Graham Field 760000-INS-LAB-RevA08 User's Manual

image

Contents

1. Graham Field Brand AltaDyne Plus 760000 ALTERNATING PRESSURE LOW AIR LOSS MATTRESS SYSTEM USER MANUAL Important Do not operate the Mattress System without first reading and understanding this manual Save this manual for future use CONTENTS INTRODUCTION 004m 3 INTENDED USE OF THIS DEVICE 3 IMPORTANT SAFETY PRECAUTIONS 3 yQlczlm c 4 la Lislicme 4 CAUTION M 5 EQUIPMENT SYMBOLS 5 PRODUCT DESCRIPTION srira 6 PUMP AND MATTRESS 6 PUMP FRONT 5uecciccessvasdrecasus sup caua E 6 PUMP REAR eec 7 PUMP FRONT PANEE iii nono Cami abuse cud pedi naissance do nra 7 FRONT PANEL FIRST ROW CYCLE TIME ADJUSTMENT 7 FRONT PANEL SECOND ROW PRESSURE ADJUSTMENT 7 FRONT PANEL THIRD ROW THERAPY MODE ADJUSTMENT 7 FRONT PANEL FOURTH ROW 2 2 8
2. Power failure indicator The power failure indicator illu minates and the alarm sounds when there is a power failure Depress the alarm mute button to disable both alarm and LED 8 760000 INS LAB RevA08 Service indicator Mechanical failure service required contact your Graham field equipment provider INSTALLATION UNPACKING 1 Before unpacking check for obvious damage to the car ton or its contents If damage is evident please notify the carrier and your Graham Field equipment provider 2 Remove all loose packing from the carton 3 Carefully remove all components from the carton 4 Inspect all components If damage is evident please notify the carrier and your Graham Field equipment provider PUMP AND MATTRESS INSTALLATION 1 Remove all sheets blankets and pillows from the bed 2 Place the mattress on top of the bed frame ensure that the foot end symbol faces up at foot end as shown at right Secure mat tress to bed with straps 3 Fold open the wire hangers on the back of the pump as shown at right 4 Either hang the pump on the bed s foot end as shown at right or place the pump on a flat sur face easily accessible to the care giver and or doctor 760000 INS LAB RevA08 9 Caution Place device in a position where giver and or doctor can access it easily 5 Connect the mattress pump air hose quick connector ensure that the connector has clicked securely into p
3. 9 UNPACKING aa aaa aa 9 PUMP AND MATTRESS INSTALLATION 9 OPERATION B 10 PRESSURE 11 cima agua a dg 11 TRANSPORT CBP e 12 GPR FUNCTION renean 12 uhi apz i pect 12 eHI C jc 12 GENERAL MAINTENANCE sossun ira adiuncta Rud wn auia 13 FUSE REPLACEMENT ca eodera ga aas cati aa uuuua i tai OE ag 13 AIR FILTER REPLACEMENT cran aco 13 SS POP AG IE ce 14 SPECIFICA NONS arinin 14 WARRANTY P 15 GF Graham Field Lumex and AltaDyne are registered trademarks of GF Health Products Inc GF Health Products Inc is not responsible for typographical errors Packaging warranties specifications and products are subject to change without notice 2 760000 INS LAB RevA08 INTRODUCTION Use this manual for set up and operation of the Alternating Pressure Low Air Loss Mattress System Read all instruc tions before using the mattress system Save this manual for future reference INTENDED USE OF
4. heavier patient will need a higher firmer setting but pressure adjustment must ultimately be based on the patient s weight distribution short stocky person may be the same weight as a tall thin person but their settings will not be the same The mattress should be firm enough to support the patient yet soft enough so the patient will sink down to eliminate pressure points Initially inflate the mattress to maximum pressure Then when patient lies on the mattress test the pressure by placing four fingers below the mattress under the patient s but tock area your fingers should just fit If the fit is too loose decrease the setting If your fingers do not fit well increase the setting two notches firmer Wait ten minutes and then check again You should only have to adjust this setting one time per patient Please consult your physician for an appropriate setting QUICK DISCONNECT Pull the power plug from the wall connector to disconnect the device quickly 760000 INS LAB RevA08 11 TRANSPORT CAP QUICK Connector In case of power failure or trans port Disconnect the quick connec tor and cap the mattress with the transport cap to slow deflation CPR FUNCTION If CPR must be performed on the patient while the mattress is in use to immediately deflate the mat tress either 1 Pull the CPR valves from the mattress in the location of the arrow shown at right or m 2 Disconnect the quick connector e from the p
5. of button increase pressure button therapy adjust button therapy mode indicators at right of button panel lock button panel lock 22 la indicator at right of button service indicator low pressure indicator upper j power ON OFF button ON OFF power failure indicator lower indicator at right of button alarm mute button Front panel first row cycle time adjustment There are four cycle time settings Press the cycle time adjust button to cycle between the settings The selected cycle time setting s indicator will illuminate Front panel second row pressure adjustment There are ten pressure firmness settings Press the de crease pressure button at the left end of the row to de crease pressure softer and press the increase pressure button at the right end of the row to increase pressure firmer The selected pressure level setting s indicator will illuminate Front panel third row therapy mode adjustment There are four therapy mode settings Autofirm to fill mattress quickly or to fully inflate mattress during patient ingress egress for better support Alternating Pressure every other tube fills then deflates Static all tubes remain full and Seat Inflation when the head is raised to 760000 INS LAB RevA08 7 increase pressure under buttocks Press therapy adjust button at the left end of the row to cycle between the settings The selected therapy mode setting s indicator wil
6. to 90 non condensing Shipping 10 to 90 non condensing Classification Class Type BF IPXO Applied Part Air Mattress Not suitable for use in the presence of a flammable anesthetic mixture No AP or APG protection 14 760000 INS LAB RevA08 Dimensions L x W x H 78 7 x 35 4 x 8 200 x 90 x 20 3 cm Weight 16510 7 5 kg WARRANTY GF Health Products Inc warrants the Lumex AltaDyne Plus 160000 Alternating Pressure Low Air Loss Mattress Sys tem for a period of twelve months for defects in workman ship and materials During the warranty period defective items will be repaired or replaced at GF Health Products option at no charge Mattress Specification 760000 INS LAB RevA08 15 USA Corporate Headquarters Graham Field Health Products 2935 Northeast Parkway Atlanta Georgia 30360 telephone 800 347 5678 770 447 1609 fax 800 726 0601 678 291 3232 www grahamfield com 760000 INS LAB RevA08 2008 GF Health Products Inc
7. THIS DEVICE The intended use of the Alternating Pressure Low Air Loss Mattress System is e Thetreatment of pressure ulcers while optimizing user comfort Pain management as prescribed by a physician As described above in either a homecare or long term care setting IMPORTANT SAFETY PRECAUTIONS The safety statements presented in this chapter refer to the basic safety information that should be observed by those using the Alternating Pressure Low Air Loss Mat tress System There are additional safety statements in other chapters or sections which may be the same as or similar to the following or specific to the operations DANGER Indicates an imminent hazard situation that if not avoided will result in death or serious injury N WARNING Indicates a potential hazard situation or unsafe practice that if not avoided could result in death or serious injury A Caution Indicates a potential hazard or unsafe prac tice that if not avoided could result in minor person al injury or product property damage Note Provides application recommendations or other useful information to ensure that you get the most from your product 760000 INS LAB RevA08 3 DANGER To reduce the risk of electrocution N DANGER Always unplug this product immediately after using N DANGER Do not use this product while bathing N DANGER Do not place or store this product where it can fall or be pulled into a tub
8. e cover of the fuse holder fuse locations are shown in picture at right Insert a new fuse of the correct rating T1A 250V VDE approved Replace the fuse holder cover Ensure that fuse holder cover is securely installed tool needed small screwariver Use a small screwdriver to remove the air filter plate located at rear of pump filter location shown in picture at right filter Remove the filter The filter is reusable if not torn and can be washed gently with a mild detergent and water Dry the filter before use Replace the filter and cover Ensure that filter cover is securely installed Replace the air filter regularly if it is gray torn or the environment is dirty 760000 INS LAB RevA08 13 1 Lay the mattress out flat and upside down 2 Roll from the head end toward the foot end 3 Stretch the foot end strap around the rolled mattress to prevent unrolling 4 Store mattress and pump in a dust free environment with no exposure to direct sunlight A Caution Do not fold crease or stack mattress SPECIFICATIONS Specification Power Supply 120V System AC 120V 60 Hz Note see rating label on product Oyde me 10 15 20 or 25 minutes Airinput fsm Weight 510 22k Temperature Operation 50 F to 104 F 10 C to 40 C Storage 5 F to 122 F 15 C to 50 C Shipping 5 F to 158 F 15 C to 70 C Operation 10 to 90 non condensing Storage 10
9. ght capacity for this product is 400 Ib 189 kg CAUTION Caution The pump can be used only with the accom panying mattress Do not use it for any other purpose EQUIPMENT SYMBOLS Attention Consult accompanying documents this manual x Indicates that this product is according to the degree of protection against electric shock for type BF equipment Functional earth for UL only n Class II for UL only Disposal of Electrical amp Electronic Equipment WEEE Do not treat this product as household waste For more detailed information with regard to returning and recycling this product please consult your local city office household waste disposal service or Graham Field Health Products equipment provider 760000 INS LAB RevA08 5 For U S and Canada only Medical Equipment Air Pump E228589 S with respect to electrical shock fire and 53DG C US mechanical hazards only in accordance with UL60601 1 and CAN CSA C22 2 No 601 1 PRODUCT DESCRIPTION PUMP AND MATTRESS SYSTEM foot end symbol mattress f mattress pump connectors pump and mattress system PUMP FRONT M ick comme quick connector 6 760000 INS LAB RevA08 PUMP fuse main power 4 switch power cord VA receptacle filter PUMP FRONT PANEL cycle time adjust button cycle time decrease pressure button indicators at right of button pressure level indicators at right
10. l illuminate Front panel fourth row Panel lock button The panel lock protects the panel set tings from accidental change Press to lock press for three seconds to unlock When the panel remains untouched for three minutes it locks automatically When the panel is locked the panel lock indicator will illuminate Power button ON OFF The power button turns power to the mattress system ON or OFF Press to turn ON press to turn OFF When power is on the power ON OFF indicator will illuminate When power is turned OFF mattress system will slowly deflate Note Main power switch on side of pump must be ON in order for the panel power button to function Alarm mute button The alarm mute button temporarily suspends the LED indicator and buzzer when either the low pressure alarm or power failure alarm is activated Should the problem not be resolved within five minutes the alarm will resume Low pressure indicator The low pressure indicator illumi nates and the alarm sounds when pressure level is lower than the selected pressure If the low pressure indicator persistently illuminates inappropriately 1 Check to ensure all connections are properly and se curely connected per installation instructions 2 Check for any leakage tubes or connecting hoses If necessary contact your Graham Field equipment pro vider to replace any damaged tubes or hoses 3 If problem persists contact your Graham Field equip ment provider
11. lace before continuing Caution Ensure that the air hoses are not kinked or tucked under the mattress N WARNING Locate all cords so that they will not be stepped on tripped over or otherwise subjected to damage or stress N WARNING Do not use a household extension cord if the electrical cord does not reach the power outlet Use of an improper extension cord could result in fire and electric shock If an extension cord must be used use a three conductor cord with ground properly wired in good electrical condition and keep it as short as possible N WARNING Ensure that the local power voltage is appropriate for the pump unit 6 Plug the power cord into a properly grounded electrical outlet OPERATION l Turn the main power main M switch to the ON position power switch 2 Press the front panel power button The power indica tor will illuminate power button 10 760000 INS LAB RevA08 Note The mattress system will automatically enter Autofirm mode for the quickest inflation at set up initial inflation takes approximately twenty minutes 3 The pump will begin delivering air into the mattress 4 When the selected pressure level is reached the pump will enter alternating mode The mattress can then be adjusted to the desired pressure 5 Replace all sheets blankets and pillows on the bed PRESSURE ADJUSTMENT Generally a lighter patient will need a lower softer set ting while a
12. or sink N DANGER Do not place this product in or drop into water or other liquid N DANGER Do not reach for a product that has fallen into water Unplug it immediately WARNING To reduce the risk of burns electrocution fire or personal injury WARNING Do not leave this product unattended when plugged in N WARNING Always use close supervision when this product is used by on or near children or those who require close supervision N WARNING Use this product only as intended and described in this manual Do not use attachments or accessories not recommended by Graham Field N WARNING Never operate this product if a it has a damaged cord or plug b itis not working properly c it has been dropped or damaged d it has been dropped into water Return the product to your Graham Field equipment provider N WARNING Keep the cord away from heated surfaces 4 760000 INS LAB RevA08 N WARNING Never block the air openings of this product or place it on a soft surface such as a bed or couch where the openings may be blocked Keep the air opening free of lint hair and other similar particles N WARNING Never drop or insert any object into any opening or hose N WARNING Do not use this product in the presence of flammable anesthesia or where oxygen or nitrous oxide is being administered N WARNING Connect this product only to a properly grounded power outlet N WARNING Maximum patient wei
13. ump in the location of 2 arrow shown at right MAINTENANCE CLEANING Perform the following cleaning procedures before use Pump Wipe the pump with a clean damp cloth and mild detergent and keep it away from dust Air dry A Caution Do not use corrosive or powdered cleansers to clean the pump Caution Do not immerse or soak the pump A Caution Do not clean with phenolic products Mattress Wipe the mattress unit with a clean damp cloth and mild detergent The mattress may also be cleaned us ing a 1076 solution of sodium hypochlorite bleach diluted in water Air dry all parts thoroughly before use A Caution Do not clean with phenolic products A Caution After cleaning air dry the mattress without direct exposure to sunlight 12 760000 INS LAB RevA08 Fuse replacement Air filter replacement GENERAL MAINTENANCE 2 Check main power cord for abrasion or excessive wear Check mattress cover for signs of wear or damage En sure mattress cover and tubes are connected correctly Check airflow from the air hose connector The airflow should alternate between each connector every half cycle time if it s in alternating mode Check the air hoses for any kink or break For replace ment please contact your GF equipment provider tool needed small screwariver If you suspect a blown fuse discon nect the plug from the wall outlet immediately Use a small screwdriver to remove th