Rayocomp PS 10
Contents
1. 3 6 Understanding the device Page 7 3 6 Understanding the Device In the chapters of this user manual you will often find instructions referring to keys or segments of the Rayocomp PS 10 In the following you can see pictures of the device accompanied by short descriptions in order to simplify the assignment 3 6 1 Front View Front view of the Rayocomp PS 10 Caption 1 Display 2 Keyboard 3 Three exit hubs for connecting the hand foot or Velcro detectors and the measuring and protection cups 4 Sensor connection For a correct connection please insert the sensor plug and then fasten it by twisting the knurled screw 5 Entry hub 6 Memory card 7 Memory card insert 3 6 2 Rear View Page 8 3 6 2 Rear View Rear view of the Rayocomp PS 10 Caption 1 ON OFF switch 2 Charging socket 3 Charging plug which is inserted in the charging socket for the charging process 4 Interface for program updates 5 Sticker indicating the serial number year of construction etc 3 6 3 Keyboard Page 9 3 6 3 Keyboard Ce DOR Keyboard of the Rayocomp PS 10 Caption Mode key for selecting the menu item IF1 IF3 keys for determining frequently used fundamental frequency values 10 0 key for the rapid access to this special value and for activating the M10 test protocol Key for loading the values from a memory card into the device Key for storing values from the device on a memory card
2. OI 2010 Setup precision of 0 05 54 integrated harmonising programs 14 Acupuncture oscillation therapy 22 Programmes according to Dr Ulrich 9 special programmes 06 2070 11 Test programmes 2080 Rayonex Analysis and Harmonising programme RAH Module 8 om 2090 Rayonex Analysis and Harmonising programme RAH Module 9 ele A RAH Module 10 Let me mention a detail concerning module 1 the standard step length precision of the device is 0 25 for instancel2 5 12 75 etc If the module 1 is activated the step lengths between the fundamental frequency values can be adjusted to the higher precision of 0 05 i e 12 5 12 55 etc 4 7 Menu Item into Page 60 10 Here the current service state of the device is indicated After approx 6000 operation hours or when your device hasn t been in service for about three years it will indicate 100 service is due The device should be sent to us for service then Here the currently set polariser mode is indicated This is no special information the polariser setup is visible in other menus too But for the sake of completeness this is also displayed here This number informs you about the hardware version of the device In this case it is version 02 Devices including the version 02 have already been prepared in such a way that the RAH module can be activated by means of an activating code Devices with the version 01 would first have to be modified To this p
3. Start key for starting F1 programs for example Stop key for interrupting an automatic run or for deleting the function N key of the polariser to switch the polariser to neutral key of the polariser to switch on the bipolar function key of the polariser to activate a polarisation gyrating to the right key of the polariser to activate a polarisation gyrating to the left Arrow left key i e for selecting the previous value during an automatic program course Arrow right key i e for selecting the following value during an automatic program course Numerical block for entering digits numbers fundamental frequency values C deletes the decimal point indicates a precise value i e 12 75 3 6 4 Displa Page 10 3 6 4 Display oe ze OO Sc 88 88 5688 58 58 42 Display of the Rayocomp PS 10 Caption 1 2 Numbering the modules All in all the Rayocomp PS 10 is intended for the activation of 12 different modules If you select the menu item Info the modules already activated in the device are displayed 3 Here the current service state of the device is indicated After approx 6000 operation hours or when your device hasn t been in service for about three years it will indicate 100 service is due 4 In the menu item EI you can read the remaining total harmonising time here 5 This displays the polarisation type which is currently set 6 Here you can read which menu item
4. 5 User Manual 6 Transport case 7 Cardboard packing box It is absolutely necessary to keep the packing in case you have to return the device for maintenance e g for servicing please send it back in its transport case and the cardboard box We cannot take over the guarantee for insufficient packing If the old packing material gets lost we can replace it on request liable to costs When sending back the device please make sure that all additional parts and accessories are secured inside the case In case you intend to send back the PS10 please always include the appurtenant charging set 3 2 Technical Data Weight of the Rayocomp PS 10 1 6 kg Weight including case and accessories 9 7 kg Height 100 mm Width 160 mm Depth 260 mm Power supply for charging 100 240 V alternating current 17V 8 4 V secondary direct current Max reception capacity for charging 24 W Operation time with a fully charged power pack 8 hours at least Range of the operation temperature IS C 35 C The device should not be exposed to great thermal fluctuations otherwise you run the risk of condensation build up inside If the instrument has cooled down i e during transportation to very low temperatures below 5 C you have to make sure it is sufficiently acclimatised within the working temperature range mentioned above before switching it on 3 Basics of the Rayocomp PS 10 Page 6 3 3 Safety Instructions as to Oper
5. Background illumination is permanently switched off 1 Background illumination is permanently switched on 2 Background illumination is set to automatic 30 seconds after the last keyboard action the background illumination is switched off The background illumination turns on again as soon as the keyboard is used This saves energy the battery pack lasts longer With this setup the background illumination is switched off at night if you are harmonising while you are resting Your sleep is not disturbed by the bright light of the illumination in the background On delivery the standard setup of the Rayocomp PS 10 is number 2 4 5 2 Deleting Double Values in the Fl Programs Setup Number 11 If you are arranging different BnPS programs it is inevitable that several of the fundamental frequency values are repeated If you prefer not to harmonise fundamental frequency values twice you can set this up here This is also valid for RAH programm compilation Here double entered programmes are disregarded and are not considered multiply during harmonisation While testing the RAH programmes double values e g from the memory card are always rejected and are only selected once Enter the setup number 11 and you can add one of the following numbers 0 Do not delete double values when loading the Fl programmes respectively RAH programmes 1 Delete double values when loading the Fl programmes respectively RAH programmes On delive
6. deleting double values is also valid for double RAH programmes e Please regard The harmonising program groups listed above must be enabled in other words the corresponding modules must be activated in the Rayocomp PS 10 More information is available in chapter 4 7 4 3 2 Composing and Harmonising Various Harmonising Programs Page 27 The following example will show you how to combine the harmonising program 1 51 pre control with the harmonising program 1 52 immune system in F1 and then harmonise Activate the key for MODE until the menu item FIT appears Rayonex Rayocomp PS 10 cy j LN r Now you can enter the first number of the harmonising program 1 51 Press the following keys HBD The display shows Rayonex Rayocomp PS 10 P d E l Gp Note Here you can see the number of funda 7 mental frequency values already stored in this N case there are 77 After this the Rayocomp PS 10 is ready for the entry of the next harmonising program in this example it is the program 1 52 Press the following keys 4 3 2 Composing and Harmonising Various Harmonising Programs Page 28 The display shows Rayonex Rayocomp PS 10 CH j cus Now that the two desired harmonising programs have been entered please press the key and you can enter the harmonising time The Rayocomp PS 10 suggests a standard of 30 seconds But the time can also be altered individually
7. you can look into the list of test compositions 4 4 Menu Item Test Page 36 POI Testwith range values POG Check uptest Z Acid base t P10 Quick test of the 14 channels P11 Vital substance test If you enter 01 for example you can carry out a range value test with 02 the fundamental frequency values of a memory card can be read into the device for subsequent testing With 03 you can test all programs integrated in your device as a module The test programs 04 to 11 are special compilations of fundamental frequency values You can use them if module 6 test programs is activated in your Rayocomp PS 10 Important information The fundamental frequency values tested with resonance can be saved on an inserted memory card at any time by pressing the key You can read in the fundamental frequency values from a memory card test them and save the so determined resonances on the same card again for instance This makes sense for subsequent testing of the resonances of a range value test Please keep in mind that you overwrite all values previously stored on the memory card when you press the key So you see that the program number determines the fundamental frequency values which are going to be tested As soon as the values intended for testing are available the test run is always the same 4 4 Menu Item Test Page 37 You will see the following picture on your display if for exa
8. Rayocomp PS 10 with this setup function If your Rayocomp has been subject to jostling and vibration perhaps during long transportation you can make sure that the frequency module is still in perfect working order with this function After you have entered the setup number 50 and confirmed your entry with the OK key the Rayocomp PS 10 starts a trial run You can read the progress of the trial run from your display At the end the device should display the following message Rayonex Rayocomp PS 10 EE i el LI mine LI 0 The 1 displayed behind the 50 indicates an optimal functioning of the frequency module If you see a O in the same place however the trial run was faulty Please repeat the trial run in this case If the O appears again your Rayocomp PS 10 requires servicing and has to be sent back to us 4 6 Menu item Info Page 59 4 6 Menu Item into In the menu item Info you can look into all important details concerning the device quickly and conveniently You arrive in the menu item Info by pressing the key MODE several times until the display below appears Ga Rayonex Rayocomp PS 10 4 a gt lt a 3 a ot l l Ce Jl G Caption 1 At present the Rayocomp PS 10 has ten modules You may recognise which of the modules are activated in the device by verifying whether a checkmark is shown following the number 1 to 10 Here is a list of the modules Module
9. Upper respiratory system After you have confirmed the entry with OK the total time changes The total time is the sum of harmonising times for the already selected programs and the currently selected program Upper respiratory system 4 3 5 Utilising the Harmonising Programs of the new RAH Page 33 Rayonex Rayocomp PS 10 RI Ta UI TU C D 00 38 00 If you intend to harmonise program 70 16 for 15 minutes too confirm by pressing the OK button Rayonex Rayocomp PS 10 Now that all programs have been selected harmonising can commence Harmonising begins as soon as you have actuated the start button This triggers the first program The percent display shows the progress After 100 is achieved the polariser switches over to and the timer for harmonising starts counting backward The countdown of the total harmonising time proceeds in steps of one second while the single programs are counted backward at minute intervals The display shows the program number When ever you apply the RAH module please make sure that no card is inserted in the card slot Should this ever happen the user is informed to remove the card with error message Er 40 This applies to all kinds of memory cards 4 3 5 Utilising the Harmonising Programs of the new RAH Page 34 Pause Function and Terminating the Harmonising The harmonising time is interrupted when you press the stop button once The start button restarts th
10. in Rayocomp Im PS 10 please press the d CT CT cnn e until you see under the menu item test the display shown here on the left The display shows rAH The user is then informed that he can test RAH programmes 4 4 5 RAH Module 10 Level test Page 41 After a few seconds the display changes and you can enter a RAH programme number Please press now the following keys The Rayocomp PS 10 now w adjusts the RAH programme number 04 00 The display shows The RAH programme number 04 00 1 Www gf e fig 2 is a main RA programme 2 j IR f II j Therefore the figure 1 shows Level 1 3 R l jt If like in this example level 2 and 3 also appear the user knows that there are respective programmes on sub and detail levels under the current selected programme On the highest RAH programme level 45 programmes are available fig 6 The programme 04 00 is the fourth fig 5 of 45 programmes of the highest level For the test the polariser of the Rayocomp PS 10 is automatically switched on N fig 3 There are currently no programmes selcted that have to be harmonised visible under fig 4 Now using the sensor the test can begin The functions are as follows 4 4 5 RAH Module 10 Level test Page 42 If the sensor button is pushed forward the second level of the device is selected fig 1 The sub programme 04 10 appears fig 2 As the third level is not shown you can see immediately that t
11. into 4 7 Special Function Keys 4 7 1 Using the F Keys 4 7 2 Using the 70 0 Key 5 Appendix 5 1 Summary of the Integrated Test Programs 5 2 Summary of the Integrated Rayonex Programs 5 3 Summary of the Integrated Acupuncture Programs 5 4 Summary of the Integrated Special Programs 5 5 Summary of the Integrated Programs by Dr Ulrich 5 7 Summary of the Modules for Integration 5 8 Summary of the Possible Error Messages 6 Private Documents 63 64 65 66 68 70 2 Conformity Declaration Page 4 EU Conformity Declaration Rayocomp PS 10 EC Conformity Declaration For the following medicinal product grade Article Rayocomp PS 10 Article number 2000 we declare that it fulfils the corresponding basic conditions appendix of the Medicinal Product Recommendation 93 42 EWG included changed EC Directive 2007 47 EG in the version valid on the date of delivery Valid harmonised standards were regarded and applied In responsibility for the manufacturer RAYONEX Schwingungstechnik GmbH Sauerland Pyramiden 1 D 57368 Lennestadt this declaration is made by Dipl Ing Dietmar Heimes Business Manager SiB MP i S Lennestadt 07 04 2010 D S VW Town Date Signature 3 Basics of the Rayocomp PS 10 Page 5 3 Basics of the Rayocomp PS 10 3 1 Volume of delivery The delivery volume consists of 1 The RAYOCOMP PS 10 2 Integrated polariser 3 Integrated power pack 4 Battery charger
12. item Direct provides the Rayocomp PS 10 with the functions of the former Rayometer But now with the new Rayocomp PS 10 it is no longer necessary to adjust the fundamental frequency value manually You simply enter the value with the help of the keyboard and confirm with the OK key The device adjusts the value itself Example setting the fundamental frequency value 56 00 After switching on your device you arrive in the menu item Direct The display shows Note Rayonex Rayocomp PS 10 geg mg eg Please note that entering e Zer ht a f j f f i f eet NE the be helpful wh bel e Se Ben entering one digit values U That s why the entry of N is interpret as fundemanetal frequency value of 6 00 Now press the following keys HOBO G The Rayocomp PS 10 sets the fundamental frequency value 56 00 and displays Rayonex Rayocomp PS 10 Beneath the fundamental frequency value you can see numerals indicating the running time here 1 minute and 12 seconds Now please work according to the explanations in chapter 4 2 1 to test the value with the polariser or to harmonise it A new fundamental frequency value can be set at any time 4 2 3 Automatic programme sequence when using an RAH Green Card Page 24 4 2 3 Automatic programme sequence by using a RAH Green Card In chapter 4 1 6 the advantages of an RAH Green Card have already been described Under the menu item D
13. the Rayocomp PS 10 as described in this chapter and in case there is any need for harmonisation can be stored You can either first choose a main level or start right in the middle If you had selected right in the beginning RAH number 04 21 the display would have shown you the same Very Important Please also read the chapter 4 4 8 Important tips and tricks for testing with the RAH Module 10 These listed additional handling advices will faciliate your use of the module 10 4 4 6 RAH Module 10 Test with integrated test protocols Page 44 4 4 6 RAH Module 10 Test with integrated test protocols The idea of the test protocols was born in the Heilpraktikerschule der Paul Schmidt Akademie There you can learn among other things which organ structures and regulation ranges you have to consider regarding the respective disease This is also valid for energetic testings by means of RAH For example in the case of hypertension you have to necessarily consider the kidney as it produces the enzyme renin which increases the blood pressure The hormonal system is important too as it has a big influence on the metabolism and the blood pressure The new analysis support considers exactly these coherences For example if you select the frequency structure of the RAH programme 39 60 hypertension all RAH programmes that are linked up with the disease pattern will be recommended to you directly for the testing at the push of o
14. with a long beep signal When an RAH Green Card is read in for testing the total resonance memory will be deleted automatically beforehand as the user would get confused otherwise The same is valid for the test with test protocols 4 4 8 Important tips and tricks for the test with the RAH Module 10 Page 48 5 Storing the necessary RAH programmes on a RAH Green Card During the RAH testing all necessary RAH programmes and their harmonisation times can be stored on an RAH Green Card at any time most reasonably at the end of the testing Therefore you have to insert an RAH Green Card and press the PS After pressing the key the number of RAH programmes and their harmonisation times indicated on the display fig 1 will be stored Storing the RAH programme numbers can also be performed under the menu item Fl This programme item is best suited for entering the individual RAH programme numbers and individual harmonisation times Please keep in mind Whenever RAH programmes are stored the RAH Green Card gets deleted and will then be loaded with the new RAH programme numbers Thus all RAH programm numbers stored on the RAH Green Card up to then will be deleted Furthermore RAH programmes can only be read in and stored on the RAH Green Card when Module M9 or Module M10 is avtivated 4 5 Menu Item Setup Page 49 4 5 Menu Item Setup The menu item Setup w
15. 6 Private Documents Page 70 6 Private Documents 6 Private Documents Page 71 6 Private Documents ARN NIE Rayonex Schwingungstechnik GmbH Sauerland Pyramiden 1 e 57368 Lennestadt Tel 49 2721 6006 0 e Fax 49 2721 6006 66 Internet www rayonex de e E mail info rayonex de FB 146 e Rev28 e Stand 15 September 2011
16. 7 The Rayocomp PS 10 changes to the menu item Setup and shows on the display Rayonex Rayocomp PS 10 Li eg Se Once again press the key 0 The Rayocomp PS 10 changes over to the menu item Info and displays Rayonex Rayocomp PS 10 1 it it ke 2 l di CG 3 L c s l 4 v no nn d No IT OU Uc If you press the key you will move back to the menu item Direct Once more the display will show Rayonex Rayocomp PS 10 wert LI 000 U ie 4 1 5 Operating the Integrated Polariser and Sensor Page 18 4 1 5 Operating the Integrated Polariser and Sensor Just like the Rayocomp PS 1000 polar the Rayocomp PS 10 has an integrated polariser to improve comfortable operation The functions of the integrated polariser can be activated in two different ways with the keyboard or the sensor Because the use of the keyboard is rather bothering when performing tests we recommend using the button of the sensor to carry out the functions of the integrated polariser right from the start Basically the use of the polariser in connection with the Raycocomp PS 10 posesses the following advantages e While testing the polariser helps you to find out whether a disturbance is strong or weak and shows you the optimal polarity e Specially when harmonising the bi polar function intensifies the effect These four keys enable the utilisation of the polariser functions via k
17. BnPS in the display Rayonex Rayocomp PS 10 Note The Analysis and I CITT Harmonising system RAH will CH CH on emm only work when there is no a an memory card in the memory card Slot before Starting the harmonising In case this is forgotten occasionally the user will be adverted by the error message Er 40 If the card is extracted then the transmission Starts automatically The submenu item BnPS is indicated by the prompt BnPS in the upper part of the display Now the mode key is actuated once more to select the RAH module The display will show the following image Rayonex Rayocomp PS 10 LI os CO rAH is the short index for RAH and means in this case the harmonisation with the Analysis and Harmonising system If the mode key is not pressed again the display changes over to the input mask of the RAH module 4 3 5 Utilising the Harmonising Programs of the new RAH Page 31 The illustration below shows the display of the input mask The program number is entered here Rayonex Rayocomp PS 10 The PS 10 is waiting for the entry of a program number now We choose the programs Vitalisation complete and Upper respiratory system in this case Looking at the respective lists you will see that the program Vitalisation complete has the number 01 00 and the program Upper respiratory system is assigned to the number 70 16 In order to determine the treat
18. Composing and Harmonising Various Harmonising Programs 4 3 3 Loading Fundamental Frequency Values from Memory Cards and Harmonising 4 3 4 Fundamental Frequency Values from the Menu Item Test Taking Over and Harmonising 4 3 5 Utilising the Harmonising Programs of the new RAH 2 4 VI CO wO DJs OF On On VI vu mm mess 22 23 24 25 25 26 29 29 30 1 Contents Page 3 4 4 Menu Item Test 4 4 1 Testing Range Values 4 4 2 Testing with Memory Card Values 4 4 3 Testing with Integrated Programs 4 4 4 Testing with Integrated Test Programs 4 4 5 RAH Module 10 Level test 4 4 6 RAH Module 10 Test with integrated test protocols 4 4 7 RAH Module 10 Test with the RAH Green Card 4 4 8 Important tips and tricks for testing with the RAH Module 10 4 5 Menu Item Setup 4 5 1 Set Up of Background Illumination 4 5 2 Deleting Double Values in the Fl Programs 4 5 3 Adjusting the Attention Sound Beep 4 5 4 Pre Setting the Test Range Values 4 5 5 Presetting the harmonisation time for Fundamental Frequency Values 4 5 6 Activating and Deactivating Modules 4 5 7 Storing Fundamental Frequency Values on Memory Cards 4 5 8 Displaying Memory Card Information 4 5 9 Providing Read Protection for a Memory Card 4 5 10 Providing Write Protection for a Memory Card 4 5 11 Deleting a Memory Card Completely 4 5 12 Initialising a Memory Card 4 5 13 Motor Test Run of the Frequency Module 4 6 Menu Item
19. Green Card In addition a special function was created under the main menu item of the Rayocomp PS 1000 polar and the Rayocomp PS 10 If you insert an RAH Green Card an automatic harmonisation sequence begins At first the single frequencies stored on the RAH Green Card are harmonised with the time according to the setup and after that the stored RAH programmes with their individual harmonising time Thus an RAH Green Card compiled by the therapist can easily be used by the patient at home After inserting the RAH Green Card in the menu Direct the harmonisation run starts automatically 4 1 7 Error Display Page 21 4 1 7 Error Display Operation of the Rayocomp PS 10 is very simple indeed Still you may accidently press a button which has no function for this particular part of the program In this case the user will receive an acoustic signal The Rayocomp PS 10 has three different error messages or acoustic signals 1 One long sound With this signal the device confirms a successfully completed action i e an entry or storage 2 Two short sounds The key you are pressing has no current function 3 Two long sounds This signals a severe error The device will supply an error code In the appendix chapter 5 8 you can look up the Current error For example if an error has been detected the device might indicate the following Rayonex Rayocomp PS 10 D OU LI LU If you
20. Rayocomp PS 10 Portable High Performance Bio Resonator User Manual Art Nr 91OOEN In spite of all the information advice and suggestions for solving problems this book cannot replace a visit to an alternative practitioner or a doctor orientated to natural healing Also we must point out that the classic orthodox medicine still does not accept or acknowledge the interrelations presented in this manual 1 Contents Page 2 l 2 Contents Conformity Declaration Basics of the Rayocomp PS 10 3 1 3 2 3 3 3 4 3 5 3 6 Volume of Delivery Technical Data Safety Instructions as to Operation Guarantee Cleaning Understanding the Device 3 6 1 Front View 3 6 2 Rear View 3 6 3 Keyboard 3 6 4 Display User Manual 4 1 4 2 4 3 General Directions for Operation 4 1 1 Literature Relating to the Bio Resonance According to Paul Schmidt 4 1 2 Charging the Integrated Power Pack 4 1 3 Switching the Device On and Off 4 1 4 Selecting the Menu with the Mode Key 4 1 5 Operating the Integrated Polariser and Sensor 4 1 6 Basic Information for RAH Green Card 4 1 7 Error Display Menu Item Direct 4 2 1 Testing and Harmonising a Fundamental Frequency Value with the Integrated Polariser and Sensor 4 2 2 Setting Up Individual Fundamental Frequency Values 4 2 3 Automatic programme sequence by using a RAH Green Card Menu Item EI 4 3 1 Basics Concerning the Use of Integrated Programs 4 3 2
21. am groups you can take them from the appendix chapters 5 2 to 5 5 Here is a basic summary 01 to 1 54 Rayonex harmonising programs 01 to 2 14 Harmonising programs of acupuncture oscillation medicine 01 to 3 09 Special programs 01 to 4 17 Harmonising programs according to Dr Ulrich The function F1 BnPS helps you to compile harmonising programs from the different groups For example you can combine the harmonising program 1 51 pre control with the program 3 02 scar interference suppression The Rayocomp PS 10 will then automatically set up the fundamental frequency values of the pre control first followed by the values for suppressing scar interference Before giving detailed examples for the entry let us summarise important information concerning the application of the F1 BnPS programs first e You can compile a total of up to 500 fundamental frequency values in F1 BnPS If the first program includes 77 fundamental frequency values and the second 126 values you have 203 values altogether Now there is space left for further 297 values e If you are arranging different programs it always happens that several fundamental frequency values are repeated If you generally want to avoid harmonising the same fundamental frequency values more than once this can be predetermined in the setup menu item 4 5 2 This may shorten harmonising time The original set up from works allows double harmonising values The set up
22. aranteed a blip can be heard several times In this case the power pack must be recharged immediately to avoid damage Important information You cannot damage your device by switching it off arbitrarily Neither will the integrated harmonising programs be lost nor will your hardware suffer damage 4 1 4 Selecting the Menu with the Mode Key Page 14 4 1 4 Selecting the Menu with the Mode Key The Rayocomp PS 10 has a selection of menu items Every menu item provides certain application possibilities Therefore partitioning into single menu items serves as a means for simplifying operation The following menu items are available 1 Menu Item Direct 2 Menu Item EI 3 Menu Item fest 4 Menu Item Setup 5 Menu Item into Use this menu item for the rapid output of single fundamental frequency values You could say this menu item covers the functions of the old Rayometer As a matter of principle the Menu Item F1 includes two separated programmes On the one hand the BnPS programme which is indicated by Bnb on the display After a short time the representation will switch over to P_ _ By means of the BnPS programmes you can automatically run the harmonising programmes installed in the device or prepared by yourself on the memory card provided the modules have already been activated in the device The RAH programme which is also found under the Menu Item F1 is shown only if it has eff
23. as designed to adapt the Rayocomp PS 10 to your individual intentions by means of presettings Available besides the alteration of presets are test functions and important and useful functions in connection with memory cards Select the menu item Setup with the key MODE The display shows Rayonex Rayocomp PS 10 TTT U At this stage the device is waiting for a setup number to address the desired setup function For exmple if you want to enter the setup number 10 please press 1 then O and confirm your entry with the OK key Further actions are described in the following subchapters To simplify orientation the layout of the setup numbers is large and marked in red in the corresponding subchapters Please note You run no risk of ruining the device or deleting anything permanently if you alter the setup Every setup of the device can be returned to its original state Have courage and try it out It is the best way to become familiar with your Rayocomp PS 10 4 5 1 Set Up of Background Illumination Page 50 4 5 1 Set Up of Background Illumination Setup Number 10 For reasons of improved legibility the display of the Rayocomp PS 10 was provided with background illumination Because of this feature the display remains perfectly visible even if the light is dim You can select one of three different types of background illuminations Enter the setup number 10 and you can add one of the following numbers 0
24. ation The Rayocomp PS 10 is charged with a direct current The charging rate is 800mA Please note that the Rayocomp PS 10 may only be used witha battery charger supplied by Rayonex e Therefore make sure that no humidity can penetrate the power pack short circuit risk e Prevent any persons i e children from poking sharp metal items into the device power stroke risk e Make sure that the connection cables belonging to the delivery volume of the detectors are only used for connecting the device with the detectors e Use the Rayocomp PS 10 in power pack mode only and never with the charging cable plugged in When using the detectors especially the pointed detector for acupuncture points be sure to handle them with care risk of eye injury 3 4 Guarantee The warranty period runs for 1 year from the date of delivery on and is only valid for instruments that have not been opened with the intact guarantee seal as evidence Improper use of the device invalidates the warranty claim In conformity to general conditions for inspection service is liable to costs even if this becomes due within the warranty period 3 5 Cleaning The surface of the device should be cleaned with a soft cloth and a small amount of alcohol once a week Never treat the device with strong cleaning agents The detectors which come into close contact with the patients i e hand and foot detectors need cleaning after every use preferably with alcohol
25. command with the OK key 4 5 7 Storing Fundamental Frequency Values on Memory Cards Page 54 4 5 7 Storing Fundamental Frequency Values on Memory Cards Setup Number 30 A great advantage for anyone working with the Rayocomp PS 10 is the use of memory cards They allow you to store individual frequency compilations which can be used for testing in the menu Test or for harmonising in the F1 BnPS menu Generate your own memory cards with this setup function The frequency compilations can be stored on a normal memory card or a Green Card Insert the memory card in your device and enter the setup number 30 Every entered fundamental frequency number will now be stored on your memory card as soon as you have confirmed it with the OK key The display shows Rayonex Rayocomp PS 10 The red arrow points to the numeral which indicates the number of the currently stored values on your memory card With the key 3 you can delete the last value entered on the card Should the card contain left over values which you want to delete please use the setup number 43 see also chapter 4 5 10 to delete memory card completely 4 5 8 Displaying Memory Card Information Page 55 4 5 8 Displaying Memory Card Information Setup Number 40 This setup can help you if you have an unknown memory card and have no idea how many values are stored on it what kind of a memory card it is and SO on As a Green Card can comp
26. damental frequency values on the physical level as well as pathological fundamental frequency values of the corresponding organ These are concluded by a value 21 50 to stabilise the body polarities In practice the course of the use of the integrated harmonising programmes will be that based upon a diagnosis made e g based upon the results of a check up test see Chapter 4 4 4 the organs that are the most affected ones can be harmonised by means of the integrated programme Sure enough you can use the classical therapy programs of the bio resonance according to Paul Schmidt with the Rayocomp PS 10 devices but now we have a brand new addition the harmonising programs of the new Rayonex Analysis and Harmonising System RAH in the menu item F1 This requires a Special hardware in your device If your Rayocomp PS 10 bears the inscription Fvolution on the side this new hardware is already integrated in the device For a detailed description of the utilisation of the new RAH system look into chapter 4 3 5 4 3 2 Composing and Harmonising Various Harmonising Programs Page 26 4 3 2 Composing and Harmonising Various Harmonising Programs The menu item F1 BnPS provides a very efficient function for the Rayocomp PS 10 enabling the composition of harmonising programs integrated in the device and harmonising their fundamental frequency values with these compositions The Rayocomp PS 10 offers a variety of different harmonising progr
27. dule enabling code 18 Final value must be larger than initial value KIE Ne e P ee ae ei If this is a frequently occurring malfunction the device must be sent in If this is a frequently occurring malfunction the device must be sent in 21 It is not possible to adjust the zero position correctly If this is a frequently occurring malfunction the device must be sent in Card is write protected Module is not activated Number of selected resonance values is too large for the card It is not possible to provide read and write protection for this card Maximum capacity of the memory full This module has already been enabled temporarily No Green Card inserted The Polariser does not switch over to the selected mode If this is a frequently occurring malfunction the device must be sent in More than 200 RAH programmes were selected Double value Double value S 5 7 Summary of the Possible Error Messages Page 69 No Description _ 86 A maximum of 200 RAH programs can be loaded successively 89 90 No values on the memory card No valid memory card inserted Please check or if necessary reinitialise the card see Setup Item 40 or 44 respectively More than 500 values selected mt This Fl program is non existent see program list This module cannot be activated by the present hardware version In 92 case you are interested in the RAH Module please contact the Rayonex company Program error
28. e Rayocomp PS 10 posseses an integrated battery pack to enable the application of the device without needing to connect it to the mains supply As a matter of principle the Rayocomp PS 10 should never be used while itis still being charged The Rayocomp PS 10 is charged in the quickest way when switched off An empty battery pack requires approx 4 hours for complete charging In order to avoid unnecessary electro magnetic pollution please unplug the charging unit when it is not being used And this is how you charge the Rayocomp PS 10 Switch off the device 2 Connect the charging unit to the mains supply 3 Plug the charging cable into the rear of the device see chapter 3 6 2 4 The charging unit indicates the various charging processes When the LED turns green the charging process is complete and you can use the device again 5 Please disconnect both the plug from the rear panel of the device and the charging set from the electrical outlet Function CSCC ello green Standby charging device in stand by mode ready to charge SE blinks Pre charge tests whether charging is required slowly a blinks Rapid charge power pack is charged with constant current rapidly Maintain charging power is reduced Le slowly blinks Error charging cannot start SE a Ready the Rayocomp PS 10 is ready for use blinks blinks Wait charging device is not ready yet Note The charging status of the power pac
29. e harmonising time In order to stop harmonisation and set back the RAH module the stop button is pressed twice Repeating RAH Programs or Advancing to the Next Position If you have selected several programs at once you can choose programs within the set sequence directly This is performed with the arrows The arrow key pointing to the right or alternatively the button of the Rayotensor moved to the right lets you leap to the next program Actuating the arrow pointing to the left or the Rayotensor button pushed left selects the previous harmonising program The sequence of the program run corresponds with the sequence determined when the program flow was selected Actuating the button to the left or the Rayotensor button to the left sets the harmonising time of the current program back to the start value The previous program is only called up again if you repeat this step Testing a RAH Harmonising Program with the Rayotensor The harmonising programs of the Rayocomp PS 10 cannot be applied for harmonising alone you can also use them for testing Simply select a harmonising program as described at the beginning and press the start button After the program has been called up you can test with the Rayotensor to find out whether harmonisation is required or not The polariser is therefore set to neutral N A rotation of the Rayotensor cage shows that harmonisation is not required a linear motion shows the necessity to harmonise T
30. e is loaded completely and is ready for testing When an RAH programme was stored as a resonance a long beep signal sounds after successful storage The acoustic signals are very helpful because the user doesn t have to look on the display The user can concentrate completely on the testing 2 It is also possible to perform the handling functions of the sensor with the keyboard 4 4 8 Important tips and tricks for the test with the RAH Module 10 Page 47 3 Directly Altering the harmonisation time regarding a stored resonance value After storing a resonance value it is always possible to give it an individual harmonisation time directly Therefore you press the key START Afterwards the RAH programmes can also be related with an individual time fig 1 In front of the double point the hours are entered the minutes behind it Later on when you store all found RAH programmes on a RAH Green Card the individual time will be stored under this programme Rayonex Rayocomp PS 10 1 foes as LI 4 Very important Deleting the resonance memory Always when a new patient is going to be analysed the resonance memory has to be deleted After switching on the device the memory is empty and while testing with RAH programmes and their harmonisation times the memory gets filled During the RAH testing the resonance memory can be deleted by pressing the key for a longer time The device confirms this
31. ected in the last testing and stored on the RAH Green Card could now be checked again It is also possible to compile your own RAH programmes to store them on the RAH Green Card and then to test them Please insert a programmed RAH Green Card into the Rayocomp PS 10 and press the following key The RAH programmes stored on the card will now be read into the Rayocomp PS 10 and are then available for testing The display shows then for example Rayonex Rayocomp PS 10 Currently the RAH programme 04 22 is selected It is the first fig 2 of 10 RAH programmes to be tested fig 3 The note Card fig 1 appears on the display to show the user immediately that currently the values of a RAH Green Card are tested As already described in chapter 4 4 5 all respective RAH programmes can now be tested and stored Please also read the chapter 4 4 8 Important tips and tricks for testing with the RAH Module 10 4 4 8 Important tips and tricks for testing with the RAH Module 10 Page 46 4 4 8 Important tips and tricks for testing with the RAH Module 10 The following listed handling advices tips and tricks are valid for the level test chapter 4 4 5 the test with integrated test protocols chapter 4 4 6 and the test with the RAH Green Card chapter 4 4 7 l Acoustic signals The device always creates a double beep when a RAH programme was selected for testing Only after that the programm
32. ectively been activated The RAH module is marked by the indication RAH in the display After a short time the RAH prompt _ _ _ will appear In the following explanations the menu item RAH will be indicated with F1 RAH in the same way the Fundamental Frequency Values are harmonised in F1 BnPS This menu item provides all possibilities for testing fundamental frequency values If the M10 module is activated RAH programmes can also be tested The display is identic to menu item F1 The manual differentiates between test BnPS and test RAH With this menu item you can alter pre settings of the device call up information concerning the memory cards activate or de activate modules and perform a device test Select this menu item to gain information concerning the serial number of the device It will also indicate the current state of the integrated software and the existing hardware Particularly important you can see directly which modules are integrated in the Rayocomp PS 10 4 1 4 Selecting the Menu with the Mode Ke Page 15 When the Rayocomp PS 10 is switched on the display of the device shows the menu item Direct Rayonex Rayocomp PS 10 gege geg Now press the key oa eu 00 LI MODE The Rayocomp PS 10 changes over to the menu item Fl BnPS and TE m o WI displays Rayonex Rayocomp PS 10 LN GE If now the Mode key is not pressed amn P am am agai
33. en Card to the patient for his home therapy with the Rayocomp PS 10 Both devices work together hand in hand As the RAH programmes are often harmonised with different times an RAH Green Card can not only store the RAH programmes but also its individual harmonising times The RAH Green Card works perfectly with the new test possibilities of the module 10 Just imagine you want to select and test all relevant RAH programmes dealing with hypertension All RAH programmes that need to be harmonised can be analysed and can be stored on the RAH Green Card after the test While storing both the PS 1000 polar and the PS 10 add automatically well proven harmonising times Furthermore the values stored on the RAH Green Card can be read in for a further test This Saves a lot of time and avoids entering errors Up to 200 RAH programmes with their individual harmonsing times can be stored on an RAH Green Card Of course you can delete the RAH Green Card store new values on it and alter it with write protection But that s not all An RAH Green Card combines the Bioresonance according to Paul Schmidt and RAH seamlessly As the use of RAH programmes is often accompanied by single frequencies of the Bioresonance according to Paul Schmidt e g from an individual range value test the Rayonex engineers found a way to store single frequencies up to 500 different values as well as RAH programmes on the RAH
34. esting with Memory Card Values With the Rayocomp PS 10 it is very easy to save fundamental frequency values on a memory card The values stored on the card can be used for harmonisation via menu item F1 The values on the card can also be used for testing And this is the way it works Select the menu item Test with the key MODE Now insert the memory card and according to the table Chapter 4 4 enter the number 02 The Rayocomp PS 10 reads in the values from the memory card and changes over to the test mode as described before 4 4 3 Testing with Integrated Programs The memory cards represent individual compilations of fundamental frequency values but you can also test with the programs integrated in your Rayocomp PS 10 One example if the programs of the acupuncture oscillation therapy are integrated in your Rayocomp PS 10 you can test a complete channel for instance the kidney channel Simply carry out the following steps Select the menu item Test with the key MODE 4 4 3 Testing with Integrated Programs Page 39 According to the table Chapter 4 4 enter the number 03 and confirm the input with the OK key Then you can choose the program you want to test In this special case enter 2 08 for the kidney channel compare chapter 5 3 Now you can test the fundamental frequency values belonging to this channel Don t forget Only programs activated as a module can be tested If the module for the acupu
35. eyboard nee mcomoga 200 Boo EOEAERE noe 4 1 5 Operating the Integrated Polariser and Sensor However the functions of the polariser can also be set to the sensor directly while testing Moving the sensor button in direction of the connection cable backward the current polarisation of the frequencies is altered In the following you will find a summary of the sensor functions green Configuration of the sensor while testing Fundamental Frequency Values yy This function is explained under the menu item TEST BnPS You can take the exact course of a test and harmonisation from the menu item 4 2 1 Look at the display to see the currently set polarisation Rayonex Rayocomp PS 10 aa HU OO A OO OO ap lt M 808 G8 88 se 688 68 86 4129 5M Dk 4 1 6 Basic Information for the Rayonex Green Card Page 20 4 1 6 Basic Information for the Rayonex Green Card The idea of the new RAH Green Card is as simple as it is brilliant Previously when harmonising programmes were determined with the Rayocomp PS 1000 polar or Rayocomp PS10 by means of RAH for future use these had to be noted or printed With the new RAH Green Card these programme compilations can now be stored on a special memory card Thus it is now possible to e g store the tested programmes of the Rayocomp PS 1000 polar on an RAH Green Card and to give the RAH Gre
36. for example if you want to harmonise each fundamental frequency value for one minute and thirty seconds you only have to enter 01 30 For your convenience the Rayocomp PS 10 calculates the total harmonising time at once to be read in the line underneath in this case 1 hour 41 minutes and 30 seconds Rayonex Rayocomp PS 10 Time for each fundamental frequency value Total harmonising time With the key E J f 8m the automatic set up of all values begins This causes the Polariser to leap to With the key R g J you can interrupt the harmonising process telephon and continue with the key START 4 3 3 Loading Fundamental Frequency Values from Memory Cards and Harmonising Page 29 4 3 3 Loading Fundamental Frequency Values from Memory Cards and Harmonising We have already explained how the menu function F1 BnPS offers all possibilities to compile fundamental frequency values Of course this also includes the application of memory cards especially the RAH Grren Card For example if you have stored the resonance values of a range value test on a memory card you can use them to harmonise automatically in F1 BnPS To do so select the menu item F1 with the key MODE The display shows Rayonex Rayocomp PS 10 The Rayocomp PS 10 can now read the values stored on the memory card and harmonise with them Tip 1 You can combine the values of an integrated pro
37. gram with those of a memory card For example you read in the program 1 51 first then the memory card and then if you like a further program Tip 2 You can compile your own memory cards in the menu item Setup Look this up in the sub chapter 4 5 6 4 3 4 Fundamental Frequency Values from the Menu Item Test Taking Over and Harmonising If you have determined resonance values in the menu item Test you can use them directly for harmonisation after changing over to the menu item F1 Do so by entering 0 00 in F1 A better method however is to store these resonance values on a memory card and read them into the menu item F1 whenever they are needed 4 3 5 Utilising the Harmonising Programs of the new RAH Page 30 4 3 5 Utilising the Harmonising Programs of the new Rayonex Analysis and Harmonising System RAH An example illustrates the use of the RAH module We chose the programs Vitalisation complete and Upper respiratory system for our example The RAH module can be found in the menu item F1 BnPS together with the bio resonance programs according to Paul Schmidt Two separate submenu items BnPS and RAH were set to simplify the identification of the menus for the user These submenu items are selected with the mode key You can take the current program descriptions and program numbers from the special brochures Actuate the mode key several times until you see the menu item F1
38. has been selected 7 This symbol shows you the charge status of the battery pack If all 4 bars can be seen on the display the accumulator is full If no bar is indicated the accumulator is empty 8 During harmonisation with BnPS programmes the remaining duration of the harmonising appertaining to the current basic frequency value will be shown at this place 9 Here the currently set fundamental frequency value is indicated We have selected a large numeral display for better reading comfort 4 User Manual Page 11 4 User Manual 4 1 General Directions for Operation 4 1 1 Literature Relating to the Bio Resonance According to Paul Schmidt It cannot be the ambition of a user manual to explain all facets and possibilities of a method within such a complex area as bio resonance as Dietmal Heimes developed by Paul Schmidt Bio Reson a Schmidt a 0 According mg Application Meanwhile there is sufficient literature on the troduction DEN market from which you can take the application possibilities of the method For example we recommend the book Bio resonance as developed by Paul Schmidt with a total of 424 pages This book can be obtained from the publisher The address is Spurbuchverlag Am Eichenhugel 4 96148 Braunach Germany Phone 49 0 9544 1561 Internet www spurbuch de E mail info spurbuch de 4 1 2 Charging the Integrated Power Pack Page 12 4 1 2 Charging the Integrated Power Pack Th
39. he button of the Rayotensor button down also allows you to test with different polarities to find out whether the disorder is strong or weak If harmonising is required you write down the number of the program before Starting to test the next one Do so by pressing the stop button twice Then you can select the next harmonising program you want to test Please note that that the module 8 of the RAH only comprises a selection of the harmonising programs Only the modules 9 and 10 comprehend all available RAH programmes The functions of the module M10 are explained in detail 4 4 Menu Item Test Page 35 4 4 Menu Item Test The menu item Test offers a number of functions for testing fundamental frequency values We talk about testing in this context when we try to find out whether a certain fundamental frequency value requires harmonisation or not The test method was already explained in chapter 4 2 1 If you want to test single fundamental frequency values this is easiest to perform in the menu item Direct as we have described already It always makes sense to use the menu item Test if you want to test a large number of fundamental frequency values one after the other If you want to test electro magnetic disturbance for example you have to test six different fundamental frequency values In the menu item Direct we could enter these six values one after the other but one can hardly say this method is qu
40. here are no further sub programmes under the programme number 04 10 3 programmes are available on the shown second RAH programme level fig 6 The programme 04 10 is the first fig 5 of 3 programmes of the second level For the test the polariser of the Rayocomp PS 10 is automatically switched on N fig 3 There are currently no programmes selected that have to be harmonised visible under fig 4 If the sensor shows a linear movement while testing the sensor button is pushed down to mark the current programme that has to be harmonised This can be seen via the check mark on the right fig 7 Now the counter for stored resonance values fig 4 shows 1 LC UI If the sensor button is now pushed to the right the next RAH programme 04 20 will be selected The display shows 4 4 5 RAH Module 10 Level test Page 43 Rayon x Rayocomp PS 10 On the left of the display it is now shown that there are further detail programmes fig 1 under the programme 04 20 visible via the 3 i e for the third detail level If the sensor button is pushed forward the device selects the first detail programme 04 21 fig 2 Now the display shows Rayoy ex Rayocomp PS 10 The RAH programme 04 21 is the first of 7 detail programmes under the et RAH programme 04 20 which can be i E C found itself under the main group ES A 04 00 The whole structure of the RAH programmes can be analysed with
41. ick and easy However if you select an already activated test program i e the one for testing electro magnetic radiation from the menu item Test all six frequencies of this program are already integrated in the storage of the device and no longer need to be entered After the first frequency value has been tested you must switch to the following program value Do this by means of the keyboard key arrow to the right or use the sensor pushbutton to the right The device then automatically sets the selected test value Determined resonances can be temporarily stored in the internal storage of the device for harmonising later on This is achieved by actuating the OK key of the keyboard or by pushing the button of the sensor forward The temporarily stored values can then be harmonised via the menu item F1 with the program number P0 00 see chapter 4 3 4 Alternatively the values can be be stored on a memory card with the PS following key The Rayocomp PS 10 allows access to several of these fundamental frequency value compilations to be explained in the following Select the menu item Test BnPS with the df wopre key The display shows Rayonex Rayocomp PS 10 Just like the menu item F1 test is separated in BnPS and RAF However the menu item RAH can only be selected if the module 10 is activated Here you enter the number of the corresponding test program On the next page
42. ing them on the RAH Green Card Furthermore the module 10 supports the level test known from the Rayocomp PS 1000 polar When a main RAH programme is chosen the sub RAH programmes can be selected and tested by the use of the sensor s button IF you already use a RAH module with your Rayocomp PS 10 an upgrade to higher modules is possible e g from module 9 to module 10 The level test The level test supports the user very efficiently in the analysis of energetic disfunctions Using the example of electro smog tests we concretise it as follows If there is a general force of electro smog RAH No 04 00 you can narrow it on level 2 to find out whether these problems occur from electric and magnetic alternating fields RAH No 04 10 from pulse modulated radiation RAH No 04 20 or from high energy radiation RAH No 04 30 If there is for example a problem due to pulse modulated radiation you can switch on the third and last level the detail level Here frequency structures are available to find out in detail whether forces due to mobile communications UMTS WLAN or Bluetooth are existent This level structure is a central theme through the total RAH cause physiology pathology Special programmes like Bach flowers Schussler salts and enables indications for energetic disfunctions and a fast and effective testing of frequency ranges necessary to be harmonised Rayonex Rayocomp PS 10 If Module 10 is activated
43. ion you can delete a memory card completely thus preparing it for new writing processes Whenever you delete a memory card with this function you do not only delete the fundamental frequency values stored on it but also any read or write protection added to the contents of the card Please insert the memory card you want to delete in your Rayocomp PS 10 Enter the setup number 43 and wait until a O is displayed For safety reasons you must enter a 1 at this stage and confirm with the OK key 4 5 12 Initialising a Memory Card Setup Number 44 If you should come upon a memory card with illegible sticker you can test the card with this setup function and reinitialise it The function works for the memory cards of the bio resonance according to Paul Schmidt and also for the RAH Green Card When initialising the memory card with this function the system values of the card are newly determined and written and all the stored fundamental frequency values are deleted If the card is read or write protected this is also deleted Please insert the memory card intended for initialisation in your Rayocomp PS 10 Enter the setup number 44 and wait until a O is displayed For safety reasons you must enter a 1 at this stage and confirm with the OK key 4 5 13 Motor Test Run of the Frequency Module Page 58 4 5 13 Motor Test Run of the Frequency Module Setup Number 50 The user can test the mechanically driven frequency module of the
44. irect an RAH Green Card stored with frequency values or RAH programmes can be used for an automatic harmonising sequence Our principal object is that people who want to use this device for harmonisation can do this in a very simple way For the automatic programme sequence under the menu item Direct which is displayed immediately on switching on the device you have to do the following Switch on the device and connect it with the detectors Insert the stored RAH Green Card into the memory card slot Wait until the data have been loaded then unplug the card The automatic harmonisation sequence begins A LA PM zc At first the fundamental frequency values stored on the RAH Green Card are set up The display shows Rayonex Rayocomp PS 10 After the fundamental frequency emm ap values have been set up one after the is NW E j j other the automatic harmonisation III with the loaded RAH programmes with es their individual stored harmonisation gg 37 times begins 98 18 38 gi If only RAH programmes were loaded from the RAH Green Card only RAH programmes will be harmonised This is also valid for fundamental frequency values In the automatic harmonisation sequence of the RAH programmes the display shows The display shows rAH so that you know immediately that RAH programmes are being harmonised currently Note Using the sensor each single fundamental fre
45. k is always displayed This bar indicator allows you to estimate the length of time left for operation before it becomes necessary to recharge the device 4 1 3 Switching the Device On and Off Page 13 4 1 3 Switching the Device On and Off On the rear wall of the Rayocomp PS 10 you will find the switch for switching the device on and off When switching on please make sure that no memory card inserted in the device In case wens PS 10 is not under supervision for a longer period of time overnight 3 ch for example make sure that the power pack is charged sufficiently Total discharge even if it happens only once damages the power pack permanently After the device is switched on a short trial run is started A mechanical test is performed along with a check of the power pack If you receive the message Er 29 the power pack must be recharged before you can work with the device Then the device switches over to the menu item Direct and is immediately ready for action The display shows Rayonex Rayocomp PS 10 wert LIT Lk U It is not necessary to move to a certain menu item before switching off the device This means you can switch off the device at any time Please keep in mind that the resonance values and the found RAH programmes detected in the menu item Test are lost when the device is switched off When the Status of the power pack goes down until its functions can no longer be gu
46. look into the list in chapter 5 8 you will find the following information for the indicated error Er 88 Er is the abreviation for error you have tried to read in a memory card which has not been inserted Note The acoustic signal can be switched on or off in the menu item Setup see chapter 4 5 3 4 2 Menu Item Direct Page 22 4 2 Menu Item Direct 4 2 1 Testing and Harmonising a Fundamental Frequency Value with the Integrated Polariser and Sensor lt makes no difference whether you want to test a fundamental frequency value an allergen a parasite a Chakra etc the course of testing and harmonising is always the same Here is the description of a liver test fundamental frequency value 56 00 as an example Before beginning with the testing process please consider whether you are testing yourself or another person For a self test connect yourself to the hand and foot detectors or the Velcro detectors You hold the sensor in your right hand for the test your index finger must touch the knurled screw of the sensor The frequency spectra of the device also enter the circulation of your right arm in this way If you are testing another person your patient is connected to the hand and foot detectors or Velcro detectors too In this case your index finger must not touch the knurled screw to avoid accidental testing of your own resonances Now set the fundamental frequency value 56 00 example in the men
47. ment course with these programs please press the following keys Now the display will look like this Rayonex Rayocomp PS 10 vor LI TU III These values indicate the length of the harmonising time of the currently selected program and the total time needed for the programs already selected i I LI BIR Ug The next step is for setting the harmonising time The already set default value is 10 minutes This value can be taken over directly with the OK key As an option this value can also be altered in steps of one minute length to suit your own needs If this entry is altered the value is taken over as the new default value This simplifies the entry for the user if a harmonising time other than the given value is required 4 3 5 Utilising the Harmonising Programs of the new RAH Page 32 If the time is lengthened to 15 minutes for instance the display will change as follows Rayonex Rayocomp PS 10 oo 6 ert LI TI i 0 15 LI ney we WL DD Harmonisation of the first program will now take 15 minutes The total harmonisation time will also be 15 minutes because only one program has been selected so far Confirm the entry of the harmonisation time with the OK button The following fig shows the new display Now you can enter the second program The 2 in front of the harmonising time indicates the second program Rayonex Rayocomp PS 10 70 16 must be entered as next to select the program
48. mple the testing range values with the pre set values are selected Rayonex Rayocomp PS 10 With this display the Rayocomp PS 10 informs you about which fundamental frequency value is being tested at present 1 how many fundamental frquency values are going to be tested altogether 2 how many fundamental frequency values have already been tested 3 how many fundamental frequency values have caused a resonance 4 and whether the currently displayed fundamental frequency value is a resonance value 5 With the help of the sensor you can find out whether there is any resonance and if this resonance is strong or weak compare chapter 4 2 1 As a practical feature while testing you can set the next value or store the current value with the sensor compare chapter 4 1 5 4 4 1 Testing Range Values The idea behind this test is to find out all the fundamental frequency values in disorder at the moment then to store and harmonise them With the bio resonance devices according to Paul Schmidt it is possible to carry out a resonance test in a very wide range of fundamental frequency values In practice this is how it works beginning with the fundamental frequency value 0 00 you start testing whether this fundamental frequency value is required for regulation If this is the case it is stored in the Rayocomp PS 10 Carry on by testing the fundamental frequency value 0 50 then 1 00 and so on After the test precisely those reso
49. n for a short moment the PS10 will automatically start the input mask allowing you to enter the BnPS programmes Press the following key again Rayonex Rayocomp PS 10 LI j By pressing the Mode key again the Rayocomp PS 10 will switch over to Rayonex Rayocomp PS 10 the RAH module However this will wm only occur if the modules have an as aT effectively been activated Thus if l either the Module 8 Module 9 or Module 10 is activated RAH is indicated in the display for a short moment Subsequently the RAH programmes can be entered via the N input mask __ __ Te G 4 1 4 Selecting the Menu with the Mode Key Page 16 Rayonex Rayocomp PS 10 Rayonex Rayocomp PS 10 Vv alal UI II l o After having pressed again the Rayocomp PS 10 changes to the menu item Test and shows on the display N If the module 10 is activated the menu item Test RAH will be shown on the display after the menu item Test BnPS By pressing on the key mode once again this menu item can be selected If the key mode isn t pressed once more after having selected the menu item Test BnPS the following prompt will appear Rayonex Rayocomp PS 10 CP Here BnPS fundamental frequency eme values can be tested For further t information take a look at chapter 4 4 Please press again the key 4 1 4 Selecting the Menu with the Mode Ke Page 1
50. nance values necessary for regulating the organism of the patient have been determined In a way the detected resonance spectrum presents the energetic fingerprint The complexity of the determined fundamental frequency values reflects the complexity of the disorder Therefore a diagnosis is normally not given after the first test But the so found resonance values are used for harmonisation This is carried out by transferring the fundamental frequency values to the tested person via hand and foot detectors or Velcro detectors 4 4 1 Testing Range Values Page 38 Another possibility is to transfer the frequencies by means of the large Surface detector maybe while sleeping at night And this is the method for performing the range value test Select the menu item Test with the key MODE According to the table Chapter 4 4 enter the number 01 and confirm the input with the OK key You will see the start value of the test on your display If you have made the right choice please confirm with the OK key After this the concluding value is displayed which is confirmed with the OK key once again Now you have the chance to alter the step lengths or to confirm the indicated length with the OK key We recommend the start value 0 00 the final value 99 99 and a step length of 0 50 With this set up 200 values will be tested Now you can follow the description for testing fundamental frequency values to find resonances 4 4 2 T
51. ncture oscillation therapy is not activated you cannot test with the programs belonging to the module 4 4 4 Testing with Integrated Test Programs In addition to the integrated programs the device also contains a compilation of test programs which are very frequently required They are listed below once more for your convenience Geopathy test POG _ Check up test PO Acid basetest P10 Quick test of the 14 channel P11 Vital substance test Taking electro magnetic pollution as an example you test as follows Select the menu item Test with the key MODE Then enter the number 04 and confirm your entry with the OK key You will receive the six most important fundamental frequency values for testing electro magnetic pollution Of course the test programs 04 to 10 can only be used if the corresponding module 6 compare chapter 5 7 is activated in your Rayocomp PS 10 4 4 5 RAH Module 10 Level test Page 40 4 4 5 RAH Module 10 Level test Basics for the RAH Module M10 For the Rayocomp PS 10 the module M10 is the most effective module which has in addition to the module M9 numerous test functions The module M10 offers test protocols With these protocols all necessary RAH programme numbers can be provided for the user s test at the push of a button The corresponding RAH programme numbers don t have to be entered separately The test results are then available for harmonisations as well as for stor
52. ne button Sometimes this can be more than 100 different RAH programmes An RAH programme with an available test protocol is marked with a 10 fig 1 on the right of the display Rayonex Rayocomp PS 10 This shows immediately that a test 1 am protocol is available The test protocol 2m l Up l I itself is activated by pressing the key a ot LIL The display then changes to the RAH programmes of the test protocol You can leave the test protocol again by pressing the Stop key Rayonex Rayocc Pore vet LIT UU 2960 we ko 53 A check mark will appear behind the 10 on the display fig 5 after activating the test protocol Then the current test protocol is shown In the middle of the display the number of the test protocol is indicated fig 2 In this case 39 60 The figures 3 and 4 show that the current RAH programme 00 00 is the first of a total of 53 RAH programmes As already described in chapter 4 4 5 all respective RAH programmes can now be tested and stored l Please also read the chapter 4 4 8 Important tips and tricks for testing with the RAH Module 10 4 4 7 RAH Module 10 Test with the RAH Green Card Page 45 4 4 7 RAH Module 10 Test with the RAH Green Card In the module 10 of the Rayocomp PS 10 it is not only possible to harmonise the RAH programmes stored on a RAH Green Card but also to test them As an example the RAH programmes that were sel
53. otal of four special keys equipped with certain functions Important These four keys can only be operated in the menu Direct PER DE c HitwoaeG Ep ooo Ceges TE elclot stant stop LI E E ad a e 1 three IF keys IF1 IF2 and IF3 2 the 10 0 key Caption 4 8 1 Using the F Keys You can use the F keys Individual Fundamental Frequency Value to store often used fundamental frequency values They work like the station selection buttons of a radio First you tune in to the radio station then you keep the key pressed for a while until the station is stored The function of the IF keys is exactly the same Enter the desired fundamental frequency value in the menu item Direct Confirm the entry with OK Then press the desired IF key until you hear a signal Set the desired fundamental frequency value in the menu item Direct Then press the desired F key until you hear a signal Now the value is stored and can be called up at any time by pressing this F key You can store three different fundamental frequency values in this way IF1 IF2 and IF3 4 8 2 Using the 70 0 Key The key 70 0 is permanently linked to this fundamental frequency value This often needed fundamental frequency value allergen harmonising energy duplicating is called up rapidly with the special key whenever you want Furthermore this key has a special function in the module M10 5 Appendix Page 62 5 Appendix 5 1 S
54. ou can enter the start value first After hitting the OK key you can enter the final value 4 5 4 Pre Setting the Test Range Values Page 52 As soon as this has been confirmed with the OK key the step length can be pre set perhaps to a smaller value of 0 25 As always the step length is confirmed with the OK key Here is the standard setup of the Rayocomp PS 10 on delivery Initial value 00 00 Final value 99 99 Step length 00 50 4 5 5 The pre setting of the harmonisation time for fundamental frequency values Setup Number 14 The pre setting of the harmonisation time for fundamental frequency values can be very helpful for the use of a BnPS F1 programme as well as for the use of a Rayonex Green Card If e g single frequencies and also RAH programmes were stored on an RAH Green Card the individual harmonisation times for the RAH programmes can be stored but not for the single frequencies If an automatic programme sequence is started by the means of an RAH Green Card the default for the single frequencies can be regulated by the pre setting of the harmonisation time Standard is 30 seconds If the harmonisation time has been changed in this setup item to e g minute every existing fundamental frequency value of the Rayocomp PS 10 is automatically harmonised with 1 minute when using a RAH Green Card In the same way when using the BnPS Programmes the pre set harmonisation time of this setup i
55. quency value or RAH programme can be switched forward if e g the sensor shows that the harmonisation has been finished ahead of schedule 4 3 Menu Item EI Page 25 4 3 Menu Item EI 4 3 1 Basics Concerning the Use of Integrated Programs When the Rayocomp devices were first introduced in 1993 it was possible to set groups of fundamental frequency values the harmonising programs automatically at last The first Rayocomp PS 1000 devices already had 51 integrated harmonising programs Later on three further programs were added their validity remains to this very day For example if such a program consists of 100 different fundamental frequency values the Rayocomp PS 10 sets one fundamental frequency value exactly 100 times Due to their simple application the harmonising programs integrated in the Rayocomp devices are the ideal means for utilising the potential of many thousands of individual measurements because the harmonising programs contain the resonance values frequently found in a certain regulation area For example if you test many individuals with liver problems the comparison of the so found resonance spots comes up with conspicuous incidences this discovery finally lead to the development of the control plan The set up within the harmonising programs always follows the same pattern The first fundamental frequency values are the pre control values taken from the control plan Then the programs receive the fun
56. rehend BnPS Fundamental Frquency values as well as RAH programmes BnPS and RAH storage values are displayed by turns A Green Card can comprehend maximum 500 BnPS and 200 RAH values You can detect the number of RAH programmes from that figure that is shown in the display with the 200 at the same time e g 34 200 Insert the memory card and enter the setup number 40 The display may show Rayonex Rayocomp PS 10 Caption 1 This number indicates the amount of values already stored on the memory card 2 Here you can read the number of values which can be stored on the memory card An exception is the green Card The Green Card can contain 500 BnPS and 200 RAH values which are shown by turns 3 if you see a 0 the memory card is faulty A number 1 indicates a memory card suitable for use when working with the bio resonance according to Paul Schmidt If the displayed numeral is a 2 this memory card can be used for the B A T acute programs according to Manfred Denecke A Green Card is indicated with a 3 4 5 9 Providing Read Protection for a Memory Card Page 56 4 5 9 Providing Read Protection for a Memory Card Setup Number 41 This setup function provides a read protection for the inserted memory card If you use a read protected memory card with a Rayocomp device i e in Fl the fundamental frequency values are set up but not displayed This may make sense when developing and testing new fundamen
57. ry the standard setup of the Rayocomp PS 10 is number O 4 5 3 Adjusting the Attention Sound Bee Page 51 4 5 3 Adjusting the Attention Sound Beep Setup Number 12 The Rayocomp PS 10 has been equipped with an acoustic signal This may signalise the use of a wrong key or confirm an entry The Rayocomp PS 10 has three different error messages or acoustic signals 1 One long sound With this signal the device confirms a successfully completed action i e an entry or storage 2 Two short sounds The key you are pressing has no current function 3 Two long sounds This signalises a severe error The device will supply an error code In the appendix chapter 5 8 you can look up the current error This setup function enables or disables the signal Enter the setup number 12 and you can add one of the following numbers 1 The acoustic signal is switched on 0 The acoustic signal is switched off On delivery the standard setup of the Rayocomp PS 10 is number 1 4 5 4 Pre Setting the Test Range Values Setup Number 13 One of the most successful methods of bio resonance according to Paul Schmidt is the range value test Beginning with a start value usually 00 00 and closing with a final value usually 99 99 in step lengths of 00 50 i e every single fundamental frequency value is tested for a resonance and then harmonised These three values can be pre set in this setup function After entering the setup number 13 y
58. tal frequency values Please insert the memory card intended for read protection in your Rayocomp PS 10 On entering the setup number 41 the display will either show O card is not read protected or 1 if the card already has read protection A RAH Green Card cannot be modified with a read protection If the card is still without read protection please enter 1 and confirm your entry with the OK key Please keep in mind the read protection of a memory card can only be reversed by deleting it see chapter 4 5 10 4 5 10 Providing Write Protection for a Memory Card Setup Number 42 This setup function provides a write protection for the inserted memory card A memory card with write protection cannot be deleted by accident nor can other fundamental frequency values be stored on it Please insert the memory card intended for write protection in your Rayocomp PS 10 On entering the setup number 42 the display will either show O card is not write protected or 1 if the card already has write protection If the card is not write protected yet please enter the numeral 1 and confirm with the OK key to provide write protection for your memory card If the number 1 is displayed to show that your memory card is write protected simply enter O and confirm with the OK key to remove write protection 4 5 11 Deleting a Memory Card Completely Page 57 4 5 11 Deleting a Memory Card Completely Setup Number 43 With this setup funct
59. tem will be used Rayonex Rayocomp PS 10 After selecting setup number 14 you confirm it with the key OK Then the display shows the illustration as demonstrated here on the left 00 30 means 00 minutes and 30 seconds At first the cursor blinks in the input box of the minutes For example you can enter 0100 and confirm this with the key OK and so 1 minute is pre set The entry 0040 would mean a pre setting of 40 seconds 4 5 6 Activating and deactivating modules Page 53 4 5 6 Activating and Deactivating Modules Setup Number 20 At present a total of ten modules can be activated or deactivated respectively in the Rayocomp PS 10 If for example the harmonising programs of the acupuncture oscillation therapy are not activated on your device yet you can use them immediately after activating them with an enabling code This sub item of the menu has the function of activating or deactivating modules After having entered the setup number 20 you can select the number of the module intended for activation or deactivation Please confirm each input with the OK key Here is a list of the module numbers 01 _ Setup precision of 005 11 Test programs 06 Denecke e n RAH Module 8 mm Rayonex Analysis and Harmonising programme RAH Module 9 nm Rayonex Analysis and Harmonising programme RAH Module 10 Your final action is to enter the seven place code for activating or deactivating and confirm your
60. u item Direct to the Rayocomp PS 10 The polariser integrated in the device is set to N Test with the sensor to find out whether you achieve a rotation left or right motion or a linear movement A rotation means that the field in the organism linked to the fundamental frequency value is stable and therefore does not require treatment If however the achieved motion is linear this patient is suffering from an energetic blockade If you want to know more precisely which type of polarisation is most Suitable for helping the organism or if you want to find out whether the disturbance is strong or weak press the button backward after having tested a linear movement on N Now the polariser is switched over to polarisation If the sensor starts rotating the disturbance is not very strong However if the linear movement continues when set to polarisation the disturbance is stronger Press the button backward and test whether the polarisation which is now set leads to a rotation Rotation indicates that the polarisation is most suitable for harmonisation If still no rotation is achieved press the button of the sensor once more backward Now you will see a rotation on the polarisation Harmonisation has been completed as soon as the sensor performs a rotation when the polariser is set to N 4 2 2 Setting Up Individual Fundamental Frequency Values Page 23 4 2 2 Setting Up Individual Fundamental Frequency Values The menu
61. ummary of the Integrated Test Programs Display POI _ Testing with range values Test with memory card values Test with integrated programs ER EENG P 06 Check up test PO Acid baset S P10 Quick test of the 14 channels P11 Vital substance test 5 2 Summary of the Integrated Rayonex Programs Page 63 5 2 Summary of the Integrated Rayonex Programs 5 3 Summary of the Integrated Acupuncture Programs Page 64 5 3 Summary of the Integrated Acupuncture Programs P 2 13 Channel of the governing vessel P 2 14 Channel of the conception vessel 5 4 Summary of the Integrated Special Programs Page 65 5 4 Summary of the Integrated Special Programs 5 5 Summary of the Integrated Programs by Dr Ulrich Page 66 5 5 Summary of the Integrated Programs by Dr Ulrich Learning program DOG Ned ed P 4 11 Constants of physics program Spiritual 3 protection Heart program 5 6 Summary of the Modules for integration Page 67 5 6 Summary of the Modules for Integration Module KEE 01 2010 Step precision adjustment of 0 05 Rayonex Analysis and Harmonising programme RAH Module 8 2090 Rayonex Analysis and Harmonising programme RAH Module 9 Rayonex Analysis and Harmonising programme RAH Module 10 vacant vacant 5 7 Summary of the Possible Error Messages Page 68 5 7 Summary of the Possible Error Messages No Description gt BE E 11 Wrong memory card type S 16 Invalid mo
62. urpose please contact the Rayonex company This number informs you about the software version of the device In this case it is version 28 This indicates the currently active menu item This symbolises the load condition of the battery pack This number stands for the construction month of the device in this case December This is the construction year of the device It was the year 2006 in our example These four numerals are the serial number of the device The number in our case 22 39 is important for the definite classification of the device Rayonex maintains a database with all the information concerning the device i e the activated modules the previous service date etc If you are thinking of enabling a module in the Rayocomp PS 10 you can call us and tell us your serial number to receive an enabling code After entering this code you can utilise the module immediately A really useful tip in this context If you wish to try out a certain module call us to receive a so called temporary enabling code This code which can only be used once allows the use of the respective module for about 10 hours The RAH module can even be tested for 100 hours The determinant is the runtime of the device not the runtime of the module During this time you can find out for yourself whether the module is useful for your purposes or not 4 8 Special Function Keys Page 61 4 8 Special Function Keys The Rayocomp PS 10 has a t