Model 4046/4146 General Purpose Pump Calibrators User's Guide

image

Contents

1. 7 Dampening Module ee oeaaneemsiaanmmesiaanmmniiannnnsiannneri 8 4 o MIN ON ANCE as sasesesenncnerccsnencnsennsneneeanancassonssensesnaneusaansesennenenesaae 11 Flow Sensor eddies thee tan facies aaah 11 RG COMINCALION aiviet actinides sica aa et iaia etat 11 CAS OS agre aK AA anti menis ta 11 SIOrage iii citi NAB eae al ie dea Nil 11 5 Troubleshooting s ecccecnnncecceneesencenensnnnennseneensnseneenenseaeenenseaeenens 13 APPENDIXES A Specifica ONE INAABANGAN HANDAAN 15 B Primary vs Secondary Standard aeemaameeaaenemaennenen 17 This page intentionally left blank Chapter 1 Unpacking and Parts Identification Carefully unpack the instrument and accessories from the shipping container Check the individual parts against the list of components in Table 1 If any parts are missing or damaged notify TSI immediately Table 1 List of Components see figures 1 1 and 1 2 Part Model Qty Item Description 1 Calibrator 4046 or 4146 1 Filter Model 4046 22 mm 1S0 taper 1602292 1 Model 4146 63 mm HEPA filter with 0 375 inch barbed 1602342 fittings Tubing Kit Model 4046 with connectors and battery pack mounting TK4000 lugs Model 4146 with connectors and battery pack mounting TK4100 lugs 1 Battery Pack 4199 1 Soft Carrying Case Model 4046 1319403 Model 4146 1319404 Model 4046 4146 User s Guide 1980512 Dampening Module Model 4146 only 1040
2. Direction of Flovv Calibrator Outlet Inlet of Active Sampling Instrument i e DustTRAK Figure 2 2 Typical Setup Configurations Setting Up Active Sampling Instrument without internal dampening chamber e g TSI SIDEPAK AM510 Personal Aerosol Monitor Figure 2 3 Typical Dampening Module Set Up Use the 6 in 15 cm tubing to connect the dampening chamber in line between the Model 4146 Calibrator and the instrument For best results position Dampening Module with openings to the side off the work surface 6 Chapter 2 Chapter 3 Operation Overview The Model 4046 4146 Calibrator measures volumetric flow rate inside the flow tube All measurements made by the Models 4046 4146 are NIST traceable ON OFF Switch Slide the switch to the ON position The LCD displays volumetric flow in units of liters per minute L min Flow Rate Measurement Flow rate data will be displayed on the Model 4046 4146 through the LCD Volumetric flow is displayed in units of liters per minute L min Setting Flow Rates and Calibrating Instruments The Model 4046 4146 Calibrator may be used to set flow rates and calibrate sampling pumps or other instruments which draw an active flow sample Be sure to follow all instrument manufacturer s flow setting and calibration procedures 1 Start up the Model 4046 4146 Calibrator and let it warm up for 1 minute prior to setting any flow rates or conducting
3. any calibrations 2 Configure your sampling with the desired sampling media and sampling train to be used 3 Connect the outlet of the Model 4046 4146 Calibrator to the inlet of the sample media used with sampling pumps or to the inlet of the active sampling instrument using the tubing supplied 4 Follow the instrument manufacturer s flow setting and or calibration procedures The volumetric flow in units of liters per minute L min will be continuously displayed on the LCD of the calibrator Dampening Module The Dampening Module is included for use with systems that do not have internal dampening chambers The Dampening Module P N 1040229 can be added in line between the end of the sampling train and the Model 4146 calibrator The Dampening Module reduces flow pulsation with insignificant affect on flow rate or back pressure Figure 3 1 Dampening Module vvith Tubing Attached All diaphragm pumps produce flovv pulsations during operation Most commercially available 3 L min and 5 L min sampling pumps are designed with internal dampening chambers to minimize flow pulsations Some common air sampling instruments that may not have internal flow pulsation dampening chambers are listed below Contact your instrument manufacturer if you have questions This list is just an example and is not intended to be all inclusive e Low flow personal sampling pumps e Confined space meters e Photometers e Optical Particl
4. power requirements Power Supply 7 5 VDC 1 5 V 300 mA maximum DC INPUT 0 0 0 Connecting Filter and Flow Tubes The Model 4046 4146 has a thermal sensor exposed to the air flow that must be protected from foreign matter and particles TSI supplies a filter that should be connected to the inlet of the calibrator however any filter will work as long as it has a minimum efficiency of 99 9 Always use a filter on the inlet of the calibrator Failure to filter the air flow may change the calibration and or permanently damage the sensor Note Flow direction is identified by the large arrow printed on the bottom side of the calibrator and on the bottom of the battery pack Attach the filter to the inlet of the calibrator using supplied tubing and or adapters Connecting a tube to the outlet of the calibrator will create back pressure See Appendix A for calibrator accuracy specifications when operating at various pressures In general minimize back pressure on the calibrator by using shorter lengths of tubing to maintain highest level of accuracy 4 Chapter 2 Sampling Pump with internal dampening chamber Calibrator Filter Calibrator Inlet Ny Direction of Flow Calibrator Outlet Sampling Media Filter Cassette Sorbent Tubes Etc Sampling Pump 2 Active Sampling Instrument with internal dampening chamber Calibrator Filter Calibrator Inlet
5. 229 Table 2 Optional Accessories Qty Item Description AC Adapter 115 V North America ungrounded 2613033 100 240 V NEMA 5 15 plug grounded 8918 NA 100 240 V Europlug CEE 7 16 grounded 8918 EC 100 240 V Great Britain grounded fused 8918 GB 100 240 V Australia NZ 8918 AT 1 2 Soft Sided Carrying Case Figure 1 1 Model 4046 Components Calibrator 3 Mounting Lugs 4 Tubing a pb GN a 5 6 AA size Batteries 6 Adapter 7 User s Guide 8 High Efficiency Inlet Filter Figure 1 2 Model 4146 Components Calibrator Soft Sided Carrying Case Mounting Lugs Latex Tubing 6 AA size Batteries 6 High Efficiency Inlet Filter 7 User s Guide 8 Adapters 9 Dampening Module Chapter 1 Chapter 2 Setting Up Assembly Instructions for the Calibrators The calibrator can be quickly attached to the battery pack by screwing the mounting lugs into the threaded holes in the base of the calibrator The lugs should then be placed over the receiving holes in the top of the battery pack and snapped gently into place Insert the connector from the battery pack to the receptacle in the back of the calibrator The calibrator is now ready to use Figure 2 1 Calibrator Assembly Supplying Power The calibrator can be powered in one of two ways through the power jack using the supplied 4199 battery pack or using an optional AC adapter The DC power input connector is shown below along with the
6. PRIMARY CALIBRATORS MODEL 4046 4146 USER S GUIDE P N 1980512 REVISION E OCTOBER 2013 START SEEING THE BENEFI las EP REGISTERING TODAY Thank you for your TSI instrument purchase Occasionally TSI releases information on software updates product enhancements and new products By registering your instrument TSI will be able to send this important information to you http register tsi com As part of the registration process you will be asked for your comments on TSI products and services TSI s customer feedback program gives customers like you a way to tell us how we are doing O UNDERSTANDING ACCELERATED TSI Incorporated Visit our website www tsi com for more information USA Tel 1 800 874 2811 India Tel 91 80 67877200 UK Tel 44 149 4 459200 China Tel 86 10 8219 7688 France Tel 33 491118764 Singapore Tel 65 6595 6388 Germany Tel 49 241 523030 2013 TSI Incorporated Printed in U S A Copyright TSI Incorporated 2004 2013 All rights reserved Address TSI Incorporated 500 Cardigan Road Shoreview MN 55126 USA TSI calibrators employ a heated platinum sensor They should not be used in the presence of flammable or explosive gases or mixtures TSI calibrators are not medical devices under FDA 510k and in no situation should they be utilized for human respiration measurements LIMITATION OF WARRANTY AND LIABILITY effective June 2011 For country spe
7. cific terms and conditions outside of the USA please visit www tsi com Seller warrants the goods sold hereunder under normal use and service as described in the operator s manual shall be free from defects in workmanship and material for 24 months or if less the length of time specified in the operator s manual from the date of shipment to the customer This warranty period is inclusive of any statutory warranty This limited warranty is subject to the following exclusions and exceptions a Hot wire or hot film sensors used with research anemometers and certain other components when indicated in specifications are warranted for 90 days from the date of shipment Pumps are warranted for hours of operation as set forth in product or operator s manuals Parts repaired or replaced as a result of repair services are warranted to be free from defects in workmanship and material under normal use for 90 days from the date of shipment Seller does not provide any warranty on finished goods manufactured by others or on any fuses batteries or other consumable materials Only the original manufacturer s warranty applies Unless specifically authorized in a separate writing by Seller Seller makes no warranty with respect to and shall have no liability in connection with goods which are incorporated into other products or equipment or which are modified by any person other than Seller oo 2 o The foregoing is IN LIEU OF all other warra
8. e Counters OPCs e Condensation Particle Counters CPCs e Photo lonization Detectors PIDs e Flame lonization Detectors FIDs 8 Chapter 2 Note The TSI DustTRAK and SiDEPAK Sampling Pumps contain internal dampening chambers The Dampening Module is not needed to measure flow rates with these instruments The TSI SinEPax AM510 Personal Aerosol Monitor does not contain an internal dampening chamber TSI recommends using the Dampening Module when using the Model 4146 Calibrator to measure AM510 flow rates Setting Up This page intentionally left blank Chapter 2 Chapter 4 Maintenance Flow Sensor Periodically inspect the flow sensor by looking into the outlet of the calibrator Remove dust particles and fibers from the sensor with clean dry compressed air The flow sensor may break if touched Never run liquids through the calibrator and never touch the sensor with a brush Dust or other deposits on the flow sensor will degrade the 4046 4146 s flow accuracy The calibrator must be switched off for cleaning Only use clean dry compressed air when attempting to remove contamination from the sensor Re certification To maintain a high degree of confidence in the measurements made by the Model 4046 4146 TSI recommends that the instrument be returned to TSI every 12 months for re certification The calibrator will be recalibrated and returned with a certificate of calibration to US National Institut
9. e of Standards Technology NIST traceable standards Cases If the instrument case or storage case needs cleaning wipe it off with a soft cloth dipped in isopropyl alcohol or mild detergent Never submerge the calibrator or allow liquids to enter the flow tube Storage When storing the calibrator always cover the ends of flow tubes with the caps provided to prevent dust or other foreign matter from entering the tube This page intentionally left blank Chapter 4 Chapter 5 Troubleshooting Table 3 lists the symptoms possible causes and recommended solutions for common problems encountered with Model 4046 4146 Calibrators If the symptom is not listed or if none of the solutions solves the problem please contact TSI Technical Support at 1 800 874 2811 or 651 490 2811 Table 3 Troubleshooting Symptom Possible Causes Corrective Action No display Unit not switched on Switch on the unit No power to Check power instrument connection Temperature Temperature sensoris The temperature value reads high at being heated from the will track the actual air low or zero flow sensor temperature once the flows flow rate through the calibrator exceeds 1 Std L min Flow readings The flow is fluctuating Improve inlet fluctuate greatly conditions or increase display averaging time Display shows The sensor may be Return calibrator to flows over range damaged or broken TSI for service with no flow passi
10. ime There is no primary standard for air flow rate since the measurement is dependent on all three primary standards of mass length and time Hence all air flow calibrators are secondary standards in the strictest sense of the definition In the industrial hygiene arena the terms primary and secondary have become generally accepted terms over time and should not be confused with actual primary and secondary standards as prescribed by the National Institute of Science and Technology NIST Some manufacturers of calibrators by their own definition refer to their products as primary standards referring to the manner they are applied and generally accepted for use in the industry Other manufacturers adequately refer to their instruments as calibrators There are no industry standards established for such designations based on the measurement technology being employed What is important is accuracy repeatability and regular documented verification to a higher accepted standard such as NIST Both primary and secondary calibrators must be verified on a regular basis and that testing recorded Some guidelines recommend annual calibration to a higher accepted standard while others suggest as frequently as every 30 days or before a significant survey It is important to have a certificate of traceability to NIST or other industry recognized testing organization indicating that an instrument has been calibrated against some
11. in x 2 1 in 18 3 cm x 6 4 cm x 5 3 cm Model 4146 5in x 2in x 1 25 in 12 7 cm x 5cm x 3 2 cm Tube Adapters Model 4046 3 0 Inlet amp Outlet Model 4046 22 mm male ISO taper Model4146 0 375 inch O D straight Weight Model 4046 1 2 Ibs 0 5 kg Model 4146 1 7 Ibs 0 8 kg Flow Body Material Polycarbonate Battery pack 7 5 VDC 1 5 V 300 mA maximum Notes 1 Accuracy stated at standard conditions of 21 1 C and 101 3 kPa 2 Accuracy stated with air temperature and flow body temperature within 10 C of one another 3 Accuracy stated measuring dry air less than 10 R H Includes 40 576 of reading repeatability 5 Volumetric flow rate is calculated from the mass flow measurement Add an additional 0 25 of reading to the flow accuracy to account for the uncertainty in measuring air temperature and pressure gt Specifications subject to change without notice 15 L 4 04 1 maw L 180 8 mm 4 Figure A 1 Model 4046 Dimensions o Hr CI a 10 LL 4044 mm H 126 9 mm Figure A 2 Model 4146 Dimensions 16 Appendix A Appendix B Primary vs Secondary Standard By definition a primary standard is one whose value is without relation or comparison to any other reference There are three primary standards mass length and t
12. mother device verified by NIST This page intentionally left blank Appendix B O UNDERSTANDING ACCELERATED TSI Incorporated Visit our website www tsi com for more information USA Tel 1 800 874 2811 India Tel 91 80 67877200 UK Tel 44 149 4 459200 China Tel 86 10 8219 7688 France Tel 33 4 91 11 87 64 Singapore Tel 65 6595 6388 Germany Tel 49 241 523030 P N 1980512 Rev E 2013 TSI Incorporated Printed in U S A
13. ng through flow tube Technical Contacts If you have any difficulty installing the Model 4046 4146 or if you have technical or application questions about this instrument contact an applications engineer at TSI Incorporated 651 490 2811 or contact answers tsi com If the Model 4046 4146 fails or if you are returning it for service visit our website at http rma tsi com or contact TSI at TSI Incorporated 500 Cardigan Road Shoreview MN 55126 USA Phone 1 800 874 2811 USA or 1 651 490 2811 E mail answers tsi com TSI GmbH Neukollner Strasse 4 52068 Aachen GERMANY Telephone 49 241 52303 0 Fax 49 241 52303 49 E mail tsigmbh tsi com Web www tsiinc de TSI Instruments Ltd Stirling Road Cressex Business Park High Wycombe Bucks HP12 3ST UNITED KINGDOM Telephone 444 0 149 4 459200 Fax 44 0 149 4 459700 E mail tsiuk tsi com Web www tsiinc co uk Chapter 5 Appendix A Specifications Specifications Flow Measurement Measurement Range Model 4046 2 5 to 300 L min Model 4146 0 01 to 20 L min Accuracy Model 4046 2 of reading or 0 05 std L min Air whichever is greater Model 4146 2 of reading or 0 005 std L min whichever is greater at standard conditions 21 1 C and 101 3 kPa See notes 1 through 5 below Instrument Temp Range Operation Ambient 0 to 50 C Storage Ambient 20 to 60 C Physical Dimensions External Dimensions Model 4046 7 2 in x 2 5
14. nties and is subject to the LIMITATIONS stated herein NO OTHER EXPRESS OR IMPLIED VVARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE WITH RESPECT TO SELLER S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS BUYER S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER S OPTION REPLACEMENT OF THE GOODS WITH NON INFRINGING GOODS TO THE EXTENT PERMITTED BY LAW THE EXCLUSIVE REMEDY OF THE USER OR BUYER AND THE LIMIT OF SELLER S LIABILITY FOR ANY AND ALL LOSSES INJURIES OR DAMAGES CONCERNING THE GOODS INCLUDING CLAIMS BASED ON CONTRACT NEGLIGENCE TORT STRICT LIABILITY OR OTHERWISE SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE OR AT THE OPTION OF SELLER THE REPAIR OR REPLACEMENT OF THE GOODS IN THE CASE OF SOFTWARE SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS OR ANY SPECIAL CONSEQUENTIAL OR INCIDENTAL DAMAGES SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION DISMANTLING OR REINSTALLATION COSTS OR CHARGES No Action regardless of form may be brought against Seller more than 12 months after a cause of action has accrued The goods returned under warranty to Seller s factory shall be at B
15. uyer s risk of loss and will be returned if at all at Seller s risk of loss Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY which contains the complete and exclusive limited warranty of Seller This LIMITATION OF WARRANTY AND LIABILITY may not be amended modified or its terms waived except by writing signed by an Officer of Seller Service Policy Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers our service policy is designed to give prompt attention to any problems If any malfunction is discovered please contact your nearest sales office or representative or call TSI s Customer Service department at 800 874 2811 USA or 001 651 490 2811 International or visit www tsi com Trademarks TSI and TSI logo are registered trademarks of TSI Incorporated CONTENTS CHAPTERS 1 Unpacking and Parts Identification 2 ee22eeeee2ee 1 2 Seting Upaan aaa aeaa DA LAAN a 3 Assembly Instructions for the Calibrators e 3 Supplying POWE endin eraai aaa aaia aa anaa aiaiai iaeiaiai 4 Connecting Filter and Flow Tubes 20nn anawaanawawanawwananana 4 Bo Op ratio maa NAN ENNA Oaa a aa aao a mili 7 OQVEIVIE Mies asistencia ei dives 7 ON OFF SWICK naam cattails 7 Flow Rate Measurement nananana wa wnna ww anawannanwwana no wnnsananana 7 Setting Flow Rates and Calibrating Instruments