User manual CuroCell® Nova

image

Contents

1. Care of sweden Supporting Life User manual 2015 11 17 CuroCell Nova Alternating pressure mattress Item number user manual 95 001014 9 22 L345 C1325 Ce Contents Important safety precautions Introduction Product description Assembly Installation Operation Cleaning and reconditioning Storage Maintenance Troubleshooting Technical description Spare parts Options and accessories Warranty Other information Own notes Oo oOo O A WwW 11 12 13 14 15 17 17 18 18 19 IMPORTANT SAFETY PRECAUTIONS Read all instructions prior to use Warning To minimize the risk of fire electric shocks or bodily injury 1 The product must only be used for its intended purpose as described in this manual The product may not be combined assembled or repaired with other parts e g pump and mattress accessories or spare parts other than those described in this manual or other documentation from Care of Sweden 2 Do not use the product in wet rooms or where there is a risk of the pump coming into contact with water or another liquid Never handle a pump that has come into contact with water liquid Pull the plug out of the electrical socket immediately and send the product to an authorized service technician for servicing 3 DO NOT USE in the vicinity of or in contact with fire sources hot surfaces such as fire burning cigarettes or hot lamps Although the product has underg
2. 1 Optionals PRIMA 6 An elevating pad that is placed underneath a mattress overlay for example CuroCell Nova air mattress 11 2 Accessories Art no Description CC 9091 Sturdy transport bag 11 3 Other accessories To ensure the best level of comfort we recommend use of accessories from Care of Sweden since our products are designed to work well together Please contact Care of Sweden or your local dealer for accessories or more information on our products 17 12 Warranty 12 1 Scope This CuroCell system is covered by a 2 year warranty for manufacturing defects The warranty does not apply to normal wear and tear or product impairment or damage stemming from negligence or improper handling care 12 2 Service and repairs This CuroCell mattress is a medical device in accordance with MDD 93 42 EEC that is used as an aid in preventing and treating pressure ulcers For this reason we recommend that the pump be regularly serviced and inspected to maintain functionality and performance Like other technical devices a pump that is properly cared for works better and lasts longer Service and maintenance must always be performed by Care of Sweden or one of its authorised technicians Only use spare parts that have been approved by Care of Sweden Use of non approved spare parts invalidates the warranty A warranty claim must be submitted to Care of Sweden before the product is returned Please contact Care of Sweden or y
3. 9 2 Labelling The mattresses are labelled with production date article number size care and wash instructions EAN bar code as well as space for own labelling The product also displays which side is up down and head foot 15 9 3 Symbol Definitions a RE Article No a Read the user manual Not in domestic refuse follow recycling instructions Counteract shear force Machine wash at 95 C m Class Il equipment Manufacturer _ GRO anna ARo 7 O 4nna Group of function Recommended user weight K Heel function gt lt 1 Chlorine X Do not iron 4 3 gt 3 gt gt gJ XXXX XX XX Year Month Day CE CE marked acc to MDD 93 42 EEC Category Patient information category Do not wash Tumble dry Ry Do not dry clean Foot placement ey Y ue Recycling Place on top of existing mattress Machine wash at 70 C Rol Do not tumble dry Temperature range Double Insulated Ingress Protection Rating ES The mattress should only be used lying in the lengthwise direction on the mattress 16 10 Spare parts Art no Description CC 9025 Hygienic cover Olivia CC 9022 Hygienic cover Steel CC 8217 Fuse set of 2 CC 8202 Filter If other spare parts are needed please contact Care of Sweden or your local dealer for further assistance 11 Options and accessories 11
4. the pump is in the OFF position Connect the plug to an approved electrical socket 220 V 4 Operation Note Always read the user manual prior to use 4 1 Start up 1 Start the pump by pressing 1 On 2 Air starts to fill up the mattress It takes about 20 minutes to inflate the mattress 3 If necessary the patient can lie on the mattress even when not inflated in this case the underlay will carry the load When using a mattress with an air underlay this must be inflated Note We recommend that the comfort dial be set to max when inflating a deflated mattress 4 2 Choice of programs There are two different programs see section 2 3 for details We recommend alternating mode by default 4 3 Comfort setting U alil an 2 The air pressure in the mattress can be adjusted 0 40 1 2 according to the patient s desire for comfort Always 90 60 2 3 consult the prescriber or other authorised personnel for 70 80 3 4 appropriate setting 90 100 4 5 1 1 Use the comfort dial to set a suitable air pressure B n 130 140 6 7 based on the table on the pump 150 7 8 2 When the lamp the symbol A lit the mattress is ready for use www careofsweden se NOTE e f the head of the bed is being raised pressure must be increased at least one 1 level e When only parts of the mattress are under load e g if the patient is an amputee the pressure should be increased 1 or 2 levels e Hand c
5. f Sweden AB 2015 20
6. g time Alternating pressure is when the pressure regularly alternates between the mattress cells to replicate body movement Variable pressure cycles are set up where there is little or no pressure This is optimal for skin improvement or to prevent skin damage Always read the user manual prior to use 1 1 General information This CuroCell mattress is a medical device in accordance to MDD 93 42 EEC According to statutory regulations made by the authorietes regarding medical devices the manufacturer is required to report all accidents or incidents involving there products We would be very grateful for all information involving accidents or incidents relating to our products by reporting them immediately to us here at Care of Sweden The system has been tested and approved according to the following standards IEC 600601 1 IEC 60601 1 2 SS EN ISO 12182 SS EN ISO 14971 SS EN ISO 10993 EN 597 1 EN 597 2 EN 14126 1 2 Intended use CuroCell Nova is an alternating pressure mattress system that can be used in all kinds of caring environments to prevent pressure ulcers as well as treatment up to and including category 2 CuroCell Nova is a mattress overlay system to be placed on top of an existing mattress We recommend that position changes are made regularly The time interval must be evaluated by the responsibel personnel depending on the patient s status diagnosis and general condition The mattress is meant for use for persons
7. hat the mattress cover is intact and that the cover and air cells are correctly assembled 3 Check that the air flow alternates between each outlet on the side of the pump 4 Check for damaged or pinched tubes and connectors Contact the manufacturer or your local dealer for spare parts 7 2 Changing fuses 1 Pull the plug out of the electrical socket before replacing fuses 2 Remove the cover for the fuses with a screwdriver Pull out the fuse holder 3 Fit a new fuse of the same type T1AL 250 V and screw the protective cover back in place There are two fuses 1 Loosen the small round cover plate at the back of the pump 2 Lean the pump back and the filter slides out of its holder 3 Place the new filter in the holder and put back the cover plate O The filter should be cleaned regularly if the pump is used ina very dirty environment 13 8 Troubleshooting Problem Solution Pump does not start Alarms for low pressure light mattress loses air Patient is bottoming Mattress moves around Some cells have less air Pump is noisy Check that the power cord is connected to the power socket Check the fuses Check that the air tubes are connected to the pump Check that the CPR valves are closed Open the mattress and look for any leakage Comfort knob may be set up Adjust the pressure slightly falls through and wait a few minutes Check that the CPR valve is closed C
8. heck for leaks from air cells or tubes Check again with the hand check see section 4 3 Check that the air mattress is secured to the underlying mattress with the elastic straps This is normal since the air flow alternates between every other cell for a set cycle time one cycle 10 minutes Check how the pump is hung on the bed Resonance can occur in vibrations are felt parts of the bed Remove the pump and listen to see whether this makes a difference Can be remedied by putting the pump on a flat steady surface or by placing a towel between the pump and bed If your questions are not answered by the above information please contact Care of Sweden or your local dealer for further assistance 14 9 Technical description 9 1 Technical specification PUMP SPECIFICATIONS CuroCell Nova 220 V 240 V 50 60 Hz 0 05 A ee 25 x 13 5 x 9 5 cm Type of fuse po aao Storage 15 C 50 C Shipping 15 C 70 C peration 10 90 non condensing Storage 10 90 non condensing umidi Atmospheric pressure OOhPa 1013 25hPa 7 Class Il Type BF IP2X Classification MATTRESS SPECIFICATION Model Ff Curae ll Nova Dimension LxWxH Po 85x200x13 cm Weight Pd e NOTE Care of Sweden retains the right to modify the product specification at any time a WARNING Follow operating and maintenance instructions carefully to prevent injury as well as damage to the product or other equipment
9. heck the pressure at all times Please refer to section 4 4 e Recommended user weight up to 150 kg 4 4 Hand check u Hand check is used to ensure that the mattress is s properly inflated This should be done regularly once per shift after changing the position in the bed or changing the comfort settings SF Place a flat hand under the air cells directly underneath the user s pelvic region and make sure that the user does not lay directly on the underlying mattress There should be about 1 2 fingers margin If you can feel the user s body resting on your hand you need to increase the comfort settings to next highest step Wait 10 minutes and perform a new hand check 4 5 CPR In case of an emergency where CPR cardiopulmonary resuscitation is necessary Rapidly pull the CPR band Remove the tube connections on the side of the pump for even faster deflation NOTE Make sure that the CPR connector is properly connected after use 4 6 Low pressure At an abnormally low air pressure in the mattress the LED lights on the symbol rae The lamp goes out when the pressure returned to normal Please refer to chapter 8 for troubleshooting WARNING If the air pressure remains low check the mattress and tubes for damage Repair as necessary or contact Care of Sweden or one of its authorised repair technicians z 3 10 5 Cleaning and reconditioning It is important to follow the instructions below before us
10. ing the product for another user Consult the hygiene responsible or the supplier for help and instructions when unsure 5 1 Cleaning PUMP WARNING Do not open the pump housing risk of electric shock Service and maintenance must be performed by authorised technicians Do not get the pump wet or submerge it in any liquid Wipe of the pump with a damp cloth and a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 If using a different agent choose one that does not affect the plastic of the pump housing MATTRESS COVER Covers can be wiped off with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 Excrements and blood spots must be removed immediately with cold water Carefully follow local instructions and usage instructions for the cleaner 357 Q Ul Bet R Multi layer covers should be divided before washing Wash with like colors Chlorine and phenol based cleaners could have a negative impact on the PU surface TTA and should be avoided If chlorine is used we recommend a mixture of max 1 INNER COVER AND DISTANCE FABRIC accessories oy Q Ul Bx K Clean the affected area with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 11 5 2 Recond
11. itioning PUMP See section 5 1 for instructions MATTRESS 1 Disconnect the tubes and let the air out also open the CPR valve 2 Remove the cover See instructions in chapter 5 1 for cleaning covers If the product is equipped with a foam core this should be removed and cleaned separately 3 Move the mattress to a clean area which is suitable for cleaning 4 Wipe off the cells all tubes and the CPR valve with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 5 Wipe off and disinfect all the work surfaces with an alcohol based cleaning agent intended for this purpose 6 Put the mattress together by putting back the cell package in the cover and place the pump tubes by the hole on the short side 7 Fasten the cell package in its holder in all corners Adjust and close the zipper Poe OC Re Cee RE CCC CR RE CCE Or rere Cre Ce ere on Crore eC ere NOTE Check the hygiene cover each time the product is cleaned If damaged it must be replaced or repaired Also check the pump tube connectors and power cord during cleaning Damaged parts must be replaced or repaired 6 Storage The mattress and pump are best stored in the product bag optional extra Handle the packed product with care Do not put anything heavy on top otf it 12 7 Maintenance 7 1 General 1 Check the power cord and plug for damage and scratches 2 Check t
12. liance Prevention and Treatment of Pressure Ulcers Quick Reference Guide Emily Haesler Ed Cambridge Media Osborne Park Western Australia 2014 1 3 Other information Action plan Pressure ulcers To achieve the best possible result for this product it is recommended that a well planned and well structured working environment be used In our guideline Action plan you will find examples which can be applied and put into practice when using our products Mattress guide This practical guide gives you as a care giver help in choosing the correct mattress from the Care of Swedens range CuroCell Nova is a mattress in function group C This information can be found on our website www careofsweden com or you can order a copy from our Customer services by sending your details to info careofsweden se 2 Product description 2 1 Pump and mattress 1 Mattress 2 CPR rapid deflation 3 Tubes 4 Pump connector 5 Pump Air cells 2 2 Pump Heel function 1 Connection tubes 2 On Off 3 Power cord 4 Fuses 2 5 Suspension hook 6 Filter 2 3 Pump functions control panel CUD or Ooo G A 00000 0o00 Button for switching between static and alternating function Oo00D Mode 1 The pressure change between cells in a cycle of ten minutes We recommend that this mode is used by default 00000 Mode 0 Same pressure in all cells In static mode the alternatating cycle is discontinued and the
13. of a recommended minimum length of 120 cm The recommended user weight is up to 150 kg The mattress should only be used lying in the lengthwise direction on the mattress with the feet at the end marked with the feet symbol The mattress must always be cleaned between patients NOTE e For certain patients such as amputees are the recommended minimum weight not valid These users may however have other requirements as the entire mattress is not being loaded Please refer to sections 4 2 4 3 and 4 4 for operating control e When used together with positioning pillows read the user s manual carefully to ensure correct positioning of the patient e f you are using this product with an evacuation sheet it is the user s responsibility to ensure that it is safe to use for evacuation e Be careful when using bed rails or other protection on the bed so that the mattress does not become trapped or damaged e The mattress must not be lifted with a person lying on it or used for any other type of transport other than a possible evacuation using an evacuation sheet e This mattress is unsuitable for use during x ray examinations because of the risk of blurred images or artefacts that may lead to diagnostic errors Lee eee ee eee eee eee eee ee eee eee eee eee eee eee eee eee eee eee ee eee ee eee eee eee eee eee ee 1 National Pressure Ulcer Advisory Panel European Pressure Ulcer Advisory Panel and Pan Pacific Pressure Injury Al
14. one fire testing it could become damaged if it comes into contact with fire sources 4 Position and use the product in a manner that prevents it from becoming trapped or damaged NOTE Be particularly aware of trapping damage when using bed rails 5 Never use the product if the pump power cord or plug is defective the pump casing is damaged or it is not functioning properly Contact a service technician for examination and repair 6 Regularly check product functionality through a hand check see section 4 4 7 Continual monitoring is required when the product is used for individuals needing special supervision such as children 8 The mattress are protected with a hygiene cover Please try to avoid the simultaneous use of multiple hygiene cover with this mattress Because of the risk vapour permeable can be destroyed 9 The hygiene cover does not allow liquid or air to penetrate but is vapor permeable Make sure that the patient is positioned correctly to avoid the risk of suffocation 10 Prevent damage to the hygiene cover from sharp objects 11 If the cover are supplied with handles on the long sides theses handles are only to be used for lifting and moving the mattress itself All other use will be at your own responsibility and is not covered by the product warranty 1 Introduction CuroCell air mattresses operate according to the alternating pressure principle a technique that is well proven and been used for a lon
15. our local dealer if you have questions regarding warranty and service 12 3 Life expectance This product s expected life is estimated to be 3 years 13 Other information 13 1 Disassembly and recycling Except for certain parts of the pumps energy recovery is possible for almost all material in CuroCell products through incineration in waste incineration facilities Pump A used CuroCell pump must not be disassembled It must instead be taken to a recycling centre and sorted as electronic waste Mattress A used CuroCell mattress must be taken to a recycling centre and sorted as combustible waste For more information please contact Care of Sweden or your local dealer 13 2 Returns and complaints Complaints are to be reported to the supplier Return of a product shall follow the supplier s instructions and can only be done after the complaint has been reported Damages due to the transportation should be reported to the transportation agent 18 14 Own notes 19 abi Manufactured by Care of sweden Supporting Life Contact Phone 46 771106600 Fax 46 325 12840 Email info careofsweden se Internet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns vag 4A SE 514 33 Tranemo SWEDEN Supplied by Care of Sweden and CuroCell are trademarks of Care of Sweden AB Care o
16. pressure is equalized over the cells This function makes all patient movements as well as nursing and care of the patient easier 4 Power off on 1 0 am alil Suitable air pressure is set with the comfort dial 1 8 Guidelines based on user weight is given in the table on the pump NOTE In all uses where the head end of the bed is raised the pressure must be increased at least one 1 step Lights when the mattress is ready for use Alarms for low pressure In order to solve it see chapter 8 troubleshooting 3 Assembly Installation When unpacking the mattress check that none of the parts are damaged In the event of damage contact the manufacturer or your local dealer before using the product 1 Place the mattress on the existing mattress and fasten it with the elastic bands around the underlying mattress Note e Make the bed with sheets for added comfort e Check that the CPR valve is closed 2 Hang the pump at the foot end of the bed or place it on a flat sturdy surface Route the power cord so that there is no risk of it causing someone to trip being run over by the bed wheels or getting crushed when the bed is raised or lowered Note Make sure that the CPR valve is closed 3 Connect the mattress air hoses to the side of the pump If the pump and hoses are equipped with quick connectors an audible click will be heard when these have been connected correctly 4 Check that the switch on