Troubleshooting Guide

image

Contents

1. 150 600023 Rev 1
2. Troubleshooting Guide Symptom Possible Cause Remedy LabWave 9000 inoperative Ensure that instrument is properly connected to power outlet and that instrument is switched on Check fuses and replace if nec essary Balance will not tare Balance overloaded use less sample Balance not opearting Check 0 5A fuse in instrument pedestal and replace if necessary No microwave drying Door not closed properly Check 15A fuses 8A fuses for 240V units and replace if necessary Sample burning Sample too large and or unevenly applied to pad Sample not homo geneous Power level too high Results not reproducible Sample applied unevenly to pad Sample not homogeneous and absorbs microwaves unevenly Keep sample container closed when not extracting sample Work quickly when applying sample to pads and placing on balance pan Incorrect results Verify proper sample handling and method parameters At the end of the test the Final weight greater than display reads Data Results initial weight Repeat test Invalid Check balance calibration Door Open Message Ensure that instrument door is properly closed A qualified service technician using the schematics located in the Appendix and standard troubleshooting techniques should be able to identify and resolve operational problems If a problem persists call the CEM Corporation Service Department or your local distributor for assistance Labwave 9000

Download Pdf Manuals

image

Related Search

Related Contents

BEDIENUNGSANLEITUNG MANUALE DELL'UTENTE  Untitled - Insieme per la Scuola  IMC Networks iMcV-LIM  SCOUT User`s Guide  MASTER-K10S1 - LS Control Argentina    Rosco Diagnostica  731 lankorep - vieille materiaux  MANUALE DI ISTRUZIONI Tipo SS  Rack Mount Remote Controller (1RU) I&O Manual  

Copyright © All rights reserved.
Failed to retrieve file