        Talyrond 30 (with Ultra) Service Manual TH-SM-30U PAGE 2
         Contents
1.      1 1 7 2 Use a spirit level on the homed C amp L table surface and check that the instrument is level to within 0 1mm m in both planes  by careful fine adjustment of the four adjustable feet  levelling may need refining after being allowed to settle for 24 hours     Talyrond 30  with Ultra  Service Manual TH SM 30U PAGE 2  4 Issue 1 September 2001    1 1 8 Connecting the Instrument    1 1 8 1 Connect the instrument as per the connection diagram in Fig 2 1 below          PSU  265 957    POWER IN    I  GAUGE    K509 1181 01      O    x      U     ee ee ae ae ae a a a l  u u  o u u  a a J    PC  K508 1400 01 265 966       K509 1130 03 K509 1466 01  K510 907 SPINDLE    MO LORZALK A    Figure 2 1 Connecting the instrument    Talyrond 30  with Ultra  Service Manual TH SM 30U PAGE 2 5 Issue 1 September 2001    1 1 9 Connecting the instrument to a compressed air supply    Warning    Indiscriminate use of compressed air can lead to serious injury  Use extreme caution when connecting or dis   connecting compressed air supplies to and within the instrument  Always ensure that the air supply is switched  off or is disconnected before attempting any dismantling  Under no circumstances allow air to be blown directly    onto the skin     Note    The main air supply pipe should be angled down in the direction of the airflow and the supply pipe to the instrument rise  vertically from it  By this means  the oil and moisture in the air delivered to the instrument will be minimized  In addit
2.  i e 0 1mm for the 0 4mm range or 0 25mm for the 2mm range  Select OK     1 5 1 4 In the Instrument Status dialog  right mouse click on the top of the gauge icon and from the drop down menu select  Cali   brate    and from the Calibrate drop down menu select  New   the  New  dialog is displayed  In the  New  dialog select the name of the  Calibration Standard that is currently in use or select a  New Standard  if one does not already exist  Select OK   If a new standard is  selected the  New Standard    dialog is displayed  select the calibration component type and give it an appropriate name  Select OK the  Define Standard dialog is displayed  check that the correct standard type is displayed and enter the correct Step Height    Select OK this  will display a Perform Calibration dialog with a Perform Calibration tab  choose the required Gauge Range  and select OK     Talyrond 30  with Ultra  Service Manual TH SM 30U PAGE 2 9 Issue 1 September 2001    Note        Do not check the Contact Before Measurement checkbox    1 5 1 5 This will start the calibration  a Confirm dialog is displayed asking the user to  Move gauge to first step  select OK  a further  Confirm dialog is displayed asking the user  Move gauge to other step   slide the glass so that the stylus drops onto the second surface   select OK  When the calibration is complete another  Confirm  dialog is displayed follow the instructions on the dialog  Selecting  Yes   will store the calibration constant     1 5 1 6 Ca
3. WARNING   THE INSTRUMENT WEIGHS 38kg AND IS UNSTABLE WHEN CARRIED     THEREFORE  IN ORDER TO LIFT IT FROM ITS PALLET AND PLACE IT IN ITS WORKING LOCATION  THE INSTRU   MENT SHOULD BE LIFTED VERTICALLY  THE SERVICE ENGINEER MUST DECIDE WHETHER TO USE MANUAL  OR MECHANICAL LIFTING DEPENDING UPON THE ENVIRONMENT AND THE DISTANCE TO BE MOVED   Cautions  Do not use the instrument worktable to lift the base unit out ofthe shipping crate   To avoid damage to the precision rotary spindle parts  DO NOT rotate the worktable until the instrument is connected    to the air line and the correct air pressure is applied  The weight of the worktable is usually sufficient to prevent rota   tion when pressurised air is not present     The workbench must be strong enough to support 140kg  300lbs    he Transit bolts  Note   The Talyrond 30 is not fitted with column transit bolts or spindle transit bolts   1 1 6 Setting up the instrument    1 1 6 1 When all of the units are assembled and the TRANSIT BOLT ASSEMBLIES REMOVED  ensure that the equipment does  not rock by adjusting the four feet on the base until they are all in solid contact with the worktable surface  The instrument must then be  levelled  System interconnections can then be made and the air and mains power connected     1 1 7 Levelling    1 1 7 1 If high accuracy roundness measurements are to be made  then the instrument must be level  Any roundness measure   ments taken without proper levelling will contain an elliptical component
4. bar  select  User    and from the drop down menu select the  Select Instrument    option  The     Select Instrument    dialog is displayed     1 4 1 2 Ensure that the  Off Line  check box is unchecked  and select  New On Line Instrument     If an    On Line Instrument    has not  previously been configured  this will be the only option available  Select OK  the  New On Line Instrument    dialog will be displayed     1 4 1 3 Select the Instrument Type e g  TR30 from the drop down menu and give the instrument an appropriate name  Select OK   an    Initial Configure    dialog is displayed     1 4 1 4 Ensure that the hardware appropriate to the Instrument is selected e g  Manual or Automatic C  amp  L Table  Select OK  a     Configure    dialog for the Gauge is displayed   Currently only the Talymin Inductive gauge is available      1 4 1 5 Select OK  a    Configure    dialog for the Stylus is displayed  From the drop down menu select the appropriate stylus     Talyrond 30  with Ultra  Service Manual TH SM 30U PAGE 2 8 Issue 1 September 2001    1 5    1 4 1 6 Select OK an    Initial Configure    dialog is displayed  Enter the scale positions for the Arm  R  and the Column  Z   these are  taken from the Fiducial Scales on the Arm and Column     1 4 1 7 Select OK  the   lt Instrument Name gt   dialog is displayed  it contains information on the selections made  Select OK  the  Ultra software will now be connected to the newly created instrument  Check that the   lt Instrument Name 
5. d Printer    1 1 10 2 With the exception of the mains inlet  connect the system as shown in the diagram in Connecting the RIM  1 1 10 3 Ensure that all connections are made in accordance with the following list   11 11 Switching on    1 1 11 1 The power to the instrument is supplied via the external 12V power supply  265 957   The electrical power to the PC is  controlled via the mains on off switch on the front panel of the PC     1 1 11 2 Once the instrument has been fully connected it may be switched on  Switch on the mains power to the instrument and the  PC and press the on off switch on the computer  Once the computer has run its self check routines  the software can be started using  the on screen Ultra icon     1 1 11 3 To switch off the system  switch off the computer  then switch off the mains power switches to the instrument and the PC     1 1 11 4 The instrument is now ready for operation     1 2 Positioning the gauge on workpiece    1 2 1 The workpiece to be measured is placed on the worktable and the axis of the workpiece aligned with the axis of rotation  A stylus  contacts the workpiece which is rotated about the vertical axis     1 3 Installing Upgrading Ultra software  1 3 1 Ensure that the Ultra Software is completely shut down     1 3 2 Using Windows Explorer  navigate to the directory where Ultra was installed  The default location is  C  Program Files Taylor  Hobson Ultra database     1 3 3 Copy the two files  Licence txt   from C  Program Files Taylor H
6. e correct one for the Gauge Range in use  i e    for the 0 4mm and 2 0mm Ranges use the 300mm Flick Standard and for the 0 8mm Range use the 20mm Flick Stan    dard   1 5 2 2 When the set up is complete  from the Talyrond drop down menu  select Instrument Status  The Instrument Status dialog  will be displayed  right mouse click on the top of the gauge icon and from the drop down menu select  Calibrate    and from the Calibrate  drop down menu select  New   In the  New  dialog select the name of the Standard that is currently in use or select a  New Standard    if one  does not already exist  Select OK the  New Standard    dialog is displayed select  the calibration component type and give it an appropri   ate name  Select OK the Define Standard dialog is displayed  check that the correct standard type is displayed and enter the correct Flat    Talyrond 30  with Ultra  Service Manual TH SM 30U PAGE 2  10 Issue 1 September    Depth  Select OK this will display a Flick Standard dialog with a Perform Calibration tab choose the required Gauge Range  select OK   This will then start the calibration which when complete will display a window showing a Raw Profile view     1 6 Setting Resistive and Phase Balances  1 6 1 Ensure that the Gauge has been calibrated     1 6 2 Place a component centrally on the Centring and Levelling table  using the concentric lines on the table as a guide     1 6 3 From the Talyrond drop down menu select    Instrument Status    the Instrument Status dial
7. gt   is now displayed in the bot   tom RH side on the Status Bar     Calibrating the gauge  Note        Calibration of the Gauge can be carried out using a Flick Standard or by using Slip Blocks   1 5 1 Calibration Using Slip Blocks    Note        When using slip blocks to calibrate the gauge  it is good practice to lower the stylus from the null position by half the   step height e g  For Range 1  the gauge reading below the gauge icon on the Instrument Status dialog is  250mm     This is to ensure that the Gauge is calibrated symmetrically about the null position  1 5 1 1 Ensure that the table is in the  Home  position and apply the spindle clamp  Set the attitude and orientation to  Horizontal  and Down  Wring the slip blocks  2 8mm  3mm and 2 5mm  onto the glass flat and place under the gauge  bring the stylus into position  just above and towards the edge of the 3mm slip block     1 5 1 2 From the Talyrond drop down menu  select  Instrument Status     The Instrument Status dialog will be displayed  Right  mouse click on the top of the gauge icon and from the drop down menu select  Properties   Select the appropriate gauge range i e  2mm  or 0 4mm  Select OK    1 5 1 3 Using the Device Control dialog for the column  select contact and check that the stylus is in contact with the 3mm slip  block  On the Device Control dialog  select Move Relative Z  Negative  the Move Relative Z  Negative  dialog will be displayed  Enter  half the step height for the range being calibrated
8. ion   the supply pipe must be free of sharp bends to avoid airflow restriction    In cases where the quality of the air supply is poor  it may be necessary to install filters  air dryers and automatic drains  into the supply lines     It is essential that the air supply is connected before the instrument is switched on  The airline is connected to the rear  of the base of the instrument  Ensure that there is no leakage from the connection  Slide the red sleeve valve towards    the font of the instrument to switch it on     1 1 9 1 The regulator should always be left in the  locked  position  To adjust the regulator  first release the knob by pulling it  upwards  set the pressure and then push the knob home to lock it again     1 1 9 2 On the front panel of the compressed air enclosure  switch on the air supply  Remove the side panel of the compressed air  enclosure and  observing the pressure gauge  ensure that the regulator is set to 50 Ibf in2 gauge  3 4 bar   If necessary  check and rec   tify any obvious air leaks and adjust the pressure as described in the note above     1 1 10 Interconnection of Instrument  Host Computer and Printer    Warning  Do not make any connections to the mains supply yet     CAUTION    When interconnecting the PC to the VDU and Printer  ensure correct connections are made     Talyrond 30  with Ultra  Service Manual TH SM 30U PAGE 2 6 Issue 1 September 2001    1 1 10 1 Follow the manufacturer s interconnection details supplied with the PC  VDU an
9. librating the 0 08mm Range  Ensure that the table is in the  Home  position and apply the spindle clamp  Set the attitude  and orientation to Horizontal and Down  Wring the slip blocks  3mm and 2 96mm  onto the glass flat and place under the gauge  bring  the stylus into position just above and towards the edge of the 3mm slip block  From the Talyrond drop down menu  select  Instrument  Status   The Instrument Status dialog will be displayed  Right mouse click on the top of the gauge icon and from the drop down menu  select  Properties   Select the gauge range 0 4mm  Select OK     1 5 1 7 Using the Device Control dialog for the column  select contact and check that the stylus is in contact with the 3mm slip  block  On the Device Control dialog  select Move Relative Z  Negative  the Move Relative Z  Negative  dialog will be displayed  Enter   1mm and select OK  On the Device Control dialog  select Move Relative Z  Positive  the Move Relative Z  Positive  dialog will be dis   played  Enter  08mm     1 5 1 8 Repeat paras 4 and 5    1 5 1 9 If a previously used standard is selected  Select OK this will display a Perform Calibration dialog with a Perform Calibration  tab  choose the required Gauge Range  check the Contact Before Measure box and select OK  This will then start the calibration     1 5 2 Calibration Using The Flick Standard  1 5 2 1 Centre and Level the Centring and Levelling Table   Note        When using a Flick Standard to calibrate the gauge  ensure that it is th
10. obson Ultra config  and  Ultra mdb   from C  Program Files Taylor  Hobson Ultra database  to a safe location for example  C       1 3 4 Follow these instructions to Uninstall the existing Ultra Software     1 3 4 1  Click on the Windows START button     1 3 4 2  Select the  Settings  option from the menu   Talyrond 30  with Ultra  Service Manual TH SM 30U PAGE 2 7 Issue 1 September 2001    1 3 4 3  Select the  Control Panel    from the sub menu    1 3 4 4  Double click on the  Add Remove Programs  icon    1 3 4 5  Select Ultra from the list and click on the  Add Remove     button    1 3 5 Using Windows Explorer  navigate to the    Taylor Hobson directory  The default location is  C  Program Files Taylor Hobson  1 3 6 Delete the  Taylor Hobson  directory    1 3 7 Install the new Ultra Software by putting the CD into the CD drive and following the instructions of the auto run install program   1 3 8 Using Windows Explorer  navigate to the directory where you stored the files in step 3  Copy the file  Licence txt  from this backup    location to the  config directory and  Ultra mdb  to the new  database directory  The default location is  C  Program Files Taylor Hob   son Ultra     1 4 Configuring the instrument  Note        The software comes with an Off Line instrument already configured and it is selected by default     1 4 1 To create and configure a new On     Line Instrument that reflects the hardware carry out the following procedure     1 4 1 1 From the Main Window menu 
11. og will be displayed  Right mouse click  towards the top of the gauge icon and from the drop down menu select  Calibrate   From the drop down menu select Resistive Balance   the Resistive Balance dialog will be displayed follow the instructions on the dialog     1 6 4 Right mouse click towards the top of the gauge icon and from the drop down menu select    Calibrate     From the drop down menu  select Phase Balance  the Phase Balance dialog will be displayed follow the instructions on the dialog     Talyrond 30  with Ultra  Service Manual TH SM 30U PAGE 2  11 Issue 1 September 2001    
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
日本ルワンダ学生会議 第三回本会議 活動報告書 2009 年 12 月 18 日  「PITREC on PC」 NPC100 取扱説明書    PASCO Specialty & Mfg. PS-2116A User's Manual  CFA AN0125 5 - CFA AN013 30 - CF AN0140  USER MANUAL  Manual - Munters  Samsung SN-208BB Инструкция по использованию  mopera U 設定マニュアル Android™5.x共通版(PDF形式:1059KB)      Copyright © All rights reserved. 