1. A WARNING Patient and sling MUST be removed from the lift during ALL installation procedures 1 Position the load cell assembly of the Jasmine Scale into the boom mounting bracket Detail A of FIGURE 2 2 Secure the Jasmine Scale to the boom mounting bracket with a bolt washers bushings nylon washers locknut and the top rubber cap Securely tighten Install the small rubber cap NOTE Refer to Detail A of FIGURE 2 for washer and small rubber cap orientation NOTE Ensure the shaft of the bolt passes through both sides of the boom mounting bracket Insert swivel bar pin with an o ring and a washer through the swivel bar Detail B of FIGURE 2 aa Align the mounting holes in the swivel bar pin with the mounting holes in the load cell assembly Detail B of FIGURE 2 Secure swivel bar pin to the load cell assembly with a screw and locknut Tighten securely Detail B of FIGURE 2 g A WARNING After ANY adjustments repair or service and BEFORE use make sure all attaching hardware is tightened securely otherwise injury or damage may occur 6 Weigh the patient Refer to Operating the Scale and Weighing the Patient in the Jasmine Patient Lift Owner s Manual Invacare P N 1150704 DETAIL A Top Rubber Boom Jasmine DETAIL B Mounting Scale Locknut Bracket Swivel Bar Nylon Small Washers kt Rubber Cap Jasmine Scale Load Cell Assembly Swivel Bar Pin FIGURE 2 Installing th
2. from the display NOTE The scale is now active and continually updating the weight display 4 Place the patient in the sling For patient placement instructions refer to Operation in the Jasmine Patient Lift Owner s Manual Invacare P N 1150704 9I Activate the lift mechanism to raise the patient until they are completely supported by the lift For patient lifting instructions refer to Lifting the Patient in the Jasmine Patient Lift Owner s Manual Invacare P N 1150704 6 Note the weight display 7 When the weight display becomes stable press the LOCK button to lock the weight display This will be indicated by the word LOCK appearing in the display window NOTE Should it be necessary to unlock the weight while the patient is still supported by the lift the UNLOCK button may be pressed The weight will unlock and the word LOCK will disappear from the display window The weight value will then be updated Once the weight becomes stable the weight can be locked again by pressing the LOCK button NOTE Stable being defined as the weight fluctuating two tenths of a pound For example a patient weighing one hundred lbs the scale will fluctuate between 99 8 and 100 2 until the LOCK key is pressed Fluctuation of the weight displayed is normal as noted above Press the LOCK button to lock the weight NOTE The UNITS button can be pressed to toggle between units of pounds and kilograms This is indicated by lb or kg appearing in t
3. Jasmine Scale JLS5 TD Assembly Installation and Operating Instructions Yes you can SAVE THESE INSTRUCTIONS NOTE Check all parts for shipping damage In case of shipping damage DO NOT use Contact carrier dealer for further instruction To ensure the safe installation operation and use of the Jasmine Scale these instructions MUST be followed A WARNING DO NOT use this equipment without first reading and understanding these instructions If you are unable to understand the Warnings Cautions and Instructions contact distributor before attempting to use this equipment otherwise injury or damage may result The weight capacity is limited to the lowest rated capacity of any one of the components in use i e Patient Lift Sling or Scale The patient s weight MUST NOT exceed the lowest rated capacity of any component After ANY adjustments repair or service and before use make sure that all attaching component parts are secure A ACCESSORIES WARNING Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products Kit Description DESCRIPTION QUANTITY Load Cell Assembly Swivel Bar Pin Locknut Mounting Screw l Nylon Washers 2 l l l l Small Steel Washer Specifi
4. PLIED WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER SOME STATES DO NOT ALLOW EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE OR LIMITATION ON HOW LONG AN IMPLIED WARRANTY LASTS SO THE ABOVE EXCLUSIONS AND LIMITATIONS MAY NOT APPLY TO YOU THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE OR PROVINCIAL LAWS AND REQUIREMENTS Invacare Corporation www invacare com USA Canada All rights reserved Trademarks are identified One Invacare Way 570 Matheson Blvd E Unit 8 by and All trademarks are owned by or Elyria Ohio USA Mississauga Ontario licensed to Invacare Corporation unless 44036 2125 L4Z 4G4 Canada otherwise noted 800 333 6900 800 668 5324 2008 Invacare Corporation Yes you can Part No 1152408 Rev B 10 08
5. acity is limited to the lowest rated capacity of any one of the components in use e g Patient Lift Sling or Scale The patient s weight MUST NOT exceed the lowest rated capacity of any component 1 Attach sling straps to the swivel bar For proper attachment instructions refer to Attaching a Sling to the Swivel Bar in the Jasmine Patient Lift Owner s Manual Invacare P N 1150704 NOTE For improved accuracy on the patient lift attach all sling straps to the center hooks on both sides of the swivel bar However patient comfort may require a different configuration of sling straps 2 Press the ON OFF key Jasmine Scale JLS5 4 Part No 1152408 A WARNING DO NOT operate key strokes with pointed objects e g pencils pens fingernails etc Otherwise damage to the key pad will result NOTE The display will indicate the last weight that was measured The word LOCK will be seen in the display box NOTE The ZERO key is pressed in order to avoid capturing the weight of the sling and the hardware If the ZERO key is not pressed the weight of the sling and the weight of the hardware will be included in the weight displayed NOT ZERO ING OUT WILL GIVE A FALSE READING OF THE USER S TRUE WEIGHT 3 Press the ZERO key When the ZERO key is pushed the following will happen A The scale will reset to zero and the word ZERO will appear in the display B The lock function will then be turned off and the word LOCK will disappear
6. ation in the Jasmine Patient Lift Owner s Manual Invacare P N 1150704 for further instructions 2 For sling detachment instructions refer to Lifting the Patient in the Jasmine Patient Lift Owner s Manual Invacare P N 1150704 3 Turn the scale display power on by pressing the ON OFF key 4 Rotate the Swivel Bar until the back of the scale display is visible 5 Insert the tip of a paper clip or similar tool into the small hole on the back of the scale display and press the CAL button NOTE Once the CAL button is pressed CAL1 will be seen in the display window 6 Press the UNITS key The display reads CAL2 7 Press the LOCK UNLOCK key The display reads UNLD 8 Press the ZERO key The display reads LOAD 9 Hang 200 pounds of calibrated weight from the scale and press the ZERO key 10 Press the ZERO key The display reads DONE Jasmine Scale JLS5 6 Part No 1152408 11 Remove the weight Press the LOCK UNLOCK key NOTE The scale is now in test mode 12 Press the ZERO key The display reads 0 00 Ib NOTE The scale will now be in normal operation BACK VIEW OF SCALE DISPLAY CAL button A n pe Display Window FIGURE 5 Calibrate the Jasmine Scale Troubleshooting SYMPTOM PROBABLE CAUSE SOLUTION Battery has been replaced and unit still does Contact Invacare NOT
7. cations Weight Range Up to 500 Ibs 226 7 Kg Resolution 0 2 Ibs 0 1 Kg Display Liquid Crystal Digital Automatic Power Down Two minutes Size 5 3 4 L X 3 5 8 W X 4 H Weight 55 Ibs Power 9 Volt Alkaline battery included Battery Life Approximately 1500 readings Temperature Range 50 To 104 F 10 To 40 C Part No 1152408 l Jasmine Scale JLS5 Removing the Swivel Bar Cradle Attachment NOTE For this procedure refer to FIGURE 1 1 Remove the side rubber caps NOTE Support the swivel bar or cradle attachment from underneath while removing the attaching hardware 2 Remove the socket head screw washer bushings top rubber cap nylon spacers washer and nut that attach the swivel bar or cradle attachment to the boom NOTE When the attaching hardware is removed the pin washer and o ring will become loose NOTE Save the socket screw small steel washer and locknut to secure the Jasmine Scale to the boom NOTE Save the remaining hardware for future use of the swivel bar or cradle attachment without the scale Top Rubber Cap Washer Side Rubber Cap l Nut Optional Cradle Bushing Bushing Attachment Washer Socket Head Screw Side Rubber Cap Nylon Spacer Swivel Bar w Pin Bottom Rubber Cap FIGURE Removing the Swivel Bar Cradle Attachment Jasmine Scale JLS5 2 Part No 1152408 Installing the Jasmine Scale NOTE For this procedure refer to FIGURE 2
8. e Jasmine Scale Part No 1152408 3 Jasmine Scale JLS5 Operating the Scale Keypad Functions INDICATOR INDICATOR KEY DISPLAYED LOCATION DEFINITION ON OFF Center of Display Window Pressing this key will apply power to the OFF scale and turn the unit on When the scale is already on pressing the button will turn the unit off ZERO ZERO Lower Left Corner of the Pressing this key when the scale is on will Display Window reset the weight shown in the display window to zero UNITS lb or kg Upper Right Corner of the The function of this key is to change the Display Window unit of measurement from pounds Ib to kilograms kg LOCK LOCK Lower Right Corner of the Display This key is used to LOCK or UNLOCK a UNLOCK Window weight value in the display window NOTE Weight displayed is a STORED weight and not the current weight when the key is used in the lock position N A LO BAT Center of Display Window Indicator is shown in the display window to notify when battery is low Display Window HT RE ZERO Lock B j Y DETECTO a BS pace O i 1 Atach sling peration 7 2 Press ON key a Keys Jasmine Scale 3 Press ZEROkey ERO Model JLS5 4 Liftpatient 5 Press LOCK key TOOK CAPACITY 500 LB 227 KG Lowerpatent 7 Read Weight WEIGHT FIGURE 3 Operating the Scale Keypad Functions Weighing the Patient A WARNING The weight cap
9. he display window 8 The lift may now be lowered and the sling removed from the patient NOTE The patient s weight will continue to be seen in the display window The display will turn off automatically after a two minute period of non use no changes in weight exceeding five pounds 2 kilograms You can NOT adjust the time delay for automatic shut off After the display has turned off the weight may be recalled by pressing the ON OFF button The unit can be turned off by pressing the ON OFF button a second time Part No 1152408 5 Jasmine Scale JLS5 Replacing the Battery NOTE For this procedure refer to FIGURE 4 NOTE The scale is powered by a 9 volt alkaline battery that should provide approximately 1500 readings before needing replacement When battery replacement is needed LO BAT will appear on the display Perform the following 1 Slide the battery door open in the direction of the arrow 2 Remove existing battery 3 Install the NEW battery 4 Reinstall the battery door Assembled 9V Battery Battery FIGURE 4 Replacing the Battery Calibrate the Jasmine Scale NOTE For this procedure refer to FIGURE 5 on page 7 NOTE The Jasmine Scale will be pre calibrated at the factory with the load cell Should it be necessary to re calibrate the scale follow the instructions outlined below 1 The patient and the sling MUST be removed from the scale to properly calibrate the Jasmine Scale Refer to Oper
10. rchase indicate nature of the defect Invacare Corporation will issue a serialized return authorization The defective unit or parts MUST be returned for warranty inspection using the serial number when applicable as identification within 30 days of return authorization date DO NOT return products to our factory without our prior consent C O D shipments will be refused please prepay shipping charges LIMITATIONS AND EXCLUSIONS THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE PRODUCTS MODIFIED WITHOUT INVACARE S EXPRESS WRITTEN CONSENT INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE OR TO A PRODUCT DAMAGED BY CIRCUMSTANCES BEYOND INVACARE S CONTROL AND SUCH EVALUATION WILL BE SOLELY DETERMINED BY INVACARE THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO ADHERE TO THE INSTRUCTIONS IN THIS MANUAL THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES IMPLIED WARRANTIES IF ANY INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESSED WARRANTY PROVIDED HEREIN AND THE REMEDY FOR VIOLATIONS OF ANY IM
11. work properly Display Codes CAL CALIBRATION REQUIRED Indicates improper stored calibration data calibration is necessary OVER CAPACITY Indicates a weight exceeding the capacity has been loaded on the scale Part No 1152408 7 Jasmine Scale JLS5 LIMITED WARRANTY PLEASE NOTE THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4 1975 This warranty is extended only to the original purchaser user of our products This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state Invacare warrants the products manufactured excluding the 9V battery to be free from defects in materials and workmanship for a period of two years from the date of purchase If within such warranty period any such product shall be proven to be defective such product shall be repaired or replaced at Invacare s option This warranty does not include any labor or shipping charges incurred in replacement part installation or repair of any such product Invacare s sole obligation and your exclusive remedy under this warranty shall be limited to such repair and or replacement For warranty service please contact the dealer from whom you purchased your Invacare product In the event you do not receive satisfactory warranty service please write directly to Invacare at the address on the back cover provide dealer s name address date of pu

Download Pdf Manuals

image

Related Search

Related Contents

pdf  取扱説明書  Intellinet 560849 network switch  SYMMETRISCHE VOEDING 1A ............................................ 3  普通便座  Samsung 920NW User Manual  Normes et Procédures des établissements résidentiels pour adultes  Samsung GT-S5360 คู่มือการใช้งาน  manuel pédagogique - Fédération européenne des écoles de  Owner`s Manual - Buy Two Way Radios  