Home
OPERATING INSTRUCTIONS DENTAL UNIT
Contents
1. AC230V SOHZ 1 6A Air main pressure 0 45 to 0 SMPa Water main pressure 0 1 to 0 2MPa Mass Chair Mount Type 94 7 kg Without Dental Light Pedestal Type 111 7 kg Without Dental Light Dental light AL 720S AL 720M AL 520 900 Dental Light Type 920 Working environment Temperature Operating 5 C 40 C Storage 10 C 50 C Humidity 10 80 Atmospheric pressure 600hPa 1060hPa Service Life 10 Years Classification of foot controller IPX1 applicable standard IEC60529 Protection class against electric shock Class I equipment Applied parts Type B applied parts Handpiece for unit List of compatible handpieces Equipment that is not suitable for use in air flammable anesthetic gas oxygen or nitrous oxide 3 OPERATING INSTRUCTIONS 3 1 DOCTOR TABLE SECTION Turn off the master switch after daily operation and for long term interval 1 Master Switch Fig 3 1 Fig 3 2 amp Fig 3 3 Turn on the master switch located under the right hand side of the d
2. NCAUTION Refer to handpiece manufacturers operating instructions 14 3 2 CUSPIDOR UNIT SECTION Dental Light switch for E Type Cupfiller Switch Bowl Flush Switch gt Chair Auto Mode Switches Assistant Side Control Panel Dental Light Mode Selection Switch 105000 900 Dental Light Type 920 Water Heater Switch ON OFF Service Water Control Knob Service Air Service Water Outlet Outlet Cuspidor Unit Control Panel Fig 3 22 Assistant Side Control Panel and Cuspidor Unit Control Panel 1 Assistant Side Control Panel Fig 3 22 Cupfiller switch bowl flush switch dental light switch and chair auto mode switches are located on the assistant side control panel Refer to 3 1 2 2 OAO 41 2 Cuspidor Unit Control Panel Fig 3 22 A Dental Light Mode Selection Switch IO 5000 900 Dental Light Type 920 Dental light can be operated ON OFF either by the touchless switch located on the light head or by the manual switch on the cuspidor unit control panel To operate by the touchless switch Set the switch lever to Left side SENSOR Side To operate manually Set the switch lever to Right side Manual Side Set the switch lever to Centre for OFF B Water Heater Switch Fig 3 22 Water heater switch is located on the cuspidor unit control panel Turn on the water heater switch and the cupfiller water will warm up
3. 21 6 MAINTANANCE AND INSPECTION 6 1 Guide for daily maintenance and inspection Maintenance and inspection by user Management of maintenance and inspection of medical equipment should be implemented by the user medical institution In case the user does not implement such management it is permitted that such management is outsourced to a qualified entity such as a medical equipment repair company e For safe use of this product it is necessary that inspection should be conducted in the specified frequency on the tems described below No tem Frequency nspection method Influence if inspection Maintenance required in case and diagnosis not conducted of nonconformity ie Make sure the chair start movement stops by any of the following actions when foot controller pedal is depressed During chair auto movement depress of any operation switch While setup is in progress with function switch on the doctor membrane switch panel When the spittoon bowl is turned to patient side Pedestal type Before Leakage of water and air start shall not be observed around the product Cupfiller Before When a paper cup is placed start on the cupfiller the cup shall be detected and cupfilling shall be executed Malfunction may arise if the cup is of another material grade such as stainless steel and plastics or if the paper cup is of dark color or pattern Unexpected personal Contact your dealer or our injury and
4. C Service Water Outlet Fig 3 22 The service water outlet provides a quick connection for water D Service Water Outlet Control Knob Fig 3 22 The water volume from the service water outlet can be adjusted by the service water outlet control knob Turning the knob counterclockwise will increase the flow volume and turning clockwise will decrease E Service Air Outlet Optional Fig 3 22 Service air outlet provides a quick connection for air 15 3 Cuspidor Unit Body A Assistant s Syringe Flow Control Knobs Fig 3 23 Assistant s syringe flow control knobs are located in the cuspidor unit body The yellow capped knob is to adjust the assistant s syringe air flow volume and the blue capped knob is to adjust water flow volume Cupfiller Flow Control Knob B Cupfiller Flow Control Knob Fig 3 23 Bowl Flush Flow Control Knob Assistant s Syringe Cupfiller flow volume can be controlled by the Water Flow Control Knobs Assistant s Syringe cupfiller flow control knob Pinch valve system Air Flow Control Knobs Loosen the lock nut and adjust cupfiller water S Fig 3 23 Flow Control Knobs in flow volume by turning the knob Cuspidor Unit Body Tighten the lock nut after adjustment C Bowl Flush Flow Control Knob Fig 3 23 Bowl flush flow volume can be controlled by the bow flush flow control knob Pinch valve system Loosen the lock nut and adjust bowl flush water flow volume by turnin
5. gt 95 dip in U gt 95 dip in Ur for 5s for 5s Power frequency 3 A m 3 A m Power frequency magnetic fields 50 60 Hz should be at levels characteristic of a magnetic field typical location in a typical commercial IEC 61000 4 8 or hospital environment NOTE Uy is the a c mains voltage prior to applications of the test level 25 Guidance and manufacture s declaration electromagnetic immunity The CLESTA II Unit is intended for use in the electromagnetic environment specified below The customer or the user of the CLESTA II Unit should assure that it is used in such an environment Immunity test IEC 60601 test level ee Electromagnetic environment guidance Portable and mobile RF communications equipment should be used no closer to any part of the CLESTA II Unit including cables than the recommended separa tion distance calculated from the equation applications to the Frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms 3 Vrms d 1 2VP IEC 61000 4 6 150 kHz to 80 MHz outside ISM bands Radiated RF 3V m 3 V m d 1 2VP 80 MHz to 800 MHz IEC 61000 4 3 80 MHz to 2 5 GHz d 2 3VP 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site
6. Balance Arm Air Brake Fig 3 18 Balance arm air brake button is located on the handle Balance Arm When the master switch is ON the balance arm is locked Grasp the handle and press the air brake button to adjust the table height Release the air brake button at the desired table position the balance arm is locked Note Do not load over 3 kg on the table Air Brake Button Fig 3 18 Balance Arm Air Brake 8 Dental Size Film Viewer Fig 3 19 Film viewer ON OFF switch is located right side of the film viewer Press the switch the film viewer turns on A Press again the film viewer turns off f Film Viewer ON OFF Switch Fig 3 19 Dental Size Film Viewer 9 Panorama Size Film Viewer Optional Fig 3 20 Film viewer ON OFF switch and film viewer dimmer Film Viewer are located on the right side of the film viewer ON OFF Switch Film viewer brightness can be adjusted by the film viewer mi 4 ON i OFF intensity control wheel Bright Dark Film Viewer Dimmer Fig 3 20 Panorama Size Film Viewer Holder Support Arm Handpiece Holder Fig 3 21 A CAUTION Do not adjust the holder support arm and handpiece holder Because the angle adjustment of the holder is fixed at the g Handpiece Holder point of installation the holder support arm will be Holder Support Arm damaged in case it is moved too hard Fig 3 21 Holder Support Arm Handpiece Holder Handpieces
7. Device Fig 3 9 Safety Lock Indicator 8 Coolant Water ON OFF Switch Fig 3 10 E When a handpiece is picked up and this switch is pressed both LED A air and LED W water lights up the coolant water and air comes out from the handpiece In case of air motor or air turbine switching between spray 6 both of LED A and LED W are lit and OFF occurs when this switch is pressed In case of electric scaler switching between water only LED W Fig 3 10 Coolant Water ON OFF Switch is lit and OFF occurs when this switch is pressed regardless of the mode In case of micromotor either the 2 mode or the 4 mode can be selected by mode select setup When this switch is pressed in the 2 mode setup switching between spray and OFF occurs In case of 4 mode setup switching occurs in the sequence indicated below each time when this switch is pressed Spray to Water only to Air only to OFF As for the mode setting Please refer to 3 1 3 h Electric Motor Speed Set Switch Optional Fig 3 11 E Two different modes limit mode and preset mode are available as micromotor rotation speed modes Pressing this switch each time changes the speed mode Limit speed gt SET1 gt SET2 gt SET3 gt Limit speed The indicator indicates the selected mode Fig 3 11 Electric Motor Speed Set Switch 1 Switching to limit rotation speed limit mode Pick up the micromotor from the holder and press this switch t
8. E Momentarily press the bowl flush switch cB water flushes for 6 seconds and stops automatically Timer Mode Press the bowl flush switch for 2 seconds water flushes continuously Continuous mode O While the bowl flush is working by momentarily pressing the bowl flush switch the bowl flush will stop Fig o gt Bowl Pusti Switeh Note 2 Clesta II unit can be set to timer mode standard setting and continuous mode for bowl flush The bowl flush water volume can be adjusted by the bowl flush flow control knob located inside the cuspidor body Refer to 3 2 3 4 The Light Pack Switch Optional Fig 3 6 A E Pick up the fiber optic handpiece from the holder momentarily press the light pack switch the indicator illuminates in green and the fiber optic power turns on Sn o To switch off the light pack simply press the light pack switch again Fig 3 6 Light Pack Switch 7 Chair Manual Control Switches Fig 3 7 A E a Seat Lifting Press the switch until the seat A E is lifted up to the desired position b Seat Lowering Press the switch until the seat is lowered to the desired position c Backrest Reclining Press the Q switch until the backrest is reclined to the desired position d Backrest Raising Press the switch unti
9. Excessive handpiece oil or the like shall not be attached to the motor section Troubles such as injury in patient s oral cavity and equipment failure may arise The bar will not normally work and troubles may arise The tip will not normally work and troubles may arise Troubles may arise if the nut comes off The motor section will not work normally and troubles may arise 22 Control the water flow in accor dance with Control of compo nents described in the instruction manual If any other abnormality arises refer to the instruction manual attached to individual equipment Contact your dealer or our office if recovery is not achieved If abnormality such as flaw and deformation is found on the bar replace the bar in accordance with the instruction manual attached to individual equipment If the tip was worn or deformed replace the tip in accordance with the instruction manual attached to the scaler Contact your dealer or our office if any other trouble arises Turn and positively retighten the nut that fixes the nozzle Execute care in accordance with the instruction manual attached to individual micromotor ol Item Frequency nspection method Influence if inspection Maintenance required in case and diagnosis not conducted of nonconformity Care Vacuum and saliva ejector handpiece 0 Care Cuspidor section 11 Care Solid collector 2 Care Exterior After clos
10. Switch Decrease Switch Increase Switch Store Switch 0 Switch 1 Switch 2 Switch LP Switch Four preset time can be set 0 1 2 LP Momentarily press function switch and set the time by pressing decrease switch and increase switch Press the store switch then press the 0 1 2 or LP switch to store in memory 3 Preset time operation 0 6 Q Function Switch 0 Switch 1 Switch 2 Switch LP Switch Start Switch Electric Motor Speed Setting Switch Press the function switch then press the 0 1 2 or LP switch to choose desired preset number Press start switch to start timer 4 Cancel the timer during time countdown Function Switch Start Switch Electric Motor Speed Setting Switch Momentarily press function switch then press start switch to cancel timer b Group Selection Mode The group selection mode is a convenient function for the dental clinic where a multiple up to four groups of dentists work with one unit The following functions can be set for each group Preset position for chair Preset rotation speed for micro motor To set the group 1 Momentarily press the function switch twice on main control panel and the function indicator will indicate group number 2 Momentarily press the 0 1 2 or LP switch on main control panel to set one of 4 group 0 Groupl 1 Group2 2 Group3 LP Group4 Function Switch 0 Switch 1 Switch 2 Switch LP Switch 10 c Flush out system
11. accordance with Method for care described in the instruction manual Vacuum suction will become weak Discoloration and deterioration to the exterior and corrosion and rusting to metallic components may arise Execute wiping in accordance with Method for care described in the instruction manual Contact your dealer or our office if the main switch cannot be turned off or if the main valve cannot be closed Product failure and troubles may arise Contact your dealer or our office if any abnormality arises The product will not normally work and troubles may arise Drain the water from the air filter in accordance with Method for care described in the instruction manual Drain the water from the air filter drain valve Contact your dealer or our office if the set pressure is abnormally high or low The product will not normally work and troubles may arise Injury caused by falling of goods located on the table and other troubles may arise Contact your dealer or our office if any abnormality arises Normal output will not be Discharge the oil in accordance produced due to inferior with Method for care described handpiece exhaust in the instruction manual 23 6 2 Guide for Periodical Check up Some parts and components of the products are degraded or deteriorated depending on the frequency of use Annual check up and maintenance as well as replacement of consumable parts are
12. amber and green LEDs Fig 3 12 Electric Motor Direction Indicator in green Forward Rotation Control Switch Indicator in amber Reverse Rotation Note Do not change the electric motor direction while the motor is running When the electric motor with setting Reverse Rotation is returned the holder and picked up again a buzzer sounds O Dental Light ON OFF Switch Fig 3 13 E Switch for on off the dental light Fig 3 13 Dental Light ON OFF Switch a Function Switch Fig 3 14 E Use this switch for setting various working conditions Q As for the setup procedures please refer to 3 1 3 a i F Fig 3 14 Function Switch 3 Function Swicth Setup Procedure E a Timer Timer can be set maximum 90 mins 50 secs in 10 secs segment Function Switch Decrease Switch Increase Switch Start Switch Electric Motor Speed Setting Switch 1 To set a timer Momentarily press the function switch and set the time by pressing decrease switch and increase switch Minimum setting time by switch is 10 seconds Minimum setting time by switch is minute The setting time is indicated on the function indicator Momentarily press the start switch to start timer The end of setting time is informed by electronic sounds Example Setting time 3 minutes 30 seconds is indicated as 03 30 in the function indicator 2 Preset time setting Oo oOo amp Function
13. equipment The equipment or system should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the equipment or system should be observed to verify normal operation in the configuration in which it will be used Guidance and manufacture s declaration electromagnetic emissions The CLESTA II Unit is intended for use in the electromagnetic environment specified below The customer or the user of the CLESTA II Unit should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions The CLESTA II Unit uses RF energy only for its internal CISPR 11 function Therefore its RF emissions are very low and are Group 1 P not likely to cause any interference in nearby electronic equipment RF emissions The CLESTA II Unit is suitable for use in all establishments Class B CISPR 11 including domestic establishments and those directly Harmonic emissions Class A connected to the public low voltage power supply network TEC 61000 3 2 si that supplies buildings used for domestic purposes Voltage fluctuations Flicker emissions Complies IEC 61000 3 3 Guidance and manufacture s declaration electromagnetic immunity The CLESTA II Unit is intended for use in the electromagnetic environment specified below The customer or the user of the CLESTA II Unit should assure that it is used in su
14. in the presence of electromagnetic interference waves Do not place this product around equipment generating electromagnetic waves including communications equipment elevators etc as incorrect operation of this product may occur in the presence of electromagnetic interference waves Do not use equipment generating electromagnetic waves such as mobile phones around this product N WARNING 8 Be sure to turn off the main switch when electrocautery is in use Be sure to turn off the main switch when electrocautery is in use because noise may cause incorrect operation of this product 9 Ensure the maintenance of this product e Failure to maintain this product may lead to physical injury or property damage e Refer to the section of maintenance 10 Do not place objects weighing 3 kg or more on the Doctor s table Do not place objects weighing three kilograms or more on the Doctor s table This could cause damage to the Doctor s table defective function or accidents 11 Be sure to use the mirror cover e Be sure to use the mirror cover of the dental light when the light is turned on Direct contact with lamps may cause burns e See the Instruction Manual of the dental light for further information 12 Be sure to turn off the power when replacing lamps e Be sure to turn off the power when replacing the dental light This could result in electric shock e Use only dedicated halogen lamps e Immediately after a halogen lamp
15. troubles may Office if any abnormality arises arise due to motion of the chair during medical treatment and due to pinching between doctor section and chair functions Check for leakage of and The product will not normally work and troubles may arise Contact your dealer or our office if any abnormality arises Execute re inspection in accor dance with Method for operation described in the instruction manual Contact your dealer or our office if recovery is not achieved as a result of re inspection Cupfilling may not be executed Check of motions of equipment For each patient Check of air turbine bar Check of scaler tip For each patient Check of tightness of syringe nut For each patient After closing Matters attached to micromotor Air turbine revolution water flow air flow and so Before start a forth shall be free of bnormality Micromotor revolution water flow and so forth shall be free of abnormality Scaler vibration water flow and so forth shall be free of abnormality Appropriate bar shall be positively mounted Make sure to refer to the instruction manual attached to individual equipment Appropriate tip shall be positively mounted and be correctly used Make sure to refer to the instruction manual attached to the scaler The nut for fixing the nut of Type 77 3 way syringe shall be positively tightened
16. Bowl Optional Fig 3 26 The cuspidor bowl can be rotated either 90 to the inside and 90 to the outside ZA CAUTION Do not hold the cupfiller nozzle when rotating the cuspidor bowl be sure to hold the bottom of the spittoon bowl Rotating via the cupfiller nozzle may result in damage to the cupfiller nozzle 7 Height Adjustable Assistant Arm Optional Fig 3 27 Press the lock release button and raise arm to adjust the assistant holder height Position at desired height and release the lock button after that Note Support the arm with your hand until it is positioned at the desired height 8 Dental Light Fig 3 28 Inside Chair Side Bottom of the cuspidor bowl Fig 3 26 Cuspidor Bowl Rotation Assistant Holder Lock Release Button Balance Arm So Fig 3 27 Height Adjustable Assistant Arm Please refer to operating instruction for dental light 3 3 FOOT CONTROL SECTION 1 Foot Control Type A2 Fig 3 29 A Drive Air Pedal Pick up a handpiece from the handpiece holder and depress the drive air pedal the handpiece starts running Switch B Coolant Water Switch Coolant water switch allows handpiece coolant water to be turned on or off C Chip Blower Pedal Chip Blower Button Fig 3 29 Foot Control By depressing the chip blower button the chip air will Type A2 come out from handpiece without the bu
17. DENTAL UNIT CLESTA II OPERATING INSTRUCTIONS IMPORTANT This manual provides operating instructions for the CLESTA II The instructions contained in this booklet should be thoroughly read and understood before operating the unit and chair After the installation has been completed keep this manual in a safe place and refer to it for future maintenance If you have any questions about this Manual or this product please contact us If manual becomes unreadable or is lost please request a new manual by contacting your dealer Installation should be conducted by authorized personnel only Follow instructions on installation manual 1Bamont CE TABLE OF CONTENTS Page SAFETY PRECAUTIONS 1 1 OVERVIEW AND MAJOR COMPONENTS 5 2 DIMENSIONS AND SPECIFICATIONS 2 1 DIMENSIONS 6 2 2 SPECIFICATIONS 6 3 OPERATING INSTRUCTIONS 3 1 DOCTOR TABLE SECTION 7 3 2 CUSPIDOR UNIT SECTION 15 3 3 FOOT CONTROL SECTION 17 4 SAFETY LOCK DEVICE 18 5 CARE AND MAINTENANCE 19 6
18. MAINTANANCE AND INSPECTION 6 1 GUIDELINE FOR DAILY MAINTENANCE AND INSPECTION 22 MAINTENANCE AND INSPECTION BY USER 6 2 GUIDELINE FOR PERIODICAL CHECK UP 24 7 ELECTROMAGNETIC COMPATIBILITY EMC 25 8 LIST OF COMPATIBLE HANDPIECES 27 Intended Use of the Product This product is an active therapeutic device intended for the exclusive use for diagnoses treatments and relative procedures of dentistry The product must be operated or handled by the qualified dentists or by dental staffs under the supervision of the dentist Such dentists or dental staffs should instruct and or assist the patients to approach to and leave from the product Patients should not be allowed to operate or handle the product unless he she is so instructed The product is supplied together with the handpieces like electric micromotor air turbine and or motor scaler and so on Compatibility of Handpieces Use the compatible handpieces as shown on the attached list for this unit List of compatible handpieces Important Notes In case of the troubles please contact Takara Belmont offices or your dealers Do not disassemble or attempt to repair Disassembly repair or modifications shoud only be done by a qualified repair technician Attempts at disassembly repair or modifications may lead to abnormal operation and accidents Disposal of residue mate
19. NI 3 way Minilight w Light LUZZANI 6 way Minilight w Light DCI 3 way SYRINGE Turbine BIEN AIR BORA 836L UNIFIX with LIGHT NSK Ti Max X Air motor BIEN AIR Aquilon 830 UNIFIX with LIGHT PM1132 NSK EX 203 EX 6 Micromotor BIEN AIR MC3LK PLMP021PCB PM1132 BIEN AIR MC3LK PL970PCB PM1132 BIEN AIR MX DMX PCB PM1132 BIEN AIR MX2 DMX2 PRO PCB PM1132 BIEN AIR ISOLITE LK 40 IR E PLMP021PCB PM1132 BIEN AIR ISOLITE LK 40 IR E PL970PCB PM1132 NSK NL 400 NL 400SB PCB EX 6 NSK TIM 40J DA 290N PCB EX 6 NSK NLX PLUS Scaler SATELEC SP4055 w Light SATELEC SP4055 NEWTRON w Light NSK VARIOS 150 LUX w light NSK VARIOS 170 w light EMS SCALER CAVITRON 27 NOTE ee a UK Ltd Hemot TAKARA BELMONT CORPORATION Belmont House pull 2 1 1 Higashishinsaibashi Chuo ku Osaka 542 0083 Japan One St Andrews Way Bow TEL 81 6 6213 5945 FAX 81 6 6212 3680 London E3 3PA U K Tel 44 20 7515 0333 Fax 44 20 7987 3596 BOOK NO FEFA23M0 Printed in Japan 2012 04
20. Optional The CLESTA II is equipped with two types of flush out system Short time flush out is for cleaning handpiece water lines Long time flush out is for handpiece water lines bowl flush water line and cupfiller water line B a Function Switch Decrease Switch Increase Switch 1 Short time flush out Momentarily press the function switch three times and momentarily press the decrease switch Pick up the handpieces from the holder and set them in the cuspidor bowl By momentarily pressing the foot controller this starts short time flush out Water comes out from the handpiece and stops automatically after 40 seconds During flush out by momentarily pressing any one of unit control switch or foot controller will cancel flush out immediately 2 Long time flush out Momentarily press the function switch three times and momentarily press the increase switch Pick up the handpieces from the holder and set them in the cuspidor bowl By momentarily pressing the foot controller this starts long time flush out for 5 minutes Then cupfiller and bowl flush out starts and stops automatically in another 5 minutes During flush out momentarily pressing the any one of unit control switch or foot controller will cancel flush out immediately d Control panel switching sound on off Pressing a switch on the control panel makes an electronic sound This sound can be eliminated as follows gt Function Switch Decrease Switch Increase Swi
21. ater Otherwise it may break 12 Pay attention to water discharge when attach detach the spittoon bowl Make sure to turn the main switch off when you attach detach the spittoon bowl Cup filler sensor may react to hands Resulting water spill 13 Precautions for cleaning the resin cover For cleaning do not use cleaning agents containing solvent or abrasives thinners or oil based alcohol butanol and isopropyl alcohol which may cause cracks 14 Immediately wipe off drug solution when it comes into contact with this unit Should drug solution or water comes into contact with this unit immediately wipe it off with a dry soft towel etc This could result in defective function or electric leakage as well as spotting or rusting 15 Close the water main valve and turn off the main switch upon completion of work Be sure to close the water main valve and the main switch at the end of each work day to prevent water leakage from occurring 16 Be sure to operate switches with your hands Be sure to operate switches with your hands except the foot controller which is operated with your foot Operation with body parts other than hands may cause damage or incorrect operation 17 Pay attention during the headrest operation Do not allow hands fingers or hair to become entangled in the moving parts of the headrest during operationg 18 Be careful not to heat an empty water heater Exercise caution as heating of an empty water heater may re
22. atient during operation e Confirm that the patient is seated in the proper position Keep your eyes on the patient during the operation e Pay special attention to surroundings at automatic operation of the dental treatment table Damage to the backrest stool or Doctor s table may occur 6 Pay attention during movement of the Doctor s table e Pay attention to surroundings when you move the Doctor s table Injury by the tips of handpieces etc may occur e Be sure to move the Doctor s table by holding the handle of the unit e Be sure to move up amp down the Doctor s table by releasing the balance arm brake of the unit 7 Do not place anything hot on the Doctor s table Do not place anything hot on the Doctor s table This could cause deformation or discoloration 8 Cautions when adjusting the height of the cart type instrument table Cart type only After adjusting the height be sure to fasten the loosened lock screw Failure to heed this warning may cause the table to drop resulting in an accident 9 Observe the cart hose Cart type only Do not tread on the cart hose Failure to heed this warning may result in damage to the hose as well as people tripping over 10 Do not smack or rub this product Do not smack or rub this product forcefully This could cause damage to covers or defective function 11 Precautions for cleaning a spittoon bowl The spittoon bowl is made of glass Handle with care Do not wash it with hot w
23. breakers for equipment in the clinic when this product will not be used for a long period of time Be sure to turn off breakers for equipment in the clinic when this product will not be used for a long period of time following the completion of work during the suspension of business etc Insulation degradation may cause electrical fire 2 Be sure to turn off the main switch upon completion of work or during work breaks Be sure to turn off the main switch upon completion of work or during work breaks This prevents incorrect operation due to accidental contact and associated hazards 3 Do not sit on other than seat When the backrest is at the forward position do not sit on or place an undue load on the headrest or legreat of dental chair This could cause the unit to topple or could damage the unit 4 Do not place an undue load on the arm Do not get on or place an undue load on the arm of this unit or dental chair armrest This could cause the unit to topple or other accidents 5 Be sure to establish a grounding connection Be sure to establish a proper grounding connection Refer to a vendor for grounding connection Failure or electric leakage may lead to electric shock 6 Never disassemble repair or modify this product Individuals other than certified repair technicians should not disassemble or attempt to repair and modify this product This could lead to an accident failure electric shock or fire 7 Use with caution
24. ch an environment TEC 60601 Electromagnetic environment Immunity test Compliance level test level guidance Electrostatic 6 kV contact 6 kV contact Floors should be wood concrete or discharge ESD 8 kV air 8 kV air ceramic file If floors are covered IEC 61000 4 2 with synthetic material the relative humidity should be at least 30 Electrical fast 2 kV for power 2 kV for power Mains power quality should be that transient burst supply lines supply lines of a typical commercial or hospital IEC 61000 4 4 1 kV for input output 1 kV for input output environment lines lines Surge 1 kV differential mode 1 kV differential mode Mains power quality should be that TEC 61000 4 5 2 kV common mode 2 kV common mode of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations lt 5 Ur gt 95 dip in U for 0 5 cycle lt 5 Ur gt 95 dip in Ur for 0 5 cycle Mains power quality should be that of a typical commercial or hospital environment If the user of the on power supply 40 Ur 40 Ur CLESTA II Unit requires continued input lines 60 dip in Uy 60 dip in U operation during power mains IEC 61000 4 11 for 5 cycle for 5 cycle interruptions it is recommended that 10 Uy 10 Uz the CLESTA II Unit be powered from 30 dip in Uy 30 dip in U an uninterruptible power supply or a for 25cycle for 25cycle battery lt 5 Ur lt 5 Ur
25. e due to a power failure or other reasons O Cupfiller Nozzle Cupfiller Base 3 Bowl Flush Nozzle D Spittoon Bowl Solid Collector 6 Cuspidor Unit Control Panel Assistant Holder With Membrane Switch Vacuum Handpiece O Assistant Side Syringe 1 OVERVIEW AND MAJOR COMPONENTS OGO OG Fig 1 1 Overview and Major Components 0 Saliva Ejector Handpiece O Assistant Holder Arm D Cuspidor Top 43 Cuspidor Body Light Pole E First Arm Horizontal Arm O Balance Arm D Instrument Tray Main Control Panel 9 Dr Table Indicator Tray Handle With Arm Air Brake Button 1 Dr Side Handpiece Holder 2 Dr Side Syringe 3 Waste Receptacle Film Viewer 5 Foot Controller Dental Light 7 Handpieces Micromotor Air Turbine Motor Scaler and etc Manufacturers recommend to use the handpieces 2 DIMENSIONS AND SPECIFICATIONS 2 1 DIMENSIONS Chair Mount Type Pedestal Type mm z 3 28 3 a gl 3 A dL dada ps N So E 592 700 30130 e af f z 28 BE ae z Fig 2 1 Dimensions Ld 2 2 SPECIFICATIONS Rated power supply
26. g the knob Tighten the lock nut after adjustment Note Turning a knob counterclockwise increase flow volume and turning clockwise will decrease 4 Assistant Instrument Holder Fig 3 24 When picking up an instrument Saliva ejector or Vacuum handpiece from the assistant holder this starts the instrument working automatically Returning the instrument to the holder stops Open automatically Close Saliva ejector handpiece and vacuum handpiece have stop valves to close and adjust suction power Vacuum Saliva Ejector Handpiece Handpiece Fig 3 24 Vacuum Handpiece and Saliva Ejector Handpiece 5 Sensor Cupfiller Optional Fig 3 2 Fig 3 3 amp Fig 3 25 Place the cup paper cup on the cupfiller base water comes out from the cupfiller nozzle Paper Cup fills up the cup and stops automatically When cupfiller starts the bowl flush also starts and will run for about 6 sec and it stops automatically While filling the cup by momentarily pressing the cupfiller switch this will cancel the cup filler water flow Bowl Flush Nozzle During bowl flush momentarily press the bow flush switch y this will cancel the bowl flush water flow Fig 3 25 Sensor Cupfiller Note Use only suitable disposable paper cup dental paper cup Use only an empty cup using a cup with some water left can cause an over flow The sensor cupfiller needs over 2 seconds interval between cup filling 16 6 Cuspidor
27. has burnt out the lamp and the lamp holder are still hot Replace the lamp after they cool down e Do not touch halogen lamps with bare hands e See the Instruction Manual of the dental light for further information 13 Immediately wipe off any water spills or leakage on the floor Immediately wipe off any water spills or leakage on the floor Decreased strength of the floor may lead to physical injury including fall or property damage 14 Use with caution on patients with a cardiac pacemaker Use this product with extreme caution on patients with a cardiac pacemaker In the case of any abnormalities in patients during use immediately turn off this product and discontinue use 1 Only experienced personnel should use this product Only dentists or other dental professionals should use this product 2 Confirm safety before use Before use confirm that the parts are correctly and safely operating and that there are no obstacles around this product 3 Pay attention to patients and children Keep your eyes on patients especially children so that mischief or inadvertent operation of equipment will not lead to unexpected accidents 4 Discontinue use if you feel that something is wrong Always be careful to inspect this product for looseness rattling tilting wobbling sounds temperature odors etc Immediately discontinue use at the first feeling that something is wrong N CAUTION 5 keep your eyes on the p
28. hinly and evenly to the two O rings after sterilization Tighten the nut firmly in direction B to undo the nozzle ZA CAUTION Sterilization pouch Sterilization by class B cycles Fig 5 12 Sterilization Belmont Sterilization temperature is 135 degrees or less 77 Syringe Nozzle Skip the drying process if the temperature is to exceed 135 C If damage occurs to the sterilization pouch discard and sterilize again using a new pouch e Before use make sure that the nut is firmly tightened Air Filter Main Water Storage Maintenance Lid Drain Valve Valve Optional After cleaning the nozzle keep it in the clean place Open 4 Tubings and hoses Tubings and hoses can be cleaned with DURR FD333 cleaner Lose Pump 5 Air Filter Drain Valve Fig 5 13 Cover Air filter drain valve is located under the maintenance lid Once a week open the drain valve and drain off water from the air line Peo Air Fiter Dram Valye and Main Water Valve 6 Main Water Valve Optional Fig 5 13 Main water valve is located under the maintenance lid Turn off the main water valve after daily operation and for long term intervals 7 Filter Replacement Fig 5 14 The water filter in the junction box needs to be replaced at least once a year The air filter in the junction box needs to be replaced at least once every three years Contact your local service representative for replacement Fig 5 14 Filter Air Filter Water Filter
29. ic treatment Ultrasonic periodontal treatment Pick up the handpiece of scaler and set the range by pressing increase switch or decrease switch 5 Doctor Table Section Control Knobs A Handpiece Spray Water Control Knobs Fig 3 15 The handpiece spray water control knobs are located under the doctor table Each handpiece spray water control knob is marked 1 4 from the left side HP1 HP2 HP3 The handpiece spray water volume can be controlled independently The HP4 is optional g Doctor s Syringe Flow Control Knobs B Doctor s Syringe Flow Control Knobs Fig 3 15 Doctor s syringe flow control knobs are located on Handpiece Spray Water the facing right side under the doctor table The flow control knobs adjust the doctor s syringe air Fig 3 15 Doctor Table Section Control Knobs and water flow volume The yellow capped knob is the air flow control knob the blue capped knob is the water flow control knob Note Turning the control knob counterclockwise will increase the flow volume and turning clockwise x r Increase Decrease will decrease Fig 3 16 Stop Fig 3 16 Control Knob 6 Handpiece Pressure Gauge Fig 3 17 Handpiece drive air pressure gauge is located on the rear side of the doctor table While a handpiece is working the handpiece drive air pressure is indicated on the handpiece pressure gauge G Handpiece Drive Air Pressure Gauge Fig 3 17 Handpiece Pressure Gauge 13 7
30. ily press the function switch seven times and press the increase switch To return to original setting Momentarily press the function switch seven times and press the decrease switch Coolant Water ON OFF Switch In case of micromotor either the 2 mode or the 4 mode can be selected by mode select setup When this switch is pressed in the 2 mode setup switching between spray and OFF occurs In case of 4 mode setup switching occurs in the sequence indicated below each time when this switch is pressed Spray to Water only to Air only to OFF Function Switch Decrease Switch Increase Switch Coolant Water ON OFF Switch To set 2 mode Momentarily press the function switch eight times and press the decrease switch To set 4 mode Momentarily press the function switch eight times and press the increase switch Cupfiller and bowl flush Cupfiller and bowl flush are set to operate together when the cupfiller switch is activated bowl flush also starts To make these operate independently amp Function Switch Decrease Switch Increase Switch Momentarily press the function switch nine times and press the decrease switch To return to original setting Momentarily press the function switch nine times and press the increase switch 2 4 Scaler for SATELEC SP4055 Optional The setting range of ultrasonic scaler can be selected in 3 ranges Scaling prosthesis removal amalgam plugging Ultrasonic endodont
31. ing After closing After 13 Check of main switch closing and main valves 4 Product s moving parts 5 6 Check of water pressure and air pressure 17 Check of conditions of table section 8 Care Oil mist separator After closing After closing Flush the suction line and then clean the filter of the vacuum or saliva ejector handpiece Clean the cuspidor and dust filter Clean the filter of the solid collector Chemical filthy water and so forth shall not be found attached or remaining on the product exterior The product main switch shall be off and water air main valves shall be closed No abnormal noise or the like shall be produced from product s moving parts when the product is operated Water may enter the air line and equipment failure may arise Check the water pressure and air pressure by reading pressure gauges in the U Box section Basic set pressure Water 0 1 to 0 2 MPa Air 0 45 to 0 5 MPa The table shall be free of inclination and water shall not flow on the table The oil level in the oil mist separator shall be lower than the red line Clean the suction line and filter in accordance with Method for care described in the instruction manual Faulty suction may arise Execute dust removal and cleaning in accordance with Method for care described in the instruction manual Faulty water drainage may arise Clean the filter in
32. ing the coolant water switch until buzzer sound about 2 sec The electric motor rotation direction is indicated on the control panel by LED Please see page 9 0 Electric Motor Rotation Direction Control Switch 4 SAFETY LOCK DEVICE In the following cases the safety lock device to lock the chair movement is activated 1 When the pedal of the foot controller is depressed 2 When any switch on the doctor control panel or the assistant control panel is depressed while the chair is moving 3 During setting with the function switch on the doctor control panel 4 When the cuspidor bow is rotated toward the patient side Pedestal type Note Please refer to page 8 T Fig3 9 18 5 CARE AND MAINTENANCE Z CAUTION Turn OFF the master switch after daily operation and for long term intervals Cleaning Unit CAUTION All surfaces can be cleaned with DURR FD333 cleaner Spray the cleaner DURR FD333 on cloth and wipe the surfaces with the cloth Do not drench the chair and unit Wipe all surfaces dry after cleaning Drain ap 1 Cuspidor Bowl Fig 5 1 Basket q Strainer Take out the drain cap and the basket strainer located in the center of cuspidor bowl and clean them Remove the cupfiller base and pull out the cuspidor bowl and clean it Cuspidor 7 Bowl After cleaning secure the cuspidor bowl firmly A CAUTION Fig 5 1 Cuspidor Bowl Drain Cap Turn off the main switch before removing the c
33. ion distances between Portable and mobile RF communications equipment and the CLESTA II Unit The CLESTA II Unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the CLESTA II Unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the CLESTA II Unit as recommended below according to the maximum output power of the communications equipment Rated masimumi ouipul Separation distance ae to frequency of transmitter DENE E AEA 150 kHz to 80 MHz 80MHzto800MHz 800 MHz to 2 5 GHz w d 1 2VP d 1 2VP d 2 3VP 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by adsorption and reflection from structures objects and people 8 List of Compatible Handpieces DESCRIPTION Syringe LUZZA
34. l the backrest is raised up to the desired position Fig 3 7 Chair Manual Control Switches 6 Chair Auto Mode Switches Fig 3 8 A E A a Preset Operation G 0 Momentarily press the preset 1 switch O x the chair moves to the preset 1 position and stops automatically Fig 3 8 Chair Auto ae Switches Preset 2 position operated by the preset switch 2 Note For preset position adjustment refer to chair Manual b Auto Return Operation Momentarily press the auto return switch 0 the chair returns to the initial position the seat is the lowest position and the backrest is the upright position and stops automatically g Last Position Memory Operation Momentarily press the last position memory switch at the treatment position the backrest raises up to the rinsing position upright position and stops automatically Momentarily press the last position memory switch again the backrest returns to the previous treatment position and stops automatically a Emergency Stop Safety Stop During automatic movements preset auto return and last position memory by momentarily pressing any chair control switch this will cancel the automatic movement immediately T Safety Lock Indicator Fig 3 9 A E The safety lock indicator illuminates umber when the safety lock device LOCK af is working Note Please refer to page 18 4 Safety Lock
35. le to reach Fig 5 6 19 B Rinse thoroughly by distilled water at ordinary temperature or by clean water for more than minute fill HH Nh UN Continue the cleaning if contamination is remained D Immersed with an alkaline detergent for 5 minutes A Keown We recommend to use D212 made by DURR E Rinse thoroughly by distilled water at ordinary C Check whether contamination is removed or not after cleaning Fig 5 4 Surface cleaning temperature or by clean water for more than minute ZA CAUTION A Cleaning must be done within 1 hour after use Fitting cleaning brush Y A or Tooth brush RN Throw out and do not take autoclave handpiece for A following cases Slide groove e Any waste material can not be removed by clogged hole e Contamination and solid material attached to handpiece can not be removed Fig 5 5 Cleaning of hole and sliding part Sterilization Fig 5 7 Vacuum Tip Vacuum Cap Vacuum Handpiece Body Saliva Ejector Handpiece Body can be autoclave Vacuum handpiece body and saliva ejector body have to assemble before autoclave Fig 5 6 Cleaning the area A Insert the handpiece in a sterilization pouch and seal it brush is unable to reach B Autoclave for 20 min at 121 C or 4 min at 135 C Sterilization with autoclave is permitted up to 250 times CAUTION e Sterilization by class B cycles e Sterilization temperature is 135 degrees or less e The cap filter and b
36. ne Scaler Sterilize the handpiece according to manufacturer s operating manual Seca yringe body 3 Belmont 77 Syringe Fig 5 10 Disassembly Remove the nozzle from syringe by turning it in direction A Cleaning by hand A Wipe off the surface contamination by a cloth while rinsing the Nozzle surface by running clean warm water at 40 5 degrees Scrub the tip and joint part of nozzle by a cleaning brush or by a tooth brush with running clean warm water at 40 5 degrees Fig 5 11 B Rinse thoroughly by distilled water at ordinary temperature or by clean water for more than minute C Check whether contamination is removed or not after cleaning Continue the cleaning if contamination is remained D Immersed with an alkaline detergent for 5 minutes We recommend to use ID212 made by DURR E Rinse thoroughly by distilled water at ordinary temperature or by clean water for more than 1 minute Cleaning brush Z CAUTION or Tooth brush Cleaning must be done within 1 hour after use Fig 5 11 Cleaning Belmont Throw out and do not take autoclave handpiece for following cases 77 Syringe Nozzle e Contamination and solid material attached to handpiece can not be removed Sterilization Fig 5 12 The nozzle can be sterilized with autoclave A Insert the handpiece in a sterilization pouch and seal it B Autoclave for 20 min at 121 C or 4 min at 135 C Sterilization with autoclave is permitted up to 250 times Apply Vaseline t
37. o select limit mode For selecting the upper limit in the limit mode press either plus switch or minus switch The upper limit of the micromotor rotation speed changes in three steps or 5 steps Upper limit of rotation speed in case of 3 steps 10000 20000 40000 min rpm Upper limit of rotation speed in case of 5 steps 5000 10000 20000 30000 40000 min rpm See item g on page 7 for setting in 5 steps The micromotor rotation speed can be varied in the range of up to the selected upper limit by sliding the foot controller pedal right or left The rotation speed range varies by the micromotor type 2 Switching to preset rotation speed preset mode Pick up the micromotor out of the holder press this switch and then select preset mode SETI to 3 The rotation speed in this mode can be changed by pressing plus switch or minus switch Press store switch for storing the changed rotation speed When the foot controller is depressed upon selection of preset mode SET 1 to 3 the micromotor runs at the fixed rotation speed indicated on the indicator O Electric Motor Rotation Direction Control Switch Optional E Fig 3 12 Amber LED Green LED Reverse Rotation Forward Rotaition After picking up the electric motor from the holder the electric motor rotation direction can be changed by momentarily pressing Cid this switch E9 the rotation direction will be indicated by the
38. octor table the power indicator on the main control panel will then illuminate green Fig 3 2 amp 3 3 O 2 Main Control Panel A Spec Control Panel A E Spec Control Panel E oo 8 QQ RRE un HO 020 Fig 3 2 A Spec Main Control Panel Fig 3 3 E Spec Main Control Panel Q Power Indicator Q Safety Lock Indicator 3 Decreasing Switch 2 Cupfiller Switch Coolant Water ON OFF Switch Increasing Switch G Bowl Flush Switch Electric Motor Speed Setting Switch 5 Store Switch Light Pack Switch O Electric Motor Direction Control Switch 6 Function Indicator Chair Manual Control Switches Dental Light Switch Chair Auto Mode Switches Function Switch 2 Cupfiller Switch Fig 3 4 A E Momentarily press the cupfiller switch ily water will come out from gt the cupfiller nozzle for 3 seconds and stops automatically A E Also the bowl flush starts and will run for 6 seconds and stops automatically While the cupfiller is working by momentarily pressing the cupfiller switch w the cupfilling will cancel Also when the cupfiller starts the spittoon water flushes 6 seconds Fig 3 4 Cupfiller Switch and stops automatically Synchronized Bowl Flush Note The cupfiller water volume can be adjusted by the cupfiller flow control knob located inside the cuspidor body Refer to 3 2 3 3 Bowl Flush Switch Fig 3 5 A
39. ody are made of resin They may become deteriorated if they are sterilized in an autoclave many times After autoclave sterilization the cap filter body and valve are subject to discoloration which does not have a negative effect on performance Sterilization pouch The slide knob can be autoclave 100 times and is expendable supplies Skip the drying process if the temperature is to exceed 135 C If damage occurs to the sterilization pouch discard and sterilize again using a new pouch Fig 5 7 Vacuum Handpiece and Saliva Ejector Handpiece Storage After cleaning the vacuum tip keep it in the clean place 1 2 Cleaning vacuum and saliva ejector lines Fig 5 8 The sucking unit comes into contact with secretions spit and blood that contain bacteria every day Be sure to clean and sterilize it at the end of each work day Recommended cleaner Orotol Plus made by DURR Fig 5 8 Cleaning Vacuum Saliva Ejector Lines CAUTION Hose Ce Connector bd 1 3 Vacuum hose and saliva ejector hose are detachable from the cuspidor unit Fig 5 9 Turn the hose connector 90 counterclockwise the hose can be removed from the cuspidor unit Insert the hose connector and turn through 90 clockwise to reconnect Note After daily operation run two cup of clean water through handpieces to clean inside Release Vacuum Hose Saliva Ejector Hose Fig 5 9 Vacuum and Saliva Ejector Hose 20 2 Micro Motor Turbi
40. r rotating 17 2 Electric Motor Foot Control Type SE Fig 3 30 A Drive Air Pedal Pick up a handpiece from the handpiece holder and depress Coolant Water the drive air pedal the handpiece starts running Switch Chip Blower Button B Coolant Water Switch Momentarily depressing the coolant water switch is changed handpiece coolant water and air situation 7 s Drive Air Pedal The situation is shown on the main control panel I Electric Mot Refer to page 8 Coolant Water ON OFF Switch Rotation Speed Fig 3 30 Electric Motor Foot Control Type SE C Chip Blower Button By depressing the chip blower button the chip air will come out from handpiece without the bur rotating D Electric Motor Rotation Control Optional Pick up the electric motor from the handpiece holder and while pressing down slide drive air pedal horizontally to right and the electric motor will start running The rotation speed increases by sliding the drive air pedal further to the right The speed control by the foot control is within the limits of the electric motor speed setting E Coolant Water Switch Electric Motor Rotation Direction Switch Optional The coolant water switch can be changed for electric motor rotation direction switch To change to electric motor rotation direction switch Keep depressing the coolant water switch until buzzer sound about 2 sec To return to original coolant water switch setting Keep depress
41. required The required parts including consumable parts are listed below It may be different from the following list depending on the option of the unit For check up and repair call a technician of our authorized dealer Parts and components that require periodical check up Parts Description eae No Parts Description e Vacuum handpiece body 3years 8 Regulator 2 Saliva ejector handpiece body 3years o vavs years 3 4 Water supply hose Film viewer body part 5 6 Air supply hose Arm section of moving part Electric wiring of moving part Control PCBs EE La Footcontvoter 5years 10 Swicnes 5years EE ES Consumable parts F No Parts Despi J No Pars Descipton 1 Valve for vacum handpiece body 6 Fiter for oll mist separator Fiter Ar 8 Water a Handpiece wbings o o ring Packing Diaphragm afvee CCS 5 saiva gjectornoss ss N WARNING Execute the maintenance in accordance with this instraction manual and operating manual attached to each individual equipment Dental light Handpiece etc Failure to maintain this product may lead to physical injury or property damage 24 7 ELECTROMAGNETIC COMPATIBILITY EMC Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual Portable and mobile RF communications equipment can affect medical electrical
42. rial Please request a special contractor when you dispose amalgam In case of disposal of equipment When disposing the chair appropriately dispose complying with all current applicable regulations and local codes In EU area EU directive 2002 96 EC on waste electrical and electronic equipment WEEE is applied on this product In this directive environment conscious recycling abandonment is obligated SYMBOLS In this manual on the labels on the control panel of CLESTA I following symbols are used Confirm the meaning of each symbol Protective earth ground Chair last position To raise the chair Chair last position Chair manual control To Recline the backrest Fiber optic handpiece light on off mode selection Service outlet water flow control Water heater Authorized representative in the European community Type B Applied Parts Functional earth ground ON power Chair preset To Recline the backrest Chair auto return To lower the chair Chair auto return Chair preset1 Chair manual Chair auto control control To raise To lower the chair the backrest Handpiece coolant spray on off Store contol Bowl flush Cupfiller Minus T Plus sec Autoclave Symbol This symbol on component means that warnings or the component can be possibility to sterilised with an danger autoclave at 135 C max Rotation mode select Rotation speed Hehh Service o
43. sult in burning of the heater leading to fire N CAUTION 19 Precautions when using handpieces etc In order to ensure safety be sure to confirm that rotation has completely stopped before changing the rotation speed of the micromotor or inserting removing the bar of handpieces See the Instruction Manual accompanying various handpieces for further information 20 Precautions for cleaning the operation panel membrane switches Penetration of droplets of sanitizing spray into the back of the operation panel may be associated with switch failure Use a paper towel soaked with sanitizing solution to clean the surface of the operation panel 21 Precautions for cleaning e Never use sandpaper metal scrub brushes and abrasive cleaning agents to clean the unit e Do not use strongly acidic cleaning agents or alkaline pipe cleaning agents to avoid corrosion of metals etc 22 Practice of flush out Practice the flush out of water retained in the unit before the start of clinical practice at the beginning of each work day to maintain the quality of water for dental treatment and ensure a steady supply of water to handpieces Refer to the procedure for flush out 23 Set the pressure of the water tank at 200 kPa or less Adjust the air supply pressure for the water tank to 200 kPa or less An excessively high pressure may cause damage to the water tank 24 Do not use water other than purified water distilled water or pure water for
44. survey should be less than the compliance level in each frequency range gt Interference may occur in the vicinity of equipment marked with the following symbol D NOTE 1 At 80 MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by adsorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the CLESTA II Unit is used exceeds the applicable RF compliance level above the CLESTA II Unit should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the CLESTA II Unit b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3V m Essential performance purpose of IMMUNITY testing Unless operated by the switches for chair control the chair connected to CLESTA II does not make any movements except for sounding a buzzer and switching on off the indicator 26 Recommended separat
45. tch Momentarily press the function switch four times and momentarily press the decrease switch To return to original setting Momentarily press the function switch four times and momentarily press the increase switch e Fiber optic handpiece lighting mode Optional In case that fiber optic handpiece is installed the fiber optic turns on when the handpiece is taken out of the holder and turns off when the handpiece is returned to the holder This could be changed to fiber optic turns on when the handpiece is taken out of the holder and drive air Function Switch Decrease Switch Increase Switch pedal of foot control is activated Momentarily press the function switch five times and press the decrease switch To return to original setting Momentarily press the function switch five times and press the increase switch a f Electronic sound for timer i Electronic sound for timer can be changed QO Oe wo Function Switch Chair Auto Mode Switch Momentarily press the function switch six times Momentarily press one of chair auto mode switch 0 1 2 LP then the new electronic sound is to be memorized Micro motor maximum speed setting Optional The maximum rotation speed of the micro motor can be selected in 3 steps 10000 20000 40000min rpm This function can be changed to 5 steps 5000 10000 20000 30000 40000min rpm as follows Q Function Switch Decrease Switch Increase Switch Momentar
46. the water tank The water tank is intended only for use with purified water distilled water and pure water Do not use mouthwash or electrolyzed water such as ConCool or povidone iodine as they may cause clogged tubing or affect internal valves and equipment 25 Read the documents accompanying the various pieces of equipment Before use be sure to carefully read the package inserts and Instruction Manuals accompanying the various pieces of equipment including optional articles to ensure proper use NOTICE 1 Troubleshooting and contact information In the case of any problems discontinue use turn off the main switch and contact the dealer or our company 2 Check operation of the compressor With no air supplied this product does not operate even after turning on the main switch Turn on the power of the compressor before operating this product 3 Precautions when using water other than tap water The water unit is intended for use with tap water Caution should be exercised as the use of water other than tap water water through a sterilizer of water systems etc may result in failure of equipment 4 Use the turbine with a water check valve Use the turbine with a water check valve Contact the dealer or our company when a turbine without a water check valve will be used 5 Handling of equipment in the case of a power failure Put the handpiece in the holder and turn off the main switch if equipment stops working during us
47. uspidor bowl 2 Solid Collector Fig 5 2 Pull out the solid collector filter and clean it and Basket Strainer Solid Collector Filter After cleaning refit the solid collector firmly A Fig 5 2 Solid Collector Filter 3 Handpiece 1 1 Vacuum Handpiece and Saliva Ejector Handpiece For effective sterilization washing for removing Vacuum Tip contamination and immersion by a cleaning agent are ow 1 required Then rinse by water in order to remove residual cleaning agent on medical device Take following Upper Vacuum Cap 7 U Sali procedures from cleaning to sterilization Elector Cap Use the disposable saliva ejector tip rer Lower Saliva Lower Vacuum Cap Ejector Cap gt 20d PPP Disassembly Fig 5 3 Slide Knob Slide Knob Disassemble the handpiece for the preparation of cleaning TE Saliva Eject as the following figures show Pull the hose connector to Hendpieca Hose Connector Hand co disconnect the vacuum hose S S z Cleaning by hand 5 eee ofc ies zs Vacuum Saliva Eject A Wipe off the surface contamination by a cloth while rinsing the Hose Hose surface by running clean warm water at 40 5 degrees Fig 5 4 Fess Vacuum Handpiece and Scrub the intubation or hole slide groove and filter by a cleaning i Saliva Ejector Handpiece brush or by a tooth brush with running clean warm water at 405 degrees Fig 5 5 Wipe off by a cloth for the area which brush is unab
48. utlet air Serial number Ez Caution It means caution w nN aa wy A Refer to instruction Alternating current manual booklet OFF power To raise the backrest Chair preset2 To raise the chair Handpiece Setting Micro motor Forward Reverse select Scaler power control Dental light on off Service outlet water Date of manufacture Separate collection for electrical and electronic equipment Non ionizing radiation SAFETY PRECAUTIONS E Before use read the Safety precautions carefully to ensure proper use E The following information is designed to ensure safe use of this product and to prevent injury and damage to you and others The precautions contained here are classified depending on the severity and degree of imminence of possible injury or damage resulting from improper use Be sure to follow all the information which is important for safety Classification of precautions Severity and degree of imminence of possible injury or damage This symbol indicates that ignorance of these precautions may lead to severe injury or even death as a result of improper use This symbol indicates that ignorance of these precautions may lead to mild or moderate physical injury or damage to property as a result of improper use NOTICE This symbol indicates that it is recommended to follow these precautions for safety N WARNING 1 Be sure to turn off
Download Pdf Manuals
Related Search
Related Contents
入 札 公 告 取扱説明書 700-TAP017 elliptimill e8/e9 home & commercial service manual PROPERTY OF TVH® Portable Multi-logger FP3000 user manual - Champagne PC Services Manuel d`installation et d`entretien Electrodistributeurs à trois Page 1 Page 2 ・ご使用前に、 この「安全上のご注意」をよくお読 みの FH-2500 - Dolmar libretto d`istruzioni per l`uso e manutenzione Copyright © All rights reserved.
Failed to retrieve file