Troubleshooting - medistore.com.au
Contents
1. Alarm Mute button A too high 2 Disconnect the air hoses to reduce pressure reconnect when pressure has decreased once 3 Check for twists in the air hoses between Mattress Overlay and Control Unit 4 Switch on Power Audible alarm only Power 1 Press the Alarm Mute button to silence the audible alarm failure 2 Brlolelolclele Check the power cable is firmly plugged into the mains outlet and the Control Unit and check the mains power is switched on Check the mains plug fuse 3 AMP then check both Control Unit fuses 1 AMP fuses can be released using a screwdriver to push and turn Lights audible alarm Boo ele o No alternation wW N Reset the alarm turn off Power and press the Alarm Mute button Disconnect the air hoses to reduce pressure Reconnect when pressure has decreased Patient is sinking on the mattress Pressure may be set incorrectly 2 Increase the pressure setting by pressing up the Pressure arrow To check effective system performance conduct a bottoming out test NOTE Ifthe problem is not resolved please contact Medistore Australia on 1300 882 194 WARNING Donottryto open the Control Unit Opening the unit could cause personal injury or equipment damage freecall 1300 882 194 op medistore supporting a better life freefax 1300 882 197 www medistore com au Mute Alarm will continue to sound until fault is corrected Refer to trouble shooting guide to di
2. Pro Car Troubleshooting ALARM FAULT CAUSE SOLUTION Control Unit does Power or 1 Check the Control Unit is connected to mains power outlet with the correct voltage not operate fuse issue 2 Check the Control Unit is switched on switch off and unplug the unit before restarting No display lights 3 Check the mains plug fuse 3 AMP then check both Control Unit fuses 1 AMP illuminate fuses can be released using a screwdriver to push and turn Lights audible alarm Initial 1 Reset the alarm turn off Power and press the Alarm Mute button fl failure 2 Check all air hoses along the inside of the mattress each should be firmly connected Brslelclelele Check each air cell is securely attached to its connecting air pipe 3 Check all cells pipes and hoses for any air leakage 4 Switch on Power Lights audible alarm Pressure 1 Reset the alarm turn off Power and press the Alarm Mute button A too low 2 Check the CPR tagis intact ensuring both sealing connectors are firmly fitted to the internal Bilelclelle Mattress Overlay air hoses 3 Check all air hoses along the inside of the mattress each should be firmly connected Check each air cell is securely attached to its connecting air pipe 4 Check all cells pipes and hoses for any air leakage 5 Check that the air filter cover is correctly secured and the air filter is clean 6 Switch on Power Lights audible alarm Pressure 1 Reset the alarm turn off Power and press the
3. agnose fault Mode selections Press to alternate between static or alternating mode Automatically returns to alternating after 30 minutes on static Pro Car Interface interpretation Lockout Hold for five 5 seconds until beep is heard to lock or unlock es annscuenone A SJaaaastii Power button Hold to turn on or off Always turn off before unplugging unit Max inflate Push for auto set up rapid inflate to maximum pressure Comfort controls Increase or decrease based on patient weight Refer to comfort guide for correct settings BA medistore supporting a better life freecall 1300 882 194 freefax 1300 882 197 www medistore com au Feb13
Download Pdf Manuals
Related Search
Related Contents
Star Micronics 347F Printer User Manual StarTech.com 10 ft Computer Power Cord - C19 to C20 DP-50 digitalpiano シリンダーキット(146cc SCUT) 取扱説明書 Sony WX-GT77UI Operating Instructions Classic Accessories 72982 Instructions / Assembly ZR-12-27 - OPUS 4 Release note MAK 3003 Sampling Stand Alone Menus e configuração Copyright © All rights reserved.
Failed to retrieve file