessceseessceneeseeeseees 6 Setting the output pulse to use the internal or external pull up resistor cccscssccesseseceeeseecseees 6 Return Goods Authorization eeseeeeeeeetteeterteetsersesresresresrrerrerrsrssessessessessessssssessessensensesseseesesseeseeseesersereereerereeereerereere 7 Waste Electrical and Electronic Equipment WEEE Directive seessesessesssssssesssrssesssssssssssssssssssseneeneeneeneeneenese 7 Statement of Warranty 55sec sas ieee Sea Sel eat ee haere eee aa 8 Page 2 Form FLIT 555 11 10 K Factor Scaler Installation amp Operating Instructions Flio tech INTRODUCTION The Flo tech K Factor Scaler is a field adjustable frequency divider which interfaces the output signal from a turbine meter with a magnetic pick up to the input of a PLC RTU CPU data acquisition card or similar totalizer device The adjustable frequency divisor referred to as the K factor allows the pulses being sent from a turbine meter to be divided into a recognizable unit that an end device such as a PLC can count and display OPERATING PRINCIPLE Fluid moving though a turbine flow meter causes the rotor to rotate in relation to the flow rate The rotation of the rotor
8. he K Factor board consists of the following four items Setting the K factor divider Setting the output pulse width Setting the output level normally high or normally low Setting the pulse to use the internal or external pull up resistor Setting the K factor The K factor is the ratio of input pulses per each output pulse and can be viewed as a divisor The minimum K factor can be set to 1 where each input pulse yields an output pulse The maximum K factor can be set to 999 999 999 where it would take this many input pulses to yield one output pulse The K factor is set by entering it into the Program Values column of the software under K Factor The K factor will be programmed into the board when the Program button is pressed but note that all values must be entered before programming is allowed by the software Setting the output pulse width The output pulse width is the length of time the pulse remains active before resetting to its resting state The K Factor has a total of six different pulse widths to choose from Some end devices require that the pulse be a certain length or longer in order for proper detection of each incoming pulse For these devices it s important to select a pulse width that is long enough for the end device to recognize The pulse width option is set by selecting the desired pulse width radio button in the Program Values column of the software The Pulse Width option will be programmed into t
9. he board when the Program button is pressed but note that all values must be entered before programming is allowed by the software In addition to six pre set pulse widths to choose from another option is also available called Auto mode This mode acts in the same manor but does not restrain the output pulse to a specific length Instead it varies and is dependent on output frequency The higher the output frequency the shorter the pulse width output The lower the frequency output the longer the pulse width output This option turns off the Pulse Output option because it does not apply in this mode Setting the output level normally high or normally low Most end devices will be unaffected by this setting but the K Factor Scaler has the ability to invert the output pulse level This option is set by selecting the desired pulse output radio button in the Program Values column of the software The Pulse Output option will be programmed into the board when the Program button is pressed but note that all values must be entered before programming is allowed by the software When the Pulse Output option High is selected the output level is normally low and the duration of the selected pulse width is high When the Pulse Output option Low is selected the output level is normally high and the duration of the selected pulse width is low Setting the output pulse to use the internal or external pull up resistor Either the
10. internal pull up resistor or an external resistor must be used for the K Factor Scaler to provide an output pulse This option is controlled by the onboard jumper and not by software With the jumper installed the internal 3 6K pull up resistor is connected to the input voltage of the board With the jumper uninstalled the internal pull up resistor is disconnected and an external pull up resistor and supply voltage is required Page 6 Form FLIT 555 11 10 K Factor Scaler Installation amp Operating Instructions Flio tech Return Goods Authorization When returning equipment for service a Returned Goods Authorization RGA number must be obtained from our Service Department Please contact them by phone at 800 433 5263 or 262 639 6770 or by e mail to flo techtechhelp racinefed com All returns go to the following address and must include the RGA number on the outside of the box Flo tech Division of Racine Federated Inc 8635 Washington Avenue Racine WI 53406 3738 USA Attn RGA xxx xxxx Waste Electrical and Electronic Equipment WEEE Directive In the European Union this label indicates that this product should not be disposed of with household waste It should be deposited at an appropriate facility to enable recovery and recycling For information on how to recycle this product responsibly in your country please visit www racinefed com recycle Form FLIT 555 11 10 Page 7 Flo tech Division of Racine Federated I
11. ity and the implied warranty of fitness for a particular purpose or application are expressly disclaimed and shall not apply to any products sold or services rendered by Flo tech The above warranty supersedes and is in lieu of all other warranties either expressed or implied and all other obligations or liabilities No agent or representative has any authority to alter the terms of this warranty in any way E lo tec lh 8635 Washington Avenue Racine WI 53406 3738 USA Tel 800 433 5263 or 262 639 6770 fl o tec h CoO m Fax 800 245 3569 or 262 639 2267 E Mail flo techsales racinefed com Materials and specifications subject to change without notice RFD Racine Flow Meter Group Meal DYNASONICS Flo tech HEDTAND PRESO racine 2010 Racine Federated Inc Flo tech is a trademark of Racine Federated Inc All rights reserved WINDOWS is a registered trademark of Microsoft Corp Printed in USA UL is a registered trademark of Underwriters Laboratories Form FLIT 555 11 10
12. nc Limited Warranty and Disclaimer Flo tech Division of Racine Federated Inc warrants to the end purchaser for a period of one year from the date of shipment from the factory that all flow meters manufactured by it are free from defects in materials and workmanship This warranty does not cover products that have been damaged due to misapplication abuse lack of maintenance or improper installation Flo tech s obligation under this warranty is limited to the repair or replacement of a defective product at no charge to the end purchaser if the product is inspected by Flo tech and found to be defective Repair or replacement is at Flo tech s discretion A returned goods authorization RGA number must be obtained from Flo tech before any product may be returned for warranty repair or replacement The product must be thoroughly cleaned and any process chemicals removed before it will be accepted for return The purchaser must determine the applicability of the product for its desired use and assumes all risks in connection therewith Flo tech assumes no responsibility or liability for any omissions or errors in connection with the use of its products Flo tech will under no circumstances be liable for any incidental consequential contingent or special damages or loss to any person or property arising out of the failure of any product component or accessory All expressed or implied warranties including the implied warranty of merchantabil
13. onnections are non polarized Shielded twisted pair wire is recommended for this connection to the turbine meter Pulse Output Either the internal or external pull up resistor must be used for the K Factor Scaler to provide an output pulse This option is controlled by the onboard jumper With the jumper installed the internal pull up resistor is connected Internal Pull up Resistor The internal pull up resistor allows for simple installation but care must be taken to ensure that the device being connected to by the pulse output can handle voltage levels as high as the supply feeding the K Factor board Another important setup consideration when using the internal pull up resistor is to make sure the output pulse from the K Factor board can supply enough current for the receiving device to be able to read the pulse The available current that the K Factor board can supply to the receiving device can be calculated with the following equation Input Voltage 0 7 V 3600 Q 470 Available Current Verify that the receiving device input current requirement is below this value for proper operation Otherwise an external pull up resistor less then 3 6K will have to be used External Pull up Resistor Using an external pull up resistor offers the end user greater flexibility of controlling the output pulse provided by the K Factor board to the receiving device Since power sources and receiving devices differ between users different resistor v

Download Pdf Manuals

image

Related Search

Related Contents

AV-IP03 : MAC - aquila vizion  Kärcher Nass-Trockensauger  V7 Projector Lamp for selected projectors by DUKANE, NEC  第5 機械の導入、利用、管理等に関する事項  取扱説明書 - DXアンテナ  Mode d`emploi Déchaumeur à dents Smaragd 7  EN-MOTION - ProTerapi A/S  Gebrauchsanweisung discovery® delight  Ewent EW9523 networking cable  : 13Jdo1 - Assembleia Municipal de Lisboa  