        Service Manual for Digital Manometer DTG-35
         Contents
1.    Service Manual  for  Digital Manometer  DTG 35    DTG 35    1 Instructions       Please read this service manual carefully before unpacking and setting the unit  for operation  and follow the instructions precisely as described herein    These devices may only be installed  used and maintained by skilled personnel  who are familiar with this service manual and can observe applicable regulations  regarding industrial safety and accident prevention        2  Contents  1    Instructions oo 2  2  GORE NES sierici 2  3    Regulation usage NEN alka RENT 3  4  PrinciplecofOperation  3 allea 3  5  IE CANO CK UD sata id alti 3  6  Mechanical Connection                       ennen ennen en nner 4  7  Electrical Connection                         erre nie iene iene 5  8  Function Keys ee iaia liano 6  9  Adjustments ice leda  6  10  GonroL FUNCtionsani aa 7  11  Technical Dates iaia 11  2  Maintenance salata a aaa nr enkes 11  Toa DIMENSIONS  nananana RR ES 12  14  Ordering C006   ciilirirai iene 13  15  Declaration of Compliance    14    Manufactured and Marketed by     DRUCK  amp  TEMPERATUR Leitenberger GmbH  Bahnhofstr  33   72138 Kirchentellinsfurt   Tel   49 7121 90920 0   Fax  49 7121 90920 99   E Mail  DT Export Leitenberger de  www druck temperatur de   www LR Cal com    Page 2 DTG 35 manual    DTG 35    3  Regulation usage       These units of type DTG 35 serve to measure  monitor and remote transfer of  pressure dependent operational processes in machines and systems     T
2.  input EH OOO  Password    DHD set on 0005  factory setting     Cancel P  Access to further Menu Options    Automatic null point setting  E On confirmation  automatic  la  amp  vi P  null point adjustment succeeds  E Cae    Password PSEL o le UH    Lastseting  example   y la   Sl Passwod change  P           Return to Display    DTG 35 manual Page 7    DTG 35    10 2 Unit with peak value detector    and unit with peak value detector   analogue output      P   1xP   Unit Switched on    come gi    2xP   Unit switched off    Peak value memory Last Setting   Register max  value  Register min  value  Peak value detector Off  Select    Confirm selection and goto       next Menu Option    Password input   PA   a HEHHE Password  a Set on 0005  Factory setting     Cancel P  Access to further Menu Options    Automatic null point setting  zero  On confirmation  automatic    null point setting succeeds    Password Last setting  example     Password change    Return to Display       Page 8 DTG 35 manual    DTG 35       10 3 Unit with switching output  relay output     cons TRI    2xP   Unit Switched off    Password  Set on 0005  factory setting     Password input    Access to further Menu Options    Switching point Last switching point setting    Set Sw  point on 07 50  example     Hysteresis Last setting hysteresis    Example   0 02  minus 2 Digit     Automatic null point adjustment  zero  On confirmation  automatic    null point setting succeeds    Password Last setting  example     Pass
3. A wrong connection can result in the destruction of device s e   lectronics                             Cable   With relay output     with analogue output    No    1   2   Contact output  Relay in    3 GND  OV Reference    4 Analogue output 0 2Vpc   5  Contact output  Relay out                 Plug M12x1 5 core cable       View of plug contacts cable wires with numbers    DTG 35 manual    Page 5    DTG 35    8  Function Keys       For the selection of menu options  the following functions keys are available     Y Next Menu Option   M Previous Menu Option  P 1 x operate Switch on  P 2 x operate Switch off    P amp  Jump to Function    Adjustment and Function     Y Value adjustment upwards   a Value adjustment downwards   P Confirm input for next Menu Option   MAV Reject input  Return to Menu Option  9  Adjustments       Possible adjustments of the device       Zero point    Password  factory preset  0005     Peak value memory  option     Relay and Hysteresis  option    factory preset  switching point on 50  of measuring range     A VIN    Factory presets       Battery symbol on  voltage under 7 V     Switch off time  default  0   inactive      Conversion rate  default  5 measurements per second      Analogue output  linear  within measuring range  option     ROD        Page 6 DTG 35 manual    DTG 35    10  Control Functions       10 1 Standard version DTG 35    and unit with analogue output    P  1xP   Unit Switch on             LCD Display P      2xP Unit Switch off  Password
4. ce       We  DRUCK  amp  TEMPERATUR Leitenberger GmbH  Germany  decla   re  that the product    Digital Manometer Type  DTG 35  complies with the standards given below     EN 50081 1 2 1994 03  Electromagnetic compatibility   Generic emission standard    EN 50082 2 1996 02  Electromagnetic compatibility   Generic emission standard    EN 61010 1 1994 03    Safety requirements for electrical measurement  control  and laboratory use    Also  following EWG guidelines are fulfilled     VAI    H  Peters M  Wenzel    Signature     Date  29 11 00    Page 14 DTG 35 manual    
5. hese devices are equipped as follows     e 4 digit LCD display   e Three Programming keys   e Process connection  St  Steel    e Power Supply via 9V block battery  e Limit relays  option    e Peak value memory  option    e Analogue output  option     4  Principle of Operation       A ceramic sensor picks up the pressure  to be measured  and sends the signal  to the display via evaluation electronics  Parallel to this  an output is provided as  analogue signal for the purpose of remote transfer of measured pressure    A relay output is also made available     5  Unit Check up       These devices are checked before despatch and sent away in perfect condition   Should the damage to a device be visible  we recommend a thorough inspection  of the delivery packing  In case of damage  please inform your parcel service     forwarding agent immediately  since they are responsible for damages during  transit     Scope of delivery    Standard scope of delivery applies on   e Transmitter with display   e Service manual    DTG 35 manual Page 3    DTG 35    6  Mechanical Connection       Page 4    Before installation     e Ensure that the max  pressure in your system is within the prescribed range of  digital manometer  The measuring range can be read from the type label of    device     e Make sure that the permitted max  operational temperature of device is not    exceeded     e Ascertain that the electrical power supply to the unit is in agreement with the    operational data of the uni
6. t     e Please confirm that there are no parts of packing material present inside the    unit     Installation     e Ensure that the piping has no pressure inside   e The digital manometer should be mounted just like a mechanical manometer   e With standard thread connection  sealing is achieved by means of a suitable    gasket  flat seal or seal ring  DIN 16    258      e While screwing in the device  rotate the hexagonal screw  SW 24  and not the    housing     e If possible  please check after the mechanical installation  whether the con     nection joint is adequately sealed or        SW 24    a        poca          ee                                              Flat seal ring  acc  to  DIN 16 258    not     G1 4                     10  13 0 5  18       T                SY                   __05 5 0 2       DTG 35 manual    DTG 35    7 Electrical Connection          Attention  Please ensure that the voltage of your power supply corresponds  with the admissible voltage level for this device        e Make sure that the power supply lines are not active    only with option  limit contacts or analogue output    e Open the battery enclosure on the back side of the unit and connect the 9V  block battery with the connection plug    e Place the 9V block battery in the enclosure and close it with the lid    e Terminate the connection wires on the plug  cable   as shown in the illustrati   on below     e Power supply conductor  area of cross section   min  0 5 mm         Warning  
7. times  Dia 90 min  Only factory adjustable    auto off  0   inactive  not recommended with analogue or  switching output   Zero point correction   lt t25   Parts in contact with medium  St Steel 1 4571  Ceramic  NBR  Connection  G    4  Manometer   Ya  NPT  Option   Medium temperature   30    85   C  Ambient temperature  0    60  C  Storage temperature   30    80  C  Permissible relative humidity    lt 90   not condensing  Protection cat   IP 65    Limit value relay  Option   N O   bistable  arbitrary adjustable  adjustable hysteresis    Max  switching power  30 V AC DC  2A  Analogue output  Option   0 2 VDC   Load   gt  100 kQ   Electrical connection Plug M12 x 1 or 0 5m cable     for Relay or analogue output   Peak value memory  Option   Min or Max value   Resetting via keypad    12  Maintenance       In case  the medium to be measured is not polluted  the unit is maintenance free     DTG 35 manual Page 11    DTG 35    13  Dimensions                                     Optional  PG7  LCD                   O74 14 51                         I      M12 x1  Se  I  I     LY I                                     24                                     Rotatable  Display                      89     4                                                                                                                                     G1 4A  a    38    Page12 O DTG manual    DTG 35    14  Ordering code       see datasheet    DTG 35 manual Page 13    DTG 35    15  Declaration of Complian
8. word change    Cancel       Return to Display    DTG 35 manual Page 9       DTG 35    10 4 Unit with switching  relay  output   peak value detector      P   1xP   Unit Switched on    corse  234   2xP   Unit Switched off  EJ  Peak value memory P    _oFF   Last setting    Regidter max  value    res   vl min Register min  value    o F F Peak value detector Off  GT TE sa  abbrechen Confirm selection and goto    next Menu Option    Password  Set on 0005  factory setting     Password input PA S 5 P     abbrechen         Access to further Menu Options    Switching point Last switching point setting    Adjust switching point    Hysteresis Last setting hysteresis    Example   0 03  3 Digit     Automatic null point adjustment  zero  On confirmation  automatic    null point setting succeeds    Password Last setting  example     Password change       Return to Display    Page 10 DTG 35 manual    DTG 35    11  Technical Data       Nominal size  74 mm   Accuracy class 0 5   Display  4 digit LCD  digit height 12 7 mm   Measuring range   1   0  0   1  0   1 6  0   2 5  0   4  0   6  0   10   0   16  0   25  0   40  0   60  0   100  0   160   0   250  0   400   Overload range  3 x Pn  upto 40 bar     2 x Py  60 bis 250 bar   1 5 x Py  ab 250 bar    Power supply  9 VDC  block battery    Service life at conversion rate 5 s  5000 h  block battery 600mAh    10000 h  Lithium blockbattery 1200 mAh     Conversion rate  5 per Sec   standard    1 to 10 per sec  Factory adjusted   Automatic switch off 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Sans titre 1 - Pointe Ardennes Parapente  Samsung WB100 Uživatelská přiručka  EZ28 - Minipelle Zéro Déport 2,8T (Wacker Neuson)  American Standard CHAMPION 2585.000ST User's Manual  Worldwide Lighting W83132C6 Installation Guide  90ー2 XRS System  まちの話題 赤ちゃんからのピアノコンサート (6月9日)宣  Esse 900 & 905 W35 Users Manual  Turbo-V 2300 TwisTorr Pumps  Samsung SEK-2000 Manual de Usuario    Copyright © All rights reserved. 