Teledyne P/NM70946 User's Manual

image

Contents

1. 500 375 250 Cell Output uA 125 5 50k 100 k 150 k 200 k Oxygen Concentration ppm Figure 2 3 Characteristic Input Output Curve for a Micro Fuel Cell Teledyne Analytical Instruments Operational Theory 2 2 5 2 Operational Theory Model Ultra Trace 3000 2 3 Sample System The sample system delivers gases to the Micro Fuel Cell sensor from the analyzer rear panel inlet Depending on the mode of operation either sample or calibration gas is delivered The Model Ultra Trace 3000 sample system is designed and fabricated to ensure that the oxygen concentration of the gas is not altered as it travels through the sample system The sample system for the standard instrument incorporates 1 4 VCR for sample inlet and outlet tube connections at the rear panel The sample or calibration gas that flows through the system is monitored by a flowmeter downstream from the cell Figure 2 4 shows the piping layout and flow diagram for the standard model SAMPLE IN EXHAUST o TOP RIGHT SIDE VCR FITTINGS ae CELL BLOCK NEEDLE VALVE FLOWMETER Figure 2 4 Piping Layout 2 6 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operational Theory 2 Sample
2. 3 3 2 2 50 Pin Equipment Interface Connector Figure 3 4 shows the pin layout of the Equipment Interface connector The arrangement is shown as seen when the viewer faces the rear panel of the analyzer The pin numbers for each input output function are given where each function is described in the paragraphs below lt 17 o o oo o o o o 33e o o o e 18 o o o o o o o o o o o o o 90 Figure 3 4 Equipment Interface Connector Pin Arrangement Analog Outputs There are four DC output signal pins two pins per output For polarity see Table 3 1 The outputs are 0 1 V de of Range Voltage rises linearly with increasing oxygen from 0 V at 0 ppm to 1 V at full scale ppm Full scale 100 of programmable range 0 1 V de Range ID 0 25 V Low Range 0 5 V Medium Range 0 75 V High Range 1 V Air Cal Range 4 20 mA de Range Current increases linearly with increasing oxygen from 4 mA at 0 ppm to 20 mA at full scale ppm Full scale 100 of programmable range 4 20 mA de Range ID 8 mA Low Range 12 mA Medium Range 16 mA High Range 20 mA Air Cal Range Table 3 1 Analog Output Connections Pin Function Range ID 4 20 mA floating Range ID 4 20 mA floating Range 4 20 mA floating Range 4 20 mA floating Range ID 0 1 V dc 23 Range ID 0 1 V dc negative ground 24 Range 0 1 V de 7 Range 0 1 V dc negative grou
3. OPERATING INSTRUCTIONS FOR Model Ultra Trace 3000 Oxygen Analyzer W Teleg lyne Analytical Instrument is v LTRA 7 TRACE orvan ACE 2000 R ANALYZE SYSTEM ZERO SPAN ALARMS RANGE SET Flow DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND OR SERVICING BEFORE CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR MANAGER P NM70946 12 10 99 ECO 99 0483 Teledyne Analytical Instruments Model Ultra Tace 3000 Copyright 1999 Teledyne Analytical Instruments All Rights Reserved No part of this manual may be reproduced transmitted tran scribed stored in a retrieval system or translated into any other language or computer language in whole or in part in any form or by any means whether it be electronic mechanical magnetic optical manual or otherwise without the prior written consent of Teledyne Analytical Instruments 16830 Chestnut Street City of Industry CA 91749 1580 Warranty This equipment is sold subject to the mutual agreement that it is warranted by us free from defects of material and of construction and th
4. the analog range ID outputs Signals from the power supply are also monitored and through the microprocessor the system failure alarm is activated if a malfunction is detected 2 10 DP Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Installation 3 Installation E SS C Installation of the Model Ultra Trace 3000 Analyzer includes 1 Unpacking 2 Mounting 3 Gas connections 4 Electricalconnections 5 Installing the Micro Fuel Cell 6 Testing the system 3 1 Unpacking the Analyzer Although the analyzer is shipped complete certain of the parts such as fuses and sensors are wrapped separately to be installed on site as part of the installation Carefully unpack the analyzer and inspect it for damage Imme diately report any damage or shortages to the shipping agent 3 2 Mounting the Analyzer The Model Ultra Trace 3000 is for indoor use in a general purpose area It is NOT for hazardous environments of any type The standard model is designed for flush panel mounting Figure 3 1 is an illustration of the Ultra Trace 3000 standard front panel and mounting bezel There are four mounting holes one in each corner of the rigid frame The drawings section in the rear of this manual contains outline dimensions and mounting hole spacing diagrams On special order a 19 rack mounting panel can be provided For rack mounting one Ultra Trace 3000 series analyzer is flush panel mounted o
5. 2 Alarm Relay Contact Pins Pin Contact 45 Threshold Alarm 1 normally closed contact 28 Threshold Alarm 1 moving contact 46 Threshold Alarm 1 normally open contact 42 Threshold Alarm 2 normally closed contact 44 Threshold Alarm 2 moving contact 43 Threshold Alarm 2 normally open contact 36 System Alarm normally closed contact 20 System Alarm moving contact 37 System Alarm normally open contact Digital Remote Cal Inputs Accept 0 V off or 24 V dc on inputs for remote control of calibration See Remote Calibration Protocol below See Table 3 3 for pin connections 3 6 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Installation 3 Zero Span Floating input 5 to 24 V input across the and pins puts the analyzer into the Zero mode Either side may be grounded at the source of the signal 0 to 1 volt across the terminals allows Zero mode to terminate when done A synchronous signal must open and close the external zero valve appropriately See Remote Probe Connector The C option internal valves operate automatically Floating input 5 to 24 V input across the and pins puts the analyzer into the Span mode Either side may be grounded at the source of the signal 0 to 1 volt across the terminals allows Span mode to terminate when done A synchronous signal must open and close external span valve appropriately See Figure 3 5 Remote Probe Connector The C option internal
6. Select and enter More to get a new screen with additional subfunctions listed e Self Test The instrument performs a self diagnostic test to check the integrity of the power supply output boards and amplifiers e Version Displays Manufacturer Model and Software Version of instrument e Neg The operator selects whether display can show negative oxygen readings or not e TRAK HLD The operator sets whether the instrument analog outputs track the concentration change during calibration and sets a time delay for the concentration alarms after calibration e Filter This is to set the response time of the digital filter in the LO range 4 3 1 Tracking the Oxygen Readings during Calibration and Alarm delay The user has the option of setting the preferenc as to whether the analog outputs track the display readings during calibration or not To set the prefer ence press the System key once and the first System menu will appear in the VFD display TRAK HLD Auto Cal PSWD Logout More TRAK HLD should be blinking To enter this system menu press the Enter key once Output Sttng TRACK Alarm Dly 10 min Or Output Sttng HOLD Alarm Dly 10 min 4 4 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operation 4 In the first line TRACK or HOLD should be blinking The operator can toggle between TRACK and HOLD with the Up or Down keys When TRACK is selected the analog outputs 0 1 VDC and 4 20 ma and the r
7. USE 250V 1 0A T FUSE SAMPLE IN SEE MANUAL FOR INTERCONNECTION DETAIL J EXHAUST OUT Figure 3 3 Rear Panel of the Model Ultra Trace 3000 3 3 1 Gas Connections The unit is manufactured with sinch VCR fittings For a safe connec tion SAMPLE IN In the standard model gas connections are made at the SAMPLE IN and EXHAUST OUT connections Calibration gases must be Tee d into the Sample inlet with appropriate valves A VCR fitting is pro vided for the inlet connection The inlet gas pressure should be reasonably regulated Pressures be tween 1 and 50 psig are acceptable as long as the pressure once established will keep the front panel flowmeter reading in an acceptable range 0 5 to 2 0 SLPM For non pressurized sample or very low pressure less than psig vacuum service plumbing is recommended See next section Vacuum Service Teledyne Analytical Instruments 3 3 3 Installation Model Ultra Trace 3000 If greater sample flow is required for improved response time install a bypass in the sampling system upstream of the analyzer input VACUUM SERVICE If the sample pressure is at atmospheric or very low pressure the instrument must be ordered with the vacuum service option This will ensure that the flow control valve is located on the exhaust side of the Micro Fuel Cell EXHAUST OUT Exhaust connections must be consistent with the hazard level of the constituent gases Check Local State and Feder
8. done and the sample is being analyzed Note The Remote Valve connections described below provides signals to ensure that the zero and span gas valves will be controlled synchronously Range ID Relays Four dedicated Range ID relay contacts The first three ranges are assigned to relays in ascending order Low range is as signed to Range 1 ID Medium range is assigned to Range 2 ID and High range is assigned to Range 3 ID The fourth range is reserved for the Air Cal Range 25 Table 3 4 lists the pin connections Table 3 4 Range ID Relay Connections Pin Function 21 Range 1 ID Contact 38 Range 1 ID Contact 22 Range 2 ID Contact 39 Range 2 ID Contact 19 Range 3 ID Contact 18 Range 3 ID Contact 34 Range 4 ID Contact Air Cal 35 Range 4 ID Contact Air Cal Network I O A serial digital input output for local network protocol At this printing this port is not yet functional It is to be used for future options to the instrument Pins 13 and 29 Remote Valve Connections The Ultra Trace 3000 is a single chassis instrument which has no Remote Valve Unit Instead the Remote Valve connections are used as a method for directly controlling external sample zero span gas valves See Figure 3 5 3 8 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Installation 3 SAMPLE hot J ina m Solenoid 2 hot ZERO hot i Solenoid 3 hot SPAN hot Matching 9 Sol
9. for alarm 1 have been set press Alarms again and repeat this procedure for alarm 2 AL 2 e To reset a latched alarm go to Dft and then press either A two times or V two times Toggle it to Y and then back to N OR Go to Ltch and then press either A two times or V two times Toggle it to N and back to Y DP Teledyne Analytical Instruments 4 19 4 Operation Model Ultra Trace 3000 Alarm Hysterisis There is alarm hysterisis to prevent chatter of the alarm contacts It is set to 0 2 ppm for alarms set above 1 ppm and 10 ppb for alarms set below 1 ppm 4 7 The Range Function The Range function allows the operator to program up to three concen tration ranges to correlate with the DC analog outputs If no ranges are defined by the user the instrument defaults to Low 0 250 ppb Med 0 1 ppm High 0 10 ppm The Model Ultra Trace 3000 is set at the factory to default to autoranging In this mode the microprocessor automatically responds to concentration changes by switching ranges for optimum readout sensitivity If the current range limits are exceeded the instrument will automatically shift to the next higher range If the concentration falls to below 90 of full scale of the next lower range the instrument will switch to that range A corresponding shift in the DC percent of range output and in the range ID outputs will be noticed The autoranging feature can be overridden so that analog output stays on a
10. function Decide how your alarms should be configured The choice will depend upon your process Consider the following four points 1 Which if any of the alarms are to be high alarms and which if any are to be low alarms Setting an alarm as HIGH triggers the alarm when the oxygen concentration rises above the setpoint Setting an alarm as LOW triggers the alarm when the oxygen concentration falls below the setpoint Decide whether you want the alarms to be set as e Both high high and high high alarms or e One high and one low alarm or e Both low low and low low alarms 2 Are either or both of the alarms to be configured as failsafe In failsafe mode the alarm relay de energizes in an alarm condition For non failsafe operation the relay is energized in an alarm condition You can set either or both of the concentration alarms to operate in failsafe or non failsafe mode 3 Are either of the alarms to be latching In latching mode once the alarm or alarms trigger they will remain in the alarm mode even if process conditions revert back to non alarm conditions This mode requires an alarm to be recognized before it can be reset In the non latching mode the alarm status will terminate when process conditions revert to non alarm conditions 4 Are either of the alarms to be defeated The defeat alarm mode is incorporated into the alarm circuit so that maintenance can be performed under conditions which would norma
11. function when they have completed their assigned operations Pressing the Escape button in many cases also switches the analyzer back to the Ana lyze function Alternatively you can press the Analyze button at any time to return to analyzing your sample 4 9 Signal Output The standard Model Ultra Trace 3000 Oxygen Analyzer is equipped with two 0 1 V de analog output terminals accessible on the back panel one concentration and one range ID and two isolated 4 20 mA dc current outputs one concentration and one range ID See Rear Panel in Chapter 3 Installation for illustration The signal output for concentration is linear over the currently selected analysis range For example if the analyzer is set on range that was defined as 0 100 ppm O then the output would be Voltage Signal Current Signal ppm O Output V de Output mA de 0 0 0 4 0 10 0 1 5 6 20 0 2 7 2 30 0 3 8 8 40 0 4 10 4 50 0 5 12 0 60 0 6 13 6 70 0 7 15 2 80 0 8 16 8 90 0 9 18 4 100 1 0 20 0 The analog output signal has a voltage which depends on the oxygen concentration AND the currently activated analysis range To relate the signal output to the actual concentration it is necessary to know what range the instrument is currently on especially when the analyzer is in the autoranging mode 4 22 DN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operation 4 To provide an indication of the range a second pair of analog output ter
12. the power source 2 Open the front panel door by pressing the release button on the top right corner of the door all the way in with a narrow gauge tool 3 With one hand hold the top of the cell block while unscrewing the plastic ring holder Once the plastic ring is loose remove the top of the cell block CAUTION The cell is shipped separately inside two bags filled with oxygen free inert gas Extreme care must be taken to ensure that the cell is exposed to air for the very minimum amount of time including during installation As an example an air exposure of 3 to 5 minutes may require 24 hours or longer before the cell recovers to less than 0 5 ppm on an oxygen free sample gas On the other hand if the air exposure is limited to less than 10 seconds the recovery time will be reduced to two hours or less 3 5 Testing the System Before plugging the instrument into the power source e Check the integrity and accuracy of the gas connections Make sure there are no leaks e Check the integrity and accuracy of the electrical connections Make sure there are no exposed conductors e Check that inlet sample pressure is within the accepted range se section 3 3 1 Power up the system and test it by performing the following operations 1 Repeat the Self Diagnostic Test as described in chapter 4 section 4 3 5 3 10 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Installation 3 COLLAR CELL BL
13. the sensor to determine when the output is within the accept able range for zero In the manual mode the operator determines when the reading is within the acceptable range for zero Make sure the zero gas is connected to the instrument If you get a CELL FAILURE message skip to section 4 4 1 3 4 4 1 1 Auto Mode Zeroing Press Zero to enter the zero function mode The screen allows you to select whether the zero calibration is to be performed automatically or manu ally Use the AV arrow keys to toggle between AUTO and MAN zero set tling Stop when AUTO appears blinking on the display Zero Settling AUTO lt ENT gt To Begin Press Enter to begin zeroing tHE PPM Zero Slope ppm s The beginning zero level is shown in the upper left corner of the dis play As the zero reading settles the screen displays and updates information on Slope unless the Slope starts within the acceptable zero range and does not need to settle further Then and whenever Slope is less than 0 08 for at least 3 minutes instead of Slope you will see a countdown 5 Left 4 Left and so fourth These are five steps in the zeroing process that the system must complete AFTER settling before it can go back to Analyze 4 12 PN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operation 4 HHEH PPM Zero 4 Left ppm s The zeroing process will automatically conclude when the output is within the acceptable range for a good ze
14. to use interface from operator to analyzer It displays values options and messages that give the operator immediate feedback NeedleValve To adjust flow of gas sample Flowmeter Monitors the flow of gas past the sensor Readout is 0 2 to 2 4 standard liters per minute SLPM of nitrogen C Standby Button The Standby turns off the display and outputs but circuitry is still operating CAUTION The power cable must be unplugged to fully disconnect power from the instrument When chassis is exposed or when access door is open and power cable is connected use extra care to avoid contact with live electrical circuits Access Door For access to the Micro Fuel Cell the front panel swings open when the latch in the upper right corner of the panel is pressed 1 4 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Introduction 1 all the way in with a narrow gauge tool Accessing the main circuit board requires unfastening rear panel screws and sliding the unit out of the case 1 6 Rear Panel Equipment Interface The rear panel shown in Figure 1 2 contains the gas and electrical connectors for external inlets and outlets Some of those depicted are op tional and may not appear on your instrument The connectors are described briefly here and in detail in chapter 3 Installation PR Teledyne Analytical Instruments CAUTION RISK OF ELECTRIC SHOCK DO NOT OPEN NO USER SERVICEABLE PARTS INSIDE SERVICE ONLY B
15. values are in effect Ranges LO 250ppb ppm MED 1 ppm HI 10 ppm Auto Ranging ON Alarm Relays Defeated Alarm 1 at10 000 ppm Alarm 2 at 1 000 ppm HI Not failsafe Not latching Zero Auto every 0 days at 0 hours Span Auto at 008 00 ppm every 0 days at 0 hours If you choose not to use password protection the default password is automatically displayed on the password screen when you start up and you simply press Enter for access to all functions of the analyzer DP Teledyne Analytical Instruments 4 1 4 Operation Model Ultra Trace 3000 4 2 Using the Data Entry and Function Buttons Data Entry Buttons The lt gt arrow buttons select options from the menu currently being displayed on the VFD screen The selected option blinks When the selected option includes a modifiable item the AV arrow buttons can be used to increment or decrement that modifiable item The Enter button is used to accept any new entries on the VFD screen The Escape button is used to abort any new entries on the VFD screen that are not yet accepted by use of the Enterbutton Figure 4 1 shows the hierarchy of functions available to the operator via the function buttons The six function buttons on the analyzer are e Analyze This is the normal operating mode The analyzer monitors the oxygen content of the sample displays the percent of oxygen and warns of any alarm conditions e System The system function consists of six subfuncti
16. 2 8 Installation 3 1 Unpacking the Analyzer nnen neen 3 1 3 2 Mounting the Analyzer nnen eeneennenneennennenn 3 1 3 3 Rear Panel Connections nnee 3 3 3 3 1 Gas Connections tene en 3 3 3 3 2 Electrical Connections nennen 3 4 3 3 2 1 Primary Input Power nennen 3 4 3 3 2 2 50 pin Interface Connector ccc 3 5 3 3 3 Remote Probe Connector enen 3 8 3 4 Installing the Micro Fuel Cell eneen 3 10 3 5 Testing the System ccc cca her secs Meck ks 3 10 Operation 4 1 Introduction Sese ne ornata tertia earn dk inte Enait hete tet 4 1 4 2 Using the Data Entry and Function Buttons 4 2 4 3 The System Function nennen neeeeenenenneneneen 4 3 4 3 1 Tracking the O Readings during CAI amp Alarm 4 4 4 3 2 Setting up an Auto Cal nnen 4 5 4 3 3 Password Protection annen eenenen 4 6 4 3 3 1 Entering the Password een 4 7 DP Teledyne Analytical Instruments Trace Oxygen Analyzer 4 3 3 2 Installing or Changing the Password 4 8 ASA LOGOU Zoniet edt deedeetee as 4 9 4 3 5 System Self Diagnostic Test nnn 4 9 4 3 6 Version Screen serine codes eer eaistelecpexsasebear yee 4 10 457 Filter FUNCION ennen dte 4 11 4 4 Calibration of the Analyzer nnen eenen 4 11 AAN Zero Alseen aa EA EEES EAE ENES 4 11 4 4 1 1 Auto Mode Zeroing nen 4 12 4 4 1 2 Manual Mode Zeroing nnen 4 13 44153 Cell Failure zet ens daens nte nnn 4 13 442 Span Caliente nennen 4 14 4 4 2 1 A
17. Define Analysis Confrigure Mode anges of Alarm Set Password Operation ba Logout v Neg Filter Figure 4 1 Hierarchy of Functions and Subfunctions 4 3 The System Function The subfuctions of the System function are described below Specific procedures for their use follow the descriptions e Auto Cal Used to define an automatic calibration sequence and or start an Auto Cal e PSWD Security can be established by choosing a 5 digit password PSWD from the standard ASCII character set See Installing or Changing the Password below for a table of ASCII characters available Once a unique password is assigned and activated the operator MUST enter the UNIQUE password to gain access to set up functions which alter the instrument s operation such as setting the instrument span or zero setting adjusting the alarm setpoints or defining analysis ranges Teledyne Analytical Instruments 4 3 4 Operation Model Ultra Trace 3000 After a password is assigned the operator must log out to activate it Until then anyone can continue to operate the instrument without entering the new password Only one password can be defined Before a unique password is assigned the system assigns TETAI by default This allows access to anyone After a unique password is assigned to defeat the security the password must be changed back to TETAI e Logout Logging out prevents unauthorized tampering with analyzer settings e More
18. OCK SENSOR CELL HOLDER Figure 3 8 Installing the Micro Fuel Cell Teledyne Analytical Instruments 3 11 3 Installation Model Ultra Trace 3000 IMPORTANT In the event of loss of flow through the analyzer if the vent is vented to a location of high oxygen content oxygen will back diffuse through the vent line and in most cases quickly saturate the cell with oxygen which can then require a quite long purge down time for the sensor when then exposed to low oxygen concentrations In the event that flow is to be interrupted into the analyzer it is suggested that the user do one of the following 1 Bag the sensor in nitrogen during this time 2 Install a shut off valve on the vent port of the ana lyzer or somewhere within the users sample system 3 12 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operation 4 Operation 4 1 Introduction Once the analyzer has been installed it can be configured for your application To do this you will e Setsystem parameters e Establish a security password if desired requiring Operator to log in e Establish and start an automatic calibration cycle if desired e Calibrate the instrument e Define the three user selectable analysis ranges then choose autoranging or select a fixed range of analysis as required e Set alarm setpoints and modes of alarm operation latching failsafe etc Before you configure your Ultra Trace 3000 these default
19. Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operational Theory 2 Operational Theory 2 1 Introduction The analyzer is composed of three subsystems 1 Micro fuel Cell Sensor 2 Sample System 3 Electronic Signal Processing Display and Control The sample system is designed to accept the sample gas and transport it through the analyzer without contaminating or altering the sample prior to analysis The Micro fuel Cell is an electrochemical galvanic device that translates the amount of oxygen present in the sample into an electrical current The electronic signal processing display and control subsystem simplifies operation of the analyzer and accurately processes the sampled data The microprocessor controls all signal processing input output and display functions for the analyzer 2 2 Micro Fuel Cell Sensor 2 2 1 Principles of Operation The oxygen sensor used in the Model Ultra Trace 3000 series is a Micro fuel Cell Model B 2CXL designed and manufactured by Analytical Instruments It is a sealed plastic disposable electrochemical transducer The active components of the Micro fuel Cell are a cathode an anode and the aqueous KOH electrolyte in which they are immersed The cell converts the energy from a chemical reaction into an electrical current in an external electrical circuit Its action is similar to that of a battery There is however an important difference in the operation of a battery as compared
20. Trace Oxygen Analyzer Operation 4 4 3 7 Filter Function The response time on the most sensitive range ppb range is user definable from approximately 1 60 minutes The adjustable filter allows the user to tune the response of the analyzer to best balance sensor noise and response time requirements The factory default setting is 5 minutes The actual response time will depend on the user s sample system the length and size the tubing of tubing as well as the sample flow rate The filter setting can be accessed by selecting SYSTEM on the keypad followed by MORE on the display with the lt gt keys The filter function is then selected and changed using the arrow keys Press ENTER and ANA LYZE to return to analyze mode In the event of an over range condition the filter rate will automatically switch to a faster setting approximately 45 sec response time for the duration of the over range or upset condition This feature allows the ana lyzer to quickly respond to and track an upset condition 4 4 Calibration of the Analyzer The analyzer must be calibrated prior to its use For most applications where the desired range of measurement is 0 to 10 ppm or less we recommend the analyzer be calibrated using a span gas as detailed below Span Gas Calibration Before the cell is ready for calibration it must be purged with sample gas to a low oxygen level preferably below 0 1 ppm However if the oxygen contact of the sample gas is
21. Under Pressure 2 20 psig Flow Control l l i Flowmeter i Needle l l Valve Exhaust Out l L Figure 2 5 Flow Diagram Sample Under Pressure Standard Model Ultra Trace 3000 ea a eee l l Sample at l 0 lt 2 psig Flow Control Needle Valve Flowmeter Vacuum Do not exceed 10 Hg Vacuum l Figure 2 5 1 Flow Diagram Sample at Zero Pressure Model Ultra Trace 3000 V Teledyne Analytical Instruments 2 7 2 Operational Theory Model Ultra Trace 3000 Figure 2 5 is the flow diagram for the sampling system In the standard instrument calibration gases can be connected directly to the Sample In port by teeing to the port with appropriate valves 2 4 Electronics and Signal Processing The Model Ultra Trace 3000 Oxygen Analyzer uses an 8031 microcon troller with 32 kB of RAM and 128 kB of ROM to control all signal pro cessing input output and display functions for the analyzer System power is supplied from a universal power supply module designed to be compatible with any international power source Figure 2 6 shows the location of the power supply and the main electronic PC boards The signal processing electronics including the microprocessor analog to digital and digital to analog converters are located on the motherboard at the bottom of the case The preamplifier board is mounted on top of the motherboard as shown in the figure These boards are accessible after re moving the back panel Figu
22. Y QUALIFIED PERSONNEL 85 250Vac 47 63Hz 0 9 A MAX USE 250V 1 0A T FUSE SAMPLE IN SEE MANUAL FOR INTERCONNECTION DETAIL EXHAUST OUT Figure 1 2 Model Ultra Trace 3000 Rear Panel e Power Connection Universal AC power source e Gas Inlet and Outlet One inlet and one exhaust out e Analog Outputs 0 1 V de oxygen concentration plus 0 1 V de range ID and isolated 4 20 mA de oxygen concentration plus 4 20 mA dc range ID e Alarm Connections 2 concentration alarms and system alarm DN Teledyne Analytical Instruments 1 5 1 Introduction Model Ultra Trace 3000 Note 1 6 e RS 232 Port Serial digital concentration signal output and control input e Remote Probe Used in the Ultra Trace3000 for controlling external solenoid valves only e Remote Span Zero Digital inputs allow external control of analyzer calibration e Calibration Contact To notify external equipment that instrument is being calibrated and readings are not monitoring sample e Range ID Contacts Four separate dedicated range relay contacts Low Medium High Cal e Network I O Serial digital communications for local network access For future expansion Not implemented at this printing If you require highly accurate Auto Cal timing use external Auto Cal control where possible The internal clock in the Model Ultra Trace 3000 is accurate to 2 3 Accordingly internally scheduled calibrations can vary 2 3 per day DP
23. a tion related to it The safety of personnel is ultimately the responsibility of those who control process conditions While this instrument may be able to provide early warning of imminent danger it has no control over process conditions and it can be misused In particular any alarm or control systems installed must be tested and understood both as to how they operate and as to how they can be defeated Any safeguards required such as locks labels or redun dancy must be provided by the user or specifically requested of Teledyne at the time the order is placed Therefore the purchaser must be aware of the hazardous process conditions The purchaser is responsible for the training of personnel for providing hazard warning methods and instrumentation per the appropriate standards and for ensuring that hazard warning devices and instrumentation are maintained and operated properly Teledyne Analytical Instruments the manufacturer of this instrument cannot accept responsibility for conditions beyond its knowledge and control No statement expressed or implied by this document or any information disseminated by the manufactur er or its agents is to be construed as a warranty of adequate safety control under the user s process conditions DP Teledyne Analytical Instruments Trace Oxygen Analyzer Specific Model Information The instrument for which this manual was supplied may incorporate one or more options not supplied in
24. al Instruments 16830 Chestnut Street City of Industry CA 91749 1580 Phone 626 934 1500 Fax 626 961 2538 TWX 910 584 1887 TDYANYL COID Web www teledyne ai com or your local representative Teledyne Analytical Instruments A 3 Appendix Model Ultra Trace 3000 A 3 Drawing List B 71228 Piping Diagram A 4 19 inch Relay Rack Panel Mount Figure A 1 Single 19 Rack Mounts dimensions in mm A 4 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Appendix A 5 Ultra Trace 3000 SERIES ANALYZERS APPLICATION NOTES ON PRESSURES AND FLOW RECOMMENDATIONS 3000 series analyzers require reasonably regulated sample pressures While the 3000 analyzers are not sensitive to variations of incoming pressure provided they are properly vented to atmospheric pressure The pressure must be maintained as to provide auseable flow rate trough the analyzer Any line attached to sample vent should be 1 4 or larger in diameter FLOW RATE RECOMMENDATIONS Ausable flow rate for a 3000 series analyzer is one which can be measured on the flowmeter This 1s basically 0 5 2 0 SLPM The optimum flow rate is 1 SLPM mid scale Note response time is dependent on flow rate alow flow rate will result in slow response to O2 changes in the sample stream The span flow rate should be the approximately same as the sample flow rate CELL PRESSURE CONCERNS The sensors used in 3000 series analyzers are optimized t
25. al laws and ensure that the exhaust stream vents to an appropriately controlled area ifrequired 3 3 2 Electrical Connections For safe connections no uninsulated wiring should be able to come in contact with fingers tools or clothing during normal operation CAUTION Use Shielded Cables Also use plugs that provide excellent EMI RFI protection The plug case must be connected to the cable shield and it must be tightly fastened to the analyzer with its fastening screws Ultimately it is the installer who ensures that the connections provide adequate EMI RFI sielding 3 3 2 1 Primary Input Power The power cord receptacle and fuse block are located in the same assembly Insert the power cord into the power cord receptacle CAUTION Power is applied to the instrument s circuitry as long as the instrument is connected to the power source The red _ switch on the front panel is for switching power on or off to the displays and out puts only The universal power supply requires a 85 250 V ac 47 63 Hz power source Fuse Installation The fuse block at the right of the power cord receptacle accepts US or European size fuses A jumper replaces the fuse in whichever fuse receptacle is not used Fuses are not installed at the factory Be sure to install the proper fuse as part of installation See Fuse Replace ment in chapter 5 maintenance 3 4 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Installation 3
26. ange ID contacts will track the instrument readings during calibration either zero or span TRACK is the factory default When HOLD is selected the analog outputs 0 1 VDC and 4 20 ma and the range ID contacts will freeze on their last state before entering one of the calibration modes When the instrument returns to the Analyze mode either by a successful or an aborted calibration there will be a three minute delay before the analog outputs and the range ID contacts start tracking again The concentration alarms freeze on their last state before entering calibration regardless of selecting HOLD or TRACK But when HOLD is selected the concentration alarms will remain frozen for the time displayed in the second line of the TRAK HLD menu after the analyzer returns to the Analyze mode The factory default is three minutes but the delay time is program mable To adjust to delay time use the Left or Right arrow keys When the time displayed on the second line blinks it can be adjusted by Pressing the Up or Down keys to increase or decrease its value The minimum delay is 1 minute the maximum is 30 This preference is stored in non volatile memory so that it is recovered if power is removed from the instrument 4 3 2 Setting up an Auto Cal When proper automatic valving is connected see chapter 3 installa tion the Analyzer can cycle itself through a sequence of steps that automati cally calibrates the instrument Note If you req
27. ange the password Change Password lt ENT gt Yes lt ESC gt No Press Enter to change the password either the default TETAI or the previously assigned password or press Escape to keep the existing pass word and move on If you chose Enterto change the password the password assignment screen appears TE TAI lt ENT gt To Proceed or A A A A A lt ENT gt To Proceed Enter the password using the lt gt arrow keys to move back and forth between the existing password letters and the AV arrow keys to change the letters to the new password The full set of 94 characters available for pass word use are shown in the table below Characters Available for Password Definition A B C D E F G H J K L M N O P Q R S T U V W X Y Z n a b c d e f g h i j k 1 m n o p q T S u v wW X y Z gt n amp f j 0 1 2 3 4 5 6 7 8 9 3 gt ip When you have finished typing the new password press Enter A verification screen appears The screen will prompt you to retype your password for verification A A A A A Retype PWD To Verify 4 8 DN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operation 4 Wait a moment for the entry screen You will be given clearance to proceed A A A A A lt ENT gt TO Proceed Use the arrow keys to retype your password and press Enter when finished Your password will be stored in the microprocessor and t
28. at our liability shall be limited to replacing or repairing at our factory without charge except for transportation or at customer plant at our option any material or construction in which defects become apparent within one year from the date of shipment except in cases where quotations or acknowledgements provide for a shorter period Components manufactured by others bear the warranty of their manufacturer This warranty does not cover defects caused by wear accident misuse neglect or repairs other than those performed by Teledyne or an autho rized service center We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse We reserve the right to employ any suitable material in the manufacture of our apparatus and to make any alterations in the dimensions shape or weight of any parts in so far as such alterations do not adversely affect our warranty Important Notice This instrument provides measurement readings to its user and serves as a tool by which valuable data can be gathered The information provided by the instrument may assist the user in eliminating potential hazards caused by his process however it is essential that all personnel involved in the use of the instrument or its interface with the process being measured be properly trained in the process itself as well as all instrument
29. ce the cell on cell holder with the screen side facing down 3 Place the cell block on top of the cell holder 4 Hold the cell block with one hand and rotate the collar clockwise 5 To determine if the assembly is fully seated rotate the cell block slightly 6 When fully seated the cell block can be rotated with three fingers but considerable drag will be felt 5 2 5 Cell Warranty The Class B 2CXL Micro Fuel cell is used in the Model Ultra Trace 3000 This cell is warranted for 6 months from the date of shipment Note any Addenda attached to the front of this manual for special information applying to your instrument With regard to spare cells warranty period begins on the date of ship ment The customer should purchase only one spare cell per section 5 2 1 Do not attempt to stockpile spare cells The B 2CXL cell is not designed for applications where CO is a major component in the sample however concentrations of 1 000 ppm or less will not adversely effect the cell performance Consult TAI for the appropriate cell for measuring trace oxygen in the presence of CO If a cell was working satisfactorily but ceases to function before the watranty period expires the customer will receive credit toward the purchase of a new cell If you have a warranty claim you must return the cell in question to the factory for evaluation If it is determined that failure is due to faulty work manship or material the cell will be repla
30. ced at no cost to you Note Evidence of damage due to tampering or mishandling will render the cell warranty null and void 5 4 PN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Maintenance 5 5 3 Fuse Replacement 1 Place small screwdriver in notch and pry cover off as shown in Figure 5 2 Fuse Block Cover Figure 5 2 Removing Fuse Block from Housing 2 To change between American and European fuses remove the single retaining screw flip Fuse Block over 180 degrees and replace screw 3 Replace fuse as shown in Figure 5 3 4 Reassemble Housing as shown in Figure 5 2 A Jumper Bar ee 23 Fuse Block 4 ah Cover American Fuses European Fuses Figure 5 3 Installing Fuses Teledyne Analytical Instruments 9 9 5 Maintenance Model Ultra Trace 3000 5 4 System Self Diagnostic Test 1 Press the System button to enter the system mode 2 Use the lt gt arrow keys to move to More and press Enter 3 Use the lt gt arrow keys to move to Self Test and press Enter The following failure codes apply Table 5 1 Self Test Failure Codes Power 0 OK 1 5 V Failure 2 15 V Failure 3 Both Failed Analog 0 OK 1 DAC A 0 1 V Concentration 2 DAC B 0 1 V Range ID 3 Both Failed Preamp 0 OK 1 Zero too high 2 Amplifier output doesn t match test input 3 Both Failed 5 5 Major Internal Components The Micro Fuel cell is accessed by unlatching an
31. choose Aut o Cal and press Enter again When the Span Zero values screen appears use the lt gt arrows to blink the Span or Zero OFF ON field Use AV arrows to set the OFF ON field to ON You can now turn these fields ON because there is a nonzero span interval defined 4 3 3 Password Protection If a password is assigned then setting the following system parameters can be done only after the password is entered span and zero settings alarm setpoints analysis range definitions switching between autoranging and manual override setting up an auto cal and assigning a new password However the instrument can still be used for analysis or for initiating a self test without entering the password If you have decided not to employ password security use the default password TETAI This password will be displayed automatically by the microprocessor The operator just presses the Enter key to be allowed total access to the instrument s features NOTE If you use password security it is advisable to keep a copy of the password in a separate safe location 4 6 DN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operation 4 4 3 3 1 Entering the Password To install a new password or change a previously installed password you must key in and ENTER the old password first If the default password is in effect pressing the ENTER button will enter the default TETAI pass word for you Press System to enter the Sys
32. d a system failure alarm e Extensive self diagnostic testing at startup and on demand with continuous power supply monitoring e Two way RFI protection e RS 232 serial digital port for use with a computer or other digital communication device e Four analog outputs two for measurement 0 1 V de and Isolated 4 20 mA dc and two for range identification e Convenient and versatile steel flush panel or rack mountable case with slide out electronics drawer 1 4 Model Designations Ultra Trace 3000 Standard model for sample under pressure Ultra Trace 3000 V Instrument configured for Vacuum Service 1 2 DP Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Introduction 1 1 5 Front Panel Operator Interface The standard Ultra Trace 3000 is housed in a rugged metal case with all controls and displays accessible from the front panel See Figure 1 1 The front panel has thirteen buttons for operating the analyzer a digital meter an alphanumeric display and a window for viewing the sample flowmeter Function Keys Six touch sensitive membrane switches are used to change the specific function performed by the analyzer Door Latch Telec lyne Analytical Instrument is Digital Meter ULTRA 7 ss TRACE orvan ACE 2000 A ANALYZE SYSTE F En sp Alphanumeric z AN zero Display ALARMS RANGE SCHH E SCAPE SET Foy Sample System Flow Indicator Standby Switch Data Entry Buttons Fu
33. d swinging open the front panel as described earlier Other internal components are accessed by removing the rear panel and sliding out the entire chassis See Figure 5 4 below The gas piping is illustrated in Figure 2 4 and the major electronic components locations are shown in Figure 2 5 in chapter 2 WARNING SEE WARNINGS ON THE TITLE PAGE OF THIS MANUAL The Ultra Trace 3000 contains the following major components e Analysis Section Micro Fuel Cell B 2CXL Cell block with stainless steel wetted parts Sample system 5 6 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Maintenance 5 e TEC cooler assembly e Power Supplys e Microprocessor e Displays 5 digit LED meter 2 line 20 character alphanumeric VFD display e RS 232 Communications Port e TEC temperature controller PCB See the drawings in the Drawings section in back of this manual for details X X X Pe Teledyne Analytical Instruments CAUTION 8 S oF ELECTRIC snoek DO NOT OPEN No USER SERVICEABLE PARTS INSIDE SEFVICE ONLY BY QUALIFIED PERSONNEL 85 250Vac 47 63HZ 0 9 A MAX USE 250V 1 0A T FUSE SAMPLE IN SEE MANUAL FOR INTERCONNECTION DETAIL Ml Ze EXHAUST OUT Figure 5 4 Rear Panel Screws To detach the rear panel remove only the 14 screws marked with an X 5 6 Cleaning If instrument is unmounted at time of cleaning disconnect the instru ment from the power source Close and latch the front panel access
34. door Clean outside surfaces with a soft cloth dampened slightly with plain clean water Do not use any harsh solvents such as paint thinner or benzene For panel mounted instruments clean the front panel as prescribed in the above paragraph DO NOT wipe front panel while the instrument is controlling your process Teledyne Analytical Instruments 5 7 5 Maintenance Model Ultra Trace 3000 5 7 Troubleshooting Problem Erratic readings of the Oxygen concentration as reported by the analyzer Possible Cause The analyzer may have been calibrated in an inaccurate fashion Solution Turn the analyzer off then turn back on again Press the System key when prompted by the analyzer Press System for default Values This will return the analyzer to its default settings in calibration and zero values If erratic behavior continues clean sensor contacts or replace the sensor Possible Cause Atmospheric Oxygen may be diffusing in through the vent and affecting the oxygen level which the sensor sees Solution Increase flow rate and or length or vent tubing in order to dilute or minimize the diffusion of oxygen from the vent back to the sensor Problem Inaccurate zero operation i e the user has zeroed the analyzer accidentally on gas much higher than one would normally use for a zero gas Solution Turn the analyzer off then back on again Press the System key when prompted by the analyzer Press System for default Va
35. e cell is a diffusion membrane of Teflon whose thickness is very accurately controlled Beneath the diffusion membrane lies 2 2 DP Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operational Theory 2 the oxygen sensing element the cathode with a surface area almost 4 cm The cathode has many perforations to ensure sufficient wetting of the upper surface with electrolyte and it is plated with an inert metal The anode structure is below the cathode It is made of lead and has a proprietary design which is meant to maximize the amount of metal available for chemical reaction At the rear of the cell just below the anode structure is a flexible membrane designed to accommodate the internal volume changes that occur throughout the life of the cell This flexibility assures that the sensing mem brane remains in its proper position keeping the electrical output constant The entire space between the diffusion membrane above the cathode and the flexible rear membrane beneath the anode is filled with electrolyte Cathode and anode are submerged in this common pool They each have a conductor connecting them to one of the external contact rings on the contact plate which is on the bottom of the cell 2 2 3 Electrochemical Reactions The sample gas diffuses through the Teflon membrane Any oxygen in the sample gas is reduced on the surface of the cathode by the following HALF REACTION O 2H O 4e 40H cat
36. ell a Check your span and zero gas to make sure it is within specifications b Check for leaks up stream from the cell where oxygen may be leaking into the system c Check if more purging time with zero calibration gas is needed If there are no leaks and the span gas is within specification replace the cell The failure alarm and the failure message will reset after re entering the zero mode again 5 2 PN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Maintenance 5 5 2 3 Removing the Micro Fuel Cell The Micro Fuel cell is located inside the stainless steel cell block behind the front panel see Figure 5 1 To remove an existing cell 1 Remove power to the instrument by unplugging the power cord at the power source 2 Open the front panel door by pressing the release button on the top right corner of the door all the way in with a narrow gauge tool 3 With one hand hold the top of the cell block while unscrewing the plastic ring holder Once the plastic ring is loose remove the top of the cell block COLLAR CELL BLOCK SENSOR O RING kh CELL HOLDER Figure 5 1 Removing the Micro Fuel Teledyne Analytical Instruments 5 3 5 Maintenance Model Ultra Trace 3000 5 2 4 Installing a New Micro Fuel Cell Before installing a new cell check the O ring in the base of the cell holder Replace if worn or damaged 1 Verify that the O ring is properly located 2 Pla
37. enoid 1 hot EXHAUST hot Circuitry Solenoid 4 hot SAMPLE return If m Solenoid 2 return ZERO return Necessary Solenoid 3 return SPAN return Solenoid 1 return EXHAUST return i Solenoid 4 return Degre Sed wo en Figure 3 5 Remote Probe Connections The voltage from these outputs is nominally 0 V for the OFF and 15 V de for the ON conditions The maximum combined current that can be pulled from these output lines is 100 mA If two lines are ON at the same time each must be limited to 50 mA etc If more current and or a different voltage is required use a relay power amplifier or other matching circuitry to provide the actual driving current In addition each individual line has a series FET with a nominal ON resistance of 5 ohms 9 ohms worst case This can limit the obtainable voltage depending on the load impedance applied See Figure 3 6 LOAD 1 gt gt Figure 3 6 FET Series Resistance Teledyne Analytical Instruments 3 9 3 Installation Model Ultra Trace 3000 3 4 Installing the Micro Fuel Cell The Micro Fuel Cell Model B 2CXL is not installed in the cell block when the instrument is shipped Install it before the analyzer is placed in service The Micro Fuel cell is located inside the stainless steel cell block behind the front panel see Figure 3 8 To install the cell 1 Remove power to the instrument by unplugging the power cord at
38. fixed range regardless of the oxygen concentration detected If the concentration exceeds the upper limit of the range the DC output will saturate at 1 V dc 20 mA at the current output However the digital readout and the RS 232 output of the concentra tion are unaffected by the fixed range They continue to read accurately with full precision See Front Panel description in Chapter 1 The automatic fourth range is always 0 1000 ppm and is not program mable 4 7 1 Setting the Analog Output Ranges To set the ranges enter the range function mode by pressing the Range button on the front panel L 250 ppb M 1 ppm H 10 ppm Mode AUTO Use the lt gt arrow keys to blink the range to be set low L medium M or high H 4 20 DN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operation 4 Use the AV arrow keys to enter the upper value of the range all ranges begin at 0 Repeat for each range you want to set Press Enter to accept the values and return to Analyze mode See note below Note The ranges must be increasing from low to high for example if range 1 is set as 0 500 ppb and range 2 is set as 0 10 ppm range 3 cannot be set as 0 5 ppm since it is lower than range 2 Ranges and alarms are set in ppb or ppm units depending on concen tration All concentration data outputs change from ppb units to ppm when the concentration is above 1 0 ppm Range Low L is always a ppb range and cannot be se
39. he system will immediately switch to the Analyze screen and you now have access to all instrument functions If all alarms are defeated the Analyze screen appears as 0 0 ppm Anlz Range 0 100 If an alarm is tripped the second line will change to show which alarm itis 0 0 ppm Anlz AL 1 NOTE If you log off the system using the logout function in the system menu you will now be required to re enter the password to gain access to Span Zero Alarm and Range functions 4 3 4 Logout The Logout function provides a convenient means of leaving the analyzer in a password protected mode without having to shut the instrument off By entering Logout you effectively log off the instrument leaving the system protected against use until the password is reentered To log out press the System button to enter the System function TRAK HLD Auto Cal PSWD Logout More Use the lt gt arrow keys to position the blinking over the Logout function and press Enterto Log out The screen will display the message Protected Until Password Reentered 4 3 5 System Self Diagnostic Test The Model 3000 has a built in self diagnostic testing routine Pre programmed signals are sent through the power supply output board and sensor circuit The return signal is analyzed and at the end of the test the A Teledyne Analytical Instruments 4 9 4 Operation Model Ultra Trace 3000 status of each function is displayed on the screen eithe
40. higher than 0 1 ppm a zero gas such as nitrogen with oxygen level below 0 1 ppm may be required The recommended span gas concentration is between 7 0 to 9 0 ppm oxygen in nitrogen and will require calibration be performed in the 0 10 ppm analyzer range 44 1 Zero Cal The B 2CXL cell has a zero offset of less than 0 1 ppm oxygen Nor mally the offset slowly decreases during the first 7 to 10 days of operation and is expected to reach a steady value after this time DP Teledyne Analytical Instruments 4 11 4 Operation Model Ultra Trace 3000 Generally the value of the zero offset is part of the oxygen reading of the sample gas as shown by the analyzer readout As an example a reading of 0 5 ppm oxygen may include 0 4 ppm oxygen in the sample gas and a 0 1 ppm zero offset The determination of the zero offset requires the use of oxygen free gas to the analyzer We recommend the use of nitrogen gas with a scrubber to assure oxygen levels below 20 ppb The user may decide to eliminate the zero offset for improved accu racy If so desired the analyzer is equipped to provide this function How ever we do not recommend carying out the cal zero during the first 10 days of the operation of the cell The Zero button on the front panel is used to enter the zero calibration function Zero calibration can be performed in either the automatic or manual mode In the automatic mode an internal algorithm compares consecutive readings from
41. hode Four electrons combine with one oxygen molecule in the presence of water from the electrolyte to produce four hydroxy ions When the oxygen is reduced at the cathode lead is simultaneously oxidized at the anode by the following HALF REACTION Pb 20H Ph H O 2e anode Two electrons are transferred for each atom of lead that is oxidized Therefore it takes two of the above anode reactions to balance one cathode reaction and transfer four electrons The electrons released at the surface of the anode flow to the cathode surface when an external electrical path is provided The current is propor tional to the amount of oxygen reaching the cathode It is measured and used to determine the oxygen concentration in the gas mixture The overall reaction for the fuel cell is the SUM of the half reactions above or 2Pb O gt 2PbO W Teledyne Analytical Instruments 2 3 2 Operational Theory Model Ultra Trace 3000 These reactions are specific to oxygen as long as no gaseous components capable of oxidizing lead such as iodine bromine chlorine and fluorine are present in the sample In the absence of oxygen no current is generated 2 2 4 The Effect of Pressure In order to state the amount of oxygen present in the sample in ppb or parts per million of the gas mixture it is necessary that the sample diffuse into the cell under constant pressure If the total pressure increases the rate that ox
42. ing inert gas blanketing Air separation and liquefaction Chemical reaction monitoring Semiconductor manufacturing Petrochemical process control Quality assurance Gas analysis certification 3 Main Features of the Analyzer The Model 3000 Ultra Trace Oxygen Analyzer is sophisticated yet simple to use The main features of the analyzer include A 2 line alphanumeric vacuum fluorescent display VFD screen driven by microprocessor electronics that continuously prompts and informs the operator High resolution accurate readings of oxygen content from low ppm levels through 25 Large bright meter readout Stainless steel cell block wetted surfaces Teledyne Analytical Instruments 1 1 1 Introduction Model Ultra Trace 3000 e Advanced Micro Fuel Cell designed for trace analysis has a 0 250 ppb low range with less than a 100 ppb offset and six months warranty and an expected lifetime of one year e Versatile analysis over a wide range of applications e Microprocessor based electronics 8 bit CMOS microprocessor with 32 kB RAM and 128 kB ROM e Three user definable output ranges from 0 250 ppb through 0 1000 ppm allow best match to users process and equipment plus a fixed 1000 ppm over range e Auto Ranging allows analyzer to automatically select the proper preset range for a given measurement Manual override allows the user to lock onto a specific range of interest e Two adjustable concentration alarms an
43. lly activate the alarms 4 18 DN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operation 4 The defeat function can also be used to reset a latched alarm See procedures below If you are using password protection you will need to enter your password to access the alarm functions Follow the instructions in section 4 3 3 to enter your password Once you have clearance to proceed enter the Alarm function Press the A arm button on the front panel to enter the A arm function Make sure that AL 1 is blinking AL 1 AL 2 Choose Alarm Set up alarm 1 by moving the blinking over to AL 1 using the lt gt arrow keys Then press Enter to move to the next screen AL 1 10 000 ppm HI Dft N Fs N Ltch N Five parameters can be changed on this screen e Value of the alarm setpoint AL 1 ppm oxygen value can be set from 0 to 999 ppb 1 000 1000 00 ppm e Out of range direction HI or LO e Defeated Dft Y N Yes No e Failsafe Fs Y N Yes No e Latching Ltch Y N Yes No e To define the setpoint use the lt gt arrow keys to move the blinking over to AL 1 Then use the AV arrow keys to change the number Holding down the key speeds up the incrementing or decrementing Remember the setpoint units are ppm O e To set the other parameters use the lt gt arrow keys to move the blinking over to the desired parameter Then use the AV arrow keys to change the parameter e Once the parameters
44. lues This will return the analyzer to its default settings in calibration and zero values Now proceed to carefully calibrate and zero the analyzer 5 8 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Appendix Appendix A 1 Specifications Packaging General Purpose e Flush panel mount Standard e Relay rack mount Contains one instrumentsin one 19 relay rack mountable plate Optional Sensor Teledyne B 2CXL trace analysis Micro Fuel Cell Cell Block 316 stainless steel Sample System All wetted parts of 316 stainless steel 90 Response Time Less than 30 seconds at 25 C 77 F on 10 ppm range 120 seconds on 250 ppb range programmable response in ppb range Ranges Three user definable ranges from 0 250 ppb to 0 1 000 ppm plus over range of 0 1 000 ppm Autoranging with range ID output Alarms Onesystem failure alarm contact to detect power failure or sensor zero and span failure Two adjustable concentration threshold alarm contacts with fully programmable setpoints Displays 2 line by 20 character VFD screen and one 5 digit LED display Digital Interface Full duplex RS 232 communications port Teledyne Analytical Instruments A 1 Appendix Model Ultra Trace 3000 Power Universal power supply 85 250 V ac at 47 63 Hz Operating Temperature 5 40 C Accuracy 2 of full scale for all ranges except 0 250 ppb range at constant temperature For 0 250
45. minals are used They generate a steady preset voltage or current when using the current outputs to represent a particular range The following table gives the range ID output for each analysis range Range Voltage V Current mA LO 0 25 8 MED 0 50 12 HI 0 75 16 0 1000ppm 1 00 20 s Teledyne Analytical Instruments 4 23 4 Operation Model Ultra Trace 3000 4 24 PN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Maintenance 5 Maintenance 5 1 Routine Maintenance Aside from normal cleaning and checking for leaks at the gas connec tions routine maintenance is limited to replacing Micro Fuel cells and fuses and recalibration For recalibration see Section 4 4 Calibration WARNING SEE WARNINGS ON THE TITLE PAGE OF THIS MANUAL 5 2 Cell Replacement The B 2CXL Micro Fuel Cell is a sealed electrochemical transducer with no electrolyte to change or electrodes to clean When the cell reaches the end of its useful life it is replaced The spent fuel cell should be dis carded according to local regulations This section describes fuel cell care as well as when and how to replace it 5 2 1 Storing and Handling Replacement Cells To have a replacement cell available when it is needed TAI recom mends that one spare cell be purchased 9 10 months after commissioning the Ultra Trace 3000 or shortly before the end of the cell s one year warranty period CAUTION Do not stockpile cells The warrant
46. n the rack panel See Appendix for dimensions of the mounting panel PN Teledyne Analytical Instruments 3 1 3 Installation Model Ultra Trace 3000 Latch 6 7 lt gt 1 0 Figure 3 1 Front Panel of the Model Ultra Trace 3000 All operator controls are mounted on the control panel which is hinged on the left edge and doubles as the door that provides access to the sensor and cell block inside the instrument The door is spring loaded and will swing open when the button in the center of the latch upper right corner is pressed all the way in with a narrow gauge tool less than 0 18 inch wide such as a small hex wrench or screwdriver Allow clearance for the door to open in a 90 degree arc of radius 7 125 inches See Figure 3 2 Provide 1 inch of air clearance at the bottom and sides of the instrument for proper air flow for TEC fun Failure to provide such a clearance will cause damage to the TEC system Do not restrict this openings Figure 3 2 Required Front Door Clearance 3 2 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Installation 3 3 3 Rear Panel Connections Figure 3 3 shows the Model Ultra Trace 3000 rear panel There are ports for gas inlet and outlet power communication and both digital and analog concentration output Pe Teledyne Analytical Instruments CAUTION RISK OF ELECTRIC stock OPEN No USER SE S INSIDE SERVICE ON 7 ol 85 250Vac 47 63Hz 0 9 A MAX
47. ncentration reading is acceptable and manually exits the function 4 4 2 1 Auto Mode Spanning Press Span to enter the span function The screen that appears allows you to select whether the span calibration is to be performed automatically or manually Use the AV arrow keys to toggle between AUTO and MAN span settling Stop when AUTO appears blinking on the display Span Settling AUTO lt ENT gt For Next Press Enter to move to the next screen Calib Holding time Cal hold 5 min 4 14 DN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operation 4 This menu allows the operator to set the time the analyzer should be held in the AUTO span mode after the readings of the analyzer settle Five minutes is the default but it could be adjusted anywhere from 1 to 60 min utes by using the UP or DOWN keys Press Enter to move to the next screen Span Val 008 00 ppm lt ENT gt Span lt UP gt Mod Use the AV arrow keys to enter the oxygen concentration mode 999 99 ppm is maximum value of span gas allowed Use the lt gt arrow keys to blink the digit you are going to modify Use the AV arrow keys again to change the value of the selected digit When you have finished typing in the concentration of the span gas you are using press Enter to begin the Span calibration HH HH ppm Span Slope ppm s The beginning span value is shown in the upper left corner of the display As the span reading settle
48. nction Buttons Figure 1 1 Model Ultra Trace 3000 Front Panel e Analyze Perform analysis for oxygen content of a sample gas e System Perform system related tasks described in detail in chapter 4 Operation e Span Span calibrate the analyzer e Zero Zero calibrate the analyzer DN Teledyne Analytical Instruments 1 3 1 Introduction Model Ultra Trace 3000 e Alarms Setthe alarm setpoints and attributes e Range Set up the 3 user definable ranges for the instrument Data Entry Keys Six touch sensitive membrane switches are used to input data to the instrument via the alphanumeric VFD display e Left amp Right Arrows Select between functions currently displayed on the VFD screen Up amp Down Arrows Inerement or decrement values of functions currently displayed e Enter Moves VFD display on to the next screen in a series If none remains returns to the Analyze screen e Escape Moves VFD display back to the previous screen in a series If none remains returns to the Analyze screen Digital Meter Display The meter display is a Light Emitting Diode LED device that produces large bright 7 segment numbers that are legible in any lighting It produces a continuous readout from 0 999 ppb and then switches to a continuous ppm readout from 0 9999 9 ppm It is accurate across all analysis ranges without the discontinuity inherent in analog range switching Alphanumeric Interface Screen The VFD screen is an easy
49. nd on U A Q Alarm Relays The nine alarm circuit connector pins connect to the internal alarm relay contacts Each set of three pins provides one set of Form C relay contacts Each relay has both normally open and normally closed Teledyne Analytical Instruments 3 5 3 Installation Model Ultra Trace 3000 contact connections The contact connections are shown in Table 3 2 They are capable of switching up to 3 amperes at 250 V ac into a resistive load The connectors are Threshold Alarm 1 Threshold Alarm 2 System Alarm e Can be configured as high actuates when concen tration is above threshold or low actuates when concentration is below threshold e Can be configured as failsafe or nonfailsafe e Can be configured as latching or nonlatching e Can be configured out defeated e Can be configured as high actuates when concen tration is above threshold or low actuates when concentration is below threshold e Can be configured as failsafe or nonfailsafe e Can be configured as latching or nonlatching e Can be configured out defeated Actuates when DC power supplied to circuits is unacceptable in one or more parameters Permanently configured as failsafe and latching Cannot be de feated Actuates if self test fails Reset by pressing button to remove power Then press C again and any other button EXCEPT Systemtoresume Further detail can be found in chapter 4 section 4 5 Table 3
50. o function at atmospheric pressure At pressures other than atmospheric the diffusion rate of O2 will be different than optimum value Higher pressures will produce faster O2 diffusion rates resulting in higher O2 reading and shorter cell life To use a 3000 series analyzer at acell pressure other than atmospheric the analyzer must be calibrated with a known calibration gas at the new cell pressure to adjust for the different diffusion rate Cell pressures below 2 3 atmospheric are not recommended because they tend to cause excessive internal expansion which may resultin seal failure For operation at cell pressures other than atmospheric care must be taken not to change the sample pressure rapidly or cell damage may occur For cell pressures above atmospheric caution must be exercised to avoid over pressuring the cell holder FLOW CONTROL VALVE All 3000 series analyzers have a flow control device This device is a needle valve and serves two functions in the sample path The first function is to limit the flow rate of the sample through the analyzer The needle valve is chosen to operate over a specific range of pressures and provide a useable flow rate over thatrange The second function that the needle valve provides is a pressure drop This device is selected to provide the only significant pressure drop in the sample path Teledyne Analytical Instruments A 5 Appendix Model Ultra Trace 3000 BY PASS To improve the system resp
51. ocessor and the instrument automatically returns to the Analyze mode 4 4 1 3 Cell Failure Cell failure in the Ultra Trace 3000 is usually associated with inability to zero the instrument down to a satisfactorily low ppm reading correspond ing to a current of 2 nanoamps approx 1 ppm When this occurs the A Teledyne Analytical Instruments 4 13 4 Operation Model Ultra Trace 3000 instrument returns back to analyzer mode without taking the zero calibration The Ultra Trace 3000 system alarm trips and the LCD displays a failure message ppm Anlz CELL FAIL ZERO HIGH Before replacing the cell a Check your span gas to make sure it is within specifications b Check for leaks up stream from the cell where oxygen may be leaking into the system c Check if more purging time with Zero calibration gas is needed If there are no leaks and the span gas is within specification replace the cell as described in chapter 5 Maintenance The failure alarm and failure message will reset after entering the Zero mode 4 4 2 Span Cal The Span button on the front panel is used to span calibrate the ana lyzer Span calibration can be performed using the automatic mode where an internal algorithm compares consecutive readings from the sensor to determine when the output matches the span gas concentration Span cali bration can also be performed in manual mode where the operator deter mines when the span co
52. ons that regulate the internal operations of the analyzer e Auto Cal setup e Show negative readings e Password assignment e Set digital filter e Self Test initiation e Checking software version e Logging out e Zero Used to set up a zero calibration e Span Used to set up a span calibration e Alarms Used to set the alarm setpoints and determine whether each alarm will be active or defeated HI or LO acting latching and or failsafe e Range Used to set up three analysis ranges that can be switched automatically with auto ranging or used as individual fixed ranges Any function can be selected at any time by pressing the appropriate button unless password restrictions apply The order as presented in this manual is appropriate for an initial setup Each of these functions is described in greater detail in the following procedures The VFD screen text that accompanies each operation is repro duced at the appropriate point in the procedure ina Monospaced type style Pushbutton names are printed in Oblique type 4 2 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operation 4 ANALYZE SYSTEM Perform Oxygen Analysis 0 the Sample v TRAK HLD ZERO Set l sbument Y pan Perform ALARMS Self Diagnostic ezt Set Instrument Zero Vv Initiate Automatic Set Alarm Calibration Setpoints
53. onse a by pass can be added to increase the sample flow rate to the analyzer by a factor of ten A by pass provides a sample flow path around the analyzer of 2 18 SCFH typically CONVERSIONS 1 PSI 2 04 INCHES OF MERCURY in Hg 1 SCFH 0 476 SLPM NOTE The MSDS on this material is available upon request through the Teledyne Environmental Health and Safety Coordinator Contact at 626 934 1592 A 6 DP Teledyne Analytical Instruments
54. ppb range accuracy is as follows 5 ppb at constant temperature 10 ppb over temperature variation of 20 30 C 15 ppb over temperature variation of 15 35 C All accuracy specifications are contingent upon the completion of zero ans span calibra tion All accuracy is established at constant pres sure and equilibrium has been established Analog outputs 0 1 V dc percent of range 0 1 V dc range ID 4 20 mA dc isolated percent of range 4 20 mA dc isolated range ID Dimensions 19 cm high 24 9 cm wide 31 cm deep 6 96 in high 8 7 in wide 12 2 in deep A 2 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Appendix A 2 Recommended 2 Year Spare Parts List Qry PART NUMBER DESCRIPTION 1 CP2037 TEC Cooler 1 C65507 Back Panel Board 1 C62371 B Front Panel Board 1 C71410 Preamplifier Board Instruction 1 C70941 TEC Controller PCB 1 C62365 A Main Computer Board 3 F9 Fuse 1A 250V 3AG Slow Blow 3 F1275 Fuse 1A 250V 5x20mm European Slow Blow 1 0286 O ring 1 C6689 B2CXL Micro Fuel Cell do not store for more than 6 months 1 CP1798 50 pin D sub interface connector 1 F1498 Fan 50 CP1799 Pins for CP1798 connector A minimum charge is applicable to spare parts orders Note Orders for replacement parts should include the part number if available and the model and serial number of the instrument for which the parts are intended Orders should be sent to Teledyne Analytic
55. r as OK or as a num ber between 1 and 3 See System Self Diagnostic Test in chapter 5 for number code The self diagnostics are run automatically by the analyzer whenever the instrument is turned on but the test can also be run by the operator at will To initiate a self diagnostic test during operation Press the System button to start the System function TRAK HLD Auto Cal PSWD Logout More Use the lt gt arrow keys to blink More then press Enter Version Self Test Neg N Filter 5 Use the lt gt arrow keys again to move the blinking to the Self Test function The screen will follow the running of the diagnostic RUNNING DIAGNOSTIC Testing Preamp 83 During preamp testing there is a countdown in the lower right corner of the screen When the testing is complete the results are displayed Power OK Analog OK Preamp 3 The module is functioning properly if it is followed by OK A number indicates a problem in a specific area of the instrument Refer to Chapter 5 Maintenance and Troubleshooting for number code information The results screen alternates for a time with Press Any Key To Continue Then the analyzer returns to the initial System screen 4 3 6 Version Screen Move the lt gt arrow key to More and press Enter With Version blinking press Enter The screen displays the manufacturer model and software version information 4 10 DN Teledyne Analytical Instruments Ultra
56. rature control PCB a separate power supply a thermistor and a special insulated cellblock The system is used to regulate the cell temperature at 11 degrees C Operating the Micro fuel Cell at a low 2 4 Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer temperature minimizes the cell offset typically less than 75 PPB A second benefit of the TEC system is that by regulating the cell temperature the analyzer becomes tolerant of thermal transients The TEC module is a solid state semiconductor heat pump Passing DC current through the TEC module produces heat flow though the device One side of the device will become hot and the other side will become cold The TEC module is attached to a heat sink which is cooled by a fan This is required to maintain the hot side at an acceptable temperature The hot side temperature limits the overall performance of the TEC module The fan draws air into the bottom of the analyzer this air is forced over the heat sink and exits through the left side of the analyzer The power for the TEC device is supplied by a Pulse Width Modulated PWM proportional switching temperature controller The temperature controller PCB supplies a 12 volt pulse whose duty cycle is proportional to the cooling required The tempera ture controller PCB uses a thermistor to monitor the temperature of the of cell block Power for the fan and the temperature controller PCB is supplied by a separate 12VDC power supply
57. re 2 7 is a block diagram of the Analyzer electronics Universal Power Supply Front Panel Display Board k TEC Power Supply eee oe ee ee oe oe Slide out Electronics Drawer Temperature Controller Board Motherboard Preamplifier PCB Figure 2 6 Electronic Component Location Inside the Model Ultra Trace 3000 2 8 DN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operational Theory 2 Current Second to Voltage Stage Amplifier Amplifier Ato D Converter aem System __ Micro Processor Displays Self Test DtoA Converter Figure 2 7 Block Diagram of the Model Ultra Trace 3000 Electronics W Teledyne Analytical Instruments 2 9 2 Operational Theory Model Ultra Trace 3000 In the presence of oxygen the cell generates a current A current to voltage amplifier converts this current to a voltage which is further amplified in the second stage amplifier The output from the second stage amplifier is sent to an 18 bit analog to digital converter controlled by the microprocessor The digital concentration signal along with input from the control panel is processed by the microprocessor and appropriate control signals are directed to the display alarms and communications port The same digital information is also sent to a 12 bit digital to analog converter that produces the 4 20 mA de and the 0 1 V dc analog concentration signal outputs and
58. ro Then the analyzer automatically returns to the Analyze mode Because the reading of the slope is not very sensitive it is recom mended that zero gas be purging a few minutes before starting the Auto mode zeroing This will ensure cell stability on the new Zero settings 4 4 1 2 Manual Mode Zeroing Press Zero to enter the Zero function The screen that appears allows you to select between automatic or manual zero calibration Use the AV keys to toggle between AUTO and MAN zero settling Stop when MAN appears blinking on the display Zero Settling Man lt ENT gt To Begin Press Enter to begin the zero calibration After a few seconds the first of five zeroing screens appears The number in the upper left hand corner is the first stage zero offset The microprocessor samples the output at a prede termined rate It calculates the differences between successive samplings and displays the rate of change as Slope a value in parts per million per second ppm s HHEH ppm Zero Slope ppm s NOTE It takes several seconds for the true Slope value to display Wait about 10 seconds Then wait until Slope is sufficiently close to zero before pressing Enter to finish zeroing Generally you have a good zero when Slope is less than 0 05 ppm s for about 30 seconds When Slope is close enough to zero press Enter Ina few seconds the screen will update Once zero settling is completed the information is stored in the micropr
59. s the screen displays and updates informa tion on Slope Spanning automatically ends when the span output corre sponds within tolerance to the value of the span gas concentration Then the instrument automatically returns to the analyze mode 4 4 2 2 Manual Mode Spanning Press Span to start the Span function The screen that appears allows you to select whether the span calibration is to be performed automatically or manually Span Settling MAN lt ENT gt For Next Use the AV keys to toggle between AUTO and MAN span settling Stop when MAN appears blinking on the display Press Enter to move to the next screen Press Enter to move to the next screen Calib Holding time Cal hold 5 min This menu allows the operator to set the time the analyzer should be held in the AUTO span mode It does not have any effect in the MANual mode Just press Enter key to continue W Teledyne Analytical Instruments 4 15 4 Operation Model Ultra Trace 3000 Span Val 008 00ppm lt ENT gt Span lt UP gt Mod Press A lt UP gt to permit modification Mod of span value Use the arrow keys to enter the oxygen concentration of the span gas you are using 999 99 is maximum value of span gas The lt gt arrows choose the digit and the AV arrows choose the value of the digit Press Enter to enter the span value into the system and begin the span calibration Once the span has begun the microprocessor samples the o
60. seful life the zero offset begins to rise until the analyzer finds the zero unsatisfactory Nevertheless feeding the wrong span 4 16 DN Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Operation 4 gas or electronics failure could set this feature off at the end of the span Consider this before replacing the cell 4 5 Switching of Sample Streams The Model Ultra Trace 3000 may be used to monitor more than one type of sample gas such as nitrogen and helium or two streams of nitrogen containing low less than 1 ppm and high 500 to 1000 ppm concentrations of oxygen For these applications we recommend the following parameters and conditions be followed 1 While switching streams such as between nitrogen and helium both the gas streams must be made to flow continuosly even through only one stream is going through the analyzer at one time Failure to do so may result in spikes in output when the dead ended stream is switched in For more discussion of this read the Technical Bulletin TB00001 dated 3 98 This Bulletin is available upon request from TATAI 2 The analyzer is capable of handling the switching of streams without effecting the sensitivity of the cell This switching of streams between nitrogen and helium will not require SPAN adjustment However care must be taken to keep sample flow rates in the desired levels as the flowmeter is calibrated for air and nitrogen streams only 3 The analyzer can wi
61. t higher than 1000 ppb nor lower than 10 ppb Ranges medium M and High H can only be set in ppm The Medium M range can be set between and 10 ppm while the high H range can be set between 10 and 1000 ppm 4 7 2 Fixed Range Analysis The autoranging mode of the instrument can be overridden forcing the analyzer DC outputs to stay in a single predetermined range To switch from autoranging to fixed range analysis enter the range function by pressing the Range button on the front panel Use the lt gt arrow keys to move the blinking over AUTO Use the AV arrow keys to switch from AUTO to FX L FX M or FX H to set the instrument on the desired fixed range low medium or high L 250ppb M 1 ppm H 10 ppm Mode FX L or L 250 ppb M 1 ppm H 10 ppm Mode FX M or L 250 ppb M 1 ppm H 10 ppm Mode FX H Press Escape to re enter the Analyze mode using the fixed range NOTE When performing analysis on a fixed range if the oxygen concentra tion rises above the upper limit or default value as established by the operator for that particular range the output saturates at 1 V dc or 20 mA However the digital readout and the RS 232 output continue to read the true value of the oxygen concentration regardless of the analog output range Teledyne Analytical Instruments 4 21 4 Operation Model Ultra Trace 3000 4 8 The Analyze Function Normally all of the functions automatically switch back to the Analyze
62. tem mode TRAK HLD Auto Cal PSWD Logout More Use the lt gt arrow keys to scroll the blinking over to PSWD and press Enter to select the password function Either the default TETAI password or AAAAA place holders for an existing password will appear on screen de pending on whether or not a password has been previously installed TETATI Enter PWD or A A A A A Enter PWD The screen prompts you to enter the current password If you are not using password protection press Enter to accept TETAI as the default pass word If a password has been previously installed enter the password using the lt gt arrow keys to scroll back and forth between letters and the AV arrow keys to change the letters to the proper password Press Enter to enter the password If the password is accepted the screen will indicate that the password restrictions have been removed and you have clearance to proceed PSWD Restrictions Removed In a few seconds you will be given the opportunity to change this password or keep it and go on Change Password lt ENT gt Yes lt ESC gt No Press Escape to move on or proceed as in Changing the Password below Teledyne Analytical Instruments 4 7 4 Operation Model Ultra Trace 3000 4 3 3 2 Installing or Changing the Password If you want to install a password or change an existing password proceed as above in Entering the Password When you are given the oppor tunity to ch
63. tes on Pressures and Flow A 5 Teledyne Analytical Instruments V Model Ultra Tace 3000 DANGER COMBUSTIBLE GAS USAGE WARNING This is a general purpose instrument designed for use in a nonhazardous area It is the customer s responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks always exist The customer should ensure that the principles of operation of this equipment is well understood by the user Misuse of this product in any manner tampering with its components or unau thorized substitution of any component may adversely affect the safety of this instrument Since the use of this instrument is beyond the control of Teledyne no responsibility by Teledyne its affiliates and agents for damage or injury from misuse or neglect of this equipment is implied or assumed vi Teledyne Analytical Instruments Ultra Trace Oxygen Analyzer Introduction 1 Introduction 1 1 Overview The Teledyne Analytical Instruments Model 3000 Ultra Trace Oxygen Analyzer is a versatile microprocessor based instrument for detecting oxygen at the parts per billion ppb level in a variety of gases This manual covers the Model Ultra Trace 3000 General Purpose flush panel and or rack mount units only These units are for indoor use in a nonhazardous environment 1 2 Typical Applications A few typical applications of the Model Ultra Trace 3000 are 1 Monitor
64. the standard instrument Commonly available options are listed below with check boxes Any that are incorporated in the instrument for which this manual is supplied are indicated by acheck mark in the box Instrument Serial Number Options Included in the Instrument with the Above Serial Number U Ultra Trace 3000 V Instrument configured for Vacuum Service L 19 Rack Mnt The 19 Relay Rack Mount units are available with one Ultra Trace 3000 series analyzers installed in a standard 19 panel and ready to mount in a standard rack Teledyne Analytical Instruments iii Model Ultra Tace 3000 Table of Contents Introduction TOA SOV EINIGW rete ereen ha weds Aa stede 1 1 1 2 Typical ApplicatiOnS assente ade 1 1 1 3 Main Features of the Analyzer nnn 1 1 1 4 Model DSsiGMAONS aaneen nt ndhelen 1 2 1 5 Front Panel Operator Interface 1 3 1 6 Rear Panel Equipment Interface ennen 1 5 Operational Theory 2d SIMPOGUIGION nnen nerede ewes 2 1 2 2 Micro Fuel Cell Sensor nnen neven enneeren eenn 2 1 2 2 1 Principles of Operation nennen 2 1 2 2 2 Anatomy of a Micro Fuel Gell 2 2 2 2 3 Electrochemical Reactions nnen 2 3 2 2 4 The Effect of Pressure nanne 2 4 2 2 5 Calibration Characteristics nnen 2 4 2 2 6 TEC Cooling System tete 2 4 2 3 Sample System eooneeeeeeooeeeeeneeeoeeternnnrrrrrrnnnrrreerernn e 2 5 2 4 Electronics and Signal Processing nnen
65. thstand repeated switching of streams between low oxygen levels of less than 1 ppm and high oxygen levels of 5000 to 1000 ppm Again care must be taken to ensure proper and continuous flows at all times 4 5 1 Special Notes on Hydrogen Gas Stream The standard oxygen cell Model B 2CXL though tested in nitrogen gas streams is expected to perform well in hydrogen streams However for certain applications such as continuous monitoring of very low levels of oxygen in hydrogen the user may request cells that have been specifically tested in hydrogen streams Consult Teledyne for further details 4 6 The Alarms Function The Model Ultra Trace 3000 is equipped with 2 fully adjustable con centration alarms and a system failure alarm Each alarm has a relay witha DP Teledyne Analytical Instruments 4 17 4 Operation Model Ultra Trace 3000 set of form C contacts rated for 3 amperes resistive load at 250 V ac See Figure 3 5 in Chapter 3 nstallation and or the Interconnection Diagram included at the back of this manual for relay terminal connections The system failure alarm has a fixed configuration as described in chapter 3 Installation The concentration alarms can be configured from the front panel as either high or low alarms by the operator The alarm modes can be set as latching or non latching and either failsafe or non failsafe or they can be defeated altogether The setpoints for the alarms are also established using this
66. to the Micro fuel Cell In the battery all reactants are stored within the cell whereas in the Micro fuel Cell one of the reactants oxygen Teledyne Analytical Instruments 2 1 2 Operational Theory Model Ultra Trace 3000 comes from outside the device as a constituent of the sample gas being analyzed The Micro Fuel Cell is therefore a hybrid between a battery and a true fuel cell All of the reactants are stored externally in a true fuel cell 2 2 2 Anatomy of a Micro Fuel Cell The Micro Fuel Cell is a cylinder only 1 inches in diameter and 1 inches thick It is made of an extremely inert plastic which can be placed confidently in practically any environment or sample stream It is effectively sealed although one end is permeable to oxygen in the sample gas The other end of the cell is a contact plate consisting of two concentric foil rings The rings mate with spring loaded contacts in the sensor block assembly and provide the electrical connection to the rest of the analyzer Figure 2 1 illustrates the external features Top View Bottom View Concentric Sensing LZ FoilContact Surface Rings Figure 2 1 Micro fuel Cell Refer to Figure 2 2 Cross Section of a Micro Fuel Cell which illus trates the following internal description Dress Ring Diffusion Membrane Electrolyte Cathode Anode Contact Plate Flexible Membrane Figure 2 2 Cross Section of a Micro Fuel Cell not to scale At the top end of th
67. uire highly accurate Auto Cal timing use external Auto Cal control where possible The internal clock in the Model 3000 XL is acurate to 2 3 Accordingly internally scheduled calibrations can vary 2 3 per day To setup an Auto Cal cycle CAUTION We do not recommend frequent Zero adjustments of the cell A newly installed cell may take 7 10 days of operation to reach a steady Zero typically less than 0 2 ppm If required the instrument may be zeroed W Teledyne Analytical Instruments 4 5 4 Operation Model Ultra Trace 3000 after this initial stabilizing period and may be checked again after a addi tional 7 10 day frequency of zero adjustment is at the discretion of the user once a month is suggested Choose System from the Function buttons The LCD will display five subfunctions TRAK HLD Auto Cal PSWD Logout More Use lt gt arrows to blink Auto Cal and press Enter A new screen for Span Zero set appears Span OFF Nxt Zero OFF Nxt Od Oh Od Oh Press lt gt arrows to blink Span or Zero then press Enter again You won t be able to set OFF to ON if a zero interval is entered A Span Every or Zero Every screen appears Span Every 0 d Start 0 h from now Use AV arrows to set an interval value then use lt gt arrows to move to the start time value Use AV arrows to set a start time value To turn ON the Span and or Zero cycles to activate Auto Cal Press System again
68. uto Mode Spanning nnee 4 14 4 4 2 2 Manual Mode Spanning nnen 4 15 AAS Span Pane erin eene ereen eenen 4 16 4 5 Switching of Sample Streams nanne 4 17 4 5 1 Special notes on hydrogen gas stream 4 17 4 6 The Alarms Function aaneen eneevenenenseerenensen 4 17 4 7 The Range Function nnneneenneneneneennnnenen 4 20 4 7 1 Setting the Analog Output Ranges 4 20 4 7 2 Fixed Range Analysis nnee 4 21 4 8 The Analyze Function nn neeneneeeennneenneeeen 4 23 4 9 Signal Output se te As Se hea ee te ed lg 4 23 Maintenance 5 1 Routine Maintenance annen eneen 5 1 5 2 Gell Replacement tansnennensebeerenvennndannder een 5 1 5 2 1 Storing and Handling Replacement Cells 5 1 5 2 2 When to Replace a Gell nnn 5 2 5 2 3 Removing the Micro Fuel Cell eee 5 2 5 2 4 Installing a New Micro Fuel Gell 5 4 5 2 5 Cell WAAN seinen rereeebenknen benee itanseshekied 5 4 5 3 Fuse Replacement annen enenneeneneeeeeeenn 5 5 5 4 System Self Diagnostic TESt ne vaneenn 5 5 5 5 Major Internal Component aaneen 5 6 96 GIS AIM eenst Bette eee entend cect a Stace be me 5 7 5 7 TROUDIESNOOLING Zenner ae eae vette eet 5 8 Appendix A 1 Model Ultra Tace 3000 Specifications A 1 A 2 Recommended 2 Year Spare Parts List A 3 A 3 Drawing DISK sm tranen E a nnee A 4 A 4 19 Inch Relay Rack Panel Mount en A 4 A 5 Application No
69. utput at a predetermined rate It calculates the difference between successive samplings and displays this difference as a Slope on the screen It takes several seconds for the first Slope value to display Slope indicates the rate of change of the Span reading Itis a sensitive indicator of stability Hitt Span Slope ppm s When the Span value displayed on the screen is sufficiently stable press Enter Generally when the Span reading changes by 1 or less of the full scale of the range being calibrated for a period of ten minutes it is sufficiently stable Once Enteris pressed the Span reading changes to the correct value The instrument then automatically enters the Analyze func tion 4 4 3 Span Failure The analyzer checks the output of the cell at the end of the span If the raw output of the cell is less than 0 5 uA ppm O the span will not be accepted The analyzer will return to the previous calibration values trigger the System Alarm and display in the VFD Span Failed This message will be shown for five seconds and the instrument shall return to the Analyze mode In the upper right hand corner of the VFD display FCAL will be shown This message flag will help the operator troubleshoot in case calibration was initiated remotely To reset the alarm and the flag message the analyzer must be properly spanned A trace cell is unlikely to fail span As explained before when the sensor reaches the end of its u
70. valves operate automatically Cal Contact This relay contact is closed while analyzer is spanning and or zeroing See Remote Calibration Protocol below Table 3 3 Remote Calibration Connections Pin Function 9 Remote Zero 11 Remote Zero 10 Remote Span 12 Remote Span 40 Cal Contact 41 Cal Contact Remote Calibration Protocol To properly time the Digital Remote Cal Inputs to the Model Ultra Trace 3000 Analyzer the customer s controller must monitor the Cal Relay Contact When the contact is OPEN the analyzer is analyzing the Remote Cal Inputs are being polled and a zero or span command can be sent When the contact is CLOSED the analyzer is already calibrating It will ignore your request to calibrate and it will not remember that request Once a zero or span command is sent and acknowledged contact closes release it If the command is continued until after the zero or span is complete the calibration will repeat and the Cal Relay Contact CRC will close again For example PN Teledyne Analytical Instruments 3 7 3 Installation Model Ultra Trace 3000 1 Test the CRC When the CRC is open Send a zero command until the CRC closes The CRC will quickly close 2 When the CRC closes remove the zero command 3 When CRC opens again send a span command until the CRC closes The CRC will quickly close 4 When the CRC closes remove the span command When CRC opens again zero and span are
71. y period starts on the day of shipment The spare cell should be carefully stored in an area that is not subject to large variations in ambient temperature 75 F nominal or to rough handling Teledyne Analytical Instruments 5 1 5 Maintenance Model Ultra Trace 3000 WARNING THE SENSOR USED IN THE MODEL 3000 TRACE OXYGEN ANALYZER USES ELECTROLYTES WHICH CONTAIN TOXIC SUBSTANCES MAINLY LEAD AND POTASSIUM HYDROXIDE THAT CAN BE HARMFUL IF TOUCHED SWALLOWED OR INHALED AVOID CONTACT WITH ANY FLUID OR POWDER IN OR AROUND THE UNIT WHAT MAY APPEAR TO BE PLAIN WATER COULD CONTAIN ONE OF THESE TOXIC SUBSTANCES IN CASE OF EYE CONTACT IMMEDIATELY FLUSH EYES WITH WATER FOR AT LEAST 15 MINUTES CALL PHYSI CIAN SEE APPENDIX MATERIAL SAFETY DATA SHEET CAUTION Do not disturb the integrity of the cell package until the cell is to actually be used If the cell package is punctured and air is permitted to enter the cell will require an excessively long time to reach zero after installation 1 2 weeks 5 2 2 When to Replace a Cell Cell failure in the Ultra Trace 3000 is usually characterized by a significant increase in the zero offset of more than 1 0 ppm 2 nanoamps sensor output approximately oxygen or the inability to calibrate the sensor When this occurs the Ultra Trace 3000 will still Zero but the system alarm trips and the LCD displays a failure message dt ppm Anlz CELL FAIL ZERO HIGH Before replacing the c
72. ygen reaches the cathode through the diffusing membrane will also increase The electron transfer and therefore the external current will increase even though the oxygen concentra tion of the sample has not changed It is therefore important that the sample pressure at the fuel cell usually vent pressure remain relatively constant between calibrations 2 2 5 Calibration Characteristics Given that the total pressure of the sample gas on the surface of the Micro Fuel Cell input is constant a convenient characteristic of the cell is that the current produced in an external circuit is directly proportional to the rate at which oxygen molecules reach the cathode and this rate is directly propor tional to the concentration of oxygen in the gaseous mixture In other words it has a linear characteristic curve as shown in Figure 2 3 Measuring circuits do not have to compensate for nonlinearities In addition since there is zero output in the absence of oxygen the characteristic curve has close to an absolute zero less than 0 1 ppm oxygen Depending upon the application zeroing may still be used to compensate for the combined zero offsets of the cell and the electronics 2 2 6 TEC Cooling System Ultra Trace 3000 analyzers include an advance Thermal Electric Cooler TEC system This system enhances the performance of the Micro fuel Cell by cooling it and regulating its operating temperature The TEC system includes a TEC module a tempe