Home
        Eiki Service Manual
         Contents
1.               May 1 1980    EIKI INDUSTRIAL       LTD   C P O  BOX 1229 OSAKA JAPAN   06  3311 9475        EIKI INTERNATIONA  INC    27882 Camino   Laguna Niguel  California    92677  cun 831  251 1       When the show must go on             MODEL    N    SERIES      Authorized          Distributor or Dealer        TABLE OF CONTENTS  N  Series    Title  PAGE    INTRODUCTION  314 1  PRINCIPLES OF OPERATIONS  T    Introduction ss c sare due a edet URS EX AE E E coy 15  1 2 Mechanical System                                15  1 3 Electrical System                                 15  1 4 Sound System    vss age pues B    15  314 2  SERVICE PROCEDURES  2                                      CE         ICI 17  2 2 Tools And Test Equipment                         17  2 3 Lubricants And Lubrication Charts                   18  2 4 Troubleshooting Hints                             2  314 3  MODULE REMOVAL AND INSTALLATION  3 1 Cam Tank Module                       wes CE ex ey esr 25  3 2 Amplifier                                             25  P 3 3 Motor Module                                   26  3 4 Transformer                                       26  3 5 Lens Holder And Gate Module                       26  314 4  MODULE REPAIR AND ADJUSTMENTS  4   Cam Tank Module                                27                                5                                   27  Be  Disassembly              lama nane        ehe 27       ResAssembly  osse it bad RE HARALD A RT     28  D  A
2.         Stretch or replace     See Sec 5 4 B       Relocate  amp  check     See Sec 5 4    Check  amp  tighten      Check  2 sprocket timing     See Sec 5 4 B        Check     See Sec 4 1 D1       Check  amp  replace   Check  amp     lubricate take up  spindle cork   See Sec 5 1       Adjust clearance     See Sec 5 4       Replace    Stretch or replace            N  Series     MECHANICAL SYSTEM        SYMPTOM PROBABLE CAUSE REMEDY  30  Film continues to flap on 1  Loop setter gear shaft is    1  Check  amp  lubricate  loop setter roller     in binding  or tight in hub    forward  2  Loop setter gear spring ten    2  Adjust spring tension  or re   sion too strong place                31 Lower loop is lost in   Reverse rubber roller is not   1  Check reverse rubber roller s     reverse    driving the flywheel set collar function    See Sec 5 3    2  Reverse rubber roller binding 2  Clean  amp  remove the cause of       binding  3   2 sprocket not transporting   3  Check  amp  adjust  film   See Sec 5 4   Et 32  Upper loop is lost       1  Damaged  or poor film 1  Repair or replace   forward  2   1 sprocket teeth plate loose 2  Check  amp  tighten               1 sprocket shoe not seating   3  Check  amp  adjust   properly   See Sec 5 5 B1    4  Loop setter roller continues 4  See Symptom No  29   to activate or activates er              ratically   33  Upper loop is lost in 1  Damaged  or poor film 1  Repair  or replace  2  reverse  2   1 sprocket shoe clearance is   2  Check  
3.        Check  amp  replace     Check      Repair or replace   See Sec 4 2  amp  3 2        1  Exciter lamp cover missing or  not installed correctly    2  Incorrect grounding when  connecting the projector to  an external amplifier or  sound system       Defective exciter lamp supply    4  Defective solar cell or con   nections to amplifier               Install cover      Check for ground loop con   ditions       Check voltage and repair    See Sec 4 2     Check  amp  repair       ze    SYMPTOM        SOUND SYSTEM   PROBABLE CAUSE             Series    REMEDY             54  Excessive amplifier hum   optical       continued        5    Front or rear cover speaker  jacks not insulated from the  chassis properly    6  Defective amplifier module      Check  amp  repair       Repair or replace     See Sec 4 2  amp  3 2           55  Excessive amplifier hum   magnetic     56  Distorted sound    1   2   3     Poor film recording   Motor shield not installed  Improper connection to an  external amplifier or sound  system       Magnetic head in poor con     tact with the film       Poor shielding to the head or  the head coil shorted to the    projector s frame    6  Defective amplifier module    tn bw                       Incorrect exciter lamp     Exciter lamp cover not com     pletely installed      Amplifier module is defective    Defective speaker    Magnetic    sound recorded    poorly      Optical sound lens not aligned    correctly      Dirt on the sound drum or    on the s
4.       49  5 3 Reverse Projection  A  Description xi s      esse e Grn bare Wa cp BER ens 49  B  Adjustments    ese ludo Fu des Ve Sa el ee 49  5 4 Lower Loop Setter System  A  Description    us vere n Exp epe Pete ed ha de 50  B  Adjustments  cus               rer      Eres  50  5 5 Self Threading Mechanism       Descriptiot        oes ES uber eee ree atc venae    52  B  Adjustments      esse vue cil rr RE TRE IV RI 55  1   1 Sprocket shoe assembly                      55  2   1 Film guides                         Ret emm 55  3  Retraction of the claw                          56  4   2                                                 56  5  Sound drum rubber pinch                             56  6  Tension guide and roller assembly                 57  7   2 Sprocket and shoe assembly                   58  5 6 Rewind Mechanism  A                                                             59  B                                                            59  5 7 Sound Drum And Flywheel                                                               59      Adjustments 4        u papa               60  314 6  SOUND PICK UP SYSTEM  6 1 Optical Sound Focusing Procedure  A  Description   ail css oue ee dque E pS SUES           S 61  B  Alignment   25s ii             61  1  Tools and equipment                           61  2  Set up procedure                              61  3  Sound focus                                             61  4  Buzz track adjustment                          
5.    the wrong   2  Check  amp  switch  position  NST NT 2  3 models  only   3 Rear cover speaker not 3  Check  amp  connect   plugged in  not applicable on  NST NT 3 models   or exten     sion speakers not connected     4  Defective rear cover speaker    4  Repair or replace   or extension speaker  or  speaker jack   5  Amplifier fuse blown 5  Replace  6  Defective solar cell or con    6  Repair or replace  nections  See Sec 4 2   7  Dirt or foreign object in the 7  Clean   optical sound lens  8  Defective amplifier module 8  Locate  amp  repair  or replace   See Sec 3 2  amp  4 2   1  Incorrect  or defective exciter   1  Check  amp  replace  lamp  or sound lens   2  Dirty exciter lamp  or sound   2  Clean  or replace   lens  or dirt  foreign object in  the optical sound lens  3 Sound optics incorrectly 3  Check  amp  re align sound pick   aligned  up system  See Sec 6 1   4  Low exciter lamp voltage  or   4 Check  amp  repair amplifier  low AC supply voltage  exciter lamp supply   See Sec  4 2   Check wall outlet   5  Weak or defective solar cell  5  Replace   6  Defective speakers  or poor   6  Check  amp  replace  speaker connection  7  Poor film quality 7  Check with another film   8  Defective amplifier module 8  Repair or replace      See Sec 3 2   amp  4 2                             SYMPTOM     SOUND SYSTEM     PROBABLE CAUSE             Series    REMEDY          50  No sound   magnetic only    NST NT 2   3 models     1  Mag Opt switch in the wrong    position      2  
6.   flat response  amplifier  lighter weight  etc  With few exceptions   such as color  most parts from the          series  substitute directly into the previous          series   This manual also covers the NT manual threading  models when excluding the sections on the thread   ing mechanism     1 2  MECHANICAL SYSTEM    The EIKI Self threading projector is controlled by   a single function switch for forward  reverse and   high low lamp positions    A  THREADING   Threading is accomplished by setting the  threading guide lever to thread  function  switch to forward and inserting the film  into the threading channel  The threading  mechanism is automatically released and  engages the claw by a light tug of the film  when attaching it to the take up reel    B  FORWARD LAMP   Advance the switch from forward to lamp  low or high    C  REVERSE   To reverse the projector  return the function  switch to    OFF    and then to reverse and  lamp    D  REWIND   To rewind the film  engage the rewind lever   Attach the film to the empty supply reel  and advance the function switch to  For   ward         1 3  ELECTRICAL SYSTEM                      series projectors are available in voltages  of 100V  110V  120V  220V and 240V  110 220V  and 120 240V dual voltage models are also available   To comply with electrical safety regulations of  various countries  UL  USA   CSA  Canada   VDE   Germany   SEV  Switzerland   SAA  Australia    SEMKO  Sweden   NEMKO  Norway   FEMKO   Finland   DEMKO  
7.  MODULE      Remove the fly wheel     Unplug the rear cover speaker cord     Remove the volume  bass and treble knobs     NOTE  on magnetic models remove the mag opt  knob       Unplug the MT 9 pin plug     Unplug the solar cell and magnetic lead miniature    plug       Remove the two phillips screws  one at each end    of the amplifier       The amplifier can now be removed by sliding    out       To re install  reverse the procedure above  Be    sure that the indicator on the controls lines up  and all the connectors are secured     Fig  425                  3 3  MOTOR MODULE    1  Remove motor belt    2  Disconnect motor nylon connector s     3  Unscrew 3 screws    4  Remove motor module    5  Motor modules are exchanged less pulley    6  To re install  reverse the above procedures    7  Motors with plastic fan housings use the ground  strap  NST after S N 10484  and NT after S N  18550  with metal fan housing  320 12141      the ground strap has been omitted     GROUND STRAP  FOR PLASTIC FAN  HOUSING OF EARLY PRODUCTION     Fig   26       3 4  TRANSFORMER MODULE      Unplug nylon connectors    Unscrew 3 screws    3  To re install  reverse the above procedures    Care should be taken that the wires are routed 3   away from any moving parts of the projector                 3 5  LENS AND          MODULE    1  Swing open the gate and remove the lens   Anti   theft screw on U S A  models     2  Remove the two phillips screws  40  securing  the film gate assembly to the main ca
8.  RUBBER PINCH ROLLER       p  FUNCTION ROTARY SWITCH Fig  17 3 Fig   18  32 N    34  LOOP SAFETY  ROLLER ARM  24  _                              2 4  TROUBLE SHOOTING HINTS      There are four basic steps to trouble shooting   this projector   a  Analyze the symptom  b  Localize the trouble to a functional system  or module  c  Replace or repair that system or module      Isolate the trouble within the module  b  Locate and repair the specific trouble  Checking Semiconductors With A VOM   a  Set the ohms scale to R x 10  b  The forward resistance should be low  c  The reverse resistance should be high          TRANSISTOR NPN TYPE  25    250           19           TRANSISTOR                   25    258  Fig   20          Typical Amplifier Test Set Up       TO PROJECTOR        OSCILLATOR        AMP UNDER TEST             Series   2 4 D   Fig   21  DIODES    NOTE  Forward and reverse resistance LOW  amp   HIGH is only a suggested quick and easy check  of components out of circuit  This test is only  for shorted and open junction test  A VOM will  not test the quality of a semi conductor accu   rately    IC s are best checked by checking the signal  input and output condition  This can be done  by inserting a low level audio tone into the  MIC jack and the signal path from the input  of IC 1 through IC 2 and to IC 3          Amplifier test cables can be easily made from    locally available parts  A nine pin miniature  tube socket and male plug can be wired as an  extension power
9.  cable  allowing the amplifier  to be operated away from the projector  The  solar cell and exciter lamp connection can also  be extended if so desired              Fig  422  EXTENSION CABLE      OSCILLOSCOPE   X Je            jo   peu    SPEAKER           MIC INPUT        24              314   3  MODULE REMOVAL AND INSTALLATION PROCEDURE     N  Series   3 2                    CAM TANK  MODULE       TRANSFORMER  MODULE             Fig   23       MOTOR MODULE                                     AMPLIFIER MODULE    3 1  CAM TANK MODULE    No    PU    Q3 tA              10       Unplug the projector and open the rear cover     Remove the 3 transformer mounting screws    and unplug the transformer   See Sec 3 4       Remove the motor belt     Remove the main driving belt by releasing the  tension gear       Retract the claw by turning the inching knob     Open the lamphouse and remove the lamp and    holder assembly by unscrewing the knurled  nut       Unscrew the two cam tank mounting screws     Remove the cam tank slowly and carefully     Care should be taken to avoid damaging the  claw by striking it against the main casting       Re installation can be done by the reverse    procedure  Care should be taken that the cam  tank worm gear and main drive fiber gear do  not bind  A small amount of gear lash is  required to prevent abnormal wear    To check the timing and claw adjustments   refer to         5 1 and 6 3        Un          M    oo               25       3 2  AMPLIFIER
10.  is not    clutch cover      releasing  or  pulley defective    Clutch cover pulley has no      end play on the  1 sprocket  drum shaft           b Check  amp  adjust  or replace       N  Series    REMEDY     See Sec 5 3       Check  amp  adjust      See Sec 5 1           3  Clutch cam seated incorrect    3  Check  amp  reposition  or re   or defective place   See Sec 5 1   4  Clutch cam binding between   4  Check  amp  adjust  clutch cover pulley         Spacer  28  Weak back tension of the  1  Drive gear is binding  or has  1  Check  amp  clean  or adjust   take up reel in    reverse    no end play on the drive  See Sec 5 3   pulley shaft   2  Clutch cam seated incorrect  2  Check  amp  reposition  or binding between drive gear  See Sec 5 1   and clutch cover assy  or  defective clutch cam  3  Drive gear  or clutch cover   3  Check  amp  replace  defective  29  Loop setter roller continues   1  Damaged or poor film 1  Repair or replace    to activate  or activates         2     ratically in  forward          Loop setter    Insufficient gear spring ten   sion  roller in the    wrong position        2 sprocket plate loose    Lower loop is too small      Insufficient claw protrusion     or claw pitch       Broken claw    Too much take up tension      Incorrect clearance between    the loop setter gear and main  drive belt     Loop setter gear  or main    drive belt defective  or broken      Insufficient tension of the    film shoe springs            w                       11 
11. 62  6 2 Magnetic Sound Playback System  As Descriptio z wey wei npe EN OR CUR oak ERI ao 63  B  Alignment                                      63  1  Tools and equipment                           63  2  Set up procedure                               63  314 7  LAMP CIRCUIT  Tl Lamp Circuit os dou ea Ao ool 9 ed REN RR FOE 65      Descrption xi   lode e ea pace acte Bee 65  B  Replacement and alignment                         65  314 8  ELECTRICAL SYSTEM  8 1 Electrical System                                 65  A  Function Control Switch                           65  B  Electrical Circuit                                         65  314 9  SERVICE UPDATES AND MODIFICATIONS  Attachment  CIRCUIT DIAGRAMS                         67   79    my                     Series    INTRODUCTION   1  This Service Manual provides the necessary informations for the repair  adjustment  and maintenance of  EIKI NST NT series projectors  model NST 0  NST 1  NST 2  NST 3  NT 0  NT 1  NT 2  and      3  and  also EX 2000 N series except for the Xenon lamp supply    2  This service manual contains some part numbers for convenience in identification only  When ordering  replacement parts  refer to NST NT replacement parts list    3  EIKI NST NT series projectors may in the future be improved or modified    Modifications made after the issue of this manual will be covered by Service Updates     A copy of all of the pertinent diagrams are attached at the end of this manual     CAUTION  Care must be exe
12. Denmark   specific models are        N  Series   1 4   manufactured to meet such regulations  including  the option of 50Hz  60Hz  50 60Hz operation   Power transformers vary according to the voltage  range and also to the various electrical safety  requirements  The secondary windings of all trans   formers provide 8V AC to the Pilot Lamp and  Exciter Lamp circuits  24   AC  HIGH  and 22V  AC  LOW  to the Halogen Projection Lamp and  40V or 36V AC to the Amplifier    See Sec  4 4 TRANSFORMER MODULE    Motor ON OFF and Lamp ON OFF and reverse  is controlled by the Function Switch which consists  of a Cam and 5 micro switches   The projection lamp is a Halogen ELC type 24V  250 watt  The  High Low  position will extend  the lamp life  depending upon operating conditions   NOTE   The Halogen Lamp EJL type 24V  200W may also be used with some  reduction in light output   The motor is an induction type with capacitor   Motors are available for all the voltage ranges  above  Optional Synchronous motor kits are also  available for precise film speeds and tele cine  conversions   Transformers and motors are simple and easy to   replace modules  AC power cords  line terminals   and all other electrical parts are designed to meet  the safety requirements of the countries listed        1 4  SOUND SYSTEM    EIKI          series models are designated according to  the sound playback and record capabilities with or  without front cover extension speakers    Optical Playback only  Model NS
13. Magnetic head not in contact  with the sound track   3  Defective or dirty magnetic  head   4  Mag Opt switch defective          bad connection     5  Defective speakers  or speaker    connections   6  Defective amplifier module    1    2    3    4    5    6      Check  amp  switch     Check  amp  adjust     Clean  or replace     Check  amp  replace    Check  amp  replace       Repair or replace   See Sec 4 2  amp  3 2        51  Poor sound or low volume   magnetic only    NST NT 2   3 models     52  Exciter lamp fuse blows       1  Poor sound track      2  Dirty  or defective head     3  Head not making good con  i  tact with the film     4  Incorrect sound head align   ment   5  Defective speakers  or poor  speaker connection   6  Defective amplifier module    3    4     wa    an      Check with another film     Clean or replace      Adjust    See Sec 6 2    Align    See Sec 6 2      Check  amp  replace      Repair or replace   See Sec 4 2  amp  3 2        1  Excessive AC line voltage   2  Incorrect fuse   3  Incorrect  or defective exciter  lamp   4  Defective exciter lamp power  supply   5  Defective exciter lamp socket       53  Amplifier fuse blows     1  Incorrect fuse   2                   connection to an  external speaker system   3  Defective amplifier module       54  Excessive amplifier hum   optical         to be continued                         Check wall outlet     Check  amp  replace     Check  amp  replace       Check  amp  repair   See Sec 4 2      Replace
14. T 0   1    amp  NT 0   1   Optical  amp  Magnetic Playback  Model NST 2  NT2  Optical  amp  Magnetic Record Playback    Model NST 3  NT 3   Model NST 1   2  amp   3 and      1  NT 2  amp  NT 3 are  standard with two  4 ohm  12 5cm  speakers built   in the front cover  Models NST 1  amp  2  and NT 1   amp  2 are also standard with one  8 ohm  12 5cm   speaker built in the rear cover    Models NST 0  and NT O are standard with one  speaker only  8 ohm  12 5cm  built in the rear  cover    The standard amplifier modules for the NST O            15            amp  NT 0 and NST 1  amp  NT 1 are optical sound  reproduction only    Amplifier modules for the NST 2 and NT 2 are  capable of both optical and magnetic sound re   production    Amplifier modules for the NST 3 and NT 3 are  capable of both optical playback and magnetic  record playback              Series   1 4     320   2  SERVICE PROCEDURES               Un    2 1  PRECAUTIONS      EIKI          series projectors have been designed    for the ultimate in simplicity and ease of service  and repair  Each screw is very important  and  when servicing or re assembling the projector   screws should not be omitted or carelessly lost        screws should be firmly tightened to assure  reliable projector operation after re assembly      When lubricating the projector s plastic parts     silicon oil or grease should be used  Other types  of lubricants may harm plastic parts  Avoid  using any solvents such as Trichloroethylene   whic
15. amp  adjust  too great   See Sec 5 5 B1   3  Claw protrusion incorrect 3  Check  amp  adjust   See Sec 4 1 D1   34  Excessive noise in the film  1  Upper loop too small 1  Check  1 sprocket timing  gate in  forward  with     2 Film contacting the loop   2 Check  2 sprocket and loop  good undamaged film setter roller  setter timing  See Sec 5 4 B   3  Dirty film gate 3  Clean  S 4  Loose claw 4  Tighten  5  Incorrect claw protrusion 5  Check  6  Inner guide rail binding    6  Check  amp  adjust  7 Film shoe bent  worn       7  Check  amp  replace  binding    8  Claw position incorrect 8  Check  amp  adjust     See Sec 4 1 D2   9  Weak or broken cam follower   9  Replace  spring  See Sec 4 1 B        A      See Symptom No  34  noise  in the film gate    35  Unsteady picture         Incorrect shutter blade posi  1  Check  amp  adjust  tion  See Sec 4 1 F     36  Travel ghost             SYMPTOM     MECHANICAL SYSTEM   PROBABLE CAUSE     N  Series    REMEDY       37  Excessive noise in  reverse     only    1  Claw position incorrect           Check  amp  adjust     See Sec 4 1 D2                                2  Claw angle is incorrect 2  Check  amp  adjust   See Sec 4 1 E   38  Insufficient framing 1  Claw position incorrect 1  Check  amp  adjust   See Sec 4 1 D2   2  Worn cam follower  or glid    2  Replace  ing pin   See Sec 4 1 B   39  Excessive noise when the still 1  Motor pulley misaligned 1  Adjust  picture lever is depressed 2  Shutter pulley binding on the 2  Remove  c
16. bricate the cam area with a small amount  of molybdenum disulfide grease and moisten the    felt with    a few drops of molybdenum oil      When mounting the shutter blade and hub  assembly  the hub must line up with the mating  hub in the cam tank    In the event the shutter has been   removed from the hub  See Sec  320     NOTE     4 1 F Changing Shutter Blades              28       Adjustments                                                                                                                                N  Series   4 1 D1     Fig   32                   CLAW    a  Claw protrusion can be adjusted by the screw   B  on the fulcrum collar  As the sliding pin        31  wears  the protrusion wil    increase re     quiring the adjustment of the protrusion     SCREW            SCREW  C     dicated in Fig  432  amp  33     clockwise  increases the claw    Fig   33    b  Loosen set screw  A  by 1 8 of turn as in     c  Turn screw  B   Loosening screw  B   counter     protrusion     Tightening the screw  B   clockwise  decreases    the claw protrusion        SMALL LARGE   v  NOTCH        d  Checking claw protrusion using Tool No  320    O1T  fig   34     1  Open lens holder and gate assy     2  Attach the tool between the inner guide  rail and the outer guide rail     3  With the small  V  notch toward the  outer guide rail  the claw should just  contact the tool     4  With the large          notch toward the  outer guide rail  the claw should move  freely without con
17. djustments       1  Claw Protrusion                                28  2  Claw Position And Framing                       29  E  Claw Replacement                                30  F  Changing Shutter                                        31      Still Picture                                            32  4 2 Amplifier Module      Specifications     ccelis      Ok hh oes 33  B  Power Supply Circuit                              33  C  Audio Amplifier         0                               33    D  Magnetic Record Amplifier                         34  E  Amplifier Circuit                                          34  4 3 Motor Module  Aw Specifications         RT EY REP 39  B  Motor Circuit Diagram                             39  C  Connector Wiring Code                            39  D  Silent Film Operation And 50   60Hz Conversion        40  4 4 Transformer Module  A  Transformer Circuit Diagram                        41  B  Connector Wiring Code                            41  4 5 Lens Holder And Gate Module      Film Gate Assy so aa               race            Ag pd ede be ao 43  B  Film Shoe and Lens Holder Assy                     43  314 5  GENERAL MECHANICAL SERVICING AND ADJUSTMENTS  kn 5   Take Up Arm  Ao IDescnptions cdd ARN reed E Ra 47  B                    in                  hEESO Na panes 47       TABLE OF CONTENTS          Series  Title  PAGE  5 2 Supply         A  Description        pet        NERA 48  B                      5                              
18. e  faces away from the cam tank  See Fig   44   45    Fig   44  3     7  HUB        31          MISSING  SEMI CIRCLE  i    INCORRECT        CALLY LINE UP CORRECTLY              Series   4 1 F   Fig   45    INCORRECT  NOTE  3 BLADED SHUTTERS WILL AUTOMATI              Series   4 1 G    G  Still Picture Clutch   The still picture clutch consists of the shutter  pulley  50   the shutter blade and hub  46  and the  stil  picture clutch lever  55  and tension spring   12   The friction between the shutter blade hub s  coned surface and the coned surface of the shutter  pulley is maintained by the tension spring  12   when the still picture clutch is in the up position   driving the cam tank and the rest of the projector  mechanism  In the down position the still picture  lever releases the spring tension  slightly separating  the shutter blade hub and the shutter pulley  To  adjust the still position clearance  spacer  54  may  be added or deleted  as necessary  Lubricate the  bushing with a small amount of molybdenum  disulfide grease  In    the still position the heat  filter  57  allows a single frame to be shown  without burning the film  however  the projector  is not designed to operate in this position for  extended periods of time  Long still operations will  cause premature wear to the shutter pulley bushing  and cam tank shaft    Fig   46        32        
19. ess than 7 64mm  will cause the claw to engage the film between  perforations possibly causing film damage or  unsteady pictures         pitch more than 7 67mm will cause excessive  gate noise  Typical film perforation should  have a pitch of 7 605     7 635mm  but older  films may have a smaller pitch due to shrink   age      The claw as shown in  Fig  440  has a 5  degree angle at the top tooth  and should the  claw pitch become larger than the pitch of the  perforations  this would help prevent any film  damage             n    Fig   40                Series   4 1 E3   2  Adjusting The Claw Pitch   a  Moving the fulcrum control assy  The ful   crum control assy  Fig   41  horizontally  adjusts the contact point between the cam  follower and the cam which in turn effects  the claw pitch  Moving the fulcrum control  assy  to the left decreases the pitch and   moving it to the right increases the pitch   Fig   41           CLAW LEVER  TENSION SPRING       FULCRUM  CONTROL ASSY    CAM FOLLOWER    b  The fulcrum control assy  should only be  moved slightly  Too much adjustment will  cause the claw to hit the sides of the film  perforations causing film jitter      Correct claw angle is shown in Fig   42      Under some circumstances a bent claw lever  as shown in Fig   42    and 42 C may have to  be corrected by straightening out the claw    Fig   42       e ex    3  Claw Lever Tension  See Fig   41    The claw lever tension force should be from 1 2  to 1 25kg when the spring i
20. et shoe See Sec 5 4   2   2 sprocket shoe spring weak  2  Check  amp  replace  or missing See Sec 5 5   7   18  Self threading mechanism not   1  Release bracket spring is too  1  Check  amp  adjust  released with a tug on the  weak or broken  See Sec 5 5      film 2  Release bracket pin binding 2 Lubricate pin        adjust  i linkage   See Sec 5 5 B   19  Film does not cross the auto  1  Auto take up guide not align   1  Check  amp  adjust   take up guide   ed properly   NST only  2  Film twisted  or curled 2 Stretch  or replace the film       leader                       SYMPTOM        MECHANICAL SYSTEM    PROBABLE CAUSE             Series    REMEDY       20 Pilot lamp on  motor runs    but film does not advance    1  Still picture lever engaged    2  Broken or defective motor  belt                  Release to normal run posi     tion      Check  amp  replace       3  Cam tank plate washer loose 3  Tighten  See Sec 4 1 B   4  Motor pulley loose 4  Tighten  e 5  Main drive belt broken 5  Replace  21  Film speed is too slow or too   1  Belt is installed incorrectly 1  Check  amp  re install  fast 2  Incorrect motor and shutter 2  Replace with correct pulleys    pulley combination     See Sec 4 3 D        22  Film comes out of the path    near the sound drum      Rubber pinch roller is bind   ing     Rubber roller and tension   guide out of alignment   82 sprocket shoe not seating  properly                 Remove  lubricate and adjust     See Sec 5 5 B5       Check  amp  adju
21. h Roller Petroleum 5631 21  Cam Tank Molybdenum 5628  Pivot Shaft Oil Disulfide  Rewind Lever Shaft Petroleum 5631 Grease  Oi 21 A Cam Tank Felt Molybdenum 5632  Dampening Roller Petroleum 5631 Disulfide Oil   Pivot Shaft       or Grease  5528   Tension Gear Arm Petroleum 5631 22  Rewind Gears  Shaft Silicone Oil 5629  Pivot Pin Oil  100  Tension Gear Shaft Silicone Oil 5629 23  Flywheel Hole Petroleum 5631   100          1 Sprocket Gear  amp  Petroleum 5631 24  Loop Safety Roller Silicone Oil 5629  Worm Gear Oil  100    Reverse Rubber Roller Petroleum 5631 25  Main Interlocking Bracket Silicone 5625    Shaft Oil Grease  Sound Drum Ball Bearings  Factory OB  26  Reel Guide Bracket Hole Petroleum 5631  sealed 6082            18           LUBRICANTS  amp  LUBRICATION CHARTS     ITEM    27   28     29   30     DESCRIPTION  Reel Guide Wire  Anchorage Fulcrum Pin  Shutter Pulley Bushing    Cam Tank Fulcrum Pin       Ball Bearings             SUGGESTED EIKI    LUBRICANT PART      Petroleum 5631  Oil   Molybdenum 5628  Disuifide    Grease   Petroleum 5631   Oi     Factory      6082   sealed  08 60822  08 62002  08 626   FILM PATH    4    31     32     33     2    17  PIN        19     ITEM      DESCRIPTION    Function Switch Cam  Bracket    Function Switch Shaft    Loop Safety Roller Arm  Pin   Clutch Cover             Bushing             Series    SUGGESTED EIKI  LUBRICANT PART      Molybdenum 5628          6  PIN        3 16  PIN                                                  Di
22. h will harm most plastic or painted parts       EIKI projectors require a minimum of special    tools  The most important is an ordinary ISO  Phillips screw driver set       To avoid damage to screw heads  it is important    to remember the adage    7096 push  3096 turn     It is also important to select the right size screw  driver blade  A rule of thumb is to use the largest  blade possible       To avoid possible electrical shock  always dis     connect the projector from the power source  when servicing     2 2  TOOLS AND TEST EQUIPMENT      Tools     When servicing EIKI    N    series  ordinary  150  Phillips type  screw drivers and single   bladed screw drivers should be enough    EIKI screw driver kits  P N 5615 are availa   ble      Molex extractor tool is most useful  when replacing the pins of Nylon connectors  to the transformer or motor       Special Tools     EIKI    N    series have been designed so that  no special tools are required to service the  projector  However  a common claw protru   sion gauge  Tool No  320 01T  is most  helpful for accurately setting the claw pro   trusion  Standard 16mm film may be used as  a thickness gauge where necessary   Most  film is about 0 15mm or  005  thick               Series  Tool No  320 01T    Cam Claw Protrusion Tool   see sec  4 1 D     LARGE   V  NOTCH 77          SMALL   V  NOTCH           C  Test Equipment      limited amount of test equipment is required  for routine maintenance and modular replace   ment  Howeve
23. ket  holes  thread a strip of good film  Remove  the film shoe and bracket assy  With a  standard 50mm  2     lens installed  look  through the lens  Focus and rotate the  inching knob while observing the claw  position in the film sprocket holes   Do  not turn on the lamp     Fig   37 indiciates the correct position of   the claw just before the start of the pull    down  Fig   38 indicates the position after  completion of the pull down     5  To adjust the claw position  move the ful   crum assy  horizontal and slightly vertical  until the distances a  b  c  d are equal as  indicated in Fig   37  amp  38  Upon compie   tion of the position adjustment  tighten  the fulcrum assy  screws and re check  the framing range with the projector  running    Fig   37 Fig   38   AFTER PULL DOWN           4       M     a    k    PERFORATION    1       CLAW  d                                               E  Replacing The Claw   When replacing a worn or defective claw  it is  important to mount the claw correctly before  securing the mounting screws  Incorrect claw  mounting may result in excessive film gate noise  or unsteady picture    1  Correct Claw Pitch       The International dimensions of the 16mm film          39                       0 90   0 05                         M                                      2                   F       7 95   0 10 8           a  The International dimensions of the 16mm  film are shown in Fig  439  Claw pitch is set  at 7 64     7 67mm  A pitch l
24. ket connection       2  Cooling is restricted    3  Defective lamps  or incorrect      lamp other than EIKI ELC  type   4  Excessive or fluctuating AC  line voltage       1  Lamp not seated properly   2 Foreign object in the light  path   3  Lamp not centered horizon   tally   4  Defective lamps  or incorrect  lamp other than EIKI ELC  type   5  Function switch       position   6  Slow or defective lens   7  Low AC line voltage   8  Improper shutter                            Replace lamp socket    Locate  amp  remove the cause     Check  amp  replace       Check AC line       Check  amp  reseat    Remove      Adjust knurled knob      Check  amp  replace      Switch to    high    position       Try another lens    Check AC line    Check or replace           SYMPTOM    V  SOUND SYSTEM       N    Series       47 No sound and the exciter    lamp is not on       48  No sound  exciter lamp is on    49  Poor sound or low volume                PROBABLE CAUSE REMEDY  1  Amplifier is not turned on  1  Turn on   2 9 pin amplifier plug defective 2  Check  amp  repair  or replace   3  Defective exciter lamp 3  Replace  4  Defective exciter lamp socket   4  Repair or replace  5  Exciter lamp fuse blown 5  Check  amp  replace  6  Defective amplifier module 5  Replace  or repair   See Sec 3 2   amp  4 2   7  Defective exciter lamp power   7  Locate  amp  repair  or replace   supply of the transformer  See Sec 4 4   1  Amplifier volume is too low  1  Check  amp  adjust   2  Mag Opt switch   
25. lean  inspect and  shaft lubricate   See Sec 4 1 G   40  Film transport does not stop 1  Shutter pulley seized 1  Remove cam tank  clean  amp   in still picture operation lubricate pulley   See Sec 4 1 G   2 Still picture lever shoulder   2  Tighten screws  screws loose  See Sec 4 1 G   41  Film burns when still picture   1  Still picture lever not com    1  Fully depress  lever is depressed pletely depressed  2 Heat filter misaligned or   2  Check  realign or replace     broken  42  No rewind or poor rewind 1  Take up spindle cork too dry   1  Remove and lubricate  or tight  See Sec 5 1   2  Broken or stretched or oily   2  Check and replace  supply arm belt  3  Defective rewind gears 3  Inspect  amp  replace  4  Broken rewind arm tension   4  Check  amp  replace  spring  See Sec 5 6   5  Take up pulley binding on   5  Clean  amp  lubricate  the shaft  43  Noisy rewind 1  Rewind gears not fully en  yq  Adjust the gear position  gaged   See Sec 5 6   2 Worn or defective rewind   2  Replace  gears     See Sec 4 5   44  Uneven focus 1  Dirty film gate 1  Clean  2  Film shoe binding and not 2  Check  amp  realign  completely seated  See Sec 4 5 B   3  Inner guide rail binding 3  Check   See Sec 4 5   4  Lens holder misaligned 4  Check  amp  adjust           See Sec 4 5           10          SYMPTOM    IV  LAMP CIRCUIT    PROBABLE CAUSE             Series    REMEDY       45  Lamp life is abnormally short    46  Uneven or insufficient screen  illumination          1  Poor lamp soc
26. nd shutter  blade with hub    Remove the shutter blade and hub assembly   Position the still picture clutch where it exposes  all three cam tank cover screws  45   Remove  the screws and cover plate assembly       The curved plate spring  24  fits over fulcrum    control pin  Unscrew the screw  22  and  remove curved plate spring       Unhook the claw lever spring  36   and remove    claw lever assy   28    To remove cam  10  and cam plate  9   un   screw the three set screws  11     To remove cam shaft assy   4   remove inching  knob  8  and worm gear  7     To replace the cam tank bearings  the inner  bearing is pressed on the cam tank shaft and is  replaced as part of the cam shaft with bearing  assy    Clean all the old dried molybdenum grease  from the cam tank     Fig   30       C                  FIBER WASHER    G1 080120    Fig   31       B      3 3 d    17   1       d   h A    L _ ao             20 2322   19      P 1 217   22               a  A 3       30          7    29           D     1  Claw Protrusion    Reassembly of Cam Tank By Reversing The       Above Procedure  1  Make sure curved plate spring  24  is not jammed  between the end of fulcrum pin  18  and washer     23      The cam      Worm gear  7  is mounted without any clearance z    shaft should have no end play     between the cam tank bearing     No end play is allowed for the ball bearings on  the cam shaft   5  When overhauling the cam tank  it is suggested  that the felt oil pad be replaced     Re lu
27. ng thread    1         is hitting the shutter   1  Check  amp  adjust  ing blade See Sec 5 5 B3   13  Curled film goes over the  1           setter position is too  1  Check  amp  adjust  loop setter roller   low See Sec 5 4   _ 2  Severely curled film 2  Straighten the lead end  14  Film goes under the  3 film    1   2 Film guide defective 1  Replace  guide or comes out   2   3 Film guide defective 2  Replace  3   2 and  3 film guides        not   3  Check  amp  adjust  aligned  See Sec 5 5 B4   4  Curled film 4  Straighten the lead end  15  Film will not thread over the 1  Rough surface on the lamp 1  Remove  amp  inspect the cast   sound drum house casting  restricting the ing guides  film travel  2  Rubber pinch roller not com  2  Check  amp  adjust   pletely released from the  See Sec 5 5 B5   sound drum  3  Rubber pinch roller bound   3  Check  adjust and lubricate  IAE up or out of alignment  See Sec 5 5 B5   16  Film stops at the  2 sprocket 1  Rough surface or restriction 1  Remove the lamp house cast   to the film path on the lamp ing and inspect  house casting  2   2 sprocket teeth or cover  2  Tighten and re time  plate loose  See Sec 5 4   3  Tension guide and roller  3  Check for correct film feed   assembly out of alignment ing from the sound drum to  the  2 sprocket   See Sec 5 5 B6   4   2 sprocket shoe clearance 15   4  Check and adjust  too tight Sce Sec 5 5 B7   17  Film comes out of the  2  1   2 sprocket plate is loose 1  Tighten and check timing  sprock
28. not trimmed 1  Trim and repeat threading  guide properly  2  End of film is severely curled    2 Straighten and re thread  22 or twisted   3  Very soft leader 3  Replace leader  4   1 film guide shaft loose or   4  Re align and check   bent  See Sec 5 5 B2   5  Aperture plate assembly mis    5  Re align and check  aligned  See Sec 4 5    9 Tip of the leader enters   1  Inner guide rail bent or not   1  Align or replace  underneath the inner guide aligned correctly    See Sec 4 5   rail   2  Loose guide rail screws 2  Tighten  See Sec 4 5      3 Film leader        severely   3  Trim and re thread    curled or twisted   10  Film does not thread through   1 Film leader not trimmed 1  Trim  the gate  and jams   properly      2  Leader is severely curled or   2  Trim  straighten or replace  twisted  or too soft    3         much  1 film shoe   3  Check  amp  adjust  clearance    See Sec 5 5 B1   4  Dirty and obstructed film   4  Clean  amp  check  gate    5  Insufficient side travel of the   5  Check  amp  adjust  inner guide rail  See Sec 4 5   6  Side pressure spring tension 6  Check  amp  adjust    too strong  See Sec 4 5   7 Film shoe is not retracted 7  Check  amp  adjust       during threading           See Sec 4 5        ES        SELF THREADING SYSTEM      N  Series                                        SYMPTOM PROBABLE CAUSE REMEDY  11  Chattering noise during 1         15 not completely 1  Check  amp  adjust  threading retracted See Sec 5 5   3   12  Clicking noise duri
29. olar cell      See Symptom No  57 also        57  Excessive wow  amp  flutter    un      Rubber pinch roller binding      Sound drum bearings defec     tive      Reverse rubber drive roller in      contact with the set collar        Flywheel not installed     Flywheel rubbing      the  power cord   Incorrect alignment or ten     sion of the tension guide and  roller assembly       58 Sound not stabilized soon     after starting         Weak    Insufficient fly wheel plate    spring tension    rubber pinch roller    spring              Check with another film     Check  amp  install     Check  amp  re connect properly        Adjust     See Sec 6 2       Repair      Repair or replace     See Sec 4 2  amp  3 2          Check  amp  replace    2  Check  amp  rc install    Un   US        Replace      Replace    Try a known good recording      Check  amp  realign     See Sec 6 1       Check  amp  clean      Sce Symptom No  57       un      Check       Check    Check  amp  adjust    clean  lubricate  amp   adjust  See Sec 5 5       Check  amp  replace     See Sec 5 7       Check  amp  adjust     See Sec 5 3       Check  amp  adjust     See Sec 5 5 B6        EM            Check  amp  adjust     See Sec 5 7       Check  amp  replace     See Sec 5 5 B5            14           314   1  PRINCIPLES OF OPERATIONS    1 1  INTRODUCTION    The                   series self threading projector is very  similar to the earlier    R    series model with added  improvements such as silent threading
30. r  when servicing the individual  modules such as the amplifier  the following  equipment and test films would be essential   a  Vom  Voltage Ohm meter   b  Oscilloscope    Audio AC VTVM    Wow  amp  Flutter Meter    400Hz SMPTE Test Film    3150Hz Wow  amp  Flutter SMPTE Test  Film      Multi Frequency SMPTE Test Film    7000Hz Sound Focus SMPTE Test Film  7000Hz Mag  Azimuth SMPTE Test Film  Buzz Track SMPTE Test Film     Audio Oscillator                                          gt           2 3  LUBRICANTS  amp  LUBRICATION CHARTS  TN Senn    Apply a few drops after every 500 operating hours  The items marked with   would require more frequent    lubrication  ER  CAUTION  DO NOT OVERLUBRICATE     12 AN       Fig   1  28 ig          25  24    23    12    21             11    SUGGESTED EIKI    SUGGESTED EIKI    EM   DESCRIPTION LUBRICANT PART   ITEM   DESCRIPTION LUBRICANT PART     1 Sprocket Hub Petroleum 5631 15   Take Up Pulley Cork Silicone 5625  Oil Liner Grease   2 Sprocket Hub Petroleum 5631 16  Loop Setter Shaft Petroleum 5631  Oil Cil  Duracon Guide Rollers Silicone Oil 5629 17   3 Film Guide Pivot Pin Silicone Oil 5629   100  100 E  Duracon Guide Rollers Silicone Oil 5629 18  Tension Guide Roller  amp  Silicone Oil 5629     100 Bracket Pin  100   Rubber Pinch Roller Molybdenum 5628 19     Take Up Pulley  amp  Shaft Molybdenum 5628  Disulfide Disulfide  Grease Grease  Self Thread Lever Petroleum 5631 20   Supply Arm Spindle Shaft Petroleum 5631  Oil Cil  Rubber Pinc
31. rcised to avoid electrical shock while servicing the projector                         N    Series                             N SERIES TROUBLE SHOOTING CHART  I  ELECTRICAL SYSTEM  SYMPTOM PROBABLE CAUSE REMEDY  1  Pilot lamp not on 1       power to the AC wall 1  Check outlet  outlet  2  Defective power cord 2  Check  amp  repair or replace  3  Faulty transformer or con    3  Repair or replace  nection  See Sec 4 4   4  Defective pilot lamp 4  Replace   2 Pilot lamp on  motor does  1  Defective motor module 1  Replace or repair   not run in    forward    position See Sec 4 3      1  2  Defective or open motor           2  Check  amp  repair  or replace  nections See Sec 4 3 C   3  Faulty motor capacitor 3  Check  amp  replace   See Sec 4 3   4  Switch cam not activating   4  Check  amp  adjust  micro switch  1 See Sec 8 1 A   5  Defective micro switch  1 5  Check  amp  replace         Sec 8 1 A     6  Motor thermal switch open  6  Allow motor to cool and   UL  amp  CSA type  check again    3  Pilot lamp on  motor runs 1  Defective lamp 1  Replace lamp  but the lamp does not come 2  Lamp not seated properly in 2  Check  amp  reseat  on in    forward lamp  posi  the socket  tion  4      3  Defective lamp socket 3  Repair or replace   4  Defective micro switch  4        4  Check  amp  replace   5 on the function control  See Sec 8 1 A   5  Open transformer connection   5  Check  amp  repair   See Sec 4 4 B   6  Defective transformer module   6  Replace   See Sec 4 4    4  Pilo
32. s stretched to the  maximum travel of the claw lever arm       the  tension is too weak  the cam follower may float  off the cam surface causing excessive gate noise  and an unsteady picture  On the other hand   if the spring is too strong the cam follower may  wear out prematurely or cause a slight hesitation  of the claw lever when the projector is initially  started  To obtain the correct tension it may be  necessary to replace the spring  A slight adjust   ment can be made by stretching the spring if  necessary                           30           F  Changing Shutter Blades                                        Fig   43                                     UPPER GHOST    Two  three and five bladed shutters can be mounted  on the shutter pulley  The mounting holes used to  secure the blades to the pulley allow a small adjust   ment for shutter blade timings  Incorrect shutter  timing results in what is commonly called  travel  ghost   The adjustment is accomplished using the  SMPTE test film and adjusting the blade position    LOWER GHOST    for minimum upper or lower image movement as  shown in  Fig   43   Since the adjusting screws  are only accessible with the cam tank removed   this becomes a trial and error adjustment  However   the skilled technicians can accomplish this in one    or two adjustments     When mounting the 2 blade shutter  the missing    PRIME BLADE     m semi circle on the hub must be positioned toward  one of the blades  The curved edge of the blad
33. st     See Sec 5 5 B6       Check  amp  adjust     See Sec 5 5 B7        23  Excessive take up torque in     forward        24  Excessive take up torque        reverse            Dry cork of spindle assy    2  Dirty or sticky take up arm  belt  or take up pulley       1  Reverse clutch spring too  strong       25  Take up poor or not at all in   forward            Take up arm belt broken  or  stretched  or oily belt     Defective ball bearing       Take up pulley binding    Excessive grease on take up    pulley  6  Clutch cam defective     2  3  Loose spindle shaft  4  5    M      Too loose film on the take up  reel              Ue         M      Lubricate cork     See Sec 5 1       Clean or replace      See Sec 5 1       Check  amp  adjust     See Sec 5 3       Replace or clean       Replace       Tighten screw     Clean or replace    Avoid over lubrication     Check  amp  replace  or reposi     tion clutch cam    See Sec 5 1        26  Take up poor or not at all in   reverse             supply arm belt      Reverse belt broken or oily    Clutch cam not engaging  Reverse clutch spring weak               Broken or stretched or oily                Check  amp  absorb extra film  slack           Replace     Clean or replace      Check  amp  repair  See Sec 5 3     Tighten knurled nut     See Sec 5 3           SYMPTOM     MECHANICAL SYSTEM     PROBABLE CAUSE       27  Weak back tension of the 1  supply reel in    forward         Reverse clutch cover pulley  binding and clutch cam
34. sting    Fig   29        26                 Series   3 5     Fig   28                                   Fig   29    To re install the film gate assembly  follow the  reverse procedure  It is also necessary to check  and adjust the claw travel in the aperture plate  assembly   See section 4 2 claw position and  framing                  Series    314   4  MODULE REPAIR AND ADJUSTMENTS  4 1        4     4 1  CAM TANK MODULE    A  Specifications    Revolution  24 FPS  1440 RPM 5   18 FPS  1080 RPM 6     Cam Claw Protrusion MIN  1 0mm     MAX  1 2mm    040  to  045      Claw Pitch 7 64     7 67        Tension of Claw Lever  Spring 312 11161  NOTE  Tension of Claw Lever Spring is meas   ured with a tension scale pulling on Cam Claw         1 2     1 25kg             and the Claw Lever Spring stretched to maxi  9   mum   B  Disassembly of Cam Tank  Fig  430  amp  31  10        1  Set the still picture clutch to the still position    with the shutter pulley  50  rotating freely  11     2  Unscrew  53  at the end of the cam shaft and  shutter pulley  50    3  Remove shutter pulley  50  and the plate    washer  51   Care should be taken not toloose 12     the slim washer  54  behind the plate washer   The slim washer is selected where necessary  to provide the correct clearance between the  shutter pulley  50  and the shutter blade and  hub  46         d es    While holding the shutter blade  46  slowly  raise the still picture clutch  releasing the  tension of the clutch spring  12  a
35. sulfide  Grease  Molybdenum 5628  Disulfide  Grease  Silicon 5625  Grease  Petroleum 5631  Oil  Fig   2  1  HUB      8  PIN   3  Fig   3    SELF THREADING MECHANISM   Main Interlocking Bracket          Series     LUBRICANTS  amp  LUBRICATION CHARTS     Fig   4                CAM TANK MODULE                                                                                                                                           CAM TANK MODULE  amp  FILM GATE  Fig   5                                                        20               N  Series     LUBRICANTS  amp  LUBRICATION CHARTS                                                                                      1 SPROCKET GEAR  amp  HUB  Fig   6  CLUTCH COVER PULLEY FR  12  1 34  1  I I      I I T      Fig   7  SOUND DRUM  HUB    amp  FLYWHEEL     TAKE UP PULLEY  amp  SHAFT Fig   8  LOOP SETTER  3 16 Fig   9  e e T seas xa                                         N  Series     LUBRICANTS  amp  LUBRICATION CHARTS   Fig   10 Fig   11    REWIND GEARS  amp  CONTROL  LEVER TAKE UP ARM DRIVE  PULLEY                                                                         SUPPLY ARM DRIVE PULLEY TENSION GEAR Fig   13  Fig   12                      UU                                 gt 22                  Series     LUBRICANTS  amp  LUBRICATION CHARTS                                                                                   SUPPLY ARM SPINDLE Fig   14 REVERSE RUBBER ROLLER Fig   15       SHAFT PIN           SS RR  20   
36. t lamp on  motor runs in 1  Micro switch  2 and or  3 1  Check or replace     forward    but not      defective  See Sec 8 1 A      reverse    position   tp  2  Defective motor module 2  Replace    See Sec 4 3    5  Function switch does not 1  Loose function switch knob 1  Re locate and tighten   follow the indicated sequence 2  Loose switch cam 2  Re position and tighten  3  Reverse or take up clutch 3  Check and repair       cam   s installed incorrectly        Sec Sec 5 1   amp  5 3           SYMPTOM    II  SELF THREADING SYSTEM    PROBABLE CAUSE             Series    REMEDY       6      2 film guide  or self thread  lever  does not set the mecha   nism to thread           Cam bracket is loose on the     2 film guide shaft      Main interlocking bracket is    binding or not latching with  the release bracket hook    M2         Re position and tighten     See Sec 5 5 B4       Check  amp  adjust     See Sec 5 5 B          Re insert the film correctly                7  Leader or film is dimpled in 1  Film is inserted with the   1  the first few inches by the sprocket holes opposite the    sprocket teeth sprocket teeth    2         enough clearance be    2  Check 4 adjust  tween the sprocket plate and    See Sec 5 5 B1    the film shoe    m 3  Film path and  2 sprocket   3  Check alignment of the ten   drive out of alignment sion guide rollers   See Sec 5 5 B6   4  Film leader too soft 4 Check with another film  SERRE    leader   8  Leader jams      the  1 film 1  Film leader 
37. tacting the tool    Fig   34           v  NOTCH         Hi CLAW MOVES  CLAW STOPS FREELY    e  When using another type of claw protrusion  gauge of similar specifications as  320 4 1    disregard items 2  3  4 and follow the instruc   tions associated with that gauge     2  Claw Position And Framing Adjustments   a  If the claw does not enter the center of the       film perforations  or if the framing adjustment   is insufficient  the claw position should be   adjusted  This adjustment can be either hori   zontal or vertical  To adjust the fulcrum assy   slightly loosen the two screws  C  shown in   Fig   32 4 33     1  Checking the Framing Adjustment  This is  best accomplished with the projector run   ning  showing a shop test film  With the  framing control lever in the up position   the frame bar of the film should appear as  in  Fig   35  and in the down position the  frame bar should appear as in  Fig   36               29       BEFORE PULL DOWN    LN                Series   4 1 02     2  If the conditions in step 1 are not correct   adjust the vertical or up and down posi   tion of the fulcrum assy   Fig  432  amp  33   by slightly loosening screws  C  on the  fulcrum assembly    Only a very small movement is required to   effect the framing position    NOTE     framing range as indicated can   not be reached  check for a worn  cam follower  32 of Fig   31     3  Checking Position For Correct Alignment  With The Sprocket Holes   To view      claw position in the sproc
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
MARKETINg-INNOVATION / RELATION CLIENT  User`s Manual  American Telecom ATS 4121CCL User's Manual  Classe d`eau        Copyright © All rights reserved. 
   Failed to retrieve file