        Operating instructions Groundwater datalogger OTT CTD
         Contents
1.            The combination of  Only observer mode  with a password protects the operating    program against unauthorized input of operating parameters     The  Advanced operation      option displays an additional button in the window    for setting the operating parameters     25    Fig  15  Display input window  OTT CTD    26    operating program     observer      The window displays the current instanta   neous value for all measurements   Furthermore  the window displays the  window sensor number name  the  current battery voltage  the amount   of power used from the batteries up   to now  and the date and time     Call the observer function as follows    Start the OTT CTD operating program    Set up the PC OTT CTD communication link  see Chapter 7 2     Depending on the option set  In the start window of the OTT CTD operating  program  either select  View Instantaneous Values  or  Setup device  and in  the subsequent window select the  View Instantaneous Values  function in the   OTT CTD  menu  gt  the OTT CTD starts an instantaneous value measurement  and the  Observer  window opens            fi Orpheus Mini   OTT CTD operating program   Observer  x     CTD 1   0000254892 45V 0 0204h 11 11 2009 16 40 55       Water level   0001 1 25 1 51 m    Observer registration with value change    Temperature   0002 20 34   C  Specific Conductivity   0004 26 72 mS cm  Salinity   0005 18 21 PSU  TDS   0006 17 10 mg l  Supply voltage   0003 4 5 Da    Store Refresh   Exit      I Auto
2.    ZOTT    Operating instructions  Groundwater datalogger  OTT CTD                            These operating instructions  version  01 1009   cover the OTT CTD software  versions    OTT CTD firmware from V 1 00 0  OTT CTD operating program from V 1 50 0    The OTT CTD firmware version can be found in the  Advanced operation  mode    OTT CTD  window of the operating program  The version of the operating  program can be found via the  Info  function in the  Help  menu     We reserve the right to make technical changes and improvements without notice     Table of contents    10    12    13    14    15    16    17    18    19    20    21    Scope of supply   Order numbers   Safety information   Introduction   Installing  checking  and exchanging batteries    Installing the OTT CTD    6 1 Installing in 1  observation wells  6 2 Installing in 2   3   4   5  or 6  observation wells   top cap with adapter plate recess  6 3 Installing in observation wells beginning at 2  diameter   top cap without adapter plate recess  6 4 Installing in observation wells beginning at 2  diameter without a top cap    Setting OTT CTD operating parameters  7 1 Installing the OTT CTD operating program    7 2 Establishing a communication link from PC to OTT CTD   7 3 Setting OTT CTD operating parameters   7 4 Saving loading OTT CTD configuration   7 5 Importing exporting OTT CTD configuration   Determining and displaying instantaneous values  observer function   Reading out data   Exporting data   Display
3.   Feed the communication unit into the observation well until the O ring sits on  the observation well  see Figure 6     Raise and lower the pressure probe in quick succession approx  30 cm     this  removes any air bubbles in the conductivity sensor    Set operating parameters  see Chapter 7     Close the observation well cover carefully  Caution  Do not damage the  infra red interface  maintain correct spacing         Fig  6  Installing the OTT CTD    in 1  observation well     Observation well cover   example          Infrared interface    O ring              1  Observation well                   Fig  7  Installing the OTT CTD in 2    3   4   5  or 6  observation wells     Example  2  observation well    6 2 Installing in 2   3   4   5  or 6  observation wells     top cap with adapter plate recess    Required accessories  OTT top cap with appropriate sized recess and the  matching adapter plate     Install the OTT CTD as follows    Preliminary work  If not already done  install batteries  see Chapter 5    Open the top cap cover    Insert correctly sized adapter plate into top cap    Pass pressure probe through the hole in the adapter plate    Slowly and carefully lower the pressure probe into the observation well  with the pressure probe cable  see Fig  8     Raise and lower the pressure probe in quick succession approx  30 cm     this  removes any air bubbles in the conductivity sensor    Pass the communication unit through the hole in the adapter plate until the  O ring 
4.   based on the specific conductivity     The pressure probe uses the hydrostatic pressure of the water column above a  relative pressure measuring cell to determine the water level     pressure compen   sation capillary in the pressure probe cable gives the measuring cell the current  ambient air pressure as a reference  Erroneous measurement results due to atmos   pheric air pressure fluctuations are thus eliminated  The OTT CTD measures the  specific electrical conductivity using a 4 electrode conductivity sensor with inte   grated temperature sensor  The measurement electrodes are made of graphite     The temperature compensation process for the conductivity measurement and the  reference temperature used can be chosen as well as the calculation method for  the salinity     The OTT CTD is available with five water level measuring ranges      m water column  0     0 4 bar      10 m water column  0     1 bar      20 m water column  0     2 bar     DOGG      40 m water column  0     4 bar   0    100 m water column  0     10 bar     With the help of a reference value that is input during startup  the OTT CTD s  standard setting provides measurement results in the form of depth values  Alter   natively  measured values can be levels or pressures  The measurement intervals   sample intervals  can be preselected as necessary     The operating parameters are adjusted with the  OTT CTD Operating program   PC software  This software allows the system to be conveniently and flexibly 
5.  Deutsch    directory  e g  double  click on the file symbol      the Setup Assistant opens and guides you through  the installation    Follow the installation instructions on the screen     1 Alternatively    English  or   Frangais  or  VEspafiol     7 2 Establishing a communication link from PC to OTT CTD    In the following chapters  establishing a communication connection between the  OTT CTD and a PC is a pre requisite for the subsequent steps  The following  description illustrates the various methods of setting up this communication link     The communication between the OTT CTD and a PC is established without contact  via invisible infrared light  IrDA interface      Required accessories     OTT Duolink reading head   or  OTT IrDA Link USB reading head    1  with PCB version  b   see label on the connection line    Note    Together with a modem and the OTT ITC intelligent top cap  it is possible to  establish a remote communication connection  See online help     How to establish a communication link    Start the OTT CTD operating program    Change the language for the OTT CTD operating program as required    Press function key  F3   multiple times  until the required language appears   In the start window  press the  Setup device  button     the operating program  displays the  Basic operation  window  No operating parameters for the   OTT CTD are visible yet         Setting up a communication link with the OTT Duolink reading head     Connect the OTT Duolink reading
6.  L   PUR  ABS  PC  stainless steel 1 4539  904 L   1 4462  UNS  31803     IP 67  submersion depth max  2 m  submersion duration max  24 h   IP 68    complies with EN 61000 4 2  4 kV contact discharge   complies with EN 61000 4 3  10 V m    complies with EN 61000 4 4  2 kV    complies with EN 61000 4 5  4 kV    complies with EN 61000 4 6  10 V    complies with EN 55022 Class B  30     1000 MHz     41    Appendix A   Declaration of conformity for OTT CTD          OTT       Konformit  tserkl  rung  Declaration of Conformity  Declaration de Conformit      Wir  We  Nous OTT Messtechnik GmbH  amp  Co  KG  Anschrift  Address  Adresse Ludwigstra  e 16  D 87437 Kempten    erkl  ren  da   das Produkt  declare  that the product  declarons  que le produit  Bezeichnung  Name  Nom OTT CTD    Artikel  Nr   Article No   No  d  Article 55 445 001 9 0    mit den Anforderungen der Normen fulfills the requirements of the standard  satisfait aux exigences  des normes    EG  2004 108 EG     national  international    EN 61000 6 4 IEC 61000 6 4  St  raussendung  emission    mission   Klasse  class  classe B class  classe B  St  rfestigkeit  noise immunity  immunit     EN 61000 6 2 IEC 61000 6 2   EN 61000 4 2  4 kV 8 kV  IEC 61000 4 2  4 kV 8 kV    EN 61000 4 3  10 V m  IEC 61000 4 3  10 V m    EN 61000 4 4  1 kV 2 kV  IEC 61000 4 4  1 kV 2 kV    EN 61000 4 6  10 V  IEC 61000 4 6  10 V     und den hinterlegten Pr  fberichten   bereinstimmt und damit den Bestimmungen entspricht   and the taken test 
7.  OTT CTD as follows  example     Preliminary Work  If not already done  install batteries  see Chapter 5    Make a hole     6 5 mm  laterally at the upper end of the observation well   Attach the suspension bracket with an M6 hex bolt and M   hex nut  both  stainless steel  to the observation well  see Fig  11     Pass the pressure probe through the holes in the suspension bracket   Slowly and carefully lower the pressure probe into the observation well  with the pressure probe cable    Raise and lower the pressure probe in quick succession approx  30 cm     this  removes any air bubbles in the conductivity sensor    Pass the communication unit through the holes in the suspension bracket until  the O ring sits on the suspension bracket  see Fig  10     Determine the current depth with a contact gauge and make a note of it    Set operating parameters  see Chapter 7         Hex nut A     Plate se          Hexagon bolt    Suspension    bracket        Observation    well                   17    18    7 Setting OTT CTD operating parameters    To set the OTT CTD operating parameters you need the PC software  OTT CTD  Operating program   WBSPLO exe   This software is contained on the  OTT CTD  Software  CD ROM  accessory      Hardware and software requirements  see CD insert     7 1 Installing the OTT CTD operating program    Install the OTT CTD operating program as follows    Insert the OTT CTD Software CD ROM into the drive of the PC    Start the  setup exe  file in the   Software
8.  OTT CTD operating program    Set up the PC OTT CTD communication link  see Chapter 7 2     Press the  Download Data  button in the start window     the operating pro   gram shows the available sensors and the possible read period  If these para   meters are not visible  press the  Connect  button    Select the required sensors or  All sensors     Select the required read period or  All     Click the  Read out  button     the operating program copies the measured  values from the OTT CTD to the PC  The data are then available in the program  directory of the operating program    If necessary  Display measurements via the  View export data  function  and or export to a software application    Click on the  Back  button     Fig  16  Read out data  ER OTT Orpheus Mini   OTT CTD operating program   Download data    op x   File Help    0000254892       All sensors  11 01 2009 1171272009  11711720098   Tal 1171272009 v    Read I  View   Export data    OTT Orpheus Mini   OTT CTD     IrDA   OTT DuoLink    Connect Back  comi     5 4       Alternative possibilities for readouts     With the OTT CTD operating program and an OTT ITC   In conjunction with a modem and the OTT ITC intelligent top cap  it is also  possible to establish a remote communication link  See online help    With the PC application  OTT Hydras 3  Basic      Select the required station in the tree display by double clicking  make the  required settings in the subsequent window and click on the  Start  button   For  fu
9.  and water   Clean graphite electrodes and temperature sensors for the conductivity sensor  with soapy water and cotton buds  Lime scale deposits can be removed using  a common household scale remover  Make sure to follow the use and safety  instructions of the scale remover   Rinse the pressure probe thoroughly with clear water   Reattach the black protective cap   Recommendation  Calibrate the conductivity sensor  see Chapter 16 4    Reinstall OTT CTD  see Chapter 6    Determine current reference value with a contact gauge and enter it   see Chapter 7 3 or 8    Close the top cap observation well cover           Pressure probe with Temperature sensor  conductivity sensor  and datalogger         Black  protective cap       Graphite Conductivity  electrodes sensor    Pressure Sensor          Fig  23  Replacing the desiccant capsules    16 2 Replacing the desiccant capsules    Recommended interval  Depending on the level of air humidity at the station  every 12 to 24 months and when replacing the batteries     How to replace the desiccant capsules  Open the top cap observation well cover   Pull the communication unit approximately 80 cm out of the observation well  and hold  a second person would be useful    Slide the pipe casing of the communication unit approximately 30 cm in the  direction of the pressure probe cable   The rubber stop located on the pressure  probe cable  see Fig  1  prevents the pipe casing from falling    Remove used desiccant capsules   Install 2 new des
10.  head to a serial PC interface  COM1  etc     Place the OTT Duolink reading head on the OTT CTD infrared interface    see Fig  12     In the OTT CTD operating program  select the communication path  IrDA    OTT Duolink  and the serial interface used  COMI  etc         Setting up a communication link with the OTT IrDA Link USB reading head     Connect the OTT IrDA Link USB reading head to a USB interface on the PC    USB interface drivers must be installed  see separate installation instructions    Place the OTT IrDA Link USB reading head on the OTT CTD infrared interface   see Fig  12     Select the communication path  IrDA   OTT IrDA Link  in the operating  program     Notes    The integrated infrared interface of a PC  standard for many notebooks  can  only be used with the Windows 95 and Windows NT operating systems    The OTT CTD infrared interface has a radiation angle of approx   15      Caution  OTT CTD connected to OTT ITC  Establishing a communication  connection can take up to a minute   This is always the case if the OTT CTD  attempts to communicate with the OTT ITC  e g  SMS message  and the OTT ITC  is open at the time         Fig  12  Establishing a communication  link PC OTT CTD                                         19    7 3 Setting OTT CTD operating parameters    How to set the OTT CTD operating parameters    Click the  Read  button  alternatively  Menu  OTT CTD   function    Read       the operating program reads the current operating parameters of  the OT
11.  tailored to a wide range of measurement requirements  The software can be set  to provide a basic or an advanced operator interface  The basic operator inter   face allows all settings to be adjusted within a single program window  In the  advanced operator interface  the sample interval can be controlled with limit  events  for example  The software also supports the execution of pump tests     The stored measured values are made available through an infrared interface   IrDA  for wireless readout by a PC with OTT CTD Operating program or OTT  Hydras 3 or by a PDA with OTT Hydras 3 Pocket     Together with an OTT ITC intelligent top cap  accessory   remote data transfer  and remote parameter input is possible via the GSM mobile telephone network   GSM   global system for mobile communications   The remote data transfer can  be optionally carried out by SMS text messages or using the packet oriented  mobile radio transmission service GPRS  general packet radio service      The OTT CTD is installed simply by hanging it in observation wells of 1  in diame   ter and larger  Various adapters suspension brackets are available as accessories  for this purpose  Three 1 5 V lithium batteries  type AA  provide an operating life  of over five years  sample interval  1 hour  system length  50 m   Alternatively   alkaline batteries with a limited working life can be used     The communication unit of the OTT CTD can withstand temporary flooding  for  details  see chapter 21   Technical 
12.  to archive the OTT CTD configuration  on a PC  For example  you can provide multiple OTT CTDs with the same confi   guration     To store a configuration  Select the  Save configuration  function in the  File  menu  or use the button       the operating program stores the configuration under the name and number  of the OTT CTD  When changes are made to a configuration that has already  been saved  confirm the window that appears   Warning  this configuration  name already exists  Overwrite   with  Yes   otherwise save the configuration  under a different station number      To load a configuration  proceed as follows  Select the  Load configuration  function in the  File  menu  or use the button    For an already opened configuration  confirm  Ignore changes   message with   Yes   if necessary  save configuration previously    In the  Saved OTT CTD Configurations  window  select the configuration by  double clicking it   the operating program loads the configuration     7 5 Importing exporting an OTT CTD configuration    To transfer an OTT CTD configuration  via diskette or E mail for example  the func   tions Import Export are available  Likewise  a configuration can be sent via Export  as an XML file to the OTT Hydras 3 user software  In the process  the OTT Hydras 3  user software completely applies all of the measurement stations   sensor configu   rations to an OTT Hydras 3 operating range     Available export import formats  Export   BIN  file   Export to a text file  
13.  to measure groundwater levels  the water temperature  and the specific conductivity of the groundwater  Only use the OTT CTD as  described in these operating instructions    For further information      see Chapter 4   Introduction      Note all the detailed safety information given within the individual work steps   All safety information in these operating instructions are identified with the  warning symbol shown here     Ensure the electrical  mechanical  and climatic specifications listed in the  technical data are adhered to   For further information     see Chapter 11   Technical data      Handle the pressure probe cable carefully  Do not kink the cable or pull it  across sharp edges     Do not make any changes or retrofits to the OTT CTD  If changes or retrofits  are made  all guarantee claims are voided     Have a faulty OTT CTD inspected and repaired by our repair center  On no  account carry out repairs yourself   For further information     see Chapter 19   Repair      Dispose of the OTT CTD properly after taking out of service  On no account put  the OTT CTD into the normal household waste   For further information     see Chapter 20   Notes about the disposal of old units      4 Introduction    The OTT CTD groundwater datalogger provides precise measurement and record   ing of groundwater levels and temperatures  as well as the specific electrical  conductivity of the groundwater  The OTT CTD also calculates the salinity and a  TDS value  Total Dissolved Solids
14.  value  with one minute added to the stored time  The date and time are set using  the OTT CTD operating program     Set the date and time as follows    Start the OTT CTD operating program    Set up the PC OTT CTD communication link  see Chapter 7 2     Click on the  Setup device  button in the start window    Select the  Date time  function in the  OTT CTD  menu     the operating  program reads the date and time out from the OTT CTD and opens the   OTT CTD   date time  window        gjort Orpheus Mini Z OTT CTD   Date   time  x          OTT CTD 12 11 2009 15 49 02 Refresh        PC 12 11 2009 15 49 04  Set date time     E 11 2009     15 5002        Exit            If necessary  click on the  Refresh  button   the OTT CTD reads out the date  and time again    If necessary  set the values required in the two input fields    Click on the  Set date time  button     the operating program sets the OTT CTD  date and time to the PC time date or the values set     Caution  If the PC is on summer time mode  ID  PC  DST    the operating pro   gram automatically uses the standard time without summer correction  winter  time   To receive continuous time series  it is a good idea not to use summer time  in the OTT CTD    In connection with a GPRS remote data transfer and the  Time synchronization   function  it is necessary to refer the time to UTC GMT and to set the time zone of  the station in the  Advanced operation     OTT CTD  window   see online help     Click on the  Exit  button  
15. 2 within 10 minutes  Ensure that the polarity is correct    Slide the pipe casing of the communication unit back on until it stops    Slowly and carefully place the communication unit back into the observation well   Close the top cap observation well cover     Notes    The OTT CTD stores the measured values in a non volatile memory  This pre   vents any data loss when replacing the batteries  This also applies to storage  over a long time period with the batteries removed    If it takes longer than approx  10 minutes to replace dead batteries  the time   and possibly the date  will have to be re entered  see Chapter 13   Date and  time settings    Furthermore  it takes   dependent on how much data is stored      up to 7 minutes until the OTT CTD begins measurements again    When replacing the batteries  we recommend changing the desiccant capsules  at the same time  see Chapter 16 2   Replacing the desiccant capsules       Fig  3  Installation accessory set     Part 1     10    1   hole for OTT CTD  2   hole for contact gauge  3   recess for screws on the top cap    6 Installing the OTT CTD    System length up to 100 meters   The installation of the OTT CTD is carried out by suspending it in the observation well   In the process  the necessary accessories and type of installation varies depending  on the diameter of the observation well and the design of the top cap     1  observation wells see 6 1  2   3   4   5  or 6  observation wells  see 6 2  OTT top caps with recess fo
16. Export for the OTT Hydras 3 user software    Import   BIN  file     The operating program stores all the necessary data of a configuration depending  on the export type in its own    BIN      TXT  or    XML  file     Export a configuration as follows     BIN  file    This export type is required to transfer an OTT CTD configuration as a    BIN   file  The configuration can be read back into the OTT CTD operating program via  the  Import configuration  function     Select the  Export configuration  function in the  File  menu    In the  OTT CTD export configuration  window  edit the file name if necessary   select the memory location and click on  Save      the operating program  stores the configuration ina    BIN  file     Note    The file name of the    BIN  file is arbitrary  The operating program suggests a  combination of station number and station name     23    24    To export a configuration  text file    This export type is required to transfer an OTT CTD configuration for documentation  purposes as a    TXT  file  This    TXT  file can be opened with any text editor or  text processing program     Select the  Export text file  function in the  File  menu    In the  OTT CTD export configuration  window  edit the file name if necessary   select the memory location and click on Save     the operating program stores  the configuration in a    TXT  file     Note    The file name of the    TXT  file is arbitrary  The operating program suggests a  combination of station nu
17. OTT CTD  operating program is located    Set up the PC OTT CTD communication link  see Chapter 7 2     Select the  Firmware update  function in the  OTT CTD  menu    Confirm message with  Yes      the operating program copies the new  firmware to the OTT CTD  Afterwards the OTT CTD determines and saves the  water level  the water temperature and the specific conductivity again in accor   dance with the sample interval set    Caution  During the update  avoid breaking the communication link  e g by  accidentally removing the OTT Duolink from the infrared interface of the  OTT CTD   If the communication link is broken  the firmware will no longer run   In the same way  no other programs should be started or files opened during  the copying process     Notes    If there are multiple   bin  files in the directory  you will have to manually select  the required file   The measurements saved in the OTT CTD are not lost after an update     33    34    Fig  22  Cleaning the pressure probe     16 Maintenance work    16 1 Cleaning the pressure probe    Recommended interval  every 12 months  with difficult local measuring conditions  heavy deposits    as required every 4 to 6 months    Measurements that are inaccurate or not plausible  indicate a soiled pressure sensor      How to clean the pressure probe  Open the top cap observation well cover   Completely remove the OTT CTD from the observation well   Remove the black protective cap   Clean the pressure sensor carefully using a brush
18. T CTD    Adjusting operating parameters  see overview on page 21     If necessary     Set up cyclical data transfer with SMS text message in combi   nation with an OTT ITC intelligent top cap    or    Set up cyclical data transfer  using GPRS in combination with an OTT ITC intelligent top cap      see below   Set time  adjust the proposed PC date time or individually set date time using   Set date time   see also Chapter 13     Caution  The operating program automatically corrects any summer time PC  adjustments    Click the  Program  button  Warning  Reset the OTT CTD and delete data  memory additionally   Confirm with  Yes   recommended for initial installation  and reinstallations     Caution  All measured values collected until now will be permanently lost   Check the level of the adjusted measured value  pressure sensor   See Chapter 8   Remove OTT Duolink    Close the top cap observation well cover     If necessary  Set up cyclical data transfer with SMS text message  in combination with an OTT ITC intelligent top cap      Activate check box  Modem ITC connected     Activate check box  SMS data transmission active     Click on the  ITC settings  button     the operating program starts an Assistant  for setting all the necessary operating parameters    Enter the receiver phone no  for the data SMS   phone number of a large  account or for a PC with GSM modem     Click on the  Next  button    Enter SIM PIN  4 digit  of the SIM card    Click on the  Next  button    Set or 
19. TD  operating program password     Click on the  Program  button    Warning   Reset OTT CTD and delete data memory additionally   Confirm  with  No      the OTT CTD is now protected against unauthorized operation             Communication interface  General   Modem   ITC    Password OTT CTD             Caution  If the password is lost  you can no longer configure or parameterize  the OTT CTD  If this occurs  contact the OTT HydroService     How to release a password protected OTT CTD    Start the OTT CTD operating program    Set up the PC OTT CTD communication link  see Chapter 7 2     Click on the  Setup device  button in the start window    Select the  Enter password  function in the  OTT CTD  menu    Enter password    Click on the  OK  button    Confirm message  Password accepted  OTT CTD unlocked  with  OK    Click on the  Read  button     the operating program reads in the current  OTT CTD operating parameters        OTT CTD x   Enter password  0  9 4  Z        Abbrechen               31    32    Fig  21  OTT CTD date and time settings     13 Date and time settings    The internal clock of the OTT CTD is a high accuracy  realtime clock  It runs as  soon as batteries are installed in the OTT CTD  After the batteries are removed   the clock will continue to run for approximately 10 more minutes  For longer periods  of power interruption  the OTT CTD loses the date and time  When batteries are  re installed  the OTT CTD assumes the date and time of the last stored measured 
20. The  OTT CTD   date time  window closes     14 Deleting data memory    Caution  The stored measured values in the OTT CTD are permanently lost when  deleting the data memory  If necessary read out measured values before deleting     Delete the data memory as follows    Start the OTT CTD operating program    Set up the PC OTT CTD communication link  see Chapter 7 2     Click on the  Setup device  button in the start window    Select the  Delete data memory  function in the  OTT CTD  menu   Acknowledge the warning  Are you sure you want to delete data memory    with  Yes      the operating program deletes the complete data memory of the  OTT CTD  all measurement channels including the info channel   Afterwards the  OTT CTD determines and saves the water level  the water temperature and the  specific conductivity again in accordance with the sample interval set     15 Update OTT CTD firmware    If necessary  you have the possibility to update the OTT CTD firmware  operating  system   This makes sense if  for example  devices delivered at different times are  to receive the same operating system version  Updating is carried out via the  OTT CTD operating program  According to availability  an updated version of the  OTT CTD firmware can be found on the internet site  www ott com      How to update the firmware    Download the new version of the firmware  file  e g   OTT CTD V1 00 3 bin    from the Internet site    Copy the file  OTT CTD VX XX X bin  to the directory in which the 
21. anufacturers    Do not mix lithium and alkaline batteries    Properly dispose of dead batteries  Do not include in household waste     Suitable battery types    3x 1 5 V AA cells  LR6  FR6    alkaline or lithium design  LiFeS  Energizer L91     Insert the batteries as follows    Slide the pipe casing of the communication unit approximately 30 cm in the  direction of the pressure probe cable    Insert 3 batteries  LR6 FR6   AA  in the battery compartment   as shown in Figure 2  Ensure that the polarity is correct    Screw the pipe casing of the communication unit back on     Notes    The OTT CTD begins measurements within a few seconds of the batteries being  inserted  there is no on off switch     If the OTT CTD is to be shut off     remove the batteries  This will prevent the  batteries from draining prematurely and will stop any recording of unusable  measurements    When bringing back into operation it takes   dependent on how much data is  stored     up to 7 minutes until the OTT CTD begins measurements again                Battery compartment Backstop    Pipe casing  communication  unit  LR6 FR6   AA   Battery          Battery life    For a 1 hour sample interval an 50 m system length  without ITC      Lithium batteries  at least 5 years  Alkaline batteries  at least 1 5 years  high quality battery types   Notes    The operating program has a calculation function that determines the approxi   mate battery life based on the operating parameters currently set  The basis fo
22. asuring range      m  3 or 2  cm  1 or 0  feet  2 or 1  inch  1 or O      bar  4 or 3  psi  3 or 2    C  2or I   F  Zor 1    for the pressure output type   7  input reference    for the depth output type   7  input contact gauge value  for the level output type   7  enter staff gauge value  generate reference to the level zero     pressure    558    24h  time delay at which the OTT CTD records measured values and stores them   store only if sample interval   mean interval   The sample interval setting is at a fixed  time raster  e g  sample interval of 10 minutes  time raster      00 10  00 20  00 30         55    24h  time delay during which the OTT CTD calculates an arithmetic mean from  the sample interval values and stores it  The storage interval must be equal to or larger  than the sample interval  The sample interval must divide exactly into the storage inter   val  e g  sample interval  10 minutes  storage interval  1 hour     the OTT CTD stores an  arithmetic mean from 6 sample interval values each hour      21    22    Additional settings for conductivity  salinity and TDS    Temperature compen  the mathematical algorithm on which the calculation   sation  Conductivity  of the specific conductivity at a defined reference  temperature is based  Possibilities   Freshwater    Saltwater   Standard method 2510    ISO  7888 EN 27888          none   For the  Standard  method 2510   the reference temperature can be  selected  20   C or 25   C    Calculation method the mathe
23. com   www ott com    
24. data       Fig  1  Setting up a groundwater measure   ment station with the OTT CTD     The OTT CTD essentially consists of three com   ponents  Communication unit  pressure probe  cable and pressure probe with datalogger     OTT CTD system length   length of  communication unit   cable length    length of pressure probe with datalogger    The system length is required when  ordering an OTT CTD  When setting   the OTT CTD operating parameters the  system length is not required       The rubber stop attached to the pressure  probe cable prevents the pipe casing from  falling when the communication unit is  open  Do not move the rubber stop            Top cap                                        Observation well                         Pressure probe cable                         Pressure probe with  datalogger unit                    77   7777777777  7 777 7777777771777717 777777171                Communication unit      O  Oo         r Rubber stop      D  E  2  E     2  N  S  a  Y Mine  T      A    E       E2  o 8  R  g  u 2 5  a  5       o 3  gt   DE            POS 58  Ej  3   aii    BLO  Se      No un  IT N Conductivity sensor  oO   o                   I         Pressure Sensor             7777717          Fig  2  Installing batteries     5 Installing  checking  and exchanging batteries    Please note    Only use the battery types indicated  no rechargeable batteries    Always use brand new batteries  Do not mix used and new batteries   Do not mix batteries of different m
25. dow  with the graphical view of the selected data      Show ruler  function key  F10   Move ruler  arrow keys            Zoom in  draw the required area by dragging with the mouse      Zoom out  function key  F12     Print graphic  Select the  Print  function in the  File  menu    Close graphic  Select the  Exit  function in the  File  menu  or X      How to display the data numerically    Click on the  View export data  button in the start window    Select the required station sensor in the list  multiple selection possible     Click on the  Table  button     the operating program displays the window with   the numerical view of the data in the OTT data protocol      Only display measured values  select  Values       Highlight special areas  only with information data   Select  Observer  amp   operation    Alarm  amp  limit    Communication  amp  error    Error  amp  service log     Print numerical display  Select the  Print  function in the  File  menu    Close numerical display  Select the  Exit  button  or RI      How to delete the data    Select the required station sensor in the list  multiple selection possible    Click on the  Delete  button   Confirm the confirmation message with  OK         gjort Orpheus Mini   OTT CTD operating program   View   Export data  File Help                    Station FROM Date    0000254892 11 11 2009 12 11 2009  0000254892 0002 11 11 2009 12 11 2009 Graphic  0000254892 0003 11 11 2009 12 11 2009  0000254892 0004 11 11 2009 12 11 2009 Tab
26. easured values    Temperature values erroneous       check configuration   for advanced operation  see online help     Pressure sensor values erroneous       check configuration   for advanced operation  see online help         check the pressure compensation capillary in the communication unit   blocked    If necessary  clean        check pressure sensor for contamination  Clean if necessary       The standard factory configuration is located on the  OTT CTD software  CD ROM      37    38    Conductivity values erroneous      check configuration   for advanced operation  see online help        check conductivity sensor for contamination  If necessary  clean and then  recalibrate  Calculated salinity values erroneous  Err 05       check configuration   for advanced operation  see online help        input value is outside valid range     salinity by standard method  1 0   C  lt  t  lt  35 0   C for salinity 2 0 PSU  lt  salinity  lt  42 0 PSU    Salinity by  USGS 2311  method  0   C  lt  t  lt  30 0   C for salinity 2 0 ppt  lt  salinity  lt  60 0 ppt  Observer does not display calculated salinity values      values lie outside the valid range     salinity according to standard method  2 0 PSU  lt  salinity  lt  42 0 PSU    salinity according to  USGS 2311  method  2 0 ppt  lt  salinity  lt  60 0 ppt    No measured values in database    Configuration incorrect        check configuration   for advanced operation  see online help    Battery voltage lower than about 3 6 V2       
27. enter the SMS C phone no  of the network operator   Phone number of  the SMS service center  The SMS C phone number is normally already saved  on the SIM card     in this case  set to  SIM card      Click on the  Next  button    Select the SMS transfer mode    Click on the  Next  button    Set the interval for transmission of the storage values    Click on the  Next  button    Set the offset time for the transmission interval    Click on the  Exit  button    Check the settings afterwards and correct as necessary    Click on the  OK  button     If necessary  Set up cyclical data transfer using GPRS in com   bination with an OTT ITC intelligent top cap      Activate check box  Modem ITC connected     Activate check box  GPRS data transmission active     Click on the  ITC settings  button     the operating program opens a window  with several tabs for setting all the necessary operating parameters    Make the required settings in the  General    Operator    FTP  and  Time  sync   tabs  Caution  The SIM PIN for the SIM card used should be deactivated   Otherwise enter the SIM PIN in the  Modem ITC  tab  advanced operation  mode     Click on the  OK  button     1 For further information  refer to the online help       org Orpheus Mini   OTT CTD operating program          Meas type   Meas range    Unit    Sample interval    Storage interval    Measuring range  System length  Date   time    File OTTCTD ITC Help  OTT CTD  Number  0000254892 Name nr t    lt    iC   s   s   sS Battery 
28. es OTT OrpheusMini_CTD Export    No path can be  set for the  Excel  format  the operating program automatically opens Microsoft  Excel during the export and displays the data in a new worksheet  In format   Hydras 3  the  2  button can be used to automatically find the installation  directory of OTT Hydras 3    In format  CSV   set the  Field separator    Dec  separator    Date format   and  Time format   For the date and time formats  the typical windows place   holders can be used     Step 2   Exporting data     Select the required station sensor in the list  multiple selection possible    Click on the  Export      button    Confirm message about successful export with  OK     If necessary  delete the exported data  click on the  Delete  button   Confirm the message with  Yes      Fig  17  Display export data     11 Displaying data    After reading out into the OTT CTD operating program  the measurement and  information data can be found in the the  RAWDATA  sub directory of the pro   gram directory  standard setting   C  Program  files OTT OrpheusMini_CTD RAWDATA    For an initial check  you can display  the data graphically and numerically and print them out if necessary     How to display the data graphically    Click on the  View export data  button in the start window    Select the required station sensor in the list  multiple selection possible   info  data can only be displayed numerically     Click on the  Graphic  button  gt  the operating program displays the win
29. esistance to electrostatic discharge  ESD       Resistance to electromagnetic fields       Resistance to transient fields  burst        Resistance to surge       Resistance to HF  asymmetric     Line borne and radiated interference    4 MB   approx  500 000   9   1 Info channel  Water level pressure  Temperature  Conductivity   Supply voltage   5 seconds     24 hours  5 seconds     24 hours        Simple or advanced operation       5 extended sample intervals with start stop time indication  support of  pump fests       Selection of units       Pressure level measurement or depth measurement       With consideration of the local gravitational acceleration       Compensate for water density using salinity and temperature       Temperature compensation for conductivity  Options  Freshwater   Saltwater  Standard method 2510  reference temperature  25   C or  20   C   ISO 7888   EN 27888  none       Salinity  standard method or USGS 2311      Display of instantaneous values with level monitoring function       Password protection       Measured value processing  Calculation of mean  delta storage  scal   ing function  extreme value storage  limit control of the sample interval       Virtual sensor virtual terminal       Together with OTT ITC  alarm management  remote data transfer    1   2  3   4    4 5    5   6      gt  2     400 mm x 22 mm  317 mm x 22 mm  1 5     200 m  1    5 cm    approx  0 410 kg  approx  0 430 kg  approx  0 082 kg m    ABS  POM  stainless steel 1 4539  904
30. his  removes any air bubbles in the conductivity sensor    Pass the communication unit through the holes in the suspension bracket until  the O ring sits on the suspension bracket  see Fig  10     Determine the current depth with a contact gauge and make a note of it    Set operating parameters  see Chapter 7     Close the top cap           Suspension  bracket                15    16    Fig  10  Installing the OTT CTD in obser   vation wells beginning at 2  diameter  with a top cap without a recess     Example  2  observation well                                               Top cap      upper part         Screw  Communication  2 ae Infrared  unit     interface   Suspension        O ring    bracket             Caution  For top caps with installed suspension brackets  never completely  remove the screw in the top cap  Otherwise  the suspension bracket and the  OTT CTD will fall into the observation well     Fig  11  Example installation of an  OTT CTD in an observation well  beginning at 2  diameter     Proceed similarly for  other local conditions     6 4 Installing in observation wells beginning at 2  diameter  without a top cap    For this installation case  an individual solution to fasten the OTT CTD must be  found depending on the measurement station  Example  fix a suspension bracket  with an M   hex bolt nut laterally at the upper end of the observation well     Required accessories  Suspension bracket  for example  M   hex bolt  plate  and M   hex nut     Install the
31. iccant capsules   Slide the pipe casing of the communication unit back on until it stops   Slowly and carefully replace the communication unit into the observation well   Close the top cap observation well cover           Pipe casing  Communication unit    Dessicant capsules             16 3 Checking replacing the batteries  See Chapter 5    16 4 Calibrating the conductivity sensor    The goal of the calibration is to determine the so called  cell constant  for the con   ductivity sensor  To do this  it is necessary to check the zero point and to carry out  a conductivity measurement with a standardized calibration solution     The cell constant   as well as the unchanging geometric measurements  the mate   rials used and the construction of the conductivity sensor     takes account of the  aging process of the electrodes     Recommended interval  every 12 months  after every cleaning   with difficult local measuring conditions  heavy deposits    as required every 4 to 6 months   Required accessories  standardized calibration solution  calibration container   see Chapter 2   Order numbers      35    Caution  For an optimum calibration process  the OTT CTD and the calibration  solution must be at the surrounding temperature     How to calibrate the conductivity sensor   Carefully clean the pressure probe and dry well  see Chapter 16 1     Start the OTT CTD operating program    Set up the PC OTT CTD communication link  see Chapter 7 2     Press the  Setup device  button   Butto
32. ing data   Protecting the OTT CTD and OTT CTD operating program with a password  Date and time settings   Deleting the data memory    Installing new OTT CTD firmware    Maintenance work   16 1 Cleaning the pressure probe   16 2 Replacing the desiccant capsules  16 3 Checking replacing the batteries  16 4 Calibrating the conductivity sensor  Error messages  Troubleshooting fault correction  Repair    Notes about the disposal of old units    Technical data    Appendix A   Declaration of conformity for OTT CTD    28    29    30    31    33    34    34    38    40    40    41    43    1 Scope of supply    OTT CTD      I groundwater datalogger consisting of a communication unit with installed  O ring  pressure probe cable with pressure compensation capillary and  Kevlar core for longitudinal stabilization  2 desiccant capsules  pressure probe    with integrated conductivity sensor and datalogger     3x 1 5 V batteries  alkaline  LR6   AA  or lithium design  FR6   AA       1 brief Instructions    I factory acceptance test certificate  FAT     2 Order numbers    OTT CTD    Accessories    Groundwater datalogger  Information needed for order      Measuring range  0     A m     0    10m   0    20 m   0    40 m   0    100 m      System length  1 5    200 m   1    5 cm       Battery type  alkaline  lithium    Installation kit  consisting of  adapter ring 1   adapter  plates 2   4   6   suspension bracket    Adapter plates 3  and 5     Suspension brackets  for top caps starting at 2  with
33. insert new batteries       The standard factory configuration is located on the  OTT CTD software  CD ROM      19 Repair    With a problem with the device  use Chapter 18  Troubleshooting  fault correction  to see if you can resolve the problem yourself   In case of device defects  please contact the repair center of OTT     OTT Hydromet GmbH  Repaircenter   Ludwigstrasse 16   87437 Kempten   Germany  Telephone  49 831 5617 433  Fax  49 831 5617 439  repairQott com    Caution  Only have a defective OTT CTD checked and repaired by the OTT  repair center  Under no circumstances carry out any repairs yourself   Any repairs or attempted repairs carried out by the customer will result  in the loss of any guarantee rights     20 Notes about the disposal of old units    Within the member countries of the European Union   In accordance with the European Union guideline 2002 96 EC  OTT takes back  old devices within the member countries of the European Union and disposes of  them in an appropriate way  The devices concerned by this are marked with the  symbol shown aside     For further information on the return procedure  please contact your local sales  contact  You will find the addresses of all sales partners in the internet on   www ott com   Please take into consideration also the national implementation    of the EU guideline 2002 96 EC of your country     For all other countries  Dispose of the OTT CTD properly after taking out of service   Observe the regulations valid in you
34. le  0000254892 INFO 11 11 2009 12 11 2009   Delete    Export CSV  Export Excel  Export OTT MIS  Export Hydras 3    Export Options       Back            29    Fig  17  Protecting the OTT CTD operating  program with a password     input window General    A password protected OTT CTD operating Pacer PASSWORT  program only makes sense if the operating  program starts in observer mode  activate    30    12 Protecting the OTT CTD and OTT CTD operating  program with a password  To prevent any unauthorized input of operating parameters  you can    protect the OTT CTD and  OTT CTD operating program    with a password for each     How to protect the OTT CTD operating program with a password    Start the OTT CTD operating program    Select the  Options  function in the  File  menu    Enter a maximum eight digit password in the  Password  input field  Permitted  characters 0    9  A     Z   This password is independent of the OTT CTD  password     Activate check box  Only observer mode     Click on the  OK  button    Close the operating program  gt  the operating program is now protected  the   Setup device  button in the start window is no longer visible        to establish a password  I    Advanced operation          bserver    check box  Only observer mode    IV Only observer mode  I Do not calculate manual input value with scaling    I Suppress display of instantaneous value before input    OK                Caution     If the password is lost  you can no longer configure or paramete
35. lifetime    Water level   Pressure   Temperature Conductivity   Salinity   TDS    Number jon     fooo2 fooos     Water level v   Depth v     m  0 01  v     Depth to water   set newly    m    Dynamic density compensation    fc  0 01  v      01 00 00 v    01 00 00 v             0 10    100 00 mS cm v    ms om v     Temperature compensation     Freshwater 3   at Reference temperature             0 1 bar  5 00 m  1171172009 16 19 53  PC  11 11 2009 16 19 55      Set date time            I Modem   ITC connected          OTT CTD       Program    com1  x     Read    iba   OTT DuoLink v        Configuration  Load    Save      Back                 Download successful    OTT CTD  Number  Name    Station number  10 characters  alphanumeric        Station name  max  40 characters  alphanumeric     Water level   Pressure   Temperature   Conductivity    Number  Name    Meas  type Meas range    Units    Decimal places    Depth to water     set newly  Water level     set newly  Pressure value     set newly    Sample interval    Storage interval    Sensor number  4 characters  alphanumeric     Sensor name  max  40 characters  alphanumeric       Water level  Depth or level pressure      Conductivity 0 001     2 000 mS cm   0 10     100 00 mS cm      Salinity  2     42 00 PSU   2    60 00 ppt       Pressure sensor  m  cm  feet  inch   bar  psi      Temperature    C     F     Conductivity  mS cm   pS cm     Salinity  PSU   ppt     TDS  mg l   fixed specification  depending on the units and    me
36. matic refresh       If necessary  Input observer registration into the entry field of the pressure  sensor and click on the  Save  button    If necessary  Start new instantaneous value measurement  click on the  Refresh   button    Z  Automatic refresh  automatically starts an instantaneous value  measurement every 5 seconds     End the observer function  click on the  Exit  button and close the OTT CTD  operating program     Notes    The OTT CTD stores each time the observer function is called in an info channel  with the date and time of day  After reading in and accepting into the OTT  Hydras 3 user software  this information can be displayed in the evaluation  window of a sensor using the  Info Data   Station   Displays  function    Observer registration general    They are likewise visible via the  View data     Table  function of the OTT CTD operating program    If the  Do not calculate manual input value with scaling  check box in the   Options  window is activated  the OTT CTD also stores the input check value   as well as the current instantaneous value  After reading in and accepting into  the OTT Hydras 3 user software  this information can be displayed in the evalu   ation window of a sensor using the  Info Data     Sensor     Displays  func   tion   Observer registration with check value    They are likewise visible via the   View data    Table  function of the OTT CTD operating program     9 Reading out data    Read out data as follows  on location     Start the
37. matical algorithm on which the calculation    Salinity  of the salinity is based  Possibilities   Standard  method    USGS 2311   With the calculation  method USGS 2311  the unit is  ppt   with the  standard method the unit  PSU  is fixed     Store salinity With the check box activated  the datalogger saves   Store TDS the values  With the check box deactivated  the  instantaneous values are still visible in the observer  window     Other displays in the  Basic operation  window    Measuring range Measuring range of the pressure probe   see Fig  1   System length Cable length including communication unit   pressure  probe  see Fig  1    Date time Internal date time of the OTT CTD    Detailed information on the  Advanced operation  function can be found in the  online help     Factory settings    OTT CTD   Number Serial number  Name OTT CTD I   Water level Pressure Temperature  Number 0001 0002  Measurement type Water level Temperature  Type of output Water level depth  Unit m  EG  Decimal places 3 2 2  Sample interval Th Th  Storage interval Th Th   Conductivity Salinity TDS  Number 0004 0005 0006  Name Specific Conductivity Salinity TDS  Unit mS cm PSU mg l  Decimal places 2  Measuring range 0 10    100 00 mS cm 2    42 00 PSU  Temperature compensation Fresh water  Calculation method Standard method  Sample interval Th Th Th  Storage interval Th Th 1h  Modem ITC deactivated  connected    7 4 Saving loading OTT CTD configuration    The functions  Load  and  Save  are provided
38. mber and station name     To export a configuration  OTT Hydras 3    This export type is required to transfer an OTT CTD configuration to the OTT  Hydras 3 user software  This    XML  file is read in via the Hydras 3 function   File    Import Station Configuration  XML       Select the  Export Hydras 3  XML   function in the  File  menu    In the  OTT CTD XML configuration  window  check the file name  select the  memory location and click on Save     the operating program stores the config   uration ina    XML  file     Note    The file name of the    XML  file is arbitrary  To avoid affecting later data  transfer  it should not be changed  The operating program proposes a combi   nation of measurement station number and measurement station name along    with the ending    STATION XML      To import a configuration  Select the  Import Configuration  function in the  File  menu   In the  OTT CTD Import Configuration  window  select the appropriate    BIN   file and click on Open     the operating program reads in the configuration  from the   BIN  file   Prior to this  save any configuration that is already open      Note    The standard factory configuration can be found on the CD ROM  OTT CTD  Software      Fig  14  Input window to establish  options for the observer     For further information on protecting  the OTT CTD operating program with  a password  see Chapter 12     8 Determining and displaying instantaneous values     observer function     The operating program ha
39. n not visible      deactivate the  Only   observer mode  check box in the  Options  function in the  File  menu and   restart the operating program     Select the  Calibrate conductivity sensor  function in the  OTT CTD  menu    Press the  Start zero point test  button  The operating program starts an instan    taneous value measurement and updates this every 5 to 6 seconds  If the mea    surement result displayed  blue triangle  is not in the green area     clean the  conductivity sensor again and dry it well    Press the  End zero point test  button    Select the calibration solution used  Alternatively  select  Other      and enter   the appropriate conductivity value    Add the calibration solution to the calibration container  filling height approx  3 4     Place the OTT CTD into the calibration container  screw on cover in place  and   tighten the screw on cover by hand  Caution  There must not be any air bub    bles in the slot of the conductivity sensor  If necessary  lightly shake the calibra   tion container    Press the  Start calibration  button  The operating program starts the instan    taneous value measurement again and continuously updates it    Wait until the values for conductivity and temperature do not change for sever    al updates  The operating program displays the current and newly calculated   cell constant  Caution  Temperature values shown in red     the conductivity  sensor is not at the temperature of the calibration solution  New cell constant  
40. out recess  and for universal installation    Cable suspension unit  for OTT CTD system lengths of  gt  100 m    CD ROM   OTT CTD Software     OTT CTD operating program for PC  Top cap   with integrated fastening hook     for 2  observation wells     for 4  observation wells     for 6  observation wells    Intelligent top cap OTT ITC  for GSM remote data transfer    5 sided key  to lock OTT top caps    Optical OTT Duolink reading head  Optical OTT IrDA link USB reading head    Calibration container    55 445 001 9 0    55 440 025 9 2    55 440 444 4 1  55 440 450 4 1    on request    56 571 000 9 7    24 220 052 9 5  24 220 054 9 5  24 220 057 9 5    55 530 0xx 3 2    20 250 095 4 1    55 520 017 4 2  55 520 026 9 2  55 445 025 9 2    Replacement Alkaline battery  parts  LR6   AA  3 units required  Consumable    materials Lithium battery    FR6   AA  3 units required    Desiccant capsules  2 in aluminum bags    Conductivity calibration solution    0 1 mS cm  1000 ml    0 5 mS cm  946 ml    1 412 mS cm  1000 ml   12 856 mS cm  946 ml    47 6 mS cm  1000 ml    3 Basic safety information    96 800 004 9 5    97 800 008 9 5    97 100 280 9 5    55 495 350 9 5  55 495 351 9 5  55 495 352 9 5  55 495 353 9 5  55 495 354 9 5    Read these operating instructions before using the OTT CTD for the first time   Become completely familiar with the installation and operation of the OTT CTD  and its accessories  Retain these operating instructions for later reference     The OTT CTD is used
41. r  this are lithium batteries    Due to their design  at 0  C  alkaline batteries drop to 50   of their original  20   C capacity and at  10   C they drop to approximately 35    If ambient  temperatures are expected to drop below 0   C at the measurement station  it   is recommended that lithium batteries be used     Check the battery voltage as follows    Set up the PC OTT CTD communication link  see Chapter 7 2     Select the  OTT CTD  menu   View Instantaneous values  function      the  OTT CTD starts an instantaneous value measurement     the  Observer   window indicates the current battery voltage and the energy withdrawn from  the batteries so far in Ah    If the battery voltage is  lt  3 6 to 3 7 volts     replace the batteries    Click on the  Exit  button    Close the operating program     1 With suitable settings  Menu  File   Function  Options    the operating program starts with the   Observer  window     Replace dead batteries as follows    Open the top cap observation well cover    Pull the communication unit approximately 80 cm out of the observation well  and hold  a second person would be useful     Slide the pipe casing of the communication unit approximately 30 cm in the  direction of the pressure probe cable   The rubber stop located on the pressure  probe cable  see Fig  1  prevents the pipe casing from falling  Do not move the  rubber stop     Remove dead batteries    Insert 3 new batteries  LR6 FR6   AA  into the battery compartment as shown  in Figure 
42. r adapter plate  Observation wells beginning at 2  in diameter  see 6 3  top caps without recess for adapter plate  Special case  Observation wells beginning at see 6 4    2  in diameter without top cap    System length over 100 meters  For installation of an OTT CTD with a system length of more than 100 meters   there is a special cable suspension unit available  see accessories   This suspension    unit is described separately        Adapter plate 6       Adapter plate 2                 Fig  4  Installation accessory set   Part 2     Fig  5  Adapter plate accessories 3  and 5      1   hole for OTT CTD  2   hole for contact gauge  3   recess for screws on the top cap       Adapter ring 1     Suspension bracket                         Adapter plate 5     Adapter plate 3                11       6 1 Installing in 1  observation wells    Required accessories  Adapter ring for 1  observation wells     Install the OTT CTD as follows    Preliminary work  If not already done  install batteries  see Chapter 5    Open the observation well cover    Determine the current depth with a contact gauge and make a note of it   Slide O ring from the communication unit onto the pressure probe cable   Place adapter ring over the pressure probe cable and slide it until it comes  to rest against the communication unit    Slide O ring back onto the communication unit    Place the pressure probe in the observation well    Slowly and carefully lower the pressure probe with the pressure probe  cable  
43. r country for the disposal of electronic  devices   Never put the OTT CTD into the normal household waste     Materials used  See Chapter 21  Technical data    39    21 Technical data    Water level  Measuring range    Resolution     Accuracy  linearity   hysteresis    Long term stability  linearity   hysteresis    0 Point   Overload safe without permanent mechanical damage  0    0 4 bar    0    I bar   0    2 bar   0    4 bar   0    10 bar  Units    Pressure sensor  Temperature compensated operating range    Temperature  Measuring range  Resolution  Accuracy   Units    Conductivity  Measuring ranges    Resolution  0 001    2 000 mS cm  0 10    100 00 mS cm  Accuracy  0 001    2 000 mS cm  0 10    100 00 mS cm  Units  0 001    2 000 mS cm  0 10    100 00 mS cm    Power supply    Current consumption  active  measurement  active  communication  passive  Lifetime  1 h sample interval  50 m system length  w o ITC   with lithium batteries  with alkaline batteries    Clock  Design  Accuracy  Buffer period for battery replacement    Interface    Storage temperature        Am water column  0     0 4 bar      10 m water column  0     1 bar       20 m water column  0     2 bar       40 m water column  0     4 bar       100 m water column  0     10 bar   0 001 m  0 1 cm  0 01 ft  0 1 inch  0 0001 bar  0 001 psi     0 05   full scale     0 1   a full scale     0 1   full scale    ooo0o0    4 bar   10 bar   15 bar   25 bar   40 bar   m  cm  ft  inch  bar  psi   ceramic  temperature compen
44. reports and therefore corresponds to the regulations of the Directive   et les rapports d essais notifies et  ainsi  correspond aux r  glement de la Directive     Ort und Datum der Ausstellung  Kempten  den OS Toro      Place and Date of Issue   Lieu et date d    tablissement       Name und Unterschrift des Befugten    Name and Signature of authorized person    Nom et signature de la personne autoris  e Dr  Anton Felder   CEO     OTT MESSTECHNIK GmbH  amp  Co  KG Gesch  ftsf  hrer  Dr  Ing  Anton Felder   Pers  nlich haftende Gesellschafterin  OTT MESSTECHNIK Verwaltungs GmbH    Postfach 21 40   87411 Kempten Sitz der Ges   Kempten   Registergericht Kempten HRB 7687 und HRA 3807   USt  ID  Nr  DE 128 780 710   Steuer Nr  127 171 51206  Ludwigstra  e 16   87437 Kempten   Tel    49 0 831 56 17 0 Commerzbank AG M  nchen  BLZ 733 400 46    Kto Nr  775 0649 00   BIC  COBADEFF733   IBAN  DE13 7334 0046 0775 0649 00  Fax   49  0 831 56 17 209 LBBW Kempten  BLZ 600 501 01    Kto Nr  4546443   BIC  SOLADEST   IBAN  DE81 6005 0101 0004 5464 43  info ott hydrometry de Sparkasse Allg  u  BLZ 733 500 00    Kto Nr  18 861   BIC  BYLADEM1ALG   IBAN  DE24 7335 0000 0000 0188 61  www ott hydrometry de Postbank M  nchen  BLZ 700 100 80    Kto Nr  933 53 809   BIC  PBNKDEFF   IBAN  DE57 7001 0080 0093 3538 09       42       43    Document number    55 445 001 B E    02 0411    OTT Hydromet GmbH    Ludwigstrasse 16   87437 Kempten   Germany  Tel   49 831 5617 0  Fax  49 831 5617 209    info ott 
45. rize the   OTT CTD on this PC  If this occurs  contact the OTT HydroService    The password assigned here only pertains to the OTT CTD operating program  installed on this PC  The OTT CTD itself is not protected by this     How to release a password protected OTT CTD operating program    Start the OTT CTD operating program   Select the  Options  function in the  File  menu   Enter the password in the  OTT CTD  window        Fig  18  OTT CTD operating program    release   password input window   Enter password  0  9 4  Z     ss ooo    Abbrechen         Click on the  OK  button    Deactivate the  Only observer mode  check box    Remove the entry in the  Password  field    Click on the  OK  button     the operating program is released     Fig  19  Protect the OTT CTD  with a password   input window  to establish a password     Fig  20  OTT CTD release    password input window     How to protect the OTT CTD with a password     Start the OTT CTD operating program    Set up the PC OTT CTD communication link  see Chapter 7 2     Click on the  Setup device  button in the start window    Click on the  Advanced operation      button     Button not visible      activate the check box  Advanced operation      in the   Options  function in the  File  menu     Click on the  Read  button    In the tree view  select  Communication Interface     Enter a maximum eight digit password in the  Password  input field  Permitted  characters 0    9  A     Z   This password is independent of the OTT C
46. rther information  see the online help for the OTT Hydras 3  Basic  software   With a Pocket PC and the  OTT Hydras 3 Pocket  software    For further information  see the OTT Hydras 3 Pocket operating instructions     27    28    10 Exporting data    After reading out into the OTT CTD operating program  the measurement and  information data can be found in the  RAWDATA  sub directory of the program  directory  standard setting   C  Program files OTT OrpheusMini_CTD RAWDATA     To transfer to an external application  you can export the data in various formats     CSV   structured text file in CSV format  Comma Separated Values   Excel   Microsoft Excel spreadsheet program   OTT Hydras 3   raw data format for the OTT Hydras 3 user software  OTT MIS   OTT specific file format for automatically importing into the  measurement database of the OTT Hydras 3 user software    Using the  Export Options  dialog window  you can set the location for saving  and  with CSV format  various export parameters also  The data remain in the  sub directory after export  If necessary  you can specifically delete them     Info data can only be exported to the OTT Hydras 3 user software   How to export the data    Step 1   Make export settings     Click on the  View export data  button in the start window    In the  View export data  window  click on the  Export Options  button  see  Figure 17     Enter the path to the storage location of the required export format  standard  setting   C  Program fil
47. s a so called  Observer  to enable the determination    and display of instantaneous values  The observer also makes it possible to enter  a manually determined measurement  e g  contact gauge value  into the datalog     ger  observer registration      Various options can be set concerning the observer     Selection possibility in the start window of the OTT CTD operating program     direct and sole call of the observer  or     calling a window to set the operating parameters   Effect of an optionally entered observer registration  pressure sensor       observer registration is used solely as a check value  or     observer registration leads to a change in the value  setting offset   Suppress the display of the instantaneous value  pressure sensor  prior to the  input of an observer registration    Set the observer options as follows    Start the OTT CTD operating program    Select the  Options  function in the  File  menu   If the option dialog cannot  be called  a password must first be entered  See Chapter 12    Activate the check box es  for the observer option s  required       Only observer mode     Do not calculate manual input value with scaling   only check value no value change      Suppress previous instantaneous value display   Click on the  OK  button            General      I Advanced operation        Password    Observer  I Only observer mode  I    Do not calculate manual input value with scaling    I Suppress display of instantaneous value before input    OK  
48. sated     5   C      45   C  ice free      25   C      70   C  0 01   C     0 1   C     C    SF    0 001     2 000 mS cm  0 10     100 00 mS cm    0 001 mS cm  0 01 mS cm     0 5   of measured value  at least  0 001 mS cm    1 5   of measured value  at least  0 01 mS cm     mS cm   pS cm  mS cm    3 x 1 5 V batteries  LR6   AA  FR6   AA   alkaline or lithium design   LiFeS  Energizer L91     55 mA  65 mA  23 pA    at least 5 years  at least 1 5 years  high quality battery types     real time clock   1 minute month  at  25   C   approx  10 minutes    Infrared  IrDA    40   C      85   C      at a value range of  32 750 m   3275 0 cm   327 50 ft   3275 0 inch   3 2750 bar   32 750 psi    40    Data memory   Measurement memory   Number of measurements  Number of logical channels  Physical channels  input signals     Sample interval  Storage interval  mean interval     Individually configurable functions    Mechanical Data  Can be installed in observation wells    with adapter ring    with adapter plates for top caps  with recess  OTT  HT     with suspension brackets for top caps  without recess universal installation  Dimensions  Communication unit L x     Pressure probe  L x      System length  cable length including  communication unit pressure probe     Weight  communication unit  incl  batteries   pressure probe  Pressure probe cable  Material  Pressure probe housing  Cable jacket  Communication unit  Protection class  Communication unit  Pressure probe    EMC limits       R
49. shown in red     the cell constant lies outside the valid range    Press the  End calibration  button    Confirm the resulting question   Save new cell constant        cell constant is within the tolerance range  gt   No     conductivity value is in the green area        Cell constant is outside the tolerance range and within the valid range  gt    Yes   conductivity value is in red area   current cell constant is shown in  black    If the cell constant is outside the valid range     error message    Close the OTT CTD operating program    Rinse the pressure probe thoroughly with clear water    Reinstall OTT CTD  see Chapter 6     Determine current reference value with a contact gauge and enter it    see Chapter 7 2 or 8     Close the top cap observation well cover    Dispose of the used calibration solution     Notes    Only use the calibration solution once    Always store the calibration solution in a closed container   The carbon dioxide  in the surrounding air and evaporation can change the conductivity value    Avoid water entering the calibration solution  Always dry the calibration con   tainer carefully after calibration    Calibration solution can be disposed of in the normal public sewer system    The OTT CTD saves the last three cell constants in a history  see  Advanced  operation       channel  Conductivity       17 Error messages    If erroneous measurements occur  or if an operating fault occurs  the OTT CTD  stores one of the following error messages in the da
50. sits on the adapter plate  see Fig  7     Determine the current depth with a contact gauge and make a note of it   Set operating parameters  see Chapter 7     Close the top cap                     _    OTT Top cap             Adapter plate 2   __       Recess    Communication _  unit       13             O ring    Communication unit    Recess          Adapter plate       Top cap                Fig  8  Installing the OTT CTD     14    Fig  9  Installing the suspension bracket in  observation wells beginning at 2  diameter  with a top cap without a recess     Example  4  observation well    6 3 Installing in observation wells beginning at 2  in diameter   top cap without adapter plate recess    Required accessories  Top cap with attachment screw for the top of the top cap   Suspension bracket     Caution  There must be sufficient space in the top cap for the infrared interface  to not be damaged when the top of the top cap is closed     Install the OTT CTD as follows    Preliminary work  If not already done  install batteries  see Chapter 5    Open the top cap cover    Remove the upper part of the top cap  see Fig  9     Place the suspension bracket on the screw  see Fig  9     Reattach the upper part of the top cap  see Fig  9     Pass the pressure probe through the holes in the suspension bracket   Slowly and carefully lower the pressure probe into the observation well  with the pressure probe cable    Raise and lower the pressure probe in quick succession approx  30 cm     t
51. ta memory instead of a  measured value     Err 00 internal error  automated measurement    Err O1 internal error  AD conversion error    Err 03 Exceeded measuring range   Err 05 An input value required for the measurement calculation is not in the  valid range  example  the temperature value needed for the calculation  of the salinity  standard method  is outside of the valid range    0   C  lt  t  lt  35 0   C    Err 06 An input value for the measurement calculation is missing   Err 10 Measured value  still  not recorded    Internal errors indicate a device defect if they occur repeatedly     18 Troubleshooting fault correction    No communication possible  operating program OTT CTD     Password programmed        input correct password    Batteries installed        install batteries    Battery voltage lower than about 3 6 V2       insert new batteries    Battery contacts corroded        carefully clean the battery contacts    Pressure probe cable damaged        send OTT CTD to the factory for repair    Infrared interface dirty        clean the infrared interface carefully with a damp  soft cloth    Setting measured value not possible        scaling module missing  Check configuration    for advanced operation  see online help      Communication starts and then breaks off    Battery voltage lower than about 3 6 V        insert new batteries    Distance from readout unit IrDA interface to OTT CTD too small large       maintain correct distance  see Chapter 7 2      Erroneous m
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
SoftBank 202K 取扱説明書  User Manual - IVB 995-OH_M SD XC TYPE 22  Fedders CH60ACZ1VF Heat Pump User Manual  (TSP) CF633R20 Unit Level Logistics System (ULLS)  Dossier de Presse  DMS 800 PSG.indb  La quête du Set T0.5    VITROCERAMIC HOB - User instructions VAŘIE ZE SKLOKERAMIKY    Copyright © All rights reserved. 