Invacare IRC5P User's Manual
Contents
1. Clean Replace Cabinet Filters Check outlet HEPA Filter Check Compressor Inlet Filter Check Oxygen Concentration SensO models Check Power Loss Alarm As Required Clean Heat Exchanger Replace Exhaust Muffler Every 180 Days for IRC5P Check oxygen Concentration NOTE Refer to Preventive Maintenance section of Service Manual Part No 1143482 27 Perfecto Series SECTION 7 TROUBLESHOOTING GUIDE SECTION 7 TROUBLESHOOTING GUIDE SYMPTOM Alarm PROBABLE CAUSE Main Power Loss SOLUTION Short beeps long pause Power cord not plugged in Insert plug into outlet Concentrator not operating power switch On 2 No power at outlet 2 Inspect house circuit breakers or fuses If problem recurs use a different outlet Concentrator not operating Power Switch On Beeeeeeeep Beep Beep 3 Tripped circuit 3 Call Service Provider breaker Alarm System Failure la Remove and clean cabinet Continuous l Unit overheating due filters to blocked air intake 2 Insufficient power at outlet 3 Internal repairs required Ib Move oxygen concentrator at least three inches away from walls draperies or furniture 2 DO NOT use extension cords Move to another electrical outlet or circuit 3 Call Service Provider Perfecto Series 28 Part
2. IRC5P Model Indicator Light Explanation RED light Total Unit Shut Down Switch IMMEDIATELY to a back up oxygen supply Call supplier IMMEDIATELY GREEN light I O On Off System okay Auto Shut Down e a RED TT Call Supplier Use Backup Normal e a GREEN FIGURE 5 7 Indicators If Your Unit Does Not Feature the O Sensor Perfecto Series 24 Part No 1143482 SECTION 6 MAINTENANCE SECTION 6 MAINTENANCE NOTE The Invacare concentrators are specifically designed to minimize routine preventive maintenance In places with high dust or soot levels maintenance may need to be performed more often Only qualified personnel should perform preventive maintenance on the concentrator Power should be disconnected before beginning preventive maintenance on the Invacare concentrator NOTE At a minimum preventive maintenance MUST be performed according to the maintenance record guidelines Refer to Preventive Maintenance Record on page 27 Routine Maintenance NOTE The following routine maintenance should be performed more frequently See recommended intervals in each section and preventive maintenance record Cleaning the Cabinet Filter NOTE For this procedure refer to FIGURE 6 1 on page 26 NOTE There is one cabinet filter located on the back of the cabinet 1 Remove the filter and clean at least once a week depending on environmental conditions NOTE Environmental conditions that may req
3. has fallen into water Unplug IMMEDIATELY A product should NEVER be left unattended when plugged in This device is to be used only in accordance with the prescription of a physician and this Owner s Manual If at any time the patient or attendant conclude that the patient is receiving an insufficient amount of oxygen the supplier and or physician should be contacted IMMEDIATELY DO NOT change the L min setting on the flowmeter unless a change has been prescribed by your physician or therapist Close supervision is necessary when this product is used near children or physically challenged individuals Use this product for only intended use as described in this manual DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices Part No 1143482 9 Perfecto Series SECTION GENERAL GUIDELINES WARNING Use of some administration accessories or certain humidifiers not specified for use with oxygen concentrator may impair the performance In certain circumstances oxygen therapy can be hazardous Invacare recommends that you seek medical advice before using this machine This equipment is to be used as an oxygen supplement and is not considered life supporting or life sustaining Avoid creation of any spark near medical oxygen equipment This includes sparks from static electricity created by any type of friction Polarized Plug Instruction As a safety
4. that will burn in air and some that will not are easily ignited and burn rapidly in high concentrations of oxygen For safety concerns it is necessary that all sources of ignition be kept away from the product and out of the room in which it is being used NO SMOKING signs should be prominently displayed WARNING A spontaneous and violent ignition may occur if oil grease or greasy substances come in contact with oxygen under pressure These substances MUST be kept away from the oxygen concentrator tubing and connections and all other oxygen equipment DO NOT use any lubricants unless recommended by Invacare For optimum performance Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time Shorter periods of operation may reduce maximum product life If the concentrator has a damaged cord or plug if it is not working properly if it has been dropped or damaged or dropped into water call qualified technician for examination and repair Keep the oxygen tubing cord and unit away from heated or hot surfaces including space heaters blankets stoves and similar electrical appliances DO NOT move or relocate concentrator by pulling on the cord NEVER drop or insert any object into any opening Part No 1143482 7 Perfecto Series SECTION GENERAL GUIDELINES WARNING NEVER block the air openings of the product or place it on a soft surface such as a bed or
5. 5 3 Perfecto 5 44 2 dBA Average Perfecto 5 Up to 6 000 ft 1828 meters above sea level without degradation of con centration levels From 6 000 ft 1828 meters to 3 129 ft 4000 meters below 9076 efficiency All IRC5PO2 IRC5P models 95 6 to 87 at 0 5 to 5 L min Perfecto 5 5 psi 0 5 psi 34 5 kPa 3 45 kPa Perfecto 5 0 5 to 5 L min For flowrates less than L min we recom mend the use of the Invacare Pediatric Flow meter IRCPFI6 Accessory Perfecto 5 0 L min to 0 5 L min Rapid Audible Alarm Beeping No Accessories Connected Perfecto 5 280 W Operating 3 L min 35 psi 3 5 psi 241 kPa 24 1 kPa 0 7 L min Cabinet Outlet HEPA and Compressor Inlet Part No 1143482 13 Perfecto Series SECTION 4 TYPICAL PRODUCT PARAMETERS Safety System Current overload or line surge shutdown High temperature compressor shutdown High Pressure Alarm w compressor shut down Low Pressure Alarm w compressor shutdown Battery Free Power Loss Alarm SensO Oxygen System SensO Model Low Flow Alarm Operating Ambient 50 F 95 F 10 C 35 C at 20 60 relative Temperature humidity Exhaust Perfecto 5 Less than Ambient 35 F 19 C Oxygen Output Perfecto 5 Less than Ambient 6 F 3 C Cabinet Impact Resistant flame retardant plastic cab inet that conforms to UL 94 0 Regulatory Listing Double Insulated Product ETL certified complying with UL 1097 ETL
6. 6 MAINTENANCE 00000000000000000000000000 25 Routine 25 SECTION 7 TROUBLESHOOTING GUIDE 28 LIMITED WARRANTY 000000000000000000000000000000000000 3 Perfecto Series 4 Part No 1143482 SPECIAL NOTES SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage Refer to the following table for definitions of the signal words SIGNAL WORD MEANING Danger indicates an imminently hazardous situation which if not avoided will result in death or serious DANGER injury Warning indicates a potentially hazardous situation which if not avoided could result in death or WARNING serious injury Caution indicates a potentially hazardous situation which if not avoided may result in property CAUTION damage NOTICE The information contained in this document is subject to change without notice DANGER DO NOT SMOKE while using this device Keep all matches lighted cigarettes or other sources of ignition out of the room in which this product is located and away from where oxygen is being delivered NO SMOKING signs should be prominently displayed Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air Failure to observe this warning can result in severe fire proper
7. No 1143482 SECTION 7 TROUBLESHOOTING GUIDE SYMPTOM YELLOW or RED light Illuminated Alarm Continuous On RED light only Only applies to 5PO models PROBABLE CAUSE Low oxygen purity SOLUTION Clean or Replace filters 2 Kinked or blocked tubing cannula or humidifier 2 Inspect for kinks or blockages Correct clean or replace item Once corrected turn power Off for 60 seconds and then turn power back On 3 Flowmeter set at 0 5 L min for Perfecto 5 3 Check flowmeter is set to 1 0 L min or more Refer to Recommended Guidelines for Optimum Performance on page 17 4 Unit overheating due to blocked air intake 4a Remove and clean cabinet filters 4b Move concentrator at least three inches from walls draperies and furniture 5 Insufficient power at outlet 5 DO NOT use extension cords Move to another electrical outlet or circuit 6 Internal repairs required 6 Call Service Provider GREEN light with YELLOW Light Flashing Internal repairs required Call Service Provider Part No 1143482 29 Perfecto Series SECTION 7 TROUBLESHOOTING GUIDE SYMPTOM PROBABLE CAUSE SOLUTION Alarm LOW FLOW ALARM la Inspect for kinks or Rapid la Kinked or blocked blockages Correct clean or tubing cannula or replace item Once Beep Beep humidifier corrected turn power Off for Bee
8. Operator s Manual Perfecto Series Oxygen Concentrators without SensO Oxygen Concentrators with SensO HomeFill Compatible Model IRC5PO2 Model IRC5P Dealer This manual MUST be given to the end user User BEFORE using this product read this manual and save for future reference For more information regarding Invacare products parts and services please visit www invacare com Yes you can WARNING WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS CAUTIONS AND INSTRUCTIONS CONTACT INVACARE TECHNICAL SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE SERIOUS INJURY OR PROPERTY DAMAGE MAY RESULT ACCESSORIES There are many different types of humidifiers oxygen tubing cannulas and masks that can be used with this device You should contact your local home care provider for recommendations on which of these devices will be best for you They should also give you advice on the proper usage maintenance and cleaning WARNING Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products NOTE Updated versions of this manual can be found at www invacare com Perfecto Series 2 Pa
9. USION AND LIMITATION MAY NOT APPLY TO YOU THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE PROVINCIAL LAWS AND REQUIREMENTS Invacare Corporation www invacare com USA Invacare Yes you can SensO and the One Invacare Way Medallion Design are registered Elyria Ohio USA trademarks of Invacare Corporation 44036 2125 Perfecto HomeFill Sieve GARD Precise 800 333 6900 Technical Services B3 Corporation 800 832 4707 9 2007 Invacare Corporation Yes you Ry are trademarks of Invacare Part No 1143482 Rev A 2 28 07
10. certified to CSA C22 2 No 68 Placement No closer than three inches from any wall furniture draperies or similar surfaces Tubing 7 ft cannula with a maximum 50 ft of Crush Proof Tubing DO NOT pinch Relative Humidity 20 to 60 Time of Operation Up to 24 hours per day Recommended Storage and 20 F to 150 F 29 C to 65 C at 15 95 Shipping Temperature relative humidity NOTE Stated concentration levels achieved after initial warm up period approximately 30 minutes Perfecto Series 14 Part 1143482 SECTION 4 TYPICAL PRODUCT PARAMETERS IRC5PO2 INDICATORS LABEL O PURITY INDICATOR LIGHTS SYMBOL LED SYSTEM OKAY GREEN Indicator Light over 85 2 Between 73 3 YELLOW Indicator light to 85 2 YELLOW Solid B YELLOW Flashing Sensor Failure Call a qualified technician SYSTEM FAILURE RED Indicator Light Below 73 376 Continuous Audible Alarm Sieve GARD Compressor Shut down IRC5P INDICATORS LABEL STATUS INDICATOR LIGHTS SYMBOL LED SYSTEM OKAY GREEN Indicator Light SYSTEM FAILURE RED Indicator light A Continuous Audible Alarm Sieve GARD Compressor Shutdown Call a qualified technician Part No 1143482 15 Perfecto Series SECTION 5 OPERATING INSTRUCTIONS SECTION 5 OPERATING INSTRUCTIONS Introduction Your oxygen concentrator is intended for individual use in the home It is an electr
11. couch where the air opening may be blocked Keep the openings free from lint hair and the like Fill humidifier with water to the level shown by the manufacturer DO NOT overfill Keep unit at least three inches away from walls draperies furniture and the like Invacare recommends that Crush Proof oxygen tubing be used with this product and not exceed 50 ft 15 2m in length Maintenance The Invacare Oxygen Concentrator was specifically designed to minimize routine preventive maintenance Only professionals of the healthcare field or persons fully conversant with this process such as factory trained personnel should perform preventive maintenance or performance adjustments on the oxygen concentrator Radio Frequency Interference Most electronic equipment is influenced by Radio Frequency Interference RFI Caution should be exercised with regard to the use of portable communications equipment in the area around such equipment Perfecto Series 8 Part No 1143482 SECTION GENERAL GUIDELINES WARNING To Reduce The Risk Of Burns Electrocution Fire Or Injury To Persons Avoid using while bathing If continuous usage is required by the physician s prescription the concentrator MUST be located in another room at least seven feet from the bath DO NOT come in contact with the concentrator while wet DO NOT place or store product where it can drop into water or other liquid DO NOT reach for product that
12. e and soot free No confined spaces Example No closets Relative Humidity 20 to 60 Time of Operation Up to 24 hours per day Flow Rate Perfecto 5 From 0 5 L min to 5 L min For flow rates less than L min use the Pediatric Flowmeter Accessory NOTE Ensure that your concentrator is at least three inches away from walls draperies or furniture to assure sufficient air flow Avoid deep pile carpets and heaters radiators or hot air registers Plug in Power Cord 1 Plug in power cord to an electrical outlet Part No 1143482 17 Perfecto Series SECTION 5 OPERATING INSTRUCTIONS Connect Humidifier If So Prescribed WARNING For optimum performance Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time Shorter periods of operation may reduce maximum product life DO NOT overfill humidifier DO NOT reverse the oxygen input and output connections Water from the humidifier bottle will travel through the cannula back to the patient NOTE For this procedure refer to FIGURE 5 1 FIGURE 5 2 on page 19 FIGURE 5 3 on page 20 1 Remove cap from bottle 2 Fill humidifier with distilled water to the level indicated by the manufacturer Replace the humidifier cap and securely tighten Humidifier Humidifier Bottle w o Cap Bottle with Cap FIGURE 5 1 Filling the Humidifier 3 Insert a flathead screwdriver in the plate groove on the top edg
13. e of the filter access door and gently pry the filter access door off FIGURE 5 2 4 Pull up and remove the humidifier bottle adapter FIGURE 5 2 Perfecto Series 18 Part No 1143482 SECTION 5 OPERATING INSTRUCTIONS FIGURE 5 2 Humidifer Bottle Adapter Attach the humidifier bottle adapter to the humidifier bottle by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened See Detail A in FIGURE 53 Place the humidifier bottle adapter assembly in the humidifier compartment on the concentrator See Detail B in FIGURE 5 3 Attach oxygen tubing from the humidifier bottle adapter assembly to oxygen outlet connector on the concentrator See Detail B in FIGURE 5 3 Attach the cannula patient supply tubing to the humidifier bottle outlet See Detail B in FIGURE 5 3 After assembly ensure that oxygen is flowing through the cannula Part No 1143482 19 Perfecto Series SECTION 5 OPERATING INSTRUCTIONS FIGURE 5 3 Humidifier Compartment Perfecto Series 20 Part 1143482 SECTION 5 OPERATING INSTRUCTIONS Power Switch NOTE For this procedure refer to FIGURE 5 4 1 Press power switch to on position All the panel lights and the audible alarm will come on for one second indicating that the unit is functioning properly After one second only the GREEN system ok power light will stay on FIGURE 5 4 Power Switch Flow
14. en designed to assure maximum protection of the concentrator If the concentrator is to be reshipped by common carrier additional cartons are available from Invacare Unpacking NOTE For this procedure refer to FIGURE 3 1 1 Check for any obvious damage to the carton or pm its contents If damage is n evident notify the carrier or your local dealer 2 Remove allloose packing from the carton 3 Carefully remove all the components from the ieu tpe ad carton FIGURE Unpacking NOTE Unless the oxygen concentrator is to be used IMMEDIATELY retain containers and packing materials for storage until use of the concentrator is required Inspection 1 Inspect examine exterior of the oxygen concentrator for nicks dents scratches or other damages Inspect all components Storage 1 Store the repackaged oxygen concentrator in a dry area 2 DO NOT place objects on top of repackaged concentrator Perfecto Series 12 Part 1143482 SECTION 4 TYPICAL PRODUCT PARAMETERS SECTION 4 TYPICAL PRODUCT PARAMETERS Electrical Requirements Rated Current Input Sound Level Altitude Oxygen Output Concentration Levels Maximum Outlet Pressure Flow Range Low Flow Alarm Average Power Consumption Pressure Relief Mechanism Operational at Change in maximum recom mended flow when back pres sure of 7kPa is applied 120 VAC x 10 132 VAC 108 VAC 60 Hz Perfecto
15. feature this appliance has a polarized plug one blade is wider than the other This plug will fit in a polarized outlet only one way If the plug does not fit fully in the outlet reverse the plug If it still does not fit contact a qualified electrician DO NOT attempt to defeat this safety feature Perfecto Series 10 Part 1143482 SECTION 2 FEATURES SECTION 2 FEATURES Oxygen Outlet Lights Fault and Power Indicator Lights Flowmeter Elapsed Time Power Switch Meter L of ACCESSORIES NOT SHOWN PreciseRX Pediatric Humidifier Flowmeter Accessory IRCPFI6 Perfecto 5 Only HomeFill home oxygen compressor 200 REAR VIEW Cabinet Filter NOTE This outlet fitting is to be used only for filling oxygen cylinders with the HomeFill home oxygen compressor The outlet fitting does not affect concentrator performance Refer to the HomeFill owner s manual part number 1100873 for connection and operating instructions When not in use the plug provided with the concentrator should be inserted into the outlet fitting For more information about the HomeFill Power contact your Invacare dealer Cord Fitting Part No 1143482 Perfecto Series SECTION 3 HANDLING SECTION 3 HANDLING The concentrator should ALWAYS be Kept in the upright position to prevent cabinet damage while being transported The shipping container has be
16. ill be refused please prepay shipping charges Part No 1143482 3l Perfecto Series LIMITATIONS AND EXCLUSIONS THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE SOOT OR SMOKE FILLED ENVIRONMENTS OR OTHER THAN NORMAL APPLICATION USE OR SERVICE OR TO PRODUCTS MODIFIED WITHOUT INVACARE CORPORATION S EXPRESS WRITTEN CONSENT INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS OR TO PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE CORPORATION OR TO PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE CORPORATION S CONTROL THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER WHETHER EXPRESS OR IMPLIED INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS SO THE ABOVE EXCL
17. ll be proven to Invacare Corporation s satisfaction to be defective such product shall be repaired or replaced at Invacare Corporation s option This warranty only applies to the labor for repairs performed by the Invacare Service Department or Invacare Authorized Service Centers It does not apply to the labor performed by the purchaser or user This warranty does not include normal wear and tear or shipping charges Invacare Corporation s sole obligation and your exclusive remedy under this warranty shall be limited to such repair or replacement Routine maintenance items such as filters are excluded from this warranty For warranty service please contact Invacare Corporation s Service Department at the toll free number on the back cover Upon receiving notice of an alleged defect in a product Invacare Corporation will issue a serialized return authorization It shall then be the responsibility of the purchaser to return the entire unit or remove at purchaser s cost the defective component part s identified pack the component part s in a manner to avoid shipping damage and to ship the component part s to either Invacare Corporation s plant or service center as specified by Invacare Corporation in advance Defective component part s MUST be returned for warranty inspection using the serial number as identification within thirty days of return authorization date DO NOT return products to our factory without prior consent C O D shipments w
18. onically operated device that separates oxygen from room air It provides high concentration of oxygen directly to you through a nasal cannula Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems Your provider will show you how to use your oxygen concentrator He She should be contacted with any questions or problems regarding your oxygen concentrator This owner s manual will tell you about your concentrator and will serve as a reference as you use your concentrator Select a Location You may select a room in your house where using your oxygen concentrator would be most convenient Your concentrator can be easily rolled from room to room on its casters Your oxygen concentrator will perform best when operated under the following conditions Refer to Recommended Guidelines for Optimum Performance on page 17 Usage in environments other than those described may result in the need for increased equipment maintenance The air intake of the unit should be located in a well ventilated area to avoid airborne pollutants and or fumes Perfecto Series 16 Part 1143482 SECTION 5 OPERATING INSTRUCTIONS Recommended Guidelines for Optimum Performance Placement No closer than three inches from the wall furniture draperies or similar surfaces Tubing and Cannula 7 ft cannula with a maximum 50 ft of Crush Proof Tubing DO NOT pinch Environment Smok
19. p Beep 60 seconds and then turn Ib Flowmeter set at 0 5 power back ON L min for Perfecto 5 Ib Reset flowmeter to prescribed flowrate If prescribed flowrate is below 0 5 L min use the Pediatric Flowmeter IRCPFI6 Only with Perfecto 5 Perfecto Series 30 Part No 1143482 LIMITED WARRANTY LIMITED WARRANTY NOTE THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4 1975 This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer This warranty is not extended to any other person or entity and it is not transferable or assignable to any subsequent purchaser or owner Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person This warranty gives you specific legal rights and you may also have other legal rights which may vary from state to state Invacare Corporation warrants its Perfecto Oxygen Concentrators when purchased new and unused to be free from defects in materials and workmanship for a period of five years from date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty DO NOT OPEN OR ATTEMPT TO SERVICE THE 4 WAY VALVE BODY This will void any and all warranty applicable to the valve body If within such warranty periods any such product sha
20. rate NOTE For this procedure refer to FIGURE 5 5 on page 22 1 Turn the flowrate knob to the setting prescribed by your physician or therapist A WARNING DO NOT change the L min setting on the flowmeter unless a change has been prescribed by your physician or therapist NOTE To properly read the flowmeter locate the prescribed flowrate line on the flowmeter Next turn the flow knob until the ball rises to the line Now center the ball on the L min line prescribed Part No 1143482 21 Perfecto Series SECTION 5 OPERATING INSTRUCTIONS 2 Iftheflowrate on the flowmeter ever falls below 0 5 L min for Perfecto 5 for more than about one minute the LOW FLOW alarm will be triggered This is a rapid beeping of the audible alarm Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle After rated flow is restored more than 0 5 L min for Perfecto 5 the LOW FLOW audible alarm will go off NOTE The use of some accessories such as the pediatric flowstand and the HomeFill compressor will deactivate the Low Flow Alarm Flow Knob Flowmeter 0 5 L min Ball FIGURE 5 5 Flowrate SensO Oxygen Purity Indicator If Your Unit Features the O Sensor Feature This feature monitors the purity of oxygen generated by the oxygen concentrator If purity falls below factory preset standards indicator lights on the control panel will illuminate Perfecto Series 22 Par
21. rt No 1143482 Table of Contents TABLE OF CONTENTS ACCESSORIES SPECIAL NOTES 00000000000000000000000000000000000000000000000 5 SECTION GENERAL GUIDELINES 6 Dcos oen rra dt 8 Radio Frequency 8 To Reduce The Risk Of Burns Electrocution Fire Or Injury TO e NE e ae 9 Polarized Plug Instruction 10 SECTION 2 FEATURES 0000000000000000000000000000000000 SECTION 3 HANDLING 000000000000000000000000000000000 12 Unpacking 12 heran 12 12 SECTION 4 TYPICAL PRODUCT PARAMETERS 000000000000000000000000000000000000000000000000 3 SECTION 5 OPERATING INSTRUCTIONS 16 M c eO cO l6 Select a Location sn l6 Recommended Guidelines for Optimum Performance 17 Plugin Power 17 POWER 21 ossa 21 SensO Oxygen Purity Indicator If Your Unit Features the O Sensor Feature usas nana 22 Initial Startup of the Concentrator 23 Explanation of Oxygen Purity Indicator Lights Model IR SPO seca aia ani denda QD diam 23 Part No 1143482 3 Perfecto Series Table of Contents TABLE OF CONTENTS Indicators If Your Unit Does Not Feature the Sensor 24 SECTION
22. t No 1143482 SECTION 5 OPERATING INSTRUCTIONS Initial Startup of the Concentrator NOTE Concentrator may be used during the initial start warm up time approximately 30 min while waiting for the O purity to reach maximum When the unit is turned on the GREEN light will come on SYSTEM OK O greater than 85 After five minutes the oxygen sensor will be operating normally and will control the indicator lights depending on oxygen concentration values Explanation of Oxygen Purity Indicator Lights Model IRC5PO2 NOTE For this procedure refer to FIGURE 5 6 GREEN light O Normal Operation YELLOW light Call supplier IMMEDIATELY You may continue to use the concentrator unless instructed otherwise by your supplier Be certain that backup oxygen is nearby RED light Total unit shutdown Switch IMMEDIATELY to backup oxygen supply Call supplier IMMEDIATELY GREEN light with YELLOW light flashing Call supplier IMMEDIATELY Oxygen sensor malfunctioning you may continue to use the concentrator OXYGEN PURITY Auto Shut Down A e Use Backup RED Below S SFR cai Supplier Norma 6 YELLOW Normal Oo 4 GREEN FIGURE 5 6 SensO Oxygen Purity Indicator If Your Unit Features the O Sensor Feature Part No 1143482 23 Perfecto Series SECTION 5 OPERATING INSTRUCTIONS Indicators If Your Unit Does Not Feature the O Sensor NOTE For this procedure refer to FIGURE 5 7
23. ty damage and cause physical injury or death Part No 1143482 5 Perfecto Series SECTION GENERAL GUIDELINES SECTION GENERAL GUIDELINES In order to ensure the safe installation assembly and operation of the Perfecto concentrator these instructions MUST be followed CAUTION Caution Federal law restricts this device to sale or rental by or on order of a physician or any other practitioner licensed by the law of the State in which he she practices to use or order the use of this device WARNING SECTION GENERAL GUIDELINES contains important information for the safe operation and use of this product DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as Owner s Manuals Service Manuals or Instruction Sheets supplied with this product or optional equipment If you are unable to understand the Warnings Cautions or Instructions contact a healthcare professional dealer or technical personnel before attempting to use this equipment otherwise injury or damage may occur DANGER Risk of electric shock DO NOT disassemble Refer servicing to qualified service personnel Perfecto Series 6 Part No 1143482 SECTION GENERAL GUIDELINES DANGER The use of oxygen therapy requires that special care be taken to reduce the risk of fire Any materials
24. uire more frequent cleaning of the filters include but are not limited to high dust air pollutants etc 2 Clean the cabinet filter with a vacuum cleaner or wash in warm soapy water and rinse thoroughly 3 Dry the filter thoroughly before reinstallation CAUTION DO NOT operate the concentrator without the filter installed Part No 1143482 25 Perfecto Series SECTION 6 MAINTENANCE FIGURE 6 1 Cleaning the Cabinet Filter Cabinet A WARNING UNPLUG the concentrator when cleaning To avoid electrical shock DO NOT remove cabinet 1 Clean the cabinet with a mild household cleaner and non abrasive cloth or sponge Humidifier NOTE To clean the oxygen humidifier follow the instructions provided by the manufacturer If none are provided follow these STEPS 1 Clean the humidifier every day 2 Wash it in soapy water and rinse it with a solution of ten parts water and one part vinegar 3 Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier WARNING DO NOT overfill humidifier bottle Perfecto Series 26 Part No 1143482 SECTION 6 MAINTENANCE PREVENTIVE MAINTENANCE RECORD Model No IRC Serial No On Each Inspection Record Date Of Service Record Elapsed Hours on Hour Meter Clean Cabinet Filters Check Prescribed L min Flow Rate Annually During Preventive Maintenance Schedule Or Between Patients
Download Pdf Manuals
Related Search
Related Contents
Red Hat Enterprise Linux ES (v. 3) 4119 Rev D RTD Sensor Installation Manual 003-2417-00 - Artizan Wall-hung Cabinet Installation Global Door Controls GLA10SSUN-626 Installation Guide MBS598154 - MyBioSource Control de nivel rotatorio de la serie MAXIMA+ Thermal transfer Printer for heat shrinkable sleeve Operator`s Manual MANUAL TECNICO Manual de serviços 99 ASUS ROG SW-MA1 Copyright © All rights reserved.