Home
        SynthesisCell™ user manual
         Contents
1.    Antec  Industrieweg 12       2382 NV Zoeterwoude  The Netherlands  TM  SynthesisCell  User manual  206 0010  Edition 1  2014   ISO  9001    T  31 71 5813333   F 31715813334   info myantec com   www myantec com        Copyright   2014  Antec  The Netherlands  Contents of this publication may not be reproduced in  any form or by any means  including electronic storage and retrieval or translation into a foreign  language  without prior agreement and written consent from the copyright of the owner  The  information contained in this document is subject to change without notice     ROXY  ALEXYS  DECADE  DECADE Il  INTRO  Flexcell  ReactorCell  SenCell  VT 03   SynthesisCell  SynthesisCell  ISAAC  HyREF are trademarks of Antec  Whatman     word and  device  and Whatrnan     word only  are trademarks of Whatman International Ltd  SOLVENT  IFD    and AQUEOUS IFD    are trademarks of Arbor Technologies  Inc  Clarity    DataApex   are  trademarks of DataApex Ltd  Microsoft   and Windows    are trademarks of Microsoft Corporation   Excel is a registered trademark of the Microsoft Corporation     The software and the information provided herein is believed to be reliable  Antec shall not be  liable for errors contained herein or for incidental or consequential damages in connection with the  furnishing  performance  or use of software or this manual  All use of the software shall be entirely  at the user   s own risk     INTRODUCTION Table of contents 3    Intended use    For researc
2.   Oxidation of heteroatom  N  S  S methyltiopurine     lidocaine  phenothiazines 2 7    under basic conditions    i d    R C NH R  gt  R C  N R ZA    R R     4    Acetaminophen  amodiaquine  amsacrine   mitoxantrone 3 8         o        So    OH   gt      A NSR  NS    Dehydrogenation    Table 2  Examples of solvents used for electrochemical oxidation   Compound Mobile phase Electrode    Acetaminophen  20mM ammonium acetate in   Glassy Carbon  Magic  25  methanol Diamond    Amodiaquine  1   20mM ammonium formate  pH   Glassy Carbon  Magic  7 4 adjusted with ammonium   Diamond  hydroxide  in 50  acetonitrile    Amodiaquine  2    50  methanol Glassy Carbon  and 50  10 mM aqueous  lrinotecan 20mM ammonium formate with   Glassy Carbon  Magic  0 1  formic acid  pH 3 3  in Diamond  50  acetonitrile  on both different oxidation  profiles were obtained     Angiotensin 0 1  formic acid in 50  Magic Diamond  acetonitrile    Adenosine 20mM ammonium formate Magic Diamond   OH 7 3 adjusted with  ammonium hydroxide  in 50   acetonitrile  Toremifene 20mM ammonium formate  pH   Glassy Carbon  7 4 adjusted with ammonium  hydroxide  in 50  methanol  LYL  LWL 90 10 1  v v v  Glassy Carbon  water acetonitrile formic acid oe       References    1  A  Baumann et al   J  Chromatogr  A  121 6  3192 3198  2009     2  T  Johansson  L  Weidolf and U  Jurva  Rapid Commun  Mass Spectrom    21  2323 2331  2007     S M  van Leeuwen et al   Anal  Bioanal  Chem   382  742   750  2005     W  Lohmann and U  Kar
3.  19    Storage 21  Specifications SynthesisCell 22    Accessories SynthesisCell 23    Ordering Information SynthesisCell 23    CHAPTER 1    The SynthesisCell    Introduction    Congratulations on your purchase of the SynthesisCell     The SynthesisCell is a bulk electrochemical cell  designed for small scale  electrosynthesis of metabolites  reactive intermediates and other oxidation and  reduction products in milligram quantities  The SynthesisCell is available with  different working electrodes  WE   Tubular Reticulated Glassy Carbon  RGC      Flat Smooth Magic Diamond  MD  and Tubular Mesh Platinum  Pt   Moreover   there are two types of reference electrodes the Pd H2  HyREF  and a salt bridge  Ag AgCl electrode  SB REF   A coiled platinum wire is used as an auxiliary  electrode  AUX      For successful and efficient conversion  the parameters as the potential   composition of the solution  Supporting electrolyte  organic solvent concentration   pH   need to be optimized depending on the type of compounds or desired  reaction  In general  the samples can be oxidized in the solutions containing  supporting electrolyte at concentrations of 10mM or higher  The higher  concentration of supporting electrolyte  ca  100mM  can improve conversion    In case MS is used for analysis  the solvent also affects the ESI response   ionization suppression   In that case the solvent must be compatible with ESI MS  and to fulfill this requirement ammonium acetate  ammonium formate  formic a
4.  Fig  1                  pH  Fig  1  Potential difference between HyREF and Ag AgCl  REF versus pH   So  if an Ag AgCl REF is compared to a HyREF  the pH effect on HyREF     compared to Ag AgCl must be taken into account  The pH voltage relation is  described by     Exyrer   Eagagci   328   29 9 pH  1   Example    If an Ag AgCl REF is exchanged for a HyREF a working potential of 800 mV  vs   Ag AgCl  at pH 3  has to be changed to     Enyrer   800   328   29 9 3   560 mV  vs  HyREF     As a rule of thumb  the working potential should be chosen such that a similar  background signal  I cell  is measured with the HyREF as in using the Ag AgCl  REF     Working electrodes    Electrochemical conversion puts high demands on the working electrode material   The working electrode should be made of a  electro  chemically inert material   Furthermore  to avoid an irregular flow profile over the electrode  it should have a  very well defined surface  In the table   the working potential limits are listed of the  WE materials available for SynthesisCell  The values presented in the table are    12    SynthesisCell    user manual    only the estimation and will depend on mobile phase composition  pH  supporting  electrolyte  and the analyzed compound itself  At high positive working potentials  the water in the mobile phase electrolyses and results in decrease in metabolites  formation  In case of electrolysis of the water mobile phase the cell current  Icell   readout will display the mes
5.  needs maintenance  the cell has to be disassembled        Before disassembling the flow cell read General precautions     1  Switch off the cell and disconnect the cell cable    2  Remove the AUX and the HyREF    3  To remove the WE  Gently push the WE downwards using the white  Teflon contacts  than pull it up to remove the retaining rings  Gently rest  the WE on the bottom of the cell  Remove the vessel cap  and remove  WE from the vessel     N Use proper eye and skin protection when working with solvents     17    Maintenance Magic Diamond electrode    The Magic Diamond working electrode consists of an ultra thin crystalline  Diamond layer deposited on a carrier metal  Therefore  such electrode cannot be  polished to restore the electrode surface in case of contamination    An effective method to restore the performance is by electrochemical reactivation  of the electrode surface under acidic conditions     Activation of the electrodes using pulse mode  This activation procedure is suitable for both MD and GC electrodes     1  Refill the bottle with clean solvent containing the supporting electrolyte   e g   the same solvent used for oxidation of the sample   2  Go to the Options pull down menu and choose    Detection mode     Fig  17         ROXY Dialogue     File Edit  Options  Help   Data f Devices  Timed Events Connect  Not connected       m Pulse        Oven   Detection Mode Pulse settings  Step T   Settings e1  200 v  1000 ms  Temp f FW update wizard E2   2 00 y 211
6.  the diamond  electrode was seen within days of operation  delamination blister    formation of the  MD layer      21    Storage    If the cell is not in use  switch off the cell  wash and store with water  If not in use  for longer time  several days  weeks   wash all parts with water and store clean    and dry        Before disconnecting the cell from the detector  turn off the cell first      22 SynthesisCell    user manual    CHAPTER 4    Specifications SynthesisCell    Cell type    Cell volume    Working  electrodes  WE     Reference  electrode  RE     Auxiliary  electrode  AUX     Port plug    Electric  connections    Three electrode synthesis cell  consisting of working  electrode  WE   reference electrode  RE  and auxiliary  electrode  AUX     Up to 80 mL of sample solution in glass reaction vessel  Optional   Water jacketed reaction vessel  for cooling exothermic  reactions     Tubular Reticulated Glassy Carbon  RGC   Optional    Flat Smooth Magic Diamond  MD   Tubular Mesh Platinum  PT     Pd H2 reference electrode  HyREF  Optional   Ag AgCl reference electrode    Coiled platinum wire in glass isolation tube    Access port for sample collection  dispensing of reagents   or venting of cell    Cell cables for use with ROXY Potentiostat    CHAPTER 5    23    Accessories SynthesisCell    The Antec SynthesisCell  p n 206 4300  is shipped together with a number of parts  listed below  in shipkit   Use these part numbers when re ordering parts     Part no    206 0037    20
7. 000 ms    No sense Users Hake   E3  0 00 y 310 ms    Cell PQ   Noise Test      ts  20 ms    Cell    of OQ System A   One Off    cell a Start Run   Stop Run     Clear All Shift Del       Sequence F2    D2D System Suitability          Fig  2  Pull down menu  Options  gt  Detection Mode     3  Select Pulse mode  Fig 18  and click OK        F  Select mode   cell 1    Mode       Fig  3  Select mode window     4  Set the parameters in the cell control window  Fig  19      18 SynthesisCell    user manual    Parameter Value  E1  2V  E2  2V  E3 OV  t1 1000 ms  t2 1000 ms  t3 0 ms  Ts 20 ms        It is recommended to set the Output Range to 200 uA      5  Set Run time value to 5 min and turn on the cell    6  Go to Options pull down menu and click Start Run  Fig  20    Program will ask to save the data  The excel file will be created   And start the  acquisition        ROXY Dialogue     File Edit Options Help    Data   Celi   Devices  Events  Connect  Not connected    Pulse  Oven Acquisition     Output Pulse settings    Step T g JSk Run time   500 min Range  200uA   FS E1  2 00 Von  1000 ms    Delay   0 00 min E2  200 v  100 ms  Sr ly      Filter   off    Hz    Marker   E3 0 00 Yy t3  o ms      Cell 1     ts   20 ms  Cel B        Onc Of             Temp  C LM35    No sensor selected                   ROXY Dialogue     File Edit Help    Devices       Timed Events    Connect  Not connected      r Pulse settings      m Pulse        Oven Detection Mode      Step T         Temp J    Settin
8. 6 0300  206 0306  206 0322  206 0314    206 0304  206 0310  206 0900  206 0340    Description    Complete SynthesisCell  consisting of 80 mL reaction vessel with  Tefloncap  WE  Tubular Reticulated Glassy Carbon   RE  HyREF   and AUX electrodes  stir bar  electrode cables  etc   all parts included  for immediate use with ROXY Potentiostat     Optional   Water jacketed reaction vessel   Flat Smooth Magic Diamond  MD  working electrode  Tubular Mesh Pt  PT  working electrode   Ag AgCl reference electrode   Spare Parts   Tubular Reticulated Glassy Carbon  RGC  working electrode  Auxiliary Pt electrode in glass tube   Glass reaction vessel  80 mL    Glass coated magnet stir bar  length 8mm    
9. and  monitoring and control instruments  Nevertheless  we have taken steps to  eliminate all restricted substances from our products     Antec Leyden is an ISO 9001 2008 certified company        4    SynthesisCell    user manual    ZN          Symbols    The following pictograms are used in this user manual     Warning caution sign  It calls attention to a procedure or practice which  if not  adhered to  could result in severe injury or damage to parts or all of the equipment   Do not proceed beyond a warning sign until the indicated conditions are fully  understood and met     The attention sign signals relevant information  Read this information  as it might  be helpful     The note sign signals additional information  It provides advice or a suggestion that  may support you in using the equipment     INTRODUCTION Table of contents 5    Safety practices    Perform periodic leak checks on LC tubing and cell connections  Do not allow      flammable and or toxic solvents to accumulate  Follow a regulated  approved  waste disposal program  Never dispose of such products through the municipal  sewage system     LC equipment should be used by trained laboratory personnel only  Use     proper eye and skin protection when working with solvents  Additional  safety requirements or protection may be necessary depending on the  chemicals used in combination with this equipment  Make sure that you  understand the hazards associated with the chemicals used and take    appropriate measure
10. ation procedure event table  The syringe pump commands are ignored  as no pump is connected     In case that no significant improvement of the cell performance is seen     A  For MD electrode use the electrochemically reactivation of the  electrode surface under acidic conditions   B  For RGC electrode  replace the electrode for a new one     Activation of the electrodes using scan mode in acidic conditions    1  Disassemble the SynthesisCell as described on page 12  Keep the MD  electrode disc fixed inside the working electrode assembly    2  Wipe the electrode surface with a tissue wetted with methanol or acetone  and subsequently with demi water to clean the electrode surface  Wear  gloves  Under all circumstances try to avoid direct contact of the  electrode surface with fingers  The skin contains fatty substances which  will foul the electrode     20 SynthesisCell    user manual    3  Assemble the SynthesisCell with MD electrode as described in the  previous section    4  Prepare a solvent of 0 5 M Nitric acid  HNOs   fill and install the  synthesisCell     f Make sure that all parts that are not acid resistant are not connected in the    system during this step        5  Set the ROXY in SCAN mode with the following settings  E1    3 00 Volt   E2    3 Volt  Scan rate  50 mV s  scan cycle  continuous and range 200  uA V    6  Start scanning under acidic conditions for 1 hours    7  After scanning switch off the cell and the flow rate  replace the solvent by  HPLC grade wa
11. cid  or acetic acid can be used  It is not recommended to use acetic acid and its  derivatives with Magic Diamond electrode     The information how to acquire the MS Voltammogram and use the Scan or  DC mode for metabolite synthesis are described in the ROXY Potentiostat       User manual  210 7010  and Dialogue for ROXY user guide  210 7017      Table 1  Typical enzymatic CYP reactions that can be done electrochemically   Compound indicated with asterisk is drawn as an example  Table is adapted from  Lohmann et al  2010  The electrochemical oxidation reactions are among others   S oxidation  N dealkylation  hydroxylation and dehydrogenation     8 SynthesisCell    user manual    Enzyme catalyzed Electrochemically simulated Reference    Allylic and aliphatic hydroxylation Tetrazepam  requires high potential   O    OH     R CH2 CH3  gt   R CH CH3       Benzylic hydroxylation Metoprolol  only in low yields   2  R R  O     Desalkylation of amines Amodiaquine   lidocaine  clozapine 3 4  C Z     N   CH2 R  gt  NH   R CHO om    NH  phe  Desalkylation of ethers Metoprolol  dibutylsulphide    low yield  only under acidic conditions  2 5  R O CH2 R     gt  R OH   R CHO ae  Hydroxylation of aromatics Metoprolol  mephenytoin  diclofenac    especially for aromatics 2 6  with electron donating groups       gt       a  P  N  R R H OH  Cl  Epoxide formation Benzo a pyrene   Not mimicked electrochemically f  4 LN TT a  N ALN f   C C  gt  C C  gt     C C    or      C C  gy  a TD a i i  R i R R
12. ctrode  without reference electrode   the working potential would  continuously change due to polarisation effects at the electrodes  resulting in  highly unstable working conditions    If the working potential would be applied only over the reference electrode versus  the working electrode  without auxiliary electrode   the working   potential would be very well defined  However  the potential of a reference  electrode is only well defined if the current drawn is extremely low  pico amperes   which woud make this configuration useless for our purpose     A three electrode configuration combines the best of both configurations  The  reference electrode stabilises the working potential and the auxiliary electrode can  supply high currents  This results in the tremendous dynamic range of a three   electrode system  with sufficient yield in electrosynthesis     HyREF reference electrode    The SynthesisCell is standard equipped with a HyREF  Pd Hz2  reference  electrode  The HyREF can be used under harsh conditions and is free of  maintenance under normal use  An important characteristic of the HyREF is the  pH dependence of the reference potential     It is important to realise that if the pH of the solvent is changed  also the  optimum conversion potential changes  In such case it is advisable to re     determine the optimum potential by creating a voltammogram        11    HyREF versus Ag AgCl Reference electrode  The reference potential of an Ag AgCl and HyREF are different 
13. gs    FW update wizard    E1  2 00 y    E2   2 00 y    E3  0 00 y    ti  1000 ms  t2   1000 ms  t3   0 ms    Users Marker      PQ   Noise Test  OQ System A   One Off     I cell E Start Run    ts   20 ms          Stop Run Ctrl Q  Clear All Shift Del    Sequence F2  D2D System Suitability          Fig  5  Start the activation procedure     19    After the activation procedure turn OFF the cell and leave the flow for additional 1   2min     The activation procedure is available also as a program    Activation _pulse_rev01 evt  and can be executed automatically via events  window  Fig  22   The method can be found in My Documents and in the  subfolders  Dialogue Templates     The detailed background information about the supplied events files and relevant  Dialogue settings are provided in the Dialogue for ROXY User guide  210 7017            r     ROXY Dialogue          File Edit Tools Options Help      Data    Cell 1   Devices l Events  e   Events   Activation_pulse_rev01 evt Connect_ Not connected    time  min   parameter i Sort       0 00 Syr Flow Rate 20 ulmin Syr Pump Set the Detection mode to Add row   0 00 Syr Pump Start infuse Syr Pump Pulse mode first    0 05 Pulse E1  3  t1  3 ts 2 00   2 00  0 00  10C 00001   0 05 Cell on off on 00001   5 00 Cell on off off 00001   6 00 Syr Pump Stop Syr Pump status Output       6 10 End events      Repeats    0     M Start stop with run          C  Users Agnieszka Kraj D ocuments    Dialogue 10M  ADC  2  Max  250 n      Fig  6  Activ
14. h purposes only  While clinical applications may be shown  this  instrument is not tested by the manufacturer to comply with the In Vitro  Diagnostics Directive     WEEE directive    Antec Leyden is a Business to Business producer of analytical analysis equipment  which fall under WEEE Annex IA categories 8 and 9  includes medical devices  and monitoring and control instruments   All equipment of Antec Leyden which are  subjected to the WEEE directive  shipped after August 13  2005  are labelled with  the    crossed out wheelie bin           The symbol on the product indicates that the product must not be disposed  as unsorted municipality waste     Collection  amp  recycling information  business to business   Antec Leyden offers the possibility for disposal and recycling of their  instrument at an appropriate recycling facility if requested  there may be    costs involved with this service   Please contact Antec Leyden for more  information about this service and to register the return and disposal of end   of life instruments  info myantec com   To assure hygienic  amp  personal  safety all instrument must be returned with a signed decontamination form  which is available on the website        Shipping address for end of life products     Antec Leyden  Industrieweg 12  2382NV Zoeterwoude  The Netherlands    ROHS directive   Our instruments are currently exempt from the RoHS directive because they fall  under WEEE Annex IA categories 8 and 9  which includes medical devices 
15. isCell with RGC WE and HyREF  1  WE contact  2   AUX contact  3  HyREF contact  4  Additional port for gas purging or  sampling  5  SynthesisCell cap  6  RGC WE   7  Glass reaction vessel     This hardware should be used by trained laboratory personnel only  Use  proper eye and skin protection when working with solvents  Additional  safety requirements or protection may be necessary depending on the    chemicals used in combination with this equipment  Make sure that you  understand the hazards associated with the chemicals used and take  appropriate measures with regards to safety and protection        The SynthesisCell is delivered pre assembled however some protectors for safe  transportation must be removed before use     14 SynthesisCell    user manual    Never switch ON the flow cell when     the cell cable is not correctly connected    the cell is not filled with buffer electrolyte    because substantial damage to the working electrode or electronics may  occur        Preparations    1  Take the SynthesisCell from the shipping box and remove the protecting  foam parts    2  Prepare a solution containing the substance s  and a supporting  electrolyte  ammonium formate  formic acid  or acetic acid  0 1   1  in  water     3  Download and install the latest version of Dialogue software   http   www myantec com support documents and downloads download   software          Picture 2  A  RGC WE fixed in the vessel cap  B  WE fixed in the glass reaction  vessel  C  AUX in the f
16. pending  on the substance s  and the applied potential  the cell current  I Cell  can  go up to 2     20 mA    Monitor the formation of product by taking aliquots from the vial through  the sampling port    A 0 1 mmol L  1 5 mg  MOPEG solution is for 90  converted in about 40  min under optimized conditions    When finished  switch off the cell and collect the solution in another vial   When mixing of electrode solutions is not allowed  first take out the AUX  without mixing     16 SynthesisCell    user manual  CHAPTER 3    Maintenance    Working electrode maintenance    Activation of the working electrode is necessary if the electrode surface has been  electrochemically changed  This may be due to fouling by oxidation  reduction   reaction products  Excessively high currents also may change the electrode  surface  This is noticed by a strongly decreased sensitivity after prolonged use     Decreased cell performance    Several actions can be taken at decreased flow cell performance  Take the next  step only if the previous was not successful     1  Rinse the electrodes with clean solvent  water with ca 50  organic solvent    Clean  wipe  the REF and AUX with a wetted tissue with solvent    2  Electrochemical activation of the GC and MD electrodes using pulse mode   see separate section of this manual     3  The  re activation procedure for Diamond electrodes using scan mode  see  separate section of this manual      Disassembly of the SynthesisCell    If the working electrode
17. ritted glass tube  D  AUX assembled in the vessel cap     Assembling the SynthesisCell    4  Fill the glass reaction vessel with a solution containing electrolyte s  and  substance s   The maximum allowed volume is 80mL  Put the stirrer in the  vessel    5  Place the vessel cap on the WE and fix the WE position  height  with the  retaining rings as indicated in the picture 2A    6  Place the WE in the vessel  Picture 2B   Adjust the position of WE if  needed by careful pushing or pulling the white Teflon contacts  The RGC    Operation    11     12     13     14     15     15    WE is fragile and must be handled with care  Keep the wires connecting  the WE with the Teflon tubes dry    Place the AUX electrode in the fritted glass tube  Picture 2C  and mount it  in the vessel cap  Picture 2D   The AUX tube should be filled with mobile  phase which takes a minute  The fritted glass tube keeps the AUX  separated from the WE  reaction products on both electrodes are not  mixed  An additional tube without frit is available for cases where the  mixing of products is allowed    Place the HyREF in the vessel cap  Picture 2D      Place the SynthesisCell on the magnetic stirrer and adjust the rotation       Connect the cell cable as shown in the picture 4  Red  WE  blue  AUX and    black  REF    Set the measurement conditions using Dialogue  The user manual   210 7017     Dialogue for ROXY  is available on our website    The reaction is started as soon as the SynthesisCell is started  De
18. s with regards to safety and protection     Use of this product outside the scope of this guide may present     a hazard and can lead to personal injury        Spare parts and service availability    Manufacturer provides operational spare parts of the instrument and current  accessories for a period of five years after shipment of the final production run of  the instrument  Spare parts will be available after this five years period on an    as  available    basis     Manufacturer provides a variety of services to support her customers after  warranty expiration  Repair service can be provided on a time and material basis   Contact your local supplier for servicing  Technical support and training can be  provided by qualified chemists on both contractual and as needed basis     6 SynthesisCell    user manual    Table of contents    Safety practices 5    Spare parts and service availability 5  Table of contents 6    The SynthesisCell 7  Introduction 7  References 9  Three electrode configuration 10  HyREF reference electrode 10  HyREF versus Ag AgCl Reference electrode 1 1  Example 11    Working electrodes 11    Installation 13  Unpacking 13  Preparations 14  Assembling the SynthesisCell 14  Operation 15    Maintenance 16  Working electrode maintenance 16  Decreased cell performance 16  Disassembly of the SynthesisCell 16  Maintenance Magic Diamond electrode 17  Activation of the electrodes using pulse mode 17  Activation of the electrodes using scan mode in acidic conditions
19. sage    overload    and auxiliary potential  Eaux  will have  the extreme value   9 9V     Furthermore  in case of EC MS the electrospray signal will become very unstable  because of gas formed in the cell and loss of the signal can be observed  When  such a phenomenon is observed it is recommended to adjust the potential to a  lower value     Table 3  Working potential limits for electrodes  WE  used with SynthesisCell     WE material potential limits vs  HyREF  V        Glassy carbon  2 5V  1 5V  Magic diamond    3 5V  2 5V    The MD electrode consisting of an ultra thin film of doped diamond material  deposited on a carrier  The special properties of doped diamond electrodes  such  as a wide potential window  their inertness and excellent response stability  makes  them well suited for electrochemical conversion of a wide variety electro analytical  application  The MD electrode is perfect choice if the oxidation of the analyte  requires high potential in aqueous electrolytes     13    CHAPTER 2    Installation    Unpacking    Inspect the transport box for possible damage as it arrives  Immediately inform the  transport company in case of damage  otherwise she may not accept any  responsibility  Contact your supplier in case of damage or if not all marked items  on the checklist are included    Prior to shipment  your SynthesisCell has been thoroughly inspected and tested to  meet the highest possible demands  The results of all tests are included        Picture 1  Synthes
20. st  Anal  Chem   79  6831   6839  2007     K G  Madsen et al   Chem  Res  Toxicol   21  1107   1119  2008     S M  van Leeuwen  H  Hayen and U  Karst  Anal  Bioanal  Chem   378   917 925  2004     B  Blankert et al   Electroanalysis  17  1501   1510  2005     8  W  Lohmann and U  Karst  Anal  Bioanal  Chem   386  1701   1708  2006      oF i e w           10 SynthesisCell    user manual    Three electrode configuration    In the SynthesisCell a three electrode configuration is used  Error  Reference  source not found    The working potential is set between the working electrode  and the auxiliary electrode  AUX   The auxiliary electrode is kept at a precisely  defined reference electrode potential by means of the so called voltage clamp   This is an electronic feedback circuit that compensates for polarisation effects at  the electrodes     At the working electrode  which is kept at virtual ground  the electrochemical  reaction takes place  i e  electrons are transferred at the working electrode  This  results in an electrical current to the I E converter  which is a special type of  operational amplifier  The output voltage can be measured by an integrator or  recorder     Essentially  for the oxidation or reduction reaction it would be sufficient to use only  two electrodes  However  the three electrode configuration has several  advantages over a two electrode configuration    If the working potential would be applied only over an auxiliary electrode versus  the working ele
21. ter and flush  wash  the cell to remove the acidic solution    8  You can start up the synthesis again     In the case there is no significant sensitivity improvement observed  repeat step 1  to 10 and increase the total scan time  step 7     One can also try to perform the activation procedure under the mobile phase  conditions of the application  Furthermore  in literature anodic treatments are  reported  for example the application of a static potential of   3 Volt vs  Ag AgCl for  5     10 minutes  to restore the electrode response  Such procedure could be  executed as an alternative  if the above mentioned procedure does not lead to  satisfactory results     Electrochemical reactivation procedure for MD electrode is part of the ROXY  firmware and available via display panel  Diag Activate   The scan settings and  the duration are as described in points 6 and 7  and only    push button    action is  needed to execute it  See ROXY Potentiostat user manual  210 7010       Replace the Magic Diamond electrode disc when no improvement is seen after  repeated reactivation conditioning attempts     Chemical compatibility  The MD electrode exhibits an excellent inertness and can  be used with a large variety of mobile phase and chemicals  However it has been  observed that the MD electrode operational lifetime is strongly reduced when  exposed to fluorinated acids  such as tri fluoroacetic acid  Even at relatively low  concentrations  2  in aqueous solution  significant damage of
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Modello "L-1 "  Macally PHR-100SU Computer Drive User Manual  607508_5540-5560 Operator & Parts Manual  Manuale d`uso  KOHLER K-R37051-BN Instructions / Assembly  Mantis® Deluxe XP Tiller Operation  [U4.92.04] Procédure TEST_FICHIER  X-VA-3809-3819-eng Cover.pmd  Fiesta EZT34545-P310 User's Manual  Samsung Monitor TV de 28 inchi pentru fanii sporturilor Manual de utilizare    Copyright © All rights reserved. 
   Failed to retrieve file