Home
        Telex 550 Headphones User Manual
         Contents
1.    e The measuring cell where catalytic conversion occurs   e Anelectronic amplifier circuit for converting the output of the  cell to a DC signal    e  Athermostatic assembly for temperature control in the cell  compartment   e  Areservoir for providing make up water to the humidifier     29 Teledyne Analytical Instruments 1 1    1 Introduction Model 356           1 2 Required Equipment    For proper operation  the analyzer may require accessory equipment   particularly in the area of sample conditioning  The need for additional  equipment is dictated by the conditions of each application     1 2 1 Sample Conditioning    The sample must be free of entrained solids and condensable vapors   and be at a relatively constant pressure between 1 and 100 psig  However   more efficient operation is obtained with pressures in the range of 5 to 10  psig  Pressure surges can carry fluid from the humidifier into the cell and  impair cell operation  Filters  scrubbers  or pressure regulators are often  necessary  depending on local conditions     1     Filters  If filters are necessary  they should be conveniently  located near the analyzer  and installed in a fashion which permits  easy removal for periodic cleaning or replacement     Scrubbers  Ifthe sample contains small quantities of acidic  anhydrides  50     etc   or mercaptans  H S  etc   they will react  with the electrolyte or the cathode  and must be removed  A  caustic scrubber is usually effective     Pressure regulators  Whil
2.   Troubleshooting    If there is no rise in oxygen reading when the sample flow is restarted in  step 5 9 1 b  the measuring cell should be checked for leaks  The two most  likely leaks locations are at the two terminal seal O rings  or at the large      ring in the cell block mounting base     Check that the terminals are screwed tightly into place  Frequently   when a leak occurs at a terminal connection  a greenish crystalline deposit  will be found inside the cell around the terminal mounting screw  This is lead  carbonate  which is formed by carbon dioxide in the air reacting with lead  ions in the electrolyte     If the large O ring at the mounting plate is leaking  the oxygen indica   tion will begin rising almost immediately after the sample flow is stopped  It  will continue to rise until sample flow 15 restarted  at which time it will begin  to gradually decrease             Teledyne Analytical Instruments    5       5 Maintenance  amp  Troubleshooting    Model 356WA       Symptom    No analyzer response to  oxygen     Cell lacks sensitivity     5 8    Teledyne Analytical Instruments    g    Cause    Poor electrical connection   or F1 fuse  a standard 2  amp Slo Blo fuse  has  blown     Cell electrolyte level too  low     Dirty cell     Short between cell cathode  and anode  the screen to  lead  the base material  is  shorted     If d  above corrects the  problem  the cell has  probably been poisoned   probably by fluid flowing into  the cell humidifier column     C
3.   appropriate part  P N P31   To replace humidifier  heater element    Shut off all power and  remove the AC power  cord from the power  source    Refer to wiring diagram C   18593  left hand side   bottom of the cell com   partment     Pull out the block assem   bly from the bottom of the  humidifier column  pull out  the heater element  and  unsolder the wires   Replace with new heater  element  P N R139    Solder wires  turn the  terminals down into the  holder  and replace the  holder     Verify that the power switch  is ON   Check condition of 2A fuse        Remove the       compart   ment cover  Refer to  Interconnection diagram  A 18594  upper left hand  corner     e Connect voltmeter  set to  high AC  to terminals 1  and 2 of terminal strip  TS4  Schematic diagram  D 18632  lower right hand  corner  Wiring diagram C   18593  upper left hand  corner      5 10    Teledyne Analytical Instruments    Trace Oxygen Analyzer Maintenance  amp  Troubleshooting 5       Symptom Cause What To Do   Cell lacks sensitivity  contin  Faulty triac  Q6  in heater    Disconnect the thermo    ued   control or faulty triac driver  A4  switch by removing one of   continued   the wires from TS4    terminal 3  Meter should  show voltage    If voltage shows  short  terminals 3 and 4 with a  test jumper  Meter reading  should drop to about zero   less than 10 VAC     If power fails to go on and  off as jumper is alter   nately installed and  removed between termi   nals 3 and 4  the triac or  tr
4.   temperature somewhat above ambient  so a suitable trap to remove conden   sate without plugging the vent line will be required  The vent line should  also be arranged so that it cannot become plugged by dirt or dust        9 Teledyne Analytical Instruments 3 3    3 Installation Model 356WA                18 1 4                    1 4  Female NPT       115 V 50 60 Hz Purge   7 8  dia  holes Power In   for 1 2  conduit Signal Out             FIGURE 5 ANALYZER OUTLINE DIAGRAM    Figure 5  Gas Connections to Back of Analyzer    3 4 PN Teledyne Analytical Instruments    Trace Oxygen Analyzer Operations 4       Operations       4 1 Filling the Reservoir    The reservoir is located on the right side of the analyzer case   1  Insure that the cap on the drain spout is securely tightened   2  Remove cap from fill port on top of reservoir     3  Pour distilled water into reservoir until it is half full  about one  quart   The water will automatically flow into the humidifier  column     4  Replace cap on fill port and securely tighten  A missing or loose  cap will permit the sample to vent into the analyzer case     4 2 Detector          The cell is located in the heated compartment on the left side of the  analyzer case  as shown in Figure 4 1  To open the compartment  unscrew  the captive knurled knobs at the top and bottom of the compartment and  remove the plastic window     4 2 1 Cell Packaging  The cell is packaged separately from the analyzer  It is filled with  distilled water
5.  Carefully remove the screw which holds the screen assembly in  place  Remove the screen assembly  Use clean tweezers and do    5 2       Teledyne Analytical Instruments    Trace Oxygen Analyzer Maintenance  amp  Troubleshooting 5       Cell Mounting    Nuts  4  N   ash    Cell Block    Terminal seal  O rings  2     Mounting Bolts  4        Lead Electrode    Cathode Screen    Assembly    O Ring    Cover       Figure 5 1  Cell Assembly       9 Teledyne Analytical Instruments 5 3    5       Maintenance  amp  Troubleshooting Model 356WA    not touch it with your fingers  Avoid any possible  contamination ofthe screen     c  Cleanse the cell thoroughly in electrolyte solution    d  Insert the new lead electrode in place in the cell body and secure  with lock washers and screws        Carefully install screen assembly and secure with mounting  screw    f  Immerse the cell assembly in hot cleaning solution and then rinse  in distilled water     g  Install prepared cell assembly as described in Section 4 2 3          Installation     5 6 Screen Assembly    The screen assembly will become discolored after prolonged use due to  contamination  When this occurs  and if the cell no longer displays adequate  sensitivity  the entire cell assembly must be replaced  Refer to Section 4 2 3   Cell Installation for cell installation instructions     5 7 Reservoir and Humidifier Column    Approximately once each year the reservoir and humidifier column  should be drained and cleaned  Use the fol
6.  NOTE  Specify cell class and range of analyzer when ordering   2 O5 O ring  cell terminal  1 025 O ring  cell seal  1 026 O ring  calibrator cover plate  2 O9 O ring  humidifier column  2   8 O ring  reservoir cap  2 0204 O ring  flowmeter  1 A33748 Thermistor assembly  1 A3042 Humidifier column assembly  1 R2454 Humidifier column heater  110V   1  R2453 Humidifier column heater  220V   1 A5267 Reservoir cap  5 F10 Fuse  3AG 2A  5 F6 Fuse  3AG 1 4A  Slo Blo  2 H2 Heater  1 B6274 Flowmeter asm   specify background gas   1 B30868 PC Board  Proportional temp  control   For applications 10 PPM or higher   1 B36026 PC Board  Proportional temp  control  220     1   14449      Board  Proportional temp  control   For applications less than 10 PPM        41274      Board  Proportional temp  control  220     1  B29600 PC Board       converter  isolated  O option   1  A9309 PC Board  Alarm comparator  dual   2 opt      optional    A minimum charge is applicable to spare parts orders     IMPORTANT  Orders for replacement parts should include the model num   ber  serial number  and range of the analyzer for which the  parts are intended     A 2         Teledyne Analytical Instruments    Trace Oxygen Analyzer    Appendix       Orders should be sent to     Teledyne Analytical Instruments    16830 Chestnut Street    City of Industry  CA 91749 1580    Phone  626  934 1500  FAX  626  961 2538    TWX  910  584 1887 TDYANLY         Web  www teledyne ai com    or your local representative    Dra
7.  a particular gas  e g  oxygen  will depend on the  relative partial pressure of that gas on each side of the leak  In a sample  having only a few parts per million oxygen  there will be a net flow of  oxygen inward unless the sample pressure is many thousands of pounds     5 9 1 Leak Detection Procedure  The procedure outlined here is based on the premise that the leak rate is  independent of sample flow rate     a  Stop the sample flow to permit oxygen to accumulate at the point  of the leak     b  After approximately one minute  restart the sample flow  It is  advisable to practice establishing the flow rate to 150                the  reference flow indication on the flowmeter  with one quick turn  of the throttle valve     c  Simultaneous to restarting the flow  start a stopwatch to measure  the time required for the recorder meter to respond to the  accumulated oxygen     d  The following are approximate times for the accumulated oxygen  to reach the cell from various points in the sample path through  the analyzer at 150 cc min     1 5 to 2 seconds Calibrator   3 5 to 3 75 seconds Base of humidifier column    5 5 seconds Flowmeter   6 5 seconds Metering valve   7 5 seconds Gas connection at rear of analyzer    gt  than 7 5 seconds Sample connection lines leading to   analyzer     A leak at this location may indicate a leak in the column or in the   reservoir system     5 9 2 Cell Leak    5 6       Teledyne Analytical Instruments    Trace Oxygen Analyzer Maintenance  amp
8.  cathode screens   initiating an electrochemical reaction  Four electrons are generated by the  oxidation of the lead anode  and are then used to reduce oxygen at the  cathode  The flow of electrons between the anode and cathode creates an  electric current which is directly proportional to the oxygen concentration in  the sample stream  In the absence of oxygen  no oxidation or reduction  takes place  and no current is produced     In simplified form  the reaction may be described as follows  oxygen is  reduced at the cathode by the mechanism             O    2H 0  gt  40H     This cathodic half reaction occurs simultaneously with the anodic half   reaction    Pb   20H        PbO   H O   Ze     The overall reaction is         2Pb  gt  2PbO    2 2 Humidifier    It is necessary to maintain a film of electrolyte on the screens of the  electrode assembly  This means that the humidity of the sample as it flows  through the cell must be such that the water vapor pressure of the electro         Teledyne Analytical Instruments 2 1    2 Operational Theory Model 356WA       lyte is equal to the water vapor pressure in the sample gas  If the humidity  of the sample is too low  water will evaporate from the electrolyte  drying  the cell  If the sample humidity is too high  water will condense out into the  electrolyte  flooding the cell     The sample is humidified by bubbling it through water in the humidifier  column just before it enters the cell         humidifier column is in the s
9.  to prevent oxidation of the electrodes from exposure to the  atmosphere  The cell should be left filled with the distilled water until the  analyzer is installed and ready for operation  The cell should not be exposed  to the atmosphere for any prolonged duration     4 2 2 Electrolyte  The cell electrolyte is Teledyne Type A  used in applications where  there is a complete absence of acidic anhydrides         SO   in the sample  gas  Type A electrolyte is a 10  solution  w v  of reagent grade potassium  hydroxide  KOH  in distilled water        Teledyne Analytical Instruments 4 1    4 Operations Model 356WA       Heater  Cell Compartment Terminal Strip    Control Assembly    10 20    30    Interconnect  Terminal Strip       Span Range Humidifier  Control          6    Current Adj Alarm 1                               Cell  Assembly  Bis  123 45678 91011 12 13 14                Flowmeter Throttle  Valve  Sample  System  Humidifier Assembly  Column with  Assembly Reservoir  Sample  System  Assembly          Figure 4 7  Typical Model 306WA With 2 Alarm Option    WARNING           A electrolyte is caustic  Use extreme care       handling  Protective equipment including but not  limited to gloves and safety glasses should be worn  while handling electrolyte  Refer to the Material Safety  Data Sheet in the Appendix regarding potential hazards  and corrective action in case of accident     Type B electrolyte is a 20   solution  w v  of potassium carbonate and  should be used when t
10. 5 6  292 16 6  293 17 7  294 18 8  295 20 0  296 21 2  297 23 9  298 23 9  299 25 4  300 26 9  301 28 6  302 30 7  303 32 1  304 34 0  305 35 9  306 38 0  307 40 2  308 42 5    NOTE     The MSDS on this material is available upon request  through the Teledyne Environmental  Health and  Safety Coordinator  Contact at  626  934 1592    DN Teledyne Analytical Instruments    
11. ATION  2  en autre ain tete nn nier 3 1  3 2 Electrical Connections                   0  00    3 1  3 3 Sample Connections                                                 3 2    4 Operations    4 1 Filling the Reservoir    4 1  4 2 Detector Bella                     4 1   4 2 1 Cell Packaging cola            ee 4 1   4 2 2         rasteren are 4 1   4 2 3 Cell                                                               4 3  4 3 Throttle Valve          een 4 5  4 4 Humidity            ener nn eel 4 5  ADP OWS        na                         4 7  4 6 Warm Up and Stabilization         nennen 4 7  4 7 Calibration           4 7    5 Maintenance  amp  Troubleshooting    5 1 Flowmeter and Humidifier                                         5 1  5 2 Cell Electrolyte Level    5 1  5 3                                            5 1  54 VallbDralion                     ee 5 1                               ee een 5 2   5 5 1 Electrolyte                                                       5 2   5 5 2 Lead  Electrode anna nt ss 5 2  5 6 Screen Assembly    5 4    Teledyne Analytical Instruments il       Model 356WA    5 7 Reservoir and Humidifier Column                             5 4  5 8 Leak Detection           5 6   5 8 1 Leak Detection Procedure                             5 6   5 8 2                      bide 5 7                                                                                                          1                             anne nungen    2  Brawingikist  na
12. Trace Oxygen Analyzer       Model 356WA    Trace Oxygen Analyzer  Instruction Manual    HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM   PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM          HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FORA  TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED     ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND OR SERVICING  BEFORE  CONDUCTINGANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR MANAGER     P NM356WA  09 14 99  ECO 99 0373          Teledyne Analytical Instruments    Model 356WA       Copyright    1999 Teledyne Analytical Instruments    All Rights Reserved  No part of this manual may be reproduced  transmitted   transcribed  stored in a retrieval system  or translated into any other language or computer  language in whole or in part  in any form or by any means  whether it be electronic   mechanical  magnetic  optical  manual  or otherwise  without the prior written consent of  Teledyne Analytical Instruments  16830 Chestnut Street  City of Industry  CA 91749   1580     Warranty    This equipment is sold subject to the mutual agreement that it is warranted by us  free from defects of material and of construction  and that our liability shall be limited to  replacing or repairing at our factory  without charge  except for transportation   or at  customer plant at our option  any material or construction in 
13. UTION     150 cc min  for the specified sample gas     Excessive flow rate may cause water in the humidifier  to be carried to the flowmeter causing moisture to  accumulate  This can cause the ball to stick in the  flowmeter  To remove moisture  remove the flowmeter  and allow to air or blow dry  Refer to the detailed in   structions in Figure 5 2 for removal and installation of  the column     44 Humidity Control    The humidity control is located on the front panel of the control unit   and is adjusted to maintain a constant electrolyte level in the detector cell   In effect  the control governs the humidity of the sample which is directed    to the cell        Teledyne Analytical Instruments 4 5    4 Operations Model 356WA       Outside Electrode Connected to   Y  Terminal Post      Oxygen Generated Here    Ne   Electrical Terminal Board  Sub Assembly       Wick Saturated With KOH  Electrolyte Solution    Inside Electrode Connected  to      Terminal Post      Hydrogen Generated Here    Figure 8  Calibrator Assembly    4 6    Teledyne Analytical Instruments    Trace Oxygen Analyzer Operations 4       1  At start up  the humidity control knob should be set to 30  Note  the cell electrolyte level as areference  Approximately 3 32  of  the bottom edge of the screen assembly should be immersed in  the electrolyte     2  Operate the analyzer for 24 hours and compare the electrolyte  level with the    reference    established in Step 1  If the level is  lower than the reference  
14. adjust the knob a few divisions  clockwise  if higher  adjust a few divisions counterclockwise     3  Operate another 24 hours and repeat Step 2   4  Continue adjustments at ever increasing intervals until a constant  electrolyte level is attained in the cell     Once the analyzer is suitably located  all components serviced and  installed  and sample and electrical connections made  the instrument is  ready for operation     45 Power    When power is turned on  power is applied to the cell compartment  heater  The cell operates without applied power  but its output will vary  with changes in ambient temperature     46 Warm Up and Stabilization    When the analyzer is initially put into operation  the air in the lines and  sample passages will drive the output indication to the top of the scale  The  time required to sweep out this residual air may be several hours before an  on scale indication is reached  During this time the cell compartment is  heating and reaching its controlled temperature     4 7 Calibration    The analyzer is calibrated by adding a known amount of oxygen into  the analyzer by means of a calibrated span gas  The sensitivity of the ana   lyzer is adjusted until the change indicated by the analyzer 15 equal to the  amount of oxygen known to be present      the span gas  The span gas must  be composed of the same elements as the specified sample gas 1f the  analyzer s calibrated flowmeter is to be used  TBE AI recommends that the  oxygen content of the sp
15. ame  heated compartment as the cell and so is held at the same temperature  The  water in the column  however  is cooled by evaporation into the sample gas   Thus  the sample gas will normally have a humidity that is too low for  equilibrium with the cell  It is assumed here  of course  that since the cell  component is heated above ambient temperature  the sample gas 15 less than  saturated at the compartment temperature when it enters the analyzer     The humidity of the sample is increased to be in equilibrium with the  cell electrolyte by heating the water in the humidifier column  The humidifer  heater is in the base of the column  and the amount of heating is adjusted  with the humidity control that is located on the panel of the control unit     The amount of heating required depends on the sample flow rate  the  sample humidity  and the specific heat of the sample  The correct adjust   ment for the operating conditions of any particular installation is obtained by  checking the cell electrolyte level periodically and replenished when neces   sary according to the instructions in Section 4 2 3  Cell Installation     The humidifier column also contains baffles to stop water from splas   hing up into the line to the sample cell at high flow rates     2 3 Flow System    The analyzer flow system is shown schematically in Figure 2 1  It  includes a needle valve for adjusting the sample flow rate  a flowmeter to  indicate the sample flow required for calibration  the humidif
16. an gas be between 70 90  of full scale on the  range of interest      s    Teledyne Analytical Instruments 4 7    4 Operations Model 356WA       4 8    To calibrate the analyzer     1     Shut the sample gas off  disconnect the sample in line and  connect the the span gas to the sample inlet  Allow the span gas  to flow through the system  Adjust the span gas flow rate until  the flowmeter float is centered in the flowmeter reference  indicator  Note the reading of the external recorder or meter     Turn the range selector switch to the range which encompases  the oxygen concentration of the span gas  The meter or recorder  should move upscale and indicate the exact concentration of  oxygen in the span gas     If the external recorder meter does not indicate the proper  amount of oxygen  adjust the span dial until it does     After the recorder or meter has stabilized on the proper readout   turn off the span gas flow  disconnect the span gas  reconnect  the sample gas and establish the proper flow        Teledyne Analytical Instruments    Trace Oxygen Analyzer Maintenance  amp  Troubleshooting 5       Maintenance  amp  Troubleshooting       After the analyzer has been put into operation and calibration has been  accomplished  routine inspection will be required for normal operation     5 1 Flowmeter and Humidifier    The flowmeter and humidifier column must be checked daily to insure  proper flow  and corrected as necessary  Refer to Section 2 2  Zumidifier  and 4 4  Throttl
17. anges should the instrument be moved to a higher  elevation  in this instance a higher span setting will be required since the  sensing element is sensitive to the partial pressure of oxygen      The ambient temperature does not affect the flow rate through the  analyzer since the flow tube is located      the same temperature controlled  compartment as the cell assembly     A 4         Teledyne Analytical Instruments    Trace Oxygen Analyzer Appendix       When span gases are used or when sample gas is being analyzed the  flow can vary  10 20  and more without changing the reading  It is best   however  to keep the sample flow so that the float in the flow tube is at or  near the center of the target  Otherwise  a different humidifier setting may be  required     If positioning the float in the center of the target is in error by plus or  minus one quarter of the float    s diameter  an error of approximately  1 5   of reading    will be produced     There are instances when it will be necessary to reset or check the exit  flow of sample gas and set or reset the position of the target on the flow tube     e When a different background gas is being analyzed       When the target has been accidently repositioned on the flow tube   The vent flow should be measured either with a volume displacement   flow device  e g  a    bubble o meter     or a calibrated rotometer with correc   tion factors for ambient temperature and pressure  and sample gas density  and viscosity  When usin
18. any  alarm or control system installed must be tested and understood  both as they operate and  as they can be defeated  Any safeguards required such as locks  labels  or redundancy  must be provided by the user or specifically requested of Teledyne     The purchaser must be aware of the hazardous conditions inherent in the process es   he uses  He is responsible for training his personnel  for providing hazard warning  methods and instrumentation per the appropriate standards  and for ensuring that hazard  warning devices and instrumentation are maintained and operated properly     Teledyne Analytical Instruments  the manufacturer of this instrument  cannot accept  responsibility for conditions beyond its knowledge and control  No statement expressed  or implied by this document or any information disseminated by the manufacturer or  his agents is to be construed as a warranty of adequate safety control under the user s  process conditions        Teledyne Analytical Instruments    Trace Oxygen Analyzer       Table of Contents       1 Introduction    1 1 Method of Operation                                                   1 1  1 2 Required Equipment    1 2  1 2 1 Sample Conditioning                                     1 2  1 2 2 Recorder  Meter Readout        1 2    2 Operational Theory    2 Sensor tiennie meren etterende Le      2 1  2 2                                                                                    2 1  2 3          ass 2 2    3 Installation    3 1  LOC
19. e Valve     5 2 Cell Electrolyte Level    The level of electrolyte in the cell must be checked daily and adjusted  as necessary  Refer to Section 4 2 3  Cell Installation  The proper level for  the electrolyte is when the bottom edge of the sensor screen is wetted but not  immersed  Approximately   32   of the silver electrode should extend into the  pool of electrolyte     5 3 Reservoir  The water level in the reservoir should be checked at least every two  weeks  Follow this procedure   a  Remove the cap from the filler spout     b  Obtain a clean glass tube about 4 to 5 inches long and about 4    outer diameter        Lower the tube into the tank through the filler spout until it  touches bottom     d  Place    finger over the end of the tube  and withdraw the tube  from the tank  The height of water in the tube is the height  of the water level in the tank         If the water level in the reservoir is below 1   add a quart of  distilled water     f  Replace the filler spout cap securely after verifying that the      ring seal is in good condition        Teledyne Analytical Instruments 5 1    5 Maintenance  amp  Troubleshooting Model 356WA       5 4 Calibration    The sensitivity of the unit should be checked at two to four week  intervals  Calibration procedures are described in Section 4 7  Calibration     5 5 Cell    The electrochemical reaction in the analytical process results in the  accumulation of lead ions in the electrolyte  making the replacement of the  lead e
20. e the analyzer will accept pressures to  100 psig  a range of 5 to 10 psig is recommended  In addition   pressure surges can affect instrument operation  In either case  the  use of a pressure regulator is advisable  Install the regulator as  close to the sample point as possible to reduce sample travel time  to a minimium  The regulator should incorporate a metallic  diaphragm to prevent the diffusion of atmospheric oxygen into  the sample     1 2 2 Recorder  Meter Readout    The meter installed on the 356WA is either analog or digital  The  recorder used for analyzer signal readout is usually of the self balancing  potentiometric type  It should have an input inpedance of 10 Q or higher   Output is 0 to 1VDC or less  optional  1 to 5  4 to 20       10 to 50 mADC      1 2       s Teledyne Analytical Instruments    Trace Oxygen Analyzer Operational Theory 2       Operational Theory       2 1 Sensor    The sensor is an open cathode cell  an electrochemical transducer  specific to oxygen  The cathode of the cell is composed of silver screen  elements with a large surface area  The screen assembly is mounted in an  acrylic block  with the lower edges of the screens immersed in potassium  hydroxide electrolyte  A thin layer of electrolyte is maintained on the sur   faces of the screens by capillary action  A lead disk is positioned under the  screens and serves as the anode  An exploded view of the cell is showm in  Figure 5 1     The sample gas stream is passed directly over the
21. ell has been poisoned by a  component in the sample     Cell electrolyte level too low  due to misadjusted or faulty  humidity control     Cell electrolyte too high   due to misadjusted or faulty  humidity control     Faulty humidity control     What To Do    Verify proper connection  from the cell through the  control unit to the external  recorder or meter  check for  blown fuse  Replace as  necessary    Inspect electrolyte level and  add as necessary  See  Section 4 2 3  item 4   Remove cell and clean  thoroughly  See Section  4 2 3    Correct short  Refill with  fresh electrolyte as needed     Check for excessive flow   rate    Check for excessive foam   ing in the humidifier column   Drain and clean reservoir as  necessary  Refer to Section  5 8    If there has been no change  in the normal sample  composition  a scrubber  may be required to remove  the offending component     Add electrolyte as neces   sary  Adjust humidity  control  See Section 4 5     Drain electrolyte and refill   Adjust humidity control  See  Section 4 5       Shut off main power       Disconnect orange wire  from TS1 terminal 5   place a voltmeter  set to  AC current  range 0 100  mA  in series with TS1 5  and the disconnected  orange wire     CAUTION  High voltage AC    present     Trace Oxygen Analyzer Maintenance  amp  Troubleshooting 5       Symptom Cause What To Do    Cell lacks sensitivity  contin  Faulty humidity control  contin     Turn humidifier control  ued   ued   completely counter   c
22. g a volume displacement flow device it will be  necessary to correct the 150 cc min flow rate for ambient temperature and  pressure  It will also be necessary  using either type of flow measuring  device  to compensate for the increase in flowrate due to humidifying the  sample gas   The flow tube inside the analyzer is measuring the dry gas  flowrate      To determine the corrected vent flow rate it will be necessary to know  the ambient temperature  in   K   the ambient pressure  in mm Hg  and the  vapor pressure of water at ambient temperature     Ambient temperature can be measured in   Centigrade or   F and con   verted to   K     Degrees      5    F  32   9  Degrees        C   273    Ambient pressure can be measured with an accurate barometer            If the actual oxygen concentration were  for example  8 0 ppm  the  resultant reading would be 7 9  float high  or 8 1  float low              Teledyne Analytical Instruments A 5    Appendix Model 356WA       or by knowing the altitude    P mm Hg    760    2 50 per 100 ft  of altitude        The vapor pressure of water at ambient temperature can be obtained  from Table 1     To determine the corrected flow  F       substitute the ambient tem   perature  pressure        vapor pressure of water      the following formula           corrected        150    x T    760 mm               294                  Where    T   ambient temperature  in   K    P   ambient pressure  in mm Hg    P ae   Vapor pressure of water at ambient te
23. he CO  level in the background gas is between 500    ppm and 1    This narrow range is rarely encountered  It is  however  available from the factory  Safety related information for this electrolyte can    be found in the Appendix   Type C electrolyte is a 20   solution  w v  of potassium bicarbonate    and should be used when the CO  level in the background gas is between 1  and 100    Safety related information for this electrolyte can be found in    the Appendix     4 2         Teledyne Analytical Instruments    Trace Oxygen Analyzer Operations 4       Sufficient electrolyte is provided for initial servicing of the cell  Elec   trolyte for future service should be ordered from Teledyne  When ordering   specify type and quantity     4 2 3 Cell Installation    Prior to servicing and installing the cell  inspect the lead electrode in  the acrylic base for signs of oxidation  indicated by a reddish brown or  yellow discoloration  If discoloration is noted  clean the cell as directed in  section 5 5 2 before placing it in service     WARNING  Type    electrolyte is caustic  Use extreme care in    D    2     3     4     handling  Protective equipment including but not  limited to gloves and safety glasses should be worn  while handling electrolyte  Refer to the Material Safety  Data Sheet in the Appendix regarding potential hazards  and corrective action in case of accident   Remove the four cell mounting bolts which secure the plastic  cover  Pour out the distilled water     Po
24. iac driver is probably at  fault and should be  replaced       If power does      on         off  the thermoswitch  P N  T22  is faulty and should  be replaced    If none of the above  improve the sensitivity   replace the cell        Teledyne Analytical Instruments 5 11    5 Maintenance  amp  Troubleshooting Model 356WA       5 12       Teledyne Analytical Instruments    Trace Oxygen Analyzer Appendix       Appendix       Specifications    Standard Ranges  Three ranges between 0 10 PPM and  0 5000 PPM O   0 2 PPM O  avail   able as an option     SamplingSystem  Wetted parts  304 welded stainless  steel    Sensitivity  1  of low range   Accuracy   2  of low range   Response and Recovery Time  90  in less than 1 minute  for lowest  range    Operating Temperature Range   40   F to  120   F   5   C to  49   C    Sample Requirement  Flow  150cc min    Pressure  1 to 150 psig   Temperature   60   F to  100   F    15   C to  38   C   Power Requirement  115 VAC  50 60 Hz  150 W  other  voltages available    Alarms  Adjustable single or dual alarm  op   tional  setpoints   Recorder Signal Output  Voltage  0 1 VDC or less   Current  1 5  4 20  10 50 mADC   optional    Local Readout  Digital or analog meter       Specified flow rate required only during calibration  Measuring  cell is not sensitive to changes in flow rate           Teledyne Analytical Instruments A 1    Appendix Model 356WA       Spare Parts List  QTY  PART NO  DESCRIPTION  1 B1473 Lead electrode  1 C1372 Cellassembly 
25. ic pressure  A   s   female NPT fitting is installed on the back of the instrument for making  sample and vent line connections  Thepurge line is fitted with a   4  tube  fitting    1  Connectors  Use straight tube connectors where possible   This facilitates removal of the analyzer section from the case  during maintenance or service     2  Lines  Lines should consist of metallic tubing  since oxygen can  diffuse through plastic  Use continuous tubing where possible     3  Vent  The analyzed sample is vented through the back of the  unit as shown in Figure 2 3     3 2    9 Teledyne Analytical Instruments          Output    GND    Current  Output       Trace Oxygen Analyzer Installation 3       HOT   AC Power In  NEUT 150 Watt Max  GND Load   NO        Alarm 2   NC   NO        Alarm 1   NC       Figure 2 2  General Connection Diagram    See the specific Interconnection Diagram for your instrument in the drawing  package located at the back of the manual  See also any Addenda that may be  Included with this manual     The analyzer should have a vent line of 44  diameter tubing at least two  feet long  running downward from the vent connection  This is to prevent  air from diffusing into the reservoir and dissolving into the humidifier make   up water     If it is not desirable to vent the sample into the atmosphere  a vent line  to carry the sample to a suitable venting area will be required  The sample  leaves the vent connection of the analyzer saturated with water vapor ata
26. ier  the measur   ing cell  and an automatic level control system for the water in the humidi   fier     As can be seen from Figure 2 1  the sample enters the humidifier  column against the pressure of a water column from the base of the humidi   fier to the water level in the reservoir  which is approximately 4 inches   This determines the minimum sample pressure at which any sample can flow  through the analyzer  In practice  the sample pressure must be somewhat  greater than this in order to have an adequate flow rate     The automatic level control in the humidifier column is accomplished  by connecting the sample outflow from the cell to the bottom of the reser     2 2    Teledyne Analytical Instruments    Flowmeter    Sample In          gt   Throttle  Valve    Trace Oxygen Analyzer    Operational Theory 2                                 Humidifier  Column    Cell    Pressure  ATM    b a        Reservoir    L Drain         Cell Compartment    Figure 2 7  Flow System Schematic       9 Teledyne Analytical Instruments 2 3    2    Operational Theory Model 356WA       voir  This places a back pressure on the sample in the cell and upper portion  of the humidifier column equal to the water column from the bottom of the  reservoir to the water level in the reservoir  Thus  the water level in the  humidifier column is held even with the sample connection at the bottom of  the reservoir  There will be a slight additional pressure in the top of the  humidifier column depending on 
27. l Strip  No  4    LIS       Cu       Sample Flow    Control Valve       Cell Block Humidifier Column Heater    Humidifier Column    Figure 4 2  Cell Compartment Components    4 4    Teledyne Analytical Instruments    Trace Oxygen Analyzer Operations 4       6  Connect the red lead to the center terminal and the black lead to  the outer terminal     NOTE  The silver screens in the cell have been specially treated to pro   vide proper detection characteristics  They must be kept clean and  MUST NOT be touched  Even clean fingers secrete natural oils  which contaminate the screens  If the screens need straightening   wash a small pair of tweezers thoroughly to remove any grease   rinse them in distilled water  and use them to carefully bend the  screens back into place     4 3 Throttle Valve    The throttle valve is located at the top of the reservoir tank  Refer to    Figure 4 1     1  Gently turn the valve counterclockwise  A stream of bubbles    should appear at the base of the humidifier column  and the float  of the flowmeter should rise in its tube     2  Adjust the valve so that the flowmeter float is centered in the    CAUTION     flow rate reference indicator     Open the throttle valve carefully  Excessive flow rate  may cause water in the humidifier column to be carried  into the detector cell  This can cause erratic readings  and may require disassembly  cleaning  and refilling of  the sensor     3  The flowmeter indicator has been factory set to a flow rate of    CA
28. lectrode or the electrolyte necessary     5 5 1 Electrolyte Replacement  Replace the electrolyte at least every two months  or even more fre   quently if foreign matter is accumulating in the cell  Remove the cell from is  compartment  and drain  clean  rinse  and refill as described in Section 4 2 3   Cell Installation  After the cell is serviced or replaced  calibrate the analyzer  as outlined in Section 4 7  Calibration     5 5 2 Lead Electrode  If electrode is discolored when new  or has obviously deteriorated from  use  it may be necessary to clean or replace the electrode  Use the following  procedure while referring to Figure 5 1     If the lead electrode is simply discolored  clean the entire cell according  to the following procedure     a  Heat a quart of Teledyne cleaning solution to slightly below the  boiling point  and completely fill the cell cavity with the heated  solution         Let the solution stand for approximately five minutes  Drain and  dispose of solution in an approved manner         Repeat steps a  and b      d  Rinse the cell with distilled water and then fill with electrolyte   Let stand for approximately two minutes and then drain and  dispose of electrolyte in an approved manner         Refill the cell with electrolyte  immersing the lower edge of the  silver screens to about a 3 32   depth     If the lead electrode is obviously beyond repair  it must be replaced     a  Remove the terminal nuts by removing them from the mounting  screws     b 
29. lockwise     Turn on main power   While watching the  voltmeter  turn the  humidifier control clock   wise  the meter reading  should go from 0 50  milliamps  If not  replace  the humidifier heater  element or the humidifier  heater control  To deter   mine which should be  replaced  see paragraph  below     If there is no current  either  the humidifier element or  control is defective  To  determine which it is        Turn the humidifier control  knob fully cew   Disconnect the voltmeter  from TS1 5 and orange  wire  Reconnect the  orange wire    Change the voltmeter  function to 200V        Place the voltmeter  across TS1 5 and termi   nal TS1 4    Watching the meter  turn  the humidifier control  knob cw and watch the  voltmeter read 0 110 to  120V AC    If not  replace R2  behind  the control knob   see  paragraph below    If there is voltage  replace  the heater element  see  paragraph below             Teledyne Analytical Instruments 5 9    5 Maintenance  amp  Troubleshooting       Symptom Cause    Cell lacks sensitivity  contin  Faulty humidity control  contin     d  Cell compartment not at d   proper temperature   120   F      e  Faulty triac  Q6  in heater e   control or faulty triac driver   A4      Model 356WA    What To Do    To replace humidifier  control    Shut off all power and  remove the AC power  cord from the power  source    Refer to wiring diagram D   18633  upper right hand  corner of the control unit    Remove the wiring and  controls  Replace R2 with
30. lowing procedure     a  Reduce sample flow to approximately 50 cc min     b  Refer to Figure 4 1  Place a small funnel with attached tubing  beneath the drain spout which 15 located on the underside of the  reservoir  Remove the drain spout cap and allow the reservoir  and humidifier column to drain thoroughly     c  Replace the drain spout cap    turn the sample flow off  and  remove the fill cap on the top of the reservoir         Add a small amount of electrolyte to a pint of warm distilled  water  Pour the solution into the reservoir and replace the fill            e  Gently open the throttle valve and permit the sample to flow for  five to ten minutes     f  Reduce the sample flow to 50 ec min  and drain the solution  from the reservoir  When drained  replace the drain cap and turn  off the sample flow     g  Rinse by filling with distilled water and draining several times     h  If the humidifier column still retains deposits on its walls  it should  be removed and cleaned with a brush and suitable cleaner  See       9 Teledyne Analytical Instruments    Trace Oxygen Analyzer Maintenance  amp  Troubleshooting 5       Figure 5 2  After cleaning   thoroughly rinse the column in  distilled water before  reinstalling     NOTE   felecirolyte has severely etched the  column  It should be replaced     i  After the reservoir and column  have been cleaned  refill the  reservoir as outlined in Section  4 1  Filling the Reservoir     Removing the Humidifier  Column    1  Grasp the 
31. mp   in mm Hg    By way of example  suppose that the target   s position on the flow tube   is in question and it has been determined that the ambient temperature is 77    F and the altitude where the instrument is being used is 5000 ft  above sea  level     Ambient Temp    K   57    32  273 298  K    Ambient Pressure  P mm Hg    760    2 50   50    635 mm Hg    Vapor Pressure of Water  at 298   K   23 9 mm Hg       Corrected Flow        150x 298 X 760  a    corr   189  lt          294 635         This approximation is within  0 5  of the ICAO Standard Atmo   sphere over the range 0 7500 ft  above sea level  Ref     Fluid Mechanics for  Engineering Technology    by Irving Granet  Prentice Hall  pp 83 84            Reference ambient temperature   294   K  21   C  was the ambient  temperature used in the Faradaic calculations     A 6         Teledyne Analytical Instruments    Trace Oxygen Analyzer Appendix       From these computations  the exit flow rate should be set to 189 cc min   using an appropriate flow measuring device and the target respositioned if  necessary so that the float is centered within the target opening     Subsequently  it should not be necessary to measure the exit flow unless  another circumstance  of the type listed above  occurs             Teledyne Analytical Instruments A 7    Appendix    Model 356WA       Table 1  Vapor Pressure of Water   From 288 308 K   Ambient Temperature    K  Vapor Pressure of Water  mm Hg   288 12 9  289 13 7  290 14 6  291 1
32. nce of sample lines and power     NOTE  Since the level of the electrolyte in the measuring cell is critical and  the water level control system for the humidifier is gravity sensitive   THE ANALYZER MUST BE MOUNTED SO THAT THE BOTTOM OF  THE CASE IS LEVEL     3 2 Electrical Connections    A diagram of the necessary electrical connections is shown in Figure 2 2     Note  See the Interconnection Diagram  drawing A 37526  included in the  back of this manual  as well as any Addenda that may be included  with this manual for information specific to your instrument             Teledyne Analytical Instruments 3 1    3 Installation Model 356WA         Percent Oxygen Sample Line  Recorder Signal Leads Analyzer  1 4  or 1 8  Metal Tubing    22 ga  Twisted  Pair Shielded  Cable     Pressure Shut Off  Regulator Valve   5 10 psig Output        Sample    115 VAC Power and Ground       2 Power Leads  1 Ground Wire      16       Insulated  Condensate Trap    Figure 2 1 Typical System Layout    The connections include a terminal for grounding the analyzer case and  chassis in accordance with accepted industrial practices  The maximum  power requirement is less than 112 amperes at 115 VAC     3 3 Sample Connections    The sample line is connected at the back of the analyzer case as de   picted      Figure 2 3  Use care in assembling any part of the sampling system  to avoid leaks  Oxygen can diffuse into the system through small leaks even  when sample pressure is much greater than atmospher
33. sse    3  Calibration Data  a ae    4  Material Safety Data                                                         A 9       Teledyne Analytical Instruments    Trace Oxygen Analyzer Introduction 1       Introduction    The Teledyne Analytical Instruments Model 356WA Trace Oxygen  Analyzer is designed to detect oxygen concentrations in process streams  It  utilizes Teledyne   s patented electrochemical sensor which requires minimal  maintenance and exhibits a 90  response in less than one minute  Cell  output is insensitive to flow rate changes within the operating range of the  analyzer   sflowmeter     The Model 356WA features a welded stainless sampling system for  long term  leak free operation     While the analyzer is offered in several configurations  they are virtually  identical with the exception of housing or options such as special meters  For  purposes of clarity  this manual will discuss the unit in general  The differ   ences between the configurations are minor and will be obvious to the user     1 1 Method of Operation    Gas from the process stream is fed through a sample line to the sample  inlet port of the analyzer  The sample is directed through the analyzer   s  sample system  where oxygen concentration is detected by the sensor  The  sensor generates an output signal which is read out on a suitable recorder or  meter     The analyzer components include     e    throttle valve and flowmeter to control sample flow   e A humidifier to condition the sample
34. the flow rate  the pressure needed to push  the sample through the cell and associated tubing   but at normal flow rates  this merely slightly lowers the level in the humidifier column     The sample bubbles through the water in the reservoir on its way to the  outlet port  Some of the water vapor will re condense  so that the sample  flows out of the outlet port saturated at the reservoir temperature  which is  slightly above ambient  The sample bubbling through the make up water  will scrub out any oxygen which may be dissolved in it  This assures that the  sample will not pick up any oxygen as it passes through the humidifier  column        Teledyne Analytical Instruments    Trace Oxygen Analyzer Installation 3       Installation    3 1 Location    With proper shielding of the leads  the analyzer and the readout device  can be separated by as much as 1 000 feet  However  they should be placed  as close together as possible  For the most convenient operation  the read   out recorder or meter should be within view of the controls  particularly  when the unit is being calibrated  Figure 2 1 depicts a typical system layout   Other location considerations     1  The analyzer should be sheltered from the elements    2  Ambient temperature must be within 30 to 120   F    3  The unit should not be subject to excessive shock or vibration   4  Itshould be as close as possible to the sample point     5  There must be access to the back and side of the unit for  connection or maintena
35. tube  and with a  twisting motion  work it up into  the top humidifier block until it  clears the bottom block  completely     2  Angle the bottom of the tube  towards you and twist it free of  the top block        3  The top and bottom O rings will  remain captive in their blocks     To Reinstall    1  Engage the top of the tube in the  top block at an angle and twist  up into the cavity of the top  block unit the bottom of the tube  clears the bottom block     2  Swing the tube into line  and  twist down into the bottom  block until the tube seats     NOTE  77   flowmeter is similarly constructed  and can be removed and reinstalled  following the same procedure outlined  above        Figure 5 2  Removing Replacing the  Humidifier    Refer to the Spare Parts List in the Appen   dix for replacement part numbers for the       9 Teledyne Analytical Instruments 5 5    5 Maintenance  amp  Troubleshooting Model 356WA       flowmeter column or humidifier column     5 9 Leak Detection    The most frequent cause of trouble in trace measurement is leakage   Tiny leaks which may be unnoticeable can cause serious errors in trace  measurements  One of the principal problems is that air can diffuse into a gas  line through a small leak  even though the gas pressure in the line may  greatly exceed atmospheric pressure     When a leak occurs in a system where the mass flow velocity is less  than the molecular velocity  gas molecules move in both directions through  the leak  The net flow of
36. ur about half the furnished electrolyte into the cell and slosh  until all components within the cell are wetted by the solution   Drain and dispose of the solution     Wipe the top of the cell and the           with a clean  disposable  tissue to remove solution from the exterior  DO NOT touch the  interior of the cell     Carefully pour in electrolyte until it just touches the bottom edge  of the silver screen assembly at all points  This 15 indicated by     definite wicking of electrolyte onto the screen assembly at every  point along ithe bottom edge  It is essential at this point that the  bottom edge of the screen assembly be wetted at all points  as  seen by the wicking action   but not over immersed  as large a  surface area as possible of the screen assembly must remain  above the electrolyte  while every point of the bottom edge must  be wetted   The level is correct when the bottom edge of the  sensor screen is wetted but not immersed  approximately 7 52  of  the silver electrode extends into the pool of eelectrolyte     NOTE  The electrolyte level in the cell is critically related to its sensitivity  due to the change in the cathode surface area exposed to the  eledctrolyte     5     Carefully place the cell under the cell mounting plate with the  outer terminal toward the front  Secure in place with four bolts  supplied with the cell  Refer to Figure 4 2             Teledyne Analytical Instruments 4 3    4 Operations Model 356WA    Cell Mounting  Bolts    Termina
37. which defects become  apparent within one year from the date of sale  except in cases where quotations or  acknowledgements provide for a shorter period  Components manufactured by others bear  the warranty of their manufacturer  This warranty does not cover defects caused by wear   accident  misuse  or neglect  We assume no liability for direct or indirect damages of any  kind and the purchaser by the acceptance of the equipment will assume all liability for  any damage which may result from its use or misuse     We reserve the right to employ any suitable material in the manufacture of our  apparatus  and to make any alterations in the dimensions  shape or weight of any parts  in  so far as such alterations do not adversely affect our warranty     Important Notice    This instrument is intended to be used a tool to gather valuable data  The informa   tion provided by the instrument may assist the user in eliminating potential hazards  caused by the process that the instrument is intended to monitor  however  it is essential  that all personnel involved in the use of the instrument or its interface with the process  being measured be properly trained in the process itself  as well as all instrumentation  related to it     The safety of personnel is ultimately the responsibility of those who control process  conditions  While this instrument may be able to provide early warning of imminent  danger  it has no control over process conditions  and can be misused  In particular  
38. wing List       5855      47143  D 47142  B 30364  B 14718  B 15016  D 22297  D 47139  B 21916    Outline  Piping  Schematic  Schematic  Schematic  Schematic  Wiring  Wiring    Interconnection         Teledyne Analytical Instruments A 3    Appendix Model 356WA       Calibration Data    Range  The ranges of this analyzer are     Range Switch Position No  1 PPM O   Range Switch Position No  2             Range Switch Position No  3 PPM        Output Signal  The output signal is D C     Background Gas    This analyzer is intended to measure oxygen in a background of     The flowmeter has been set to indicate a flow of 150 cc min  of this gas   If any other type of gas is to be analyzed  the flowmeter must be reset for that  gas  using a displacement type flowmeter  when the flow 15 set to 150 cc   min     Cell Class     Electrolyte Type  Type A  10  potassium hydroxide in distilled water  Type B  10  potassium carbonate in distilled water  Type C  20  potassium bicarbonate in distilled water    Calibration Considerations    In order to calibrate the Model 356  all that is required is to center the  float of the flow tube in the target  add a known quantity of oxygen  0 100  PPM  via a calibrated span gas and adjust the span potentiometer to match  the known oxygen concentration of the span gas  It should not be necessary  to compensate for changes in altitude or ambient temperature     The flow tube is a mass flow device and therefore automatically com   pensates for altitude ch
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
  AVEA Auto-FIO2 Option      :150kg以 :鋼 鉄製  Manuel d`utilisation de la maquette CENTRALE  Bedienungsanleitungen - Wenger Notebooktasche  Manual del Usuario - Aula Virtual V1  Baladeur MP3-CD - Instructions Manuals  BSM370 USER`S MANUAL Thank you for purchasing the BSM370    Copyright © All rights reserved. 
   Failed to retrieve file