Invacare 1148114 Oxygen Equipment User Manual

image

Contents

1. or death High pressure oxidizing gas vigorously accelerates combustion DO NOT alter or modify cylinder or related components DO NOT use Oil or Grease Keep cylinder cylinder valve all equipment and connections free of oil and grease to avoid violent ignition resulting in serious bodily injury or death For medical applications Use only as directed by a licensed practitioner Uninterrupted use of high concentrations of oxygen over a long duration without monitoring its effect on oxygen content of arterial blood may be harmful Oxygen Cylinders filled by the Venture HomeFill shall be used for personal use only Not to be filled for resale or use by professional users This cylinder is to be used with the Invacare Venture HomeFill system This cylinder contains 93 3 oxygen produced by an oxygen concentrator using the pressure swing adsorption process DO NOT handle cylinder or use contents until you are professionally trained including emergency procedures Use in accordance with Venture HomeFill Operators Manual Open valve slowly Close valve after each use and when empty Secure cylinder during storage and use No smoking in cylinder area Keep away from heat flame and spark Keep out of reach of children DO NOT drop Transfilling of this gas is performed by Venture HomeFill Only Cylinders that have been refinished and or exposed to elevated temperatures must be hydrostatically tested before filling Cylinder temper
2. repair and or replacement For warranty service please contact the Invacare service department during normal business hours at the toll free telephone number identified on the back cover of this manual Upon receiving notice of an alleged defect in a product Invacare will issue a serialized return authorization It shall be the responsibility of the purchaser to pack the product s or part s in an appropriate manner to avoid shipping damage and return the product s or part s at the purchaser s expense to either Invacare s plant or service center as specified by Invacare in advance The defective product s or component part s must be returned for warranty inspection using the serial number as identification within 30 days of the return authorization date DO NOT return any product s or part s to our factory or service center without prior consent C O D shipments will be refused please pre pay shipping charges HomeFill 22 Part No 1148114 LIMITED WARRANTY LIMITATIONS AND EXCLUSIONS THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE PRODUCTS MODIFIED WITHOUT INVACARE S EXPRESS WRITTEN CONSENT INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIF
3. the cylinder is full press the compressor power switch to the Off O position FIGURE 4 6 12 Grasp the cylinder assembly FIGURE 4 6 13 With the other hand grasp the outer ring sleeve of the connector fillport and push DOWN FIGURE 4 6 14 Lift up on the cylinder assembly to remove from the connector fillport FIGURE 4 6 PUSH DOWN Outer Ring Sleeve Cylinder 1 Cylinder m l Fillport Assembly Connector Compressor Fillport FIGURE 4 6 Disconnecting the Cylinder from the Compressor 15 When the cylinder fillport is disconnected from the connector fillport release the outer ring sleeve of the connector fillport FIGURE 4 6 16 Remove the cylinder from the compressor cradle FIGURE 4 6 Part No 1148114 17 HomeFill SECTION 4 INSPECTION AND COMPRESSOR INTERFACE A WARNING The fillport cover on the connector fillport and the post valve fillport MUST be replaced after filling and whenever not in use If either of the fillport covers are missing contact your healthcare provider before using 17 Place the fillport covers onto the connector fillport located on the compressor and post valve fillport located on the cylinder FIGURE 4 7 18 Install a CGA870 post valve compatible conserver or regulator according to the manufacturer s instructions 19 Open the toggle valve by turning counterclockwise and examine the conserver or regulator gauge to ensure cylinder is full
4. IC CONSENT OF INVACARE PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE S CONTROL OR PRODUCTS REPAIRED BY ANYONE OTHER THAN INVACARE OR AN INVACARE SERVICE CENTER SUCH EVALUATION SHALL BE SOLELY DETERMINED BY INVACARE THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER WHETHER EXPRESSED OR IMPLIED INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS SO THE ABOVE EXCLUSION AND LIMITATION MAY NOT BE APPLICABLE THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE PROVINCIAL LAWS AND REQUIREMENTS Part No 1148114 23 HomeFill Invacare Corporation www invacare com MET e USA INVACARE One Invacare Way Elyria Ohio USA 44036 2125 Yes you can 800 333 6900 Technical Services 800 832 4707 All rights reserved Trademarks are identified by the symbols and All trademarks are owned by or licensed to Invacare Co
5. Operator s Manual HomeFill Oxygen Post Valve Cylinder HF2POST6 HF2POST9 HF2POSTD Caution Federal law restricts this device to sale or rental by or on order of a physician or any other practitioner licensed by the law of the State in which he she practices to use or order the use of this device Dealer This manual MUST be given to the end user User Before using this cylinder read this manual and save for future reference For more information regarding Invacare products parts and services INVACARE 221 please visit www invacare com Yes you can A WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS CAUTIONS AND INSTRUCTIONS CONTACT A HEALTHCARE PROFESSIONAL DEALER OR A QUALIFIED TECHNICIAN BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE INJURY OR DAMAGE MAY RESULT A WARNING Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories Accessories designed by other manufacturers have not been tested by Invacare A DANGER DO NOT SMOKE while using this device Keep all matches lighted cigarettes candles or other sources of ignition out of the room in which this product is located and away from where oxygen is being delivered Textiles and other materials that normally would not burn are easily ignited and burn with great inten
6. anger indicates an imminently hazardous situation which if not avoided will result in death or serious injury WARNING Warning indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION Caution indicates a potentially hazardous situation which if not avoided may result in property damage RECEIVING INSPECTION Remove product from package and inspect for damage If there is any damage DO NOT use Contact your dealer Home Care equipment provider INTENDED USE To store supplemental oxygen for use with a CGA 870 post valve compatible conserver or regulator to be used by patients who may have difficulty extracting oxygen from the air they breathe NOTE Updated versions of this manual are available on www invacare com HomeFill 4 Part No 1148114 SECTION GENERAL GUIDELINES SECTION GENERAL GUIDELINES A WARNING SECTION IMPORTANT SAFEGUARDS contains important information for the safe operation and use of this product DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as Owner s Manuals Service Manuals or Instruction Sheets supplied with this product or optional equipment If you are unable to understand the Warnings Cautions or Instructions contact a health care professional dealer or technical personnel bef
7. ature should not exceed 125 F 52 C Aluminum cylinders subject to the action of fire or heated to temperatures in excess of 350 F must be withdrawn from service and condemned by trained personnel DO NOT use caustic paint strippers Valve and safety relief devices must be removed and replaced by trained and authorized personnel DO NOT alter or change valve assembly Invacare Corporation DO NOT REMOVE THIS PRODUCT LABEL E Lac May Not for Commercial Resale Sanford FL 32773 1 800 832 4707 Part Number 1125090 Rev B 11 04 FIGURE 4 1 External Examination Checking Cylinder Pressure A WARNING NEVER use tools of any kind to connect disconnect the cylinder and the compressor Otherwise severe injury and or damage may occur DO NOT drop oxygen cylinders Use two hands when handling transporting oxygen cylinders Otherwise injury or damage may occur DO NOT use a liquid leak detector to test for leaks NOTE For this procedure refer to FIGURE 4 2 on page 14 1 Installa conserver or regulator onto the cylinder according to the device manufacturer instructions HomeFill 12 Part No 1148114 SECTION 4 INSPECTION AND COMPRESSOR INTERFACE 2 Ensure toggle valve is open Open the toggle valve by rotating counterclockwise 3 Examine the gauge on the connected conserver or regulator Perform one of the following If the cylinder pressure is less than 1500 psig fill the cylinder Connect the cylinder
8. ble explosive substances vapors or atmosphere NOTE For this procedure refer to FIGURE 4 1 on page 12 The Hydrostatic Testing Date Aluminum cylinders MUST undergo testing every five years A WARNING DO NOT fill cylinders that have not been tested in the past five years Otherwise serious personal injury may result Contact your dealer for replacement HomeFill 10 Part 1148114 SECTION 4 INSPECTION AND COMPRESSOR INTERFACE External Examination NOTE For this procedure refer to FIGURE 4 1 on page 12 1 Examine the outside of the cylinder for the following conditions and replace the cylinder if they exist Dents or dings Arc burns Oil or grease Any other signs of damage that might cause a cylinder to be unacceptable or unsafe for use 2 Examine the cylinder for evidence of fire or thermal damage Evidence includes charring or blistering of the paint or other protective coating or heat sensitive indicator If fire or thermal damage is found replace the cylinder 3 Inspect the cylinder fillport for the following and replace if found Debris oil or grease e Noticeable signs of damage e Signs of corrosion inside the valve e Signs of excessive heat or fire damage Part No 1148114 HomeFill SECTION 4 INSPECTION AND COMPRESSOR INTERFACE DANGER Improper filling or abuse of this cylinder or failure to follow these safety instructions may cause serious bodily injury
9. by the oxygen patient and are not to be distributed to any other individual for any purpose DO NOT allow oil from your hands or other sources to come into contact with the cylinder fittings These substances may become flammable in the presence of oxygen and cause injury Avoid touching the cylinder fitting whenever possible Example of items to avoid are hand creams or lotions cooking oils suntan oils or sunblocks and similar oily products If you use these types of products Invacare suggests you wash your hands prior to using this product If the cylinder fitting should come in contact with such substances it should be cleaned off with a damp cloth prior to connecting it to the compressor HomeFill 6 Part No 1148114 SECTION GENERAL GUIDELINES A WARNING NEVER oil or lubricate the cylinder fittings NEVER use any penetrating oil or lubricant such as WD 40 or 3 in One If connection difficulty is experienced verify that the cylinder fitting is being inserted straight down into the compressor coupler If continued difficulty is experienced contact your supplier for assistance DO NOT allow cylinders to tip or fall Secure gas cylinders so they cannot fall Ensure cylinder is always in upright position DO NOT store cylinders near sources of heat or flame DO NOT use while sleeping without consulting your healthcare provider DO NOT place the unit under clothing while in use When this product is in use a small amount o
10. f oxygen is vented Wearing the unit under clothing may saturate fabrics with oxygen and cause them to burn rapidly if exposed to sparks or flame It may take several hours for oxygen levels in fabrics to return to normal Part No 1148114 7 HomeFill SECTION 2 FEATURES SECTION 2 FEATURES Toggle Valve Cylinder Fill Port FIGURE 2 1 Product Diagram HomeFill 8 Part No 1148114 SECTION 3 TYPICAL PRODUCT PARAMETERS SECTION 3 TYPICAL PRODUCT PARAMETERS DIMENSIONS HF2POST6 Weight 4 0 Ibs 1 81 kg HF2POST6 Overall Length 1 1 in 2 8 cm HF2POST6 Diameter 4 4 in 11 1 cm HF2POST Weight 5 0 Ibs 2 27 kg HF2POST9 Overall Length 14 in 35 5 cm HF2POST9 Diameter 4 4 in 11 1 cm HF2POSTD Weight 6 30 Ibs 2 85 kg HF2POSTD Overall Length 20 8 in 52 8 cm HF2POSTD Diameter 4 4 in 11 1 cm STORAGE CONDITIONS TEMPERATURE 40 F 40 C to 140 F 60 C MAXIMUM HUMIDITY 95 Part No 1148114 9 HomeFill SECTION 4 INSPECTION AND COMPRESSOR INTERFACE SECTION 4 INSPECTION AND COMPRESSOR INTERFACE NOTE Refer to the HomeFill Compressor Owner s Manual for compressor operating instructions Cylinder Prefill Inspection A WARNING All cylinders MUST be inspected before attempting to fill otherwise injury or damage may occur NEVER smoke in an area where oxygen is being administered NEVER use near any type of flame or flamma
11. ggle Valve closed 4 Conserver not sensing breath SOLUTION Set to prescribed setting 2 Replace fill cylinder 3 Open toggle valve 4a Check position of cannula in nose Do not breathe through mouth 4b Check to see if the cannula is kinked Replacement Parts Accessories HF2POST6BAG or HF2POST9BAG Carrying Bag HomeFill 20 Part No 1148114 Part No 1148114 21 HomeFill LIMITED WARRANTY LIMITED WARRANTY PLEASE NOTE THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4 1975 This warranty is extended only to the original purchaser user of our products This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state Invacare warrants this product to be free from defects in materials and workmanship for a period of three years from date of purchase If within such warranty period any such product shall be proven to Invacare s satisfaction to be defective such product shall be repaired or replaced at Invacare s option This warranty does not apply to problems arising from normal wear and tear or failure to adhere to these instructions nor does it include shipping charges incurred in replacement part installation or repair of any such product Invacare s sole obligation and your exclusive remedy under this warranty shall be limited to such
12. needle pointing into GREEN area If the cylinder is not full perform the following Repeat Filling the Cylinder on page 13 If cylinder is not full after second attempt to fill contact the provider Toggle Valve Connector Fillport Cover Fillport Cover FIGURE 4 7 Connector Fillport HomeFill 18 Part No 1148114 SECTION 5 SECTION 5 MAINTENANCE 1 After each use clean exterior of the product with a dry lint free cloth only 2 Store product in a clean area free from grease oil and other sources of contamination CAUTION DO NOT use cleaning solutions DO NOT immerse product in any kind of liquid All repairs MUST be done by Invacare Corporation Part No 1148114 19 HomeFill SECTION 6 CYLINDER FILL TIMES SECTION 6 CYLINDER FILL TIMES NOTE All filling times are approximate and may vary on environmental conditions Cylinder Concentrator Flow Rate to Patient Filling Platinum 5 up to 3 L min and Platinum 10 up to 5 Times L min HF2POST6 HF2POST9 HF2POSTD NORMAL hr 25 min 2 hr 20 min 3 hr 20 min MAXIMUM hr 50 min 2 hr 45 min 3 hr 50 min Troubleshooting If post valve cylinder fails to function try the following solutions If problem cannot be corrected consult your dealer home care equipment provider PROBLEM No flow POSSIBLE CAUSE Regulator or conserver is in Off position 2 Cylinder empty 3 To
13. ore attempting to use this equipment otherwise injury or damage may occur A DANGER This product is not intended as a life sustaining or life supporting device A WARNING Only personnel instructed and trained in the use of the HF2POST6 HF2POST9 and HF2POSTD Cylinder Assembly should operate this product Product contains magnetic ferrous material that may affect the results of an MRI Be sure all connections are tight and leak free DO NOT use a liquid leak detector to test for leaks DO NOT autoclave DO NOT gas sterilize with ethylene oxide DO NOT clean with aromatic hydrocarbons Part No 1148114 5 HomeFill SECTION GENERAL GUIDELINES A WARNING DO NOT attempt to repair product DO NOT immerse product in any kind of liquid Store product in a clean area when not in use Only use an Invacare Corporation HF2POST6BAG or HF2POST9BAG carrying bag NEVER smoke in an area where oxygen is being administered ALWAYS confirm prescribed dose before administering to patient and monitor on a frequent basis DO NOT use if dirt or contaminants are present on or around cylinder valve product or connecting devices NEVER use near any type of flame or flammable explosive substances vapors or atmosphere ALWAYS follow ANSI and CGA standards for Medical Gas Products and Flowmeters E 7 and Oxygen Handling G 4 The use of this device is limited to the oxygen patient Cylinders MUST be used only
14. rporation unless otherwise noted 2007 Invacare Corporation Part No 1148114 Rev B 9 07
15. s visible to accept the cylinder fillport PUSH Outer Ring Sleeve DOWN Connector Fillport GREEN Dots BEFORE amp COUPLING CYLINDER GREEN DOTS FIGURE 4 4 Resetting Connector Fillport 6 Grasp the cylinder assembly as shown in Detail A in FIGURE 4 5 7 Position the cylinder in the compressor cradle as shown in Detail C in FIGURE 4 5 8 Align the cylinder fillport with the connector fillport as shown in Detail C in FIGURE 4 5 9 Pull UP on the outer ring sleeve of the connector fillport Detail B while pushing DOWN on the cylinder assembly to couple the cylinder fillport into the connector fillport as shown in Detail B in FIGURE 4 5 NOTE The cylinder is properly connected when an audible click is heard 10 Press the compressor power switch to the On 1 position Part No 1148114 15 HomeFill SECTION 4 INSPECTION AND COMPRESSOR INTERFACE NOTE Refer to Cylinder Fill Times on page 20 for the length of time it will take to fill the cylinder DETAIL A DETAIL B Cylinder Assembly Cylinder Fillport Outer Ring Sleeve Compressor DETAIL Power Switch Cylinder Fillport Outer Ring Sleeve Compressor Cylinder Connector Cradle Fillport virg FIGURE 4 5 Connecting the Cylinder to the Compressor HomeFill 16 1148114 SECTION 4 INSPECTION AND COMPRESSOR INTERFACE 11 After
16. sity in oxygen enriched air Failure to observe this warning can result in severe fire property damage and cause physical injury or death NOTICE The information contained in this document is subject to change without notice HomeFill 2 Part No 1148114 TABLE OF CONTENTS TABLE OF CONTENTS SPECIAL NOTES 4 RECEIVING INSPECTION 4 INTENDED USE 4 SECTION I GENERAL GUIDELINES 5 SECTION 2 FEATURES 8 SECTION 3 TYPICAL PRODUCT PARAMETERS 9 SECTION 4 INSPECTION AND COMPRESSOR Cylinder Prefill Inspection 0 9 10 Filling the Cylinder gt 13 SECTION 5 MAINTENANTCE 19 SECTION 6 CYLINDER FILL TIMES 20 u cerei 20 Replacement Parts 20 LIMITED WARRANTY 22 Part No 1148114 3 HomeFill SPECIAL NOTES SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage Refer to the following table for definitions of the signal words SIGNAL WORD MEANING DANGER D
17. to the compressor Refer to Filling the Cylinder on page 13 e Ifthe cylinder pressure is greater than 1500 psig DO NOT attempt to fill this cylinder It may not fill Use cylinder as normal Filling the Cylinder NOTE For this procedure refer to FIGURE 4 2 thru FIGURE 4 7 1 Check the cylinder pressure before attempting to fill cylinder Refer to Checking Cylinder Pressure on page 12 2 Close the toggle valve by rotating clockwise FIGURE 4 2 3 Remove the conserver or regulator according to the device manufacturer instructions FIGURE 4 2 Part No 1148114 13 HomeFill SECTION 4 INSPECTION AND COMPRESSOR INTERFACE CONSERVER REGULATOR CONSERVER REGULATOR INSTALLED REMOVED Gauge Toggle Valve Conserver Regulator Cylinder FIGURE 4 2 Component Identification 4 Remove the post valve cover located on the cylinder and the connector fillport covers located on the compressor FIGURE 4 3 Connector Post Valve Cover Fillport Cover FIGURE 4 3 Fillport Covers 5 Onthe compressor unit momentarily push DOWN on the outer ring sleeve of the connector fillport to reset the connector FIGURE 4 4 HomeFill 14 Part No 1148114 SECTION 4 INSPECTION AND COMPRESSOR INTERFACE NOTE If the outer ring sleeve is in the UP position GREEN dots not visible the connector fillport will not be able to accept the cylinder fillport Pushing DOWN momentarily will reset the connector fillport GREEN dot